TENDER ENQUIRY DOCUMENT FOR PURCHASE OF MEDICAL EQUIPMENT FOR AND ON BEHALF OF EMPLOYEE STATE INSURANCE CORPORATION UNDER THE ADMINISTRATIVE CONTROL OF MINISTRY OF LABOUR AND EMPLOYMENT GOVT. OF INDIA HLL/PCD/ESIC-20/10-11 BY HLL Lifecare Limited (A GOVT. OF INDIA ENTERPRISE & Formerly HINDUSTAN LATEX LTD.) Procurement & Consultancy Services Division B-14A, Sector-62, Noida-201 307, Uttar Pradesh, India Ph: 0120-4071500; Fax: 0120-4071513 URL: www.lifecarehll.com Email: [email protected]

Upload: others

Post on 25-Sep-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

TENDER ENQUIRY DOCUMENT FOR PURCHASE OF

MEDICAL EQUIPMENT

FOR AND ON BEHALF OF

EMPLOYEE STATE INSURANCE CORPORATION UNDER THE ADMINISTRATIVE CONTROL OF MINISTRY OF LABOUR AND EMPLOYMENT

GOVT. OF INDIA

HLL/PCD/ESIC-20/10-11

BY

HLL Lifecare Limited

(A GOVT. OF INDIA ENTERPRISE & Formerly HINDUSTAN LATEX LTD.)

Procurement & Consultancy Services Division

B-14A, Sector-62, Noida-201 307, Uttar Pradesh, India

Ph: 0120-4071500; Fax: 0120-4071513

URL: www.lifecarehll.com

Email: [email protected]

Page 2: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited

HLL/PCD/ESIC-20/10-11 Page 2 of 166 01.09.2010

INDEX Section Topic Page No.

Section I – Notice inviting Tender (NIT) ---------------------------------------------------------- 03

Section II – General Instructions to Tenderers (GIT) --------------------------------------------- 07

Section III – Special Instructions to Tenderers (SIT) ---------------------------------------------- 24

Section IV – General Conditions of Contract (GCC) ---------------------------------------------- 25

Section V – Special Conditions of Contract (SCC) ----------------------------------------------- 40

Section VI – List of Requirements ------------------------------------------------------------------- 41

Section VII – Technical Specifications --------------------------------------------------------------- 45

Section VIII – Quality Control Requirements ------------------------------------------------------- 141

Section IX – Qualification Criteria ------------------------------------------------------------------- 142

Section X – Tender Form ----------------------------------------------------------------------------- 144

Section XI – Price Schedules -------------------------------------------------------------------------- 145

Section XII – Questionnaire ---------------------------------------------------------------------------- 149

Section XIII – Deleted

Section XIV – Manufacturer’s Authorisation Form -------------------------------------------------- 150

Section XV – Bank Guarantee Form for Performance Security /CMC Security ---------------- 151

Section XVI – Contract Form (A & B) ------------------------------------------------------------------152

Section XVII – Proforma of Consignee Receipt Certificate ----------------------------------------- 156

Section XVIII– Proforma of Final Acceptance Certificate by the Consignee --------------------- 157

Section XIX – Details of Shipping arrangement for Liner Cargoes in respect of C&F/CIF/

Turnkey/F.O.R. Contracts for Import-------------------------------------------------- 159

Section XX – Check List for the Tenderers ----------------------------------------------------------- 163

Section XXI – Consignee address ---------------------------------------------------------------------- 166

Page 3: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited NIT

HLL/PCD/ESIC-20/10-11 Page 3 of 166 01.09.2010

SECTION I

NOTICE INVITING TENDERS (NIT) HLL LIFECARE LIMITED

(A GOVERNMENT OF INDIA ENTERPRISE)

Procurement & Consultancy Services Division

B-14A, Sector-62, Noida-201 307, Uttar Pradesh, India

Ph: 0120-4071500; Fax: 0120-4071513

Email: [email protected]

Tender Enquiry No. HLL/PCD/ESIC-20/10-11 Date: 01.09.2010

1. Procurement & Consultancy Services Division of HLL Lifecare Limited (Formerly Hindustan Latex

Limited), for and on behalf of Director General of Employee State Insurance Corporation (ESIC),

invites sealed tenders, from eligible and qualified tenderers for supply of following Medical

Equipment for ESI Hospitals at different locations within India:

Sl.

No. Item Description

Total

Quantity

(Nos.)

EMD Amount

Rs.

1 Biphasic Defibrilator 8 56,000.00

2 Fiber Optic Bronchoscope with Camera 6 144,000.00

3 Pulse Oximeter 3 3,000.00

4 Anaesthesia Workstation 2 100,000.00

5 ECG Machine 12 channel 2 6,000.00

6 Pharmaceutical Refrigerator for O.T. 1 3,000.00

7 O T Tables for Orthopaedics 5 290,000.00

8 O T Tables for General Surgery & Gynaecology 13 520,000.00

9 Electro Surgical Unit 5 45,000.00

10 O T Table for Minor O T 1 7,000.00

11 Boyle's Appratus 1 20,000.00

12 Multipara Monitor (6Nos) with Central Station (1No) 2 144,000.00

13 Portable Ventilator 1 10,000.00

14 BIPAP 2 12,000.00

15 Binocular Microscope 6 27,000.00

16 Urine Analyzer 1 8,500.00

17 ELISA Reader with Washer 5 50,000.00

18 Cell Counter(Fully automated 5 part differential

analyser) 2 28,000.00

19 Biochemistry Semi- Auto Analyser 2 6,000.00

20 Coagulation Analyzer 2 20,000.00

21 Electrolyte Analyzer 1 8,000.00

22 500 M.A. X-ray machine 1 18,000.00

23 Portable X-ray machine (Mobile) 3 18,000.00

24 Colour Doppler Digital 1 40,000.00

25 Infant Resuscitator 1 10,000.00

26 Neonatal Monitor 6 48,000.00

27 Ventilator Critical Care 1 26,000.00

28 Open Care System 6 84,000.00

Page 4: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited NIT

HLL/PCD/ESIC-20/10-11 Page 4 of 166 01.09.2010

Sl.

No. Item Description

Total

Quantity

(Nos.)

EMD Amount

Rs.

29 Transport Incubators 1 14,000.00

30 Plaster cutter with Aspirator 1 10,500.00

31 Tourniquet Electronic 1 3,800.00

32 Hot Pack Unit 1 6,200.00

33 Cold Pack Unit 1 6,200.00

34 Traction Unit 1 18,000.00

35 Operating Laparoscope Set for Gynaec 1 70,000.00

36 Hysteroscope System 1 20,000.00

37 Cryo Surgical Unit 4 52,000.00

38 Vacuum Extractor 4 36,000.00

39 Fetal Monitor 7 56,000.00

40 ENT Examination Unit 4 200,000.00

41 Operating Laparoscope Set for General Surgery 2 140,000.00

42 Multipara Bed side Monitor 1 14,000.00

43 Emergency Resuscitation Kit-Adult(imported) 6 66,000.00

44 Portable Spirometer 1 12,000.00

45 Infusion Pump Syringe 5 6,000.00

46 PFT Machine(Spirometer & Computerised) 2 48,000.00

47 Fully Automated Biochemistry Analyzer 1 14,000.00

48 Centrifuge Machine 24-36 tubes 3 8,400.00

49 Computed Radiography System 1 100,000.00

50 Advancved Colour Doppler Ultrasound Scanner 1 120,000.00

51 Donor Chair 2 24,000.00

52 Blood Collection Monitor 3 12,000.00

53 Blood Bank Refrigerator 3 15,000.00

54 Tube Sealer 2 4,800.00

55 Cell Counter - 3 Parameter 1 8,000.00

56 Refrigerated Centrifuge 1 4,000.00

57 Plasma Expressor(E) 2 8,000.00

58 Dielectric Tube Sealer Table Top 1 3,000.00

59 PLATELET INCUBATOR & AGITATOR 1 4,000.00

60 AUTOMATED PLASMA THAWING

EQUIPMENT 1 6,000.00

61 LAMINAR FLOW- VERTICAL 1 8,000.00

62 Deep Freezer -40 Deg. 2 8,000.00

63 Deep Freezer -80 Deg. 1 6,000.00

64 Cell Separator 1 8,000.00

65 ICU Ventilator (High End) 1 24,000.00

66 ABG Analyzer 2 28,000.00

67 Arthroscopy System 1 90,000.00

68 Color Doppler Ultrasound Unit 1 36,000.00

69 Video Bronchoscopy System 1 80,000.00

70 High Frequency Mobile X- Ray Unit 1 20,000.00

Page 5: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited NIT

HLL/PCD/ESIC-20/10-11 Page 5 of 166 01.09.2010

Sl.

No. Item Description

Total

Quantity

(Nos.)

EMD Amount

Rs.

71 Digital Mamography System 1 180,000.00

72 Holter Monitoring System 2 32,000.00

73 Stress Test System(TMT) 1 2,400.00

74 Volumetric Infusion Pump 10 12,000.00

75 Biosafety Cabinet 1 16,000.00

76 Laprocator 2 20,000.00

77 Ultrasound Machine 3D with Color Doppler 2 108,000.00

78 Phototherapy Machine 1 15,000.00

79 Microscope 6 3,600.00

80 Yag Laser 1 50,000.00

81 Cell Separator 1 40,000.00

82 Cell Washer 1 14,000.00

83 Donor Couch 1 3,000.00

84 Platelet Incubator & Agitator 1 10,000.00

85 Sterile Tube Connecting Device 1 19,000.00

86 General Instruments Set 24 696,000.00

87 Surgery Instrument Set for Minor O.T. 4 60,000.00

88 Orthopaedic Surgery Instrument Set 1 70,000.00

89 Vaginal Hysterectomy 4 20,000.00

90 Abdominal Hysterectomy set 1 3,200.00

91 Delivery and Ceasarian set 4 24,000.00

92 General Surgical Instruments for O T 1 40,000.00

93 Instruments for Dental Dept. 1 14,000.00

94 Instruments for Eye Implant 1 20,000.00

2. Tender No.: HLL/PCD/ESIC-20/10-11

Sl

No. Description Schedule

i. Dates of sale of tender enquiry

documents

03.09.2010 to 04.10.2010, in all working

days, during 1000 Hrs. to 1600 Hrs. (IST)

ii. Place of sale of Tender Enquiry

Documents

HLL Lifecare Limited

Procurement & Consultancy Services Divn.

B-14A, Sector-62, Noida -201 307

iii. Cost of the Tender Enquiry Document Rs. 3,000.00/ USD 75.00

iv. Pre Tender Meeting Date & Time 10.09.2010, 1100 Hrs. (IST)

v. Pre Tender Meeting Venue Same as given in 2 (ii)

vi. Closing date & time for receipt of

Tender 05.10.2010, 1430 Hrs. (IST)

vii. Time and date of opening of Techno-

Commercial tenders 05.10.2010, 1500 Hrs. (IST)

viii. Venue of Opening of Techno-

Commercial Tender Same as given in 2 (ii)

Page 6: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited NIT

HLL/PCD/ESIC-20/10-11 Page 6 of 166 01.09.2010

3. Interested tenderers may obtain further information about this tender from the office of Head

(P&CD), HLL Lifecare Ltd., Noida. Tender Enquiry Documents may be purchased on payment of

non-refundable fee of Rs. 3,000.00/ USD 75.00 per set in the form of account payee Demand

Draft/Pay Order/Cashier’s Cheque/Banker’s Cheque, drawn on a scheduled bank in India, in favour

of “HLL Lifecare Limited” payable at New Delhi.

4. If requested, the Tender Enquiry Documents will be mailed by Registered Post/Speed Post to the

domestic tenderers and by international airmail to the foreign tenderers, for which extra expenditure

per set will be Rs 100.00 for domestic post and USD 50.00 for international airmail. The tenderer is

to add the applicable postage cost in the non-refundable fee mentioned in Para 3 above. However,

HLL Lifecare Ltd. shall not be responsible for any postal loss/delay.

5. Tenderer may also download the tender enquiry documents from the web site www.esic.nic.in or

www.lifecarehll.com and submit its tender by utilizing the downloaded document, along with the

required non-refundable fee as mentioned in Para 3 above.

6. All prospective tenderers may attend the Pre Tender meeting. The venue, date and time indicated in

the Para 2 above

7. Tenderers shall ensure that their tenders, complete in all respects, are dropped in the Tender Box

located at HLL Lifecare Limited, Procurement and Consultancy Division, B-14A, Sector-62,

Noida -201307, Uttar Pradesh on or before the closing date and time indicated in the Para 2 above,

failing which the tenders will be treated as late tender and rejected. The tenders sent by post/ courier

must reach the above said address on or before the closing date & time indicated in Para 2 above,

failing which the tenders will be treated as late tender and rejected.

8. In the event of any of the above mentioned dates being declared as a holiday / closed day for the

purchase organisation, the tenders will be sold/received/opened on the next working day at the

appointed time.

9. The Tender Enquiry Documents are not transferable.

10. All Tenders must be accompanied by EMD as mentioned against each item. Tenders without EMD

shall be rejected.

For and on behalf of Employee State Insurance Corporation

Head (P & CD)

HLL Lifecare Limited,

Procurement and Consultancy Division

B-14A, Sector -62, Noida -201307,

Uttar Pradesh.

Page 7: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 7 of 166 01.09.2010

SECTION - II

GENERAL INSTRUCTIONS TO TENDERERS (GIT)

CONTENTS

Sl.

No. Topic

Page

No.

A PREAMBLE

1 Definitions and Abbreviations 9

2 Introduction 10

3 Deleted --

4 Language of Tender 11

5 Eligible Tenderers 11

6 Eligible Goods and Services 11

7 Tendering Expense 11

B TENDER ENQUIRY DOCUMENTS

8 Contents of Tender Enquiry Documents 11

9 Deleted --

10 Clarification of Tender Enquiry Documents 12

C PREPARATION OF TENDERS

11 Documents Comprising the Tender 12

12 Tender Currencies 13

13 Tender Prices 13

14 Indian Agent 15

15 Firm Price / Variable Price 16

16 Deleted --

17 Documents Establishing Tenderer’s Eligibility and

Qualifications 16

18 Documents Establishing Good’s Conformity to Tender

Enquiry Document

16

19 Earnest Money Deposit (EMD) 17

20 Tender Validity 17

21 Signing and Sealing of Tender 17

D SUBMISSION OF TENDERS

22 Submission of Tenders 18

23 Late Tender 18

24 Alteration and Withdrawal of Tender 18

Page 8: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 8 of 166 01.09.2010

E TENDER OPENING

25 Opening of Tenders 19

F SCRUTINY AND EVALUATION OF TENDERS

26 Basic Principle 19

27 Preliminary Scrutiny of Tenders 19

28 Deleted --

29 Discrepancies in Prices 20

30 Discrepancy between original and copies of Tender 20

31 Qualification Criteria 20

32 Conversion of Tender Currencies to Indian Rupees 20

33 Deleted --

34 Comparison of Tenders 21

35 Additional Factors and Parameters for Evaluation and

Ranking of Responsive Tenders 21

36 Tenderer’s capability to perform the contract 21

37 Contacting the Purchaser 21

G AWARD OF CONTRACT

38 Purchaser’s Right to Accept any Tender and to Reject

any or All Tenders

21

39 Award Criteria 22

40 Variation of Quantities at the Time of Award 22

41 Notification of Award 22

42 Issue of Contract 22

43 Non-receipt of Performance Security and Contract by

the Purchaser/Consignee

22

44 Return of EMD 22

45 Publication of Tender Result 23

46 Corrupt or Fraudulent Practices 23

Page 9: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 9 of 166 01.09.2010

A. PREAMBLE

1. Definitions and Abbreviations

1.1 The following definitions and abbreviations, which have been used in these documents shall have

the meanings as indicated below:

1.2. Definitions:

(i) “Purchaser” means the organization purchasing goods and services as incorporated in

the Tender Enquiry document.

(ii) “Tender” means Bids / Quotation / Tender received from a Firm / Tenderer / Bidder.

(iii) “Tenderer” means Bidder/ the Individual or Firm submitting Bids / Quotation / Tender

(iii) “Supplier” means the individual or the firm supplying the goods and services as

incorporated in the contract.

(iv) “Goods” means the articles, material, commodities, livestock, furniture, fixtures, raw

material, spares, instruments, machinery, equipment, medical equipment, industrial plant

etc. which the supplier is required to supply to the purchaser under the contract.

(v) “Services” means services allied and incidental to the supply of goods, such as

transportation, installation, commissioning, provision of technical assistance, training,

after sales service, maintenance service and other such obligations of the supplier

covered under the contract.

(vi) “Earnest Money Deposit” (EMD) means Bid Security/ monetary or financial guarantee

to be furnished by a tenderer along with its tender.

(vii) “Contract” means the written agreement entered into between the purchaser and/or

consignee and the supplier, together with all the documents mentioned therein and

including all attachments, annexure etc. therein.

(viii) “Performance Security” means monetary or financial guarantee to be furnished by the

successful tenderer for due performance of the contract placed on it. Performance

Security is also known as Security Deposit.

(ix) “Consignee” means the Hospital/Dispensaries/Institute/Medical College/ person to

whom the goods are required to be delivered as specified in the Contract. If the goods

are required to be delivered to a person as an interim consignee for the purpose of

despatch to another person as provided in the Contract then that “another” person is the

consignee, also known as ultimate consignee.

(x) “Specification” means the document/standard that prescribes the requirement with

which goods or service has to conform.

(xi) “Inspection” means activities such as measuring, examining, testing, gauging one or

more characteristics of the product or service and comparing the same with the specified

requirement to determine conformity.

(xii) “Day” means calendar day.

1.3 Abbreviations:

(i) “T E Document” means Tender Enquiry Document

(ii) “NIT” means Notice Inviting Tenders.

(iii) “GIT” means General Instructions to Tenderers

(iv) “SIT” means Special Instructions to Tenderers

(v) “GCC” means General Conditions of Contract

(vi) “SCC” means Special Conditions of Contract

Page 10: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 10 of 166 01.09.2010

(vii) “DGS&D” means Directorate General of Supplies and Disposals

(viii) “NSIC” means National Small Industries Corporation

(ix) “PSU” means Public Sector Undertaking

(x) “CPSU” means Central Public Sector Undertaking

(xi) “LSI” means Large Scale Industry

(xii) “SSI” means Small Scale Industry

(xiii) “LC” means Letter of Credit

(xiv) “DP” means Delivery Period

(xv) “BG” means Bank Guarantee

(xvi) “ED” means Excise Duty

(xvii) “CD” means Custom Duty

(xviii) “VAT” means Value Added Tax

(xix) “CENVAT” means Central Value Added Tax

(xx) “CST” means Central Sales Tax

(xxi) “RR” means Railway Receipt

(xxii) “BL” means Bill of Lading

(xxiii) “FOB” means Free on Board

(xxiv) “FCA” means Free Carrier

(xxv) “FOR” means Free On Rail

(xxvi) “CIF” means Cost, Insurance and Freight

(xxvii) “CIP (Destinations)” means Carriage and Insurance Paid up to named port of

destination. Additionally the Insurance (local transportation and storage) would be

extended and borne by the Supplier from ware house to the consignee site for a period

including 3 months beyond date of delivery.

(xxviii) “DDP” means Delivery Duty Paid named place of destination (consignee site)

(xxix) “INCOTERMS” means International Commercial Terms as on the date of Tender

Opening

(xxx) ”ESIC” means Employee State Insurance Corporation.

(xxxi) “DG” means Director General of ESIC under the administrative control of Ministry of

Labour and Employment

(xxxii) “CMC” means Comprehensive Maintenance Contract (labour, spare and preventive

maintenance)

(xxxiii) “RT” means Re-Tender.

2. Introduction

2.1 The Purchaser has issued these TE documents for purchase of goods and related services as

mentioned in Section – VI – “List of Requirements”, which also indicates, interalia, the required

quantity, delivery schedule, terms and place of delivery.

2.2 This section (Section II - “General Instructions to Tenderers”) provides the relevant information

as well as instructions to assist the prospective tenderers in preparation and submission of

tenders. It also includes the mode and procedure to be adopted by the purchaser for receipt and

opening as well as scrutiny and evaluation of tenders and subsequent placement of contract.

2.3 The tenderers shall also read the Special Instructions to Tenderers (SIT) related to this purchase,

as contained in Section III of these documents and follow the same accordingly. Whenever there

is a conflict between the GIT and the SIT, the provisions contained in the SIT shall prevail over

those in the GIT.

2.4 Before formulating the tender and submitting the same to the purchaser, the tenderer should read

and examine all the terms, conditions, instructions, checklist etc. contained in the TE documents.

Failure to provide and/or comply with the required information, instructions etc. incorporated in

these TE documents may result in rejection of its tender.

Page 11: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 11 of 166 01.09.2010

3. Deleted

4. Language of Tender

4.1 The tender submitted by the tenderer and all subsequent correspondence and documents relating

to the tender exchanged between the tenderer and the purchaser, shall be written in the English

language, unless otherwise specified in the Tender Enquiry. However, the language of any

printed literature furnished by the tenderer in connection with its tender may be written in any

other language provided the same is accompanied by a notarised English translation and, for

purposes of interpretation of the tender, the English translation shall prevail.

4.2 The tender submitted by the tenderer and all subsequent correspondence and documents relating

to the tender exchanged between the tenderer and the purchaser, may also be written in the Hindi

language, provided that the same are accompanied by notarised English translation, in which

case, for purpose of interpretation of the tender etc, the English translations shall prevail.

5. Eligible Tenderers

This invitation for tenders is open to all suppliers who fulfil the eligibility criteria specified in

these documents.

6. Eligible Goods and Services

All goods and related services to be supplied under the contract shall have their origin in India or

any other country with which India has not banned trade relations. The term “origin” used in this

clause means the place where the goods are mined, grown, produced, or manufactured or from

where the related services are arranged and supplied.

7. Tendering Expense

The tenderer shall bear all costs and expenditure incurred and/or to be incurred by it in

connection with its tender including preparation, mailing and submission of its tender and for

subsequent processing the same. The purchaser will, in no case be responsible or liable for any

such cost, expenditure etc regardless of the conduct or outcome of the tendering process.

B. TENDER ENQUIRY DOCUMENTS

8. Content of Tender Enquiry Documents

8.1 In addition to Section I – “Notice inviting Tender” (NIT), the TE documents include:

Section II – General Instructions to Tenderers (GIT)

Section III – Special Instructions to Tenderers (SIT)

Section IV – General Conditions of Contract (GCC)

Section V – Special Conditions of Contract (SCC)

Section VI – List of Requirements

Section VII – Technical Specifications

Section VIII – Quality Control Requirements

Section IX – Qualification Criteria

Section X – Tender Form

Section XI – Price Schedules

Section XII – Questionnaire

Section XIII – Deleted

Section XIV – Manufacturer’s Authorisation Form

Section XV – Bank Guarantee Form for Performance Security/CMC Security

Section XVI – Contract Forms A & B

Section XVII – Proforma of Consignee Receipt Certificate

Page 12: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 12 of 166 01.09.2010

Section XVIII – Proforma of Final Acceptance Certificate by the consignee

Section XIX – Details of Shipping arrangement for Liner Cargoes in respect of

C&F/CIF/Turnkey/F.O.R. Contracts for Import

Section XX – Check List for the Tenderers

Section XXI – Consignee List

8.2 The relevant details of the required goods and services, the terms, conditions and procedure for

tendering, tender evaluation, placement of contract, the applicable contract terms and, also, the

standard formats to be used for this purpose are incorporated in the above-mentioned documents.

The interested tenderers are expected to examine all such details before submission of the tender.

9. Deleted

10. Clarification of TE documents

A tenderer requiring any clarification or elucidation on any issue of the TE documents may take

up the same with the purchaser in writing. The purchaser will respond to such request provided

the purchaser receives the same minimum 1 (one) hour prior to the scheduled time of pre-bid

meeting. Clarification for the same will be published on the websites as mentioned in para 5 of

NIT (Section I) within 4 days after the prebid meeting.

C. PREPARATION OF TENDERS

11. Documents Comprising the Tender

11.1 The Two Tender System, i.e. “Techno – Commercial Tender” and “Price Tender” prepared by

the tenderer shall comprise the following:

A) Techno – Commercial Tender (Un priced Tender)

i) Earnest money furnished in accordance with GIT clause 19.

ii) Tender Form as per Section X.

iii) Documentary evidence, as necessary in terms of clauses 5 and 17 establishing that the

tenderer is eligible to submit the tender and, also, qualified to perform the contract if its

tender is accepted.

iv) Tenderer/Agent who quotes for goods manufactured by other manufacturer shall furnish

Manufacturer’s Authorisation in the prescribed format as per Section - XIV.

v) Power of Attorney/Authorisation in favour of signatory of TE documents.

vi) Documents and relevant details to establish in accordance with GIT clause 18 that the

goods and the allied services to be supplied by the tenderer conform to the requirement of

the TE documents.

vii) Performance Statement as per section IX along with relevant copies of purchase orders

and end users’ satisfaction certificate.

viii) Price Schedule(s) as per Section XII filled up with all the details including Qty., Make,

Model, Contry of origin, etc. of the goods offered with prices blank (without indicating

any prices).

ix) Certificate of Incorporation of the bidder.

x) Checklist as per Section XX.

xi) Statement of deviations parameter wise from tendered technical specifications, if any.

B) Price Tender:

The information given at clause no. 11.1 A)- viii) above should be reproduced with the prices

indicated.

Page 13: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 13 of 166 01.09.2010

N.B.

1. All pages of the Tender should be page numbered and indexed.

2. It is the responsibility of tenderer to go through the TE document to ensure furnishing all

required documents in addition to above, if any.

11.2 The authorized signatory of the tenderer must sign the tender duly stamped at appropriate places

and initial all the remaining pages of the tender.

11.3 A tender, which does not fulfil any of the above requirements and/or gives evasive

information/reply against any such requirement, shall be liable to be ignored and rejected.

11.4 Tender sent by fax/telex/cable/electronically shall be ignored.

12. Tender currencies

12.1 The tenderer supplying indigenous goods or already imported goods shall quote only in Indian

Rupees.

12.2 For imported goods if supplied directly from abroad, prices shall be quoted in any freely

convertible currencies say USD, Euro, GBP or Yen. As regards price(s) for allied services, if any

required with the goods, the same shall be quoted in Indian Rupees only if such services are to be

performed /undertaken in India. Commission for Indian Agent, if any and if payable shall be

indicated in the space provided for in the price schedule and will be payable in Indian Rupees

only.

12.3 Tenders, where prices are quoted in any other way shall be treated as non-responsive and

rejected.

13 Tender Prices

13.1 The Tenderer shall indicate on the Price Schedule provided under Section XI all the specified

components of prices shown therein including the unit prices and total tender prices of the goods

and services it proposes to supply against the requirement. All the columns shown in the price

schedule should be filled up as required. If any column does not apply to a tenderer, it should be

clarified as “NA” by the tenderer.

13.2 The tenderer has the option to submit its quotation for any one or more item (s) in the List of

Requirements. However, separate sealed cover to be used for each item for price bid.

13.3 The quoted prices for goods offered from within India and that for goods offered from abroad are

to be indicated separately in the applicable Price Schedules attached under Section XI.

The price quoted by the tenderer for indigenous goods shall not be higher than the lowest price

charged for the goods of the same nature, class or description to an individual/ firm/ organisation

or department of Govt. of India.

For imported goods, the price quoted shall not be higher than the lowest price charged by the

tenderer for the goods of the same nature, class or description to a purchaser, domestic or foreign

or to any organisation or department of Govt. of India.

If it is found at any stage that the goods as stated have been supplied at a lower price, then that

price, with due allowance for elapsed time will be applicable to the present case and the

difference in cost would be refunded by the supplier to the purchaser, if the contract has already

been concluded.

13.4 While filling up the columns of the Price Schedule, the following aspects should be noted for

compliance:

Page 14: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 14 of 166 01.09.2010

13.4.1 For domestic goods or goods of foreign origin located within India, the prices in the

corresponding price schedule shall be entered separately in the following manner:

a) the price of the goods, quoted ex-factory/ ex-showroom/ ex-warehouse/ off-the-shelf, as

applicable, including all taxes and duties like sales tax, CST VAT, CENVAT, Custom Duty,

Excise Duty etc. already paid or payable on the components and raw material used in the

manufacture or assembly of the goods quoted ex-factory etc. or on the previously imported

goods of foreign origin quoted ex-showroom etc;

b) any sales or other taxes and any duties including excise duty, which will be payable on the

finished goods in India if the contract is awarded;

c) charges towards Packing & Forwarding, Inland Transportation, Insurance, Loading/Unloading and other local costs incidental to delivery of the goods to their final destination as specified in the List of Requirements and Price Schedule;

d) the price of Incidental Services, as mentioned in List of Requirements and Price Schedule;

e) the prices of Turnkey ( if any), as mentioned in List of Requirements, Technical Specification

and Price Schedule; and

f) the price of CMC, as mentioned in List of Requirements, Technical Specification and Price

Schedule.

13.4.2 For goods offered from abroad, the prices in the corresponding price schedule shall be

entered separately in the following manner:

a) the price of goods quoted FOB port of shipment, as indicated in the List of Requirements and

Price Schedule;

b) the amount of freight and insurance and price of goods quoted CIP port of entry in India as

indicated in the List of Requirements and Price Schedule;

c) the price of goods quoted should be on DDP basis at consignee site in India as indicated

in the List of Requirements, Price Schedule and Consignee List;

d) wherever applicable, the amount of custom duty with CDEC applicable on CIP value on the

goods to be imported; ESIC will issue Customs Duty Exemption Certificate (CDEC) where

applicable.

e) the charges for Loading/Unloading, Inland transportation, Insurance and other local costs,

Incidental cost to delivery of the goods from the port of entry in India to Consignee Site, as

specified in the List of Requirements and Price Schedule;

f) the charges for Incidental Services, as in the List of Requirements and Price Schedule;

g) the prices of Turnkey ( if any), as mentioned in List of Requirements, Technical Specification

and Price Schedule; and

h) the price of CMC, as mentioned in List of Requirements, Technical Specification and Price

Schedule.

13.5 Additional information and instruction on Duties and Taxes:

13.5.1 If the Tenderer desires to ask for excise duty, sales tax/ VAT, Service Tax, Works Contract Tax

etc. to be paid extra, the same must be specifically stated. In the absence of any such stipulation

the price will be taken inclusive of such duties and taxes and no claim for the same will be

entertained later.

13.5.2 Excise Duty:

a) If reimbursement of excise duty is intended as extra over the quoted prices, the supplier must

specifically say so also indicating the rate, quantum and nature of the duty applicable. In the

absence of any such stipulation it will be presumed that the prices quoted are firm and final

and no claim on account of excise duty will be entertained after the opening of tenders.

Page 15: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 15 of 166 01.09.2010

b) If a Tenderer chooses to quote a price inclusive of excise duty and also desires to be

reimbursed for variation, if any, in the excise duty during the time of supply, the tenderer

must clearly mention the same and also indicate the rate and quantum of excise duty included

in its price. Failure to indicate all such details in clear terms may result in rejection of that

tender.

c) Subject to sub clauses 13.5.2 (a) & (b) above, any change in excise duty upward/downward

as a result of any statutory variation in excise duty taking place within contract terms shall be

allowed to the extent of actual quantum of excise duty paid by the supplier. In case of

downward revision in excise duty, the actual quantum of reduction of excise duty shall be

reimbursed to the purchaser by the supplier. All such adjustments shall include all reliefs,

exemptions, rebates, concession etc. if any obtained by the supplier.

13.5.3 Sales Tax:

If a tenderer asks for sales tax/ VAT, Service Tax and Works Contract Tax to be paid extra, the

rate and nature of sales tax applicable should be shown separately. The sales tax / VAT, Service

Tax and Works Contract Tax will be paid as per the rate at which it is liable to be assessed or has

actually been assessed provided the transaction of sale is legally liable to sales tax / VAT,

Service Tax and Works Contract Tax and is payable as per the terms of the contract. If any

refund of Tax is received at a later date, the Supplier must return the amount forth-with to the

purchaser.

13.5.4 Octroi Duty and Local Duties & Taxes:

Octroi duty, terminal tax and other levies of local bodies (like town body, municipal body etc.) if

not exempted shall be reimbursed to the supplier on production of such proof of payment along

with the final bill.

13.5.5 Customs Duty:

The tenderer shall quote the Customs duty wherever applicable, considering availability of

CDEC. The applicable rates and amount of the Custom Duty and the corresponding Indian

Customs Tariff number should be shown separately in the price schedule. In case of non-

availibility of CDEC, the supplier has to custom cleared the goods paying the applicable Custom

Duty and any difference of duty from the applicable CDEC rate shall be reimbursed separately

on production of such proof of payment alongwith the final bill.

13.6 For transportation of imported goods offered from abroad, relevant instructions as incorporated

under GCC Clause 10 shall be followed.

13.7 For insurance of goods to be supplied, relevant instructions as provided under GCC Clause 11

shall be followed.

13.8 Unless otherwise specifically indicated in this TE document, the terms FCA, FOB, FAS, CIF,

CIP, DDP etc. for imported goods offered from abroad, shall be governed by the rules &

regulations prescribed in the current edition of INCOTERMS, published by the International

Chamber of Commerce, Paris

13.9 The need for indication of all such price components by the tenderers, as required in this clause

(viz., GIT clause 13) is for the purpose of comparison of the tenders by the purchaser and will no

way restrict the purchaser’s right to award the contract on the selected tenderer on any of the

terms offered.

14. Indian Agent 14.1 If a foreign tenderer has engaged an agent in India in connection with its tender, the foreign

tenderer, in addition to indicating Indian agent’s commission, if any, in a manner described under

GIT sub clause 12.2 above, shall also furnish the following information:

Page 16: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 16 of 166 01.09.2010

a) The complete name and address of the Indian Agent and its permanent income tax

account number as allotted by the Indian Income Tax authority.

b) The details of the services to be rendered by the agent for the subject requirement.

c) Details of Service outlets in India, nearest to the consignee(s), to render services during

Warranty and CMC period.

15. Firm Price

15.1 Unless otherwise specified in the SIT, prices quoted by the tenderer shall remain firm and fixed

during the currency of the contract and not subject to variation on any account.

15.2 However, as regards taxes and duties, if any, chargeable on the goods and payable, the conditions

stipulated in GIT clause 13 will apply.

16. Deleted

17 Documents Establishing Tenderer’s Eligibility and Qualifications

17.1 Pursuant to GIT clause 11, the tenderer shall furnish, as part of its tender, relevant details and

documents establishing its eligibility to quote and its qualifications to perform the contract if its

tender is accepted.

17.2 The documentary evidence needed to establish the tenderer’s qualifications shall fulfil the

following requirements:

a) in case the tenderer offers to supply goods, which are manufactured by some other firm, the

tenderer has been duly authorised by the goods manufacturer to quote for and supply the

goods to the purchaser. The tenderer shall submit the manufacturer’s authorization letter to

this effect as per the standard form provided under Section XIV in this document.

b) the tenderer has the required financial, technical and production capability necessary to

perform the contract and, further, it meets the qualification criteria incorporated in the

Section IX in these documents.

c) in case the tenderer is not doing business in India, it is duly represented by an agent stationed

in India fully equipped and able to carry out the required contractual functions and duties of

the supplier including after sale service, maintenance & repair etc. of the goods in question,

stocking of spare parts and fast moving components and other obligations, if any, specified in

the conditions of contract and/or technical specifications.

d) in case the tenderer is an Indian agent/authorized representative quoting on behalf of a

foreign manufacturer for the restricted item, the Indian agent/authorized representative is

already enlisted under the Compulsory Enlistment Scheme of Ministry of Finance, Govt. of

India, operated through Directorate General of Supplies & Disposals (DGS&D), New Delhi.

18. Documents establishing Good’s Conformity to TE document.

18.1 The tenderer shall provide in its tender the required as well as the relevant documents like

technical data, literature, drawings etc. to establish that the goods and services offered in the

tender fully conform to the goods and services specified by the purchaser in the TE documents.

For this purpose the tenderer shall also provide a clause-by-clause commentary on the technical

specifications and other technical details incorporated by the purchaser in the TE documents to

establish technical responsiveness of the goods and services offered in its tender.

18.2 In case there is any variation and/or deviation between the goods & services prescribed by the

purchaser and that offered by the tenderer, the tenderer shall list out the same in a chart form

without ambiguity and provide the same along with its tender.

Page 17: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 17 of 166 01.09.2010

18.3 If a tenderer furnishes wrong and/or misguiding data, statement(s) etc. about technical

acceptability of the goods and services offered by it, its tender will be liable to be ignored and

rejected in addition to other remedies available to the purchaser in this regard.

19. Earnest Money Deposit (EMD)

19.1 Pursuant to GIT clauses 8.1 and 11.1(A) the tenderer shall furnish along with its tender, earnest

money for amount as shown in the List of Requirements. The earnest money is required to

protect the purchaser against the risk of the tenderer’s unwarranted conduct as amplified under

sub-clause 19.7 below.

19.2 Deleted

19.3 The earnest money shall be denominated in Indian Rupees or equivalent currencies as per GIT

clause 12.2. The earnest money shall not be accepted in any other form except the following:

i. Account Payee Demand Draft or

ii. Banker’s cheque

19.4 The demand draft or banker’s cheque shall be drawn on any commercial bank in India or country

of the tenderer, in favour of the “HLL Lifecare Limited” payable at New Delhi.

19.5 Deleted.

19.6 Unsuccessful tenderers’ earnest money will be returned to them without any interest, after expiry

of the tender validity period, but not later than thirty days after conclusion of the resultant

contract. Successful tenderer’s earnest money will be returned without any interest, after receipt

of performance security from that tenderer.

19.7 Earnest Money is required to protect the purchaser against the risk of the Tenderer’s conduct,

which would warrant the forfeiture of the EMD. Earnest money of a tenderer will be forfeited, if

the tenderer withdraws or amends its tender or impairs or derogates from the tender in any

respect within the period of validity of its tender or if it comes to notice that the

information/documents furnished in its tender is incorrect, false, misleading or forged without

prejudice to other rights of the purchaser. The successful tenderer’s earnest money will be

forfeited without prejudice to other rights of Purchaser if it fails to furnish the required

performance security within the specified period.

20. Tender Validity

20.1 If not mentioned otherwise in the SIT, the tenders shall remain valid for acceptance for a period

of 120 days (One hundred and twenty days) after the date of opening of techno-commercial

tenders prescribed in the TE document. Any tender valid for a shorter period shall be treated as

unresponsive and rejected.

20.2 In exceptional cases, the tenderers may be requested by the purchaser to extend the validity of

their tenders up to a specified period. Such request(s) and responses thereto shall be conveyed by

surface mail or by fax/ email followed by surface mail. The tenderers, who agree to extend the

tender validity, are to extend the same without any change or modification of their original tender

and they are also to extend the validity period of the EMD accordingly. A tenderer, however,

may not agree to extend its tender validity without forfeiting its EMD.

20.3 In case the day up to which the tenders are to remain valid falls on/ subsequently declared a

holiday or closed day for the purchaser, the tender validity shall automatically be extended up to

the next working day.

21. Signing and Sealing of Tender

21.1 The tenderers shall submit their tenders as per the instructions contained in GIT Clause 11.

Page 18: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 18 of 166 01.09.2010

21.2 Unless otherwise mentioned in the SIT, a tenderer shall submit two copies of its tender marking

them as “Original” and “Duplicate”. Duplicate tenders may contain all pages including Technical

Literature/Catalogues as in Original tenders.

21.3 The original and duplicate copies of the tender shall either be typed or written in indelible ink

and the same shall be signed by the tenderer or by a person(s) who has been duly authorized to

bind the tenderer to the contract. The letter of authorization shall be by a written power of

attorney, which shall also be furnished along with the tender.

21.4 All the copies of the tender shall be duly signed at the appropriate places as indicated in the TE

documents and all other pages of the tender including printed literature, if any shall be initialled

by the same person(s) signing the tender. The tender shall not contain any erasure or overwriting,

except as necessary to correct any error made by the tenderer and, if there is any such correction;

the same shall be initialled by the person(s) signing the tender.

21.5 The tenderer is to seal the original and copy of the tender in separate envelopes, duly marking the

same as “Original”, “Duplicate” and so on and writing the address of the purchaser and the

tender reference number on the envelopes. The sentence “NOT TO BE OPENED” before

__________ (The tenderer is to put the date & time of tender opening) are to be written on these

envelopes. The inner envelopes are then to be put in a bigger outer envelope along with envelope

containing EMD, which will also be duly sealed, marked etc. as above. If the outer envelope is

not sealed and marked properly as above, the purchaser will not assume any responsibility for its

misplacement, premature opening, late opening etc.

21.6 TE document seeks quotation following two Tender System, in two parts. First part will be

known as ‘Techno - Commercial Tender’, and the second part ‘Price Tender’ as specified in

clause 11 of GIT. Tenderer shall seal ‘Techno - Commercial Tender (along with envelope

containing EMD)’ and ‘Price Tender’ separately and covers will be suitably super scribed.

Both these sealed covers shall be put in a bigger cover and sealed and procedure prescribed in

Paras 21.1 to 21.5 followed.

D. SUBMISSION OF TENDERS

22. Submission of Tenders

22.1 Unless otherwise specified, the tenderers are to deposit the tenders in the tender box kept for this

purpose at HLL Lifecare Limited, Procurement and Consultancy Division, B-14A, Sector -

62, Noida -201307, Uttar Pradesh. In case of bulky tender, which can not be put into tender

box, the same shall be submitted by the tenderer by hand to Head (P&CD) or his nominee, HLL

Lifecare Limited, Procurement and Consultancy Division, B-14A, Sector -62, Noida -201307, Uttar Pradesh. The officer receiving the tender will give the tenderer an official receipt

duly signed with date and time.

22.2 The tenderers must ensure that they deposit their tenders not later than the closing time and date

specified for submission of tenders. It is the responsibility of the tenderer to ensure that their

Tenders whether sent by post or by courier or by person, are dropped in the Tender Box by the

specified clearing date and time. In the event of the specified date for submission of tender falls

on / is subsequently declared a holiday or closed day for the purchaser, the tenders will be

received up to the appointed time on the next working day.

23. Late Tender

23.1 A tender, which is received after the specified date and time for receipt of tenders will be treated

as “late” tender and will be ignored. 24. Alteration and Withdrawal of Tender

24.1 The tenderer, after submitting its tender, is permitted to alter / modify its tender so long as such

alterations / modifications are received duly signed, sealed and marked like the original tender,

Page 19: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 19 of 166 01.09.2010

within the deadline for submission of tenders. Alterations / modifications to tenders received

after the prescribed deadline will not be considered.

24.2 No tender should be withdrawn after the deadline for submission of tender and before expiry of

the tender validity period. If a tenderer withdraws the tender during this period, it will result in

forfeiture of the earnest money furnished by the tenderer in its tender.

E. TENDER OPENING

25. Opening of Tenders

25.1 The purchaser will open the tenders at the specified date and time and at the specified place as

indicated in the NIT.

In case the specified date of tender opening falls on / is subsequently declared a holiday or closed

day for the purchaser, the tenders will be opened at the appointed time and place on the next

working day.

25.2 Authorized representatives of the tenderers, who have submitted tenders on time may attend

the tender opening provided they bring with them letters of authority from the corresponding

tenderers.

The tender opening official(s) will prepare a list of the representatives attending the tender

opening. The list will contain the representatives’ names & signatures and corresponding

tenderers’ names and addresses.

25.3 Two-Tender system as mentioned in para 21.6 above will be as follows. The Techno-

Commercial Tenders are to be opened in the first instance, at the prescribed time and date as

indicated in NIT. These Tenders shall be scrutinized and evaluated by the competent committee/

authority with reference to parameters prescribed in the TE document. During the Techno -

Commercial Tender opening, the tender opening official(s) will read the salient features of the

tenders like brief description of the goods offered, delivery period, Earnest Money Deposit and

any other special features of the tenders, as deemed fit by the tender opening official(s).

Thereafter, in the second stage, the Price Tenders of only the Techno-Commercially acceptable

offers shall be opened at a latter date which will be notified to such tenderers. The prices, special

discount if any of the goods offered etc., as deemed fit by tender opening official(s) will be read

out.

F. SCRUTINY AND EVALUATION OF TENDERS

26. Basic Principle

26.1 Tenders will be evaluated on the basis of the terms & conditions already incorporated in the TE

document, based on which tenders have been received and the terms, conditions etc. mentioned

by the tenderers in their tenders. No new condition will be brought in while scrutinizing and

evaluating the tenders.

