drager patient monitor pm8060 - user manual

104
PM 8060 vitara Haemodynamic Patient Monitor Instructions for Use Software version 2.n 2-893-96 D Dräger Medical "Only for internal use by Draeger. Not applicable as Instructions for Use!"

Upload: wasanhibrahim

Post on 19-Dec-2015

189 views

Category:

Documents


20 download

DESCRIPTION

monitor

TRANSCRIPT

  • PM 8060 vitaraHaemodynamic Patient Monitor

    Instructions for Use

    Software version 2.n

    2-89

    3-96

    DDrger Medical"O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Working with these Instructions for Use

    In the top header line the subject...of the main chapter. The second line contains the title of the sub-chapter foreasy reference and finding your way quickly about theInstructions for Use.

    On the page body ...the Instructions for Use in combined text/graphical format. The information isexplained directly in terms of actions that teach the userhow to operate the machine by hands-on experience.

    Left-hand column the text ...contains explanations and instructs the user in theoperation of the product, with brief, unconfusinginstructions in ergonomic sequence.The bullets indicate separate actions, and, if severalactions are described for the same accompanyingillustration, the numbers provide clear cross-referencingbetween text and graphics and identify the sequence ofoperations.

    Right-hand column the illustrations ...provide the visual reference to the text and help locatethe various parts of the equipment. Parts mentioned inthe text are highlighted, and inessential details have beendeliberately omitted.Screen displays also guide the user and confirm correctperformance of the required operations.

    Working with these Instructions for Use

    2

    Xxxxxxxxx

    l Xxxxxxxxxxxxxxxxxxxxxxxxxxxx:

    Xxxxxxxxxx: Vxxxxxxxxx?

    Xxxxxxxxxxxxx: Hxxxxxxx ?, Fxxxxxxx ?, Dxxxxxx ?

    Xxxxxxxxx: Sxxxxxxxxxxxxx ?

    Xxxxxxxxxx Xxxxxxx xxxxxxxxxxxxxxx

    1 Xxxxxxxxxxx Xxxxxxxxxx xxxxx -

    l Xxxxxxxxxxxxx XXXXXXXXXX cccxxxxxxxxxxxxx xxxxx x

    2 Xxxxxxxxxxxxxxxxxx xxxxxxxxx (X xxxxxxxxx).

    Xxxxxxxxxxxxxxxxxxx

    l Xxxxxxxxxxxxxxx xxx Xxxxxxxxxxx xxxxxxxxx xxxxxxxx -

    l Xxxxxxxxxx - xxxxxx xxxxxxxxxxxx ! - xxx xxx Xxxxxxxxxx

    xxxxxxxxxxxx xxxxx !

    l Xxxxxxxxxx xxxxxxxx xxxxxx xxx xxxxxxxxx Xxxxxxxxxx xxxxxx

    xxx xxx Xxxxxxxxxxxxx:

    xxxx.

    (xxxxxxxx X xxxxxxxx Xxxxxxxxxxx)

    l Xxxxxx xxxxxxxxx xxxx xxxxxx Xxxxxxx xxxx Xxxxxxxxx.

    l Xxxxxx xxx Xxxxxxx xxx Xxxxxxxx xxxxxxxxxx.

    O2-Xxxxxxx

    l Xxxxxxxxxxxxx xxx Xxxxxx xxxxxx xxxxxcxxx !

    Xxxxxxxx xxx xxx Xxxxxxxxxxxxxxx xxx Xxxxxxxx xxxxxxxx -

    l xxxxxxxxxxxxxxxx xxxxxxxxxxxxx xxxxxxxxxxxx

    Xxxxxxxxx xxxxxxxxx

    l Xxxxxxxxxxxx xxxxxxxx

    3 Xxxx xx Xxxxx xxx xx Xxxxx xxxxxxxxxx xcxxxxxxxx - xxx xxx

    Xxxxxxxcxx -

    xxxxxx Xxxxxx Xxxxxxxxxxxxxxx xxxxx xxxxxxxxxxxxxx !

    l Xxxxxxxxxxxxx xx Xxxxxxxx xxxxx xxx xxxxxxxxxxxxx

    4 Xxxxxxxx xxx xxx Xxxxxxxxxxx xcxx xxxxxxx xxxxxxxxx

    (xxxxxxxxxxxx) -

    Xxxxxxx Yxxxxxxxxx

    1. EKG

    1. EKG (Modul doppelter Hhe)

    XYZ

    10:3400:00

    01-10-96

    standard screen - adultAdvisory:Caution:Warning:

    65pulse def. 0 ST + 0.05

    122/ 89(100)

    ARTiBP 1

    4.5mean

    CVP

    99pulse 65

    pulse 65

    15min M (102)

    SysmmHgDia

    SpO2

    122/ 89NiBP

    36.8 37.2T1 T2 C

    iBP 2

    auto setlimits

    alarminfo

    screenconfig.

    para-meter

    timerstart

    limits

    config.

    1. ECGII

    1mV

    Pleth

    ART

    CVP

    130

    10

    80

    0

    HR/Pulse (ECG)

    Self-test

    Technology for life

    PM 8060 VitaraVersion 2.n 02-12-97

    Modulseite (deutsch):36.8 37.2

    T1 T2 C

    15min M (102)

    SysmmHgDia

    122/ 89NiBP

    Sys

    man 10:34 M (102)

    man 10:35 M (11.1)

    mmHgDia

    122/ 89NiBP

    117 M (102)

    SysmmHgDia

    122/ 89NiBP

    SyskPaDia

    15.2/ 9.3NiBP

    99Puls 65

    SpO2 %

    (EKG)

    1/min

    mmHg mmHg

    122/ 89(100)

    ARTiBP 1

    Puls 65

    122/ 89(100)

    AORTAiBP 1

    Puls 65

    mmHg

    122/ 89(100)

    A.PulmiBP 1

    Puls 65

    mmHg

    122/ 89(100)

    iBP ?iBP 1

    Puls 65

    65Pulsdef. 0 ST + 0.05

    HR/Puls

    21RR

    mmHg mmHg

    4.5mean

    ZVDiBP 2 iBP 2

    4.5mean

    ICP

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Contents

    3

    Contents

    For your safety and that of your patients 4

    Intended use 5

    Operating concept 7

    Preparation for use 13

    Monitor functions 19

    Parameter Box 49

    Messages Cause Remedy 73

    Care 79

    What's what 83

    Technical data 89

    Abbreviations and symbols 98

    Index 101

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • For Your Safety and that of YourPatients

    Strictly follow the Instructions for Use

    Any use of the apparatus requires full understanding andstrict observation of these instructions.The apparatus is only to be used for purposes specifiedhere.

    Maintenance

    The apparatus must be inspected and serviced regularlyby trained service personnel at six monthly intervals. Repair and general overhaul of the apparatus may only becarried out by trained service personnel.We recommend that a service contract be obtained withDrgerService and that all repairs also be carried out bythem.Only authentic Drger spare parts may be used for main-tenance. Observe chapter "Maintenance Intervals".

    Accessories

    Do not use accessory parts other than those in the orderlist.

    Not for use in areas of explosion hazard

    This apparatus is neither approved nor certified for use inareas where combustible or explosive gas mixtures arelikely to occur.

    Safe connection with other electrical equipment

    Electrical connections to equipment which is not listed inthese Instructions for Use should only be made followingconsultations with the respective manufacturers or anexpert.

    Liability for proper function or damage

    The liability for the proper function of the apparatus isirrevocably transferred to the owner or operator to theextent that the apparatus is serviced or repaired bypersonnel not employed or authorized by DrgerServiceor if the apparatus is used in a manner not conforming toits intended use.Drger cannot be held responsible for damage causedby non-compliance with the recommendations givenabove.The warranty and liability provisions of the terms of saleand delivery of Drger are likewise not modified by therecommendations given above.

    Drger Medical AG & Co. KGaA

    For your safety and that of your patients

    4

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Contents

    Intended use..............................................................................................6

    5

    Intended useContents

    Intended use

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Intended use

    Haemodynamic Patient Monitor PM 8060 vitara

    When connected to the parameter box, the essentialhaemodynamic parameters are displayed together withtheir monitoring features:

    The ECG curve with heart rate and ST segmentanalysis.

    Two channels for invasively measured blood pressure(iBP), with the pressure values for systolic, diastolicand mean pressure and the pressure curve.

    The measured values for the non-invasively measuredblood pressure (NiBP) with the pressure values forsystolic, diastolic and mean pressure.

    Two channels for body temperature.

    The functional O2 saturation (SpO2) with the pulserate.

    The plethysmogram.

