ct scanners for coronary ct angiography (ccta) in...

129
Report CT scanners for coronary CT angiography (CCTA) in challenging patient groups King's Technology Evaluation Centre (KiTEC) Department of Medical Engineering and Physics King's College Hospital NHS Foundation Trust Denmark Hill London, SE5 9RS phone: +44 (0) 203 299 1626 fax: +44 (0) 203 299 3314 Certificate No. FS36209

Upload: others

Post on 22-Sep-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

Report

CT scanners for coronary CT angiography

(CCTA) in challenging patient groups

King's Technology Evaluation Centre (KiTEC) Department of Medical Engineering and Physics King's College Hospital NHS Foundation Trust Denmark Hill London, SE5 9RS phone: +44 (0) 203 299 1626

fax: +44 (0) 203 299 3314

Certificate No. FS36209

Page 2: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

1

Table of Contents Abbreviations ............................................................................................................5

Glossary ...................................................................................................................7

1. REPORT SUMMARY ............................................................................................. 10 1.1 Comparative technical specifications ................................................................. 10

1.2 Approaches used by the manufacturers to address the challenges in CCTA of

difficult to image patients ............................................................................................ 11

1.2.1 Patients with high heart rates .......................................................................... 11

1.2.2 Arrhythmia ...................................................................................................... 12

1.2.3 High calcium scores ........................................................................................ 12

1.2.4 Coronary artery stents .................................................................................... 13

1.2.5 Patients with coronary artery bypass grafts (CABG) ....................................... 13

1.2.6 Obese patients ............................................................................................... 14

1.3 Additional challenges in scanner selection ......................................................... 15

1.4 Advantages, uncertainties and risks of comparing scanners using technical

specifications .............................................................................................................. 15

2. INTRODUCTION .................................................................................................... 17 3. CLINICAL CHALLENGES IN CARDIAC IMAGING ................................................. 18

3.1 Patients groups posing additional challenges .................................................... 19

3.2 Key technical requirements in cardiac CCTA imaging ........................................ 20

4. PRINCIPLES OF CCTA SCANNING ...................................................................... 22 4.1 Essential components of a CT scanner and key scanner parameters ................ 22

4.2 Scan modes used in CCTA ................................................................................ 24

5. RADIATION DOSE IN CCTA .................................................................................. 31 5.1 Justification of the CCTA examination ............................................................... 33

5.2 Limiting Z-axis scan length ................................................................................ 33

5.3 Use of small scan field of view ........................................................................... 33

5.4 Tube current modulation .................................................................................... 33

5.4.1 Automatic (spatial) tube current modulation (ATCM) ....................................... 34

5.4.2 Temporal (ECG-gated) tube current modulation ............................................. 35

5.5 Optimisation of tube potential ............................................................................ 36

5.6 Prospectively ECG-triggered scan modes ......................................................... 38

5.7 Development of standardized protocols ............................................................. 39

5.8 Iterative reconstruction (IR) ................................................................................ 42

Page 3: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

2

5.9 Z-axis overbeaming and overscanning .............................................................. 43

6. CT SCANNERS FOR CARDIAC IMAGING ............................................................ 45 6.1 CT scanner models included in the report .......................................................... 45

6.2 Brief description of scanner models ................................................................... 46

6.2.1 GE Healthcare ................................................................................................ 49

6.2.2 Philips Healthcare ........................................................................................... 49

6.2.3 Siemens Healthcare ....................................................................................... 50

6.2.4 Toshiba Medical Systems ............................................................................... 51

6.3 Comparison of technical specifications .............................................................. 53

6.3.1 Temporal resolution ........................................................................................ 53

6.3.2 Spatial resolution ............................................................................................ 55

6.3.3 Volume coverage ............................................................................................ 60

6.3.4 X-ray flux ........................................................................................................ 63

7. TECHNICAL APPROACHES ADOPTED IN CURRENT SYSTEMS TO OPTIMISE IMAGE QUALITY IN ‘DIFFICULT TO IMAGE’ PATIENT GROUPS ............................... 65

7.1 Patients with high heart rates (> 65 bpm) .......................................................... 65

7.2 Patients with arrhythmia .................................................................................... 72

7.3 Patients with high calcium scores (>400) ........................................................... 74

7.4 Patients with stents ............................................................................................ 78

7.5 Patients with coronary artery bypass grafts (CABG) .......................................... 84

7.6 Obese patients (> 30 kg/m2).............................................................................. 85

8. ADVANTAGES, UNCERTAINTIES AND RISKS OF COMPARING SCANNERS USING TECHNICAL SPECIFICATIONS ....................................................................... 95

8.1 Bias ................................................................................................................... 95

8.2 Impact on clinical performance .......................................................................... 95

8.3 Access to technical specifications ...................................................................... 95

8.4 Quality of the technical data ............................................................................... 95

8.5 Expertise required ............................................................................................. 96

8.6 Manufacturer-specific technical features ............................................................ 96

8.7 Software upgrades ............................................................................................ 96

8.8 Multi-factorial effects .......................................................................................... 96

8.9 Technological advances .................................................................................... 97

8.10 Patients with multiple conditions ...................................................................... 97

9. FURTHER WORK .................................................................................................. 98 APPENDIX 1: CLINICAL EVIDENCE ............................................................................ 99

Page 4: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

3

APPENDIX 2: QUESTIONNAIRE USED TO COLLECT TECHNICAL SPECS OF CT SCANNERS ................................................................................................................ 112 APPENDIX 3: REFERENCES ..................................................................................... 115 APPENDIX 4: SNOMED-CT CODES .......................................................................... 126

Page 5: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

4

Report Details

Produced by KiTEC - King's Technology Evaluation Centre

Department of Medical Engineering and Physics

King's College Hospital NHS Foundation Trust

Denmark Hill

London, SE5 9RS, UK

phone: +44 (0) 203 299 1626

fax: +44 (0) 203 299 3314

Authors (alphabetical) Keevil, Stephen

Lewis, Cornelius

Lewis, Maria

McMillan, Viktoria

Pascoal, Ana

Correspondence to Professor Stephen Keevil, 0201 7188 3054

[email protected]

Dr Cornelius Lewis, 0203 2991646

[email protected]

Website version November 2014

Acknowledgements

KiTEC appreciate the co-operation of the following CT manufacturers in providing

CT scanner specification data for the production of this report:

GE Healthcare, Chalfont St Giles, Buckinghamshire, UK

Philips Healthcare, Guildford, Surrey, UK.

Siemens Healthcare, Frimley, Surrey, UK.

Toshiba Medical Systems, Crawley, West Sussex, UK

Declaration of interest

KiTEC is commissioned by the NICE Medical Technologies Evaluation

Programme to deliver evidence preparation and assessment services. The

design, development and authoring, of this report, and any opinions expressed,

are the sole responsibility of the authors.

Page 6: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

5

Abbreviations

AF Atrial fibrillation

BMI Body mass index

bpm beats per minute

CABG Coronary artery bypass graft

CAD Coronary artery disease

CCS Canadian Cardiovascular Society

CEP Centre for evidence-based purchasing

CCA Catheter coronary angiography

CCTA Coronary computed tomography angiography

CT Computed tomography

CTA Computed tomography angiography

CTDIw Computed tomography dose index weighted

CTDIvol Volume computed tomography dose index

CNR Contrast-to-noise ratio

CV Cardiovascular

CVD Cardiovascular death

DAR Diagnostic Assessment Report

DG3 Diagnostic Guidance 3

DSCT Dual-source computed tomography

EAC External Assessment Centre

ECG Electrocardiogram

HCS High calcium score

HD High definition

HDCT High definition computed tomography

HHR High heart rate

HR Heart rate

HRF Heart rate frequency

HRQoL Health-related quality of life

HRV Heart rate variability

ICA Invasive coronary angiography

ICER Incremental cost-effectiveness ratio

ImPACT Former UK centre for CT scanner technical evaluations, based at St. George Hospital, London, (disbanded 2011)

Page 7: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

6

KHP King’s Health Partners

KiTEC Kings Technology Evaluation Centre

kV Kilovolt

kW Kilowatt

LAD Left anterior descending artery

LCA Left coronary artery

LCX Left circumflex artery

mA milliampere

MI Myocardial infarction

MSCT Multi-slice computed tomography

MSR Multi-segment reconstruction

MTF Modulation transfer function

NA Not applicable

NFE Non-fatal event

NFMI Non-fatal myocardial infarction

NHS National Health Service

NICE National Institute for Health and Care Excellence

NIHR National Institute for Health Research

NR Not reported

NGCCT New generation cardiac computed tomography

OR Odds ratio

PCI Percutaneous coronary intervention

PTA Prospectively ECG-triggered axial

PTH Prospectively ECG-triggered helical

QALY Quality-adjusted life year

RCA Right coronary artery

RGH Retrospectively ECG-gated helical

ROC Receiver operating characteristic

Se Sensitivity

SNOMED-CT Systematized nomenclature of medicine clinical terms

Sp Specificity

SROC Summary receiver operating characteristic

UK United Kingdom

Page 8: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

7

Glossary

Arrhythmia Any disturbance of the normal rhythmic beating of the heart or

myocardial contraction.

Artefact Error in the representation of any visual information introduced by

the technique(s) or equipment.

Blooming artefact High-attenuation structures, namely calcified plaques or stents,

with enlarged appearance due to partial-volume effects,

obscuring the adjacent lumen.

Calcium scoring A technique by which the extent of calcification in the coronary

arteries is measured and scored.

Contrast resolution Ability to differentiate between different tissue types in an image.

Contrast-to-noise

ratio

A measure related to image quality. Defined as the difference

between the signal in the region of interest and the surrounding

background divided by the average variation in the background.

Coronary

angiography

A diagnostic imaging procedure which provides anatomical

information about the degree of stenosis (narrowing) in a

coronary artery.

Coronary artery An artery that supplies the myocardium.

Coronary artery

disease

A condition in which atheromatous plaque builds up inside the

coronary artery leading to narrowing of the arteries which may be

sufficient to restrict blood flow and cause myocardial ischemia.

CTDIw A radiation dose index used in CT to represent the average

absorbed dose in the scan plane of a standard perspex phantom.

CTDIvol As CTWIw but taking into account the exposure variation in the z

axis for non-contiguous scans.

Double sampling Sampling technique that utilises two measurements per detector

in order to prevent deterioration of resolution and the generation

of aliasing artefacts. It can be performed in x-y plane and also z-

axis.

Dual-source multi-

slice computed

tomography

A dual source scanner has two pairs of x-ray sources and multi-

slice detectors mounted at approximately 95 degrees to each

other.

Page 9: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

8

Dynamic focal spot

(“flying focal spot”)

An electromagnetic shifting of the location of the focal spot on the

x-ray tube anode relative to the detectors. This technique is used

to increase the data sampling density (see ‘double sampling’).

ECG gating Technique used to improve temporal resolution and reduce

artefacts due to cardiac motion. The patient’s ECG provides real-

time information of the cardiac cycle and is used to identify the

point to trigger image acquisition or to reconstruct images.

Focal spot The area on the anode of an x-ray tube that is struck by electrons

and from which the resulting x-rays are emitted.

Gantry Part of the CT scanner that accommodates the x-ray source,

detectors and rotating mechanism to allow cross-sectional views

of patient anatomy.

Graft Surgical procedure to move tissue from one site to another on

the body, or from another person, without bringing its own blood

supply with it.

Invasive Coronary

Angiography

Invasive procedure used to image the blood vessels of the heart.

It involves manipulation of cardiac catheters from an artery in the

arm or top of the leg. A contrast medium is injected into the

coronary arteries, and the flow of contrast in the artery is

monitored with a rapid series of x-rays. It is considered the “gold

standard” for providing anatomical information and defining the

site and severity of coronary artery lesions.

Inter-observer

variability

The differences occurring between individuals performing the

same and especially a visual task.

Isotropic resolution Equal resolution in all three dimensions of the voxel, i.e. in the

scan plane (x-y), and in the z-axis.

Iterative

reconstruction

Algorithms used to reconstruct 2D and 3D images by performing

repeated reconstructions to improve image fidelity.

Modulation

transfer function

Illustrates the fraction (or percentage) of an object’s contrast that

is recorded by the imaging system, as a function of size (i.e.

spatial frequency) of the object. The MTF is a used as a

complete description of the resolution properties of an imaging

system.

Multi –segment

reconstruction

An approach used in cardiac gated scans whereby data from two

or more heart beats is used to reconstruct an image .

Multi-slice CT

coronary

angiography

A non-invasive investigation that provides anatomical information

about the degree of stenosis (narrowing) in the coronary arteries.

The scanner requires a high rotation time and as the technology

Page 10: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

9

has advanced the number of slices in each rotation has

increased.

Negative predictive

value

Indicates the likelihood that a patient does not have a disease

when the test result is negative.

Padding Padding – additional x-ray beam on time; the x-ray exposure

window may be extended to cover a longer phase of the cardiac

cycle to allow for greater a range of cardiac phases available for

image reconstruction.

Pitch The ratio of table feed per rotation to slice collimation. A pitch

less than 1 represents overlapping scans.

Positive predictive

value

Indicates the likelihood of disease in a patient when the test

result is positive

Revascularisation The restoration of blood supply to the affected tissues.

Sampling rate In CT, the frequency at which data are collected as the gantry

rotates around the patient.

Sensitivity Proportion of people with the target disorder who have a positive

test result.

Spatial resolution Ability of an imaging system to accurately depict anatomical

features in an image.

Specificity Proportion of people without the target disorder who have a

negative test result.

Stenosis A narrowing of the arteries leading to a reduction in blood flow.

May be due to the build-up of atherosclerotic deposits of fibrous

and fatty tissue or may be a congenital defect.

Stent An implantable device designed to be inserted into a vessel or

passageway to keep it open.

Temporal

resolution

In CT, it is the period of time over which medical images are

captured or recorded. The ‘intrinsic spatial resolution’ is that

which can be achieved by the scanner without using special

modes such as multi-segment reconstruction or motion

correction techniques.

Volume coverage Refers to the length of the detector array along the z-axis and

can be calculated from the sum of the dimensions of each

detector row along this axis.

z-axis The direction that the scanning table travels in (i.e. head to toe or

vice versa).

Page 11: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

10

1. Report summary

This report discusses the technical approaches used to address the challenges of CCTA

imaging and compares the relevant technical specifications of scanners available on the

market suitable for cardiac imaging of challenging patient groups.

A full comparison of eleven CT scanner models is included in this report from the four

main CT scanner manufacturers (GE, Philips, Siemens and Toshiba) contacted. All these

CT scanners are available on the UK market. An additional two scanners are only

described briefly due to lack of data at the time this report was compiled. The vast

majority of the data included in the report were provided by the manufacturers via

responses to a questionnaire1 (Appendix 2).

The principal performance parameters compared in this report are: temporal

resolution, spatial resolution, anatomical volume coverage, and x-ray flux, because

they are considered key in CCTA imaging in the difficult to image subgroups. This

approach was adopted because it helps to clarify the impact of individual technical

features on the quality of the images produced. However, it should be noted that

technical specifications do not provide a complete description of the performance of a

CT scanner and therefore should not be used to draw final conclusions on the

comparative clinical performance of the different scanners. Other technical and non-

technical factors such as post-processing tools, patient characteristics and disease

state also have an impact on the overall clinical performance of the scanner.

1.1 Comparative technical specifications

The key technical specification data for each scanner are tabulated in chapter 6. All

models are third generation CT scanners and have between 64 and 320 detector rows

with dimensions of between 38.4 mm and 160mm in the craniocaudal (z-axis) direction.

Minimum gantry rotation times range from 250 ms to 420 ms. Two models are dual

source systems2. Other specifications tabulated and discussed are: gantry bore diameter,

maximum scan field of view, weight allowed on couch, x-ray generator power, tube

voltage range and tube current range.

1 At the time of the exercise we were informed by some manufacturers that not all data requested

was available, particularly on their newest systems. Additionally, for some of the claims made,

published peer-reviewed evidence is still not available.

2 Although the manufacturer of dual source systems now offers a second dual source model, at

the time this report was compiled sufficient technical data was not available to include this

model in the full comparative specifications.

Page 12: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

11

1.2 Approaches used by the manufacturers to

address the challenges in CCTA of difficult to

image patients

CT manufacturers have taken different approaches to meeting the challenges posed by

the various difficult to image patient groups and these are discussed in full in chapter 7.

They are summarised in the paragraphs below.

Whilst each group is discussed separately it is certainly possible that patients will present

with more than one challenge and when scanning such patients this will need to be

considered.

1.2.1 Patients with high heart rates

High temporal resolution and selection of the most static cardiac phase for image

reconstruction are essential for obtaining CCTA images that are free from cardiac motion

artefacts.

The scan mode used is often dependent on the patient’s heart rate. Current

recommendations are to scan patients with lower heart rates with prospectively ECG-

triggered axial (PTA) mode, or prospectively ECG-triggered helical (PTH) mode where

this mode is offered, as these tend to be lower dose modes. However, these modes

generally allow less flexibility in the cardiac phases available for image reconstruction.

Dual source systems, such as the Siemens Somatom Definition Flash Stellar and

Siemens Somatom Force, have a high intrinsic temporal resolution, of particular value for

imaging patients with high heart rates, and on the these systems CCTA scans can be

performed using PTA scan mode even at heart rates above 85 bpm.

Scanning using the same scan mode regardless of a patient’s heart rate is possible on

scanners with a large z-axis coverage, for example the GE Revolution CT and Toshiba

Aquilion ONE models, where scanning in PTA Volume mode is recommended even at

heart rates greater than 75 bpm.

On the Philips Brilliance iCT it has been shown that diagnostic CCTA scans can reliably

be achieved in PTA mode at heart rates up to 75 bpm but in practice diagnostic images

have been obtained at higher heart rates.

A software solution for dealing with motion artefacts is offered on GE systems. The

SnapShot Freeze (SSF) motion correction algorithm is claimed to provide an effective

temporal resolution of 29 ms although peer-reviewed evidence is not yet available.

Toshiba have also recently implemented Adaptive Motion Correction, an algorithm that

compensates for motion in the vessels, myocardium and valves.

Page 13: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

12

1.2.2 Arrhythmia

The issue of whether CCTA is a suitable examination for excluding CAD in patients with

atrial fibrillation remains contentious (Vorre et al, 2013; Schuetz et al, 2013). Limited

evidence is available, largely due to the relatively recent technological developments that

have allowed improvements in this area.

GE claims that its adaptive gating algorithm can predict the heart rate of the next cardiac

cycle allowing the scanning window to be adapted accordingly. In addition GE

recommends the use of the SSF motion correction algorithm and multi-segment

reconstruction techniques for improving the effective TR. The Revolution CT scanner

provides a ‘Smart arrhythmia management’ tool that avoids scanning during an irregular

beat.

The Philips iCT Elite can cover the cardiac volume in two to three heart beats in PTA

scan mode and studies (on the Brilliance iCT) have shown that patients with atrial

fibrillation can be successfully imaged (Muenzel et al, 2011, Chao et al, 2010). All Philips

CT scanners supporting card CCTA have automated arrhythmia handling tools that

enable diagnostic quality scans through detection or rejection of ectopic beats.

A meta-analysis has shown that the high intrinsic temporal resolution of the Siemens

dual source scanner makes it suitable for ruling out CAD in patients with atrial fibrillation

(Sun G et al, 2013). On all Siemens cardiac scanner models the ‘adaptive sequence’

axial mode will omit or repeat a scan when an ectopic beat is detected. For patients with

known arrhythmia Siemens recommends the use of retrospectively ECG-gated helical

(RGH) mode with automatic temporary suspension of ECG-dose modulation if arrhythmia

is detected.

Toshiba’s solution to scanning patients with arrhythmia on the Aquilion PRIME scanner is

to switch from PTH to RGH mode if arrhythmia is detected (and vice versa if the heart

rate returns to normal). On the Aquilion ONE scanners, exposure is delayed if an

arrhythmia occurs and will not take place until the heart returns to its normal rhythm. A

recent study which included a small number of patients with heart rhythm irregularities

(Uehara, 2013) showed that equivalent results could be obtained in these patients as on

those with regular heart rhythm.

1.2.3 High calcium scores

A high spatial resolution can reduce the amount of calcium ‘blooming’ and will be

particularly beneficial in the presence of high calcium levels. Manufacturers therefore

generally recommend the use of sharper reconstruction filters in these cases, although

this will result in increased image noise.

Page 14: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

13

Other than a high intrinsic spatial resolution, a good temporal resolution is important to

reduce blurring due to motion, and other features such as scanning at higher tube

potentials (kVs) are recommended. Additionally, the use of iterative reconstruction

algorithms, calcium subtraction techniques and dual energy scans, where available, are

proposed.

GE recommends the use of the High Definition (HD) mode available on the Discovery

CT750 HD3 and the Revolution CT. On the CT750 HD use of the dual energy cardiac

mode (GSI Cardiac) to obtain monochromatic (keV images) is also recommended.

Philips claims that the use of iterative algorithms, iDose4 or IMR (Iterative Model-based

Reconstruction) results in improved spatial resolution and therefore reduced blooming

(Philips, 2011).

Siemens recommends using SAFIRE or ADMIRE iterative reconstruction for reduced

blooming in patients with high calcium scores.

On Toshiba scanners, the iterative reconstruction algorithm AIDR 3D which includes

artefact reduction software is recommended. Toshiba also recommends using SURESubtraction Coronary for removing calcium, an approach claimed to have fewer

drawbacks than thresholding techniques.

1.2.4 Coronary artery stents

In common with high calcium scores, the presence of stents can also lead to beam

hardening artefacts and blooming that can make the diagnosis of in-stent restenosis

difficult. Different types of stent materials result in varying levels of blooming but

generally image interpretation becomes problematic for stent diameters less than 3 mm.

Manufacturers generally use the same approaches for dealing with stent imaging as for

high calcium levels. Toshiba’s SURESubtraction Coronary can also be used to remove

stents and check patency, an approach claimed not to be possible with thresholding

techniques alone.

1.2.5 Patients with coronary artery bypass grafts (CABG)

The challenges in the assessment of native coronary arteries in patients after CABG (as

discussed in section 7.5) are related to poor run-off, more extensive calcification and

3 The GE Discovery CT750 HD has been replaced by the Revolution GSI and Revolution HD

CT scanner models. The GSI has dual energy capability whereas the HD does not.

Page 15: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

14

diffusely narrowed arteries with small dimensions (Sun et al, 2012). Additionally scanning

of longer lengths to include the whole thorax may be required.

On the Revolution CT scanner GE suggests the use of two table positions with ‘Smart

collimation’ to enable one beat acquisition of the heart and avoid a volume boundary over

the heart.

On the Siemens Somatom Definition Flash Stellar the use of high pitch helical ‘Flash’

mode is claimed to offer diagnostic quality when extended coverage is required.

The results of a small study with the Philips Brilliance iCT scanner showed that CCTA

with a low dose protocol achieved a high diagnostic accuracy in patients with CABG

(Aunt Minnie, 2014).

On the Toshiba Aquilion ONE, upon setting the desired range, Wide Volume mode

automatically calculates the two acquisitions volumes required to cover the heart and

graft. In addition Toshiba claims that AIDR 3D iterative algorithm and a large range of

reconstruction filters available will aid in providing good image quality in CCTA on

patients with CABG.

1.2.6 Obese patients

Manufacturers generally suggest the use of a higher tube potential if an increased photon

flux is required at the detectors. However, the introduction of iterative reconstruction

algorithms has somewhat reduced the problem of photon flux limitation even in obese

patients as, with the same exposure settings, lower noise values are achieved than with

traditional filtered back projection (FBP) methods. The approaches suggested by

manufacturers to scan obese patients are summarised in the following paragraphs.

GE promotes the use of the ASIR iterative reconstruction algorithm, or the more

advanced ASIR-V on the Revolution CT. Another suggested approach for achieving an

increased photon flux in obese patients is to decrease rotation time, which will result in a

higher tube-current-time product (mAs). However, the reduced noise will be achieved at

the expense of temporal resolution. On the Revolution CT automatic exposure control

(AEC) is available for CCTA scans and so the tube potential (kV Assist) and tube current

(Auto mA/Smart mA) will automatically be optimised for patient size.

Philips proposes the use of the iterative algorithm iDose4 and claims that it facilitates

noise reduction whilst maintaining diagnostic image quality in PTA CCTA scans

performed on obese patients.

Siemens has AEC technology available for CCTA on all the cardiac scanner models.

CAREDose 4D automatically adjusts the mA and CAREkV automatically selects the

optimal kV for the patient size. Siemens claims that the Somatom Edge Stellar and

Page 16: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

15

Somatom Flash Stellar systems will be advantageous for scanning obese patients as the

Stellar detectors, which have low electronic noise, are optimised for low-signal imaging.

Toshiba’s approach to performing CCTA on obese patients is to use AIDR 3D iterative

reconstruction. The AEC system (SUREExposure) automatically adjusts the mA for patient

size and also takes account of the fat in the patient, taking advantage of inherent contrast

when calculating the exposure required. SUREkV suggests the optimal tube potential for

optimising contrast and image quality. Toshiba claims that their new PUREVision detector

provides a 40% increase in light output and a 28% decrease in electronic noise (Toshiba,

2014). Toshiba CT scanners also have the unique feature of lateral couch movement to

facilitate accurate patient positioning, although restrictions in scan field of view need to

be considered.

From a practical viewpoint, scanners with a couch that supports a high weight and those

with a large gantry bore will be desirable.

1.3 Additional challenges in scanner selection

A diagnostic CCTA scan at the lowest possible dose for each patient requires not only

suitable equipment but also an optimised protocol. CCTA protocols are complex and

multi-factorial. The choice of protocols for challenging patients is also influenced by user

preferences that will vary between centres.

Introduction of new features into clinical protocols should be approached carefully and

there is a need for education of users by the manufacturer on how these features can be

best used. Appropriate use of the tools provided on the CT scanner requires an

understanding of how the system works and how variations in scan parameters,

reconstruction methods and post processing tools affect image quality and patient dose.

The discussions in this report mainly address the hardware features of CT equipment.

