Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items at the end of the Scopes Report) These reports are compiled by our Medical Engineering Team & are for guidance only; all lots are sold as seen, as per our Conditions of Sale. We recommend that wherever possible you inspect lots prior to purchase. Lot Description Condition 100 FUJINON EG-530FP Videoscope 1000x8.9mm Insertion Tube Very good condition Bending mechanism & Angulation Control Good movement and locks Light Guide Cable & Light Transmission Good Objective Lens and Focusing No processor to test with Remote Control Switches No processor to test with Distal End Good condition Image Quality No processor to test with Working Channel (s) Clear Forceps Elevation Mechanism N/A Leak Test Pass Overall Condition Good physical condition, unable to fully test scope’s working condition 101 FUJINON EC-530FL Videoscope 1600x12.5mm Insertion Tube Good condition Bending mechanism & Angulation Control Good movement and locks Light Guide Cable & Light Transmission Good Objective Lens and Focusing No processor to test with Remote Control Switches No processor to test with Distal End Good Image Quality No processor to test with Working Channel (s) Clear Forceps Elevation Mechanism N/A Leak Test Pass Overall Condition Good physical condition, unable to fully test scopes working condition 102 FUJINON EG-530FP Videoscope 1000x8.9mm Insertion Tube Good condition Bending mechanism & Angulation Control Good movement and locks Light Guide Cable & Light Transmission Good Objective Lens and Focusing No processor to test with Remote Control Switches No processor to test with Distal End Good condition Image Quality No processor to test with Working Channel (s) Clear Forceps Elevation Mechanism N/A Leak Test Pass Overall Condition Good physical condition, unable to fully test scopes working condition 103 OLYMPUS No Identification Insertion Tube Good condition Bending mechanism & Angulation Control Good movement and locks Light Guide Cable & Light Transmission Good but only manual

Upload: others

Post on 24-Feb-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015

(Condition Reports for other items at the end of the Scopes Report)

These reports are compiled by our Medical Engineering Team & are for guidance only; all

lots are sold as seen, as per our Conditions of Sale. We recommend that wherever possible

you inspect lots prior to purchase.

Lot Description Condition

100 FUJINON

EG-530FP

Videoscope

1000x8.9mm

Insertion Tube – Very good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – No processor to test with

Remote Control Switches – No processor to test with

Distal End – Good condition

Image Quality – No processor to test with

Working Channel (s) – Clear

Forceps Elevation Mechanism – N/A

Leak Test – Pass

Overall Condition – Good physical condition, unable to fully test scope’s

working condition

101 FUJINON

EC-530FL

Videoscope

1600x12.5mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – No processor to test with

Remote Control Switches – No processor to test with

Distal End – Good

Image Quality – No processor to test with

Working Channel (s) – Clear

Forceps Elevation Mechanism – N/A

Leak Test – Pass

Overall Condition – Good physical condition, unable to fully test scopes

working condition

102 FUJINON

EG-530FP

Videoscope

1000x8.9mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – No processor to test with

Remote Control Switches – No processor to test with

Distal End – Good condition

Image Quality – No processor to test with

Working Channel (s) – Clear

Forceps Elevation Mechanism – N/A

Leak Test – Pass

Overall Condition – Good physical condition, unable to fully test scopes

working condition

103 OLYMPUS

No

Identification

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good but only manual

Page 2: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Videoscope

1000x11.6mm

brightness control is working

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional

Distal End –Fair condition

Image Quality – Satisfactory

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Satisfactory overall condition

104 Olympus

HYF-XP

Direct view

200x3.23mm

Insertion Tube –Satisfactory condition, large dent

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Satisfactory optics

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Satisfactory image (more than 50 fibres missing)

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition –Satisfactory physical condition, large dent and most

of the fibres missing from image

105 OLYMPUS

GIF-Q230

Videoscope

1000x11.87m

Insertion Tube –Satisfactory condition with few minor dents

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good focusing

Remote Control Switches – Fully funcional

Distal End – Good condition

Image Quality –Good image quality

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

106 OLYMPUS

CYF- V2

Videoscope

250x5.61mm

Insertion Tube – Satisfactory condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – No processor to test with

Remote Control Switches – No processor to test with

Distal End – Good condition

Image Quality – No processor to test with

Working Channel (s) – Clear

Leak Test –Pass

Overall Condition – Good overall condition

107 OLYMPUS

CYF- V2

Videoscope

270x5.66mm

Insertion Tube – Good condition with minor dents

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – No processor to test with

Remote Control Switches – No processor to test with

Distal End – Good condition

Image Quality – No processor to test with

Working Channel (s) – Clear

Page 3: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Leak Test –Pass

Overall Condition – Good overall condition

108 OLYMPUS

CYF- V2

Videoscope

250x5.61mm

Insertion Tube – Good condition with minor dents

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – No processor to test with

Remote Control Switches – No processor to test with

Distal End – Good condition

Image Quality – No processor to test with

Working Channel (s) – Clear

Leak Test –Pass

Overall Condition – Good overall condition

109 PENTAX

EC-380IL

Videoscope

1400x12.10mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional

Distal End – Satisfactory condition

Image Quality – Reasonable image

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Satisfactory physical condition

110 OLYMPUS

OSF-2

Direct View

500x12.20mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Poor movement

Light Guide Cable & Light Transmission – Good light guide cable

Objective Lens and Focusing – Poor focussing

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Poor image

Working Channel (s) – Clear

Leak Test –Leaking

Overall Condition – Poor overall condition, dark image is seen and the

objective lens has been come off the probe.