    The respirogram.

    When connected to an anaesthetic workstation, datafrom an airway monitor can also be displayed, but inthis case airway alarms are not transferred to thePM 8060 vitara.

    Precautions for useFire risk. Use only non-flammable anaesthetic agentsconforming to EN 740.

    Any add-on electrical equipment must be earthed bymeans of an equipotential bonding conductorconnected to the base unit.

    Electromagnetic fields that exceed the value specifiedin EN60601-1-2 may impair the operation of theapparatus (for example during nuclear spin tomography -MRT, NMR and NMI).

    Do not use mobile phones within a distance of 10 metres from the machine.Mobile phones can cause interference to electrical andelectronic medical appliances, thereby putting patients atrisk.*

    Use the workstation under the supervision of qualifiedmedical personnel, so that assistance can be providedimmediately in the event of any malfunction!

    * Drger medical appliances comply with the interference immunityrequirements of the specific standards for the products or EN 60601-1-2 (IEC 601-1-2). However, depending on the designof the mobile phone and situation of use, field strengths may occurin the immediate environment of a mobile phone that exceed thevalues of the standards quoted and therefore cause interference.

    Intended use

    6

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Contents

    User interface........................................................................................... 8Keys with dedicated function......................................................................8Keys with variable function......................................................................... 8Rotary control.............................................................................................9Indicator lamps.........................................................................................10Structure of the screen display................................................................. 10Menu structure......................................................................................... 10Screen saver............................................................................................11Colour concept........................................................................................ 11The various screen displays......................................................................11

    Operating conceptContents

    7

    Operating concept

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • User interface

    Keys with dedicated function (Hardkeys)

    The right-hand side is reserved for operating elements,the left-hand side for displays.

    G This key is used to deactivate the alarm tone fortwo minutes. The alarm is reactivated by pressingthe key again. The yellow LED in the key lights upwhile alarms are suspended.

    Any new messages are signalled once by theappropriate tone sequence.

    E This key switches the monitor from standby tomeasuring mode and vice versa.

    The dark area contains four keys that act directly on thescreen contents:

    S This key is used to select the desired screen(standard screen, trend screen and list screen).

    Q This key is always used to call up the last standardscreen used.

    X This key is used to freeze the haemodynamiccurves on the screen so that they can be viewed inmore detail.

    U This key is used to generate a manual entry in thelist screen.

    Keys with variable function (Softkeys)An unlabelled membrane keypad is positioned next to theright-hand edge of the screen.

    The function of each of these keys at any one time issoftware-driven and is displayed in the correspondingposition on the screen.

    The screen only displays the keys that can be activated atthat moment.

    Operating conceptUser interface

    8

    D

    1-2--

    Stop

    D

    1-2--

    Stop

    0022

    9094

    0032

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Rotary controlA single "turn-and-push" control is used to select andconfirm settings.

    Turn to select:Turning the rotary control moves a cursor frame acrossthe screen or changes a numeric value.

    Press to set: When the rotary control is pressed, the value selected byturning the control is accepted as the valid parameter, ora process is started or stopped.

    Example: To adjust the volume of the audible alarm

    1 Call up the configuration menu from standby modeby pressing the config softkey.

    A menu of default settings is displayed.

    l Select = turn the rotary control.

    Move the cursor frame to the acoustic field.

    l Confirm = press the rotary control.

    2 The selected option acoustic is highlighted ininverse video (dark type on light background see colour concept).

    The cursor frame is positioned over the arrowsymboll ( z ) in the open acoustic field.

    3 This menu is used to:

    set the volume of the alarm and of the pulse tone,

    select between a plain pulse tone and a pulse toneof variable frequency modulated by the degree ofoxygen saturation (SpO2).

    select the tone system (Drger or ISO standard).

    l Turn the rotary control = move the cursor frame intothed alarm sound field.

    l Confirm = press the rotary control.

    The confirmed selection alarm sound is high-lighted in inverse video (dark on light). The tone isplayed back at the selected volume.

    The dashed selection window now contains volumelevels from 1 (soft) to 9 (loud).

    The original value (here: 5) is set against anorange background.

    l To set a new value = turn the rotary control.

    Move the cursor to the desired volume. The tone isplayed back at the selected volume.

    l Confirm = press the rotary control.

    The new value is displayed against an orangebackground. The cursor-frame returns to the arrowsymbol ( z ).

    l Press the rotary control again (if necessary severaltimes). The cursor moves back one menu level at atime, in each case to the arrow symbol ( z ), untilyou exit the configuration menu.

    9

    Operating conceptUser interface

    Standby/Configuration

    10:3400:00

    01-10-96Alarm sound and pulse tone configuration menu. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    alarm sound 1 2 3 4 5 6 7 8 9

    pulse tone 0 1 2 3 4 5 6 7 8 9

    pulse tone ECG Pleth.SpO2 modulated

    tone sequence DRGER Euro-Standard

    alarms inactive !

    deletetrend

    config.

    Switch on with

    Standby

    Data ManagerSW-Version 2.n

    02-12-97

    ParameterboxSW-Version 2.n

    PM 8060 Vitara

    E

    1

    2

    3

    0092

    9094

    0102

    9094

    0112

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Indicator lamps

    Two bar-shaped indicator lamps are located above theG key: these lamps continue to indicate the alarmstates even when the acoustic alarm is suspended:

    Red (upper) lamp, flashing: Warning !!!Yellow (lower) lamp, flashing: Caution !!Yellow lamp, constant: Advisory !

    Structure of the screen display

    The screen is made up of grids in which the individualtraces and blocks of measured values (modules) canbe defined as required.

    The fields for date/time, equipment status, warning,caution or advisory messages, the screen keys (softkeys)and the user advisory field are always located in the sameplace in all screens.

    Menu structure

    The hierarchy and sequence of menu levels is made clear by their staggered arrangement across the screen.The active menu is always green with a yellow cursor.Previous menu levels are grey with a black cursor.

    The fields shown against an orange background indicatethe current configuration. The procedure for opening andclosing menus is always the same.

    It is always possible to return directly to the standardscreen by pressing the Q key.

    Operating concept User interface

    10

    Standby/Configuration alarms inactive !

    10:3400:00

    01-10-96

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    ECG

    NiBP

    iBP locations

    iBP channels

    SpO2 / Pleth

    Resp.

    Select a menu position. Please confirm !Change the settings. Please confirm again !

    ECG

    function on off

    numberof leads 3 5

    ST-segmentanalysis on off

    pulse deficit on offindication

    pacemaker on off

    filter on off

    amplitude 2

    1.ECG2.ECG 3.ECG

    derivation I IIIII

    State Warning / Caution / Advisory

    Modules (Curves / Graphs)

    Date / Time User advisory field

    D

    1-2--

    Stop

    0042

    9094

    0122

    9094

    0132

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Screen saver

    If none of the controls or keys on the Monitor is operatedfor about 5 minutes while in Standby mode, the screenswitches to a screen saver program to prevent screenburn-in. The yellow LED in the Standby key lights up, anda Drger logo ( D) appears at different positions onthe screen.

    The screen display is restored immediately any useroperation is detected.

    Colour concept

    Certain colours are used recurrently with the samestandard meaning when operating the device:

    Screen element: Display colour:

    Active menu level .................... Cyan

    Non-active menu level ............. Grey

    Cursor .................................... Yellow frame

    Selected menu item ................ Inverse (white on black)

    Selected alternative................. Orange background

    Alarm messages are displayed in clearly distinct fieldsaccording to the urgency of the message:

    1 Warning Symbol !!! Text on red field2 Caution Symbol !! Text on yellow field3 Advisory Symbol ! Text on white field

    The colours of the individual measurement parametersand displays and the choice of bright or dark backgroundcan be freely configured by the user (see page 30).

    The various screen displays

    4 The other screen pages are called up by pressing

    S.

    A selection menu is then displayed for the followingalternative screen displays:

    Standard screen 1,

    Standard screen 2,

    List screen and

    Trend screen.

    11

    Operating conceptScreen displays

    D

    Stop

    1-2-- 4

    0062

    9094

    10:3400:00

    01-10-96 The limit indicated has been violated. Please select a new value and confirm !

    standard screen - adultAdvisory:Caution:Warning:TEMP2 CALIB !TEMP1 !!SPO2 !!!

