construction of simultaneous spr and qcm sensing platform
TRANSCRIPT
ORIGINAL PAPER
Construction of simultaneous SPR and QCM sensing platform
JongMin Kim Æ SeongHoon Kim Æ Tatsuya Ohashi ÆHiroshi Muramatsu Æ Sang-Mok Chang ÆWoo-Sik Kim
Received: 17 May 2009 / Accepted: 5 August 2009 / Published online: 26 August 2009
� Springer-Verlag 2009
Abstract To construct a novel simultaneous SPR and
QCM sensing system, an AT-cut quartz crystal is fabri-
cated by sputtering 250 nm of ITO on one side of the
quartz plate over a 5-nm thick underlay of titanium, while a
50-nm thick layer of gold is sputter-deposited on the other
side to induce a total light reflection of an incident laser
beam on the thin gold layer. The signals of the sensing
system are detected using a Handy-SPR and QCA922 when
dropping 200 lL of Milli-Q water into the sensing cell.
A decrease in the SPR reflected light intensity is clearly
identified. In the same experiment, the changes in the
resonant frequency and resistance are about 2 kHz and
200 X, respectively. Furthermore, the oscillation stabilities
of the resonant frequency and resistance are about 50 Hz
and 2 X, respectively, which are sufficient to detect a large
mass change on the QCM/SPR chip.
Keywords SPR � QCM � Resonant frequency �Resonant resistance � SPR angle � ITO � Titanium �Index matching oil
Introduction
Chemical sensing systems are widely used in the fields of
health care, environmental monitoring and preservation,
and the agricultural and chemical industries. However, in
the real world, chemicals are often mixed and the con-
centrations of the target chemicals are invariably low,
thereby requiring complicated procedures and a long
examination time for the accurate detection of such
chemicals. Thus, a simple sensing device for detecting
target chemicals in chemical mixtures need to be devel-
oped, especially for use in health care, environmental
preservation, and process control. Several research groups
have already tested different combinations of receptors and
transducers. Yet, the selectivity and sensitivity are not
enough and the reliability is questionable, as the sensing
signal is usually obtained based on one analytical variable
only. Therefore, attempts have also been made to develop
sensing systems that obtain multiple sets of analytical
variables simultaneously.
The current authors previously developed simultaneous
sensing systems to analyze the viscoelastic characteristics
of polymer blend thin films using the quartz crystal reso-
nance (QCR) and differential scanning calorimetry (DSC)
[1], the local viscoelasticity and surface morphology of
polystyrene thin films using the QCR and atomic force
microscopy (AFM) [2], the electrochromic and viscoelastic
property changes of polypyrrole thin films during the redox
process using the QCR and UV–visible spectroscopy [3],
the capacitor characteristics and surface morphology of
polymer thin films using the QCR and AFM [4, 5], and the
morphology and physical properties of cultured cells using
a micro camera and quartz crystal [6, 7].
Furthermore, simultaneous sensing systems combining
the QCR and surface plasmon resonance (SPR) have
J. Kim � S. Kim � S.-M. Chang
Department of Chemical Engineering, Dong-A University,
840 Hadan-Dong, Saha-Gu, Pusan 604-714, Korea
T. Ohashi � H. Muramatsu
Department of Bioscience and Biotechnology,
Tokyo University of Technology, 1404-1, Katakura, Hachioji,
Tokyo 192-0982, Japan
W.-S. Kim (&)
Department of Chemical Engineering, ILRI,
Kyunghee University, Seochun 1, Kiheung, Yongin,
Kyungki-Do 449-701, Korea
e-mail: [email protected]
123
Bioprocess Biosyst Eng (2010) 33:39–45
DOI 10.1007/s00449-009-0370-5
recently been explored to improve the sensitivity and
reliability of both sensor systems [8–12]. In these appli-
cations, two types of response based on the piezoelectricity
and surface plasmons are used as the main information.
