    Comparative specifications

    128 to 320 slice CT scanners

    CEP08028

    March 2009

  • 8/18/2019 CT Scanners comparison

    2/61

    Contents 2

    CEP08028: March 2009

    Introduction ............................................................................................. 3 

    Technical specifications .......................................................................... 5 

     Acknowledgements ............................................................................... 58 

    References............................................................................................ 59 

     Appendix 1: Supplier contact details ..................................................... 60 

     Author and report information................................................................ 61 

  • 8/18/2019 CT Scanners comparison

    3/61

    Introduction 3

    CEP08028: March 2009

    Purpose of this document

    This supplementary report to the CEP buyers’ guide to multislice CT scanners [1] provides comparative specifications of CT scanners that are capable of acquiringmore than 64 slices. In this category the number of slices is generally considered inconjunction with the length of coverage. These are specialist scanners primarilyaimed at cardiac or perfusion imaging, and length of coverage along the z-axis is ofparticular importance for these studies. A greater number of slices may mean longercoverage, or may be as a result of a higher slice acquisition density due to thedynamic focal spot.

    Comparative specifications are also available for other categories of scanner, asshown below.

    Report number Report title

    CEP08024 CT clinical applications software [6] 

    CEP08025 16 slice CT scanners [4] 

    CEP08026 32 to 40 slice CT scanners [2] 

    CEP08027 64 slice CT scanners [3] 

    CEP08028 128 to 320 slice CT scanners (this report)

    CEP08029 Wide bore CT scanners [5] 

    Scanners covered in this report

     At the time of writing, there are four manufacturers of medical CT scanners sellingsystems in the UK; three supply scanners offering more than 64 slices.

    Manufacturer Scanner model

    Philips Brilliance iCT

    Siemens Somatom Definition AS+

    Toshiba Aquilion ONE

    The data given in this report are representative of the scanners as of August 2008,and are liable to change. Features such as workstations and software packages maybe listed as cost options here, but such additional costs are generally negotiable.

    Specifications are presented side-by-side to facilitate comparison, and grouped into a

    series of topics relating to different aspects of the scanner, such as gantry, tube and

  • 8/18/2019 CT Scanners comparison

    4/61

    Introduction 4

    CEP08028: March 2009

    detectors etc. The data are presented in 28 tables, one for each aspect. Indexnumbers are provided for each question to aid referencing.

    Whilst every effort has been made to ensure the accuracy of this report, we have notverified the data provided. Specific responses should be confirmed with themanufacturers

  • 8/18/2019 CT Scanners comparison

    5/61

    Technical specifications

    CEP08028: March 2009

    List of topic areas

    Table 1. Gantry...................................................................................................... 6 

    Table 2. X-ray generator ....................................................................................... 7 

    Table 3. X-ray tube................................................................................................ 7 

    Table 4. Detection system..................................................................................... 9 

    Table 5. Couch.................................................................................................... 10 

    Table 6. Scanning capabilities............................................................................. 12 

    Table 7. Scanned projection radiography (SPR)................................................. 13 

    Table 8. Acquisition parameters.......................................................................... 13 

    Table 9. Automatic exposure control................................................................... 16 

    Table 10. Image reconstruction............................................................................. 17 

    Table 11. Data processing .................................................................................... 20 

    Table 12. System start-up and calibration............................................................. 21 

    Table 13. Manufacturer's performance data.......................................................... 23 

    Table 14. Operator console................................................................................... 25 

    Table 15. Main computers..................................................................................... 26 

    Table 16. Data management................................................................................. 27 

    Table 17. Workstation ........................................................................................... 28 

    Table 18. Workflow / connectivity .......................................................................... 32 

    Table 19. Ancillary equipment ............................................................................... 35 

    Table 20. Advanced reconstruction & display ....................................................... 36 

    Table 21. Clinical applications software ................................................................ 38 

    Table 22. Cardiac applications .............................................................................. 39 

    Table 23. Radiotherapy applications..................................................................... 41 

    Table 24. Advanced scanning facilities ................................................................. 42 

    Table 25. Installation requirements ....................................................................... 44 

    Table 26. Maintenance & support.......................................................................... 47 

    Table 27. Supplementary information.................................................................... 49 

    Table 28. Sustainability ......................................................................................... 51 

    Table 29. Notes to Siemens replies....................................................................... 55 

  • 8/18/2019 CT Scanners comparison

    6/61

    Technical specifications

    CEP08028: March 2009

    Technical specification comparison: 128 to 320 slice CT scanners

    Table 1. Gantry

    Philips Brilliance iCT Siemens SOMATOM Definition AS+ Toshib

    1.1 Scanner type 3rd generation 3rd generation 3rd ge

    1.2 Aperture [cm] 70 78 72

    1.3Focus-isocentredistance [mm]

    570 595 600

    1.4Focus-detector distance[mm]

    1040 1085 1072

    1.5Slip ring for powersupply

    Low voltage Low voltage Low vo

    1.6Slip ring for datatransmission

    Optical RF RF

    1.7 Tilt range [degrees] No ± 30 (0.5° increment) ± 22

    1.8 Tilt for helical scanning No No Not av

    1.9Positioning lights -planes

    Transaxial (internal and external),sagittal & coronal

    Transaxial, sagittal, coronal Transa

    1.10 Positioning lights - type Laser Laser Laser

    1.11Positioning lights -accuracy

    (at centre of gantry)[mm]

    ± 0.5 ± 1 ± 1

  • 8/18/2019 CT Scanners comparison

    7/61

    Technical specifications

    CEP08028: March 2009

    Table 2. X-ray generator

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    2.1 Type High frequency High frequency High fr

    2.2 Location Rotating frame Gantry In gant

    2.3Method of generatorcooling

     Air cooled Air or Water Air

    2.4 Power rating [kW] 120 100 70

    2.5 kV settings 80, 120, 140 80, 100, 120, 140 80, 100

    2.6mA range (and step

    size)

    10 - 1000 (in 1mA steps)20 - 800

    (1 mA steps)

    10 - 50

    2.7Maximum mA for eachkV

    80 kV: 650mA 120 kV: 1000mA 140 kV:750mA

    80 kV: 550 mA

    100 kV: 650 mA

    120 kV: 800 mA

    140 kV: 714 mA

    80 kV:

    100 kV

    120 kV

    135 kV

    Table 3. X-ray tube

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    3.1 Type and make iMRC Siemens Straton CXB-7

    3.2 Focal spot size(s) [mm],quoted to IEC 336/93standard

    0.6 x 0.7 1.1 x 1.2 0.7 x 0.70.9 x 1.1

    0.9 x 01.6 x 1

    3.3Moving (dynamic /flying) focal spot, xy-plane

    Yes Yes No

  • 8/18/2019 CT Scanners comparison

    8/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    3.4Moving (dynamic /flying) focal spot, z-axis

    Yes Yes No

    3.5Settings at which focalspot changes

    Small focal spot in HR/UHR mode;

    Not kW limitedNot power dependant

    80 kV:

    100 kV

    120 kV

    135 kV

    3.6

    Total filtration at max kV(inherent + beamshaping filter) on centralaxis

    [mm Al equivalent@120 kV]

    4.8 @ 140kV 6.8

    4.8 (L w

    3.7Half value layer of x-raybeam at 120 kV (ornearest) [mm Al]

    6.9 ± 0.5 @ 120 KV 8.7 Confid

    3.8 Anode heat capacity[MHU]

    30 MHU equivalent with real-timecooling

    30 equivalent 7.5

    3.9

    Maximum anode coolingrate

    [kHU / min]

    1608 7300 1386

    3.10 Method of tube cooling Oil-to-air Direct oil Liquid

    3.11 Expected tube life [amp-hours]

    1 year unlimited guarantee 3 years (see note 1) 300000

  • 8/18/2019 CT Scanners comparison

    9/61

    Technical specifications

    CEP08028: March 2009

    Table 4. Detection system

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    4.1 Detector type Solid state array Solid state array Solid s

    4.2 Detector material Gadolinium oxysulphide (GOS) Siemens UFC Gadoli

    4.3 Decay constant [μs] Info. not available 3 3

    4.4

    Maximum number ofsimultaneously acquireddata sets (number of'slices')

    128 128 320

    4.5Sampling frequency for

    0.5 s scan [Hz]9,600 (up to 4,800 samples/rev) 4608 2400

    4.6Sampling frequency forfastest scan [Hz]

