detection by hplc
TRANSCRIPT
-
8/9/2019 Detection by Hplc
1/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Selecting the Right Detector:Selecting the Right Detector:
Types of Detectors in HPLCTypes of Detectors in HPLC
Shulamit LevinShulamit Levin
/HPLCwww.forumsci.co.ilhttp://
UV/VISRefractive index
Fluorescence
Electrochemical
Conductivity
Mass-spectrometric (LC/MS)
Evaporative light scattering
Appendix:
Cutoff of solvents UV
Troubleshooting of RI detector as
an example
The Detector is theThe Detector is the EyeEye of the HPLC Systemof the HPLC System
HPLC Column
in OvenAuto
Sampler
Pumpflows 50-5000L/min)
Fraction
Collector
Waste
DetectorControl &
DataProcessing
1.Fuc os e2.Galactosami ne
3.Glucosami ne4.Galactose
5.Glucose6.Mannos e
20.00Minutes5.00
mV
0.00
300
12
34
5
6
a bc d
DetectorsDetectors
UV/VISUV/VIS
Refractive indexRefractive index
FluorescenceFluorescence
ElectrochemicalElectrochemical
ConductivityConductivity
MassMass--spectrometric (LC/MS)spectrometric (LC/MS)Evaporative light scatteringEvaporative light scattering
Optical Bench of UVOptical Bench of UV--VIS DetectorVIS Detector
DeuteriumArc Lamp
Rotating
DiffractionGrating
190 to 600nm
Taper-Cell
FlowCell
Beam Splitter
Mirrors
Dual
Photodiode
ApertureSlit
Illumination
Lens
Beam-Defining
Apparatus
Optical Light Path
Sample side
Reference side
-
8/9/2019 Detection by Hplc
2/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Beer's LawBeer's Law
Reduce Pathlength Reduce Concentration
Absorbance = Extinction Coefficient x Pathlength x Concentration
UVUV ChromophoresChromophores
UVUV ChromophoresChromophores UVUV--VisVis chromophoreschromophores
max Emx 10-3 @max
Adenine 260.5 E = 13.4
Guanine 275 E = 8.1Cytosine 267 E = 6.1
Thymine 264.5 E = 7.9
Uracil 259.5 E = 8.2NADH 340 E = 6.23NAD 260 E = 18
-
8/9/2019 Detection by Hplc
3/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
UV spectrum of 10UV spectrum of 10 nMnM mobile phasemobile phase
AU
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
nm
210.00 220.00 230.00 240.00 250.00 260.00 270.00 280.00 290.00
pH 0.94
TFA
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
nm210.00220.00 230.00240.00 250.00 260.00270.00280.00 290.00
265.1 295.8
pH 2.0
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
nm
210.00 220.00 230.00 240.00 250.00 260.00 270.00 280.00 290.00
205.2
pH 2.26
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
nm210.00220.00230.00240.00250.00260.00270.00280.00290.00
pH 2.78
Sodium PhosphateFormic Acid
Acetic Acid
U.V. CutU.V. Cut--offs for some Common Solventsoffs for some Common Solvents
SolventSolvent UV CutoffUV Cutoff SolventSolvent UV CutoffUV Cutoff
WaterWater 180180 N-HeptaneN-Heptane 197197MethanolMethanol 205205 CyclohexaneCyclohexane 200200N-PropanolN-Propanol 205205 Carbon tetrachlorideCarbon tetrachloride 265265AcetonitrileAcetonitrile 190190 ChloroformChloroform 245245THFTHF 225225 BenzeneBenzene 280280AcetoneAcetone 330330 TolueneToluene 285285Methyl acetateMethyl acetate 260260 Methylene chlorideMethylene chloride 232232Ethyl AcetateEthyl Acetate 260260 TetrachloroethyleneTetrachloroethylene 280280NitromethaneNitromethane 380380 1,2-Dichloroethane1,2-Dichloroethane 225225
All wavelengths reported in nm.All wavelengths reported in nm.
Remember Solvents chosen can affect detection!!Remember Solvents chosen can affect detection!!