There are many software features available for CCTA scanning which can enhance

image quality at the stages of data acquisition and image reconstruction, as well as at the

reporting stage. To adequately compare these software features would require further

input from the manufacturers, and would be extremely challenging as these aspects of

scanner technology are changing rapidly.

1.4 Advantages, uncertainties and risks of

comparing scanners using technical

specifications

Comparing technical specifications of CT scanners has value but also has limitations and

these are discussed in Chapter 8. In summary, technical specification data provides

Page 17: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

16

useful preliminary information to help with the selection of a suitable scanner for CCTA

but their limitations must be considered.

Page 18: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

17

2. Introduction

Coronary computed tomography angiography (CCTA) is in increasingly widespread use

for non-invasive evaluation of coronary artery disease. CCTA is now a well-established

technique with accuracy similar to that of invasive coronary angiography. However,

certain patient groups are more challenging and may require specialised protocols or

equipment for successful imaging.

In this report we describe the technical developments in CT technology which address

the challenges of CCTA and list the CT scanners currently available on the market for

cardiac imaging in the challenging patient groups i.e. patients with high heart rates,

arrhythmia, high coronary calcium scores, stents and/or grafts and obese patients.

Technical specifications and features of the relevant CT scanners are presented

alongside a discussion of their impact on CCTA with a focus on the challenging patient

groups.

Permission has been obtained for including images from relevant publications with

acknowledgement to the source of the material.

Page 19: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

18

3. Clinical challenges in cardiac imaging

The gold standard for evaluating coronary anatomy is invasive coronary angiography

(ICA) because of its excellent spatial resolution (Lim et al, 2013). However, significant

advances in coronary CT angiography (CCTA) using multiple-row detectors (MDCT)

have made it possible to evaluate the heart and coronary arteries non-invasively.

CCTA is a non-invasive procedure used to image the heart, great vessels and coronary

arteries. It provides anatomical and functional information and is usually aimed at

diagnosing coronary artery disease. Recent advances in CT hardware and software have

promoted a rapid expansion in the clinical use of CCTA.

Various studies have demonstrated the consistently excellent negative predictive value

(NPV) of CCTA imaging (99%) for 64-row CT scanners (Mowatt, 2008). This is useful in

the care pathway, particularly to exclude the need for patients to undergo ICA

procedures.

The scanner’s technical features, the clinical protocol used and the patient characteristics

all influence the diagnostic performance. Varying values for the specificity of CCTA (with

64-row systems) have been reported in the literature, including 64% (Meijboom et al,

2008), 83% (Budoff et al, 2008) and 90% (Miller et al, 2008).

CCTA imaging poses additional challenges compared to CT imaging of other parts of the

anatomy for the following reasons:

the cardiac arteries are constantly moving in a complex, cyclical pattern;

the cardiac arteries taper down to small diameters and follow tortuous routes;

the cardiac anatomy to be imaged is typically 120 – 140 mm in length and this

volume must be scanned within a patient’s breath-hold time; and

sufficient radiation needs to be delivered in a very short period of time in order

to achieve adequate diagnostic quality and allow contrast differentiation

between tissue types.

Page 20: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

19

3.1 Patients groups posing additional challenges

Patients with the following characteristics pose further challenges for CCTA:

High heart rate: the heart’s normal rhythm (sinus rhythm) at rest is between 60 and 100

beats per minute (bpm) (British Heart Foundation, 2012). In this report a heart rate

elevated above 65 bpm is defined as high (NICE, 2012).

Arrhythmia: a disturbance of the normal cardiac rhythm (faster, slower or irregular)

(British Heart Foundation, 2012).

High calcium score: quantified using the Agatston score, a CT-based score derived

from the product of the measured density and area of the calcium in the artery (Pelber,

2007). In this report an Agatston score of > 400 Agatston Units is defined as high (NICE,

2010).

Obesity: quantified using the Body Mass Index (BMI) derived from the patient’s weight

and height. In this report, patients with a BMI of > 30 kg/m2 are defined as obese (NICE,

2010).

Cardiac stents: these devices have become a mainstay in coronary revascularization

therapy. Despite major advances in stent materials and CT scanner technology, stents

may not be well defined in CT images mostly due to blooming or motion artefacts

(Mahnken, 2012).

Coronary artery bypass grafts: coronary artery bypass graft (CABG) surgery may be

performed in coronary revascularization therapy (Jones et al, 2008).

In the UK approximately 500,000 inpatient episodes of coronary heart disease were

recorded in 2010/11 (405,000 in England, 50,200 in Scotland, 24,300 in Wales and 14,600

in Northern Ireland) (British Heart Foundation).

No literature was found with accurate estimates of the number of people that fit within each

patient subgroup for whom CT imaging may be challenging however, some sources can

be used to obtain estimates of the target population.

According to the Health Survey for England (2012) 67% of men and 57% of women were

either overweight (25 kg/m2 ≤BMI<30 kg/m2) or obese (BMI ≥30 kg/m2) (Heath survey,

2014). Overweight was more common than obesity, with 42% of men and 32% of women

being overweight but not obese (compared to a total of 25% obese).The same study

reported a mean BMI of 27.3 kg/m2 for men and 27.0 kg/m2 for women and this rose with

age from youth to late middle-age, before falling again in old age.

Central obesity (assessed by waist circumference) was relatively common. 34% of men

and 45% of women had a raised waist circumference (over 102 cm for men and over 88

cm for women). Raised waist circumference increased with age and continued to increase

into the oldest age group to 52% of men and 64% of women aged 75 and over.

Page 21: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

20

Considering that raised BMI increases the risk of coronary heart disease, these numbers

suggest that a significant proportion of the patients that could benefit from CCTA may be

overweight and/or obese.

The same source also reported 104,000 admissions to hospitals in England due to atrial

fibrillation, flutter and other cardiac arrhythmias.

An additional 23,000 admissions were due to complications of cardiac and vascular

prosthetic devices, implants and grafts some of which may be eligible for CCTA.

These statistics suggest that a significant proportion of the population may fall into

groups that have been defined as “challenging to image” to image.

Radiation exposure to patients undergoing CCTA can be relatively high in diagnostic

terms, which poses some risk. Repeated exposures potentially required by patients with

chronic conditions will increase this risk.

Application of the fundamental principles of radiation protection, justification and

optimisation, are essential when using CCTA. Radiation dose and risk in CCTA is further

discussed in Chapter 5.

3.2 Key technical requirements in cardiac CCTA

imaging

In recent years manufacturers of CT scanners have taken different approaches in

developing scanner technology (Fleischmann et al, 2011) (Hassan et al, 2011). Some

have aimed their developments particularly at cardiac applications, whereas others have

focussed on improvements that have more general applications.

Cardiac CT imaging pushes the limits of temporal resolution, spatial resolution, x-ray

generator power, data acquisition and processing and has been the driving application

for many developments in CT. Table 1 summarises the specific challenges for CCTA

imaging in the defined patient subgroups and the key technical requirement to reduce the

problem. Of course it is important to recognise that all these features will have an impact

on the overall image quality for the different subgroups of cardiac patients.

Page 22: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

21

Table 1 – Relationship between the specific challenges of patients in whom CCTA imaging is

difficult and the key CT scanner technical requirement (*criteria as defined in NICE, 2012).

Specific challenges/Patient subgroup Key technical requirements

High heart rate (> 65 bpm*) High temporal resolution

Arrhythmia (irregular cardiac motion) Fast volume coverage & high temporal resolution

High calcium score (Agatston > 400*) High spatial resolution

Obesity ( >30kg/m2*) High x-ray flux

Stents High spatial resolution

Grafts Fast volume coverage & high spatial resolution

In chapter 4 the principles of CCTA scanning are introduced with particular focus on the

various scan modes.

Chapter 5 focuses on describing radiation dose from various scan protocols and of the

technical features available for dose reduction.

Chapter 6 introduces CT scanners suitable for CCTA that are currently available on the

market. Data for each key technical specification is presented in tables.

In chapter 7 the approaches adopted by the manufacturers to meet the challenges posed

by the ’challenging to image’ patient subgroups are discussed.

Relevant clinical studies demonstrating the impact of technology developments and

comparing alternative approaches to meet the challenges are referenced in the text and

summarised in appendix 1.

Page 23: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

22

4. Principles of CCTA scanning

CCTA requires intravenous injection of an iodine-based contrast medium to enhance the

visualisation of the coronary arteries followed by CT scanning at high rotation speed with

concurrent monitoring of the patient’s ECG signal.

As discussed in chapter 3 imaging the heart poses various challenges and currently

manufacturers offer a variety of equipment and imaging modes suitable for performing

CCTA. The following brief overview of the main components of a CT scanner and most

common scan modes used in CCTA is provided to facilitate clarification and

understanding of subsequent discussions.

4.1 Essential components of a CT scanner and key

scanner parameters

During a cardiac CT scan, the x-ray tube and detector assembly rotates around the

patient delivering a radiation beam that is selectively attenuated by the patient’s anatomy

according to its composition and thickness. The attenuated beam emerges through the

patient and interacts with a detection system composed of thousands of individual x-ray

detector elements. The x-ray energy absorbed in each individual detector is converted

into an electrical signal that is spatially codified, amplified and transmitted to a computer.

Signal data are then processed by dedicated algorithms to reconstruct an image of the

patient anatomy.

Three key scanning parameters that affect image quality and patient dose in CT

scanning are tube current (mA), tube potential (kV) and x-ray exposure time (as defined

by the gantry rotation time). A brief description of each parameter follows:

Tube current (mA): affects the quantity of photons produced by the x-ray tube.

Scanning at higher mA results in a proportional increase in radiation dose to the patient

and reduced image noise (assuming all other parameters are held constant).

Tube potential (kV): affects both the energy and the total quantity of photons produced

by the x-ray tube. Scanning at higher kV increases the radiation dose to the patient and

reduces image noise (assuming all other parameters are held constant).

Decreasing the kV increases iodine enhancement and can result in an improved CNR.

Gantry rotation time (ms): is the time taken for the scanner gantry to complete a 360o

rotation. Shorter gantry rotation times offer a better temporal resolution thereby reducing

the possibility of coronary motion artefacts.

Page 24: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

23

The product of tube current and gantry rotation time is ‘mAs per rotation’. Higher mAs

values result in higher dose but lower image noise (assuming all other parameters are

held constant).

Modern CT scanners offer features to vary the tube current and tube potential

automatically to optimise dose and image quality.

Figure 1 illustrates essential components of a CT scanner system.

The spatial resolution of a scanner fundamentally depends on the focal spot size as well

as the size of the detector elements in the x-y plane and the z-axis direction.

(a) (b)

Figure 1 - Schematic of CT scanner main components (a) in x-y plane and (b) in z-axis direction.

Diagnostic CT scanners currently in clinical use in the UK are all multi-detector row CT

(MDCT) scanners, and the current standard for performing CCTA is considered to be

scanners with a minimum of 64 slices(4) (Mark et al, 2010) (BSCI, 2012). Figure 2 shows

the different z-axis detector array configurations on the CT scanners discussed in this

report.

4 It should be noted there is a distinction between the specification of a CT scanner in terms of

slices (or channels) and that given in terms of detector rows (or banks). This will be discussed

further in Section 6.3.3.

x-ray fan beam (Scan FOV~50 cm)

x-ray tube

detectors elements

beam shaping (bow-tie) filter

x-y plane z

x-ray beam

Multiple detector rows

Page 25: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

24

Figure 2 - Schematic of detector array configurations for state of the art multi-detector row CT

(MDCT) scanners offered by the main manufacturers. The number of detector rows is shown as

well and the maximum length of the detector in the z-direction (e.g. Philips iCT Elite has 128

individual detector elements each 0.625 mm in size resulting in a total z-axis length of 80 mm).

Somatom Definition Flash Stellar and Somatom Force have two sets of detector systems, each

with the configuration shown.

4.2 Scan modes used in CCTA

In CCTA the patient’s ECG-signal (Figure 3) is monitored throughout the scan and

synchronised with the CT data acquisition so that gated CCTA images can be

reconstructed at a selected phase of the cardiac cycle, ideally the phase in which there is

least motion.

38.4 mm

64 x 0.6 mm

160 mm

320 x 0.5 mm

Toshiba Aquilion One & One Vision

80 mm

128 x 0.625 mmPhilips Ingenuity & IQon

Siemens Somatom Definition Flash Stellar,

Definition Edge Stellar & Definition AS+

GE Revolution CT

256 x 0.625 mm

160 mm

Siemens Somatom

Force

57.6 mm

96 x 0.6 mm

GE CT750 HD, Revolution GSI & Revolution HD

64 x 0.625 mm

40 mm

Toshiba Aquilion Prime

40 mm

80 x 0.5 mm

Philip iCT Elite

64 x 0.625 mm

40 mm

Page 26: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

25

Figure 3 - Representation of the ECG wave form and the phases of the cardiac cycle. Adapted

from http://zone.ni.com/reference/en-XX/help/373698A-01/bioapps/hrv_analyzer/ , accessed on

31 March 2014)

CT scanning can be broadly categorised into two scan modes. The first is axial (or

sequential) scan mode where the patient couch remains stationary throughout a gantry

rotation while a single slab of anatomy is imaged. Unless the x-ray beam is wide enough

to image the whole cardiac anatomy in a single rotation, the table is then moved so that

the adjacent slab can be imaged. This process is repeated until the required craniocaudal

scan length (z-axis) of the patient’s anatomy is covered.

The second mode is helical (or spiral) scan mode where the couch moves through the

gantry bore in the z-direction during x-ray exposure. The speed at which the couch

moves through the gantry relative to the z-axis x-ray beam width defines the ‘pitch’ of the

helical scan (the ratio of the table feed per rotation to x-ray beam width). Scanning at

higher pitches results in more anatomy covered per rotation.

Page 27: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

26

In CCTA these two scan modes can be further subdivided, and the various modes are

summarised in

Figure 4 and Table 2 .The advantages and disadvantages of these scan modes are given

in Table 3.

C. Retrospectively ECG gated helical (PGH) scan mode

(i) Without ECG-gated tube current modulation

(ii) With ECG-gated tube current modulation

D. Prospectively ECG-triggered helical (PTH) scan mode

(i) Low pitch

(ii) High pitch

Page 28: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

27

C. Retrospectively ECG gated helical (PGH) scan mode

(i) Without ECG-gated tube current modulation

(ii) With ECG-gated tube current modulation

D. Prospectively ECG-triggered helical (PTH) scan mode

(i) Low pitch

(ii) High pitch

A. Prospectively ECG-triggered axial (PTA) scan mode

(i) No padding

(ii) With padding

B. Prospectively ECG-triggered axial ‘Volume’ (PTAV) scan mode

(i) No padding

(ii) With padding

Page 29: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

28

Figure 4 Scan modes employed in coronary CT angiography A & B show axial (sequential) scan modes

and C & D show helical scan modes

Page 30: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

29

Table 2 - Description of scan modes used in CCTA.

Scan mode Description

A Prospectively ECG-triggered axial

(PTA) (Fig. 4A)

The scan mode is based on prospective sampling of the patient’s ECG to predict the

optimal phase of the cardiac cycle for data acquisition, determined by the heart rate.

The x-rays beam is triggered at the pre-defined phase of the cardiac cycle and x-ray

exposure only occurs during the period required for data acquisition. The x-ray

exposure window may be extended to cover a longer phase of the cardiac cycle

(referred to as padding) to allow for a greater range of cardiac phases available for

image reconstruction.

Following the exposure of a slab of the cardiac volume, the couch position is

incremented during the subsequent heartbeat, and then the adjacent slab is exposed

during the same cardiac phase as previously. This process is repeated over various

beats until the whole cardiac volume is imaged.

Prospectively ECG-triggered axial

Volume (PTA V) (Fig. 4B)

This mode is similar to the PTA mode described above but no couch incrementation

is required. It is a mode available on CT scanners with an x-ray beam of sufficient

width to acquire the whole cardiac volume in a single rotation within one heartbeat.

Similarly to PTA this mode can operate with or without padding. This scan mode has

a lot of flexibility; the whole cardac cycle can be exposed with or with ECG-gated tube

current modulation. Alternatively data from multiple heartbeats can be acquired and

combined for improved temporal resolution.

Helical

Low pitch (~0.15 – 0.3)

Retrospectively ECG-gated helical

Without or with ECG gated tube

current modulation (RGH) (Fig. 4C)

In this mode the cardiac volume is continuously irradiated while the couch moves at a

low pitch through the gantry bore. Data are acquired throughout the whole cardiac

cycle for image reconstruction in any cardiac phase.

More commonly, this mode is employed with ECG-gated tube current modulation.

The ECG signal is used to select the cardiac phase for viewing the coronary arteries

and outside this phase the mA is reduced The width of the maximum mA window can

be increased for more flexibility of cardiac phases available at full image quality.

There is also flexibility in the % reduction of mA outside the maximum mA window.

Low pitch (~0.15 – 0.3)

Prospectively ECG-triggered helical

(PTH – low pitch) (Fig. 4D(i)

This mode is similar to prospectively ECG-triggered step-and-shoot mode in that

irradiation occurs only during a predetermined phase of the cardiac cycle, but with the

couch continuously moving through the gantry at a low pitch. Padding can be used for

greater flexibility in cardiac phases available for image reconstruction.

High pitch ( >3)

Prospectively ECG-triggered helical

(PTH – high pitch) (Fig. 4D(ii)

This mode is only available on dual source CT scanners. The couch moves

continuously at a high pitch through the gantry and the x-rays are triggered at a pre-

defined phase of the cardiac cycle. The whole cardiac volume is acquired within a

single heartbeat.

Page 31: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

30

Table 3 - Advantages and disadvantages of various CCTA scan modes. Qualitative comparisons of dose are given. They provide guidance of relative dose in different scan modes for a given manufacturer, assuming the same patient characteristics

Scan mode Advantages Disadvantages

Axial

Prospectively ECG-triggered axial

(PTA)

Significantly lower radiation dose than

retrospectively gated helical.

Elimination of helical artefacts where

relevant.

Not recommended for high heart rates

Functional cardiac data not usually

available.

Slightly longer scan time than

equivalent helical scan.

Prospectively ECG-triggered axial

Volume

(PTA V)

Single heart beat acquisition eliminates

misregistration and step artefacts

particularly for irregular heart rhythm.

Short overall scan time.

Elimination of helical artefacts where

relevant.

Use of multisegment reconstruction can

improve temporal resolution

Low dose mode

Cone beam artefacts may be an issue.

Helical

Low pitch (~0.15 – 0.3)

Retrospectively ECG-gated helical

Without ECG gated tube current

modulation

(RGH)

Functional data available.

Use of multisegment reconstruction can

improve temporal resolution

Highest radiation dose.

Low pitch (~0.15 – 0.3)

Retrospectively ECG-gated helical

with ECG-gated tube current

modulation

(RGH + ECG mA modulation)

Functional data available.

Where a low minimum mA is available

in ECG-gated tube current modulation,

doses can approach PTA scans.

Increased radiation dose.

Low pitch (~0.15 – 0.3)

Prospectively ECG-triggered helical

(PTH – low pitch)

Shorter overall scan time than PTA

mode scan with equivalent radiation

dose.

Functional cardiac data not available.

High pitch ( >3)

Prospectively ECG-triggered helical

(PTH – high pitch)

Single heart beat acquisition eliminates

misregistration and step artefacts

Short overall scan time.

Low dose mode.

Not suitable for high heart rates due to

cardiac phase difference in images

acquired within one heartbeat.

Functional cardiac data not available.

Page 32: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

31

5. Radiation dose in CCTA

MDCT has been one of the most significant areas of progress in medical imaging over

the past 10 years. Manufacturers of CT scanners have invested effort in research and

development of CT technology and each generation of scanners has offered improved

hardware and software features aimed at increasing the accuracy of cardiac imaging.

As discussed in the following chapters the major advantages of modern CT scanner

systems are high temporal resolution and high spatial resolution combined with fast

volume coverage. These features meet the challenges of cardiac imaging and provide

fast acquisition of non-invasive images of the heart and coronary arteries with

exceptionally good quality and accuracy. However, the radiation dose to patients

undergoing CCTA examinations is an additional challenge that needs to be considered.

Dose in CCTA depends on various factors such as the type of scanner and dose

reduction features available, the protocol used, including patient preparation, and the

patient body habitus and stage of disease.

Major contributions to dose reduction have been provided by manufacturers of CT

scanners and also by users and clinical research groups who have tested the available

tools and further developed optimised protocols for CCTA. Important progress has been

achieved in reducing dose in CCTA with manufacturers claiming dose reductions in

CCTA down to sub milliSievert levels with no detrimental effects on image quality. Figure

5 shows example CCTA images with diagnostic quality obtained at standard and ultra-

low dose. Despite this, doses to patients in the ‘challenging to image’ subgroups can be

higher than average as the lowest dose protocols cannot always be used.

In this report a comparison of patient doses for the different scanner models has not

been carried out. It was considered that it would not be possible to perform a fair inter-

system comparison using evidence from the literature or from information provided by the

manufacturers. For this reason the remainder of this section focuses on describing the

effect on dose from various scan protocols and on the technical features available for

dose reduction, but makes no comparisons between doses achieved on different

systems.

Page 33: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

32

Figure 5 Three-dimensional volume rendered images from standard (A-C) and ultra-low dose (D-

F) coronary computed angiography of the same patient acquired with 1.2mSv and 0.2mSv. CCTA

was performed on a 64 slice CT scanner (Discovery CT750 HD, GE Healthcare) using prospective

ECG-gating. Adaptive statistical iterative reconstruction (ASIR; GE Healthcare) was used to

reconstruct standard dose images and ultra-low-dose images were reconstructed using a new IR

algorithm (MBIR5; GE Healthcare). The effective radiation dose from CCTA was calculated as the

product of the dose-length product (DLP) and a conversion coefficient for chest [k=0.014

mSv/(mGy.cm)] (Fuchs et al, 2014).

There is a growing body of evidence that the appropriate use of dose reduction strategies

in CCTA can result in substantial dose reduction whilst maintaining diagnostic image

quality. Some of these strategies are available to users on all manufacturers’ CT scanner

systems, whereas a subset is manufacturer specific.

5 It should be noted that MBIR (known as Veo on commercial GE systems) is not currently

available in cardiac scanning on commercial GE CT scanners.

Page 34: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

33

5.1 Justification of the CCTA examination

The first step to limit the radiation burden of CCTA is to adhere to appropriate use criteria

and guidelines for the procedure (Taylor AJ, 2010). The CCTA examination is generally

recommended for diagnosis and risk assessment in patients with low or intermediate risk

or pre-test probability of coronary artery disease. It is also appropriate for structural and

functional evaluation.

NICE recommends CCTA (using a 64-slice scanner or above) to assess arteries and

identify significant stenosis in people with an estimated likelihood of coronary artery

disease (CAD) of 10–29% and a calcium score of 1–400 (NICE, 2010). CCTA is

recommended as an option for first-line evaluation of disease progression, to establish

the need for revascularisation, in people with known CAD.

Various other indications have been investigated and shown to have great promise such

as the follow-up of bypass grafts and the evaluation of ventricular function and cardiac

valves. New indications are currently being investigated to extend cardiac CT to the

assessment of myocardial perfusion and viability. CCTA is not recommended as a first

line option in situations in which immediate revascularisation is being considered and

should not be used as a screening tool.

5.2 Limiting Z-axis scan length

A simple and effective way to reduce radiation dose associated with cardiac CT is to limit

the coverage of the anatomy along the craniocaudal axis (z-axis) to the minimum

required. The extent of the region of interest of the coronary arteries needing to be

imaged may be selected (and reduced) using, for example, calcium scoring images

acquired before the contrast-enhanced CCTA examination.

5.3 Use of small scan field of view

In CCTA protocols a small scan field of view is frequently specified as this utilises the

‘small’ beam shaping filter if one is available. These beam shaping filters have a greater

thickness of attenuating material in the periphery of the full field of view, thereby reducing

dose outside the central region of interest.

5.4 Tube current modulation

The main principle of tube current modulation (TCM) techniques is to vary the tube

current (mA) depending on patient attenuation characteristics. The mA modulation can

be done across the patient anatomy or in specific intervals of the cardiac cycle (temporal

Page 35: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

34

modulation), or both. Currently, manufacturers provide a range of tube current

modulation techniques and an understanding of their operation principles and pitfalls is

essential for their proper use.

5.4.1 Automatic (spatial) tube current modulation (ATCM)

In this approach, rather than delivering a constant mA throughout the entire scan, the mA

is varied according to body size, shape and attenuation properties of the anatomy.

Typically, mA is increased for thicker (more attenuating) regions of the anatomy

compared to thinner (less attenuating) regions. The tube current can be modulated in the

x-y plane (angular modulation), the z-axis (longitudinal modulation), or a combination of

the two.

To produce images with more uniform noise levels, and consequently adequate

diagnostic quality at reduced dose some form of image quality index has to be defined in

the protocol by the user. This image quality index is termed “noise index” by GE,

“mAs/rotation” or “dose index” by Philips, “quality reference mAs” by Siemens and

“standard deviation” by Toshiba.

All manufacturers enable ATCM on their scanners (termed AutomA/SmartmA by GE,

DoseRight by Philips, CAREDose 4D by Siemens and SureExposure by Toshiba). GE

currently does not allow automatic tube current modulation for CCTA scans on the

Optima 660 or Discovery CT 750HD scanners models, but currently only on the

Revolution CT. On the former two scanners, mA and kV are set from a user prescribed

mA/kV table through CT localiser radiograph scout-based BMI or through user input BMI.

Most ATCM techniques are based on calculations obtained from the CT localiser

radiograph and when using this technique it is very important to ensure that the patient’s

anatomy is well centred (at the isocentre) in the gantry.

The appropriate image quality may vary according to patient size, anatomic region, tube

potential, and diagnostic task and the image quality index should be set accordingly.

Recommended indices are usually provided by the manufacturer, and optimisation

studies are important to investigate their adequacy for non-standard patients (e.g. those

with a high BMI). Considering that user input is required, this technique is in fact not truly

“automatic” and its outcomes are significantly dependent on the user’s understanding

and experience of the technique.

