dio-562 brief manual

12
V Zahradách 24/836 180 00 Prague 8 Tel: +420-2-8384-2063, -2064, -3022 Fax: +420-2-8384-1067 e-mail: [email protected] Ultrasonic Flaw Detector DIO 562 Brief Operation Manual

Upload: dylkanwings

Post on 20-Nov-2015

132 views

Category:

Documents


32 download

DESCRIPTION

Starmans DIO 562 Intructions manual - Brief vers.

TRANSCRIPT

  • V Zahradch 24/836 180 00 Prague 8 Tel: +420-2-8384-2063, -2064, -3022 Fax: +420-2-8384-1067 e-mail: [email protected]

    Ultrasonic Flaw Detector

    DIO 562

    Brief Operation Manual

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 2 -

    CONTENTS

    1. DIO 562 MENUS ....................................................................................... 3

    1.1. OPERATION MENUS ........................................................................................................ 3

    1.2. OPERATION SUB-MENUS ................................................................................................ 3

    1.3. AUXILIARY MENUS ......................................................................................................... 3

    2. BASIC PRINCIPLES OF OPERATION................................................ 4

    2.1. CONNECTION OF ULTRASONIC TRANSDUCERS ............................................................... 4

    2.2. SWITCHING ON............................................................................................................... 4

    2.3. SWITCHING OFF............................................................................................................. 4

    2.4. MAIN MENU ..................................................................................................................... 4

    2.5. POWER SUPPLY AND ACCUMULATOR CHARGING ........................................................... 4

    2.6. MEMORY CHIP ................................................................................................................ 5

    2.7. MAINTENANCE ................................................................................................................ 5

    2.8. OPERATION OF THE FLAW DETECTOR............................................................................ 5

    2.9. PROGRAMMING OF THE FUNCTION KEYS F1 F5.......................................................... 5

    3. BRIEF DESCRIPTION OF INSTRUMENT FUNCTIONS BY MENU6

    3.1. OPERATION MENU .......................................................................................................... 6 3.1.1. PARAM Menu.......................................................................................................................6 3.1.2. GATES Menu........................................................................................................................6 3.1.3. MEMO Menu........................................................................................................................7 3.1.4. TOOLS Menu........................................................................................................................8

    3.2. OPERATION SUB-MENUS ................................................................................................. 8 3.2.1. STORE Sub-Menu ................................................................................................................8 3.2.2. CALIB Sub-Menu .................................................................................................................8 3.2.3. SETUP Sub-Menu ................................................................................................................9 3.2.4. MISCEL. Sub-Menu .............................................................................................................9

    3.3. AUXILIARY MENUS........................................................................................................ 10 3.3.1. HWSET Menu .....................................................................................................................10 3.3.2. AVG/DGS Menu .................................................................................................................11 3.3.3. DAC Menu..........................................................................................................................12

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 3 -

    1. DIO 562 Menus

    1.1. Operation Menus

    1.2. Operation Sub-Menus

    1.3. Auxiliary Menus

    HWSET (Hardware Setting) PRB/AVG (Probe parameters for AVG/DAC) Setting of main parameters List of probes MAIN MENU (described below in 1.4) allows setting of special modes. Access to the Main Menu is possible only during switching the instrument ON by simultaneous depressing and ON.

    1 2

    3 4

    5

    1 2 3 4

    6

    5 6

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 4 -

    2. Basic principles of operation

    2.1. Connection of ultrasonic transducers Normal and angle probes (generally, single element transducers) should be connected always to

    the BNC IN connector (IN 1 for two channel version). Dual element transducers shall be connected to BNC IN and OUT connectors (IN 1 and OUT 1 for two channel version). These connectors shall be used also for separate transducers in through-transmission mode.

    2.2. Switching ON Press the ON key to switch the instrument ON. will appear on the

    display. Then press any arrow (cursor) key to continue. The instrument is ready and PARAM menu is displayed.

