ChemStation E-Seminars
Page 1ChemStation E-Seminar
Dec, 2008
![Page 2: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/2.jpg)
ChemStation E-Seminar Agenda Schedule Short description AgilentSchedule Short description Agilent
Speaker
15. Oct. 2008 ChemStation Data Storage Concept Steven Brown
Ortrud Emde
12. Nov. 2008 ChemStation Navigation TableHow to improve Review and Reprocess
Steven Brown
Ortrud EmdeOrtrud Emde10. Dec. 2008 Integrating My Results in ChemStation Steven Brown
Ortrud Emde
14. Jan. 2009 How to set up a calibration table in ChemStation Steven Brown
Bob Guiffre
04. Feb. 2009 Various Reporting in Chemstation Steven Brown
Ortrud Emde
18. Feb. 2009 Updated Compliance 21 CFR Part 11 solution with Steven Brown
Page 2ChemStation E-Seminar
Dec, 2008
18. Feb. 2009 Updated Compliance 21 CFR Part 11 solution with Agilent ChemStation & OpenLAB ECM
Steven Brown
Ortrud Emde
![Page 3: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/3.jpg)
Seminar 4: How to set up a calibration Table in ChemStationChemStation
Topics
OpenLAB Overview
Calibration and peak Identification Overview
Demo Simple Calibration table
C lib ti S ttiCalibration Setting
Multilevel calibration
Recalibration Sequence and Manual
Questions
Page 3ChemStation E-Seminar
Dec, 2008
![Page 4: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/4.jpg)
Q tifi ti C lib tiQuantification - Calibration
Page 4
![Page 5: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/5.jpg)
In this Section, We Will Discuss:
The steps necessary to build a calibration table.
How to select calibration settings.
Calibration table options.
Peak identification tools.
Page 55
![Page 6: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/6.jpg)
Building a Calibration Table
1. Inject standards with concentrations that bracket yourunknowns.
2. Set-up appropriate integration events for the low standard.Integrate to obtain the peak height or area. Save the integration events to the method.
3 Begin a new calibration table with level one the lowest3. Begin a new calibration table with level one, the lowestlevel. Fill in compound names and amounts.
4 I t t h dditi l t d d l l d dd t th4. Integrate each additional standard level and add to the calibration table.
Page 66
5. Examine the calibration curve and save it to the method.
![Page 7: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/7.jpg)
Example Calibration
mAU100
DAD1 A, Sig=254,4 Ref=550,100 (DEMO\005-0101.D)
0.7
47
021
40
60
80 1.0
.565
7
StandardISTD
min1 2 3 4 5 6
0
20
2
5.8
37
ng/μL ng/μL
3 Levels
1. Standard 30.0 ISTD 50.02. Standard 50.0 ISTD 50.03 Standard 75 0 ISTD 50 0
Page 77
3. Standard 75.0 ISTD 50.0
![Page 8: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/8.jpg)
Signal Details
Define signals to be evaluated during a method
Page 88
![Page 9: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/9.jpg)
Calibration Settings
Page 99
![Page 10: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/10.jpg)
Load the Signal(s) for the Low Level Standard and IntegrateIntegrate
Page 1010
![Page 11: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/11.jpg)
Create a Calibration Table: Level One
Page 1111
![Page 12: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/12.jpg)
Fill in:
Calibration Table Overview: Fill in the DialogFill in:Compound NameAmount
Check if:Reference PeakISTDIndicate ISTD #
Page 1212
![Page 13: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/13.jpg)
Demo CalibrationDemo Calibration
Page 13ChemStation E-Seminar
Dec, 2008
![Page 14: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/14.jpg)
Load the Signal - Add Level Two
Fill in Amount for Level 2
Page 1414
Fill in Amount for Level 2
![Page 15: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/15.jpg)
Add Additional Levels
Add Level 3Add Level 3 and fill in theamount
Page 1515
![Page 16: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/16.jpg)
Calibration Table Options
Compound, Amt, Area, Rsp Factor,Ref, ISTD#
Compound, Grp, Amt, Low LimitHigh Limit
Compound, Amt, Area, Def, CurveType, Origin, Weight
Compound, Area, Rsp%, Def, +-, Pk Usage
Page 1616
![Page 17: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/17.jpg)
Compound Details
Use Low and High Limits as qualifiers to confirm peakidentification.
Page 1717
![Page 18: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/18.jpg)
Peak Details
Check Def to apply thedefault curve type and origin from the Calibration Settingsdi l bdialog box.
Curve TypesPiece wise Exponent
Origin TreatmentsIgnore
Linear QuadraticLog CubicPower Average Rsp/Amt
IncludeForceConnect straight segment
Page 1818
Power Average Rsp/Amt Connect straight segment
![Page 19: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/19.jpg)
Identification Details
By default the largest peak of a compound is the Main peakBy default, the largest peak of a compound is the Main peak. Others are set to Ignore.
