comprehensive two dimensional

Upload: missbliss84

Post on 05-Apr-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/2/2019 Comprehensive Two Dimensional

    1/24

    Comprehensive two-dimensional

    gas chromatography for theanalysis of organohalogenatedmicro-contaminantsPeter Korytar, Peter Haglund, Jacob de Boer, Udo A.Th. Brinkman

    We explain the principles of comprehensive two-dimensional gas chroma-

    tography (GC GC), and discuss key instrumental aspects with emphasison column combinations and mass spectrometric detection. As the main item

    of interest, we review the potential of GC GC for the analysis of organo-

    halogenated micro-contaminants, and highlight its superiority over conven-

    tional 1D-GC. We present results for 12 compound classes, including

    polychlorinated biphenyls, dibenzo-p-dioxins and furans, and n-alkanes,

    toxaphene and polybrominated diphenyl ethers. We draw attention to target

    analysis as well as within-class and between-class separations.

    2006 Elsevier Ltd. All rights reserved.

    Keywords: Comprehensive two-dimensional gas chromatography; GC GC; Mass

    spectrometry; Organohalogenated compound; Separation

    1. Introduction

    In 1966, the Swedish chemist Jensen [1]

    announced that he had identified a group

    of organochlorine compounds that was

    accumulated to toxic concentrations in

    nature. Unlike the closely related chlori-

    nated pesticides, such as DDT and the

    various drins which had, so far,

    attracted most attention of environmental

    chemists the newly detected compounds,

    the polychlorinated biphenyls (PCBs), had

    entered the environment essentially

    unintentionally. Since that time, these twoclasses of persistent organic pollutants,

    which shared high annual production,

    widespread usage, long persistence and

    serious toxic effects, have been the subject

    of a rapidly increasing number of funda-

    mental as well as applied studies. Over the

    years, many related classes of compounds

    have been added to the list; Table 1 gives

    an overview, and also lists the number of

    theoretically possible congeners, which

    gives an impression of the degree of com-

    plexity that can be expected in technical

    mixtures and, thus, in environmental and

    food samples. For the rest, we assume that

    the general reader will be familiar with the

    main characteristics and usage of the

    various classes of organohalogens and, at

    least in some cases, with the catastrophes

    or otherwise, which led to their being

    considered priority pollutants by many

    governments and international (UN, EU)

    bodies.

    During the entire past half-century, gas

    chromatography with electron-capture

    Peter Korytar*

    Netherlands Institute for Fisheries Research,

    P.O. Box 68, NL-1970 AB IJmuiden,

    The Netherlands

    Free University, Department of Analytical Chemistry and Applied Spectroscopy,

    de Boelelaan 1083, NL-1081 HV Amsterdam,

    The Netherlands

    Peter Haglund

    Umea University, Department of Chemistry, Environmental Chemistry,

    SE-901 87 Umea, Sweden

    Jacob de BoerNetherlands Institute for Fisheries Research,

    P.O. Box 68, NL-1970 AB IJmuiden,

    The Netherlands

    Udo A.Th. Brinkman

    Free University, Department of Analytical Chemistry and Applied Spectroscopy,

    de Boelelaan 1083, NL-1081 HV Amsterdam,

    The Netherlands

    *Corresponding author.

    E-mail: [email protected]

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    0165-9936/$ - see front matter 2006 Elsevier Ltd. All rights reserved. doi:10.1016/j.trac.2005.12.003 3730165-9936/$ - see front matter 2006 Elsevier Ltd. All rights reserved. doi:10.1016/j.trac.2005.12.003 373

    mailto:[email protected]:[email protected]
  • 8/2/2019 Comprehensive Two Dimensional

    2/24

    detection (GCECD) was the predominant analytical

    technique for selective and sensitive determination of

    organohalogens. In the early part of that period, when

    packed-column GC was the only means of separation

    available, total-PCB determination (based on comparing

    an environmental sample with a technical PCB mixture)

    was all that could be achieved. The introduction of

    (fused-silica) capillary columns was a real breakthrough,

    which, all of a sudden, enabled congener-specific deter-

    mination of the PCBs and, of course, also of the other

    classes of organohalogens. However, it also became clear,

    after this major step forward, that capillary GC was notthe solution for all, or even most, problems. Admittedly,

    single-column (or one-dimensional, 1D) GC can provide

    the required resolution when the number of target ana-

    lytes is restricted (e.g., with the organochlorine pesticides

    (OCPs) or polychlorinated naphthalenes (PCNs)).

    However, even for the PCBs with their still moderate

    number of 209 congeners, or for the about 140 CB

    congeners present in a technical mixture, no satisfactory

    1D-GC solution could be found. And, not surprisingly,

    correspondingly more serious problems were encoun-

    tered with complicated mixtures, such as toxaphene,

    polychlorinated terphenyls (PCTs) and, specifically,

    polychlorinated alkanes (PCAs); the PCAs typically show

    up in 1D-GC as an essentially unresolved broad band

    covering a major part of the baseline. The magnitude of

    the problem becomes even clearer when one considers

    that the present discussion is about within-class separa-

    tions only: interferences caused by other organohalogens

    and/or matrix constituents or problems due to widely

    different concentrations of (partly) co-eluting congeners

    have not been taken into account.

    Over the years, several approaches have been used to

    alleviate the problems. One of these was sample frac-

    tionation (e.g., by size-exclusion or adsorption liquid

    chromatography). Another was to filter out interferences

    by selective mass spectrometric (MS) detection, which is

    powerful but fails to resolve co-elutants with closely

    similar mass spectra, as is often the case for congeners of

    the same compound class, but also of different classes.

    On the GC side, one way to go was the parallel use of

    several stationary phases and the combination of infor-mation from the GC runs on different columns to solve

    specific separation problems. It was a step further to

    combine two different columns in a single set-up: in so-

    called multi-dimensional GC (MDGC), selected fractions

    from the first column are subjected to a second GC run

    that uses a completely different separation mechanism.

    Many successful applications have been reported, but one

    should keep in mind that, in essentially all cases, they are

    of a heart-cutting nature: only a single, or at best a few,

    small fractions are transferred to the second column for

    further separation. That is, it is an excellent solution

    when information is required about a few target analytes,

    but not about the entire sample. In the latter instance (i.e.with complex samples and/or when unknowns have to be

    traced), MDGC becomes much too complicated and time-

    consuming. It is in these situations and their number

    can confidently be said to be increasing daily that a so-

    called comprehensive approach is needed: MDGC, or GC

    GC, is now replaced by GC GC, with which, instead of a

    few selected fractions, the entire sample is subjected to

    separation on two different columns. One immediate

    advantage is that the information content is much

    greater, and another is that the GC GC run is ready once

    the first-dimension run is finished; that is, GC GC is

    much more efficient than MDGC.It is the goal of this review briefly to explain the

    principle of GC GC and to discuss key aspects, such as

    column-to-column interfacing or modulation, detection

    and detector requirements (including a comparison of

    time-of-flight MS (TOF-MS) and fast-scanning quadru-

    pole MS (qMS) instruments) and the selection of properly

    matched column combinations.

    We will devote most attention to applications in the field

    of organohalogenated micro-contaminants, which will

    also be used to highlight the added value of GC GC

    compared with 1D-GC. Readers who are interested in a

    more detailed review of the various aspects of GC GC and

    in applications to compound classes other than organo-

    halogens, should consult two extensive reviews [2,3].

    2. GC GC: general principle

    In GC GC, the entire sample is subjected to two GC

    separations that are based on different separation

    mechanisms. Fig. 1 shows a schematic of a GC GC

    system. In most instances, the sample is first separated

    on a high-resolution capillary GC column typically a

    1530 0.250.32 mm ID, 0.11 lm df column

    Table 1. Main classes of polyhalogenated micro-contaminants

    Name Acronym Maximumnumber ofcongenersa

    Polychlorinated biphenyls PCBs 209Polychlorinated dibenzo-p-dioxins PCDDs 75

    Polychlorinated dibenzofurans PCDFs 135Toxaphene components Toxaphene 61 696

    bornane congeners 16 640camphene congeners 12 288dihydrocamphene congeners 32 768

    Organohalogenated pesticides OCPs ca. 300Polychlorinated terphenyls PCTs 8149Polychlorinated diphenylethers PCDEs 209Polychlorinated naphthalenes PCNs 75Polychlorinated alkanes PCAs Very highPolybrominated biphenyls PBBs 209Polybrominated diphenylethers PBDEs 209

    aEnantiomers not included.

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    374 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    3/24

    containing a non-polar stationary phase. As will be

    explained in Section 3, an interface called a modulator is

    used to separate the first-column eluate into a very large

    number of adjacent small fractions. To maintain the

    first-column separation, these fractions should be no

    larger than one quarter of the peak width, or r, in that

    dimension. In order to meet this so-called modulation

    criterion, temperature programming in GC GC is

    slower than in 1D-GC, and typically occurs at a rate of

    13C/min. Each individual fraction is trapped, re-

    focused and, next, launched into the second GC column,

    which is much shorter and narrower than the first one

    typical dimensions are 12 m 0.1 mm ID 0.1 lm df.

    The second-column separation generally is of a polar

    or shape-selective nature. That is, the separation

    mechanisms are indeed different or, in other words,

    orthogonal separation conditions have been created. The

    separation in the second column is extremely fast and

    usually takes only 28 s as against 45120 min for the

    first-dimension separation. Consequently, it is performed

    under essentially isothermal conditions. The fast sepa-

    ration in the second dimension causes the analyte peaks

    to be very narrow with widths of, typically, 100600 ms

    at the baseline. These narrow peaks require fast detectors

    with a small internal volume and a short rise time in

    order to achieve a proper reconstruction of the (second-

    dimension) chromatograms. In some systems, the second

    column is housed in a separate oven to allow more

    flexible, independent temperature programming.