    17,778 (up to 4,800 sample/rev) 4608 2572

    4.7Data transmission rate[Mbps]

    5280 8500 20000

    4.8

    Number of detectorelements per row(excluding referencedetectors)

    672 736 896

    4.9Number of referencedetector elements perrow

    1n.a. (reference detector is in tubecollimator)

    1 (tota

    4.10Number of detectorelements along z-axis

    128 64 320

    4.11Effective length ofdetector elements in z-axis (at isocentre) [mm]

    128 x 0.625mm 64 x 0.6 320 x 0

  • 8/18/2019 CT Scanners comparison

    10/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    4.12Total effective length ofdetector array atisocentre [mm]

    80 38.4 160

    4.13 Quarter detector shift No Yes Yes

    4.14Hardware features toimprove xy-planeresolution

    Dynamic Focal Spot (DFS)

     Attenuating comb (UHR)Flying Focal Spot (xy) No

    4.15Hardware features toimprove z-axisresolution

    Spiral Groove Bearing and Smart FocalSpot

    z-sharp / z-UHR comb No

    4.16

    Option for slice upgrade('Forklift' essentiallymeaning scannerreplacement)

    No Yes (Forklift) No

    Table 5. Couch

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    5.1 Couch top material Carbon fibre composite Carbon fibre Carbon

    5.2Couch top length andwidth [cm]

    243 x 41Standard: 240 x 45

    (MultiPurpose Table option: 240 x 53)246 x 4

    5.3Horizontal movementrange [cm]

    190Standard: 160

    (Multipurpose Table option: 200)239 (st

    5.4Couch controls - ongantry

    Yes (left/right, front/back)Left & Right (standard)

    Back (Option)

    Front,

    Rear, L

  • 8/18/2019 CT Scanners comparison

    11/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    5.5Couch controls - inroom

    Optional table side control

    Elevation control pedals on both sidesof the couch. Additional controls (i-Control panel) at the table if CT Flurooption installed

    Foot sw

    5.6Couch controls - onconsole

    Yes Yes Yes

    5.7 Accuracy/reproducibilityof table positioning [mm]

    ± 0.25 ± 0.25 ± 0.25

    5.8Horizontal range(without extensions)[cm]

    130Standard: 160

    (Multipurpose Table option: 200)203

    5.9Scannable horizontalrange with table topextension(s) [cm]

    175Standard: 160

    (Multipurpose Table option: 200)195

    5.10Vertical movementrange out of gantry [cm]

    57.8 - 102.8Standard: 48 - 92

    Multipurpose Table: 55 - 9233 - 99

    5.11Vertical movementrange in gantry [cm]

    85.0 - 102.8 72 - 92 78 - 99

    5.12Maximum clearance ingantry [cm]

    17.8 69 58

    5.13

    Minimum couch top

    height outside gantry[cm] 57.8

    Standard & 300 kg option: 48

    MPT option: 53 33

    5.14Maximum weightallowed on couch [kg]

    204 200 (standard) or 280 (option) 300

  • 8/18/2019 CT Scanners comparison

    12/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    5.15

    Maximum weight [kg] onstandard couch whichstill achieves statedperformancespecificationseg.accuracy (mm)

    204 200 (standard) or 280 (option) 230

    Table 6. Scanning capabilities

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    6.1Maximum continuous

    scan time [s] Info. not available 100 100

    6.2

    Maximum number ofrotations in one helicalrun at standardabdomen parameters

    Info. not available 72 250

    6.3

    Maximum helical scancoverage using a 1 mmimaged slice thicknessand a pitch of 1.5 [mm]

    Info. not available 2000 1950

    6.4Overscan required forsequential scanning[degrees].

    Info. not available No Max 35

    6.5Overbeaming requiredfor helical scanning[rotations]

    Info. not available Siemens Adaptive Dose Shield utilised2

    0 for vo

  • 8/18/2019 CT Scanners comparison

    13/61

    Technical specifications

    CEP08028: March 2009

    Table 7. Scanned projection radiography (SPR)

    Philips Brilliance iCT Siemens SOMATOM Definition AS+ Toshib

    7.1 Proprietary SPR name Surview Topogram Scano

    7.2Maximum SPR length[mm]

    1750 1574 (1974 option)1950 (S

    7.3SPR dimensions [mm xmm]

    500 x 1750 1574 (1974 option) Freely selectablewidth:

    length:

    7.4 Angular positions of X-ray tube available forSPR [degrees]

    90° (Lateral) 180º (AP) a.p., p.a., lateral0, 90, 1

    each 5

    7.5Maximum number ofSPR views available forscan planning

    2 2 2

    7.6 Real time image display Less than 5 s after scan Yes Yes

    7.7 Accuracy of sliceprescription from SPR[mm]

    ± 0.5 ± 0.5 ± 2

    7.8 Accuracy of distancemeasurements fromSPR, at isocentre [mm]

    ± 0.25 ± 0.5 ± 2

    Table 8. Acquis ition parameters

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    8a Key settings

    8.1 kV steps 80, 120, 140 80, 100, 120, 140 80, 100

  • 8/18/2019 CT Scanners comparison

    14/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    8.2 mA range and step size 10 - 1000 (1mA steps)20 - 800

    (1 mA steps)

    10 - 50

    8.3 Max. mA (for each kV)80 kV: 650mA 120 kV: 1000mA 140 kV:750mA

    80 kV: 550 mA

    100 kV: 650 mA

    120 kV: 800 mA

    140 kV: 714 mA

    80 kV:

    100 kV

    120 kV

    135 kV

    8.4 Scan fields of view [cm] 25, 50 5 - 50 18, 24,

    8b Sequential mode

    8.5 Rotation times forsequential scanning [s] 0.27 (option), 0.33, 0.375, 0.4, 0.5,0.75, 1.0, 1.5 0.3, 0.33, 0.5, 1 0.35, 01.0, 1.5

    8.6Partial rotation scantimes [s]

    0.18 0.22, 0.24 0.28

    8.7

    Nominal z-axis x-raybeam collimations,sequential scanning[mm]

    1.25, 12.5, 40, 60, 70, 80 2, 5, 10, 14.4, 19.2, 361, 2, 4,120, 12

    8.8

    Maximum z-axis x-raybeam collimation withsub-millimetre slices[mm]

    80 36 160

    8.9

    Thinnest acquisitionwidth available at full z-axis x-ray beamcollimation [mm]

    0.625 0.6 0.5

  • 8/18/2019 CT Scanners comparison

    15/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    8.10

    Sequential acquisitionoptions

    (beam collimation:available dataacquisition widths)

    80: 0.625, 2.5 70: 0.625 60: 0.625 40:0.625 12.5: 0.625 1.25: 0.625

    38.4: 0.6

    19.2: 1.2

    10: 10

    5: 5

    4.8: 0.6

    2: 1

    320: 0

    280: 0

    256: 0

    240: 0

    160: 0

    120: 0

    80: 0.5

    32: 8

    24: 6

    16: 4

    12: 3

    8: 2

    4: 1

    2: 0.5

    1: 1

    8c Helical mode

    8.11Rotation times forhelical scanning [s]

    0.27 (option), 0.33, 0.375, 0.4, 0.5,0.75, 1.0, 1.5

    0.3, 0.33, 0.5, 10.35, 01.5

    8.12Nominal z-axis x-raybeam collimations,

    helical scanning [mm]

    1.25, 12.5, 40, 60, 70, 80 38.4, 12, 6, 4.8 32, 16,

    8.13

    Maximum z-axis x-raybeam collimation withsub-millimetre slices[mm]

    80 38.4 32

  • 8/18/2019 CT Scanners comparison

    16/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    8.14

    Thinnest acquisitionwidth available at full z-axis x-ray beamcollimation [mm]

    0.625 0.6 0.5

    8.15

    Helical acquisitionoptions

    (beam collimation:available dataacquisition widths)

    80: 0.625, 2.5 70: 0.625 60: 0.625 40:0.625 12.5: 0.625 1.25: 0.625

    38.4: 0.6, 1.2

    12: 0.6

    6: 0.6

    4.8: 0.3, 0.6

    32: 0.5

    16: 0.5

    8.16Pitches available for

    routine scanning0.04 - 1.5, User selectable 0.35 - 1.5 freely selectable

    64-slic

    32-slic

    8.17Recommended pitchesfor optimal image quality

    0.04 - 1.5, Optimal valuesrecommended based on protocol

    Not Applicable due to SureViewalgorithms

    0.641

    0.828

    1.484

    Table 9. Automatic exposure control

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    9.1

     Automatic mA control

    (AEC / mA modulation)

    software

    DoseRight CARE Dose 4D SURE

    9.2- mA adjustment forpatient size

    Yes, ACS Yes Yes

    9.3- mA adjustment alongthe z-axis

    Yes, Z-DOM Yes Yes

  • 8/18/2019 CT Scanners comparison

    17/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    9.4- mA modulation duringrotation

    Yes, D-DOM Yes Yes

    9.5Can z-axis androtational modulation beemployed concurrently?

