Determination of Bacteria Retention in the Thermo Scientific Barnstead Smart2Pure Water Purification System Verena Ruckstuhl, Thermo Fisher Scientific, Langenselbold, Germany and Julie Foster, Ph.D., Thermo Fisher Scientific, Asheville, NC Application Note Key Words Water purification, lab water, ultrapure water, bacteria, 0.2 micron filter, Thermo Scientific Barnstead, Smart2Pure Abstract The bacteria retention of a Thermo Scientific™ Barnstead™ Smart2Pure™ water purification system was evaluated using the membrane filtration method based on the European Pharmacopoeia method described in Chapter 2.6.12. 1 Introduction Bacteria are single celled organisms that can be found nearly everywhere in a busy laboratory. Although many of these bacteria are harmless to a person’s health, they can create unwelcome variability in an experiment. Because of their abundance and ability to be easily transferred, precautions such as sterile technique were employed. Using bacteria-free water during steps such as sample preparation, system rinsing, or buffer preparation is an easy method of reducing the chance of bacterial contamination. Water purification systems are a reliable source for bacteria-free water. On average, bacteria such as Escherichia coli (E. coli), which are widely used in the laboratory, have an length of about 2 µm, and an average diameter of 0.5 µm. 2 A 0.2 µm absolute membrane filter at the end of the system is used to remove any particles or bacteria that are larger than the pore size of the filter 3 . Proper maintenance of the water system, including filter replacement as specified in the manual, helps to ensure the water remains bacteria-free. Ultrapure water from a Barnstead Smart2Pure water system was analyzed for the presence of aerobic bacteria. The Barnstead Smart2Pure 3 system was chosen from the family of Smart2Pure systems, which also includes Smart2Pure UV and UF models. All of these systems have the same feed water requirements, basic flow path and all dispense water through a 0.2 µM final filter. The systems are fed by tap water and utilize two purification steps to create 18.2 megohm lab water. The first step purifies the feed water using a reverse osmosis (RO) membrane to remove the majority of impurities in the water. The second step of purification utilizes an ultrapure cartridge to remove any ions in the water. Additionally, an optional ultraviolet (UV) lamp and/or an ultrafilter (UF) can be added to the system configuration. Methods Bacteria retention of the Smart2Pure system using tap water as feed water The bacterial content of tap water was determined after disinfecting the water outlet with 70% ethanol and rinsing with 1 L water. Afterwards three 1 L sample volumes were collected and tested for bacterial growth using the membrane filtration technique. 1 Three different volume samples were generated from the three 1 L samples: a 1 L, 10 mL, and a 1 mL sample. These three samples were filtered through a 0.2 µm cellulose nitrate (CN) membrane. The membrane was afterwards transferred aseptically to a R2A-Agar and incubated at 35°C for 5 days in a Thermo Scientific™ Heratherm™ compact microbiological incubator (model IMC18). The agar plates with sample volumes of the membrane filtration (1 L and 10 ml) showed after 5 days incubation at 35°C too many colonies to count. Only the 1 ml sample volumes gave results that could be quantitated. After installing and rinsing of the Smart2Pure 3 system according to the Smart2Pure water purification system operating instructions, the system was ready for sampling. 0.2 L of water from the system was used to rinse the 0.2 µm filter of the Smart2Pure 3 system and then 1 L samples were collected in sterile flasks.

Upload: others

Post on 14-Sep-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Determination of Bacteria Retention in the Barnstead Smart2Pure … · 2020. 8. 12. · volume samples were generated from the three 1 L samples: a 1 L, 10 mL, and a 1 mL sample

Determination of Bacteria Retention in the Thermo Scientific Barnstead Smart2Pure Water Purification System Verena Ruckstuhl, Thermo Fisher Scientific, Langenselbold, Germany and Julie Foster, Ph.D., Thermo Fisher Scientific, Asheville, NC

Ap

plica

tion

No

te

Key WordsWater purification, lab water, ultrapure water, bacteria, 0.2 micron filter, Thermo Scientific Barnstead, Smart2Pure

Abstract The bacteria retention of a Thermo Scientific™ Barnstead™ Smart2Pure™ water purification system was evaluated using the membrane filtration method based on the European Pharmacopoeia method described in Chapter 2.6.12.1

IntroductionBacteria are single celled organisms that can be found nearly everywhere in a busy laboratory. Although many of these bacteria are harmless to a person’s health, they can create unwelcome variability in an experiment. Because of their abundance and ability to be easily transferred, precautions such as sterile technique were employed. Using bacteria-free water during steps such as sample preparation, system rinsing, or buffer preparation is an easy method of reducing the chance of bacterial contamination.

Water purification systems are a reliable source for bacteria-free water. On average, bacteria such as Escherichia coli (E. coli), which are widely used in the laboratory, have an length of about 2 µm, and an average diameter of 0.5 µm.2 A 0.2 µm absolute membrane filter at the end of the system is used to remove any particles or bacteria that are larger than the pore size of the filter3. Proper maintenance of the water system, including filter replacement as specified in the manual, helps to ensure the water remains bacteria-free.

