Page 1: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the
Page 2: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the
Page 3: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

3

2. Whenmeasuring flammable liquids,observeprecautionsagainstfireorexplo-sion.

3. Whenhandlinghazardousliquids,alwaysfollow the liquidmanufacturer’s safetyprecautions.

4. Whenworking in hazardous environ-ments,alwaysexerciseappropriatesafetyprecautions.

5. Forbestresults,alwaysverifyaccuracybeforeuse.

Product DescriptionThese computer electronics are designedspecificallyforuseonEDMIndustrialGradeTurbineHousings.Theyarealsodesignedtoworkwithseveralaccessoryoutputmodules.TheCMOS,microprocessor-basedelectronicshaveextremelylowpowerrequirementsanddataretentioncapabilitiesinbothRAMandROM. Information is clearly displayed ona large6-digitLCDreadoutwith two-pointfloatingdecimalfortotalsfrom.01to999,999.Alloperationsareeasilyaccessedwiththetwobuttonsonthefrontpanel.Liquid flows through the turbine housingcausinganinternalrotortospin.Astherotorspins,anelectricalsignalisgeneratedinthepickupcoil.Thispulsedataistranslatedfromtheturbineintocalibratedflowunitsshownonthecomputer’sreadout.Upon receipt, examine your equipment forvisibledamage.Thecomputerisaprecisionmeasuringinstrumentandshouldbehandledassuch.Ifanyitemsappeardamagedormiss-ing,contactyourdistributor.Makesureyourcomputermodelmeetsyourspecific needs.Refer to the SpecificationsSection to confirm required features.Themodelnumberofyourcomputerisdisplayedonthelowerfrontsideofthecomputerandalsounderneathabattery.

Ifyouorderedyourcomputerelectronicswithaturbinehousing,itisinstalledatthefactory.

INSTALLATION

Diagram 1

Ifyouorderedyourcomputerseparatelyfromyourturbine,simplymountthecomputerontheturbinewiththefourscrewsatthecornersofthefaceplate.Makesuretheseal isfullyseatedbeforetighteningthescrews.Ifyouorderedthecomputerwithturbineandanaccessorymodule,pleasereviewandthor-oughlyunderstandallinstallationinstructionsbeforeproceeding.AllEDMturbinesaredesignedtomeasureflowinonlyonedirection.Thedirection is indi-catedbythearrowcast-moldedintheturbineoutlet.Ifthecomputerdisplayisupsidedown,removethefourscrews,turnthedisplay180°andreinstallthescrews.SeeDiagram1.

Avoid electronically “noisy” environments.Installatleast6inches(15.2cm)awayfrommotors,relays,ortransformers.Inaddition,EDMmetershaveNEMAType4enclosures.Toensureaccuratemeasurement,removeallairfromthesystembeforeuse.Itisstronglyrecommendedthataccuracybeverifiedprior touse.Todo this, removeallairfromthesystem,measureanexactknownvolumeintoanaccuratecontainer,andverifythevolumeagainstthereadoutorrecordingequipment. If necessary, use a correctionfactortofigurefinalvolume.Forbestresults,accuracy should be verified periodically aspartofaroutinemaintenanceschedule.

Page 4: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

4

CALIBRATION

OPERATION

Computer DisplayAlloperationsarereflectedintheLCDreadout.Thelargecenterdigitsindicateamounts,wheresmallerwordsor“icons”locatedaboveandbelowindicatespecificinformationregardingtotals,flow,calibrationandunitsofmeasure.

Activate the MeterComputerisoncontinuouslyandalwaysreadytoperform.Thecomputerispoweredbyfieldreplaceablebatteries.Whendisplaybecomesdim,fadedorthelowbatterymessageappears(seebelow),thebatteriesneedtobereplaced.ReferencetheMaintenanceSectionfordetails.

Batch and Cumulative TotalsThecomputermaintainstwototals.TheCumu-lativeTotalprovidescontinuousmeasurementandcannotbemanuallyreset.TheBatchTotalcanberesettomeasureflowduringasingleuse.TheCumulativeTotalislabeledwithTOTAL1,BatchTotalislabeledTOTAL2BATCH.WhentheCumulativeTotalreachesadisplayreadingof999,999thecomputerwillhighlightanX10icon.Thisindicatestotheoperatorthatazeromustbeaddedtothe6digitsshown.Whenthenextrolloveroccurs,thecomputerwillhighlightanX100icon.Thisindicatestotheoperatorthattwozeroesmustbeaddedtothe6digitsshown.PresstheDISPLAYbuttonbrieflytoswitchbetween the TOTAL 1, TOTAL 2 BATCHand FLOWRATE. Press DISPLAY brieflyto display theTOTAL2BATCH.Hold theDISPLAYbutton for3 seconds to reset theBatchTotaltozero.Whenfluid is flowing through themeter, asmallpropellericonishighlighted.NOTE:Totalizationcountstotalunitswithout

differentiatingbetweengallons,litresorfieldcalibratedunits.

Flowrate Feature (some models)Tousethisfeature,pressandreleaseDISPLAYuntilFLOWRATEiconappears.Thefactory

settimebasewillbehighlightedtotherightof FLOWRATE (M=minutes,H = hours,D=days).WhenFLOWRATEisinvoked,thedisplaywillbeindicatingrateofflow.

Factory and Field CalibrationAll calibration information is visible to theuseras iconson the top lineof thedisplay,abovethenumericdigits.All units are configured with a “factory”calibration.Bothgallonsandlitresareavail-able(“GL”or“LT”willbedisplayed).WhileholdingtheCALIBRATEbutton,brieflypressDISPLAYtotogglebetweengallonsandlitres.Thisfactorycalibration(indicatedwithFAC)ispermanentlyprogrammedintothecomputerandisnotuseradjustable.NOTE:Yourcomputermayhaveotherunitsof

measureprogrammedintoit.Ifso,holdingtheCALIBRATEbuttonandmomentarilypressingtheDISPLAYbuttonwilltogglethroughallfactorysetunits.Otherpos-sibleunitsare:IGL(imperialgallon),QT(quart),CF(cubicfeet),CM(cubicmeter),BL(42gal.barrel),CC(cubiccentimeter)orOZ(ounce).

