Dr. Sandra Klein, 06/20071
Design and Calibration of aDesign and Calibration of a
Dissolution Test EquipmentDissolution Test Equipment
Training Workshop on Dissolution, Training Workshop on Dissolution,
Pharmaceutical Product Interchangeability Pharmaceutical Product Interchangeability
and Biopharmaceutical Classification System.and Biopharmaceutical Classification System.
Kyiv, Ukraine, June 25 - 27 2007 Kyiv, Ukraine, June 25 - 27 2007
Dr. Sandra Klein, Institute of Pharmaceutical Technology, Dr. Sandra Klein, Institute of Pharmaceutical Technology,
Johann Wolfgang Goethe University FrankfurtJohann Wolfgang Goethe University Frankfurt
![Page 2: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/2.jpg)
Dr. Sandra Klein, 06/20072
Dosage forms to be tested
• immediate release dosage forms
• powders, granules / beads, tablets, capsules
• controlled release dosage forms
• powders, granules / beads, tablets, capsules
• transdermal systems
• implants
![Page 3: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/3.jpg)
Dr. Sandra Klein, 06/20073
Official Dissolution Monographs
United States Pharmacopeia
• USP XXX (30)
European Pharmacopoeia
• Ph. Eur. 5th Edition, Supplement 5.3
British Pharmacopoeia
• BP 2007
Japanese Pharmacopoeia
• JP XIV (14)• http://jpdb.nihs.go.jp/jp14e/contents.html
![Page 4: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/4.jpg)
Dr. Sandra Klein, 06/20074
Official dissolution apparatus
USP 30 classification
1. Rotating Basket (Ph.Eur./BP/JP)
2. Paddle (Ph.Eur./BP/JP)
3. Reciprocating Cylinder (Ph.Eur.)
4. Flow Through Cell (Ph.Eur./BP/JP)
5. Paddle Over Disk (Ph.Eur.)
6. Rotating Cylinder (Ph.Eur.)
7. Reciprocating Holder
![Page 5: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/5.jpg)
Dr. Sandra Klein, 06/20075
Which type of dissolution apparatus ?
Depends on intention
1. Quality control•examining batch homogeneity•examining batch to batch conformity•examining stability
2. Research & Development • examining drug release behavior of preformulations• in vitro simulation of the gastrointestinal passage
3. IVIVC
![Page 6: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/6.jpg)
Dr. Sandra Klein, 06/20076
Apparatus 1 - Basket
Useful for
• capsules• beads• delayed release / enteric
coated dosage forms• floating dosage forms• surfactants in media
Standard volume
• 900/1000 ml• 1, 2, 4 liter vessels
![Page 7: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/7.jpg)
Dr. Sandra Klein, 06/20077
Apparatus 1 - Basket
Advantages
• breadth of experience(more than 200 monographs)
• full pH change during the test
• can be easily automated which is important for routine investigations
![Page 8: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/8.jpg)
Dr. Sandra Klein, 06/20078
Apparatus 1 - Basket
Disadvantages
• disintegration-dissolution interaction
• hydrodynamic „dead zone“
under the basket
degassing is particularly
important
• limited volume
sink conditions for poorly
soluble drugs ?
![Page 9: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/9.jpg)
Dr. Sandra Klein, 06/20079
Apparatus 1 - Basket
![Page 10: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/10.jpg)
Dr. Sandra Klein, 06/200710
Apparatus 2 - Paddle
Useful for
• tablets• capsules• beads• delayed release / enteric
coated dosage forms
Standard volume
• 900/1000 ml
Method of first choice !!!
![Page 11: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/11.jpg)
Dr. Sandra Klein, 06/200711
Apparatus 2 - Paddle
Advantages
• easy to use
• robust
• can be easily adapted
to apparatus 5
• long experience
• pH change possible
• can be easily automated
which is important for
routine investigations
![Page 12: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/12.jpg)
Dr. Sandra Klein, 06/200712
Apparatus 2 - Paddle
Disadvantages
• pH/media change is often difficult
• limited volume sink conditions for poorly soluble drugs ?
