level transmitter / controller / indicator 2...shuttle® level transmitter shuttle® level...

6
Data Sheet As our products are continuously improved, we reserve the right to make any change in the stated specifications and dimensions without prior notice. US 2.1 Shuttle 0708 2.1 Level Transmitter / Controller / Indicator 3-in-1 package that is simple to install and use. With expanded ranges and improved chemical resistance. Class 1, Division 1 Hazardous Areas Approved Sensors NEW

Upload: others

Post on 04-Aug-2020

15 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

Asourproductsarecontinuouslyimproved,wereservetheright tomakeanychangeinthestatedspecificationsanddimensionswithoutpriornotice.US2.1Shuttle0708

2.1 Level Transmitter / Controller / Indicator

3-in-1 package that is simple to install and use.

With expanded ranges and improved chemical resistance.

Class 1, Division 1 Hazardous Areas

Approved SensorsNEW

Page 2: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

2

General

Features Shuttle®isdesignedtomeasurewaterlevelsaswellasotherfluidsandaggressivemedia,wheretheultrasonicsensorisnotincontactwiththemedia.• Ultrasonicmeasuringsystemwith3

sensorsmeasuringrangesfromfrom0-4in.to80ft.

• Learning-functionallowstheShuttle®LevelTransmittertolearnhowthetankisarrangedwithpiping,pumpsandotheritemsthatmightinterferewiththemeasu-rement.Thesystemalsoignoresturbu-lenceandelectricalinterference,soevenaslowmovingmixercanpassthroughtheultrasonicbeamwithoutanyinfluenceonthemeasurement.

• 2outputrelaysforlevel/systemalarmoralternatingpumpcontrolwithdutyand

standbycycles.• 4-20mAoutputwithzeroandspan

adjustment.• Indicationofthestrengthoftheecho

signalwithabargraph.• Verynarrowultrasonicpulsesignal

concentratessignalstrengthforminimumsensitivitytofoamandsludgeonthesurfaceoftheliquid.

• EasytoreadLCD-displayforbothsetupandindication.Actuallevelisindicatedwithbothabargraphandwithdigitsinmillimeters,centimetres,metres,feetandinches.Thedigitalreadoutcanbechosenaseitherrelativelevelorabsolutelevel.

• 10-30VDCor115VACsupply.• FMapprovedsensorsforhazardousareas.

Shuttle®isanadvancedmicro-processorbasedultrasoniclevelmeterwithanintel-ligentandveryprecisemeasuringsystemutilizingthenewesthardwareandsoftwaretechnologies.Shuttle®providessimpleandlogicalopera-tion,usingadvancedtechnologyresultinginahighqualityproductatacompetitiveprice.

Shuttle®isfullyautomaticafterinitialsetup.9functionbuttonsbehindthefrontpanelareusedtoconfigure:• Unitselectionineitherfeet,inchesme-

tres,centimetresandmillimetres.• Adjustmentofzerolevelofthemeasure-

mentandselectionofrelativeorabsolutereference.

• ZeroandspanforthemA-output.• Functionforthetwo2outputrelays-off,

levelalarm,systemalarmorpumpcontrolwithalternation.

• Learningfunction,sothelevelmeterrecognizeshowthewetwellisarrangedwithpumps,pipingandotherelementssothemeasurementwillnotbeaffectedbyfalseechos.

Operation

Shuttle®isequippedwith2outputrelayswhichcanbeconfiguredforlevelorpumpcontrol,systemalarmorlevelalarm.TherelayfunctioncanbesettoeitherNOorNC

Control functions functionwithanoptionaldelayfunction.Therelayscanbeusedforstartandstopofpumpswithalternation.

Shuttle® Level Transmitter / Controller / Indicator

Page 3: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

3

Application Shuttle®iswellsuitedforanyultrasoniclevelmeasurement.Theleveltransmittermayalsobeusedformeasuringandcontroloflevelsintankswithcontaminatedand/oraggressivemedia.Shuttle®has4-20mAoutputincreasingordecreasingwiththelevel,wherebothzerolevelandspancanbeadjustedforthewholerangeofthesensor.

Specifications

Electrical connections

Ultrasonic Sensor

Shuttle® Level Transmitter / Controller / Indicator

Typicalapplicationsaremeasurementandcontrolin:- stormflowtanks- pumpstations- tanksinsewagetreatmentplants- tanksinwaterplants- tanksandvesselsinprocessand

foodindustries

Shuttle® Level Transmitter

Shuttle® Level Transmitter

Range: 0-35ft,0-45ft,0-75ft

Span: From0-4in.to0-75ft

Powersupply: 115VAC±10%10-30VDC

Consumption: Approx.2W

Temperature: -5to+150°F

Input: Fromultrasonicsensor

Accuracy: ±0.2%

Outputs: Analog:4-20mA,max.load500Ω.Digital:2xrelayswithprogrammableNO/NCfunction.Max.1Aresistiveload@50V.

Indication: LCDdisplaywith4digitsandsymbols.

Operation: Functionkeypadbehindfrontpanel.

Enclosure: NEMA4X

Approvals: CE:EN50081-1,EN50082-1UL:FileE194021

Page 4: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

4

Shuttle®Ultrasonic sensors

Shuttle® Ultrasonic Sensor 200570 / 200771

Range: 45ft.(fluids)15ft.(solids)

Frequency: 30kHz

Spread: 3°

Deadband: 14"

Temperature: -5to+150°F

Materials: PP,POM

Cable: Shielded,oilresistantPVCinsulation,length39ft.(Max.300ft.with690010cable.)

Enclosure: NEMA6P,watertight,withstandsmax.30ftsubmersion.