Page 36: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

35

5.4.2 Temporal (ECG-gated) tube current modulation

The general principle of retrospectively ECG-gated helical (RGH) scan acquisitions with

ECG-gated tube current modulation consists of decreasing the tube current (mA) outside

the cardiac phase in which data for reconstructing the coronary arteries are acquired.

Accurate evaluation of coronary arteries is typically dependent upon diagnostic

information obtained in late diastole (usually at 70% or 75% of the R-R interval) however,

at higher heart rates, the end systole is often the most diagnostic and motion-free phase

for evaluation of the coronary arteries (Hausleiter et al, 2006). Using this knowledge,

temporal tube current modulation techniques based on the patient’s ECG, have been

developed with the aim of reducing dose to the patient.

The user can define the percentage level of mA reduction and the width of the maximum

mA window within the cardiac cycle (Figure 6). The mA can generally be reduced to 20%

of the maximum mA value set, or on some scanners, down to 5%. Individual patient

characteristics (e.g. heart rate, arrhythmia) influence the adequacy and effectiveness of

this technique to reduce patient dose. It is more efficient in patients with lower and stable

heart rates as lower levels of mA and narrower window widths can be used.

For patients with a high heart rate (e.g. > 80 bpm) and unstable heart rhythm RGH with

ECG-gated tube current modulation is usually not recommended because CCTA images

may need to be reconstructed outside this phase.

Studies using ECG-gated tube current modulation reported overall potential dose savings

of approximately 20% to 50%, depending on the exact protocol used (Entrinkin et al,

2011).

Figure 6 - Diagram illustrating ECG-based modulation of the X-ray tube current. The width of the

temporal window with maximum tube current (Max mA) can be selected by the user, while the

temporal width of the image reconstruction window is fixed. The level of the minimum mA can also

be varied down to 5% 0r 20% depending on the scanner model.

Max mA

Time (s)

Min mA

X-rays ‘on’

Image reconstruction phase

Page 37: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

36

5.5 Optimisation of tube potential

It has been routine to scan adult patients using a tube potential of 120 kV. The principle

behind the benefit of lower kV in some clinical applications is that the superior

enhancement of iodine achieved at lower energies potentially results in an improved

contrast to noise ratio. At the same time, if all other parameters are held constant, a

reduction in tube potential from 120 kV to 100 kV results in a reduction in dose of

approximately 30%. This has promoted the investigation of scanning at lower tube

potentials.

Reducing tube potential from 120kV to 100kV has been recommended in CCTA primarily

in non-obese patients (body weight ≤85 kg or BMI ≤30) due to the increase in image

noise above acceptable levels in larger patients. (Hausleiter et al, 2006) (Kalendar et al,

2009) (Blankstein et al, 2011).

A non-randomised study of CCTA with DSCT on 294 patients (both prospectively ECG-

triggering and retrospectively ECG-gating protocols were included) also showed that

lowering the kV from 120kV to 100kV resulted in a substantial reduction in radiation dose

of approximately 45% (Figure 7) with no compromise in diagnostic performance. Despite

higher image noise being observed in the images produced at 100 kV compared to 120

kV, the contrast-to-noise ratio (CNR) and the signal-to-noise ratio (SNR) (Table 4) were

considerably higher at the lower potential.

Figure 7 - Radiation dose of CCTA using 100 kV versus 120 kV. The observed dose reduction was

statistically significant for all scans acquired with a retrospectively gating and prospectively

triggering acquisition protocols (Blankstein et al, 2011).

Page 38: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

37

Table 4 - Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) was calculated using the

mean density of the contrast-filled left ventricular chamber and mean density of the left ventricular

wall (adapted from Blankstein et al, 2011).

Parameter 100 kV 120 kV p-value

CNR* 6.9 6.0 0.0176

SNR 9.4 8.3 0.0244

At present there is growing interest in lowering the kV even further, to 80 kV in paediatric

patients and in selected groups of smaller adult patients. Promising results have been

reported. Jun et al, 2012 showed a 70% and 88% dose reduction with an 80 kV protocol

and with 80 kV and ECG-based tube current modulation, respectively, compared to the

120 kV protocol in adults with normal BMIs. Importantly, the benefits of lowering the kV in

CCTA from 100 kV to 80 kV have been demonstrated across multiple CT systems

(Labounty et al 2011). Additional benefits of using 80 kV protocols include the potential to

significantly reduce contrast agent doses in CCTA (Cao et al, 2014).

The methods reported for selection of tube potential for individual patients vary but

usually include a metric of body habitus (BMI, physical inspection by a clinician and/or

visualization of scout and axial bolus test images prior to the scan). This is a subjective

process and it is important to develop more objective recommendations for selection of

tube potential as this seems to be a key aspect to promote the effectiveness of the

technique.

All Siemens cardiac scanners models included in this report have automatic kV selection,

CAREkV, available. More recently GE and Toshiba have also introduced automatic kV

selection on their scanners. These are named kV Assist and SUREkV respectively.

iPatient is available on all Philips scanners which have 64 slice capability or greater

utilizing Exam Cards claimed to facilitate patient-specific dose management.

Further evidence is required to better understand the utility and limitations of using lower

kV techniques in CCTA and the potential additional benefits of combining this with other

dose reduction techniques.

Page 39: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

38

5.6 Prospectively ECG-triggered scan modes

As described in section 4.2 various scan modes are available for performing CCTA scans

and the advantages and disadvantages of the various modes were summarised in Table

3.

In general, prospectively ECG-triggering, whether in axial or helical scan mode (Figure 4

A, B and C) will result in a significant reduction in radiation dose compared to

retrospectively gated scan mode (Figure 4,C) as only the cardiac phase to be used for

image reconstruction is irradiated. In the past these scan modes were limited to patients

with low heart rates, but with increases in temporal resolution and z-axis detector

coverage the limiting heart rate for prospectively ECG- triggered scan modes is being

raised.

An important consequence of the prospectively ECG-triggered techniques is that the

radiologist may be entirely dependent on images acquired from a single cardiac phase to

make the diagnosis as no data are acquired in the other phases. However, users can

define an additional amount of exposure time (padding) just before and just after the

selected phase that allows for greater flexibility in phase reconstruction. Padding results

in additional exposure to the patient (Figure 4 A ii).

In prospectively ECG-triggered axial mode (PTA) all cardiac CT scanners provide a user-

definable padding parameter. For example if data for image reconstruction can be

obtained in 220 ms and the user defines 50 ms of padding the total exposure time will be

320 ms (padding is added before and after the selected cardiac phase). If motion

artefacts are present in the primary phase selected for reconstruction, the other phases

may have no such limitations and may therefore provide diagnostic information. The

amount of padding should be selected taking into consideration the patient

characteristics and diagnostic task. Low heart rates are essential if using little or no

padding. A systematic review showed that use of PTA leads to a significant reduction in

radiation dose compared to RGH while offering comparable image quality and diagnostic

value (Sun and Ng, 2012).

Scanners with a detector extent of 160 mm in the z-axis generally employ the so-called

Volume scan mode. This is a fundamentally similar approach to conventional PTA

scanning but the whole cardiac anatomy is generally acquired within one heart beat with

no requirement for couch incremantation to cover successive portions of the heart. This

mode results in a very short scan time and does not suffer from banding artefacts due to

spatial misregistration of data from successive heartbeats.

The high pitch, prospectively ECG-triggered helical (PTH) mode involves scanning at a

pitch of approximately 3.4 which is a factor of more than 10 times higher than pitch

values typically used in retrospectively ECG-gated (RGH) mode (ranging from ~ 0.15 to

0.3). This allows the whole cardiac volume to be acquired within a single heartbeat

(Figure 4 D ii) and Figure 8). The implementation of this technique requires a dual source

Page 40: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

39

CT scanner with a table technology that allows extremely fast movement and specific

reconstruction algorithms that combine data acquired by the two sets of detectors. To

achieve the required image quality this scan mode is generally limited to a low, regular

heart rate (stable HR ≤60 bpm) typically requiring the administration of beta blockers

(Entrikin et al, 2011).

.

Figure 8 - Prospectively ECG-triggered helical acquisition at high pitch minimises dose through a

combination of extremely short exposure time (280 ms in the example), dynamic pre-patient

collimation (eliminating overscanning) and near complete elimination of overlap in the cone beam

thereby minimising extraneous patient exposure (adapted from Entrinkin et al).

One of the most recently introduced scan modes is prospectively ECG-triggered helical

mode at a low-pitch (ranging from 0.15 to 0.3) with radiation exposure only occurring

during the phase required for imaging the coronary arteries. Similarly to padding in PTA

mode, the width of the exposure window can be increased to allow for more flexibility in

the cardiac phase available for image reconstruction. Toshiba is currently the only

manufacturer that enables this CCTA scan mode. However, no evidence was identified in

the literature search undertaken on clinical use of this mode.

5.7 Development of standardized protocols

The techniques briefly described above offer opportunities for dose reduction in CCTA.

However it is through the combination of various strategies that the greatest potential for

dose reduction can be achieved.

Page 41: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

40

LaBounty and colleagues compared the use of standardized and non-standardised

CCTA protocols at three sites equipped with single-source 64-slice CT scanners (GE

750HD) (LaBounty et al, 2010 - a). Standardized protocols considered patient heart rate,

BMI and clinical indications to systematically define the use of prospective (versus

retrospective) triggering, tube potential, tube current, padding duration and ECG-based

tube current modulation. The scanners offered iterative reconstruction (IR) but this was

not used. Therefore, the results are more generalizable to others scanners.

The standardized protocols resulted in an overall reduction in estimated effective dose of

65% when compared with the non-standardized approach. Additionally, no differences in

the sensitivity, specificity and accuracy were found between the two protocols.

The authors recommended that tube current should be specified based on chest wall

diameter, BMI or estimated noise measures and should vary on a case by case basis.

Tube potential should also be based on a patient’s BMI, and z-axis coverage should be

determined by the coverage clinically required. The authors proposed standardization of

scan parameters (Table 5) for CCTA on patients with controlled heart rates (≤65 bmp) for

use on most 64-MDCT scanners with cardiac capabilities. It should be noted that whilst

the methodology reported in this study may be applicable to the optimisation of protocols

on other CT scanner models the specific optimised protocol parameters must not be

transferred to other models without validation. In particular the tube current (mA) settings

will vary between scanner models.

The above approach is recommended on scanners that do not have the capability for

automatic tube current modulation and/or automatic tube potential selection. The

automated approach is considered to be superior to manual selection.

Table 5 – Example of a proposed standardised CCTA protocol. The authors of the study

recommend that prospectively ECG-triggered axial (step and shoot) should be utilised in all

patients with regular heart rates (≤65 bpm). If heart rate is consistently ≤ 55 bpm and regular, use

of padding is not recommended. *80kV could be considered in patients with a small thoracic

diameter and body mass index <25 kg/m2. PSS – prospective step and shoot (or PTA, using the

terminology adopted in this report) (adapted from LaBounty et al, 2010 a).

Page 42: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

41

Figure 9 - Examples above are curved multiplanar reformats through the right coronary artery

generated from PSS (i.e. PTA) scans obtained on a 64-slice scanner without the use of IR

techniques in three different patients. Both A and B were scanned with identical imaging

parameters. However, in B the entire desired z-axis coverage was obtained in three axial steps

resulting in nearly 25% lower dose compared with A. In C, the sub milisievert dose estimate was

achieved through a combination of PSS technique, decreased mA, no padding and decreased

scan length (shows the improvement in image quality following the application of a standard dose

reduction protocol) (Entrinkin et al, 2011).

As discussed, a good understanding of the impact of scan parameters and scan modes

on image quality and patient dose is important to optimise CT protocols.

Centres that wish to optimise their protocols may start by establishing standard protocols

for groups of patients with similar characteristics. A further step in the optimisation

process is tailoring the CT scanner parameters to the individual characteristics of each

patient. To help users in this task, manufacturers provide features on their scanners to

perform automatic selection of various key scanning parameters and scanning modes

based on the characteristics of individual patients. Examples of these features are:

kV selection according to patient size and clinical application;

mA selection and mA modulation according to patient attenuation;

ECG-triggering phase according to heart rate;

Width of maximum mA window in RGH mode according to heart rate;

Amount of padding according to heart rate.

Automatic selection of scanning parameters in CCTA may be particularly useful for

patients presenting with non-standard characteristics (e.g. high BMI, HR pattern)

providing images with adequate quality consistently, whilst taking patient dose into

consideration.

Page 43: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

42

5.8 Iterative reconstruction (IR)

Traditionally filtered back projection techniques have been employed to transform the

raw image data obtained on a CT scanner into a final reconstructed image. However,

these techniques employ various assumptions regarding the underlying physics

processes and CT geometry that typically result in increased noise and less accuracy in

the reconstructed images (Entrinkin et al, 2011). In contrast IR techniques use a

statistical model of photon noise and may also include modelling of scanner

characteristics to allow a more accurate integration of the physics and scanner geometry

(Moscariello et al, 2011).

A drawback of IR techniques is the computer power and processing time required which

have the penalty of significantly longer reconstruction times and have delayed

introduction of IR in clinical routine.

Despite these challenges all manufacturers have succeeded in developing simplified IR

techniques feasible for clinical use and they are currently available on every CT scanner

platform. The original algorithms used have now been further developed and provide

improved performance in terms of dose and artefact reduction.

IR itself does not reduce patient dose. However, the noise reduction it provides allows for

the reduction of scan parameters (e.g. mAs, kV) during image acquisition allowing patient

dose savings.

Each manufacturer has developed specific IR algorithms that operate in different ways

and are implemented at different stages of the reconstruction process. Additionally

different ‘generations’ of IR algorithms are available from each manufacturer. It is

important that the radiologist and CT scanner operator have an understanding of how IR

works on the scanner in question, and that they are satisfied with the resulting image

appearance that may differ from that with which they are familiar.

Manufacturers often use blending of iterative reconstruction and FBP to provide images

that are more clinically acceptable. For example, a high percentage of IR may be used

when scanning obese patients and the mAs is kept constant so that image noise is

reduced without an increase in dose. Alternatively, when scanning patients with normal

or low BMI a certain percentage of IR is used and the mAs reduced so that image quality

is maintained at a decreased radiation dose.

Page 44: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

43

5.9 Z-axis overbeaming and overscanning

Overbeaming is the additional dose beyond the edge of the detector rows of a MDCT,

which is required because the x-ray beam penumbra must lie outside the active detector

length (Figure 10). With wider z-axis x-ray beam coverage, the percentage dose

contribution from this penumbra region decreases therefore the contribution of this effect

to unnecessary patient dose has reduced on modern scanners with wide z-axis beam

coverage.

Figure 10 - Diagram illustrating overbeaming, overscanning and adaptive section collimation

technology during helical (spiral) scanning (adapted from Goo, 2009).

Overscanning (also known as overranging) is applicable to helical scan mode where

extra rotations are required beyond the planned scan length to reconstruct the first

and the last images (Figure 10).

To reduce the dose to the patient due to overscanning, some manufacturers have

incorporated adaptive dynamic collimation on their systems (also referred to as

adaptive collimation). This feature involves the use of collimators that move

automatically in the z-axis into and out of the x-ray beam at the beginning and ending

of a helical acquisition and by doing so filter unnecessary radiation (Figure 11).

Page 45: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

44

Figure 11 - Illustration of (a) conventional and (b) adaptive section collimation CT scanning

protocols. For adaptive section collimation the shape of the x-ray cone beam at the beginning and

end if spiral acquisition is controlled by two collimators made of absorbent material (adapted from

Deak et al, 2009).

Overscanning is independent of the planned scan length and is proportional to beam

collimation, reconstructed slice width, and pitch. The contribution of this effect to the

total patient dose is therefore more important for shorter scan ranges such as are

used in cardiac CT (and paediatric CT) (Deak et al, 2009).

Page 46: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

45

6. CT scanners for cardiac imaging

6.1 CT scanner models included in the report

A questionnaire (appendix 2) was sent to the four major CT scanner manufacturers

requesting specification data on CT scanner models considered suitable for performing

CCTA in the ‘challenging to image’ patient groups defined in section 3.1.

The CT scanner manufacturers approached are listed alphabetically below and the order

does not reflect the merits of the systems. This approach is used throughout the report

GE Healthcare, Chalfont St Giles, Buckinghamshire, UK

Philips Healthcare, Guildford, Surrey, UK.

Siemens Healthcare, Frimley, Surrey, UK.

Toshiba Medical Systems, Crawley, West Sussex, UK.

Each manufacturer responded with between two and four scanner models of varying

specifications (6). Responses on eleven scanner models were received from the

manufacturers and these are tabulated in Table 6.

Siemens and Philips made contact shortly before the deadline for submission of the

report with requests to include two additional scanners (Siemens Somatom Force and

Philips IQon Spectral CT). Limited information was provided for these systems and

further clarifications could not be obtained within the timeframe of the project. A short

description of these two scanners is provided and the key specification parameters on

the Siemens Somatom Force and Philips IQon Spectral CT are provided in Table 7 but

these two scanners are not included in further discussions.

All scanners included have at least 64 x-ray detector rows, providing a minimum of

approximately 40 mm z-axis coverage (Figure 1). CT scanner models commonly referred

to as ‘64-slice scanners’ have been shown to give improved accuracy in the diagnosis of

coronary artery disease compared to scanners with fewer ‘slices’ (Mark et al, 2010), and

are now considered as the minimum standard for enabling successful CCTA scans on

patients with a wide range of characteristics.

6 At the time of the exercise KiTEC was informed by some manufacturers that not all data

requested was available, particularly on their most recent systems. Additionally, for some of the

claims made, published peer review evidence is still not available.

Page 47: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

46

6.2 Brief description of scanner models

The key technical specifications of the CT scanners included in this report are shown in

Table 7.

The Siemens Somatom Force scanner and Philips IQon Spectral CT are also listed, as

the manufacturer provided summary data. However, these systems are not included in

the majority of comparison tables and charts as no detailed technical specifications were

available at the time this report was being compiled.

The GE Discovery CT750 HD CT scanner model has been recently replaced by the

Revolution GSI and Revolution HD, models that have very similar technical specifications

to the CT750 HD. The text, and data in the tables and graphs, that refer to the Discovery

CT750 HD, can be taken to also apply to the Revolution GSI & HD.

The scanners in this report all have between 64 and 320 detector rows providing

coverage of between 38.4 mm and 160 mm in the z-axis direction. The z-axis detector

dimension ranges from 0.5 mm to 0.625 mm. Minimum gantry rotation times range from

250 ms to 420 ms. The maximum allowable weight on the couch is 227 kg on some

models and approximately 300 kg on other models, the latter being either standard or

available as an option. Other specifications given in Table 7 are: gantry bore diameter;

maximum scan field of view; x-ray generator power; range of tube potential and tube

current. More detailed comparisons of the performance parameters essential for

successful CCTA scans, particularly in the difficult to image patient groups, are provided

in section 6.3.1.

All the scanners in this report are available with iterative reconstruction options and

automatic tube current modulation to support dose optimisation.

Page 48: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

47

Table 6 - CT scanner models considered suitable for CCTA (as per information provided by the

manufacturers) included in report. The specifications included in the following chapters of the

report are for the newest generation CT scanners.

Scanner model Year of launch Scanner type

GE

Healthcare

Optima 660 2011 3rd generation; 64 detector row

Revolution GSI/HD

(Discovery CT750 HD)

2014

(2008)

3rd generation; 64 detector row:

Revolution GSI & Discovery HD 750

CT – alternating kV systems

Revolution CT 2013 3rd generation; 256 detector row

Philips

Healthcare

Ingenuity 2010 3rd generation; 64 detector row

Brilliance iCT (first launch)

iCT Elite (new generation)

2007

2012 3rd generation; 128 detector row

IQon Spectral CT 2013 3rd deneration; 64 detector row; dual

detector layer system

Siemens

Healthcare

Somatom Definition AS+ 2008 3rd generation; 64 detector row

Somatom Definition Edge

Stellar

2011 3rd generation; 64 detector row

Somatom Definition Flash

Stellar

2011 3rd generation 64 detector row; dual x-

ray source - detector system

Somatom Force 2013 3rd generation 96 detector row; dual x-

ray source - detector system

Toshiba

Medical

Systems

Aquilion PRIME 2012 (first launch)

2013 (new generation)

3rd generation; 80 detector row

Aquilion ONE 2007 (first launch)

2013 (new generation)

3rd generation; 320 detector row

Aquilion ONE Vision 2012 3rd generation; 320 detector row

Page 49: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

48

Table 7 - Key CT scanner specification parameters as per information provided by the

manufacturers.

1 Dual x-ray source- detector systems

Scanner model

Min.

gantry

rotation

time

(ms)

Z-axis

coverage

(mm)

Z-axis

detector

config. (#

x mm)

Bore

dia-

meter

(mm)

Max

scan

FOV

(mm)

Max

weight

on

couch

(kg)

x-ray

generator

power

(kW)

Tube

potential

range

(kV)

Tube

current

range

(mA)

GE

Healthcare

Optima 660 350 40 64 x 0.625 700 500 227 72 80 - 140 10 - 600

Discovery

CT750 HD 350 40 64 x 0.625 700 500

306

(option) 107 80 - 140 10 - 835

Revolution CT 280 160 256 x 0.625 800 500 227 103 70 - 140 10 - 740

Philips

Healthcare

Ingenuity 420 40 64 x 0.625 700 500 295 80 80 - 140 10 - 655

iCT Elite 270 80 128 x 0.625 700 500 295 120 80 - 140 10 - 1000

iQon Spectral

CT

270 40 64 x 0.625 700 500 204 120 80 – 140 10 - 1000

Siemens

Healthcare

Somatom Definition AS+

300 38.4 64 x 0.6 780 500 227 (307

option)

80 (100

option)

70 - 140 20 - 666 Somatom

Definition Edge

Stellar

280 38.4 64 x 0.6 780 500 227

(307

option)

100 70 - 140 20 - 800

Somatom

Definition

Flash Stellar (1)

280 38.4 64 x 0.6 780 500/3

30

227

(307

option)

2 x 100 70 - 140 20 – 800

(per tube)

Somatom

Force (1)

250 57.6 96 x 0.6 780 500/3

50

227

(307

option)

2 x 120 70 - 150 20 – 1300

(per tube)

Toshiba

Medical

Systems

Aquilion PRIME 350 40 80 x 0.5 780 500 300 72 80 - 135 10 - 600

Aquilion ONE 350 160 320 x 0.5 780 500 300 72 80 - 135 10 - 600

Aquilion ONE

Vision

275 160 320 x 0.5 780 500 300 100 80 - 135 10 - 900

Page 50: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

49

6.2.1 GE Healthcare

Optima 660

The Optima 660 is the lowest specification model of the GE scanners included in this

report. It has a minimum gantry rotation time of 350 ms. The detector array consists of 64

x 0.625 mm detector rows providing a z-axis coverage of 40 mm.

Revolution GSI & Revolution HD (replace Discovery CT750 HD)

The Revolution GSI launched in 2014 is a direct replacement for the GE Discovery

CT750 HD. The Revolution HD, also launched in 2014, has the same specifications as

the GSI but without dual energy capability. Both scanner models have new features

including, a touch screen interface in the scanner gantry for improved patient workflow,

cardiac motion correction and software for automatic selection of x-ray tube potential.

Both scanner models have the same minimum gantry rotation time and detector array

configuration as the Optima 660, but a different detector material, Gemstone™, with a

faster response time (Jiang et al, 2008). The scanners can operate in high definition

mode (HD) that provides 2.5 times the number of views per rotation as in its standard

mode. On the Revolution GSI, the faster detector material also enables dual energy

scanning with ‘fast kV switching’, acquiring data at the high and low energies in alternate

views during a gantry rotation.

Revolution CT

The Revolution CT is the highest specification scanner offered by GE. It was launched at

the Radiological Society of North America (RSNA) annual meeting in December 2013.

The system has 256 x 0.625 mm detector rows giving a z-axis coverage of 160 mm

which provides full organ coverage in a single gantry rotation. It uses the same detector

material, Gemstone™, as the CT750 HD and so can achieve the same spatial resolution

through the use of HD mode. However, it is not yet have dual energy capability using

rapid kV switching. The scanner allows a minimum x-ray tube potential of 70 kV

compared to the minimum 80 kV available on other GE systems.

6.2.2 Philips Healthcare

Ingenuity

The Ingenuity CT scanner has a minimum gantry rotation time of 420 ms. The detector

array consists of 64 x 0.625 mm detector rows providing a z-axis coverage of 40 mm.

iCT Elite (Referred to as Brilliance iCT in following figures)

The iCT Elite has a minimum gantry rotation time of 270 ms enabled by the AirGlide

bearing. The detector array consists of 128 x 0.625 mm rows providing a z-axis coverage

of 80 mm. The scanner has z-axis flying focal spot technology enabling the acquisition of

Page 51: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

50

256 slices per rotation. The imaging chain on the iCT Elite is supported by technologies

like the 2nd generation NanoPanel Elite integrated and modular detector, Eclipse

collimation, 2D Antiscatter Grid and iterative algorithms such as iDose4 and model-based

IMR.

IQon Spectral CT

Introduced in 2013, the IQon Spectral CT has a novel dual-layer detector. Dual energy

data can be obtained retrospectively from all scans without the need for special modes.

Due to this detector-based approach there are no special limitations in temporal

registration, field of view, or the use of dose modulation.

IQon Spectral CT is based on the NanoPanel Prism detector which uses an Yttrium-

based top scintillator with 25% higher light output and Elite electronics adapted from the

NanoPanel Elite detector. Philips claim that one of the benefits of a detector-based

approach to Dual Energy CT is that simultaneous detection allows projection-domain

processing without temporal misregistration for high-motion procedures such as cardiac

CT.

The IQon Spectral CT has a minimum gantry rotation time of 270 ms and the dual-layer

detector array provides a z-axis coverage of 40 mm. The IQon includes Philips’ latest

innovations in iterative reconstruction, including IMR (Iterative Modelled Reconstruction).

Due to the incomplete technical data this scanner is not included in further discussions or

comparisons within this report.

