emmaandpulseox

11
EMMA Emergency Capnometer & Fingertip Sats Monitor Greater Sydney HEMS ASNSW Training Package

Upload: nswhems

Post on 21-Jan-2015

410 views

Category:

Business


0 download

DESCRIPTION

 

TRANSCRIPT

Page 1: Emmaandpulseox

EMMA Emergency Capnometer&

Fingertip Sats MonitorGreater Sydney HEMS

ASNSWTraining Package

Page 2: Emmaandpulseox

EMMA Monitor

• Quantative IR Capnometer.

• Inline, with disposable sensor airway adapter. Single Patient Use.

• Powered by 2x AAA batteries, 8 hours normal use.

• Weight 60g.

• Alarms:

o Battery low, no adapter, check adapter.

o Apnoea (Default 3bpm)

o ET CO2 Low / High. (Default: Low = off / High = 50mmHg.)

Page 3: Emmaandpulseox

Buttons and Display

On/ Off button.

Alarm silence.(and upper limit EtCO2

adjustment)

CO2 Status Bar

Battery Status.(and lower limit ETCO2

Adjustment)

Alarm silence indicator,Yellow (not displayed here)

p ETCO2

mmHg

Disposable Airway adaptor

Respiratory Ratebpm

Page 4: Emmaandpulseox

Battery Change

• Cover Unlock.

Page 5: Emmaandpulseox

On / Off

• Press on/off button.

• Insert airway adapter.

• Powers off automatically.o 15 sec after airway adapter disconnect.o In apnoea situation, 2 mins after apnoea alarm

cancelled, if no breath detected.

Page 6: Emmaandpulseox

Alarms

• Battery low – blinks with 30 mins life left.

• No breath – blinking BPM display and 3 beeps every ~30 sec.

o Default low RR = 3bpm.

• ET CO2 limits exceeded – CO2 display blinking display and single beep.

o Default: Hi = 50mmHg, Lo = 0mmHg.

Page 7: Emmaandpulseox

USE

• Backup capnometer for the LP15 monitor in pre-hospital RSI.

• Backup capnometer for the Propaq in transport and In Hospital intubations.

• Spontaneously breathing patients with LMA, following failed intubation etc.

• The LP15 and Propaq remain our primary monitors.

Page 8: Emmaandpulseox

Fingertip Pulse Oximeter

• Standard absorbance pulse oximetry @ 660/940 nm.

• Displays HR, Sats, plethysmograph wave or bar.

• Low power indicator on main screen when < 30 min Battery life.

On Button

Page 9: Emmaandpulseox

Use

• On button.

• Repeated pushes scrolls thru different orientations, and pleth waveforms / bar graph.

• Auto off after 15 sec of no signal.

Pleth wave

Pleth bar

Page 10: Emmaandpulseox

Specifications

• Sp O2: 70-99%

o Accuracy: (+/- 2%: 80 – 100%, +/- 3%: 70-80%).

• Pulse rate: 3-325bpm.

• Power consumption: 2xAAA batteries, 30 hours normal use. Low voltage indicator.

• Operating range: 5-40°C

• Weight 50g.

Page 11: Emmaandpulseox

Stowage• All equip held within

EMMA hard case.

• Kept in the monitoring pouch in Pre-Hospital pack 1.

• Contents:o 4 spare AAA batteries

o 1 EMMA on lanyard

o 1 Pulse ox on lanyard

o 1 sealed airway adapter• Extra airway adapter in

monitoring pouch.