    The outcome of a GC GC run is a large series of high-

    speed, second-dimension chromatograms, which are

    usually stacked side by side to form a two-dimensional

    Figure 1. Schematic of a GC GC system with different modulator types.

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 375

  • 8/2/2019 Comprehensive Two Dimensional

    4/24

    (2D) chromatogram with one dimension representing the

    retention time on the first column and the other the

    retention time on the second column. The most conve-

    nient way to visualize these chromatograms is as contour

    plots, where peaks are displayed as spots in a 2D plane

    using colors and/or shading to indicate signal intensities.

    Apex plots, which use specific symbols to indicate theposition of the peak apexes, are another frequently used

    means to display GC GC chromatograms; the overall

    presentation becomes much simpler, and that is espe-

    cially advantageous when ordered structures are studied

    (in this review, apex plots are used (e.g., Figs. 4 and 8)

    and are combined with contour plots (e.g., Figs. 9 and

    10); contour plots are also displayed (e.g., Figs. 2 and 5).

    There is general agreement regarding the main

    advantages of GC GC over 1D-GC. Most strikingly, the

    peak capacity is much higher, and that yields a dramat-

    ically improved separation of the analytes of interest from

    each other but also and this is often even more impor-

    tant from interfering matrix constituents. In addition, amain benefit of the trapping-plus-refocusing occurring

    during modulation is, typically, a 310-fold improved

    signal-to-noise ratio compared with 1D-GC. Finally,

    compound identification is more reliable in GC GC be-

    cause each substance now has two identifying retention

    values rather than one. Specifically, when orthogonal

    conditions are used, chemically related compounds show

    up as so-called ordered structures (i.e. as clusters or

    bands). This phenomenon greatly facilitates group-typeanalysis, fingerprinting studies and the provisional clas-

    sification of unknowns. It is particularly important in the

    study of classes of organohalogens, as will become clear

    from most of the figures included in Sections 6 and 7.

    In the following sections, we will discuss in some detail

    three topics of general analytical interest: modulation;

    detection and analytical performance; and, column

    selection.

    3. Modulation

    The key component of a GC GC instrument is themodulator that joins the two columns. It serves three

    main goals:

    Figure 2. GC GClECD chromatograms of PCA-60 technical mixture on DB-1 in the first dimension and each of the six columns indicated inthe second dimension [24].

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    376 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    5/24

    (i) collecting and focusing of each of the fractions elut-

    ing from the first-dimension column;

    (ii) re-injecting/launching of each collected fraction

    into the second-dimension column; and,

    (iii) trapping of the next eluent fraction from the first

    column during the launch of the preceding

    fraction.Over the years, different ways of modulating the first-

    column effluent have been reported (Fig. 1). Initially,

    heating was preferred, with a rotating slotted heater, the

    sweeper, rapidly moving over a thick-film modulation

    capillary, heating it locally. Despite its frequent use in

    the early years of GC GC, because of the vulnerability

    of the set-up, the time-consuming optimization of the

    multi-parameter design and the restricted application

    range, the sweeper and related modulator types are

    obsolete today.

    The first major step forward was the introduction of

    the longitudinally modulating cryogenic system (LMCS).

    This modulator uses expanding CO2 (liquid) for trappingof the analytes at the top of the second-dimension col-

    umn. By subsequently moving the trap rapidly to an

    upstream position, the re-focused zone is exposed to the

    GC oven air and instantaneously volatilized and laun-

    ched. Today, jet-based modulators with no moving

    parts at all with either CO2 or liquid N2 for cooling

    are generally preferred. Single-, dual- and quad-jet

    modulators have been introduced, and several of these

    have been extensively compared in a recent study [4].

    The main conclusion was that all cryogenic modulators,

    if properly optimized, can satisfactorily be used for most

    applications, and certainly for those in the field oforganohalogen analysis. One caveat should be added. In

    order to ensure proper modulation of high-boiling com-

    pounds (i.e. highly substituted congeners), cooling

    should be just enough to trap the target compounds

    safely. Cooling that is too strong can cause remobiliza-

    tion to be inefficient and may lead to distorted and/or

    tailing peaks [5]. In addition, severe cooling will lead to

    more bleed from the first-dimension column being

    accumulated and that, in turn, will yield noisier second-

    dimension chromatograms. Some studies have indicated

    that modulation of high-boiling compounds can easily be

    achieved by air cooling [6]. If found to be true in further

    studies, this will make GC GC of organohalogens less

    expensive and simpler to operate.

    4. Detection

    As briefly mentioned in Section 2, detectors with a high

    data-acquisition rate and a negligible internal volume

    are required to describe properly the very narrow peaks

    that are the outcome of a GC GC run. Flame ionization

    detectors (FIDs) have data-acquisition rates up to 200 Hz

    and dead volumes that are effectively zero. It does not

    therefore come as a surprise that virtually all early

    GC GC studies were carried out with an FID, including

    those dealing with organohalogens [7]. However, as

    soon as real-life studies and analyte detectability became

    an issue of interest, it became clear that FID would have

    to be replaced by ECD to obtain sufficient selectivity and

    sensitivity. Conventional ECDs have data-acquisitionrates up to 50 Hz, but the main problem is their 1.5-ml

    cell volume, which causes severe peak broadening [8].

    Kristenson et al. [4] showed that, from amongst the

    miniaturized ECDs marketed in recent years, only the

    Agilent micro-ECD (lECD) with an internal volume of

    150 ll gives acceptable peak widths. The best results

    were obtained when working at the maximum flow of

    make-up gas (150 ml/min) and at temperatures above

    300C. However, as is to be expected, even under opti-

    mum conditions, the lECD delivers about 2-fold broader

    peaks than an FID [9]. All recent GC GClECD studies

    use an Agilent detector.

    The main problem of element-selective detection ingeneral and, therefore, also of ECD is that no struc-

    tural information is provided. In other words, MS is

    indispensable for identification/confirmation of the

    numerous separated compounds, whether target ana-

    lytes or unknowns. Today, the preferred choice is a TOF-

    MS that can, typically, acquire up to 50500 mass

    spectra per second (with unit mass resolution). The

    coupling of GC GC separation and TOF-MS detection

    presents no difficulties, and there is no additional peak

    broadening. This is true for both conventional electron

    impact (EI) ionization and the recently introduced elec-

    tron-capture negative ionization (ECNI) mode. An addi-tional advantage of a TOF-MS is that the sensitivity is

    higher than that of the full-scan mode of conventional

    scanning MS detectors, while the high acquisition rate

    prevents spectral skewing, and deconvolution conse-

    quently becomes a very powerful tool.

    Unfortunately, TOF-MS instruments are very expen-

    sive. From the start therefore, the potential and limita-

    tions of qMS instruments, which are available in

    essentially all GC laboratories, were investigated by

    several groups of workers [1013]. Briefly, the outcome

    of these studies was that, for restricted mass ranges of

    150200 Da, acquisition rates of 2030 Hz can be ob-

    tained, and that tentative identification is then indeed

    often possible, but that proper quantification cannot be

    achieved on the basis of the 35 data points registered

    across a peak. Recently, the situation has improved (i.e.

    when two rapid-scanning qMSs were marketed, the

    Shimadzu QP 2010 and the Perkin Elmer Clarus 500,

    which allow scanning up to 10,000 Da/s [14,15]).

    Table 2 provides an illustrative comparison of these two

    instruments and a state-of-the-art conventional qMS. A

    clear indication of the improved performance is that, if

    the lower limit of the peak width at the baseline is set at

    300 ms, the conventional qMS is restricted to a mass

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 377

  • 8/2/2019 Comprehensive Two Dimensional

    6/24

    range of about 100 Da, whilst 23-fold larger ranges can

    be handled by rapid scanning detectors. This is fullysatisfactory for many, and certainly for most, organo-

    halogen applications. In the selected ion monitoring

    (SIM) mode, acquisition rates of up to 90 Hz can be used

    when only one ion is monitored (i.e. seven data points

    can be collected even for peaks having a baseline width

    of only 80 ms). While the present conclusions are grat-

    ifying, they do not imply that rapid-scanning qMS

    instruments can replace TOF-MS in all instances, as,

    particularly when a wide mass range has to be covered

    (e.g., in searching for unknowns), TOF-MS has to be

    used.

    4.1. Analytical performance data

    Analyte detectability and linearity are influenced by

    quite a number of parameters such as GC conditions

    and set-up, and the number of modulations but the

    key factor is the detector used. Table 3 shows instru-mental limits of detection (iLODs) and dynamic ranges

    for various detectors and compound classes quantified

    so far by GC GC. The data can be called fully

    satisfactory, specifically the iLODs, which are 35-fold

    lower than in 1D-GC due to peak re-focusing in the

    modulator.

    5. Columns and column combinations

    To summarize, and simultaneously extend, what has so

    far been said about columns or, more appropriatelycolumn combinations to be used in GC GC, in

    essentially all early studies, a conventional 1530 m

    non-polar first-dimension column was combined with a

    Table 3. Analytical performance data for various detectors and classes of organohalogens

    Detector (mode) Class/compound iLOD [pg] Linearity* Reference

    Range [pg] R 2

    lECD PCBs 0.010.07 1400 >0.999 [9,19,20]PCDD/Fs 0.040.15 0.1200 >0.998 [9,19]2,3,7,8-TCDD 0.09 0.140 >0.998 [9,19]

    TOF-MS (EI) PCBs 0.110 0.51000 >0.993 [22,37]PCDD/Fs n.a. 0.2500 >0.996 [36,38]2,3,7,8-TCDD 0.20.5 0.5200 >0.996 [36,38,63]PBDEs 5 0.22000 >0.994 [37]OCPs 510 51000 >0.991 [37]

    qMS (EI, scan 50 Da) PCBs 12 101000 >0.997 [15]

    qMS (ECNI, SIM) PBDEs 1040 n.a. n.a. [14]PCDD/Fs 10710 n.a. n.a. [14]2,3,7,8-TCDD 710 n.a. n.a. [14]

    *n.a., not available.