    2.3. Switching OFF Press the OFF key to switch the instrument OFF from within any mode.

    2.4. Main menu

    While the instrument is switched OFF, press simultaneously and ON keys. Main menu will

    appear on the display. Use and keys to move the cursor and or key to select the appropriate item (command).

    Since these functions are not needed for routine operation of the flaw detector, their description

    is given in the section Special functions. The use of particular items should be avoided if operator is not sure or does not have appropriate skills. Otherwise, data on the memory chip may be lost and/or principal changes of internal parameters of the instrument might occur!

    2.5. Power supply and accumulator charging NiMH (LiIon optional) accumulator is used for power supply. Before charging, it is

    recommended to fully discharge the accumulator at least once a week. Charger supplied with the instrument should be used. Charging time of fully discharged NiMH accumulator is approx. 16 hours.

    A battery symbol in the bottom right corner of the display indicates a capacity (voltage) of the accumulator. When the voltage has been decreased below the preset level, the battery symbol starts blinking, with audible signal.

    Information about the accumulator voltage in Volts is available in the operation menu TOOLS MISC. The voltage value should be within 11 V to 13.9 V for proper operation.

    The instrument may be normally used during the charging cycle.

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 5 -

    2.6. Memory chip Internal memory chip (capacity 1 MB) allows saving 100 setups, 100 AVG (DGS) and 100 DAC

    curves, 780 echographs of A-scans and 100 B-scans, including 10 minutes of voice records. All data and setups may be saved to a PC (using standard delivered software and connecting cable) and load from PC.

    2.7. Maintenance DIO562 does not require any special maintenance, except timely charging the accumulator.

    2.8. Operation of the flaw detector There is an operation menu with sub-menu for routine operation of the flaw detector next to

    the right edge of the display. Use and keys to move the cursor within the appropriate menu. Selected item or its value is always highlighted by blinking cursor and inverse display. Use or key to change the value of the item (function), to switch function On and OFF or move to the next menu or operation sub-menu see DIO 562 Menu.

    In order to switch to the next operation menu, use or key when the cursor is on the first (upper) position (name) of the operation menu.

    Keys F1 to F5 (so called function keys) are user programmable to individual need of the operator.

    2.9. Programming of the function keys F1 F5 The programmable keys F1 F5 have two functions. Short keystroke moves the cursor on the

    preprogrammed menu item (within any operation menu or sub-menu). Long press (longer than approx 3 seconds) recalls the instrument setup (i.e. saved settings of the instrument parameters).

    Programming of function keys for selected menu items is very simple. If the function key had not been assigned yet, simply position the cursor on the selected item and then press the desired function key. Single acoustic signal sounds as soon as the function key has been pressed and double tone sounds after the key release. The cursor will remain on the selected item. If a single tone sounds after the key release and the cursor moves to another item, then this key had been already assigned. Deleting of the function key assignments is described in the chapter Sub-menu MEMORY. Setting (assignment) of the function keys is saved to the instruments memory with saved setups.

    The second function of F1 F5 keys serves for recall of the saved settings (setup). This function cannot be changed. In order to recall the setup from the memory, press the function key for longer than approx. 3 seconds (until the sound signal). Press two function keys simultaneously for numbers 6 to 9 as follows:

    F Key F1 F2 F3 F4 F5 F1+F2 F1+F3 F1+F4 F1+F5

    Setup number within group 1 2 3 4 5 6 7 8 9

    This allows the direct access to the nine setups within a group. The group means choice from ten

    setups (1-9, 11-19, 21-29 91-99). The desired group of setups shall be set in the TOOLS SETUP sub-menu. For example, if the memory number #35 had been set, then setup number 31 will be recalled by pressing F1 key (not setup number 1) see section Sub-menu SETUP.

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 6 -

    3. Brief description of instrument functions by menu

    3.1. Operation Menu DIO 562 offers 4 operation menus (PARAM, GATES, MEMO, TOOLS). When the instrument

    had been switched ON, the PARAM menu is displayed as the first.