Piezoelectricity is a phenomenon related to the pro-
duction of an electrical potential when certain materials
(mainly crystals) are compressed. This process is also
reversible, as when an electric potential is applied to such
materials this produces deformation. Therefore, when an
alternating electric potential is applied to a piezoelectric
material, this generates an acoustic wave that is affected by
changes in temperature, pressure, and most importantly, by
changes in the physical properties at the interface between
the material surface and a foreign fluid or solid. As such,
piezoelectric crystal sensors are passive solid-state elec-
tronic devices that can respond to such changes. A selec-
tive sensor is then obtained when the sensor surface is
coated with a selectively interacting thin film, which was
introduced by King in 1964 based on a piezoelectric
sorption detector. A quartz crystal microbalance (QCM) is
also well known as a sensitive mass detecting device and
thin film microrheology monitoring device, and its theory
and applications have also been well documented in pre-
vious reviews [13]. In the case of a 9 MHz quartz crystal,
the mass sensitivity is normally of a nano gram order, but
applying a high frequency mode [14, 15] and an overtone
mode [16–18] produces better sensitivity within a sub-nano
gram order.
Meanwhile, surface plasmons are longitudinal charges
that propagate waves along the interface of a metal and a
dielectric. Surface plasmon can be excited using the
experimental systems developed by Otto and Kretchmann,
yet most SPR instruments use Kretchmann’s method,
which is an attenuated total reflectance (ATR) configura-
tion. Since the SPR occurs at the boundary of a metal and
external medium, these oscillations are very sensitive to
any change in the boundary, such as the adsorption of
molecules onto the metal surface or changes in the mor-
phology and structure. Its theory and applications have
already been well described by Rebecca et al. [19]. The
SPR response angle is related to the refractive index of a
material which is proportional to the mass change of the
target material. While the resonant frequency change of a
QCM is also related to the mass change of the target
material, it is additionally affected by the viscoelasticity of
the target material. Therefore, these complementary char-
acteristics mean that simultaneous detection using both the
SPR and a QCM makes the sensing system more accurate
and reliable by avoiding the individual limitations of each
sensor system. Thus, various studies have attempted to
analyze the sensing responses of materials using a con-
secutive combination of the SPR and a QCM. The direct
mass sensitivity comparison between QCM and SPR is
reported elsewhere [20]. Accordingly, this study fabricated
a QCM system that allows a simultaneous QCM and SPR
investigation. The responses of the SPR angle, resonant
frequency, and resonant resistance are all simultaneously
measured using various index matching oils, plus the fea-
sibility of practical applications is also examined.
Experiment
Materials and instruments
A 9 MHz AT-cut quartz plate (7.9 9 7.9 9 0.2 mm, QA-
A9M-AU, SEIKO E.G.&G), Ag paste (Dotite FA-705BN,
Fujikura Kasei), immersion oil(nd = 1.516, Olympus),
silicon oil (KF-56, Shinetsu Chemical), cover glass (No.1S
(18 9 18 9 0.18 mm), Matsunami), and Milli-Q water
(Milli-Q integral, Millipore) were all used to fabricate the
QCM/SPR sensing system. Plus, a batch-type sputter (CFS-
4ES-232, Shibaura), temperature-controlled incubator
(CN-25A, Mitsubishi), quartz crystal oscillation measuring
system (QCA922, SEIKO E.G.&G), probe-type film
thickness measurement instrument (Dektak 8, Veeco Co.
Inc.), and Handy-SPR (PS-0109, NTT-AT) were also used
as fabricating and measuring instruments.
Fabrication of simultaneous QCM/SPR chip
The choice of the electrode material is important to provide
a sharp SPR peak and good stability. The index matching
oil is also important to remove any dissonance between the
quartz plate and the bottom of the hemispherical prism
when the quartz crystal is oscillated, and should be a low
viscous medium to reduce the oscillating energy of the
quartz crystal. Furthermore, the index matching oil allows
an optical coupling of the quartz plate to the bottom of the
hemispherical prism to eliminate any gap between their
refractive indices [19].
One side of the AT-cut quartz crystal plate was sputtered
with 250 nm of ITO over a 5-nm thick underlay of tita-
nium, while the other side of the plate was sputtered with
50 nm of gold. The original diameter of the quartz crystal
electrode was 5 mm, and the thickness of the sputtered
electrodes was evaluated using a thickness measuring
instrument (Dektak 8, Veeco Co. Inc.). After sputtering, the
ITO electrode was annealed at 350 �C for 2 h to reduce any
inherent resonant resistance, and the final resonant resis-
tance was measured to be about 1.5 9 10-4 X cm. The
quartz crystal was conductively connected to a cover glass
with a gold surface and sealed using the same cover glass.