27. Preliminary Scrutiny of Tenders

27.1 The Purchaser will examine the Tenders to determine whether they are complete, whether any

computational errors have been made, whether required sureties have been furnished, whether

the documents have been properly signed stamped and whether the Tenders are generally in

order.

27.2 Deleted.

Page 20: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 20 of 166 01.09.2010

27.3 Deleted

27.4 The tenders will be scrutinized to determine whether they are complete and meet the essential

and important requirements, conditions etc. as prescribed in the TE document. The tenders,

which do not meet the basic requirements, are liable to be treated as non – responsive and will be

summarily ignored.

27.5 The following are some of the important aspects, for which a tender shall be declared non-

responsive and will be summarily ignored;

(i) Tender form as per Section X (signed and stamped) not enclosed

(ii) Tender is unsigned.

(iii) Tender validity is shorter than the required period.

(iv) Required EMD have not been provided.

(v) Tenderer has quoted for goods manufactured by other manufacturer(s) without the required

Manufacturer’s Authorisation as per Format given in Section XIV.

(vi) Tenderer has not agreed to give the required performance security.

(vii) Goods offered are not meeting the tender enquiry specification.

(viii) Tenderer has not agreed to other essential condition(s) specially incorporated in the tender

enquiry like terms of payment, liquidated damages clause, warranty clause, DDP clause,

Delivery period clause, dispute resolution mechanism applicable law.

(ix) Poor/ unsatisfactory past performance.

(x) Tenderers who stand deregistered/banned/blacklisted by any Govt. Authorities.

(xi) Tenderer is not eligible as per GIT Clauses 5.1 & 17.1.

(xii) Tenderer has not quoted for the entire quantity as specified in the List of Requirements for

the quoted item (s).

28. Deleted

29 Discrepancies in Prices

29.1 If, in the price structure quoted by a tenderer, there is discrepancy between the unit price and the

total price (which is obtained by multiplying the unit price by the quantity), the unit price shall

prevail and the total price corrected accordingly, unless the purchaser feels that the tenderer has

made a mistake in placing the decimal point in the unit price, in which case the total price as

quoted shall prevail over the unit price and the unit price corrected accordingly.

29.2 If there is an error in a total price, which has been worked out through addition and/or subtraction

of subtotals, the subtotals shall prevail and the total corrected; and

29.3 If there is a discrepancy between the amount expressed in words and figures, the amount in

words shall prevail, subject to sub clause 29.1 and 29.2 above.

30. Discrepancy between original and copies of Tender

In case any discrepancy is observed between the text etc. of the original copy and that in the

other copies of the same tender set, the text etc. of the original copy shall prevail.

31. Qualification Criteria

Tenders of the tenderers, who do not meet the required Qualification Criteria prescribed in

Section IX, will be treated as non-responsive and will not be considered further.

32. Conversion of tender currencies to Indian Rupees

In case the TE document permits the tenderers to quote their prices in different currencies, all

such quoted prices of the responsive tenderers will be converted to a single currency viz., Indian

Rupees for the purpose of equitable comparison and evaluation, as per the exchange rates

established by the Reserve Bank of India for similar transactions, as on the date of ‘Price

Tender’ opening.

Page 21: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 21 of 166 01.09.2010

33. Deleted

34. Comparison of Tenders

34.1 Unless mentioned otherwise in Section – III – Special Instructions to Tenderers and Section – VI

– List of Requirements, the comparison of the responsive tenders shall be carried out on Delivery

Duty Paid (DDP) consignee site basis. The quoted turnkey prices and CMC prices will also be

added for comparison/ranking purpose for evaluation.

35. Additional Factors and Parameters for Evaluation and Ranking of Responsive Tenders

35.1 Further to GIT Clause 34 above, the purchaser’s evaluation of a tender will include and take into

account the following:

i) In the case of goods manufactured in India or goods of foreign origin already located in India,

sales tax & other similar taxes and excise duty & other similar duties, Customs Duties, Service

Tax, Works Contract Tax etc which will be contractually payable (to the tenderer), on the

goods if a contract is awarded on the tenderer; and

ii) in the case of goods of foreign origin offered from abroad, customs duty and other similar

import duties/taxes, which will be contractually payable (to the tenderer) on the goods if the

contract is awarded on the tenderer.

35.2 The purchaser’s evaluation of tender will also take into account the additional factors, if any,

incorporated in SIT in the manner and to the extent indicated therein.

35.3 Deleted

36. Tenderer’s capability to perform the contract

36.1 The purchaser, through the above process of tender scrutiny and tender evaluation will determine

to its satisfaction whether the tenderer, whose tender has been determined as the lowest evaluated

responsive tender is eligible, qualified and capable in all respects to perform the contract

satisfactorily. If, there is more than one item in the List of Requirements, then, such

determination will be made separately for each item.

36.2 The above-mentioned determination will, interalia, take into account the tenderer’s financial,

technical and production capabilities for satisfying all the requirements of the purchaser as

incorporated in the TE document. Such determination will be based upon scrutiny and

examination of all relevant data and details submitted by the tenderer in its tender as well as such

other allied information as deemed appropriate by the purchaser.

37. Contacting the Purchaser

37.1 From the time of submission of tender to the time of awarding the contract, if a tenderer needs to

contact the purchaser for any reason relating to this tender enquiry and / or its tender, it should do

so only in writing.

37.2 In case a tenderer attempts to influence the purchaser in the purchaser’s decision on scrutiny,

comparison & evaluation of tenders and awarding the contract, the tender of the tenderer shall be

liable for rejection in addition to appropriate administrative actions being taken against that

tenderer, as deemed fit by the purchaser.

G. AWARD OF CONTRACT

38. Purchaser’s Right to accept any tender and to reject any or all tenders

The purchaser reserves the right to accept in part or in full any tender or reject any or more

tender(s) without assigning any reason or to cancel the tendering process and reject all tenders at

Page 22: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 22 of 166 01.09.2010

any time prior to award of contract, without incurring any liability, whatsoever to the affected

tenderer or tenderers.

39. Award Criteria

Subject to GIT clause 38 above, the contract will be awarded to the lowest evaluated responsive

tenderer decided by the purchaser in terms of GIT Clause 36.

40. Variation of Quantities at the Time of Award/ Currency of Contract

40.1 At the time of awarding the contract, the purchaser reserves the right to increase or decrease by

up to twenty five (25) per cent, the quantity of goods and services mentioned in the schedule (s)

in the “List of Requirements” without any change in the unit price and other terms & conditions

quoted by the tenderer.

40.2 If the quantity has not been increased at the time of the awarding the contract, the purchaser

reserves the right to increase by up to twenty five (25) per cent, the quantity of goods and

services mentioned in the contract without any change in the unit price and other terms &

conditions mentioned in the contract, during the currency of the contract or within one-year from

the date of Notification of Award..

41. Notification of Award

41.1 Before expiry of the tender validity period, the purchaser will notify the successful tenderer(s) in

writing, by registered / speed post or by fax/ telex/cable (to be confirmed by registered / speed

post) that its tender for goods & services, which have been selected by the purchaser, has been

accepted, also briefly indicating therein the essential details like description, specification and

quantity of the goods & services and corresponding prices accepted. The successful tenderer

must furnish to the purchaser the required performance security within twenty-one (21) days

from the date of dispatch of this notification, failing which the EMD will be forfeited and the

award will be cancelled. Relevant details about the performance security have been provided

under GCC Clause 5 under Section IV.

41.2 The Notification of Award shall constitute the conclusion of the Contract.

42. Issue of Contract

42.1 Promptly after notification of award, the Purchaser/Consignee will mail the contract form (as per

Section XVI) duly completed and signed, in duplicate, to the successful tenderer by registered /

speed post.

42.2 Within twenty one days from the date of the contract, the successful tenderer shall return the

original copy of the contract, duly signed and dated, to the Purchaser by registered / speed post.

42.3 The Purchaser reserves the right to issue the Notification of Award consignee wise.

43. Non-receipt of Performance Security and Contract by the Purchaser/Consignee

Failure of the successful tenderer in providing performance security and / or returning contract

copy duly signed in terms of GIT clauses 41 and 42 above shall make the tenderer liable for

forfeiture of its EMD and, also, for further actions by the Purchaser/Consignee against it as per

the clause 24 of GCC – Termination of default.

44. Return of E M D

The earnest money of the successful tenderer and the unsuccessful tenderers will be returned to

them without any interest, whatsoever, in terms of GIT Clause 19.6.

Page 23: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GIT

HLL/PCD/ESIC-20/10-11 Page 23 of 166 01.09.2010

45. Publication of Tender Result

The name and address of the successful tenderer(s) receiving the contract(s) will be mentioned in

the notice board/bulletin/web site of the purchaser.

46. Corrupt or Fraudulent Practices

It is required by all concerned to observe the highest standard of ethics during the procurement

and execution of such contracts. In pursuance of this policy, the Purchaser: -

(a) defines, for the purposes of this provision, the terms set forth below

as follows:

(i) “corrupt practice” means the offering, giving, receiving or soliciting of any thing of

value to influence the action of a public official in the procurement process or in

contract execution; and

(ii) “fraudulent practice” means a misrepresentation of facts in order to influence a

procurement process or the execution of a contract to the detriment of the Purchaser, and

includes collusive practice among Tenderers (prior to or after Tender submission)

designed to establish Tender prices at artificial non-competitive levels and to deprive the

Purchaser of the benefits of free and open competition;

(b) will reject a proposal for award if it determines that the Tenderer recommended for award has

engaged in corrupt or fraudulent practices in competing for the contract in question;

(c) will declare a firm ineligible, either indefinitely or for a stated period of time, to be awarded a

contract by the purchaser if it at any time determines that the firm has engaged in corrupt or

fraudulent practices in competing for, or in executing the contract.

Page 24: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited SIT

HLL/PCD/ESIC-20/10-11 Page 24 of 166 01.09.2010

SECTION - III

SPECIAL INSTRUCTIONS TO TENDERERS

(SIT)

The following Special Instructions to Tenderers will apply for this purchase. These special instructions

will modify/substitute/supplement the corresponding General Instructions to Tenderers (GIT)

incorporated in Section II. The corresponding GIT clause numbers have also been indicated in the text

below:

In case of any conflict between the provision in the GIT and that in the SIT, the provision contained in the

SIT shall prevail.

Page 25: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 25 of 166 01.09.2010

SECTION - IV

GENERAL CONDITIONS OF CONTRACT (GCC)

TABLE OF CLAUSES Sl No. Topic Page

1 Application 26

2 Use of contract documents and information 26

3 Patent Rights 26

4 Country of Origin 26

5 Performance Security 26

6 Technical Specifications and Standards 27

7 Packing and Marking 27

8 Inspection, Testing and Quality Control 28

9 Terms of Delivery 28

10 Transportation of Goods 28

11 Insurance 29

12 Spare parts 29

13 Incidental services 30

14 Distribution of Dispatch Documents for Clearance/Receipt of Goods 30

15 Warranty 31

16 Assignment 32

17 Sub Contracts 32

18 Modification of contract 32

19 Prices 33

20 Taxes and Duties 33

21 Terms and mode of Payment 33

22 Delay in the supplier’s performance 35

23 Liquidated Damages 36

24 Termination for default 36

25 Termination for insolvency 37

26 Force Majeure 37

27 Termination for convenience 37

28 Governing language 38

29 Notices 38

30 Resolution of disputes 38

31 Applicable Law 38

32 General/Miscellaneous Clauses 38

Page 26: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 26 of 166 01.09.2010

GENERAL CONDITIONS OF CONTRACT (GCC) 1. Application The General Conditions of Contract incorporated in this section shall be applicable for this

purchase to the extent the same are not superseded by the Special Conditions of Contract

prescribed under Section V, List of requirements under Section VI and Technical Specification

under Section VII of this document.

2. Use of contract documents and information

2.1 The supplier shall not, without the purchaser’s prior written consent, disclose the contract or any

provision thereof including any specification, drawing, sample or any information furnished by or

on behalf of the purchaser in connection therewith, to any person other than the person(s)

employed by the supplier in the performance of the contract emanating from this TE document.

Further, any such disclosure to any such employed person shall be made in confidence and only so

far as necessary for the purposes of such performance for this contract.

2.2 Further, the supplier shall not, without the purchaser’s prior written consent, make use of any

document or information mentioned in GCC sub-clause 2.1 above except for the sole purpose of

performing this contract.

2.3 Except the contract issued to the supplier, each and every other document mentioned in GCC sub-

clause 2.1 above shall remain the property of the purchaser and, if advised by the purchaser, all

copies of all such documents shall be returned to the purchaser on completion of the supplier’s

performance and obligations under this contract.

3. Patent Rights

The supplier shall, at all times, indemnify and keep indemnified the purchaser, free of cost,

against all claims which may arise in respect of goods & services to be provided by the supplier

under the contract for infringement of any intellectual property rights or any other right protected

by patent, registration of designs or trademarks. In the event of any such claim in respect of

alleged breach of patent, registered designs, trade marks etc. being made against the purchaser, the

purchaser shall notify the supplier of the same and the supplier shall, at his own expenses take

care of the same for settlement without any liability to the purchaser.

4. Country of Origin

4.1 All goods and services to be supplied and provided for the contract shall have the origin in India

or in the countries with which the Government of India has trade relations.

4.2 The word “origin” incorporated in this clause means the place from where the goods are mined,

cultivated, grown, manufactured, produced or processed or from where the services are arranged.

4.3 The country of origin may be specified in the Price Schedule

5. Performance Security

5.1 Within twenty-one (21) days from date of the issue of notification of award by the

Purchaser/Consignee, the supplier, shall furnish performance security to the Purchaser/Consignee

for an amount equal to ten percent (10%) of the total value of the contract, valid up to sixty (60)

days after the date of completion of all contractual obligations by the supplier, including the warranty obligations.

5.2 The Performance security shall be denominated in Indian Rupees or in the currency of the contract

as detailed below:

a) It shall be in any one of the forms namely Account Payee Demand Draft drawn from any

Scheduled bank in India or Bank Guarantee issued by a Scheduled bank in India, in the

Page 27: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 27 of 166 01.09.2010

prescribed form as provided in section XV of this document in favour of the

Purchaser/Consignee.

b) In the case of Bank Guarantee furnished from banks outside India (i.e. foreign Banks), it

should be authenticated and countersigned by any nationalised bank in India by way of

back-to-back counter guarantee.

5.3 In the event of any failure /default of the supplier with or without any quantifiable loss to the

consignee/purchaser including furnishing of consignee wise Bank Guarantee for CMC security as

per Proforma in Section XV, the amount of the performance security is liable to be forfeited. The

Administration Department may do the needful to cover any failure/default of the supplier with or

without any quantifiable loss to the Government.

5.4 In the event of any amendment issued to the contract, the supplier shall, within twenty-one (21)

days of issue of the amendment, furnish the corresponding amendment to the Performance

Security (as necessary), rendering the same valid in all respects in terms of the contract, as

amended.

5.5 The supplier shall enter into Comprehensive Maintenance Contract as per the ‘Contract Form - B’

in Section XVI with respective consignees, 3 (three) months prior to the completion of Warranty

Period. The CMC will commence from the date of expiry of the Warranty Period.

5.6 Subject to GCC sub – clause 5.3 above, the Purchaser/Consignee will release the Performance

Security without any interest to the supplier on completion of the supplier’s all contractual

obligations including the warranty obligations & after receipt of Consignee wise bank guarantee

for CMC security in favour of Head of the Hospital/ Institute/ Medical College of the consignee as

per the format in Section XV.

6. Technical Specifications and Standards

6.1 The Goods & Services to be provided by the supplier under this contract shall conform to the

technical specifications and quality control parameters mentioned in ‘Technical Specification’ and

‘Quality Control Requirements’ under Sections VII and VIII of this document.

7. Packing and Marking

7.1 The packing for the goods to be provided by the supplier should be strong and durable enough to

withstand, without limitation, the entire journey during transit including transhipment (if any),

rough handling, open storage etc. without any damage, deterioration etc. As and if necessary, the

size, weights and volumes of the packing cases shall also take into consideration, the remoteness

of the final destination of the goods and availability or otherwise of transport and handling

facilities at all points during transit up to final destination as per the contract.

7.2 The quality of packing, the manner of marking within & outside the packages and provision of

accompanying documentation shall strictly comply with the requirements as provided in

Technical Specifications and Quality Control Requirements under Sections VII and VIII and in

SCC under Section V. In case the packing requirements are amended due to issue of any

amendment to the contract, the same shall also be taken care of by the supplier accordingly.

7.3 Packing instructions:

Unless otherwise mentioned in the Technical Specification and Quality Control Requirements

under Sections VII and VIII and in SCC under Section V, the supplier shall make separate

packages for each consignee (in case there is more than one consignee mentioned in the contract)

and mark each package on three sides with the following with indelible paint of proper quality:

a. contract number and date

b. brief description of goods including quantity

c. packing list reference number

d. country of origin of goods

Page 28: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 28 of 166 01.09.2010

e. consignee’s name and full address and

f. supplier’s name and address

8. Inspection, Testing and Quality Control

8.1 The purchaser and/or its nominated representative(s) will, without any extra cost to the purchaser,

inspect and/or test the ordered goods and the related services to confirm their conformity to the

contract specifications and other quality control details incorporated in the contract. The purchaser

shall inform the supplier in advance, in writing, the purchaser’s programme for such inspection

and, also the identity of the officials to be deputed for this purpose. The cost towards the

transportation, boarding & lodging will be borne by the purchaser and/or its nominated

representative(s).

8.2 The Technical Specification and Quality Control Requirements incorporated in the contract shall

specify what inspections and tests are to be carried out and, also, where and how they are to be

conducted. If such inspections and tests are conducted in the premises of the supplier or its

subcontractor(s), all reasonable facilities and assistance, including access to relevant drawings,

design details and production data, shall be furnished by the supplier to the purchaser’s inspector

at no charge to the purchaser.

8.3 If during such inspections and tests the contracted goods fail to conform to the required

specifications and standards, the purchaser’s inspector may reject them and the supplier shall

either replace the rejected goods or make all alterations necessary to meet the specifications and

standards, as required, free of cost to the purchaser and resubmit the same to the purchaser’s

inspector for conducting the inspections and tests again.

8.4 In case the contract stipulates pre-despatch inspection of the ordered goods at supplier’s premises,

the supplier shall put up the goods for such inspection to the purchaser’s inspector well ahead of

the contractual delivery period, so that the purchaser’s inspector is able to complete the inspection

within the contractual delivery period.

8.5 If the supplier tenders the goods to the purchaser’s inspector for inspection at the last moment

without providing reasonable time to the inspector for completing the inspection within the

contractual delivery period, the inspector may carry out the inspection and complete the formality

beyond the contractual delivery period at the risk and expense of the supplier. The fact that the

goods have been inspected after the contractual delivery period will not have the effect of keeping

the contract alive and this will be without any prejudice to the legal rights and remedies available

to the purchaser under the terms & conditions of the contract.

8.6 The purchaser’s/consignee’s contractual right to inspect, test and, if necessary, reject the goods

after the goods’ arrival at the final destination shall have no bearing of the fact that the goods have

previously been inspected and cleared by purchaser’s inspector during pre-despatch inspection

mentioned above.

8.7 Goods accepted by the purchaser/consignee and/or its inspector at initial inspection and in final

inspection in terms of the contract shall in no way dilute purchaser’s/consignee’s right to reject the

same later, if found deficient in terms of the warranty clause of the contract, as incorporated under

GCC Clause 15.

8.8 Principal/ Foreign supplier shall also have the equipment inspected by recognised/ reputed agency

like SGS, Lloyd or equivalent (acceptable to the purchaser) prior to despatch at the supplier’s cost

and furnish necessary certificate from the said agency in support of their claim.

9. Terms of Delivery

Goods shall be delivered by the supplier in accordance with the terms of delivery specified in the

contract.

10. Transportation of Goods

Page 29: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 29 of 166 01.09.2010

10.1 Instructions for transportation of imported goods offered from abroad:

The supplier shall not arrange part-shipments and/or transhipment without the express/prior

written consent of the purchaser. The supplier is required under the contract to deliver the goods

under DDP at consignee site; the shipment shall be made by Indian flag vessel or by vessels

belonging to the conference lines in which India is a member country through India’s forwarding

agents/coordinators. In case the forwarding agent/coordinators are unable to provide timely

adequate space in Indian flag vessel or by vessels belonging to the conference lines, the supplier

shall arrange shipment through any available vessel to adhere to the delivery schedule given in the

contract.

In case of airlifting of imported goods offered from abroad, the same will be done only through

the National Carrier i.e. Air India wherever applicable. In case the National Carrier is not

available, any other airlines available for early delivery may be arranged.

Goods will be custom cleared by the supplier/ Indian agent and transported to the consignee’s site

as per the contract terms.

Instructions for transportation of domestic goods including goods already imported by the supplier

under its own arrangement:

In case no instruction is provided in this regard in the SCC, the supplier will arrange

transportation of the ordered goods as per its own procedure.

The goods shall be custom cleared by the Supplier/ Indian Agent and transported to the

consignee’s site as per contract terms. The supplier shall be responsible for safe and timely

delivery of ordered goods under his own arrangement.

11. Insurance:

Unless otherwise instructed in the SCC, the supplier shall make arrangements for insuring the

goods against loss or damage incidental to manufacture or acquisition, transportation, storage and

delivery in the following manner:

i) in case of supply of domestic goods on Consignee site basis, the supplier shall be responsible

till the entire stores contracted for arrival in good condition at destination. The transit risk in

this respect shall be covered by the Supplier by getting the stores duly insured. The insurance

cover shall be obtained by the Supplier and should be valid till 3 months after the receipt of

goods by the Consignee.

ii) in case of supply of the imported goods on DDP Basis, the supplier shall arrange and pay for

marine/ air insurance making the consignee as beneficiary. The additional extended Insurance

(local transportation and storage) would also be borne by the Supplier from the port of entry to

the consignee site for a period including 3 months beyond date of delivery.

12. Spare parts

12.1 If specified in the List of Requirements and in the resultant contract, the supplier shall

supply/provide any or all of the following materials, information etc. pertaining to spare parts

manufactured and/or supplied by the supplier:

a) The spare parts as selected by the Purchaser/Consignee to be purchased from the supplier,

subject to the condition that such purchase of the spare parts shall not relieve the supplier

of any contractual obligation including warranty obligations; and

Page 30: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 30 of 166 01.09.2010

b) In case the production of the spare parts is discontinued:

i) Sufficient advance notice to the Purchaser/Consignee before such discontinuation

to provide adequate time to the purchaser to purchase the required spare parts etc.,

and

ii) Immediately following such discontinuation, providing the Purchaser/Consignee,

free of cost, the designs, drawings, layouts and specifications of the spare parts, as

and if requested by the Purchaser/Consignee.

12.2 Supplier shall carry sufficient inventories to assure ex-stock supply of consumable spares for the

goods so that the same are supplied to the Purchaser/Consignee promptly on receipt of order from

the Purchaser/Consignee.

13. Incidental services

Subject to the stipulation, if any, in the SCC (Section – V), List of Requirements (Section – VI)

and the Technical Specification (Section – VII), the supplier shall be required to perform the

following services.

i) Installation & commissioning, Supervision and Demonstration of the goods

ii) Providing required jigs and tools for assembly, minor civil works required for the

completion of the installation.

iii) Training of Consignee’s Doctors, Staff, operators etc. for operating and maintaining

the goods

iv) Supplying required number of operation & maintenance manual for the goods

14. Distribution of Dispatch Documents for Clearance/Receipt of Goods

The supplier shall send all the relevant despatch documents well in time to the

Purchaser/Consignee

Unless otherwise specified in the SCC, the usual documents involved and the drill to be followed

in general for this purpose are as follows.

A) For Domestic Goods, including goods already imported by the supplier under its own

arrangement

Within 24 hours of despatch, the supplier shall notify the purchaser, consignee, and others

concerned if mentioned in the contract, the complete details of despatch and also supply the

following documents to them by registered post / speed post (or as instructed in the contract):

(i) Four copies of supplier’s invoice showing contract number, goods description,

quantity, unit price and total amount;

(ii) Two copies of packing list identifying contents of each package;

(iii) Inspection certificate issued by the nominated Inspection agency, if any.

(iv) Certificate of origin;

(v) Insurance Certificate as per GCC Clause 11.

(vi) Manufacturers/Supplier’s warranty certificate & In-house inspection certificate.

B) For goods imported from abroad

Within 24 hours of despatch, the supplier shall notify the purchaser, consignee, and others

concerned if mentioned in the contract, the complete details of despatch and also supply the

Page 31: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 31 of 166 01.09.2010

following documents to them by airmail/ registered post / speed post (or as instructed in the

contract).

(i) Four copies of supplier’s invoice showing contract number, goods description,

quantity, unit price and total amount;

(ii) Original and four copies of the negotiable clean, on-board Bill of Lading/Airway

bill, marked freight pre paid and four copies of non-negotiable Bill of

Lading/Airway bill;

(iii) Four Copies of packing list identifying contents of each package;

(iv) Insurance Certificate as per GCC Clause 11.

(v) Manufacturer’s/Supplier’s warranty certificate;

(vi) Inspection Certificate for the despatched equipments issued by recognized/ reputed

agency like SGS, Lloyd or equivalent (acceptable to the purchaser) prior to despatch

(vii) Manufacturer’s own factory inspection report;

(viii) Certificate of origin

(ix) Port of Loading;

(x) Port of Discharge and

(xi) Expected date of arrival.

15. Warranty

15.1 The supplier warrants comprehensively that the goods supplied under the contract is new, unused

and incorporate all recent improvements in design and materials unless prescribed otherwise by

the purchaser in the contract. The supplier further warrants that the goods supplied under the

contract shall have no defect arising from design, materials or workmanship or from any act or

omission of the supplier that may develop under normal use of the supplied goods under the

conditions prevailing in India.

15.2 This warranty shall remain valid for 2 (Two) years in general, after the goods or any portion

thereof as the case may be, have been delivered to the final destination and installed and

commissioned at the final destination and accepted by the Purchaser/Consignee in terms of the

contract, unless specified otherwise in the SCC.

a. No conditional warranty like mishandling, manufacturing defects etc. will be acceptable.

b. Warranty as well as Comprehensive Maintenance contract will be inclusive of all

accessories and Turnkey work and it will also cover the following items:-

i. X-ray and CT tubes and high-tension cables.

ii. Helium replacement

iii. Any kind of motor

iv. Plastic & Glass parts

v. All kinds of sensors including oxygen sensors

vi. All kinds of coils, probes and transducers including ECG cable, BP transducers,

SpO2 Probes, Ultrasound and Color Doppler Transducers/probes, BP Cuffs,

Defribrillator internal paddles, chart recorders, ventilator reusable patient circuits,

servo humidifier with chamber, electrodes and probes for blood gas analyser, MRI

coils.

vii. All kinds of flat panel sensors and cassettes for Digital Radiography & Computer

Radiography systems and patients handling trolleys, etc.

viii. Printers and imagers including laser and thermal printers with all parts.

ix. UPS including the replacement of Batteries.

x. Air-conditioners

c. Replacement and repair will be under taken for the defective goods.

d. Proper marking has to be made for all spares for identification like printing of installation

and repair dates.

Page 32: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 32 of 166 01.09.2010

15.3 In case of any claim arising out of this warranty, the Purchaser/Consignee shall promptly notify

the same in writing to the supplier. The period of the warranty will be as per G.C.C clause number

15.2 above irrespective of any other period mentioned elsewhere in the bidding documents.

15.4 Upon receipt of such notice, the supplier shall, within 8 hours on a 24(hrs) X 7 (days) X 365

(days) basis respond to take action to repair or replace the defective goods or parts thereof, free of

cost, at the ultimate destination. The supplier shall take over the replaced parts/goods after

providing their replacements and no claim, whatsoever shall lie on the purchaser for such replaced

parts/goods thereafter. The penalty clause for non-rectification will be applicable as per tender

conditions

15.5 In the event of any rectification of a defect or replacement of any defective goods during the

warranty period, the warranty for the rectified/replaced goods shall be extended to a further

period as mentioned under clause 15.2 from the date such rectified / replaced goods starts

functioning to the satisfaction of the purchaser.

15.6 If the supplier, having been notified, fails to respond to take action to repair or replace the

defect(s) within 8 hours on a 24(hrs) X 7 (days) X 365 (days) basis, the purchaser may proceed to

take such remedial action(s) as deemed fit by the purchaser, at the risk and expense of the supplier

and without prejudice to other contractual rights and remedies, which the purchaser may have

against the supplier.

15.7 During Warranty period, the supplier is required to visit at each consignee’s site at least once in 6

months commencing from the date of the installation for preventive maintenance of the goods

15.8 The Purchaser/Consignee reserve the rights to enter into Annual Comprehensive Maintenance

Contract between Consignee and the Supplier for the period as mentioned in General Points for

Technical Specifications, Section VII (para-4), after the completion of warranty period.

15.9 The supplier along with its Indian Agent and the CMC provider shall ensure continued supply of

the spare parts for the machines and equipments supplied by them to the purchaser for 10 years

from the date of installation and handing over.

15.10 The Supplier along with its Indian Agent and the CMC Provider shall always accord most

favoured client status to the Purchaser vis-à-vis its other Clients/Purchasers of its

equipments/machines/goods etc. and shall always give the most competitive price for its

machines/equipments supplied to the Purchaser/Consignee.

16. Assignment

16.1 The Supplier shall not assign, either in whole or in part, its contractual duties, responsibilities and

obligations to perform the contract, except with the Purchaser’s prior written permission.

17. Sub Contracts

17.1 The Supplier shall notify the Purchaser in writing of all sub contracts awarded under the contract

if not already specified in its tender. Such notification, in its original tender or later, shall not

relieve the Supplier from any of its liability or obligation under the terms and conditions of the

contract.

17.2 Sub contract shall be only for bought out items and sub-assemblies.

17.3 Sub contracts shall also comply with the provisions of GCC Clause 4 (“Country of Origin”).

18. Modification of contract

18.1 If necessary, the purchaser may, by a written order given to the supplier at any time during the

currency of the contract, amend the contract by making alterations and modifications within the

general scope of contract in any one or more of the following:

a) Specifications, drawings, designs etc. where goods to be supplied under the contract are

to be specially manufactured for the purchaser,

b) Mode of packing,

c) Incidental services to be provided by the supplier

Page 33: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 33 of 166 01.09.2010

d) Mode of despatch,

e) Place of delivery, and

f) Any other area(s) of the contract, as felt necessary by the purchaser depending on the

merits of the case.

18.2 In the event of any such modification/alteration causing increase or decrease in the cost of goods

and services to be supplied and provided, or in the time required by the supplier to perform any

obligation under the contract, an equitable adjustment shall be made in the contract price and/or

contract delivery schedule, as the case may be, and the contract amended accordingly. If the

supplier doesn’t agree to the adjustment made by the Purchaser/Consignee, the supplier shall

convey its views to the Purchaser/Consignee within twenty-one days from the date of the

supplier’s receipt of the Purchaser’s/Consignee’s amendment / modification of the contract.

19. Prices

19.1 Prices to be charged by the supplier for supply of goods and provision of services in terms of the

contract shall not vary from the corresponding prices quoted by the supplier in its tender and

incorporated in the contract except for any price adjustment authorised in the SCC.

20. Taxes and Duties

20.1 Supplier shall be entirely responsible for all taxes, duties, fees, levies etc. incurred until delivery

of the contracted goods to the purchaser.

20.2 Further instruction, if any, shall be as provided in the SCC.

21. Terms and Mode of Payment

21.1 Payment Terms

Payment shall be made subject to recoveries, if any, by way of liquidated damages or any

other charges as per terms & conditions of contract in the following manner.

A) Payment for Domestic Goods Or Foreign Origin Located Within India.

Payment shall be made in Indian Rupees as specified in the contract in the following manner:

a) On delivery:

90 % payment of the contract price shall be paid on receipt of goods in good condition and

upon the submission of the following documents:

(i) Four copies of supplier’s invoice showing contract number, goods description, quantity,

unit price and total amount;

(ii) Consignee Receipt Certificate as per Section XVII in original issued by the authorized

representative of the consignee;

(iii) Two copies of packing list identifying contents of each package;

(iv) Inspection certificate issued by the nominated Inspection agency, if any;

(v) Insurance Certificate as per GCC Clause 11;

(vi) Certificate of origin.

b) On Acceptance:

Balance 10 % payment would be made against ‘Final Acceptance Certificate’ as per

Section XVIII of goods to be issued by the consignees subject to recoveries, if any, either

on account of non-rectification of defects/deficiencies not attended by the Supplier or

otherwise.

B) Payment for Imported Goods:

Page 34: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 34 of 166 01.09.2010

Payment for foreign currency portion shall be made in the currency as specified in the contract in

the following manner:

a) On delivery:

Ninety (90) % of the net CIP price (CIP price less Indian Agency commission) of the goods

shipped shall be paid through irrevocable, non-transferable Letter of Credit (LC) opened in

favour of the supplier in a bank in his country and upon submission of documents specified

hereunder:

(i) Four copies of supplier’s invoice showing contract number, goods description, quantity,

unit price and total amount;

(ii) Original and four copies of the negotiable clean, on-board Bill of Lading/ Airway bill ,

marked freight pre paid and four copies of non-negotiable Bill of Lading/Airway bill;

(iii) Four Copies of packing list identifying contents of each package;

(iv) Insurance Certificate as per GCC Clause 11 and documents also to be submitted for

payment of LC confirming that dispatch documents has already been sent within 24

hours to all concerned as per the contract;

(v) Manufacturer’s/Supplier’s warranty certificate;

(vi) Manufacturer’s own factory inspection report and

(vii) Certificate of origin by the chamber of commerce of the concerned country;

(viii) Inspection Certificate for the despatched equipments issued by recognized/ reputed

agency like SGS, Lloyd or equivalent (acceptable to the purchaser) prior to despatch.

(ix) Consignee Receipt Certificate as per Section XVII in original issued by the authorized

representative of the consignee

b) On Acceptance:

Balance payment of 10 % of net CIP price of goods would be made against ‘Final Acceptance

Certificate’ as per Section XVIII to be issued by the consignees through irrevocable, non-

transferable Letter of Credit (LC) opened in favour of the Foreign Principal in a bank in his

country, subject to recoveries, if any.

c) Payment of custom duty amount with Custom Duty Exemption Certificate (CDEC), if

applicable, customs clearance and handling charges, loading/ unloading, inland

transportation, incidental costs till consignee site & incidental services (including installation & commissioning, supervision, demonstration and training) will be paid in

Indian Rupees to the Indian agent at actual not exceeding the quoted rates after 100 %

payment to the foreign principal.

d) Payment of Indian Agency Commission:

Indian Agency commission will be paid to the manufacturer’s agent in the local currency for

an amount in Indian rupees indicated in the relevant Price Schedule (as per prevailing rate of

exchange ruling on the date of Contract) and shall not be subject to further escalation /

exchange variation. Payment shall be made after 100% payment to the Foreign Principal.

C) Payment of Turnkey, if any:

Turnkey payment will be made to the manufacturer’s agent in Indian rupees indicated in the

relevant Price Schedule (as per prevailing rate of exchange ruling on the date of Contract) and

shall not be subject to further escalation / exchange variation. Payment shall be made to the Indian

Agent after 100 % payment to the Foreign Principal.

Page 35: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 35 of 166 01.09.2010

D) Payment for Annual Comprehensive Maintenance Contract (CMC) Charges:

The consignee will enter into CMC with the supplier at the rates as stipulated in the contract. The

payment of CMC will be made on six monthly basis after satisfactory completion of said period,

duly certified by the consignee on receipt of bank guarantee for an amount equivalent to 2.5 % of

the cost of the equipment as per contract in the prescribed format given in Section XV valid till 2

months after expiry of entire CMC period.

21.2 The supplier shall not claim any interest on payments under the contract.

21.3 Where there is a statutory requirement for tax deduction at source, such deduction towards income

tax and other tax as applicable will be made from the bills payable to the Supplier at rates as

notified from time to time.

21.4 Irrevocable & non-transferable LC shall be opened by ESIC/ Purchaser. However, if the supplier

requests specifically to open confirmed LC, the extra charges would be borne by the supplier. If

LC is required to be extended and/or amended for reasons not attributable to the purchaser/

consignee, the charges thereof shall be borne by the supplier.

21.5 The payment shall be made in the currency / currencies authorised in the contract.

21.6 The supplier shall send its claim for payment in writing, when contractually due, along with

relevant documents etc., duly signed with date, to the purchaser.

21.7 While claiming payment, the supplier is also to certify in the bill that the payment being claimed

is strictly in terms of the contract and all the obligations on the part of the supplier for claiming

that payment has been fulfilled as required under the contract.

21.8 While claiming reimbursement of duties, taxes etc. (like sales tax, excise duty, custom duty) from

the Purchaser/Consignee, as and if permitted under the contract, the supplier shall also certify that,

in case it gets any refund out of such taxes and duties from the concerned authorities at a later

date, it (the supplier) shall refund to the Purchaser/Consignee forthwith.

21.9 In case where the supplier is not in a position to submit its bill for the balance payment for want of

receipted copies of Inspection Note from the consignee and the consignee has not complained

about the non-receipt, shortage, or defects in the supplies made, balance amount will be paid by

the paying authority without consignee’s receipt certificate after three months from the date of the

preceding part payment for the goods in question, subject to the following conditions:

(a) The supplier will make good any defect or deficiency that the consignee (s) may report within

six months from the date of despatch of goods.

(b) Delay in supplies, if any, has been regularized.

(c) The contract price where it is subject to variation has been finalized.

(d) The supplier furnishes the following undertakings:

“I/We, __________ certify that I/We have not received back the Inspection Note duly receipted by the

consignee or any communication from the purchaser or the consignee about non-receipt, shortage or

defects in the goods supplied. I/We ______ agree to make good any defect or deficiency that the

consignee may report within three months from the date of receipt of this balance payment.

22. Delay in the supplier’s performance

22.1 The supplier shall deliver the goods and perform the services under the contract within the time

schedule specified by the Purchaser/Consignee in the List of Requirements and as incorporated in

the contract.

22.2 Subject to the provision under GCC clause 26, any unexcused delay by the supplier in maintaining

its contractual obligations towards delivery of goods and performance of services shall render the

supplier liable to any or all of the following sanctions:

(i) imposition of liquidated damages,

Page 36: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 36 of 166 01.09.2010

(ii) forfeiture of its performance security and

(iii) termination of the contract for default.

22.3 If at any time during the currency of the contract, the supplier encounters conditions hindering

timely delivery of the goods and performance of services, the supplier shall promptly inform the

Purchaser/Consignee in writing about the same and its likely duration and make a request to the

Purchaser/Consignee for extension of the delivery schedule accordingly. On receiving the

supplier’s communication, the Purchaser/Consignee shall examine the situation as soon as

possible and, at its discretion, may agree to extend the delivery schedule, with or without

liquidated damages for completion of supplier’s contractual obligations by issuing an amendment

to the contract.

22.4 When the period of delivery is extended due to unexcused delay by the supplier, the amendment

letter extending the delivery period shall, interalia contain the following conditions:

(a) The Purchaser/Consignee shall recover from the supplier, under the provisions of the clause 23

of the General Conditions of Contract, liquidated damages on the goods and services, which

the Supplier has failed to deliver within the delivery period stipulated in the contract.

(b) That no increase in price on account of any ground, whatsoever, including any stipulation in

the contract for increase in price on any other ground and, also including statutory increase in

or fresh imposition of customs duty, excise duty, sales tax/ VAT, Service Tax and Works

Contract Tax or on account of any other tax or duty which may be levied in respect of the

goods and services specified in the contract, which takes place after the date of delivery

stipulated in the contract shall be admissible on such of the said goods and services as are

delivered and performed after the date of the delivery stipulated in the contract.

(c) But nevertheless, the Purchaser/Consignee shall be entitled to the benefit of any decrease in

price on account of reduction in or remission of customs duty, excise duty, sales tax/ VAT,

Service Tax and Works Contract Tax or any other duty or tax or levy or on account of any

other grounds, which takes place after the expiry of the date of delivery stipulated in the

contract.

22.5 The supplier shall not dispatch the goods after expiry of the delivery period. The supplier is

required to apply to the Purchaser/Consignee for extension of delivery period and obtain the same

before despatch. In case the supplier dispatches the goods without obtaining an extension, it would

be doing so at its own risk and no claim for payment for such supply and / or any other expense

related to such supply shall lie against the purchaser.

23. Liquidated damages

23.1 Subject to GCC clause 26, if the supplier fails to deliver any or all of the goods or fails to perform

the services within the time frame(s) incorporated in the contract, the Purchaser/Consignee shall,

without prejudice to other rights and remedies available to the Purchaser/Consignee under the

contract, deduct from the contract price, as liquidated damages, a sum equivalent to 0.5% per

week of delay or part thereof on delayed supply of goods and/or services until actual delivery or

performance subject to a maximum of 10% of the contract price. Once the maximum is reached

Purchaser/ Consignee may consider termination of the contract as per GCC 24.

During the above-mentioned delayed period of supply and / or performance, the conditions

incorporated under GCC sub-clause 22.4 above shall also apply.

24. Termination for default

24.1 The Purchaser/Consignee, without prejudice to any other contractual rights and remedies available

to it (the Purchaser/Consignee), may, by written notice of default sent to the supplier, terminate

the contract in whole or in part, if the supplier fails to deliver any or all of the goods or fails to

Page 37: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 37 of 166 01.09.2010

perform any other contractual obligation(s) within the time period specified in the contract, or

within any extension thereof granted by the Purchaser/Consignee pursuant to GCC sub-clauses

22.3 and 22.4.

24.2 In the event of the Purchaser/Consignee terminates the contract in whole or in part, pursuant to

GCC sub-clause 24.1 above, the Purchaser/Consignee may procure goods and/or services similar

to those cancelled, with such terms and conditions and in such manner as it deems fit and the

supplier shall be liable to the Purchaser/Consignee for the extra expenditure, if any, incurred by

the Purchaser/Consignee for arranging such procurement.

24.3 Unless otherwise instructed by the Purchaser/Consignee, the supplier shall continue to perform the

contract to the extent not terminated.

25. Termination for insolvency

25.1 If the supplier becomes bankrupt or otherwise insolvent, the purchaser reserves the right to

terminate the contract at any time, by serving written notice to the supplier without any

compensation, whatsoever, to the supplier, subject to further condition that such termination will

not prejudice or affect the rights and remedies which have accrued and / or will accrue thereafter

to the Purchaser/Consignee.

26. Force Majeure

26.1 Notwithstanding the provisions contained in GCC clauses 22, 23 and 24, the supplier shall not be

liable for imposition of any such sanction so long the delay and/or failure of the supplier in

fulfilling its obligations under the contract is the result of an event of Force Majeure.

26.2 For purposes of this clause, Force Majeure means an event beyond the control of the supplier and

not involving the supplier’s fault or negligence and which is not foreseeable and not brought about

at the instance of, the party claiming to be affected by such event and which has caused the non –

performance or delay in performance. Such events may include, but are not restricted to, acts of

the Purchaser/Consignee either in its sovereign or contractual capacity, wars or revolutions,

hostility, acts of public enemy, civil commotion, sabotage, fires, floods, explosions, epidemics,

quarantine restrictions, strikes excluding by its employees, lockouts excluding by its management,

and freight embargoes.

26.3 If a Force Majeure situation arises, the supplier shall promptly notify the Purchaser/Consignee in

writing of such conditions and the cause thereof within twenty-one days of occurrence of such

event. Unless otherwise directed by the Purchaser/Consignee in writing, the supplier shall

continue to perform its obligations under the contract as far as reasonably practical, and shall seek

all reasonable alternative means for performance not prevented by the Force Majeure event.

26.4 If the performance in whole or in part or any obligation under this contract is prevented or delayed

by any reason of Force Majeure for a period exceeding sixty days, either party may at its option

terminate the contract without any financial repercussion on either side.

26.5 In case due to a Force Majeure event the Purchaser/Consignee is unable to fulfil its contractual

commitment and responsibility, the Purchaser/Consignee will notify the supplier accordingly and

subsequent actions taken on similar lines described in above sub-paragraphs.

27. Termination for convenience

27.1 The Purchaser/Consignee reserves the right to terminate the contract, in whole or in part for its

(Purchaser’s/Consignee ’s) convenience, by serving written notice on the supplier at any time

during the currency of the contract. The notice shall specify that the termination is for the

convenience of the Purchaser/Consignee. The notice shall also indicate interalia, the extent to

which the supplier’s performance under the contract is terminated, and the date with effect from

which such termination will become effective.