UV Detection ofUV Detection ofAccQAccQ--Tag Amino Acid DerivativesTag Amino Acid Derivatives
0.000
0.002
0.004
0.006
0.008
0.010
0.012
0.014
0.016
0.018
0.020
0.022
0.024
15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00
AU
Minutes
AsparticAcid
GlutamicAcid
Hydroxypro
line
Serine
Asparagine
AMQ
Glycine
Glutamine H
istidine
NH3
Threonine
Arginine
Alanine
Proline
Alpha-aminobutyricacid
Tyrosine
CysteicAcid
Vaine
Mtehionine
Ornithin
e
Leucine
Lysin
e
Phenylalanine
Tryptophan
Isoleuc
ine
Diode Array DetectorDiode Array Detector
-
8/9/2019 Detection by Hplc
4/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Extraction of 3D DataExtraction of 3D Data
Time
Absorbance
Chromatogram
1
2
Wavelength
Absorbance Spectrum
PDA Spectrum Index PlotPDA Spectrum Index PlotDNPH Derivatives 0.25 ng Each PeakDNPH Derivatives 0.25 ng Each Peak
260.00 260.00
280.00 280.00
300.00 300.00
320.00 320.00
340.00 340.00
360.00 360.00
380.00 380.00
400.00 400.00
420.00 420.00
440.00 440.00
nm
0.00000.0000
0.00020.0002
0.00040.0004
0.00060.0006
AUAU
Millennium PDA Spectrum Index Plot - SampleWeight 0.25 ng - PDA 360.0 nm
CoelutionCoelution of 2 Peaksof 2 Peaks
AU
A
B
Coelutiondetection at asingle wavelength
Coelution is the sum ofabsorbance of 2 peaksA and B
Peak Purity MeasurementPeak Purity Measurement
2.20 2.40 2.60 2.80 3.00
200.00
250.00
300.00
0.00
0.20
0.40
Maximum
Impurity
Spectra at apex andinflection points aredisplayed
Spectrum atmaximum impurityis different
AU
nm
-
8/9/2019 Detection by Hplc
5/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Maximum Impurity DetectionMaximum Impurity Detection
260.00 260.00
280.00 280.00
300.00 300.00
320.00 320.00
340.00 340.00
360.00 360.00380.00 380.00
400.00 400.00
420.00 420.00
440.00 440.00
nm
Minutes
nm
-0.00010 0.00010
0.00000 0.00000
0.00010 0.00010
0.00020 0.00020
0.00030 0.00030
18.40 18.60 18.80 19.00 19.20
AU AU
Millennium PDA Spectrum Index Plot - SampleWeight 0.25 ng360nm 996PDA 360.0 nm
Hexaldehyde 2,5-Dimethylbenzaldehyde
Coelutionof DNPHHexaldehyde and
2,5-Dimethylbenzaldehyde
Determination of Peak PurityDetermination of Peak Purity
Absorbance
Time
S t a n d a r d
T i m e
U n k n o w n
Peak Purity analyzes all spectra (minimum
15) within a peak against the apex spectrum
of the peak itself.
Peak PurityPeak Purity Spectral MatchingSpectral Matching
Spectral match of apex spectrum of the
unknown against the apex spectrum of a
standard, stored in a users library.
Different SpectraDifferent Spectra 53 deg53 deg
200.00 240.00 280.00 320.00
nm
EthylPabaEthylparaben
Absorbance
10 deg of Spectral Contrast10 deg of Spectral Contrast
Similar spectra for
structurally related
compounds
230.00 250.00 270.00 290.00 310.00
nm
Theophylline
Dyphylline
Absorbance
-
8/9/2019 Detection by Hplc
6/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Spectral Contrast 0.5 DegreesSpectral Contrast 0.5 Degrees
200.00 240.00 280.00 320.00nm
MethylparabenEthylparaben
Absorbance
Very similar spectra,CH2 difference
Spectral Contrast candifferentiate thesespectra
Spectra of nonSpectra of non--UV Active CompoundsUV Active Compounds
210.00 230.00 250.00 270.00 290.00nm
Absorbance