111 OLYMPUS

CYF-3

Direct View

270x5.8mm

Insertion Tube – Satisfactory condition with few minor dents

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good focussing

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with few fibres missing

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Average overall condition

112 PENTAX

F1-10BS

Direct View

Insertion Tube – Good condition with few minor dents

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – No light guide cable

Objective Lens and Focusing – Satisfactory optic

Page 4: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

550x3.65mm

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Reasonable image (25+ fibres missing)

Working Channel (s) – N/A

Leak Test – Leaking

Overall Condition – Reasonable overall condition, 25+ fibres missing

113 OLYMPUS

ENF-P3

Direct View

200x4.0mm

Insertion Tube – Good condition with few minor dents

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good focussing

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with 5+ fibres missing

Working Channel (s) – N/A

Leak Test – Pass

Overall Condition – Good overall condition

114 OLYMPUS

CF-240AL

Videoscope

1600x12.80mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good focussing

Remote Control Switches – Fully functional

Distal End – Good condition

Image Quality – Good image quality with dim colour and slightly grey

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Satisfactory overall condition

115 OLYMPUS

GIF-XQ240

Videoscope

1000x10.10mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – Good focussing

Remote Control Switches – Fully functional

Distal End – Good condition

Image Quality – Good image but the image colour is slightly grey

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Satisfactory overall condition

116 PENTAX

EB-1830T2

Videoscope

500x6.60mm

Insertion Tube – Good condition with few minor dents

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional

Distal End – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

Page 5: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

117 OLYMPUS

No

Identification

Direct View

500x4.40mm

Insertion Tube – Satisfactory condition with few dents

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Reasonable optics

Remote Control Switches – N/A

Distal End – Good

Image Quality – Reasonable image with few pixels missing

Working Channel (s) – N/A

Leak Test – Leaking

Overall Condition – Satisfactory overall condition

118 OLYMPUS

CYF-200

Videoscope

270x5.60mm

Insertion Tube – Good condition with few minor dents

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Reasonable optics

Remote Control Switches – Fully functional

Distal End – Good

Image Quality – Reasonable image

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Satisfactory overall condition

119 OLYMPUS

CF-240DL

Videoscope

1600x12.10mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional

Distal End – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Good overall condition

120 OLYMPUS

BF-P260F

Videoscope

450x4.53mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional

Distal End – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

121 OLYMPUS

BF-200

Videoscope

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good but very dim light

output

Objective Lens and Focusing – Satisfactory optics

Remote Control Switches – Fully functional

Distal End – Good

Page 6: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

450x6.4mm Image Quality – Satisfactory image

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

122 Fujinon

NAP-F

Direct View

420x4.65mm

Insertion Tube – Poor condition with half of the insertion tube’s outer

insulation is ripped, exposing inner metal insulation

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Light guide cable is in poor

condition with its outer insulation worn and torn. Good light output

Objective Lens and Focusing – Good focussing

Remote Control Switches – N/A

Distal End – Satisfactory condition

Image Quality – Good image with few pixels missing

Working Channel (s) – N/A

Leak Test – N/A

Overall Condition – Poor overall condition

123 PENTAX

FNL-10P2

Direct View

260x3.95mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good light output

Objective Lens and Focusing – Reasonable optics and focussing

Remote Control Switches – N/A

Distal End – Good

Image Quality – Reasonable image

Working Channel (s) – N/A

Leak Test – Pass

Overall Condition – Good overall condition

124 OLYMPUS

LF-TP

Direct View

450x5.05mm

Insertion Tube – Good condition with few minor dents

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – No light guide cable

Objective Lens and Focusing – Satisfactory optics and focussing

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Reasonable image with few pixels missing

Working Channel (s) – Clear

Leak Test –Pass

Overall Condition – Good overall condition

125 OLYMPUS

TJF-240

Videoscope

1200x12.75mm

Insertion Tube – Good condition with the distal end cap torn and worn

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good Light output but auto

brightness control is not working

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional

Distal End – Poor condition

Image Quality – Good image

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

Page 7: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

126 Richard Wolf

No

Identification

Direct View

270x3.66mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good Light output

Objective Lens and Focusing – Good optics and focussing

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with more than 100 pixels missing

Working Channel (s) – N/A

Leak Test – Leaking

Overall Condition – Good overall condition

127 PENTAX

ECY-15P2

Direct View

300x5.60mm

Insertion Tube – Good Condition

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good light output

Objective Lens and Focusing – Good optics and focussing

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with few pixels missing

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

128 Fujinon

EG-450WR5

Videoscope

1000x10.05mm

Insertion Tube – Good Condition

Bending mechanism & Angulation Control – Good movement and lock

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

129 Fujinon

EC-450WL5

Videoscope

1600x12.55mm

Insertion Tube – Good Condition

Bending mechanism & Angulation Control – Good movement and lock

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

130 Fujinon

No

Identification

Videoscope

Insertion Tube – Good Condition

Bending mechanism & Angulation Control – Good movement and lock

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Page 8: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

1600x13.82mm

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Good overall condition

131 Fujinon

No

identification

Videoscope

1600x12.55mm

Insertion Tube – Good Condition

Bending mechanism & Angulation Control – Good movement but locks

are not functional

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Good overall condition

132 Fujinon

EG-450WR5

Videoscope

1000x9.65mm

Insertion Tube – Good Condition

Bending mechanism & Angulation Control – Good movement and lock

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

133 Fujinon

No

Identification

Videoscope

1000x9.90mm

Insertion Tube – Good Condition with few minor dents

Bending mechanism & Angulation Control – Good movement and lock

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

134 Fujinon

EG-450WR5

Videoscope

1000x10.0mm

Insertion Tube – Good condition with few minor dents

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

135 Olympus

BF-P60

Insertion Tube –Good condition with minor dent

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good

Page 9: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Direct view

520x4.95mm

Objective Lens and Focusing – Satisfactory optics

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with one pixel out

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition –Good physical condition