For this purpose, both sides of the cover glass were sput-
tered with 120 nm of gold over a 20-nm thick underlay of
chromium to transmit the QCM signals with a sputtering
40 Bioprocess Biosyst Eng (2010) 33:39–45
123
area of 7 9 7 mm2. Figure 1 shows a schematic illustration
of the preparation procedures for the novel simultaneous
QCM/SPR chip. The quartz crystal was optically coupled
to the prism, and any air removed by inserting the index
matching oil between the quartz crystal and the cover glass.
Novel cell assembly for simultaneous QCM/SPR
system
Figure 2 shows a schematic diagram of the principle of the
SPR system (a) and the proposed measuring cell (b) for the
simultaneous SPR and QCM sensing system. Normally,
SPR uses an irradiation light source of a 65–75� incidence
angle to produce evanescent wave on the surface of a gold
chip. Then, this evanescent wave and a surface plasmon
wave occurring on the surface cause a surface plasmon
resonance. As a consequence, the total reflected light
power will be particularly decreased at the incident angle.
This incident angle is called as the SPR angle, thus the
change of the refractive index caused by a certain kind of
mass loading is useful as a sensing response. Our purpose
is replacing the SPR chip with a QCM to enable multi-
functional sensing or analytical system. To enable the
purpose, a QCM should be oscillated in the SPR environ-
ment, and a purpose specific cell design is required. Thus,
only one reflection site of the QCM is necessary for
operating the SPR, but QCM requires both the electrodes
on both the quartz surfaces to oscillate. Thus, we used a
transparent metal electrode (ITO) as shown in Fig. 2b and
Fig. 1. In Fig. 2b, only one side of the quartz crystal
electrode was exposed to the liquid sample, while the other
electrode was mounted onto the bottom of a hemispherical
prism. The cell was filled with the index matching oil to
eliminate any air between the quartz crystal and the cover
glass. The cell was made of an acrylic plate and the height
controlled using a soft lithography technique with a PDMS
(polydimethysiloxane) stamp. The SPR instrument uses a
software provided by the manufacturer (Sprwin 3 for
Handy-SPR PS-0109, NTT-AT). The software supports a
continuous measuring mode with a gate time 3 s. In the
case of the QCM instrument, the possible gate time is about
0.1 s with a frequency resolution of 0.1 Hz and a resistance
resolution of 0.1 X. Thus, continuous measuring of both
the systems can be possible with a gate time 3 s.
Results and discussion
SPR response of QCM/SPR chip
The test cover glass used was No.1S (Matsunami, thick-
ness = 0.18 mm), as its thickness was similar to the
thickness of the quartz crystal. The glass was prepared by
ultrasonic cleaning in acetone for 5 min and Milli-Q water
for another 5 min, then dried using a vacuum pump. A gold
electrode was then fabricated on the cover glass using a
sputtering technique. Thereafter, 5 ll of the immersion oil
(nd = 1.516), as the index matching oil, was dropped on
the bottom of the hemispherical prism, and the prepared
cover glass placed over the top. The SPR response
according to the function of the gold layer thickness was
then investigated using SPR instruments, and the maximum
resonant peak obtained with a 50 nm gold layer. As the
adhesion of pure gold to the cover glass was not strong
enough to be used as a measuring device, the gold was
sputtered over another adhesive metal layer with a strong
adhesion to the cover glass to prevent the gold layer
peeling off the cover glass. A titanium film was used as the
adhesive metal layer over the cover glass, and the effect of
the titanium film on the SPR response was examined.
An adhesive metal film of one tenth the thickness of the
target electrode was found to be sufficient. Thus, a 50-nm
gold layer was sputtered over a 5-nm titanium layer. The
SPR response was then measured by dropping 100 ll of
Milli-Q water on the gold layer and the response compared
with that of a standard commercial SPR chip (Gold chip
011992, NTT-AT). The results are presented in Fig. 3.