Switching between different units will notcorrupttheTotal’scontents.Forexample,inGLmode,thecomputertotalizes10.00gallons,iftheuserswitchestoLTmode,thedisplaywillread37.85litres(thesamevolume,dif-ferentunit).The“field”calibrationmaybesetbytheuser,andcanbechangedormodifiedatanytimeusingthecalibrationproceduredescribedinthe Calibration Section. Totals or flowratederivedfromthefieldcalibrationareinvokedwhentheFACiconisnolongervisibleonthetoplineofthedisplay.

Verify Accuracy BeforeBeginning Field CalibrationFor themost accurate results,dispenseat aflowratewhichbestsimulatesyouractualoper-atingconditions.Avoid“dribbling”morefluidorrepeatedlystartingandstoppingtheflow.Thiscanresultinlessaccuratecalibrations.

Page 5: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

5

2. The +/– position appears either as an“underscore” character for plus, or asa “hyphen” character for minus. TheDISPLAY button selects the positionandtheCALIBRATEbuttontogglesthischaracter.

3. TheDISPLAYbuttoncanthenbepushedtoselectthenumericpositions.PresstheCALIBRATEbuttontoscrollfrom0to9. Enter the percentage of change youwantthedisplaytocorrect.Whensatisfiedwiththevalue,pressbothCALIBRATEand DISPLAY buttons simultaneously.CALEndwillbedisplayedandunitwillgobacktonormaloperation,lesstheFAC(factorycalibration)icon.

4. Allenabledunits-of-measureremainvis-ibleandselectable–theenteredcorrectionwillbeappliedtoallenabledunits.

5. To return to factory calibration (FAC),press and hold both CALIBRATE andDISPLAY buttons for about 3 secondsuntilFAcCALisdisplayed.Thenreleasebuttons. Unit should return to normaloperationandFACiconisvisible.

The “09” seriesEDM display has been pro-grammedwithmanynew features,most ofwhichcanbeenabledbytheenduserbywayofaconfigurationprocess.Bydisabling“un-necessary”features,day-to-dayflowmeterop-erationcanbegreatlysimplified,makingtheuniteasiertouse.Thereareseveralfeaturesthatare disabled bydefaultwhenshippingstandardmeters.(Forexample,K-FactorEntryFieldCalibration,describedbelow.)Formoreadvancedusers,itmaybedesirabletoenableALL possible features.User configurablefeaturesinclude:• Totalizers/Modes Enabled (CumulativeTotal,Batch2Total,FlowrateMode)

• FlowrateTimebase (Units perMinutes,HoursandDays)

Make sure youmeet themeter’sminimumflowraterequirements: 1/2inchmeter 1GPM(3.8LPM) 3/4inchmeter 2GPM(7.6LPM) 1inchmeter 5GPM(18.9LPM) 1-1/2inchmeter 10GPM(37.9LPM) 2inchmeter 20GPM(75.7LPM)Theuseofauniformlydependable,accuratecalibrationcontainerisrecommendedforthemost accurate results. For best results, themeter shouldbe installedandpurgedofairbeforefieldcalibration.Duetohighflowrateonthe2inchmeter,itisstronglyrecommendedthatFieldCalibrationbecompletedwithacombinationofvolumeandweight determinedwith fine resolutionscales.

Field Calibration NecessityField Calibration and Factory Calibrationare defined in the Operation Section. Fac-torycalibrationsettingsareprogrammedintoeachcomputerduringmanufacturing,usingstoddardtestsolventat70°F(21°C)forlowflow and 1-inchmeters andwater at 70° F(21°C)for2-inchmeters.Settingsarecorrectfor light liquids such as water, gasoline ordiesel.ReadingsusingtheFactoryCalibration(FAC)maynotbeaccurateinsomesituations,forexample,“heavy”liquidssuchasmotoroil, under extreme temperature conditions,non-standardplumbingconfigurationsorwithfluidsotherthanmentionedabove.Forimprovedaccuracyundersuchconditions,thecomputerallowsfor“field”calibration,thatis,userentryofcustomcalibrationparameters.A“singlepoint”calibrationmayyieldaccept-able accuracywhenused in anon-standardapplication.

Field Calibration Procedures(Correction Factor Method)1. To calibrate, press andhold theCALI-

BRATEandDISPLAYbuttonsforabout3secondsuntilyouseeFLdCAL.ReleasebothbuttonsandyouwillseeCF-00.0.Youarenowinthefieldcalibrationmodeandvaluesfrom-99.9%to+99.9%canbeentered.

USER CONFIGURATION

Page 6: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

6

• FactoryCalibrationCurveUnitsEnabled(Gallons,ImperialGallons,Litres,Quarts,Ounces,CubicFeet,CubicCentimeters,CubicMetersorBarrels(42gal.)

• Dispense/DisplayorK-FactorEntryCali-bration

Changing ConfigurationSettingsAccesstotheconfigurationprocessisrestrict-edforsecurityuntila“password”isentered.Contact your distributor orKOBOLD to get thepasswordandinstructionstounlockandresetconfigurationsettings.ThisinformationisalsoavailableontheWebsite.All configurationsare entered and stored as six-digit “codes”whereeachdigitrepresentsasettingforoneoftheconfigurationoptions.Newconfigurationsettingsarestoredinthecomputer’slong-termmemory andwill not be lost either inOFFmodeorduringbatterychange.

K-Factor Entry FieldCalibrationPresentlyallEDMcomputersareprogrammedwiththreedifferentfieldcalibrationmethods,onlyoneofwhichisactive,the“correction/factor”calibrationproceduredescribedabove.It is possible to activate “K-Factor entry”or “dispense/display” field calibration bychangingconfigurationsettings.ContactyourdistributororKOBOLDtogetthecorrectpassword,configurationcode,andinstructionsfor thiscalibrationmethod.ThisinformationisalsoavailableontheWebsite.