136 Fujinon

No

Identification

Videoscope

1600x12.70mm

Insertion Tube – Good condition with few minor dents

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good light transmission

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Satisfactory condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

137 Richard Wolf

No

Identification

Videoscope

250x5.00mm

Insertion Tube – Good condition with

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Not tested

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

138 Olympus

PCF-240S

Videoscope

900x12.30mm

Insertion Tube –Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional

Distal End – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition –Good physical condition

139 Olympus

GIF-2T20

Direct view

1000x12.65mm

Insertion Tube –Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with two pixels out

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition –Good physical condition

Page 10: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

140 Olympus

TJF-240

Videoscope

1200x12.82mm

Insertion Tube –Good condition with minor dents and missing distal end

cap

Bending mechanism & Angulation Control – Satisfactory movement and

locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Satisfactory optics

Remote Control Switches – Fully functional

Distal End – Poor condition with missing end cap

Image Quality – Good but dull coloured image

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition –Good physical condition

141 PENTAX

FCY-15P2

Videoscope

(No case)

300x5.10m

Insertion Tube –Good condition

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – N/A

Distal End – Satisfactory condition

Image Quality – Good image with 2 pixels missing

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Good physical condition

212 Olympus

CHF-P60

Direct view

270x5.60mm

Insertion Tube –Good condition with minor dent

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Satisfactory optics with few stains on the

lens

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with 15+ pixels out

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition – Satisfactory physical condition

213 Olympus

BF-B3

Direct view

470x6.40

Insertion Tube – Poor condition with large dents and the outer insulation

is ripped off exposing the inner metal insulation

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Cable has few dents but good

light transmission

Objective Lens and Focusing – Poor optics with stains on the lens

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Poor image with lots of pixels missing

Working Channel (s) – Clear

Leak Test – N/A

Overall Condition – Poor physical condition

214 Olympus

CHF-P60 Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement

Page 11: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Direct view

270x5.0mm

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with 100+ pixels missing

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition –Good physical condition

215 PENTAX

FB-15X

Direct view

510x4.95mm

Insertion Tube –Good condition

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Good image with two pixels out

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good physical condition

216 Olympus

CYF-240

Videoscope

240x5.60mm

Insertion Tube –Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional except for one

Distal End – Good condition

Image Quality – Satisfactory image

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good physical condition

217 Olympus

LF-V

Videoscope

510x4.30mm

Insertion Tube – Good condition

Bending mechanism & Angulation Control – Good movement

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Good condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition –Good physical condition

218 Olympus

CF-1T200L

Videoscope

1600x14.65mm

Insertion Tube –Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – Good optics

Remote Control Switches – Fully functional

Distal End – Good condition

Image Quality – Satisfactory image

Working Channel (s) – Clear

Leak Test – Leaking

Page 12: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Overall Condition –Good physical condition

222 PENTAX

ED-3410

Videoscope

520x12.20mm

Insertion Tube –Good condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Satisfactory condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Leaking

Overall Condition –Good physical condition

223 PENTAX

EG-3840TK

ED-3440TK

ED-3440TK

Videoscope

1000x12.05mm

1200x12.96mm

1200x12.40mm

Insertion Tube –Good condition, Both ED-3440TK are missing its distal

end cap

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good except for one (EG-

3840TK) with dim light

Objective Lens and Focusing – No processor to test with

Remote Control Switches – Not tested

Distal End – Satisfactory condition

Image Quality – Not tested

Working Channel (s) – Clear

Leak Test – Leaking (EG-3840TK), Pass

Overall Condition – Reasonable physical condition

224 GYRUS ACMI

DUR-8

Direct View

1000x12.05mm

Insertion Tube –Good condition

Bending mechanism & Angulation Control – Poor movement

Light Guide Cable & Light Transmission – No light guide cable

Objective Lens and Focusing – Below average

Remote Control Switches – N/A

Distal End – Good condition

Image Quality – Poor

Working Channel (s) – Clear

Leak Test – N/A

Overall Condition – Good physical condition

254 Pentax

ED-3410

ES-3840

EG-3440

Videoscope

1200x12.36mm

600x12.54mm

1000x12.55mm

Insertion Tube – Good Condition

Bending mechanism & Angulation Control – Good movement and locks

Light Guide Cable & Light Transmission – Good light output

Objective Lens and Focusing – Good optics and focussing

Remote Control Switches – Fully functional

Distal End – Good condition

Image Quality – Image is slightly yellow

Working Channel (s) – Clear

Leak Test – Pass

Overall Condition – Good overall condition

Page 13: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Condition Report – Sale: 1505 Date: 12 / 05 / 2015

These reports are compiled by our Medical Engineering Team & are for guidance only; all

lots are sold as seen, as per our Conditions of Sale. We recommend that wherever possible

you inspect lots prior to purchase.

Lot

Number

Description Report

1 GE T2000 Treadmill

with GE Case Stress

Test Unit

The unit powers up successfully. The Case software can be

accessed. Treadmill powers up ok. Acquisition module not

included therefore unable to test the system and connection cable

RS232 to the treadmill and ECG unit is not included. No further

test carried out. Few keys are missing from the keyboard. Unit is

in satisfactory physical condition.