6.2.3 Siemens Healthcare

SOMATOM Definition AS+

The Siemens Somatom Definition AS+ is the lowest specification Siemens scanner

included in this report. It has a minimum gantry rotation time of 300 ms. The detector

array consists of 64 x 0.6 mm detector rows giving a z-axis coverage of 38.4 mm. The

scanner utilizes a flying-focal spot in the z-axis (z-sharp) to double-sample the detector

rows, so 128 slices are acquired per gantry rotation. As for all other Siemens scanners

included in this report, this model has a minimum tube potential of 70 kV.

SOMATOM Definition Edge Stellar

The Siemens Somatom Definition Edge Stellar has many design aspects in common with

the Definition AS+ but with some upgraded features. These include a faster minimum

gantry rotation time of 280 ms and Stellar detectors. The Stellar detector combines the

photodiode with the analogue to digital converter (ADC) in one integrated circuit to

reduce electronic noise and provide sharper slice profiles through reduced cross-talk

between neighbouring detector rows (Ulzheimer et al, 2012).

Page 52: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

51

At RSNA 2014 Siemens introduced an upgrade to the Definition Edge Stellar which

enables Dual Energy scanning by splitting the x-ray beam (in the z-axis direction) into a

high and low energy component

SOMATOM Definition Flash Stellar

The Siemens Somatom Definition Flash Stellar is, like its predecessor the Definition

Flash, a dual source CT scanner with two detector-tube assemblies each with 64 x 0.6

mm detector rows, enabling the acquisition of 128 slices per rotation using z-sharp (flying

focal spot) technology, but with the new stellar detector. This scanner also has a

minimum gantry rotation time of 280 ms It has dual energy capability, through operating

the two x-ray tubes at different tube potentials, and allows high helical pitch scanning

(Flash mode) for fast volume coverage. Dual source technology halves the intrinsic

temporal resolution compared to a single source scanner with the same rotation time.

SOMATOM Force

The SOMATOM Force was launched at RSNA December 2013 and is a dual source CT

scanner with two detector-tube assemblies. The system has a reduced gantry rotation

time of 250 ms and therefore an intrinsic temporal resolution of 66 ms. Other upgraded

features include 2 x 96 x 0.6 mm detector rows acquiring 192 slices per rotation using z-

sharp technology, a larger field of view FOV (350 mm) on the second assembly, an x-ray

tube potential range from 70 kV to 150 kV, and a new design of tin filter (Selective

Photon Shield II) which, it is claimed, improves energy separation by 30% for dual energy

scanning.

The scanner also includes a new iterative reconstruction algorithm (ADMIRE) also

available as an upgrade on the Somatom Definition Flash Stellar and Somatom Definition

Edge Stellar models. The system is designed to be a high-end general and cardiac

scanner with a focus on research activities and future applications (e.g. cardiac, dual

energy and preventive medicine).

Due to the incomplete technical data this scanner is not included in further discussions or

comparisons within this report.

6.2.4 Toshiba Medical Systems

Aquilion PRIME

The Aquilion PRIME is the lowest specification Toshiba scanner presented in this report.

It has a minimum gantry rotation time of 350 ms. The detector array consists of 80 x 0.5

mm detector rows providing a z-axis coverage of 40 mm.

Page 53: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

52

Aquilion ONE

The Aquilion ONE has the same minimum gantry rotation time as the PRIME, but its

detector array consists of 320 x 0.5 mm detector rows providing a z-axis coverage of

160 mm. This wide detector coverage enables whole organ coverage in a single gantry

rotation.

Aquilion ONE VISION Edition

The Aquilion ONE Vision edition has many specifications in common with the original

Aquilion ONE but with some upgraded features principally a minimum rotation time of

275 ms and a larger, 100 kW, x-ray generator.

Page 54: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

53

6.3 Comparison of technical specifications

The principal performance parameters considered key in investigations of coronary artery

disease (CAD) in the challenging patient groups are:

Temporal resolution

Spatial resolution

Volume coverage

X-ray flux

This sub-section includes updated comparisons of these performance parameters. The

data presented was obtained from manufacturers’ responses to the KiTEC specification

questionnaire (Appendix 2). The responses were checked for accuracy and consistency

and additional questions were forwarded to the manufacturers where clarification was

required. Where a manufacturer could not provide a full explanation to a given question

within the timeframe of this project this has been highlighted.

6.3.1 Temporal resolution

A high temporal resolution in conjunction with ECG-gating to reconstruct images in the

optimal phase of the cardiac cycle is essential for sharp depiction of the coronary arteries

with minimum blurring from cardiac motion. This requirement is particularly critical in

patients with high heart rates. In recent years manufacturers have placed strong

emphasis on increasing the temporal resolution of CT scanners through both hardware

and software approaches.

The temporal resolution (TR) of a CT scanner can be considered in terms of the intrinsic

TR and the effective TR. In this report the intrinsic TR is defined as the time taken to

acquire 180° of data, the minimum usually necessary for image reconstruction. The

effective TR is that achieved when various methods for improving the intrinsic TR (e.g.

motion correction algorithms, multi-segment reconstruction) are applied. The intrinsic and

effective TR of the scanner models are tabulated in Table 8 and presented graphically in

Figure 12 and Figure 13.

Page 55: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

54

Table 8 - Temporal resolution (TR) for CCTA scans as per information provided by the

manufacturers.

Scanner model

Min.

gantry

rotation

time

(ms)

Intrinsic

TR

(ms)

Maximum

number of

segments -

axial mode

Effective TR -

axial mode (1)

(ms)

Maximum

number of

segments -

helical

mode

Effective TR

- helical mode (1)

(ms)

GE

Optima 660 350 175 1 29 with SSF(2) 4 44 – 175

(29 with SSF(2))

Discovery CT750 HD 350 175 1 29 with SSF(2) 4 44 – 175

(29 with SSF(2))

Revolution CT 280 140 1 24 with SSF(2) N/A N/A

Philips

Ingenuity 420 210 1 210 4 53 - 210

iCT Elite 270 135 1 135 4 34 - 135

Siemens

Somatom Definition

AS+ 300 150 1 150 2 75 - 150

Somatom Definition

Edge Stellar 280 142 1 142 2 71 - 142

Somatom Definition

Flash Stellar (4) 280 75 1 75 2(3)

38– 75 (low pitch)

75 (high pitch)

Toshiba

Aquilion PRIME 350 175 N/A N/A 5 35 - 175

Aquilion ONE 350 175 5 35 - 175 N/A N/A

Aquilion ONE Vision 275 137 5 27 - 137 N/A N/A

1 Range of TRs: minimum is optimal value achieved with maximum available number of segments in multi-

segment reconstruction; maximum value given is for single segment reconstruction 2 SSF – SnapShot Freeze (manufacturer specific term) motion correction software is claimed to remove

blurring due to motion and achieve an effective TR of 29 ms 3 Multi-segment reconstruction is not available in high pitch helical mode 4 Dual source system

N/A Not applicable - scanner not used in this mode for CCTA scans

Page 56: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

55

Figure 12 - Intrinsic temporal resolution (TR) of CT scanners (lower values represent a better

temporal resolution).

^ On GE scanners axial mode TR includes use of SnapShot Freeze algorithm. On Toshiba scanners axial mode TR is the optimal value with multi-segment reconstruction using 5 segments.

* In helical mode the optimal TR with maximum number of segments available is given. On GE scanners helical mode TR is with SnapShot Freeze algorithm.

Figure 13 - Effective temporal resolution (TR) of CT scanners (lower values represent a better

temporal resolution).

6.3.2 Spatial resolution

Spatial resolution is another key parameter for successful CCTA imaging. Intrinsic spatial

resolution values quoted are usually measured in stationary phantoms. However, the

effective spatial resolution will also be dependent on the scanner’s temporal resolution as

this affects the degree of additional blurring due to cardiac motion.

0

20

40

60

80

100

120

140

160

180

200

220

GE Optima

660

GE

Discovery

CT 750HD

GE

Revolution

CT

Philips

Brilliance

Ingenuity

Philips

Brilliance

iCT

Siemens

Definition

AS+

Siemens

Defintion

Stellar Edge

Siemens

Definition

Flash Stellar

Toshiba

Aquilion

PRIME

Toshiba

Aquilion

ONE

Toshiba

Aquilion

ONE Vision

Intr

insi

c TR

(m

s)

0

20

40

60

80

100

120

140

160

180

200

220

GE Optima

660

GE Discovery

CT 750HD

GE

Revolution

CT

Philips

Brilliance

Ingenuity

Philips

Brilliance

iCT

Siemens

Definition

AS+

Siemens

Defintion

Stellar Edge

Siemens

Definition

Flash Stellar

Toshiba

Aquilion

PRIME

Toshiba

Aquilion

ONE

Toshiba

Aquilion

ONE Vision

Effe

ctiv

e T

R (

ms)

Temporal resoution: Axial scan mode^

Temporal resoution: Helical scan mode*

Page 57: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

56

Spatial resolution must be considered in three dimensions, therefore in this section the

axial (x-y plane) and the z-axis direction resolution are discussed separately to highlight

key developments and facilitate comparison of the various systems. Broadly speaking,

intrinsic x-y plane spatial resolution on CT scanners has not increased significantly in

recent years, as it has always been relatively high compared to that along the z-axis.

The emphasis has been on matching the resolution in the z-axis to that in the x-y plane

to achieve isotropic (equal in all dimensions) resolution. Multi-slice scanners, particularly

those with 64 and more detector rows, are largely capable of meeting this aim.

Axial (x-y plane) spatial resolution

Table 9 compares the principal scanner features that affect the limiting x-y plane spatial

resolution. It also presents data for the spatial resolution achieved with two different

reconstruction kernels recommended by each manufacturer: a standard resolution kernel

for general CCTA scans and a higher resolution (sharp) kernel for CCTA in patients with

stents and/or high calcium scores.

The spatial resolution is defined as the minimum distance (mm) between two small, high

contrast objects that allows the two objects to be resolved in the image. They have been

calculated from the 2% modulation transfer frequency (MTF) values provided by the

manufacturer for the two kernels (standard and high resolution). Where the 2% MTF

value was not given, data was interpolated between the 10% and 0% MTF values

provided. This was a practical approach considered adequate to provide a reasonable

estimate and allow comparison of the data. The spatial resolution values are presented

graphically in

Figure 14.

Some manufacturers may recommend a slightly sharper kernel for improved spatial

resolution at the expense of increased noise.

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

GEOp ma660GEDiscoveryCT750HD

GERevolu onCT

PhilipsIngenuity

PhilipsBrillianceiCT

SiemensDefini onAS+

SiemensDefin onStellarEdge

SiemensDefini onFlashStellar

ToshibaAquilionPRIME

ToshibaAquilionONE

ToshibaAquilionONE

Vision

Axialplanespa

alresolu

on(mm)

Standardkernel

Sharpkernel

Page 58: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

57

Table 9 - Spatial resolution in x-y (scan) plane as per information provided by the manufacturer.

Scanner models Focal spot

sizes(1) (mm)

Number detectors per row

Sampling density (No. of views per

360, per slice for min. gantry rotation time without FFS(2))

Sampling density (No. of views per

360, per slice for min. gantry rotation

time with FFS(2))

x-y plane spatial resolution (3)

(standard kernel) (mm)

x-y plane spatial resolution(3)

(high res. kernel) (mm)

GE Optima 660

S: 0.7 x 0.6 L: 0.9 x 0.9

888 861 N/A 0.61 (Standard) N/P.

Discovery CT750 HD S: 0.7 x 0.6 L: 0.9 x 0.9

888 984 2496 (HD mode) 0.61 (Standard) 0.34 (Edge-HD mode)

Revolution CT S: 0.7 x 0.6 L: 0.9 x 0.9

N/P N/P. 2496 (HD mode) N/P 0.34 (Edge-HD mode)

Philips Ingenuity

S: 0.6 x 0.7 L: 1.1 x 1.2

672 2320 N/A 0.63 (standard) 0.48 (high resolution)

iCT Elite S: 0.6 x 0.7 L: 1.1 x 1.2

672 2400 N/A 0.63 (standard) 0.48 (high resolution)

Siemens

Somatom Definition AS+

S: 0.7 x 0.7 L: 0.9 x 1.1

736 1152 N/A in cardiac mode 0.68 (B30f/I30f) 0.53 (B46/I46f)

Somatom Defintion Stellar Edge

S: 0.7 x 0.7 L: 0.9 x 1.1

736 1152 N/A in cardiac mode 0.68 (B30f/I30f) 0.53 (B46/I46f)

Somatom Definition Flash Stellar (4)

S: 0.7 x 0.7 L: 0.9 x 1.1

736 / 480 1152 N/A in cardiac mode 0.68 (B30f/I30f) 0.53 (B46/I46f)

Toshiba Aquilion PRIME

S: 0.9 x 0.8 L: 1.6 x 1.4

896 900 N/A 0.45 (FC03) 0.39 (FC30)

Aquilion ONE S: 0.9 x 0.8 L: 1.6 x 1.5

896 900 N/A 0.45 (FC03) 0.39 (FC30)

Aquilion ONE Vision S: 0.9 x 0.8 L: 1.6 x 1.5

896 786 N/A 0.45 (FC03) 0.39 (FC30)

1 Focal spot sizes quoted according to IEC 336/93. Dimensions for small (S) and large (L) focal spot sizes are given 2 Flying focal spot (FFS) in x-y plane, available in cardiac mode on some scanners, oversamples detector elements for increased spatial resolution. On GE scanners this mode is referred to as High Definition (HD) mode. 3 Calculated from 2% MTF values provided by manufacturers for kernel quoted. High resolution kernels may be recommended for patients with stents or high calcium scores to reduce blooming artefacts. 4 Dual source detector system.

N/P data not provided by the manufacturer; N/A – not applicable

Page 59: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

58

Figure 14 - Axial (x-y) plane spatial resolution of CT scanners (smaller values represent a better

spatial resolution).

z- axis spatial resolution

The z-axis dimensions and z-spatial resolution of the CT scanners are shown in Table 10.

The z-axis spatial resolution is defined primarily by the z detector dimension but can be

enhanced in a number of ways described in Table 10.

As for the x-y plane, z-axis spatial resolution values are quoted in terms of the minimum

distance (mm) between two small high contrast objects that allows the two objects to be

resolved in the image. These values were derived from the 2% MTF values provided by

the manufacturers. Where the 2% value was not given, data was interpolated from the

10% and 0% MTF values provided. This was a practical approach considered adequate

to provide a reasonable estimate and to allow comparison of the data. The specifications

are compared graphically in Figure 15.

Figure 15 - z-axis spatial resolution of CT scanners (lower values represent a better spatial

resolution).

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

GEOp ma660GEDiscoveryCT750HD

GERevolu onCT

PhilipsIngenuity

PhilipsBrillianceiCT

SiemensDefini onAS+

SiemensDefin onStellarEdge

SiemensDefini onFlashStellar

ToshibaAquilionPRIME

ToshibaAquilionONE

ToshibaAquilionONE

Vision

Axialplanespa

alresolu

on(mm)

Standardkernel

Sharpkernel

0

0.1

0.2

0.3

0.4

0.5

0.6

GE Optima

660

GE Discovery

CT 750HD

GE

Revolution

CT

Philips

Ingenuity

Philips

Brilliance iCT

Siemens

Definition

AS+

Siemens

Defintion

Stellar Edge

Siemens

Definition

Flash Stellar

Toshiba

Aquilion

PRIME

Toshiba

Aquilion ONE

Toshiba

Aquilion ONE

Vision

z-ax

is s

pat

ial r

eso

luti

on

(m

m)

Axial mode

Helical mode

Page 60: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

59

Table 10 - Spatial resolution in z-axis direction as per information provided by the manufacturer

1 Calculated from 2% MTF values provided by manufacturer

N/P. data not provided by manufacturer

N/A Not applicable - scanner not used in this mode for CCTA scans

Scanner model

z-width

of

detector

row

(mm)

z-spatial resolution

(mm)(1)

Approach used to improve z-resolution Axial

mode

Helical

mode

GE

Optima 660 0.625 0.27 0.27

Reflector material and construction of detector

allow for improved z-resolution

Overlapping reconstructions in helical mode

Discovery CT750 HD 0.625 0.27 0.27

Reflector material and construction of detector

allow for improved z-resolution

Overlapping reconstructions in helical mode

Revolution CT 0.625 0.27 N/A

Reflector material and construction of detector

allow for improved z-resolution

Overlapping image recons (2 per slice) in axial

Philips Ingenuity 0.625 0.42 0.42 N/P

iCT Elite 0.625 0.42 0.42 z-flying focal spot (ZFS)

Siemens Somatom Definition

AS+ 0.6 0.53 0.53 z-flying focal spot (z-sharp)

Somatom Definition

Edge Stellar 0.6 0.29 0.29

z-flying focal spot (z-sharp)

Stellar detectors deliver sharper slice profile

Somatom Definition

Flash Stellar 0.6 0.29 0.29

z-sharp technology

Stellar detectors deliver sharper slice profile

Toshiba Aquilion PRIME 0.5 N/A 0.45

ConeXact 3D algorithm

Overlapping image recons every 0.1 mm in helical

Aquilion ONE 0.5 0.43 N/A ConeXact 3D algorithm

Overlapping image recons every 0.25 mm in axial

Aquilion ONE Vision 0.5 0.43 N/A ConeXact 3D algorithm

Overlapping image recons every 0.25 mm in axial

Page 61: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

60

6.3.3 Volume coverage

Scanners with a wide z-axis detector can acquire the cardiac volume in a smaller number

of heartbeats (ideally a single heartbeat) and CCTA images therefore do not suffer from

misregistration artefacts that may result from irregular heart rates. Dual source scanners

are capable of acquiring the cardiac volume in a single heartbeat by using a high pitch

(>3) (Flash mode) and therefore also have the advantage of eliminating misregistration

artefacts, although the use of this high pitch mode is generally not recommended for high

heart rates.

The z-axis detector length of the various scanner models is presented in Table 11. It is

pertinent to note that quoting the ‘number of slices’, the approach often used to define

the level of a CT scanner, does not necessarily provide an indication of the length of z-

axis coverage.

The number of heartbeats required to scan a cardiac length of 140 mm with various scan

modes is tabulated in Table 12 and shown graphically in Figure 16. In general the

number of heartbeats required the cover a given volume is related to the z-axis

dimension of the detector, with scanners with a z-axis detector dimension of 160 mm

capable of covering the volume within a single heartbeat. The exception to this is if multi-

segment reconstruction is used for high heart rates. The Siemens Dual source scanners

can also cover the cardiac volume within a single heartbeat in their high pitch Flash

mode that is limited to lower heart rates.

Page 62: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

61

Table 11 - Length of z- axis detector coverage, number of detector rows, number of slices

acquired per rotation and number of reconstructed slices per rotation as per information provided

by the manufacturer

Scanner models No. of detector rows

No. of slices acquired per rotation

No. of slices reconstructed per rotation (axial mode)

Length of coverage in z-axis (mm)

GE

Optima 660 64 64 64 40

Discovery CT 750HD 64 64 64 40

Revolution CT 256 256 512(1) 160

Philips

Ingenuity 64 64 (2) 64 40

iCT Elite 128 256 (3) 256 80

Siemens

Somatom Definition AS+

64 128 (3) 384 38.4

Somatom Definition Edge Stellar

64 128 (3) 384 (4) 38.4

Somatom Definition Flash Stellar

64 per detector -

tube assembly

128 (3) per detector - tube assembly

384 (4) per detector - tube assembly

38.4

Toshiba

Aquilion PRIME 80 80 160 (1) 40

Aquilion ONE 320 320 640 (1) 160

Aquilion ONE Vision 320 320 640 (1) 160

1 Two overlapping slices per detector row reconstructed in axial mode from raw data

2 Raw data sampling is increased using a high-order interpolator to provide twice the

number of rows of detector data for reduction of ‘windmill’ artefact in helical mode

3 z-axis flying focal spot provides increased sampling by acquisition of two slices per

detector row

4 Three overlapping slices per detector row reconstructed in axial mode from raw data

Page 63: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

62

Table 12 - Volume coverage: number of heartbeats required to cover cardiac volume at heart

rates of 60 bpm and 80 bpm as per information provided by the manufacturer.

1 Calculated for pitch recommended in manufacturer’s protocols. Additional rotations required for image reconstruction from helical overscan have not been included in calculations.

N/A. - Not applicable - scanner not used in this mode for CCTA scans: n.a - this mode not available.

Scanner models

No of heartbeats within time taken to scan 140 mm cardiac length

Prospectively ECG-

triggering axial (PTA)

Retrospectively ECG-gated

helical low pitch

(RGH low pitch) (1)

Prospectively ECG-triggered

helical (High or low pitch)(1)

(PTH)

60 bpm 80 bpm 60 bpm 80 bpm 60 bpm 80 bpm

GE

Optima 660 7 N/A N/A 7

(pitch 0.2) n.a. n.a.

Discovery CT750

HD 7 N/A

7

(pitch 0.2) n.a. n.a.

Revolution CT 1 1 N/A N/A n.a. n.a.

Philips

Ingenuity 7 10 10

(pitch 0.16)

13

(pitch 0.16) n.a. n.a.

iCT Elite 3 4 3 (pitch 0.16) 4

(pitch 0.16) n.a. n.a.

Siemens

Somatom

Definition AS+ 7 7 N/A

7

(pitch 0.18) n.a. n.a.

Somatom

Definition Edge

Stellar

7 7 N/A 7

(pitch 0.18) n.a. n.a.

Somatom

Definition Flash

Stellar

N/A 7 N/A N/A 1 (pitch 3.4) n.a.

Toshiba

Aquilion PRIME N/A N/A 5

(pitch 0.22)

7

(pitch 0.2)

5

(pitch 0.22) N/A

Aquilion ONE 1 1 N/A N/A N/A N/A

Aquilion ONE

Vision 1 1 N/A N/A N/A N/A

Page 64: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

63

Figure 16 – Number of heartbeats required to scan cardiac 140 mm length in various scan modes.

6.3.4 X-ray flux

The available x-ray flux can be an important consideration in CCTA scanning as the short

gantry rotation times needed for good temporal resolution require a high tube current

(mA) to deliver an adequate signal at the detectors for sufficiently low image noise level.

Achieving a high enough x-ray flux may be a particular problem when scanning obese

patients, although the recent implementation of iterative reconstruction algorithms in CT

(see section 5.8) reduces this problem.

X-ray generator power is often used as a measure of a scanner’s x-ray flux capabilities

as this determines the maximum mA available at a particular tube potential. However, a

better measure of x-ray flux is the dose in a standard phantom (CTDIW) obtained with the

maximum mA and minimum gantry rotation time as presented in Table 13. The caveat to

the approach used here is the assumption that the scanner models all have a similar

dose efficiency. If the scanners vary significantly in terms of dose efficiency then this

measure is not appropriate as a given level of dose will not result in the same image

noise.

Calculations of CTDI were made for two x-ray tube potentials, 100 kV, commonly

recommended in standard patient size CCTA protocols, and 120 kV, the tube potential

typically used in obese patient protocols.

0

2

4

6

8

10

12

GEOp ma660

GEDiscoveryCT

750HD

GERevolu on

CT

PhilipsIngenuity

PhilipsBrillianceiCT

SiemensDefini on

AS+

SiemensDefin onStellarEdge

SiemensDefini onFlashStellar

ToshibaAquilionPRIME

ToshibaAquilionONE

ToshibaAquilion

ONEVision

No.o

fheratbeatsfor140m

mcardiaclength

axial(60bpm)

Axial(80bpm)

lowpitchhelical(60bpm)

lowpitchhelical(80bpm)

highpitchhelical(60bpm)

Page 65: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

64

Table 13 - CTDIw for minimum rotation time and maximum tube current available at tube

potentials of 100 kV and 120 kV.

(CTDIw is the absorbed radiation dose in a 32 cm diameter cylindrical Perspex phantom which is approximately

equivalent, in terms of x-ray attenuation, to a large patient. Assuming all the scanner models have a similar dose

efficiency, scanners with a higher CTDIw value will have lower image noise. The CTDIw figures do not rank the scanners

in terms of dose efficiency, but indicate the maximum dose they are able to achieve if used at maximum mA settings in

cases where this might be required e.g. for very large patients.)

Scanner models

X-ray

generator

power

(kW)

x-ray tube potential

100 kV

x-ray tube potential

120 kV

Max tube

current

(mA)

CTDIW (1)

(mGy)

Max tube

current

(mA)

CTDIW (1)

(mGy)

GE

Optima 660 72 480 8.1 600 16.0

Discovery CT750 HD 107 800 12.4 835 20.5

Revolution CT 103 720 N/P 740 17.8.

Philips Ingenuity 80 667 10.6 600 16.6

iCT Elite 120 1000 9.2 1000 15.9

Siemens

Somatom Definition

AS+ 80 650 6.1 666 10.7

Somatom Definition

Edge Stellar 100 650 5.5 800 11.6

Somatom Definition

Flash Stellar

100

(per tube) 600 5.8 800 11.9

Toshiba

Aquilion PRIME 72 600 13.9 600 22.5

Aquilion ONE 70 600 13.2 600 21.4

Aquilion ONE Vision 100 900 15.5 750 21.0

1 CTDIw/100 mAs values were provided by the CT manufacturers

N/P. data not provided by manufacturer

Page 66: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

65

7. Technical approaches adopted in current

systems to optimise image quality in ‘difficult

to image’ patient groups

This chapter describes the approaches used by each manufacturer on their cardiac CT

scanner models for imaging the various groups of challenging patients. Although each

patient group is discussed separately it must be noted that some patients will present

with two or more challenges e.g. high heart rate and high calcium levels.

To achieve a diagnostic CCTA scan at the lowest possible dose for each patient requires

not only suitable equipment but also an optimised protocol. A CCTA protocol is

comprised of numerous factors, including patient preparation and iodine contrast

administration, as well as the scan protocol. In CCTA scans the key element of the scan

protocol is the scan mode used. The various scan modes available for CCTA were

described in section 4.2 along with a summary of their comparative advantages and

disadvantages. Choice of scan mode is determined by the scanner model and also by

patient characteristics, primarily the heart rate. Other factors that must be considered in

the scan protocol include the scan range, scan field of view, x-ray beam collimation,

gantry rotation time, x-ray tube potential and tube current, as well as reconstruction

parameters such as slice width, reconstruction algorithm and reconstruction kernel. The

following sections present each manufacturer’s approach for meeting the challenges of

the various patient groups and these are summarised in Table 14 a - f.

