doc and tn determination: a comparison of instruments from...
TRANSCRIPT
DOC and TN determination: acomparison of instruments from
Elementar and Shimadzu –High Temperature Catalytic Oxidation
(HTCO) AnalyzersZuzanna Cie śla
Anna KowalskaForest Research Institute
Poland
3rd Meeting of the Heads of the Laboratories, 16-17 June 2011, Arcachon, France
Elementar LiquiTOC High Temperature TOC/TNb
General specifications:Injection volume 0,05 – 4 ml
Carrier gas Oxygen or Synthetic air
Flow rate 200 ml/min
Catalyst 1% Pt on TiO2
Analysis mode TIC/NPOC Mode ⇒ automatic dosing of HCl, integration of TIC peak
followed by NPOC peak
TNb detector Electrochemical detection (EC cell)
CO2 detector IR
Reactor temperature 800 °C
Calibration standards mixture of :
-Potassium hydrogen phthalate (KHP)
-Sodium carbonate (Na2CO3)
-Ammonium chloride (NH4Cl)
-Sodium nitrate (NaNO3)
QC standards Glycine (5 mg/l C; 2,91 mg/l N)
AccuStandards : 10 mg/l
Total Organic Carbon (KHP in water+H2SO4)
Total Organic Nitrogen (NH4Cl in water+H2SO4)
Analysis time approx. 18 minutes3rd Meeting of the Heads of the Laboratories, 16-17 June 2011, Arcachon, France
Problems (1):
– First problem at start-up : the electrochemical
TNb detector was not functional (as the
instrument had been used only sporadically
during 4 years) and had to be replaced
– Carry-over (memory) effect for DOC : observed
when water samples and also standards with DOC
content ≥ approx. 10 mg/l are followed by
samples with lower concentration:
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
Problems (2):
• ⇒ this problem may be caused by the poor efficiency of the catalyst and by the design of the instrument itself, in which DI water is not only used for rinsing the tubing and the sampling needle but also for sample delivery into the reactor,
• Problems with nitrogen recovery (TN < NH4+ +
NO3-) but only in case of some throughfall
samples with high levels of nitrogen (>15mg/l) not in case of standards,
• Very low sensitivity when using standard injection volume (0,5 ml),3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
Problems (3):
• Poor accuracy in the NPOC/TNb mode (only NPOC peak integration) which requires external acidification and Inorganic Carbon removal but allows short analysis time (approx. 8 min.),
• Quadratic characteristic of the calibration curve for TNb and few fitting options possible in software,
• No possibility to automatically select the optimal calibration range ⇒ an additional injection needs to be performed.
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
Solutions:
• 1 calibration range was chosen for both DOC and TNb
determination in routine analysis ⇒ 1 – 50 mg/l with
an injection volume of 0,8 ml,
• Samples with concentration < 1 mg/l are analyzed in a
0 - 2 mg/l range with a 2 ml injection volume,
• To minimize the carry-over effect for DOC, samples
are grouped in the analysis sequence according to
their type (bulk open field, throughfall, stemflow, soil
solution) and the plot of origin and separated by two
injections of a DI water:
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
Shimadzu TOC-VCPN+TNM-1 Analyzer with TNb unit
with chemiluminescence detection.
General specifications:Injection volume 10 – 2000 μl
Carrier gas Synthetic air
Flow rate 125 ml/min
Catalyst 0,5% Platinum on Alumina Pellets
Analysis mode NPOC method ⇒ automatic HCl addition and sparging
TNb detector Chemiluminescence detector
CO2 detector Non-dispersive Infrared detector (NDIR)
Reactor temperature 720 °C
Calibration standards TC/IC standard:
-Potassium hydrogen phthalate (KHP)
-Sodium carbonate (Na2CO3)
-Sodium hydrogen carbonate (NaHCO3)
TNb standard:
-Potassium nitrate (KNO3)
QC standards Glycine ( 5 mg/l C; 2,91 mg/l N)
AccuStandards : 10 mg/l
Total Organic Carbon (KHP in water+H2SO4)
Total Organic Nitrogen (NH4Cl in water+H2SO4)
Analysis time approx. 10 minutes (3 replicates)
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
Properties (1):
– No carry-over effect observed for DOC even with significant concentration differences among samples in the sequence ⇒
• Automatic assessment of the analysis results based on the coefficient of variation: if the entered CV limit (2%) is exceeded additional injections are performed and first results excluded,
• Different construction: DI water is used only to wash the flow line to the injection tube and for auto dilution,
• Possibility to periodically conduct a catalyst regeneration procedure which restores the catalyst’s performance.
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
Properties (2):
– Very good sensitivity and repeatability with a 50
µl injection volume,
– Very high correlation and linearity of the
calibration curves within the analytical range, but
still the calibration needs to be adjusted more
precisely for the lowest concentrations ( 0 – 1
mg/l),
– Automatic selection of optimum calibration curve
( up to 3 calibration curves in a method).
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France
Method Validation Characteristics Elementar Shimadzu
DOC
Calibration and
analytical range
With every sample
sequence
1 – 50 mg/l
Monthly
0 – 50 mg/l
Regression coefficient
(R2)
0,9988 0,9997
Precision 6,47 % 1,30 %
LOQ 0,82 mg/l 0,24 mg/l
TNb
Calibration and
analytical range
With every sample
sequence
1 – 50 mg/l
Monthly
0 – 50 mg/l
Regression coefficient
(R2)
0,9954 (c = -0,0322) 0,9988
Precision 9,25 % 0,99 %
LOQ 0,97 mg/l 0,14 mg/l
3rd Meeting of the Heads of the
Laboratories, 16-17 June 2011, Arcachon,
France