certificate - mettler toledo€¦ · this certificate may only be reproduced in its entirety and...

3
DEKRA Certification B.V. Ë M. Erdhuizen Certification Manager Page 1/3 © Integral publication of this certificate and adjoining reports is allowed. This Certificate may only be reproduced in its entirety and without any change. DEKRA Certification B.V. Meander 1051, 6825 MJ Arnhem P.O. Box 5185, 6802 ED Arnhem The Netherlands T +31 88 96 83000 F +31 88 96 83100 www.dekra-certification.com Registered Arnhem 09085396 CERTIFICATE (1) EC-Type Examination (2) Equipment and protective systems intended for use in potentially explosive atmospheres - Directive 94/9/EC (3) EC-Type Examination Certificate Number: DEKRA 15ATEX0015 X Issue Number: 1 (4) Equipment: Load Cell model SLC611 (5) Manufacturer: Mettler-Toledo (Changzhou) Measurement Technology Ltd. (6) Address: No. 111 West Taihu Road, Xinbei District, Changzhou, Jiangsu, 213125, P.R. China (7) This equipment and any acceptable variation thereto is specified in the schedule to this certificate and the documents therein referred to. (8) DEKRA Certification B.V., notified body number 0344 in accordance with Article 9 of the Council Directive 94/9/EC of 23 March 1994, certifies that this equipment has been found to comply with the Essential Health and Safety Requirements relating to the design and construction of equipment and protective systems intended for use in potentially explosive atmospheres given in Annex II to the directive. The examination and test results are recorded in confidential test report number 217429900 issue 1. (9) Compliance with the Essential Health and Safety Requirements has been assured by compliance with: EN 60079-0 : 2012 + A11 EN 60079-11 : 2012 EN 60079-26 : 2007 EN 60079-15 : 2010 EN 60079-31 : 2009 (10) If the sign ”X” is placed after the certificate number, it indicates that the equipment is subject to special conditions for safe use specified in the schedule to this certificate. (11) This EC-Type Examination Certificate relates only to the design, examination and tests of the specified equipment according to the Directive 94/9/EC. Further requirements of the directive apply to the manufacturing process and supply of this equipment. These are not covered by this certificate. (12) The marking of the equipment shall include the following: II 1 G Ex ia IIC T6 … T4 Ga II 1 D Ex ia IIIC T100 °C Da II 3 G Ex ic IIC T6 … T4 Gc II 3 G Ex nA IIC T6 … T4 Gc II 3 D Ex tc IIIC T100 ºC Dc This certificate is issued on 8 May 2015 and, as far as applicable, shall be revised before the date of cessation of presumption of conformity of (one of) the standards mentioned above as communicated in the Official Journal of the European Union.

Upload: others

Post on 12-Oct-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: CERTIFICATE - Mettler Toledo€¦ · This Certificate may only be reproduced in its entirety and without any change. DEKRA Certification B.V. Meander 1051, 6825 MJ Arnhem P.O. Box

DEKRA Certification B.V.

ËM. ErdhuizenCertification Manager Page 1/3

©Integral publication of this certificate and adjoining reports is allowed. This Certificate may only be reproducedin its entirety and without any change.

DEKRA Certification B.V. Meander 1051, 6825 MJ Arnhem P.O. Box 5185, 6802 ED Arnhem The NetherlandsT +31 88 96 83000 F +31 88 96 83100 www.dekra-certification.com Registered Arnhem 09085396

CERTIFICATE(1) EC-Type Examination(2) Equipment and protective systems intended for use in

potentially explosive atmospheres - Directive 94/9/EC

(3) EC-Type Examination Certificate Number: DEKRA 15ATEX0015 X Issue Number: 1

(4) Equipment: Load Cell model SLC611

(5) Manufacturer: Mettler-Toledo (Changzhou) Measurement Technology Ltd.

(6) Address: No. 111 West Taihu Road, Xinbei District, Changzhou, Jiangsu, 213125, P.R. China

(7) This equipment and any acceptable variation thereto is specified in the schedule to this certificate and the documents therein referred to.

(8) DEKRA Certification B.V., notified body number 0344 in accordance with Article 9 of the Council Directive 94/9/EC of 23 March 1994, certifies that this equipment has been found to comply with the Essential Health and Safety Requirements relating to the design and construction of equipment and protective systems intended for use in potentially explosive atmospheres given in Annex II to the directive.

