cbpst, kochi testing technical specifications.pdf
TRANSCRIPT
S.
No.
List of Equipments for CBPST - Kochi Qty
Testing Equipment
1. Digital Vernier Calipers
Each One
No
2. Digital Ball Ended micrometer
3. Digital Thickness Gauge
4. Pi Tapes
5. Analytical Weighing Balance
6. Moisture Analyzer
7. Vacuum Oven with Pump
8. Hydrostatic Pressure testing System for Pipes
(a) Hydrostatic Pressure Tester
(b) Hot Water bath
(c) Low temperature Cabinet
(d) End fittings for Pipes
9. Constant Temperature Oil Bath for Reversion Test
10. Falling Weight Impact Tester for Pipe
11. Environmental Stress Cracking resistance Tester
12. Free Falling Dart Impact Tester with Built- In Vacuum
System, for Plastics Films Testing
13. Izod and Charpy Impact tester
14. Ultrasonicator
15. Compact Microprocessor-Controlled Bench-Top Laboratory
Fermenter-Bioreactor
16. Autoclave
17. Incubator
18. Optical Microscope with Programmable Hot Stage
19. FTIR with Microscope
20. Ultra Violet Visible Double Beam spectrophotometer
21. Universal Testing Machine
22. Gel Permeation Chromatography
23. CHN Analyzer
24. Horizontal Laminar air flow Specification
25. Water Vapour Transmission Rate
26. Gas Permeability Tester
PROCESSING MACHINERY
1. Extrusion Blown Film Unit with Carona Treater
S. No. 1
Digital Vernier Calipers
Technical Specifications
Range : 0 - 150 mm & 0 – 300 mm Accuracy : ± 0.01 mm Repeatability : ± 0.01 mm Display : LCD Construction material : Carbide-tipped jaws for outside and inside measurement Power Supply : Lithium battery Calibration certificate traceable to NIST should be provided
S. No. 2
Digital Ball Ended micrometer
Technical Specifications
Range : 0 - 25 mm Resolution : ± 0.001 mm Measuring surface (Ball) : 3.2 mm diameter Measuring force : 7 to 9 N Measuring system : Electromagnetic induction type sensor Power supply : Lithium battery (CR2032) x 1 Calibration certificate traceable to NIST should be provided
S. No. 3
Digital Thickness Gauge
Technical Specifications
Measuring Range : 0 – 10 mm / 0.4 inch Resolution : 0.001 mm / 0.0001 inch Depth of Jaw : 50 mm Display : LCD Measuring force : < 1.4 N Lifting device : Operating by lifting device – self closing Power Supply : Lithium battery Calibration certificate traceable to NIST should be provided
S. No. 4
Pie Tapes
Technical Specifications
Size : 20-300, 300-700, 700-1100 mm L.C : 0.1 mm Construction material : High grade of SS material Calibration certificate traceable to NIST should be furnished
S. No. 5
Analytical Weighing Balance
Technical Specifications
Capacity : 220 g Accuracy : 0.1 mg Display : Backlit display Size : 20 (W) x 35 (H) x 30 (D) cm Repeatability : 0.1 mg Power Supply : 230 V, 50 hz Pan Size : 8 – 10 cm Weighing units : mg, g, & kg Stabilization time : 2 sec Stability Filters : Three filters to match weighing environment Output : RS 232 Accessories : Calibration Weights & Density measurement kit
S. No. 6
Moisture Analyzer
Technical Specifications
Sample weight : 0.01 g to 200 g
Reproducibility : ± 0.05 mg for (1 mg water content) and ±0,5 mg for
(20 mg water content)
Display via PC : Absolute value of water content of sample in mg
Measuring range : 0.2 mg - 30 mg
Test temperatures : 50°C up to 210 °C in 1°C-steps
Ambient air temperature : in °C in 0.1 °C steps
Ambient air pressure : 700 - 1100 mbar in steps of 0.1 mbar
Relative air humidity : 0 - 100 % in steps of 0.1 %
Power supply : 230 V, 50 Hz
Dimensions : 285 mm (H) x 170 mm (W) x 250 mm (D)
Interface : Bi-directional RS 232 interface
Required PC : Windows 98, XP; Pentium 233 or equivalent; RAM:128 MB,
serial interface RS 232
S. No. 7
Vacuum Oven with Pump
Technical Specifications
Capacity (Vacuum Pump) : 50 Litres/min
Heating Load : 2 KW
Power Supply : 230V / 50 Hz AC
Size (inner Chamber) : 12 inch Diameter X 15 inch Depth
Construction : Inside – SS 304, Outside - MS Body with Powder
coating with proper insulation. Vacuum pump with
0.5 HP, double stage, shutter with glass window
and heat resistant (silicone) Gasket.