    1. ECGII

    1mV

    ART

    CVP

    Pleth

    65pulse def. 0 ST + 0.05

    HR/Pulse (ECG)

    122/ 89(100)

    ARTiBP 1

    4.5mean

    CVP

    94pulse 65

    pulse 65

    15min M (102)

    SysmmHgDia

    SpO2

    122/ 89NiBP

    36.8 37.2T1 T2 C

    iBP 2

    160

    20

    0

    0

    auto setlimits

    alarminfo

    timerstart

    limitsHR 72 50 120Puls 72 50 120NiBP

    sys -- 80 160SpO2 92 94 --ST 0.00 0.10T1 37.5 -- --T2 32.2 -- --RR -- 5 25iBP1

    sys 122 80 160iBP2

    Mean 4.5 80 160

    0542

    9094

    1 2 3

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Standard screen (1 or 2)containing the configured graph and measured valuefields.

    List screencontaining all measured values and alarms; therebymaking it easier to complete the anaesthesia record. This list can also be printed out to a connected printer(optional).

    Trend screenfor displaying the trend curves of the measured valuesover time.

    12

    11:3400:00

    01-10-96

    list screen - adultAdvisory:Caution:Warning:

    intubation

    start ofsurgery

    positionchange

    extuba-tion

    start ofanaesthesia

    end ofsurgery

    event

    print

    1. ECGII

    1mV

    65pulse def. 0 ST + 0.05

    Manual list entry by

    time

    10:55 interval 2 01-10-9610:55 IND-210:55 11:0011:0511:08 intubation11:1011:1311:1511:20 interval 1 01-10-9611:20 OP-2 11:20 11:25 11:3011:32 start of surgery

    HR/Pulse1/min

    6467

    636263

    6362

    NiBP mmHgSys/M/Dia

    120/100/90123/103/88

    120/100/90123/103/88128/107/93

    120/100/90122/102/89

    SpO2%

    90 97

    96 90 97

    95 97

    CO2Vol. %Fet

    3537

    313433

    3537

    PAW mbar peak/PEEP

    32/3 31/3

    33/3 31/3 30/3

    32/3 31/3

    Warning

    SPO2

    SPO2

    !!!

    !!!

    MVL/min

    6.4 6.3

    6.2 6.1 6.0

    6.4 6.3

    Cpatml/mbar

    1212

    121212

    1212

    O2Vol. %Fi

    2131

    323133

    3031

    anaesth.agentFi

    ENF1.41.4 1.21.21.1

    ISO1.41.4

    tempCT1

    36.836.9

    36.736.836.9

    36.836.9

    config. list entry next page preceding page

    HR/Pulse (ECG)

    10:3400:00

    01-10-96 Select a trend sector and magnify by confirming !

    NiBP SpO2

    HR/RRIBP 1200

    100

    0

    100

    60

    80

    200

    100

    0

    11:00 12:00 13:00

    Sys

    MDia

    Sys

    MDia

    200

    100

    0

    11:00 12:00 13:00

    1. ECG II

    1mV

    fulltrend

    config.

    trend screen - adultAdvisory:Caution:Warning:

    65pulse def. 0 ST + 0.05

    HR/Pulse (ECG)

    0152

    9094

    0162

    094

    10:3400:00

    01-10-96

    standard screen - adultAdvisory:Caution:Warning:

    65pulse def. 0 ST + 0.05

    122/ 89(100)

    ARTiBP 1

    4.5mean

    CVP

    99pulse 65

    pulse 65

    15min M (102)

    SysmmHgDia

    SpO2

    122/ 89NiBP

    36.8 37.2T1 T2 C

    iBP 2

    auto setlimits

    alarminfo

    screenconfig.

    para-meter

    timerstart

    limits

    config.

    1. ECGII

    1mV

    Pleth

    ART

    CVP

    130

    10

    80

    0

    HR/Pulse (ECG)

    0142

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Contents

    Connecting the System ......................................................................... 14Connecting the power supply .................................................................. 14Connecting external devices ....................................................................14

    Switching the monitor on and off .......................................................... 18

    13

    Preparation for useContents

    Preparation for use

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Connecting the System

    Connectingthe power supply:

    When used in combination with an anaestheticmachine Julian:

    1 Connect the CAN port (marked CAN 1) at the backof the screen to the CAN port on the back of theanaesthetic unit (port marked CAN 1).

    When used as stand-alone Monitor or with Drger-Cato:

    1 Connect the CAN port (marked CAN 1) at the backof the screen to the Vitara power adapter.

    l Plug the power cord from the Vitara power adapterinto a 230 V mains socket. With anaesthetic units, thepower is supplied via a socket that is switched on andoff by the system master switch.

    2 CAN 2 is designed for add-on units.

    3 For intracranial and intracardial operations,equipotential bonding must be ensured by connectinga grounding cable from the ground stud on the backof the screen (marked P ) to the ground stud in theoperating theatre!

    Connecting external devices Interface configuration, see page 34.

    Via the RS 232 printer interface:

    4 Using a printer data cable (serial interface orconverter for the parallel interface)

    l Secure the device connectors by tightening thescrews provided!

    Via the Drger RS 232 MEDIBUS interface:

    5 Using a data cable for connecting a Medibus device(e.g. PM 8050 Monitor, Julian AnaestheticWorkstation or a PC).

    l Secure the device connectors by tightening thescrews provided!

    Via the RGB interface:

    6 Using the monitor cable to connect an external RGBscreen (VGA or SVGA, Multisync.).

    l Secure the device connectors by tightening thescrews provided!

    Preparation for use Connecting the System

    14

    RGB RS 232 C RS 232 C CAN 2 CAN 1

    m P

    4 56 12 3

    0182

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • 15

    Preparation for use Connecting the System

    PM 8060 vitara to Julian

    Julian CAN 1/2...................to.....Vitara CAN 1

    Julian CAN 1/2...................to.....Parameter box holder

    Julian COM .......................to.....Vitara RS 232 C

    Printer................................to.....Vitara RS 232 C wThe authorised cable types are listed in the Accessorieslist.

    PM 8060-vitara with external power adapteras stand-alone monitor or connected toCato:

    Power adapter CAN1/2......to.....Vitara CAN 1

    Power adapter CAN1/2......to.....Parameter box holder

    Airway monitor ...................to.....Vitara RS 232 C

    Printer................................to.....Vitara RS 232 C wThe authorised cable types are listed in the Accessorieslist.

    Never connect a CAN connector to an RS 232 Cconnector: especially in PCs, a wrong connection ofthis kind can damage the RS 232 C interface.

    COM 3

    COM 2CAN 1

    CAN 2

    COM 1

    m P

    0192

    9094

    Medibus-Device

    Printer

    Vitara

    Power Supply Parameterbox

    option

    option

    0172

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Fitting the PM 8060 vitara to the swivel armof a Drger Anaesthetic Unit

    l Push the arm on to the rail and fix with two screws.

    1 The arm's resistance to movement can be adjusted atthe hinge (Allen screw - A/F 6 mm).

    2 Screw the PM 8060 vitara securely to the arm. Ifnecessary, fix with thread-locking compound (Allenscrew - A/F 12 mm).

    Fitting the PM 8060 vitara to the swivel armof a standard rail

    l Check the load-bearing capacity of the rail.The weight of the PM 8060 vitara complete withparameter box is approx. 5 kg.

    3 Screw the PM 8060 vitara securely to the arm. If necessary, fix with thread-locking compound (Allen screw - A/F 12 mm).

    4 Place the arm on the rail and fix with the rail clamp.

    l Also fix the power adapter to the rail.

    PM 8060 vitara stand-alone on pedestal

    5 Screw the PM 8060 vitara securely to the pedestal(Allen screw - A/F 12 mm) and stand on a suitablystrong, horizontal and non-slip surface.

    Turning and tilting

    You can adjust the viewing angle of the PM 8060 vitaraand adapt it to the ergonomic requirements of yourworkplace.

    6 The tilt resistance can be adjusted with two cross-slottet screws.

    7 The swivel resistance can be adjusted by the screwon the inside of the fixing column (Allen screw - A/F 5 mm).

    Preparation for useConnecting the System

    16

    D

    D

    1-2--

    Stop

    D

    1-2--

    Stop

    D

    1-2--

    Stop

    12

    3

    4

    5

    6

    7

    0202

    9094

    0212

    9094

    0222

    9094

    0232

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • 17

    Preparation for useConnecting the System

    Fitting the parameter box

    The parameter box can be operated either in its blueholder or in the slot-in housing.

    The slot-in housing can be installed in an anaestheticsystem (e.g. Julian or Cato) by Drger Service.

    The blue holder can be attached by the user to astandard rail (10 x 25 millimetres) in the work area.

    With the blue holder, the parameter box can be pivotedup and down. In the slot-in housing, its position is fixed.

    In both cases, the synchronisation output for anexternal defibrillator is accessible from the back ofthe machine!