Thecomputerelectronicsispoweredbylith-iumbatteries.Removingthebatteriesbeforestoringthemeterwillextendbatterylife.Ifthemeter’sreadoutshouldbecomedim,blankorthelowbatterymessageappears(seebelow),thebatteriesshouldbereplaced.Replacementbatteriescanbeorderedfromyourdistributororthefactory.SeedetailsinthePartsSection.

Whenbatteriesaredisconnectedorfail, theBatchandCumulativeTotalswillmaintainthevaluetheyhad.FactoryandFieldCalibrationCurvesareretainedinthemeter’scomputerwhenpowerislost.Itisstronglyrecommendedthatbatterycheckandterminalcleaningbeapartofaroutinemaintenance schedule. Battery terminalsshouldbecleanedannually.Batteriescanbereplacedwithoutremovingthemeterfromthepipingsystem.

Replace Batteries1. Removethecornerscrewsfromthemeter

face and lift the computer electronicsfromtheturbine.

2. Removethebatteries.3. Checkthebatteryterminalsandremove

anycorrosion.4. Install thenewbatteriesandmakesure

thepositivepostsarepositionedcorrectly.Whenthebatteriesareinstalledcorrectly,the computer powers on automaticallyandthereadoutdisplaysinformation.

5. Makesurethesealisfullyseatedbeforeplacingthecomputerelectronicsontheturbine.Tightenthefourscrews.

6. Do not clean exterior of computer as-semblywithIsopropylAlcohol.

MAINTENANCE

Page 7: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

7

Symptom Probable Cause Corrective ActionMeterisnot 1.FieldCalibrationnotper- FieldcalibrateagainorselectFactoryaccurate formedproperly Calibration. 2.FactoryCalibrationnot PerformaFieldCalibrationaccording suitableforliquidbeing toCalibrationSection. measured 3.Meteroperatedbelow Increaseflowrate. minimumflowrate 4.Meterpartiallyclogged Removemeter.Cleancarefully.Make withdriedliquid surerotorspinsfreely. 5.Turbinebearingspartially Removemeter.Cleancarefully.Make cloggedwithdriedliquid surerotorspinsfreely. 6.Sealantmaterialwrapped Removemeter.Makesurerotorspins aroundrotor freely. 7.Installedtoocloseto Installcorrectly. fittings 8.Installedtoocloseto Installcorrectly. motorsorelectrically “noisy”environment

Readoutfaded 1.Batteriesweak,dead,or Removecomputer,checkandreplaceorblank notconnected batteriesifnecessary. 2.Computerdefective Contactthefactory.

Normalflowrate 1.FieldCalibrationnot FieldCalibrateagainorselectFactorybutmeterdoes performedcorrectly Calibration.notcount 2.Rotorstuckordamaged Removemeter.Makesurerotorspins freely. 3.Sealantmaterialwrapped Removemeter.Makesurerotorspins aroundrotor freely. 4.Computerdefective Contactthefactory.

Reducedflowrate 1.Metercloggedwithdried Removemeter.Cleancarefully.Makeandmeterdoes liquids surerotorspinsfreely.notcount 2.Belowminimumflowrate Increaseflow.

Cannotgetmeter 1.Wrongbuttonsequence Proceedwithcalibrationaccordingtointofieldcalibra- theCalibrationSection.tion 2.Computercircuitboard Replacecomputer.Contactthefactory. defective 3.Buttondefective Replacecomputer.Contactthefactory.

TROUBLESHOOTING

Page 8: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

8

SPECIFICATIONS

Standard Features Include: • 2TotalizingRegisters • 1FactoryCalibrationCurve • 1FieldCalibrationCurve • RateofFlowFeature • FlowrateTimeBaseinMinutes

Input Pulse Rate: MinimumPulseIn: DC MinimumCoilInput: 10Hz MaximumRaw: 1,000Hz

K-Factor: Minimum: .01pulses/unit Maximum: >999,999pulses/ unit

Field Calibration Correction: Minimum: -99.9% Maximum: +99.9%

Readout Totals: MinimumDisplay: 0.01 MaximumDisplay: 999,999(x100)

Temperatures: Operational: +0°to+140°F (-18°to+60°C) Storage: -40°to+158°F (-40°to+70°C) Ifwideroperatingtemperatureranges aredesired,referenceinformationon EDMRemoteKits.

Power: InternalPowerSupply: 2Lithium Batteriesat 3voltseach BatteryLife: 5years OptionalExternal PowerModule: 7-30VDC

J-1 Reset Programminginterfaces.Notacces-

sibletouser.

J-2 Pulse Signal Output Thissuppliesahigh-levelamplified

open collector signal.Outputwillwithstand amaximumopen-circuitvoltageof60voltsDCandamaxi-mumclosed-circuitof100mA.

J-4 Pulse Signal Input Requiresasineorsquarewavewith

open-circuit voltage of 3-30 voltsP-P, amaximum rise/fall rate of0.01V/µ second and amaximumfrequencyof750Hz.

J-5 Power Input When used with Ground (J1-6),

thishasreversepolarityprotection,butnoon-boardvoltageregulation.Suppliedvoltagemaybe5voltsto10voltsDC.

J-6 Ground

J-7, 8, Programminginterfaces.Notacces-sibletouser.

9, 10

NOTE:Safety approvals are void if anyexternal connections aremade tocomputerelectronics.

J 1

J 10

Computer Electronics Terminal Connections

Page 9: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

9

PARTS

Thefactory,whenprovidedwithmodelnum-berandserialnumber,canreplaceyourentireComputerElectronicsAssembly.Orderreplacementkits,parts,andaccessorieswiththepartnumbersgivenhere.

Part No. Description

EDM-P113520-1 BatteryReplacementKit

Forwarranty consideration, parts, or otherservice information,pleasecontactyour lo-caldistributor.Ifyouneedfurtherassistance,call theCustomerServiceDepartment in Pittsburgh, PA,duringnormalbusinesshours.

