electrospraying/electrospinning equipment in 66-353 safety...

6
SAFETY SLIDES Electrospraying/Electrospinning equipment in 66-353 1

Upload: others

Post on 06-Jan-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Electrospraying/Electrospinning equipment in 66-353 SAFETY ...web.mit.edu/cohengroup/safety/electrospinning102010.pdfElectrospinning/s praying of other hazardous solvents in fume hood

SAFETY SLIDESElectrospraying/Electrospinning equipment in 66-353

1

Page 2: Electrospraying/Electrospinning equipment in 66-353 SAFETY ...web.mit.edu/cohengroup/safety/electrospinning102010.pdfElectrospinning/s praying of other hazardous solvents in fume hood

Overview of electrospinning apparatus

2

Page 3: Electrospraying/Electrospinning equipment in 66-353 SAFETY ...web.mit.edu/cohengroup/safety/electrospinning102010.pdfElectrospinning/s praying of other hazardous solvents in fume hood

Safety Slides – High voltage electrical equipment

3

Maximum voltage ~ 40,000 V

Low current discharge

Rubber Mat prevents any accidental shortingDo not open case while plugged!

Unplug in case of power loss

Page 4: Electrospraying/Electrospinning equipment in 66-353 SAFETY ...web.mit.edu/cohengroup/safety/electrospinning102010.pdfElectrospinning/s praying of other hazardous solvents in fume hood

Collecting plate

4

-Grounded conducting plate-Insulated from fume hood

- Asahiklin AK-225 is non-flammable

-Electrospinning/spraying of other hazardous solvents in fume hood

Page 5: Electrospraying/Electrospinning equipment in 66-353 SAFETY ...web.mit.edu/cohengroup/safety/electrospinning102010.pdfElectrospinning/s praying of other hazardous solvents in fume hood

Nozzle, high voltage clip

5

• Alligator clip connected to high voltage output

• The plate is isolated from lab stand and fume hood

• Care when using flammable solvents

• Wear long-sleeved labcoat/safety googles while operating

Page 6: Electrospraying/Electrospinning equipment in 66-353 SAFETY ...web.mit.edu/cohengroup/safety/electrospinning102010.pdfElectrospinning/s praying of other hazardous solvents in fume hood

Risks from exposure to electrical current

• major hazards associated with electricity are electrical shock and fire

• Wet skin is has significantly less resistance (wear gloves!)• The effect of the shock may range from a slight tingle to

severe burns to cardiac arrest.

6

Current Reaction

1 Milliampere Perception level

5 Milliamperes Slight shock felt; not painful but disturbing

6-30 Milliamperes Painful shock; "let-go" range

50-150 Milliamperes Extreme pain, respiratory arrest, severe muscular

contraction

1000-4,300 Milliamperes Ventricular fibrillation

10,000+ Milliamperes Cardiac arrest, severe burns and probable death

Source: EHS

Current drawn ~ 10-20 microamperes