    Table 2. Comparison of performance of two rapid-scanning and a conventional qMS in GC GC [11,1315,62]

    Mass range (Da) ornumber of ions monitored

    Perkin Elmer Clarus 500 Shimadzu QP 2010 Agilent MSD HP 5973

    maximumacquisition (Hz)

    minimumpeak widtha (ms)

    maximumacquisition (Hz)

    minimumpeak widtha(ms)

    maximumacquisition (Hz)

    minimumpeak widtha(ms)

    Full scan

    400 17 410 20 350 12 540300 23 300 25 280 15 470200 31 230 33 210 20 350100 63 110 50 140 n.a. n.a.

    SIM1 ion 91 80 n.a. 33 2102 ions 45 160 n.a. 18 3903 ions 30 230 n.a. 12 580

    n.a., not available.aCalculated for seven points per peak from: minimum peak width = 7/maximum acquisition.

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    378 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    7/24

    much shorter, narrower (medium-)polar or shape-

    selective second-dimension column. Using such a

    non-polar more polar column set has three distinct

    advantages:

    A wealth of information on the 1D-GC separation of

    the compound classes of interest is available in the

    literature and can be used for the first-dimensionseparation.

    The non-polar columnsalso have a high thermal stabil-

    ity and, consequently, cause little bleeding. As was

    shown in one study [19], use of some polar columns

    in the first dimension (e.g., an LC-50or a DB-Dioxin sta-

    tionary phase) can cause LODs to be 10 times higher.

    While in the first dimension there is a truly volatility-

    based boiling-point separation, for all other, more

    selective columns, separation will indeed be mainly

    governed by specific analyte/stationary phase interac-

    tions with, however, also a volatility-based contribution.

    Fortunately, the extremely rapid second-dimension

    separation will be essentially isothermal. This impliesthat, for sample constituents present in each individual

    first-dimension effluent fraction (which will have clo-

    sely similar boiling points), there will be no volatility

    contribution. In other words, the two separation

    mechanisms are indeed independent, and the condi-

    tions are orthogonal.

    In the early literature on GC GC, most of the above

    was already clearly understood and, possibly because of

    this and also because of a rather limited availability of

    short and narrow-bore second-dimension columns, not

    too much attention was devoted to optimizing the

    selection of column combinations. In recent years, this

    attitude has changed and, today, several interesting

    studies on, specifically, second-dimension column selec-

    tion are available notably in the context of organo-halogen analysis [1923]. We include illustrative

    examples in Sections 6 and 7 below. The present dis-

    cussion is therefore limited to two comments:

    (i) when aspects such as nature of substituents (Cl vs.

    Br), molecular shape (planar vs. non-planar) or

    parent-compound nature (aromatic vs. non-

    aromatic) play a role, even a brief study of

    second-column characteristics can be most reward-

    ing; and,

    (ii) the column combination providing the best struc-

    tural ordering (often considered the Number

    One criterion) does not always offer the best

    overall resolution of a class of compounds (highlyrelevant when unraveling the composition of rela-

    tively unknown mixtures such as toxaphene or

    PCAs).

    As an example, Fig. 2 shows what is found when a

    non-polar DB-1 first-dimension column is combined with

    six different second-dimension columns for the study of

    PCAs [24]. In this case, using 007-65HT as the second

    column yielded optimal structural ordering (less band

    Table 4. First- and second-dimension columns used for GC GC of organohalogens

    Code Phase References

    First-dimension columnsDB-1, HP-1, VF-1ms 100% methylpolysiloxane [19,20,2224,33,35,37,54]Rtx-Dioxin 2 Proprietary [36]Rtx-500 Proprietary (carborane) [38]HT-5 5% phenyl-methylpolysiloxane (carborane) [19]HT-8 8% phenyl-methylpolysiloxane (carborane) [22]DB-XLB Proprietary [9,19,21,22]DB-Dioxin Proprietary (44% methyl, 28% phenyl, 20% cyanopropyl,

    8% polyoxyethylene-polysiloxane)[19]

    LC-50 50% liquid crystalline-methylpolysiloxane [7,19]Chirasil-Dex 2,3,6-tri-O-methyl-b-cyclodextrin [2527]BGB-172 25% 2,3,6-tert.-butyldimethylsilyl-b-cyclodextrin [28]BGB-176SE 20% 2,3-di-O-methyl-6-O-tert.-butyldimethyl-b-cyclodextrin [28]

    Second-dimension columnsBPX-5 5% phenyl-methylsilphenylene [7]HT-8 8% phenyl-methylpolysiloxane (carborane) [19,20,2224,35,37,54]Rtx-500 Proprietary (carborane) [36]DB-17 50% phenyl-methylpolysiloxane [35]BPX-50 50% phenyl-methylpolysiloxane (silphenylene) [19,22,38]007-65HT 65% phenyl-methylpolysiloxane [19,23,24]OV 1701, DB-1701 14% cyanopropyl-phenyl-methylpolysiloxane [19,35]BPX-70 70% cyanopropyl polysilphenylene-siloxane [21]SP-2340 100% biscyanopropyl polysiloxane [21]VF-23ms Proprietary (7090% cyano-containing polymer) [19,23,24,26,27]007-210 50% trifluoropropyl-methylpolysiloxane [19,23,24]LC-50 50% liquid crystalline-methylpolysiloxane [9,19,21,2326]SupelcoWax-10, CP-WAX-52CB Polyethylene glycol [19,23,24,27,33,35]

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 379

  • 8/2/2019 Comprehensive Two Dimensional

    8/24

    overlap than with SupelcoWax-10 and LC-50), but, with

    VF-23ms, a much larger part of the 2D-separation space

    was used and overall resolution was, consequently,

    better.

    The only exception to what was said above is in the

    area of chiral separations. Since separations such as

    those of the pairs of atropisomeric PCBs require longcolumns and long run times for these applications,

    the more discriminating (i.e. in this case shape-selective)

    column is used in the first dimension [2528]. As can be

    seen in Table 4, substituted b-cyclodextrins were the

    preferred stationary phases in these studies. Relevant

    examples are included in Section 6 below.

    6. Within-class separations

    6.1. Polychlorinated biphenyls (PCBs)

    PCBs are products of anthropogenic activity and com-

    prise a class of chlorinated aromatic compounds with110 chlorine atoms attached to a biphenyl backbone.

    In total, there are 209 CB congeners; the presence of

    about 140 of these has been confirmed in technical

    formulations, such as Aroclors and Clophens, and also in

    environmental samples.

    For regulatory and monitoring purposes, seven CBs

    (CBs 28, 52, 101, 118, 138, 153 and 180) have been

    selected because of their abundance in humans and in

    foodstuff of animal origin, and they are often called the

    EU indicator CBs.

    In the past two decades, much attention has been paid

    to the toxicology of PCBs, particularly to the congenersthat show the same type of toxicity as polychlorinated

    dibenzo-p-dioxins and dibenzofurans (PCDD/Fs), the

    so-called dioxin-like or WHO CBs. They include four

    non-ortho (CBs 77, 81, 126 and 169) and eight mono-

    ortho (CBs 105, 114, 118, 123, 156, 157, 167 and 189)

    substituted CBs.

    Due to restricted rotation around the central CC bond

    of biphenyl, some CB congeners occur as enantiomers.

    Separation of such enantiomer pairs is interesting, since

    it enables the study of enantioselective bioaccumulation

    and biodegradation. There are 19 such pairs (CBs 45,

    84, 88, 91, 95, 131, 132, 135, 136, 139, 144, 149,

    171, 174, 175, 176, 183, 196 and 197), which exist as

    stable atropisomers at ambient or physiological temper-

    atures.

    The separation of any of the groups of CBs listed above

    and, specifically, of all CB congeners from each other and

    from the plethora of matrix constituents is a challenging

    task. Even though 1D-GC (with ECD or MS detection)

    can separate some 100150 CB congeners, there still is

    no unambiguous chromatographic separation of the 12

    WHO CBs, or even the seven EU indicator CBs, from the

    209 congeners or, even, the 140 present in technical

    formulations [29]. To achieve the intended goal, it is

    necessary to use sample fractionation, multiple injec-

    tions on different stationary phases, heart-cut multidi-

    mensional GC or columns coupled in series. It should,

    therefore, come as no surprise that, since the introduc-

    tion of GC GC, there have been many attempts to use

    this technique for improving separation of PCBs.

    There are over 20 papers [4,5,79,15,1923,2528,3038] that discuss PCB separation by GC GC.

    However, most of these use PCBs merely as test analytes

    to optimize and/or demonstrate modulator tuning,

    proper modulator or detector performance, and instru-

    ment set-up. These studies have clearly contributed to

    the development of robust GC GC procedures but

    provide little information on the separation itself. The

    present review therefore mainly discusses papers aimed

    at improving separation and/or detection of CB cong-

    eners [7,2022,32] and the atropisomeric CB pairs

    [2528].

    The first study on the GC GC separation of PCBs was

    presented by Phillips and Xu [32], who used theretention database of all 209 CBs published for 20 sta-

    tionary phases [39] to construct 2D gas chromatograms.

    The authors did not discuss the separation of individual

    congeners, but they predicted structured chromato-

    grams for the non-polar semi-polar DB-1 CNBP

    column combination. The CBs in the 2D plane were

    grouped together according to the number of chlorine

    substituents, and the position of the congeners within a

    homologue group was determined by the number of

    ortho chlorines, with the retention sequence being:

    4 < 3 < 2 < 1 < 0.

    In 2001, Haglund et al. [7] analyzed PCBs in technicalClophen A50. The authors used a rather unconventional

    column set-up, with a highly shape-selective liquid

    crystal LC-50 (10 m 0.15 mm 0.1 lm) column in the

    first dimension and a non-polar BPX-5 (0.25 m 0.1

    mm 0.1 lm) column in the second dimension. The

    non-ortho CBs were most strongly retained by the LC-50

    column, followed by the mono-, di- and multi-ortho CBs.

