Classifying Chemicals1. Reagent Grade: Reagent grade chemical conform to the minimum standards set forth by the Reagent Chemical committee of the American Chemical Society (ACS) and are used wherever possible in analytical work.2. Primary Standard Grade: Extraordinary purity is required for a primary standard. Primary standard reagent is carefully analyzed and the assay is printed on the container label. 3. Special-Purpose Reagent: chemicals that have been prepared for a specific application. Included among these are solvents for spectrophotometry and high-performance liquid chromatography.

Page 3: Chapter 2

1. Select the best grade of chemical available for analytical work.2. Replace the top of every container immediately after removal of the reagent.3. Hold the stoppers of reagent bottles between your fingers.4. Never return any excess reagent to a bottle.5. Never insert spatulas, spoons, or knives into a bottle that contains a solid chemicals.6. Keep the reagent shelf and the laboratory balance clean and neat.7. Observe regulations concerning the disposal of surplus reagents and solutions.

Page 4: Chapter 2

Use liquid detergent (preferably hot) Finish up with distilled water Use deionized water (if available) For water insoluble solution, use alcohol or

acetone Rinsing with reagent – needed especially in

titration“It is seldom necessary to dry the interior

surface of glassware before use: it can be a waste of time and potential source of contamination”

Page 5: Chapter 2

Marking Laboratory Ware

A chemical analysis is ordinarily performed in duplicate or triplicate. Each vessel that holds a sample must be marked so that its content can be positively identified. Flask, beaker and some crucibles have small etched areas on which semi permanent mark can be made with a pencil.

Saturated iron(III) chloride can be used for markings.

Page 6: Chapter 2

Problem: Overheating Bumping cause partial loss of the solution

Ans: Careful and apply gentle heating

glass beads minimize bumping

Page 7: Chapter 2

An analytical balance is a weighing instrument with a maximum capacity that ranges from 1 g to a few kilograms with a precision of at least 1 part in 105 at maximum capacity.Macrobalances have a maximum capacity ranging between 160 and 200 g; measurement can be made with a standard deviation of 0.1mg.Semimicroanalytical balances have a maximum load of 10 to 30 g with a precision of 0.01mg.Microanalytical balance has a capacity of 1 to 3 g and a precision of 0.001mg.

Page 8: Chapter 2
Page 9: Chapter 2
Page 10: Chapter 2

Bouyancy error - density of the sample is very much different from the standard masses.

Temperature Effect – Failure to allow sufficient time for a heated object to return to room temperature.

Page 11: Chapter 2

Top Loading – 150 to 200 g25,000 g

Triple Beam balance -

Page 12: Chapter 2

Oven drying is the most common way of removing moisture from solids. This approach is not appropriate for substances that decompose or for those from which water is not removed at the temperature of the oven.Dried material are stored in desiccator while they cool so as to minimize the uptake of moisture. The base section of the desiccator contains a chemical drying agent (desiccants) such as anhydrous calcium chloride, calcium sulfate, magnesium perchlorate or phosphorus pentoxide.

Page 13: Chapter 2
Page 14: Chapter 2

Weighing by DifferenceWeighing by difference is a simple method for determining a series of sample weights. First the bottle and its contents are weighed. One sample is then transferred from the bottle to a container; gentle tapping of the bottle with its top and slight rotation of the bottle control over the amount of sample removed. Following transfer, and its residual contents are weighed. The mass of the sample is the difference between the two weighings.

Page 15: Chapter 2
Page 16: Chapter 2

Simple CruciblesSimple crucibles serve only as containers. Porcelain, aluminum oxide, silica and platinum crucibles maintain constant mass and are used principally to convert a precipitate into a suitable weighing form. The solid is first collected on filter paper. The filter and contents are then transferred to a weighed crucible, and the paper is ignited.

Filtering CruciblesFiltering crucibles serve not only as containers but also as filters. A vacuum is used to hasten the filtration, a tight seal between crucible and filtering flask is accomplished with any of the several types of rubber adapters.

Page 17: Chapter 2

Sintered-glass CruciblesSintered-glass crucibles are manufactured in fine, medium, and coarse porosities. The upper temperature limit for sintered glass crucible is ordinarily about 200oC. Filtering crucibles made entirely of quartz can tolerate substantially higher temperatures.

Filter PaperPaper is an important filtering medium. Ashless paper is manufactured from cellulose fibers that have been treated with hydrochloric and hydrofluoric acids to remove metallic impurities and silica, ammonia is then used to neutralize the acids. It is necessary to destroy the paper by ignition if the precipitate collected on it is to be weighed.

Page 18: Chapter 2
Page 19: Chapter 2
Page 20: Chapter 2
Page 21: Chapter 2

Heating EquipmentMany precipitate can be weighed directly after being brought to constant mass in a low temperature drying oven. Such an oven is electrically heated and capable of maintaining a constant temperature to within 1oC. The maximum attainable temperature ranges from 140 to 260oC, depending on make and model, for many precipitate 110oC is a satisfactory drying temperature.Microwave laboratory ovens are currently appearing on the market. Where applicable, these greatly shorten drying cycles.Muffle furnace (a heavy duty electric furnace) is capable of maintaining controlled temperatures of 1100oC or higher. Long handled tongs and heat resistance gloves are needed for protection.

Page 22: Chapter 2

Pipets

Burets

Volumetric flask

Measuring cylinder

Page 23: Chapter 2
Page 24: Chapter 2
Page 25: Chapter 2
Page 26: Chapter 2
Page 27: Chapter 2
Page 28: Chapter 2

Volumetric glassware is calibrated by measuring the mass of a liquid (water) of known density and temperature that is contained in the volumetric ware.

Page 29: Chapter 2

1. Record all data and observations 2. Supply each entry with a heading3. Date each page of the notebook4. Never attempt to erase an incorrect entry,

cross it out with single horizontal line5. Never remove a page from the notebook6. Do not overcrowd entries7. Keep first few pages for table of contents

Page 30: Chapter 2

Take a picture of 3 instruments present in the laboratory, paste or print directly in the bond paper with description or usage of the instruments.