7.1 Patients with high heart rates (> 65 bpm)

Obtaining CCTA images that are free from cardiac motion artefacts, particularly on

patients with high heart rates, requires a scanner with a high temporal resolution and

also the selection of the most static cardiac phase for image reconstruction. At lower

heart rates this is generally in diastole (~70% R-R interval), whereas for higher heart

rates end–systole may be the optimal phase (~45% R-R interval) (see Figure 17).

The scan mode used largely determines the flexibility in selection of the cardiac phase

for image reconstruction. Although the scan mode of choice is principally based on the

patient’s heart rate, the cut-off heart rate for a given mode varies with scanner model and

is mainly dependent on the scanner’s temporal resolution. The scan modes

recommended by the manufacturers for different heart rates are given in Table 14.

Generally, the current recommendations are to scan patients with lower heart rates with

prospectively ECG-triggered axial (PTA) mode, or prospectively ECG-triggered helical

(PTH) mode where available, as these are lower dose scan modes. However, these

Page 67: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

66

modes generally allow less flexibility in the cardiac phase available for image

reconstruction.

Figure 17 - Coronary CTA on dual source CT scanner with retrospectively ECG-gated helical

mode in a patient with a heart rate of 82 bpm (a) Curved MPR of the RCA at end-systole (45 % of

the RR interval) with excellent image quality (b) Curved MPR of the RCA during diastole (75 % of

the RR interval) displaying poor image quality (Kim HY et al, 2012).

Scanning using the same scan mode regardless of a patient’s heart rate is possible with

certain scanners, and this reduces the possibility of selecting a non-optimal scan mode.

Scanners that have this flexibility are those with a large z-axis coverage, namely the GE

Revolution CT and Toshiba Aquilion ONE models, where scanning in PTA mode is

recommended even for heart rates greater than 75 bpm.

On the GE Revolution CT, at higher heart rates data can be acquired in two cardiac

phases in a single heartbeat through ‘dual-peak phase’ triggering of x-rays in systole and

diastole, with the tube current being reduced to 20% of the maximum value outside these

cardiac phases. The cardiac phase in which the x-rays are triggered can be automatically

selected according to heart rate or else the user can over ride this if preferred. GE also

recommends the use of the SnapShot Freeze motion correction algorithm and multi-

segment reconstruction, techniques for improving the effective TR.

On the Philips Brilliance iCT it has been shown that diagnostic CCTA scans can be

reliably achieved in PTA mode at heart rates up to 75 bpm (Figure 18), but diagnostic

images have been obtained in this mode at higher heart rates.

Page 68: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

67

Figure 18 - Philips Brilliance iCT: Receiver-operator characteristic (ROC) curve to establish a cut

off mean heart rate value up to which diagnostic image quality can be consistently achieved using

prospectively triggered axial mode. The curve suggests a cut off of 75 bpm beyond which

coronary image quality was affected (Area under curve [AUC] = 0.92; 95% CI: 0.87, 0.98; P <

0.05) (Hou et al, 2012).

Dual source scanners such as the Siemens Somatom Definition Flash Stellar have a

very high intrinsic temporal resolution. On this system the temporal resolution for all heart

rates and in all scan modes is 75 ms. This scanner can acquire the 180° of data required

for image reconstruction in a quarter of a rotation instead of the usual half of a rotation.

This allows scanning in PTA mode up to heart rates in excess of 85 bpm without the

need to revert to retrospectively ECG-gated helical (RGH) mode (Figure 19) (Xu, 2010;

Sun, 2011).

On Siemens single source systems, the Somatom Definition AS+ and the Somatom

Definition Edge Stellar, diagnostic image quality can be obtained in PTA modes on

patients with heart rates up to 70 bpm and also a slight heart rate variability.

Page 69: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

68

Figure 19 - CCTA images acquired on a dual source CT scanner in adaptive sequential mode.

Mean HR = 96 bpm (range 86 – 105 bpm). Image quality was rated as score 1 (absence of motion

artefacts or noise-related blurring, excellent vessel opacification, and no structural discontinuity).

(a) Volume rendered image of coronary arteries (b) Curved multiplanar reformations show right

coronary (c) Curved multiplanar reformations show left anterior descending artery (Xu et al, 2010).

On Siemens scanners the amount of padding used in PTA scans is automatically

selected according to the patient’s heart rate (adaptive sequential mode). When using

RGH scan mode they utilise ‘adaptive ECG pulsing’ with ECG-gated tube current

modulation. This automatically adjusts the width of the maximum mA window according

to heart rate.

Toshiba is currently the only manufacturer with a low pitch prospectively ECG –triggered

helical (PTH) scan mode available, and this is recommended on the Aquilion PRIME up

to heart rates in excess of 65 bpm. As in prospectively triggered axial scanning,

increased padding is recommended with increasing heart rate. Toshiba have also

recently implemented Adaptive Motion Correction, an algorithm that compensates for

motion in the vessels, myocardium and valves.

On Toshiba’s Aquilion scanners, the cardiac phase in which the x-rays are triggered in

PTA Volume scan mode is automatically selected according to heart rate and the amount

of padding is increased for added flexibility. On the Aquilion ONE, it has been shown that

although the image quality decreases with increasing heart rate, good image quality is

still achieved (Figure 20), with the image quality score maintained at between 1

Page 70: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

69

(excellent) and 2 (good) for heart rates between 60 bpm and 100 bpm (Sun G et al,

2012).

(a) (b)

(c) (d)

Figure 20 - Images of a 63-year-old patient with an HR of 77.1 bpm and an HRv of 17.1. Excellent

and good images could be reconstructed. (a, b) Curve planar reconstruction showed stenosis in

right coronary artery (RCA) and left anterior descending artery (LAD) (arrows). (c, d) Invasive

coronary angiography confirmed the diagnosis (adapted from Sun G et al, 2012).

As mentioned in section 6.3.1 various methods are used for boosting a scanner’s intrinsic

temporal resolution. One of these is the use of multi-segment reconstruction (MSR),

whereby data from a number of consecutive heartbeats is taken to reconstruct images at

a given z-axis position. In this way the time over which data is acquired within a singe

heartbeat is reduced and so the temporal resolution is improved. On the majority of

systems it is possible to use MSR in low pitch RGH scan mode. On some scanners,

namely the Toshiba Aquilion ONE and ONE Vision models, MSR is also available in PTA

scan mode. MSR is particularly suited to scanners with wide z-axis coverage as fewer

cardiac cycles are needed to acquire the cardiac volume so there is a lower likelihood of

misregistration artefacts. However, the MSR approach does suffer from various

drawbacks (Tomizawa, 2012). On the Revolution CT, GE does not recommend multi-

segment reconstruction, but instead, for heart rates above 65 bpm advises ‘dual peak

phase’ acquisition and/or the use of SnapShot Freeze (SSF) motion correction software.

Because in the ‘dual phase peak’ method, x-rays are triggered in two different phases of

the cardiac cycle, the optimal phase for image reconstruction can be selected for each

Page 71: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

70

coronary artery. The SSF algorithm uses information from adjacent cardiac phases within

a single cardiac cycle to characterise vessel motion and determine the actual vessel

position at the prescribed target phase (Leipsic et al, 2012). SSF can be applied in both

PTA and RGH scan modes and is available on all GE’s cardiac CT scanner models. GE

claims that its implementation results in images with an effective temporal resolution of

29 ms (Figure 21). Early results with the SSF algorithm are promising (Carrascosa et al,

2015) and a prospective, international, multicentre trial (ViCTORY) is currently in

progress. Its primary aim is to determine the diagnostic accuracy of SSF for the

diagnosis of CAD, compared to ICA (Min et al, 2013).

Figure 21 - Case study of SnapShot Freeze motion corrected CCTA. 54 year old male patient,

BMI = 36, HR = 72 - 77 bpm. Prospectively triggered axial (PTA) scan mode with padding. Scan

shows mild coronary artery disease in the LAD with no stenosis (GE Healthcare, 2012).

Page 72: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

71

Table 14 - Scan modes used at different heart rates as per recommendations provided by the manufacturer. Scanner models Low heart rate (bpm) Medium heart rate (bpm) High heart rate (bpm)

GE

Optima 660 <65

Prospective axial - Step and shoot (PTA (~15% padding)

65 -75 Retrospective helical single segment (RGH)

(Snapshot Plus)

>75 Retrospective helical Multi-segment (RGH)

(Snapshot Burst)

Discovery CT750 HD <65

Prospective axial - Step and shoot (PTA) (~15% padding)

65 -75 Retrospective helical single segment (RGH)

(Snapshot Plus)

>75 Retrospective helical Multi-segment (RGH)

(Snapshot Burst)

Revolution CT <65

Prospective Volume axial (PTA V) (Single peak phase )

65 – 75 Prospective Volume axial (PTA V)

(Dual peak phase)

>75 Prospective Volume axial (PTA V)

(Dual peak phase)

Philips

Ingenuity <65

Gated Step/Shoot Cardiac (PTA) <65

Gated Step and Shoot Cardiac (PTA_

>65 Retrospective adaptive (RGH) Multi-segment reconstruction

iCT Elite <75

Gated Step/Shot (PTA) <75

Gated Step/Shoot Cardiac(PTA)

>75 Retrospective adaptive (RGH) Multi-segment reconstruction

Siemens

Somatom Definition AS+ <65

Prospective axial step and shoot (PTA) (Auto % padding)

65 -85 User preference: Prospective Step and shoot or

Retrospective helical (PTA or RGH)

>85 Retrospective helical (RGH)

with ECG mA modulation (min. mA 4%/20%)

Somatom Defintion Stellar Edge

<65 Prospective axial step and shoot (PTA)

(Auto % padding)

65 -85 User preference: Prospective Step and shoot or

Retrospective helical (PTA or RGH)

>85 Retrospective helical (RGH)

with ECG mA modulation (min. mA 4%/ 20%)

Somatom Definition Flash Stellar (Dual source)

<65 High pitch prospective helical (PTH-high pitch)

(Flash mode)

65 – 85 Prospective axial - Step and shoot (PTA)

(Auto % padding)

>85 Prospective axial - Step and shoot (PTA)

(Auto % padding)

Toshiba

Aquilion PRIME <60

Low pitch prospective helical (PTH-low pitch) (20% padding)

60 – 65 Low pitch prospective helical (PTH-low pitch)

(30% padding)

>65 Low pitch prospective helical (PTH-low pitch)

(60% padding)

Aquilion ONE <60

Prospective Volume axial ) (PTA V) (Auto phase selection with 10% padding

60 – 65 Prospective Volume axial (PTA V)

(Auto phase selection with 10% padding)

>65 Prospective Volume axial (PTA V)

(Auto phase selection with 50% padding)

Aquilion ONE Vision 30 - 55

Prospective volume axial (PTA V) (Auto phase selection with 4% padding)

56 – 75 Prospective Volume axial (PTA V)

(Auto phase selection with 10% padding)

>75 Prospective Volume axial (PTA V)

(Auto phase selection with 20% padding)

Page 73: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

72

7.2 Patients with arrhythmia

The issue of whether CCTA is a suitable examination for excluding CAD in patients with

atrial fibrillation remains contentious (Vorre et al, 2013; Schuetz et al, 2013). Limited

evidence is available, largely due to the relatively recent technological developments that

allow improvements in this area.

In patients with arrhythmia, the length of the cardiac cycle is variable and so a high

intrinsic temporal resolution is beneficial as it allows more latitude in the phase used for

image reconstruction whilst maintaining diagnostic image quality. CCTA scanning of

patients with arrhythmia is also facilitated on scanners that can acquire the cardiac

volume within a single heartbeat as the problem of beat-to-beat cardiac cycle variations

is eliminated (West et al, 2010).

Generally, scanners deal with arrhythmia by increasing the percentage of the cardiac

cycle irradiated to give flexibility in reconstruction phase. All manufacturers also provide

software to edit the ECG signal post acquisition by deleting, adding or removing R-

waves. When scanning in PTA mode, exposure can be temporarily suspended if an

ectopic beat is detected.

GE claims that for patients with arrhythmia the adaptive gating algorithm can predict the

heart rate of the next cardiac cycle and adapt the padding accordingly. In addition GE

recommends the use of the SnapShot Freeze motion correction algorithm and multi-

segment reconstruction, techniques for improving the effective TR. On the GE Revolution

CT, data are acquired during a single cardiac cycle and ‘Smart arrhythmia management’

is available. The manufacturer claims that the arrhythmia management capabilities on

the system allow avoiding scanning during irregular beats.

The Philips iCT Elite has a detector z-axis length of 8 cm and so can acquire the cardiac

volume over three heart beats in PTA scan mode, so patients with atrial fibrillation can be

successfully imaged (Muenzel et al, 2011, Chao et al, 2010). All Philips CT scanners

supporting card CCTA have automated arrhythmia handling tools that enable diagnostic

quality scans through detection or rejection of ectopic beats.

Page 74: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

73

Figure 22 - Coronal and sagittal reformats of CCTA scans on Brilliance iCT 128-row (256-slice) 60

year-old man after aortic valve replacement. Mean heart rate on CCTA was 67 bpm and high HR

variability of 53 bpm. (Muenzel et al 2011).

A meta-analysis has shown that the high intrinsic temporal resolution of the Siemens

dual source scanners makes them suitable for ruling out CAD in patients with atrial

fibrillation (Sun G et al, 2013). An example of a CCTA scan on a Siemens Somatom

Definition Flash of a patient in atrial fibrillation is shown in Figure 23. Two recent studies

have shown that more consistent radiation exposure and image quality across a wide

range of rates and rhythms could be achieved with PTA in systolic rather than diastolic

phase (Lee et al, 2013, Srichai et al, 2013). Scanning in the comparatively short systolic

phase is facilitated by the good intrinsic temporal resolution of dual source systems.

Another recent study has shown the successful use of double phase (60% and 30% R-R

cycle) high pitch PTH acquisition over two heartbeats on the Siemens Somatom

Definition Flash in patients with atrial fibrillation (Wang et al, 2013).

Figure 23 - CCTA on Siemens Somatom Definition Flash DSCT on a 51-year-old woman with

atrial fibrillation. The heart rate was irregular (mean, 136 beats per minute; range 83–200 beats

per minute). Prospectively ECG-triggered sequential imaging was performed with absolute phase

acquisition (200 ms to 400 ms of the R-R interval). Imaging was obtained at 100 kV and 370 mA.

Four slabs were used. Curved multiplanar reformation image of the right coronary artery (a) shows

excellent to good image quality (b) ECG recording made during data acquisition (Xu et al, 2013).

Page 75: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

74

On all Siemens cardiac scanner models the ‘adaptive sequence’ axial mode will omit or

repeat a scan when an ectopic beat is detected. For patients with known arrhythmia

Siemens recommends the use of RGH mode with automatic temporary suspension of

ECG-dose modulation if arrhythmia is detected.

Toshiba’s solution to scanning patients with arrhythmia on their Aquilion PRIME scanner

is to switch from PTH to RGH mode if arrhythmia is detected and vice versa if the heart

rate is returns to regularity. In other words, rather than acquiring data from a preselected

cardiac phase, data are acquired from the whole cardiac cycle for increased flexibility. On

the Aquilion ONE scanners, where data acquisition generally occurs during only a single

cardiac cycle, exposure is delayed if an arrhythmia occurs and will not take place until

the heart returns to its normal rhythm. Toshiba’s Real Beat Control Technology monitors

the patient’s heart rate up to the point of acquisition to ensure that the exposure window

is correct. In a study which included a small number of patients with chronic atrial

fibrillation, it was shown that on the Aquilion ONE equivalent diagnostic accuracy may be

obtained on patients with high heart rates and rhythm irregularities as on those with low

heart rates and normal sinus rhythm (Uehara, 2013).

7.3 Patients with high calcium scores (>400)

Severe calcifications in the coronary arteries are problematic because they result in

‘blooming artefacts’ caused by x-ray beam hardening and can lead to overestimation of

the degree of coronary stenosis. (Abdulla et al, 2012). A high spatial resolution can

reduce the amount of blooming and will be particularly beneficial in the presence of high

calcium levels. Manufacturers therefore generally recommend the use of sharper

reconstruction kernels in these cases, although this will result in increased image noise.

A good temporal resolution is also important as any additional blurring from cardiac

motion, will be minimised. Additionally the use of higher tube potentials, 120 KV or

above, may be recommended with high calcium levels to reduce beam hardening effects

and associated blooming.

On their Revolution range of scanners, GE recommends the use of High Definition (HD)

mode with a sharp reconstruction kernel (EDGE) for patients with high calcium scores.

HD mode provides a high number of views per rotation which is enabled by the fast

response time of the GemstoneTM detectors (Jiang, 2008). This results in a high sampling

density and improved x-y plane spatial resolution. All others manufacturers also achieve

a better x-y plane spatial resolution through the use of the sharper kernels in patients

with high calcium scores.

On all the scanner models in this report, z-axis resolution values of around 0.5 mm and

below are quoted. GE claims to achieve a z-axis resolution much smaller than the

detector’s z-dimension through their new detector construction (proprietary information).

On the Philips iCT Elite and all the Siemens scanners improved z-axis resolution is

Page 76: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

75

achieved with using a flying focal spot (termed z-sharp by Siemens and ZFS by Philips).

The Stellar detectors on the Siemens Somatom Edge Stellar and Flash Stellar scanners

provide an improved z-axis spatial resolution of 0.29 mm compared to 0.53 mm on

Siemens scanner models without the Stellar detectors. Toshiba has the smallest physical

z-axis detector dimension for achieving a good z-axis resolution and further improvement

is obtained by reconstruction of overlapping images enabled by the 3D ConeExact

algorithm.

Other than a high intrinsic spatial resolution, other techniques for improving the image

quality in patients with high calcium scores are recommended. Namely the use of

iterative reconstruction algorithms, calcium subtraction techniques and dual energy

scans where available.

In a study to assess of the impact of IR algorithms on calcium blooming, CCTA scans on

ex-vivo donor hearts showed no difference in terms of blooming artefacts between GE’s

FBP and its two IR reconstruction algorithms, ASIR and MBIR7 (Figure 24). However, use

of MBIR improved image quality, reduced image noise and increased CNR as compared

to the other two reconstruction methods. Therefore the authors concluded that MBIR may

have an important clinical role in the evaluation of the coronary artery tree which should

be validated in vivo (Scheffel et al, 2012).

Figure 24 - GE Discovery CT750 HD: Cross-sectional CT image reconstructed with (a) FBP, (b)

ASIR, and (c) MBIR technique. There is no difference in blooming of calcified component of the

coronary artery plaque in the three different reconstruction modes (Scheffel et al, 2012).

GE recommends the use of the dual energy cardiac mode (GSI Cardiac) on the

CT750 HD. Simulated monochromatic (keV) images obtained from dual energy scans

have been shown to reduce the effect of calcium blooming and give percentage lumen

7 It should be noted that MBIR (known as Veo on commercial GE systems) is not currently

available in cardiac scanning on commercial GE CT scanners.

Page 77: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

76

stenosis values more comparable to conventional angiography than those obtained from

single energy CCTA images (Figure 25).

Figure 25 - Percentage decrease in perceived level of stenosis in monochromatic images

obtained on a GE Discovery CT750 HD operating in dual energy mode. Plaques are largest at 40

keV and smallest at 140 keV with little change above 100 keV. Images courtesy Dr.James Earls

(AuntMinnie, 2012).

Philips claims that use of their iterative algorithm iDose4 or IMR (Iterative Model-based

Reconstruction) results in improved spatial resolution and therefore reduced blooming

(Philips, 2011). In a small study of 10 patients quantitative assessment indicated that

blooming artefact was reduced with iDose4 compared to FBP reconstruction (Figure 26).

Page 78: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

77

Figure 26 - Calcium blooming reduction at routine dose on the Philips Brilliance iCT with iDose4

iterative reconstruction compared to FBP. iDose4 reconstructions were performed in high

resolution mode while keeping image noise at the same level as FBP reconstruction (Philips,

2011).

Siemens recommends using SAFIRE iterative reconstruction (IR), or their more recent

ADMIRE IR for reduced blooming in patients with high calcium scores. A 2011 study on a

Definition Flash comparing Siemens’ original IR algorithm, IRIS, with FBP reconstruction

of CCTA on 55 consecutive patients with a calcium score ≥ 400, concluded that IR

reduces image noise and blooming artefacts from calcifications. This leads to improved

diagnostic accuracy of CCTA in patients with heavily calcified coronary arteries (Figure

27).

Figure 27 Siemens Somatom Definition Flash: Contrast-enhanced prospectively ECG-triggered

coronary CT angiographic images in a 73-year-old man with chest pain. Curved multiplanar

reformations of the left anterior descending coronary artery show heavy calcifications (arrows) in

the proximal vessel. Blooming artefacts limit evaluation of adjacent vessel lumen on (a) FBP

reconstructions, mimicking a substantial stenosis, but are reduced on (b) iterative reconstructions,

which enabled correct classification of the lesion as not significantly stenotic, as confirmed with (c)

subsequent coronary catheterization (Renker et al, 2011).

Toshiba recommends using the iterative reconstruction algorithm AIDR 3D that includes

artefact reduction software. Toshiba also recommends SURESubtraction Coronary for

eliminating calcium, an approach claimed to have fewer drawbacks than using

thresholding techniques. .SURESubtraction Coronary, can be obtained with a near dose

neutral scanning protocol by subtracting a routine Calcium Score dataset from a

Coronary CT Angiography dataset, using sophisticated segmentation and registration

algorithms (Toshiba, 2014). The Calcium Score scan is used as the non-contrast mask

for subtraction, thereby effectively removing calcium from the images (Figure 28).

Page 79: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

78

Figure 28 - Toshiba Aquilion ONE: A case example of a patient with coronary artery calcifications

and coronary artery stenosis (59-year-old male, coronary calcium score = 583). (a) Conventional

CCTA, (b) Subtraction CCTA, (c) Invasive coronary angiography. Stenotic lesions were observed

in the left diagonal branch (arrows). While the lumen patency was not clearly visible on the

conventional CCTA images, subtraction CCTA clearly depicted the stenosis confirmed by invasive

coronary angiography. Also, subtraction CCTA denotes no obvious stenosis in the left main trunk

and left anterior descending artery (arrowheads) (Tanaka et al, 2013).

7.4 Patients with stents

Similarly to high calcium levels, the presence of coronary artery stents can also lead to

beam hardening artefacts resulting in blooming that obscures the lumen and can make

diagnosis of in-stent restenosis difficult. Different types of stent materials lead to varying

levels of blooming but generally image interpretation becomes problematic for stent

diameters less than 3 mm. As in the presence of high levels of calcium, for stent imaging

a high spatial resolution improves diagnostic image quality, so a scanner with a high

limiting spatial resolution and the use of sharper reconstruction kernels will reduce

blooming artefacts (Figure 29). However, the use of sharper kernels results in increased

noise levels that may impact on diagnostic image quality. Although claims have been

made for improved spatial resolution with iterative reconstruction (IR) algorithms, no

published clinical studies could be identified to substantiate this. However, IR will result in

less noisy images and so in combination with high resolution reconstruction kernels

should provide improved diagnostic accuracy in the assessment of in-stent stenosis.

Again the importance of a good temporal resolution must be recognised to minimise

blurring due to motion. Manufacturers generally recommend the same approaches for

improving spatial resolution in stent imaging as for imaging patients with high calcium

levels, recommending use of sharper kernels, IR reconstruction algorithms and setting

higher tube potentials (kVs) for blooming artefact reduction.

Page 80: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

79

B26f B30f B45f B46f

Figure 29 - Siemens Somatom Definition dual source CT: Multiplanar reformations of the Pro

Kinetic 3 mm diameter stent. Comparison of the four different reconstruction kernels, B26f, B30f,

B45f, B46f (numbers increase with increasing sharpness of kernel). Note the improved lumen

visibility using the B46f reconstruction kernel (Maintz et al 2009).

A clinical study of 180 consecutive patients compared the assessment of coronary in-

stent restenosis (ISR) on the GE Discovery CT750 HD using ASIR (Group 1) and the GE

LightSpeed VCT (Group 2). Group 1 showed a higher stent evaluability than Group 2

(99% vs 92%, P = .0021). An example is shown in Figure 30. However for stent-based

analysis, sensitivity, specificity and accuracy values were not statistically different

(Andreini et al, 2012).

Figure 30 - GE Discovery CT750 HD with ASIR (a) Multiplanar reconstruction from high-spatial-

resolution multidetector CT examination shows the presence of diffuse ISR in three bare-metal

stents implanted in the right coronary artery (RCA), with severe stenosis at the level of the second

Page 81: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

80

and third stent segments (arrows) (b) Image obtained at ICA confirms the presence, distribution,

and severity of ISR (arrows) (Andreini et al, 2012).

A recent pilot clinical study on the Philips Brilliance iCT compared the diagnostic

performance in the evaluation of the in-stent lumen of three different reconstruction

modes. Images were reconstructed with (1) an FBP algorithm using a standard cardiac

kernel (CB), (2) an FBP algorithm with high resolution cardiac kernel (CD), and (3) iDose4

(HIR) with CD kernel The iDose4 algorithm with the CD kernel significantly reduced the

image noise compared to FBP with the CD kernel and significantly reduced coronary

stent blooming artefacts compared to FBP with CB kernel (Figure 31). Therefore the

combination of the HIR algorithm for reduced noise and the CD kernel for improved

resolution significantly improved diagnostic performance for the detection of in-stent

stenosis (Figure 32) (Oda, 2013).

Figure 31 - Philips Brilliance iCT: CT voxel attenuation profiles across the stent for FBP with the

CB kernel (standard cardiac), FBP with the CD kernel (high resolution cardiac) and HIR and the

CD kernel using multi - planar reformation images (Oda et al, 2013).

