chapter 4 spectral characteristics of organic...
TRANSCRIPT
94
CHAPTER 4
SPECTRAL CHARACTERISTICS OF ORGANIC DYES
4.1 INTRODUCTION
Materials with large and fast nonlinear optical response are required for the development of photonics devices such as optical switches and optical limiters .The essential requirements of good photonic materials are high non linearity, synthetic flexibility and ease of processing. Extremely large numbers of organic compounds with delocalized electron, conjugated double bond systems and a large dipole moment have been synthesized to realize the susceptibilities far larger than the inorganic optical materials. It is thus important to understand the influence of various photo processes of the dyes on their nonlinear response. To study the nonlinear optical processes of the dyes, their lasing behavior must be elucidated. The spectral characteristics of these dyes give an insight about their lasing characteristics.
Solvents play an important role in the absorption and emission
characteristics of dyes. The solvent medium can cause displacement of the
fluorescence and absorption spectra of the dyes. Hence, the position and
structure of the absorption and emission spectra of molecules in solutions
are dependent on the solvent medium used. The solvent medium can force the
solute particles to get distributed in a spectral fashion in space through solute-
solvent interactions (Fleming and Wolynes 1990). So by recording the
absorption spectra, fluorescence spectra and the lifetime of the dyes, the effect
of medium on their spectral characteristics can be studied (Dyumaev et al
1992).
95
Time- resolved measurements are widely used in fluorescence spectroscopy because they contain more information than the steady- state data. Information on the number of fluorescence entities and number of different sites of binding a single fluorophore is available through time-resolved measurements. Lifetime τ, is measured as the average time the molecule spends in the excited state prior to the return to the ground state. Generally, fluorescence lifetimes are in the range of nano to pico seconds. The fluorescence lifetime can function as a molecular stopwatch to observe a variety of molecular events which occur in the above timescale.
In this chapter a study of the absorption and fluorescence spectra and lifetimes of the dyes in liquid environment are studied. The absorption curve does not mention the nonlinear effects as the light source in the spectrophotometer is insufficient to cause these effects, yet the absorption measurements is used to determine the suitable wavelength of the light source for which the optical material can act as a optical limiter. The molecular structures of dyes chosen are shown in Figure 2.1(a-f). Therefore, thorough investigations of these dyes in liquid media are necessary to understand the lasing characteristics dyes in solvent. The solvent medium can force the solute particle to get solute-solvent interactions. Physical properties of the solvents that have been used in this study are shown in Table 4.1.
Table 4.1 Physical properties of the solvents that have been used in
this study
Solvent Polarity
index Dipole
moment Linear ref. index (no)
Dielectric constant
Density gm/lit
Viscosity
Ethanol 5.2 1.69 D 1.361 24 0.789 0.949
Methanol 5.0 1.69 D 1.328 33 0.791 0.593
Acetonitrile 5.8 3.44D 1.344 37.5 0.786 0.380
96
4.2 SPECTRAL CHARATERISTICS OF ANTHRAQUINONE
DYES
The anthraquinone dyes ( alizarin, alizarin red and alizarin cyanin
green) are obtained from Loba Chemie, India. The molecular structures of
these dyes are given in Figure 2.1. (a) - (i-iii). Thin layer chromatography
(TLC) test confirms the absence of any impurities. These dyes were used as
received. Acetonitrile was chosen as solvent since these dyes were readily
soluble in it.
4.2.1 Absorption and Fluorescence Spectra of Anthraquinone Dyes
The absorption spectra of the anthraquinone dyes of 0.01mM
concentration in acetonitrile are recorded and shown in Figure 4.1.The
fluorescence spectra of the dyes in acetonitrile for 0.01mM concentration are
recorded and shown in Figure 4.2 a-c. The peak wavelengths of the absorption
and fluorescence spectra are measured and the spectral parameters such as
molecular absorption coefficient ((λ)), oscillator strength (f), absorption
bandwidth ()1/2, fluorescence bandwidth (FWHM) and Stoke’s shift are
calculated and given in Table 4.2.
4.2.2 Fluorescence Lifetime Measurements
The fluorescence decay measurements of anthraquinone dyes
were recorded using IBH time correlated single photon counting spectrometer
(TCSPC). The fluorescence decay profile and the residuals of the dye alizarin
red, alizarin cyanin green and alizarin in the solvent acetonitrile are shown in
Figure 4.3 a-c. The residuals shown in the Figure are well within the error
limits. Alizarin red, alizarin cyanin green exhibits bi-exponential decay
while alizarin exhibits a tri – exponential decay. The average life time values
of the anthraquinone dyes are shown in Table 4.2.