    Holder for operating table:

    Alternatively, the "Op. Table" parameter box holder canbe used. This holder is especially suitable for fixing closeto the patient. It can remain on the operating table or bedwhen transporting the patient. The parameter box isconnected to the PM 8060 vitara by means of the quick-release cable and is mounted securely in the housing.

    This system is not compatible with the holders withclip-in mechanism.

    Slot-in housing:

    l Push the parameter box as far as it will go into theslot-in housing. The parameter box must lock intoposition, thereby connecting up its electricalconnections.

    To remove the parameter box, pull the blue handle onthe parameter box to release the locking mechanism.The parameter box can then be pulled out of the slot-in housing.

    Blue holder:

    l Fix the blue holder in a suitable position.

    l Connect the holder to the power supply (i.e. thepower adapter in the stand-alone version or the backof the Julian or Physio Flex) using the cable supplied.

    l Hang the parameter box onto the holder from above.The plug connector must engage.

    l To remove the parameter box, pull the blue handle ofthe box. This releases the box so that it can beremoved from the blue holder.

    1 Defibrillator connection.

    D

    D

    1-2--

    Stop

    1

    0242

    9094

    0252

    9094

    0262

    9094

    1

    0982

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Tilting the parameter box

    After overcoming an initial holding resistance, theparameter box can be easily tilted to an ergonomicallycomfortable angle.

    The holding resistance can be adjusted at the hinge. The adjusting screws are covered by white caps (Allenscrew - A/F 6 mm).

    Switching the monitor on and off

    When used in combination with an anaesthetic system(e.g. Julian):

    l The PM 8060 vitara is switched on and off by themaster switch of the anaesthetic system.

    When used as stand-alone monitor:

    l The power adapter has its own master switch.

    After switching on

    The Monitor runs through its start-up and self-testroutines:

    The internal program memories are tested. All LEDsand display elements light up for about 2 seconds.

    The LED in the Standby key E remains lit. An alarm tone is generated.

    The screen is activated and displays a clock symbol(u = "please wait", duration of delay for the self-test).

    The keys are not yet activated.

    Shortly after this initial period, the Monitor automaticallyenters Standby mode.

    Preparation for useSwitching the machine on and off

    18

    Self-test

    Technology for life

    PM 8060 VitaraVersion 2.n 02-12-97

    deletetrend

    config.

    Switch on with

    Standby

    Data ManagerSW-Version 2.n

    02-12-97

    ParameterboxSW-Version 2.n

    PM 8060 Vitara

    E

    0282

    9094

    0292

    9094

    0272

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Contents

    Configuring the screen .......................................................................... 20Mode....................................................................................................... 22Alarm limits ..............................................................................................23Parameters...............................................................................................23Configuring the standard screen with modules ........................................ 27Configuring the screen colours ................................................................30Acoustic ..................................................................................................31List entry ................................................................................................. 32Transport function.................................................................................... 32RS 232 MEDIBUS................................................................................... 34RS 232 Printer ........................................................................................ 34Basic configuration...................................................................................35

    Screen functions during operation ........................................................ 35Starting up the screen ............................................................................. 35Standard screen ......................................................................................37List screen .............................................................................................. 37Trend screen with zoom function ............................................................. 39Alarm limits ..............................................................................................40Autosetting the patient alarms ..................................................................41Alarm information .................................................................................... 42Configuring the screen structure in operation ...........................................42Parameters ..............................................................................................42Timer .......................................................................................................43Configuration .......................................................................................... 43

    Alarm concept ........................................................................................45Alarm priority ...........................................................................................45

    19

    Screen functionsContents

    Screen Functions

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • 0302

    9094

    0102

    9094

    Configuring the screen

    Basic configuration in standby mode:

    The basic configuration is protected by a password andcan only be modified in standby mode. This configurationis reactivated whenever the Monitor is started (i.e.switched on or operated from standby mode). It can alsobe reactivated by selecting Call standard duringoperation.

    Temporary configuration during operation:

    The screen settings entered during operation remainactive until the screen is next switched off.

    The alarm limits set during operation only remain activeuntil the screen is next switched to standby mode.

    In both cases, after switching off, the password-protected basic configuration is reactivated. Thisconfiguration can also be activated at any time byselecting Call standard.

    If the screen has been blacked out by the screen-saverfunction, it will be reactivated as soon as any activity isdetected on the machine!

    The Standby screen contains two screen keys(softkeys) for calling up menus:

    1 Delete trend; e.g. for a new patient:

    The Standby screen contains two screen keys(softkeys) for calling up menus for

    deleting the trend and list memory

    configuring the screen.

    The trend memory and list are both cleared at the sametime. If a parameter box is connected, the data in theparameter box will also be deleted!

    l Press the delete trend softkey. The system asks you again to confirm you are sureyou want to delete the trend.

    l Press the delete key to confirm.

    If you press do not delete, the original screen willbe restored unchanged.

    2 Calling up the standby configuration menu

    l Press the config softkey to obtain the configurationmenu.

    Configurations set in standby mode remain activewhenever the machine is switched on again.

    By contrast, configurations entered during operation areonly valid until the machine is next switched off (alarmlimits remain valid until the next Standby).

    Screen functionsConfiguring the screen

    20

    deletetrend

    config.

    Switch on with E

    Standby

    Data ManagerSW-Version 2.n

    02-12-97

    ParameterboxSW-Version 2.n

    PM 8060 Vitara

    deletetrend

    config.

    Switch on with

    Standby

    Data ManagerSW-Version 2.n

    02-12-97

    ParameterboxSW-Version 2.n

    PM 8060 Vitara

    E

    1

    2

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • The screen illustrated opposite is displayed:

    l Turn the rotary control until the cursor frame ispositioned over default values.

    l Press the rotary control.

    Changing default values

    The password-protected default settings are activatedwhenever the machine is switched on. Before thesesettings can be changed, the correct password must beentered:

    Entering the password:

    l The password consists of four digits. The numbersare selected one-by-one from the number bar byturning the rotary control and pressing to confirm ineach case.

    After you have confirmed the fourth correct digit, you willbe allowed to access the "default values" pull-downmenu.

    The menu for configuring the default values and settingsis displayed:

    mode Switches over between adult andneonatal mode (toggle).

    alarm limits Configure the defaults for adult orneonatal alarm limits.

    parameters Default settings for thehaemodynamic parameters.

    screen Configures the various screens.

    acoustic Sets the volume and type of tonesequence.

    transport Defines the data exchange betweenparameter box and system monitorduring patient transfer.

    RS 232 (MEDIBUS) Sets the interface parameters

    RS 232 (Printer) Configures the printer interface orthe second Medibus interface.

    basic config. Date time and language selectionand location of the machine.

    21

    Screen functionsConfiguring the screen

    Standby/Configuration alarms inactive !

    10:3400:00

    01-10-96Check system configuration. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    Standby/Configuration alarms inactive !

    10:3400:00

    01-10-96

    Default values

    These default settings are passwordprotected.The default settings are activeon start-up.

    To change the default settingsenter the password:

    0 1 2 3 4 5 6 7 8 9

    0332

    9094

    0322

    9094

    Standby/Configuration alarms inactive !

    10:3400:00

    01-10-96Check system configuration. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    0332

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • These settings are always called up and changed by thesame procedure:

    l Turn the rotary controlto activate the option marked by the cursor frame.

    l Press the rotary controlto make a selection inside the dashed border.

    l Fields with orange backgroundrepresent the current settings. The illustrations thatfollow always show the factory-set defaults.

    l Fields with black or white backgroundshow the previous menu steps used to reach thecurrently selected field.

    l The arrow ( z )refers back to the previous menu level.

    Mode

    For the following features, the PM 8060 vitara hasseparate data sets in adult and neonatal modes

    alarm limits

    curve speed,

    These data sets can be separately configured and storedby the user:

    The user can switch between these data sets byselecting:

    adult = adult mode

    neon. = neonatal mode.

    Various settings for NiBP measurement are also switchedover at the same time.

    Screen functionsConfiguring the screen

    22

    Standby/Configuration alarms inactive !

    10:3400:00

    01-10-96Select the required mode (adult or neonate). Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    0342

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Alarm limits

    This menu defines:

    the default alarm limits for the two different operatingmodes (adults and neonates).

    These alarm limits are automatically active after

    switching on the PM 8060 vitara

    leaving Standby mode

    selecting default under Alarms.

    If the adult/neonate (adult/neon.) mode is changed later,the alarm limits preset for the relevant mode will remainvalid.