    3.1.1. PARAM Menu This menu allows setting the basic parameters of the flaw detector and control of

    most frequently used functions.

    GAIN gain adjustment (0.0 80.0 dB)

    STEP G setting of the gain step (0.1; 0.5; 1.0 or 6.0 dB)

    MAT.VEL switches two sound propagation velocities, which had been set in xxxx m/s

    xxxx m/s setting of sound propagation speed in the material (1000 9999 m/s)

    >0 s delay (0.00 50.00 s); adjustment of the delay, e.g. within the probe shoe or wedge

    dB-REF setting of the reference echo for generation of

    AVG/DGS curve (name changes to REF) or

    DAC (name changes in sequence to REF-01 to REF-05)

    GATE1 fast and simple adjustment of the first gate when the cursor positioned on , setting the threshold level, when the cursor positioned on , setting the gate position within the range

    battery symbol indicates the remaining accumulator capacity

    mm symbols displayed under the time base, which allow: 1) zoom, when the cursor is positioned on 2) setting the time base (range), when the cursor is positioned on 3) time base scale setting

    - for normal probe in the sound path (millimeters, mm, or microseconds, s); - for angle probe (the angle shall be set in TOOLS CALIB menu) in the

    sound path (Smm), projection distance or shortened projection distance (Pmm). Time base may be calibrated in inches as well in TOOLS CALIB menu

    3.1.2. GATES Menu The menu is used for setting the parameters for operation with gates (sometimes

    called monitors). The instrument can operate with three gates simultaneously. Parameters of each gate may be set for each gate individually. The appropriate gate is assigned by number.

    switches ON/OFF the enlarged display of the chosen value (chosen in TOOLS MISC. BIG P), instead of the menu. Echograph may be enlarged also (no menu on the display)

    GAIN gain adjustment (0.0 80.0 dB). NOTE: when the GATE 3 had been chosen on the GATE position, the gain adjustment will be applied only within this gate range the general gain of the instrument remains unchanged!

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 7 -

    GATE 1,2,3 selection of the gate to be adjusted

    gate threshold level (10 95 % of the full screen height FSH)

    > front edge of the gate (numeric value indicates the position in mm or s, depending on the setting of the instrument)

    < rear edge of the gate (numeric value indicates the position in mm or s, depending on the setting of the instrument)

    ON / OFF switch the selected gate ON and OFF

    selection of the gate triggering logic: (triggers above positive, or below negative; switch the sound signalization ON and OFF

    MEASURE selection of the measuring mode within the GATE 1 (also THC. FF, or THC. PP in the GATE 2, which is switched ON automatically); OFF THC. FF = measuring echo-echo (flank to flank) THC. PP = measuring echo-echo (peak to peak) THC. 0F = measuring zero-echo (time base zero to echo flank) THC. PP = measuring zero-echo (time base zero to echo peak)

    mm symbols displayed under the time base, which allow zoom and time base settings, the same as in the PARAM menu

    3.1.3. MEMO Menu This menu may be used to save the echographs (A-scans and/or B-scans) to the

    memory, review saved records (incl. voice records - memos).

    GAIN setting the gain (same as in PARAM menu)

    RECORD selection of the record; each record (echograph) is assigned by letter A to Z and number 00 to 29 for saving/loading echographs to/from the memory

    Save saving the echographs to the memory. Short press of the key will save the single echograph (long press activates the voice memo recording). When the is pressed and hold, the echographs will be saved in sequence (approx. 20 echographs per 1 second), while the key is depressed, assuming that there are free memory locations available

    View viewing the echograph on the display, without change of the instrument settings

    Recall loading the echograph from the memory, including settings, which were active during the saving of the echograph

    Freez freezing the echograph. This function depends on the arrow key used: 1) key switches functions

    ALL (instantaneous freezing of entire A-scan), B-scan (B-scan mode switched ON) or OFF (freeze OFF)

    2) key switches functions MAX (freezing the maximum values in the echograph, with continuous loading subsequent max. amplitudes. Repeated pressing the key will erase the record, but the function remains active. Press the key to switch this function OFF.