While the resonant angle of the proposed chip was about 2�
Fig. 1 Fabrication procedure
for a simultaneous QCM/SPR
chip. The electrode construction
is based on 3-metal layers to
enable both the QCM and the
SPR function. a Preparing cover
glass, b Fixing QCM, c Final
QCM/SPR chip
Bioprocess Biosyst Eng (2010) 33:39–45 41
123
different from that of the commercial chip, the sharpness of
the SPR response was almost the same. The difference in
the resonant angle may have originated from the refractive
index gap between the two kinds of metal electrode. Thus,
the results confirmed that the proposed SPR chip can be
used as a measuring chip in the new QCM/SPR system.
QCM stability tests with QCM/SPR system under
atmospheric conditions
The stability of the QCM in the simultaneous sensing
system was first tested in an incubator at 25 �C. The res-
onant frequency and resonant resistance were measured
using a QCA922, and the results are presented in Fig. 4.
After 10 min, the signal responses reached a steady state,
while the deviations of the resonant frequency and resonant
resistance were about 5 Hz and 1 X in the stable area,
respectively, thereby confirming that an AT-cut quartz
crystal fixed with cover glasses can be used as a simulta-
neous sensing chip under atmospheric conditions.
Effect of matching oil on QCM response
The stability of the QCM in the simultaneous sensing
system was examined after injecting an index matching oil,
such as immersion oil (nd = 1.516, Olympus Co.) or sili-
con oil (KF-56, Shinetsu Chemical Co.), between the
quartz crystal and the cover glass. Figure 5 shows the
results obtained when using the high viscous immersion oil
as the index matching oil, while Fig. 6 shows the results
when using the low viscous silicon oil.
When the high viscous immersion oil was injected, the
resonant frequency decreased to about 17 kHz when
compared with the oscillation frequency in air (compare
Figs. 4 and 5). After 30 min, the response of the resonant
frequency reached a stable area with a frequency deviation
of about 300 Hz. Meanwhile, the resonant resistance
Fig. 2 Schematic illustration of
SPR principle (a) and proposed
measuring cell (b). The
proposed measuring cell is
fabricated to fulfill the
requirement of both the SPR
and the QCM operation
Fig. 3 SPR responses of standard SPR chip (A) and proposed SPR
chip (B) under atmospheric conditions Fig. 4 Typical resonant frequency response (A) and resonant resis-
tance response (B) of QCM in simultaneous sensing system under
atmospheric conditions
Fig. 5 Typical resonant frequency response (A) and resonant resis-
tance response (B) of QCM in simultaneous sensing system after
injecting immersion oil (nd = 1.516) between quartz crystal and
cover glass
42 Bioprocess Biosyst Eng (2010) 33:39–45
123
increased about 10-fold when compared with the resonant
resistance value in air (compare Figs. 4 and 5). In the stable
area, the deviation was about 10 X, as shown in Fig. 5.
When the low viscous silicon oil was injected, the res-
onant frequency decreased to 6.8 kHz when compared with
the oscillation frequency in air, as shown in Fig. 6. After
30 min, it reached a stable area with a frequency deviation
of about 10 Hz. The resonant resistance increased about
4-fold when compared with the resonant resistance value in
air, while the deviation was about 1 X in the stable area,
which was similar to that in air.
Therefore, the results revealed that a low viscous
matching oil was more effective for the proposed QCM/
SPR simultaneous sensing system.
Effect of proposed sensing cell on resonant response
of QCM
The influence of the proposed QCM/SPR cell on the
resonant oscillation properties of the QCM was measured.
In the designed cell, a screw and O-ring were used to
enable just one side of the quartz crystal electrodes to be
exposed to the liquid samples. The influence was then
measured in an incubator at 25 �C after injecting 5 ll of
silicon oil into the cell as the index matching oil. The
measured results are presented in Fig. 7, and show that the
deviations of the resonant frequency and resonant resis-
tance were about 20 Hz and 2 X after reaching a stable
area, respectively. These deviations were about 2-fold
higher than those in the incubator without the measuring
cell, thereby confirming the effectiveness of the proposed
sensing cell.
Water dropping experiments for obtaining simultaneous
QCM/SPR signals
Finally, the efficacy of using an At-quartz crystal for the
proposed SPR/QCM simultaneous sensing system was
examined by measuring the SPR, resonant frequency, and
resonant resistance using a Handy-SPR and QCA922
simultaneously when dropping 200 ll of Milli-Q water on
the gold electrode. The results are presented in Fig. 8.