27.2 The goods and services that are complete and ready in terms of the contract for delivery and

performance within thirty days after the supplier’s receipt of the notice of termination shall be

Page 38: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 38 of 166 01.09.2010

accepted by the Purchaser/Consignee following the contract terms, conditions and prices. For the

remaining goods and services, the Purchaser/Consignee may decide:

a) To get any portion of the balance completed and delivered at the contract terms, conditions

and prices; and / or

b) To cancel the remaining portion of the goods and services and compensate the supplier by

paying an agreed amount for the cost incurred by the supplier towards the remaining portion

of the goods and services.

28. Governing language

28.1 The contract shall be written in English language following the provision as contained in GIT

clause 4. All correspondence and other documents pertaining to the contract, which the parties

exchange, shall also be written accordingly in that language.

29. Notices

29.1 Notice, if any, relating to the contract given by one party to the other, shall be sent in writing or by

cable or telex or facsimile and confirmed in writing. The procedure will also provide the sender of

the notice, the proof of receipt of the notice by the receiver. The addresses of the parties for

exchanging such notices will be the addresses as incorporated in the contract.

29.2 The effective date of a notice shall be either the date when delivered to the recipient or the

effective date specifically mentioned in the notice, whichever is later.

30. Resolution of disputes

30.1 If dispute or difference of any kind shall arise between the Purchaser/Consignee and the supplier

in connection with or relating to the contract, the parties shall make every effort to resolve the

same amicably by mutual consultations. The jurisdiction for the settlement of disputes will be at

New Delhi, India.

30.2 If the parties fail to resolve their dispute or difference by such mutual consultation within twenty-

one days of its occurrence, then, unless otherwise provided in the SCC, either the

Purchaser/Consignee or the supplier may give notice to the other party of its intention to

commence arbitration, as hereinafter provided the applicable arbitration procedure will be as per

the Arbitration and Conciliation Act, 1996 of India. In the case of a dispute or difference arising

between the Purchaser/Consignee and a domestic Supplier relating to any matter arising out of or

connected with the contract, such dispute or difference shall be referred to the sole arbitration of

an officer, appointed to be the arbitrator by the Director General of ESIC. The award of the

arbitrator shall be final and binding on the parties to the contract subject to the provision that the

Arbitrator shall give reasoned award in case the value of claim in reference exceeds Rupees One

Lac (Rs. 1,00,000/-)

30.3 Venue of Arbitration: The venue of arbitration shall be New Delhi, India.

31. Applicable Law

The contract shall be governed by and interpreted in accordance with the laws of India for the

time being in force.

32. General/ Miscellaneous Clauses

32.1 Nothing contained in this Contract shall be constructed as establishing or creating between the

parties, i.e. the Supplier/its Indian Agent/CMC Provider on the one side and the Purchaser on the

other side, a relationship of master and servant or principal and agent.

32.2 Any failure on the part of any Party to exercise right or power under this Contract shall not

operate as waiver thereof.

32.3 The Supplier shall notify the Purchaser/Consignee of any material change would impact on

performance of its obligations under this Contract.

Page 39: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited GCC

HLL/PCD/ESIC-20/10-11 Page 39 of 166 01.09.2010

32.4 Each member/constituent of the Supplier/its Indian Agent/CMC Provider, in case of consortium

shall be jointly and severally liable to and responsible for all obligations towards the

Purchaser/Consignee for performance of contract/services including that of its Associates/Sub

Contractors under the Contract.

32.5 The Supplier/its Indian Agent/CMC Provider shall at all times, indemnify and keep indemnified

the Purchaser against all claims/damages etc. for any infringement of any Intellectual Property

Rights (IPR) while providing its services under CMC or the Contract.

32.6 The Supplier/its Agent/CMC Provider shall, at all times, indemnify and keep indemnified the

Purchaser/Consignee against any claims in respect of any damages or compensation payable in

consequences of any accident or injury sustained or suffered by its employees or agents or by any

other third party resulting from or by any action, omission or operation conducted by or on behalf

of the supplier/its associate/affiliate etc.

32.7 All claims regarding indemnity shall survive the termination or expiry of the contract

Page 40: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited SCC

HLL/PCD/ESIC-20/10-11 Page 40 of 166 01.09.2010

SECTION – V

SPECIAL CONDITIONS OF CONTRACT (SCC)

The following Special Conditions of Contract (SCC) will apply for this purchase. The corresponding

clauses of General Conditions of Contract (GCC) relating to the SCC stipulations have also been

incorporated below.

These Special Conditions will modify/substitute/supplement the corresponding (GCC) clauses.

Whenever there is any conflict between the provision in the GCC and that in the SCC, the provision

contained in the SCC shall prevail.

Page 41: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited List of Requirement

HLL/PCD/ESIC-20/10-11 Page 41 of 166 01.09.2010

SECTION – VI

LIST OF REQUIREMENTS

Part I: Consignee wise Qty.

Sl. No.

Item Description Adityapur Indore Joka Baddi Manesar Bhiwadi Ludhiana

Basai Darapur

Total Qty

1 Biphasic Defibrilator 2 1 5 8 2 Fiber Optic Bronchoscope with Camera 1 1 1 1 1 1 6 3 Pulse Oximeter 2 1 3 4 Anaesthesia Workstation 2 2 5 ECG Machine 12 channel 1 1 2 6 Pharmaceutical Refrigerator for O.T. 1 1 7 O T Tables for Orthopaedics 1 1 1 1 1 5

8 O T Tables for General Surgery & Gynaecology 2 3 3 1 4

13

9 Electro Surgical Unit 2 3 5 10 O T Table for Minor O T 1 1 11 Boyle's Appratus 1 1

12 Multipara Monitor (6Nos) with Central Station (1N0) 1 1

2

13 Portable Ventilator 1 1 14 BIPAP 2 2 15 Binocular Microscope 1 3 2 6 16 Urine Analyzer 1 1 17 ELISA Reader with Washer 1 1 1 1 1 5

18 Cell Counter(Fully automated 5 part differential analyser) 1 1

2

19 Biochemistry Semi- Auto Analyser 1 1 2 20 Coagulation Analyzer 1 1 2

21 Electrolyte Analyzer 1 1 22 500 M.A. X-ray machine 1 1 23 Portable X-ray machine (Mobile) 1 2 3 24 Colour Doppler Digital 1 1 25 Infant Resuscitator 1 1 26 Neonatal Monitor 1 2 2 1 6 27 Ventilator Critical Care 1 1

28 Open Care System 3 1 1 1 6 29 Transport Incubators 1 1 30 Plaster cutter with Aspirator 1 1 31 Tourniquet Electronic 1 1 32 Hot Pack Unit 1 1 33 Cold Pack Unit 1 1 34 Traction Unit 1 1 35 Operating Laparoscope Set for Gynaec 1 1 36 Hysteroscope System 1 1 37 Cryo Surgical Unit 1 1 1 1 4

38 Vacuum Extractor 1 1 1 1 4 39 Fetal Monitor 1 2 2 2 7 40 ENT Examination Unit 1 1 2 4

41 Operating Laparoscope Set for General Surgery 1 1

2

42 Multipara Bed side Monitor 1 1

Page 42: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited List of Requirement

HLL/PCD/ESIC-20/10-11 Page 42 of 166 01.09.2010

Consignee wise Qty. Sl. No.

Item Description Adityapur Indore Joka Baddi Manesar Bhiwadi Ludhiana

Basai Darapur

Total Qty

43 Emergency Resuscitation Kit-Adult(imported) 1 2 2 1

6

44 Portable Spirometer 1 1 45 Infusion Pump Syringe 1 4 5

46 PFT Machine(Spirometer & Computerised) 1 1

2

47 Fully Automated Biochemistry Analyzer 1 1 48 Centrifuge Machine 24-36 tubes 3 3 49 Computed Radiography System 1 1

50 Advancved Colour Doppler Ultrasound Scanner 1

1

51 Donor Chair 2 2 52 Blood Collection Monitor 2 1 3 53 Blood Bank Refrigerator 1 2 3 54 Tube Sealer 1 1 2 55 Cell Counter - 3 Parameter 1 1 56 Refrigerated Centrifuge 1 1 57 Plasma Expressor(E) 2 2 58 Dielectric Tube Sealer Table Top 1 1 59 PLATELET INCUBATOR & AGITATOR 1 1

60 AUTOMATED PLASMA THAWING EQUIPMENT 1

1

61 LAMINAR FLOW- VERTICAL 1 1 62 Deep Freezer -40 Deg. 1 1 2 63 Deep Freezer -80 Deg. 1 1 64 Cell Separator 1 1

65 ICU Ventilator (High End) 1 1 66 ABG Analyzer 1 1 2 67 Arthroscopy System 1 1 68 Color Doppler Ultrasound Unit 1 1 69 Video Bronchoscopy System 1 1 70 High Frequency Mobile X- Ray Unit 1 1 71 Digital Mamography System 1 1

72 Holter Monitoring System 2 2 73 Stress Test System(TMT) 1 1

74 Volumetric Infusion Pump 10 10 75 Biosafety Cabinet 1 1 76 Laprocator 2 2

77 Ultrasound Machine 3D with Color Doppler 2

2

78 Phototherapy Machine 1 1

79 Microscope 6 6 80 Yag Laser 1 1 81 Cell Separator 1 1 82 Cell Washer 1 1 83 Donor Couch 1 1 84 Platelet Incubator & Agitator 1 1 85 Sterile Tube Connecting Device 1 1 86 General Instruments Set 4 8 8 4 24 87 Surgery Instrument Set for Minor O.T. 1 1 1 1 4 88 Orthopaedic Surgery Instrument Set 1 1 89 Vaginal Hysterectomy 1 1 1 1 4 90 Abdominal Hysterectomy set 1 1 91 Delivery and Ceasarian set 1 1 1 1 4

Page 43: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited List of Requirement

HLL/PCD/ESIC-20/10-11 Page 43 of 166 01.09.2010

Consignee wise Qty. Sl. No.

Item Description Adityapur Indore Joka Baddi Manesar Bhiwadi Ludhiana

Basai Darapur

Total Qty

92 General Surgical Instruments for O T 1 1 93 Instruments for Dental Dept. 1 1 94 Instruments for Eye Implant 1 1

Part II: Required Delivery Schedule:

a) For Indigenous goods or for imported goods if supplied from India: Within 60 days from date of Notification of Award to delivery at consignee site. The date of

delivery will be the date of delivery at consignee site (Tenderers may quote earliest delivery

period).

b) For Imported goods directly from abroad: Within 90 days from date of opening of L/C. The date of delivery will be the date of delivery at

consignee site (Tenderers may quote earliest delivery period).

Part III: Scope of Incidental Services:

Installation & Commissioning, Supervision, Demonstration, Trial run and Training etc. as specified

in GCC Clause 13. Installation & Commissioning shall be completed within 15 days of handing

over the site of installation, complete in all respect by the consignee. The date of handing over of

the site has to be intimated by the supplier to the purchaser. The delay on the part of the supplier to

install and commission the equipment will attract the provisions as contained in the liquidated

damage clause.

Part IV:

Turnkey (if any) as per details in Technical Specification. The tenderer shall also specify the time

schedule for completion of Turnkey work.

Part V:

Annual Comprehensive Maintenance Contract (CMC) as per details in General Technical

Specifications para 4. The cost of CMC need not be quoted for items in sl. no. 86 to 94 and these

items will be evaluated for ranking purpose without the CMC price.

Part VI:

Required Terms of Delivery and Destination.

a) For Indigenous goods or for imported goods if supplied from India: Delivery required at Consignee Site.

Insurance (local transportation and storage) would be borne by the Supplier from warehouse

to the consignee site for a period including 3 months beyond date of delivery

b) For Imported goods directly from abroad:

The foreign tenderers are required to quote their rates on DDP at consignee’s site basis

giving break up of the price as per the Proforma prescribed in the Price Schedule.

Custom clearance, handling, unloading & loading and transportation to the consignee’s site

shall be the responsibility of the supplier/ Indian agent.

Insurance (local transportation and storage) would be extended and borne by the Supplier

from ware house to the consignee site for a period including 3 months beyond date of

delivery.

Page 44: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited List of Requirement

HLL/PCD/ESIC-20/10-11 Page 44 of 166 01.09.2010

c) Destination/Consignee details are given in Section XXI

Part VII:

Inspection:

a) For Indigenous goods or for imported goods if supplied from India: At consignee site by the respective In-charge of the Hospital or his authorised representative.

b) For Imported goods directly from abroad: Pre-despatch inspection at manufacturer’s premises as per GCC clause 8.8 and inspection

after arrival in India at consignee site by the respective In-charge of the Hospital or his

authorised representative.

Page 45: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 45 of 166 01.09.2010

Section – VII

Technical Specifications

Item No. 1

Biphasic Defibrilator • Biphasic • Waveform display (min 3) • External energy selection from 2 J to 200 J, biphasic • Charging time less than 8 seconds @ 200 J (with a charged battery) • Synchronizer and cardio version • Unique disarm button (in addition to automatic time delay) • Should come with high resolution monitor. • Should come with 3 lead ECG that can be measured from cables, adult external paddles, paediatric paddles. • Heart rate: 20 to 300 bpm BPM with user selectable alarms. • Should come with external pacing, demand and asynchronous modes • Should display CPR in real time • Should have large internal memory that stores and prints 25 ECG events. • Battery to last not less than 2 hours of continuous ECG monitoring or 30 full energy discharges. • Battery indicator on display and self test on battery Accessories needed • Paddles with remote energy selection, charge and discharge buttons on paddles. • Pediatric paddles adapters (set of 2) • A/c 240 v V 50 hz charger and mains power source. Adult paddles and test paddles. • 3 lead patient cable • One spare battery • Roll of 50mm recording paper x 10 rolls • 5 oz tube of defibrillation gel x 5 tubes • ECG cable with leads- 2 sets. • External disposable pacing pads-2 sets • Operation manual, service manual complete.

Item No. 2

Fiber Optic Bronchoscope with Camera

Bronchoscope should have : Field of view : 120deg. Depth of field : 3-50mm Tip deflection up/down : 180/130deg. Distal diameter : 4.9 mm Insertion tube diameter : 4.8mm

Diameter of working channel : 2.2mm

Insertion tube working length : 600mm

Total length : 900mm

Sharp, smooth image

Ergonomic control section

Easy insertion and excellent maneuverability

Page 46: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 46 of 166 01.09.2010

Compatible to electrosurgical treatment Light Source should have : Compatible light source (Halogen) with fibre-optic cable & back-up lamp Standard Accessories like : Foreign Body Retriever Forceps Tempo-nation Balloons Magnet Forceps Biopsy Forceps Single Chip camera of high resolution > 500 lines Incorporate 1/3" high sensitivity CCD and should have anti moire filter C-mount camerahead to attach bronchoscope and 3 windows size. 14" colour Medical grade LCD monitor for images

Item No. 3

Pulse Oximeter · It monitors from -neonate to adult. . To track oxygen saturation even during low perfusion condition. · Set seconds alarm management technology avoids unnecessary alarm & distraction during monitoring of severe motion patients. . Display for better monitoring. · Dual colour LED for differentiating normal mode & alarm mode. · Internal battery provides minimum of 8 hours backup. · 10 segment LED bar graph for pulse strength monitoring. · Facility for computer interface & printer connectivity · Individual alarm settings for Sao2 & pulse rate · Compatible with the complete range of disposable & reusable sensors. · Built-in battery charger with charger indication.

Item No. 4 Anaesthesia Workstation Technical specifications: Anesthesia system should be high end three gas system with three gas Oxygen, Nitrous Oxide and Medical Air double scale flowmeter with high and low flow and minimal flow provisions.

Should have an independent Oxygen flow meter for Oxygen delivery and an integrated variable flow suction unit. System should have at least three drawers and an additional writing surface that can be easily accessed. Drawers shall have the ability to lock , and shall be easily removed for the purposes of cleaning and sterilisation. Pipeline, cylinder and Airway pressures should all be displayed on colour coded gauges and be visible at all times during operation. Should have provision to attach 2 cylinders 1 each for O2 and N2O. Should have facility of delivering basal flow of oxygen on switiching on the machine.

System should have a second user accessible port for extraction of Anesthetic gas when using a nonrebreathing patient circuit. System should also provide the option of returning sample gas to the scavenging system with a dedicated port. A single pneumatic/electric on/off switch should activate the gas flow and vaporization.

Page 47: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 47 of 166 01.09.2010

The unit should have a battery back up facility for the ventilator in the event of power failure and should operate for a minimum of one hour. In the event of complete power loss and battery failure it shall still be possible to manually ventilate and deliver anaesthetic agent. System ahould have easily accessible common gas outlet in the event of an emergency and for use of alternate breathing circuits. Should have unlockable Oxygen flush to deliver oxygen flow of approximately 40l/min. Should have built in safety features like O2 failure alarm, N2O cutoff, Low O2 pressure etc., Should have motion sensitive back lighting for vaporizer dial adjustment. Should also have mandatory illumination of the writing table. The frame should have integrated power outlets to supply a minimum of four external devices. Should have locking of the front castors by a single central brake mechanism. Gas Flow The unit shall have a mechanical hypoxic guard system to control the ratio of Oxygen and Nitrous oxide to ensure a minimum of 25% of oxygen delivery at all times to avoid delivery of hypoxic mixture. It shall be possible to deliver Air with only basal flow oxygen independent of the above mentioned hypoxic control. Gas flow shall be controlled mechanically to avoid errors during power failure and electronic malfunction. Visual display of the gas flow shall be by physical means independent of electrical power. Cascade or dual flow tubes should be available for all gases to allow suitable resolution and accurate control at low total fresh gas flows. Flow meters should have backlight and antiglare illumination. The unit should have an independent measurement and display of fresh gas flow offering safety for low and minimal flow anaesthesia. A bag arm with height and positional adjustment shall be available as an option. Vaporizers The unit should accommodate two vaporizers for anesthetic agent delivery to allow easy selection of agent to be used. A third vaporiser storage area shall be available as an option. Vaporiser should be selectatec type, tool free installation and vaporiser of user choice can be mounted at will with interlocking facility to allow operation of only one vaporiser at one time. Vaporizers supplied with the unit shall be routine maintenance free for the life of the product. Should provide Isoflurane and Sevoflurane key filled vaporisers. Breathing System All parts of the breathing system that are in contact with patient gas should be latex free and autoclavable. Should not require tools when dismantled for cleaning and sterilization. Should accept large and small volume absorber canisters. The ventilator bellows shall be clearly visible and should ascend on expiration to provide a quick visual indicator for system leaks. Breathing system should have the option of C02 Absorber bypass control that will allow the absorber canisters to be removed without introducing system leaks.

Page 48: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 48 of 166 01.09.2010

Should have bag / vent selecting valve integrated onto the absorber and should automatically turn on the ventilator when positioned to vent mode. Ventilatior

Ventilator should be pneumatically driven, electronically controlled and should be ascending bellows type. Ventilator should automatically change drive gas should there be a gas depletion. Ventilator shall have a color display with touch screen user interface.

Ventilator should have the following ventilation abilities, volume control, decelerating flow pressure control, SIMV with pressure support and pressure support.

Ventilator should be capable of ventilating diverse range of patient groups from neonates to adult patients with restictive airways with tidal volume range between 20 ml to 1500 ml with single bellows system. Assisted modes of breathing should be flow triggered.

Ventilator should have an active proportional exhalation valve to prevent the potential of over delivery during pressure modes of ventilation. Ventilator should have a leak and compliance test that can be done independently of the full system check. On switching on, the ventilator system should be able to and shall give the user a choice of doing a unit test or bypassing in the case of an emergency. Ventilator shall compensate for fresh gas flow and compliance of the entire circuit dynamically. Measurement at the patient end of the circuit ( sensor at the patient end) should be provided to compensate for small leakages and compressible volume variability that occur during ventilation.

User should also have the option of setting a pre set compliance correction where similar circuits are used constantly. Should provide constant fresh gas flow into the breathing circuit during the inspiratory phase as mandatory. Ventilator should have the ability to set and store a hospital default as well as individual user preferences for easy selection of ventilation parameters and include screen layout, alarm preferences and ventilation settings. User should be able to set their own password.

Apnea alarms must be user adjustable to allow for all operating conditions and phases during Anesthesia. Ventilator should have the ability to display and store Patient Spirometery loops including Flow-Volume and Pressure Volume curves. Ventilator should also display waveforms for flow and airway pressure. Ventilator should display measured fresh gas independent of the flow meters. Ventilator should display a dynamic compliance measurement. Integrated Monitoring system: Anesthesia Monitoring system should be of modular type and capable of monitoring adult, pediatric and neonatal patients. Should be from the same manufacturer as of the anesthesia system.

Monitor should have minimum 19” independent flat panel display with multi color touch screen user interface to ensure all parameters are visible simultaneously. Module rack / housing should be independent and should be able to be placed near to the patient. Should be capable of 8 traces display. Should have facility to monitor: ECG, NIBP, SpO2, Respiration,

Page 49: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 49 of 166 01.09.2010

Invasive pressures (3), temperatures (2), Capnography and Bispectral index. Should have Cardiac output port enabled. Should have automatic identification and measurement of anesthetic agents, EtCo2, O2 and N2O and MAC value. Should have depth of anesthesia monitoring using Bispectral index. Cardiac output monitoring facility with all accessories. ECG should have capability for 3, 5 and / or 10 lead monitoring and should have built in arrhythmia monitoring on all 12 leads Inbuilt ST segment analysis and arrhythmia detection for all the leads should be available. Should have haemodynamic, oxygenation and drug dose calculations. EtCO2 should have both mainstream and side stream in one module. Respiration should be available with Cardio Vascular Artifact filter. OCRG(oxy cardio respiro gram) should be available for monitoring neonates. Alarm parameter should flash red in the presence of high priority alarms (e.g. ventricular fibrillation and asystole) and flash yellow in the presence of medium or low priority alarms (e.g. noisy signal, etc.) 24 hours trend data should be displayed. All monitors including central station should have similar user interface for usage among all clinicians. Modules should be compatible with transport monitors. Monitor shall provide capability to remote view of real time waveforms via the internet. Should be able to upgrade to softwares for electronic flow sheet and full dislcosure of all waveforms. On-screen keyboard for entering this data should have USB ports to connect mouse, key board, bar code scanner. Alarm limit status (ON/OFF) must be indicated on-screen for each parameter and actual parameter alarm settings must be displayed on-screen when alarms are on. Position of the displayed waveforms and color of the waveform must be user configurable. Monitor shall permit the optional ability to receive and display information from other patient devices such as ventilators, infusion pumps and other standalone devices. All modules should be compatible with all monitors quoted. Should be supplied with necessary accessories for adult , pediatric and neonatal accessories. Should be US FDA Approved Should be compatible with HIS and Should be HL7 compliant

Monitor should have capability to accommodate remote viewing of real time waveforms through internet. Accessories and spares

ECG / respiration: 5 lead ECG cable and lead wire set and 10 lead ECG cable and lead wire set per monitor NIBP: Adult: 2 sizes and Pediatric 2 sizes and neonatal, 1 size per monitor SPo2 Sensor: Adult sensor with cable, pediatric sensor with cable and neonatal sensor with cable per monitor IBP: Include 10 nos of disposable pressure transducer with bracket and interface cable per monitor Temperature: Skin and nasopharyngeal probes per monitor BIS: 25 nos of disposable sensors per monitor Environmental factors: Safe disposal system : AGSS – Anesthetic Gas Scavenging System, should be in place

Page 50: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 50 of 166 01.09.2010

Item No. 5

ECG Machine 12 channel · Simultaneous acquisition of up to 12 leads. · Real time continuous recording of 3, 6 and 12 channel. · Recording speeds of 10, 25 or 50 mm/sec · Extensive ECG quality control by AC Noise Filtering and Baseline. · A4 size reports for convenient reading and filing. · Colour coded keys for case of operation. · Convenient battery operation for greater mobility. · Versatile report formats and speed options to provide auto reports or rhythm reports. · User configurable filters. · Preview signal quality prior to printing. Saving time and paper. · Keyboard entry for patient ID information. · Capability to generate any number of ECG copies possible for filing and distribution. · Adult and paediatric analysis programs std. · Automatic interpretations of ECG data. · Availability of adult, paediatric and neonatal accessories. · Accessories e.g., stand, cables, electrodes etc. Should be quoted separately. · ECG Paper roll for 1000 patients · Service and operation manual complete.

Item No. 6

Pharmaceutical Refrigerator for O.T. Good quality double steel walled cabinet Capacity 500 litre Forced air circulation to maintain chamber temp. uniformity. Temp. in range of 2 to 8 degree centigrade Low noise level Automatic defrosting & digital temp.display Audiovisual alarm for low & high temp. Input voltage 220 to 240 volts Inbuilt electronic voltage stabilizer

Item No. 7

O T Tables for Orthopaedics Electro-hydraulic battery operated mobile operating and extension table for orthopaedic interventions including open and intra-medullary nailing and shall be fully compatible with C-Arm Image Intensifier. Should include back plate adjustment, with SFC padding, including. basic equipment - 2 traction bars short and long, 1 rotation-tilt clamp, 2 elongation bars, 1 rotation and traction stirrup clamp, 1 screw tension device, 1 counter traction post for neck of femur, 1 foot plate support, 1 side rail elongation. Should have the following features: - Battery-powered or mains operated electrically controlled hydraulic drive.

Page 51: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 51 of 166 01.09.2010

- Base covers and column casing, table top frame, traction bars and basic accessories made of CrNi steel. Operating table with easy-rolling castors for longitudinal and lateral movement. Operating table top subdivided into 9 sections: head rest , 3-section back plate with 2 detachable shoulder segments, seat plate with 2 detachable buttock supports, pair of leg plates, detachable. - 2 traction bars beneath the seat plate with 2 separately adjustable pivot joints each for extension of the extremities with unimpeded use of the image intensifier. - Guide rails beneath the seat plate allowing X-ray cassette insertion from either the right- or the left-hand side for the hip area. - Detachable pads made of special foam core, 80 mm thick, radio translucent and electrically conductive on plastic supporting plates. Should meet the following Electrical standards: Special-design batteries with a capacity for approx. 2 weeks ( 50 - 80 operations). - Battery charge level: electronic monitoring with optical indicator and acoustic signal - Recharging the batteries via mains supply 100 - 240 V AC (adjustable), - 50-60Hz, via mains cable - Class II, Type B; the enclosure leakage current meets the requirements of the patient leakage current for CF conditions according to EN 60601-1. Should meet following Technical data: - Length of table top incl. head rest : 1965 mm (With standard back plate) - Width of seat / back plate without side rails : 530 mm Permissible load approx : 180kg Should have following Electro hydraulic adjustments: Control of the electro hydraulic motions from outside the sterile area via cordless IR remote control optional hand control. - Height without padding : 800 – 1175 mm - Trendelenburg / reverse Trendelenburg : + 25° / - 30° - Lateral tilt : 20° - Back plate: + 60 ° - 0-position (horizontal alignment of the table top) Should supply the following with Basic table: - IR remote control comprising: cordless, battery-powered hand control with film keypad and mobile charging station (to recharge the battery of the hand control) with mains cable; wall mounting should be possible. - Cable-connected hand control, detachable, with film keypad; for connection to the operating table via flexible coiled cable - Head rest - Foot plates, width adjustable, with Velcro strap and heel pad, pair - Arm board - Anaesthesia screen - All necessary clamps or adaptors

Page 52: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 52 of 166 01.09.2010

- Accessory stand Should supply the following with Extension accessories: FEMUR NAILING IN SUPINE POSITION: a) Counter traction post for femur: 1 No. FOR TIBIA & FIBULA NAILING a) Counter traction post for tibia : 1 No. b) Guide bar : 1 No. c) Condoyle fixation : 1 No. FOR TOTAL HIP END PROSTHESIS a) Fixtures for body support : 2 Nos. b) Back- buttock support with pad : 1 No c) Lateral support with pad : 1 No. KNEE REPLACEMENTS a) Motorised Knee positioning device including radial clamp : 1 No SPINE IN KNEE ELBOW POSITION a) Positioning device for spinal surgery, height adjustable and horizontal adjustment : 1 No. b) Trolley : 1 No. c) Sitting bracket with lateral supports : 1 No d) Radial setting clamps : 2 Nos. CERVICAL SPINE SURGERIES a) Connection Fixture : 1 No b) Horse shoe shaped head rest ( two pc.) : 1 No. c) Guiding roller for head side traction : 1 No. INTRAMEDULLARY NAILING OF HUMERUS a) Humerus positioning device : 1 No. b) Countertraction post for humerous : 1 No c) Weinberger hand traction device : 1 No. OPERATIONS ON THE HAND. a) Hand operating table : 1 No. b) Radial setting clamp. : 1 No.

Item No. 8

O T Tables for General Surgery & Gynaecology - Should be Mobile, modular operating table for any surgery, electro-hydraulic drive, with corded hand control, with upper back plate, with leg plates, with head rest, with SFC padding of 80 mm. Should have following features: - Battery (1 week/ 60-70 operations on Single Charge) and mains operation. - Stable base construction with 4 double swivel castors for easy motion and manoeuvring (locking of the double swivel castors via a foot pedal) - Base cover and cover for the override control panel made of CrNi steel, resistant to impact, breakage and disinfectants - SFC padding (Special Foam Core) with multi-layer construction for exceptional comfort during patient positioning and utmost patient safety

Page 53: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 53 of 166 01.09.2010

- Safe patient positioning due to optimum covering of table top joints - Column casing made of CrNi steel - Supporting bars for the seat and back section, leg plate sockets, joint covers and side rails made of CrNi steel - Operating table top subdivided into 5 sections: head rest, upper back plate, lower back plate, seat plate & leg plates - If there are any malfunctions or if the corded hand control is defective, the operating table can be controlled using the override panel.Entire table top without crossbars, allowing intra-operative fluoroscopy Operating elements: corded hand control, override panel, foot switch Should meet following Electrical Standards: - Battery charge level: electronic monitoring with optical indicator Recharging of the batteries and supply of the operating table by means of a mains - cord - Nominal mains voltage /220/230-240 V frequency 50/60 Hz - IEC 60601-1 Should meet following Technical data: -Length of table top : 2075 mm -Width of seat/lower back plate without side rails : 510 mm -Width of seat/lower back plate with side rails: 560 mm Max. Patient Weight -180 Kg Should have following Electro hydraulic adjustments: Via corded hand control or override control panel, with status indicators for battery charge level and base brake feature, located at the column head. Adjustment options via corded hand control or IR remote control: -Height with padding: 600 – 950 mm -Trendelenburg/reverse Trendelenburg: + 25° / - 25° -Lateral Tilt :15° -Lower back plate: 60° / 50° -Leg plates: 0° / 95 ° -Longitudinal Shift: 290 mm -Body Elevator: 100 mm -0-position (horizontal alignment of the table top) Should have following Manual adjustments: -Standard back plate : + 90° / - 45° -Standard head rest : + 45° / - 45° -Base braking (lock / unlocking) Should be supplied with Basic equipment:

- Corded hand control - Head rest, inclinable, SFC padding - Retrofit kit, side rails, for head rest for subsequent lateral mounting - Arm Boards pair - Anesthesia Screen

- Clamps or Adaptors - Leg plates with SFC padding - Gopel knee crutch with clamp pair - Side supports pair

Page 54: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 54 of 166 01.09.2010

Item No. 9 Electro Surgical Unit Should be suitable for all types of surgeries. Digital system with automatic patient plate monitoring. Display: Digital Monopolar cut: 300 to 400 W Not less than two blend modes Provision for Spray, Dessication Bipolar Coagulation Facility for underwater cutting Facility for simultaneous coagulation Audio visual alarm for breakage of contact between patient and plate Accessories: Double pedal foot switch Single Pedal Foot switch Patient plate with cable x1 Autoclavable handles: 3 sets Electrodes: 3 sets Bipolar forceps with cord x 1 All accessories should be from same manufacturer to ensure compatibility. All instruments should be autoclavable or Single Use. Single Use Disposables if offered should be sufficient for 20 surgeries. The equipment should be US FDA Approved Complete instruction and service manual should be supplied.

Item No. 10

O T Table for Minor O T

O T Table with following specs Electro Mechanical Adjustments Height 760 mm to 960 mm, Trendlenberg/Rev Trendelberg,25/25 deg Lateral Tilt.-15 deg Manual Adjustments Backrest with gas spring up/down-65/25 deg Headrest Detachable Legrest Detachable Standard. Four Section Mattress – 1 Set Accessories 1-Arm Board 02 2-Anaesthesia Screen 01 3-Body Strap 01 4- Radial setting clamp 03 5- Goepel Knee crutches 02 6-Perineal cutout filter 01

Page 55: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 55 of 166 01.09.2010

Item No. 11

Boyle's Appartus Anaesthesia system should be high end three gas system with three gas Oxygen, Nitrous Oxide and Medical Air with double scale flowmeter for O2 & N2O. System should have at least two drawer. Pipeline, cylinder and Airway pressures should all be displayed on colour coded gauges and be visible at all times during operation. Should have provision to attach 2 cylinders 1 each for O2 and N2O. The unit should have a battery back up facility for the ventilator in the event of power loss and should operate for a minimum of one hour. In the event of complete power loss and battery failure it shall still be possible to manually ventilate and deliver anaesthetic agent. Should have unlockable Oxygen flush to deliver oxygen flow of approximately 40l/min. Should have built in safety features like O2 failure alarm, N2O cutoff, Low O2 pressure etc., The unit shall have a hypoxic guard system to control the ratio of Oxygen and Nitrous oxide to ensure a minimum of 25% of oxygen delivery at all times to avoid delivery of hypoxic mixture. The unit should have an independent measurement and display of fresh gas flow offering safety for low and minimal flow anaesthesia. The unit should accommodate two vaporizers for anesthetic agent delivery to allow easy selection of agent to be used. Vaporiser should be selectatec type, tool free installation and vaporiser of our choice can be mounted at will with interlocking facility to allow operation of only one vaporiser at one time. Vaporizers supplied with the unit shall be routine maintenance free for the life of the product. Should provide Isoflurane key filled vaporisers. All parts of the breathing system that are in contact with patient gas should be latex free and canister and bellow should be autoclavable. The ventilator bellows shall be clearly visible and should ascend on expiration to provide a quick visual indicator for system leaks. Breathing system should have C02 Absorber. Should have bag / vent selecting valve integrated onto the absorber and should automatically turn on the ventilator when positioned to vent mode. Ventilator should be pneumatically driven, electronically controlled and should be ascending bellows type. Ventilator should automatically change drive gas should there be a gas depletion. Ventilator shall have a large display with touch screen user interface. Ventilator should have the following ventilation abilities, volume control, SIMV with pressure support . Ventilator should be capable of ventilating diverse range of patient groups from neonates to patients with restictive airways with tidal volume range between 20 ml to 1500 ml with single bellows system. Assisted modes of breathing should be flow triggered. Ventilator shall have an active proportional exhalation valve to prevent the potential of over delivery during pressure modes of ventilation. Ventilator shall compensate for fresh gas flow and compliance of the entire circuit dynamically. Measurement at the patient end of the circuit ( sensor at the patient end) should be provided to compensate for small leakages and compressible volume variability that occur during ventilation. User should also have the option of setting a pre set compliance correction where similar circuits are used constantly.

Page 56: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 56 of 166 01.09.2010

Should provide constant fresh gas flow into the breathing circuit during the inspiratory phase as mandatory. Apnea alarms must be user adjustable to allow for all operating conditions and phases during Anesthesia. Ventilator should also display waveforms for flow and airway pressure. Ventilator shall display measured fresh gas independent of the flow meters. Ventilator shall display a dynamic compliance measurement.

Item No. 12 Multipara Monitor with Central Monitoring Station

Patient monitor system should be of modular type and capable of monitoring adult, pediatric & neonatal patients. Monitor should have 17” independent flat panel display. Touch screen user interface . Module rack / housing should be independent and shall be able to be placed near to the patient. Should be capable of 8 traces display. Monitor must be capable of simultaneously monitoring the following parameters which should be present as standard: ECG, NIBP, SpO2, invasive pressures (2), temperatures (2) Should be compatible with Capnography, Cardiac output, EEG, and BIS and prices to be offered as optional ECG should have capability for 3, 5 and / or 10 lead monitoring and should have built in arrhythmia monitoring on all leads. Inbuilt ST segment analysis and arrhythmia detection for all the leads should be possible. Haemodynamic and drug dose calculations should be available. Arrhythmia should be grouped based on classifications – and should show no of arrhythmias occurred. Respiration should be available with Cardio Vascular Artifact filter. ICP monitoring should be possible. Alarm parameter should flash red in the presence of high priority alarms (e.g. ventricular fibrillation and asystole) and flash yellow in the presence of medium or low priority alarms (e.g. noisy signal, etc.) 24 hours trend data should be displayed.

All monitors including central station should have similar user interface for easy usage among all clinicians. Monitor shall provide the capability to interact with alarms at remote bedsides. Monitor shall provide the capability to receive and display real-time waveforms, trended data and alarm status from other bedside or telemetry units on the patient monitoring network. Monitor shall provide the capability enter patient information at the bedside or central monitor. On-screen keyboard for entering this data is preferable. Should have USB ports to connect mouse, key board, bar code scanner. Alarm limit status (ON/OFF) must be indicated on-screen for each parameter and actual parameter alarm settings must be displayed on-screen when alarms are on. Position of the displayed waveforms must be user configurable. Waveform color changing should be user configurable. Monitor shall permit the optional ability to receive and display information from other patient devices such as ventilators, infusion pumps and other standalone devices. All modules should be compatible with all monitors quoted. Bed to bed communication between the monitors should be possible with out a central

Page 57: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 57 of 166 01.09.2010

station.

Networking to central station should be possible and price of central station should be offered as optional Patient monitoring network shall use standard TCP/IP protocol and be capable of residing on hospital’s network infra-structure. Should be compatible with HIS and should be HL7 compliant. Monitor should provide remote viewing of real time waveforms through internet. Patient monitoring network shall be able to support up to 1,000 monitoring nodes. Should be supplied with necessary accessories for adult , pediatric and neonatal accessories. Accessories and spares 1. ECG / respiration: 5 lead ECG cable and lead wire set per monitor 2. NIBP: Adult: 2 sizes and Pediatric 2 sizes and neonatal, 1 size per monitor

3. SPo2 Sensor: Adult sensor with cable, pediatric sensor with cable and neonatal sensor with cable per monitor 4. IBP: Include 10 nos of disposable pressure transducer with bracket and interface cable per monitor 5. Temperature: Skin and nasopharyngeal probes per monitor. The equipment should be CE & US FDA Approved. Central Monitoring Station for multipara monitor System should have minimum 16 beds capability. Central station should have 17” color display. Should have drug dose and hemodynamic calculations. It should have possible to view information such as vital signs, alarm status, arrhythmia analysis, trended parameters, patient data etc for any selected bed from the central station. Should have separate computer keyboard and 4 channel thermal array recorder. should have default alarm limits and customizable parameter settings. Central station should have full bed review capability. Central station should be able to be configured as a bedside monitor if required. Should have 24 hours trends. Should have capability for HL7 interface. Should be capable of monitoring telemetry modules. All system should have CE & US FDA certifications. Should be supplied with a On-line suitable UPS Note: Price of MULTI PARAMETER MONITOR and CENTRAL MONITORING STATION should be quoted separately.

Item No. 13

Portable Ventilator The Ventilator should be portable. The ventilator should be capable of ventilating from pediatric patients to adults. The modes should include Assist Control, SIMV in Volume control and Pressure control modes. It should also have CPAP and PSV available. The tidal volume should range from 50ml to 2000ml.

Page 58: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 58 of 166 01.09.2010

Peak Inspiratory Pressure - 5 to 55 mbar Pressure Support – 5 to 55 mbar Insp. Time – 0.3 to 2.4 secs Breath Rate – 0-60 BPM Insp. Sensitivity – 1 to 5 Exp. Sensitivity – 5 – 95% Flow Pattern should be Square, descending and Sinusoidal PEEP should be 0.5mbar to 20mbar Rise Time should be from 1 to 4 with increments of 1. I:E ratio should be 1:4 to 1:1 I / T should be 20% to 50% The Apnea time should be 1- 60 secs and backup rate of 5 to 40 BPM. Vt SIGH should be in single to double multiplier of Vt. The ventilator should have bright display with backlight The ventilator should have Waveform display of Pressure Vs. Time and Flow Vs. Time. Ventilator should be able to be used invasively through an artificial airway, or noninvasively through a mask or other noninvasive interface. The ventilator should have choice to select the type of breathing circuit (Paed. OR Adult) for Circuit volume compensation.

The ventilator should have internal battery back up for at least 11 hours with Real-time battery life indicator The ventilator noise level should be < 30 dB at 1 m. The ventilator should have Automatic atmospheric adjustment The ventilator should have recording facility of patient data trends through USB for 3 months to 1 year time with following parameters; Monitoring: pressure, inspired flow, exhaled flow and leak measurement Trends: leaks, VTI, VTE, Rate, I/T, M. Vol, P MAX and PEEP measurements The Machine should be operable with oxygen flow from B type oxygen cylinders as well as central oxygen supply.