5 Olympus Stack with

Olympus Lucera CV-

260 Camera

Controller, CLV-U40

Light Source and

OVC 200 Training

Scope

The processor and light source are able to power up with its

function keys and indicators in working order. The light source is

able to produce a good light output. The brightness adjustment is

in working order. Monitor is not included. Camera head included

and was tested on another processor and monitor and seems to be

working. The stack is in good physical condition.

6 Sonosite iLook

Portable Ultrasound

with Probe, Power

Supply and User

Guide

Unit powers up ok. Display produces a good image. Function

buttons are in working condition. Linear probe is in good physical

condition. No further tests carried out. Sonosite power supply

included. Overall unit is in reasonable physical condition with

wear and tear around buttons.

7 Sonosite iLook

Portable Ultrasound

with Probe, Power

Supply and User

Guide

Unit powers up ok. Display produces a good image. Function

buttons are in working condition. Linear probe is in good physical

condition. No further tests carried out. Sonosite power supply

included. Overall unit is in good physical condition.

9 Datex Ohmeda S/5

Avance Anaesthetic

Machine

The unit powers up and completes the system self test

successfully. Display panel, function keys and rotary control knob

are all in good working order. All ventilation modes which

include VCV, PCV, SIMV/PSV, PSV Pro, SIMV-PC & PCV-VG

are functional and the back up battery is also working. Overall the

unit is reasonable physical and cosmetic condition (the surface

layer on some of the front panel buttons have been worn out

slightly). Bellows, CO2 canister and medical gas hoses are in

pace. The unit does not include a gas module. O2 sensor needs

Page 14: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

replacing. SN: ANBH00184. Software version: 06.10.

10 Datex Ohmeda S/5

Patient Monitors x 2

with Some Leads

Both units are able to power up and pass through its initial self-

test successfully. Both of the unit’s display and function buttons

are working in order. One unit has only gas module. Second unit

has multi-parameter and gas module. ECG and SpO2 leads

included and were tested using ECG simulator and the values

were within limit. NIBP not tested but seem to be working.

Pressure, temperature and gas not tested. Both units are in fair

physical condition with broken parts.

12 Olympus CLV-S20

Lightsource

CLV S20 is able to power up and produce a good light output. All

the function buttons and indicators are functioning. Its spare bulb

is in working order. The brightness adjustment for manual mode

is working. Overall unit is in fair physical condition.

14 Olympus CLH-250

Light Source - 28128,

Olympus OTV-S5

Camera Control -

28510, Olympus

OEV142 Trinitron

Monitor, Olympus

OTV-S5 Camera

MH999E - 28129,

522383

The processor, light source and the monitor powers up ok. Light

source failed to produce a light output. The camera head was

tested using the processor and a good image was obtained on the

monitor. No cable or connections included. All units are in

satisfactory physical condition.

15 GE Marquette Dash

3000 Patient Monitor

x 2

Both units are able to power up and pass through its initial self-

test successfully. Both of the unit’s display and function buttons

are working in order. ECG, NIBP, Temperature, Pressure and

SpO2 leads are not included and therefore not tested. CO2

measurement cable in included. Both units are in fair physical

condition with scuffs and marks.

16 Olympus CLK-4 Light

Source

CLK-4 is able to power up however failed to produce light output.

Unit’s air pump is functioning. Overall unit is in fair physical

condition.

18 ConMed Hyfrecator

2000 ESU

Unit power up ok, the numerical display and RF power selector

knob is working fine. The monopolar and bipolar modes was

tested using an ESU analyser and all of the RF powers measured

were the set values. The unit is in good physical condition but

come with no mobile stand.

21 Eschmannn TD 830

Electrosurgical Unit

Unit power up ok, the numerical display and RF power selector

keys are working fine. The unit was tested using an ESU analyser.

The bipolar RF outputs were well within the set values. However,

Page 15: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

the monopolar mode CUT has no RF power output. COAG output

is working fine. The unit is in average physical condition and no

monopolar and bipolar footswitch included.

27 Carl Zeiss 150 FC

Colposcope

Lamp is powering up and produces a good light output with its

brightness adjustment in working order. Reasonable optic system.

Magnification is working. The objective lens is f=300. Eyepiece

included. The mechanical movements and locks are in working

order. The physical condition is average.

Eyepiece: 10x

36 Eschmann TD850

Electrosurgical Unit

Unit power up ok, the numerical display and RF power selector

knob is working fine. The monopolar mode was tested using an

ESU analyser and some of the RF power measured were slightly

low than the set values for CUT; COAG was ok. The bipolar RF

outputs were within the set values. The unit is in fair physical

condition and monopolay and bipolar footswitches are included.

44 Camlab ALC 4232

Centrifuge

ALC 4232 centrifuge power up ok, with its function buttons in

working order. The lid lock is functioning. Timer and speed

controls are working. Unit completed a cycle without any errors.

Rotor and test tube holders in fair condition. Overall lot is in

satisfactory physical condition with some scuffs and marks.

48 Penlon Prima SP

Anaesthetic Machine

with Penlon AV900

Touch Screen

Ventilator

Unit powers up ok. The pressure gauges and flow meters are

working fine. O2 flush is functional. Basal flow and anti-hypoxic

feature is working in order. The AV900 ventilator powers up and

passed the initial self-test successfully. The touch screen and

display are working. The re-breathing circuit and absorber are all

intact. The ventilator is working. All the set parameters and alarm

system is functional. Oxygen sensor is not included. The medical

gas hose is included for the O2, Air and N2O. AGSS hose

included. Overall the lot is in satisfactory physical condition with

scuffs and marks.