97
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
400 500 600 700 800
A
B
C
Wavelength (nm)
Ab
sorb
anc
e
Figure 4.1 Absorption spectra of anthraquinone dyes in acetonitrile
(A) Alizarin (B) Alizarin Red (C) Alizarin Cyanin Green
0
2000
4000
6000
8000
10000
12000
14000
16000
600 650 700 750 800 850
Wavelength (nm)
Fluo
resc
ence
Int
ensi
ty
Figure 4.2.a Fluorescence spectra of alizarin dye in acetonitrile
98
0
5000
10000
15000
20000
25000
500 600 700 800 900Wave length (nm)
Fluo
resc
ence
Int
ensi
ty
Figure 4.2.b Fluorescence spectra of alizarin red dye in acetonitrile
0
0.2
0.4
0.6
0.8
1
1.2
650 700 750 800 850Wavelength (nm)
Fluo
resc
enc
e In
tens
ity
Figure 4.2.c Fluorescence spectra of alizarin cyanin dye in acetonitrile
99
-6
-4
-2
0
2
4
6
4 9 14 19
Residuals
Time (ns)
Sta
ndar
d D
evia
tion
-3
-2
-1
0
1
2
3
3 8 13 18
Residuals
Time (ns)Sta
ndar
d D
evi
ati
on
1
10
100
1000
10000
100000
0 10 20 30 40 50 60
Prompt
Decay
Fit
ALIZARIN
Log
(Cou
nts)
Time (ns)
-6
-4
-2
0
2
4
2 12 22 32 42
Residuals
Time (ns)
Sta
ndar
d D
evi
atio
n
1
10
100
1000
10000
0 10 20 30 40
Prompt
Decay
Fit
ALIZARIN RED
Time (ns)
Log
(Cou
nts)
1
10
100
1000
10000
0 10 20 30 40 50
Prompt
Decay
Fit
ALIZARIN CYANIN GREEN
Time (ns)
Log
(Co
unts
)
Figure 4.3.a Fluorescence decay profile and residuals of alizarin dye
Figure 4.3.b Fluorescence decay profile and residuals of alizarin red dye
Figure 4.3.c Fluorescence decay profile and residuals of alizarin cyanin
green dye
100
Table 4.2 Spectral parameters of antharquinone dyes in acetonitrile
Dye
Absorption spectra
Fluorescence spectra Average
life time (ns)
Peak wave- length
nm
(λ) 104 L mol-1 cm-1
()1/2 cm-1
Oscillator Strength fx10-24 L mol-1
cm-2
Peak wave- length
nm
FWHM nm
Stoke’s shift cm-1
Alizarin 643 6.65 1100 0.317 715 113 1513 1.00 Alizarin Red
600 5.59 2500 0.605 659 133 1485 0.52
Alizarin Cyanin Green
588 6.29 4000 1.089 689 36 1000 0.68
4.2.3 Results and Discussion
The longest wavelength absorption band of the anthraquinone dyes
are shown in Figure 4.1. The absorption maxima in nm are shown in Table
4.2. The absorption wavelength maxima decrease in the order of alizarin >
aizarin red > alizarin cyanin green. This variation is related to the conjugation
of π electronic cloud in the fused aromatic system, which is further altered by
the substituents on anthraquinone nucleus. In alizarin, two – electron releasing
OH- groups are bound to anthraquinone nucleus. In alizarin red there is an
electron withdrawing SO3Na grouping. As a result the absorption maximum is
blue shifted to a lower wavelength value (600 nm) in alizarin red compared to
alizarin (643nm). Hence, in alizarin red, the electron cloud will be more
tightly held than in alizarin. In alizarin cyanin green, the electron releasing
hydroxyl groups of alizarin are replaced by aniline groupings, but both the
aniline groupings carry electron withdrawing SO3Na grouping. As a result the
electron release from aniline groupings to anthraquinone nucleus is largely
101
retarded by SO3Na grouping so the electronic cloud is still more tightly held
than in alizarin red. Hence, the absorption maximum is shifted to shorter
wavelength (588 nm) than in alizarin red.