1-412-788-2830

Toobtain prompt, efficient service, alwaysbepreparedwiththefollowinginformation:1. Themodel number of your computer

electronics.2. Theserialnumberormanufacturingdate

codeofyourcomputerelectronics.3. Specificinformationaboutpartnumbers

anddescriptions.Forwarrantyworkalwaysbepreparedwithyouroriginalsalessliporotherevidenceofpurchasedate.

Returning PartsPlease contact the factory before returningany parts. Itmay be possible to diagnosethe trouble and identify needed parts in atelephone call.We can also informyouofany special handling requirements youwillneedtofollowcoveringthetransportationandhandlingofequipmentwhichhasbeenusedtotransferhazardousorflammableliquids.

SERVICEWEEE DIRECTIVE

The Waste Electrical andElectronic Equipment (WEEE)directive (2002/96/EC) wasapproved by the EuropeanParliament and the Council ofthe European Union in 2003.This symbol indicates that thisproduct contains electrical andelectronic equipment that may

include batteries, printed circuit boards, liquid crystaldisplays or other components that may be subjectto local disposal regulations at your location. Pleaseunderstand those regulations and dispose of thisproduct in a responsible manner.

Do not return computer electronics or meters without specific authority from the Customer Service Department. Due to strict regulations governing transportation, handling, and disposal of hazardous or flammable liquids, we will not accept computer electronics or meters for rework unless they are completely free of liquid residue.

Meters not flushed before shipment can be refused and returned to the sender.

CAUTION

CAUTION

Page 10: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

Rev. A 920739-1

04/05

Kobold Instruments, Inc.1801 Parkway View DrivePittsburgh, PA 15205Phone: 412/788-2830Fax: 412/788-4890www.koboldusa.com

Industrial GradeTURBINE HOUSINGOwner’s ManualIncludes Aluminum, Brass and Stainless Steel Housings andStainless Steel Housings with ANSI Flange Connections

For FM Approved Units: Look on the Computer Electronics for the FM model number.

Page 11: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

GENERAL INFORMATION

This manual will assist you in installing andmaintaining your turbine housing. (See Fig-ure 1) Information on computer electronicsand accessory modules are contained in othermanuals. Please reference those as necessary.

For best results, take the time to fully ac-quaint yourself with all information about allcomponents of your Electronic Digital Me-tering System prior to installation and use. Ifyou need assistance, contact the distributorfrom whom you purchased your turbine.

This symbol is used throughout themanual to call your attention to safetymessages.

Warnings alert you to the potential for per-sonal injury.

Cautions call your attention to practices orprocedures which may damage yourequipment.

Notes give information that can improve ef-ficiency of operations.

It is your responsibility to make sure all opera-tors have access to adequate instructions aboutsafe operating and maintenance procedures.

2

Read Me!For your safety, review the major warningsand cautions below before operating yourequipment.

1. This equipment is approved to handleonly fluids which are compatible with thehousing material. Use only fluids that arecompatible with the housing material andthe wetted components of your turbine.

2. When measuring flammable liquids, ob-serve precautions against fire or explo-sion.

3. When handling hazardous liquids, alwaysfollow the liquid manufacturer’s safetyprecautions.

4. When working in hazardous environ-ments, always exercise appropriate safetyprecautions.

5. Always dispose of used cleaning solventsin a safe manner according to the solventmanufacturer’s instructions.

6. During turbine removal, liquid may spill.Follow the liquid manufacturer’s safetyprecautions for clean up of minor spills.

7. Do not blow compressed air through theturbine.

8. Do not allow liquids to dry inside the tur-bine.

9. Handle the rotor carefully. Even smallscratches or nicks can affect accuracy.

10. When tightening the turbine, use a wrenchonly on the wrench flats.

11. For best results, always verify accuracybefore use.

Product DescriptionThese Industrial Meter Turbines are identifiedby the internal diameter of the inlet and outlet.

EDM-4__01 - 1/2 inch (Mid Flow)EDM-4__02 - 3/4 inch (Mid Flow)EDM-4__03 - 1 inch (Mid Flow)EDM-4__04 - 1-1/2 inch (High Flow)EDM-4__05 - 2 inch (High Flow)

NOTE: See page 6 for complete model number.

Each of these turbines are designed to workwith on-board computer electronics and/orwith one of several accessory modules thatcan interface to a wide variety of reportingand collecting devices.

General Information ................................... 2Installation .................................................. 3Maintenance ............................................... 4Troubleshooting ......................................... 6Catalog Number Chart ............................... 6Specifications ............................................. 7Parts .......................................................... 15Service ...................................................... 15

TABLE OF CONTENTS

Figure 1

Turbine Housing

Computer Electronics(sold separately)

Page 12: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

Liquid flows through the turbine housing caus-ing an internal rotor to spin. As the rotor spins,an electrical signal is generated in the pickupcoil. The electrical signal provides the outputnecessary to operate the on-board computerelectronics for local indication directly on theturbine or one of several accessory modulesthat transmit the signal to external equipment.

Upon receipt, examine your meter for visibledamage. The turbine is a precision measuringinstrument and should be handled as such.Remove the protective plugs and caps for athorough inspection. If any items are damagedor missing, contact your distributor.

Make sure the turbine model meets your spe-cific needs. Refer to the Specifications Sec-tion and confirm the following:

1. The flowrate is within the limits of yourmodel.

2. The liquid is compatible with the turbine’swetted components.

3. The system’s pressure does not exceed theturbine’s maximum pressure rating.

Information specific to your particular turbine,including serial number, model number,manufacturing date, and K-factor is found onthe surface of the turbine.

SN = Serial Number, a 6-digit number thatidentifies this particular turbine.

MODEL = Model Number begins with aletter indicating the housing material.

A for AluminumB for BrassS for Stainless Steel

Two digits follow the material code indicat-ing the size.

05 = 1/2 inch 15 = 1-1/2 inch07 = 3/4 inch 20 = 2 inch10 = 1 inch

The final letter indicates the type of thread.F for FlangeN for NPTI for ISO

MFG DATE = Manufacturing Date indi-cating the week and year of manufacture.