    No Yes Yes

    9.6Method for prescribing AEC image quality

    Reference imageNoise index

    (Reference mAs set by user)Standa

    9.7Method for systemcontrol of mA

    Single or dual Surview SPR & online control Using s

    9.8 Operator review of AEC Yes display of min, mean, max mAs No Yes

    9.9 AEC manual adjustment Yes Yes - max / min mA Yes

    Table 10. Image reconstruction

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    10a Parameters

    10.1Minimum reconstructionfield of view [cm]

    Std & high res: 5 UHR: 2.5 5 5

    10.2

    Maximumreconstruction field ofview at full image quality[cm]

    Std & high res: 50 UHR: 25 50 50

    10.3Extended scan field ofview [cm]

    No Option 78 Not av

    10.4 Reconstruction matrices 512², 768², 1024² 512 x 512 512 x 5

  • 8/18/2019 CT Scanners comparison

    18/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    10.5

    Sequential image sliceoptions

    (beam collimation : slicewidths) [mm]

    80: 0.625, 2.5

    70: 0.625

    60: 0.625

    40: 0.625

    12.5: 0.625

    1.25: 0.625

    36: 2.4, 3, 3.6, 6, 7.2, 9, 12

    19.2:

    10:

    5:

    4.8:

    2:

    320: 0

    280: 0

    256: 0

    240: 0

    160: 0

    120: 0

    80: 0.5

    32: 8

    24: 6

    16: 4

    12: 3

    8: 2

    4: 1

    2: 0.5

    1: 1

    10.6

    Helical image sliceoptions (beamcollimation: slice widths)[mm]

    1.25 to 80 collimation:

    0.55 to 7.5 slice width

    38.4(128x0.6*): 0.6, 0.75, 1, 1.5, 2, 3,4.5, 6, 7, 8, 10

    38.4 (32x1.2):1.5, 2, 3, 4.5, 6, 7, 8, 10

    12:

    6:

    4.8:

    * double-sampled

    Equal o(max 1

    10.7Minimum reconstructioninterval of images,helical scanning [mm]

    0.1 0.1 0.1

  • 8/18/2019 CT Scanners comparison

    19/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    10 b Functionality

    10.8Number of prospectivereconstructions that canbe set up in protocol

    10 8 780

    10.9

     Any delay in eitherscanning orreconstruction whenperformed concurrently

    No No No

    10.10Simultaneous scanningand routine analysis

    Yes Yes with 2nd monitor option Yes

    10.11Simultaneous scanningand archiving and/orhard copying

    Yes Yes Yes

    10.12Simultaneous scanningand transfer to secondconsole/workstation

    Yes Yes Yes

    10 c Performance

    10.13

    Standard reconstructionrate for 512 x 512images - head [imagesper second (ips)]

     Approx. 20 40 32

    10.14Standard reconstructionrate for 512 x 512images - body[ips]

     Approx. 18 40 32

    10.15Reconstruction upgradeoptions

    RapidView 20 is standard Not applicable No

  • 8/18/2019 CT Scanners comparison

    20/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    10.16Fastest reconstructionrate [ips]

    Up to 20 40 32

    Table 11. Data processing

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    11.1Reconstructionalgorithm - sequentialscanning

    Filtered Back Projection Filtered Back Projection ConeX

    11.2

    Reconstruction

    approach - sequentialscanning

    2D 2D/3D Volum

    11.3Cone beamreconstruction -sequential scanning

    Yes, with Step & Shoot Cardiac option For 3D recon modes Yes

    11.4Reconstructionalgorithm - helicalscanning

    Cone Beam Reconstruction Alogirthm(COBRA)

    WFBP (Weighted Filtered BackProjection)

    T-COT

    (t

    11.5Reconstructionapproach - helicalscanning

    3D Back Projection 3D WFPB Volum

    11.6 Cone beamreconstruction - helicalscanning

    Yes Yes, always Yes

    11.7 Low signal correction Yes, Adaptive Filtering Yes Yes

    11.8 Motion correctionYes, with overscan (in sequentialmode)

    No No

  • 8/18/2019 CT Scanners comparison

    21/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    11.9Beam hardeningalgorithms

    Yes, IBC (iterative bone correction)

    Standard beam hardening algorithm(water correction) in all modes,

    PFO (iterative bone correction) inselected modes

    Yes

    11.10 Beam hardening trigger Yes, Protocol dependent PFO is on in all soft-tissue head modes Kernel

    11.11Noise reductiontechniques - raw databased

    Yes, Adaptive Filtering 3D adaptive filter Yes

    11.12Noise reductiontechniques - image

    based

    Yes, Image Enhancement 3D advanced smoothing algorithm Yes

    11.13Other artefact reductionmethods

    TBA Z-sharp technology No

    Table 12. System start-up and calib ration

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    12.1Total time from stand-bymode to scanning [mins]

  • 8/18/2019 CT Scanners comparison

    22/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    12.5Tube warm-up time[mins]

    3 0 5

    12.6 Warm-up frequencyDaily (or when idle for prolongedperiods)

    Daily After 3

    No wa

    12.7Detector calibration atwarm-up

    Yes, air cals Part of warm up Autom

    12.8Time to performdetector calibrations[mins]

    25 6 2

    12.9Recommendedfrequency for detectorcalibration

    Weekly Daily Daily

    12.10 Automated QC tests Std QA Protocols Yes No

    12.11 Self-check frequency Daily Daily Not ap

    12.12 Any additionalcalibration by theradiographer

    No Not required Not req

    12.13Time to perform theseadditional calibrations[mins]

    Not available Not required Not ap

  • 8/18/2019 CT Scanners comparison

    23/61

    Technical specifications

    CEP08028: March 2009

    Table 13. Manufacturer's performance data

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    13.1In plane spatialresolution for sharpestclinical algorithm [lp/cm]

    MTF0: 24 MTF2: 23 MTF10: 16 MTF50:8 (0.75s, 2.5mm, UHR, 25cm FOV E)

    MTF2: 24.0

    (120 kV, 160 mA, 0.5 s, 8x0.6 mm,UHR-Comb, U95u)

    MTF2:

    (120 kV0.5s, 2focus)

    13.2

    Longitudinal (z-axis)spatial resolution forsharpest clinicalalgorithm [lp/cm]

    Info. not available MTF2: 21 MTF2:

    13.3

    Contrast resolution:

    smallest rod sizediscernable (in plane)[mm]

    2mm @ 0.3 % contrast

    4mm @ 0.3% contrast

    5mm @ 0.3% contras

    5 mm @ 0.3% at 16 mGy

    (120 kV, 180 mAs, 1.0 s, 10mm slice,Sequence scan, B31s)

    Info no

    13.4Method for contrastresolution determination

    2mm: 120 kVp, 350 mAs, 10 mm,40mGy

    4mm: 120 kVp, 250 mAs, 10 mm, 27mGy

    5mm: 120 kVp, 200 mAs, 10 mm,19mGy

    (Determined Visually)

    Observer method Observ

    13.5CT number accuracy(water) [+/-HU]

    ±4 2 5

    13.6CT number uniformity[HU]

    ±3 ± 2 (20 cm Water phantom) ± 5

    13.7CT number accuracy(air) [+/-HU]

    +/- 4HU Centre - 10HU Periphery 10 Info no

    13.8CT number accuracy(bone) [+/-HU]

     Acrylic (120 kV) = 140±15HU of non-water materials depends onkV

    Info no

  • 8/18/2019 CT Scanners comparison

    24/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    13.9Scanner settings for stdhead scan, used forquoting CTDI100 values