The reflected light intensity was observed to change,
where the sharpness of the resonant angle was about 85%
of that of the cover glass sputtered with a 50-nm gold layer,
as shown in Figs. 3 and 8. The choice of the electrode
material is considered to be important to provide a sharp
SPR resonance peak. In Fig. 8, the QCM was sputtered
with 250 nm of ITO on one side of the quartz plate over a
5-nm thick underlay of titanium, and a 50-nm gold film on
the other side. This probably induced the reflection of light
Fig. 6 Typical resonant frequency response (A) and resonant resis-
tance response (B) of QCM in simultaneous sensing system after
injecting silicon oil between quartz crystal and cover glass
Fig. 7 Typical resonant frequency response (A) and resonant resis-
tance response (B) of QCM in simultaneous sensing system after
fixing proposed measuring cell
Fig. 8 Simultaneous sensing responses of a SPR and b QCM when
dropping water. In panel b, (A) denotes resonant frequency response,
and (B) denotes resonant resistance response
Bioprocess Biosyst Eng (2010) 33:39–45 43
123
not only from the gold layer but also from the ITO or
titanium layer, thereby weakening the intensity of the light
that reached the gold layer and resulting in broad SPR
peaks. Thus, if the thickness of the sputtered metal layer
can be controlled more precisely to be thinner, the sharp-
ness of SPR peak could be improved for use as a sensing
kit.
The dropping of the water produced a decrease in the
resonant frequency and increase in the resonant resistance
of about 2 kHz and 100 X (compare Figs. 7 and 8),
respectively. These changes are within the normal range
that occurs when a QCM comes into contact with water
[13]. However, the deviations in the resonant frequency
and resonant resistance were only about 50 Hz and 2 Xafter reaching a stable area (time after 1 h), respectively,
which are not enough to detect a small mass change and
only effective for detecting a large mass change, such as
living cells under certain conditions [6, 7]. The increase in
the response deviation of the QCM was mainly caused by
the viscosity of the index matching oil. Thus, the problem
of signal noise can be solved if a lower viscous index
matching oil is used in the future. Plus, further develop-
ment is required applying a high frequency [14, 15] or
overtone [16–18] modes of the QCM to enhance the sen-
sitivity. Various applications, which are normally treated as
a possible area of QCM application, may be possible in
biosensor with these developments [13]. Finally, the data is
reproducible within the standard deviation range of 5% in
our 5 experiments continuously with a similar response
shape.
Conclusion
This study fabricated and investigated a QCM for a
simultaneous SPR and QCM sensing system. In a water
dropping experiment, the deviations in the resonant fre-
quency and resonant resistance were about 50 Hz and 2 X,
respectively, after reaching a stable area, which are not
enough to detect a small mass change. Meanwhile, the
sharpness of the SPR angle was shown to depend on the
kind of electrode material used and its thickness, and the
response stability of the QCM to depend on the viscosity of
the index matching oil. As such, a more sensitive mea-
suring system can be designed by selecting an adequate
electrode material combination and lower viscous match-
ing oil. Therefore, the results confirmed the possibility of
developing a simultaneous SPR/QCM sensing platform.
Acknowledgment This research was supported by Basic Science
Research Program through the National Research Foundation of
Korea (NRF) funded by the Ministry of Education, Science and
Technology (KOSEF, 2009-0064245).
References
1. Kim JM, Chang SM, Muramatsu H (1999) Monitoring changes in
the viscoelastic properties of thin polymer films by the quartz
crystal resonator. Polymer 40:3291–3299
2. Kim JM, Chang SM, Muramatsu H (1999) Scanning localized
viscoelastic image using a quartz crystal resonator combined with
an atomic force microscopy. Appl Phys Lett 74:466–468
3. Kim JM, Chang SM, Muramatsu H (1999) In situ optoelectro-
chemical approach for the dynamic property study of polypyrrole
thin film by quartz crystal combined with UV–visible advanced
design. J Electrochem Soc 146:4544–4550
4. Song SH, Han DS, Lee HJ, Cho HS, Chang SM, Kim JM,
Muramatsu H (2001) Application of quartz crystal for the
supercapacity studies of Ppy/anion composites. Synthetic Met
117:137–139
5. Kim JM, Park JY, Song SH, Lee BJ, Muramatsu H, Chang SM
(2001) Analysis of Im-SH self-assembled monolayer formation
and its interaction with Fe2? and Zn2? using quartz chemical
analyzer. Sens Actuators B Chem 76:74–79
6. Kang HW, Ida K, Yamamoto Y, Muramatsu H (2008) Monitoring
of morphology and physical properties of cultured cells using a
micro camera and a quartz crystal with transparent indium tin
oxide electrodes after injections of glutaraldehyde and trypsin.