Working range : 0 to 200°C
Temperature Accuracy : ± 2°C
Temperature controller : PID action digital temperature controller
S. No. 8: Hydrostatic Pressure testing System for Pipes
S. No. 8 (a)
Hydrostatic Pressure Tester
Technical Specifications
Variable Working pressure Range Digital type : 0-10MPa Display Regulating accuracy : 0.001MPa Testing station : 5 Stations (for Acceptance & Type test) : [2 for acceptance (0-10 MPa) & : 3 for Type test (0-4 MPa)] This instrument must be suitable to carryout Hydrostatic Pressure Test as per: IS 4984 & 4985 Suitable pump has to be provided to developing water pressure as required for Acceptance and Type test separately. Fine Filter has to be provided for water inlet Timer has to be provided to record the starting time, Time of failure, over pressure time, pressure compensating time etc.
S. No. 8 (b)
Hot Water bath
Technical Specifications
Size (LXWXH) (Inner Dimension) : 1m X 1m X 1.2m Temperature control system : Ambient to 100⁰C
Accuracy : ± 1⁰C (Throughout the bath)
Should conform to the requirements of standard IS 4984.
In built water circulation pump. Well insulated SS body and top cover. Inside the water bath SS stand / grid should be provided.
S. No. 8 (c)
Low temperature Cabinet
Technical Specifications
Chamber size inside : 750 X 500 x 900 mm (LxWxH) No of Racks : 3 Racks Chamber material : SS 304 Door : Top opening single door with proper insulation
Temperature range : -10°c to 30°c
Accuracy : ± 0.1⁰C
S. No. 8 (d)
End fittings for Pipes
Technical Specifications Sizes (mm) : 12,16,20, 25, 32, 40, 50, 63, 75, 90, 110 - Each 2 sets
125, 140, 160, 180, 200, 225, 250, 280,300 & 355 - Each one set
Fitting must be suitable for UPVC/HDPE Pipes as per IS 4985/IS4984 respectively Should be light weight End Fitting Material : Non rusting metal ‘O’ ring material should be silicone Bolts/ Nuts/ Allen screws/ connection Nipples must be stainless steel / copper / brass.
S. No. 9
Constant Temperature Oil Bath for Reversion Test
Technical Specifications
Temperature Range : Ambient to 200°C
Temperature Controller : PID, LED Display
Temperature Accuracy : ± 1°C
Internal Dimensions (LxWxH) : 600 mm x 600 mm x 450 mm
Body Construction
: Inner SS & Outer MS Powder
Coated, double wall
construction with good quality
insulation with lid.
Conform to the standards :
: IS : 4985, IS : 12235,
IS:13592, IS:15328, IS:4984
with latest amendments.
S. No. 10
Falling Weight Impact Tester for Pipe
Technical Specifications Standards : IS 4985, IS 12818, IS 13592, IS 12235 Falling weight testing machine : Main frame with guide rail or tube fixed in a true vertical
Position Release Mechanism : Release mechanism with free fall from variable height up
to at least 2 Meter with an accuracy o ± 10 mm Max. Fall Height : 2050mm ± 10 mm With calibrated scale (preferable) Mass of the Striker (kgs) : 0,25, 0.5,0.8,1.0,1.25,1.6,2.0,2.5,3.2,4.0,5.0,6.3 Specimen holder for pipes : V Block with adjustable height. :’V’ Block base size 300 x 300 mm : ‘V’Block base thickness 25 mm : ‘V’ Angle : 120⁰
S. No. 11
Environmental Stress Cracking resistance Tester
Technical Specifications
Constant temperature bath with S S lid.
