Transcript
Page 1: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

Asourproductsarecontinuouslyimproved,wereservetheright tomakeanychangeinthestatedspecificationsanddimensionswithoutpriornotice.US2.1Shuttle0708

2.1 Level Transmitter / Controller / Indicator

3-in-1 package that is simple to install and use.

With expanded ranges and improved chemical resistance.

Class 1, Division 1 Hazardous Areas

Approved SensorsNEW

Page 2: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

2

General

Features Shuttle®isdesignedtomeasurewaterlevelsaswellasotherfluidsandaggressivemedia,wheretheultrasonicsensorisnotincontactwiththemedia.• Ultrasonicmeasuringsystemwith3

sensorsmeasuringrangesfromfrom0-4in.to80ft.

• Learning-functionallowstheShuttle®LevelTransmittertolearnhowthetankisarrangedwithpiping,pumpsandotheritemsthatmightinterferewiththemeasu-rement.Thesystemalsoignoresturbu-lenceandelectricalinterference,soevenaslowmovingmixercanpassthroughtheultrasonicbeamwithoutanyinfluenceonthemeasurement.

• 2outputrelaysforlevel/systemalarmoralternatingpumpcontrolwithdutyand

standbycycles.• 4-20mAoutputwithzeroandspan

adjustment.• Indicationofthestrengthoftheecho

signalwithabargraph.• Verynarrowultrasonicpulsesignal

concentratessignalstrengthforminimumsensitivitytofoamandsludgeonthesurfaceoftheliquid.

• EasytoreadLCD-displayforbothsetupandindication.Actuallevelisindicatedwithbothabargraphandwithdigitsinmillimeters,centimetres,metres,feetandinches.Thedigitalreadoutcanbechosenaseitherrelativelevelorabsolutelevel.

• 10-30VDCor115VACsupply.• FMapprovedsensorsforhazardousareas.

Shuttle®isanadvancedmicro-processorbasedultrasoniclevelmeterwithanintel-ligentandveryprecisemeasuringsystemutilizingthenewesthardwareandsoftwaretechnologies.Shuttle®providessimpleandlogicalopera-tion,usingadvancedtechnologyresultinginahighqualityproductatacompetitiveprice.

Shuttle®isfullyautomaticafterinitialsetup.9functionbuttonsbehindthefrontpanelareusedtoconfigure:• Unitselectionineitherfeet,inchesme-

tres,centimetresandmillimetres.• Adjustmentofzerolevelofthemeasure-

mentandselectionofrelativeorabsolutereference.

• ZeroandspanforthemA-output.• Functionforthetwo2outputrelays-off,

levelalarm,systemalarmorpumpcontrolwithalternation.

• Learningfunction,sothelevelmeterrecognizeshowthewetwellisarrangedwithpumps,pipingandotherelementssothemeasurementwillnotbeaffectedbyfalseechos.

Operation

Shuttle®isequippedwith2outputrelayswhichcanbeconfiguredforlevelorpumpcontrol,systemalarmorlevelalarm.TherelayfunctioncanbesettoeitherNOorNC

Control functions functionwithanoptionaldelayfunction.Therelayscanbeusedforstartandstopofpumpswithalternation.

Shuttle® Level Transmitter / Controller / Indicator

Page 3: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

3

Application Shuttle®iswellsuitedforanyultrasoniclevelmeasurement.Theleveltransmittermayalsobeusedformeasuringandcontroloflevelsintankswithcontaminatedand/oraggressivemedia.Shuttle®has4-20mAoutputincreasingordecreasingwiththelevel,wherebothzerolevelandspancanbeadjustedforthewholerangeofthesensor.

Specifications

Electrical connections

Ultrasonic Sensor

Shuttle® Level Transmitter / Controller / Indicator

Typicalapplicationsaremeasurementandcontrolin:- stormflowtanks- pumpstations- tanksinsewagetreatmentplants- tanksinwaterplants- tanksandvesselsinprocessand

foodindustries

Shuttle® Level Transmitter

Shuttle® Level Transmitter

Range: 0-35ft,0-45ft,0-75ft

Span: From0-4in.to0-75ft

Powersupply: 115VAC±10%10-30VDC

Consumption: Approx.2W

Temperature: -5to+150°F

Input: Fromultrasonicsensor

Accuracy: ±0.2%

Outputs: Analog:4-20mA,max.load500Ω.Digital:2xrelayswithprogrammableNO/NCfunction.Max.1Aresistiveload@50V.

Indication: LCDdisplaywith4digitsandsymbols.

Operation: Functionkeypadbehindfrontpanel.

Enclosure: NEMA4X

Approvals: CE:EN50081-1,EN50082-1UL:FileE194021

Page 4: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

4

Shuttle®Ultrasonic sensors

Shuttle® Ultrasonic Sensor 200570 / 200771

Range: 45ft.(fluids)15ft.(solids)

Frequency: 30kHz

Spread: 3°

Deadband: 14"

Temperature: -5to+150°F

Materials: PP,POM

Cable: Shielded,oilresistantPVCinsulation,length39ft.(Max.300ft.with690010cable.)

Enclosure: NEMA6P,watertight,withstandsmax.30ftsubmersion.