The maximum absorption wavelength of anthraquinone dyes is
largely influenced by the substituents bounded to anthraquinone nucleus. The
electron releasing substituents make the electron cloud of anthraquinone
nucleus less tightly held, thus shifting the absorption maximum to a longer
wavelength, while electron withdrawing grouping shift the absorption
maximum towards shorter wavelength.
The fluorescence spectra of the dyes are shown in Figure 4.2 a-c.
All the three dyes show fluorescence emission at longer wavelength than their
respective absorption maxima. Hence, loss of energy by vibrational cascade
leads to occupation of lower energy level. Generally, the vibrational cascade
will be more significant for flexible molecular systems than rigid molecular
systems. It is evident from the fluorescence spectrum of alizarin cyanin green,
as it is more flexible than other two dyes, due to two bulky anilino derivatives
bound to anthraquinone nucleus.
The average lifetime of anthraquinone dyes are shown in Table 4.2.
Alizarin with its rigid molecular structure shows larger lifetime than other two
dyes. Alizarin cyanin green shows longer lifetime than alizarin red. Hence,
the anilino derivatives bound to anthraquinone nucleus in alizarin cyanin
green has less influence on the lifetime of the excited species, compared to
alizarin red, where the substituents, SO3Na is directly bound to the
anthraquinone nucleus. The solvent acetonitrile can solvate SO3Na of alizarin
red and aids rapid vibrational cascade. In alizarin cyanin green SO3Na groups
are very far from anthraquinone nucleus and hence, it has less influence on
the lifetime of anthraquinone nucleus.
102
4.3 SPECTRAL CHARACTERISTICS OF AZO DYES
The azo dyes (oil red o and metanil yellow) are obtained from loba
chemie, India. The molecular structure of these dyes were given in Figure 2.1.
(b) - (i-ii). Thin layer chromatography (TLC) test confirms the absence of any
impurities. These dyes were used as received. Acetonitrile was chosen as
solvent, since these dyes were readily soluble in it.
4.3.1 Absorption and Fluorescence Spectra of Azo Dyes
The absorption spectra of azo dyes in acetonitrile for 0.01mM
concentration are recorded and shown in Figure 4.4.The emission spectra of
azo dyes in acetonitrile for 0.01mM concentration are recorded and shown in
Figure 4.5.The peak wavelengths of the absorption and fluorescence spectra
are measured and the spectral parameters are calculated and are shown in
Table 4.3.
4.3.2 Fluorescence Lifetime Measurements
The fluorescence decay profile and the residuals of the dyes oil red
o and metanil- yellow in solvent acetonitrile are shown in Figure 4.6 a-b. The
residuals shown in the Figure are well within the error limits. The average life
time values of these dyes are shown in Table 4.3. Oil red o exhibits bi-
exponential decay while metanil yellow exhibits a tri – exponential decay.
103
0
0.2
0.4
0.6
0.8
300 400 500 600 700 800 900Wavelength (nm)
Abs
orba
nce
Oil Red O
0
0.4
0.8
1.2
300 400 500 600 700Wavelength (nm)
Ab
sorb
an
ce
Metanil Yellow
Figure 4.4 Absorption spectra of azo dyes in acetonitrile
0
20
40
60
80
540 560 580 600 620 640Wavelength (nm)
Fluo
rese
cnce
Inte
nsit
y
Oil Red O
0
20
40
60
80
550 575 600 625 650 675
Wavelength (nm)
Flu
ores
cenc
e In
tens
ity
Metanil Yellow
Figure 4.5 Fluorescence spectra of azo dyes in acetonitrile
104
1
10
100
1000
10000
100000
0 10 20 30 40 50
Prompt
Decay
Fit
METANILYELLOW
Log
(Cou
nts)
Time (ns)
-6
-4
-2
0
2
4
6
3 8 13 18 23
Residuals
Time (ns)Sta
ndar
d D
evia
tio
n
-4
-2
0
2
4
3 13 23 33
Residuals
Time (ns)
Sta
ndar
d D
evia
tion
Figure 4.6.a Fluorescence decay profile and residuals of oil red o dye
Figure 4.6.b Fluorescence decay profile and residuals of metanil yellow
dye
Table 4.3 Spectral parameters of azo dyes in acetonitrile
Dye
Absorption spectra Fluorescence spectra Average
life time (ns)
Peak wave-length
nm
(λ) 104 L mol-1 cm-1
()1/2 cm-1
Oscillator Strength f x 10-24 L mol-1
cm-2
Peak wave-length
nm
FWHM nm
Stoke’s shift cm-1
Oil red O 362, 523 6.40 3900 1.10 556 35 1120 1.00 Metanil- yellow
352, 521 11.05 4900 2.34 589 58 2136 3.00
1
10
100
1000
10000
100000
0 10 20 30 40 50 60
Prompt
Decay
Fit
OIL RED O
Log (C
ounts
)
Time (ns)
105
4.3.3 Results and Discussion
The absorption spectra of azo dyes are shown in Figure 4.4. The
absorption maxima in nm are shown in Table 4.3. The absorption wavelength
maxima decrease in the order of oil red O > metanil yellow. In each spectrum
there are two absorption maxima, the longer wavelength absorption maxima
is due to the excitation of electron in azo group (π -π* electronic transition).
