Advanced Heart Failure Center of Excellence Diagnosis of Device Thrombosis Andrew Civitello MD, FACC Medical Director, Heart Transplant Program Director, Advanced Heart Failure Fellowship Co-Director, Advanced Heart Failure Center of Excellence Baylor St. Luke's Medical Center / Texas Heart Institute

Upload: others

Post on 06-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Diagnosis of Device Thrombosis

Andrew Civitello MD, FACCMedical Director, Heart Transplant ProgramDirector, Advanced Heart Failure Fellowship

Co-Director, Advanced Heart Failure Center of ExcellenceBaylor St. Luke's Medical Center / Texas Heart Institute

Page 2: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Adapted from Mehra et al, J Heart Lung Transplant 2014;33:1-11

Trends in Patient Profiles, LVAD Volumes and Management Strategies

Secu

lar

Tren

ds

Implant Year 2008 2009 2010 2011 2012 2013

GI bleeding and vWF deficiencyLow reported thromboembolic events

Relaxed anticoagulation targetsReduced emphasis on post-op bridging

Aortic valve pathology emergesLower pump speed for AV openingM

anag

emen

t Co

nsid

erat

ions

Spike in HeartMate IIThrombosis

Page 3: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

• Development of clot within the flow path of any the pump components:

• Inflow cannula

• Pump housing / rotor

• Outflow graft

• Thrombus may originate in the pump, or

• from the left atrium or ventricle, or

• from right sided chambers and travel through a septal defect

• And lodge in any, or all, of the pump components

Definition of Pump Thrombosis

Goldstein, J Heart Lung Transplant 2013

Page 4: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

LVAD Thrombosis

Rotor

Outflow Cannula

Inflow Cannula

Page 5: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Scandroglio, et al J Am Coll Cardiol. 2016;67(23):2758–68

Diagnosis of Pump Thrombosis May be Elusive

Tchantchaleishvili et al Ann Cardiothorac Surg 2014;3(5);490-495

Page 6: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Adapted from Goldstein et al, J Heart Lung Transplant 2013:32;667-670

Algorithm for the Diagnosis of Suspected LVAD Thrombosis

Page 7: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Diagnosis of VAD Thrombosis - 2018

Adapted from Rame and Birati, J Am Coll Cardiol HF 2015;3 (11); 857-858

Assess other causes of HF or change in pump parameters

Normal

Admit to HospitalConsider intravenous anticoagulation

>3X Upper Normal Limit

Check LDH

Clinical Suspicion of VAD Thrombosis (At least one of the following):• Signs of Symptoms Suggestive of VAD thrombosis• Evidence of hemolysis• Change in pump parameters

Page 8: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Thrombus Formation

Hemolysis, but NoHemodynamic Compromise

Incomplete Thrombosis

Hemolysis, withAbnormal PumpFunction

Complete Thrombosis

Pump Stop WithCardiogenic Shock

The Relationship Between Hemolysis and VAD Thrombosis

Blitz, Ann Cardiothorac Surg 2014;3(5):450-471

Page 9: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Elevated LDH Levels Within 3 Months After HM II Implantation

Starling et al, N Engl J Med 2013;370:33-40

Page 10: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Shah et al, J Heart and Lung Transplant 2014:33;102-104

LDH and Serum Free Hb as Markers of Device Thrombosis

Page 11: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Consider Pump Exchange or OHTNo

Diagnosis of VAD Thrombosis - 2018

Adapted from Rame and Birati, J Am Coll Cardiol HF 2015;3 (11); 857-858

Diagnosis:Echo Ramp studyHigher peak powerConsider CT angioConsider RHC

Assess other causes of HF or change in pump parameters

Normal

HM II HVAD

Yes

Hemodynamically Stable?

Admit to HospitalConsider intravenous anticoagulation

>3X Upper Normal Limit

Check LDH

Clinical Suspicion of VAD Thrombosis (At least one of the following):• Signs of Symptoms Suggestive of VAD thrombosis• Evidence of hemolysis• Change in pump parameters

Page 12: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Echocardiography for Device Thrombosis

Increased LV size by LVIDd after LVAD thrombosis

Stainback et al, J Am Soc Echocardiogr 2015;28:853-909.)

Page 13: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Echocardiography Ramp Test - Diagnosis of Device Thrombosis

Increased AV Opening Time by M-mode

Stainback et al, J Am Soc Echocardiogr 2015;28:853-909.)

Page 14: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Echocardiography Ramp Test - Diagnosis of Device Thrombosis

Low Velocity Inflow Cannula Peak velocity

Stainback et al, J Am Soc Echocardiogr 2015;28:853-909.)

Page 15: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Echocardiography Ramp Test - Diagnosis of Device Thrombosis

Abnormal Inflow Cannula Colorflow

Stainback et al, J Am Soc Echocardiogr 2015;28:853-909.)