CE-approvals: EN50081-1,EN50082-1

Standard range version

Shuttle® Ultrasonic Sensor: 200640 / 200641

Range: 35ft.(fluids),15ft.(solids)

Frequency: 40kHz

Spread: 7°

Deadband: 14"

Temperature: -5to+150°F

Materials: PTB,GlassreinforcedEpoxy

Cable200640:Cable200641:

Shielded,oilresistantPVCinsulation,length39ft.Shielded,oilresistantPVCinsulation,length164ft.(Max.300ft.with690010cable.)

Enclosure: NEMA6P,watertight,withstandsmax.30ftsubmersion.

CE-approvals: EN50081-1,EN50082-1

Ex-approvals: FMClass1,Div.1,GroupsA-G

Standard rangeversion

Extended range version

Shuttle® Ultrasonic Sensor: 200630 / 200631

Range: 80ft.(fluids),30ft.(solids)

Frequency: 30kHz

Spread: 6°

Deadband: 32"

Temperature: -5to+150°F

Materials: PTB,GlassreinforcedEpoxy

Cable200630:Cable200631:

Shielded,oilresistantPVCinsulation,length39ft.Shielded,oilresistantPVCinsulation,length164ft.(Max.300ft.with690010cable.)

Enclosure: NEMA6P,watertight,withstandsmax.30ftsubmersion.

CE-approvals: EN50081-1,EN50082-1

Ex-approvals: FMClass1,Div.1,GroupsA-G

Shuttle® Ultrasonic Sensor: 200660

Range: 30ft.(fluids),15ft.(solids)

Frequency: 50kHz

Spread: 6°

Deadband: 14"

Temperature: -5to+150°F

Materials: PP,PVDF

Cable: Shielded,oilresistantPVCinsulation,length39ft.(Max.300ft.with690010cable.)

Enclosure: NEMA6P,watertight,withstandsmax.30ft.submersion.

CE-approvals: EN50081-1,EN50082-1

Chemical resistent version

Page 5: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

5

Shuttle®Ultrasonic sensors

9090

6060

3030

0

-40

-30

-20

-10dB

Ultrasonic sensor type 200570 / 200771

Frequency: 30KHz

Range: Upto45ft.influidsand15ft.insolids.

Directivity: Seefigurebelow.

Standard range version

9090

6060

3030

0

-40

-30

-20

-10dB

Ultrasonic sensor type 200640 / 200641

Frequency: 40KHz

Range: Upto35ft.influidsand15ft.insolids.

Directivity: Seefigurebelow.

Standard range version

9090

6060

3030

0

-40

-30

-20

-10dB

Ultrasonic sensor type 200630 / 200631

Frequency: 30KHz

Range: Upto80ft.influidsand30ft.insolids.

Directivity: Seefigurebelow.

Extended range version

Ultrasonic sensor type 200660

Frequency: 50KHz

Range: Upto30ft.influidsand15ft.insolids.

Directivity: Seefigurebelow.

Chemical resistent version

Page 6: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet Shuttle® Level Transmitter / Controller / Indicator

Dimensions

Universal bracket

Sensor bracket, standardSensor bracket, short

Shuttle® Level Transmitter

Ultrasonic sensor200570 / 200660 /

200571

Ultrasonic sensor200640 / 200630 / 200641 / 200631

Order numbers Shuttle® Level Transmitter complete with Ultrasonic Sensor

201209 Shuttle®LevelTransmitterwithoutsensor,powersupply115/230VACor10-30VDC

201214 Shuttle®LevelTransmittercompletewithUltrasonicSensor200570and39ft.ofcable,powersupply115/230VACor10-30VDC

201224 Shuttle®LevelTransmittercompletewithUltrasonicSensor200630and39ft.ofcable,powersupply115/230VACor10-30VDC

201234 Shuttle®LevelTransmittercompletewithUltrasonicSensor200640and39ft.ofcable,powersupply115/230VACor10-30VDC

201244 Shuttle®LevelTransmitter,completewithShuttle®UltrasonicSensor200660and39ft.ofcable,powersupply115/230VACor10-30VDC

Accessories

200205 Universalbracket

200219 ShortSensorbracketforUltrasonicSensor

200220 StandardSensorbracketforUltrasonicSensor

200590 ConnectionboxforUltrasonicSensorcable

690010 AdditionalcableforUltrasonicSensor.Canbeextendedto300ft.,specifylength

200570 UltrasonicSensor,30kHz,39ft.ofcable,range:45ft.influids,15ft.insolids

200571 UltrasonicSensor,30kHz,150ft.ofcable,range:45ft.influids,15ft.insolids

200630 UltrasonicSensor,30kHz,39ft.ofcable,range:75ft.influids,30ft.insolids

200631 UltrasonicSensor,30kHz,150ft.ofcable,range:75ft.influids,30ft.insolids

200640 UltrasonicSensor,40kHz,39ft.ofcable,range:35ft.influids,15ft.insolids

200641 UltrasonicSensor,40kHz,150ft.ofcable,range:35ft.influids,15ft.insolids

200660 Ultrasonicsensor,50kHz,39ft.ofcable,range:30ft.influids,15ft.insolids

840010 Shuttle®CommunicationSoftwareforsettingsincl.InterfaceandRS232cable

840011 Shuttle®CommunicationSoftwareforsettingsandFirmware,incl.InterfaceandRS232cable


Top Related