Analyte and 2 Impurities
DetectorsDetectors
UV/VISUV/VIS
Refractive indexRefractive index
FluorescenceFluorescence
ElectrochemicalElectrochemical
ConductivityConductivity
MassMass--spectrometric (LC/MS)spectrometric (LC/MS)Evaporative light scatteringEvaporative light scattering
Refractive Index DetectorRefractive Index Detector
-
8/9/2019 Detection by Hplc
7/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Refractive Index DetectorRefractive Index DetectorDifferential Refractive Index DetectorDifferential Refractive Index Detector
LAMP
SLED
R
RS
To Amplifier
No sample = n
With sample = n+
X = Const x n
Sugar AnalysisSugar Analysis
-160.00
-140.00
-120.00
-100.00
-80.00
-60.00
-40.00
-20.00
0.00
20.00
40.00
60.00
80.00
100.00
120.00
5.00 6.00 7.00 8.00 9.00 10.00 11.00
mV
Minutes
Fructose
Dextrose Sucrose
Maltose Lactose
SampleName: Sugars D Vial: 1 Inj: 1 Ch: SATIN Type: Standard
Bagel Extract
Polymer AnalysisPolymer Analysis
0.00
50.00
100.00
150.00
200.00
250.00300.00
350.00
400.00
450.00
500.00
550.00
600.00
650.00
700.00
750.00
800.00
18.00 20.00 22.00 24.00 26.00 28.00 30.00
MV
Minutes
1260000
96400
5570
SampleName: GPC STDS
Dow 1683
28
90000
190000
10300
192300
-
8/9/2019 Detection by Hplc
8/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
LipidsLipids
DelRIU
5.0 6.0 7.0 8.0
Minutes
1 2
250 ng on column1=Tristearin
2=Myristic acid
Styragel HR 0.5,4.6 x 300 mm,35C, 0.35 mL/min
dRI sensitivity =32X, 32C
DetectorsDetectors
UV/VISUV/VIS
Refractive indexRefractive index
FluorescenceFluorescence
ElectrochemicalElectrochemical
ConductivityConductivity
MassMass--spectrometric (LC/MS)spectrometric (LC/MS)
Evaporative light scatteringEvaporative light scattering
Fluorescence ProcessFluorescence Process
Excitation
Emission
Energy Levels
Ground State
Excited States
ExcitationExcitation--Emission SpectraEmission Spectra
Maximum ofExcitation Spectrum
Maximum of
Emission Spectrum
Stokes shift
Lifetime= 10-9 10-15 sec
-
8/9/2019 Detection by Hplc
9/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Fluorescence DetectorsFluorescence Detectors
Excitation filter
LAMP
Cell
Emission filter
Photomultiplier
Fluorescence Detector Optical BenchFluorescence Detector Optical Bench
Photomultipliertube
Emission
Grating
ExcitationGrating
Mirror
Torroidal Mirror
Beam Splittter
FlowCell
Mirror
Torroidal Mirror
Photodiode
Excitation
Slit
Emission
Slit
UVUV vsvs Fluorescence SensitivityFluorescence Sensitivity
Minutes
AMQ
20.00 40.00 60.00
Response
AccQ-Tag amino acidanalysis
Fluorescence
Excitation=250 nm
Emission=395 nm
UV 254 nm
DetectorsDetectors
UV/VISUV/VIS
Refractive indexRefractive index
FluorescenceFluorescence
ElectrochemicalElectrochemical
ConductivityConductivity
MassMass--spectrometric (LC/MS)spectrometric (LC/MS)Evaporative light scatteringEvaporative light scattering
-
8/9/2019 Detection by Hplc
10/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Electrochemical DetectorElectrochemical Detector
Reference Electrode
Working Electrode
Electrolyte (mobile phase)
Auxiliary Electrode
Analyte is oxidized or reduced
+ -
As compounds are oxidized or reduced, a current proportional to concentration is produced.