Page 1919
![Page 20: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/20.jpg)
Qualifiers
The ratio of 230 to 254 is 38% The ratio of 230 to 254 is 224%The ratio of 230 to 254 is 38%for Biphenyl
The ratio of 230 to 254 is 224%.The peak would be rejected
Page 2020
![Page 21: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/21.jpg)
Edit Calibration Table Options
Define the columns included in your Calibration Table
Page 2121
![Page 22: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/22.jpg)
Quantification Tools
C lib i T k
New Calibration Table
Add New Level Edi C lib iCalibration TaskTool
Add New Level
Signal DetailsAdd New Peaks
Edit CalibrationTable Options
Recalibrate withCurrent Chromatogram
Select compoundsfrom chromatogramg
Select Calibrated peak(s) fromChromatogram and delete from
lib ti t blSelect peak(s) from h t d dd
from chromatogram and recalibrate
Page 2222
calibration table. chromatogram and addto calibration table
![Page 23: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/23.jpg)
To Save the Calibration Table as Part of a Method...
Save Table to Method
Page 2323
![Page 24: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/24.jpg)
Save Method
• Be sure to know which method you are ki ith d h t th tworking with and save changes to the correct
method.
M th d D l t h (i lib ti• Method Development changes (i.e. calibration table creation / updates) should be saved to the Current Method (Master Method).
• Integration events and other parameters specific to the individual data file should be
d t th I di id l M th d f D t Filsaved to the Individual Method from Data File (DA.M)
Page 2424
![Page 25: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/25.jpg)
Information on ChemStationDocumentation can be found on the Agilent websiteghttp://www.chem.agilent.com/Scripts/PDS.asp?lPage=61233
• Product Datasheet• Specification• Application Notes• Manuals
Manuals can be found your ChemStation Installation DVD
• Getting Started with New ChemStation WorkflowPartNo.G2170-90042
U d t di Ch St ti• Understanding your ChemStationPartNo. G2070-91125
• OpenLAB OptionPartNo.G2170-90233
Software Status Bulletinhttp://www.chem.agilent.com/en-US/Support/Downloads/Patches/Pages/default.aspx
Customer Trainings (NorthAmerica)g ( )
http://www.chem.agilent.com/cag/training/northam/index.asp
Page 25ChemStation E-Seminar
Dec, 2008March, 2008
![Page 26: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/26.jpg)
Learning Products – North America Course CatalogCatalog
Users needing to increase productivity in the lab by utilizing both standard and advanced features available in the Agilent GC/LC ChemStation may want to attend one of the following courses:
H2606A - ChemStation for GC Data Analysis and Reporting (2 days)
H5928A - Agilent HPLC (2D) Data Analysis and Reporting (2 days)
H4039A - Agilent HPLC (3D) Data Analysis and Reporting (3 days)
Course Features
Data acquisition and method creationData analysis including integration and calibrationSequencingReportingReporting
For more information concerning course content, dates and locations, please visit:
http://www chem agilent com/en-US/education/en-us/classroomtraining/Pages/Courselisting aspx
Page 26ChemStation E-Seminar
Dec, 2008
http://www.chem.agilent.com/en US/education/en us/classroomtraining/Pages/Courselisting.aspx
![Page 27: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/27.jpg)
Learning Products – European Course Catalog
Users needing to increase productivity in the lab by utilizing both standard and advanced features available in the Agilent GC/LC ChemStation may want to attend one of the following courses:
H4033A – Agilent HPLC (3D) Method&Run Control, Data Analysis and Reporting (4 days)
H8718A - Agilent HPLC (3D) Data Analysis and Reporting (2 days)
H5928A – Advanced User Training, Quantification and Result Reporting (2 days)
Course Features
Data acquisition and method creationData analysis including integration and calibrationSequencingReportingReporting
For more information concerning course content, dates and locations, please visit:
htt // h il t / US/ d ti / /h
Page 27ChemStation E-Seminar
Dec, 2008
http://www.chem.agilent.com/en-US/education/pages/homepage.aspx
![Page 28: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/28.jpg)
Who to contact for more information about OpenLAB and Agilent software productsOpenLAB and Agilent software products1. Agilent Account Managers (sales)
2 Agilent Informatics Product Specialists2. Agilent Informatics Product Specialists
Eastern US - Marc Mandelbaum – 302-683-7149
Pacific Coast (CA, WA, OR) - Spencer Tse 408-553-3655
Midwest and South - Johnathon McSayles 847-944-6019
Midwest /West/Latin America – Stephen Brown 847-944-6461
Page 28
![Page 29: Como configurar una Calibración ChemStation](https://reader037.vdocument.in/reader037/viewer/2022100218/568c4b901a28ab49169cae1b/html5/thumbnails/29.jpg)
QUESTIONS?
Page 29ChemStation E-Seminar
Dec, 2008