: Inside-SS, outside –MS with heat insulation Lid: S S.
Heating System : Heater with digital temperature controller (PID) Temperature Thermometer : RT to 100°C. Accuracy : ±0.10°C To maintain uniformity in temperature throughout the bath
: Stirrer should be provide.
Blanking die for cutting specimen
: 38±2.5 mm by 13±0.8mm with ejector
Nicking Jig of lever arm with dial gauge
: 0-10 mm
Notch depth adjustment : 0.30, 0.40, 0.50, 0.65 mm with notch length 19 mm
Bending cum transferring device : 1 No.
Specimen Holder : S S 10 Nos. Test Tubes (Glass) : 32 mm dia and 200 mm length – 10 Nos Corks fit to the glass tubes : 10 Nos. Interchangeable notching blades : 5 Nos. Thermometer : 1)50°C (20 to 60°C range)
2)100°C (60 to 105°C) Conform to : ASTM D 1693
S. No. 12
Free Falling Dart Impact Tester with Built- In Vacuum System, for Plastics
Films Testing
Technical Specifications
Clamp : Inside diameter of 127 mm, vacuum operated clamps with
rubber gasket on both side. (Lower/Upper)
Dart holding device : Electromagnetic releasing device
Masses : SS masses of diameter 30 mm with a centre hole of
diameter 6.6 mm and height varying from 2.5 to 25 mm
(5g,10g,15g,60g - each 4 Nos and 90g - 2 Nos)
Dart : Hemispherical with diameter of 38.1 mm made out of
aluminum
Aluminum Shaft : Length of 115 mm and diameter of 6.4 mm. Tip of length
12.7 mm is made out of steel
Weight of dart including
shaft : 100.0 gm
Reference standard : ASTM D 1709, Method A & B, IS 2508.
S. No. 13
Izod and Charpy Impact tester
Technical Specifications Model : Digital Model Confirms to Standard : ASTM D 256 & ASTM D 6110, ISO 180 & ISO 179
Pendulum Impact Energy : 0-25 Joules Least Count of Indication : 0.1 Joules Release Angle of Pendulum : 150⁰ Striking velocity range of Pendulum : 2.5 – 3.5 m/s. Accessories : Standard detachable weights corresponding to (2J, 4J, 7J,
15J,and 25J)to obtain the desired test range. Standard fixtures for fitting Izod /Charpy test and other standard accessories.
Auto Notch Cutter & Sample : Suitable for notching plastics specimens for Izod & Mounting accessories Charpy impact tests. As per ASTM D 256, D 6110 and ISO
180, ISO 179 Power : 230V, 50Hz, Single phase Range of 4 seaters : 0-2.71J Resolution 0.02J 0-5.42J, Resolution 0.05J 0-10.84J Resolution 0.10J 0-21.68J Resolution 0.20J 0-25.0 Resolution 0.20J
S. No. 14
Ultrasonicator
Technical Specifications Frequency : 20 kHz Processing volume : 200 µL to 2000 mL Probe Tip diameter : Max. 25.4 mm Probe material : Ti-Al Alloy, autoclave Power output : 0 – 300 Watts Timer : 1 second to 24 hrs Pulser : 1 sec. To 99 hrs Temperature controller : 0 - 100°C
Temperature Accuracy : 1°C
Display : 145 mm LCD
S. No. 15
Compact Microprocessor-Controlled Bench-Top Laboratory Fermenter-
Bioreactor
Technical Specifications Power Supply : 230 V /50-60 Hz Operating Temperature : 0 – 60°C