    As a consequence, the non- and mono-ortho congeners

    eluted at the highest temperature within each homo-

    logue group, and the toxic non-ortho CBs 77, 126 and

    169 and the mono-ortho CBs 105, 118 and 156 all

    showed up in the lower right-hand corner of the contour

    plot. In order to achieve reasonable second-dimension

    elution times (10C/min) had to be used.

    Under these conditions, all six planar CBs quoted above

    were successfully separated from other CBs present in

    the mixture within a run time of only 17 min. In addi-

    tion, the seven EU indicator CBs were also adequately

    resolved. In our opinion, despite the excellent separation

    of the non-ortho CBs from the other congeners, their

    quantification may be problematic, because the excessive

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    380 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    9/24

    bleed of the LC-50 phase at the elevated temperatures at

    which non-ortho CBs elute will also be modulated and

    dramatically increase the noise level and, thus, increase

    the LODs. In addition, the column combination does not

    allow the determination of the many other CBs, because,

    due to the fast temperature programme required, the

    separation power of the first-dimension column cannotbe used fully and many CBs co-elute.

    In 2002, Korytar et al. [20] tested three column

    combinations for the separation of 90 CBs with emphasis

    on the separation of the 12 WHO CBs. They preferred

    a classical set-up, with a non-polar HP-1 (30 m

    0.25 mm 0.25 lm) separating solely on the basis of

    volatility in the first dimension, and three more polar

    phases, BPX-50, HT-8 and SupelcoWax-10 (all 1 m 0.1

    mm 0.1 lm), in the second dimension. A complete

    separation of the WHO CBs from each other and from the

    other CBs was obtained with the latter two columns.

    With HP-1

    SupelcoWax-10, only six congeners wereinvolved in co-elutions, as against 12 congeners for

    HP-1 HT-8. However, the latter combination provided

    the highest information content, because structured

    chromatograms were obtained. The CBs were found to

    be grouped together according to the number of chlorine

    Table 5. Co-eluting PCBs on six column combinations [21,22]*,**

    DB-1 HT-8 HT-8 BPX-50 DB-XLB BPX-50 DB-XLB SP-2340 DB-XLB LC-50 DB-XLB BPX-70

    Non-ortho CB congeners 77/144

    Mono-ortho CB congeners 118/131a

    Marker CB congeners 52/69 153/168 101/90 101/90 153/168138/163/164

    Other CB congeners 23/54a 132/179a 38/47/62a 21/33 4/10 31/53a

    16/32 160/175a 20/21/33 47/62/65 20 /33 47/62/6520/21/33 20/33 66/155 42/59 43/69 42/5943/49 47/48 84/89 37/40a 62/65 57 /94a

    48/75 93/95/98 90 /101 57/94a 58/67 86/11242/59 112/119 107/123 58/67 63/76 106 /10741/64 97/117 63/76 88 /95 175/18261/70 108/107 86/125 84/8956/60 163/164 107/134a 83/119

    98/102 182/187 160/163 86/12588/91 175/182 160 /16383/112 201/204 175/182a

    115/116 196/203 201/204108/107 196/203139/149134/143146/165182/187196/203

    No. of CBs resolved byGC GC

    163 188 191 176 181 194

    No. of CBs resolved byGC GCMS

    165 192 192/194 184 183 198

    Temperature ramp (C/min) 1 1 1 1.5 1.5 0.5

    Modulation time (s) 3 3 3 4 5 5

    Run time (min) 140 146 144est. 100est. 90 240

    , all congeners separated.aResolved by MS.est.Run time not explicitly mentioned in paper; estimated from available information.*CB congener classification according to their presence in any of the Aroclors 1242, 1254 or 1260: Bold, >1.0 wt.%; Bold, 0.051.0 wt.%;italics, trace or undetected.**IUPAC numbering is used. Numbering of [21] was corrected accordingly.

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 381

  • 8/2/2019 Comprehensive Two Dimensional

    10/24

    substituents and, within a group, the number of ortho

    chlorines. This is the structure predicted by Phillips and

    Xu [32], but now verified experimentally. The ordered

    structure is very useful because it helps to predict the

    position of other CB congeners in the 2D plane. This

    enabled the prediction, on the basis of published 1D-GC

    retention data, that the WHO CBs will be separated fromall other congeners present in Aroclor mixtures. In

    addition, a large number of unknown peaks detected in a

    cod-liver extract could be provisionally identified as CB

    congeners.

    The effort to analyze PCBs by means of GC GC cul-

    minated in two recent studies [21,22], in which attempts

    were made to separate all 209 CBs. In these studies,

    seven column combinations were evaluated. The sepa-

    ration for six of these is visualized in Table 5. The

    DB-XLB HT-8 set-up is not included because it pro-

    vided a very limited improvement over 1D-GC due to the

    very similar separation mechanisms applied in both

    dimensions. When comparing the data of Table 5, oneshould keep in mind that the results also depend on run

    time or, more precisely, the temperature ramp during

    elution, and on modulation time. In general, slower

    elution and a shorter modulation time yield more effi-

    cient separation. Table 5 therefore includes these two

    parameters. As for the DB-1 HT-8 column combina-

    tion, the ordered structure quoted above has been con-

    firmed for all CBs (Fig. 3). The separation of the

    homologue groups was so clear cut that only one

    co-elution of two congeners, CBs 23 and 54, was caused

    by the overlapping of two (tri- and tetra-substituted)

    homologue series. All other co-elutions occurred within

    these series. For the rest, this column set performed

    much poorer, in terms of number of congeners resolved,

    than the other column combinations. Because most

    co-eluting compounds have closely similar mass spectra,

    additional separation by means of MS detection is pos-sible in a limited number of cases only such as here for

    CBs 23 and 54.

    As regards the other column sets, the best result was

    found for DB-XLB BPX-70, which separated 198

    congeners. However, the run time was as high as 4 h, or

    almost double that of most other procedures and it is

    questionable whether, in several instances, a limited

    overall increase of resolution justifies such an excessive

    demand of time. In other words, from a practical point of

    view, HT-8 or DB-XLB combined with BPX-50 or

    DB-XLB LC-50 are to be preferred with the first two

    sets separating 192 congeners in 2.5 h, and the latter a

    lower number (183) of analytes, but in a very short time(1.5 h). Two of these combinations separate all WHO

    CBs and EU indicator CBs. The HT-8 BPX-50 set has

    the added advantage of yielding structured chromato-

    grams. For the penta-CBs, this is shown in the apex plot

    of Fig. 4, which also clearly visualizes the non-ortho-

    number-based sub-division.

    One should consider that the above separations were

    studied with the congener concentrations in the stan-

    dard mixture being essentially the same and with

    RS > 0.5 as the separation criterion. As is well known,

    Figure 3. GC GCTOF-MS chromatogram of all CB congeners on DB-1 HT-8 column set [22].

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    382 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    11/24

    this resolution will not be satisfactory for quantifying

    partly co-eluting congeners with widely different con-

    centrations, as is often true for PCBs and other classes oforganohalogens in real-life samples. Even if RS > 1, the

    determination of the WHO CBs in one run together with

    predominant congeners may be difficult.

    Another problem was encountered when the target

    analytes had to be determined in a seal-blubber sample

    [21]. After the usual sample treatment, some of them

    were found to be present in concentrations too low to be

    detected by GC GClECD. One solution would be to

    concentrate the extract. However, some of the major

    congeners would then overload the second-dimension

    column and, in addition, it would be difficult to keep all

    congeners within the working range of the detector.Therefore, two separate injections will be required.

    6.1.1. Chiral analysis. With atropisomeric CBs, the

    preferred method is heart-cut MDGC [29]. Most

    reported methods use two-oven systems and comprise

    an achiral precolumn and a chiral main column. Heart

    cuts (from the precolumn) containing the atropisomers

    of interest and, inevitably, other PCBs or interferences

    are directed towards the chiral column. With typical

    precolumns, such as DB-5 or DB-XLB, three or four out

    of the 11 atropisomeric pairs present in technical for-

    mulations co-elute with some of the 140 CBs [29] and

    must be separated in the second dimension. Chirasil-

    Dex (permethylated 2,3,6-tri-O-methyl-b-cyclodextrin)

    is the most commonly used chiral column that can

    resolve nine out of the 19 stable, and seven out of the

    11 interesting pairs of chiral congeners [29]. The

    column sequence used in heart-cut MDGC has to be

    changed for GC GC, because the second-dimension

    column must be short and, if a short chiral column is

    used, no separation of the enantiomers would be

    achieved (i.e. the chiral column is used in the first

    dimension, and a polar or shape-selective column in the

    second dimension).

    In all published papers [2528], Chirasil-Dex was used

    as the first-dimension column; in one study [28],

    BGB-172 (25% 2,3,6-tert.-butyldimethylsilyl-b-cyclo-dextrin) and BGB-176SE (20% 2,3-di-O-methyl-6-O-

    tert.-butyldimethyl-b-cyclodextrin) were also evaluated.

    As second-dimension columns, Bordajandi et al. tested

    HT-8, VF-23ms and SupelcoWax-10 for 65 [28] or 90

    [27] CBs, whilst Harju et al. [25,26] used VF-23ms and

    LC-50 for all 140 Aroclor CBs. The results of the latter,

    more comprehensive study are displayed in Table 6. It

    shows that all nine atropisomeric pairs, which are

    resolved on Chirasil-Dex, partly or completely co-elute

    with one or more CBs in 1D-GC, while all but two or

    three of the co-elutions are resolved in the second

    dimension. As an application, grey-seal tissue was ana-lyzed on both column combinations. Seven out of the

    nine atropisomeric pairs were detected, and enantio-

    meric fractions could be determined for six of these (CBs

    91, 95, 132, 135, 149 and 174) by using lECD and/or

    TOF-MS detection; the second eluting enantiomer of CB

    84 co-eluted with CB 56. The most abundant

    Table 6. Co-elutions for nine atropisomeric CB pairs in 1D-GCand GC GC with 140 Aroclor CBs [26]*

    Atropisomeric

    CBs

    Co-elutants on

    Chirasil-Dex (CD) CD VF-23ms CD LC-50

    84 56, 90, 99, 101 56 9991 63 95 93 93

    132 176, 141 141 141 (0.7)a

    135 110, 82 136 115 149 77, 124 174 202 176 132, 141

    aRS in second dimension.*For classification code, see Table 5.