    120 kV, 200 mAs, 64x1.25mm, 6mmST, 250 cm FOV, EB Filter

    120 kV, 350 mA, 1.0 s, 4.8 mm slice(coll. 12x1.2 mm), Standard ShapedBeam Filter

    CTDI1

    13.10CTDI100 at isocentre inair [mGy / 100mAs]

    Info. not available 22.3 Info no

    13.11CTDI100 at centre ofhead phantom [mGy /100mAs]

    10.36 16.4 CTDI1

    13.12CTDI100 at periphery ofhead phantom [mGy /100mAs]

    11.24 16.7 CTDI1

    13.13Scanner settings for stdbody scan, used forquoting CTDI100 values

    120 kV, 250 mAs, 32x2.5mm, 5mm ST,350 cm FOV, B Filter

    120 kV, 420 mA, 0.5 s, 4.8 mm slice(coll. 32x1.2 mm), Standard ShapedBeam Filter

    CTDI1

    13.14CTDI100 at isocentre inair [mGy / 100mAs]

    Info. not available 19.5 CTDI1

    13.15CTDI100 at centre ofbody phantom[mGy/100mAs]

    3.36 4.4 CTDI1

    13.16CTDI100 periphery ofbody phantom[mGy/100mAs]

    6.76 7.9 CTDI1

    13.17Dose profile FWHM,measured at isocentre[mm]

    ±10% for all collimation in air

    Std Body Collimation:

    Nominal FWHM: 43.0 mm

    Nominal Base Width (FilmMeasurement of Acceptance Test):43.5 mm

    Confid

  • 8/18/2019 CT Scanners comparison

    25/61

    Technical specifications

    CEP08028: March 2009

    Table 14. Operator console

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    14.1Dimensions of desktop[mm x mm]

    330 x 900 (W x L) Site dependant 450 x 7

    14.2 Input devices KB, Mouse, Bar code reader Keyboard / Mouse Keybo

    14.3 Scan controls Separate control device Keyboard / Mouse / Scan controller Keybo

    14.4Recorded instructionsystem options

    Yes, 9 pre-recorded languages, alsouser programmable

    Customisable recording Custom

    14.5Visual breath-holdindicator

    Yes, scan timing bar on console No Yes

    14.6Number of monitors atconsole

    21 (standard)

    2 (option)2

    14.7

    Function of eachmonitor (state if anyfunction is restricted toonly one monitor)

    Scanning on left monitor

     Advanced Processing on right monitor

     Acquisiiton control, patient data base,image viewing, clinical softwareevaluation packages, filming

    Scan a

    (Recon

    14.8 Display type (LCD/CRT) LCD LCD LCD

    14.9Diagonal dimension ofimage screen [cm]

    48 48 48.3

    14.10Image area matrixdimensions

    1280 x 1024 1024 x 1024 1024x1

    14.11Usual range of CTNumber displayed [HU]

    Dependent on application and anatomy –1,024 to +3,071 -1024 t

    14.12Extended range of CTnumber display [HU]

    -1024 to +3072 –10,240 to +30,710 -32768

  • 8/18/2019 CT Scanners comparison

    26/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    14.13 Accuracy of distancemeasurements in xy-plane [mm]

    ± 0.25 2 x pixel size 0.01 to

    14.14CTDI displayed onconsole

    Yes CTDIvolCTDIw

    (User d

    14.15Dose Length Product(DLP) displayed onconsole

    YesYes predicted before scan / actual afterscan

    Planne

     Actual

     After S

    14.16Geometric Efficiencydisplayed on console

    Yes Yes - if < 70% Yes - if

    Table 15. Main computers

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    15 a Console Computer

    15.1 Location Console tower In console / under desk Under

    15.2 Make and model Precision 690 Siemens Fujitsu TOSH

    15.3 Operating system Windows XP Windows XP Window

    15.4Number and type of

    CPUOne Dual Core 2 x Intel Xeon Confid

    15.5 CPU speed [GHz] 2 x 2.66 2 x 3.0 GHz Confid

    15.6Standard amount ofcomputer RAM [GB]

    4 2 Confid

    15.7Maximum amount ofcomputer RAM [GB]

    4 Not applicable Confid

  • 8/18/2019 CT Scanners comparison

    27/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    15 b Reconstructor

    15.8 Location Stand-alone server Location dependant In scan

    15.9 Make and model Proprietary Siemens TOSH

    15.10 Operating system Windows Windows XP Confid

    15.11Number and type ofCPU

    6 Xeon Quad Core 4 x Dual Xeon Confid

    15.12 CPU speed [GHz} 2.66 8 x 3.6Ghz Confid

    15.13Standard amount of

    computer RAM [GB]

    4 4 x 4 Confid

    15.14Maximum amount ofcomputer RAM [GB]

    8 Not applicable Confid

    Table 16. Data management

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    16 a Storage (at scanner)

    16.1Standard total hard disccapacity [GB]

    292 1679 3800

    16.2 Max total hard disccapacity [GB] 292 1679 3800

    16.3Hard disc capacity forimage storage [GB]

    292 292 500

  • 8/18/2019 CT Scanners comparison

    28/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    16.4Hard disc capacity forimage storage [#images]

    500000 520000 800000

    16.5Capacity for raw datastorage [GB]

    882 1168 3300 (

    16.6 Local compression User selectable No No

    16.7Image compressionalgorithms and rates

    Info. not available Not applicable Not ap

    16 b Image archive

    16.8 External media options CDR, DVD-RAM DVD, CD, USB-HDD, Memory-StickCD-R

    DVD

    16.9 Disc burning facility Yes, standard Dicom part 10 standard Standa

    16.10 Data archive options Info. not availableImages, Recons, Raw Data, Doserecords, Dicom Structured Reports (e.g.MS Word, Adobe pdf)

    Raw D

    Images

    16.11 Protocol archive options Info. not availableScan Protocols, Utilisation Data,DicomStructured reports

    Protoc

    Table 17. Workstation

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    17 a Standard workstation

    17.1Workstation(s) providedas standard

    No OptionalMC (St

    4D WS

  • 8/18/2019 CT Scanners comparison

    29/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    17.2Workstation softwareupgrade options

     All clinical applications Details on application

    Cerebr

    Calcium

    Cardia

    Dental

    SURE

    Lung C

    Fusion

    EP Ele

    Vesse

    17.3Workstation hardwareupgrade options Dual Monitor/Hard disc Upgrade PC / RAM 4D WS

    17.4Computer make andmodel

    Dell Precision 690 Siemens FujitsuTOSHI

    HP (4D

    17.5 Operating system Windows XP Windows XP Window

    17.6Number and type ofCPU

    Two Dual Core Intel Xeon Confid

    17.7 CPU speed [GHz] 2 x 2.66 GHz 2 x 3.0 GHz Confid

    17.8 Amount of computerRAM supplied asstandard [GB]

    4 6 Confid

    17.9Maximum amount ofcomputer RAM [GB]

    4 8 Confid

    17.10Disc storage capacitysupplied as standard[GB]

    146 2203800 (M

    2396 (4

  • 8/18/2019 CT Scanners comparison

    30/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    17.11Maximum total hard discstorage capacity [GB]

    438 (optional) 2203800 (M

    2396 (4

    17.12Speed of connections tonetworks [Mbits/s]

    10 / 100 / 1000 10001000 (M

    1000 (3

    17.13 External media options DVD-RAM, CD DVDCD-R

    DVD

    17.14 Disc burning facility Yes DICOM "part 10" standard Standa

    17.15 Data archive optionsSurview, Axial images, MPRs, VR,SSD, MIP, clinical applications, save

    states

    Raw / ImageRecon

    (Raw D

    17.16 Other archive options Info. not available Not applicable Report

    17 b Image disp lay

    17.17 Number of monitors1 standard

    2 (option)1 (2 option)

    1 (MC)

    1 (4D W

    17.18 Display type (LCD/CRT) LCD LCD LCD

    17.19 Screen size [cm] 48 48.3 cm (19 inches)48.2 (M

    76.2 (4

    17.20 Colour or greyscale Colour Colour Colour

    17.21 Monitor orientation Landscape Landscape Landsc

    17.22 Resolution [megapixels] 1.27 1.31.3, 12

    4.1, 25

  • 8/18/2019 CT Scanners comparison

    31/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    17.23Luminance [cd/m2] \

    Contrast [nnn:1]

    Luminance:250

    Contrast: 350:1

    Clinical Setting:

    Luminence:137

    Contrast: 450:1

    Lumina

    Contra

    17.24Calibrated displaycontrast

    No CIE / DICOM160 (M

    150 (4

    17.25Image area matrixdimensions

    1280 x 1024 1024 x 10241024 x

    2160 x

    17.26Standard CT numberrange [HU]

    -1024 to 3072 -1024 to +3071 -1024 t

    17.27Extended CT numberrange [HU]

    No -10240 to +30710 standard -32768

    17.28 Accuracy of distancemeasurements in xy-plane [mm]

    ±0.25 2 x pixel size 0.01 to

    17 c Multi-user facilitation

    17.29 Shared image database Yes, images reside on server Optional : Satellite console (CTWP) Not av

    17.30 Thin client access Yes, via thin client software Expert i Not av

    17.31Remote access toworkstation

    Yes, via thin client software Expert i : configurable access Yes (4

    17.32External access toworkstation

    Yes, via VPN connection Expert i : configurable access Not av

    17.33 Additional systemrequirements

    Minimum 512 MB Ram 1280 x 1024display GB speed network adaptorMinimum processor speed - 1GHz

    Expert i : Network Not ap

  • 8/18/2019 CT Scanners comparison

    32/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    17.34 Multi-use licensing

    Unlimited licenses - up to the capacityof the server

    10k, 30k, 60k slices simultaneously

    Per seat Not av

    Table 18. Workflow / connectivity

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    18 a Image transfer

    18.1 Scanner to workstationYes - preset in protocol, edit during

    planning or manual

     Automated as part of protocol or

    operator triggered

     Autom

    18.2Rate of image transferscanner to workstation[ips]

    Info. not available

    Background function duringreconstruction from scanner to Siemensworkstation - otherwise 25 ips with nocompression applied

    Maxim

    18.3 Scanner to PACSYes - preset in protocol, edit duringplanning or manual

     Automated as part of protocol oroperator triggered

     Autom

    18.4Rate of image transferscanner to PACS [ips]

    Info. not available

    Background function duringreconstruction from scanner to SiemensPACS - otherwise 25 ips with nocompression applied

    Maxim

    18.5 Workstation to PACS Yes - manual Operator triggered Manua

    18.6Rate of image transferworkstation to PACS[ips]

    Info. not available 25 ips uncompressed Maxim

    18.7 PACS to workstation Yes - manual Operator triggered Manua

  • 8/18/2019 CT Scanners comparison

    33/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    18.8Rate of image transferPACS to workstation[ips]

    Info. not available

    25 ips if Siemens PACS andWorkstation are used. This rate mayvary considerably if other PACSsystems or workstations are used.

    Maxim

    18 b Interoperabi lit y

    18.9 Scanner with RIS StandardStandard assuming Dicom WORKLISTon RIS

    Option

    18.10 Scanner with PACS Standard Standard Option

    18.11 Workstation with RIS Not available No Standa

    18.12 Workstation with PACS Standard Standard Standa

    18.13 Reporting tools Yes, as DICOM encapsulated PDF No Yes (4

    18.14Structured reportsavailable

    Not available Clinical Software dependant Not av

    18 c IHE

    18.15Which IHE profilessupported

    Scheduled Workflow

    Portable Data for Imaging

     Access to Radiology Information

    SCANNER: ScheduledWorkflow,Patient InformationReconciliation, Access to RadiologyInformation, and ConsistentPresentation of ImagesWORKSTATION:Image Display,

    Eveidence Creator, Print Composer,Report Creator, Report Reader, imageDisplay, Portable Media Creator,Report Reader

    Sched

    Patient

    Portab

    Eviden

    ConsisPresen

    Charge

  • 8/18/2019 CT Scanners comparison

    34/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    18.16 Actors implemented -scanner

    1) Acquisition Modality + options:

    - Assisted Acquisition Protocol Setting

    - PPS Exception Management

    - Patient Based Worklist Query

    - Broad Worklist Query

    2) Image Display

    3) Portable Media Creator and Reader

     Acquisition Modality ( Patient BasedWorklist Query, Broad Worklist Query,Modality Group Case), ImageDisplay,Evidence Creator,AcquisitionModality, Image Display,PrintComposer

     Acquis

    18.17 Actors implemented -workstation

    1) Image Display

    2) Portable Media Creator and Reader

    Image Display, Evidence Creator, PrintComposer, Report Creator, ReportReader, image Display, Portable Media

    Creator, Report Reader

    Image

    18 d DICOM objects

    18.18CT image object (1993)supported

    Yes Yes Yes

    18.19CT 'enhanced'multiframe object (2005)supported

    No Standard Standa

    18.20CT object supported asstandard

    Image Object (1993) CT image object (1993) 2005

    18 e DICOM services on scanner / main conso le

    18.21 Storage SCU & SCP Yes SCU, S

    18.22 Print SCU Yes SCU

    18.23 Query / retrieve SCU Yes SCU, S

    18.24 Modality worklist SCU Yes SCU

  • 8/18/2019 CT Scanners comparison

    35/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    18.25Performed procedurestep

    SCU Yes SCP

    18.26 Storage commitment SCU Yes SCU

    18 f DICOM services on workstation

    18.27 Storage SCU & SCP Yes SCU/S

    18.28 Print SCU Yes SCU

    18.29 Query / retrieve SCU Yes SCU

    18.30Modality worklist

    management

    No No Not av

    18.31Performed procedurestep

    NO MPPS service as SCU Not av

    18.32 Storage commitment SCU Yes SCU

    Table 19. Ancillary equip ment

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    19.1 Contrast injector Option Single or dual headed (option) Local P

    19.2 Insufflators Option CO2 (option) Local P

    19.3 Head RestElevated and Reclined (standard); Flat(option)

    Straight head rest (standard), coronalheadrest (option)

    StraighTilt (op

    Tilt & P

    Corona

    Ratche

    Infant (

  • 8/18/2019 CT Scanners comparison

    36/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    19.4Stereotactic headrestraint

    Option Option Local P

    19.5 Restraining straps Velcro Velcro (standard) Velcro

    19.6 Slicker mattress cover Standard Standard Standa

    19.7 Bariatric couch option? Option Yes No

    19.8Bariatric couch weightlimit [kg]

    295 330 Not ap

    19.9CT fluoroscopy

    hardware

    Monitor, Exposure Control, foot switch Option

    Monito

    Expos

    Couch

    19.10Hard-copy imagingdevice

    No Option Local P

    19.11 Gantry mounted ECG Yes (with cardiac gating package) Option No

    Table 20. Advanced reconstruc tion & d isplay

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    20 a Operator conso le

    20.1

    Maximum and minimum

    intensity projections(MIP, MINIP)

    Yes Yes Yes

    20.2Multi-planarreconstruction (MPR)

    Yes Yes Yes

    20.3 Planes available in MPR Any Yes Yes

  • 8/18/2019 CT Scanners comparison

    37/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definition AS+ Toshib

    20.4Shaded surface display(SSD)

    Yes Yes Yes

    20.5Volume renderingsoftware

    Yes Yes Yes

    20.6Segmentation andsculpting

    Manual and automatic Manual (standard), Automated (option) Yes

    20 b Standard workstation

    20.7Maximum and minimumintensity projections(MIP, MINIP)

    Yes Yes Yes

    20.8Multi-planarreconstruction (MPR)

    Yes Yes Yes

    20.9 Planes available in MPR Any Yes Yes

    20.10Shaded surface display(SSD)

    Yes Yes Yes

    20.11Volume renderingsoftware

    Yes Yes Yes

    20.12Segmentation andsculpting

    Manual and automatic Manual (standard), Automated (option) Yes

  • 8/18/2019 CT Scanners comparison

    38/61

    Technical specifications

    CEP08028: March 2009

    Table 21. Clinical applications sof tware

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    (OC - OPERATOR CONSOLE, WS - WORKSTATION)

    21.1CT angiography: Basicpackage

    CT Viewer (OC & WS) 3D (OC & WS ) Not ava

    21.2CT angiography: Advanced package

     AVA (OC & WS) InSpace & InSpace AVA (OC & WS) Not ava

    21.3Virtual endoscopy:Basic package

    CT Viewer (OC & WS) Fly Through (OC, & WS)Fly-Thr

    Colon (

    21.4Virtual colonoscopy: Advanced package Virtual colonoscopy (OC & WS) Syngo Colonography CT (OC & WS) Colon C