KF = K-Factor given in pulses per gallon(PPG).

For your future reference, it might be usefulto record this information in the manual in caseit becomes unreadable on the turbine.

3

Turbines are designed to measure flow in onlyone direction. The direction is indicated bythe arrow cast-molded in the turbine outlet. Ifthe opposite direction is desired, and you areusing on-board computer electronics, rotatethe computer electronics 180 degrees prior toinstallation.

Flow altering devices such as elbows, valves,and reducers can affect accuracy. The follow-ing guidelines are given to enhance accuracyand maximize performance. Distances givenhere are minimum requirements; double themfor desired straight pipe lengths.

Upstream from the turbine, allow a minimumstraight pipe length at least 10 times the inter-nal diameter of the turbine. For example, withthe 1 inch turbine, there should be 10 inches(25.4cm) of straight pipe immediately up-stream. The desired upstream straight pipelength is 20 inches (50.8cm).

Downstream from the turbine, allow a mini-mum straight pipe length at least 5 times theinternal diameter of your turbine. For example,with the 1 inch turbine, there should be 5inches (12.7cm) of straight pipe immediatelydownstream. The desired downstream dis-tance is 10 inches (25.4cm).

A typical back pressure of 5 to 50 PSI (0.34to 3.4 bar) will prevent cavitation. Create backpressure by installing a control valve on thedownstream side of the meter at the properdistance detailed above.

Foreign material in the liquid being measuredcan clog the turbine’s rotor and adversely af-fect accuracy. If this problem is anticipated orexperienced, install screens to filter impuri-ties from incoming liquids.

1/2 inch, 3/4 inch and 1 inch TurbinesMaximum Particulate Size

Inches: 0.005 Microns: 125Mesh: 55Standard Sieve: 125 µmAlternative Sieve: No. 120

1-1/2 inch and 2 inch TurbinesMaximum Particulate Size

Inches: 0.018 Microns: 500Mesh: 28Standard Sieve: 500 µmAlternative Sieve: No. 35

INSTALLATION

Page 13: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

4

These turbines are tested and calibrated atthe factory using state-of-the-art calibrationprocedures and test equipment.

To ensure accurate measurement, remove allair from the system before use. To purge thesystem of air:

1. Ensure some back pressure exists on theturbine in the line.

2. Open the discharge valve or nozzle andallow fluid to completely fill the system.Make sure the stream is full and steady.

3. Close the discharge valve or nozzle.

4. Start normal operations.

Each turbine contains a removable backcoverplate. Leave the coverplate installedunless accessory modules specify removal.

Connections1. To protect against leakage, seal all

threads with an appropriate sealing com-pound. Make sure the sealing compounddoes not intrude into the flow path.

2. Make sure the arrow on the outlet ispointed in the direction of the flow.

3. Tighten the turbine onto the fittings. Usea wrench only on wrench flats.

NOTE: If connecting to new male threads,burrs and curls can adversely effect ac-curacy. Correct the problem prior to tur-bine installation.

It is strongly recommended that accuracy beverified prior to use.

Flange ConnectionUse a gasket between the meter flange andmating flange. Determine the material of thegasket based on the operating conditions andtype of fluid.

NOTE: Do not over tighten the flange bolts.This may cause the gasket to be com-pressed into the flow stream and maydecrease the accuracy of the meter.

Verify AccuracyBefore use, check the turbine’s accuracy andverify calibration.

1. Make sure there is no air in the system.

2. Measure an exact known volume into anaccurate container.

3. Verify the volume against the readout orrecording equipment.

NOTE: If necessary, use a correction factorto figure final volume.

For best results, accuracy should be verifiedperiodically as part of a routine maintenanceschedule.

Remove the Turbine

1. Drain all liquid from the turbine. Wearprotective clothing as necessary.

2. Loosen both ends of the turbine. Use awrench only on the turbine’s wrenchflats.

3. If the turbine is not immediately installedagain, cap lines as necessary.

Replace Internal Parts

1. Remove the turbine from the system asdetailed above.

NOTE: Carefully notice the orientation of allinternal parts as they are removed, espe-cially the orientation of the rotor to theflow direction arrow. See Figure 2.

SupportRotorSupport

Retaining Ring Retaining Ring

Figure 2

! ! ! WARNING ! ! !During turbine removal, liquid may spill.Follow the liquid manufacturer’s safetyprecautions for clean up of minor spills.

MAINTENANCE

Page 14: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

5

2. Using a small tool such as a screwdriveror awl, gently pry one retaining ring fromits groove. Remove the support. If nec-essary, use needle nose pliers. Little orno force should be required.

3. Carefully remove the rotor.

CAUTION: Handle the rotor carefully.Even small scratches or nicks can affectaccuracy.

4. Turn the turbine over and remove theother retaining ring. Remove the othersupport.

5. Clean, as detailed below, or discard asnecessary.

6. Replace one support and retaining ring.Parts should drop easily into place withlittle or no force.

7. Install the rotor. Make sure the wide endof the rotor’s blades faces the flow di-rection. See Figure 3.

8. Turn the turbine over and drop the sec-ond support into place. Put the final re-taining ring into position.

9. Reinstall the turbine, purge the systemof air, and verify accuracy before use.

Clean the TurbineDuring use, the turbine should be kept fullof liquid to ensure that drying does not occurinside the turbine. If drying or caking shouldoccur, the rotor will stick or drag, affecting ac-curacy. To determine if the rotor is stuck or drag-ging, gently blow air through the meter andlisten for the quiet whir of the rotor.

CAUTION: Never blow compressed airthrough the meter. It could damage therotor.

1. Remove the turbine from the system fol-lowing the directions above.

2. Carefully clean residue off all parts. Re-move internal parts as detailed above.Note orientation carefully for correct as-sembly. Internal parts can be soaked for10 to 15 minutes in compatible cleaningsolutions. Use a soft brush or small probeto carefully remove residue from the ro-tor.