Electrochemical Detection ofElectrochemical Detection ofCatecholaminesCatecholamines & Related Compounds& Related Compounds
1 . N or ep in ep he ri ne 1 50 pp b
2. Epinepherine 200 ppb
3. Normetanepheri ne 50 ppb
4. Dopamine 200 ppb5. Me tan ep he ri ne 200 p pb
6 . 3 -M e th o xy ty r am i ne 7 5 p pb
7 . 4 - Me t ho x yt y ra m in e 5 00 pp b
2.00 4.00 6.00 8.00 10.00 12.00
Minutes
0.00
nAmps
12
3
45
6
7
PulsedPulsed AmperometricAmperometric Detection ofDetection ofMonosaccharidesMonosaccharides
1. Fucose
2. Galactosamine
3. Glucosamine
4. Galactose
5. Glucose
6. Mannose
20.00Minutes5.00
mV
0.00
300
12
34
5
6
DetectorsDetectors
UV/VISUV/VIS
Refractive indexRefractive index
FluorescenceFluorescence
ElectrochemicalElectrochemical
ConductivityConductivity
MassMass--spectrometric (LC/MS)spectrometric (LC/MS)Evaporative light scatteringEvaporative light scattering
-
8/9/2019 Detection by Hplc
11/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Conductivity DetectorConductivity Detector
Mobile phase
Mobile phase plus sample
++ --
Conductivity EquationsConductivity Equations
Ohms Law
Conductivity DetectorConductivity Detector
Mobile phase
Mobile phase plus sample
Anion Analysis by ICAnion Analysis by IC
0.70
1.05
1.40
S
0.00 5.00 10.00 15.00 20.00 25.00Minutes
1
2
3
45
6 7
1. Fluoride2. Chloride
3. Nitrite
4. Bromide5. Nitrate
6. Phosphate
7. Sulfate
1 ppm2 ppm
4 ppm
4 ppm4 ppm
6 ppm
4 ppm
Column:
Eluent:
Flow rate:Injection vol.:
Detection:
Waters IC-Pak Anion HC
Borate/Gluconate2.0 mL/min
100L
Direct Conductivity
-
8/9/2019 Detection by Hplc
12/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Anion analysis by ICAnion analysis by IC
S
0.00
0.40
0.80
1.20
1.60 12
3
45
6
7
0.00 4.00 8.00 12.00 16.00 20.00 24.00
Minutes
0.00
0.01
0.02
0.03
0.04
0.05
AU
3
4
5 Column:Eluent:
Flow rate:
Injection vol.:
Waters IC-Pak Anion HR
1.2 mM Sodium Carbonate/
1.2 mM Sodium Bicarbonate
1.0 mL/min
50L
Detection: Direct Conductivity after
Suppression
Detection: UV (PDA) at 214 nm
1. Fluoride
2. Chloride
3. Nitrite
4. Bromide
5. Nitrate
6. Phosphate
7. Sulfate
1 ppm
2 ppm
4 ppm
4 ppm
4 ppm
6 ppm
4 ppm
ApplicationsApplications
Sensitivities for compounds such as phenol, catecholamines,nitrosamines, and organicacidsare in thepicomole (nanogram)
range.
Themobilephasemustbe made electrically conductive, usuallyby the addition of a suitable salt:
Ion Exchange
Reversed Phase and Ion-Pair RP
No normal phase separations
DetectorsDetectors
UV/VISUV/VIS
Refractive indexRefractive index
FluorescenceFluorescence
ElectrochemicalElectrochemical
ConductivityConductivity
MassMass--spectrometric (LC/MS)spectrometric (LC/MS)Evaporative light scatteringEvaporative light scattering DATA SYSTEM
DETECTION OF IONS
+
+
SAMPLE DESOLVATION
AND IONIZATION
LC/MSINTERFACE
SOURCE
ANALYZERION DETECTOR
HPLCMASS SPECTRUM
SORTING OF IONS
+
++
++ ++ +
+
-+
+
++
++
-
+
+
+
CHROMATOGRAM
Typical LC/MS System ProgressionTypical LC/MS System Progression
-
8/9/2019 Detection by Hplc
13/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
DATA SYSTEM
DETECTION OF IONS
+
+
SAMPLE DESOLVATIONAND IONIZATION
LC/MSINTERFACE
SOURCE
ANALYZERION DETECTOR
HPLCMASS SPECTRUM
SORTING OF IONS
+
++
++ ++ ++
-+
+
++
++
-
+
+
+
CHROMATOGRAM
Typical LC/MS System ProgressionTypical LC/MS System Progression Transition from LC to MSTransition from LC to MS
State of Matter: LiquidLiquid to GasGas
Charge State: NeutralNeutral to IonIon
Pressure: 760760 torrtorr to 1010--55 to 10to 10--88 torrtorr
APCI MechanismAPCI Mechanism
Ionization producessolvent ions
The solvent ions reactwith analyte moleculesforming clusters
CoronaNeedle
X = Solvent Molecules e.g.H 2O, MeCN
M = Sample Molecule
Heated Nebulizer
xx
xH+ M
xx
xH+
M
x
x
MMH+
xx
xx
XH+x
xH+
MH+
ElectrosprayElectrospray IonizationIonization
-
8/9/2019 Detection by Hplc
14/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Positive or Negative?Positive or Negative?
