Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems
Matthew L. Paden, MDDivision of Pediatric Critical Care
Emory UniversityChildren’s Healthcare of Atlanta at Egleston
![Page 2: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/2.jpg)
2
The Clinical Problem
Multiple ECMO/CVVH patients who are 10+ liters positive on paper, yet are clinically dehydrated
Hypothesis : Both IV pump and Braun Diapact driven CVVH systems when used with ECMO have error rates greater than 5% / hour.
![Page 3: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/3.jpg)
3
Methods for CRRT on ECMO
Commercially available CRRT machines• B Braun Diapact• Edwards Aquarius• Fresenius• Gambro
Prisma/Prismaflex
Advantage – Standardized equipment, “built for the purpose”
![Page 4: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/4.jpg)
4
Methods for CRRT on ECMO
IV Pump Driven• Described by Weber, et al 1998• IV pumps used to create ultrafiltrate and
deliver replacement fluid• Ultrafiltrate is measured using a
urometer Advantage – low cost, no additional training
for ECMO specialists, less extracorporeal blood volume
![Page 5: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/5.jpg)
5
![Page 6: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/6.jpg)
6
![Page 7: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/7.jpg)
7
The Problem
Inaccuracy has been described in CRRT in patients not on ECMO
The pumps are the problem• PCRRT 2000 –
7% extra ultrafiltrate removed 2% less replacement fluid delivered
• ASAIO 1992 – Up to 12.5% error rate Error correlates with pressure drop
across membrane and pump type
![Page 8: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/8.jpg)
8
Alaris Pump Accuracy
Sucosky et al. Awaiting publication. 2008
![Page 9: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/9.jpg)
9
Methods
Two identical saline primed ECMO circuits• Stockert S3, ½ inch drain, 3/8 inch return• 4.5 m2 Medtronic oxygenator• PAN 6 hemofilter
CVVH prescribed : IV pump vs. Diapact• Net even fluid balance• Varying ultrafiltration rates (0.5 – 2 L /
hour) Hourly weights of UF/Replacement bags,
circuit pressures
![Page 10: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/10.jpg)
10
Results
48 hourly measurements• 26 hours Alaris• 22 hours Diapact
No correlation between error rate and• Prescribed ultrafiltration rate• ECMO flow rate• CVVH blood flow rate
![Page 11: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/11.jpg)
11
Results - Alaris
Created a median 0.8% less ultrafiltrate per hour than prescribed (+7% to -12%)
Delivered a median of 4.3% less replacement fluid per hour than prescribed (+3% to -25%)
NET 3.5% of prescribed UF per hour dehydration to patient
![Page 12: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/12.jpg)
12
Examples with Alaris
4 kg neonate on ECMO with 100 ml/hour prescribed UF rate and even fluid balance• 84 ml (21 ml/kg) fluid negative per day
10 kg child on ECMO with 300 ml/hour prescribed UF rate and even fluid balance• 252 ml (25 ml/kg) fluid negative per day
45 kg child on ECMO with 2000 ml/hour prescribed UF rate and even fluid balance• 1.68 L (37 ml/kg) fluid negative per day
![Page 13: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/13.jpg)
13
Results - Diapact
Created a median 1% more ultrafiltrate per hour than prescribed (+6% to -8%)
Delivered a median of 1% more replacement fluid per hour than prescribed (+10% to -7%)
![Page 14: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/14.jpg)
14
Examples Braun
Accurate when looking at medians over time
Hourly variation could be important in a hemodynamically unstable patient
Hourly Fluid Balance
-5
35
10
10
35
-40
-5
0
-10
-30
![Page 15: Accuracy of CVVH on ECMO : A comparison of IV pump versus Braun Diapact driven systems](https://reader036.vdocument.in/reader036/viewer/2022062501/56816047550346895dcf6bc6/html5/thumbnails/15.jpg)
15
Conclusions
In this in vitro CVVH/ECMO model• Both the Alaris and Diapact methods had
error rates that could be clinically significant
• Careful physical assessment of the patient’s volume status is necessary
Further work is needed to develop more a accurate fluid management system for CRRT on ECMO