Sotware version : 3.22

49 Datex Ohmeda Light

Patient Monitor x 2

Both units are able to power up and pass through its initial self-

test successfully. Power adaptor not included but one was used to

test. Both of the unit’s function buttons are working in order.

LCD of both units are defective however one unit can still display

some information whereas the second one is fully blank. SRAM

error shown on one of the unit. ECG for one of the unit was tested

using test leads and ECG simulator and the values were within

limit and the recorder is working. No leads or accessories

included. Both units are in good physical condition with scuffs

Page 16: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

and marks.

52 Mitek VAPR 3 ESU Unit power up with error message ‘Faulty Accessories’ The

footswitch is included but hand piece.

53 Pulse Oximeter Ohmeda Biox 3740 x3 does not have any mains adapters or finger

probe; unable to test these units. The Nellcor N-300 power up ok

display and function keys working but no probe – in fair physical

condition. Datex Ohmeda 3800 – power up but the LCD remains

bright and unable to adjust contrast.

66 HP Pagewriter XLS

ECG Recorder with

ECG Leads

The unit power up ok and the initial screen is displayed but the

system failed to proceed to the function screen menu. The lot

comes with ECG leads. Unit not tested. fair physical condition.

67 Datex Ohmeda Excel

210SE Anaesthetic

Machine 7800

Ventilator

Unit powers up ok. The pressure gauges and flow meters are

working fine. O2 flush is functional. Basal flow and anti-hypoxic

feature is working. The 7800 ventilator powers up and passed the

initial self-test successfully. The re-breathing circuit and absorber

are all intact. The ventilator is functional. All the set parameters

and alarm system is working. The medical gas hose is included

for the O2, Air and N2O. AGSS hose included. Overall the lot is

in satisfactory physical condition with scuffs and marks.

68 Olympus Stack

Trolley with Sony

Trinitron Monitor

UHI-2 Insufflator,

CLV 530 Light Source

OTV-56 Camera

Controller, MH317

Footswitch, Sony VCR

and Sony Colour

Video Printer, OTV

5X2 Control

The camera processor and light source are able to power up with

its function keys and indicators are working in order. Camera

head is not included therefore unable to fully test. The light

source is able to produce a good light output. The brightness

adjustment is in working order. The monitor powers up. The

insufflation unit, color video printer and video cassette recorder

power up but not tested. Footswitch for insufflator included. The

stack system is in good physical condition.

70 Medtronic Lifepack

12 Defibrillator with

ECG and Patient

Lead

Defibrillator powers up and is able to pass through its initial self-

test successfully. Function buttons, selector knob and display are

in working order. AED cable and ECG lead included and was

used to test with Defibrillator analyser. Unit showed to be

functional in AED, Pacer and manual modes when tested. Unit is

able to charge and deliver shock and output was within limit. User

test passed. Audio voice prompts and recorder was functional. AC

adaptor included. Good physical condition.

73 Philips Intellivue

MP70 Monitor with

Monitor powers up ok with its touch screen and display in

working order. ECG was tested with test leads and ECG simulator

Page 17: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

M3105A Module x 2 and is within limit. SpO2, NBP, Temperature and pressure not

tested. No leads included. Satisfactory physical condition.

YOM: 2005

74 Philips Intellivue MP

70 Monitor with

M3105A Module x 2

Monitor powers up ok with its touch screen and display in

working order. ECG was tested with test leads and ECG simulator

and is within limit. SpO2, NBP, Temperature and pressure not

tested. No leads included. Satisfactory physical condition with

some cracks.

YOM: 2005

76 EX Medispin

Centrifuge

IEC Medispin centrifuge powers up ok ALC 4232 centrifuge

power up ok. The lid lock is functioning. Unit completed a cycle

without any errors. Rotor and holders are in fair condition.

Overall lot is in good physical condition.

86 GE Trasport Pro

Patient Monitor and

GE Dash 4000 Patient

Monitor

One of the units does not have the power adaptor therefore unable

to test. Second unit powers up ok with its display and function

buttons in working order. No leads included therefore unable to

test. Both units are in reasonable physical condition.

152 Olympus 7° Rigid

Scope

Image is blurry (Optic not clear)

153 MEA Applicator Both devices appear to intact, no physical damage found. Device

not tested.

175 Dyonics 30° Rigid

Scope x 2

In good physical condition but one rigid scope has a crack in the

optics and the image from the other rigid scope is not clear. The

focus controls are working.

180 Storz Laparoscopic

Instruments x 4

In satisfactory physical condition. The handle of one of the

forceps has been damaged.

182 Olympus OTV-S4

Camera, Video

Adaptor x 2 (ARTF-2,

ARTL08E)

The camera head is in fair physical condition. No processor

therefore unable to test the image quality.

183 Olympus OTV-S5 x 2

with Camera Head

and Vista Camera

Head

All three camera heads are in fair physical condition. No

processor therefore unable to test the image quality.

232 Datex Ohmeda Cardio

Cap/5 Monitors x 4 -

10364, 15162, 13644,

All four units are able to power up with its display and function

buttons in working order except for one with damaged LCD. 2

units have Back-up battery failure error. No leads are included but

Page 18: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

10361 one was used to test. ECG was tested using ECG simulator and a

good waveform can be obtained on all of the units. SpO2 was

working in all units except for one. All units are in poor physical

condition with corroding paint work and broken parts.

249 Datascope Passport 2

Patient Monitor with

Gas Module x 4

All 4 units are able to power up. All units pass through its initial

self-test successfully except for one in which display does not

come on and an alarm sounds. Three of the unit’s display and

function buttons are working in order. All units have gas module

but not tested. No leads or accessories included therefore not fully

tested. All units are in reasonable physical condition.