Relative humidity : 0 – 100% Display : Backlit LCD Dimension : 22 cm (W) x 38 cm (H) x 40 cm (D) Working volume : 35 mL to 5 L Vessel Material : Pyrex borosilicate glass, 6 or 9 threaded necks Mixing : Mixer with the speed of 0 – 1200 rpm. Temperature control : +5°C to 60°C Temperature accuracy : 0.1°C Precision : +/- 0.2°C Thermo sensing : High precision Pt 100 built in the glass electrode bulb pH control : Sterilizable glass electrode, range 0 – 13 pH Resolution : 0.01 pH with automatic temperature compensation Pump : High quality pumps with flow rates 0 – 10 Litre/hr. Pump connection : Upto 4 pumps with progressive proportional flow rate control O2 measuring regulation : 0 – 25 mg/lt. Probe: PEEK Membrane: Glass reinforced PTFE Resolution: 0.1 mg/ml Gas input : Max. air pressure 0.2 MPa Airflow meter : Range: 0 – 5 lt./min. Resolution: 0.1 lt./min. Reproducibility: ± 0.5% Remote control & Data Processing : Software for fermenters
S. No. 16
Autoclave
Technical Specifications
Chamber Size : 500 mm (W) x 500 mm (L) x 950mm (H) Electrical Power : 8 KW Or Sufficient wattage of industrial immersion type water heater to generate steam within a reasonable period of time on 3 phase 440V/ 50 Hz AC supply Working pressure : 1.2 to 2.2 Kg/sq.cm Working Temperature : at 121°C
Material of Construction : Inner chamber, Jacket, Door : SS 316.(5mm-10mm) Outer Chamber: SS 304 (Insulated properly)
Steam Generator : Non corrosive SS / Chromium plated Brass Heater Plate : Brass / Stainless Steel Pipe Line : Complete with SS Stand : Stainless Steel / High quality non corrosive steel Safety devices : Pressure switch and safety valve, self locking of door when chamber is under pressure
Vacuum breaker for jacket Steam generator with gauge glass valves and Low water protection with audio visual indicator.
Features : The equipment should have ISI mark on quality. Temperature, Pressure and Vacuum gauges : Steam
traps, vacuum driers, water level indicator on steam generator
S. No. 17
Incubator
Technical Specifications
Volume : 85 Litres Temperature Range : 5 - 60°C Temperature Accuracy : ± 0.5°C
Working space : Width – Min. 45.0 cm Depth – Min. 40.0 cm Height – Min. 55.0 cm External material : MS with powder coating with proper insulation No. of SS perforated Shelves : 02 Voltage Stabilizer : 1 kVA Chamber internal material : SS 304
S. No. 18
Optical Microscope with Hot Stage
Technical Specifications
Microscope Mains Operations : 220 Volts,50Hzs,Single Phase AC supply. Screen : 360° Rotatable Graduated Screen Light Control : Variable Controlled Light Intensity Meter Objective : 5x, 10x, 20x & 40x Magnification : 125x to 1000x.
Hot stage Temperature range : 50 - 400°C Tempearture accuracy : 0.1°C Heating rate : Min. 0.1°C / hr and Max. 100°C / min.