    0.7

    1

    1.3

    1.6

    1.9

    4200 4700 5200 5700 6200 67001tR (s)

    2tR

    (s)

    103 100

    94102

    88

    91

    84

    89989593

    127

    126

    104

    96

    111 120

    124

    105

    122114

    106118

    123

    108

    107

    121

    92

    90

    113

    99

    101

    112,119

    10983

    12587

    8611085

    115116

    97,117

    82

    0 50 50 5

    989593

    1098386116

    Figure 4. GC GCTOF-MS apexplotof thepenta-CBson HT-8 BPX-50columnset. Pink, tetra-orthoCBs; green, tri-orthoCBs;red,di-orthoCBs;light blue, mono-orthoCBs; dark blue, non-orthoCBs. Boxes represent co-eluting congeners [22].

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 383

  • 8/2/2019 Comprehensive Two Dimensional

    12/24

    atropisomers, CBs 135, 149 and 174, had enantiomeric

    fractions (EFs) close to the racemic value, while the EFs

    for CBs 91, 95 and 132 invariably deviated significantly

    from that value.

    6.2. Polychlorinated dibenzo-p-dioxins/furans

    (PCDD/Fs)PCDD/Fs are highly toxic compounds formed as

    by-products during a variety of chemical and combus-

    tion processes [40]. Notorious examples are their (trace-

    level) presence in technical PCB mixtures, chlorinated

    phenoxyalkanoic acid pesticides and fly ash. The number

    of chlorine substituents can vary from one to eight to

    produce up to 75 CDD and 135 CDF positional isomers.

    There is a pronounced difference in toxic and biological

    effects amongst these CDD/F congeners and toxicities

    vary 1,00010,000-fold. Seven 2,3,7,8-substituted

    CDDs and 10 CDFs are generally considered the most

    toxic, since they have toxic properties similar to 2,3,7,8-

    tetrachlorodibenzo-p-dioxin (2,3,7,8-TCDD), which isthe most toxic congener of the group. Consequently, the

    analytical challenge is to detect, and quantify, these 17

    priority congeners and, also, the 12 WHO CBs discussed

    above (which also show a dioxin-like effect). Such ultra-

    trace analyses are expensive, both because of the time-

    consuming pre-treatment and clean-up, and the need to

    use GChigh-resolution MS (GCHRMS) with its unri-

    valled sensitivity and selectivity [18]. In other words,

    sample throughput is low and cost is high. Not surpris-

    ingly, therefore, the potential of GC GC with its

    improved selectivity is eagerly explored to find less

    demanding but equally rewarding analytical ap-proaches.

    The first attempt to analyze dioxins and dioxin-like

    CBs by GC GC was made by Grainger et al. [16,17].

    They used an early sweeper for modulation and HRMS

    (resolution power 3000) for detection of a 24-compound

    mixture containing the non-ortho CBs 77, 81, 126 and

    169 and a suite of PCDD/Fs. Rather unconventional

    column dimensions were used: a very short first-

    dimension DB-5 column (2 m 0.25 mm 0.25 lm)

    and a rather long second-dimension OV-1701 column (3

    m 0.1 mm 0.05 lm). With this set-up, the separation

    took only 18 min, with an impressive iLOD of 335 attog

    (S/N 9) for 2,3,7,8-TCDD. Unfortunately, no further

    information was provided.

    As a part of their PCB study quoted above, Korytar

    et al. [20] tried to separate the priority CDD/Fs and WHO

    CBs from each other and the bulk of 90 CBs on

    HP-1 HT-8. The outcome was successful for all except

    one pair of target analytes, 1,2,3,7,8-PeCDD/CB 169.

    This stimulated a further search for a column combi-

    nation that would separate all target compounds from

    each other and, as importantly, from matrix

    co-extractants. In a subsequent study [19], therefore,

    seven first-dimension and eight second-dimension

    columns were tested. With a 100% methylpolysiloxane

    stationary phase (DB-1) in the first dimension to create

    orthogonal conditions, all congeners with different toxic

    equivalency factor (TEF) values could be separated if

    VF-23ms or LC-50 were used in the second dimension.

    When other types of first-dimension column were used

    (and orthogonality was partly sacrificed), a DB-XLBcolumn combined with 007-65HT, VF-23ms or LC-50

    gave a complete separation of all 29 priority congeners.

    With a spiked, fractionated milk extract, DB-XLB LC-

    50 was found to be the most powerful column combi-

    nation, because of the good separation of all priority

    congeners from each other as well as matrix constitu-

    ents. Analytical performance was satisfactory with a

    close to 3-order linearity, and iLODs of 30150 fg

    injected mass.

    Two of the quoted column sets the preferred

    DB-XLB LC-50 and also VF-1 LC-50 were used by

    Danielsson et al. [9,41], who studied the quantification

    of dioxins and dioxin-like CBs by GC GClECD. Fish oilfrom herring, spiked cows milk, vegetable oil and an eel

    extract were analyzed by two GC GC and four

    GCHRMS laboratories, with the latter serving as refer-

    ences. Fig. 5 shows typical GC GClECD chromato-

    grams of the mono-ortho-CB, and non-ortho-CB and

    CDD/F fraction of herring oil on DB-XLB LC-50, and

    clearly demonstrates the efficiency of the LC-50 phase to

    separate the target analytes from matrix co-extractants.

    The quantification data for WHO CBs obtained on both

    GC GClECD systems were closely similar and agreed

    very well with the GCHRMS data. As for the dioxins,

    data produced on DB-XLB LC-50 also showed goodagreement, but this was not true for VF-1 LC-50 due to

    interferences co-eluting with some congeners. For the

    rest, the CVs were somewhat higher for GC GC than for

    HRMS (540% vs. 227%). The total toxic equivalent

    (TEQ) data obtained on the preferred GC GC system

    compared well with those obtained by GCHRMS, with

    CVs for both techniques being below 10%. These results

    show that the proposed method meets the EC require-

    ments for a WHO CDD/F-plus-CB screening method,

    which should have a false negative rate of less than 1%

    and a TEQ CV of less than 30%. However, more samples

    need to be analyzed to confirm fully the criterion for the

    false negative rate. For the rest, as Danielsson et al. noted

    [9,41], before GC GClECD can be considered a con-

    firmatory method, the EC Directives will have to be

    revised; currently, MS is the only mode of detection

    allowed for this purpose.

    Focant et al. [36,38] devoted two studies to GC GC

    TOF-MS. In their first report [36], Rtx-Dioxin 2 was

    employed as the first-dimension column because it is

    known for its excellent separation of the priority CDD/Fs

    from each other and also from non-2,3,7,8-substituted

    congeners. For the second dimension, they selected a

    relatively non-polar Rtx-500 column (similar to HT-8),

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    384 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    13/24

    because they expected the second dimension to bringmore in terms of signal enhancement after modulation

    than added resolution retention on that column

    should, consequently, be rather weak. An iLOD of 0.5 pg

    was obtained for 2,3,7,8-TCDD. As for the separation of

    the 17 CDD/Fs and the four non-ortho CBs analyzed, one

    co-elution persisted in both dimensions (viz between CB

    126 and 2,3,7,8-TCDD). This co-elution is especially

    problematic because CB 126 is usually present in con-

    centrations one or two orders of magnitude higher than

    2,3,7,8-TCDD, causing a serious over-estimation of the

    sample TEQ due to the much higher toxicity of 2,3,7,8-

    TCDD. The problem was solved by deconvoluting the

    masses involved. The congener-specific data found by

    GC GCTOF-MS and GCHRMS for various environ-

    mental (sediment, fly ash) and biological (vegetation,

    fish) samples showed good agreement for 2,3,7,8-TCDD

    and most other congeners.

    However, in their next study [38], the authors used

    another column combination, Rtx-500 BPX-50. All

    priority CDD/Fs and four non-ortho CBs were then

    separated and, in a next run, all EU indicator and

    mono-ortho CBs were also separated from each other

    and other CBs. The iLOD of 2,3,7,8-TCDD improved to

    an impressive 0.2 pg. The system was used to analyze

    fish, pork, and milk samples and the results werecompared with conventional GCHRMS. Not unex-

    pectedly, the earlier conclusions on CBs were con-

    firmed. For the CDD/Fs, the results were strongly

    concentration dependent. When the average congener

    concentrations were rather high (above 0.4 pg/g fat

    weight, which corresponds to 1.1 pg of compound in-

    jected), as for fish, and a large sample intake was used

    (15 g), the two techniques showed good agreement.

    But, again, the CVs for GCHRMS were 714%, as

    against 1060% for GC GCTOF-MS. With pork and

    milk, for which the CDD/F concentrations were much

    lower (above 0.030.1 pg/g fat weight), there were

    many over-estimations with CVs up to 90%. However,

    the congener distribution was still well defined in all

    instances and can be used (e.g., for tracking sources of

    contamination). Despite the problems, the TEQ results

    for the CDD/Fs and CBs compared favorably with those

    of GCHRMS because a rather good description of the

    main TEQ contributors (2,3,7,8-TCDD, 1,2,3,7,8-

    PeCDD and 2,3,4,7,8-PeCDF) was achieved.