    3) pressed twice on the position OFF will switch the B-scan ON

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 8 -

    3.1.4. TOOLS Menu This menu allows the access to other sub-menus.

    STORE switches to the sub-menu STORE (memory)

    CALIB switches to the sub-menu CALIB (for angle probe settings) SETUP switches to the sub-menu SETUP (save/load setups)

    MISC. switches to the sub-menu MISC. HWSET hardware setting; switches to the sub-menu HWSET

    PRB/AVG setting the parameters for generation of the evaluation curves; the curve type (AVG/DGS or DAC) shall be chosen in the submenu MISC. The name of this item will be displayed depending on the chosen curve type AVG or DAC

    ON (OFF) switches the AVG/DGS or DAC curve ON/OFF HELP brief operation manual. It is saved on the memory chip and may be edited or

    amended, as needed. NOTE: The HELP file may be edited on the PC using the delivered software

    DEMO switches do the DEMO version

    3.2. Operation sub-menus The operation sub-menus are accessible from the TOOLS menu.

    3.2.1. STORE Sub-Menu It allows work with echographs and voice records saved in the memory and erase the

    function keys (Fx). Serie selection of the memory location for transfer of the echographs,

    deleting data and deleting the voice memo Send transfer of the selected echograph from the instrument memory to PC

    CLR erase the selected echograph; unwanted erase is prevented by the key symbol

    Voice recording of the voice memo; F 99/0 free memory for voice record. The number after slash indicates unavailable memory. NOTE: some memory remains unavailable even after deleting of sound records)

    CLR erase all voice memos; unwanted erase is prevented by the key symbol CLR Scn erase C-scan; not used in two-channel model CLR Fx erase the function keys assignments; the function keys assignments may be erased

    individually or all at once

    3.2.2. CALIB Sub-Menu The CALIB sub-menu allows setting the parameters for angle probes. This is

    required for proper display of the position co-ordinates of defects (defect depth and distance from the probe). Co-ordinates computing and display shall be switched ON in the GATES menu.

    Angle sound beam angle in the metal (10 - 90); OFF = function switched OFF for straight beam

    Thc.mat. setting the thickness of the material under inspection (0.1 999.9 mm)

    X-value setting the distance from the probe index point to the probe case edge (0.0 - 999.9 mm), for time base readings in the shortened projection distance

    UNIT mm selection of the measuring units for the time base: millimeters, mm, or inches

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 9 -

    3.2.3. SETUP Sub-Menu The SETUP sub-menu allows saving and loading the instrument setup parameters

    to/from the memory.

    LOAD recall the saved setup parameters from the memory by the selected memory location number

    SAVE save setup parameters to the memory; the memory location number (optionally with description any five characters, letters and/or numbers entered using the cursor keys) shall be selected before saving

    #### 4 selection of the memory location number (0 99)

    ***** description of the memory location for future identification of the setup (e.g. K15K = probe type K15K)

    3.2.4. MISCEL. Sub-Menu The MISCEL. (MISC.) sub-menu allows switch ON/OFF, selection or setting of

    various additional parameters or functions.