    From this menu, the familiar procedure of selecting(turning the rotary control) and confirming (pressingthe rotary control) can be used to open the following sub-menus

    iBP

    ECG / SpO2 / NiBP / Temp / ST / RR

    complete with the corresponding settings for the upperand lower alarm limits.

    Two dashes ( ) in the table mean that this alarm limitis inactive and not monitored. It can be activated by turning the rotary control beyond the maximum orminimum value possible and confirming it.

    Parameters

    The settings for the following parameters are defined inthis menu:

    ECG Electrocardiogram

    NiBP Non-invasive blood pressure

    iBP measuring site Preconfiguration for measuringsites

    iBP channels Preconfiguration for channels

    SpO2/Pleth Pulsoximetry/plethysmogram

    Resp Respiration

    23

    Screen functionsConfiguring the screen

    Standby/Configuration alarms inactive !

    10:3400:00

    01-10-96Standard alarm limits configuration menu. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    standard adult

    standard neon.

    Standby/Configuration alarms inactive !

    10:3400:00

    01-10-96Standard alarm limits configuration menu. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    iBP

    standard - adult

    ECG/SpO2/NiBP/Temp/

    Aorta sys

    dia

    A.Pulm sys

    mean

    dia

    CVP mean

    ICP mean

    ? sys

    mean dia

    HR/Pulse

    Pulsedef.

    SpO2

    NiBP/ART sys

    dia

    RR

    T1

    T2

    ST

    --

    120 50

    160 80

    -- --

    -- --

    92

    100 50

    10

    160 80

    100 50

    24 8

    24 8

    20 --

    8 0

    16 0

    160 80

    100 50

    24 8

    25 5

    0.10+ -

    Standby/Configuration alarms inactive !

    10:3400:00

    01-10-96

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    ECG

    NiBP

    iBP locations

    iBP channels

    SpO2 / Pleth

    Resp.

    Select a menu position. Please confirm !Change the settings. Please confirm again !

    ECG

    function on off

    numberof leads 3 5

    ST-segmentanalysis on off

    pulse deficit on offindication

    pacemaker on off

    filter on off

    amplitude 2

    1.ECG2.ECG 3.ECG

    derivation I IIIII

    0352

    9094

    0362

    9094

    0372

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • ECG settings

    With the parameters:

    measuring function Switching ECG measurement onor off.

    number of leads Choice of 3-core or 5-lead cable.

    derivation Choice of displayed derivation forthe 1st, 2nd and 3rd ECG curve.

    amplitude Measurement sensitivity.

    filter Switches the filter on or off.

    pacemaker Switches pacemaker detection onor off (see page 59).

    pulse deficit indicationSwitches the indication on or off.

    ST-segment analysis Switches the function on or off.

    If a 3-lead ECG cable is selected, only one ECG can bedisplayed. The options of a 2nd and 3rd ECG aretherefore removed.

    NiBP settings

    With the parameters:

    time interval Sets the time between measure-ments.

    active alarms Selects which criterion is to bemonitored (diastolic or systolic).

    initial pressure Sets the starting pressure for theselected mode (value depends on the selectedmode - adult / neonate).

    punction pressure Sets the punction pressure for theselected mode(value depends on the selectedmode - adult / neonatal).

    units Sets the measurement unit.

    interlock Selects whether the interlockfunction should be totally deacti-vated or coupled to either SpO2measurement or iBP measurement(thereby suppressing the relevantalarms and measured values duringNiBP measurement).

    Screen functionsConfiguring the screen

    24

    ECG

    function on off

    numberof leads 3 5

    1.ECG2.ECG 3.ECG

    derivation

    ST-segmentanalysis on off

    pulse deficit indication on off

    pacemaker on off

    filter on off

    amplitude 2

    I IIIII

    NiBP

    auto on off

    punction Start Stop

    measure Start Stop Turbo

    time interval 15 min

    active alarms sys dia

    initial pressure 180 mmHg

    punction pressure 40 mmHg

    units mmHg kPa

    interlock:

    NiBP-SpO2 on off

    NiBP-iBP on off

    0732

    9095

    0732

    9095

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • iBP locations

    This function can be used to preset all the measuringsites independently of the measuring channel. Thefollowing abbreviations are used:

    ART Arterial

    AORTA Aorta

    A. Pulm. Pulmonary artery

    CVP Central venous pressure

    ICP Intracranial pressure

    ? Any other site

    The following parameters can be set:

    active alarms Selects the monitoring criterion.For example, should systolic, diastolic ormean pressure be monitored?

    graph Selects the curve scale.Fixed values or automatic amplitudeadjustment.

    amplitude Selects the curve scale.Not selectable with automatic amplitudeadjustment.

    pulse Switches pulse rate display on/off.

    iBP channels

    defines the input channels P1 and P2 of the parameterbox:

    measurement functionSwitches the channel on and off.

    location Enters the locations of this channel.

    sensitivity Selects the sensor sensitivity.42.5 or 50 m V/V/mmHg.

    SpO2/Plethysmogram

    With the parameters:

    measuring function Switches SpO2 measurement on andoff.

    pulse Switches pulse rate display on and off.

    C-Lock Switches ECG synchronisation on andoff.

    mode Sets the measurement response speed.(Fast measurements are moresusceptible to disturbance).

    25

    Screen functionsConfiguring the screen

    sysabsol.160onsysabsol.160onsysabsol.40meanabsol.20meanabsol.20sysabsol.160

    ART

    AORTA

    A. Pulm

    CVP

    ICP

    ?

    active alarmsgraphamplitudepulseactive alarmsgraphamplitudepulseactive alarmsgraphamplitudeactive alarmsgraphamplitudeactive alarmsgraphamplitudeactive alarmsgraphamplitude

    mean

    mean

    diamorph.

    offdiamorph.

    offdiamorph.

    morph.

    morph.

    diamorph.

    iBP 1

    iBP 2

    function

    location

    sensitivity

    function

    location

    sensitivity

    on

    AORTA

    42.5

    on

    CVP

    42.5

    off

    50

    off

    50

    2

    function

    pulse

    C-Lock(ECG synchr.)

    mode

    on

    on

    on

    slownormalfast

    off

    off

    off

    SpO2 / Pleth

    0392

    9094

    0392

    9094

    0392

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Respiration

    With the parameters:

    function Switches respiration measurement onand off.

    mode Automatic or manual adjustment of therecorded level

    amplitude Selects the curve size. The minimumsize must correspond to the length ofthe displayed bar.

    trigger Activates/deactivates the superimposeddisplay of respiration recording pulseson the respiration curve (to check thatrespiration recording is correctly inprogress).

    apnea Time interval in seconds after which anapnoea alarm is triggered in theabsence of respiration measurementsignal.

    Screen functionsConfiguring the screen

    26

    Resp.

    function on off

    mode auto man

    amplitude 1 E/cm

    trigger on off

    apnea 10 s

    0382

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Configuring the standard screen withmodules

    This menu is used to configure the screen. The machinestores the configuration and screen-related settings:

    standard screen

    colour

    curve speed

    general settings

    When selected, the standard screen is immediatelyavailable with the desired setting.

    The screen is made up of modules, which can beselected from the menu by means of the rotary control.

    Configuring standard screen

    Example: The screen structure is shown schematically onthe left-hand side of the screen. Alongside it, there is atable of graphical modules and numerical modules forselection.

    Modules which are already selected are highlightedagainst an orange background, and their designations arecontained in the module blocks.

    l Move the cursor frame through the module table tothe module you want to change (in this example: iBP 1 and iBP 2) by turning the rotary control.

    l Press the rotary control to confirm.

    The module block in the table goes grey and appearsin the schematic layout on the left, bordered by acursor.

    l The block can now be moved to the desired positionin the schematic layout by turning the rotary control.

    l When the rotary control is pressed to confirm, thedesired module is set in its current position on thescreen.

    l To delete a module, the relevant position on thescreen must be overwritten with a blank module.

    27

    Screen functionsConfiguring the screen

    S h /EM 6/Stb f / BS A fb 1 R

    Standby/Configuration

    10:3400:00

    01-10-96Select a graph or a module by turning the rotary knob. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    screen

    colours

    curve speed

    general settings

    1. ECG HR

    1. ECG x2 HR x 2

    2. ECG Temp

    3. ECG NiBP

    iBP 1 iBP 1

    iBP 2 iBP 2

    iBP1+BP2 NiBP x2

    Pleth SpO2

    Pleth x 2 SpO2 x2

    Resp. Resp.

    Resp x 2 RR x 2

    BLANK

    Graph Module

    screen

    1. ECG

    Temp NiBP

    HR

    Pleth SpO2

    iBP 1

    iBP 2iBP 1 + 2

    alarms inactive !