    6.3. Toxaphene

    Toxaphene is an organochlorine pesticide mixture of

    complex composition. The major constituents are

    Figure 5. GC GClECD of the (A) mono-ortho-CB and (B) non-ortho-CB and CDD/F fractions of a fish oil on DB-XLB LC-50. Assignment for(B): 1, CB 77; 2, CB 126; 3, CB 169; IS1, 1,2,3,4-TCDD; 4F1, 2,3,7,8-TCDF; 4D1, 2,3,7,8-TCDD; 5F1, 1,2,3,7,8-PeCDF; 5F2, 2,3,4,7,8-PeCDF;5D1, 1,2,3,7,8-PeCDD; 6D3, 1,2,3,7,8,9-HxCDD; IS2, 1,2,3,4,6,7,9-HpCDD; 7D1, 1,2,3,4,6,7,8-HpCDD; 8D1, OCDD [9].

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 385

  • 8/2/2019 Comprehensive Two Dimensional

    14/24

    chlorobornanes (ca. 75 wt.%), with chlorocamphenes

    in second place, while chlorodihydrocamphenes are

    present as minor components [42]. Individual members

    of these compound classes have been isolated from the

    technical mixture, and they are likely products of

    synthesis. The presence of chlorobornenes, which has

    also been reported [42] as a possible class of constitu-ents, is discussed at the end of this section. The number

    of congeners that can be found theoretically exceeds

    60,000 (cf. Table 1). 1D-GC, which is the method of

    choice to study the composition of technical toxaphene,

    cannot create a satisfactory separation. Typically, about

    100 peaks show up in a GCECD chromatogram. Better

    results were obtained by combining GC with other

    separation techniques, such as adsorption chromatog-

    raphy on silica [43] or active carbon [44], normal-

    phase LC [45], or heart-cut MDGC [46]. With these

    techniques, the number of compounds in technical

    toxaphene was estimated to be at least 177 [43], 246

    [45], 300 [46], and 675 [44]; all quoted methods werevery time-consuming (e.g., the 675-peak experiment

    required pre-fractionation into no less than 160 frac-

    tions with a subsequent 30-min GC analysis of each

    fraction). In the environment and particularly in higher

    organisms, many toxaphene constituents are degraded,

    and only a few are bio-accumulated. This leads to a

    significantly simpler toxaphene-residue pattern com-

    pared to the technical mixture. One problem with the

    quantification of individual congeners is the limited

    availability of the standards. Today, only 23 standards

    are commercially available (Table 7). Since various

    nomenclature rules were proposed for the toxaphene

    components [4753], each congener in Table 7 has

    several names.

    In view of the separation problems outlined above,

    expectations were high when GC GC became available

    as a tool to unravel the composition of complex mixtures

    [54]. Indeed, the use of an HP-1

    HT-8 column com-bination yielded highly structured chromatograms and

    easily revealed a complex mixture of over 1000 com-

    pounds in a run of less than 3 h. Subsequent analysis of

    the 23-standard mixture and EI-TOF-MS evaluation of

    technical toxaphene showed that the 2D chromatogram

    is structured according to the number of chlorine sub-

    stituents in a molecule (see Fig. 6), with little, if any,

    dependence on the class of compounds (bornanes or

    camphenes). The range of chlorination found with

    GC GCEI-TOF-MS was 511 substituents per mole-

    cule. Using home-made software to calculate the total

    area for each iso-substitution band, hexa- to nona-

    chlorinated compounds were found to be the majorcomponents of toxaphene and represented some 97% of

    the total toxaphene mass.

    In a more recent study [23], six column combina-

    tions were tested for the separation of 12 classes of

    organohalogenated compounds (see Section 7); one

    of these was technical toxaphene. Ordered structures

    were then also found for DB-1 combined with 007-210,

    007-65HT or LC-50. Somewhat surprisingly, another

    column combination, DB-1 VF-23ms, did not deliver

    ordered structures but offered the best overall separa-

    tion; the entire GC GC plane was used and visual

    Table 7. List of 23 commercially available standards for toxaphene and their various codes

    IUPAC name Parlar[47,48]

    Nikiforov[49]

    Wester et al.[50,51]

    AV-code[52]

    OK-code[53]

    2,2,3-exo,8,9,10-hexachlorocamphene 11 C[032001]-(11)2-exo,3-endo,8,8,9,10-hexachlorocamphene 12 C[021001]-(21)2-exo,3-endo,7,8,9,10-hexachlorocamphene 15 C[021011]-(11)2,2,5,5,9,10,10-heptachlorobornane 21 HpCB-6533 B[30030]-(012) B7-499 99-0132,2,3-exo,8,8,9,10-heptachlorocamphene 25 C[032001]-(21)2-endo,3-exo,5-endo,6-exo,8,8,10,10-octachlorobornane 26 OCB-4921 B[12012]-(202) B8-1413 198-3032,2,3-exo,8,8,9,9,10-octachlorocamphene 31 C[032001]-(22)2,2,5-endo,6-exo,8,9,10-heptachlorobornane 32 HpCB-6452 B[30012]-(111) B 7-515 195-1112,2,5,5,9,9,10,10-octachlorobornane 38 OCB-6535 B[30030]-(022) B8-789 99-033

    2,2,3-exo,5-endo,6-exo,8,9,10-octachlorobornane 39 OCB-6964 B[32012]-(111) B8-531 199-1112-endo,3-exo,5-endo,6-exo,8,9,10,10-octachlorobornane 40 OCB-4917 B[12012]-(112) B8-1414 198-1132-exo,3-endo,5-exo,8,9,9,10,10-octachlorobornane 41 OCB-3223 B[21020]-(122) B8-1945 41-1332,2,5-endo,6-exo,8,8,9,10- octachlorobornane 42a OCB-6460 B[30012]-(211) B8-806 195-3112,2,5-endo,6-exo,8,9,9,10-octachlorobornane 42b OCB-6454 B[30012]-(121) B8-809 195-1312-exo,5,5,8,9,9,10,10-octachlorobornane 44 OCB-2455 B[20030]-(122) B8-2229 97-0332-endo,3-exo,5-endo,6-exo,8,8,9,10,10-nonachlorobornane 50 NCB-4925 B[12012]-(212) B9-1679 198-3132,2,5,5,8,9,10,10-octachlorobornane 51 OCB-6549 B[30030]-(112) B8-786 99-1132,2,5-endo,6-exo,8,8,9,10,10-nonachlorobornane 56 NCB-6461 B[30012]-(212) B9-1046 195-3132,2,3-exo,5,5,8,9,10,10-nonachlorobornane 58 NCB-7061 B[32030]-(112) B9-715 103-1132,2,5-endo,6-exo,8,9,9,10,10-nonachlorobornane 59 NCB-6455 B[30012]-(122) B9-1049 195-1332,2,5,5,8,9,9,10,10-nonachlorobornane 62 NCB-6551 B[30030]-(122) B9-1025 99-0332-exo,3-endo,5-exo,6-exo,8,8,9,10,10-nonachlorobornane 63 NCB-3261 B[21022]-(212) B9-2206 169-3132,2,5,5,6-exo,8,9,9,10,10-decachlorobornane 69 DCB-6583 B[30032]-(122) B10-1110 227-133

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    386 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    15/24

    inspection showed that, here, the highest number of

    congeners was observed. The column sets that yielded

    ordered structures indicated the presence of a cluster of

    unknown compounds (Fig. 6). This group of com-

    pounds was not observed in the earlier study [54],

    because the temperature programme then used was

    much slower. This caused much more spreading of the

    toxaphene congeners in the 2D plane; wrap-around

    occurred and led to co-elution with the unknown

    compounds. Combined information on the peak shapes

    and the ECNI-TOF mass spectra added as inserts dem-

    onstrated that the unknowns were decomposition

    products formed during the first-dimension GC run (i.e.

    chlorinated bornenes, probably formed by HCl elimi-

    nation). So far, not a single polychlorinated bornene

    has been isolated from technical mixtures and the

    assumption about their presence was based only on

    GCMS and GCFTIR data [42]. The observations cited

    created serious doubts about the presence of bornenes

    in technical toxaphene.

    6.4. Polychlorinated alkanes (PCAs)

    PCAs are complex mixtures with a degree of chlorina-

    tion of 3070 wt.%, and carbon chain lengths of

    C10C13 (short-chain PCAs), C14C17 (medium-chain

    PCAs) or >C17 (long-chain PCAs). They are used as

    extreme-pressure additives in industrial cutting fluids,

    plasticizers and flame retardants for polyvinyl chloride

    (PVC) and other plastics and rubbers, and as additives

    in paints and sealants. PCAs are persistent and non-

    biodegradable, and they accumulate in the food chain.

    The global production of PCAs has been reduced since

    the early 1980s [55], but is still in the range of

    380,000 tons [56]. Short-chain PCAs cause particular

    concern due to the high amounts released into the

    environment, and their toxicity, which is higher than

    that of other PCAs.

    Analysis of PCAs is difficult because of the extreme

    complexity of the mixtures and the lack of quantification

    standards, and semi-quantification is all that can be

    achieved. In many environmental studies dealing with

    Figure 6. Total-ion GC GCECNI-TOF-MS chromatogram of technical toxaphene on DB-1 HT-8, with the mass spectra of two peaks asindicated. Lines indicate position of iso-substitution bands [23].