    21.5Brain perfusionsoftware

     Advanced Brain Perfusion (OC & WS) Syngo Neuro Perfusion CT (OC, & WS)CBP (M

    CBP (3

    21.6Body perfusionsoftware

    Functional CT (OC & WS) Syngo Body CT Perfusion (OC & WS) Not ava

    21.7Computer assistedreading: Lung

    Lung Nodule CAD (OC & WS) Syngo Lung CAD (OC & WS) Lung C

    21.8Computer assistedreading: Colon

    Colon CAR (OC & WS) Syngo Colonography PEV (OC & WS) Colon C

    21.9Bone mineral

    densitometry

    Workstation Syngo Osteo (OC & WS) Not ava

    21.10 Dental OC & WS Syngo Dental (OC & WS ) Dental

    21.11 Image fusion OC & WS Syngo Image Fusion (OC & WS) Fusion

  • 8/18/2019 CT Scanners comparison

    39/61

    Technical specifications

    CEP08028: March 2009

    Table 22. Cardiac applications

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    22.1

    Prospective ECG-triggered cardiacscanning for calciumscoring

    Yes, Heartbeat-CS Pro HeartView option Yes

    22.2Calcium scoringsoftware

    Yes, Heartbeat-CS, Agatston, Mass, Average

    Syngo Calcium Scoring Option

    22.3

    Prospective ECG-triggered cardiacscanning for coronary

    angiography

    Yes, Rate Responsive CV Toolkit HeartView Yes

    22.4

    Retrospective ECG-gated cardiacscanning for coronaryangiography

    Yes, Rate Responsive CV Toolkit HeartView option Yes

    22.5Coronary angiographyvessel analysissoftware

    Yes, Comprehensive Cardiac Analysis Syngo Circulation option Yes

    22.6Temporal resolution forsingle sectorreconstruction [ms]

    135 150 175

    22.7Rotation times for

    cardiac scanning [sec]0.27 0.3 0.35 , 0

    22.8 Automatic adaptedselection of rotationtime?

    Yes No Yes

  • 8/18/2019 CT Scanners comparison

    40/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    22.9Maximum mAavailable for cardiacscanning

    1000 800 580

    22.10

    ECG-gated mAmodulation (maxreduction from peakmA) [%]

    80 96 100

    22.11Pitches available forcardiac scanning

    0.2 - 0.3 0.18 - 1.5Not app

    (scanni

    22.12 Automatic adapted

    selection of pitch?Yes, based on heart rate Yes - Heart Rate

    Not app

    (scanni

    22.13Multi-sectorreconstruction

    Yes, up to 5 sectors Yes - 2 5

    22.14 Automatic adaptedselection of multi-sector reconstruction

    Yes, MaxCycle - system automaticallyuses the maximum number of thesectors for each voxel.

    YesYes

    Heart R

    22.15 ECG editingYes, R-Tag can be added, removed andmoved.

    Delete, Disable, Insert, RR intervalreconstruction ( time and %)

    Yes

  • 8/18/2019 CT Scanners comparison

    41/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    22.16

    For 70 kg patient, 60bpm give typical scanparameters and

    (a) effective dosewithout ECGmodulation

    (b) effective dose withECG modulation

    Info. not available

    70%-70% window for ECG pulsing

    Retrospective gating (helical - singlesegment recon):

    14.5 mSv, w/o ECG-gated currentmodulation

    5.9 mSv, with ECG-gated currentmodulation (to 20% max)

    3.5 mSv, with ECG-gated currentmodulation (to 4% max)

    Prospective gating (sequential):

    2.7 mSv

    120 kV

    240 * 0

    Effectiv

    Modula

    Prospe

    Table 23. Radiotherapy appli cations

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    23.1Radiotherapy planningtable top

    Option Option Not ava

    23.2Carbon fibre breastboard

    No Option Not ava

    23.3External positioninglasers

    Option Option Not ava

    23.4 External positioninglaser options

    LAP 1 Red LAP 3 Red LAP 1 GreenLAP 3 Green

    Option Not ava

    23.5Respiratory gatingsoftware

    Option OptionNot ava

    (Not ne

    23.6Respiratory gatinghardware

    Bellows - option Pressure SensorNot ava

    (Not ne

  • 8/18/2019 CT Scanners comparison

    42/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    23.7Radiotherapy CTsimulation software

    Not available Option Option

    Table 24. Advanced scanning facilities

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    24.1Dual-energy / multi-energy capabilities

    Not available No No

    24.2Dual-energy methodused

    Not available No Not app

    24.3Dual-energy analysissoftware

    Not available No Not app

    24.4Contrast media bolustracking software

    Yes - BolusPro UltraCARE Bolus - standard

    SURES

    24.5Contrast media bolustracking; scan timing

    Yes Region of interest set by the user1 fps -

    User de

    24.6Contrast media bolustracking; triggermechanism

     Automatic/manualRegion of interest threshold set by theuser

    HU

    24.7

    Contrast media bolus

    tracking; usermonitoring Yes Region of interest graph Image a

    24.8Contrast media bolustracking; manualoverride

    Yes Yes, to force the scan Abort o

  • 8/18/2019 CT Scanners comparison

    43/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    24.9CT fluoroscopy: out-of-room capability

    ExposureYes, provided Fluoro option has beenpurchased

    Not app

    24.10CT fluoroscopy: in-room capability

    FootswitchYes, provided Fluoro option has beenpurchased

    Not app

    24.11CT fluoroscopy, imagerate [fps]

    8 i-fluro option 10 fps Not app

    24.12CT fluoroscopy; beamcollimations [mm]

    Info. not available6 x 1.2, 12 x 1.2, 24 x 1.2, 1 x 5,

    1 x 10Not app

    24.13CT fluoroscopy; slice

    thicknesses [mm]

    Info. not available6 x 1.2, 3 x 2.4, 6 x 2.4, 3 x 4.8,

    6 x 4.8, 1 x 5, 1 x 10

    Not app

    24.14CT fluoroscopy; otherviewing options

    Info. not availableFull 2D and 3D Volume imagingavailable (option)

    Not app

    24.15CT fluoroscopy; in-room scan controls

    Info. not available Foot switch and button on the gantry Not app

    24.16CT fluoroscopy; in-room couch control

    Info. not availableJoy stick gantry controls, floating topcontrol and I control (option)

    Not app

    24.17CT fluoroscopy; scanalarms

    Info. not available Visual display on the monitor Not app

    24.18CT fluoroscopy,additional hardware

    Info. not availableIn room monitor single or dual, ceiling orflorr mounted, Joy stick, foot pedal, I

    control wired or wireless

    Not app

    24.19CT fluoroscopy, dosecontrol

    Info. not available Hand CARE: switch off dose for 100° Not app

    24.20 CT perfusion capability Info. not availableYes, syngo Neuro Perfusion, syngoBody perfusion and Adaptive 4D spiraloptions

    CBP (C

  • 8/18/2019 CT Scanners comparison

    44/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    24.21

    Maximum scancoverage for perfusionscans - sequentialscanning [mm]

    Info. not available 38 (Head & Body) 160

    24.22

    Maximum scancoverage for perfusionscans - helicalscanning [mm]

    Info. not available 140 (Head)/270 (Body) Adaptive 4DSpiral Option

    Not app

    24.23 CT perfusion analysis CT Perfusion packageYes, syngo Neuro Perfusion, syngoBody perfusion and Adaptive 4D spiraloptions

    CBP (C

    Table 25. Installation requirements

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    25.1Power supplyrequirements (gantry)

    200 - 500 V AC

    50/60 Hz

    225 kVA

    Three-phase distribution source

    380 - 440 V, 3Phase, N & E 50Hz

    140 kVA

    Supplied from Siemens Line ConnectionCabinet

    3 phase

    25.2Power supplyrequirements (other)

    Info. not available 380 - 440 V , 3Phase, N & E 50Hz Not app

    25.3 Plumbing requirements None Site dependant Not app

    25.4Minimum floor load-bearing [kg/m2]

    729.7 Up to 333.33 ±10% 1070

    25.5Minimum floor arearequired for gantry &couch [m²]

    26 18 Info not

  • 8/18/2019 CT Scanners comparison

    45/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    25.6Minimum floor arearequired for other in-room system [m²]