3. When the rotor turns freely, assembleand install it again following the instruc-tions above.

Flow Direction Arrow

Figure 3Wide End of Rotor Blade

! ! ! WARNING ! ! !Follow the liquid manufacturer’s instruc-tions for the disposal of contaminatedcleaning solvents.

Page 15: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

6

TROUBLESHOOTING

Symptom Probable Cause SolutionMeasurement is 1. Turbine operated below Increase flowrate. See Specifica-not accurate. minimum rate. tions.

2. Turbine partially clogged Remove turbine. Clean carefully.with dried liquid. Make sure rotor spins freely.

3. Turbine bearings partially Remove turbine. Clean carefully.clogged with dried liquid. Make sure rotor spins freely.

4. Sealant wrapped around Remove turbine. Clear material fromrotor. rotor. Make sure rotor spins freely.

5. Installed too close to Install correctly. See Installationfittings. Section.

6. Improper connections to Check all electrical connections.recording device. Reference appropriate installation

instructions.

7. Accuracy needs verification. Complete normal accuracy verifica-tion procedures. Repeat periodically.

8. Flange bolts too tight, Replace gasket with less torquecausing gasket to compress on bolts.into flow stream (FlangeMeters only).

Normal Range Normal Range Fitting StainlessGPM Water LPM Water Size NPT Aluminum Brass Steel

1 - 10 3.8 - 37.9 1/2" EDM-4301 EDM-4701 EDM-4201

2 - 20 7.6 - 75.7 3/4" EDM-4302 EDM-4702 EDM-4202

5 - 50 18.9 - 190 1" EDM-4303 EDM-4703 EDM-4203

10 - 100 38 - 380 1-1/2" EDM-4304 EDM-4704 EDM-4204

20 - 200 76 - 760 12" EDM-4305 EDM-4705 EDM-4205

Options

Description Suffix

Add-on remote mounting kit for displays (See Note 1) -RD

Add-on pulse frequency transmitter kit (See Note 1) -PM

Add-on pulse frequency transmitter kit in place of display (See Note 1) -SC

Add-on 4-20 mA analog transmitter kit (See Note 1) -WM

ISO threads -B

Intrinsically safe remote mounting kit for display -ERD

Intrinsically safe pulse frequency transmitter -EX

150 lb. ANSI Flange connection for stainless steel flowmeters only (EDM-4203, -04, 05) -F

NOTE 1: Use of this option invalidates the intrinsic safety rating of the flowmeter.

CATALOG NUMBER CHART

Page 16: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

7

SPECIFICATIONS – ALUMINUM

All data on 1/2", 3/4", and 1" turbines determined with 1 centipoise Kermac solvent testfluid at 70°F (21°C). Data on 1-1/2" and 2" turbines is determined with water at 70°F (21°C).

Catalog Number EDM-4301 EDM-4302 EDM-4303 EDM-4304 EDM-4305Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.

Linear Flow RangeGallons/minute (GPM) 1-10 2-20 5-50 10-100 20-200Litres/minute (LPM) 3.8-37.9 7.6-75.7 18.9-190 38-380 76-760

Maximum Flow 1Gallons/minute (GPM) 15 30 75 150 300Litres/minute (LPM) 56.8 113.6 284 568 1,136

Maximum Pressure Dropin 10:1 Range

PSIG 8 7.5 5 4 4bar 0.55 0.5 0.34 0.28 0.28

Frequency Range inLinear Flow Range 45-450 Hz 37-370 Hz 45-475 Hz 35-350 Hz 33-330 Hz

ConnectionsNPT or ISO Threads Yes Yes Yes Yes YesFemale Yes Yes Yes Yes YesInlet/Outlet Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.Wrench Size: Inch 1-1/16 in. 1-5/16 in. 1-5/8 in. 2-3/8 in. 3 in. Millimeter 27mm 33mm 41mm 60mm 75mm

Weight*Pounds 0.6 lbs. 0.7 lbs. 0.8 lbs. 2.0 lbs. 3.1 lbs.Kilograms 0.8kg 1.0kg 1.1kg 1.8kg 2.9kg

* Computer electronics add 0.2 lbs. (0.1kg) to total weight.1 The meter can operate up to this flowrate without damage. Continuous operation will severely

degrade meter life and performance.

Page 17: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

8

Specifications – Aluminum (cont’d.)

PerformanceLinear Range for 1/2 in. 10:1 @ ±2.0% of readingLinear Range for 3/4 in. and 1 in.: 10:1 @ ±1.5% of readingLinear Range for 1-1/2 in. and 2 in.: 10:1 @ ±1.0% of readingRepeatability: ±0.1%

Pressure Rating 300 PSIG (21 bar)

Wetted ComponentsHousing: AluminumJournal Bearings: Ceramic (96% Alumina)Shaft: Tungsten CarbideRotor and Supports: PVDFRetaining Rings: 316 Stainless Steel

Temperature Range -40° to +250°F (-40° to +121°C)These temperatures apply to operations and storage. They are only for the turbine with-out computer electronics. Final operational temperature range is determined by com-puter electronics or accessory modules.

Catalog Number EDM-4301 EDM-4302 EDM-4303 EDM-4304 EDM-4305Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.A = Height:

Inches 1.8 in. 2.0 in. 2.2 in. 2.8 in. 3.2 in.Centimeters 4.6cm 5.1cm 5.6cm 7.1cm 8.2cm

B = WidthInches 2.0 in. 2.0 in. 2.0 in. 2.7 in. 3.3 in.Centimeters 5.1cm 5.1cm 5.1cm 6.9cm 8.4cm

C = LengthInches 4.2 in. 4.3 in. 4.5 in. 5.3 in. 6.3 in.Centimeters 10.7cm 10.9cm 11.4cm 13.5cm 16.0cm

Computer electronics add 0.7 in. (1.8cm) to height of turbine.

Dimensions

A

➤ B ➤

➤ C

Page 18: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

9

SPECIFICATIONS – Brass

All data on 1/2", 3/4", and 1" turbines determined with 1 centipoise Kermac solvent testfluid at 70°F (21°C). Data on 1-1/2" and 2" turbines is determined with water at 70°F (21°C).