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 387

  • 8/2/2019 Comprehensive Two Dimensional

    16/24

    halogenated micro-contaminants, an essentially

    unresolved band covering a large part of the GC baseline

    is all that indicates the presence of PCAs. Currently,

    some 40 individual standards are available, but most of

    them, as will be demonstrated below, are not present in

    the technical mixtures. Mixtures of PCAs, with various

    chlorine contents and carbon-chain lengths, are used as

    reference standards for semi-quantification. Analysis is

    usually carried out by GC coupled to HRMS in ECNI

    mode. The method is based on monitoring [MCl] ions

    to determine the concentrations of individual classes (i.e.

    congeners with the same number of carbon and chlorine

    atoms). HRMS is highly selective and eliminates inter-

    ferences caused by other polychlorinated pollutants and

    Figure 7. (A) GC GCECNI-TOF-MS extracted ion chromatogram (m/z 7073) of polychlorinated decanes with 55 wt.% Cl.(B) GC GCECNI-TOF-MS chromatograms of C10C13 technical mixture with 55 wt.% Cl. Colored lines indicate the position of apices withinthe band of polychlorinated (red) decanes, (green) undecanes, (blue) dodecanes and (black) tridecanes. (C) Overlay of GC GCECNI-TOF-MSchromatograms of (red) short-, (green) medium- and (blue) long-chain PCA mixtures with different Cl content. White numbers indicate number of(carbon + chlorine) atoms of the compounds present in the bands [24].

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    388 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    17/24

    PCAs with the same nominal mass. However, this

    detection method is not available in many laboratories

    and is too expensive for routine analysis, so low-resolution

    MS (LRMS) is also used for quantification. An interna-

    tional inter-comparison study for short-chain PCAs has

    shown that the two techniques can give comparable

    quantitative results (e.g., in biota [57,58]). Nevertheless,because of the increased risk of interferences, improved

    sample clean-up and/or the use of GC GC instead of

    1D-GC is urgently required.

    The potential of GC GC was investigated in a recent

    paper in which six column combinations were tested

    [24]. The highest number of separated congeners was

    found with the DB-1 VF-23ms combination, but, for

    characterization and quantification, DB-1 007-65HT

    was preferred because this set provided more informa-

    tion in terms of ordered structures (i.e. group and sub-

    group separation). As demonstrated for polychlorinated

    decanes in Fig. 7A, the separation of PCA congeners

    with the same chain length is based on the number ofchlorine substituents. The authors added that the

    number of congeners in the technical mixtures (formed

    by uncontrolled synthesis) was relatively rather limited.

    If a larger number were present, the iso-substitution

    bands would become broader and neighboring bands

    would start to overlap. This is demonstrated in Fig. 7A

    by the outlying positions of several individual congen-

    ers, which should be considered unlikely products of

    uncontrolled synthesis. With mixtures of PCAs of

    varying chain length, the ordered structures comprise

    compounds having the same number of carbon-plus-

    chlorine atoms for example, C10Cl8 is on the samediagonal line as C11Cl7, C12Cl6 and C13Cl5. This is

    elegantly visualized in Fig. 7B. The position of the

    various compounds in each diagonal band depends on

    the number of carbon atoms: compounds with longer

    carbon chains have lower second-dimension retention

    times. This carbon-chain-length selectivity creates a

    distinct separation of compounds that differ by at least

    three carbons; C10 and C13 compounds show no over-

    lap in Fig. 7B. For a mixture of short-, medium- and

    long-chain PCAs, the (carbon+chlorine)-based ordering

    is seen to hold over a summed-number range of at least

    14 to 26. In addition, the carbon-chain-length selec-

    tivity creates a partial separation of the three PCA

    groups (Fig. 7C). Obviously, using GC GC is a major

    step forward in PCA analysis, although, simulta-

    neously, it is also clear that additional separation power

    possibly provided by LC-based sample fractionation

    will be needed for a satisfactory overall unraveling of

    the composition of PCA mixtures.

    Two relevant examples of the added value of GC GC

    compared to 1D-GC are as follows. In the quoted study

    [24], two dust samples were analyzed, and significantly

    different GC GClECD patterns of the short- and med-

    ium-chain PCAs were easily observed. Visual evaluation

    of the chromatograms showed that one sample con-

    tained more medium- than short-chain PCAs and that

    these medium-chain PCAs had a relatively low degree of

    chlorination, while the other sample contained more

    short- than medium-chain PCAs, with the short-chain

    PCAs having a higher degree of chlorination. In other

    words, pattern recognition based on the visual evalua-tion allowed provisional identification of sample com-

    position.

    Another advantage of GC GC is the elimination of

    LRMS interferences amongst the PCAs. Recently, Reth

    and Oehme [56] discussed three critical limitations of

    LRMS for the analysis of short- and medium-chain PCAs.

    [MCl] ions of a specific PCA (e.g., [MCl + 2] of

    C11H17Cl7 with m/z 360.9) interfere with:

    (i) the [MCl] ions of a PCA with five carbon atoms

    more and two chlorine atoms less (i.e. [MCl]

    (m/z 361.1) of C16H29Cl5);

    (ii) the [MCl] ions of a PCA with two carbon atoms

    more and one chlorine atom less (i.e. [MCl + 8](m/z 361.0) of C13H22Cl6); and,

    (iii) with [M + Cl] ions of a PCA with the same number

    of carbon atoms and two chlorine atoms less (i.e.

    [M + Cl] (m/z 360.9) of C11H19Cl5).

    GC GC solves these three problems due to the

    enhanced chromatographic resolution.

    6.5. Polybrominated diphenylethers (PBDEs)

    PBDEs are widely used as flame retardants (e.g., in

    polymers, textiles, electronic boards) and, similar to the

    PCBs, there are 209 BDE congeners. However, with only

    some 2025 congeners, the composition of the technicalPBDE mixtures is rather simple. Most environmental

    monitoring programmes focus on the analysis of seven

    congeners (BDEs 28, 47, 99, 100, 153, 154 and 209),

    which are most abundant in technical mixtures and are

    conventionally considered a type of reference set, similar

    to the seven EU indicator CBs. However, in environ-

    mental and biota samples, many other BDEs are found

    than the 2025 technical congeners because of their

    photolytic and biological debromination and metabolism

    in higher animals.

    The analysis of PBDEs is carried out by GCECD or

    GCMS (EI or ECNI mode) [59]. Today, 125 BDE cong-

    eners are commercially available and their retentioncharacteristics on seven stationary phases have been

    reported [60]. No stationary phase separates all cong-

    eners and not even the seven reference BDEs from all

    others. The best separation was achieved on a DB-XLB

    column with which 70 congeners were separated, while

    55 congeners were involved in 22 co-elutions. In real-life

    samples, the situation is even worse because other bro-

    minated compounds (e.g., polybrominated biphenyls

    (PBBs), hexabromocyclododecane (HBCD), tetrabromo-

    bisphenol-A (TBBP-A) or dimethyl-tetrabromobisphenol-

    A (me-TBBP-A)) may be present in the sample extract

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 389

  • 8/2/2019 Comprehensive Two Dimensional

    18/24

    and cause further interferences, which, often, cannot be

    solved by means of MS.

    GC GC of BDEs is challenging because of the high

    boiling points and thermal instability of the higher bro-

    minated congeners. In a study to be discussed in more

    detail in Section 7, Focant et al. [35,37] found that the

    higher brominated congeners were retained too stronglyin the second dimension, with the limited thermal sta-

    bility of most stationary phases not allowing a suffi-

    ciently high temperature to speed up the separation. In

    the end, an HT-8 column was used; it was stable up to

    360C. However, even with a DB-1 HT-8 column set,

    hepta-BDE 183 and deca-BDE could not be included

    because retention was still too great.

    Korytar et al. [61] used GC GClECD to separate the

    125 BDEs recently marketed. For the first dimension,

    DB-XLB was not a good choice because nona- and deca-

    substituted congeners were decomposed completely on

    this phase and partial decomposition was observed down

    to hexa-BDEs. Fortunately, on DB-1, decomposition ofnona-BDEs was negligible and only slight decomposition

    of deca-BDE was observed. As for the second dimension,

    from amongst six columns tested, 007-65HT was found

    to add most to the selectivity of the first-dimension sep-

    aration. In contrast with the findings quoted above, no

    extreme peak broadening or trapping was observed for

    the nona-BDEs 206, 207 and 208, and deca-BDE 209.

    Possibly, the band broadening in [35] was caused by a

    low temperature of the MS transfer line, which houses a

    significant part of the second-dimension column or by

    too low a temperature of the cooling gas used for

    modulation.Fig. 8 shows the considerably improved separation on

    the DB-1 007-65HT column set, with only 17

    co-eluting pairs involving 35 congeners. In addition, the

    seven reference BDEs were separated from all other BDEs.

    When a dust extract was analyzed (insert in Fig. 8), 18

    BDE congeners were identified.

    Various other brominated flame retardants and a

    number of BDE metabolites were all found to elute within

    the BDE band, and that caused several more co-elutions

    (Fig. 8). Of course, some separations were improved,

    notably that of TBBP-A and BDE 153; the latter analyte

    can now be quantified even if a high concentration of

    TBBP-A is present, as was true for the dust sample.Finally, Fig. 8 shows that second-dimension separation

    facilitates the use of fluorinated BDEs as internal stan-

    dards, because all F-BDEs that co-elute with the parent

    compound in the first dimension are separated in the

    second dimension, appearing just below that parent.

    Figure 8. GC GClECDof ( ) BDEs,( ) fluorinated BDEs,( ) otherbrominated flame retardants and( ) BDE metabolites on DB-1 007-65HT.Insert: GC GClECD contour plot of dust extract [61].

    Trends Trends in Analytical Chemistry, Vol. 25, No. 4, 2006

    390 http://www.elsevier.com/locate/trac

  • 8/2/2019 Comprehensive Two Dimensional

    19/24

    7. Between-class separations

    For obvious reasons, the common denominator of nearly

    all papers devoted to GC GC analysis of organohalo-

    gens is the (much) improved within-class resolution that

    can be achieved, with the joint addressing of the priority

    CDD/Fs and CBs as the only, and logical, exception.Because of the marked successes so obtained, we now see

    a gradual shift of interest to the more intriguing

    between-class separations [23,35,37].

    As a first attempt, Focant et al. [35] used GC GC

    TOF-MS for the simultaneous determination of 38

    predominant CBs, 11 persistent halogenated pesticides

    (OCPs), one PBB and eight PBDEs. With the DB-1 HT-8

    column combination selected, of 58 test compounds,

    only one pair of CBs was not resolved.