• hydrodynamics are complex, they vary with site of the dosage
form in the vessel (sticking,floating) and therefore may
significantly affect drug dissolution
• „coning“
• sinkers for floating dosage forms
![Page 13: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/13.jpg)
Dr. Sandra Klein, 06/200713
Sinker types
JP/ USP / Ph. Eur. 5.3 Sinker
„a small loose piece of nonreactive material such as
not more than a few turns of wire helix may be attached
to dosage units that would otherwise float …“
„…. other validated sinker devices may be used“
![Page 14: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/14.jpg)
Dr. Sandra Klein, 06/200714
Coning
![Page 15: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/15.jpg)
Dr. Sandra Klein, 06/200715
Apparatus 2 - Paddle
![Page 16: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/16.jpg)
Dr. Sandra Klein, 06/200716
Apparatus 3 – Reciprocating cylinder
Useful for
• tablets• beads• controlled release formulations
Standard volume
• 200-250 ml per station
![Page 17: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/17.jpg)
Dr. Sandra Klein, 06/200717
Apparatus 3 – Reciprocating cylinder
Advantages
• easy to change the pH• pH-profiles• hydrodynamics can be
directly influenced by varying the dip rate
Disadvantages
• small volume (max. 250 ml)• little experience• limited data
![Page 18: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/18.jpg)
Dr. Sandra Klein, 06/200718
Apparatus 3 – Reciprocating cylinder
![Page 19: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/19.jpg)
Dr. Sandra Klein, 06/200719
Apparatus 4 – Flow-Through Cell
Useful for
• low solubility drugs• microparticulates• implants• suppositories• controlled release formulations
Variations
• open system• closed system
![Page 20: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/20.jpg)
Dr. Sandra Klein, 06/200720
Cell types
Tablets 12 mm Tablets 22,6 mm Powders / Granules Implants Suppositories /
Soft gelatine capsules
![Page 21: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/21.jpg)
Dr. Sandra Klein, 06/200721
Apparatus 4 – Flow-Through Cell
Advantages
• easy to change media pH• pH-profile possible• sink conditions• different modes
a) open systemb) closed system
Disadvantages• Deaeration necessary
• high volumes of media
• labor intensive
![Page 22: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/22.jpg)
Dr. Sandra Klein, 06/200722
Apparatus 4 – Flow-Through Cell
![Page 23: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/23.jpg)
Dr. Sandra Klein, 06/200723
Apparatus 5 – Paddle over disk
Useful for
• transdermal patches
Standard volume
• 900 ml
![Page 24: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/24.jpg)
Dr. Sandra Klein, 06/200724
Apparatus 5 – Paddle over disk
Advantages
• standard equipment (paddle) can be used, only add a stainless steel disk assembly
Disadvantages
• disk assembly restricts patch size
![Page 25: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/25.jpg)
Dr. Sandra Klein, 06/200725
Apparatus 6 – Rotating cylinder USP apparatus 7 – Reciprocating holder
most probably will be removed from the USP !!!
![Page 26: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/26.jpg)
Dr. Sandra Klein, 06/200726
Summary
Immediate release dosage forms:
apparatus 1 or 2 (preferably 2)
Controlled release dosage forms:
apparatus 1 or 2 using different media for QC apparatus 3 or 4 for R&D purposes
Beside the selection of an adequate dissolution apparatus, adequate test conditions are crucial for all purposes !
![Page 27: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/27.jpg)
Dr. Sandra Klein, 06/200727
Qualification of Dissolution Systems
Prednisone
y = 0,0432x - 0,0039
0
0,5
1
1,5
2
2,5
3
0 10 20 30 40 50 60
Concentration [mg/L]
Ab
sorp
tio
n @
242
nm
![Page 28: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/28.jpg)
Dr. Sandra Klein, 06/200728
Calibration
Why ?
• to confirm suitability of the equipment and proper operation of the apparatus
How ?
• mechanical calibration (verification of physical parameters)• chemical calibration („Apparatus Suitability Test“ – USP)
When ?