    CRV-AVG(DAC) selection of the AVG/DGS or DAC curve for evaluation

    PWS DIS display power saving mode ON/OFF; the display backlight will be switched OFF (after 15 s or 30 s), if no indication is present within any gate during the preset time

    GRF-LIN selection of the echograph display mode:

    1) GRF-LIN line display of the cerograph

    2) GRF-SOL solid display of the echograph (better readability under the bright light, e.g. sunshine)

    3) GRF-FFT frequency spectrum display (for transducer evaluation only)

    BRIGHT4 setting of the display backlight brightness (0 9)

    VOLUME2 setting the loudness of the acoustic signalization for indication within the gate (0 9)

    C +6 0 setting the auxiliary curves range, in dB, for AVG/DGS or DAC curve

    AVRG 4x setting the averaging (1x, 4x, 16x, 64x), i.e. eliminating the random echoes. NOTE: The instrument response with averaging will be slower, i.e. lower display rate

    BIG P selection of the reading value to be enlarged on the display. Enlarged reading display shall be activated in the GATES menu. If there is only one value on the display, then this value will be enlarged abstractedly from the selected item. Meaning of the particular symbols is as follows: P projection distance or shortened projection distance; S sound path; D depth of the defect; E - equivalent size of the defect (when the AVG/DGS or DAC curve is activated); A relative amplitude of the measured echo, compared with AVG/DGS (DAC) curve; enlarged value display switched OFF

    RFSH2/s refresh rate (2, 3 or 5), per second, of the numeric values on the display

    MR 500 setting the pulse repetition rate (50 Hz to 5000 Hz)

    REJ 00% setting the reject level (00 91 %)

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 10 -

    3.3. Auxiliary menus The auxiliary menus are accessible from the TOOLS menu

    3.3.1. HWSET Menu The HWSET menu (hardware settings) allows setting of

    the basic parameters of the instrument electronic circuits. PULSER label, for information only VT+W Lw setting the pulse voltage and width

    (transmitter power) (Hw high voltage and wide pulse, Hn high voltage and narrow pulse, Lw low voltage and wide pulse, Ln - low voltage and narrow pulse)

    WD 250n setting the pulse width (62 ns to 500 ns); for higher pulse number changes form WD to BF and indicates the pulse basic frequency (1 8 MHz)

    NUM. 1 number of cycles in one transmitted pulse (1 16) DAMPING damping of the instrument input (30, 33, 300, 4700 Ohm), for electrical

    matching the probe. For new version DAM IAPMP (ON 0dB, ON +20dB, OFF 0dB, OFF +20dB)

    FILTER input filter: WIDEBAND = basic setting; LC 5.0M = filter for 5MHz probe; LC 4.0M = filter for 4MHz probe; LC 2.2M = filter for 2.25MHz probe; fir 2MHz (1MHz, 5MHz, 2M5) = special digital filter for special purposes

    DETECTOR signal rectifying: FULL = fully rectified; FULL_LP = fully rectified with filter; HF = high frequency display, not rectified; FULLL_PD = fully rectified with peak detector for transducers with higher frequencies

    RANGE coarse setting the time base range (0.07 ms, 0.6 ms, 3.1 ms); for steel and pulse-echo mode 0.6 ms = time base range from 18 mm to 1500 mm; 3.1 ms = time base range from 90 mm to 7500 mm; 0.07 ms = time base range from 2,5 mm to 180 mm.

    TX-RX selection of the operating mode: single = for single element transducer; dual = for dual element transducer or through-transmission mode

    INP.SEL channel input selection (for two-channel model) first = first channel; second = second channel; altern = switching-over both, first and second channel

    ESC (F1) leave menu; press F1 to move the cursor from any item to the ESC position, then press or to leave the menu

    CLR TMR reset the instrument operation time. Displayed value indicates the total operation time since the last reset

    X-scan selection of the display mode and its parameters: B-scan, C-scan and D-scan Additional modes: Escan = communication DIO 562 with PC via RS232 connector; Fscan = thickness and amplitude data via RS232, Gscan = analog output 0-5 V (output voltage is proportional to the echo amplitude or position within the GATE 1)

    k = 100/p (1 999/p) constant for computing the display rate (for B-scan only) POS. OFF (x, xy, Ix) selection of the co-ordinates for scan EMUX OFF (IN1 IN8) selection of the channel SYNC OFF (MST, SLV, FLW, DLY) synchronization setting DLY 1ms (0 ms 10 ms) delay after synchronization 19.JAN 03 setting the internal date and time

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 11 -

    3.3.2. AVG/DGS Menu This menu allows setting the parameters for generating

    AVG/DGS curve for evaluation. The type of the curve shall be selected in the MISC. sub-menu.