The shorter wavelength, absorption maxima are due to electronic excitation in
the aromatic ring. As the π -π* is an allowed transition, the (λ) values for
both the dyes are very high.
The fluorescence spectra of azo dyes are shown in Figure 4.5. The
fluorescence maxima of oil red O, occurs at shorter wavelength than metanil
yellow dye. In metanil yellow the SO3Na grouping is more solvated than
methyl substituents in oil red O. Hence, there will be much enhanced
vibrational cascade in metanil yellow than in oil red O. As a result the
fluorescence maxima of metanil yellow lies at longer wavelength than in oil
red O. It is also reflected in the fluorescence lifetime values of the dyes.
Metanil yellow slowly relaxes and emits fluorescence, whereas, oil red O
relaxes very slowly than metanil yellow and hence, rapidly fluoresce. So the
lifetime of the oil red O is less than metanil yellow.
4.4 SPECTRAL CHARACTERISTICS OF THIAZIN DYES
The thiazin dyes (Azure I, Azure II and Toludine Blue O) are
obtained from CDH , India. The molecular structures of these dyes were
given in Figure 2.1. (c) - (i-iii).Thin layer chromatography (TLC) test
confirms the absence of any impurities. These dyes were used as received.
Acetonitrile was chosen as solvent, since these dyes were readily soluble in
it.
106
4.4.1 Absorption and Fluorescence Spectra of Thiazin Dyes
The absorption spectra of thiazin dyes in acetonitrile for 0.01mM
concentration are recorded and shown in Figure 4.7.The emission spectra of
the thiazin dyes in acetonitrile for 0.01mM concentration are recorded and
shown in Figure 4.8.The peak wavelengths of the absorption and fluorescence
spectra are measured and the spectral parameters are calculated and shown in
Table 4.4.
4.4.2 Fluorescence Lifetime Measurements
The fluorescence decay profile and the residuals of the dyes azure I,
azure II and toludine blue O in solvent acetonitrile are shown in Figure 4.9
a-c. The residuals shown in the Figure are well within the error limits. The
average life time values of these dyes are shown in Table 4.4. All the dyes
exhibit a bi-exponential decay.
