catalogo ldm-2014 layout 1 - brevetti cea · ldm uses a well-known and widely proven detection...

2
When Inspection Needs Perfection LDM

Upload: others

Post on 26-Jun-2020

14 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Catalogo LDM-2014 Layout 1 - Brevetti CEA · LDM uses a well-known and widely proven detection principle based on the measurement of electrical conductivity of sealed containers using

When Inspection Needs Perfection

LDM

Catalogo LDM-2014_Layout 1 21/03/16 17:57 Pagina 1

Page 2: Catalogo LDM-2014 Layout 1 - Brevetti CEA · LDM uses a well-known and widely proven detection principle based on the measurement of electrical conductivity of sealed containers using

conc

ept:

over

-prin

t.it

BREVETTI C.E.A. SpA

Via del Commercio, 28 - 36050 Sovizzo (VI) ItalyTel. +39 0444.551.988 - Fax +39 0444.536.241www.brevetti-cea.com - [email protected]

Rev

. 2 ·

03/2

016

Specification subject to change without notice.All rights reserved. The copying of technical data and images is forbidden. © Copyright BREVETTI C.E.A. SpA

Inspected containers: Vials, Ampoules and Cartridges

Products to be inspected Waterlike, Suspensions, Oily

Container size: Diameter: from 8mm to 32mm Height: up to 130mm

Machine nominal speed: up to 400 pcs/min (24.000 pcs/h)

Leak Detection Units 3/4 upside - down

Min Product Conductivity 1,5 μS/cm

Loading: By trays or in line

Unloading Accepted: By trays or in line

Unloading Rejected: By trays

Electric supply: 230V±10% - 50/60Hz - 1Ph + PE

Power: 3kW

Compressed air: 6bar, 200 Nl/min

Dimensions (WxDxH): 2650x1560x1750mm

Weight: 900 kg

Loading height: 970 mm ± 25 mm

Unloading height: 970 mm ± 25 mm

Noise: < 75 dB(A)

Manual loading

“Fail safe” reject verification system (Optional)

21 CFR part 11 compliance for electronic records (recipes, batches/sub-batches, audit trail)

Ethernet connection

User friendly interface

In-line connection to upstream and downstream equipments.

Leak Detection available for liquid products (High Voltage)

The High voltage leak detection technology is available as a stand-alone machine or as a com-pact modular system which can be integrated in new and existing lines.LDM uses a well-known and widely proven detection principle based on the measurement ofelectrical conductivity of sealed containers using a high voltage electric field.

The system measures the electrical conductivity of the “container + liquid product” placedbetween two electrodes. Any conductivity variation due to glass or sealing defects is detected.LDM introduces the concept of upside-down inspection, which improves precision and performances. Combined with a fast and advanced digital-processing unit, the system can easily detect defective containers (cracks, micro-holes, thin glass).

LD module is the synthesis of more than thirty years of Brevetti C.E.A. expertise in the fields ofinspection and leak detection. Used in combination with K32S, A35 and A50-450 inspectionmachines, it can be considered the most compact and performing "inspection Line" forampoules, vials and cartridges available on the market.

The information given on this data sheet is accurate as for the date of issue. Nothing here in shall bedeemed to constitute a warranty, explicitly or implicitly. It is the user’s responsibility to determine theapplicability and suitability of such material to any particular purpose.

Catalogo LDM-2014_Layout 1 21/03/16 17:58 Pagina 2