253 Sonosite Scanner ilook The unit power up ok, the LCD and function soft keys are

working fine. Good image obtained with the linear probe. A slight

damage to the corner of the probe is found but doesn’t seem to

affect the image. A selection button is missing. Sonosite charger

is included. Unit is in good physical condition.

254 Stack System to

Include Sony Monitor

Pentax EPM - 3500

Light Source, Sony

VCR, Sony Colour

Video Printer, Pentax

05-A13 White Balance

x 2, Pentax EG-3440

Video Colonoscope,

Pentax EG 3840

Videoscope, Pentaxed

341- Videoscope

All units on the stack are able to power up. The function keys and

indicators are working on the EPM 3500 video processor except

for few keys which are worn out and stiff. Videoscope (x3)

included and is mentioned under scope report lot 254. The light

source is able to produce a good light output and the air pump is

functional. The Sony monitor is working fine. The videocassette

recorder and printer power up. Keyboard included. The stack

system is in good physical condition.

255 Olympus Evis TK-E2

Procedure Cart with

Olympus CV-240

Camera Control Unit,

Olmypus CLV-U40

Light Source and

Olympus DEV191H

LCD Monitor -

507601, 2840, 2841,

2839 and Olympus

MH-977 'Pigtail'

Connector

The processor and light source are able to power up with its

function keys and indicators in working order. The light source is

able to produce a good light output. The brightness adjustment is

in working order. LCD Monitor included and is able to power up.

Camera head not included therefore unable to fully test. Keyboard

included. Overall unit is in good physical condition.

256 Drager Babylog 8000

Ventilator - 1224,

The ventilator is able to power up and complete its initial self-test

successfully. The unit’s function buttons and display are in

Page 19: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

13048 working order. All of the ventilators modes can be selected and

was tested with medical air compressor and test lung; the

ventilator was functional at the time of test. Exhalation valve is

not included. Flow sensor cable included but no flow sensor. A

test flow sensor was used for testing. Unable to calibrate O2

(100% oxygen was not being used). Air and O2 medical gas hoses

included. The ventilator is in satisfactory physical condition.

Vent modes: CPAP, IPPV/IMV, SIMV and SIPPV

Version: 5.01

Hours of operation: 51613

258 Maquet Betastar

Electric Operating

Table (No Remote)

The operating table’s column stainless steel cover with some

damages, the castors are not moving freely. The cushions are in

average physical condition with few wear & tear; and back rest

mechanical adjustment is working fine. No remote control to test

the motorised functions.

261 Olympus WM-30

Workstation with

Sony Monitor,

Olympus EU-M30

Endoscope Ultrasound

Centre, Sony Digital

Still

The monitor, processor and Sony digital still recorder all power

up ok. No camera head included therefore unable to test. Overall

stack is in satisfactory physical condition.

263 B Braun Dialog Plus

Dialysis Machine x 3 -

23156, 22030, 23431

All units are able to power up, with their function buttons and

touchscreen displays functional. All menus accessible. The units

are not fully tested. BP not tested. All of the unit’s double blood

pumps are intact, as are the heparin pumps. Units also include one

air detectors which again are also intact. Does not include water

and drainage hoses. Overall the dialysis machines are in average

physical condition.

Version: 8.28 (x3)

Operation hours: 29167, 28630, 31786

266 Drager Incubator

8000SC

Unit powers up ok with its numerical display and function button

in working order. Inop alarm appears. No further tests were

carried out. The height adjustment is in working order. Good

physical condition with some scratches.

269 Vallylab Force FX

Diathermy (No Foot

The unit powers up and passes through its initial self-test

successfully. The numerical display and function keys are

Page 20: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Switch) working. Monopolar (Cut & Coag) and bipolar outputs were

tested using Electrosurgical analyser and measured values were

within the set limit. Monopolar, bipolar footswitches and hand

control leads are not included. Unit is in good physical condition

with few scratch marks on the casing and fainted paint work.

270 Erbe ICC 200

Diathermy with Erbe

APC 300 Argon

Diathermey on Trolley

- 107104, 107105

The unit powers up and passes through its initial self-test

successfully. The numerical display and function keys are

working. Unable to test the monopolar and bipolar outputs.

Monopolar footswitches included and seem to be working.

Bipolar footswitch and hand control leads are not included. APC

300 powers up ok. Not fully tested. Overall unit is in satisfactory

physical condition.

271 Olympus WM-60

Workstation To

Include Sony Monitor,

Storz Telecam DX Pal

Processor, Circon

MV9083 Automatic

Video Light Source,

Linvatec D3000

Shaver

The monitor and processor powers up ok. Not fully tested due to

no camera head available. The light source failed to power up.

The D3000 shaver powers up but no footswitch included. Unit not

tested. Overall stack is in satisfactory physical condition.

276 Datex Ohmeda

Cardiocap/5 Monitor

on Stand with Fukuda

Denshi Module Racks

x 4

The unit is able to power up with its display and function buttons

in working order. No leads included but one was used to test.

ECG was tested using ECG simulator and the values were within

limit. SpO2 was tested and is working. NIBP and pressure not

tested. Four module racks included. Average physical condition.

293 Fujifilm Dry Pix 4000 The unit power up and system self-test completed ok. The

warming up phase completed. The menu is accessible and

functioning ok. A test print was done and good printout obtained.

Unit appears to be working fine. Good physical condition.