Table 4.4 Spectral parameters of thiazin dyes in acetonitrile
Dye
Absorption spectra Fluorescence spectra Average
life time (ns)
Peak wave- length
nm
(λ) 104 L
mol-1 cm-1
()1/2 cm-1
Oscillator Strength f x 10-24 L mol-1
cm-2
Peak wave- length
nm
FWHM nm
Stoke’s shift cm-1
Azure I 638 1.25 1999 0.11 685 62 1074 3.28
Azure II 654 1.06 1074 0.05 706 62 1125 1.70
Toluidine Blue O
624 0.81 1852 0.07 680 40 1294 1.20
107
-0.02
0
0.02
0.04
0.06
0.08
0.1
0.12
0.14
400 500 600 700 800
Azure I
Azure II
Abs
orba
nce
Wavelength (nm)
0
0.2
0.4
0.6
0.8
350 450 550 650 750
Toludine Blue O
Abs
orba
nce
Wavelength (nm)
Figure 4.7 Absorption spectra of thiazin dyes in acetonitrile
0
200000
400000
600000
800000
1000000
640 690 740 790
Azure I
Azure II
Wavelength (nm)
Fluo
resc
enc
e In
tens
ity
0
20
40
60
80
600 650 700 750 800 850 900
Wavelength (nm)
Flu
ore
sce
nc
e In
ten
sity
Toludine Blue O
Figure 4.8 Fluorescence spectra of thiazin dyes in acetonitrile
108
-3
-2
-1
0
1
2
3
3 8 13 18
Residuals
Time (ns)Sta
nda
rd D
evia
tion
-3
-2
-1
0
1
2
3
3 8 13 18 23 28
Residuals
Time (ns)
Standar
d Dev
iation
Time (ns)
-4
-3
-2
-1
0
1
2
3
4
4 8 12 16 20
Residuals
Time (ns)
Sta
nda
rd D
evi
ati
on
1
10
100
1000
10000
0 10 20 30 40 50
Prompt
Decay
Fit
AZURE I
Time (ns)
Log
(Coun
ts)
Figure 4.9. a Fluorescence decay profile and residuals of azure I dye
1
10
100
1000
10000
0 10 20 30 40 50
Prompt
Decay
Fit
AZURE II
Time (ns)
Log
(Cou
nts)
Figure 4.9.b Fluorescence decay profile and residuals of azure II dye
1
10
100
1000
10000
0 10 20 30 40 50
Prompt
Decay
Fit
Touldine Blue O
Time (ns)
Log
(Cou
nts)
Figure 4.9.c Fluorescence decay profile and residuals of toludine blue O
dye
109
4.4.3 Results and Discussion
The absorption spectra of thiazin dyes are shown in Figure 4.7. The
absorption maxima in nm are shown in Table 4.4. The absorption wavelength
maxima decrease in the order of azure II > azure I > Toludine blue O. The
absorption maximum of azure I is less than azure II. The difference between
azure I and azure II is the difference in the charge balancing ions. In azure I,
BF4- and in azure II Cl- are the charge balancing ion. As BF4
- is bulkier than
Cl- the BF4- cannot stabilize the quinonoid structure and hence, the absorption
maxima of azure I is less than azure II. The absorption maxima of toluidine
blue O is less than azure II. As the difference between azure II and toluidine
blue O is the replacement of one methyl group in azure II by hydrogen in
toluidine blue O. But the ring carries the methyl substituent.The fluorescence
lifetime of thiazin dyes illustrates a decrease from azure I > azure II >
toluidine blue O. The difference in the lifetime of thiazin dyes is attributed to
the difference in the charge balancing ions.
4.5 SPECTRAL CHARACTERISTICS OF
SULPHONAPHTHALEIN DYES
The sulphonaphthalein dyes (thymol blue and xylenol orange) are
obtained from loba chemie and fischer chemic, India. The molecular
structures of these dyes were given in Figure 2.1. (d) - ( i-ii).Thin layer
chromatography (TLC) test confirms the absence of any impurities. These
dyes were used as received. Ethanol was chosen as solvent since the dyes
were readily soluble in it.
110
4.5.1 Absorption and Fluorescence Spectra of Sulphonaphthalein
Dyes
The absorption spectra of sulphonaphthalein dyes in ethanol for
0.01mM concentration are recorded and shown in Figure 4.10.The emission
spectra of the sulphonaphthalein dyes in ethanol for 0.01mM concentration
are recorded and shown in Figure 4.11.The peak wavelengths of the
absorption and fluorescence spectra are measured and the spectral parameters
are calculated and shown in Table 4.5.
4.5.2 Fluorescence lifetime measurements
The fluorescence decay profile and the residuals of the dyes thymol
blue and xylenol orange in solvent ethanol are shown in Figure 4.12 a-b.
The residuals shown in the Figure are well within the error limits. The
average life time values of these dyes are shown in Table 4.5. Both the
dyes exhibit bi-exponential decay.