Page 16: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Page 17: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

LVAD Ramp Study: Normal Patient LVAD Ramp Study: Thrombosis Patient

Normal vs Device Thrombosis RAMP Echo Study

Uriel et al, J Am Coll Cardiol 2012;60:1764-75

Page 18: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Ramp Echo Study - Normal vs Device Thrombosis

Uriel et al, J Am Coll Cardiol 2012;60:1764-75

Page 19: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Ramp Echo Study - Normal vs. Device Thrombosis

LVEDD Slopes

LDH Values

Obstruction to Flow

No Obstruction to Flow

No Obstruction to Flow

Obstruction to Flow

Uriel et al, J Am Coll Cardiol 2012;60:1764-75

No patient with VAD Thrombosis had a

LVEDD Slope <- 0.16

Best LDH Cutoff>1103

(5X ULM)

Page 20: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

LVEDD Slope > -0.16 LVEDD Slope < -0.16RAMP echo

Check for Hemolysis: LDH, Plasma Free HemoglobinCheck Anticoagulation: INR, PTT

Using RAMP Echo Studies to Predict VAD Thrombosis

Adapted from Uriel et al, J Am Coll Cardiol:60;1764-75

Suspected Device MalfunctionLDH Elevation, Power Spikes

Device Thrombosis is Likely No Device Thrombosis or Malfunction

Page 21: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Consider Pump Exchange or OHTNo

Diagnosis of VAD Thrombosis - 2018

Adapted from Rame and Birati, J Am Coll Cardiol HF 2015;3 (11); 857-858

Diagnosis:Echo Ramp studyHigher peak powerConsider CT angioConsider RHC

Assess other causes of HF or change in pump parameters

Normal

HM II HVAD

Yes

Hemodynamically Stable?

Admit to HospitalConsider intravenous anticoagulation

>2.5X Upper Normal Limit

Check LDH

Clinical Suspicion of VAD Thrombosis (At least one of the following):• Signs of Symptoms Suggestive of VAD thrombosis• Evidence of hemolysis• Change in pump parameters

Page 22: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Computed Tomography Angiography (CTA)

• Helpful in diagnosing mechanical issues predisposing to pump thrombosis

• Important positive findings include:

- Left ventricular thrombus

- Dilated left ventricle

- Cannula malposition toward any of the left ventricular walls

- Outflow graft thrombus, kinking or twisting

- Lack of opacification of the outflow graft

Blitz, Ann Cardiothorac Surg 2014;3(5):450-471Birati et al Curr Treat Options Cardio Med (2015) 17: 2

Page 23: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Misdirection of the apical cannula toward the interventricular septum

Computed Tomography Angiography (CTA)

Milano et al, J Heart Lung Transplant 2011;30:838–40

Page 24: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Computed Tomography Angiography (CTA)

Krishnan et al, J Cardiovasc Comp Tomography 2014;8:473-474

Mural thrombus in the proximal LVAD outflow cannula

Page 25: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Consider Pump Exchange or OHTNo

Diagnosis of VAD Thrombosis - 2018

Diagnosis:• Log file analysis• Higher peak powerConsider CT angioConsider RHC

Adapted from Rame and Birati, J Am Coll Cardiol HF 2015;3 (11); 857-858

Diagnosis:Echo Ramp studyHigher peak powerConsider CT angioConsider RHC

Assess other causes of HF or change in pump parameters

Normal

HM II HVAD

Yes

Hemodynamically Stable?

Admit to HospitalConsider intravenous anticoagulation

>3X Upper Normal Limit

Check LDH

Clinical Suspicion of VAD Thrombosis (At least one of the following):• Signs of Symptoms Suggestive of VAD thrombosis• Evidence of hemolysis• Change in pump parameters

Page 26: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

RAMP Echo Slope Does Not Diagnosis HVAD Thrombosis

Uriel et al, J Cardiac Fail 2015:21;785-791

Page 27: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

HVAD Log File Analysis to Diagnosis Thrombosis

Najjar et al, J Heart Lung Transplant 2014;33:23–34

Page 28: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

HVAD Log File Analysis to Diagnosis Thrombosis

Najjar et al, J Heart Lung Transplant 2014;33:23–34

Page 29: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Consider Pump Exchange or OHTNo

Diagnosis of VAD Thrombosis - 2018

Diagnosis:Log file analysisConsider CT angioConsider RHC

Adapted from Rame and Birati, J Am Coll Cardiol HF 2015;3 (11); 857-858

Diagnosis:Echo Ramp studyIncreased peak powerConsider CT angioConsider RHC

Assess other causes of HF or change in pump parameters

Normal

HM II HVAD

Yes

Hemodynamically Stable?

Admit to HospitalConsider intravenous anticoagulation

>3X Upper Normal Limit

Check LDH

Clinical Suspicion of VAD Thrombosis (At least one of the following):• Signs of Symptoms Suggestive of VAD thrombosis• Evidence of hemolysis• Change in pump parameters

Page 30: Diagnosis of Device Thrombosis...Advanced Heart Failure Center of Excelle nce Adapted from Mehra et al, J Heart Lung Transplant 2014;33:111-T rends in Patient Profiles, LVADVolumes

Advanced Heart FailureCenter of Excellence

Questions?