    Info. not available 1.48 Info not

    25.7Recommended floorarea required forscanner [m²]

    41.4 Site dependant ≥18 43

    25.8Minimum floor arearequired for the controlconsole [m²]

    5 Site Dependant - PC towers 0.28m² Info not

    25.9

    Gantry dimensions

    (H x W x D [mm]) andweight [kg]

    1980 x 2740 x 970; 2549 1980 x 2380 x 935; 2300 2030 x

    25.10

    Couch dimensions

    (H x W x L [mm]) andweight [kg]

    1120 x 690 x 2490; 404 1000 x 750 x 2445; 500 470 x 6

    25.11

    Control consoledimensions

    (H x W x L [mm]) andweight [kg]

    760 x 1190 x 910; 56Site Dependant - PC tower 450 x 620 x220

    700 x 4

    1540 x

    1540 x

    25.12

    Power unit dimensions

    (H x W x D [mm]) andweight [kg]

    1220 x 580 x 860; 498 Not applicable 875 x 8

    25.13

    Cooling unitdimensions

    (H x W x D [mm]) andweight [kg]

    None Not applicable Not app

    25.14 Supplementary plant Air compressor Not applicable Not app

  • 8/18/2019 CT Scanners comparison

    46/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    25.15

    Supplementary plantunit dimensions

    (H x W x D [mm]) andweight [kg]

    1040 x 610 x 610; 171 Not applicable Not app

    25.16Supplementary plantlocations

    With Power unit Not applicable Not app

    25.17Peak heat output fromsystem duringscanning [kW]

    57,136 BTU/hr 17.4 Info. no

    25.18System cooling

    method

     Air cooled Air Air

    25.19 Air conditioning forscanner room

    Info. not available Only comfort cooling needed Recom

    25.20

    Environmentalrequirements (max/mintemperature, humidity)in scanner controlroom

    18 - 24°C, 35 - 70% 18 - 28 °C, 20 - 75 % 16 - 28

    25.21

    Environmentalrequirements (max/mintemperature, humidity)for control computer

    18 - 24°C, 35 - 70% 18 - 28 °C, 20 - 75 % 16 - 28

    25.22

    Environmentalrequirements (max/mintemperature, humidity)for reconstructor

    18 - 24°C, 35 - 70% 18 - 28 °C, 20 - 75 % 16 - 28

  • 8/18/2019 CT Scanners comparison

    47/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    25.23

    Environmentalrequirements (max/mintemperature, humidity)for workstation

    18 - 24°C, 35 - 70% 18 - 28 °C, 20 - 75 % 16 - 28

    25.24Connection tonetworks [Mbits/s]

    1000 MPS / 1000 Base T 1000 1000

    25.25Network connectionsrequired

    10 MPS / 10 Base T 1 per PCRJ45

    CAT5E

    25.26 Network requirements LAN LAN LAN

    25.27Static networkaddresses required Yes Yes - one per PC Yes

    Table 26. Maintenance & support

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    26 a Service level opt ions

    26.1Standard hoursengineer call-out

    Info. not available 09:00 - 17:00 Info not

    26.224 hour engineer call-out

    Info. not available Available as a costed option Info not

    26.3Weekend engineercall-out

    Info. not available Available as a costed option Info not

    26.4Out of hours plannedservicing

    Info. not available Available as a costed option Info not

    26.5 Dial-in diagnostics Info. not availableSecure remote access available (seenote 2)

    Info not

  • 8/18/2019 CT Scanners comparison

    48/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    26.6 Tube replacement Info. not available Available as a contract option Info not

    26.7 Detector replacement Info. not available Included in a Performance TOP contract Info not

    26.8 Main computers Info. not available Included in a Performance TOP contract Info not

    26.9Monitors and displaydevices

    Info. not available Included in a Performance TOP contract Info not

    26.10 Workstations Info. not available Included in a Performance TOP contract Info not

    26.11User / applicationsupport

    Info. not available Included in a Performance TOP contract Info not

    26.12 Parts Info. not available Included in a Performance TOP contract Info not

    26 b Standard service level

    26.13Median call-outresponse time

    Info. not available On site within 4 hours (see note 3)  Info not

    26.14 Median resolution time Info. not availableIf no parts are required, the mean timeto repair is typically within 2 hours of theengineer gaining access to the system.

    Info not

    26.15Median parts deliverytime

    Info. not available Same day or next day (see note 4)  Info not

    26.16 Frequency of service Info. not available 2 per annum Info not

    26.17Duration of service [+/-0.5 day] Info. not available 0.5 day Info not

    26.18 Warrantee duration Info. not available 1 year from handover of system Info not

    26.19What is the tubereplacement policy?

    Info. not available One year (see note 5)  Info not

  • 8/18/2019 CT Scanners comparison

    49/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    26.20

    What elements ofscanner / plant are notcovered in standardsupport?

    Info. not available As listed (see Notes 6 and 7) Info not

    26 c Reliabili ty

    26.21

    What is the expectedlife span of the product? (for five 8 hour daysper week)

    Info. not available Not less than 7 years (see Note 8) Info not

    26.22What is average tubelife ?

    Info. not available

    Data to date shows that the average life

    expectancy of Straton x-ray tubes life isbetween 2-3 years (see Note 9). Info not

    26.23 Guaranteed tube life Info. not available One full year (see note 5)  Info not

    26.24What is the expecteddetector life ?

    Info. not availableDetectors should typically have a lifespan of 7 years (see Note 10)

    Info not

    26.25How long arecomponent suppliesmaintained ?

    Info. not available 10 years as per CRACOE Info not

    Table 27. Supplementary information

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    27.1When first installed inUK ?

    Not applicable Not applicable Info not

  • 8/18/2019 CT Scanners comparison

    50/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    27.2How many unitsinstalled ?

    Not applicable 2 orders Info not

    27.3 Any adverse incidentreports issued on thismodel ?

    Info. not available No No

    27.4

    What is the audiblenoise output of theproduct under normaloperating conditions?[dB(A)]

    Info. not available 68 @ 1m 61

    27.5

    What is the audiblenoise output of theproduct under standbyor idle conditions?[dB(A)]

    Info. not available 65 @ 1m 61

    27.6

    Is vibration generatedby the product undernormal operatingconditions?

    Info. not available No No

    27.7

    What are the levels ofwaste output undernormal operatingconditions?

    Info. not available None None

    27.8

    Do the waste outputsfrom product undernormal operatingconditions require anyregulatory permits,consents or licences?

    Info. not available No No

  • 8/18/2019 CT Scanners comparison

    51/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    27.9 Any waterconsumption?

    Info. not available No No

    27.10

    What levels, and sizesof particulate matterare emitted from theproduct under normaloperating conditions?

    Info. not available None None

    27.11

    What features aredesigned in to theproduct to mitigateemissions to air

    Info. not available Air FiltersCoversdust frois remo

    27.12

    To what extent is theproduct designed foruse in developinghealth economies?

    Info. not available Complies Not spe

    27.13What colours areavailable

    Info. not available White or Zebra

    Table 28. Sustainabilit y

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    28.1

    What is the peak

    power consumptionwhen scanning? [kVA] Info. not available ≤140 kVA

    90(PF > 0

  • 8/18/2019 CT Scanners comparison

    52/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    28.2

    What is the typicalpower consumptionwhen not scanning?[kVA]

    Info. not available 4 kVA20

    (PF > 0

    28.3

    What is the powerconsumption when in'Stand-By' mode ?[kVA]

    Info. not available 4 kVA20

    (PF > 0

    28.4

    What power saving

    features are designedin to the product ?

    Info. not available Standby mode

    By desisuch as

    detectoand a Dregene

    28.5

    Does the productdesign incorporatefeatures to reduceoverall energydemand?

    Info. not available Yes

    VariousreductioalgorithBoost 3Exposureduced

    28.6

    What percentage ofthe packaging used isderived from recyclateas opposed to virgin

    materials?

    Info. not available As per European Union regulations 0.03

    28.7

    What percentage ofthe packaging can berecycled ? (undernormal circumstances)

    Info. not available 1 1

  • 8/18/2019 CT Scanners comparison

    53/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    28.8 Are there any knowntoxic components usedin this scanner ?

    Info. not available Mercury, LeadPb (X-r

    Hg (LC

    28.9

    Does the system useany gases classed ashazardous oroperating underpressure ?