Page 82: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

81

Figure 32 - In-stent restenosis (arrow) in the middle segment of the left anterior descending artery

in an 82-year-old man. Curved multi-planar reconstruction images with FBP and the CB kernel (a),

FBP and the CD kernel (b), and HIR and the CD kernel (c). Application of HIR with the CD kernel

(c) facilitated image noise reduction, improved visualization of the stent lumen due to a reduction

in blooming artefacts, and delineated in-stent stenosis more clearly. Conventional coronary

angiogram of the left anterior descending artery (d) confirms in-stent restenosis (arrow) (Oda et al,

2013).

In a recent small study of 37 implanted stents performed on the Siemens Somatom

Definition Flash, the IR SAFIRE algorithm was significantly superior to the FBP algorithm

in terms of noise, SNR, stent-lumen attenuation ratio and image quality, at the same

dose levels as the FBP algorithm (Figure 33) (Ebersberger et al, 2013).

Page 83: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

82

Figure 33 - Siemens Somatom Definition Flash: Cardiac CT angiography (cCTA) study of a patient

with an implanted coronary artery stent displayed as automatically generated curved multiplanar

reformat along the vessel centreline (right panels) and as cross-sections perpendicular to the

centreline (left panels). Upper panel (a) illustrates image reconstruction based on full-dose

SAFIRE (I46f kernel), whereas the lower panel (b) shows full-dose FBP reconstruction (B46f

kernel) (Ebersberger et al, 2013).

Toshiba recommend the use of SURESubtraction Coronary to improve diagnostic accuracy

of CAD in patients with in-stent restenosis (Figure 34).

Figure 34 In-stent restenosis is seen in the LAD in the subtracted image – Courtesy Dr M Chen,

NHLBI, National Institutes of Health, USA – (http://www.toshiba-medical.eu/eu/toshiba-introduces-

suresubtraction-coronary-at-esc-2014/#sthash.QWTypxrt.dpuf)

Page 84: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

83

No studies specifically on diagnosing CAD in patients with stents were identified in the

literature for the Toshiba scanners included in this report. However, a recent study on the

Toshiba Aquilion ONE showed that diagnosis of CAD in patients with in-stent restenosis

could be improved with combined CCTA and myocardial CT perfusion (CTP) compared

with CCTA alone (Rief et al, 2013).

Another recent study of CCTA on 51 consecutive patients with the Toshiba Aquilion ONE

showed that the use of the AIDR 3D iterative reconstruction algorithm reduced image

noise by 39 % compared with the FBP without affecting CT density, thus improving SNR

and CNR for CCTA. The advantages of AIDR in interpretability were also confirmed by

subjective evaluation by experts (Yoo et al, 2013). With these findings improved

diagnostic capability would also be expected for stent imaging.

Page 85: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

84

7.5 Patients with coronary artery bypass grafts

(CABG)

Assessment of the patency of bypass grafts in CCTA is usually not challenging, primarily

because of the larger size of vein grafts (typically 3 – 4 mm diameter) and their reduced

mobility, and because they are usually not heavily calcified (Mark et al, 2010). However

the assessment of native coronary arteries in patients after CABG is challenging owing to

poor run-off, (outflow of blood from the heart) more extensive calcification and diffusely

narrowed arteries with small dimensions (Sun and Choo, 2012). Additionally scanning of

longer lengths to include the whole thorax may be required.An additional challenge for

patients with internal mammary grafts is the presence of metal clips, their smaller size

(1– 2 mm) and the possible need to rule out subclavian stenosis requiring an increased

coverage (Mark et al, 2012).

When using the Revolution CT scanner for CCTA post CABG, GE suggests the use of

two table positions with smart collimation to enable one beat acquisition of the heart and

avoid a volume boundary over the heart.

On the Siemens Somatom Definition Flash Stellar the high pitch helical ‘Flash’ mode can

be used to scan a standard 300 mm thorax in 0.6 seconds if extended coverage is

required. In a study using a Somatom Definition Flash CT scanner, 50 consecutive

patients underwent CT angiography of the entire thorax, using this prospectively ECG-

triggered mode for the evaluation of graft patency after CABG surgery. The start of CT

data acquisition was automatically calculated by the CT software with the aim of imaging

the distal anastomosis (the graft section with the most potential for motion artefacts) at

the 60% R-R interval. The mean heart rate was 76 ± 19 bpm. Mean scan length was 349

± 38 mm. Diagnostic image quality was obtained in 99.4% of sections. The authors

concluded that the patency of CABG can be assessed with decreasing image quality at

high HR in high-pitch PTH thoracic CTA angiography at a low radiation dose (2.3 ± 0.3

mSv) (Goetti et al, 2010).

The results of a small study presented recently (RSNA, Chicago, 2013) using a Philips

Brilliance iCT scanner showed that CCTA with a low dose protocol (PTA mode, 100 kV

using iDose4) achieved a high diagnostic accuracy in patients with CABG (Aunt Minnie,

2014).

On the Toshiba Aquilion ONE the longer scan lengths, which may be required in CABG

scans, can be performed quickly with the wide detector array. Upon setting the desired

range, Wide Volume mode automatically calculates the two acquisition volumes required

to cover the heart and graft while SUREXposure Cardiac tailors the the dose for each

segment of the acquisition. Use of SUREStart to capture peak enhancement in the left

ventricle ensures that there are no poor out-flow artefacts. Toshiba claims that the AIDR

3D iterative algorithm and a large range of reconstruction filters available will aid in

providing good image quality in CCTA on patients with CABG.

Page 86: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

85

7.6 Obese patients (> 30 kg/m2)

As discussed in section 6.3.4 scanning obese patients can pose a challenge in CCTA

due to increased image noise. Image quality is also degraded as a result of more

scattered x-rays. All cardiac CT scanners are generally equipped with powerful x-ray

generators capable of achieving high maximum tube currents (Table 13). However, the

introduction of iterative reconstruction algorithms in CT has somewhat reduced the

problem of photon flux limitation even in obese patients as with the same exposure

settings lower noise values are achieved than with filtered back projection (FBP).

Manufacturers generally suggest use of a higher tube potential to provide an increased

photon flux at the detectors if necessary.

From a practical viewpoint, scanners with couches that supports a high weight and those

with a large gantry bore will be desirable (see Table 7 – Key CT scanner specifications).

To meet the challenges of scanning obese patients GE promotes the use of the ASIR

iterative reconstruction algorithm, or the more advance ASIR-V on the Revolution CT.

The use of ASIR in conjunction with GE’s High Definition mode on the Discovery CT750

HD has been shown to result in improved image quality and visualisation of distal

coronary segments, as compared to a standard FBP protocol (Figure 35), in overweight

and obese individuals without increasing image noise and radiation dose (Gebhard,

2013). Another approach suggested by GE for achieving an increased photon flux in

obese patients is to increase rotation time, which will result in a higher tube current-time

product (mAs). However, the reduced noise will be achieved at the expense of temporal

resolution. On the Revolution CT automatic exposure control (AEC) is available for CCTA

scans and so the tube potential (kV Assist) and tube current (Auto mA/Smart mA) will be

automatically optimised for patient size.

Figure 35 - GE Discovery CT750 HD: (A) Tertile analysis according to BMI of mean per patient

image quality score (mean ± SEM) in both groups, (B) Tertile analysis according to BMI of minimal

Page 87: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

86

area of coronary artery segments that could be visualized in both groups (mean ± SEM). HDCT

high definition computed tomography. ASIR adaptive iterative reconstruction. Data are presented

as mean ± SEM. *p < 0.05. Image Quality score: 1= Excellent - 4 = non-diagnostic (Gebhard,

2013).

A white paper by Philips on their iterative algorithm iDose4 gives a summary of the

findings of a study on 30 obese patients undergoing CCTA (Philips Healthcare, 2011). All

acquisitions were performed using PTA (Step & Shoot Cardiac) mode on the Brilliance

iCT using routine scan protocols (120 kVp, 200 - 340 mAs, average effective dose 6.3

mSv) and were reconstructed using FBP and iDose4. Image quality for both approaches

was subjectively analysed by two blinded readers. The conclusions were that iDose4

facilitates noise reduction while maintaining diagnostic image quality in PTA CCTA scans

performed on morbidly obese patients for preoperative assessment prior to bariatric

surgery. An example from the study is shown in Figure 36.

Figure 36 - Example of improved image quality with Philips iterative algorithm (iDose4) in bariatric

patients (30 patients - BMI 38.9 ± 7.1) (Philips Healthcare, 2011).

Siemens have AEC technology available for CCTA on all their cardiac scanner models.

CAREDose 4D automatically adjusts the mA and CAREkV automatically selects the

optimal kV for the patient size. The aim of CARE kV is to optimise contrast-to-noise ratio

(CNR) and dose for each clinical application.

Siemens claims that its scanners with Stellar detectors (Definition Edge and Definition

Flash) will be advantageous for scanning obese patients as they are specially optimised

for low-signal imaging due to low electronic noise and increased dynamic range.

The Somatom Definition Flash Stellar has two 100 kW generators providing a maximum

mA value of 1600 when the two x-ray tubes are used simultaneously in ‘cardio obese’

mode. When utilising these maximum mA values for reduced noise the temporal

resolution is sacrificed and becomes equivalent to that of a single source system with the

same gantry rotation time. However, the mode allows a flexible temporal resolution 75 to

Page 88: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

87

145 ms by using additional projections (90° to 180°) to obtain the required balance

between temporal resolution and improved signal-to-noise ratio.

A recent publication described the findings of a small study to evaluate the effect of a

temporal resolution improvement method (TRIM) on diagnostic image quality for

coronary artery assessment. The study was carried out on 11 obese patients (Mean BMI

36 ± 3.6 kg/m2) with a Siemens Somatom Definition Flash scanner using the XXL

acquisition protocol with a gantry rotation time of 500 ms to provide a higher tube current

– time product (mAs). The TRIM-algorithm employs an iterative approach to reconstruct

images from less than 180° of projections and uses a histogram constraint to prevent the

occurrence of limited-angle artefacts.

All data were reconstructed with a temporal resolution of 250 ms using traditional filtered

back projection (FBP) and of 200 ms using the TRIM-algorithm. All studies were deemed

diagnostic; however, there was a significant (p < 0.05) difference in the severity score

distribution of coronary motion artefacts between FBP (median = 2.5) and TRIM (median

= 2.0) reconstructions (Figure 37). The authors concluded that the TRIM algorithm

delivers diagnostic imaging quality of the coronary arteries despite the 500 ms gantry

rotation. Although this software wss not yet available on commercial Siemens scanners,

the authors noted that possible applications include improvement of cardiac imaging on

slower gantry rotation systems or mitigation of the trade-off between temporal resolution

and CNR in obese patients (Apfaltrer et al, 2013).

The TRIM algorithm is now available commercially on the Siemens Somatom

Perspective scanner but not on the Siemens scanner models considered in this report.

Figure 37 - FBP (A) and TRIM (B) reconstructions, at the same level of the right coronary artery.

The TRIM reconstruction on the right received a superior motion artefact severity score (minor =

2) than the standard reconstruction (moderate = 3) on the left, which shows more severe motion

artefacts arising from a calcified plaque (Apfaltrer et al, 2013).

Page 89: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

88

Toshiba’s approach to performing CCTA on obese patients is to use AIDR 3D iterative

reconstruction. The AEC system (SureExposure) automatically adjusts the mA for patient

size and also takes account of the fat in the patient, taking advantage of inherent contrast

when calculating the exposure required. SUREkV suggests the optimal tube potential for

optimising contrast and image quality. Toshiba claims that their new PUREVision detector

provides a 40% increase in light output and a 28% decrease in electronic noise (Toshiba,

2014). Toshiba CT scanners also have the unique feature of lateral couch movement to

facilitate accurate patient positioning although the potential of reduction in effective scan

field of view must be considered.

Page 90: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

89

Table 15 - Scanner features for difficult patient groups – information provided by manufacturers.

(a) Clinical challenge: High heart rate

Scanner model Approaches to meeting challenges of high heart rates (1)

GE

Optima 660 SnapShot Freeze motion correction algorithm; Multi-Sector Recon (SnapShot Burst/Burst Plus mode); ECG Edited Retro Reconstruction

Discovery CT750 HD

SnapShot Freeze; Multi-sector reconstruction (SnapShot Burst/Burst Plus mode).

Revolution CT Single beat modulated two-peak; Snapshot Freeze; Acquisition phases determined by the system based on patient’s heart rate variability

Philips Ingenuity Rate responsive technology and adaptive multi-cycle reconstruction.

iCT Elite Rate responsive technology and adaptive multi-cycle reconstruction.

Siemens

Somatom Definition AS+

Dual-segment acquisition protocol in axial and helical at high HR; Flex padding in axial; Adaptive ECG pulsing to widen window at high HR

Somatom Definition Edge Stellar

Dual-segment acquisition protocol in axial and helical; Flex padding in axial; Adaptive ECG pulsing to widen window at high HR.

Somatom Definition Flash Stellar

Dual source acquisition equivalent to gantry rotation time of 0.14 s; Dual-segment acquisition protocol in axial and helical; Flex padding in axial;

Adaptive ECG pulsing to widen window at high HR.

Somatom Force Dual source acquisition equivalent to gantry rotation time of 0.125s; Dual-segment acquisition protocol in axial and helical; Flex padding in axial; Adaptive ECG pulsing to widen window at high HR.

Toshiba Aquilion PRIME

The system automatically selects the appropriate factors including mA, kV, pitch and exposure phase of the cardiac cycle according to the patient heart rate. If heart rate is high this may reduce pitch or increase rotation time.

Aquilion ONE / ONE Vision

The system will automatically select the appropriate phase of the cardiac cycle for data acquisition according to the patient heart rate. If high this may suggest two rotations rather than one, but in clinical practice it is possible to narrow the exposure window to ensure one beat acquisition.

1Further clarifications on the operation of some features were not provided by the manufacturer within the timescale of the project.

Page 91: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

90

(b) Clinical challenge: Arrhythmia

Scanner model Approaches to meeting challenges of patients with arrhythmia (1)

GE

Optima 660 SnapShot Freeze motion correction algorithm; Multi-Sector Recon (SnapShot Burst/Burst Plus mode); ECG Edited Retro Reconstruction.

Discovery CT750 HD

SnapShot Freeze motion correction algorithm; Multi-Sector Recon (SnapShot Burst/Burst Plus mode); ECG Edited Retro Reconstruction.

Revolution CT Adaptive gating; Smart arrhythmia management in case of irregular heart rate.

Philips

Ingenuity Automatic arrhythmia detection and management.

iCT Elite Automatic arrhythmia detection and management

Siemens

Somatom Definition AS+

Adaptive sequence axial mode will omit/repeat scan when ectopic beat detected; Retro helical recommended for known arrhythmia – ECG-modulation temporarily suspended if arrhythmia detected; ECG editing; Phase reconstruction performed with relative (%) or absolute (ms) parameters.

Somatom Definition Edge Stellar

Adaptive sequence axial mode will omit/repeat scan when ectopic beat detected; Retro helical recommended for known arrhythmia – ECG-modulation temporarily suspended if arrhythmia detected; ECG editing; Phase reconstruction performed with relative (%) or absolute (ms) parameters.

Somatom Definition Flash Stellar

High temporal resolution of dual source allows arrhythmia imaging with single segment; Adaptive sequence axial mode will omit/repeat scan when ectopic beat detected; Retro helical recommended for known arrhythmia – ECG-modulation will temporarily suspend if arrhythmia detected;

ECG editing; Phase reconstruction performed with relative (%) or absolute (ms) parameters.

Somatom Force High temporal resolution of dual source allows arrhythmia imaging with single segment; Adaptive sequence axial mode will omit/repeat scan when ectopic beat detected; Retro helical recommended for known arrhythmia – ECG-modulation will temporarily suspend if arrhythmia detected; ECG editing; Phase reconstruction performed with relative (%) or absolute (ms) parameters

Toshiba

Aquilion PRIME Heart rate is monitored in real time and if arrhythmia is detected during prospective acquisition the system will switch into retrospective mode to ensure a diagnostic outcome.

Aquilion ONE/ONE Vision

With no need to move the table, the Aquilion ONE scanners can monitor heart rate in real time and only exposes when there is a suitable interval in the cardiac cycle. If an arrhythmia occurs, exposure will not take place until the next suitable interval. The user can set a number of arrhythmias to reject.

1Further clarifications on the operation of some features were not provided by the manufacturer within the timescale of the project.

Page 92: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

91

(c) Clinical challenge: High calcium score

Scanner model Approaches to meeting challenges of patients with high calcium scores (1)

GE

Optima 660 High-frequency kernel.

Discovery CT750 HD High Definition (HD) mode; Dual Energy cardiac (GSI Cardiac) mode.

Revolution CT High Definition (HD) mode.

Philips

Ingenuity Iterative algorithms such as IMR & iDose, Sharper edges to reduce blooming; Magic Glass visualization.

iCT Elite Iterative algoritms such as IMR & iDose, Sharper edges to reduce blooming; Magic Glass visualization.

Siemens

Somatom Definition AS+

Use of Iterative reconstruction (SAFIRE) will reduce the calcium blooming artefact through higher resolution..

Somatom Definition Edge Stellar

Use of Iterative reconstruction (SAFIRE) will reduce the calcium blooming artefact through higher resolution.

Somatom Definition Flash Stellar

Use of Iterative reconstruction (SAFIRE) will reduce the calcium blooming artefact through higher resolution.

Somatom Force Use of Iterative reconstruction (SAFIRE) will reduce the calcium blooming artefact through higher resolution.

Toshiba

Aquilion PRIME Small detector elements and high sampling rate reduce blooming artefacts; AIDR 3D includes artefact reduction software;

Aquilion ONE/ONE Vision

Small detector elements and high sampling rate reduce blooming artefacts; AIDR 3D includes artefact reduction software; SURESubtraction accurately eliminates calcium without drawbacks of thresholding;.

1Further clarifications on the operation of some features were not provided by the manufacturer within the timescale of the project.

Page 93: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

92

(d) Clinical challenge: Stents

Scanner model Approaches to meeting challenges of patients with stents (1)

GE Optima 660 High-frequency kernel.

Discovery CT750 HD High Definition (HD) mode; Dual Energy cardiac (GSI Cardiac) mode.

Revolution CT High Definition (HD) mode.

Philips Ingenuity Dedicated stent kernel, Magic Glass visualization.

iCT Elite Dedicated stent kernel, Magic Glass visualization.

Siemens Somatom Definition AS+

Acquisition using z-sharp technology will optimise spatial resolution to allow small vessel and in-stent visualization.

Somatom Definition Edge Stellar

Stellar detector with Edge technology:

- Improves in-plane spatial resolution for more accurate in-stent visualisation.

- Reduces electronic cross-talk between neighbouring detector rows delivering a spatial cross-plane resolution of 0.30 mm.

Acquisition using z-sharp technology optimises spatial resolution to allow small vessel and in-stent visualization

Somatom Definition Flash Stellar

Stellar detector with Edge technology:

- Improves in-plane spatial resolution for more accurate in-stent visualization;

- Reduces electronic cross-talk between neighbouring detector rows delivering a spatial cross-plane resolution of 0.30 mm;

Acquisition using z-sharp technology will optimise spatial resolution to allow small vessel and in-stent visualization.

Somatom Force

Stellar detector with Edge technology:

- Improves in-plane spatial resolution for more accurate in-stent visualization;

- Reduces electronic cross-talk between neighboring detector rows delivering a spatial cross-plane resolution of 0.30 mm;

Acquisition using z-sharp technology will optimize spatial resolution to allow small vessel and in-stent visualization.

Toshiba Aquilion PRIME

Small detector elements and high sampling rate give best x. y, z resolution to see fine detail in small structures. AIDR 3D and large range of reconstruction algorithms gives good image quality.

Aquilion ONE/ONE Vision

Small detector elements and high sampling rate give best x. y, z resolution to see fine detail in small structures. AIDR 3D and large range of reconstruction algorithms gives good image quality. SURESubtraction accurately eliminates calcium without drawbacks of thresholding;.

1Further clarifications on the operation of some features were not provided by the manufacturer within the timescale of the project.

Page 94: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

93

(e) Clinical challenge: Coronary artery bypass grafts

Scanner model Approaches to meeting challenges of patients with coronary artery bypass grafts (1)

GE

Optima 660 ECG triggered Step-and-Shoot mode (SnapShot Pulse) – PTA mode.

Discovery CT750 HD ECG triggered Step-and-Shoot mode (SnapShot Pulse) – PTA mode.

Revolution CT Two table positions with smart collimation to enable 1 beat acquisition of the heart and avoid volume boundary over the heart.

Philips Ingenuity Prospective Step and Shoot Complete

iCT Elite Prospective Step and Shoot Complete

Siemens

Somatom Definition AS+ The high speed scan acquisition in 128 slice acquisition and flexibility of the scan volume allows whole thorax coverage to identify graft origins to base of heart as required.

Somatom Definition Edge Stellar

The high speed scan acquisition in 128 slice acquisition and flexibility of the scan volume allows whole thorax coverage to identify graft origins to base of heart as required.

Somatom Definition Flash Stellar

The high speed scan acquisition in 128 slice acquisition and flexibility of the scan volume allows whole thorax coverage to identify graft origins to base of heart as required.

The Definition Flash can also use its unique Flash cardio spiral to acquire extended scan ranges with acquisition speeds up to 458 mm/s. This enables a standard thorax (300mm) to be imaged in 0.6 s.

Somatom Force The high speed scan acquisition in 192 slice acquisition and flexibility of the scan volume allows whole thorax coverage to identify graft origins to base of heart as required. The system can also use its Flash cardio spiral to acquire extended scan ranges with acquisition speeds up to 737 mm/s in Turbo Flash Mode. This enables a standard thorax (300 mm) to be imaged in 0.4s.

Toshiba Aquilion PRIME

Small detector elements and high sampling rate give best x. y, z resolution to see fine detail in small structures. AIDR 3D and large range of reconstruction algorithms gives good image quality.

Aquilion ONE/ ONE Vision Small detector elements and high sampling rate give best x. y, z resolution to see fine detail in small structures. AIDR 3D and large range of reconstruction algorithms gives good image quality. Wide Volume mode automatically calculates the two acquisition volumes required for coverage.

1Further clarifications on the operation of some features were not provided by the manufacturer within the timescale of the project.

Page 95: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

94

(f) Clinical challenge: Obesity

Scanner model Approaches to meeting challenges of obese patients (1)

GE

Optima 660 Slower rotation time; Cardiac filter; Higher % blended ASiR iterative reconstruction .

Discovery CT750 HD

Slower rotation time; Cardiac filter; Higher % blended ASiR iterative reconstruction.

Revolution CT AEC: Auto kV and mA adjustment based on measured patient attenuation; ASIR-V iterative reconstruction.

Philips Ingenuity

IMR, iDose4, iPatient dose and image quality management. NanoPanel Elite detector with better Signal/Noise.

iCT Elite IMR, iDose4, iPatient dose and image quality management. 120 kW power Elite detector with better Signal/Noise.

Siemens

Somatom Definition AS+

CAREDose 4D auto mA modulation and CAREkV auto kV selection; 80kW generator (option 100kW) provides the power reserves required in obese imaging to obtain the image quality required to maintaining fast rotation time.

Somatom Definition Edge Stellar

Stellar Detector with TrueSignal Technology: detector is specially optimized for low-signal imaging.

CAREDose 4D automated mA modulation and CAREkV automatic kV selection; 100kW generator provides power reserves required in obese imaging.

Somatom Definition Flash Stellar

Stellar Detector with TrueSignal Technology: detector is specially optimized for low-signal imaging.

CAREDose 4D auto mA modulation and CAREkV auto kV selection provides patient specific parameters; 100kW generator provides the power required in obese imaging; Use of two tubes simultaneously in a cardio obese mode that uses flexible temp.res.(75 to 140ms) and additional projections (90 to 180 degrees of rotation) to improve signal-to-noise ratio. This is coupled with the power of 2 x 100kW generators to provide up to 1600 mA for bariatric cardiac imaging.

Toshiba Aquilion PRIME

AIDR 3D applicable to cardiac. High capacity patient couch. Lateral couch movement aids accurate positioning. Large gantry bore. SureExposure takes account of fat in patient when calculating dose required.

Aquilion ONE/ ONE Vision

AIDR 3D applicable to cardiac. High capacity patient couch. Lateral couch movement aids accurate positioning. Large gantry bore. SureExposure takes account of fat in patient when calculating dose required.

1Further clarifications on the operation of some features were not provided by the manufacturer within the timescale of the project.

Page 96: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

95

8. Advantages, uncertainties and risks of

comparing scanners using technical

specifications

8.1 Bias

Technical specifications are objective indicators and therefore reduce the chance of bias

introduced by subjective metrics that are affected by human preference and decision

criteria. Comparing technical specifications allows experienced users of CT technology to

identify equipment that may offer the greatest potential for superior performance when

used in clinical practice.

8.2 Impact on clinical performance

The value of comparing specifications of medical technologies has been recognised and

used as the basis for selection and purchasing of medical equipment in the NHS for many

years. Specialised centres (e.g. KCARE, ImPACT, MagNET) developed relevant expertise

and fruitful cooperation with industry over many years to undertake comparative

specification studies and technology performance assessments widely used by NHS

healthcare staff in the selection of medical devices suitable for their needs.

8.3 Access to technical specifications

Detailed technical specifications may not be readily available from manufacturers as they

are sometimes reluctant to disclose details to protect their intellectual property.

Nevertheless it is reasonable to assume that it is likely to be easier to obtain technical

specifications in a shorter time than to produce diagnostic performance indicators

describing the clinical utility of the equipment.

8.4 Quality of the technical data

Technical specification data are strongly dependent on measurement methodology that

may vary between manufacturers. This may potentially lead to distorted comparisons.

However, it is relevant to point out that any one manufacturer is likely to be consistent in

the specifications provided for their different CT scanners models, allowing a direct

comparison between old and new models and between different ‘levels’ of CT scanner

model (low end and high end) from the same manufacturer.

Page 97: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

96

8.5 Expertise required

Modern CT scanners are complex items of technology and adequate expertise is required

to understand the differences between technical data for different scanners and infer how

these differences may impact on diagnostic performance for specific patient groups.