Anal Chim Acta 624:154–161
7. Kang HW, Muramatsu H (2009) Monitoring of cultured cell
activity by the quartz crystal and the micro CCD camera under
chemical stressors. Biosens Bioelectron 24:1318–1323
8. Lee BS, Chi YS, Lee KB, Kim YG, Choi IS (2007) Functional-
ization of poly(oligo(ethylene glycol) methacrylate) films on gold
and Si/SiO2 for immobilization of proteins and cells: SPR and
QCM studies. Biomacromolecules 8:3922–3929
9. Martin Y, Boutin D, Vermette P (2007) Study of the effect of
process parameters for n-heptylamine plasma polymerization on
final layer properties. Thin Solid Films 515:6844–6852
10. Zhou C, Friedt JM, Angelova A, Choi KH, Laureyn W, Frederix
F, Francis LA, Campitelli A, Engelborghs Y, Borghs G (2004)
Human immunoglobulin adsorption investigated by means of
quartz crystal microbalance dissipation, atomic force microscopy,
surface acoustic wave, and surface plasmon resonance tech-
niques. Langmuir 20:5870–5878
11. Malmstrom J, Agheli H, Kingshott P, Sutherland DS (2007)
Viscoelastic modeling of highly hydrated laminin layers at
homogeneous and nanostructured surfaces: quantification of
protein layer properties using QCM-D and SPR. Langmuir
23:9760–9768
12. Hedin J, Lofroth JE, Nyden M (2007) Adsorption behavior and
cross-linking of EHEC and HM-EHEC at hydrophilic and
hydrophobic modified surfaces monitored by SPR and QCM-D.
Langmuir 23:6148–6155
13. Muramatsu H, Kim JM, Chang SM (2002) Quartz-crystal sensors
for biosensing and chemical analysis. Anal Bioanal Chem
372:314–321
14. Furusawa H, Kitamura Y, Hagiwara N, Tsurimoto T, Okahata Y
(2002) Binding kinetics of the toroid-shaped PCNA to DNA
strands on a 27 MHz quartz crystal microbalance. Chemphy-
schem 3:446–448
15. Okahata Y, Niikura K, Furusawa H, Matsuno H (2000) A highly
sensitive 27 MHz quartz-crystal microbalance as a device for
kinetic measurements of molecular recognition on DNA strands.
Anal Sci 16:1113–1119
16. Liu S, Sun JL, Sun HS, Tan XJ, Shi S, Guo JH, Zhao J (2003)
First overtone frequency stimulated quartz tuning fork used for
shear-force scanning near-field optical microscopy. Chinese Phys
Lett 20:1928–1931
44 Bioprocess Biosyst Eng (2010) 33:39–45
123
17. Hirama K, Shoji T, Tanaka Y (2000) Suppression of lower har-
monic responses on overtone AT-cut quartz crystal resonators.
Electron Comm Jpn 3(83):92–102
18. Yoshimoto M, Tokimura S, Shigenobu K, Kurosawa S, Naito M
(2004) Properties of the overtone mode of the quartz crystal
microbalance in a low-viscosity liquid. Analytica Chimica Acta
510:15–19
19. Green RJ, Frazier RA, Shakesheff KM, Davies MC, Roberts CJ,
Tendler SJB (2000) Surface plasmon resonance analysis of
dynamic biological interactions with biomaterials. Biomaterials
21:1823–1835
20. Laricchia-Robbio L, Revoltella RP (2004) Comparison between
the surface plasmon resonance (SPR) and the quartz crystal
microbalance (QCM) method in a structural analysis of human
endothelin-1. Biosens Bioelectron 19:1753–1758
Bioprocess Biosyst Eng (2010) 33:39–45 45
123