Encompass Automation & Engineering Technologies LLC Industrial Automation & Power Distribution | Engineering | Project Management | Electrical Contracting & Installation | Turn-Key Project Services 622 Eckel Road, Perrysburg, OH 43551 USA/ Phone (419) 873-0000 / Fax (419) 872-2612 www.EAETech.com Copyright 2010 EAET-LLC. Confidential and Intellectual Property / Not For Disclosure Nor Abstraction / All Rights Reserved Scale Terminal Calibration Mettler-Toledo IND560

Upload: eaetech

Post on 05-Dec-2014

5.126 views

Category:

Documents


6 download

DESCRIPTION

 

TRANSCRIPT

Page 1: EAET Mettler-Toledo IND560 Calibration

Encompass Automation & Engineering Technologies LLC

Industrial Automation & Power Distribution | Engineering | Project Management | Electrical Contracting & Installation | Turn-Key Project Services 622 Eckel Road, Perrysburg, OH 43551 USA/ Phone (419) 873-0000 / Fax (419) 872-2612 www.EAETech.com

Copyright 2010 EAET-LLC. Confidential and Intellectual Property / Not For Disclosure Nor Abstraction / All Rights Reserved

Scale Terminal Calibration

Mettler-Toledo IND560

Page 2: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only2

How To : Traditional Calibration

Scale System Calibration

Main area of proficiency:- Mettler-Toledo IND560 Scale Terminal

Page 3: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only3

How To : Traditional Calibration

Entering Setup Mode

Press the Down Arrow to viewthe SETUP softkey.

The configuration of the IND560 terminal is accessed through the SETUP softkey

Softkeys

Page 4: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only4

How To : Traditional Calibration

With the Down Arrow navigate to Scale Sub-branch

Use the UP and DOWN navigation keys to move focus through the branches of the setup menu tree.

Press the RIGHT navigation key to expand a branch and the LEFT navigation key to collapse a branch. When the focus is on a sub-branch, focus can quickly be moved back to the main branch by pressing the LEFT navigation key.

When a branch such as Calibration is in focus, press the ENTER key to display the setup screen for that function.

Page 5: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only5

How To : Traditional Calibration

Navigate to Calibration- Press the Down Arrow twice.

When at Calibration-Press Enter

Page 6: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only6

How To : Traditional Calibration

Navigate to Calibration- Press the Down Arrow four (4) times.

Geo Code Base Serial Number

Calibration Units Linearity Adjust

Page 7: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only7

How To : Traditional Calibration

Press Zero softkey - Remove all weight from scale

- Press Start

Page 8: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only8

How To : Traditional Calibration

Press Span softkey

- Last test weight value of 100 Lb. is displayed

- Press Enter twice

- Add test weight either manually or via Automated Test Weight System

Page 9: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only9

How To : Traditional Calibration

Span- Press Start

Page 10: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only10

How To : Traditional Calibration

- Press Exit twice when done

- Unit will return to normal weighing mode

-Weigh Hopper Discharge Feeder needs to be placed back into Automatic.

Page 11: EAET Mettler-Toledo IND560 Calibration

Training Purposes Only

Encompass Automation LLC

Thank you !