Basic Compounds (-NH2) (M+H)+
Acidic Compounds (-CO2H, -OH) (M-H)-
Recognizing Multiply Charged Ions
Mass spectrometers operate on the basis of mass-to-charge ratio (m/z).Mass assignments are normally made assuming a single charge per ion(i.e. m/z = m)
Single charge Mass = (M+H)
Double charge Mass = 1/2 (M+2H)
n charge Mass = 1/n (M+nH)
Isotopes of doubly charged ions are separated by 0.5 Da
n= 20
n= 22
n= 18
n= 16, m/z = 1060
n= 23, m/z = 738
n= 21
n= 19
n= 17
Horse Heart Myoglobin
Mass RangeMass RangeMultiply Charged MoleculesMultiply Charged Molecules
Calculated MassAcquired Mass range
Hemoglobin SpectrumHemoglobin Spectrum
Presence of More Than One Charged EnvelopePresence of More Than One Charged Envelope
1000 1050 1100 1150 1200 1250 1300 1350 1400m/z0
100
%
1081.60
1009.36
1058.83
1164.52
1133.92
1261.64
1221.181376.08
1323.14
-
8/9/2019 Detection by Hplc
15/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
DeconvolutionDeconvolution byby MaxEntMaxEnt
HemoglobinHemoglobin
15000 15100 15200 15300 15400 15500 15600 15700 15800 15900 16000 16100mass0
100
%
15125.0
15857.0
15149.0
15866.0
Multiply Charged Ions How Many Charges?
(M+H)+
(M+2H)2+
1 Da
0.5 Da
DATA SYSTEM
DETECTION OF IONS
+
+
SAMPLE DESOLVATION
AND IONIZATION
LC/MSINTERFACE
SOURCE
ANALYZERION DETECTOR
HPLCMASS SPECTRUM
SORTING OF IONS
+
++
++ ++ +
+
-+
+
++
++
-
+
+
+
CHROMATOGRAM
Typical LC/MS System ProgressionTypical LC/MS System ProgressionMassMass
SpectrometerSpectrometerss
AnalyzersAnalyzers
T ime O f F l igh t Ma s s Ana ly ze rsT ime O f F l igh t Ma s s Ana ly ze rs
SOURCE
DETECTORREFLECTRON MODE
R E F LE C T R ON ONDETECTOR
LI NEAR MODEDRI FT TUBE
SOURCE
DETECTORREFLECTRON MODE
R E F LE C T R ON OF FDETECTORLI NEAR MODE
DRI FT TUBE
199
FTFT--ICRICR --SpectrometerSpectrometer
DCDC
DC
Source
Filament
Transferoptic
TrappingPlates
Transmitter Plates
ReceiverPlates
Sender
Elektroden Electrodes
Y
ZX
Magnetic FieltB
Starting with theStarting with the quadrupolequadrupole
Source
DetectorNonresonantIon
Resonant Ion
dc and Rfvol tages
V(t) = -Vdc
-Vrf
cost
V(t) = Vdc
+Vrf
cost
190
ElectronMultiplier
Inlet
RingElectrode,Rf
End Cap Electrode
AxialModulation
IonIon TrapsTraps
+ + ++++
++
IonSource
Slit
Magnetic sectorElectrostatic Sector
(ESA)
Detector
SlitNier-Johnson-Geometry(EB)
-
8/9/2019 Detection by Hplc
16/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
DATA SYSTEM
DETECTION OF IONS
+
+
SAMPLE DESOLVATIONAND IONIZATION
LC/MSINTERFACE
SOURCE
ANALYZERION DETECTOR
HPLCMASS SPECTRUM
SORTING OF IONS
+
++
++ ++ ++
-+
+
++
++
-
+
+
+
CHROMATOGRAM
Typical LC/MS System ProgressionTypical LC/MS System Progression MS DetectorsMS Detectors
++ ++
--
-
Electron Multiplier(voltage setting lower than
Dynode)
Conversion Dynode(Voltage 1- 20 kV)
Current is measured
++
Mass Analyser
dynodedynode
phosphorphosphor
photomultiplierphotomultiplier
Photomultiplier
Electron Multiplier
Mass Spectrometer 3D RunMass Spectrometer 3D Run
Mixture
2.00 4.00 6.00 8.00 10.00Time0
Int.