Table 4.5 Spectral parameters of sulphonaphthalein dyes in ethanol
Dye
Absorption spectra Fluorescence spectra Average
life time (ns)
Peak wave-length
nm
(λ) 104 L mol-1 cm-1
()1/2 cm-1
Oscillator Strength f x 10-24 L mol-1
cm-2
Peak wave- length
nm
FWHM nm
Stoke’s shift cm-1
Thymol Blue 505 6.13 4014 1.06 573.5 39 3035 4.00
Xylenol Orange 521 5.89 1641 0.42 605 84 2661 1.75
111
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
350 450 550 650 750
Thymol Blue
Xylenol Orange
Wavelength (nm)
Ab
sorb
an
ce
Figure 4.10 Absorption spectra of sulphonaphthalein dyes in ethanol
0
40000
80000
120000
160000
200000
550 600 650 700 750 800Wavelength(nm)
Fluo
resc
ence
Int
ensi
ty
Xylenol Orange
0
1
2
3
4
5
6
7
540 560 580 600 620 640
Wavelength (nm)
Fluo
resc
enc
e In
ten
sity
Thymol Blue
Figure 4.11 Fluorescence spectra of sulphonaphthalein dye in ethanol
112
-4
-2
0
2
4
3 13 23 33
Residuals
Time (ns)
Stan
dard
Devi
atio
n
-4
-2
0
2
4
3 8 13 18 23
Residuals
Time (ns)
Sta
nda
rd D
evi
ation
1
10
100
1000
10000
0 10 20 30 40 50
Prompt
Decay
Fit
Time (ns)
Log
(Cou
nts)
THYMOL BLUE
Figure 4.12.a Fluorescence decay profile and residuals of thymol blue
dye
1
10
100
1000
10000
0 10 20 30 40 50
Prompt
Decay
Fit
XYLENOL ORANGE
Log
(Cou
nts)
Time (ns)
Figure 4.12.b Fluorescence decay profile and residuals of xylenol
orange dye
4.5.3 Results and Discussion
The absorption spectra of sulphonaphthalein dyes are shown in
Figure 4.10. The absorption maxima in nm are shown in Table 4.5. The
absorption wavelength maxima decrease in the order of xylenol orange >
thymol blue. The absorption spectra of thymol blue and xylenol orange
indicates the quinonoid structure, which is due to phenolic hydroxyl group. In
the quinonoid structure the SO3 will be free and cannot be bounded to
tetrahedral carbon. The average lifetime of thymol blue is higher than that of
xylenol orange. It is attributed to direct binding of bulky iso propyl grouping
113
to the aromatic ring in thymol blue. In xylenol orange the iso propyl group is
replaced by bulky substituents, via CH2 grouping. The bulky substituents
cannot much influence on the vibrational cascade of the base nucleus as it is
separated by CH2 grouping.
4.6 SPECTRAL CHARACTERISTICS OF NATURAL DYE
The natural dye (carmine) was obtained from CDH, India. The
molecular structure of the dye was given in Figure 2.1. (e). - ( i). Thin layer
chromatography (TLC) test confirms the absence of any impurities. The dye
was used as received. Methanol was chosen as solvent since the dye was
readily soluble in it .
4.6.1 Absorption and Fluorescence Spectra of Natural Dye
The absorption spectra of carmine dye in methanol for 0.01mM
concentration was recorded and shown in Figure 4.13. The emission spectra
of the carmine dye in methanol for 0.01mM concentration was recorded and
shown in Figure 4.14.The peak wavelength of the absorption and fluorescence
spectra are measured and the spectral parameters are calculated and shown in
Table 4.6.
4.6.2 Fluorescence Lifetime Measurement
The fluorescence decay profile and the residuals of the dye carmine
in solvent methanol are shown in Figure 4.15.The residuals shown in the
Figure are well within the error limits. The average life time value for the
dye is given in Table 4.6. Carmine dye exhibit a bi-exponential decay.
114
-5
-3
-1
1
3
5
3 13 23
Residuals
Time (ns)
Sta
nd
ard
Dev
iati
on
0
0.2
0.4
0.6
0.8
1
1.2
300 400 500 600 700 800 900
Wavelength (nm)
Abs
orba
nce
Figure 4.13 Absorption spectra of carmine dye in methanol
0
20000
40000
60000
80000
100000
120000
140000
500 550 600 650 700 750 800
Wavelength (nm)
Fluo
resc
ence
Inte
nsit
y
Figure 4.14 Fluorescence spectra of carmine dye in methanol
1
10
100
1000
10000
0 10 20 30 40 50
PromptDecayFit
CARMINE
Log
(Cou
nts)
Figure 4.15 Fluorescence decay profile and residuals of carmine dye
115
Table 4.6 Spectral parameters of carmine dye in methanol
Dye
Absorption spectra Fluorescence spectra Average life
time (ns)
Peak wave- length
nm
(λ) 104 L
mol-1
cm-1
()1/2 cm-1
Oscillator Strength f x 10-24 L mol-1
cm-2
Peak wave- length
nm
FWHM nm
Stoke’s shift cm-1
Carmine 515 10.48 6000 2.72 582 89 2234 1.13
4.6.3 Results and Discussion
The absorption spectrum of the carmine dye is shown in Figure
4.13 The absorption maxima in nm are shown in Table 4.5. The carmine dye
has a absorption maximum at 515 nm, with the lifetime of 1.13 ns. As the
glucose is directly bound to anthraquinone nucleus, its lifetime is short.