296 Drager Babylog 8000

Ventilator - 1226, 1107

The ventilator is able to power up and complete its initial self-test

successfully. The unit’s function buttons and display are in

working order. All of the ventilators modes can be selected and

was tested with medical air compressor and test lung; the

ventilator was functional at the time of test. Exhalation valve is

not included. Flow sensor cable included but no flow sensor. A

test flow sensor was used for testing purpose. Unable to calibrate

O2 (100% oxygen was not being used). Air and O2 medical gas

hoses included. The ventilator is in satisfactory physical

Page 21: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

condition.

Vent modes: CPAP, IPPV/IMV, SIMV, SIPPV and PSV

Version: 5.01

Hours of operation: 15908

314 HP Pagewriter XLS

ECG Recorder with

Leads

The unit power up ok and the initial screen is displayed but the

system failed to proceed to the function screen menu. The lot

comes with ECG leads but not tested. Fair physical condition.

317 Dinamap Pro 300 V2

Patient Monitor and

Criticare 506DX

Patient Monitor with

Leads

Both unit powers up ok with its display and function button in

working order. NIBP and SpO2 of Vital care 506DX is working.

Dinamap has its NIBP working whereas SpO2 not tested as there

is no probe included. Both are in fair physical condition with

some wear and tear around the buttons.

320 Datex Ohmeda ADU

Carestation

Anaesthetic Machine

with Isoflurane

Vaporiser

Unit powers up ok. The pressure gauges are working fine. O2

flush is working. The ventilator powers up and passed the initial

self-test successfully. The re-breathing circuit is disconnected and

absorber is missing. The ventilator is working. All the set

parameters and alarm system is functional. Back up battery needs

charging. The medical gas hose is included for the O2, Air and

N2O. AGSS hose included. Patient monitor power up and

completes its initial self-test successfully. Multi parameter

module, NIBP module and gas module included but not tested. No

leads included. Overall the lot is in fair physical condition with

scuffs and marks.

328 Mac 5000 GE

Marquette ECG

Machine with Leads -

100982

Mains power indicator is on when connected and the screen

comes on. Function buttons are working. ECG waveforms can be

acquired using the included leads and the analyser. The recorder is

working and the print is clear. Back-up battery needs charging.

Unit is in satisfactory physical condition with scuffs and marks

and few cracks on the casing.

332 Philips Heartstart XL

Defib with Power

Cables x 2

Both defibrillators power up and are able to pass through its initial

self-test successfully. Function buttons, selector knob and display

are in working order. Units include AED cable (x2) and are

working fine. Units showed to be functional in both AED and

manual modes when tested with defibrillator analyser. Both are

able to charge and deliver shock within the limit. ECG part of the

unit not tested. Both units have pacer mode but not tested. Audio

voice prompts and printer were functional. Battery and ECG leads

not included. Both units are in fair physical condition.

Page 22: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

356 GE Mac 1200ST ECG

Machine (No Leads)

The unit power up ok but the initial screen cannot be displayed.

The display keeps flashing and the system failed to proceed to the

function screen menu. Unit not tested. Fair physical condition.

364 Stryker Stack trolley The processor 688i and camera head produces good image quality

on the monitor but if the camera head’s cable is move side to side

interferences can be seem on the image. This indicating broken

wires or loose contact. The Sony colour printer feed the paper

through but not printing the image. The Arthroscope Pump power

up and self-test ok. Display and function keys working fine; not

fully tested. In good physical condition. The CORE Powered

Instrument Driver power up ok and the display and touch screen

are working fine but not fully tested.

365 Valleylab Force 30

Diathermy with

Footswitches - 303041,

Valleylab Optimumm

Smoke Evacuator -

KJ01739

The unit powers up and passes through its initial self-test

successfully. The numerical display and function keys are

working. Monopolar (Cut & Coag) and bipolar outputs were

tested using Electrosurgical analyser and measured values were

within the set limit. Monopolar footswitches included and are

working fine. Bipolar footswitch and hand control leads are not

included. Smoke evacuator powers up ok. The filter needs

replacing. All the selection knobs and indicators are working. Not

fully tested. Overall unit is in good physical condition with few

scratch marks.

367 Valleylab Force 30

Diathermey with

Footswitches - 303043

The unit powers up and passes through its initial self-test

successfully. The numerical display and function keys are

working. Monopolar (Cut & Coag) and bipolar outputs were

tested using Electrosurgical analyser and measured values were

within the set limit. Monopolar footswitches included with slight

damage to the cable but are working fine. Bipolar footswitch and

hand control leads are not included. Overall unit is in satisfactory

physical condition.

370 Spacelabs Eclipse Plus

ECG Recorder (No

Leads) - CP1057

Mains power indicator is on when connected and the screen

comes on. Function buttons are working. ECG leads not included

therefore no further tests were carried out. Unit is in reasonable

physical condition with scuffs and marks and few cracks on the

casing. Unit has ‘For spare parts only’ label.

371 Drager Babylog 8000

Plus Ventilator -

ARJL - 0055

The ventilator is able to power up and complete its initial self-test

successfully. The unit’s function buttons and display are in

working order. All of the ventilators modes can be selected and

was tested with medical air compressor and test lung; the

ventilator was functional at the time of test. Exhalation valve is

not included. Flow sensor cable and flow sensor not included. A

Page 23: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

test flow sensor was used for testing purpose. Unable to calibrate

O2 (100% oxygen was not being used). Air and O2 medical gas

hoses included. The ventilator is in satisfactory physical condition

with broken front panel cover.