    S h /EM 6/Stb f / BS A fb 2 R

    Standby/Configuration

    10:3400:00

    01-10-96Please position selected parameter by turning the rotary knob. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    screen

    colours

    curve speed

    general settings

    screen

    1. ECG

    Resp. RR

    HR

    Pleth SpO2

    iBP 1

    iBP 2iBP 1 + 2

    Graph Module

    alarms inactive !

    1. ECG HR

    1. ECG x2 HR x 2

    2. ECG Temp

    3. ECG NiBP

    iBP 1 iBP 1

    iBP 2 iBP 2

    iBP1+BP2 NiBP x2

    Pleth SpO2

    Pleth x 2 SpO2 x2

    Resp. Resp.

    Resp x 2 RR x 2

    BLANK

    0392

    9094

    0402

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Examples of the selectable graph modules:

    ECG graph ECG real-time graph. The derivation,amplitude and curve speed are freelyselectable in discrete increments.

    This graph can also be displayed in adouble-height module.

    iBP curves:

    ART Displays the iBP curve for the arterialblood pressure in a separate graph ofsingle module height.

    CVP Displays the iBP curve for the centralvenous blood pressure in a separategraph of single module height.

    iBP 1 + iBP 2 Displays two iBP curves (ART and CVP)in a single graph with double the moduleheight.

    Plethysmogram curveReal-time curve of the plethysmogrammeasured by the SpO2 sensor.

    This curve can also be displayed in adouble-height module.

    Respiration curveReal-time curve of the respirationcharacteristic.

    This curve can also be displayed in adouble-height module.

    28

    Screen functionsConfiguring the screen

    1. ECG

    ZVD

    ART

    CVP

    160

    0

    20

    0

    Respiration

    iBP 1+iBP 2

    60 0

    ART130

    CVP20

    Pleth

    1. ECG (Module double height)

    Pleth (Module double height)

    Respiration (Module double height)

    0412

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Examples of the selectable numerical modules:

    Temp 1/2 Shows the temperature values of theselected channel (1 or 2).

    NiBP Shows the non-invasive blood pressure,the mean pressure and the timeremaining until the next measurement.During measurement, the bar displaysthe current cuff pressure.

    SpO2 Shows the functional O2 saturation ofthe blood and the pulse rate.

    iBP 1/2 Shows the blood pressure of theselected channel (1 or 2).

    The systolic/diastolic and meaninvasively measured blood pressurevalues are displayed, together with thepulse rate.

    For CVP and ICP, the mean pressureonly is displayed.

    Heart rate Shows the heart rate derived from the1st ECG. Additionally, the pulse deficitobtained from the difference betweenthe heart rate and the pulse ratemeasured during SpO2 measurementcan be displayed, as well as the ST segment size of the 1st ECG.

    Respiration Shows the respiration rate derived fromthe first lead of the 1st ECG.

    29

    Screen functionsConfiguring the screen

    36.8 37.2T1 T2 C

    15min M (102)

    SysmmHgDia

    122/ 89NiBP

    Sys

    man 10:34 M (102)

    man 10:35 M (11.1)

    mmHgDia

    122/ 89NiBP

    117 M (102)

    SysmmHgDia

    122/ 89NiBP

    SyskPaDia

    15.2/ 9.3NiBP

    99pulse 65

    SpO2 %

    mmHg mmHg

    mmHg

    1/min

    1/min

    (ecg)

    mmHg

    4.5mean

    CVPiBP 2

    122/ 89(100)

    ARTiBP 1

    pulse 65

    122/ 89(100)

    AORTAiBP 1

    pulse 65

    mmHg

    122/ 89(100)

    A.PulmiBP 1

    pulse 65

    mmHg

    122/ 89(100)

    iBP ?iBP 1

    pulse 65

    65pulsedef. 0 ST + 0.05

    HR/Pulse

    iBP 2

    4.5mean

    ICP

    21RR

    10 s

    (ecg)

    Asystolie

    10 sApnoe

    pulsedef. 0 ST + 0.05

    HR/Pulse

    RR

    0422

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Configuring the colour settings

    Colour settings always apply to all the featuresassociated with a parameter. The numerical modules, if configured for colour, are the same colour as theircorresponding curves.

    l Select and confirm the desired parameter.

    l Change the colour by turning the rotary control. The colours are displayed in the sequence shown inthe colour bar at the top of the menu.

    l Confirm the desired colour.

    Configuring the background settings

    dark Black background, light curves andnumbers. Preferably used in a darkenvironment.

    light White background, dark curves andnumbers. Preferably used in a brightenvironment.

    colour modules

    yes The numerical modules will be thesame colour as the correspondingcurves.

    no The numerical modules are displayedin black-and-white.

    Example:

    Screen with dark background

    Screen functionsConfiguring the screen

    30

    Standby/Configuration

    10:3400:00

    01-10-96 Select a parameter. Please confirm !Select a colour. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    screen

    colours

    curve speed

    general settings

    ARTAORTAA.PulmCVPICPiBP ?ECGNiBPSpO2Resp.Temp

    alarms inactive !

    background bright dark

    modulescoloured yes no

    Standby/Configuration

    10:3400:00

    01-10-96 Select a parameter. Please confirm !Select a colour. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    screen

    colours

    curve speed

    general settings

    ARTAORTAA.PulmCVPICPiBP ?ECGNiBPSpO2Resp.Temp

    alarms inactive !

    background bright dark

    modulescoloured yes no

    0432

    9094

    0442

    9094

    10:3400:00

    01-10-96

    standard screen - adultAdvisory:Caution:Warning:

    1. ECGII

    1mV

    Pleth

    auto setlimits

    alarminfo

    screenconfig.

    para-meter

    timerstart

    limits

    config.

    65pulse def. 0 ST + 0.05

    HR/Pulse (ECG)

    122/ 89(100)

    ARTiBP 1

    99pulse 65

    pulse 65

    4.5mean

    CVPiBP 2

    15min M (102)

    SysmmHgDia

    SpO2

    122/ 89NiBP

    36.8 37.2T1 T2 C

    ART

    CVP

    130

    10

    80

    0

    0952

    9095

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Configuring the curve speed

    The speed of the curves is determined in mm/s for thefollowing measured values:

    haemodynamic adultsHaemodynamic values in adult mode.

    haemodynamic neonatesHaemodynamic values in neonatal mode.

    Configuring the general settings

    The following units can be switched on or off:

    unit yes = ON

    no = OFF

    Acoustic

    The volume and tones are defined here.

    0 means off, while 1 is the lowest and 9 thehighest volume.

    alarm sound Volume of the alarm sound

    pulse tone Volume of the pulse tone

    ECG Tone - SpO2modulated

    Acoustic heart beat signal.If yes is selected, the pitch of the heartbeat tone varies with oxygen saturation. If no the plain heart beat tone is ofinvariable pitch.

    tone sequence Specifies the tone sequence, either toEuro- or to Drger standard.

    31

    Screen functionsConfiguring the screen

    Standby/Configuration

    10:3400:00

    01-10-96Screen configuration menu. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    screen

    colours

    curve speed

    general settings

    haemodynamic 12,5 25 50adult mm/s

    haemodynamic 12,5 25 50neonate mm/s

    alarms inactive !

    Standby/Configuration

    10:3400:00

    01-10-96Screen configuration menu. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    screen

    colours

    curve speed

    general settings

    units yes no

    alarms inactive !

    Standby/Configuration

    10:3400:00

    01-10-96Alarm sound and pulse tone configuration menu. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    alarm sound 1 2 3 4 5 6 7 8 9

    pulse tone 0 1 2 3 4 5 6 7 8 9

    pulse tone ECG Pleth.SpO2 modulated

    tone sequence DRGER Euro-Standard

    alarms inactive !

    0452

    9094

    0112

    9094

    0462

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • List entry

    This option opens a menu in which you can define thetimes (fixed intervals) or events (warning, caution, NiBPmeasurement) that generate list entries.

    In addition, this menu defines which blood pressuresource is represented.

    Transport function

    The parameter box stores the following data from thesystem monitor:

    patient data for the list screen,

    alarm limit settings and

    measurement parameter settings.

    The patient therefore remains connected to theparameter box while in transit. The patient's data iscontinuously available. There is no need to reconnect andset monitors. A complete list of transportable data isgiven in the chapter Technical Data.

    The transport function can be preset in Standby.

    Transport function on/offWhen the transport function is off, no data isexchanged between the system monitor and theparameter box (for stationary use of the parameter box).

    When on, data is communicated.