4.7 SPECTRAL CHARACTERISTICS OF FLUORONE DYE
The fluorine dye (mercurochrome) was obtained from S.D fine
chem, India. The molecular structure of the dye was given in Figure 2.1. (f)
- ( i).Thin layer chromatography (TLC) test confirms the absence of any
impurities. The dye was used as received. Methanol was chosen as solvent
since the dye was readily soluble in it.
4.7.1 Absorption And Fluorescence Spectra of Fluorone Dye
The absorption spectra of mercurochrome dye in methanol for
0.01mM concentration are recorded and shown in Figure 4.16.The emission
spectra of the mercurochrome dye in methanol for 0.01mM concentration
was recorded and shown in Figure 4.17.The peak wavelength of the
absorption and fluorescence spectra are measured and the spectral parameters
are calculated and shown in Table 4.7.
116
-5
-3
-1
1
3
5
2 12 22 32
Residuals
Time (ns)
Stan
dard
Dev
iation
Time (ns)
-0.1
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
300 400 500 600 700 800Wavelength (nm)
Abs
orb
anc
e
Figure 4.16 Absorption spectra of mercurochrome dye in methanol
0
100000
200000
300000
400000
500000
600000
550 600 650 700 750 800Wavelength (nm)
Flu
ore
sce
nt
Inte
nit
y
Figure 4.17 Fluorescence spectra of mercurochrome dye in methanol
1
10
100
1000
10000
0 10 20 30 40 50
Prompt
Decay
Fit
MERCUROCHROME
Time (ns)
Log
(Cou
nts)
Figure 4.18 Fluorescence decay profile and residuals of mercurochrome
dye
117
4.7.2 Fluorescence Lifetime Measurement
The fluorescence decay profile and the residuals of the dye
mercurochrome in solvent methanol are shown in Figure 4.18. The
residuals shown in the Figure are well within the error limits. The average
life time value of the dye is shown in Table 4.7. The dye exhibits a bi-
exponential decay.
Table 4.7 Spectral parameters of mercurochrome dye in methanol
Dye
Absorption spectra Fluorescence spectra Average
life time (ns)
Peak wave-length
nm
(λ) 104 L mol-1 cm-1
()1/2 cm-1
Oscillator Strength f x 10-24 L mol-1
cm-2
Peak wave- length
nm
FWHM nm
Stoke’s shift cm-1
Mercuro-chrome
516 5.94 2320 0.60 773 20 6433 1.40
4.7.3 Results and Discussion
The absorption spectrum of mercurochrome dye is shown in Figure
4.16. The mercurochrome dye gives a wavelength maximum at 516 nm, with
the lifetime of 1.40 ns. It also has a less lifetime similar to carmine because of
direct attachment of benzoic acid grouping attached to the main nucleus.
4.8 CONCLUSION
The spectral characteristics of the dyes were studied and the
spectral parameters were calculated. It is observed that all the dyes studied
except azo dyes, have a single band absorption and that their fluorescence
intensity profiles though sharp are weak in intensity. The azo dye molecules
studied displayed two absorption bands in the range 300nm to 650nm.
118
The fluorescence profile for anthraquinone dyes and thiazin dyes
shows a red shift which can be attributed to intermolecular charge transfer
(ICT) which occurs when these molecules are excited. Internal conversion is a
mechanism that is mostly responsible for the low fluorescence efficiency in
organic dyes. It is a direct nonradiative decay of the lower excited state singlet
state S1, to the ground state S0. As this mechanism results in heat transfer ,
which may facilitate thermal nonlinearity.
From the above studies it is to be concluded that electron releasing
substituents, bounded to the main skeleton of the dye, shift the wavelength
maxima to a longer value. But the electron withdrawing substituents, shift the
wavelength maxima to a shorter value. The bonding of the bulky substituents
and readily solvatable substituents to the main nucleus increases the lifetime.