• before using new test equipment• after relocation or major maintenance• at regular intervals („every 6 months“)
![Page 29: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/29.jpg)
Dr. Sandra Klein, 06/200729
Factors that may affect reliability of the test
Proper alignment/geometry of dissolution apparatus
– dimensions of vessels, paddles, baskets, cylinders– height, centering and wobble
Proper conditions during dissolution test
– temperature– agitation speed– degassing– sampling (sampling zone, timing, filtration, dilution) – vibration
Proper validation of analytical method
– specified in USP Chapter <1225>
![Page 30: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/30.jpg)
Dr. Sandra Klein, 06/200730
Mechanical calibration
• Verification of physical parameters specified in the pharmacopoeia: USP apparatus 1 and 2
Calibration parameter Current USP tolerance Point of measuring
Height 25 + 2 mm paddle/basket bottom
Basket wobble + 1 mm as runout bottom of basket
Rotational speed + 4 % not applicable
Vessel/shaft centering + 2 mm from center line center line
Vessel temperature 37 + 0.5 °C not applicable
Bath levelness level base plate
Shaft/paddle wobble + 1 mm as runout above top of paddle
Paddle/basket dimensions see USP see USP
Vessel dimensions see USP see USP
![Page 31: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/31.jpg)
Dr. Sandra Klein, 06/200731
Mechanical calibration - Parameters
Height – Vertical Position of the Paddle or Basket
– the vertical position of paddle or basket affects the hydrodynamics condition in the vessel
– each paddle or basket should be individually adjusted to the compendial distance
– in the pharmacopoeia, a distance of 2.5 + 0.2 cm is specified
– different kinds of height gauges can be usedto align or check* this parameter *
![Page 32: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/32.jpg)
Dr. Sandra Klein, 06/200732
Mechanical calibration - Parameters
Rotational Speed – Stirring Rate
– input variable that affects the hydrodynamics
– changes in the rotational speed result in a changing liquid-solid interface between the solvent and the dosage form
– the rotational speed can be checked by using adigital tachometer*
– the compendia specify a rotational speed tolerance of + 4 %
*
![Page 33: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/33.jpg)
Dr. Sandra Klein, 06/200733
Mechanical calibration - Parameters
Shaft Wobble – Eccentricity of Stirring Device
– assumed to alter the pattern of fluid movement in both paddle and basket apparatus and therefore may influence the dissolution rate
– can be measured with a micrometer*
– measured is the sum of distance between both sides (180°) of the axis of rotation
*
![Page 34: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/34.jpg)
Dr. Sandra Klein, 06/200734
Mechanical calibration - Parameters
Centering (Vessel / Shaft)
– the axis of the rotating shaft must coincide at allpoints with the axis of the vessel to within + 1 mm
– “the shaft has to positioned so that is axis is not more than 2^mm at any point from the vertical axis of the vessel and rotates smoothly without significant wobble”
*
![Page 35: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/35.jpg)
Dr. Sandra Klein, 06/200735
Mechanical calibration
Measurement tools
• all mechanical tools used for
calibration should be certified
to assure their reliability
• the results of mechanical
calibration have to be documented
![Page 36: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/36.jpg)
Dr. Sandra Klein, 06/200736
Apparatus suitability test (USP)
• if all parts ( apparatus, geometry, test conditions, analytical
method) are within compliance – why perform an apparatus
suitability test?
• the apparatus suitability is to check for parameters that can not be
conveniently measured (vibration, vessel cleanliness, medium
degassing ...) and also to provide an overall check of the system
![Page 37: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/37.jpg)
Dr. Sandra Klein, 06/200737
Apparatus suitability test (USP)
• first established in 1978
• routine test in most pharmaceutical laboratories
• calibration at regular intervals (every 6 months)
• standard calibrator substances according USP chapter <711>
• only the method(s) to be used have to be calibrated !