CE-approvals: EN50081-1,EN50082-1

Standard range version

Shuttle® Ultrasonic Sensor: 200640 / 200641

Range: 35ft.(fluids),15ft.(solids)

Frequency: 40kHz

Spread: 7°

Deadband: 14"

Temperature: -5to+150°F

Materials: PTB,GlassreinforcedEpoxy

Cable200640:Cable200641:

Shielded,oilresistantPVCinsulation,length39ft.Shielded,oilresistantPVCinsulation,length164ft.(Max.300ft.with690010cable.)

Enclosure: NEMA6P,watertight,withstandsmax.30ftsubmersion.

CE-approvals: EN50081-1,EN50082-1

Ex-approvals: FMClass1,Div.1,GroupsA-G

Standard rangeversion

Extended range version

Shuttle® Ultrasonic Sensor: 200630 / 200631

Range: 80ft.(fluids),30ft.(solids)

Frequency: 30kHz

Spread: 6°

Deadband: 32"

Temperature: -5to+150°F

Materials: PTB,GlassreinforcedEpoxy

Cable200630:Cable200631:

Shielded,oilresistantPVCinsulation,length39ft.Shielded,oilresistantPVCinsulation,length164ft.(Max.300ft.with690010cable.)

Enclosure: NEMA6P,watertight,withstandsmax.30ftsubmersion.

CE-approvals: EN50081-1,EN50082-1

Ex-approvals: FMClass1,Div.1,GroupsA-G

Shuttle® Ultrasonic Sensor: 200660

Range: 30ft.(fluids),15ft.(solids)

Frequency: 50kHz

Spread: 6°

Deadband: 14"

Temperature: -5to+150°F

Materials: PP,PVDF

Cable: Shielded,oilresistantPVCinsulation,length39ft.(Max.300ft.with690010cable.)

Enclosure: NEMA6P,watertight,withstandsmax.30ft.submersion.

CE-approvals: EN50081-1,EN50082-1

Chemical resistent version

Page 5: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet

5

Shuttle®Ultrasonic sensors

9090

6060

3030

0

-40

-30

-20

-10dB

Ultrasonic sensor type 200570 / 200771

Frequency: 30KHz

Range: Upto45ft.influidsand15ft.insolids.

Directivity: Seefigurebelow.

Standard range version

9090

6060

3030

0

-40

-30

-20

-10dB

Ultrasonic sensor type 200640 / 200641

Frequency: 40KHz

Range: Upto35ft.influidsand15ft.insolids.

Directivity: Seefigurebelow.

Standard range version

9090

6060

3030

0

-40

-30

-20

-10dB

Ultrasonic sensor type 200630 / 200631

Frequency: 30KHz

Range: Upto80ft.influidsand30ft.insolids.

Directivity: Seefigurebelow.

Extended range version

Ultrasonic sensor type 200660

Frequency: 50KHz

Range: Upto30ft.influidsand15ft.insolids.

Directivity: Seefigurebelow.

Chemical resistent version

Page 6: Level Transmitter / Controller / Indicator 2...Shuttle® Level Transmitter Shuttle® Level Transmitter Range: 0 - 35 ft, 0 - 45 ft, 0 - 75 ft Span: From 0 - 4 in. to 0 - 75 ft Power

Data Sheet Shuttle® Level Transmitter / Controller / Indicator

Dimensions

Universal bracket

Sensor bracket, standardSensor bracket, short

Shuttle® Level Transmitter

Ultrasonic sensor200570 / 200660 /

200571

Ultrasonic sensor200640 / 200630 / 200641 / 200631

Order numbers Shuttle® Level Transmitter complete with Ultrasonic Sensor

201209 Shuttle®LevelTransmitterwithoutsensor,powersupply115/230VACor10-30VDC

201214 Shuttle®LevelTransmittercompletewithUltrasonicSensor200570and39ft.ofcable,powersupply115/230VACor10-30VDC

201224 Shuttle®LevelTransmittercompletewithUltrasonicSensor200630and39ft.ofcable,powersupply115/230VACor10-30VDC

201234 Shuttle®LevelTransmittercompletewithUltrasonicSensor200640and39ft.ofcable,powersupply115/230VACor10-30VDC

201244 Shuttle®LevelTransmitter,completewithShuttle®UltrasonicSensor200660and39ft.ofcable,powersupply115/230VACor10-30VDC

Accessories

200205 Universalbracket

200219 ShortSensorbracketforUltrasonicSensor

200220 StandardSensorbracketforUltrasonicSensor

200590 ConnectionboxforUltrasonicSensorcable

690010 AdditionalcableforUltrasonicSensor.Canbeextendedto300ft.,specifylength

200570 UltrasonicSensor,30kHz,39ft.ofcable,range:45ft.influids,15ft.insolids

200571 UltrasonicSensor,30kHz,150ft.ofcable,range:45ft.influids,15ft.insolids

200630 UltrasonicSensor,30kHz,39ft.ofcable,range:75ft.influids,30ft.insolids

200631 UltrasonicSensor,30kHz,150ft.ofcable,range:75ft.influids,30ft.insolids

200640 UltrasonicSensor,40kHz,39ft.ofcable,range:35ft.influids,15ft.insolids

200641 UltrasonicSensor,40kHz,150ft.ofcable,range:35ft.influids,15ft.insolids

200660 Ultrasonicsensor,50kHz,39ft.ofcable,range:30ft.influids,15ft.insolids

840010 Shuttle®CommunicationSoftwareforsettingsincl.InterfaceandRS232cable

840011 Shuttle®CommunicationSoftwareforsettingsandFirmware,incl.InterfaceandRS232cable