control assignment - calibration of insulated and uninsulated thermistors

Upload: daveroberts123

Post on 04-Apr-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    1/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 1

    Introduction

    One of the most common types of physical measurements is temperature. Temperature can be

    measured in many environment and channel counts. There are various types of temperaturemeasuring sensors and components. Depending on the temperature range, desired accuracy and

    expense; appropriate sensors must be selected (National Instrument, 2011).

    There are three types of Semiconductors classifications; elemental semiconductors, compound

    semiconductors and junction semiconductors. Thermistors are compound semiconductors that are

    used for temperature measurements and identify a special resistor that varies in resistance with

    temperature (McGee, 1988).

    Background Theory

    Thermistors

    Thermistors are resistors that contain high rate of change of resistance with temperature. Typically,

    they are inexpensive ceramic compounds that are highly sensitive to temperature change.

    Regardless of the high sensitivity, they are commonly used for temperature control within the

    temperature range from approximately -100 to 500C (McGee, 1988). Mostly, themistors are

    compound semi-conductors that contain complex conduction processes. Compound semiconductors

    are usually ceramic in which solid solutions of transition metal oxides are used (McGee, 1988).

    Thermistors are employed in many electronic devices to control the temperature response or to

    compensate for temperature changes in electronic circuits. In term of temperature measurement, tominimize Joule heating which cause the thermistor to have a higher temperature than its

    surroundings, the current flows through a thermistor should be small (McGee, 1988). Sufficient

    current to casue Joule heating is desirable in the case of temperature control, switching purposes

    and fluid flow measurements or any other properties that have an effect on heat transfer from the

    thermistor. The internal heating is negligible when a small voltage is applied to a thermistor,

    therefore, its temperature will practically be the same as its surrounding temperature (McGee,

    1988). Themistors contain negative temperature coefficients of resistance, therefore, the Joule

    heating causes the resistance to drop, increases the current and increases heat transfer from the

    thermistor.

    The relationship between the Voltage and current is nonlinear due to the resistor R (Figure 1.1) is

    not constant. When the voltage increased, the heat transfer increases which will compensate for the

    Joule heating and a point will be reached where the maximum voltage will appear for any particular

    souce of current (Figure 1.1 and 1.2). The current-voltage relationship becomes unstable when the

    resistance of the thermistor drops and impulsive increases in current take place in between point B

    and C (Figure 1.3). The electrical power Nprovided to the thermistor equals to the heat dissipated to

    the surroundings P along the V-I curves. The V-I curves is extremely useful in many thermistor

    circuit applications.

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    2/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 2

    Figure 1.1: The voltage drop V across thermistor depends on the current I, the source resistance R,

    and the temperature of T (McGee, 1988). [Figure Abstracted from: McGee, T. (1988) Principles andMethods of Temperature Measurement, New York: John Wiley & Sons, p.210]

    Figure 1.2: Voltage-current response of the thermistor showing the thermistor resistance and

    temperature at different current levels (McGee, 1988). [Figure Abstracted from: McGee, T. (1988)

    Principles and Methods of Temperature Measurement, New York: John Wiley & Sons, p.211]

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    3/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 3

    Figure 1.3: Stable (C) and unstable (B) points of a simple thermistor circuit (McGee, 1988). [Figureabstracted from: McGee, T. (1988) Principles and Methods of Temperature Measurement, New York:

    John Wiley & Sons, p.211]

    In temperature measurement applications, the current must be kept at minimal to minimize self-

    heating errors. This requires a small voltage applied or current limiting. The voltage and current

    measurement errors increase as their value is decreased. In many applications, a compromise is

    needed between minimizing self-heating errors and minimizing resistance measuring errors (McGee,

    1988).

    Most thermistors contains a negative coefficient relating resistance with temperature, however,Positive-Temperature-Coefficient (PTC) thermistors are available and used due to their greater

    sensitivity. PTC thermistors contain positive coefficient only over a limited temperature range,

    beyond this range, the coefficient is negative (Leigh, 1991). The typical curve for a negative and

    positive-temperature coefficient thermistor is shown in Figure 1.4 and 1.5.

    Figure 1.4: Typical resistance/temperature curve for a negative-temperature-coefficient thermistor

    (Leigh, 1991)[Figure abstracted from:Leigh, J. (1991) Temperature Measurement & Control, Herts:Peter Peregrinus Ltd., p.25].

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    4/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 4

    Figure 1.5: Typical resistance/temperature curve for a positive-temperature-coefficient thermistor

    (Leigh, 1991)[Figure abstracted from:Leigh, J. (1991) Temperature Measurement & Control, Herts:

    Peter Peregrinus Ltd., p.25].

    Advantages and disadvantages of Thermistors

    The advantages of thermistors include; High Sensitivity,Small physical sizeand resistance at ambient

    temperature may be specified in a range of 0.1-300 k. However, the disadvantages of thermistors

    are their limited temperature range and their modest upper usable temperature (Leigh, 1991).