• if six units are tested – all have to pass
![Page 38: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/38.jpg)
Dr. Sandra Klein, 06/200738
Apparatus suitability test (USP)
Standard calibrators according to USP chapter <711>
Apparatus I, II and V:
1. disintegrating type
– USP Prednisone Tablets
2. nondisintegrating type
– USP Salicylic acid Tablets
Apparatus III:
• USP Chlorpheniramine Maleate Extended-Release Tablets
![Page 39: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/39.jpg)
Dr. Sandra Klein, 06/200739
Information supplied with calibrators
http:/www.usp.org/referenceStandards/useAndstorage/calibrators.html
![Page 40: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/40.jpg)
Dr. Sandra Klein, 06/200740
Apparatus suitability test (USP)
USP Prednisone Tablets RS – current lot P0E203
(10 mg nominal prednisone content per tablet)
• disintegrating type
• paddle and basket, 50 rpm
• 500 ml deaerated water, 37°C
• quantity of prednisone released after 30 minutes is determined
• specified ranges Lot P0E203: Apparatus 1: 47-82 % Apparatus 2: 37-70 %
![Page 41: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/41.jpg)
Dr. Sandra Klein, 06/200741
Apparatus suitability test (USP)
USP Salicylic acid Tablets RS – current lot Q0D200
(300 mg nominal salicylic acid content per tablet)
• nondisintegrating type
• paddle and basket, 100 rpm
• 900 ml deaerated phosphate buffer, 37°C
• quantity of salicylic acid, released after 30 minutes is determined
• specified ranges Lot Q0D200: Apparatus 1: 23-30 %Apparatus 2: 17-25 %
![Page 42: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/42.jpg)
Dr. Sandra Klein, 06/200742
Apparatus suitability test (USP)
Controversies regarding the current test
• the variability in the intrinsic performance of the USP calibrator
tablets is so great that it exceeds the variability in intrinsic
performance of modern test dissolution assemblies
• this variability becomes obvious in both vessel-to-vessel
variability and inter-laboratory variability of results for a given lot
of calibrators
![Page 43: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/43.jpg)
Dr. Sandra Klein, 06/200743
Calibrator Tablets:
• always check the incoming tablets !• right lot of calibrators ?• are the tablets broken, fused or severely chipped ?• particularly salicylic acid tablets are often subject to sublimation
( dust on the tablets and the inner surface of the container)
• use correct storage conditions !
• take the tablets out of the original
container immediately before test !
Troubleshooting
![Page 44: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/44.jpg)
Dr. Sandra Klein, 06/200744
Standard / Standard solution:
• USP Standard used ?
• drying procedure conducted ?
• standard solution prepared on day of test ?
• standard solution filtered in the same manner as sample ?
• amount of alcohol used in the standard < 5% ?
Troubleshooting
![Page 45: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/45.jpg)
Dr. Sandra Klein, 06/200745
Vibration
• vibration produces unwanted variation in dissolution data and
mostly results in an increased dissolution rate
• internal vibration may be caused e.g. from frayed drive belts
• external vibration may be caused by e.g. magnetic stirrers,
centrifuges, vacuum pumps, old fridges, nearby construction, ...
• inability to properly measure vibration levels at various points
within an apparatus is the main reason why calibrator tablets were
originally developed
Troubleshooting
![Page 46: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/46.jpg)
Dr. Sandra Klein, 06/200746
Vibration effects – case example:
Effect of vibration levels of the dissolution apparatus on the dissolution rate of enteric coated granules of Cefalexin. The vertical dotted line indicates 0.05 m/s2.
Kaniwa N. et al. (1998) Int J Pharm, 175, 119-129
„Low vibration“: < 0.05 m/s2
„High vibration“: > 0.05 m/s2
Troubleshooting
![Page 47: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/47.jpg)
Dr. Sandra Klein, 06/200747
Vibration:
• dissolution equipment placed planar ?
• drive chain or belt free of tension and/or dirt ?
• torn parts replaced ?
• correctly functioning gear plates ?
• individual spindles are not surging ?
• bench/table stable ?
• no sources of vibration nearby ?
Troubleshooting
![Page 48: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/48.jpg)
Dr. Sandra Klein, 06/200748
Dissolution medium:
• correctly degassed ?
• correct amount used (900/500 ml) ?
• correct amount dosed (weight/volume) ?
• dosing procedure gentle (resaturation/spillage) ?
• buffer correct (pH + 0.05 units, buffer salts, molarity) ?
• correct temperature during test (32°C / 37°C + 0.5°C)?
• evaporation during test negligible ?
Troubleshooting
![Page 49: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/49.jpg)
Dr. Sandra Klein, 06/200749
Importance of degassing:
• insufficient degassing may result in decreased dissolution rates of several drugs
• e.g. prednisone tablets but also a range of poorly soluble drugs are very sensitive to the amount of dissolved gases in the dissolution medium
• the degassing procedure should therefore be efficient and reproducible for every test
Troubleshooting
![Page 50: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/50.jpg)
Dr. Sandra Klein, 06/200750
Deaeration method USP
• heat the dissolution medium to about 41°C
• vacuum filter through a 0.45-µm-porosity membrane into a flask, stirring with a magnetic stirrer
• continue to draw a vacuum and stir for an additional 5 min
• gently transfer the medium directly into the vessel
• rotating the apparatus 2 shafts to speed equilibration to 37°C is discouraged!!!