Catalog Number EDM-4701 EDM-4702 EDM-4703 EDM-4704 EDM-4705Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.

Linear Flow RangeGallons/minute (GPM) 1-10 2-20 5-50 10-100 20-200Litres/minute (LPM) 3.8-37.9 7.6-75.7 18.9-190 38-380 76-760

Maximum Flow 1Gallons/minute (GPM) 15 30 75 150 300Litres/minute (LPM) 56.8 113.6 284 568 1,136

Maximum Pressure Dropin 10:1 Range

PSIG 8 7.5 5 4 4bar 0.55 0.5 0.34 0.28 0.28

Frequency Range inLinear Flow Range 45-450 Hz 37-370 Hz 45-475 Hz 35-350 Hz 33-330 Hz

ConnectionsNPT or ISO Threads Yes Yes Yes Yes YesFemale Yes Yes Yes Yes YesInlet/Outlet Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.Wrench Size: Inch 1-1/16 in. 1-5/16 in. 1-5/8 in. 2-3/8 in. 3 in. Millimeter 27mm 33mm 41mm 60mm 75mm

Weight*Pounds 2.0 lbs. 2.3 lbs. 2.7 lbs. 6.0 lbs. 9.6 lbs.Kilograms 0.9kg 1.0kg 1.2kg 2.7kg 4.3kg

* Computer electronics add 0.2 lbs. (0.1kg) to total weight.1 The meter can operate up to this flowrate without damage. Continuous operation will severely

degrade meter life and performance.

Page 19: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

10

Specifications – Brass (cont’d.)

PerformanceLinear Range for 1/2 in. 10:1 @ ±2.0% of readingLinear Range for 3/4 in. and 1 in.: 10:1 @ ±1.5% of readingLinear Range for 1-1/2 in. and 2 in.: 10:1 @ ±1.0% of readingRepeatability: ±0.1%

Pressure Rating 300 PSIG (21 bar)

Wetted ComponentsHousing: BrassJournal Bearings: Ceramic (96% Alumina)Shaft: Tungsten CarbideRotor and Supports: PVDFRetaining Rings: 316 Stainless Steel

Temperature Range -40° to +250°F (-40° to +121°C)These temperatures apply to operations and storage. They are only for the turbine with-out computer electronics. Final operational temperature range is determined by com-puter electronics or accessory modules.

Catalog Number EDM-4701 EDM-4702 EDM-4703 EDM-4704 EDM-4705Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.A = Height:

Inches 1.8 in. 2.0 in. 2.2 in. 2.8 in. 3.2 in.Centimeters 4.6cm 5.1cm 5.6cm 7.1cm 8.2cm

B = WidthInches 2.0 in. 2.0 in. 2.0 in. 2.7 in. 3.3 in.Centimeters 5.1cm 5.1cm 5.1cm 6.9cm 8.4cm

C = LengthInches 4.2 in. 4.3 in. 4.5 in. 5.3 in. 6.3 in.Centimeters 10.7cm 10.9cm 11.4cm 13.5cm 16.0cm

Computer electronics add 0.7 in. (1.8cm) to height of turbine.

Dimensions

A

➤ B ➤

➤ C

Page 20: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

11

All data on 1/2", 3/4" and 1" turbines determined with 1 centipoise stoddard solvent testfluid at 70°F (21°C). Data on 1-1/2" and 2" turbines is determined with water at 70°F (21°C).

Catalog Number EDM-4201 EDM-4202 EDM-4203 EDM-4204 EDM-4205Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.

Linear Flow RangeGallons/minute (GPM) 1-10 2-20 5-50 10-100 20-200Liters/minute (LPM) 3.8-37.9 7.6-75.7 18.9-190 38-380 76-760

Maximum Flow1

Gallons/minute (GPM) 15 30 75 150 300Liters/minute (LPM) 56.8 113.6 284 568 1,136

Maximum Pressure Dropin 10:1 Range

PSIG 8 7.5 5 4 4bar 0.55 0.5 0.34 0.28 0.28

Frequency Range inLinear Flow Range 45-450 Hz 37-370 Hz 45-475 Hz 35-350 Hz 33-330 Hz

ConnectionsNPT or ISO Threads Yes Yes Yes Yes YesFemale Yes Yes Yes Yes YesInlet/Outlet Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.Wrench Size: Inch 1-1/16 in. 1-5/16 in. 1-5/8 in. 2-3/8 in. 3 in. Millimeter 27mm 33mm 41mm 60mm 75mm

Weight*Pounds 1.8 lbs. 2.0 lbs. 2.4 lbs. 4.0 lbs. 6.3 lbs.Kilograms 0.8kg 1.0kg 1.1kg 1.8kg 2.9kg

* Computer electronics add 0.2 lbs. (0.1kg) to total weight.1 The meter can operate up to this flowrate without damage. Continuous operation will severly degrade meter life and performance.

SPECIFICATIONS – Stainless Steel

Page 21: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

12

Specifications – Stainless Steel (cont’d.)Performance

Linear Range for 1/2 in. 10:1 @ ±2.0% of readingLinear Range for 3/4 in. and 1 in.: 10:1 @ ±1.5% of readingLinear Range for 1-1/2 in. and 2 in.: 10:1 @ ±1.0% of readingRepeatability: ±0.1%

Pressure Rating 1,500 PSIG (103 bar)

Wetted ComponentsHousing: 316 Stainless SteelJournal Bearings: Ceramic (96% Alumina)Shaft: Tungsten CarbideRotor and Supports: PVDFRetaining Rings: 316 Stainless Steel

Temperature Range -40° to +250°F (-40° to +121°C)These temperatures apply to operations and storage. They are only for the turbine with-out computer electronics. Final operational temperature range is determined by com-puter electronics or accessory modules.