8.6 Manufacturer-specific technical features

Although the fundamental principles of operation of various CT scanners may be

identical, some features are manufacturer-specific and therefore cannot be compared

across systems based on technical specification. Two examples follow:

To achieve a good temporal resolution, one manufacturer has developed a hardware

approach using dual sources whereas another uses a software approach (SnapShot

Freeze algorithm) to correct for motion artefacts and achieve a good effective temporal

resolution.

To achieve fast anatomical coverage and image the cardiac volume within one

heartbeat, some manufacturers have developed detector arrays with long z-axis

dimensions, whereas another manufacturer achieves single beat coverage using a

narrower detector but scanning at a high helical pitch.

8.7 Software upgrades

The discussions in this report mainly address the hardware features of CT equipment.

However there are many software features available for CCTA scanning that can enhance

image quality at the stages of data acquisition and image reconstruction, as well as at the

reporting stage. Moreover, software developments are rapid and upgrades are frequently

implemented (on new and old scanner models) to improve performance.

8.8 Multi-factorial effects

Another area of uncertainty relates to the matching of a given clinical ‘difficulty’ (e.g. high

heart rate, arrhythmia, stents) to a specific key technical feature (e.g. rotation time, z-axis

x-ray beam width, spatial resolution). Although this approach is based on sound physical

considerations, a given technical improvement should result in improved image quality and

is likely to benefit all patient groups.

For example, for obese patients with normal heart rates, a better temporal resolution will,

in theory, provide no improvement in clinical performance, as the key technical

requirement is an increase in CNR. However, clinical studies show that obese patients are

Page 98: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

97

successfully imaged on systems with a good temporal resolution (Brodoefel, et al., 2008).

Two different conclusions can be drawn from this:

a) the particular patients in the study could also have been imaged successfully on a

system with a lower temporal resolution; or

b) the clearer depiction of structures due to decreased blurring from cardiac motion,

on the system with improved temporal resolution, compensates for the poorer CNR

in obese patients.

Another example relates to the fact that the spatial resolution of a CT image is not only

dependent on the technical features of the equipment but is very much influenced by the

scan protocol adopted. Variation in the acquisition or reconstruction protocols between

centres is common and might lead to noticeable differences in the spatial resolution

achieved on a given CT scanner. Furthermore, motion artefacts, whose severity will vary

with the temporal resolution of the scanner, can significantly influence the ‘effective spatial

resolution’.

8.9 Technological advances

Step changes result from rapid technological developments in CT, and may make one or

more specification features selected for comparison inadequate or obsolete. For example,

until recently, a high x-ray flux was considered an important requirement for the successful

imaging of obese patients. However, with the introduction of iterative reconstruction

algorithms it may be possible to achieve a given image quality at approximately half the

radiation dose, so the need for a high specification in this respect becomes less relevant.

8.10 Patients with multiple conditions

Finally, it should be acknowledged that the definition of subgroups of difficult to image

patients overlooks the reality that patients may present with multiple ‘difficult to image’

characteristics (e.g. obese patients with high heart rate).

In summary, comparing specifications of CT scanners has value but also has limitations. If

these are taken into consideration technical specification data may provide preliminary

useful information to help with the selection of a suitable scanner for CCTA.

The conclusions drawn from this exercise should be validated with further studies that

provide metrics more directly linked with the clinical diagnostic performance of the

technology.

Page 99: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

98

9. Further work

Perform a systematic literature review on the current clinical performance of CT

scanners suitable for CCTA focusing on studies reporting on the relevant patient

subgroups;

As good, bias-free evidence from the literature is limited, the feasibility of

performing a retrospective study of CCTA examinations in the challenging patient

groups should be explored. The study would aim to quantify the limits of the

‘difficult factors’ on different scanner models. For example, what is the heart rate at

which clinical image quality is affected on a particular scanner? This analysis would

require detailed information on the scan protocols used and of patient

characteristics;

To gain a better understanding of the fundamental performance of CT systems in

CCTA examinations and obtain a complete set of unbiased data under controlled

conditions a study could be designed using appropriate dynamic cardiac phantoms

and a standard methodology. It is acknowledged that phantoms may not

adequately simulate the complex motion and structure of the coronary arteries.

However, these studies would be feasible within a relatively short period of time

(compared to clinical studies) and their findings could be valuable to formulate

hypotheses on which robust future clinical studies could be based; and

Develop and maintain a register of CT technical specifications to accompany and

maintain the currency of this report and inform future potential users/purchasers.

The undertaking of these further activities would allow a better understanding of the level

of difficulty posed by the subgroups of challenging patients. Additionally they would

provide important evidence on which to base conclusions as to the comparative diagnostic

performance of scanners used in cardiac CT.

Page 100: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

99

Appendix 1: Clinical evidence

A literature search was performed to identify relevant studies reporting on the clinical

performance of CCTA and the impact of technology developments particularly in imaging

difficult patient groups. The tables below summarise details of some of the studies.

A – Diagnostic performance of CCTA - Impact of iterative reconstruction techniques

Record # 1

Date: 2011

Authors: Moscariello et al

Publication: European Radiology 21:2130-2138

Title: Coronary CT angiography: image quality, diagnostic accuracy, and potential for

radiation dose reduction using a novel iterative image reconstruction technique –

comparison with traditional filtered back projection (FBP).

Details of the study:

Comparison of image noise, IQ and diagnostic accuracy of CCTA for CCT using IR software

(SAFIRE) compared to traditional filtered back projection (FBP). Estimate the potential for radiation

dose savings. 65 patients underwent CCTA and ICA. Full radiation dose data was reconstructed with

FBP and 50% of the projections (to simulate the dose) were used to reconstruct the data with the

alternative IR method. CCTA was performed with a CT scanner Definition Flash (Siemens) at 120,

100 and 80 kV for patients with BMI > 25kg/m2, <25kg/m2 and BMI<20 kg/m2, respectively.

Summary of outcomes:

IR significantly reduced image noise without loss of diagnostic information and holds potential for

substantial radiation dose reduction from CCTA. The per-patient accuracy, Se, Sp, PPV and NPV

results are summarised below.

Half dose SAFIRE Full dose FBP p-value

Diagnostic. accuracy 96.9 93.8

Se (%) 100 100

Sp (%) 94.6 89.2 0.001-0.025

PPV (%) 93.3 87.5

NPV (%) 100 100

Page 101: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

100

Record # 2

Date: 2011

Authors: Moscariello et al

Publication: European Radiology 21:2130-2138

Title: Coronary CT angiography: image quality, diagnostic accuracy, and potential for

radiation dose reduction using a novel iterative image reconstruction technique –

comparison with traditional filtered back projection (FBP).

Details of the study:

Comparison of image noise, IQ and diagnostic accuracy of CCTA for CCT using an IR software

(SAFIRE) compared to traditional filtered back projection (FBP). Estimate the potential for radiation

dose savings.

Sample of 65 patients underwent CCTA and ICA.

Full radiation dose data was reconstructed with FBP and 50% of the projections (simulation of 50% of

the dose) were used to reconstruct the data with the alternative IR method.

CCTA acquired with a CT scanner Definition Flash (Siemens) at 120, 100 and 80 kV for patients with

BMI > 25kg/m2, <25kg/m2 and BMI<20 kg/m2, respectively.

Summary of outcomes:

IR significantly reduced image noise without loss of diagnostic information and holds potential for

substantial radiation dose reduction from CCTA. The per-patient accuracy, Se, Sp, PPV and NPV

results are summarised below.

Half dose SAFIRE Full dose FBP p-value

Diagnostic. accuracy 96.9 93.8

Se (%) 100 100

Sp (%) 94.6 89.2 0.001-0.025

PPV (%) 93.3 87.5

NPV (%) 100 100

Page 102: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

101

B - Diagnostic performance of CCTA - Impact of scan mode

Record # 3

Date: 2013

Authors: Neefjes LA et al

Publication: European Radiology 23 (3): 614-2

Title: Diagnostic accuracy of 128-slice dual-source CT coronary angiography: a

randomized comparison of different acquisition protocols

Details of the study:

Comparison of diagnostic performance and radiation exposure of 128-slice dual source CCTA

protocols to detect coronary stenosis (> 50% obstruction).

459 symptomatic patients randomised between 2 groups

Group A: high pitch spiral vs. narrow window sequential (HR < 65 bpm)

Group B wide window sequential vs. retrospective spiral (HR > 65 bpm).

Diagnostic performance of CCTA was compared with coronary interventional angiography in 267

patients. The mean effective dose for each protocol is compared.

high pitch spiral vs. narrow window sequential (HR < 65 bpm)

All images acquired with Somatom Definition Flash, Siemens Healthcare.

Summary of outcomes:

Sequential CCTA should be used in patients with regular heart rates using 128-slice DSCT.

Diagnostic quality was found comparable in both groups. Pool estimates (for 95% CI) of per segment

Se, Sp, PPV and NPV of prospectively ECG-gating CCTA for diagnosis of coronary stenosis with

more than 50% obstruction are summarised below.

Low heart rate

(mean:58±7bpm)

High heart rate

(mean:75±11bpm)

high pitch

spiral

narrow window

sequential

wide

window

sequential

retrospective

spiral

Se (%) 89* 97* (p=0.01) 94 92

Sp (%) 95 96 95 95

PPV 62 73 67 66

NPV 99 100 99 99

Radiation dose

(mSv)

1.16 ± 0.60

3.82 ± 1.65

(p<0.001)

*6.12 ± 2.58

8.13 ± 4.52

(p<0.001)

Page 103: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

102

Record # 4

Date: 2012

Authors: Sun Z and Ng KH

Publication: International Journal of Cardiovascular 28 (8):2109-19

Title: Diagnostic value of coronary angiography with prospectively ECG-gating in the

diagnosis of coronary artery disease: a systematic review and meta-analysis.

Details of the study:

Meta-analysis of the diagnostic value or prospectively ECG-gating CCTA in the diagnosis of

significant stenosis.

14 studies included (910 patients, 3531 coronary arteries and 12056 segments)

Studies included CCTA performed with 64-slice single source CT, 64-slice dual source CT (first and

second generation) and 320-slice CT scanners were included.

Pooled Se, Sp, PPV and NPV were are summarised below.

Summary of outcomes:

CCTA with prospective ECG-gating allowed for reduced radiation exposure without sacrifice in IQ.

per patient mean values

Se (%) 99 95

Sp (%) 91 95

PPV (%) 94 88

NPV (%) 99 98

Effective dose (mSv) 3.3

Page 104: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

103

Record # 5

Date: 2012

Authors: Menke Jan, Unterberg-Buchwald C, Staab W, Sohns JM, Hosseini AAS, Schwarz A

Publication: American Heart Journal 165 (2):154–163.

Title: Head-to-head comparison of prospectively triggered vs retrospectively gated

coronary computed tomography angiography: meta-analysis of diagnostic accuracy,

image quality, and radiation dose.

Details of the study:

Meta-analysis including 20 studies (3330 patients) with suspected or known CAD with and without

tachyarrhythmia.

Comparison of IQ, diagnostic accuracy and radiation dose of prospectively triggered CCTA and

retrospectively gated CCTA for the diagnosis of ≥ 50% coronary stenosis compared with catheter

angiography.

CCTA performed with 64-slice CT or DSCT (no scanner models were specified)

Summary of outcomes:

Prospective triggered CCTA provides IQ and diagnostic accuracy comparable with retrospective

gated CTA but at much lower radiation dose.

Percentage of CCTAs with diagnostic quality (overall and for assessment of segment), and pooled

(664 patients; 5 studies) are summarised below as well as Sensitivity (Se), Specificity (Sp) for

diagnosis of significant coronary stenosis.

Prospective

triggering

Retrospective

gating

p-value

Diagnostic quality (%) 91.3 93.3 >0.05

Se (664 patients; 5 studies) (%) 98.7 96.7 >0.05

Sp (664 patients; 5 studies) (%) 91.3 95.8 >0.05

ED (mSv) 3.5 12.3 <0.01

Page 105: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

104

Record # 6

Date: 2010

Authors: Goetti et al

Publication: European Radiology 20(11):2565-71

Title: High-pitch dual source CT coronary angiography: systolic data acquisition at high

heart rates.

Details of the study:

Assessment of the effect of systolic acquisition for ECG-triggered high pitch CCTA on motion

artefacts of coronary arteries in patients with high heart rates (HR).

80 patients (HR ≥ 70 bpm) underwent CCTA on 128-slice DSCT (ECG-triggered, pitch=3.2) at 60%

(group A; 575 segments) or 30% (group B; 579 segments) of the RR interval. CCTA image quality

was graded using a 3 points scale (1-3). Radiation dose was assessed.

Summary of outcomes:

A systolic acquisition window for high-pitch dual source CCTA in patients with high HRs (≥ 70 bpm)

significantly improves coronary artery image quality at low radiation dose.

Group A Group B p-value

Segments with non-diagnostic

image quality (score 3) (%) 2.8 (16/579) 8.3% (48/575) 0.001

Effective dose (chest) (mSv) 2.3 ± 0.3

Page 106: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

105

C – Diagnostic performance of CCTA - Impact of motion correction software

Record # 7

Date: 2012

Authors: Leipsic et al

Publication: Journal of Cardiovascular Computed Tomography 6, 164-171

Title: Effects of a novel vendor-specific motion-correction algorithm on image quality and

diagnostic accuracy in persons undergoing coronary CT angiography without rate-

control medications.

Details of the study:

Comparison of IQ and diagnostic accuracy between standard and motion-corrected (Snapshoot

freeze; GE Healthcare) reconstructions.

Sample of 36 patients with severe aortic stenosis undergoing CCTA without heart rate control and

ICA as part of an evaluation for transcatheter aortic valve replacement.

All CCTA were performed with a 64-slice Discovery HD 750 High definition scanner (GE Healthcare).

All results were interpreted with and without motion correction using both 45% and 75% of the R-R

interval for reconstructions.

Summary of outcomes:

The use of the motion correction algorithm improved IQ, interpretability and diagnostic accuracy in

persons undergoing CCTA without rate-control medication. A summary of the results obtained is

presented below:

With motion

correction

Without motion

correction

p-value

Overall IQ (grade 1-4) 2.9 ± 0.9 2.4 ± 1.0 p<0.001

Per-segment

interpretability (%) 97 (392/406) 88 (357/406)

p<0.001

Per-artery interpretability

(%)

96 (128/134) 84 (112/234) p=0.002

Per-patient interpretability

(%)

92 (33/36) 89 (32/36) p=1.0

Per-reconstruction

diagnostic quality (%)

91 (370/406) 78% (317/406) p<0.001

Per-artery reconstruction

diagnostic quality (%)

86 (115/134) 72% (317/406) p=0.007

Per patient diagnostic

quality (%)

91 (370/406) 78% (317/406) p<0.001

Page 107: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

106

Record # 8

Date: 2013

Authors: Min et al

Publication: Journal of Cardiovascular Computed Tomography 7: 200-2006

Title: Rationale and design of the ViCTORY (Validation of an Intracycle CT Motion

CORrection Algorithm for Diagnostic AccuracY) trial

Details of the study:

Ongoing prospective international multicenter trial of 218 patients to determine whether motion

correction algorithms (MCAs) improve the diagnosis of obstructive CAD in patients undergoing

coronary CCTA who are not receiving β-blockers. Conventional ICA is used as the reference

standard. CCTA performed with 64-slice or more will be included.

Summary of outcomes:

Primary outcomes are per-patient diagnostic accuracy of MCAs for the diagnosis of anatomically

obstructive CAD compared with ICA. Secondary outcomes include other per-patient, per-vessel, and

per-segment diagnostic performance metrics. Diagnostic interpretability, image quality, the upper

heart rate threshold of utility of MCAs and the additive value to traditionally reconstructed CCTA will

also be derived.

Page 108: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

107

D - Diagnostic performance of CCTA on DSCT systems (variable HRs)

Record # 9

Date: 2014

Authors: Li M, Zhang JS, Jiang ZW et al

Publication: Clinical Radiology 69(2):163-71

Title: Diagnostic performance of dual source CT coronary angiography with and without

heart rate control: systematic review and meta-analysis.

Details of the study:

Investigate the diagnostic accuracy of DSCT CCTA with and without use of -blockers.

33 studies were included (patient sample ranged from 25 to 210).

CCTA was performed with 64-row DSCT and 128-row DSCT scanners.

Summary of outcomes:

DSCT coronary angiography without HR control showed a similar diagnostic performance at the

patient level as that of heart rate control groups.

A summary of the results for Sensitivity, Specificity, Positive and negative likelihood ratios is

presented below. At per segment level a significant decrease in specificity was noticed without HR

control.

Total HR controlled HR not controlled

per

patient

per

segment

per

patient

per

segment

per

patient

per

segment

Se (%) 98 95 99 95 98 94

Sp (%) 88 97* 89 98 86 96*

(p=0.03)

PLR (%) 8.04 35.19 8.89 55.35 7.16 26.54

NLR (%) 0.02 0.05 0.01 0.05 0.03 0.06

ED (mSv) 2.6 1.6 8

Page 109: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

108

E - Diagnostic performance of CCTA - stent imaging

Record # 10

Date: 2008

Authors: Sun Z, Almutairi AMD

Publication: European Journal of Radiology 73 (2010) 266-273

Title: Diagnostic accuracy of 64 multislice CT angiography in the assessment of coronary

in-stent restenosis: a meta-analysis

Details of the study:

Meta-analysis of the diagnostic accuracy of 64-slice CT angiography for the detection of coronary in-

stent restenosis (> 50%) when compared with conventional ICA.

Summary of outcomes:

The mean value of assessable stents was 89%. 64 MDCT showed high diagnostic value (both

sensitivity and specificity) for detection of coronary in-stent restenosis (based on assessable

segments) when compared with CCTA. Stent diameter is the main parameter affecting the diagnostic

value of 64-slice CCTA.

Assessable stents only inclusion of non-assessable stents

(found in 5 studies)

Se (%) 90 79

Sp (%) 91 81

Page 110: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

109

F – Diagnostic performance of CCTA - effect of calcium score

Record # 11

Date: 2012

Authors: den Decker AM, de Smet K, de Bock GH et al

Publication: European Radiology ;22(12):2688-98.

Title: Diagnostic performance of coronary CT angiography for stenosis detection according

to calcium score: systematic review and meta-analysis.

Details of the study:

Systematic review (51 studies) and meta-analysis (27 studies) to assess the diagnostic accuracy of

CCTA for significant stenosis at different degrees of coronary calcification. ICA used as reference

standard. CCTA performed with at least 16-slices were included.

Summary of outcomes:

Sensitivity and specificity of 16-slide MDCT were significantly lower than with more modern scanners.

With 64-MDCT and newer CT scanner systems a CS cut-off for performing CCTA no longer seems

indicated.

Calcium score Se (%) Sp (%)

0 -100 95.8 91.2

101 – 400 95.6 88.2

401 – 1000 97.6 50.6*

> 1000 99.0 84.0

*significantly lower due to lack of patients with significant stenosis

Page 111: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

110

G – Diagnostic performance of CCTA performance – effect of heart rate variability

Record # 12

Date: 2013

Authors: Vorre MM and Addulla J

Publication: Radiology ;267(2):376-86

Title: Diagnostic accuracy and radiation dose of CT coronary angiography in atrial

fibrillation: systematic review and meta-analysis

Details of the study:

Systematic review to compare CCTA with conventional ICA in patients with arterial fibrillation.

Diagnostic accuracy for coronary stenosis (≥ 50%) and radiation dose were compared.

6 studies comparing CCTA and ICA and additional 7 assessed CCTA in patients with arterial

fibrillation. CCTA performed with 64-raw scanners at a minimum was considered (DSCT and 320-row

systems were included).

Summary of outcomes:

CCTA has demonstrated high diagnostic accuracy in patients with arterial fibrillation but is associated

with significantly higher radiation dose than that in patients with sinus rhythm. Results for diagnostic

accuracy of CCTA in the studied population are summarised below. No statistical significant

differences were found between the studies included in the analysis (p>0.05).

per patient

(7 studies)

per segment

(5 studies)

Se (%) 94 84.5

Sp (%) 91 98.5

PPV (%) 79 76

NPV (%) 98 99

Page 112: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

111

H – Diagnostic performance of CCTA - effect of heart rate

Record # 13

Date: 2014

Authors: Li M, Zhang JS, Jiang ZW et al

Publication: Clinical Radiology 69(2):163-71

Title: Diagnostic performance of dual source CT coronary angiography with and without

heart rate control: systematic review and meta-analysis.

Details of the study:

Investigate the diagnostic accuracy of DSCT CCTA with and without use of -blockers.

33 studies were included (patient sample ranged from 25 to 210).

CCTA was performed with 64-row DSCT and 128-row DSCT scanners.

Summary of outcomes:

DSCT coronary angiography without HR control showed a similar diagnostic performance at the

patient level as that of heart rate control groups.

A summary of the results for sensitivity, specificity, positive and negative likelihood ratios is

presented. At per segment level a significant decrease in specificity was noticed without HR control.

Total HR controlled HR not controlled

per

patient

per

segment

per

patient

per

segment

per

patient

per

segment

Se (%) 98 95 99 95 98 94

Sp (%) 88 97* 89 98 86 96*

(p=0.03)

PLR (%) 8.04 35.19 8.89 55.35 7.16 26.54

NLR (%) 0.02 0.05 0.01 0.05 0.03 0.06

ED (mSv) 2.6 1.6 8

Page 113: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

112

Appendix 2: Questionnaire used to collect technical

specs of CT scanners

Page 114: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

113

Questionnaire used to collect technical specs of CT scanners (contd)

Page 115: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

114

Questionnaire used to collect technical specs of CT scanners (contd)

Page 116: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

115

Appendix 3: References

Abdulla J, Pedersen KS, Budoff M, Kofoed KF. Influence of coronary calcification on the

diagnostic accuracy of 64-slice computed tomography coronary angiography: a

systematic review and meta-analysis. Int J Cardiovasc Imaging 2012;28:943-53

Andreini D, Pontone G, Mushtaq S, Bartorelli AL, Bertella E, Trabattoni D, Montorsi P, et

al. Coronary in-stent restenosis: assessment with CT coronary angiography, Radiology

2012;265 (2):410-7.

Apfaltrer P, Schoendube H, Schoepf UJ, Allmendinger T, Tricarico F, Schindler A, Vogt

S, et al. Enhanced temporal resolution at cardiac CT with a novel CT image

reconstruction algorithm: initial patient experience. European Journal of Radiology

2013;82 (2):270-4.

Aunt Minnie, Report from 2012 meeting of Society of Cardiovascular Computed

Tomography (SCCT),Baltimore. 2012, Dual-energy CTA may offer more accurate look at

plaque, July 24, 2012

http://www.auntminnie.com/index.aspx?sec=sup&sub=cto&pag=dis&ItemID=100032

[Accessed 11 March 2014]

Aunt Minnie, Wide-area CT, iterative recon cut dose in CABG evaluations, January 31,

2014

http://www.auntminnieeurope.com/index.aspx?sec=prtf&sub=def&pag=dis&itemId=6092

80&printpage=true&fsec=ser&fsub=def [Accessed 11 March 2014]

Blankstein R, Bolen MA, Pale R, Murphy MK, Shah AB, Bezerra HG, Sarwar A, Rogers

IS, Hoffmann U, Abbara S, Cury RC, Brady TJ. Use of 100 kV versus 120 kV in cardiac

dual source computed tomography: effect on radiation dose and image quality. Int J

Cardiovasc Imaging 2011;27(4):579-86

British Heart Foundation, Heart Rhythms. 2012

http://www.bhf.org.uk/publications/view-publication.aspx?ps=1001001 [Accessed 09

February 2014]

Bongartz G., Golding SJ, Jurik AG, Leonardi, van Meerten, E Rodríguez R, Schneider K,

Calzado A, Geleijns J, Jessen KA, Panzer W, Shrimpton PC, Tosi G, European

Guidelines for Multislice Computed Tomography Funded by the European Commission

Contract number FIGM-CT2000-20078-CT-TIP March 2004

Page 117: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

116

BSCI - British Society of Cardiovascular imaging, CT Coronary Angiography Service

Specification, May 2012.

Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E, Scherer M, Bellinger

R, Martin A, Benton R, Delago A, Min JK. Diagnostic performance of 64-multidetector

row coronary computed tomographic angiography for evaluation of coronary artery

stenosis in individuals without known coronary artery disease: results from the

prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic

Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll

Cardiol. 2008;52(21):1724-32.

Carrascosa P, Deviggiano A, Capunay C, De Zan MC, Goldsmit A, Rodriguez-Granillo

GA. Effect of intracycle motion correction algorithm on image quality and diagnostic

performance of computed tomography coronary angiography in patients with suspected

coronary artery disease. Acad Radiol. 2015;22:81-6

Cao JX1, Wang YM2, Lu JG3, Zhang Y1, Wang P1, Yang C1. Radiation and contrast

agent doses reductions by using 80-kV tube potential in coronary computed tomographic

angiography: a comparative study. Eur J Radiol.2014;83(2):309-14.

Chao P-S, Law W-L, Kuo C-J, Hung H-F et al. The diagnostic accuracy of 256-row

computed tomographic angiography compared with invasive coronary angiography in

patients with suspected coronary artery disease. European Heart Journal 2010; 31:

19916-23.

Christner JA, Kofler JM, McCollough CH Estimating Effective Dose for CT Using Dose–

Length Product Compared With Using Organ Doses: Consequences of Adopting

International Commission on Radiological Protection Publication 103 or Dual-Energy

Scanning American Journal of Roentgenology.2010;194: 881-889.

Coronary heart disease statistics. A compendium of health statistics. 2012 edition British

Heart Foundation Health Promotion Research Group Department of Public Health,

University of Oxford .

Deak PD, Langner O, Lell M, Kalender WA. Effects of adaptive section collimation on

patient radiation dose in multisection spiral CT. Radiology. 2009;252(1):140-7.

den Dekker MA, de Smet K, de Bock GH, Tio RA, Oudkerk M and Vliegenthart R.

Diagnostic performance of coronary CT angiography for stenosis detection according to

calcium score: systematic review and meta-analysis European Radiology 2012;22(12):

2688-98.