    ACTUAL SaveAs (00 - 99) number of the memory position for particular probe data saving

    PROBE probe name (type); ***** (any five characters, i.e. letters and/or numbers). The probe name will be displayed under the selected curve type in the TOOLS menu

    ESC (F1) leave menu; press F1 to move the cursor from any item to the ESC position, then press or to leave the menu

    DELAY delay in the probe wedge: c = (1000 - 9999) setting the sound velocity in the wedge

    l = (0.0 mm 99.9 mm) setting the thickness of the wedge

    SETzero computing the delay for PARAM menu

    CRYSTAL setting the transducer parameters:

    f = (0.50 MHz 20.00 MHz) frequency of the transducer D = (3.0 mm 35.0 mm) effective diameter of the transducer

    disk/a * b used for rectangular elements only; dimensions of the rectangular transducer will be automatically recalculated to effective diameter of the circular transducer (value indicated in item D)

    REF-ECHO (BW, DSR or SDH); setting the reference reflector type for AVG/DGS or DAC curve: BW back wall; DSR disc shaped reflector (flat bottomed hole); SDH side drilled hole

    REF-SIZE reference reflector size for AVG/DGS or DAC curve; diameter in mm for DSR or SDH. For BW the INFIN. (infinite surface) value will be set automatically

    ANGL (OFF, 10 - 90) beam angle

    Xv= (0.0 mm 999.9 mm) probe index value

    ATT-REF (0.0 dB/m 99.9 dB/m) attenuation in the reference block

    ATT-TEST (0.0 dB/m 99.9 dB/m) attenuation in the inspected material. May be adjusted even after creating the curve in the PARAM menu

    AMPL-CORR (-20.0 dB - +20.0 dB) gain correction, in dB, for surface roughness

    T-CORR (-20.0 dB - +20.0 dB) correction for transducer focusing, dB

  • Brief Operation Manual DIO 562

    STARMANS Electronics Ltd. - 12 -

    3.3.3. DAC Menu This menu allows setting the parameters for generating

    DAC curve for evaluation. The type of the curve shall be selected in the MISC. sub-menu.

    ACTUAL SaveAs (00 - 99) number of the memory position for particular probe data saving

    PROBE probe name (type); ***** (any five characters, i.e. letters and/or numbers). The probe name will be displayed under the selected curve type in the TOOLS menu

    ESC (F1) leave menu; press F1 to move the cursor from any item to the ESC position, then press or to leave the menu

    DELAY delay in the probe wedge: c = (1000 - 9999) setting the sound velocity in the wedge

    l = (0.0 mm 99.9 mm) setting the thickness of the wedge

    SETzero computing the delay for PARAM menu

    CRYSTAL setting the transducer parameters:

    f = (0.50 MHz 20.00 MHz) frequency of the transducer D = (3.0 mm 35.0 mm) effective diameter of the transducer

    disk/a * b used for rectangular elements only; dimensions of the rectangular transducer will be automatically recalculated to effective diameter of the circular transducer (value indicated in item D)

    REF-ECHO (BW, DSR or SDH); setting the reference reflector type for AVG/DGS or DAC curve: BW back wall; DSR disc shaped reflector (flat bottomed hole); SDH side drilled hole

    REF-SIZE reference reflector size for AVG/DGS or DAC curve; diameter in mm for DSR or SDH. For BW the INFIN. (infinite surface) value will be set automatically

    ANGL (OFF, 10 - 90) beam angle

    Xv= (0.0 mm 999.9 mm) probe index value

    R1: R5: (-99.9 0 dB) amplitude value, in dB, of the particular points of the DAC curve

    X mm (1 9999 mm) position of the particular points of the DAC curve