Page 59: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 59 of 166 01.09.2010

Item No. 14

BIPAP Non invasive ventilation/ bipap with LCD display of parameters with following features •IPAP range 6-30 cms • EPAP range 4-20 cms • Respiratory rate can be set to 4-40 bpm • Spontaneous / CPAP/ spontaneous with time/ timed • Connectors for et ET tube available for direct connection • Should have CPAP, PSV ST, PCV, PACV modes • Should have facility to set target tidal volume • Inspiratory time should be possible to set allowing critical patients to breathe out. The rise time from EPAP to IPAP can be set and varied The system should be supplied with operator’s manual, ultra mirage mask (full mask) , hose pipe and power cable

Item No. 15

Binocular Microscope 1. Antimould/Antifungal type microscope 2. Colour corrected infinity optical system 3. Nose piece Quintuple reversed inword facing. 4. Objectives Planachromatic and springloaded 4X (1pece), 10X (1perce), 20X(1piece), 40X(1piece), oil immersion 100X (2 piese). 5. Eye piece wide field, 10X, one pair each with preferably with pointer. 6. Field of view>20mm. 7. Trinocular eye piece tube to facilitate camera attachment. The eye piece tube should be siedentopf type, 30 degree inclined and rotatable by 360 degree. 8. Dioptre adjustment of both eye pieces. 9. Inbuilt arrangement of illumination with halogen lamp (6V/20W) fitted directly under file lenses (Kohler’s system) with intensity control. 10. Condenser- Bright field Abbe’s NA 1.25 with iris diaphragm and filter holder. 11. Coaxial fine and coarse adjustment with adjustable tension. 12. Double stage- Double slide holder low position and coaxial, movement 13. Power supply 220/240 volts. 14. Spare halogen lamps-6 no’s to be supplied with each microscope 15. Power cord 16. Dust cover and preferable with box

Item No. 16

Urine Analyzer Two Different strips can run on the system Combur 10M : Glucose,protein,pH, Leucocytes,Nitrite,Ketones Bilirubin,Urobilinogen,Blood,SG (10 parameter) Combur 9 :Glucose,protein,pH, Leucocytes,Nitrite,Ketones,Blood,Bilirubin,Urobilinogen

Page 60: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 60 of 166 01.09.2010

Features · Sample : Uncentrifused Fresh Urine · Analysis time – less than1 min · Sample throughput – approx. 60test strips/hours (normal Mode) · Calibration through dry strips for better results . · Totally Maintenance Free System · Light Source - LED · Wavelenght measurements - 470nm,555nm, 620 nm · Data display on Touch screen monitor · Data printout on fast low noise thermal printer · Memory for 1000 patient results with time and date,3×100 Controll · Simple Touch Screen operation · Flagging of abnormal results · Interfaces - 1 x RS 232 · 110V to 240V AC operation · CE / CB / UL/CUL approved

Item No. 17

ELISA Reader with Washer 1. Description of Function 1.1 ELISA Reader is required to Read the Colour Density known as OD(Optical Density) in ELISA (Enzyme Linked Immuno-Sorbent Assay.)Plates. 2 Operational Requirements 2.1 Only ELISA Reader is required. 3 Technical Specifications 3.1 OPTICAL SYSTEM Digital light control

8 measurement channels including 1 reference. Single and dual wavelength measurement with facility for kinetic measurement 8 s maximum measurement time for single and dual wavelength and 5 s(+/_1Sec.) for kinetic Measurement Range 400-700nm Indication Range 0-2.999 abs Accuracy Plus/Minus 2% or Plus/Minus 0.005 abs Resolution 0.001 abs Inbuilt Filters: Narrow band interference Should have the following filters – 405, 450, 492(+/_2nm), 540, 620 (+/-10nm) and 690 nm Should measure end point, curves and kinetic. 3.2 SOFTWARE: Storage of immediately preceding measurement At least 15 user programmable tests permanently stored Time programmable between each measurement. Agitation programmable before each reading Bidirectional printer interface. Data memory through computer Built in Windows based software programming software. 3.3 MEASUREMENT MODES Plate shaking mode for sample mixing selectable speed and time) Flexible blank mode setting Matrix Modes: Matrix -/x/t, Matrix-/0-0 (Range ), Matrix-/f/(Floating cut off)

Page 61: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 61 of 166 01.09.2010

Difference Mode: Absorbance of each well in even numbered subtracted from those of odd numbered columns Curve fit Modes: LIN/LIN.LIN/LOG.LOG/LOG or auto curve transformation with ability to add the standard curve; 8 to 12 way string orientation or kinetic modes Table of optical densities, Delta DD, Graphic, Reaction rate/V-Max 3.4 Adjustable for different micro plate geometrics 3.5 Halogen Lamp 20 - 40 W. 3.6 16 digit alphanumeric fluorescent display 3.7 Membrane keyboard. 3 .7 Technical Specifications for washer 3.1 Auto strip washer for 96 well plates / strips 3.2 1 x 8 strips/ 1x12 strips. 3.3 Dispensable wash volume 50 - 300 µl. 3.3a Residual wash Volume -<0.5µl 3.4 Aerosol Shield for user safety. 3.5 In built shaking facility 4 System Configuration Accessories, spares and consumables 8-12 channel manifold, all tubing sets, wash, rinse and waste bottles Maintenance kit to be provided. 4 System Configuration Accessories, spares and consumables 4.1 System as specified- 4.2 Halogen Lamps : 2 4.3 Printer inbuilt or external to be supplied along with 10 Rolls/Z Fold 4.4 Dust cover. 5 Environmental factors 5.1 The unit shall be capable of being stored continuously in ambient temperature of 0 -50deg C and relative humidity of 15-90% 5.2 The unit shall be capable of operating continuously in ambient temperature of 10 - 40deg C and relative humidity of 15-90% 6 Power Supply 6.1 Power input to be 220-240VAC, 50Hz fitted with Indian plug 6.2 Resettable over current breaker shall be fitted for protection 6.3 Suitable voltage corrector/stabilizer 6.4 Suitable UPS with maintenance free batteries for minimum one-hour back-up should be supplied with the system. 7 Standards and Safety 7.1 Comprehensive training for lab staff and support services till familiarity with the system. 7.2 Should be compliant to ISO 13485: Quality systems - Medical devices - Particular requirements for the application of ISO 9001 applicable to manufacturers and service providers that perform their own design activities. 7.3 Should be FDA or CE or ISI approved product 8 Documentation 8.1 User/Technical/Maintenance manuals to be supplied 8.2 Certificate of calibration and inspection from factory. 8.3 List of Equipments available for providing calibration and routine maintenance support as per manufacturer documentation in service / technical manual. 8.4 List of important spare parts and accessories with their part number and costing. 8.5 Log book with instruction for daily, weekly, monthly and quarterly maintenance checklist. The job description of the hospital technician and company service engineer should be clearly spelt out

Page 62: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 62 of 166 01.09.2010

Item No. 18 Cell Counter(Fully automated 5 part differential haematology analyser) · Automated hematology analyzer should include 24 parameters including histogram for RBC, WBC and platelet. · Should have impedance principle for counting and photometer for hemoglobin. · It should read at least 60 samples per hour or more. · Should have dual channel measurement. · Double dilution chamber · Sample volume less than 200 micro litres in whole blood and pre –dilute mode. · It should have various types of discrete mode and real time random access analysis to save reagent consumption and analysis time. · Sampling needle should have automatic wash from inside and outside. · LCD / VGA Monitor with graphical user interface (GUI) for easy operation. · Large illuminated colored VGA or LCD should display the result of all parameters and histogram together. · Should have sample manual and capillary mode. · Should have capacity to store at least 20000 numeric patient results and 5000 graphics. · Should have inbuilt / External graphic printer. · Should have RS232 serial and parallel port can be connected with LAN and laser printer. · Should have a membrane keyboard for routine operations and maintenance with option to attach external key board for patient demographic entry at instrument operation. · Should have three dimensional technology or Flow cytometry for differential analysis to maximize resolution, specificity and efficiency. · Should have extended analysis time for cytopenic sample. . · Should be able to integrate with optional automated slide maker and stainer. · Should have zero routine maintenance with automatic electronic aperture cleaning and back flush after each sample. · Instrument should accept all types of vacutainer tubes. · The instrument should have option for auto sampler, bar code reader. · Reagent cost per cycle including start up and shutdown if 200 & 500 samples are processed at a time should be submitted separately in the financial bid. · There should be automatic storage of calibration data and extensive quality control programme with LJ plot for at least 8 control lots and at least 25 runs per lot. Basic common necessities: · Input Voltage 230 volts 50 Hz as per Indian standard. · Service manual and technical data with all necessary passwords without any obligation. · Instruction and operational manuals without any obligation. · UPS preferably sine wave based with maintenance free batteries with duration two hours.

Item No. 19

Bio-Chemistry Semi-Auto Analyzer

Page 63: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 63 of 166 01.09.2010

Automatic flow-through Cell Analyzer Easy to use high quality precision Analyzer Monochromatic and Bichromatic measurements (340 - 630) º Interference filters 340-405-500-546-578-630mm Bandwidth 8 nm. Optical measuring pass of 10 nm. Measures in absorbances, Concentration kinetics, Fixed Time, Rate and Differential MULTI-STANDARDS Zero set: fully automatic Fully programmable directly from the keyboard Storage capacity - 100 complete tests. Reading volume need: 500 ul Disposable cuvette programmable with flow through Cell: 500- 1000ul Built in printer Thermostated block by peltier effect at 25-30-37 °C for 10 cuvettes. External interface to be connected to computer. WITH STANDARD ACCESSORIES

Item No. 20 COAGULATION ANALYZER (COAGULOMETER) Fully automated, open random access, 4 detector channel Blood Coagulation Analyzer having clotting, chromogenic and immunological assay channels. Clotting detection methods by recording change of viscosity and/or change of light transmitability and permitting both chromogenic and immunogenic assays Parameters: Provision for simultaneous random analysis for at least 15 parameters such as PT, APTT, Fibrinogen, Protein S , Protein C, APC resistance, ProC, Antithrombin III, Heparin, Plasminogen Activator inhibitor, D-Dimer Plus, von Willebrand Antigen, Lupus Anticoagulant, and coagulation factors VIII & IX . Should have throughput of at least 100 tests per hour for PT. Should be provided with a compatible printer. Should have a intra assay reproducibility of: Normal: PT<2.0% APTT<2.0% Fibrinogen<4% Patient: PT<2.0% APTT<2.0% Fibrinogen<5% Additional Accessories required :1) All standard accessories 1 set extra. 2) Compatible Online UPS with1 hour back up.

Page 64: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 64 of 166 01.09.2010

Item No. 21

Electrolyte Analyzer

Fully automated Ph/blood gas/electrolytes analyzer, measures the following parameters PH, PCO2, PCO2 and PO2 and barometric pressure, NA, K, Ca, Cl, Ct, THB, HCL • Analyses per hour should not be less than 36 • Sample volume should not exceed 110 µl for all parameter • All calibration and cleaning cycle should be fully automated with user selectable calibration times • The electrodes provide should be zero maintenance including the reference electrode. • The system should have onboard data management to store all patient result and calibrations. • A power fault protection for 20 minute to keep all calibration and programmed data • The system should have a screen to access all system software and to display the patient results. • A built –in printer should be provided to print out patient result • The system should have a patient data, new reagents and new electrodes. Startup kit 1000 samples, essential spares. • The system should work on discrete testing, i.e. selectable parameter testing

Item No. 22

500 M.A. X-ray machine A. Generator: 1. Generator should be high frequency/inverter type for constant output. 2. Output 50 kw or more. 3. Kv range 40 kv – 120 kv. 4. Output at 100 kv should be 500 ma or more. 5. It should have automatic exposure control device. 6. It should have digital display of kv & mas. 7. Anatomical programming radiography should be possible. 8. It should have over loading protection. B. X – ray tube and collimator: 1. The x-ray tube should be rotating anode high speed, compatible with the generator and must have dual focus. Focal spots of following sizes: Large focus: 1.2/2.0 mm or better. Small focus: 0.6/1.0 mm or better. Tube with anode heat storage capacity 150 khu or more. 2. Motorized collimator having additional filters (for dose reduction) and auto shut provision for the light. C. X – ray table: 1. Horizontal table with floating table top. 2. It should have transverse + 10 cm or more and longitudinal movements + 35 cm or more with electromagnetic brakes. 3 It should be thin table top. 4. It should be provided with bucky which can hold all standard sizes of cassettes upto 14”x17”.

Page 65: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 65 of 166 01.09.2010

5. Bucky should have a grid ratio 10:1 or more with 65 lines per cm. D. Vertical bucky stand: 1. It should have provision to do chest radiography with grid. E. Essential accessories: the following essential accessories to be provided with the unit. 1. Lateral cassette holder – one. Should be supplied with following: · Whole body (upper and lower) lead apron (standard make) – 8 nos. · Lead screen (with lead glass window)– 2 nos. · Lead lining for walls and door · Apron & glove hanger Heavy duty multiple glove and apron hanger wall or floor mounted. Stainless steel construction. Holds two pairs of gloves and two lead aprons. Dimensions: 76 x 46 x 25 cm approximately. All necessary aerb / barc approvals.

Item No. 23

Portable Xray unit 2.5 kw HF General purpose,single tank diagnostic mobile X-ray Equipment,suitable for radiography of standing, sitting or recumbend patients as well as for patients in bed or operation table & as stand by unit. Comparising of High Frequency X-ray generator 2.5 with the following out put- Output: 2.5 kw as per Is 7620 Kv Range : 40 kv-100kv in 24 steps mA Range :16mA-60mA, mAs Range : 0.32mA's-200mAs Exposure Time : Min 20 m sec The unit will be calibrated within the permissible limit of x-ray tube X -RAY TUBE stationary anode x-ray tube having 1.4*1.4mm focal spot Collimato:- Brightfield light beam collimator with 100w halogen lamp and auto shut-off facility Power Supply:- Single phase 195v-265v, 50Hz power supply with line resitance<0.8 ohms Mobile Stand Light weight,low height mobile stand with easy manoeuvarbility

Item No. 24

Colour Doppler Digital 1. Beam former Universal Digital broadband former accepting routine Phased array sector, Convex, Linear and Endovaginal probes. 2. Monitor

Page 66: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 66 of 166 01.09.2010

Should have high-resolution 15” medical grade Flat Panel monitor with articulating arm 3. Digital Processing Channels Atleast 1024 channels 4. Gray Scale System should have a minimum of 256 gray levels with system dynamic range to be more than 200dB 5. Display modes With B, 2B, M, Color M mode, PW, HPRF and Color Doppler with Power Doppler. Phase Inversion Tissue Harmonics should be available on Convex and Phased Array probes. steering on Color / PW modes on Linear probe should be available. Trapezoidal Image. Triplex mode . 6. Cine Review Standard cine memory providing 500 frames on 2D mode and up to 30 seconds Doppler Cine. To be confirmed. 7. Transducer technology System should be capable of handling 1 – 15 MHz Multifrequency imaging with independent selection of 2D / Color / Spectral doppler frequency should be offered. 8. Exam Presets Should be able to store atleast 30 user specific Image presets. 9. Image Optimisation on B and M modes System should have the following · Selectable frequency on probes · Up / Down & Right / Left Image rotation · Muiltiple steps of Edge Enhancement settings · Upto 25cm dept. · levels of persistence 10. Color and Spectral Doppler Optimisation System should have the following · Selectable frequency on all probes · Gate adjustments on Spectral modes. · Filter adjustments should be specified. 11. Zoom function When a more detailed view is needed, the system should reoptimises and rescans the selected region to actually increase the acoustic information content of the image upto 10 times. 12. Deep Doppler and color performance HPRF performance at depth 13. Measurements and Calculations. System should have atleast 6calipers with depth information and extensive, customizable measurement and report packages including complete Obstetrics, Gynecology studies. Distance, Area, % stenosis on B mode Distance, Time, Heart Rate, Slope on M mode Velocity, Acceleration time, Slope, PI, RI, S/D Ratio with Auto Doppler on Doppler mode 14. Image Management

System should have facilities to store atleast 100,000 images, should have the facility of performing calculations and annotations on stored images. System should have inbuilt CD / DVD writer for image transfer. 15. User Interface PC – based operating principles. Task light with back-lit illumination of control panel.

Page 67: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 67 of 166 01.09.2010

System should have on screen menu for easy and immediate access to imaging controls. Live image should be supported with side by side reference / stored images on the Flat panel monitor. 16. Power requirements

System should operative between 200 – 240 VAC, 50 Hz. Maximum power consumption should not exceed 700 VA 17. Required Accessories · Convex Probe 2 – 5 MHz (Harmonic Imaging should be available) · Linear Probe 3.0 – 9 MHz (Harmonic Imaging should be available) · Endovaginal probe 4-8 Mhz 18. Accessories · Thermal printer · Inkjet Printer

Item No. 25

Infant Resuscitator Manometer Range: 20 to 80 cm H2O (-20 to 7.8 kPa) Maximum Pressure relief should be 5 to 70 cm h2O at 8 LPM Peak inspiratory Pressure (PIP) should be 5 to 70 cm H2O at 8 LPM Positive-end expiratory pressure PEEP) should be 1 to 9 cm H2O at 8 LPM. Gas inlet flow range:- 5 LPM (min) to 15 LPM (max) If the gas inlet flow rate increases from 5 to 15 LPM, the peak inspiratory pressure typically increases approximately 8cm H2O (0.8 kPa). Delivered oxygen Concentration: - Up to 100% depending on gas supply Operating time (400 L cylinder):- @ 8 LPM 50 minutes Recommended body weight: - Up to 10 kg (22 lb) Should be supplied with all related tubing’s and consumable (Preferable reusable) for at least 50 cases (application)

Item No. 26

NEONATAL MONITOR Patient monitor system should be of modular type and capable of monitoring adult, pediatric & neonatal patients. Monitor should have 17” independent flat panel display. Touch screen user interface . Module rack / housing should be independent and shall be able to be placed near to the patient. Should be capable of 8 traces display. Monitor must be capable of simultaneously monitoring the following parameters which should be present as standard: ECG, NIBP, SpO2, invasive pressures (2), temperatures (2) Should be compatible with Capnography, Cardiac output, EEG, and BIS and prices to be offered as optional

Page 68: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 68 of 166 01.09.2010

ECG should have capability for 3, 5 and / or 10 lead monitoring and should have built in arrhythmia monitoring on all leads. Inbuilt ST segment analysis and arrhythmia detection for all the leads should be possible. Haemodynamic and drug dose calculations should be available. Arrhythmia should be grouped based on classifications – and should show no of arrhythmias occurred. Respiration should be available with Cardio Vascular Artifact filter. ICP monitoring should be possible. Alarm parameter should flash red in the presence of high priority alarms (e.g. ventricular fibrillation and asystole) and flash yellow in the presence of medium or low priority alarms (e.g. noisy signal, etc.) 24 hours trend data should be displayed.

All monitors including central station should have similar user interface for usage among all clinicians. Monitor shall provide the capability to interact with alarms at remote bedsides. Monitor shall provide the capability to receive and display real-time waveforms, trended data and alarm status from other bedside or telemetry units on the patient monitoring network. Monitor shall provide the capability enter patient information at the bedside or central monitor.

On-screen keyboard for entering this data is preferable. Should have USB ports to connect mouse, key board, bar code scanner. Alarm limit status (ON/OFF) must be indicated on-screen for each parameter and actual parameter alarm settings must be displayed on-screen when alarms are on. Position of the displayed waveforms must be user configurable. Waveform color changing should be user configurable. Monitor shall permit the optional ability to receive and display information from other patient devices such as ventilators, infusion pumps and other standalone devices. All modules should be compatible with all monitors quoted. Bed to bed communication between the monitors should be possible with out a central station. Networking to central station should be possible and price of central station should be offered as optional Patient monitoring network shall use standard TCP/IP protocol and be capable of residing on hospital’s network infra-structure. Should be compatible with HIS and should be HL7 compliant. Monitor should provide remote viewing of real time waveforms through internet. Patient monitoring network shall be able to support up to 1,000 monitoring nodes. Should be supplied with necessary accessories for adult , pediatric and neonatal accessories.

Accessories and spares 1. ECG / respiration: 5 lead ECG cable and lead wire set per monitor 2. NIBP: Adult: 2 sizes and Pediatric 2 sizes and neonatal, 1 size per monitor 3. SPo2 Sensor: Adult sensor with cable, pediatric sensor with cable and neonatal sensor with cable per monitor 4. IBP: Include 10 nos of disposable pressure transducer with bracket and interface cable per monitor 5. Temperature: Skin and nasopharyngeal probes per monitor. The equipment should be CE & US FDA Approved.

Page 69: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 69 of 166 01.09.2010

Item No. 27

Ventilator Critical Care Microprocessor controlled Ventilator for Neonates, Paediatrics and Adults patient with invasive & non- Invasive ventilation in both pressure and volume based modes. Should be expandable and up gradable. Should have the both pressure & flow trigger sensitivity. Minimum of following Modes of ventilation should be present: - • CAMV – controlled Assisted mechanical ventilation • SIMV – with pressure and volume support mode (VS) • Pressure controlled ventilation • Tube compensation • PRVC • BIPAP/Bi-level or equivalent with pressure support • CPAP • PAV+ (Proportional Assist Ventilation) or equivalent mode Should have following Parameters: - Tidal volume: 5 to 2000 ml • Frequency: 2 to 150 b/m • Pressure support: 0 to 70 cmH2O • Inspiratory time 0.2 – 8 sec. • Inspiratory flow: 3-150 L/min. • Inspiratory pressure : 5-90 cmH2O • Exhalation Sensitivity: 1-80% of Spont. Peak Flow • Oxygen cone. :- 21 to 100% • PEEP / CPAP : 0-45cmH2O Following parameters should be monitored:- a) Volume: Exp. Tidal volume & M.V b) Tinsp. c) Frequency d) FiO2 e) Pressure: peak, plateau, peep, mean f) Resistance and compliance g) Ti/ Total & RSBI, P0.1 and Vital capacity Should have user programmable Apnea back-up & should have detection of severe patient occlusion to protect patient against excessive airway pressure, terminate normal ventilation and allow patient to exhale through inspiratory limb by opening safety valve. Should have at least 10” size integrated colored touch screen. Should display: - a) Wave forms: P x t, f x t, v x t b) Loops: p x v, f x v c) Should have different color for different breath Alarms: - Audio visual Alarms for low air pressure, low oxygen pressure, low and high inspiratory pressure, low and high rate, leak rate, disconnection, apnea alarm time 15-60 sec, low battery etc. Should have reusable auto cleavable heated bacterial filter exhalation isolation system Inbuilt Battery back up for ventilator for at least 30 min. Essential Accessories:- A) Reusable heated bacterial filter/cassette exhalation isolation system- 10no. B) Humidifier- Heated temperature controlled, preferably temperature monitored with alarms- 1 no. C) Reusable humidifier chamber- 01no

Page 70: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 70 of 166 01.09.2010

D) Heater wires – 01no. E) Heater wire adapter- 01no. F) Flow sensor if applicable with flow sensor cables- 10no. G) Reusable Breathing Circuits adult = 05no. H) Reusable breathing Circuits pediatrics = 05no. I) Reusable breathing circuits neonatal = 01no. J) Nebulizer- Ultrasonic nebulizer with pore size up to 2 micro meters to deliver medicament. It should not affect Ventilator parameter delivered to pt. When in use & it has both timed and Continuous nebulisation mode. K) Compressor a. Should be of same make as of ventilator. b. Should be oil free, medical grade and silent (less then 60db at 1/meter) and flow upto 160LPM. c. Should have high temperature and low-pressure alarms. d. Compatible to be connected to compressed air from hospital central gas supply. L) Trolley The equipment should be US FDA Approved.

Item No. 28

Open Care System It should micro processed control and Temperature range should be 25’C to 38’C Temperature indication range 20’C to 45’C Power Adjustment (Manual Mode) 0-100% Display resolution Skin Temp. Should be 0.1’C Control Precision (skin Temp.) +/- 0.2’C Alarms: - Power failure High temp. - +1’C +/- 0.2’C Low temp. - -1’C +/- 0.2’C Safety High temp. 15 Min. on Maximum power Probe failure, dislodged skin probe Manual Mode – Every 10 Min. It should have indication for Heating Power bar graph, Sound inhibid, manual mode, servo mode. It should have metabolic scale with capacity 7Kg display resolution 2gm. Heater power 560 W It should have rotary reflector 180◦ Voltage 220 V 50/60 Hz Height till mattress area should be 1000mm (+/-100mm)* Mattress dimensions Approx. 565mm X 810mm Bed Inclination grade up to 12◦ It should have height adjustment mortised It should have X ray cassette tray. It should have 4 drawers with tray. It should have observation lamp. It should have blender set for O2 support with two flow meters with one humidification flask and O2 mask with 1.5 meter hose.

Page 71: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 71 of 166 01.09.2010

It should have aspirator with vacuum meter mounted on the front panel.

It should have gas panel with two O2 and two compressed air and one vacuum outlet, ABNT slandered. It should have I.V. Pole. It should have upgar timer 0 to 10 Min. It should have tray for monitor

Item No. 29

TRANSPORT INCUBATOR

· Built-in membrane pump & receiver bottle complete within the incubator chassis, with ECG, SPO2, NIBP monitor. · Oxygen flow meter & 1 lit O2 cylinder with clear front panel display of pressure completely integrated within the incubator. · Manual temp. setting with over temp (40º C) audible & visual alarm. · Internal lighting · Removable infant tray. · Special straps to fix the transport incubator securely to all stretchers in both ambulance & helicopters. · Harness to secure the infant on the tray.

· Power supply: Main supply and dry battery (12V) or mains power supply of ambulance/automobile/helicopter. · Should have a 25x10 mm rail at the back of transport incubator. · Relative humidity up to 60% C distilled water. · Weight with 1 lit O2 tank not to exceed 25 kgs. · Unit should be completely detachable from trolley for putting in plane.

· Unit should be complete with all accessories and startup kit, extra dry battery for 12 hours work, probes electrodes etc.

Item No. 30

Plaster cutter with Aspirator

1.The plaster cutter hand piece should be ergonomic design. 2. The cutter should have two speed control and with less noise 3.the cutter should have impact resistant housing and it should be versatile i.e can be used with vacuum or directly plugged into the wall/point. 4.the vacuum should have mobile on 5 castors for stability and transportation. 5.the vacuum should have facility to connect the hose with the cast cutter. 6.locking mechanism should be there in the system to lock the hose with the cutter. 7.the length of the hose should be 8 feet. 8.one complete system should be supplied with following items/accessories- I. Cast cutter hand piece-1 no. II. Cast cutter blades-20 nos(different sizes) III. Cast vacuum with hose and mobile stand-1 no. IV. Blade mounting toll-2 nos V. Small cast spreader-1 no VI. Large cast spreader-1 no.

Page 72: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 72 of 166 01.09.2010

Item No. 31

Tourniquet Electronic 1. Electrically operated system with two hoses and battery back up of upto 45 min. 2. Should have option for bier"s block and bilateral procedures.

3. Seven sizes 8'' to 42" of cylindrical cuffs should meet individual requirements of thin and fat patient for arm and thigh. Cuffs should be autoclavable. 4. Small and light weight unit that can be mounted on i.v pole or placed on a table. 5. Audible and visual alarms when pressure variation is detected 6. Unit should perform self-calibration check every time it is turn on. 7. 50 to 475 mm hg pressure setting

8. 6 to 240 minutes time setting.

9. Positive locking connectors, leak free inflation cuff sleeves, which help reduce wrinkles pinching and shearing of soft tissues. 10. Cuff with lockout features, which should ensure that the cuffs stays inflated

Item No. 32

HOT PACK UNIT · Must be full stainless steel unit with wheels with option to easily change water and heating capacity of upto 12 large packs at one time and temperature preset at 167 degree F. · Should have a PVC coated and low-water cutoff feature with special insulate to preserve heat and help conserve power consumption · Must come with built in adjustable thermostats and provide lighted on/off switches · Should have concelead elements, coved bottoms, coated racks and hospital-grade power cord · Unit must have inside mesh and hangers to hold the pads for proper stable heating to transfer · Unit should be provided with different size white clay pads and able to hold temperature upto 30minutes of deep moist heat. · Hot Packs able to be reused for hundreds of treatments and provide 12 months of warranty · Should come with durable custom size and shapes Terry covers for hot packs with different sizes for cervical, lower back, extremities and more · UL-listed, ETL/CE and CSA –approved with 220 volts

Item No. 33

COLD PACK UNIT · Must be Five Cubic Feet of Storage and able to hold 12 Gel packs · Should have adjustable thermostatic control and drain for defrosting · Dimension approx 27” deep, 34 “ high has to be a cooler and not a freezer

Page 73: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 73 of 166 01.09.2010

· Have to provide compressed cold therapy pack for extremities able to 360 degree around the injured area made out of durable Nylon outer chamber. · Must provide Body ice packs with non-freezing gel · Must be made out of PVC Vinyl exterior and available in different sizes for different body parts cervical, lumbar, and extremities · Should able to hold temperature up to 30 minutes · UL-listed, ETL/CE and CSA –approved with 220 volts option available

Item No. 34

Traction Unit · Must be FDA approved decompression-traction unit · Should able to deliver decompression therapy along with 3-channel of light therapy to muscle stimulation and blood circulation · Should able to deliver light therapy treatment and decompression simultaneously. · Should have a touch screen interactive display for treatment set-ups and easy angle selection and must come with treatment protocol manual · Must provide along with the package Angle reference chart · Should able to automatically calculate and digitally display the rope pull angle for decompression and traction as per the treatment protocol · Must provide a 8”x 10’ Infrared light Pad unit for muscles relaxing and muscle spasms · Must have built in training information for traction and light therapy protocols · Should provide protocol manual for light therapy and lumbar and cervical protocol manual · Must be a build in computerized software package and protection against accidental setting of force-must have a safety switch for emergency shut of

· Should come with Flexion stool, Knee bolsters, Cervical pillow, Ankle bolsters, and decompression belts thoracic and pelvic. · Should provide with a 4 section motorized table hi/lo with clamps, frame attachments for connecting the traction unit · Should be a certified class device with all CE mark and US FDA approved Unit, and must provide proprietary certificate and IEC 60601-1(CE) and CSA/NRTL

Item No. 35

Operating Laparoscope Set for Gynaec

Laparoscopy TROLLEY Automatic Light source 220 V,300 W. Xenon Bulb Bulb Working life more than 500 Hrs. Bulb life counter on light source

Stand By Mode via push button on light source console

Universal Jaw Assembly to adapt any make of fiber optic cable without use of any adapters.

Fiber optic Cable

High Definition Monitor

Page 74: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 74 of 166 01.09.2010

High definition monitor, resolution 1280 x 1024 DVI input.

Option for wall mounting and desktop in same unit. CO2 Insufflator Minimum 40 Liter of high flow Microprocessor controlled unit Soft Approach Pressure control for safe recovery of abdominal pressure Gas heating

HD Camera System Camera console with universal coupler & Autoclavable camera head Pure Digital signal with high definition video(1280*1024 native resolution) Progressive scan technology both on camera head & console Digital zoom & white balance on camera head Video Output 1. 2 DVI output 2. 2 SVHS & 1 RGB out put 3. One Composite out put Ideal Eye/HD Laparoscopes, Fully Autoclavable with working length 300mm Wide angled distortion free view Universal adaptor for other light cable High quality sapphires and the latest rod lens technology for excellent transmission and detail recognition 30 degree, 10mm x 1 0 degree, 10mm x 1 0 degree, 5mm x 1 Instrument set: Laparoscopic hand instruments (reusable) and rotatable Veress needle 12 cm length 1 Veress needle 15 cm length 1 Trocars sleeves 11 mm 3 Reducer 11/5 mm 2 Trocars sleeves 5.5 mm 3 Trocars (pyramidal tip) 10 mm 2 Trocars (pyramidal tip) 5 mm 2 Maryland dissector 5mm with unipolar diathermy 1 Atraumatic graspers, 5mm 2 Tooth Grasper 5mm 2 Metzenbaum scissors (5cm) with unipolar diathermy 2 Laproscopic cautery lead 2 Laproscopic bowel grasper 5mm, length 33-36 cm 1 Laproscopic suction cannuala, 10 mm 1 Laparoscopic suction cannula 5 mm 1 Bipolar Forceps 5mm Cilpenger type 1 Bipolar cable 1 Kelly Dissecting Forceps Long Jaws 5mm 1 Fenestrated Long Jaw Forceps 5mm 1 Dissecting & Grasping Forceps for ovaries 5mm 1 PCOD Dissecting Needle 5mm 1 All the Units should be US FDA Approved Cidex Tray (Indian)-2 Formalin Chamber (Indian)-2

Page 75: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 75 of 166 01.09.2010

Item No. 36

HYSTEROSCOPY SYSTEM

Hysteroscopy set – telescope 30 deg. Diameter 4mm length 30 cm, Autoclavable x 1 No. 2. Examination sheath Diameter approx 5.4 mm x 1 No. 3. Continuous flow operating sheath approx 6mm x 1 No. 4. Operating sheath approx 5.4 mm x 1 No. 5. Deleted 6. Biopsy forceps, trough cutting single action jaws, approx 5FR x 2 No. 7. Scissor, signal action jaws, approx 5FR x 2 No. 8. Biopsy and grasping forceps approx 5FR x 2 No. 9. Hysteroscope pump for Diagnostic & operative Hysteroscopy, Roller Pump working 220 – 240 VAC, pressure pre – Selection : - 150 - 200mm Hg, Flow regulation: 0 – 500ml/mm. Equipments should be supplied along with set for at least 300 procedures. x 1 No. 10. Suction Unit with following Essential accessories: i. One pedal foot switch one stage x 1 No. ii. Suction bottle 51 x 1 No. iii. Bottle cap x 1 No. iv. Bottle stand x 1 No. v. Bottle stand holder x 1 No. RESECTOSCOPE To be compatible with Hysteroscope mentioned above. 1. Working element set consisting of x 1 set i. One working element set. ii. Two cutting loops. iii. Two coagulating electrode. iv. Two high frequency cord. v. One protection tube. 2. Resectoscope sheath including connecting tubes For in and out flow, approx 26 FR oblique beak, Inner tube with ceramic insulation, rotatable x 1 No. 3. 3. Standard obturator for use with approx 26FR sheath x 1 No.

Item No. 37

Cryo Surgical Unit 1. Facility to cutoff gas line without closing cylinder while inter changing probes. 2. Twin Trigger system to frost and defrost. 3. Cleaning the inner hypodermic needle should be much easier 4. System should be operated by CO2 or NO2 gas. 5. Console to house cylinder with temperature indicator

6. Probes Tips: Exo Cervical, Endo Exo Endo cervical and long Endo Exo Cervial, HPV, Flat and Round in different shapes. 7. System high quality and reliability.

Page 76: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 76 of 166 01.09.2010

Item No. 38

Vacuum Extractor 1. Noiseless suction unit have fast vacuum built up 2. Vacuum should have maximum -90 k pa/-675 mm Hg, Suction capacity 50ltr/min/1kg/cm2 & bottle capacity 3ltrs 3. Suction system should have piston / cylinder ( self lubricating ) 4. Should have mechanical overflow protection system . 5. Set of silicon cups 50mm & 60mm-2each 6. Set of bird cups, stainless steel 40mm, 50mm & 60mm-2each 7. Machine can be operated on 220-240 V AC single phase 8. Should be provided with a gauge to display of vacuum generated 9. Should have vacuum release valve.

Item No. 39

Fetal Monitor Cardiotocograph machine with twin monitoring capability should meet the following specification and capabilities: · FHR twin monitoring using external ultrasound · Direct ECG and maternal ECG measurements. · Uterine activity using an external toco transducer or IUP catheter. · Fetal movement profile parameter to record accurately the fetal movements using the ultrasound channel without additional procedures or transducers and statistics for advance information on fetal well-being. · Low ultrasound energy to the fetus. · Audible alert indication of fetal bradycardia and tachycardia · Audible indications of paper out and NST time complete. · Should have a feature to provide more accurate and continuous fetal heart rate (FHR) thereby reducing the need for repositioning the ultrasound transducer. · Should have the facility of cross channel verification when two channels are picking up the same signal. · Should have signal quality indicators guiding to obtain the strongest and most continuous ultrasound HR signal. · Built-in multi channel high resolution thermal array recorder with visual and audible paper end detection and should annotate time of day, date and paper. Should be supplied with the following accessories: · Mobile cart with two drawers and integrated mounting rail. · 2 x ultrasound transducers · 1 x external toco transducer · 1 x ECG module with degl & MECG adapter cable · 20 x 250 g bottle of gel.

Page 77: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 77 of 166 01.09.2010

· 100 numbers of disposable signal spiral fetal scalp electrodes, quick connect type · 80 packs of recording paper. (to be supplied as per the usage; i.e. in a manner that they should not get faded without being used.)

Item No. 40

ENT Examination Unit · Hose connection for irrigation of maxillary duct. · Splash protector · Water warmer, 37 deg with electric temperature control. · Compressed air device, complete with 2 medicament sprays, one powder blower, adapter and spray holder · Jet connetion for water irrigation handle, short, medium, and angles. · Connection box for automatic switching on/off water supply by master switch. · Mains supply unit, with connection box, and installation kit. · Head light with exit source light. · Cotton swabs dispenser. · Tongue patches dispenser. · Holder for ear funnels · Water filter, complete for warm water equipment. · Cold light source, with additional rear light outlet, for operating microscope. · Operating microscope mounted on unit, with double pantograph arm, straight binocular head. · Endoscope holder and warmer. · Air flow for cold light, to prevent misting · Mirror pre warmer. Instrument cabinet all steel casing comprising of: · Suturing complete set · 4 large instrument trays · Aluminium/SS trays · Shallow drawer · Deep drawer · Used instrument deposit Swivel chair adjustable high with back rest Should be supplied complete with treatment couch and compressed air & vacuum.

Item No. 41 Operating Laparoscope Set for Gen Surgery Laparoscopy TROLLEY Automatic Light source 220 V,300 W. Xenon Bulb Bulb Working life more than 400 Hrs. Bulb life counter on light source Stand By Mode via push button on light source console Universal Jaw Assembly to adapt any make of fiber optic cable without use of any adapters. Fiber optic Cable High Definition Monitor

Page 78: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 78 of 166 01.09.2010

High definition monitor, resolution 1280 x 1024 dvi input. Option for wall mounting and desktop in same unit. CO2 Insufflator Minimum 40 Liter of high flow Microprocessor controlled unit Soft Approach Pressure control for safe recovery of abdominal pressure Gas heating HD Camera System Camera console with universal coupler & Autoclavable camera head Pure Digital signal with high definition video(1280*1024 native resolution) Progressive scan technology both on camera head & console Digital zoom & white balance on camera head Video Output 1. 2 DVI output 2. 2 SVHS & 1 RGB out put 3. One Composite out put High Definition/ Ideal Eye Laparoscopes, Fully Autoclavable with working length 300mm

Wide angled distortion free view Universal adaptor for other light sources High quality sapphires and the latest rod lens technology for excellent transmission and detail recognition 0 degree, 10mm x 1 30 degree, 10mm x 1 0 degree, 5mm x 1 Instrument set: Laparoscopic hand instruments (reusable) and rotatable Veress needle 12 cm length 1 Veress needle 15 cm length 1 Trocars sleeves 11 mm 3 Reducer 11/5 mm 2 Trocars sleeves 5.5 mm 3 Trocars (pyramidal tip) 10 mm 2 Trocars (pyramidal tip) 5 mm 2 Maryland dissector 5mm with unipolar diathermy 2 Atraumatic graspers, 5mm 2 Tooth Grasper 5mm 2 Metzenbaum scissors (5cm) with unipolar diathermy 2 Laproscopic cautery lead 2 L shaped hook electrode 5mm 2 Laproscopic bowel grasper 5mm, length 33-36 cm 1 Laproscopic suction cannuala, 10 mm 1 Laparoscopic suction cannula 5 mm 1 Clip applicator 10 mm for Large, Medium Large, Small Clips Claw Forcep 10mm 1 Babcock forcep 1 All the units should be US FDA Approved Cidex Tray (Indian)-2 Formalin Chamber (Indian)-2

Item No. 42

Multipara Bediside Monitor

Page 79: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 79 of 166 01.09.2010

Patient monitor system should be of modular type and capable of monitoring adult, pediatric & neonatal patients. Monitor should have 17” independent flat panel display. Touch screen user interface . Module rack / housing should be independent and shall be able to be placed near to the patient. Should be capable of 8 traces display. Monitor must be capable of simultaneously monitoring the following parameters which should be present as standard: ECG, NIBP, SpO2, invasive pressures (2), temperatures (2) Should be compatible with Capnography, Cardiac output, EEG, and BIS and prices to be offered as optional ECG should have capability for 3, 5 and / or 10 lead monitoring and should have built in arrhythmia monitoring on all leads. Inbuilt ST segment analysis and arrhythmia detection for all the leads should be possible. Haemodynamic and drug dose calculations should be available. Arrhythmia should be grouped based on classifications – and should show no of arrhythmias occurred. Respiration should be available with Cardio Vascular Artifact filter. ICP monitoring should be possible. Alarm parameter should flash red in the presence of high priority alarms (e.g. ventricular fibrillation and asystole) and flash yellow in the presence of medium or low priority alarms (e.g. noisy signal, etc.) 24 hours trend data should be displayed.

All monitors including central station should have similar user interface for usage among all clinicians. Monitor shall provide the capability to interact with alarms at remote bedsides. Monitor shall provide the capability to receive and display real-time waveforms, trended data and alarm status from other bedside or telemetry units on the patient monitoring network. Monitor shall provide the capability enter patient information at the bedside or central monitor. On-screen keyboard for entering this data is preferable. Should have USB ports to connect mouse, key board, bar code scanner. Alarm limit status (ON/OFF) must be indicated on-screen for each parameter and actual parameter alarm settings must be displayed on-screen when alarms are on. Position of the displayed waveforms must be user configurable. Waveform color changing should be user configurable. Monitor shall permit the optional ability to receive and display information from other patient devices such as ventilators, infusion pumps and other standalone devices. All modules should be compatible with all monitors quoted. Bed to bed communication between the monitors should be possible with out a central station. Networking to central station should be possible and price of central station should be offered as optional Patient monitoring network shall use standard TCP/IP protocol and be capable of residing on hospital’s network infra-structure. Should be compatible with HIS and should be HL7 compliant. Monitor should provide remote viewing of real time waveforms through internet. Patient monitoring network shall be able to support up to 1,000 monitoring nodes. Should be supplied with necessary accessories for adult , pediatric and neonatal accessories.

Accessories and spares 1. ECG / respiration: 5 lead ECG cable and lead wire set per monitor 2. NIBP: Adult: 2 sizes and Pediatric 2 sizes and neonatal, 1 size per monitor

Page 80: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 80 of 166 01.09.2010

3. SPo2 Sensor: Adult sensor with cable, pediatric sensor with cable and neonatal sensor with cable per monitor 4. IBP: Include 10 nos of disposable pressure transducer with bracket and interface cable per monitor 5. Temperature: Skin and nasopharyngeal probes per monitor. The equipment should be CE & US FDA Approved.

Item No. 43

Emergency Resuscitation Kit (imported)

1. To have Retromolar Intubation fiberscope for unexpected difficult airways. a. Tip Distal Bending 40°. b. To be movable eyepiece c. To have a light source connection d. With length 40-42cms and dia 5-6 cms. e. ET tube holder should be provided f. Should take min. 5.5 size of ET tube 2. Portable LED light source should be provided i. with illumination not less than 50000 Lux ii. should run on two 3v photo batteries iii. burning life should be more than 100 minutes iv. ergonomically designed and can be connected to both the fibrescopes v. life of LED should be close to 50000 hrs 3. One Laryngoscope with rechargeable battery pack and blade with fibreoptic mechanism should be provided to be used on both adult and pediatric patients with charger. 4. Other accessories like, magill forceps should be provided. 5. Should have Emergency Cricothyroidotomy for pediatric and adult i. disposable blades ii. dialator 6. Should have Combitube size 37Fr. i. with complete kit 7. Should have Intubating Laryngeal Mask Airways with Following Components: a. ILMA Sizes 3 & 4. b. ILMA Tubes ID 7mm & 7.5mm. c. Tube Stabilizing rod d. Cuff deflator 8. Should have Laryngeal Mask Airways i. sizes 1,2 and 4 9. Handy and strong brief case/bag should be provided to keep all the instruments safe. 10. Set of disposable percutaneous tracheotomy kit for adult and pediatric. 11. Should have standard AMBU bag for pediatric and adult. 12. Mechanical suction pump with suction catheter and stomach tubes. 13. Should have Aluminum Oxygen reservoir 2 Liter with oxygen tube and cather. 14. Oxygen pressure reducer, regulable 0-15 liter with coupler for respirator. 15. Ventilating bag 16. Lubricant 17. Blood pressure meter, boso K-II

Page 81: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 81 of 166 01.09.2010

18. Stethoscope 19. Rescue blanket gold/silver 20. Infusion system.

Item No 44

Portable Spirometer • The unit should be small, portable, with a built in caring handle convenient for Ward rounds and on site patient screening. • The unit should Menu driven and operate on 220V AC • The unit should incorporate a Fleisch type pneumotach autoclavable sensor with no moving Parts. • Should have built in storage up to many thousands of patients with their life time data for individual or batch printing. • The unit should incorporate with large touch screen displays with easy to follow instructions, flow/volume & volume/time graph, post test, incentive display and an inbuilt printer. • The unit should have facility to direct connectivity to PC and able to transfer the data from unit to PC and printout facility through external printer. • The unit should perform single breath test, VC, and closed circuit spirometry. • The unit should measure the following parameter : VC, FVC, FEV.5, FEV.5%, FEV1, FEV1/VC%, FEV1/FVC%, FEV/PEF, FEV3, PEF, FEF25-75, FEF25-75%, FEF25%, FEF50%, FEF75%, FMFT, FET, MVVind, FIVC, FIVC/FVC%, PIF, FIF 25%, FIF50%, TV, RV, IRV, ERV, IC, etc. • The print out should be configurable with choice of curves, parameters printed, interpretations and test quality control message . • The equipment warranty should be of one year. • The unit should be compliant to all standard like ATS 94, ERS 94, CE, FDA, and EN60601.

Item No 45

Infusion Pump Syringe

It should be Light weight , Compact & should be able to adopt syringes from 5 to 60ml horoizontal

Programming : Should be ml/hr, ml/hr x volume limit & time x volume limit

Flow rate should be 450 ml/hr with 0.1ml/hr increment

Volume Limit should be 999.9 ml

Must have Occlusion Pressure at least 3 levels

KVO should be 1ml/hr or less

Bolus rate should be 450 ml/hr

Time limit from 1min to 99hrs & 59 minutes.

Special Functions Like : Clear Volume

Should have pre alarms such as End of Battery, End of Infusion, & visual & sound alarms like Syringe

positioning error, occlusion, infusion completed, syringe error, programming error,

Block error, clamp error , low battery etc

Flow rate deviation should not be more than +/- 1% of the programmed flow rate.

Battery Rechargeable with battery life 3hrs , Should be able to work on 230 Volts

The product should have CE & ISO 13485 certifications

Item No 46

PFT Machine(Spirometer & Computerised)

Page 82: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 82 of 166 01.09.2010

The unit should be light weight, exclusively designed to carry out examination of pulmonary system and to measure & display lung function parameters, operable on USB port power , not requiring large installation space ( compact).

It should come with an incentive display for cooperation free, lung testing.

It should be designed for the following measurement protocols

Spirometery with pre/post comparison, Flow/Vol and Vol/Time loops/curves and Trend Loops.

Individual interpretation assistance.