The examination and test results are recorded in confidential test report number 217429900 issue 1.

(9) Compliance with the Essential Health and Safety Requirements has been assured by compliance with:

EN 60079-0 : 2012 + A11 EN 60079-11 : 2012 EN 60079-26 : 2007EN 60079-15 : 2010 EN 60079-31 : 2009

(10) If the sign ”X” is placed after the certificate number, it indicates that the equipment is subject to special conditions for safe use specified in the schedule to this certificate.

(11) This EC-Type Examination Certificate relates only to the design, examination and tests of the specified equipment according to the Directive 94/9/EC. Further requirements of the directive apply to the manufacturing process and supply of this equipment. These are not covered by this certificate.

(12) The marking of the equipment shall include the following:

II 1 G Ex ia IIC T6 … T4 GaII 1 D Ex ia IIIC T100 °C DaII 3 G Ex ic IIC T6 … T4 GcII 3 G Ex nA IIC T6 … T4 GcII 3 D Ex tc IIIC T100 ºC Dc

This certificate is issued on 8 May 2015 and, as far as applicable, shall be revised before the date of cessation of presumption of conformity of (one of) the standards mentioned above as communicated in the Official Journal of the European Union.

Page 2: CERTIFICATE - Mettler Toledo€¦ · This Certificate may only be reproduced in its entirety and without any change. DEKRA Certification B.V. Meander 1051, 6825 MJ Arnhem P.O. Box

(13) SCHEDULE

(14) to EC-Type Examination Certificate DEKRA 15ATEX0015 X Issue No. 1

Page 2/3

Form 100Version 5 (2013-07)

(15) Description

The Load Cell model SLC611 is used to convert a mechanical force or load into an electrical signal. The load cell is of a sealed construction and is provided with a permanently connected cable with a maximum length of 30 m.

The enclosure of the load cell provides a degree of protection of at least IP68 in accordance with EN 60079-0.

The maximum surface temperature T100 °C is based on an ambient temperature of +50 °C.Ex tc IIIC, Ex ia IIIC is applicable for a maximum layer of dust of 5 mm.

Electrical data

Signal and supply: in type of protection intrinsic safety Ex ia IIC, Ex ia IIIC and Ex ic IIC, only for connection to a certified intrinsically safe circuit, with the following maximum values (combining the parameters of all circuits):Ui = 25 V; Ii = 600 mA; Ci = 6 nF; Li = 30 µH.

The values of Ci and Li include the capacitance and inductance of the permanently connected cable for a length of maximum 30 m. For longer cables the additional capacitance and inductance has to be taken into account (0,2 nF/m and 1 µH/m).

For the type of protection Ex nA IIC or Ex tc IIIC:Signal and supply:Un = 25 V.

For gas application, the relation between maximum input power, maximum ambient temperatureand temperature class is as listed in the following table:

Temperature class

T6 T5 T4

Maximum input power

Pi = 0,51 W Pi = 0,86 W Pi = 1,25 W

Ambient temperature range

-40 °C to +45 °C -40 °C to +50 °C -40 °C to +50 °C

Installation instructions

The instructions provided with the equipment shall be followed in detail to assure safe operation.

(16) Test Report

No. 217429900 issue 1.

Page 3: CERTIFICATE - Mettler Toledo€¦ · This Certificate may only be reproduced in its entirety and without any change. DEKRA Certification B.V. Meander 1051, 6825 MJ Arnhem P.O. Box

(13) SCHEDULE

(14) to EC-Type Examination Certificate DEKRA 15ATEX0015 X Issue No. 1

Page 3/3

Form 100Version 5 (2013-07)

(17) Special conditions for safe use

The load cell and cable gland had been tested for the low risk of mechanical danger (drop height 0,4 m with 1 kg mass). The load cell shall therefore be protected against higher impact energy levels.

Provisions shall be made to prevent the rated voltage from being exceeded by transient disturbances of more than 40% for the type of protection Ex nA IIC.

(18) Essential Health and Safety Requirements

Covered by the standards listed at (9).

(19) Test documentation

As listed in Test Report No. 217429900 issue 1.