    Static and Dynamic Characteristics Respond

    The term Static characteristics of an instrument refer to the steady state reading that it gives when it

    has settled down. These types of characteristics may be stated in terms of accuracy, linearity, etc.

    The expression of Dynamic Characteristics illustrates the behaviour of an instrument in the time

    between when the measured quantity changes and a steady reading are provided (Figure 1.6)

    (Bolton, 1991).

    Figure 1.6: Dynamic responses [Figure abstracted from: Bolton, W. (1991) Industrial Controls &

    Instrumentation, Essex: Longman Group UK Limited, p.9]

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    5/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 5

    Experimental Apparatus

    Laboratory 1: Characterise the static performance of two sensors, insulated thermistor and un-insulated thermistor.

    Equipment provided:

    Hot water

    Thermocouple acting as a thermometer

    Beaker

    Digital Voltage Meter

    Experiment procedure:

    1. Fill beaker with hot water and place one of the two thermocouple wire in the beaker and

    another outside the beaker at room temperature.

    2. Slowly poor in cold water into the beaker to reduce the temperature of the water, therefore,

    changing the net voltage values on the voltmeter.

    3. Take readings and obtain the temperature and voltage values. Note that the thermocouple

    acts as a thermometer which in turns, two readings are obtained Temperature and voltage

    change.

    4. Repeat the experiment several times until desire data are obtained.

    5. Repeat the above procedures for but with form insulated thermistor.

    Figure 2.1: Experiment 1 Set-Up

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    6/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 6

    Laboratory 2: Investigate the dynamic performance of the two temperature sensors, following a

    step change in temperature.

    Equipment Provided:

    Hot Water

    Beaker

    Computer with data acquisition system

    Experiment Procedure:

    1. Prepare the computer data acquisition system.

    2. Fill in the beaker with hot water.

    3. Start the computer measurement system; wait one or two second before placing the

    thermistor inside the beaker and wait until all samples has been taken.

    4. Set the samples and the rate and time for the themistor to measure at 1000 samples at the

    rate of 50 samples per second.

    5. Repeat the test until desire results are obtained.

    6. Repeat the procedures for Insulated thermistor experiment except that the test will run at

    720 samples at the rate of 1 sample per second.

    7. Obtain all results to be further process.

    Figure 2.2: Experiment 2 Set-Up

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    7/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 7

    Results and Discussion

    Experiment 1: Static Calibration

    Graph 1.1: Static Calibration with no insulation results

    Graph 1.2: Static Calibration with Insulation results

    Graph 2 shows the sensor output (V) against the temperature (C) which indicates the values of

    voltage for the insulated and un-insulated thermistor. The information obtained may be used to

    conduct calculations which include static sensitivity (K) for un-insulated and insulated thermistor.

    R = 0.9351

    0

    0.5

    1

    1.5

    2

    2.5

    3

    3.5

    0 20 40 60 80 100

    Voltage

    Output(V)

    Temperature (C)

    Static Calibration with Insulation

    Set-1

    Set-2

    All

    Poly. (All)

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    8/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 8

    This can be achieve by the use of gradient best fit line which correspond for the static sensitivity k

    (V/C)

    Calculation:

    Ideal Straight line equation can be calculated as follow:

    minmax

    minmax)(II

    OOKslopeLineStraightIdeal (1.1)

    minmin)( IKOaInterceptLineStraightIdeal (1.2)

    )1(/041.0

    2590

    2.174.31 resultsSetCVK

    )2(/039.02590

    22.177.32 resultsSetCVK

    040.0Ave rageK

    23.0a

    63.2

    Ideal

    Ideal

    O

    aIKO

    Un-insulated Thermistor span:

    minmax IIspanInput (1.3)

    65

    2590

    spanInput

    spanInput

    minmax OOSpanPutOut (1.4)

    57.2

    2.177.3

    SpanPutOut

    SpanPutOut

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    9/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 9

    Un-insulated thermistor linearity:

    )( minminmax

    minmaxmin II

    II

    OOOO

    (1.5)

    )2565(259

    2.177.32.157.2

    58.137.1

    Insulated thermistor span:

    43.0

    33.175.1

    65

    2590

    SpanOutput

    SpanOutput

    SpanInput

    SpanInput

    Insulated thermistor Linearity:

    27.091.0

    )25.65(2590

    33.175.133.141.0

    From observation of the graph, there was error occurred during the experiment. The static

    sensitivity was calculated as 0.04 V/C. It can be said that the un-insulated thermistor was more

    suitable for the static calibration.

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    10/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 10

    Experiment 2: Dynamic Calibration

    Graph 2.1: Dynamic Calibration without insulation

    Graph 2.2: Dynamic Calibration with insulation

    0

    0.5

    1

    1.5

    2

    2.5

    3

    0 5 10 15 20 25

    Sens

    oroutput(V)

    Time (s)

    Dynamic Calibration with out Insulation

    Test 4

    Test 3

    Test 2

    1.41.45

    1.5

    1.55

    1.6

    1.65

    1.7

    0 200 400 600 800

    Voltage(V)

    Samples

    Dynamic Calibration with Insulation

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    11/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 11

    The graph of the results obtained for un-insulated indicated that the maximum value is 2.9 with the

    minimum value of 1.40.