MVV

Patient Co-operation Display

It should measure following options without using additional gas analyzer

Capno-Volumetery measurement

Emphysema diagnostic

Determination of anatomical and functional dead spaces

The system should incorporate a light weight, Ultrasonic sensor (non-heated) free from any kind of frictional inefficiencies which is absolutely insensitive to moisture. It should not require any warm-up time and calibration with following measurement ranges :-

Flow Measurement

Range : 0 to +/- 16 Ltrs/sec with accuracy better than +/- 3% or 100 ml/sec .

Volume measurement

Range : 0 to 10 ltrs with resolution of 10 ml

Corrections

Flow Volume : EGKS or ATS

Inspiratory Gas qty : BTPS (ambient module)

The machine should measure the following Parameters:

SVC , V C ex, V C max, IRV, ERV, VT, IC, Frequency, MVV, , ti, te, ti/te, FVC, FEV1,

FEV 0.5, FEV 1/ VC max, FEV 1/ FVC ex, MEF25, MEF 25-75, PEF, PIF, MVV and pre-post comparisons.

Should measure CO2 max %, Vm25-50, Vm50-75 etc.

The unit should be supplied with 50 no of disposable breathing tubes.

The unit should be supplied complete with hardware and software including a PC as per below given specifications from a respected brand like IBM, Compaq or HP.

a) CPU – Pentium IV Core 2 DUO Processor.

b) RAM – 1 GB, At least 160 GB HDD

c) At least 14” TFT Color Monitor

d) DVD R/W Drive, Keyboard, Mouse

e) Serial / Parallel interface PCI SLOT

f) Windows XP

g) Inkjet Printer.

h) Suitable UPS for the computer system

The manufacturer should be an EN ISO accredited company and the product should be duly CE marked to MDD 0124 for medical devices.

Page 83: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 83 of 166 01.09.2010

Item No 47

Fully Automated Biochemistry Analyzer

Fully automated, latest and bench top analyzer to perform the analysis of substrates, enzymes and special parameters from whole blood, serum, plasma and urine samples

System should be Discrete, fully selective random access with a provision to test STAT samples

System should have four different on-board technologies ( Photometry, Potentiometry, Fluorescence Polarization and Turbidimetry ) to measure substrates, enzymes, Homogeneous immunoassays, TDM’s and Drugs.

System should have facility for programming 125 - 150 different test parameters and the reagents should be available from the same manufacturer.

System should have a routine throughput of 400 tests / hr

Onboard sample capacity should be at least 90 or more

Flexibility to use different sample containers like primary tubes with different sizes, sample cups, micro cups and cup on tube for easy processing.

Facility to keep reagent bottles / cassettes for at least 30 common tests with on board refrigeration is must.

Sample volumes should be less than 2 - 10 ul per test.

System should have high sensitive pressure sensors to detect any incorrect pipetting even at 2 ul sample volume

Onboard sample and calibrator dilution should be available ( 1 – 100 times )

System must use disposable cuvettes to prevent any carryover without using any onboard washing

System should be used for testing special parameters like HbA1c, Lactate, hsCRP, D-Dimer, Ferritin, IgA, IgM, IgG, ASO, Cyclosporine, MPA and electrolytes ( Na, K and Cl ), TDM, DAT tests besides the routine clinical parameters

On-board reagent stability should be for at least 3 months and calibration of the parameter should be typically with lot. No daily calibration should be required by the system to save the reagents.

System should have 12 wavelength photometer with mono and bi-chromatic measurements

Light source should be 20 W halogen lamp with lamp save feature

System should have external windows NT based data control work station with flat screen monitor for programming the tests and entering the patient data.

System should external printer to take printout of patient results

Patient samples and Reagents can be scanned with barcode scanner for easy operation.

System should have 1 x RS 232 bidirectional interface and in-built modem for remote diagnostics access.

Page 84: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 84 of 166 01.09.2010

Power supply – 220 V / 50 Hz. Suitable online UPS should be supplied along with the system

Item No 48

Centrifuge Machine 24-36 tubes

1 Description of Function

1.1 Centrifuges are required in the Laboratory to separate various components of Blood for analysis.

2 Operational Requirements

2.1 Aerodynamic compact construction for vibration free performance

2.2 Table top version

3 Technical Specifications

3.1 Angle Head rotor with Tube Capacity Size 5 – 15 ml

3.2 To have capacity to hold 24-36 tubes at a time.

3.3 Should have a digital timer

3.4 Body should be made of strong fabricated & corrosion resistant steel

3.5 Control panel – for start/stop switch, dynamic brakes, step less speed regulator with zero

start switch & speed indicator with timer and protective fuses.

3.6 Door interlock

3.7 Maintenance-free brushless drive motor with exact speed pre selection and alphanumerical

interactive LCD digital display in control panel of RPM & RCF. Speed range 100 to 6000

rpm and above.

3.8 Choice of acceleration and braking profiles.

3.9 Imbalance detection and auto shut down.

4 System Configuration Accessories, spares and consumables

4.1 Tube Holders as appropriate

5 Environmental factors

5.1 Shall meet IEC-60601-1-2:2001(Or Equivalent BIS) or General Requirements of Safety for

Electromagnetic Compatibility.

5.2 The unit shall be capable of operating continuously in ambient temperature of 10 -40deg C

and relative humidity of 15-90%

5.3 The unit shall be capable of being stored continuously in ambient temperature of 0 -50deg C

and relative humidity of 15-90%

6 Power Supply

6.1 Power input to be 220-240VAC, 50Hz as appropriate fitted with Indian plug

7 Standards and Safety

7.1 The supplier should be preferably ISO certified for quality standards.

7.2 Product should preferably be FDA/CE or ISI approved

7.3 Should preferably comply with IEC/TR 61010-3-020 :Safety requirements for electrical

equipment for measurement, control, and laboratory use - Part 3-020: Conformity verification

report for IEC 61010-2-020:1992 Particular requirements for laboratory centrifuges"

7.4 Should preferably comply with IEC 61010-1 Safety requirements for electrical equipment for

measurement, control, and laboratory use - Part 1: General requirements

8 Documentation

8.1 User manual in English

Page 85: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 85 of 166 01.09.2010

8.2 Service manual in English

8.3 Certificate of calibration and inspection from factory

8.4 List of Equipments available for providing calibration and routine maintenance support as per

manufacturer documentation in service / technical manual

8.5 List of important spare parts and accessories with their part number and costing

Log book with instructions for daily, weekly, monthly and quarterly maintenance checklist. The job

description of the hospital technician and company service engineer should be clearly spelt out.

Item No 49

Computed Radiography System

Description of function

Radiography system to replace conventional Film/screen based X-Ray processing techniques with Photostimulable Phosphor Plate technology based digital technology.

Operational requirements

The system shall be able to record X-Ray images on Imaging Plates(IP)

The IP Shall be housed in CR Cassettes that have a Radio frequency Tag to store demographic data.

Operationally and functionally equivalent to and better than the present film based system

Convert these images from the IP into digital values and transfer these values to an image evaluation computer with predefined Image Processing Parameters.

Should record Paient Identification data on the on the image.

Maintain and manage data bank of all patient and image data.

Retrieve and reproduce accurate, high quality high resolution images from stored data without loss of image quality.

Present CR images on a workstation as well as on hard copies.

show full image in the X-Ray room for preview purposes.

Read and Write in CD.DVD for data Storage and review.

Appropriate technology to provide uniform and thick slice thickness.

Technical Specifications

Image Reader will have the following :

Cassete Machanism to Load and Unload IP.

Scanning mechanism to read, erase and process the images from the imaging plate. (IP)

Including autorouting newly acquired images to desired preview monitor.

-IP processing rate> 50 plates/hr.

-Panel for indicating online status of the CR Reader in case of machine malfunction.

-Emergency Mode for accepting exposed cassettes

without patient demographics for casualty trauma workflow requirements.

-Capability of retrieving at least 10 scanned images and quick check of the exam data and the image of at least the last four cassettes scanned at the X-Ray room.

- Verification of the connectivity status of configured image desination.

-Spatial resolution of digital image 6-10 pixels/mm..

-Scanning resolution for all the IP Plates should be specification in the quote.

-Should enable 12 bit images.

-X-Ray Generator compatibility with reputed manufacturers..

-Image matrix at standard resolution (14 ´ 17) - 2000 ´ 2500 Row ´ Column

-Image matrix at standard resolution (14 ´ 17) - 3000 ´ 4000 Row ´ Column

Page 86: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 86 of 166 01.09.2010

CR Workstation will have the following.:

--Capable of Archieving and printing selected images to a standard DICOM destination in DICOM 3.0 format.

--Sroring images in the local disk for predefined period.

-Mechanism for accepting new images when the local disc is full..

-Sorting of patient image based on name, date, exam etc.

-Advance Processing Software.

Using predefined parameters or user defined and stored image parameters.

-Correcting typographical in patient demographic module, in case RIS connection was down and manual data entry was done.

-Capability of changing R/L, Flipping, Rotating, Zooming, Collimating, annotating the incoming image..

-Multi image and slide formats.

-Capability of storing in CD/DVD.

-Software for Advance Image processing, applications, dispay and quality monitoring.

-Connectivity and compatibility to communicate to RIS/HIS and DICOM Compatible devices such as MR/CT/DSA Work station,

-Shouild provide for HL-7 compatible interface.

-Output grayscale resolution- 16 bits/pixel.

Remote ID and Preview station. Should have the following:

-Auto detection of cassette.

-Mechanism of writing /reading data on/from RF Tag..

-HIS/RIS/DICOM Compatibility.

-Preview scanned images on predefined preview terminal.

-Retrieving capality of last 10 patient ID on the terminal.

-Identification of overxposure on preview module.

-Mechanism for user release in case of autorouting images. To predefined DICOM destinations

-System should be able to support minimum 5 review terminals

-Preview display time <45 sec.

Color dry view imaging printer

(film based) with the following:

-Print Image from CR workstation. In DICOM 3 format.

-Mechanism to print images to 14x17 and 8x10 film sizes simultaneously.

-Docked in processor.

-Resolution> 600 DPI.

-Multiple Image and slide printing capability.

-Shall be able to swtich between Receiver Mode and Processor mode.

System Configuration Accessories, spares and consumables

IP/Cassettes size- qty

35 cmx35 cm 03

24 cmx30 cm 03

15 cmx30 cm 03

20 cmx25 cm 04

25 cmx30 cm 08

Image Reader system 01

CR Workstation 01

RIS Interface 01

Remote ID and Preview station. 01

Archieving System 01

(film based)

Page 87: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 87 of 166 01.09.2010

Black and white thermal video printer

Black and white laserjet printer for reporting

Environmental factors

The unit shall be capable of operating continuously in ambient temperature of 300 C and relative humidity of 80%

Power supply

Power input to be 220-240VAC, 50Hz, fitted with Indian plug

Resettable overcurrent breaker shall be fitted for protection

Spike protector of appropriate rating should be provided

UPS of suitable rating conforming to IS-302 shall be supplied

Standards and safety

Should be FDA or CE approaved product

Electrical safety conforms to standards for electrical safety IEC-60601 / IS-13450

Comprehensive guarantee for 5 years of complete system and all other parts for which order will be placed with uptime warranty of 98%. Comprehensive Guarantee period will be extended by double the downtime if it exceeds more than 2% in a year.

Rates of comprehensive AMC for complete system and all parts for which order will be placed with an undertaking of 98% uptime and extension of AMC period by double the downtime if it exceeds more than 2%.

Documentation

User manual in English

Service manual in English

List of important spare parts and accessories with their part number and costing.

Certificate of calibration and inspection from factory.

Log book with instruction for daily, weekly, monthly and quarterly maintenance checklist .

The job description of the hospital technician and company service engineer should be clearly spelt out

Item No 50

Advancved Colour Doppler Ultrasound Scanner

Description of function

Fir whole body Ultrasound Anatomical studies, Blood flow Studies and 3D studies.

Operational requirements

Digital Ultrasound system platform for excellent 2D, Colour & Power Doppler and 3D Imaging capability.

Technical Specification

Must have minimum 1000 digital channels*

Should have 15" or more high resolution monitor with tilt and swivel.

Line density 512 lines

Page 88: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 88 of 166 01.09.2010

Dynamic range>160dB

Penetration upto 30 cms

Upto 4 selectable frequencies in each probe

Tissue harmonic imaging with phase inversion, pulse inversion, or wide-bandwidth imaging technology.

Inbuilt 3D imaging with hand acquisition and auto sweep.

Must have minimum of 4 rendering modes with measurements.

Panaromic imaging upto 240 cms or 180 deg with colour and cine facility.

Should automatically equalize gain and brightness with touch of one button.

Ability to enhance 2D and tissue harmonic penetration and colour sensitivity momentarily to improve visualization in difficult patients.

Imaging with multiple line of sight combined to a single line of sight to imrove resolution.**

Should provide for vascular imaging enhancing by using power Doppler to enhance B Mode image.

Appropriate technology to provide uniform and thick slice thickness.

Software for various applications including Vascular, Abdomen, Foetal echo, Transcranial and cardiac studies should be available.

Accessories, spares and consumables

1-4 MHz Convex Array probe.

5-13 MHZ Liner Array Probe

4-10 MHz Endocavity probe

Volume probe

PC Based Image management system

Color dry view imaging printer

(film based)

Black and white laserjet printer for reporting

Environmental factors

The unit shall be capable of operating continuously in ambient temperature of 300 C and relative humidity of 80%

Power supply

Power input to be 220-240VAC, 50Hz, fitted with Indian plug.

Resettable overcurrent breaker shall be fitted for protection

Spike protector of appropriate rating should be provided

UPS of suitable rating conforming to IS-302 shall be supplied

Standards and safety

Should be FDA or CE approad product

Electrical safety conforms to standards for electrical safety IEC-60601 / IS-13450

Comprehensive guarantee for 5 years of complete system including all transducers and all other parts for which order will be placed with uptime warranty of 98%. Comprehensive Guarantee period will be extended by double the downtime if it exceeds more than 2% in a year.

Rates of comprehensive AMC for complete system and all prats for which order will be places with an undertaking of 98% uptime and extension of AMC period by double the downtime if it exceeds more than 2%.

Documentation

User manual in English

Page 89: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 89 of 166 01.09.2010

Service manual in English

List of important spare parts and accessories with their part number and costing.

Certificate of calibration and inspection from factory.

Log book with instruction for dail, weekly, monthly and quarterly mantenance checklist.

The job description of the holpital technician and company service engineer should be clearly spelt out

Item No 51

Donor Chair

1. Description of Function

Blood Donor Couch is a completely automatic enveloping, variable tilt chair and

specially designed to make blood withdrawals easier, safe and functional, and also for

other diagnostic and therapeutic areas

2. Operational Requirements

1) Provides a comfortable position for the donor.

2) Variable positioning for either arm with Comfortably wide armrests.

3) Armrests have swinging out as well as up and down moving facility.

4) Reclining and upright body positions with a smooth shifting to any position.

5) Both sides have supporting brackets.

6) Drawers provided for the upkeep of equipment & consumables.

7) If a vasovagal attack occurs the Donor's head needs to be lowered immediately and

his legs lifted above his heart level so that blood can flow back to the brain and other

vital organs. This facility should be available

3. Technical Specifications

3.1 Comfortable chair type with soft padding for cushioning and rexin cover.

3.2 Seat, back rest and leg rest size designed for donor comfort. It should have step

less electric remote controlled height adjustment.

3.3 Adjustable arm rest for donor’s comfort and phlebotomist friendly

3.4 Easily tilted to head low position, electrically operated

3.5 Comfortable working level for the operator. Lifting capacity - Approx 200 kg.

3.6 4 Lockable castors for easy mobility

3.7 Storage Drawers for storing consumables & Blood Collection Monitors

3.8 UP/DOWN control

3.9 OPTIONS:

(i). A paper roll holder can be fixed on the' upper part of the chair.

(ii). Melodious musical Headphone can be integrated for patient relaxation while blood

donation is in progress.

(iii). Preferable to have inbuilt trays & stands for keeping all blood collection

accessories.

3.10 Should have interface for blood collection monitor (optional)

4. System Configuration Accessories, spares and consumables

4.1 Donor Couch -01

4.2 Dust Cover -01

4.3 Power cable -01

4.4 Arm Rests (pair) -01 pair

4.5 Remote control -01

5. Environmental factors

5.1 The unit shall be capable of operating continuously in ambient temperature of 10 –

400 C and relative humidity of 15-90%

5.2 The unit shall be capable of being stored continuously in ambient temperature of 0 -

40 C and relative humidity of 15-90%

5.3 Shall meet IEC-60601-1-2: 2001 (Or Equivalent BIS) General Requirements of

Safety for Electromagnetic Compatibility.

Page 90: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 90 of 166 01.09.2010

6. Power Supply

6.1 Power input: 220-240V/ 50 Hz AC Single phase or 380-400V AC 50 Hz Three

phase fitted with appropriate Indian plugs and sockets.

6.2 Resettable over current breaker shall be fitted for protection

6.3 Suitable Servo controlled Stabilizer/CVT

Standards and Safety

7.1 Should be FDA or CE approved product

7.2 Electrical safety conforms to standards for electrical safety IEC-60601 / IS-13450

7.3 Manufacturer should have ISO certification for quality standards.

7.4 All electrical actuators and mechanisms should be housed inside the structure

making the product safer

7.5 Comprehensive warranty for 2 years and 5 years AMC after warranty

8. Documentation

8.1 User manual in English

8.2 Service manual in English

8.3 List of important spare parts and accessories with their part number and costing.

8.4 Certificate of Calibration and inspection from the factory

8.5 Log book with instructions for daily, weekly, monthly and quarterly maintenance

checklist. The job description of the hospital technician and company service engineer

should be clearly spelt out.

8.6 List of Equipments available for providing calibration and routine maintenance

support as per manufacturer documentation in service / technical manual.

8.7 Original Information Brochure should be provided.

Item No 52

Blood Collection Monitor

• Weighing range 100—999ml

• Automatic tare to zero for the bag weight.

• Adjustable low and high flow alarms.

• Adjustable donation time out up to 20 minutes.

• Adjustable default volume.

• Automatic clamp of tubing at the end of the

donation.

• Weighing accuracy +/- 2%.

• Power supply 115/230 VAC 50/60Hz

• Power consumption Max 10VA

• Dimensions - 290(L) X 253(W)X 150(H)mm

• Weigh: about approx 5kg incl battery

· Should have a data memory of approx 30000 characters

· Provision to attach bar code reader(optional) for capturing external data related to donor

· Internal fuses to be PTC-self recovery to ensure continuous operations

· Automatic Calibration Feature

Item No 53

Blood Bank Refrigerator

General

Should operate at 4 ° C with +/- 1 °C temperature uniformity

Must be designed for blood bank use. Commercial or modified commercial refrigerators are not

acceptable

Page 91: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 91 of 166 01.09.2010

Should be able to pass through standard door heights of 201 cm (79”) with casters without requiring

the unit to be tipped or laid on its side.

Should be CE marked.

Construction

Should have an interior and exterior that is constructed of minimum 20 gauge, galvanized steel.

Should incorporate)bacteria resistant, powder coated interior, exterior and door handle.

Should have a minimum non-CFC urethane insulation

Should utilize a self-closing door with full-length handle, key lock and non-CFC urethane insulation.

Should include swivel-locking casters as a standard feature.

Should incorporate a recessed interior floor to contain spills.

Should have a chamber access port in the top of the unit.

Should have an interior fluorescent light with control panel mounted switch as a standard feature.

Should have a light bulb that can be changed without removing the drawers.

Should have dual-pane, glass door and key lock with a right hand hinge.

Should have self-closing door system

Independent Temperature Controller

Should utilize an independent, microprocessor temperature controller that is programmable from +2o

C to +8o C.

Should have a Microprocessor controlled temperature readout, readable in 0.1o C increments.

Should have a stainless steel, RTD temperature probe that is located in the chamber.

Should have all functions accessible through a touch pad on the control panel.

Should have refrigeration system “On” indicator provided as a standard feature.

Independent Alarm / Monitor System

Should be able to program the high and low temperature alarms.

Should have audible and visual high and low temperature alarms as a standard feature.

Should have a stainless steel RTD temperature probe located in the top portion of the chamber in a

product simulation bottle.

Should have audible and visual door ajar alarm as a standard feature.Must have all functions

accessible through a touch pad on the control panel. Should have an alarm silence button.

Should have alarm disable switch. Should have remote alarm contacts as a standard feature.

Should have battery backup with a minimum of 2 hours life.

Should have a power fail alarm as a standard feature.

1.5 Temperature Recorder

Must have four inch, 7-day, ink-less, pressure-sensitive circular chart recorder.

Must have chart recorder temperature range of -5oC to +20oC.

Must incorporate a separate battery backup to ensure continuous operation of the chart recorder

during power failure.

Must have temperature recorder probe that is independent from other probes.

Must have power status indicator.

Must have an optional deduction of the chart recorder.

. Should have LCD Temprature Graphs that should display 24 continuous hours of data and

event logging of door openings and alarm conditions.

. Should have adjustable alarm volumes and password protected configurations.

Refrigeration System

Must incorporate a heavy-duty, air-cooled refrigeration system designed to operate on 230 volt

Page 92: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 92 of 166 01.09.2010

50/60 Hz.

Must utilize non-CFC, commercially available refrigerant.

Must have an automatic condense evaporator as a standard feature.

Must have an internal evaporator fan that shuts off when the door is opened.

Must have a compressor that can maintain required chamber temperatures when operating between

200-240 volts and 50 Hz.

Must incorporate a defrost system that requires no defrost timer, electric heaters or defrost down

time.

Must keep the refrigerator free of frost without elevating the chamber temperature.

Drawers

Must have solid bottom and liquid tight stainless steel drawers for containment of spills

Must incorporate Scratch-Guard drawer edge protectors that keep the glass from being scratched.

Must have fully extendable drawer slides.

Must have shelf standards with a clear powder-coated finish to guard against rust and corrosion.

Must have drawers that are adjustable

Must have optional drawer dividers available as an accessory.

Must have a cabinet designed to accommodate available optional half-size wire shelves, full-size

wire shelves and rollout wire baskets with no cabinet modifications.

Electrical

External transformers are not acceptable

Item No 54

Tube Sealer

1. Should be a hand held sealer for apheresis, Stem cell, leucoreduction processeas and should

have anywhere mobility for multiple application.

2. Should be supplied with one power source, hand held sealing head and one NICD

rechargeable battery pack.

3. Should do 1000 seals per fully charged battery pack and battery should fully be charged

within two hours.

4. Should be a smart sealer to adjust for different sized tubing.

5. Tear seal feature to make segments that can be separated by hand.

6. Should be certified for patient connected use.

7. Lightweight and compact for ease of mobility. Total weight approx.3kgs.

8. Should have PTC overload protection internal fuses to ensure continuous operations

Item No 55

Cell Counter - 3 Parameter

Fully automated Hematology Cell Counter 18 Parameters with Histogram for 3parameters, should have highest linearity of PLT more than 1.8million/micro litter to measure PLT Values in Platelet Reach Plasma (PRP) :-

INSTRUMENT SHOULD HAVE THE FOLLOWING ESSENTIAL FEATURES

Page 93: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 93 of 166 01.09.2010

TECHNOLOGY

: Electronic Impedance Principle for counting and HGB by photometer.

PARAMETERS

: WBC with LYM%, LYM#, MON%, MON#, GRA%, GRA#, RBC, HGB, HCT,

MCV, MCH, MCHC, RDW, PLT, MPV, PDW & PCT.

HISTOGRAM

: RBC, WBC, PLT DISPLAY AND PRINT-OUT.

CALIBRATION /QC.

: L-J graph for quality control mode with possibility to perform

automatic & manual calibration .

MESURING SYSTEM

: Dual channel measurement.

VALVES

: Minimum number of Maintenance Free Valve.

FLUIDICS

: Shortest fluidics to minimize maintenance and contamination.

APPERTURE CLEANING

: Automatic Back Flush in case of clog.

HGB MEASUREMENT

: By mean of LED with automatic blanking.

SAMPLE VOLUME/MODE

: Less than 20MCL in whole blood mode.

SAMPLING NEEDLE

: With automatic wash from inside and outside.

THRU-PUT

: 60Sample/Hour.

DISPLAY/OPERATION

: LCD Graphic Display with Touch Screen or Keypad operation.

MEMORY

: AT LEAST LAST 1000 Result memory with Histogram.

COMMUNICATION PORT

: RS-232C Serial Port or USB Port

PRINTER

: Built-in or External Printer.

REAGENT CONSUMPTION

: Technical Bid must show each Reagents consumed in ML as:-

Start-Up, Shut Down: Diluent -? Lyse -? In ML, If any other reagent needed to be mentioned.

Per Cycle: Reagent Consumed in ML such as Diluent-?

,Lyse-? Or any other reagent has to be mentioned.

Page 94: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 94 of 166 01.09.2010

Instrument must be supplied with suitable UPS & Start up reagents for at least 1000Samples.

Item No 56

Refrigerated Centrifuge

1 For separation of blood components like packed cells, platelet rich plasma, platelet concentrate,

Cryprecipitate & Buffy Coat

2 Programmable memories with tamper proof facility

3 Rotor head with swing out buckets, windshielded.

4 Oval Shaped metal buckets with removable plastic oval cups to hold double/triple/ quadruple

blood bags

5 Refrigerated centrifuge with suitable rotor head & buckets should accommodate/process

minimum 8 blood bags of 450 ml capacity simultaneously

6 Manufacturer to indicate volume & capacity of plastic bucket

7 Centrifugal Force -- Max ceiling 5000g

8 Microprocessor Controlled rotor speed to with in 10 rpm of set value

9 Different Acceleration & Deceleration profiles should be available

10 Temperature range 0 deg C to + 40 deg C, Microprocessor controlled rotor temperature with

in +/- 1deg C regardless of centrifuge speed

11 Programmable time should be from 0-99 minutes with minimum revolution of 1 minute.

12 Digital display of temperature, speed & time & other relevant parameters

13 Automatic shut down of centrifuge , if rotor load is out of balance with indicator

14 Safety Key lock to prevent unauthorized use

15 The equipment should have lockable castors

16 The Refrigerated centrifuge should have built in memory to store atleast 50 data of

centrifugation during the processing of Blood Components. The data from the centrifuge can

be transferred to a computer using a RS 232 data interface

17 Power Supply Requirement :- 220-240 Volts 50 Hz Single Phase

18 Servo Controlled heavy duty line voltage corrector should form part of the standard

configuration

19 A line voltage corrector of appropriate rating giving all the specifications should be supplied

along with the unit

20 Should be a CE marked product & from a ISO , WHO-GMP compliant manufacturer.

Item No 57

Plasma Expressor(E)

Product certification & approval

CE mark and S mark as per EN 61010-1 standard

Ensures safety against electrical shock hazards, fire hazards, mechanical hazards,

electromagnetic interference etc

Manufacturing standard

Manufactured at ISO 9001:2000 Certified facility

Power source

AC - DC power adapter supplied along with the equipment or power from another

Page 95: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 95 of 166 01.09.2010

Electronic Plasma Expressor of same make

Input for AC-DC power adapter

120-240VAC, 50/60 Hz, 0.4A

Plasma Expressor input

12VDC, 1A

Power Consumption

12W (Max)

Dimensions

Approx.(WxDxH) in mm

165 x 230 x 280

Sensor type

Infrared beam interrupt opto-sensor

Bag pressing mechanism

Stainless steel spring loaded transparent acrylic sheet

Clamping power source

Spring loaded clamp with geared DC motor

Clamp closing time

Less than 1sec

Flow rate

180 - 210ml/minute

Controls

Power Switch, Clamp Up, Clamp Down, Start

Indication lamps

Power, Start , Clamp Down

Modes of operation

Manual & electronic modes

No. of electronic plasma connected using a single power supply

4 Nos

Interface

Provision for interfacing with CopmoScale for separating a definite quantity of plasma to

the satellite bag

Item No 58

Dielectric Tube Sealer Table Top

Bench top model.

1. Compatible with all blood bags.

2. No warm up time, automatic sealing when tube is in position.

3. Seal minimum 2 mm wide.

4. Easy to clean electrodes, easily accessible & protected by cover.

5. 220-240 volts.

6. Indication for sealing progress.

Item No 59

PLATELET INCUBATOR & AGITATOR

PLATELET INCUBATOR

PLATELET INCUBATOR SHOULD HAVE THE PROVISION TO STORE THE AGITATOR FOR 48

PLATELET BAGS AGITATOR.

SHOULD HAVE CLEAR VIEW SINGLE PANE TEMPERED GLASS

AGITATOR SHOULD STOP AUTOMATICALLY ONCE THE DOOR IS OPENED.

Page 96: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 96 of 166 01.09.2010

SHOULD HAVE MICROPROCESSED CONTROLLED LED DISPLAY,TEMPERATURE GRAPH

DISPLAY,

SHOULD HAVE STAINLESSSTEEL RTD SENSOR PROBES

SHOULD HAVE PROVISION FOR 4”7DAY INKLESS CHART RECORDER WITH BATTERY

BACKUP FOR CONTINEOUS OPERATION DURING POWER FAILURE.

SHOULD HAVE ALL CONTROLS IN ONE CONVENIENT LOCATION INCLUDING CHART

RECORDER AND ALARM KEY

SHOULD BE ABLE TO MAINTAIN A TEMPERATURE OF 22 DEGREES WITH +_. 1DEGREES

VARIATION.

PLATELET AGITATOR

SHOULD BE ABLE TO STORE MINIMUM 48 RANDOM PLATELET BAGS OR APHERESIS BAGS

OR BAGS OF DIFFERENT SIZES. With GENTLE SIDE TO SIDE MOTION (1 ½” 38MM)

SHOULD HAVE SINGLE FAN FOR FORCED AIR CIRCULATION.

SHOULD BE STURDY ONE PIECE DRAWERS WITH HOLES FOR COMPELETE AIR

CIRCULATION ACROSS BOTH SURFACES OF PALATELET BAGS

Should be CE marked

Item No 60

AUTOMATED PLASMA THAWING EQUIPMENT

1. SHOULD BE ABLE TO THAW 8 – 12 PLASMA BAGS

(FFP/APHERESISOR PLASMA BAGS OF ANY SIZE or ANY MAKE)

2. SHOULD HAVE WATER BATH BASED SYSTEM WHICH SHOULD BE OPERATIONAL AT

4 DEGREE TEMPERATURE TO 37 DEGREE CELCUIS PRECISELY.

3. SHOULD BE COMPACT IN SIZE.

4. SHOULD HAVE RACK HOLDERS WITH BUILT-IN FINGERS FOR SECURELY HOLDING

THE PLASMA BAGS OF ALL SIZES.

5. SHOULD HAVE AN ALARM WHEN THE PLASMA BAGS ARE THAWED

6. SHOULD HAVE THE PROVISION FOR SELECTING PROGRAMMED TIME SETTING FOR

THE LENGTH OF THAWING CYCLE. .

7. SHOULD HAVER DIGITAL TIMER CLEARLY DISPLAYINGTHE PROGRAMMED SET TIME

OR REMAINING CYCLE IN MINUTES

8. SHOULD HAVE ALARM SYSTEM FOR ADJUSTABLE OVERTEMPERATURE ALARM

SETTING, AUDIBLE AND VISUAL ALARM WARNINGS,

9. SHOULD HAVE A DEEP THAWING CHAMBER FOR INCREASED HEAT TRANSFER

EFFICIENCIES, WHICH RESULTS IN FASTER FFP THAWING TIMES. THE CLEAN

STREAMLINED DESIGN OF THE HEAVY GAUGE STAINLESS STEEL CHAMBER SIMPLIFIES

ROUTINE CLEANING.

10. SHOULD HAVE A CHAMBER DRAIN SYSTEM WITH A HIGH FLOW RATE TO DRAIN THE

CHAMBER WITHIN 2-3MINUTS.

11. SHOULD HAVE A TERMPRATURE CONTROLLER

12. SHOULD HAVE A SERVO CONTROLLED VOLTAGE STABILIZER OF ATLEAST

Page 97: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 97 of 166 01.09.2010

3KVA 13. Should be CE approved

Item No 61

LAMINAR FLOW- VERTICAL

Hepa Filter : 99.999 %efficiency for particles >0.3 µm

Pre-Filter : 85 %efficiency for particles >0.5 µm

Particle Count : Better than US Fed Std 209B Class10 and VDI 2083 Class 3

Cabinet : Laminated High Quality Wooden Board

Work Table : AISI 304 Stainless Steel

Airflow Speed Control : Speed Controller (Three Step Speed Controller)

Blower : High efficient centrifugal type with lifetime lubricated bearings

Light : High intensity,low wattage >800 lux

Noise Level : <55 dBA

Standard Accessories : Air/gas cock and .mains power socket (16A)

Power Supply : 220-230 V,50 Hz.

Power Consumption : 400 w

Internal Work Space: 600mmx600mmx600mm

900mmx600mmx600mm

Item No 62

Deep Freezer -40 Deg.

VOLTAGE: 220V/50Hz Capacity 310 Litres

- TEMP: -20 TO -40degC, MICROPROCESSOR CONTROL

- ACCESS PORT

- HEAVY-DUTY SWIVEL CASTERS

- HEAVY-GAUGE STEEL CABINET WITH LONG-LASTING POWDER PAINT FINISH

- SINGLE DOOR +KEYLOCK, DIGITAL DISPLAY

- REFRIGERATING FLUID & INSULATION: CFC FREE.

- ROUNDED INTERIOR CORNERS

- EYE-LEVEL CONTROLS

- TEMP: ALARM

- MICROPROCESSOR ALARM

- BATTERY BACK-UP

- AUDIBLE/VISUAL OVERTEMPERATURE ALARM

- LOW VOLTAGE BOOSTER

- AIR INSULATED INNER DOORS

- 3 ADJUSTABLE Shelves

- BASE MOUNTED CONTROLS

- MULTI-POINT GASKET SEALS

- AUTOMATIC VOLTAGE BOOSTER

- SET POINT SECURITY SYSTEM

- INDEPENDENT OPERATING TEMPERATURE AND HIGH/LOW LIMIT ALARM

- AUTOMATIC VOLTAGE BOOST

- ON-BOARD MONITORING

- AIR-COOLED CASCADE REFRIGERATION SYSTEM

- EFFICIENT DOWNFEED EVAPORATOR

- HIGH CAPACITY AIR-COOLED CONDENSER

Page 98: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 98 of 166 01.09.2010

- HEATED DOOR SEALS (HOT GAS) MINIMIZES FROST BUILD-UP AROUND DOOR GASKET

- DOOR LATCH CAM ACTION WITH KEY LOCK, ONE HAND OPERATION

- DOOR HANDLE FULL LENGTH FOR EASY ACCESS

- CHART RECORDER 7 DAY 6 ”

- SERVO CONTROLLED VOLTAGE STABILIZER OF 3 KVA SHOULD BE INCLUDED

Item No 63

Deep Freezer -80 Deg.

The Deep Freezer should achieve low temperature of -80degC

Should be heavy duty refrigeration system, maintenance free, with hermetically sealed

refrigeration

compressors reliable refrigeration with minimum noise & vibration.

Integrated digital temperature cum controller with inbuilt 7 days circular chart recorder

with battery

back up of 3-4 hours to ensure display & recording of temperature even during power

failures.

Construction of double wall with efficient insulation to minimize temerature loss, inner

chamber should

be made of AISI 304 grade non corrosive stainless steel & outer made of high quality

C/R sheet

Servo controlled line voltage corrector of 5 KVAshould form part of the standard

configuration.

Should be a CE marked product & from a ISO, WHO-GMP compliant manufacturer.

The Deep Freezer should be vertical upright with a capacity to hold minimum 185-200

plasma bags

Item No 64

Cell Separator

Should be Fully automatic, Microprocessor controlled with easy access operator control panel and

has a large touch screen.

Should Perform both Single and Double needle Apheresis.

During single needle procedure the equipment shpuld continue to process the whole blood during

the return cycle to reduce the procedure time and increase the efficiency.

Should have auto elutriation separation technique to be able to collect platelets faster with high

collection efficiency to help process less blood volume.

The disposable kit for the machine should be True Closed system disposable with pre-attached

factory fitted ACD, Normal Saline and Needles as required.

Equipment should ensure all donor safety parameters before starting the procedure and all time

during operation.

Should be Capable of collecting various single donor blood components including Peripheral Blood

Stem Cells (MNC). Should be able to collect both single and double needle Platelet apheresis along

with concurrent Plasma and/or RBC

Should be capable of fully automatic PBSC (MNC) collection .. Should be Capable of

Page 99: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 99 of 166 01.09.2010

doing Prime only with Normal Saline and / or mixture of Normal Saline and ACD.

Should have a Inbuilt Cuff pressure and prompt grip for donor comfort and adequate blood flow.

Should have a Facility to use platelet additive solution and / or normal saline for re-suspension and

storage fluid in place of plasma.

Machine should have a Advance help menu available at any time during alarm conditions.

Extra corporeal volume should not be more than, 21 Oml & 205ml in case of both single and double

needle apheresis respectively ..

Should be equipped with a Yield estimator to help decide yield, volume to be processed and

suggested storage fluid and should have a optical sensor at PRP line for online monitoring of

component collection against the desired yield.

Should be Capable of downloading or printing full procedure report any time after procedure

Should be Capable to connect bar code reader if desired.

Should have rechargeable battery to store data and restart in case of power failure.

Should have Continuous monitoring of collection to avoid any contaminations through Interface

detector.

Inlet and return flow rates should be up to 100 ml./min.

The separation of blood in the machine should be able to automatically maintain a constant

hematocrit to improve collection efficiency and reduce contamination

Item No 65

ICU Ventilator (High End)

1. Latest state of art microprocessor technology based on the Biphasic Ventilation concept. 2. Can provide respiratory support to Adults, Pediatric & Infant patients with following modes, CMV (A/C), SIMV, PCV, PCMV, PSIMV, SPONT, NIV, ASV, APRV, BIPAP, APVcmv, APVsimv etc. 3. Have fully Automatic Mode ASV (Adaptive support Ventilation) effective for Passive as well for active patients 4. Have Pulmonary Function Assessments (P-V Tools) provision to have idea for optimum PEEP. 5. Have Special functions like TRC (Tube Resistance compensation), Manual Breath, Stand by & Sigh etc. 6. Can be upgraded for ETCO2 Monitoring. 7. Can be upgraded for Heliox Therapy Option. 8. Have in- built color TFT Touch screen size 15 inches for the display of the various lung parameters, curves (Volume vs time, flow vs time, pressure vs time) & Loops (P-V, F-V, P-F). DISPLAY: display Real time visualizations of Lungs with representations of Tidal Volume, Lung Compliance, resistance and patient activity along with Ventilator dependency grouped into oxygenation, CO2 elimination, Simultaneous display of at least 4 waveforms or 2 loops based on volume, flow, airway pressure, auxiliary pressure or CO2. 9. Continuously monitors 37 parameters including Rapid shallow index (RSBI), Occlusion pressure (P01), Static compliance, leakage, RC exp etc.

Page 100: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 100 of 166 01.09.2010

10. Have the following user selectable parameters: Tidal Volume: 10-2000 ml I:E Ratio : 1:9 to 4:1, Automatic Base flow: 4-30 L / min. P ramp: 25-200 ms Flow: 0-180 L/min. Apnea time: 15-60 s Expiratory Trigger Sensitivity: 5-70% of insp. Peak flow Rate: 1-120 b/min Flow Patterns: Sine, Square, 100% decelerating, 50 decelerating. 11. Have all the alarm setting for the various parameters with Altitude compensation up to 3000 m. (9843 ft). 12. Supplied with the reusable autoclavable patient circuits for Adult, pediatric & Infant patients separately.

13. Supplied with inbuilt Nebulizer & Blender and have internal battery backup of 1 hours.

14. Have the 1, 12 & 24 hours trends for the all the monitored parameters.

15. Have Storage of 1000 events with the time stamp.

16. Have upgradation for Interface connecting facility like USB, Compact Flash for screen shots & DVI with VGA Output.

Optional Accessories:

External Compressor

A) Non Invasive Pressure Support Ventilation:

1. Spontaneous / CP AP mode.

2. IPAP Range: 2-25 cm H2o with increment of 0.2 cm.

3. EPAP Range: 2-25 cm H2o with increment of 0.2 cm.

4. Reusable Full face masks.

5. Variable rise time available for weaning away patient from invasive Ventilator.

6. Respiratory Back-up upto 10 BPM.

7. Adjustable Trigger and Cycle sensitivity - High, medium and low sensitivity setting.

8. Vsync automatically minimizes any loss of pressure delivery due to leak.

B) Diagnostic Capabilities:

1. Display Respiratory Rate on breath by breath basis.

2. Display Tidal Volumes.

3. Displays Masks or Circuit Leaks.

4. Displays Delivered Pressure.

5. Alarms for Mask off.

C) Patient Safety Mechanics:

1. Inspiratary time can be setting allowing critical patient to breathe out.

2. IPAP Max Range (0.1-4 sec)

3. IPAP min available to force patient to stay at IPAP pressure for user defined time interval. IPAP Min Range (0.1-IPAP Max)

4. The Rise Time from EPAP to IPAP can be set and varied (Min, 150-900ms).

5. The world’s most comfortable Full Face masks available.

6. Ramp Time available for 45 minutes.

7. 365 nights patient compliance data can be recorded.

Page 101: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 101 of 166 01.09.2010

8. Hour meter to display total number of hours used.

9. Humidification option available, Humidifier optional-Passover or Heated.

10. Automatic smart start ON/OFF; Patient Triggered or manual.

D) Machine Fault Protection and other Mechanics:

1. SMPS for voltage regulation 95-295V built-in.

2. Power Supply Input Range 100-240V, 50-60Hz, <100 VA.

3. Carrying case included.

4. Alarms available

Item No 66

ABG Analyzer

Fully Automatic Microprocessor Controlled Latest Blood Gas Analyser with all the Necessary accessories having following features

Measured Parameters

pH, pCO2, pO2, Barometric Pressure

Na+, K+, CI-, Ca2+

Hematocrit (Hct)

tHb and SaO2

Calculated Parameters

Up to 30 calculated parameters can be programmed

pH at patient temp

pCO2 at patient temp

pO2 at patient temp

Base Excess (BEecf)

BE at actual SaO2 (BEact)

Buffer base (BB)

Actual bicarbonate (HCO3)

Standard Bicarbonate (ST HCO3)

Total CO2 (ctCO2-B & P)

Standard pH (ST pH)

Oxygen Content (cO2)

Hydrogen lon Concentration (cH+ )

AaDO2

Respiratory Indes (RI)

P50

Anion Gap (AG)

FIO2

Qs / Qt

Sample throughput - 30 / hour

Fully automatic Liquid calibration of all parameters at fixes/user defined intervals. No calibration gases required

Miniature type maintenance free electrodes

Individual switch off / on facility of whole module or parameter

Data display on built in flat color 6” LCD display screen (touch screen)

Data printout on fast low noise thermal printer with graphics capability

Page 102: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 102 of 166 01.09.2010

Storage of data on built in Microprocessor

Optional PCMCIA Card for data Storage

Built in Barcode-by facility for economical operation

Low consumption of reagents during the operation

Interfaces 2´Rs 232, 1´Ethemet Connection, 1´Barcode Scanner

Built in voltage stabilizer for the voltage range from 100-240 VAC/50Hz Only 3 types of reagents Required for calibration & Measurement. Reagent Onboard Shelf life should be minimum 45 days

Should be FDA Approved

Activating & deactivating parameters by only key-stroke

Item No 67

Arthroscopy System

Camera system

1. Camera console 220 v three chip with universal coupler & autoclavable camera head

2. Pure digital signal with high definition video(1280x1024 native resolution) 3. Resolution-2000 horizontal lines

4. 8 specialty settings

5. Integrated flexible scope filter

6. Signal to noise ratio-70 db

7. Progressive scan technology both on camera head & console

8. Brightness control on console & camera head

9. Aperture control on console

10. Inbuilt 16 step digital image enhancer on console

11. Digital zoom & white balance on camera head

12. Integrated gain/shutter/enhancement with brightness control

13. Two peripheral control on camera head

Xenon light source 300 watts

Light source xenon, 300 watts lamp,color temperature 6000 k, universal jaw for accepting any make

fiber optic cable adjustable light intensity from 0 to 100 percent, Standby mode; Fiber optic cable

High definition monitor 21"

High definition monitor, screen 21”, resolution 1280 x 1024 dvi input.

Option for wall mounting and desktop in same unit.

Arthroscope set

Telescope 4mm, 30 deg

quick latch / speed lock dual connect arthroscope sheath 5.9mm.