    In their next study [37], the authors determined the

    priority compounds in serum and milk. Single-injection

    GC

    GC was compared with a validated GCHRMSprocedure, which required three separate injections with

    three different temperature programmes and was, of

    course, much more time-consuming. The data shown for

    human serum corresponded very well between the two

    techniques. For the milk sample and the more abundant

    Figure 9. GC GClECD on DB-1 LC-50 of: (A) PCBs, PBBs, PCDEs, PBDEs, PCDTs,h PCNs, PCDD/Fs, OCPs, individualtoxaphene standards; (B) PCAs (PCA-60) as color contour plot and other classes as black dots; (C) PCTs (Aroclors 5442 + 5460) as color contourplot and dioxin-like CBs (black dots) and planar PCTs (white arrows); (D) PCDD/F fraction of a sediment extract [23].

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 391

  • 8/2/2019 Comprehensive Two Dimensional

    20/24

    congeners (>1 ng/g of lipid), per-cent deviations between

    the two methods were below 20%, which is acceptable.

    However, for some less-abundant congeners (

  • 8/2/2019 Comprehensive Two Dimensional

    21/24

    separations were found to strongly depend on the

    column combination used.

    We summarize a few relevant conclusions here:

    On DB-1 HT-8, congener separation on the basis of

    the number of halogen substituents, well-known for

    PCBs and toxaphene, was also observed for PCTs

    and PBDEs. However, there was no group-typeseparation (i.e. although the DB-1 HT-8 set can be

    successfully used for within-class separation, there is

    a high risk of interferences if congeners from other

    compound classes are present).

    Group separation based on planarity was obtained

    with the DB-1 LC-50 column set. Fig. 9AC shows

    that three groups of analytes could be distinguished.

    Three-ring planar compounds, such as PCDD/Fs,

    PCDTs and planar PCTs, were most strongly retained.

    Next in line were two-ring planar compounds, such as

    PCNs and planar PCBs. Non-planar analytes showed

    least retention and did not interfere with the planar

    compounds. The practicability of this column set isdemonstrated in Fig. 9D, which shows a chromato-

    gram of the PCDD/F fraction of a sediment sample

    after fractionation on a carbon column. Obviously, a

    properly tuned GC GC system could accommodate

    a very high number of compounds in the 2D plane,

    and could separate dioxins (indicated by black acro-

    nyms) from co-extractants (white-yellow band along

    first-dimension axis).

    Group-type separation was also delivered by

    DB-1 007-65HT (Fig. 10A-B). Here, the PBDEs were

    the strongest retained compounds in the second

    dimension and the PCAs are the least retainedorganohalogens. This is a rewarding result because

    PCAs and PBDEs are usually present in the same frac-

    tion after clean-up. As an example, the analysis of a

    dust sample is shown in Fig. 10C: the group separa-

    tion is clear-cut. Finally, Fig. 10B displays the overlay

    of three mixtures toxaphene, PCA-60 and Aroclors

    5442 plus 5460. What shows up in the 2D plane

    can be interpreted in two rather different ways:

    one is to emphasize that the presence of these three

    types of mixtures in any real-life sample will virtu-

    ally obscure all other classes of organohalogens,

    even in GC GC; but,

    it is as interesting to note that there is a striking

    separation of the three compound classes of

    interest.

    In addition, clearly ordered structures are observed for

    toxaphene and PCAs.

    The last column combination tested, DB-1 VF-

    23ms, yielded excellent within-class separation, espe-

    cially of non-aromatic compounds OCPs, toxaphene

    and PCAs but there was no group separation. This

    column set should, therefore, be preferred when

    unraveling the composition of, specifically, toxaphene

    and the PCAs, is the main aim of a study.

    8. Conclusions and perspectives

    Accurate, precise congener-specific analysis of organo-

    halogenated compounds has been a main goal of ana-

    lytical research since the early days of OCP and PCB

    determination in environmental samples in the 1960s.

    Three areas of interest can be recognized, viz. to ensure/improve:

    (i) the separation of the analytes of interest from other

    constituents present in a sample to enable their

    quantification;

    (ii) the sensitivity of the selected procedures and

    achieve the required LODs; and,

    (iii) the reliability of the identification/confirmation of

    the target analytes and/or unknowns.

    The common denominator of well-known milestones

    of analytical chemistry, such as the introduction of

    capillary GC columns, the coupling of GC with selec-

    tive detectors, such as the ECD and low- or high-

    resolution MS, and the introduction of heart-cut MDGC,is that they all simultaneously improved these three

    aspects.

    The present review shows that GC GC is effecting a

    considerable further improvement and that this tech-

    nique will no doubt become another milestone in

    improving the analysis of the various classes of

    organohalogenated compounds of interest at present. It

    is worthwhile adding that the relatively brief introduc-

    tory sections on instrumental aspects show that GC GC

    can be considered a mature technique (with the excep-

    tion of sufficiently rapid and (semi-)automated data

    handling and interpretation) and that proper instru-mentation is commercially available.

    The degree to which the overall resolution of

    organohalogens can be improved by means of GC GC

    depends on differences in the physicochemical properties

    of the target analytes but, of course, also on the (non)-

    complexity of the technical mixtures. The best results for

    within-class separations were observed for very complex

    mixtures, such as toxaphene, PCAs and PCTs, for which

    about 10 times as many peaks can easily be identified.

    For less complex mixtures, such as PCBs and PBDEs, the

    gain is self-evidently more modest, but still some 50%

    more congeners were resolved than by 1D-GC. For the

    class of PCNs, with only 75 congeners, the improve-

    ment was limited, irrespective of the stationary phase

    used. However, one should always consider that the

    separation of the target analytes from matrix constitu-

    ents is another key issue and, here, GC GC contributes

    to a satisfactory overall improvement also in the case of

    limited numbers of congeners. Between-class separations

    of the various groups of organohalogens come in more

    or less the same category, and improvement is pro-

    nounced because of the larger differences in physico-

    chemical properties between classes than amongst

    congeners of the same class. A striking example is the

    Trends in Analytical Chemistry, Vol. 25, No. 4, 2006 Trends

    http://www.elsevier.com/locate/trac 393

  • 8/2/2019 Comprehensive Two Dimensional

    22/24

    essentially complete separation of the technical mixtures

    of toxaphene, short-chain PCAs and PCTs in one

    GC GC run.

    The selection of a suitable column combination is

    essential to achieve the desired separation. Certainly for

    the fairly hydrophobic and essentially non-polar

    organohalogens, the best option is to use a non-polarstationary phase in the first dimension, because then a

    truly orthogonal system can be set up. Pure dimethyl-

    polysiloxane phases containing 08% diphenyl groups or

    those stabilized with polycarborane are preferred and are

    easily available from various producers. The separation

    mechanism in the second dimension should differ as

    much as possible from that in the first dimension. Today,

    commercial columns with geometry suitable for use in

    the second dimension are coated with one of the fol-

    lowing stationary phases polyethylene glycol or dim-

    ethylpolysiloxane with up to 70% diphenyl groups, up to

    100% cyanopropyl groups, up to 50% trifluoropropyl

    groups or up to 50% liquid crystals. Detailed experi-mental work has shown that all of these, except the

    trifluoropropyl phase, effect a significantly improved

    separation of the organohalogens.

    One problem common to all second-dimension sta-

    tionary phases, except the phenyl phase, is their low

    temperature stability. Further attention needs to be paid

    urgently to the production of low-bleeding polar/shape-

    selective phases that are stable up to 300C. Another

    challenge for the manufacturers is to produce columns

    containing two different stationary phases, with a

    discrete in-between border. Coupling columns by means

    of press-fits, which is a rather laborious and delicate job,will then become superfluous.

    A notable exception to the (non-polar) (polar/shape-

    selective) set-up is encountered in the analysis of enan-

    tiomers such as the 19 pairs of atropisomeric CBs. The

    shape-selective (i.e. chiral) column now has to be used in

    the first dimension, because the rather marginal sepa-

    ration usually obtained on such phases, makes it abso-

    lutely necessary to use a long column of, typically,

    2060 m: the 12 m length of a second-dimension

    column would have little effect.

    Analyte detectability can also be affected by the

    column combination selected. Low-bleeding columns

    should be used in the first dimension to keep noise as low

    as possible. This is another reason why non-polar sta-

    tionary phases are preferred in the first dimension. In

    addition, the second-dimension separation is performed

    under isothermal conditions. Consequently, if a study

    mainly aims to improve LODs, phases that hardly, or do

    not at all, retain the target analytes are preferred in the

    second dimension, in order to generate very narrow and,

    thus, very high peaks. In this way, an impressive iLOD of

    0.2 pg was obtained for 2,3,7,8-TCDD with TOF-MS

    detection. However, we would agree that the key aspect

    of the analytical approach is now lost it is essentially a

    1D-GC operation and there is no or very limited

    improvement of the separation.

    As regards the detection of organohalogens, today the

    lECD is the most sensitive detector available for these

    compounds with iLODs typically down to 10 fg. The

    data-acquisition rate of 50 Hz is satisfactory for nearly

    all applications, but there is considerable peak broad-ening due to the relatively large volume of the detector

    cell (150 ll). In other words, the development of a

    lECD with a 3050 ll cell volume would be a distinct

    bonus.

    For the rest, whatever the merits of lECD detection

    (and these are significant), for analyte identification/

    confirmation, one has to use a mass spectrometer.

    Today, the work-horse is the Leco TOF-MS, which can

    acquire up to 500 spectra per second over the entire

    mass range and is described in the literature for a wide

    variety of analytes and sample types. With this instru-

    ment, excellent deconvolution of mass spectra can be

    obtained. As can be seen from the present review, this isalso true for most organohalogen analyses.

    However, it is also clear that, for ultra-trace studies,

    satisfactory results cannot always be obtained. In such

    instances, the