    Transport menus yes/noIf set to no, data is transferred during operational useaccording to the default settings in Standby (fullyautomatic process).

    If yes, the user has the option of modifying the type andscope of the data transfer. The displayed transport menualways corresponds to the default settings describedhere (semi-automatic mode).

    Screen functionsConfiguring the screen

    32

    Standby/Configuration

    10:3400:00

    01-10-96Alarm sound and pulse tone configuration menu. Please confirm !

    Default values

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    alarm sound 1 2 3 4 5 6 7 8 9

    pulse tone 0 1 2 3 4 5 6 7 8 9

    pulse tone ECG Pleth.SpO2 modulated

    tone sequence DRGER Euro-Standard

    alarms inactive !

    Standby / Configuration

    11:5900:00

    01.04.96

    Default values

    alarms inactive!

    Select a menu position. Please confirm !Change the settings. Please confirm again !

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    transportfunction on off

    transport-menus yes no

    The patient is newto the:

    monitor yes no

    parameterbox yes no

    Which data do you wantto transfer from theparameter box?

    settings yes noalarm limits yes nolist yes no

    Standby / Configuration

    11:5900:00

    01.04.96

    Default values

    alarms inactive!

    Select a menu position. Please confirm !Change the settings. Please confirm again !

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    transportfunction on off

    0472

    9094

    0482

    9094

    0492

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • The patient is new for

    If a patient is connected for the first time, no data on thepatient is yet stored in the system monitor or parameterbox. This case will generally apply in the induction room.(see the example in the left-hand of the two isolatedmenus opposite).

    1 In this case, yes must be entered both for systemmonitor and for parameter box.

    Any data remaining from the previous patient willbe deleted.

    2 If the patient arrives with a connected parameter box,i.e. with a "previous history", enter no after para-meter box: the data from the parameter box will betransferred to the system monitor. This case willgenerally apply in theatre or recovery room. (See theexample in the right-hand of the two isolated menusopposite).

    The following dialogue then appears on the systemmonitor:

    Which data do you want to transfer from the para-meter box?

    Settings yes/noIf yes, the measurement parameter settings aretransferred from the box (e.g. iBP calibration, mode,NiBP interval, ECG amplitude etc.)

    If no, the current system monitor values remainvalid.

    Alarm limits yes/noIf yes, the alarm limits are transferred.

    If no, the current system screen values remain valid.

    List yes/noIf yes the data stored in the parameter box istransferred and inserted in the list.

    If no the system screen sets up a new list. The newlist is valid ("start-time") as soon as these settingshave been validated.

    In normal operation (see section on Parameter Box),this dialog only occurs when Transport Menus is set toyes. Otherwise, the instructions programmed here areexecuted immediately.

    In standardised operating theatre organisation, thetransport menus can also be switched off (TransportMenus set to no). In this case, data is transferred asspecified by the preselected settings, without any furtherdialog or confirmation.

    33

    Screen functionsConfiguring the screen

    Standby / Configuration

    11:5900:00

    01.04.96

    Default values

    alarms inactive!

    Select a menu position. Please confirm !Change the settings. Please confirm again !

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    transportfunction on off

    transport-menus yes no

    The patient is newto the:

    monitor yes no

    parameterbox yes no

    Which data do you wantto transfer from theparameter box?

    settings yes noalarm limits yes nolist yes no

    The patient is newfor the:monitor yes noparameter-box yes no

    transport-function on offtransport-menus yes no

    transport-function on offtransport-menus yes no

    The patient is newfor the:monitor yes noparameter-box yes no

    Which data do you wantto transfer from theparameter box?

    settings yes noalarmlimits yes no

    list yes no

    1 2

    0492

    9094

    0692

    9094

    0692

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • RS 232 (MEDIBUS)

    This menu is used to configure the data transferinterface:

    baud rate Transmission speed (variable, seeInstructions for Use of the device you wishto connect).1.2 or 9.6 kBaud

    The following parameters are displayed for informationonly and cannot be changed:

    parity even

    data bits 8

    stop bits 1

    RS232 (Printer)

    This menu is used to configure the interface for datatransfer to a printer (see Instructions for Use of thedevice you wish to connect).

    select protocol Selects whether the interface should beused for a printer or as an additionalMEDIBUS interface

    baud rate Transmission speed

    parity Parity of the check bits

    data bits Number of data bits

    stop bits Number of stop bits

    Screen functionsConfiguring the screen

    34

    Standby / Configuration

    11:5900:00

    01.04.96

    Default values

    alarms inactive!

    Printer/Medibus interface configuration menu. Please confirm !

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    record output printer Medibus

    baud rate(kBaud) 1.2 2.4 4.8 9.6 19.2

    Parity even odd none

    data bits 7 8

    stop bits 1 2

    Standby / Configuration

    11:5900:00

    01.04.96

    Default values

    alarms inactive!

    Medibus interface configuration menu. Please confirm !

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    baud rate 1.2 9.6(kBaud)

    parity even

    data bits 8

    stop bits 1

    0512

    9094

    0522

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • 0052

    9094

    Basic configuration

    The basic configuration comprises the following items:

    time for the current time

    date for the current date

    language for the language version. Thefollowing languages are available:

    English GBFrench FGerman DDutch NLItalian ISpanish EDanish DKSwedish SNorwegian N

    location The location of the machine can beentered with a maximum of 9 alpha-numeric characters. This locationname will then be displayed in the listafter the defined interval.

    This menu is automatically closed onconfirming the 9th character.

    Screen functions during operation

    Starting up the screen

    1 Press the E key.The standard screen is displayed.

    2 To call up other screens:Press the S key.A selection menu for the following screens isdisplayed:

    Standard screen 1,

    Standard screen 2,

    List screen and

    Trend screen.

    3 You can return directly to the Standard screen atany time by pressing the Q key.

    35

    Screen functionsConfiguring the screen

    Screen functions during operation

    Standby / Configuration

    11:5900:00

    01.04.96

    Default values

    alarms inactive!

    Language, date and time configuration menu. Please confirm !

    mode adult neon.

    alarm limits

    parameter

    screen

    acoustic

    list entry

    transport

    RS 232 (Medibus)

    RS 232 (Printer)

    basic config.

    time: 11 : 59date: 01 - 04 - 96 language: GB F D NL E I DK S Nmon. location: XXX

    D

    1-2--

    Stop

    2

    1

    0532

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Standard screencontains the graph and measured value fieldsconfigured by the user.

    The list screencontains all the measured values and alarms whichhave been saved and makes it easier to complete theanaesthetic record (see page 38). The uppermostgraph of th standard screen is shown at the top.

    The trend screenshows graphically the measured values over time. The uppermost graph of the standard screen isshown at the top.

    Screen functionsScreen functions during operation

    36

    10:3400:00

    01-10-96

    standard screen - adultAdvisory:Caution:Warning:

    65pulse def. 0 ST + 0.05

    122/ 89(100)

    ARTiBP 1

    4.5mean

    CVP

    99pulse 65

    pulse 65

    15min M (102)

    SysmmHgDia

    SpO2

    122/ 89NiBP

    36.8 37.2T1 T2 C

    iBP 2

    auto setlimits

    alarminfo

    screenconfig.

    para-meter

    timerstart

    limits

    config.

    1. ECGII

    1mV

    Pleth

    ART

    CVP

    130

    10

    80

    0

    HR/Pulse (ECG)

    11:3400:00

    01-10-96

    list screen - adultAdvisory:Caution:Warning:

    intubation

    start ofsurgery

    positionchange

    extuba-tion

    start ofanaesthesia

    end ofsurgery

    event

    print

    1. ECGII

    1mV

    65pulse def. 0 ST + 0.05

    Manual list entry by

    time

    10:55 interval 2 01-10-9610:55 IND-210:55 11:0011:0511:08 intubation11:1011:1311:1511:20 interval 1 01-10-9611:20 OP-2 11:20 11:25 11:3011:32 start of surgery

    HR/Pulse1/min

    6467

    636263

    6362

    NiBP mmHgSys/M/Dia

    120/100/90123/103/88

    120/100/90123/103/88128/107/93

    120/100/90122/102/89

    SpO2%

    90 97

    96 90 97

    95 97

    CO2Vol. %Fet

    3537

    313433

    3537

    PAW mbar peak/PEEP

    32/3 31/3

    33/3 31/3 30/3

    32/3 31/3

    Warning

    SPO2

    SPO2

    !!!

    !!!