Obturator for sheath

hook probe

straight punch, 3.4mm

15 deg upbiter 3.4mm

3.4mm 30 deg right cutting punch

3.4mm 90 deg left cutting punch

3.4mm 90 deg right cutting punch

3.4mm foreign body grasper with lock

Shaver system

Page 103: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 103 of 166 01.09.2010

Electronic control unit

Foot switch

Connecting cable for large bone handpiece

Shaver Handpiece autoclavable rpm 12000

Microdriver with drilling, wiring, micro saw and reciprocating saw attachment

Heavy Duty (Drill & Reamer) and Sagittal Saw handpieces for large bone (Arthroplasty)

Jacob's chuck,reamer (for IM & Acetabular reaming) and pin driver attachment

Aggressive cutter (6 nos.)

Full radious resector ( 6 nos.)

Oval bur ( 6 nos.)

All the items should be from one company

The equipment should be U.S. FDA approved

Imaging movable trolley – Indian

Item No 68

Color Doppler Ultrasound Unit

The system should be state of the art with full digital beam former technology and should be for whole body application which would include Cardiac, Obs/Gyn, abdominal, peripheral vascular. Small parts Imaging such as Ophthalmic Imaging, Thyroid, Intracavity application, etc.

Essential features :

· The system should have 256 Greyscale or more.

· The system should have a dynamic range of 150dB or more.

· System should have advanced image processing algorithms to reduce the speckle and artifacts for improved image quality.

· Should have Integrated Flat Panel Display with min. 17” screen size with tilt and swivel facility

General requirements :

1. The system should incorporate facility for high facility for high resolution 2D,M mode, PW, color flow imaging, power Doppler angio imaging modes.

2. All transducers should be multi frequency broad band technology for High resolution 2D and Colour Doppler imaging.

3. Zoom should be both read and write type up to 30 times

4. Imaging Depth should be atleast 32 cms or more.

5. Presets : Minimum 50

6. The system should have harmonic imaging for tissues for hard to image patients, All transducer should have Tissue Harmonic Imaging as standard

7. The system should have a full alphanumeric keyboard with illuminated keys and status display.

8. System should have cine loop review facility in individual and mixed modes with memory upto minimum 300 images and 10 seconds of Doppler / M Mode Cine

9. Should have facility of Thumbnail Menu, Support thumbnail images and digital storage and retrieval of B/W & color image data (both frozen and cine loop) on built-in and removable media (Min. 40GH HDD, DVD, CD-R/W

10. The system should have automatic real time quantification of Doppler parameters like velocity, frequency, time heart rate slope, flow volume, pulsatility index, jpeak velocity, average value, point value, area and diameter flow volume etc.

Page 104: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 104 of 166 01.09.2010

11. Directional Power Doppler angio imaging for perfusion studies should be available for visualization of flow in small vessels.

12. The system should have extensive calculation software package for generic measurement, cardiac, Ob/Gyn, vascular etc.

13. Should have direct connectivity to inkjet colour printer for printing images and report.

14. Equipment with above features to be offered with the following broad bandwidth probes

a. Broad band convex array transducer with frequency range 2-5MHz

b. Broad band linear array probe with frequency range between 2-4 MHz

c. Broad band phased array probe with frequency range between 2-4 MHz

d. Broad band TV/TR Probe with frequency Range between 5-8 MHz, (Quote optionally)

15. The system should have the following documentation devices and accessories:

a. B&W Thermal Printing, Color Inkjet Printer and suitable UPS to be provided

Item No 69

Video Bronchoscopy System

(A) THERAPEUTIC VIDEO BRONCHOSCOPY SYSTEM SHOULD HAVE FOLLOWING ITEMS:

1. Video Bronchoscope

2. Light Source (Xenon)

3. Video Processor

4. Monitor

5. Bronchical Accessories

Scope Features:

1. Remote Switche on the scope

2. Biopsy port below the control section

3. Light weight

Video Broncohscope Should:

Have latest CCD chip Technology

Be of high resolution with maximum Pixels

Be compatible with electrosurgical instruments & Laser with ceramic tip

Have grip fit design for single hand operation

Have autoclavable suction valves for reduction of crass infection possibility

Be fully immersible

Video Broncohscope specification:

CCD Type: Colour chip

Field of view: 115 - 125 Deg

Depth of view: 3-100 mm or better

Distal end: Outer diameter: 5.8 mm - 5.9 mm

Insertion tube: Outer diameter: 5.8 mm -6 mm

Bending section: Angulation range: Up 180 Deg, Down 130 Deg

Working length: 590 mm - 600 mm

Total length: 860 mm - 900 mm

Instrument channel: Inner diameter: 2.8 mm or more

(B) LIGHT SOURCE

1. Should be independent / integrated with processor

Page 105: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 105 of 166 01.09.2010

2. Illumination: Main Lamp-xenon, Emergency lamp as a stand by lamp with automatic switch over Lamp Life indicator

(C) VIDEO PROCESSOR:

Should be a latest clour chip system with following features:

Should be independent / Integrated from light source

Should be light weight

Should be capable of storing at least 40 patient’s data

Compatible with different scopes including paediatric for future upgradtion

Video output: RGB, B NC and Y/C

(D) MONITOR: Colour Monitor grade (14”) compatible with this system

(E) Computer / software: Compatible computer system with recording and printing facility for providing reports patient / storing for future intervention / reference

(F) Endoscopic trolley with good caster wheels with breaks, hanger to hang scope

(G) Bronchical Accessories: Compatible with video Bronchoscope

Biopsy Forcep - 2 No.

Biopsy forcep with needle - 2 No.

TBNA Needle - 5 No.

Cytoloty - 5 No.

Foreign Body Rat tooth - 2 No.

Foreign Body Basket - 2 No.

Tamponade Balloon - 1 No.

Injector - 2 No.

Item No 70

High Frequency Mobile X- Ray Unit

X-RAY GENERATOR:

The generator should be High Frequency with output of more than 3 kw.

It should work on constant 40 khz frequency output.

The x-ray generator should have LCD display of KV, mA and mAs.

The radiograph KV should be from 40 ti 100 KV.

The radiograph mA should be 70.

The esxposure time should be in mAs from 0.2 to 320mAs.

The generator should have be stationary anode with focal spot of 1.5mm.

It should operate on single phase 230Volts.

It should be user friendly with facility to select factors based on body anatomical chart. It should be compatible with Digital X-ray system of CR system.

It should have facility to store the X- ray factors as per the use into the memory.

STAND:

It should be ergonomically designed spring balanced stand.

It should have hand brake for locking the over all machine.

It should have easy movement or ramp, lift and form corridor.

It should have parking position of tube arm in extreme lower position.

Page 106: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 106 of 166 01.09.2010

Item No 71

Digital Mamography System Description of function 1.1 Mammography system to replace conventional Film/Screen based Mammography Studies with digital imag 2.1 Full Digital Mammography System consisting of exposure stand with attached swivel system, separate console with radiation shield, automatic exposure control and mammography X-Ray Tube. 2.2 An integrated direct-to-digital Flat Detector based on amaorphous silicon technology. 2.3 A separate workstation for image positioning and patient demographic data is required. 2.4 The workstation should be able to send, receive and print according to DICOM standards. 2.5 The workstation should also be able to obtain DICOM modality, work list from connected information system and send information about performed procedure to the connected information system 2.6 Read and Write in CD/DVD for data Storage and review. Technical Specifications 3.1 Mammography System 01 The system should consists of a tube head and detector assembly that has isocentric rotation for every positioning. The iso-centric movements should be motorized. The patient Compresion device should have automatic multi-speed variable compression system which senses the breast density and adjust the compression force. The maximum compression thickness should be 18 cm or more. The patient table should have motorised grid movement. Magnification devices of ratio 1.5 and 1.8 should offered. Digital display of compression force and compression thickness should be available. 3.2 X-Ray Generator and Tube The X-ray generator should be high frequency with the following parameters: kV range: at least 25-35 kV in steps of 1 kV mAS range: 0-750 mAS or more Exposure time: 0-700ms Maximum mA: 180mA or more X-Ray tube unit: Dual focus rotating anode tube with the following parameters: focal spot size: 0.1mm and 0.3mm Anode heat storage: 150 kHU or more Tube heat Storage: 1.3MHU or more Anode material: Molybdenum and Tungsten Please mention the filter material used in the tube 3.3 Flat Panel Detector: Type of detector: Amorphour selenium preferred Detector size: 24cmx29cm or more with two image formats Pixel size: 70i or less Image matrix in pixels: large size-3Kx 4K or more Small size: 2Kx 3K or more 3.4 Workstation for image Acquisition: The workstation should enable immediate image display for general survey for patient positioning. It should be able to store around 10000 images. The networking should on TCPIP protocol. The following image processing should be possible on the workstation: Image display:

Page 107: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 107 of 166 01.09.2010

Freely selectable screen layout Windows settings (contrast and brightness setting) Magnification, stepped and dynamic zoom Image inversion (black/white) Annotation: Left/right marking Text additions Lines Rectangles and circles Measurements: Distance Angle Density Image evaluation: contrast enhancement(with table) Display of histogram Length measurements Before /after comparison Filter

Administration: The demographic patient data should be retrieved directly from a HIS/RIS system The demographic patient data can be entered manually Retrieval of images from CD, DVD or PACS - Printing of images on DICOM – compatible printers The workstation should be fully DICOM compatible High Contrast 1Kx 1K TFT monitor should be provided with workstation.

Item No 72

Holter Recorders and Analyzers 1 It should have FDA clearance and CE certification 2 It should meet and exceed the requirements of ANSI / AAMI EC 38 Holter Recorders 1 It should be small and lightweight recorder 2 It should be capable of simultaneous real time acquisition of 3 channel and 12 channel recording. 3 Recorder should have LCD display to preview ECG waveforms during the patient hook up and have lead quality check function. 4 it should have capability of continuous recording for 24 hours and 48 hours / 7 days. 5 It should have at least 24 hours with single battery. 6 The recorder should be water resistant. 7 Recorder should be battery operated and have single AA/AAA battery or rechargeable in built sealed battery with recharging unit supplied. 8 It should have internal memory for 99 full disclosure readings. 9 it should have advanced signal processing algorithms to provide superior accuracy in beat detection, labeling and noise rejection. 10 It should be a 12 channel analyzing system. 11 it should have capacity to download 24 hours data through USB port / standard connection port in less than 3 minutes. 12 Device should be defibrillator protected.

Page 108: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 108 of 166 01.09.2010

13 Recorder should have 1000s/sec/channel digital sampling rate for standard recording and internal storage. 14 Recorder should have compact flash card memory card. 15 it should be capable of pacemaker spike detection. 16 it is preferable if it supports voice recording capability. 17 It should include 2 sets of electrodes and patient cable to enable 3 and 12 channel recording. 18 It should come in 2 sets of patient hook up pouch and hook up accessories. 19 Should include compact flash card as required for each recorder with adequate capacity to store entire recording cycle. 20 It should include cable for connecting and downloading data to PC. 21 It should include compatible flash card reader. 22 It should include users guide and technical manual as well as quick guide and patient hook up instruction posters for easy reference and use. 23 It should include 100 sets of patient event recording diary and pens. 24 It should have two years warranty on recorder / memory card and 6 months for cable. Analysis software + Hardware. 1 It should have multiple scanning options like retrospective, prospective and superimposition scanning modes, event and template review and be customizable. 2 It should have facility to convert final report into PDF and XML format that enables connectivity to data management systems. 3 It should have trend graphs for HR, RR interval, RR variance, 12 lead ST, SVPB, VPB etc and provide graphic display options. 4 It should have different beat classifications. 5 It should have various rhythm analysis and ST measurements. 6 It should have 3 and 12 lead ST segment measurement analysis. 7 It should be capable of Pacemaker analysis including atrial, ventricular and dual chamber pacing and under / over sensing and capture failure. 8 It should support customizable report format including patient data, 24 hour profile, selective printing of rhythm strips and trends, summary statistics in tabular and narrative format and support user defined acronyms for comments. 9 The PC + Printer provided should be Intel Pentium IV 2.5 GHz and above and include minimum hard disk of 250 GB / DVD+RW drive or Blue Ray disc writer / USB port/4GB RAM/10"TFT color monitor, Laser printer. Should have one year on site warranty. 10 It should have capacity to archive into hard disk and DVD+RW / High density DVD / Blue Ray disk. 11 It should include licensed operating software for ready to use status along with reliable anti virus software. These should have back up provided on CD / DVD. 12 It should include external data back up / archiving facility and hardware. 13 All trainer manuals should be supplied and on site installation to complete satisfactions of end user. 14 Training of 2 technicians to operate and use the equipment and software to be provided at the company expense. 15 Assurance of 95% uptime and prompt provision of service back up for hardware and software with replacement to maintain operational equipment as required.

Page 109: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 109 of 166 01.09.2010

Item No 73

Stress Test System(TMT) Stress testing system should be complete with PC, Software, Tread Mill, Patient acqusition module and necessary cables. System should be based on windows platform with 17" color monitor having minimum resolution of 1280 x 1024, 80 GB HDD, CD-RW, Mouse, UPS for analyser. System should acquire and analyze 12 leads and should store the full 12 lead resting / stress ECG Should show a pictogram and indicate which electrode has a bad contact before resting / Stress recording. Should be able to review retrospective average beat and ST values. Should have user settable protocols. Patient module should have USB connectivity. System should provide standard Full interpretation of Supine ECG with reasoning. Should provide display of real time 12 lead diagnostic quality ECG waveform, average complexes beat of all 12 leads with superimposed comparison along with ST level & slope and ST trend graph. Automatic detecton, display, storage and review of arrhythmia, heart rate. Should have running trends of ST available on screen during the test process. Should have ability of comprehensive auto-measurement package including RR intervals, P-wave duration, PQ interval and QRS width. System should have high quality filters for muscle and baseline noise without influence on the ST results. Should have alarm levels for ST, HR and Blood Pressure. Should have keyboard short cuts. Capability to pause ECG screen view to find a past ECG event without loosing track of current real time ECG. Capability to insert event marks during stress. Capability to change ST points during stress. Should have manual measurement cursors for P-wave, QRS and T-wave. Should allow to measure width (P, QRS, T-wave) and intervals (PR, QT etc.,) Should have alarm levels for ST, HR and Blood Pressure. Should be able to export QRS intervals from 12 lead Rhythm ECG, Stress / Resting ECG report in word / RTF format and shluld be able to export raw ECG data. Printing: Should be able to print strips during stress or after. Strips to be printed in 3 channel,s, 2x6 channels, 12 channels. Print reports should be with summary table, event markers and stage information. Should have average beat report with ST measurements. Should have Trend graphs System should provide multiple and customizable printing formats as per users choice on A4 size normal plain printing paper. Heavy duty treadmill - imported. Noise free with speed ranging from 0.8 to 19 kph and have an elevation range of 0 to 25%. Treadmill should provide smooth and safe operation. Should have automatic belt alignment. Treadmill should have 175 kgs weight(approx.) capacity with all metal chasis. Should have emergency stop button. Other Technical specs: ECG sampling rate: 1000 Hz ECG input voltage: 16mm Vpp Noise voltage: 20uV

Page 110: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 110 of 166 01.09.2010

10 lead patient cable Treadmill interface Sugnal frequency range: 0.05Hz to 250 Hz. Windows operating system. PC Pentium latest workstation. All consummables required for installation and standardisation of system to be part of the system. Should have US FDA and CE certifications Display of following parameters in all stages of exercise & recovery:- 1. METs 2. Heart Rate 3. Target Heart Rate with % of completion Heart Rate 4. ST measurement mode (Manual/Auto) 5. Current ST level & ST slope 6. Exercise protocol stage name 7. "Switch-off" facility for patient 8. Automatic BP monitoring system

Item No 74

Volumetric Infusion Pump Flow rate should be from 0.1 to 990 ml/hr with 0.1ml/hr increment It should have choice of Micro & Macro drops Volume Limit should be more than 9900ml It should use conventional tubing (4 mm), PVC any Indian Make Immediate display of choice of programming after purge Must have function for immediate forward & reverse of screen/programming parameters Programming : ml/hr, time x volume limit. Weight x concentration x dose Must have Occlusion Pressure from 0 to 120 kPa KVO should be 0.1 ml to 3ml/hr Bolus rate should be more than 900 ml/hr Must have Special functions like Titration , Fluid Balance Time limit from 1min to 99hrs & 59 minutes. Should have pre alarms – Ultrasonic sensor type

Visual / sound alarms like End of Battery Power , Incorrect flow rate, Door open, No drops, Occlusion , programming error, Running in KVO, Fluid balance etc. Flow rate deviation with standard tubing should not be more than +/- 6% of the programmed flow rate. Battery Rechargeable more than 5 hrs of operation Should be able to work on 220 Volts Unit should be CE Certified, ISO company & must have ISO 13485 certification

Item No 75

Bio-safety Cabinet

Page 111: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 111 of 166 01.09.2010

simple operation for ultimate safety with 60% less energy consumption and heat output that complies with the EN 12469 Exterior dimensions with stand Approx. (w x h x d) 1300 x 2200 x 795 mm (51.2 x 86.6 x 31.3 in) Interior dimensions Approx. (w x h x d) 1200 x 780 x 495 mm (47.2 x 30.7 x 19.5 in) Work surface with adjustable stand 750 to 960 mm (30 to 38 in) Interior work surface area Approx.0.56 m2 (930 sq. in) Working height of front window Approx.200 mm (8 in) Maximum lifting height of front window Approx.535 mm (21 in) dimensions Approx.(w x h x d) 1410 x 1700 x 925 mm Ventilation System Exhaust/inflow air volume 400 m3/h (230 CFM) Heat emission at 25°C ambient ~0.15 kW Filter Specification Supply/exhaust air filter HEPA H 14 EN 1822, Additional exhaust filter option (AEF) HEPA H 14 EN 1822, Performance Certification EN 12469; GS Nord Cert-TÜV Sound pressure level <55 dB (A) Lighting power >1200 lx Electrical Data Voltage 1/N/PE 230 V Frequency 50 Hz Power consumption 0.4 kW Current consumption 1.7 A Protection class I / IP 20 Protective measure Conductor connection Conductor connection

Individual precautions on customer side Lead fuse (slow blow) T 16 A or circuit breaker B 16. The local electrical regulations in the country of use as well as the relevant connection conditions must be observed. The national regulations for electrical engineering as well as the relevant technical connection conditions must be taken into account. Supply Management Supply requirement 230 V, 50/60 Hz standard supply. Total requirement including interior sockets 13-16 Amps. Receptacles The receptacles have a load capacity of up to 5 A and are protected with T 5 A fuses. When all receptacles are in use simultaneously, they must not exceed the maximum total load capacity of 5 A. Radio interference Circuit is interference free in accordance with EN 55 014 Service valves Up to 4 (installed through access ports) Receptacles One double, right side

Item No 76 Laprocator Laparoscope with operating channel. Dia 11 mm. Parallel eyepiece. Should be CE marked - 1 no. Trocar 12 mm with obturator – 1 no. Spare washers for trocar – 10 nos. Fiber optic light cable – 6 mm fiber bundle size, 300 mm long for high light output. Light source, Halogen 250 watt, twin bulb, constant color temperature. Light should not change whiteness on reducing intensity. Condenser lenses to focus light on fiber optic cable. Single puncture Ring applicator -2 nos Veress Needle 120 mm long – 2 nos CO2 insufflator flow rate – 30 LPM.

Page 112: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 112 of 166 01.09.2010

Item No 77

Ultrasound Machine 3D with Color Doppler It should be robust state of art, fully digital system capable of performing imaging application in abdominal, obs/gynae , musculo sketeal, cardiovascular, small part etc. It should be eaisly traspoartable and capable of performing bed side examination in operation theater, endoscopic and intervention suits. 1. System should have board band all digital beam former capable of processing singles from 2-18 MHz with tissue specific preset and enhanced spatial focus and resolution and complex pulse design and wave form excitation. 2. System should have advance image processing softwere algorithm to analyse target and artifacts so as to sharpen target anatomy and reduce/eliminate artifact and improve resoluction for example separate two objest 0.2mm apart laterally (laterl resoluction ) at 11 MHz approx. 3. System should in corporate facility for high resoluction 2D , M-mode , PW,CW,color flow imaiging modes, colour sdge processing. 4. System should have real time triplex mode facility in 2D, colour and Doppler modes with at least 20 frames/sec. or higher . 5. System shoyld have color campare mode color /color power mode and normal gray scale mode side by side or equivalent. 6.System should have quad &thumbanail display etc. 7. System should have dynamic range of 180db or higher. Higher will be preferred. 8. MI & TI should be displayed on screen. 9. System should have more then 500 frames per second and high PRF (5k) mention rate highest will be preferred. 10. System should have scan dept of 30cm or more. 11.System should have 256 shades of gray display and a frame rate of 100/Sec or higher. 12. System should have facility for real time or frozen, pan or point zoom upto 8times. 13. System should have cineloop review possible on all modes for individual and mixed loops for more then 5000 frames and more 30 seconds of spectral doppler and m-mode strip. 14. System should have 120 or more transmit channels and 20,000 or more receiving channels with on site upgradability to higher number. Full spectrum imaging where whole of band width is utillized. 15.Sytem should have automatic as well as manually adjusable touch screen guided control of :- i) Focus Transmit and dynamic recive at multiple dept individually and simultaneously. ii) Frame rate for B mode, color, compounded resolution ect. iii) Dynamic range compression. iv) Pre-processing such as edge enhancing, beam line density. v) Post processing such as image persistance, speckle reduction. vi) Real time compound resolution image no of viewing angles and lines per frame (<9) vii) Audio adjustments. viii)Adjusttable "packet" size in phase shift analyses for example 10 or move transmit and receive sequences. 16. System should have panoramic extended field of view and trapezoidal view on linear propes. 17. System should independent steering of B mode and color on linear probe. 18. System should have 3D acquisition and analysis package for e.g. free hand 3D ,static 3D, multislice view capabilities. a. Data Acquisition : free hand, mechanecal & electronic and gated techiniques. b. Transducer : matrix acquisition for faster scan rate with better damping of traansducer element for sharper focus.

Page 113: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 113 of 166 01.09.2010

19. It should have advanced issue harmonic imaging on all probes such as fundamental,phase inversion , coded Octave and Intravenous US contrast tuned imaging and quantification. 20. It should have cardiac package and probe for adult and fetal echo cardiography e.g. myocardial velocity, stress & strain quantification, quality intima media thickness, tissue velocity maping and peripheral vascular study optimization programmes. 21. It should have extensive software and automatic and user programmable calculation packeg for gray scale color Doppler, 3D applications. 22. It should have a 17"hight resolution medical grade LCD screen monitor with articulated arm. 23. It should be provided with following transducers, board band full spectrum functioning, with colour, covering 2 to 18 MHz. • Convex Abdominal 3-6 MHz Approximately. • Endocavity (TVS+TRUS)5-9 MHZ approx. with 200*R. • Linear high frequency 6-18 MHz approx. • Micro convex phased array 2-5/3-6 with cardiac capabilities. The configuration may vary according to manufacturer but all function should be served. 24. It should be capable of supporting at least three or more transducers ports with switching from consol. 25. System should have built in image Management Software, for off line application when patient has gone after examination, such as image Manipulation, Multi Planner reformating, etc. its should have hard disk memory of 360GB or more with built in CD/DVD read write and limitless connectivity e.g. USB/HDD/CD/DVD. 26. For paralle processing of imaging Data system should be provided with a separate latest Pentium i3 (2.9GHz) configuration with 1Tera Byte Hard based station with USB and serial port with 19" LCD medical grade monitor with very high quality image management Software (e.g.radion) with same capabilities as main machine such as retriving data along Demography, Zoom, Pan, retrieving information from CD/DVD wretriving and Teleradiology Software exporting JPG & AVI file format to link other state in the hospital information service (HIS), RIS & PACS server. iIt should have minimum memory storage for about 100 patients per day and cine loops for 6 month's work. 27. System should be provided with black & white tharmal printer & colour laser printer capacity of printing 3000 sheets per cartride and 3000sheet for starting work. 28. On line UPS for 2KVA or more as appropriate for supporting main equipment &all line accessosies for 30 minutes should be provided with system machine should have high inherent voltage fluctuation protection. 29. Standard accessories such as various cuffs, elbow rest, standardization phantoms, bio guides & needles etc. should also be provided. 30. Syatem should be upgradable to : a. Elastography and sound speed measurement free of cost as and when it become available with the Company. b. Comparison/super imposition/fusion with CT/MRI images obtained by teleradio.

Item No 78

PHOTOTHERAPHY MACHINE 1. Dimensions of the chamber should be at least 6 feetx 3 feet x 3feet. 2. Phototheraphy chamber of 18 UVA+18 NB UVB tubes designed for providing even irradiation of the body in the treatment area. 3. UV chockes must be provided to provide long life to the tube light and cooling fans for effective cooling of the unit. 4. integrated dosimeter system for easy calculation of irradiation levels. 5. The equipment have CE or FDA or ISI certification.

Page 114: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 114 of 166 01.09.2010

6. Advanced micro computerized electronic LCD/TFT Controller which allows setting of joules/time for UVA and milli Joules/time for NB UVB tubes. 7. Automatic computation of irradiation time from joules/time for NB UVB tubes. 8. Dose limit can be preset and cumulative dose is displayed instntaneously with provision of storage of data. Provision of 'software backup' is preferable. 9. Variation in irradiation is taken care by built in UVA/NB UVB sensors which should be able to detect all irradiation completely and uniformly. 10. Switches the system 'off' automatically with warming alarm at the end of set irradiation time. 11. Built in memory system that helps to avoide error in treatement. 12. Body to be of a metal which is rust free so as to ensure long rust free life of the unit. 13. Automated and/or mechaical safety mechanism to prevent excess irradiation to the pantients so as to avoid/prevent burns etc. 14. Electrical leakage circuit tripper/breaker in each panel to ensure maximum safety of the patient. 15. Open top unit to ensure maximum ventilation and prevent claustrophobia. 16. Mechanism to provide information to the patient regarding duration of treatment and time left for exposure during their treatment. 17. Computer for patient data management with software and interface for the phototheraphy chamber which is Rs.232 compatible. 18. To be supplied with suitable stabilizer. 19. Black UV Goggles and Eye pads cover (3 pairs each for adult and 3 pair for children ) as protective+

Item No 79 Binocular microscope a) Optical system infinity color corrected optics, antifungus treated. Eye pieces: 10X wide fied ( FV 22 or more ) with inter pupillary distance 48-75mm with dioptic adjustment both side, eye guards, eye level riser. a) Objectives: Bright field infinity color corrected optics, antifungus treated 4X,10X,40X,100X oil immersion. In changing from one objective to anther or reintroducing the same objective by rotation of the nosepiece, the center of the field should not apper displaced by more then 0.02mm in object plane. b) Nosepice : Revolving, reversed (inward) tilt c) Tubes : Siedento f tiltable Binocular tubes with minimum inclination 25-30 degrees . d) Stage : uniformly horizontal, scratch resistant, rackless, rotatable stage with right hand operation & single slide holder with a stage upper limit stopper. e) Condenser : issuing out universal with numerical aperture of 0.9/1.25 with position for bright field should have a removable filter holder, swing in, blue filter for bright field. f) lllumination system : The system should have a bulit in, variable, low voltage light source, the circuit for the light source should include a constant voltage supply. The system should be provided with a step down transformer and on/off switch and intensity control. the lamp should be provided with a lamp socket, which has the facility for easy replacement of the bulb. The housing of the microscope. Halogen bulb -12v/20-30w. The illuminator should have a built in field diaphragm for kohler illumination. Power supply : Voltage 220V, 50hZAC should have one on- off power switch, power cord with a 3pin male plug. The system should have an inbuilt protective/ safety devioce to withastand fluctuations of voltage from 140v to 280v. The fuse the halogen lamp should be easily accessible. g) Arm rest, Left and right.

Item No 80

Page 115: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 115 of 166 01.09.2010

Yag Laser A. PHOTOCOAGULATOR MODE OF OPERATION : Continuous wave TYPE : Solid state laser b WAVELEGTH : 532nm COOLING : Air cooled AIMING BEAM : Continuous wave, diode laser, 635nm B. SLIT LAMP SPOT SIZE : 50-1000µm PROTECTION FILTER FOR SAFETY MAGNIFICTION : 5step ILLUMINATION UNIT : Slit width and length 0-14mm Filters : blue, red-free, heat absorbing Halogen lamp POWER SUPPLY : 240V, 50-60Hz BASE UNIT to alloew longitudinal, lateral and vertical movements C. SLIT-LAMP ADAPTER D. Motorrized table E. Mainster and PRP lenses F. UPS 1000 VA

Item No 81 CELL SEPARATOR 1. The apheresis machine should have facility for all Blood components collection with single /double access including 1. Platelepheresis, 2. Peripheral blood stem cells, 3. Therapeutic Plasma Exchange. And 4.Red Cell Apheresis. Protocols for the above said proceduresis. 2. The cell separator should be fully automatic, microprocessor controlled with access operator control panel such as touch screen. 3. It should be based on intermittent or continuous flow technology during duble/ single needle access. 4. The Cell separator should ensure all donor safety parameters before starting the procedure and at all time during operation. 5. The cell separator should be able to perform priming with normal salineor with mixture of anticoagulant and self test as a part of donor's safety. 6. The cell separator should have inbuilt cuff pressure and promt grip for donor comfort and adequate blood flow. 7. It should have auto cuff mechanism for automatic inflation and deflation. 8. Advance help menu should be available at any time during alarm conditions. 9.Extra corporeal volume not exceed 250 ml. 10. The yield estimator to help decide yield, volume to be processed and suggested storege fiuid. 11.Interphace detector should continuously monitor the collection line to avoid any contamination during esch protocol. 12. Capable of downloading or printing full procedure report any time time after the procedure. 13. The cell separator should have rechargeble battery to store data and re-start in case of power filure. 14. Inlet and return flow rates upto 100 ml +/-15 ml per minute. 15. In case inlet line occlusion, machine should be able to re-start automatically.

Page 116: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 116 of 166 01.09.2010

16. It should have fluid leak detector for donor's safety. 17. It should have provision for saline re-infusion to donor. 18. It should be able to regulate ACD delivery, should not have bolus return. 19. It should have automatic door lock for centrifuge during the procedure. 20. Lickable castors/ trolley for easy mobillity. 21. Should have Automatic kit loading facillity. 22. The disposable kit should be self sufficeient in terms of collection bag, transfer beg, needles, anticoagulant and normal saline. 23. Should be a totally closed system. 24. Should be able to provlbe Leucoreduced platelet products 25.Power supply -220-240 volts/50Hz. 26. Suitable UPS with maintenance free batteries for minimum one hour back up. 27. FDA/CE approved 28. Five year Warranty. Five year CMC 29. Disposable kits should be available at fixed price for three years from date of installation

Item No 82 CELL WASHER 1. Microprocessor controlled fully automatic cell washer 2. Touch pad with digital LCD display 3. Programmes for wash cycles, saline volume, spin time and RPM 4. Save and store multiple programmes 5. To accommodate 12 tube of standard 12x75 mm size 6. Brushless motor for quiet operation 7. Saline detec system with audible low saline warining 8. Agitate cycle to ensure complete re-suspension of cells 9. Audible and visible alert at the end of process 10. Power 220 v/50 Hz 11. CE/FDA/BIS approved

Item No 83 DONOR COUCH 1. Provide comfortable position to donors 2. Chair type with soft padding for cushioning and rexin cover 3. Ability to adjust donor's position automatically by operating a simple switch 4. Easily tilted to head low position 5.Easily adjustable arm rest with swinging out as well as moving up and down facility 6. Arm rest wide, sturdy to hold donor's arm in a comfortable position throught the procedure of blood donation, provided on both sides 7. Lifting capacity : minimum 130 Kilograms 8. Four lockable castors for easy mobility 9. Trays and stands for keeping all blood collection accessories 10. Input power -220V/50Hz 11. CE/FDA/BIS approved

Item No 84

Page 117: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 117 of 166 01.09.2010

PLATELET INCUBATOR AND AGITATOR 1. Coountertop model for safe storage of platelets 2. Temperature set to 22+/- 2 degree C 3. Capacity : minimum 96 random bags or 32 apheresis bags 4. Power coated MS exterior, stainless stell interior 5. Transparent glass door 6. Flate bed agitator 7. Oscillations: 70 to 80 cycles per minute 8. Agitation stops when the incubator is opened and restarts automatically 9. Pull out easily removable, stainless steel, perforated doawers 10. Microprocessor controlled temperature indicator with digital display of temperature 11. Alarm for temperature deviation, door open, power failure, sensor failure 12. Built in vasible and autdible motion alarm 13. 24x7 days inkless temperature chart recorder 14.Power 220v/50 Hz 15. Voltage stabilizer of appropriate required specifications conmpatible with the unit to be provided 16.All disposables to be provided at fixed rate for three years from of installation 17. CE/FDA/BIS approved

Item No 85

STERILE TUBE CONNECTING DEVICE 1. Compatible with all standard of different types of blood bag 2. Able to weld wet to wet, dry and wet to dry tubing combinations 3. In built sensor to monitor the temperature during entering welding process 4. LCD display for continuous monitoring 5. Ensure complete sterility of transferred blood 6. No Particle or chemical residues should be formed during the welding procedure 7. Welds should have optimal tube alignment and no leakage of blood of during the process 8. Automatic replacement of wafers after each use 9. Built in wafer disposal box 10. Input power -220V/50Hz 11. CE/FDA/BIS approved

Item No. 86

General Instruments Set CATHETERISATION TRAY X 10

Sponge holder 8" 2 Artery forceps straight 9" 1 S.S. Bowl 10 cm 2 S.S.Kidney tray 10" 1 S.S. Adoptor 1 SPINAL TRAY X 10 Sponge holder 8" 2 Artery forceps curved 7" 1

Page 118: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 118 of 166 01.09.2010

S.S. Bowl 10 cm 2 S.S.Kidney tray 10" 1 BASIC SET x 10 B.P. Handle No.3 2 B.P. Handle No.4 2 Dissecting Forceps Plain 7” 2 Dissecting Forceps Plain 8” 1 Dissecting forceps Toothed 7" 1 Dissecting Forceps Toothed 8” 1 Adson Dissecting Forceps Plain 6" 1 Adson Dissecting Forceps Toothed 6" 1 Towel clip 5" 6 Cd. Scissor Suture Cutting 1 Mayo Scissor Cd. 9" TC 1 METZScissor Cd. 8" TC 1 Mayo scissor 8" st. 1 Mayo Scissor 8” 1 Needle Holder 7” Mayo hegar 1 Needle Holder 8” fine 1 Needle Holder 7” Mayo hegar 1 Artery Forceps Cd. 8” 2 Mosq. Artery Forceps Cd. 4 Mosq. Artery Forceps st. 2 Artery Forceps Cd. 7” 6 Artery Forceps St. 7" 2 Allis Forceps 7” 4 Babcock Tissue Holding Forceps 67” 2 Probe and director 1 Suction Tip No. 1, 2, 3, 4 1 Each Yaunker's Suction with detatchable tip 1 Mixture Clamp 7” 1 Langenback Ret MEDIUM 2 Langenback Ret SMALL 2 “C” Shaped Retractor ( pair ) small & med 2 Each Sponge Holder 8" 4 Skin hook sharp 2 Vein loops 2 S.S.Bowls 10 cm 4 S.S. Kidney tray 12" 2 PILES EXTRAS x 3 Sims speculum med 1 Sims speculum small 1 Sims speculum large 1 Speculam with one handle 1 Proctoscope small 2 Proctoscope med 2 Proctoscope big. 2 Artery forcep cd 8" 2 Venesection cannula 16,18,20,22 1 Each Proctoscope small 2 Proctoscope med 2 Proctoscope big. 2 Piles needle 1

Page 119: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 119 of 166 01.09.2010

Item No. 87

Minor Surgery Instrument Set

Minor Surgery Set

B.p.handle no.3 4 B.p.handle no.4 2 Dissecting forceps plain 7" 2 Dissecting forceps toothed 7" 2 Adson Dissecting Forceps Plain 2 Adson Dissecting Forceps Toothed 2 Towel Clip 4” 8 Iris Scissor st. 2 Iris Scissor cd. 2 Mosq. Artery forcep cd 6" 12 Artery Forceps Mosq. St 6" 8 Artery Forceps st. 7” 8 Artery forceps 7” Cd. 8 Allis forceps 7” 8 Babcock tissue forceps 4 Needle holder 8" mayo hegar 2 Needle holder 8" fine 2 Mayo scissor 8" st. 2 Metz. Scissors cd 8" 2 Mayo Scissor cd. 2 Venesection scissors 4" curved 2 Sponge holder 10" 8 Suction tip no1,2,3,4 2 Scoop Double Ended 2 Probe and director 2 Kilner retractor 4 Skin Hook 4 Lagenback retractor medium 4 Sinus forcep 2 Venesection cannula 16,18,20,22 2 S.S.bowls 10 cm 8 S.S.kidney tray 10" 4

CATHETERISATION TRAY

Sponge holder 8" 10 Artery forceps straight 7" 5 S.S. Bowl 10 cm 10 S.S.Kidney tray 10" 5 S.S. Adoptor 5 SPINAL TRAY

Sponge holder 8" 10 Artery forceps curved 6" 5

Page 120: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 120 of 166 01.09.2010

S.S. Bowl 10 cm 10 S.S.Kidney tray 10" 5

BASIC SET

B.P. Handle No.3 10 B.P. Handle No.4 10 Dissecting Forceps Plain 7” 10 Dissecting Forceps Plain 8” 5 Dissecting forceps Toothed 7" 5 Dissecting Forceps Toothed 8” 5 Adson Dissecting Forceps Plain 5 Adson Dissecting Forceps Toothed 5 Towel clip 5" 30 Cd. Scissor Suture Cutting 5 Mayo Scissor Cd. 8" TC 5 METZScissor Cd. 8" TC 5 Mayo scissor 8" st. 5 Mayo Scissor 8” 5 Needle Holder 7” Mayo hegar 5 Needle Holder 8” fine 5 Needle Holder 8” Mayo hegar 5 Artery Forceps Cd. 8” 10 Mosq. Artery Forceps Cd. 20 Mosq. Artery Forceps st. 10 Artery Forceps Cd. 7” 30 Artery Forceps St. 7" 10 Allis Forceps 7” 20 Babcock Tissue Holding Forceps 7” 10 Probe and director 5 Suction Tip No. 1, 2, 3, 4 5 Nos Each Yaunker's Suction with detatchable tip 5 Mixture Clamp 7” 5 Langenback Ret MEDIUM 10 Langenback Ret SMALL 10 “C” Shaped Retractor ( pair ) small & med 10 Each Sponge Holder 8" 20 Skin hook sharp 10 Vein loops 10 S.S.Bowls 10 cm 20 S.S. Kidney tray 12" 10

Item No. 88

INSTRUMENTS SETS FOR ORTHOPAEDIC SURGERY Wire Instrument Set 1 Set

Sterilising Tray with lid 1 Spare Tray 1

Page 121: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 121 of 166 01.09.2010

Lower Tray 1 Drill Bit, 2.0mm 1 Triple Drill Guide 2.0 with 3 holes, opposite side 1 hole 1 Wire Passer, 45mm bending diameter 1 Wire Passer, 70mm bending diameter 1 Wire Tightener with handle and two pegs, 1 Holding Forceps for Cerclage Wires, 1 Wire Bending Pliers, 1 Parallel Pliers, flat nosed 1 Wire Cutter, large, 1 Wire Cutter, short, 1 Bending Iron, for Kirschner Wires 1

Medullary Reamer Set 1 Set

Flexible medullary reamer 6.0mm 1 Flexible medullary reamer 6.5mm 1 Flexible medullary reamer 7mm 1 Flexible medullary reamer 7.5mm 1 Flexible medullary reamer 8.0mm 1 Flexible medullary reamer 8.5mm 1 Flexible medullary reamer 9mm 1 Flexible medullary reamer 9.5mm 1 Flexible medullary reamer 10mm 1 Flexible medullary reamer 10.5mm 1 Flexible medullary reamer 11mm 1 Flexible medullary reamer 11.5mm 1 Flexible medullary reamer 12mm 1 Flexible medullary reamer 12.5mm 1 Flexible medullary reamer 13mm 1 Flexible medullary reamer 13.5mm 1 Flexible medullary reamer 14mm 1 Flexible medullary reamer 14.5mm 1 Flexible medullary reamer 15mm 1 Flexible medullary reamer 15.5mm 1 Flexible medullary reamer 16mm 1

Bone Forcep Set For Reduction 1 Set

Verbrugge forceps 240mm 2 Verbrugge forceps 260mm 2 Verbrugge forceps 280mm 2 Reduction forceps broad,130mm 2 Reduction forceps W point,205mm 2 Reduction forceps,serrated,170mm 2 Reduction forceps,serrated,240mm 2 Aluminium Case,Red 1 Tray (Lower/upper) 1 Tray (Lower/upper) 1

DYNAMIC SCREW/DCS HIP SCREW INSTRUMENT Set 1 Set

Sterilising Tray with lid 1 Upper Tray 1 Lower Tray 1 DHS/DCS Threaded Guide Wire, 2.5mm dia., L 230/5mm 10 DHS Angled Guide 135º 1 DHS Angled Guide 150º 1

Page 122: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 122 of 166 01.09.2010

DHS/DCS Direct Measuring Device 1 DHS/DCS Wrench for one-step insertion L 230mm 1 DHS/DCS T-Handle with quick coupling, L 80mm 1 DHS Triple Reamer 1 DHS/DCS Impactor for one-step insertion, L 260mm 1 DHS/DCS Tap, L 220mm 1 DHS/DCS Centering Sleeve, locking 1 Coupling Screw, cannulated 1 DCS Angled Guide 1 DCS Triple Reamer 1

Small Fragement Instrument Set 3.5mm 1 Set

Sterilising Tray with lid 1 Lower Tray for Instruments 1 Tray for Plates 1 Upper Tray 1 Drill Bit, 2.5mm dia., L 110/85mm for quick coupling 2 Drill Bit, 3.5mm dia., L 110/85mm, for quick coupling 2 Countersink Shaft 3.5, L 72mm 1 Tap for 3.5mm Cortex Screws L 50/110mm 1 Tap for 4.0mm Cancellous Bone Screws L 110mm 2 T-Handle with quick coupling, L 80mm 1 Double Drill Sleeve 3.5/2.5 1 Insert Drill Sleeve 3.5/2.5, L 42mm Drill Bit 2.5mm dia. 1 Screwdriver, hexagonal, small, with Holding Sleeve 1 Screwdriver Shaft, hexagonal, small, L 100mm, for quick coupling 1 Screwdriver, hexagonal, small, with groove, L 200mm 1 Holding Sleeve, 1 Depth Gauge for 2.7mm to 4.0mm Screws 1 Sharp Hook, L 155mm 1 Holding Clip 4.5 - 7.0mm 2 Screw Forceps, self-retaining, L 85mm 1 DCP® Drill Sleeve 3.5 for neutral and load position 1 Bending Iron, slit widths 4.5/2.5mm, L 150mm, for Plates 2.7 and 3.5 1 Bending Iron, slit widths 2.5/4.5mm, L 150mm, for Plates 2.7 and 3.5 1 Bending Pliers for Plates 2.4 to 4.0, L 230mm 1 Bending Template for DCP® 3.5 and LC-DCP 3.5, L 87mm 1 Bending Template for DCP® 3.5 and LC-DCP 3.5, L 114mm 1 Wire Bending Pliers, L 155mm 1 Bending Iron, for Kirschner Wires 1.25 to 2.5mm dia., L 120mm 1 Reduction Forceps with points, L 130mm 1 Reduction Forceps with points, wide, ratchet lock, L 132mm 1 Reduction Forceps, toothed, ratchet lock, L 140mm 1 Bone Holding Forceps, self- centering, speed lock, L 190mm 1 Retractor, small, 8mm wide, short narrow tip, L 160mm 2 Periosteal Elevator, round edge, 6mm wide, L 200mm 1 Retractor, 15mm wide, L 160mm 2

Basic Instrument Set Large Fragment 4.5mm 1 Set

Sterilising Tray with lid 1 Upper Tray for Instruments 1 Spare Tray, without contents, 1 Lower Tray 1 Drill Bit, 3.2mm dia., L 145/120mm for quick coupling 3 Drill Bit, 4.5mm dia., L 147/120mm for quick coupling 2

Page 123: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 123 of 166 01.09.2010