    The Y value from the graph can be calculated as follow:

    VY

    Y

    22.2

    4.1632.04.17.2

    The time constant of 0.632 (value of step change) indicates in percentage form of 63.2%

    To find the X value, straight line was plotted at 90 degrees against the Y value at 2.2. The X value

    was found to be 5.5 seconds (Time Constant)

    5.5 Seconds (time constant), Sampling rate = 50 Samples/1 Second = 50 samples/sec

    The time constant ( ) is estimated at 5.5 seconds from the un-insulated graph which represent the

    dynamic performance for first order instrument.

    The results obtained from the computer software displayed that the maximum value of insulated

    test is 1.66 with the minimum value of 1.44.

    ).(115

    115

    6.1

    44.1632.044.166.1

    antconstTime

    X

    Vy

    y

    Mathlab software and Laplace Transform was used to calculate time constant in both cases. The

    time constant for un-insulated themistor was found to be less than the insulated one.

    Conclusion

    Overall, the results obtained for both Static and Dynamic experiments were fairly accurate.

    However, there were minor errors occurred which may have cause by human in term of inaccurate

    data observation, systematic or self-heating etc.

    It was observed for the static calibration that the un-insulated thermistor was the best option to be

    used for the experiment due to its least error. For the dynamic calibration, the time constants were

    compared and the un-insulated thermistor was found to be more appropriate for the dynamic

    calibration.

    Also, from the experimental observation, it can be stated that the instrument system might

    encounter some error due to chemical reaction, dirt and dust in the laboratory environment.

    Therefore, the instrument system should be examined at particular time intervals maintaining the

    international standards. The regular reports of the instrument examined should be preserved further

    accuracy

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    12/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 12

    Reference

    Bolton, W. (1991) Industrial Controls & Instrumentation, Essex: Longman Group UK Limited,

    p.8-10.

    McGee, T. (1988) Principles and Methods of Temperature Measurement, New York: John

    Wiley & Sons, p.206, 466-474, 504-511.

    Sachse, H. (1975) Semi-Conducting Temperature Sensors and Their Applications, New York:

    John Wiley & Sons, INC., p.159-165.

    Leigh, J. (1991) Temperature Measurement & Control, Herts: Peter Peregrinus Ltd., p.24-26.

    Childs, P. (2001) Practical Temperature Measurement, Oxford: Butterworth-Heinemann,

    p.39-41.

    Benedict, R. (1984) Fundamentals of temperature, Pressure, and Flow Measurements, 3rd

    ed. 1984: John Wiley & Sons, Inc., p.146.

    Figliola, R. and Beasley, D. (2006) Theory and Design for Mechanical Measurements, 4th ed.

    New York: John Wiley & Sons, Inc., p.12-15.

    National Instruments (2011) Temperature Measurement, [online] Available at:

    http://www.ni.com/temperature/ [Accessed: 10th Dec 2011].

  • 7/29/2019 Control Assignment - Calibration of Insulated and Uninsulated Thermistors

    13/13

    Daniel James Watkins [Static and Dynamic Calibration of an Insulated and Un-insulated Thermistor]

    BEng Mechanical Engineering Page 13

    Appendices

    Sensors Calibration

    Calibration consists of determining the indication or the output of a temperature sensor with

    respect to a standard sufficient number of known temperatures so that, with acceptable means of

    interpolation, the indication or output of the sensor will be known over the whole range of

    temperature used (Benedict, 1984, p146). The problem areas of calibration are immediately

    apparent; four main points must be available:

    1. Means for measuring the output of the temperature sensor

    2. A satisfactory temperature standard

    3. Controlled temperature environments

    4. A scheme for interpolating between calibrations points (Benedict, 1984).

    As Benedict (1984, p147) described, the environments for calibration distributed into two classes

    depending on the method of determining the temperature of the test sensor.

    Firstly, the sensor is exposed to a fixed-point environment so that under particular conditions,

    naturally exhibits a state of thermal equilibrium whose temperature is established numerically by

    the IPTS without recourse to a temperature standard. Secondly, the test sensor and temperature

    standard are simultaneously exposed to a controlled-temperature environment whose temperature

    is established by the standard instrument (Benedict, 1984).

    Static Calibration

    Static calibration is the most common type of calibration. For this procedure, known value is input to

    the system under calibration and the system output is recorded. The term Static implies that the

    values of the variables involved remain constant. In static calibrations, only the magnitudes of the

    known input and the measured output are important (Figliola et al. 2006).

    Dynamic Calibration

    Dynamic variables are time dependent in their magnitude and frequency content. Dynamic

    calibration determines the relationship between an input of known dynamic behaviour and the

    measurement system output. Normally, these calibrations involve applying sinusoidal signal or a

    step change as the known input signal (Figliola et al. 2006).