1:Mass Chromatogram5.054.65
3.82
0.74
8.62
5.65
8.02
10.62
Total-Ion-Current MS Chromatogram
with poor resolution
Mix
6 0 8 0 1 0 0 1 2 0 1 4 0 1 60 1 80 2 00 2 20 2 40 2 60 2 80 3 00 3 20 3 40m/z0
100
%
(10.696) 1: Scan ES+4.34e5262.87
59.99
213.90
195.98
120.8068.9298.85
76.87128.82
170.92
222.87
235.87
240.88
263.87
264.85
267.91287.01 309.02
333.84
Mixture
2.00 4.00 6.00 8.00 10.00Time0
Abs1:DiodeArray
5.054.65
3.820.74
8.62
5.65
8.02
10.62
200210220230240250260270280290300310nm0
Int.210.00
246.00
294.00
UV-Diode Array Chromatogram
with poor resolution
Mix
2.00 4.00 6.00 8.00 10.00 12.00 14.00Time0
100
%
1: Scan ES+262.874.59e5
10.60
Mixture
2.00 4.00 6.00 8.00 10.00 Time0
Int. 5.054.65
3.820.74
8.625.65
8.02
10.62
Selectivity of Mass Spectrometer DetectorSelectivity of Mass Spectrometer Detector
Extracted Ion Chromatogramof a single Component from
a mixture of Components
Mix
60 8 0 1 00 1 20 14 0 1 60 1 80 2 00 2 20 24 0 2 60 28 0 3 00 3 20 3 40m/z0
100
%
(10.696) 1: Scan ES+4.34e5262.87
59.99
213.90
195.98120.8068.92
98.8576.87 128.82
170.92
222.87
235.87
240.88
263.87
264.85
2 6 7 .9 1 2 8 7 .0 1309.02
333.84
-
8/9/2019 Detection by Hplc
17/20
-
8/9/2019 Detection by Hplc
18/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
Evaporative Light ScatteringEvaporative Light Scattering -- ELSELS
NebulizationDesolvation
DetectionTo detector cell
Nebulizer
Lamp
RayleighRayleigh ScatteringScattering Why the Sky is blueWhy the Sky is blue
Scattering is independent of the particles chemical properties, where:
N = # of particles
= Polarizability i.e. the sum of the dipoles of all the molecules in the particle.For a homogeneous particle this is proportional to the particle volume.
R = Distance of observer from scatterer
Dependence on wavelength of incident light, shorter wavelengths producegreater scattering
4
R2
I = I0
8 p4 Na2 (1 +cos2?)
Scattering ModelsScattering Models
If D/< 0.1then I = f (D6)
0.110then I = f (D2)
I = Intensity of the scattered light= wavelength of the light
Scattering is dependent on particle size D Increasing
particle size
D C1/3
(Often see solute density )I a (Cb)
With 2>b>2/32 is the limiting value for Rayleigh
symmetrical scattering
Depends on which type of scattering ispredominant
Non-linear mass detector
vuse chromatography data softwarequadratic curve or log/log curve to fitcalibration curve
ELSDELSD vsvs UVUV
-
8/9/2019 Detection by Hplc
19/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
ELSDELSD vsvs RIRI ELSD Used with Other DetectorsELSD Used with Other Detectors
ELSD to monitor allcompounds and determinepurity levels
Diode Array TIC
Min
ES+TIC
ELSD
24.5 7
Diode array TIC
PDA to monitor UV/Visfriendly compounds
Mass Spec to verify thatcompound has beensynthesized
Not a Universal DetectorNot a Universal DetectorTypically Used with Other DetectorsTypically Used with Other Detectors
UV TIC
MS ES+ TIC
ELSD
Erythromycins Separation
Min1 3 5
ELSD
Diode ArrayTIC
See NonSee Non--UV Absorbing CompoundsUV Absorbing Compounds
-
8/9/2019 Detection by Hplc
20/20
Detection in HPLC
Dr. Shulamit Levin, Medtechnica
See Your Peaks FasterSee Your Peaks FasterUse of Gradients Versus IsocraticUse of Gradients Versus Isocratic
RI Detection
ELSD Detection
Estrogen analogues and salt
mV
0.00
10.00
20.00
30.00
40.00
50.00
60.00
70.00
80.00
90.00
100.00
110.00
Minutes
0.100. 200. 300. 40 0.500. 60 0.700. 800. 901. 00 1.10 1.201. 301. 40 1.501. 601. 70 1.801. 902. 00
NaCl-0
.207
Estradiol-0
.849
Estriol-1
.210
Evaporative LightEvaporative Light--scattering Detectorscattering Detector