Ultrapure water from a Barnstead Smart2Pure water system was analyzed for the presence of aerobic bacteria. The Barnstead Smart2Pure 3 system was chosen from the family of Smart2Pure systems, which also includes Smart2Pure UV and UF models. All of these systems have the same feed water requirements, basic flow path and all dispense water through a 0.2 µM final filter. The systems are fed by tap water and utilize two purification steps to create 18.2 megohm lab water. The first step purifies the feed water using a reverse osmosis (RO) membrane to remove the majority of impurities in the water. The

second step of purification utilizes an ultrapure cartridge to remove any ions in the water. Additionally, an optional ultraviolet (UV) lamp and/or an ultrafilter (UF) can be added to the system configuration.

MethodsBacteria retention of the Smart2Pure system using tap water as feed water

The bacterial content of tap water was determined after disinfecting the water outlet with 70% ethanol and rinsing with 1 L water. Afterwards three 1 L sample volumes were collected and tested for bacterial growth using the membrane filtration technique.1 Three different volume samples were generated from the three 1 L samples: a 1 L, 10 mL, and a 1 mL sample. These three samples were filtered through a 0.2 µm cellulose nitrate (CN) membrane. The membrane was afterwards transferred aseptically to a R2A-Agar and incubated at 35°C for 5 days in a Thermo Scientific™ Heratherm™ compact microbiological incubator (model IMC18). The agar plates with sample volumes of the membrane filtration (1 L and 10 ml) showed after 5 days incubation at 35°C too many colonies to count. Only the 1 ml sample volumes gave results that could be quantitated.

After installing and rinsing of the Smart2Pure 3 system according to the Smart2Pure water purification system operating instructions, the system was ready for sampling. 0.2 L of water from the system was used to rinse the 0.2 µm filter of the Smart2Pure 3 system and then 1 L samples were collected in sterile flasks.

Page 2: Determination of Bacteria Retention in the Barnstead Smart2Pure … · 2020. 8. 12. · volume samples were generated from the three 1 L samples: a 1 L, 10 mL, and a 1 mL sample

The 1 L sample was then filtered through a 0.2 µm cellulose nitrate (CN) membrane, using a Thermo Scientific™ Nalgene™ analytical test filter funnel, as shown in Figure 2. The membrane was afterwards transferred aseptically to a R2A-Agar and incubated at 35°C for 5 days in a Heratherm compact microbiological incubator (model IMC18). The Smart2Pure 3 system was disinfected with the disinfection routine according to operating instructions, and the pretreatment cartridge, ultrapure polisher cartridge and 0.2 µm filter were replaced with new ones to retest the system.

ResultsAfter 5 days, any bacteria colonies found on the R2A-Agar plates were counted and the amount of bacteria per ml in the water were calculated. The data is summarized in Table 1.

ConclusionThe Smart2Pure system, which was directly connected to tap water, was able to remove the bacterial load of the tap water down to < 0.01 CFU/ml.

References1. European Directorate for the Quality of Medicines &

HealthCare (EDQM): European Pharmacopoeia 6.8, Microbiological Examination of Non-sterile Products: Microbial Enumeration Tests, 2010

2. Thermo Scientific Waterbook. 2014. Thermo Fisher Scientific. Retrieved 12 June, 2014 from http://www.thermoscientific.com/purewater

3. Operational Instructions for the Thermo Scientific Barnstead Smart2Pure water purification system. Thermo Fisher Scientific, Niederelbert, Germany.

thermoscientific.com/purewater © 2014 Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

ANWPS2PBACT 1014

Australia +61 39757 4300Austria +43 1 801 40 0Belgium +32 53 73 42 41China +800 810 5118 or +400 650 5118 France +33 2 2803 2180Germany national toll free 0800 1 536 376 Germany international +49 6184 90 6000

India toll free 1800 22 8374 India +91 22 6716 2200Italy +32 02 95059 552Japan +81 3 5826 1616Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Nordic/Baltic/CIS countries +358 9 329 10200

Russia +7 812 703 42 15Spain/Portugal +34 93 223 09 18Switzerland +41 44 454 12 22UK/Ireland +44 870 609 9203USA/Canada +1 866 984 3766

Other Asian countries +852 2885 4613 Countries not listed +49 6184 90 6000

thermoscientific.com

© 2013 Thermo Fisher Scientific Inc. All rights reserved. All (other) trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

Allemagne +49 6184 90 6000 Australie +61 39757 4300Autriche +43 1 801 40 0Belgique +32 53 73 42 41Chine +800 810 5118 or +400 650 5118 Espagne/Portugal +34 93 223 09 18

France +33 2 2803 2180 Inde +91 22 6716 2200Italie +32 02 95059 552Japon +81 3 5826 1616Nouvelle-Zélande +64 9 980 6700Pays Bas +31 76 579 55 55 Pays nordiques/baltes/CEI+358 9 329 10200

Russie +7 812 703 42 15Suisse +41 44 454 12 22UK/Irlande +44 870 609 9203USA/Canada +1 866 984 3766

Autres pays d’Asie +852 2885 4613 Autres Pays +49 6184 90 6000ou +33 2 2803 2180

thermoscientific.com

© 2013 Thermo Fisher Scientific Inc. Tous droits réservés. Les (autres) marques déposées sont des marques commerciales ou déposées de Thermo Fisher Scientific Inc et de ses filiales. Les caractéristiques, conditions et tarifs sont susceptibles d’être modifiés. Tous les produits ne sont pas disponibles dans tous les pays. Pour tout renseignement, veuillez vous adresser à votre distributeur local.