Temperature control method : PID controller Temperature sensor : 100 Ω Platinum RTD Min. Objective Working Distance : 6.1 mm Sample Area : 22 mm x 25 mm Chamber height : 5.0 mm Sample Viewing Aperture : 5 – 7 mm diameter
S. No. 19
FTIR with Microscope
Technical Specifications
Wave number : Minimum 7800-350 cmˉ¹ Signal to Noise Ratio : 6000:1 Wave number Accuracy : 0.02 cmˉ¹ Resolution : Minimum 0.5 cmˉ¹ Detector : FR-DTGS or LiTaO3 or MCT array detector with KBr
Beam splitter in addition to micro ATR objective FTIR Software : Window based for data manipulation and analysis and
Polymer data library. Atmosphere compensation : Automatic Atmospheric compensation Validation Standards : Automatic Validation including certified standards Instrument Calibration : Traceable standard material Humidity sensors : Physical as well as electronic humidity sensor Sample holder : Holders suitable for powder / solid samples should be
provided Accessories : Automatic optimization and calibration, Suitable work station : GC compatible : Search facility in library for plastics and additives Microscope : Microscope should have visible and infrared mode
Fully automated with motorised sample stage Integral CCD video camera and video capture card for PC Auto aperture, One touch auto focus and mapping Auto marker and auto illumination Auto view and Auto correction Simultaneous variable magnification, ATR Mapping IR image display as pixels images displays Upgradeability to IR imaging,
Signal to noise ratio for microscope : 6000:1 mrs Polarizer : Visible and IR polarizer Auto mode : Transmission, Reflectance and micro ATR
S. No. 20
Ultra Violet Visible Double Beam spectrophotometer
Technical Specifications Wavelength : 190-1100 nm Spectral Bandwidth : 1 nm (190 -1100 nm) Wavelength Display : 0.1 nm increments Wavelength Setting : 0.1 nm increments Wavelength accuracy : ± 0.1 nm at 656.1 nm D2
± 0.3 nm (190 – 110 nm) Scanning speed : Fast / medium / slow Wavelength repeatability : ± 0.1 nm Stray light : < 0.02% at 220 nm and 340 nm
< 1.0% at 198 nm Photometric System : Double beam Photometric range : Absorbance : 0-4 to 4 : Transmittance : 0 to 400% Photometric accuracy : ± 0.002 abs (0.5 abs)
± 0.004 abs (1.0 abs) ± 0.006 abs (2.0 abs)
Photometric repeatability : < ± 0.001 abs (0.5 abs) < ± 0.001 abs (1 abs) < ± 0.003 abs (2.0 abs)
Baseline stability : < 0.0003 abs/hr at 700 mm(one hr after light source turned ON)
Baseline flatness : Within ± 0.0006 abs (190 nm – 1100 nm, one hr after light source turned)
Noise Level : Within 0.00005 abs RMS value (at 700 nm) Printers : DPU,ESC/P, PCL Printers, USB I/F windows –
complaint printers are available with USB memory and PC software
Memory : USB memory saved as text and UVPC file Performance of PC : USB Memory + UV probe WinXP Detector : Dual silicon photodiode Power : AC 230 V / 50 HZ Sample holders : Square cell with stopper and cylindrical cell
made of fused silica or quartz with temperature controller
Software : Analysis software should be provided
S. No. 21
Universal Testing Machine
Technical Specifications
Maximum Load Capacity : 50 KN (Microprocessor controlled) Cross head Travel distance : Min 1000 mm Horizontal daylight : Min. 500mm Test Speed : Variable between 0.05 mm / min to 1000 mm / min. Load cells : 100 N, 1 KN, 5 KN & 50 KN Load Accuracy : ≤ 0.5 %, Fixtures : Tensile, compression, flexural, and shear fixtures. Tensile grips : Suitable to Plastic films (rubber faced grips), Rigid plastics (self lock winch grip, opening up to 12mm) woven sacks (50mm width), rubber, fibre/filament. Power : 230 Volts, 50 HZ, single phase. Conform to : ASTM, ISO, DIN, BS and IS standards. Extensometer : Contact Strain guage, Output : Software attached & data storage for sample test
methods Grips : Neumatic and Manual Accuracy for Cross head speed : ± 0.1 mm Safety lock provisions Software : Suitable for all functions such as max. load; strain,
modulus, etc
Calibration certificate for load cells traceable to NIST should be provided
S. No. 22 High Temperature Gel Permeation Chromatograph
Technical Specifications
GPC Pump
Flow rate Range: 0.001 to 5 ml / min.
Flow precision: <0.07 % RSD
Operating pressure: Upto ~ 5800 psi
Flow rate accuracy: ± 1%
Leak sensors Provided. CAN, LAN, HP-1B, RS 232, APG
remote – start are interfaced to communicated the modules with
computer
Vacuum solvent degassing module is Provided.
GPC injector
Capacity: 30 vials
Operating temperature: 220°C.