• use medium promptly after equilibration
Troubleshooting
![Page 51: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/51.jpg)
Dr. Sandra Klein, 06/200751
Alternative deaeration methods
• the USP states that „other validated deaeration techniques for
removal of dissolved gases may be used“
• other techniques include:heatingsonicationvacuumhelium sparging (expensive)
Troubleshooting
![Page 52: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/52.jpg)
Dr. Sandra Klein, 06/200752
Sampling
• take each sample at the correct time point
sampling time points (+ 2%)
• use a single glass syringe for each vessel
• sample from the right location within the vessel
between media surface and top of the
paddle blade
n.l.t. 10 mm from vessel wall
Troubleshooting
![Page 53: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/53.jpg)
Dr. Sandra Klein, 06/200753
Sampling
• always use a suitable filter check filter adsorption
• check the clearity of the filtered sample
• filter the sample immediately after sampling
• for automated sampling also check the tubings
Troubleshooting
![Page 54: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/54.jpg)
Dr. Sandra Klein, 06/200754
Physical conditions of the apparatus
• vessels scrupulously clean ?
• vessel surface smooth and curvature appropriate ?
Apparatus 1
• the conditions of the baskets, particularly of their clips is critical
• check all baskets for corrosion and blocked meshes before using them
• align the air holes to prevent air cushions emerging during the test
Troubleshooting
![Page 55: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/55.jpg)
Dr. Sandra Klein, 06/200755
Advantages
• high throughput of samples
• minimizes analyst-to-analyst variability in sampling and filtration
• reduces the average costs per analysis
• very promising for QC purposes
(Semi) Automation
![Page 56: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/56.jpg)
Dr. Sandra Klein, 06/200756
• automated mixing of water and concentrate
• preheating of medium
• deaeration(vacuum and stirring)
• media can be dispenseddirectly into the vessel
Media preparation
![Page 57: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/57.jpg)
Dr. Sandra Klein, 06/200757
Offline system
![Page 58: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/58.jpg)
Dr. Sandra Klein, 06/200758
Online system
![Page 59: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/59.jpg)
Dr. Sandra Klein, 06/200759
On- / Offline system
![Page 60: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/60.jpg)
Dr. Sandra Klein, 06/200760
HPLC - online system
![Page 61: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/61.jpg)
Dr. Sandra Klein, 06/200761
• always validate automated methods, including analytical and sampling methods
• validation should be performed using manual analysis, withdrawing samples at the same times and comparing to the automated results:
not highly variable dissolution results: two concurrent
runs
highly variable dissolution results: simultaneous
sampling
• pay attention to automated dilution and filtering processes
Automation
![Page 62: Design calibration en (1)](https://reader036.vdocument.in/reader036/viewer/2022062304/55d21d64bb61eb656e8b474a/html5/thumbnails/62.jpg)
Dr. Sandra Klein, 06/200762
• FIP Guidelines for dissolution testing of solid oral products.Dissolution Technologies 4:5-14 (1997).
• SM Diebold, JB Dressman. Dissolved oxygen as a measure for de- and re-aeration of aqueous media for dissolution testing. Dissolution Technologies 5: 13-16 (1998).
• S Qureshi. Calibration – the USP dissolution apparatus suitability test.Drug Inf. J. 30, 1055-1061 (1996).
• N Kaniwa et al. Collaborative study on the development of a standard for evaluation of vibration levels for dissolution apparatus.Int. J. Pharm. 175: 119-129 (1998).
• VA Gray, CK, Brown, JB Dressman, LJ Leeson. A new general information chapter on dissolution.Pharmacopoeial Forum 27 (6) [Nov.-Dec. 2001]
• W Brown. General information <1092> The dissolution procedure: development and validationPharmacopoeial Forum 30 (1) [Jan.-Feb. 2004]
Of general interest:
• Dissolution Technologies: http://www.dissolutiontech.com