FRENCH

Australien +61 39757 4300Belgien +32 53 73 42 41China +800 810 5118 or +400 650 5118 Deutschland nationale kostenlose Hotline 0800 1 536 376 Deutschland international +49 6184 90 6000 Frankreich +33 2 2803 2180

Indien +91 22 6716 2200Italien +32 02 95059 552Japan +81 3 5826 1616Niederlande +31 76 579 55 55Neuseeland +64 9 980 6700Nordeuropa/Baltikum/CIS+358 9 329 10200Österreich +43 1 801 40 0

Russland +7 812 703 42 15Spanien/Portugal +34 93 223 09 18Schweiz +41 44 454 12 22UK/Irland +44 870 609 9203USA/Kanada +1 866 984 3766

Andere asiatische Länder +852 2885 4613 Andere Länder +49 6184 90 6000oder +33 2 2803 2180

thermoscientific.com

© 2013 Thermo Fisher Scientific Inc. Alle Rechte vorbehalten. Alle (anderen) Warenzeichen sind Eigentum von Thermo Fisher Scientific Inc. bzw. nationalen Niederlassungen. Änderungen von Spezifikationen, Bedingungen und Preisen vorbehalten. Nicht alle Produkte sind in allen Ländern erhältlich. Ausführliche Informationen erhalten Sie von Ihrem Thermo Fisher Scientific Ansprechpartner.

GERMAN

Australia +61 39757 4300Austria +43 1 801 40 0Belgio +32 53 73 42 41Cina +800 810 5118 or +400 650 5118 Francia +33 2 2803 2180Germania +49 6184 90 6000 Giappone +81 3 5826 1616

India +91 22 6716 2200Italia +32 02 95059 552Olanda +31 76 579 55 55Nuova Zelanda +64 9 980 6700Europa del Nord/CIS+358 9 329 10200Russia +7 812 703 42 15Spagna/Portogallo +34 93 223 09 18

Svizzera +41 44 454 12 22UK/Irlanda +44 870 609 9203USA/Canada +1 866 984 3766

Altre Nazioni Asiatiche +852 2885 4613 Nazioni non in elenco +49 6184 90 6000o +33 2 2803 2180

thermoscientific.com

© 2013 Thermo Fisher Scientific Inc. Tutti i diritti riservati. Tutti gli (altri) marchi sono di proprietà di Thermo Fisher Scientific Inc. e delle sue filiali. Specifiche, condizioni e tariffe possono subire variazioni. Non tutti i prodotti sono disponibili in tutti i paesi. Per maggiori dettagli consultare il rivenditore locale.

ITALIAN

Alemania +49 6184 90 6000Australia +61 39757 4300Austria +43 1 801 40 0Bélgica +32 53 73 42 41China +800 810 5118 or +400 650 5118 España/Portugal +34 93 223 09 18

Francia +33 2 2803 2180Holanda +31 76 579 55 55India +91 22 6716 2200Italia +32 02 95059 552Japón +81 3 5826 1616Nueva Zelanda +64 9 980 6700Paises Nórdicos/Bálticos/CIS +358 9 329 10200

Rusia +7 812 703 42 15Suiza +41 44 454 12 22UK/Irlanda +44 870 609 9203USA/Canadá +1 866 984 3766

Otros países de Asia +852 2885 4613 Otros paises +49 6184 90 6000o +33 2 2803 2180

thermoscientific.com

© 2013 Thermo Fisher Scientific Inc. Reservados todos los derechos. Las (otras) marcas comerciales son propiedad de Thermo Fisher Scientific Inc. y sus filiales. Características, condiciones y precios están sujetos a posibles cambios. Algunos productos no están disponibles en todos los países. Consulte los detalles con su representante comercial local.

SPANISHFigure 1. Diagram of a Smart2Pure water system with a 6 L internal reservoir.

Tap water Smart2Pure 3 system Sterile tube

Analysis of the samples

 

 

Tap  water   GenPure  Pro    

Analysis  of  the  sample  

Sterile  flask  

 

Analysis  of  the  samples  

Test  tube  Smart2Pure  3      

Tap  Water    

Figure 2. Set up of the filtering of the 1 L sample, using a Thermo Scientific Nalgene analytical test filter funnel, through the 0.2 µm cellulose nitrate (CN) membrane.

   

   

Table 1. Bacteria count from the tap water line (1 L and 10 ml)

Tap Water Sample (10 ml)

1st Run

2nd Run

Smart2Pure System Sample (1 L)

Tap Water Smart2Pure System Water

Bacteria Count n.a. <0.01(CFU/ml)

n.a. not applicable, the colonies were too numerous to evaluate

Ap

plica

tion

No

te