    Info. not available No No

    28.10Doe the product utiliseand refrigerants?

    Info. not available No No

    28.11

    What is thepercentage ofrecyclate utilised in theproduct manufacture?

    Info. not available Info. no

    28.12

    What is theenvironmental impactof operating thisscanner ?

    Info. not available Energy consumption / heat output Use of

    28.13

    What is the carbonfootprint of themanufacturingprocesses ?[Equivalent tonnes ofCO2 emissions]

    Info. not available Unknown 0.12

    28.14

    What is the carbonfootprint of operatingthe scanner?[Equivalent tonnes ofCO2 emissions perannum]

    Info. not available Unknown 15.2

  • 8/18/2019 CT Scanners comparison

    54/61

    Technical specifications

    CEP08028: March 2009

    Philips Bril liance iCT Siemens SOMATOM Definit ion AS+ Toshib

    28.15How does the scanneraddress the needs ofthe WEEE directives?

    Info. not available

    EMAS registered & WEEE RegisteredNr. DE 23691322

    Use of

    informaavailab

    systemWEEE

    TMS Ucomplia

    28.16Do you recycle /refurbish existingscanners?

    Info. not available Yes

    The ratapproxrelevanor treat

    require

  • 8/18/2019 CT Scanners comparison

    55/61

    Technical specifications

    CEP08028: March 2009

     Additional notes

    Siemens

    The following notes are additional comments that were received in response to the questionnaire, thorder to keep the responses in the comparison tables concise.

    Table 29. Notes to Siemens replies

    Note1:

    with reference toquestion 3.11

    It is not possible to accurately predict "The lifespan" of some components such as X-ray tubes as there are tgoverning their usage, for example - the number, type and frequency of examination procedures carried out, t

    made, the parameters selected for those exposures, the intervals between exposures, etc. all factors which artime. However based on average usage and historical data X-ray tube should typically have a life span of 3 ye

    Note 2:

    Q26.5

    Siemens Remote Services utilises a secure broadband internet connection, such as N3, such that systems capro-active monitoring which detects and reports parameter deviations and potential system malfunctions to ouSupport Centre. In addition your systems can receive software updates, can be protected against viruses anda user with clinical information and e-training.

    Note 3:

    Q26.13

    When a call is received at the UPTIME Support Centre, there is immediate call clarification with all resources resolution available within the team: immediate access to remote clarification information provided via Siemencall history on SAP, technical as well as application experts, and spare parts logistics. However, if required, wengineer to site within 4 working hours.

    Note 4:

    Q26.15

    Our logistics operation is as effective as it can be; parts ordered by 6.00 pm can be delivered to site before 8.0

    In some cases it may even be possible to get spare parts delivered on the same day. Our network of spare-pastrategically placed at hospital sites throughout the country makes parts delivery to site as transparent and effcomplemented by our ‘Assisted-Fit’ programme where a larger part can be delivered by appointment with all tto install it.

  • 8/18/2019 CT Scanners comparison

    56/61

    Technical specifications

    CEP08028: March 2009

    Note 5:

    Q26.19 and Q26.23

    The STRATON X-ray tube is delivered with one year full warranty. There are no restrictions in terms of scan sThis offer is not restricted to a pro rata bases, the warranty is delivered for one full year. If the tube should faiperiod, Siemens will replace every tube during these 12 months free of charge. e.g. If a tube fails in month 10 supplied free of charge, which will come with a 2 month warranty thus fulfilling a total warranty period of 12 mo

    Note 6

    Q26.20

    Contract Exclusions:

     All Systems: The following items are not included within the Maintenance Contract (MC) and will be charged frates, unless covered by special warranty terms and/or explicitly included within the MC.

    1. Network configuration work including -

    a. Changing IP addresses

    b. Network modificationsc. Making changes or adding items to networks, e.g. moving or adding workstations

    d. All Dicom changes including adding peripheral items, e.g. printers etc

    e. Any additional hardware/software which may be required to assist networking compatibility will be charged addition to the fixed charge.

    2. UPS Batteries

    3. Any Non-Siemens badged system components supplied with the system

    4. Accessories, consumable items and positioning aids

    5. Accidental and/or wilful damage caused through mishandling or neglect.

    In addition to the items listed under “All systems” the following modality specific items, are not included within (MC) and w

  • 8/18/2019 CT Scanners comparison

    57/61

  • 8/18/2019 CT Scanners comparison

    58/61

     Acknowledgements 58

    CEP08028: March 2009

    We should like to thank the following for their contribution to this report.

    Philips Medical Systems

    Siemens Healthcare

    Toshiba Medical Systems

  • 8/18/2019 CT Scanners comparison

    59/61

    References 59

    CEP08028: March 2009

    [1] Centre for Evidence-based Purchasing (CEP). Multislice CT scanners.Buyers guide: CEP 08007; NHS PASA March 2009 http://www.pasa.nhs.uk/PASAWeb/NHSprocurement/CEP/CEPproducts.htm 

    [2] Centre for Evidence-based Purchasing (CEP). 32 to 40 slice CT scanners,technical specifications. Comparative specifications: CEP 08026; NHS PASAMarch 2009 http://www.pasa.nhs.uk/PASAWeb/NHSprocurement/CEP/CEPproducts.htm 

    [3] Centre for Evidence-based Purchasing (CEP). 64 slice CT scanners,technical specifications. Comparative specifications: CEP 08027; NHS PASAMarch 2009 http://www.pasa.nhs.uk/PASAWeb/NHSprocurement/CEP/CEPproducts.htm 

    [4] Centre for Evidence-based Purchasing (CEP). 16 slice CT scanners,technical specifications. Comparative specifications: CEP 08025; NHS PASAMarch 2009 http://www.pasa.nhs.uk/PASAWeb/NHSprocurement/CEP/CEPproducts.htm 

    [5] Centre for Evidence-based Purchasing (CEP). Wide bore CT scanners,technical specification comparison. Comparative specifications: CEP 08029;NHS PASA March 2009 http://www.pasa.nhs.uk/PASAWeb/NHSprocurement/CEP/CEPproducts.htm 

    [6] Centre for Evidence-based Purchasing (CEP). CT scanner clinicalapplications software comparison. Comparative specifications. CEP 08024;NHS PASA December 2008 March 2009 http://www.pasa.nhs.uk/PASAWeb/NHSprocurement/CEP/CEPproducts.htm 

  • 8/18/2019 CT Scanners comparison

    60/61

     Appendix 1: Supplier contact details 60 

    CEP08028: March 2009

    Supplier contact details

    Philips Medical Systems

    The Observatory

    Castlefield Road

    Reigate

    Surrey

    RH2 0FY

    tel: 01737 230400

    www.medical.philips.com/uk/ 

    Siemens Healthcare

    Faraday House

    Sir William Siemens Square

    Frimley

    Camberley

    Surrey

    GU16 8QD

    tel: 01276 696000

    www.siemens.co.uk/medical 

    Toshiba Medical Systems

    Boundary Court

    Gatwick RoadCrawley

    West Sussex

    RH10 9AX

    Tel: 01293 653700

    www.toshiba-medical.co.uk 

  • 8/18/2019 CT Scanners comparison

    61/61

     Author and report information 61

    Comparative specifications:128 to 320 slice CT scannertechnical specifications

    The ImPACT Group

    St George’s Healthcare TrustBence Jones OfficesPerimeter RoadTootingLondon

    SW17 0QT

    Tel: +44 (0)20 8725 3366

    www.impactscan.org 

     About CEP

    The Centre for Evidence-basedPurchasing (CEP) is part of the Policyand Innovation Directorate of the NHS

    Purchasing and Supply Agency. Weunderpin purchasing decisions byproviding objective evidence to supportthe uptake of useful, safe andinnovative products and relatedprocedures in health and social care.

    We are here to help you makeinformed purchasing decisions bygathering evidence globally to supportthe use of innovative technologies,assess value and cost effectiveness ofproducts, and develop nationallyagreed protocols.

    Sign up to our email alertservice

     All our publications since 2002 areavailable in full colour to downloadfrom our website. To sign up to ouremail alert service and receive newpublications straight to your mailboxcontact:

    Centre for Evidence-based Purchasing

    Room 152CSkipton House80 London RoadSE1 6HL

    Tel: 020 7972 6080Fax: 020 7975 5795Email: [email protected] www.pasa.nhs.uk/cep

    © Crown Copyright 2008