T-Handle with quick coupling, L 80mm 1 Tap for 4.5mm Cortex Screws, L 70/125mm 2 Tap for 6.5mm Cancellous Bone Screws L 195mm 1 Double Drill Sleeve 4.5/3.2 1 Insert Drill Sleeve 4.5/3.2, L 80mm 1 Double Drill Sleeve 6.5/3.2 1 Screwdriver Shaft, hexagonal, 1 Screwdriver, hexagonal, large, 1 Holding Sleeve, large, L 120mm 1 Depth Gauge for 4.5 to 6.5mm Screws 1 Sharp Hook, L 155mm 1 Tension Device, articulated 1 Combination Wrench, 1 DCP® Drill Sleeve 4.5 1 LC-DCP Drill Sleeve 4.5 1 Universal Drill Sleeve 4.5 1 Bending Template for DCP® 4.5 and LC-DCP 4.5, L 210mm 1 Bending Template for DCP® 4.5 and LC-DCP 4.5, L 120mm 1 Bending Template for DCP® 4.5 and LC-DCP 4.5, L 155mm 1

Basic Intramedullary Nailing Insts set for Femur & Tibia 1 Set

Universal Socket Wrench 1 Hex Driver for 4.5/5.0 mm screw,6,4mm,5.0mm can screw,measuring guage for nails 1 Medullary Exchange tube 1 internal fracture Alignement Device 1 Depth Guage 1 Cannulated Drill Bit,150mm,3.5mm,4.0mm,4.8mm 1 Drill Bit,150mm,2.7,3.5mm,4.0mm,4.8mm 1 Twist Bit,305mm,3.5mm,4.0mm,4.8mm 1 Trocar 4.0X240mm 1 Krischner-wire with thread 3.2X 305mm 1 Guide Wire for cannulated screw 1.8X350mm 1 cannulated wire for ILN 2.4x900mm,3.0x900,Guide Wirw for reamer 2.0x900,Guide wire with curved Tip 3.0X900mm,2.4X900mm

1

Open End Wrench SW 17/ 1 Measuring Gauge 1 Measuring Gauge for nails 1 Supine Driver 1 Handle with jacobs chuck 1 Tibial Proximal Device 1 Skin Protector 1 Curved Awl 1 Trocar 3.2mm,4.0mm,4.8mm,8.0mm 1 Tapered Reamer,Cannulated 13.5,15.5mm 1 Cannulated Reamer 9.0mm 1 Drill Sleeve 2.1mm,3.5mm,4.0mm,4.8mm,8.0mm 1 Slide Hammer 1 Tibial Extractor Bolt 1 Guide Rod For Cannulated Screws 1.8X350 1 Universal Proximal Device for Femural Nails 1 Nail Adapter Bolt 1 Length Guage 1 Twist Drill Cannulated 305mm,4.0mm,4.8mm 1 Reamer 1

Page 124: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 124 of 166 01.09.2010

General Instrument Set 1 Set

Sterilising Tray with lid 1 Spare Tray, 1 Lower Tray 1 Bone Hook, sharp, small, L 230mm 1 Bone Hook, sharp, medium, L 230mm 1 Retractor, 8mm wide, short narrow tip, L 220mm 2 Retractor, 18mm wide, short narrow tip, L 235mm 2 Retractor, 24mm wide, long and wide tip, L 270mm 1 Periosteal Elevator, curved shaft, 14mm wide, L 200mm 1 Periosteal Elevator, round edge, 6mm wide, L 200mm 1 Periosteal Elevator, straight shaft, 14mm wide, L 200mm 1 Hammer 500g, L 230mm 1 Chisel Handle, L 185mm 1 Chisel Blade, 10mm wide, thickness 0.9mm, L 81mm 1 Chisel Blade, 16mm wide, thickness 0.9mm, L 81mm 1 Chisel Blade, 25mm wide, thickness 0.9mm, L 81mm 1 Gouge, curved, for cancellous bone graft harvest., 10mm wide, L 250mm 1

Damaged screw,Broken Nail Removal Instrument set 1 Set

Aluminium Case, white, small, perforated, without contents 1 Tray, subdivided 1 Tray for Instruments, without contents 1 Hollow Reamer for 3.5/4.0mm Screws 1 Spare Reamer Tube 1 Extraction Bolt for 3.5/4.0mm Screws 1 Extraction Screw, conical, for 2.7mm, 3.5mm and 4.0mm Screws 1 Hollow Reamer for 4.5mm Screws 1 Spare Reamer Tube 1 Extraction Bolt for 4.5mm Screws 1 Extraction Screw, conical, for 4.5/6.5mm Screws 1 Hollow Reamer for 5.0/6.0/6.5/7.0mm Screws 1 Spare Reamer Tube 1 Extraction Bolt, for 5.0/6.0/6.5/7.0mm Screws 1 Aluminium Plate, anodized 1 Sharp Hook, L 155mm 1 Forceps for Screw Removal, L 205mm 1 Gouge, 10mm wide, L 205mm 1 Extract-Hook f/Nails ø9-11 1 Extract-Hook f/Nails ø11+larger 1 Extract-Hook long f/Nails ø11-14 1 Extract-Hook L480 f/Nails ø9 1 Connector f/Extract-Hook 1 T handle 1 T-Handle with quick coupling, L 80mm 1

Locking Compression Instrument set Large& Small 1 Set

Torque Screw Driver 3.5mm 1 Torque Screw Driver 5.0mm 1 Drill with Quick Coupling 2.5x150/30mm 2 Drill with Quick Coupling 2.7x200/40mm 2 Drill with Quick Coupling 3.2x150/30mm 2 Drill with Quick Coupling 4.1x200/30mm 2 Conical Screw Extractor 2.5mm 1

Page 125: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 125 of 166 01.09.2010

Conical Screw Extractor 3.5mm 1 Screw Driver Shaft 2.5mm Hex 2 Screw Driver Shaft 3.5mm Hex 2 Holding Sleever 3.5mm/ 4.0mm 1 Drill Sleever 2.8/3.5mm 2 Quick Coupling Handle 1 Universal Drill Guide 3.5mm 1 Drill Guide Handle 1 Drill Bit 4.5mm / 145mm Quick Coupling 2 Drill With Quick Coupling 4.3x 195mm 2 Screw Driver Shaft 2.5mm Hex 100mm 1 Screw Driver Shaft 3.5mm Hex 100mm 1 Drill Sleever 2.5mm - load 2 Drill Sleever 4.1mm - load 2 Drill Sleever 2.5mm with Compression 2 Drill Sleever 4.1mm with Compression 2 Universal Drill Guide 4.5mm/3.2mm 1 Locking Screw Drill Sleever 3.5mm 1 Locking Screw Drill Sleever 5.0mm 1 Tap with Quick Coupling 3.5 2 Tap with Quick Coupling 4.5 2 Locking Screw Tap with Quick Coupling 3.5mm 2 Locking Screw Tap with Quick Coupling 5.0mm 2 Hexagonal Screw Driver 2.5mm 1 Hexagonal Screw Driver 3.5mm 1 Countersink with Quick Coupling 3.5mm 1 Countersink with Quick Coupling 4.5mm 1 Depth Gauge 3.5mm 1 Depth Gauge 4.5mm/6.5mm 1 Bending Iron 2 Quick Coupling for Jacob Chuck 1 Tap / Drill Sleeve 2.5mm 1 Tap / Drill Sleeve 3.5mm 1 Tap / Drill Sleeve 4.5mm 1 Tap / Drill Sleeve 6.5mm 1 Fusion Lock Set - Case Large Fragment Instruments 1

Cannulated Screws Instruments set 4.5mm 1 Set

Sterilising Tray with lid 1 Threaded Guide Wire, 1.6mm 10 Drill Bit, 1.5mm dia. 1 Drill Bit, 3.2 2 Countersink Shaft, cannulated, 1 T-Handle with quick coupling, 1 Tap, cannulated, for 4.5mm Cannulated Screws 1 Double Drill Sleeve 4.5/3.2 1 Parallel Wire Guide, 1 Trocar 1.6mm dia. 1 Screwdriver, hexagonal, large, 1 Drill Sleeve 1 Protection Sleeve 1 Screwdriver Shaft, hexagonal, large 1 Screwdriver, hexagonal, cannulated 1 Holding Sleeve 1 Measuring Device, 1

Page 126: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 126 of 166 01.09.2010

Stylet, 1.6mm 1 Holding Clip 4.5 - 7.0mm for screw 1

Cannulated Screws Instrument Set 3.5 mm 1 Set

Vario Case for instrumrnts 1 Lid Stainless Steel size½ f/VC 1 Hold-Sleeve f/ 314.070 314.290 314.550+3 1 DoubleDrillGuide 3.5/2.5 1 ScrDriverShaft-hex-small ø2.5 1 Tap-cannul f/Cannul-Scr ø3.5+4 1 Drill Bit ø3.5/1.35 cannul L160/130 4flu 1 Drill Bit ø2.7/1.35 cannul L160/130 4flu 1 Handle w/Quick-Coupl L110 1 DoubleDrillGuide 2.7/1.25 1 Countersink cannulated f/Cannul-Scr ø3.5 1 ScrForceps self-hold L85 1 Sharp Hook L155 1 Direct Measur-Device f/Cannul-Scr ø3.5+4 1 Clean-Brush ø1.35 f/Cann-Instr 1 Hold-Clip f/Washers 1 GuideWire ø1.25 w/thread-tip w/trocar L1 1 ScrDriver-hex-cann f/Cannul-Scr ø3.5+4 1

Item No. 89

Vaginal Hysterectomy B.P.Handle No.3 1 B.P.Handle No.4 1 Diss. Forceps Plain 16cm 1 Diss. Forceps Plain 20cm 1 Diss. Forceps Tooth 20cm fine 1 Diss. Forceps Tooth 20cm 1 Diss. Forceps Tooth 16cm fine 1 Diss. Forceps Attri. 20Cm 1 Diss. Forceps Fine Plain 20cm 1 Artery Forceps St.16cm 2 Artery Forceps Curved.16cm 6 Artery Forceps Mosquito curved 5 Artery Forceps Mosquito curved 5 Artery Forceps Mosquito St. 2 Heaany's Clamp 6 Valsalum Forceps Curved 2 Allis Forceps 20cm 6 Allis Forceps 16cm 6 Babcock Forceps 20cm 2 Babcock Forceps 16cm 2 Tanaculum Forceps 25cm 3 Artery Forceps Long Curved 23cm 2

Page 127: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 127 of 166 01.09.2010

Kochers Clamp 20cm St. 2 Kochers Clamp 20cm Cu. 4 Hysterectomy Clamp 23cm 6 Needle Holder 20cm 1 Needle Holder 20cm 2 Needle Holder 15cm 1 Needle Holder 16cm 1 Scissors Mayo St. 15cm 1 Scissors Mayo St. 18cm 1 Scissor Mayo Curved 17cm 1 Scissor Mayo Curved 18cm 1 Scissor Mayo Curved 20cm 1 Sponge Holder 10" St. 2 Sponge Holder 10" Cd. 2 Avard Vaginal Spaculam Handle 1 Avard Vaginal Spaculam Weight 1 Avard Vaginal Spaculam Blade 1 Avard Vaginal Spaculam Blade 1 Sim's Spaculam 1 Sim's Spaculam Blade Guide 1 Sim Speculam Medium 1 Sim Speculam Large 1 Uterus Holding Forceps 1 Anterior Vaginal Wall Retractor 1 Jackson Retractor 2 Suction Cannula Trizer No.4 1 Towel Clip 6 S S Bowls 10 cm 2 S S Kidney tray 12" 1

Item No. 90

Abdominal Hysterectomy set

1 Sponge Holder x4 2 Needle holder x3 3 Bab cock x2 4 Artery forceps 6.1/2” x10 5 Artery forceps7.1/2” x4 6 Allis Forceps x6 7 Straight Hyst. Clamp x5 8 Curved clamp x5 9 Mosquitoes forceps 4.1/2” x6 10 Doyn’s Retractor Big x1 11 Self Retaining Forceps (Balfour) x1 12 Deverse Small x1 13 Deverse Big x1

Page 128: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 128 of 166 01.09.2010

14 Towel clip x5 15 Steel Basin x2 16 S.S Bowl x3 17 Steel Tray Big x2 18 Myo curved scissor 8” x2 19 Straight scissor 6.5” x2 20 Tooth dissecting forceps 6.5” x1 21 Non Tooth dissecting forceps 6.5” x1 22 Metzenbaum curved scissor 8” x2

Item No. 91

Delivery and Ceasarian set

B.P.Handle No.3 1 B.P.Handle No.4 1 Dissecting Forceps Tooth 15cm 2 Dissecting Forceps Plain 15cm 1 Scissor Mayo St. 18cm 1 Scissor Mayo St. & Curved 16cm 1 Each Scissor Episectomy 1 Needle Holder 18cm 2 Placenta Puncture Forceps 1 Sponge Holder 20cm 2 Sponge Holder 26cm 1 Anterior Vaginal Retractor 1 Artery Forceps Curved 16cm 2 Cord Clips 1 Sim's Speculum Medium 1 Sim's Speculum Large 1 Towel Clip 4 S. S. Bowl 6 cm 2 Big Kidney Tray 1 Allis FORCEPS 8 “ 2 Mosquito Artery cd 2 Ceasarian set B.p handle no-3 1 B.p handle no-4 1 Dissecting forceps fine toothed 14cm 1 Dissecting forceps adson plain 12.5cm 1 Dissecting forceps adson toothed 12.5cm 1 Dissecting forceps Atragrip 20cm 1 Dissecting forceps Atragrip15cm 1 Dissecting forceps plain 15cm 1 Dissecting forceps toothed 20cm 2 Cord clip 1

Page 129: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 129 of 166 01.09.2010

Towel clip backhus 5 Allis forceps 16cm 6 Allis forceps 20cm 2 Babcork forceps 16cm 2 Babcork forceps 20cm 2 Artery forcpes st 16cm 8 Artery forcpes cd 20cm 4 Artery forcpes cd 16cm 12 Needle holder 18cm 1 Needle holder 20cm 1 Needle Holder15cm 1 scissors mayo cd 16cm 1 Scissors mayo st 20cm 1 Episectomy scissors 1 McIndo scissors cd 23cm 1 Suture cutting wave scissors 18cm 1 Uterine scissors (sims) 20cm 1 kocher's clamp cd 2 Kocher's clamp st 4 Green Armitage forceps 6 Sponge holder 26cm 4 Yankurs suctions tip 2 Frizer suction tip cannula NO-3 1 C' shaped retractor 3 Doyans retractor 2 Morries retractor 1 Doyen's skin clip 1

Item No 92

General Instruments for O T

B.P.HANDLE NO.3 22 B.P.HANDLE NO.4 20 B.P.HANDLE NO.7 10 BLADDER SOUND 5 CANNULA OVONOTOMY TROCAR FOR (HYEDRO) 5 CANNULA SUCTION SET 15 SETS CERVICAL NEEDLE 4 CHIMKO 2 DILATORS MALE 2 SETS DILATORS HEGARS(UTERINE) 14 SETS DOYEN MYOMA SCREW 4 FORCEPS PLAIN ADSON 5 FORCEPS TOOTH ADSON 5 FORCEPS ALLES 5" STRAIGHT 40

Page 130: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 130 of 166 01.09.2010

FORCEPS ALLES 5" CURVED 4 FORCEPS ALLES 6" 106 FORCEPS ALLES 7" 44 FORCEPS ALLES 8" 142 FORCEPS ALLES 10" 42 FORCEPS ARTERY STRAIGHT 5" 64 FORCEPS ARTERY CURVED 5" 44 FORCEPS ARTERY CURVED 10" 75 FORCEPS ARTERY CURVED 7" 140 FORCEPS ARTERY CURVED 9" 10 FORCEPS ARTERY CURVED 10" 100 FORCEPS ARTERY CURVED 8" 120 FORCEPS ARTERY CURVED 6" 70 FORCEPS ARTERY STRAIGHT 10" 10 FORCEPS ARTERY STRAIGHT 8" 28 FORCEPS ARTERY STRAIGHT 6" 70 FORCEPS ARTERY STRAIGHT 7" 44 FORCEPS BABCOCK 10" 10 FORCEPS BABCOCK 7" 10 FORCEPS BABCOCK 8" 30 FORCEPS BABCOCK 6" 20 FORCEPS BONNEY VAGINAL 5 FORCEPS BERKELEY BONNEY HYSTERECTIOMY 8" CURVED ON FLAT 5 FORCEPS COLLIN UTERUS HOLDING 10" 5 FORCEPS CHITTLE 12" 54 FORCEPS CHITTLE 10" 10 FORCEPS DISSECTING (TOOTH) 50 FORCEPS DISSECTING TOOTH 20CM. 50 FORCEPS DISSECTING 15CM.(6") 100 FORCEPS FISTULA DISSECTING 10 FORCEPS GLAND HOLDING 4 FORCEPS GREEN ARMYTAGE 4 FORCEPS GWILLINGAM HYSTERECTOMY CORVED 8" 5 FORCEPS KOCHER ARTERY CURVD10" 50 FORCEPS KOCHER ARTERY CURVED 8" 90 FORCEPS KOCHER ARTERY CURVED 6" 50 FORCEPS KOCHER ARTERY CURVED 7" 30 FORCEPS KOCHER ARTERY CURVED 12" 15 FORCEPS KOCHER ARTERY STRAIGHT 12" 25 FORCEPS KOCHER ARTERY STRAIGHT 8" 10 FORCEPS KOCHER ARTERY STRAIGHT 6" 15 FORCEPS KOCHER ARTERY STRAIGHT 7" 10 FORCEPS LUCS BIOPSY SMALL 2 FORCEPS LUCS BIOPSY MEDIUM 2 FORCEP MAGILIS 10 FORCEPS MOSQUITO CURVED 65 FORCEPS MOSQUITO CURVED 4" 20 FORCEPS MOSQUITO CURVED 5" 30 FORCEPS MOSQUITO STRAIGHT 25 FORCEPS MOSQUITO STRAIGHT4" 10 FORCEPS MOSQUITO STRAIGHT5" 10 FORCEPS OVUM 10" 15 FORCEPS PUNCH BIOPSY 10 FORCEPS SPONGE HOLDING6" 20

Page 131: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 131 of 166 01.09.2010

FORCEPS SPONGE HOLDING 8" 10 FORCEPS SPONGE HOLDING 10" 10 FORCEPS SPONGE HOLDING 12" 20 FORCEPS SWAB HOLDING6" 25 FORCEPS SWAB HOLDING12" 50 FORCEPS TOOTH4" 10 FORCEPS TOOTH5" 55 FORCEPS TOOTH6" 35 FORCEPS TOOTH7" 60 FORCEPS TOOTH8" 10 FORCEPS PLAIN5" 15 FORCEPS PLAIN6" 50 FORCEPS PLAIN7" 55 FORCEPS PLAIN8" 50 FORCEPS HEANEYS HYSTERECTOMY CLAMP 8" DOUBLE TOOTH CURVED 5 FORCEPS HEANEYS HYSTERECTOMY CLAMP 8" DOUBLE TOOTH STRAIGHT 5 FORCEPS HEANEYS HYSTERECTOMY CLAMP 8" DOUBLE TOOTH CURVED 5 FORCEPS HEANEYS HYSTERECTOMY CLAMP 8" DOUBLE TOOTH STRAIGHT 5 FORCEPS LANCE TISSUE 5" 20 FORCEPS LANCE TISSUE 6" 20 FORCEPS LANCE TISSUE 7" 10 FORCEPS RIGHT ANGLED LONG 1 FORCEPS RIGTH ANGLED MEDIUM 1 FORCEPS SINUS 6" 30 FORCEPS UTERUS HOLDING SOMER 4 FORCEPS UTERINE PACKING 10" 4 FORCEPS VOLSELLUM 20 FORCEPS VOLSELLUM 10" 88 FORCEPS VOLSELLUM 12" 2 FORCEPS RAMPLEY SPONGE HOLDNG FORCEPS BOTH END CURRETOR(VOLKMANSCOOP) 5PAIRS FISTULA DIRECTOR5" 5 FISTULA DIRECTOR6" 5 FG- 655-01-V BONNEY UTERINE POLYPUS FORCEPS 10.1/2 (267MM) 4 FG- 660-01-E VANT2 UTERINE POLYPUS FORCEPS 10.1/2 (267MM) 4 FG- 765-01-H NOVAK SUCTION CURETTE LUER 254MM(10") 2SET(1TO12

NO.)

FG- 768-01-V RANDALL SUCTION CURETTE 4MM DIA. LEUR 10" (254MM) 2SETS (1TO12NO.)

FG- 795-01-Y LEECH WIKINSON CERVICAL BIOPSY PUNCH 8" (203MM) SHAFT 2 FG- 815-01-M VAN DOREN CERVICAL BIOPSY PUNCH WITH CURVED BASKET JAW ANGLED SHANKS 267MM(10.1/2")

4

FG-185-01-E-SIMS DUCKBELL SPECULAM SINGLE ENDED CHROMIUM PLATED SMALL13/16" X3"(21MMX76MM)7"(17MM)

2

FG-185-01-E-SIMS DUCKBELL SPECULAM SINGLE ENDED CHROMIUM PLATED MEDIUM1.1/8" X3.3/8"(29MMX86MM)7"(178MM)

2

FG-185-01-E-SIMS DUCKBELL SPECULAM SINGLE ENDED CHROMIUM PLATED LARGE1.5/8" X3.3.1/2"(419MMX89MM)7"(178MM)

2

LEECH WILKINSON INTRAUTERINE CANNULA MEDIUM BASE OF CONE 1/2" (13MM)1.3/16"(30MM),10.1/2"(267MM)

2

I.U.D.REMOVING HOOK 12.1/2" 6 SCHUMAKER CERVICAL BIOPSY PUNCH 9.1/2 SHAFT 5 HSG CANNULA SMALL 4 HERNIA RING HOLDER 2

Page 132: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 132 of 166 01.09.2010

INSTRUMENT PIN 6 M.T.P.CANNULA SET OF FOUR 5 SET MAYO OLIVE STRIPPER COMPLETE SET 2 NEEDLE HOLDER 5" 15 NEEDLE HOLDER 6" 40 NEEDLE HOLDER 7" 55 NEEDLE HOLDER 8" 55 NEEDLE HOLDER FINE9" 5 PROCTOSCOPE LARGE 10 PROCTOSCOPE MEDIUM 10 PROCTOSCOPE SMALL 10 RESPIRATORY RIGHT 1 RESPIRATORY LEFT 1 RETRACTOR ABDOMINAL WALL 4 RETRACTOR CATSPAW TWO PRONG & THREE PRONG LARGE 1 RETRACTOR CATSPAW TWO PRONG & THREE PRONG SMALL 2 RETRACTOR CATSPAW TWO PRONG & THREE PRONG MEDIUM 12 RETRACTOR RIB 1 RETRACTOR CZERNEY DYONS 15 RETRACTOR CZERNEY 8" 20 RETRACTOR DEVARS 10 RETRACTOR JERNEYS 10 RETRACTOR JOLLS THYROID 2 RETRACTOR LONDON VAGINAL BLADE 3.5'(89MMX25MM) 2 RETRACTOR MASTOID SELF RETAINING 1 RETRACTOR ANTERIOR VAGINAL WALL 50 RETRACTOR CGARNIS 6" 15 RETRACTOR LANGENBACK 10 RETARCTOR SKIN D'S(SKIN HOOK) 2 SCHUMAKER CERICAL BIOPSY PUNCH 9.1/2"SHAFT 5 SCISORS POINTED 20 SCISSORS ROND MAYO 15 SCISSORS CURVED SMALL 20 SCISSORS CURVED MEDIUM 20 SCISSORS SHARP STRAIGHT 6 SCISSORS DRESSING 6" 5 SCISSOR DRESSING SHARP & BLUNT 5" 12 SCISSOR DRESSING SHARP & BLUNT 6" 10 SCISSOR DRESSING SHARP & BLUNT 7" 10 SCISSOR DRESSING SHARP & BLUNT 8" 10 SCISSOR DRESSING SHARP 5" 10 SCISSOR DRESSING SHARP 6" 15 SCISSOR DRESSING SHARP 7" 10 SCISSORS EPISIOTOMY(INDIAN) 15 SCISSORS FINE CURVED 3" 10 SCISSORS FINE CURVED 4" 10 SCISSORS FINE CURVED 6" 10 SCISSORS FINE CURVED 8" 10 SCISSORS GAUZE CUTING 10 SCISSORS MAYO'S CURVED 5" 10 SCISSORS MAYO'S CURVED 6" 30 SCISSORS MAYO'S CURVED 7" 20 SCISSORS MAYO'S CURVED 8" 5 SCISSORS MAYO'S STRAIGHT 5" 10

Page 133: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 133 of 166 01.09.2010

SCISSORS MAYO'S STRAIGHT 6" 25 SCISSORS MAYO'S STRAIGHT 7" 15 SCISSORS MAYO'S STRAIGHT 8" 15 SCISSORS METAZOLIC SMALL CURVED 10 SCISSORS METZENBAUM CURVED BLUNT TIP 6" 20 SCISSORS METZENBAUM CURVED BLUNT TIP 7" 20 SCISSORS METZENBAUM CURVED BLUNT TIP 8" 20 SCISSORS METZENBAUM CURVED BLUNT TIP 9" 10 SCISSORS METZENBAUM STRAIGHT POINTED 4" 20 SCISSORS METZENBAUM STRAIGHT POINTED 6" 30 SCISSORS METZENBAUM STRAIGHT POINTED 7" 20 SCISSORS METZENBAUM STRAIGHT POINTED 9" 10 SCISSORS METZENBAUM STRAIGHT POINTED 8" 20 SCISSORS METZENBAUM STRAIGHT CURVED BLUNT TIP 6" 20 SCISSORS METZENBAUM STRAIGHT CURVED BLUNT TIP 7" 20 SCISSORS METZENBAUM STRAIGHT CURVED BLUNT TIP 8" 20 SCISSORS CURVED SHARP 10 SCISSORS STITCH CUTTING 30 SCISSORS STRAIGHT ROUND POINT 30 SCISSORS STRAIGHT SHARP 10 SCISSORS SYMPATHECTOMY 4 SCISSORS TAILOR 2 SCISSORS TAILOR 10" 4 SCISSORS TAILOR 12" 20 SCISSORS TAILOR 6" 2 SHIRODKAR CERVICAL CANNULA MEDIUM 2 SHIRODKAR CERVICAL CANNULA SMALL 2 SKIN GRAFTING KNIFE BLADE HANDLE 5 SPECULUM RECTAL MEDIUM 5 SPECULUM CUSKOS 30 SPECULUM RECTAL SMALL 5 SPECULUM SIM'S LARGE 15 SPECULUM SIM'S MEDIUM 30 SPECULUM SIM'S SMALL 20 SKIN HOOKS 10 TOWEL CLIP 155 UTERINE SOUND 15 UTERINE ELEVATOR 20 SPINAL TRAY 3X4 20 CLAMPS BULLDOG CURVED 8" 2 CLAMPS BULLDOG STRAIGHT 8" 2 CLAMPS INTESTINAL CRUSHNG WITH RUBBER END10" 10 CLAMPS SATENSKYS 2 CURETTE SHARP & BLUNT 25 CLAMPS COOLIES 2 NOVAK SUCTION CURETTE LEUR254MM(10") 2 RANDALL SUCTION CURETTE 4MM DIA. LEUR 10" (254MM) 2 FG- 550-01-S SIMS UTERINE CURETTE PLIABLS SHAFT SHARP 10" (254MM) 4SETS SET OF 3 SIZES FG-551-01-W SMALL 1/2"(6MM) FG-551-02-Y MEDIUM 3/2"(10MM) FG-551-03-B LARGE 1/2"(13MM) BRISTOL-DIRECTOR PROBEAMD RASPATORY 7.25"(184MM) 2

Page 134: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 134 of 166 01.09.2010

MACDONALD DISSECTOR,DOUBLE-ENDED 7.5"(191MM) 2 FISTULA DIRECTOR WITH GUARD AND FRAENUMSLIT SIN (127MM) 2 BRODIC DIRECTO PROBE-ENDED MALLEABLE 6.5"(165MM) 2 WASTON CHYNE DISSECTOR FINE WITH PROBE7.5"(191MM) 2 TRUNES WARWICK RETRACTOR BLACK FINISH 2" DIAM*1.5" WIDE 2 (51MM*38MM) SWIFT JOLY RETRACTOR BLADE 1" *4"(25MM*102MM) 5 SWIFT JOLY RETRACTOR BLADE 1" *2.5"(25MM*64MM) 5 DEAVER ABDOMINAL RETRACTOS SET OF 5 (SCALE .25) 2 DURHAM RETRACTOR 3/8 *3/4"(10MM*19MM) BLADE 4 LONDON HOSPITAL RETRACTOR BLADE 3/4*11/4(19MM*32MM) 9"(229MM) 4 VOLKMANMM RETRACTOR 2 PRONG,BLUNT 4 3/8"*1/2"(10MM*13MM)8.5"(216MM) VOLKMANMM RETRACTOR 4 PRONG SHARP 4 VOLKMANMM RETRACTOR 4 PRONG,BLUNT 6 BACKHAUS TOWEL CLIP 3.5"(89MM) 20PAIRS MAYO SAFETY PIN FORCEPS HOLDER 4.5"(114MM) 25 ANEURYSM NEEDLE 8"(203MM) 3 JOLL ANEURYSM NEEDLE SHALLOW CURVE 8"(203MM) 2 NESTED TROCARS 2,5 AND 8 AND GANGE, MEKET SILVER CANNULAC, SET 2SETS OF 3" METEL CASE 6.5"(165MM) ADSON DISSECTING FORCEPS 6" 10 POOLE SUCTION TUBE SET WITH LAWRIE EXTENSION PIECE 16.5" 2 TINKLER (SIMPSON SMITH) SUCTION TUBE WITH ON/OFF VOLVE REMOVABLE PERFORATED OUTER SHEATH AND HANDLE 5…..M*10"(150MM*255MM)

2

TINKLER (SIMPSON SMITH) SUCTION TUBE WITH ON/OFF VOLVE REMOVABLE PERFORATED OUTER SHEATH AND HANDLE 5…..M*10"(150MM*255MM)

2

TINKER SUCTION TUBE 10" (254MM) CB 250 -08-Y (TINKER SUCTION TUBE NEEDLE)

2

SAVAGE INTESTTINAL DECOMPRESSOR, TROCAR AND CIIROMIUM PLATED 2 YANKUAR SUCTION TUBE CHROMIUM PLATED 12" (305MM) 12 LAHEY CHOLECYSTOMY FORCEPS CURVED LONGITUDIND SERRATIONS 4 7,,,,,"(197MM) MOYNIIIAN CHOLECYSTOMY FORCEPS 11.5"(292MM) 2 LOWERS CHOLECYSTOMY FORCEPS 8" (293MM) 2 DESJARDINES GALL STONE FORCEPS, SMALL,SCREW JOINT 8….(220MM) 2 DESJARDINES GALL STONE FORCEPS LARGE SCREW-JOINT 9…..'(238MM) 2 OSCHER GALL BLADDER TOCAR AND CANNULA WITHH SIDE SUCTION CONECTOR 6MM DIAM 9…"(234MM) OSCHER GALL BLADDER TOCAR AND CANNULA WITH SIDE SUCTION 2 CONECTOR 4.5MM DLAM 9…."(238MM) GRAEME ANDERSON RECTO PROTOSCOPY SPECULUM DA-001-01Y 2 GABRIEL RECTAL SPECULAMDA-010-01-A 2 NAUMTON MORGEN RECTAL SPECULUM DC -065-17-Q NAUNTON MORGEN RECTAL SPECULUM WITH ABTURALOR CHILD CHROMIUM 17MM "WITH FIBES LIGHT COMM.

2

DC-015-20-R WALTON RECTAL BIOPSY FORCPS WITH LATERALLY ANGLED SHANKS TO IMPROVE VISION 8"(203MM)

2

DC-020-25L 2 YEOMAN BIOPSY FORCEPS (BASKET JAWED) TO USE THROUGH SIGMOIDASCOPE 14"(355MM)

BARRON HAEMORRHOID LIGATOR SET COMPRISING LIGATOR 8.5" (216MM) 1 MC-GIVNEY HAEMMORRHOID SET 10.5" (272MM) STEM AND NANDLE 1 ORLAY HAEMORRHOID INSTRUMENT -DUAL PURPOSE FOR 1

Page 135: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 135 of 166 01.09.2010

LACKHART MUMMERY FISLULA DIRECTORS 6.5" (165MM) NO. 1. 1 STRAIGHT NO. 2.45*CURVE NO. 3.90*CURVE NO. 4. 4 RETROGRADE DD-130-14-A 2 ST. MARK'SBLACK,NYLON,HALLOW (9/16"*3/4") (14MM*19MM) 3.5" (89MM) DD-130-16-E 5/8"*1"(16MM*25MM) DD-150-38-B 2 LORD ANALDILATOR 11/2" (38MM) DIAM, 4.5"(105MM) HED DRICK KIDNEY PEDIC CLAMP 1*2 SERRATION DOUBLE ANGLE 10.5" (267MM)

2

MAYO CLINIC PEDICLE CLAMP SCREW JOINT 9,3/4"(248MM) 2 COOLEY RENAL ARTERY FORCEPS 7,3/4"(197MM) 2 RANDALL RENAL CALCULUS 2 FORCEPS SCREW JOINT TENESTRATED JAWS STIGHT CURVE NO.1,9.5" (230MM)

RANDALL RENAL CALCULUS FORCEPS SCREW JOINT FENES TRATED JAWS DEEP CURVE NO.2, 81/4 (208MM)

2

RANDALL RENAL CALCULUS FORCEPS SCREW JOINT 2 RANDALL RENAL CALCULUS FORCEPS TAKE APART JOINT 2 TENESTRATED JAWS DEEP RETRO CURVE NO. 4 7 1/8 TURNER WARWICK STONE FORCEP, SHALLOW CURVE FENESTRATED JAWS 7.5"(184MM)

2

EA-033-02-K 2 TURNER WARWICK RENAL STONE FORCEPS DEEP CURVE FENESTRATED JUAWS 7 1/4 "(184MM)

EA-042-01-J TINKER URETEROTITHO TOMY FORCEPS 45* BASKET JAW BIN (203MM)

2

EE-085-01-Y 2 INTRODUCER FOR FOLEY BALLOON CATHETERS 18" (457MM) EF-015-01-T TOOMY BLADDER SYRING GLASS WITH METAL NOZZLE 50 EF-100-01-X 2 SUPRAPUBLE TROCAR STRAIGHT WITH NICKEL SILVER CANNULA AND METAL HANDLE 25FG 4.5"(114MM)

EF-150-01-A 2 HYDRO CELL TOCAR (SILVER CANNULA IN METAL NANDLE NO.3 NG GAUGO (2.5MM DIAM) 6" (152 MM)

EH-044-04-C TURNER WAR-WICK NEEDLE HOLDER CURVED LEFT HAND MODEL FUNSTEN CURBIDE INSERT JAWS, BLACK FINISH 8.5"(216MM)

2

LONDON VAGINAL RETRACTOR BLADE 3.5"(89MM*25MM 2 FG-185-01-E 2 SIMS DUCKBILL SPECULAM SINGLE ENDED CHROMIUM PLATED, SMALL, 13/16"*3"(21MM*76MM) 7" (17MM)

FG-185-02-G 2 SIMS DUCKBILL SPECULAM SINGLE ENDED, CHROMIUM PLATED MEDIUM 1 1/8"*3 3/8" (29MM*86MM) 7" (178MM)

FG-185-03-J 2 SIMS DUCKBILL SPECULUM PLATED, LARGE 15/8"*3.5" (41MM*89MM) 7" (178MM)

FG-655-01-V 4 BONNEY UTERINE POLYPUS FORCEPS 10.5"(267MM)

Page 136: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 136 of 166 01.09.2010

FG-660-01-E 4 VANT UTERINE POLYPUS FORCEPS 10.5"(267MM) FG-765-01-V NAVAK SUCTION CURETTE FUER 254MM (10")

1 TO 12 NO. 2 SET

FG-768-01-V RANDALL SUCTION CURETTE 4MM DIAM, LUER, 10"(254MM)

1 TO 12 NO. 2 SET

FG-795-01-Y LEECH WIKINSON CERVICAL BIOPAY PUNCH 8"(203MM) SOFT 2 FG-815-01-M 2 VAN DOREN CERVICAL BIOPAY PUNCH WITH CURVED BASKET JAW ANGLED SHANKS 267MM(10.5")

DOYAN RETACTOR FIBRE OPTIC 1SET 2" 2.1/2" 3" 3.1/2 MORRIS RETRACTOR(FIBRE OPTIC) 4 CUSCOS SPECULUM (FIBRE OPTIC) SMALL 2SETS EACH

MEDIUM LARGE B.P.INSTRUMENT (PEDESTRAL) 20 DRESSING DRUM LARGE 20 DRESSING DRUM MEDIUM 25 DRESSING DRUM SMALL 15 S.STRAY BIG WITH COVER 35 S.S TRAY MEDIUM WITH COVER 45 S.S TRAY SMALL WITH COVER 30 STEEL TRAY BIG FOR RECEIVING BABY 4 STEEL BOWL LARGE 35 STEEL BOWL MEDIUM 70 STEEL BOWL SMALL 55 KIDNEY TRAY MEDIUM 60 KIDNEY TRAY SMALL 50 INHALATION KETTLE STEEL 10 ENEMA CAN STEEL 5 STEEL TRAY WITH COVER WITH STAND 10 STEEL BASIN 20 STEEL BASIN WITH STAND 30 STEEL BOWL WITH COVER SMALL 30 STEEL BOWL WITH COVER MEDIUM 30 STEEL BOWL WITH COVER BIG 35 ELECTRIC NEEDLE SYRINGE DESTROYER 8 WEIGHING MACHINE 8 SUCTION MACHINE 1/2HP, DOUBLE JAR 8 FORMALINE CHAMBER, LARGE 5 FORMALINE CHAMBER MEDIUM 2

Page 137: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 137 of 166 01.09.2010

Item No 93

Instruments for Dental Dept.

ADSON TISSUE FORCEPS TOOTHED 1 PLAIN 1 AMALGAM CARRIER 2 ARTERY FORCEPS-STRAIGHT 1 B.P.HANDLE NO.3 2 BALL BURNISHER 2 BONE CHISEL 1 BONECURETTE 1 BONE FILE 1 BONE NIBBLER (RONGEUR) 1 CEMENT SPATULA 2 CHIP SYRINGE 1 CHITLE FORCEPS 1 COMPOSITE FILLING INSTRUMENT (TEFLON) 2 COTTON HOLDER 2 COTTON WASTE RECEIVER 2 DIAMOND CARVER 1 DRESSING SCISSORS-STRAIGHT 1 ELLIS TISSUE FORCEPS 1 EXCAVATOR DISCOID NO. 131 & 132 2EACH EXCAVATOR SPOON NO. 220 & 221 2EACH EXTRACTION FORCEPS SET 1SET ADULT PAEDIATRIC GRACY CURETTE-ANTERIOR 1 POSTERIOR 1 GUTTA PERCHA PLUGGER 2 HAND SCALERS SET 1 HOLLEN BACK'S CARVER 1 MALLET (SURGICAL HAMMER) 1 MATRIX BANDS NO. 1 & 8 1PKT EACH MATRIX RETAINER NO. 1 & 8 1EACH MOSQUITO FORCEPS-STRAIGHT 2 MOUTH MIRROR HANDLES 12 MOUTH PROP 1 NEEDLE HOLDER 1 PARALLELOGRAM CONDENSER (PF151) 2 PERISOTEAL ELEVATORL 1DOZEN (12) PLASTIC INSTRUMENT (PF21) 2 PROBE-(EXPLORER)NO.5 12 PROBE-STRAIGHT NO.9 6 ROOT ELEVATOR SET(WINTER CRIER TYPE) 3 ROOT ELEVATOR(STRAIGHT) (HOSPITAL PATTERN) 6 ROUND CONDENSER (PF153) 6 SPIRIT LAMP 1 SUTURE CUTTING SCISSORS 2 TOOTH EXTRACTION FORCEPS SET A.ADULT SIZE 2SETS B.PAEDIATRIC SIZE 1SET

Page 138: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 138 of 166 01.09.2010

TWIZZERS 12 UNIVERSAL SCALERS 6 WARD'S CARVER 2

Item No 94

Instruments for Eye Implant

UNIVERSAL MICRO SPRING CORNEAL SCISSORS-TITANIUM 10NO. MICRO NEEDLE HOLDER-TITANIUM 6 NO. LIM'S CORNEAL FORCEPS-TITANIUM 10 NO. MACPHERSON'S FORCEPS-TITANIUM 10 NO. LENS DIALER-TITANIUM 6 NO. SUTURE TYING FOR CAPS WITH PLATFORM (STRAIGHT) -TITANIUM 10 NO. VANNAS MICRO SCISSORS 4 NO. ONE TOOTH FORCEPS 10 NO. SEMCO CANNULA FOR SUCTION & IRRIGATION 23 GAUGE 10 NO. UTRATA FORCEPS FOR CAPSALORHEXIS 6 NO. AIR INJECTION CANNULAS 24 NOS REVERSE CUTTING NEEDLES 12MM 24 NOS SPATULATED NEEDLES 12MM 24 NOS CONUNCTIVAL SCISSORS (SHARP TIP)(WESTCOTT TENOTOMY) STRAIGHT & CURVED

10 NOS EACH

SMALL EYE SCISSORS4.5" (MICRO) STRAIGHT & CURVED 8 NOS EACH

EPILATION FORCEPS (CILIA) 04NO. FOREIGN BODY SPUD 04NO. IRRIGATION CANNULA (BISHOP HARMAN) 23,24,26(G) 12NO. LACRIMAL SYRINGING CANNULA (BLUNT TIP) 23G 24NO. PUNCTUM DIALATOR (WILDER) SIZE 1, 2, 3 06NO. BIPOLAR CAUTERY FORCEPS (OPH USG) 06NO. "J" SHAPED I/A CANNULA 0.3MM ASPIRATION PORT 02NO. TOOTH FORCEP (CASTROVELE A SMALL ONE TOOTH WITH TYING PLATFORM 0.12MM)

06NO.

KALT'S NEEDLE HOLDER WITH LOCK STRAIGHT 10NO. KIRBY,S LENS EXPRESSOR HOOK 06NO. BINOCULAR LOUPE WITH HEAD BAND BINOMAG TYPE 04NO CHALAZION CLAMP 02NO. NEEDLE HOLDER FOR EYE SURGERY 02NO. LACRIMAL PROBE 02NO. FOREIGN BODY NEEDLE 02NO. RETROVERTED S.S NEEDLE NO 23 10NO. EYE SHIELD PLASTIC MODE 25NO. SHIOTZ TONOMETER 02NO. OPHTHALMOSCOPE BULB HEINE MIROFLEX-2 BULB NO 030 03NO.

Page 139: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 139 of 166 01.09.2010

GENERAL TECHNICAL SPECIFICATIONS

GENERAL POINTS:

1. Warranty:

a) Comprehensive Warranty as stated in GCC clause No. 15 (in Section – IV) for complete

equipment (including X ray tubes, HT Cable, Probes, Electrodes, Detectors, Batteries for

UPS, other vacuumatic parts wherever applicable) and Turnkey Work from the date of

satisfactory installation, commissioning, trial run & handing over of equipment to Hospital/

Dispensaries/ Institute/ Medical College.

b) 98% up time Warranty of complete equipment with extension of Warranty period by double

the downtime period on 24 (hrs) X 7 (days) X 365 (days) basis.

c) All software updates should be provided free of cost during Warranty period.

2. After Sales Service:

a) After sales service centre should be available at the city of Hospital/Institution/Medical

College on 24 (hrs) X 7 (days) X 365 (days) basis. Complaints should be attended properly,

maximum within 8 hrs. The service should be provided directly by Tenderer/Indian Agent.

Undertaking by the Principals that the spares for the equipment shall be available for at least

10 years from the date of supply.

b) The service provider should have the necessary equipments recommended by the

manufacturer to carry out preventive maintenance test as per guidelines provided in the

service/maintenance manual.

3. Training:

On Site training to Doctors/ Technicians/ staff is to be provided by Principal/ Indian Agents (if

they have the requisite know-how) for operation and maintenance of the equipment to the

satisfaction of the consignee.