    MVL/min

    6.4 6.3

    6.2 6.1 6.0

    6.4 6.3

    Cpatml/mbar

    1212

    121212

    1212

    O2Vol. %Fi

    2131

    323133

    3031

    anaesth.agentFi

    ENF1.41.4 1.21.21.1

    ISO1.41.4

    tempCT1

    36.836.9

    36.736.836.9

    36.836.9

    config. list entry next page preceding page

    HR/Pulse (ECG)

    10:3400:00

    01-10-96 Select a trend sector and magnify by confirming !

    NiBP SpO2

    HR/RRIBP 1200

    100

    0

    100

    60

    80

    200

    100

    0

    11:00 12:00 13:00

    Sys

    MDia

    Sys

    MDia

    200

    100

    0

    11:00 12:00 13:00

    1. ECG II

    1mV

    fulltrend

    config.

    trend screen - adultAdvisory:Caution:Warning:

    65pulse def. 0 ST + 0.05

    HR/Pulse (ECG)

    0142

    9094

    0152

    9094

    0162

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Standard screens

    The standard screen is preconfigured by themanufacturer.

    The configuration of these screen pages is describedfrom page 26 onwards.

    The screen keys (softkeys) are described from page 38onwards.

    Example standard screen:

    List screen

    Certain measured values which have been automaticallyor manually saved are displayed here for documentation(to facilitate compilation of the anaesthetic record).

    An interval, limited by clipping the parameter box in andout, is displayed in the list by the label interval and theassociated location of the machine.

    The list can contain a maximum of 100 lines.

    This page also contains ventilation parameters importedvia the MEDIBUS interface.

    The measured values and the warning or cautionmessages triggering them are recorded line by line,together with the time, either at specified intervals and/oras a function of warning or caution messages or of anNiBP measurement.

    A list entry can be generated manually at any time bypressing U If entries are made manually, a suitabledesignation can be selected by means of the on-screensoftkeys:

    For example: start of surgerytransfer patient

    Intubation

    A paper printout of the entire list is generated on anexternal connected printer by pressing the print screenkey.

    37

    Screen functionsScreen functions during operation

    10:3400:00

    01-10-96

    standard screen - adultAdvisory:Caution:Warning:

    65pulse def. 0 ST + 0.05

    122/ 89(100)

    ARTiBP 1

    4.5mean

    CVP

    99pulse 65

    pulse 65

    15min M (102)

    SysmmHgDia

    SpO2

    122/ 89NiBP

    36.8 37.2T1 T2 C

    iBP 2

    auto setlimits

    alarminfo

    screenconfig.

    para-meter

    timerstart

    limits

    config.

    1. ECGII

    1mV

    Pleth

    ART

    CVP

    130

    10

    80

    0

    HR/Pulse (ECG)

    11:3400:00

    01-10-96

    list screen - adultAdvisory:Caution:Warning:

    intubation

    start ofsurgery

    positionchange

    extuba-tion

    start ofanaesthesia

    end ofsurgery

    event

    print

    1. ECGII

    1mV

    65pulse def. 0 ST + 0.05

    Manual list entry by

    time

    10:55 interval 2 01-10-9610:55 IND-210:55 11:0011:0511:08 intubation11:1011:1311:1511:20 interval 1 01-10-9611:20 OP-2 11:20 11:25 11:3011:32 start of surgery

    HR/Pulse1/min

    6467

    636263

    6362

    NiBP mmHgSys/M/Dia

    120/100/90123/103/88

    120/100/90123/103/88128/107/93

    120/100/90122/102/89

    SpO2%

    90 97

    96 90 97

    95 97

    CO2Vol. %Fet

    3537

    313433

    3537

    PAW mbar peak/PEEP

    32/3 31/3

    33/3 31/3 30/3

    32/3 31/3

    Warning

    SPO2

    SPO2

    !!!

    !!!

    MVL/min

    6.4 6.3

    6.2 6.1 6.0

    6.4 6.3

    Cpatml/mbar

    1212

    121212

    1212

    O2Vol. %Fi

    2131

    323133

    3031

    anaesth.agentFi

    ENF1.41.4 1.21.21.1

    ISO1.41.4

    tempCT1

    36.836.9

    36.736.836.9

    36.836.9

    config. list entry next page preceding page

    HR/Pulse (ECG)

    0142

    9094

    0152

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • The list may continue for several pages, so that it may benecessary to leaf through in order to locate a specificentry.

    Return to previous page:

    l Turn rotary control to select previous page andpress to confirm.

    Continue to next page:

    l Turn rotary control to select next page and confirm.

    Define list entry:

    l Select list entry with the rotary control and confirm.A menu appears for you to select at what times (fixedintervals) or events (warning, caution, NiBP measure-ment) list entries should be made.

    In addition, this option is used to define the bloodpressure.

    The list entries for the currently used screen can also bemodified later.

    Deleting the list and trend memory:The trend memory and list are deleted together. They canonly be deleted in Standby mode. If the parameter boxis connected, the data in the parameter box will also bedeleted!

    l Press the E key to switch over to standby mode. l Press the delete trend softkey.

    The system asks you to confirm that you are sure youwant to delete the trend.

    l Press the delete softkey to confirm.

    To restore the unchanged screen without deleting,press do not delete.

    Screen functionsScreen functions during operation

    38

    deletetrend

    config.

    Switch on with E

    Standby

    Data ManagerSW-Version 2.n

    02-12-97

    ParameterboxSW-Version 2.n

    PM 8060 Vitara

    0102

    9094

    11:3400:00

    01-10-96

    list screen - adultAdvisory:Caution:Warning:

    intubation

    start ofsurgery

    positionchange

    extuba-tion

    start ofanaesthesia

    end ofsurgery

    event

    print

    1. ECGII

    1mV

    65pulse def. 0 ST + 0.05

    Manual list entry by

    time

    10:55 interval 2 01-10-9610:55 IND-210:55 11:0011:0511:08 intubation11:1011:1311:1511:20 interval 1 01-10-9611:20 OP-2 11:20 11:25 11:3011:32 start of surgery

    HR/Pulse1/min

    6467

    636263

    6362

    NiBP mmHgSys/M/Dia

    120/100/90123/103/88

    120/100/90123/103/88128/107/93

    120/100/90122/102/89

    SpO2%

    90 97

    96 90 97

    95 97

    CO2Vol. %Fet

    3537

    313433

    3537

    PAW mbar peak/PEEP

    32/3 31/3

    33/3 31/3 30/3

    32/3 31/3

    Warning

    SPO2

    SPO2

    !!!

    !!!

    MVL/min

    6.4 6.3

    6.2 6.1 6.0

    6.4 6.3

    Cpatml/mbar

    1212

    121212

    1212

    O2Vol. %Fi

    2131

    323133

    3031

    anaesth.agentFi

    ENF1.41.4 1.21.21.1

    ISO1.41.4

    tempCT1

    36.836.9

    36.736.836.9

    36.836.9

    config. list entry next page preceding page

    HR/Pulse (ECG)

    0152

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Trend screen with zoom function

    This screen displays the histories of the followingmeasured values, in the form of value curves as afunction of time since the start of measurement:

    NiBP, iBP 1, SpO2 and HR / RR

    In the heart rate field (HR), the respiration rate (RR) isdisplayed on the same scale (cardio respirogram).

    The values can be stored for a maximum of 8 hours.

    The zoom function can be called up when the system hasbeen in operation for more than 30 minutes. A segmentof the time range can be enlarged (possibly severaltimes) with this function. The segment is identified by adashed border. The past is located on the left. Example:

    l Turn rotary control - area changes.

    l Press rotary control - the dashed area is extended tocover the full display width.

    If the monitor has been in operation for a sufficientlylong period of time, a new dashed area appears andcan then be extended as described above.

    The maximum has been reached when no additionaldashed frame appears.

    Back to full trend:

    l Press full trend in order to redisplay the completetrend.

    39

    Screen functionsScreen functions during operation

    10:3400:00

    01-10-96 Select a trend sector and magnify by confirming !

    NiBP SpO2

    HR/RRIBP 1200

    100

    0

    100

    60

    80

    200

    100

    0

    11:00 12:00 13:00

    Sys

    MDia

    Sys

    MDia

    200

    100

    0

    11:00 12:00 13:00

    1. ECG II

    1mV

    fulltrend

    config.

    trend screen - adultAdvisory:Caution:Warning:

    65pulse def. 0 ST + 0.05

    HR/Pulse (ECG)

    0162

    9094

    "O

    nly

    for i

    nter

    nal u

    se b

    y Dr

    aege

    r. No

    t app

    licab

    le a

    s In

    stru

    ctio

    ns fo

    r Use

    !"

  • Limits

    The display comprises the designation of the measuredvalue, the value actually measured on the patient (largenumerals) and the set upper and lowe