Catalog Number EDM-4201 EDM-4202 EDM-4203 EDM-4204 EDM-4205Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.A = Height:

Inches 1.8 in. 2.0 in. 2.2 in. 2.8 in. 3.2 in.Centimeters 4.6cm 5.1cm 5.6cm 7.1cm 8.2cm

B = WidthInches 2.0 in. 2.0 in. 2.0 in. 2.7 in. 3.3 in.Centimeters 5.1cm 5.1cm 5.1cm 6.9cm 8.4cm

C = LengthInches 4.2 in. 4.3 in. 4.5 in. 5.3 in. 6.3 in.Centimeters 10.7cm 10.9cm 11.4cm 13.5cm 16.0cm

Computer electronics add 0.7 in. (1.8cm) to height of turbine.

Dimensions

A

B C

Page 22: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

13

SPECIFICATIONS – Flange Fittings

All data on 1" turbine determined with 1 centipoise stoddard solvent test fluid at 70°F (21°C).Data on 1-1/2" and 2" turbines is determined with water at 70°F (21°C).

Catalog Number EDM-4203-F EDM-4204-F EDM-4205-FSize 1 in. 1-1/2 in. 2 in.Linear Flow Range

Gallons/minute (GPM) 5-50 10-100 20-200Litres/minute (LPM) 18.9-190 38-380 76-760

Maximum Flow 1Gallons/minute (GPM) 75 150 300Litres/minute (LPM) 284 568 1,136

Maximum Pressure Dropin 10:1 Range

PSIG 5 4 4bar 0.34 0.28 0.28

Frequency Range inLinear Flow Range 45-475 Hz 35-350 Hz 33-330 Hz

ConnectionsANSI 150 lb. Flange Yes Yes YesBolt Size 1/2 in. 1/2 in. 5/8 in.

Weight*Pounds 6.5 lbs. 10.6 lbs. 17.9 lbs.Kilograms 2.9kg 4.8kg 8.1kg

* Computer electronics add 0.2 lbs. (0.1kg) to total weight.

1 The meter can operate up to this flowrate without damage. Contin-uous operation will severely degrade meter life and performance.

Page 23: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

14

Specifications – Flange Fittings (cont’d.)

PerformanceLinear Range for 1 in.: 10:1 @ ±1.5% of readingLinear Range for 1-1/2 in. and 2 in.: 10:1 @ ±1.0% of readingRepeatability: ±0.1%

Pressure Rating 285 PSI (From –20° to 100°F)

Wetted ComponentsHousing: 316 Stainless SteelJournal Bearings: Ceramic (96% Alumina)Shaft: Tungsten CarbideRotor and Supports: PVDFRetaining Rings: 316 Stainless Steel

Temperature Range -40° to +250°F (-40° to +121°C)These temperatures apply to operations and storage. They are only for the turbine with-out computer electronics. Final operational temperature range is determined by com-puter electronics or accessory modules.

Catalog Number EDM-4203-F EDM-4204-F EDM-4205-FSize 1 in. 1-1/2 in. 2 in.A = Length:

Inches 6.75 in. 8.00 in. 9.50 in.Centimeters 17.14cm 20.32cm 24.13cm

B = WidthInches 4.25 in. 5.00 in. 6.00 in.Centimeters 10.80cm 12.71cm 15.24cm

Computer electronics add 0.7 in. (1.8cm) to height of turbine.

Dimensions

B

➤ A

Page 24: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

PARTS

15

Order Replacement Kits with the part numbers given here.

Kobold PartIdentifier Number Description

EDM-P 901002-52 O-RingEDM-P 125500-1 1/2 inch Rotor/Support Replacement KitEDM-P 125500-2 3/4 inch Rotor/Support Replacement KitEDM-P 125500-3 1 inch Rotor/Support Replacement KitEDM-P 125500-4 1-1/2 inch Rotor/Support Replacement KitEDM-P 125500-5 2 inch Rotor/Support Replacement KitEDM-P 125504-01 Gasket Kit (1 inch Viton)EDM-P 125505-01 Gasket Kit (1 inch Neoprene)EDM-P 125504-02 Gasket Kit (1-1/2 inch Viton)EDM-P 125505-02 Gasket Kit (1-1/2 inch Neoprene)EDM-P 125504-03 Gasket Kit (2 inch Viton)EDM-P 125505-03 Gasket Kit (2 inch Neoprene)EDM-P 904005-20 One 1/2 inch Retaining RingEDM-P 904005-21 One 3/4 inch Retaining RingEDM-P 904005-22 One 1 inch Retaining RingEDM-P 904005-23 One 1-1/2 inch Retaining RingEDM-P 904005-24 One 2 inch Retaining Ring

SERVICE

For warranty consideration, parts, or otherservice information, please contact your lo-cal distributor. If you need further assistance,contact the Kobold Instruments Customer Ser-vice Department during normal business hours.

Fax: 412-788-4890

To obtain prompt, efficient service, alwaysbe prepared with the following information:

1. The model number of your turbine.

2. The serial number or manufacturing datecode of your turbine.

3. Specific information about part numbersand descriptions.

For warranty work always be prepared withyour original sales slip or other evidence ofpurchase date.

Returning PartsPlease contact the factory before returningany parts. It may be possible to diagnose thetrouble and identify needed parts in a tele-phone call. The factory can also inform youof any special handling requirements you willneed to follow covering the transportationand handling of equipment which has beenused to transfer hazardous or flammable liq-uids.

CAUTION: Do not return turbines withoutspecific authority from the CustomerService Department. Due to strict regu-lations governing transportation, han-dling, and disposal of hazardous or flam-mable liquids, the factory will not acceptturbines for rework unless they are com-pletely free of liquid residue.

CAUTION: Turbines not flushed beforeshipment can be refused and returned tothe sender.

Page 25: EDM manual combined - KOBOLD · PDF file3 2. When measuring flammable liquids, observe precautions against fire or explo-sion. 3. When handling hazardous liquids, always follow the

Rev. A 920739-1

04/05

Kobold Instruments, Inc.1801 Parkway View DrivePittsburgh, PA 15205Phone: 412/788-2830Fax: 412/788-4890www.koboldusa.com