4. Annual Comprehensive Maintenance Contract (CMC) of subject equipment with Turnkey:

a) The cost of Annual Comprehensive Maintenance Contract (CMC) which includes

preventive maintenance including testing & calibration as per technical/ service /operational

manual of the manufacturer, labour and spares, after satisfactory completion of Warranty

period to be quoted for next 5 years on yearly basis for complete equipment (including X

ray tubes, HT Cable, Probes, Electrodes, Detectors, Batteries for UPS, other vacuumatic parts

wherever applicable) and Turnkey (if any). The supplier shall visit each consignee site as

recommended in the manufacturer’s technical/ service /operational manual, but at least once

in six months during the CMC period

b) The cost of CMC to be quoted along with taxes applicable on the date of Tender Opening.

The taxes to be paid extra, to be specifically stated. In the absence of any such stipulation the

price will be taken inclusive of such taxes and no claim for the same will be entertained later.

c) Cost of CMC will be added for Ranking/Evaluation purpose.

d) The payment of CMC will be made on six monthly basis after satisfactory completion of said

period, duly certified by end user on receipt of bank guarantee for 2.5 % of the cost of the

equipment as per Section XV valid till 2 months after expiry of entire CMC period.

e) There will be 98% uptime warranty during CMC period on 24 (hrs) X 7 (days) X 365 (days)

basis, with penalty, to extend CMC period by double the downtime period.

f) During CMC period, the supplier is required to visit at each consignee’s site at least once in 6

months commencing from the date of the successful completion of warranty period for

preventive maintenance of the goods.

g) All software updates should be provided free of cost during CMC.

Page 140: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Technical Specifications

HLL/PCD/ESIC-20/10-11 Page 140 of 166 01.09.2010

h) Failure of the above [4. e) to 4. g)] by the supplier, may lead to the forfeiture of the Bank

Guarantee for Annual CMC.

i) The payment of CMC will be made as stipulated in GCC Clause 21.

5. Turnkey:

Turnkey is indicated in the technical specification of the respective items, wherever required.

The Tenderer shall examine the existing site where the equipment is to be installed, in

consultation with HOD of Hospital/Institution/Medical College concerned. Turnkey details of

each Hospital/Institution/Medical College are given at the end of Technical Specification. The

Tenderers to quote prices indicating break-up of prices of the Machine and Turnkey Job of each

Hospital/Institution/Medical College. The Turnkey costs (wherever applicable) may be

quoted in Indian Rupee will be added for Ranking Purpose.

The taxes to be paid extra, to be specifically stated. In the absence of any such stipulation the

price will be taken inclusive of such duties and taxes and no claim for the same will be

entertained later.

The Turnkey Work should completely comply with AERB requirement, if any.

Page 141: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited QC Requirement

HLL/PCD/ESIC-20/10-11 Page 141 of 166 01.09.2010

Section – VIII

Quality Control Requirements (Proforma for equipment and quality control employed by the manufacturer(s)

Tender Reference No.

Date of opening

Time

Name and address of the Tenderer:

Note: All the following details shall relate to the manufacturer(s) for the goods quoted for.

01 Name of the manufacturer

a. full postal address

b. full address of the premises

c. Email ID

d. telephone number

e. fax number

02 Plant and machinery details

03 Manufacturing process details

04 Monthly (single shift) production capacity of goods quoted for

a. normal

b. maximum

05 Total annual turn-over (value in Rupees)

06 Quality control arrangement details

a. for incoming materials and bought-out components

b. for process control

c. for final product evaluation

07 Test certificate held

a. type test

b. BIS/ISO certification

c. any other

08 Details of staff

a. technical

b. skilled

c. unskilled

Signature and seal of the Tenderer

Page 142: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Qualification Criteria

HLL/PCD/ESIC-20/10-11 Page 142 of 166 01.09.2010

Section – IX

Qualification Criteria

01. The Tenderer must be a Manufacturer or its authorized Agent.

02. (a) The Manufacturer should have supplied and installed in last Five years from the date of

Tender Opening, atleast 100% of the quoted quantity of the similar equipment meeting major

specification parameters which is functoning satisfactorily. The foreign Manufacturer

satisfying the above criteria should also have supplied and installed in last Five years from

the date of Tender Opening, at least 50% (or one No. where the schedule of requirement is

one no.) of quoted quantity of similar model which is functoning satisfactorily any where

outside the country of manufacture.

02. (b) The Tenderers quoting as authorized representative of the manufacturer meeting the above

criteria 02 (a) should have supplied and installed in last Five years from the date of Tender

Opening, atleast 50% (or one No. where the schedule of requirement is one no. ) of the

quoted quantity of similar equipment which is functoning satisfactorily, any where in India

of the same manufacturer.

Note:

1. In support of 2 (a) & 2 (b), the Tenderer shall furnish Performance statement in the enclosed

Proforma ‘A’.

The manufacturer as well as the Tenderer/ Indian Agent shall furnish Satisfactory Performance

Certificate in respect of above, duly translated in English and duly notarized, alongwith the

tender. The performance certificate should be in accordance with the requirement of clause 2 (a) /

2 (b) as stated above.

2. The Tenderer shall furnish a brief write-up, packed with adequate data explaining and establishing

his available capacity/capability (both technical and financial) to perform the Contract (if

awarded) within the stipulated time period, after meeting all its current/present commitments. The

Tenderer shall also furnish details of Equipment and Quality Control in the enclosed Section VIII.

3. Notwithstanding anything stated above, the Purchaser reserves the right to assess the Tenderer’s

capability and capacity to perform the contract satisfactorily before deciding on award of

Contract, should circumstances warrant such an assessment in the overall interest of the

Purchaser.

4. The Purchaser reserves the right to ask for a free demonstration of the quoted equipment at a pre

determined place acceptable to the purchaser for technical acceptability as per the tender

specifications, before the opening of the Price Tender.

Page 143: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Qualification Criteria

HLL/PCD/ESIC-20/10-11 Page 143 of 166 01.09.2010

PROFORMA ‘A’

PROFORMA FOR PERFORMANCE STATEMENT

(For the period of last five years)

Tender Reference No.: _________________________________

Date & Time of opening: _________________________________

Name and address of the Tenderer: _________________________________

Name and address of the manufacturer: _________________________________

Date of

completion of

Contract

Ord

er p

lace

d b

y

(fu

ll a

dd

ress

of

Pu

rch

ase

r/C

on

sig

nee

)

Ord

er n

um

ber

and

dat

e

Des

crip

tio

n o

f

ord

ere

d g

oo

ds

and

ser

vic

es

Qu

anti

ty o

f

ord

ere

d g

oo

ds

and

ser

vic

es

Val

ue

of

ord

er

(Rs.

)

As

per

con

trac

t

Act

ual

Remarks

indicating

reasons for

delay if

any

Have the

goods been

functioning

Satisfactorily

(attach

documentary

proof)**

1 2 3 4 5 6 7 8 9

Signature and seal of the Tenderer

** The documentary proof will be a certificate from the consignee/end user with cross-

reference of order no. and date in the certificate duly notarised certification authenticating

the correctness of the information furnished. If at any time, information furnished is proved

to be false or incorrect, the earnest money and or performance security furnished will be

forfeited. Such certificates from a third party or middleman other than actual end user will

not be accepted. The satisfactory performance implies working satisfactorily without any

complaint since the date of installation, commissioning & handing over to the end user as

per the standard format enclosed.

Page 144: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Tender Form

HLL/PCD/ESIC-20/10-11 Page 144 of 166 01.09.2010

Section – X

TENDER FORM

Date__________

To,

Head (P & CD)

HLL Lifecare Limited

Procurement and Consultancy Division

B-14A, Sector -62, Noida -201307, Uttar Pradesh

Ref. Your TE document No. _________dated ___________

We, the undersigned have examined the above mentioned TE document, including

amendment/corrigendum No. __________, dated ________ (if any), the receipt of which is hereby

confirmed. We now offer to supply and deliver___________ (Description of goods and services) in

conformity with your above referred document.

If our tender is accepted, we undertake to supply the goods and perform the services as mentioned

above, in accordance with the delivery schedule specified in the List of Requirements.

We further confirm that, if our tender is accepted, we shall provide you with a performance security of

required amount in an acceptable form in terms of GCC clause 5, read with modification, if any, in

Section - V – “Special Conditions of Contract”, for due performance of the contract.

We agree to keep our tender valid for acceptance as required in the GIT clause 20, read with

modification, if any in Section - III – “Special Instructions to Tenderers” or for subsequently extended

period, if any, agreed to by us. We also accordingly confirm to abide by this tender up to the aforesaid

period and this tender may be accepted any time before the expiry of the aforesaid period. We further

confirm that, until a formal contract is executed, this tender read with your written acceptance thereof

within the aforesaid period shall constitute a binding contract between us.

We further understand that you are not bound to accept the lowest or any tender you may receive

against your above-referred tender enquiry.

We confirm that we do not stand deregistered/banned/blacklisted by any Govt. Authorities.

We confirm that we fully agree to the terms and conditions specified in above mentioned TE

document, including amendment/ corrigendum if any

(Signature with date)

(Name and designation) Duly authorised to sign tender for and on behalf of

Page 145: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Price Schedule

HLL/PCD/ESIC-20/10-11 Page 145 of 166 01.09.2010

SECTION – XI PRICE SCHEDULE

A) PRICE SCHEDULE FOR DOMESTIC GOODS OR GOODS OF FOREIGN ORIGIN LOCATED WITHIN INDIA

1 2 3 4 5 6

Price per unit (Rs.)

Item

Sl.

No.

Brief Description

of Goods (with

make & model)

Country of

Origin

Quantity

(Nos.)

Ex - factory/ Ex -

warehouse /Ex-

showroom /Off -

the shelf

(a)

Excise

Duty (if

any)

[%age &

value]

(b)

Sales Tax/

VAT (if

any)

[%age &

value]

(c)

Transportation,

loading/

unloading and

Incidental costs

till consignee’s

site

(d)

Insurance charges for

a period including 3

months beyond the

date of delivery

(e)

Incidental Services

(including Installation &

Commissioning,

Supervision,

Demonstration and

Training) at the

Consignee’s site

(f)

Unit Price (at

Consignee

Site) basis

(g)

=a+b+c+d+e+f

Total Price (at

Consignee

Site) basis

(Rs.)

4 x 5(g)

Total Tender price in Rupees: ___________________________________________________________________________________________________

In words: ____________________________________________________________________________________________________________________________

Note: - 1. If there is a discrepancy between the unit price and total price THE UNIT PRICE shall prevail.

2. The charges for Annual CMC after warranty shall be quoted separately as per Section-XI – Price Schedule C

Name________________________

Business Address________________________

Place: ___________________________ Signature of Tenderer________________________

Date: _________________________ Seal of the Tenderer________________________

Page 146: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Price Schedule

HLL/PCD/ESIC-20/10-11 Page 146 of 166 01.09.2010

B) PRICE SCHEDULE FOR GOODS TO BE IMPORTED FROM ABROAD

** to be quoted in Indian Currency

Total price at Consignee’s site

(A) In foreign currency : column (4 x e)____________________________________________________________________________ (In figures and words) plus

** (B) In Indian Rupees : column 4 x (b+f+g+h+i) Rs__________________________________________________________________(In figures and words)

________________________ _______________________________________________________________________________

Note: - 1. The Tenderer will be fully responsible for the safe arrival of the goods at the consignee site in good condition as per terms of contract.

2. The bidders break up of prices under various columns is for comparison of prices up to delivery of goods at consignee’s site for tender evaluation.

3. The quoted price should be supported with original proforma invoice from the foreign manufacturers. The proforma invoice should indicate the percentage of agency commission included in the FOB

prices. Indian Agent to be paid in Indian Currency.

4. All the components of the DDP price will be paid by the tenderer. The purchaser will make the payment of DDP price after receipt of goods at consignee’s site in good condition as per payment terms

in the contract.

5. The prices quoted in foreign currency in column (e) shall be converted in Rupees at the selling rate of exchange applicable on the date of tender opening .The customs duty amount so worked out as

percentage of net CIP value in rupees will be taken for evaluation and comparison of tenders

6. The charges for Annual CMC after warranty shall be quoted separately as per Section-XI – Price Schedule C

Name_____________________________

Business address____________________

Place:________________________________________ Signature of Tenderer___________________

Date:_____________________________________ Seal of Tenderer________________________

1 2 3 4 5

Price per unit (Currency)

Unit price on DDP

basis at consignee’s

site

Item

Sl.

No.

Brief

Description of

Goods (with

make &

model)

Country

of Origin

Quantity

(Nos.)

Gross FOB price

at sea/air port of

Lading (inclusive

of Agency

Commission)

Amount and

percentage of

Agency

Commission

**

Net FOB

(excluding

Agency

Commission)

(a-b)

Insurance

& Freight

Net

CIP by

Air/

Sea at

the

port of

entry

(c+d)

Custom

Duty

amount as

% of Net

CIP

(amount

with

CDEC as

applicable)

**

Custom

Clearance

&

Handling

Charges

**

Loading/

Unloading,

inland

transportation,

insurance as

per Clause 11

of GCC &

incidental cost

till consignee’s

site

**

Installation

commissioning

, supervision.

Demonstration

& training at

the

consignee’s

site

**

In foreign

currency

In Indian

Rupees

(a) (b) (c) (d) (e) (f) (g) (h) (i) =(e)

=(b+f+g+

h+i)

Page 147: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Price Schedule

HLL/PCD/ESIC-20/10-11 Page 147 of 166 01.09.2010

C) PRICE SCHEDULE FOR COMPREHENSIVE MAINTENANCE CONTRACT AFTER WARRANTY PERIOD

1 2 3 4 5

Comprehensive Maintenance Contract

Cost for Each Unit year wise*.

1st 2

nd 3

rd 4

th 5

th

Item

Sl.

No.

Brief Description of the

Goods

Quantity

(Nos.)

a b c d e

Total Comprehensive Maintenance

Contract Cost for 5 (or as specified) Years

[3 x (4a+4b+4c+4d+4e)]

* After completion of Warranty period

NOTE:- 1. In case of discrepancy between unit price and total prices, THE UNIT PRICE shall prevail.

2. The cost of Comprehensive Maintenance Contract (CMC) which includes preventive maintenance including testing & calibration as per technical/

service /operational manual and labour, after satisfactory completion of Warranty period may be quoted for next 5 (or as specified) years on yearly

basis for complete equipment and Turnkey (if any).

3. The cost of CMC may be quoted along with taxes applicable on the date of Tender Opening. The taxes to be paid extra, to be specifically stated. In

the absence of any such stipulation the price will be taken inclusive of such taxes and no claim for the same will be entertained later.

4. Cost of CMC will be added for Ranking/Evaluation purpose.

5. The payment of CMC will be made as per clause GCC clause 21.1 (D).

6. The uptime warranty will be 98 % on 24 (hrs) x 7 (days) x 365 (days) basis or as stated in Technical Specification of the TE document.

7. All software updates should be provided free of cost during CMC period.

8. The stipulations in Technical Specification will supersede above provisions

9. The supplier shall keep sufficient stock of spares required during Comprehensive Maintenance Contract period. In case the spares are required to

be imported, it would be the responsibility of the supplier to import and get them custom cleared and pay all necessary duties.

Name________________________

Business Address________________________

Place: ___________________________ Signature of Tenderer________________________

Date: _________________________ Seal of the Tenderer________________________

Page 148: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Price Schedule

HLL/PCD/ESIC-20/10-11 Page 148 of 166 01.09.2010

D) PRICE SCHEDULE FOR TURNKEY

Item

Sl. No.

BRIEF TURNKEY DESCRIPTION

OF GOODS

CONSIGNEE

CODE

Turnkey price

Note: -

1. The cost of Turnkey as per Technical Specification (Section VII) may be quoted on lump sum along with taxes applicable on the date of Tender

Opening. The taxes to be paid extra, to be specifically stated. In the absence of any such stipulation the price will be taken inclusive of such taxes and no

claim for the same will be entertained later.

2. Cost of Turnkey will be added for Ranking/Evaluation purpose.

3. The payment of Turnkey will be made as per clause GCC clause 21.1 (c).

4. The stipulations in Technical Specification will supersede above provisions

Name________________________

Business Address________________________

Place: ___________________________ Signature of Tenderer________________________

Date: _________________________ Seal of the Tenderer________________________

Page 149: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Limited Questionnaire

HLL/PCD/ESIC-20/10-11 Page 149 of 166 01.09.2010

SECTION – XII

QUESTIONNAIRE

Fill up the Section XX – Check List for Tenderers and enclose with the Tender

1. The tenderer should furnish specific answers (alongwith mention of relevant page nos. of

tender) to all the questions/issues mentioned in the Checklist. In case a question/issue does not

apply to a tenderer, the same should be answered with the remark “not applicable”

2. Wherever necessary and applicable, the tenderer shall enclose certified copy as documentary

proof/ evidence to substantiate the corresponding statement.

3. In case a tenderer furnishes a wrong or evasive answer against any of the question/issues

mentioned in the Checklist, its tender will be liable to be ignored.

Page 150: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 150 of 166 01.09.2010

SECTION – XIV

MANUFACTURER’S AUTHORISATION FORM

To,

Head (P & CD) HLL Lifecare Limited

Procurement and Consultancy Division

B-14A, Sector -62, Noida -201307, Uttar Pradesh

Dear Sir,

Ref. Your TE document No ____________, dated _____________

We, ___________________________________ who are proven and reputable manufacturers

of___________________________(name and description of the goods offered in the tender) having

factories at___________________________________________________________, hereby authorise

Messrs______________________________(name and address of the agent) to submit a tender,

process the same further and enter into a contract with you against your requirement as contained in the

above referred TE documents for the above goods manufactured by us.

We further confirm that no supplier or firm or individual other than Messrs.

________________________ (name and address of the above agent) is authorised to submit a tender,

process the same further and enter into a contract with you against your requirement as contained in the

above referred TE documents for the above goods manufactured by us.

We also hereby extend our full warranty, CMC as applicable as per clause 15 of the General Conditions of Contract, read with modification, if any, in the Special Conditions of Contract for the

goods and services offered for supply by the above firm against this TE document.

Yours faithfully,

[Signature with date, name and designation]

for and on behalf of Messrs___________________________

[Name & address of the manufacturers]

Note: 1. This letter of authorisation should be on the letterhead of the manufacturing firm and should

be signed by a person competent and having the power of attorney to legally bind the

manufacturer.

2. Original letter may be sent.

Page 151: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 151 of 166 01.09.2010

SECTION – XV

BANK GUARANTEE FORM FOR PERFORMANCE SECURITY/ CMC

SECURITY

To

Head of Hospital/Institute/Medical College of ESIC

WHEREAS _____________________________ (Name and address of the supplier) (Hereinafter

called “the supplier”) has undertaken, in pursuance of contract no________________________ dated

_____________ to supply (description of goods and services) (herein after called “the contract”).

AND WHEREAS it has been stipulated by you in the said contract that the supplier shall furnish you

with a bank guarantee by a scheduled commercial bank recognised by you for the sum specified

therein as security for compliance with its obligations in accordance with the contract;

AND WHEREAS we have agreed to give the supplier such a bank guarantee;

NOW THEREFORE we hereby affirm that we are guarantors and responsible to you, on behalf of the

supplier, up to a total of. ________________________ (Amount of the guarantee in words and

figures), and we undertake to pay you, upon your first written demand declaring the supplier to be in

default under the contract and without cavil or argument, any sum or sums within the limits of

(amount of guarantee) as aforesaid, without your needing to prove or to show grounds or reasons for

your demand or the sum specified therein.

We hereby waive the necessity of your demanding the said debt from the supplier before presenting

us with the demand.

We further agree that no change or addition to or other modification of the terms of the contract to be

performed there under or of any of the contract documents which may be made between you and the

supplier shall in any way release us from any liability under this guarantee and we hereby waive

notice of any such change, addition or modification.

This guarantee shall remain valid till 2 (two) months beyond the warranty period, i.e up to -----------

(indicate date)

…………………………….

(Signature with date of the authorised officer of the Bank)

………………………………………………………….

Name and designation of the officer

………………………………………………………….

………………………………………………………….

Seal, name & address of the Bank and address of the Branch

Page 152: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 152 of 166 01.09.2010

SECTION – XVI

CONTRACT FORM - A

CONTRACT FORM FOR SUPPLY, INSTALLATION, COMMISSIONING, HANDING

OVER, TRIAL RUN, TRAINING OF OPERATORS & WARRANTY OF GOODS

(Address of the Purchaser’s/Consignee’s

office issuing the contract)

Contract No___________ dated______________

This is in continuation to this office’s Notification of Award No_______ dated ______ 1. Name & address of the Supplier: ______________________________

2. Purchaser’s TE document No________ dated____________ and subsequent Amendment

No____________, dated_________ (if any), issued by the purchaser

3. Supplier’s Tender No_________ dated__________ and subsequent communication(s)

No____________ dated _________ (if any), exchanged between the supplier and the purchaser in

connection with this tender.

4. In addition to this Contract Form, the following documents etc, which are included in the documents

mentioned under paragraphs 2 and 3 above, shall also be deemed to form and be read and construed

as integral part of this contract:

(i) General Conditions of Contract;

(ii) Special Conditions of Contract;

(iii) List of Requirements;

(iv) Technical Specifications;

(v) Quality Control Requirements;

(vi) Tender Form furnished by the supplier;

(vii) Price Schedule(s) furnished by the supplier in its tender;

(viii) Manufacturers’ Authorisation Form (if applicable for this tender);

(ix) Purchaser’s Notification of Award

Note : The words and expressions used in this contract shall have the same meanings as are

respectively assigned to them in the conditions of contract referred to above. Further, the definitions

and abbreviations incorporated under clause 1 of Section II – ‘General Instructions to Tenderers’ of

the Purchaser’s TE document shall also apply to this contract.

5. Some terms, conditions, stipulations etc. out of the above-referred documents are reproduced below

for ready reference:

(i) Brief particulars of the goods and services which shall be supplied/ provided by the supplier

are as under:

Item

Sl. No.

Brief description of goods/services

Accounting unit

Quantity to be supplied

Unit Price

Total price

Terms of delivery

Page 153: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 153 of 166 01.09.2010

Any other additional services (if applicable) and cost thereof: _____________________

Total value (in figure) ____________ (In words) ___________________________

(ii) Delivery schedule

(iii) Details of Performance Security

(iv) Quality Control

(a) Mode(s), stage(s) and place(s) of conducting inspections and tests.

(b) Designation and address of purchaser’s inspecting officer

(v) Destination and despatch instructions

(vi) Consignee, including port consignee, if any

(vii) Warranty clause

(viii) Payment terms

(ix) Paying authority

____________________________

(Signature, name and address

of the Purchaser’s/Consignee’s authorised official)

For and on behalf of__________________ Received and accepted this contract

(Signature, name and address of the supplier’s executive

duly authorised to sign on behalf of the supplier)

For and on behalf of _________________________

(Name and address of the supplier)

(Seal of the supplier)

Date: _________________________

Place: _________________________

Page 154: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 154 of 166 01.09.2010

CONTRACT FORM – B

CONTRACT FORM FOR COMPREHENSIVE MAINTENANCE CONTRACT

Comprehensive Maintenance Contract No._______________________dated_________________

Between

(Address of Head of Hospital/Institute/Medical College)

And

(Name & Address of the Supplier)

Ref: Contract No___________ dated______________ (Contract No. & date of Contract for

supply, installation, commissioning, handing over, Trial run, Training of operators &

warranty of goods)

In continuation to the above referred contract

a) The Contract of Comprehensive Maintenance is hereby concluded as under: -

1 2 3 4 5 Comprehensive

Maintenance Contract

Cost for Each Unit year

wise*.

1st 2

nd 3

rd 4

th 5

th

Item

Sl. No.

BRIEF

DESCRIPTION OF

GOODS

Quantity

(Nos.)

a b c d e

Total Comprehensive

Maintenance Contract

Cost for 5 (or as

specified) Years

[3 x (4a+4b+4c+4d+4e)]

Total value (in figure) ____________ (In words) ___________________________

b) The CMC commence from the date of expiry of all obligations under Warranty i.e.

from______________ (date of expiry of Warranty) and will expire on ______________ (date

of expiry of CMC)

c) The cost of Comprehensive Maintenance Contract (CMC) which includes preventive

maintenance and labour, after satisfactory completion of Warranty period may be quoted for

next 5 (or as specified) years as contained in the above referred contract on yearly basis for

complete equipment (including X ray tubes, Batteries for UPS, other vacummatic parts, _____

& _____) and Turnkey (if any).

d) There will be 98% uptime warranty during CMC period on 24 (hrs) X 7 (days) X 365 (days)

basis, with penalty, to extend CMC period by double the downtime period.

e) During CMC period, the supplier shall visit at each consignee’s site for preventive maintenance

including testing and calibration as per the manufacturer’s service/ technical/ operational

manual. The supplier shall visit each consignee site as recommended in the manufacturer’s

manual, but at least once in 6 months commencing from the date of the successful completion

of warranty period for preventive maintenance of the goods.

f) All software updates should be provided free of cost during CMC.

g) The bank guarantee valid till ______________ [(fill the date) 2 months after expiry of entire

CMC period] for an amount of Rs. _______________ [(fill amount) equivalent to 2.5 % of the

cost of the equipment as per contract] shall be furnished in the prescribed format given in

Page 155: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 155 of 166 01.09.2010

Section XV of the TE document, along with the signed copy of CMC within a period of 21

(twenty one) days of issue of CMC failing which the proceeds of Performance Security shall be

payable to the Purchaser/Consignee.

h) If there is any lapse in the performance of the CMC as per contract, the proceeds CMC bank

guarantee for an amount of Rs. __________ (equivalent to 2.5 % of the cost of the

equipment as per contract) shall be payable to the Consignee.

i) Payment terms: The payment of CMC will be made against the bills raised to the

consignee by the supplier on six monthly basis after satisfactory completion of said period,

duly certified by the HOD concerned. The payment will be made in Indian Rupees.

j) Paying authority: ______________________ (name of the consignee i.e. Hospital/

Institute /Medical College’s authorised

official)

____________________________

(Signature, name and address of

Hospital/Institute/Medical College’s authorised official)

For and on behalf of__________________ Received and accepted this contract

(Signature, name and address of the supplier’s executive

duly authorised to sign on behalf of the supplier)

For and on behalf of _________________________

(Name and address of the supplier)

(Seal of the supplier)

Date: _________________________

Place: _________________________

Page 156: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 156 of 166 01.09.2010

SECTION – XVII

CONSIGNEE RECEIPT CERTIFICATE

(To be given by consignee’s authorized representative)

The following store (s) has/have been received in good condition:

1) Contract No. & date :______________________________

2) Supplier’s Name :______________________________

3) Consignee’s Name & Address with

telephone No. & Fax No. : ______________________________

4) Name of the item supplied :______________________________

5) Quantity Supplied :______________________________

6) Date of Receipt by the Consignee :______________________________

7) Name and designation of Authorized

Representative of Consignee :______________________________

8) Signature of Authorized

Representative of Consignee with

date :______________________________

9) Seal of the Consignee :______________________________

Page 157: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 157 of 166 01.09.2010

SECTION – XVIII

Proforma of Final Acceptance Certificate by the Consignee

No _______________ Date_______________

To

M/s____________________

_______________________

_______________________

Subject: Certificate of commissioning of equipment/plant.

This is to certify that the equipment(s)/plant(s) as detailed below has/have been received in

good conditions along with all the standard and special accessories and a set of spares

(subject to remarks in Para no.02) in accordance with the contract/technical specifications.

The same has been installed and commissioned.

(a) Contract No______________________________________ dated__________________

(b) Description of the equipment(s)/plants: _______________________________________

(c) Equipment(s)/ plant(s) nos.: ________________________________________________

(d) Quantity:_______________________________________________________________

(e) Bill of Loading/Air Way Bill/Railway Receipt/ Goods Consignment Note no___________________ dated _________________

(f) Name of the vessel/ Transporter: ___________________________________________

(g) Name of the Consignee: ___________________________________________________

(h) Date of commissioning and proving test: _____________________________________

Details of accessories/spares not yet supplied and recoveries to be made on that account.

Sl. Description of Item Quantity Amount to be recovered

No.

The proving test has been done to our entire satisfaction and operators have been trained to operate

the equipment(s)/plant(s).

The supplier has fulfilled its contractual obligations satisfactorily ## or

The supplier has failed to fulfil its contractual obligations with regard to the following:

He has not adhered to the time schedule specified in the contract in dispatching the

documents/drawings pursuant to ‘Technical Specifications’.

He has not supervised the commissioning of the equipment(s)/plant(s)in time, i.e. within the period

specified in the contract from date of intimation by the Purchaser/Consignee in respect of the

installation of the equipment(s)/plant(s).

The supplier as specified in the contract has not done training of personnel.

The extent of delay for each of the activities to be performed by the supplier in terms of the contract

Page 158: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Various Forms

HLL/PCD/ESIC-20/10-11 Page 158 of 166 01.09.2010

is______________________________________________________________________________.

The amount of recovery on account of non-supply of accessories and spares is given under Para

no.02.

The amount of recovery on account of failure of the supplier to meet his contractual obligations

is__________________ (here indicate the amount).

Signature

Name:

Designation with stamp

## Explanatory notes for filling up the certificate:

o He has adhered to the time schedule specified in the contract in dispatching the

documents/drawings pursuant to ‘Technical Specification’.

o He has supervised the commissioning of the equipment(s)/plant(s) in time, i.e. within the

time specified in the contract from date of intimation by the Purchaser/Consignee in respect

of the installation of the equipment(s)/plant(s).

o Training of personnel has been done by the supplier as specified in the contract

o In the event of documents/drawings having not been supplied or installation and

commissioning of the equipment(s)/plant(s) having been delayed on account of the supplier,

the extent of delay should always be mentioned in clear terms.

Page 159: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Shipping Arrangement

HLL/PCD/ESIC-20/10-11 Page 159 of 166 01.09.2010

SECTION – XIX

DETAILS OF SHIPPING ARRANGEMENT FOR LINER CARGOES IN RESPECT OF

C&F/CIF/TURNKEY/F.O.R CONTRACTS FOR IMPORTS

1. SHIPMENT FROM PORTS OF U.K INCLUDING NORTHERN IRELAND (ALSO EIRE), FROM

THE NORTH CONTINENT OF EUROPE (GERMANY, HOLLAND, BELGIUM, FRANCE,

NORWAY, SWEDEN, DENMARK, FINLAND AND PORTS ON THE CONTINENTAL SEABOARD

OF MEDITERRANIAN (I.E. FRENCH WESTERN ITALIAN PORTS), TO PORTS IN INDIA.

The Seller should arrange shipment of the goods by vessels belonging to the member lines of the India-

Pakistan-Bangladesh Conference. If the Seller finds that the space on the ‘Conference Lines’ vessels is

not available for any specific shipment, he should take up with India-Pakistan-Bangladesh Conference.

Conferity House, East Grinstead, Sussex (UK), for providing shipping space and also inform the Shipping

Co-ordination Officer, Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW

DELHI, Telex: VAHAN IN – 031 – 61157, 61158, 61159)

The Seller should arrange shipment through the Government of India’s Forwarding Agents, M/s Schenker

& Co., 2000-Hamburg (Cable: SCHENKER CO., HAMBURG) OR obtain a certificate from them to the

effect that shipment has been arranged in accordance with instructions of the Ministry of Surface

Transport, (TRANSCHART), New Delhi.

2. SHIPMENT FORM PORTS OF U.K. INCLUDING NORTHERN

Goods under this contract would be shipped by the national shipping companies of the Contracting Parties

operating bilateral shipping service and vessels under the flag of third countries in accordance with the

Agreement between the Government of German Democratic Republic and the Government of the Republic of India in the Field of Merchant Shipping signed on 9.1.1979, as amended up-to-date.

3. ISHIPMENT FROM ADRIATIC PORTS OF EASTERN ITALY AND YUGOSLAVIA

The seller should arrange shipment of the goods by vessels belonging to the following Indian member

lines;

1. The Shipping Purchaser of India Ltd.

2. The Scindia Steam Navigation Co., Ltd

3. India Steamship Co., Ltd

For the purpose of ascertaining the availability of suitable Indian vessels and granting dispensation in the

event of their non-availability, the Seller should give adequate notice about the readiness of each

consignment from time to time at least six weeks in advance of the required position to M/s Schenker &

Co. 2000 HAMBURG (Cable: SCHENKER CO., HAMBURG) and also endorse a copy thereof to the

Shipping Co-ordination Officer, Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART,

NEW DELHI, Telex: VAHAN IN – 031 – 61157, 61158, 61159)

The seller should arrange shipment through the Government of India’s Forwarding Agents M/s Schenker

& Co. 2000 HAMBURG (Cable: SCHENKER CO., HAMBURG) or obtain certificate from them to the

effect that shipment has been arranged in accordance with the instructions of the Ministry of Surface

Transport, (TRANSCHART), New Delhi.

4. SHIPMENT FROM POLAND & CZECHOSLOVAKIA

(i) IMPORTS FROM POLAND

Page 160: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Shipping Arrangement

HLL/PCD/ESIC-20/10-11 Page 160 of 166 01.09.2010

Shipment under this contract would be made by the National flag lines of the two parties and vessels of

the third flag conference lines, in accordance with the agreement between the Govt. of the Republic of

India and the Govt. of the Polish People’s Republic regarding Shipping Co-operation dated 27.6.1960 as

amended up-to-date.

(ii) IMPORTS FROM CZECHOSLOVAKIA

Goods under this contract would be signed by the National flag lines of the two parties and vessels of the

third flag conference lines, in accordance with the Agreement Co-operation in shipping between India and

Czechoslovakia signed on 3.11.1978 and ratified on 19.12.1979, as amended up-to-date.

Shipping arrangement should be made by the Sellers in consultation with Resident Representative of the

Indian Shipping Lines in Gdynia, Co., Morska Agencja W. Gdyniul, Pulaskiego 8, P.O. Box 246, Gdynia

(Poland) – Telex : MG PL. 054301, Tel.: 207621, to whom details regarding contract number, nature of

cargo , quantity, port of lading, discharging, name of Government consignee, expected date of readiness

of each consignment etc. should be furnish at least six weeks in advance of the required position, with a

copy thereof endorsed to the Shipping Co-ordination Officer, Ministry of Surface Transport, (Chartering

Wing), New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex: VAHAN IN – 031 – 61157, 61158,

61159)

5. SHIPMENT FROM U.S.S.R

Shipment under this contract should be made in accordance with the agreement between the Government

of the Republic of India and the Government of U.S.S.R on Merchant Shipping 1976, as amended up-to-

date, by vessels of Indo-Soviet shipping Service.

6. SHIPMENT FROM JAPAN

The shipment of goods should be made of India vessels to the maximum extent possible subject to the

minimum of 50%.

The Seller should arrange shipment of the goods in consultation with the Embassy of India in Japan,

Tokyo to whom details regarding contract number, nature of cargo, quantity, port of loading/discharge,

name of Govt. consignee, expected date of readiness of each consignment etc. should be furnished at least

six weeks in advance of the required position.

Note: The copies of such contracts are to be endorsed both to the Attached (commercial) embassy of

India in Japan, Tokyo, and the shipping Co-ordination Officer, Ministry of Surface Transport, New Delhi.

7. SHIPMENT FROM AUSTRALIA, ALGERIA, BULGARIA, ROMANIA, EGYPY

The Seller shall arrange shipment of the goods by Indian flag vessels to the maximum extent possible

subject to a minimum of 50 %. For the purpose of ascertaining the availability of suitable Indian vessels,

the seller shall give adequate notice of not less than six weeks about the readiness of each consignment to

the Shipping Purchaser of India Ltd., SHIPPING HOUSE, 245, Madame Cama Road, Bombay – 400 021

(CABLE: SHIPINDIA BOMBAY) and also endorse a copy thereof to the Shipping Co-ordination Officer, Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex:

VAHAN IN – 031 – 61157, 61158, 61159)

8. SHIPMENT FROM PAKISTAN

Page 161: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Shipping Arrangement

HLL/PCD/ESIC-20/10-11 Page 161 of 166 01.09.2010

The shipment of cargoes should be made by Indian vessels to the maximum extent possible subject to a

minimum of 50 %.

Shipment arrangement should be made by the sellers in consultation with M/s Mogul Line Ltd., 16-Bank

Street, Fort, Bombay – 400023 (Cable: MOGUL BOMBAY: Telex: 011 – 4049 MOGUL), to whom,

details regarding contract number, nature of cargo, quantity, port of lading discharging, name of

government consignee, expected date of readiness of each consignment etc. should be furnish at least six

weeks in advance of the required position, with a copy thereof endorsed to the Shipping Co-ordination

Officer, Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex:

VAHAN IN – 031 – 61157, 61158, 61159)

9. SHIPMENT FROM U.S ATLANTIC & GULF PORTS

The Seller should arrange shipment of the goods by vessels belonging to the member lines of the India –

Pakistan – Bangladesh – Ceylon and Burma Outward Freight Conference. If the Seller finds that the space

of the ‘Conference Lines’ vessels is not available for any specific shipment he should take up with India –

Pakistan- Bangladesh – Ceylon and Burma Outward Freight Conference, 19, Rector Street, New York, N.Y. 10006 USA, for providing shipping space and also inform the Shipping Co-ordination Officer,

Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex: VAHAN IN –

031 – 61157, 61158, 61159)

10. SHIPMENT FROM ST. LAWRENCE AN EASTERN CANADIAN PORTS

The Seller should arrange shipment of the goods by vessels belonging to the following shipping lines;

1. The shipping Purchaser of India Ltd. 2. The Scindia Steam Navigation Co., Ltd

If the Seller finds that the space in the vessels of these Lines is not available for any particular

consignments, he should inform the Shipping Co-ordination Officer, Ministry of Surface Transport, New

Delhi, (Cable: TRANSCHART, NEW DELHI, Telex: VAHAN IN – 031 – 61157, 61158, 61159)

immediately so that dispensation from the shipping lines concerned to use alternative lifting may be

sought.

11. SHIPMENT FROM WEST COAST PORTS OF U.S.S CANADA AND OTHER AREAS NOT

SPECIFICALLY MENTIONED ABOVE

The Seller should arrange shipment of the goods by Indian vessels to the maximum extent possible

subject to a minimum of 50 %. For the purpose of ascertaining the availability of suitable Indian vessels

and granting dispensation in the event of their non-availability, the Seller should furnish the details

regarding contract number, nature of cargo, quantity, port of lading, discharging, name of government

consignee, expected date of readiness of each consignment etc. to the Shipping Co-ordination Officer,

Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex: VAHAN IN –

031 – 61157, 61158, 61159) at least six weeks in advance of the required position.

B) BILLS OF LADING:

(i) C.I.F./C&F/TURNKEY SHIPMENTS

The Bills of lading should be drawn to indicate Shipper and ‘Consignee’ as under:

SHIPPER: The C.I.F (C&F)/TURNKEY SUPPLIERS concerned.

Page 162: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Shipping Arrangement

HLL/PCD/ESIC-20/10-11 Page 162 of 166 01.09.2010

CONSIGNEE: As per consignee’s particulars in the contract (The name an address of the

‘Port Consignee’ and ‘Ultimate’ both should be indicated).

(ii) F.O.R SHIPMENTS

The Bills of lading should be drawn indicating shipper Consignee as under:

SHIPPER: The F.O.R suppliers Concerned

CONSIGNEE: Supplier’s Indian Agent on order

Note:

1. Moreover the name of the ‘Purchaser’ and ‘Ultimate’ Consignee should appear in the body of

the Bills of Lading as the ‘Notify’ or as a remark.

2. Two non-negotiable copies of the Bills of Lading indicating the freight amount and discount, if

any allowed, should be forwarded to The Shipping Co-ordination Officer, Ministry of surface

Transport (Chartering Wing), New Delhi after the shipment of each consignment is effected. 3. The seller should avoid the use of over-aged vessels for the shipment of the goods under the

contract and if so used the cost of additional. Insurance, if any, shall be borne by the seller.

Page 163: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Check List

HLL/PCD/ESIC-20/10-11 Page 163 of 166 01.09.2010

SECTION – XX

CHECKLIST

Name of Tenderer:

Name of Manufacturer:

Sl.

No. Activity

Yes/ No/

NA

Page No. in

the Tender

document

Remarks

1. Have you enclosed EMD of required amount

for the quoted schedules?

2(a). Have you enclosed clause-by-clause technical

compliance statement for the quoted goods vis-

à-vis the Technical specifications?

2(b). In case of Technical deviations in the

compliance statement, have you identified and

marked the deviations?

3. Have you kept validity of 120 days from the

Techno Commercial Tender Opening date as

per the TE document?

4(a). Have you enclosed duly filled Tender Form as

per format in Section X?

4(b). Have you enclosed Power of Attorney/

Authorisation in favour of the signatory?

5. Have you submitted manufacturer’s

authorization as per Section XIV?

6. Have you submitted the certificate of

incorporation?

7(a). In case of Indian Tenderer, have you furnished

Income Tax Account No. as allotted by the

Income Tax Department of Government of

India?

7(b). In case of Foreign Tenderer, have you

furnished Income Tax Account No. of your

Indian Agent as allotted by the Income Tax

Department of Government of India?

Page 164: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Check List

HLL/PCD/ESIC-20/10-11 Page 164 of 166 01.09.2010

Sl.

No. Activity

Yes/ No/

NA

Page No. in

the Tender

document

Remarks

8. Have you intimated the name and full address

of your Banker (s) along with your Account

Number

9. Have you furnished Annual Report (Balance

Sheet and Profit & Loss Account) for last three

years prior to the date of Tender opening?

10. Have you submitted the Quality Control

Requirements as per Proforma given in Section

VIII of TE document?

11. Have you accepted delivery period as per TE

document?

12. Have you accepted the terms of delivery as per

‘DDP at consignee site basis’?

13. Have you accepted the warranty/CMC as per

TE document?

14. Have you accepted all terms and conditions of

TE document?

15. Have you fully accepted payment terms as per

TE document?

16(a) Have you submitted prices of goods, turnkey

(if any), CMC etc. in the Price Schedule as per

Section XI?

16(b) Have you submitted satisfactory end user

performance certificate as per the Proforma for

performance statement in Sec. IX of TE

document in respect of all orders?

17. Have you submitted copy of the order(s)

against the above end user certificate (s)?

18. Have you furnished documents establishing

your eligibility & qualification criteria as per

TE documents?

Page 165: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Check List

HLL/PCD/ESIC-20/10-11 Page 165 of 166 01.09.2010

N.B.

1. All pages of the Tender should be page numbered and indexed.

2. The Tenderer may go through the checklist and ensure that all the documents/confirmations

listed above are enclosed in the tender and no column is left blank. If any column is not

applicable, it may be filled up as NA.

3. It is the responsibility of tendered to go through the TE document to ensure furnishing all

required documents in addition to above, if any.

(Signature with date)

(Full name, designation & address of the person duly authorised sign on behalf of the

Tenderer)

For and on behalf of

(Name, address and stamp of the tendering firm)

Page 166: EMPLOYEE STATE INSURANCE CORPORATION · 62 Deep Freezer -40 Deg. 2 8,000.00 63 Deep Freezer -80 Deg. 1 6,000.00 64 Cell Separator 1 8,000.00 65 ICU Ventilator (High End) 1 24,000.00

HLL Lifecare Ltd. Consignee addresses

HLL/PCD/ESIC-20/10-11 Page 166 of 166 01.09.2010

Section – XXI

Consignee addresses

Consignee Code Consignee Address Telephone No.

Joka

Medical Superintendent

ESIC Hospital & Occupational Disease

Centre (E.Z.)

Diamond Harbour Road

Joka, Kolkata - 700 104

033-2467 1764

Basai

Medical Superintendent

ESI Hospital

Ring Road, Basaidarapur

New Delhi - 110 015

011-25100664

Baddi

Medical Superintendent

ESIC Hospital

Baddi, Himachal Pradesh

--

Bhiwadi

Medical Superintendent

ESIC Hospital

Bhiwadi, Rajasthan

--

Manesar

Medical Superintendent

ESIC Model Hospital

Manesar, Haryana

--

Ludhiana

Medical Superintendent

ESIC Model Hospital

Bharat Nagar, Ludhiana-141 001

0161-2772435/36 (T)

0161-2774357 (F)

Indore

Medical Superintendent

ESIC Model Hospital and ODC

Nanda Nagar, Indore – 452011 (M.P.)

0731-2559091 (T)

0731-2559080 (F)

Adityapur

Medical Superintendent

ESIC Hospital

Adityapur, Jamshedpur,

Jharkhand-831013

0657-2383866 (T)

0657-2383866 (F)

NB: The purchaser/consignee will ensure timely issue of CDEC, Octroi Exemption

Certificates, Road Permits & Entry Tax Exemption Certificates, wherever

applicable, to the suppliers.