GPC soft ware
PC controlled soft ware to control and monitor all functions
including pump, detector, Oven, autosampler and solvent vapour
sensors
Detector
a. UV-Visible Variable Wavelength detector
Detector type: Monochromator (UV detector)
Wavelength range: 190 -0740 nm
Wavelength accuracy: ± 3 nm
Sources: Deutirium
Noise: < 20 x 10-5
AU
Drift: 1.5 x 10-4
AU/h
Time constants: 0.1 / 0.2 / 0.5 / 1 / 2 / 5 / 10
Path length: 10 mm
b. Refractive index detector
RI range: 1 – 1.75 RIU
Cell volume: 10µl
Sensitivity: < 2 x 10-7
RIU / mV
Noise: < 3 x 10-8
RIU / hr.
S. No. 23
CHN Analyzer
Technical Specifications
Operating Principle: Based on classical technique of Pregi & Dumas. The system includes patented Pressurised Oxygen Injection System (TurboFlash Technology) for perfecting combustion in widest variety of samples matrices. Operating Mode: simultaneous (CHNS); (CNS); (S); (O); (CHN); (CN); (N). Full field up gradability from one operating mode to another through full modularity. Detection Principle: Separation of combustion product on GC Column followed by detection using Supersensitive Thermal Conductivity detector Instrument Features: Oven temperature range : From 40°c to 190°C Furnace Temperature : Up to 1100°C for all operating mode Combustion temperature : Up to 1800 °C Oxygen injection (volume) : Volume set via S/W Oxygen injection (pressure) : Set Independently from Carrier velocity via S/W Gas leak test : Automatic by keypad push button or via S/W Gas flow rate in stand-by : Zero flow in Gas Save Gas save in stand-by : Automatically selectable by keypad or S/W Catalyst & TCD protection : Automatic isolation with gas save routine Exhausted reactor change : Bayonet-lock action from outside of the instrument Sample introduction : 80 sample via electrically driven AutoSampler Sample Type : Solid, Viscous and liquid SW Instrument Control : Fully automatic control of all parameters and results reporting by user selected format. Set up guide for Windows XP/ME SW : Real time chromatogram display. Storing of an unlimited methods. Ability to reprocess individual runs and batches of samples. AutoRun & AutoCal. AutoReady & Wake up routine. Automatic Leak test. Export on Excel Calibration : Traditional K-factor or linear regression Safety : Full fail-safe operation. Sensors for high temperature compartment and transformer. Dual redundant control valving for Oxygen delivery circuit Shut-off safety interlock for users accessing the hot compartment
Instruments Performances: Practical Working Range 0,01% up to 100% Precision: Using Cystine as Reference Std Material and Sartorius Mod CP2P, MicroBalance, weighing 5-6 standard within 0,5 mg up to 1,5 for linear, regression calibration the typical results obtained are From 0,20% or up to 0,30% depending on Element % content Speed of Analysis : CHNS/CNS = 5 min; CHN = 3 minutes; for Oxygen = 2 min
S. No. 24
Horizontal Laminar air flow Specification
Technical Specifications Air Velocity : Between 0.4 to 0.5 mls +-20% HEPA Efficiency Pre-filter : 99.997% at 0.3 microns with DOP test Washable with an arrestance up to 90% at 5 microns Less than 65 dB (A) Sound Level : Lux 900 - 1300 Lux Light Intensity Cleanliness : Standard 209 E Class 100 Federal Power Supply : 220 V Single phase 50 Hz
Others:
S. S Table Top.
U. V Lights attachment
Front Acrylic Door
Gas/ Vacuum Cock
Castor wheels at the Bottom
Construction:
The cabinets are fabricated out of thick board duly sunmica clad or mild steel cabinet duly epoxy coated.
The work table is made of thick board which is sunmica clad at top or S. S Table top will be provided as
optional accessories. All joints are sealed with silastic sealant. Side panels are provided of thick
transparent acrylic sheets. The unit is fitted with pre-filter and is made to pass through highly effective
HEPA (High Efficiency Particular Air) filter having efficiency rating as high as 99.997% with hot, DOP,
thus retaining all air-borne particles of size : 0.3 micron and larger.