Page 118: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

117

Desjardins B, Kazerooni EA, ECG-Gated Cardiac CT, American Journal of

Roentgenology. 2004;182:993-1010

Ebersberger U, Tricarico F, Schoepf UJ, Blanke P, Spears JR, Rowe GW, Halligan WT,

et al. CT evaluation of coronary artery stents with iterative image reconstruction:

improvements in image quality and potential for radiation dose reduction. European

Radiology 2013;23(1):125-32.

Einstein AJ, Effects of Radiation Exposure From Cardiac Imaging. How Good Are the

Data? Journal of the American College of Cardiology 2012;59 (6)

Einstein AJ, Elliston CD, Arai AE, Chen MY, Mather R, Pearson GD, Delapaz RL,

Nickoloff E, Dutta A, Brenner DJ. Radiation dose from single-heartbeat coronary CT

angiography performed with a 320-detector row volume scanner. Radiology.

2010;254(3):698-706

Einstein AJ1, Knuuti J. Cardiac imaging: does radiation matter? Eur Heart J. 2012;

33(5):573-8

Fleischmann D, Boas FE. Computed tomography -- old ideas and new technology.

Eur Radiol. 2011 Mar;21(3):510-7.

Fuchs TA, Stehli J, Bull S, Dougoud S, Clerc OF, Herzog BA, Buechel RR, Gaemperli O

and Kaufmann PA, Coronary computed tomography angiography with model-based

iterative reconstruction using a radiation exposure similar to chest X-ray Examination,

European Heart Journal. 2014; 35:1131–1136

GE Healthcare. SnapShot Freeze Motion – corrected CCTA. CT clinical case study –

cardiac CT, 2012.

http://www3.gehealthcare.com/~/media/Downloads/us/Product/Product-

Categories/Computed-Tomography/Discovery-CT-750-HD/GEHealthcare-Case-

Study_SnapShot_Freeze_2012.pdf [Accessed 28 February 2014]

Gebhard C, Fuchs TA, Fiechter M, Stehli J et al. Image quality of low-dose CCTA in

obese patients: impact of high definition computed tomography and adaptive statistical

iterative reconstruction. Int J Cardiovasc Imaging 2013:29:1565-1574

Goetti R, Feuchtner G, Stolzmann P, Desbiolles L, Fischer MA, Karlo C, Baumueller S,

Scheffel H, Alkadhi H, Leschka S. High-pitch dual-source CT coronary angiography:

systolic data acquisition at high heart rates. Eur Radiol. 2010;20(11):2565-71

Page 119: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

118

Goetti R, Leschka S, Baumuller S, Plass A et al. Low dose high-pitch spiral acquisition

128-slice dual-source computed tomography for the evaluation of coronary artery bypass

graft patency. Investigative Radiology 2010;45(6):324-30

Goo HWG, CT Radiation Dose Optimization and Estimation: an Update for Radiologists.

Korean J Radiology 2012;13(1):1-11

Hassan A, Nazir SA, Alkadhi H. Technical challenges of coronary CT angiography: today

and tomorrow. Eur J Radiol. 2011;79(2):161-71.

Hausleiter J, Meyer T, Hadamitzky M, Huber E, Zankl M, Martinoff S, Kastrati A, Schömig

A. Radiation dose estimates from cardiac multislice computed tomography in daily

practice: impact of different scanning protocols on effective dose estimates.2006.

Circulation. 2006;113(10):1305-10.

Hausleiter J. Meyer T. Hermann F. Hadamitzky M. Krebs M. Gerber TC. McCollough C.

Martinoff S. Kastrati A. Schomig A. Achenbach S. Estimated radiation dose associated

with cardiac CT angiography. JAMA.2009;301(5):500-7

Health survey 2012 http://www.hscic.gov.uk/catalogue/PUB13218/HSE2012-Ch10-Adult-

BMI.pdf

Hou Y, Yue Y, Guo W, Feng G, Yu T, Li G, Vembar M, Olszewski ME and Guo Q.

Prospectively versus retrospectively ECG-gated 256-slice coronary CT angiography:

image quality and radiation dose over expanded heart rates. The International Journal of

Cardiovascular Imaging 2012;28 (1):153-62.

Huang B, Li J, Law MW, Zhang J, Shen Y, Khong PL. Radiation dose and cancer risk in

retrospectively and prospectively ECG-gated coronary angiography using 64-slice

multidetector CT. British Journal of Radiology 2010;83(986):152-8

ImPACT Group. Market review: advanced CT scanners for coronary angiography, s.l.:

(2010) Centre for Evidence-based Purchasing.

International Commission on Radiological Protection. 1990 recommendations of the

International Commission on Radiological Protection. Ann ICRP 1991; 21: publication no.

60.

International Commission on Radiological Protection. 2007 recommendations of the

International Commission on Radiological Protection. Ann ICRP 2007; 37: publication no.

103.

Page 120: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

119

International Commission on Radiological Protection. Recommendations of the

International Commission on Radiological Protection. Oxford, UK: Pergamon Press,

1977:ICRP publication no. 26.

Jiang H, Vartali J, Vess C. White paper. Gemstone – The Ultimate Scintillator for

Computed Tomography. November, GE Healthcare, USA, 2008

Jones CM, Chin KY, YangG-Z, Hamady M, Darzi A, Athanasiou T. Coronary Artery

Bypass Graft Imaging with 64-Slice Multislice Computed Tomography: Literature Review.

Seminars in Ultrasound CT MRI.2008;29:204-213.

Jun BR, Yong HS, Kang EY, Woo OH, Choi EJ. 64-slice coronary computed tomography

angiography using low tube potential of 80 kV in subjects with normal body mass indices:

comparative study using 120 kV. Acta Radiol. 2012;53(10):1099-106.

Kalender WA, Deak P, Kellermeier M, van Straten M and Vollmar SV. 2009. Application-

and patient size-dependent optimization of x-ray spectra for CT. Medical Physics 2009;

36(3):993-1007.

Kim HY, Lee JW, Hong YJ, Lee HJ, Hur J, Nam JE, Choi BW and Kim YJ. Dual-source

coronary CT angiography in patients with high heart rates using a prospectively ECG-

triggered axial mode at end-systole. The International Journal of Cardiovascular Imaging

2012;28(Suppl 2):101-7.

KiTEC, CT scanner technology – a discussion of its application in cardiac imaging with

particular emphasis on ‘difficult to image’ patients. November, NICE internal report, 2013.

Kleijnen Systematic Reviews Ltd Assessment Group. Diagnostic Assessment Report:

Computed tomography (CT) scanners for cardiac imaging - Somatom Definition Flash,

Aquilion One, Brilliance iCT and Discovery CT750 HD, NICE 2011.

LaBounty T., Leipsic J., Poulter R., Wood D.; Heilbron B. Srichai BM. Taylor C., Johnson

M.K. Mancini G.B.J. Mayo J. Dunning A., Lin F.Y, Gomez M.Weinsaft J.W. Min J.K. The

impact of reduced 80 kvp tube potential on coronary ct angiography by ct platform:

results from a prospective, multicenter, multivendor randomized trial. J Am Coll Cardiol.

2011;57(14s1):E671-E671.

LaBounty TM, Earls JP, Leipsic J, Heilbron B, Mancini GB, Lin FY, Dunning AM, Min JK,

Effect of a standardized quality-improvement protocol on radiation dose in coronary

computed tomographic angiography. Am J Cardiol. 2010;106(11):1663-7 (a)

Page 121: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

120

LaBounty TM, Leipsic J, Mancini GB, Heilbron B, Patel S, Kazerooni EA, Sundaram B,

Lin FY, Dunning A, Saltzman AJ, Weisz G, Weinsaft JW, Choi JH, Koduru S, Min JK.

Effect of a standardized radiation dose reduction protocol on diagnostic accuracy of

coronary computed tomographic angiography. Am J Cardiol. 2010;106(2):287-92 (b)

Lee AM, Beaudoin J, Engel LC, Sidhu MS, Abbara S, Brady TJ, Hoffmann U and

Ghoshhajra BB. Assessment of image quality and radiation dose of prospectively ECG-

triggered adaptive dual-source coronary computed tomography angiography (cCTA) with

arrhythmia rejection algorithm in systole versus diastole: a retrospective cohort study.

The International Journal of Cardiovascular Imaging 2013;29(6):1361-70

Lee TY and Chhem RK. Impact of new technologies on dose reduction in CT (2010) Eur

J Radiol. 2010 Oct;76(1):28-35

Leipsic J, Labounty TM, Hague CJ, Mancini GBJ et al. Effect of a novel vendor-specific

motion-correction algorithm on image quality and diagnostic accuracy in persons

undergoing coronary CT angiography without rate-control medications. J Cardiovasc

Computed Tomography 2012;6:164-171

Lim MJ, White CJ. Coronary angiography is the gold standard for patients with significant

left ventricular dysfunction. Prog Cardiov Dis. 2013;55 (5):504-8.

Maintz D, Burg MC, Seifarth H, Bunck AC, Ozgun M, Fischbach R, Jurgens KU and

Heindel W. Update on multidetector coronary CT angiography of coronary stents: in vitro

evaluation of 29 different stent types with dual-source CT. European Radiology 2009;19

(1): 42-9.

Mark DB, Berman DS, Budoff MJ, Carr JJ et al.

ACCF/ACR/AHA/NASCI/SAIP/SCAI/SCCT 2010 Expert Consensus Document on

Coronary Computed Tomographic Angiography: A Report of the American College of

Cardiology Foundation Task Force on Expert Consensus Documents Circulation.2010;

121:2509-2543.

McCollough, C. et al.,Strategies for Reducing Radiation Dose in CT. Radiologic Clinics of

North America, 2009;1:27-40.

McCollough CH, Primak AN, Saba O, et al. Dose performance of a 64-channel dual-

source CT scanner Radiology 2007; 43:775–784.

McCollough, C.H. Primak H.N. Braun N.Kofler J, Yu L., Christner J. Strategies for

Reducing Radiation Dose in CT. Radiologic Clinics of North America.2009;47(1):27–40.

Page 122: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

121

Meijboom WB, Meijs MF, Schuijf JD et al Diagnostic accuracy of 64-slice computed

tomography coronary angiography: a prospective, multicenter, multivendor study. J Am

Coll Cardiol 2008;52(25):2135-44.

Menke J, Unterberg-Buchwald C, Staab W, Sohns JM, Seif Amir Hosseini A, Schwarz A.

Head-to-head comparison of prospectively triggered vs retrospectively gated coronary

computed tomography angiography: meta-analysis of diagnostic accuracy, image quality,

and radiation dose. American Heart Journal 2013;165(2):154-163.e3.

Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I, Paul N, Clouse

ME, Shapiro EP, Hoe J, Lardo AC, Bush DE, de Roos A, Cox C, Brinker J, Lima JA.

Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 2008;

359(22):2324-36.

Min JK, Arsanjani R, Kurabayashi S, Andreini D et al. Rationale and design of the

ViCTORY (Validation of an Intracycle CT Motion CORrection Algorithm for Diagnostic

AccuracY) trial. J Cardiovascular Comp Tom 2013;7:200-6.

Moscariello A, Takx RA, Schoepf UJ et al. Coronary CT angiography: image quality,

diagnostic accuracy, and potential for radiation dose reduction using a novel iterative

image reconstruction technique-comparison with traditional filtered back projection.

European Radiology 2010;21:2130-8.

Mowatt G, Cook JA, Hillis GS, Walker S, Fraser C, Jia X, Waugh N. 64-Slice computed

tomography angiography in the diagnosis and assessment of coronary artery disease:

systematic review and meta-analysis. Heart 2008;94(11):1386-93.

Muenzel D, Noel PB, Dorn F, Dobritz M, Rummeny EJ and Huber A. Step and shoot

coronary CT angiography using 256-slice CT: effect of heart rate and heart rate variability

on image quality. European Radiology 2011;21(11):2277-84.

Neefjes LA, Rossi A, Genders TS, Nieman K, Papadopoulou SL, Dharampal AS, Schultz

CJ, et al. Diagnostic accuracy of 128-slice dual-source CT coronary angiography: a

randomized comparison of different acquisition protocols European Radiology 2013;23

(3):614-22.

NICE Clinical Guidance 95: Chest pain of recent onset, NICE 2010.

NICE Diagnostic Guidance 3: New generation cardiac CT scanners for cardiac imaging

in people with suspected or known coronary artery disease in whom imaging is difficult

with earlier generation CT scanners, NICE 2012.

Page 123: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

122

NICOR: National Institute for cardiovascular outcomes research. National Adult Cardiac

Surgery Audit. Annual Report 2010 – 2011.

http://www.ucl.ac.uk/nicor/audits/Adultcardiacsurgery/publications/pdfs/nacsa_report_201

1-2012 [Accessed 28th Feb. 2014).

Oda S, Utsunomiya D, Funama Y, Takaoka H, Katahira K, Honda K, Noda K, Oshima S

and Yamashita Y. Improved coronary in-stent visualization using a combined high-

resolution kernel and a hybrid iterative reconstruction technique at 256-slice cardiac CT-

Pilot study. European Journal of Radiology 2013;82(2):288-95.

Paterick TE, Jan MF, Paterick ZR, Tajik AJ, Gerber TC, Arbor, Ann, Cardiac Imaging

Modalities With Ionizing Radiation The Role of Informed Consent.Journal of the

American College of Cardiology 2012;Vol. 5, No.6

Philips Healthcare. White paper – iDose4 iterative reconstruction technique; Koninklijke

Philips Electronics N.V.The Netherlands, 2011

Renker M, Nance JW Jr, Schoepf UJ, O'Brien TX, Zwerner PL, Meyer M, Kerl JM, et al.

Evaluation of heavily calcified vessels with coronary CT angiography: comparison of

iterative and filtered back projection image reconstruction. Radiology 2011;260(2):390-9.

Rief M, Zimmermann E, Stenzel F, Martus P et al. Computed Tomography Angiography

and Myocardial Computed Tomography Perfusion in Patients With Coronary Stents. J

Am Coll Cardiol. 2013;62(16):1476-1485

Scheffel H, Alkadhi H and Leschka S. High-pitch dual-source CT coronary angiography:

systolic data acquisition at high heart rates. European Radiology 2012;20(11): 2565-71.

Scheffel H, Stolzmann P, Schlett CL, Engel LC, Major GP, Karolyi M, Do S, Maurovich-

Horvat P and Hoffmann U. Coronary artery plaques: cardiac CT with model-based and

adaptive-statistical iterative reconstruction technique. European Journal of Radiology

2012;81(3):e363-9.

Schuetz G.M., Schlattmann P. and Dewey M. Coronary CT angiography cannot be

recommended in patients with atrial fibrillation. Radiology 2013;269(3):947-948.

Scrichai MB, Barreto M, Lim RP, Donnino R et al. Prospective-triggered sequential dual-

source end-systolic coronary CT angiography for patients with atrial fibrillation:

A feasibility study. J Cardiovasc Comput Tomogr 2013;7(2):102-9.

Stefan Ulzheimer and Jan Freund. White paper. The Stellar Detector, Siemens,

Germany, Nov 2012.

Page 124: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

123

Sun G, Li M, Jiang XS, Li L, Peng ZH, Li GY and Xu L. 320-detector row CT coronary

angiography: effects of heart rate and heart rate variability on image quality, diagnostic

accuracy and radiation exposure. British Journal of Radiology 2012;85(1016):e388-94.

Sun G, Li M, Jiang ZW, Xu L, Peng ZH, Ding J, Li L and Jin ZT. 2013. Diagnostic

accuracy of dual-source CT coronary angiography in patients with atrial fibrillation: meta

analysis. European Journal of Radiology 2013;82(10):1749-54.

Sun K, Li G, Han R & Gang W, 2011. Flash Scanning of Coronary CTA with just 0.3

mSv. Available online at:

http://health.siemens.com/ct_applications/somatomsessions/index.php/flash-scanning-of-

coronary-cta-with-just-0-3-msv/

Sun ML, Lu B, Wu RZ, Johnson L, Han L, Liu G, Yu FF, et al. Diagnostic accuracy of

dual-source CT coronary angiography with prospective ECG-triggering on different heart

rate patients. European Radiology 2011; 21 (8): 1635-42.

Sun Z and Ng KH. Prospective versus retrospective ECG-gated multislice CT coronary

angiography: a systematic review of radiation dose and diagnostic accuracy. European

Journal of Radiology 2012;81(2):e94-100.

Sun Z, Almutairi AM. Diagnostic accuracy of 64 multislice CT angiography in the

assessment of coronary in-stent restenosis: a meta-analysis. European Journal of

Radiology 2010;73(2):266-273.

Sun Z, Choo GH, Ng KH> Coronary CT angiography: current status and continuing

challenges; British Journal of Radiology 2012;85:495-510

Tanaka R, Yoshioka K, Muranaka K, Chiba T, Ueda T, Sasaki T, Fusazaki T and Ehara

S. Improved evaluation of calcified segments on coronary CT angiography: a feasibility

study of coronary calcium subtraction. The International Journal of Cardiovascular

Imaging 2013;29(Suppl 2):75-81.

Taylor AJ, Cerqueira M, Hodgson JM, et al. ACCF/SCCT/ACR/

AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 appropriate use criteria for cardiac computed

tomography: a report of the American College of Cardiology Foundation Appropriate Use

Criteria Task Force, the Society of Cardiovascular Computed Tomography, the American

College of Radiology, the American Heart Association, the American Society of

Echocardiography, the American Society of Nuclear Cardiology, the North American

Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and

Interventions, and the Society for Cardiovascular Magnetic Resonance. J AmColl Cardiol

2010;56:1864 –94.

Page 125: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

124

Tomizawa N, Komatsu S, Akahane M, Torigoe R, Kiryu S and Ohtomo K. Relationship

between beat to beat coronary artery motion and image quality in prospectively ECG-

gated two heart beat 320-detector row coronary CT angiography. International Journal of

Cardiovascular Imaging 2012; 28(1):139-46.

Torres FS, Jeddiyan S, Jimenez-Juan L and Nguyen ET. Beta-blockers to control heart

rate during coronary CT angiography. Radiology 2011;259 (2):615-6

Toshiba Medical Systems. Toshiba’s PUREVision Detector. 2014. http://www.toshiba-

medical.eu/eu/wp-content/uploads/sites/2/2015/01/Toshibas-pure-ViSION-Detector-

2014-aquilion.pdf [Accessed 9th Feb 2015]

Toshiba Medical Systems Europe. Toshiba introduces ‘SureSubtractionCoronary at ESC

2014. Posted August 11 2014. http://www.toshiba-medical.eu/eu/toshiba-introduces-

suresubtraction-coronary-at-esc-2014/ [Accessed 13th Feb 2015]

Uehara M, Takaoka H, Kobayeshi Y, Funabashi N. Diagnostic accuracy of 320-slice

computed-tomography for detection of significant coronary artery stenosis in patients with

various heart rates and heart rhythms compared with conventional coronary-

angiography. Int J Cardiol. 2013;167(3):809-15.

Vorre MM and Abdulla J. Diagnostic accuracy and radiation dose of CT coronary

angiography in atrial fibrillation: systematic review and meta-analysis. Radiology 2013;

267(2):376-86.

Wang Q, Qin J, He B, Zhou Y et al. Computed tomography coronary angiography with a

consistent dose below 2 mSv using double prospectively ECG-triggered high-pitch spiral

acquisition in patients with atrial fibrillation: initial experience. Int J Cardiovasc Imaging.

2013;29:1341:9.

West, Amy and Beller, George. 256- and 320-row coronary CTA: is more better?.

European Heart Journal 2010;31(15):1823-1825.

Westwood M, Al M, Burgers L, Redekop K, Lhachimi S, Armstrong N, Raatz H, Misso K,

Severens J, Kleijnen J. A systematic review and economic evaluation of new-generation

computed tomography scanners for imaging in coronary artery disease and congenital

heart disease: Somatom Definition Flash, Aquilion ONE, Brilliance iCT and Discovery

CT750 HD Health Technol Assess 2013;17(9):1-243.

Westwood ME, Raatz HD, Misso K, Burgers L, Redekop K, Lhachimi SK, Armstrong N,

Kleijnen J. Systematic review of the accuracy of dual-source cardiac CT for detection of

arterial stenosis in difficult to image patient groups. Radiology, 2013;267(2):387-95

Page 126: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

125

Weustink AC, Neefjes LA, Kyrzopoulos S, van Straten M, Neoh Eu R, Meijboom WB, van

Mieghem CA, et al. Impact of heart rate frequency and variability on radiation exposure,

image quality, and diagnostic performance in dual-source spiral CT coronary

angiography Radiology 2009;253(3):672-80.

Wexler L, Gaining Perspective on the Risks of Ionizing Radiation for Cardiac Imaging.

Journal of the American College of Cardiology. (accepted for publication in December

2013) DOI: 10.1016/j.jacc.2013.12.035

Xu L, Yang L, Zhang Z, Li Y, Fan Z, Ma X, Lv B and Yu W. Low-dose adaptive sequential

scan for dual-source CT coronary angiography in patients with high heart rate:

comparison with retrospective ECG gating. European Journal of Radiology 2010;76(2):

183-7

Xu L, Yang L, Fan Z, Yu W, Lv B and Zhang Z. Diagnostic performance of 320-detector

CT coronary angiography in patients with atrial fibrillation: preliminary results. European

Radiology 2011;21 (5):936-43.

Xu L, Yang L, Zhang Z, Wang Y, Jin Z, Zhang L and Lu G. Prospectively ECG-triggered

sequential dual-source coronary CT angiography in patients with atrial fibrillation:

comparison with retrospectively ECG-gated helical CT. European Radiology 2013;23

(7):1822-8.

Yoo RE, Park EA, Lee W, Shim H, Kim YK, Chung JW and Park JH. Image quality of

adaptive iterative dose reduction 3D of coronary CT angiography of 640-slice CT:

comparison with filtered back-projection. The International Journal of Cardiovascular

Imaging 2013;29(3):669-76.

Page 127: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

126

Appendix 4: SNOMED-CT codes CONCEPTID TERM DESCRIPTION TYPE

49436004 Atrial fibrillation Preferred Description

49436004 Atrial fibrillation (disorder) Fully Specified Name

49436004 AF - Atrial fibrillation Synonym

60621009 Body mass index Preferred Description

60621009 Body mass index (observable entity) Fully Specified Name

60621009 Weight: body mass Synonym

60621009 BMI - Body mass index Synonym

60621009 Quetelet index Synonym

258983007 beats/min Preferred Description

258983007 beats/min (qualifier value) Fully Specified Name

258983007 beats per minute Synonym

258983007 BPM - beats per minute Synonym

232717009 Coronary artery bypass graft Preferred Description

232717009 Coronary artery bypass grafting (procedure) Fully Specified Name

232717009 CABG - Coronary artery bypass graft Synonym

232717009 CBG - Coronary bypass graft Synonym

232717009 Coronary artery bypass grafting Synonym

232717009 Coronary artery bypass graft operations Synonym

232717009 CAG - Coronary artery graft Synonym

53741008 Coronary arteriosclerosis Preferred Description

53741008 Coronary arteriosclerosis (disorder) Fully Specified Name

53741008 Arteriosclerotic heart disease Synonym

53741008 CAD - Coronary artery disease Synonym

53741008 Coronary artery disease Synonym

53741008 Coronary sclerosis Synonym

53741008 CHD - Coronary heart disease Synonym

53741008 Coronary heart disease Synonym

Page 128: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

127

77477000 Computerised axial tomography Preferred Description

77477000 Computerized axial tomography (procedure) Fully Specified Name

77477000 CAT scan Synonym

77477000 Computerised transaxial tomography Synonym

77477000 Computerised tomography Synonym

77477000 CAT - Computerised axial tomography Synonym

77477000 CT - Computerised tomography Synonym

77477000 Computerised tomograph scan Synonym

77477000 Computed axial tomography Synonym

77477000 Computed tomography Synonym

418272005 CT angiography Preferred Description

418272005 Computed tomography angiography (procedure) Fully Specified Name

418272005 Computed tomography angiography Synonym

419545005 CT angiography of coronary arteries Preferred Description

419545005

Computed tomography angiography of coronary

arteries (procedure) Fully Specified Name

419545005

Computed tomography angiography of coronary

arteries Synonym

450360000 Coronary artery calcium score Preferred Description

450360000 Coronary artery calcium score (observable entity) Fully Specified Name

364075005 Heart rate Preferred Description

364075005 Heart rate (observable entity) Fully Specified Name

364075005 Cardiac rate Synonym

33367005 Coronary angiography Preferred Description

33367005 Coronary angiography (procedure) Fully Specified Name

33367005 Angiography of coronary arteries Synonym

33367005 Coronary arteriography Synonym

33367005 Coronary angiogram Synonym

33367005 Coronary arteriogram Synonym

415070008 Percutaneous coronary intervention Preferred Description

415070008 Percutaneous coronary intervention (procedure) Fully Specified Name

17366009 Atrial arrhythmia Preferred Description

17366009 Atrial arrhythmia (disorder) Fully Specified Name

Page 129: CT scanners for coronary CT angiography (CCTA) in ...kitec.co.uk/perch/resources/cardiac-report-for-kitec-website.pdf · CT SCANNERS FOR CARDIAC IMAGING .....45 6.1 CT scanner models

128

44103008 Ventricular arrhythmia Preferred Description

44103008 Ventricular arrhythmia (disorder) Fully Specified Name

15772006 Beta 1 blocking agent Preferred Description

15772006 Beta 1 blocking agent (product) Fully Specified Name

15772006 Beta 1 adrenergic blocking agent Synonym

15772006 Cardioselective beta-blocker Synonym

15772006 Beta 1 blocking product Synonym

65818007 Stent Preferred Description

65818007 Stent, device (physical object) Fully Specified Name

65818007 Stent, device Synonym

303490004 Cardiovascular implant Preferred Description

303490004 Cardiovascular implant (physical object) Fully Specified Name

414916001 Obesity Preferred Description

414916001 Obesity (disorder) Fully Specified Name

414916001 Adiposis Synonym

414916001 Adiposity Synonym