Using a dynamic machine, the blower and motor assembly is statically and dynamically balanced.
Blowers are mounted on vibration isolation mounts to minimize vibration. The working is illuminated by
fluorescent lightings fitted to the unit. Height of the working table provides comfortable “ Sit Down”
working position for the operation to work on 220/230 volts AC. supply.
S. No. 25
Water vapour Transmission Tester
Technical Specifications
Measuring range : 0.03 – 10000 g/m2 /day
Measuring accuracy : 0.001 g/m2 /day
Sample surface area : 50cm2
Minimum sample size : 100 x 100mm
Thickness of sample : 0.01 – 12mm
Voltage settings : 230 V AC / 50 Hz
Power consumption : Max. 300 VA
Environmental temperature control : 5-40°C
Temperature control accuracy : ± 0.5°C
Communication interface : RS 232 – Interface
Warning : Hot surface (over 70°c)
Equilibrium Detector : 0 – 99.9%
Reproducibility : High permeability samples : better
than 1%
Low permeability samples better than
5%
Humidity measuring range : 40 – 95% RH in measuring chamber
(std : 90%)
Humidity accuracy : ± 1%
Keyboard : Alphanumeric
Data storage by Lithium battery : Typically >10 years
Typical time consumption for a simple
measurement
Range 1g / m2 / 24 hrs
20g / m2 / 24 hrs
>1000 g/m2/ 24 hrs
:
:
:
:
< 1 hour
< 10 minutes
< 5 minutes
Conforms to : ASTM E96, ASTM D 1653, ISO
2528, DIN 53122
S. No. 26
Gas Permeability Tester
Technical Specifications
Measuring range : 0.02- 50000 cc/m² day
Measurement accuracy : 0.01 cc/ m² day
Temperature range : 15⁰C - 45⁰C (std: 23⁰C)
Temperature accuracy : ± 0.01⁰C
Test pressure : 0.1 pa
Suitable for : O2, CO2, N2 gas operation
Air pressure : 0.1 MPa – 0.8 MPa
Specimen size : Diameter - 80 mm, Area: 50 cm²
Sample : 3 (extension up to more pieces)
Carrier gas interface (inlet) : 1/8 inch
Sample thickness : ≤ 2 mm
Power : AC 220 V / 50 HZ
Humidity control : 0%, 35-90%, 100% RH
Humidity control accuracy : ± 3% RH
Interface : RS-232 interface
Data collection & storage : Windows based software suitable for data
collection, storage and analysis.
Confirms to Standard : ASTM D 1434, IS :13360, ISO 2556
PROCESSING MACHINERY
EXTRUSION BLOWN FILM UNIT WITH CARONA TREATER
EXTRUSION BLOWN FILM UNIT WITH CARONA TREATER
S. No Items Unit Specifications Model :
1. Screw Diameter mm 30-40
2. Screw Speed rpm Please specify
3. Screw L/D Ratio - Please specify
4. Extruder Output Kg/hr 10 - 35
5. Film Thickness mm/Micron Please specify
6. Material to be process - All Thermoplastics
7. LFW (Min & Max.) mm Please specify
8. Control System - Microprocessor
9. Synchronisation of Extruder drive with Haul off unit etc
- Please specify
10. Melt temperature sensor & Melt Pressure transducers
- Please specify & quote
11. Die for HMHDPE & LLDPE - Please specify & quote
12. Die gap mm Please specify
13. Corona Treater suitable for above machine
- Please specify & quote
14. Safety
- Appropriate safety features to be provided
15. Machine dimensions (LxWxH) - Please specify
16.
Essential Spares a. Temperature controller- 02 nos. b. Thermocouples- 02 nos. c. Heaters- 01 set d. Synchronisation card -01 nos.
-
Please Specify & Quote
17. Optional a) Pipe/tube attachment b) Pelletizing attachment
- -
Please specify & quote Please specify & quote