Vent modes: CPAP, IPPV/IMV, SIMV, SIPPV and PSV

Version: 5.01

Hours of operation: 27890

377 Evis Olympus

Procedures Trolley

TK-E1 with Olympus

CV-240 Camera

Controller and

Olympus CLV-U20

Light Source -

2234013, C010963,

C004529

The camera head processor and light source are able to power up

with its function keys and indicators in working order. Camera

head is not included therefore unable to fully test. The light

source is able to produce a good light output. The brightness

adjustment is in working order. The monitor not included. The

procedure trolley is in satisfactory physical condition.

378 Datex Ohmeda S/5

Avance Anaesthetic

Machine (Rear Wheel

Broken)

The unit powers up and completes the system self test

successfully. Display panel, function keys and rotary control knob

are all in good working order. All ventilation modes which

include VCV, PCV, SIMV/PSV, PSV Pro & SIMV-PC are

functional. The back up battery is flat. Overall the unit is

reasonable physical and cosmetic condition (one of the wheels is

jammed). Bellows, CO2 canister and medical gas hoses are in

pace. The gas module (M-CAiO) is also included.

SN: ANBH00154. Software version: 06.10.

379 Datex Ohmeda S/5

Avance Anaesthetic

Machine

The unit powers up and completes the system self test

successfully. Display panel, function keys and rotary control knob

are all in good working order. All ventilation modes which

include VCV, PCV, SIMV/PSV, PSV Pro & SIMV-PC are

functional and the back up battery is also working. The system a

leak but overall the unit is reasonable physical and cosmetic

condition. Bellows, CO2 canister and medical gas hoses are in

pace. The unit does not include a gas module.

SN: ANBH001846. Software version: 06.10.

382 Drager Babytherm

8000 Incubator 24582

(Incomplete)

Unit powers up with its numerical display in working order. Error

70 is displayed. No further tests were carried out. Not a complete

set. Poor physical condition with missing button covers.

Page 24: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

383 Drager Babylog 8000

Ventilator - 0034

The ventilator is able to power up and complete its initial self-test

successfully. The unit’s function buttons and display are in

working order. All of the ventilators modes can be selected and

was tested with medical air compressor and test lung; the

ventilator was functional at the time of test. Exhalation valve is

not included. Flow sensor cable and flow sensor not included. A

test flow sensor was used for testing purpose. Unable to calibrate

O2 (100% oxygen was not being used). Air and O2 medical gas

hoses included. The ventilator is in satisfactory physical

condition.

Vent modes: CPAP, IPPV/IMV, SIMV, SIPPV and PSV

Version: 5.01

Hours of operation: 27025

385 Olympus TI-1900

Imaging Trolley with

Sony Monitor - 2843,

2842

Monitor powers up ok. Three stripes of colored lines appear at the

top of the screen. No further tests carried out. The stack and the

monitor are in reasonable physical condition.

392 Spacelabs Eclipse Plus

ECG Recorder (No

Leads) - CP1663

The unit powers up ok. The display screen and keyboard is

working fine. ECG leads not included therefore no further tests

were carried out. Good print out was produced. The unit works on

backup battery pack as well as mains connection. Satisfactory

physical condition.

393 Dinamap Pro 400

Vital Signs Monitor -

4373 (with Fingerclip)

Unit failed to power up therefore unable to fully test. SpO2 probe

included. Fair physical condition with broken printer lid.

405 Storz Tricam Camera

Head - RHE632182-H

The camera head is in good physical condition. No processor

therefore unable to test the image quality.

409 Olympus OTV-S6

Camera Head -

7003366

The camera head is in fair physical condition. No processor

therefore unable to test the image quality.

409 Olympus OTV-S6

Camera Head -

7003366

The camera head is working but the image is not very clear

(showing noise in the image). Focus control is operational.

416 GE 2000 Treadmill

with GE 2000 Stress

Unit

The unit powers up successfully. Unable to boot up due to no

operating system found. Treadmill powers up ok. Acquisition

module and connection cable RS232 to the treadmill and ECG

unit is not included. No further test carried out. Unit is in

Page 25: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

satisfactory physical condition.

425 Assorted Philips HP

and Agilent M3046A

Patient Monitors with

Multiparameter

Moduels x 5

All 5 units are able to power up successfully. All the touch

buttons and display are in working order. 5 multi-parameter

modules included. MEASSRV unsupported error message

displays on two of the units and does not recognize the modules.

Three of the units were tested with simulator and test leads and

the ECG was within limit. NIBP and SpO2 are working. No leads

or accessories are included. All units are in satisfactory physical

condition with a few worn out buttons and minor scratches on

display.

436 Olympus CLE-4E

Lightsource - 2211

CLE-4E is able to power up and produce a good light output. The

brightness adjustment is working. Air pump is functional. Overall

unit is in fair physical condition.

439 Datex Ohmeda

Cardiocap/5 Critical

Care Monitor

3 of the monitors power up and self-test completed. The displays

are working fine except for one that the screen failed to come on.

The parameters were not tested. The units are in poor cosmetic

condition with flaking body paint and corrosion on the surface.

440 Datex Ohmeda

Cardiocap/5 Critical

Care Monitor

All 4 monitors power up and self-test completed. The displays are

working fine except for one with pixel problem, with line on the

screen. The parameters were not tested. The units are in poor

cosmetic condition with flaking body paint and corrosion on the

surface.

446 Olympus Stack

System

Unit power up ok, Xenon lamp is working fine. Brightness

control, auto and manual mode fully functional. The backup

halogen lamp is also working. Unit is in good condition. The

processor power up ok and colour bar is displayed on the monitor.

No camera head to check image quality. Both Sony colour printer

comes on but with error. Stack system is in fair physical

condition.

451 Stryker 1088i HD

Camera Head -

05G032304 and

Controller - 015213

Both the video processor and the camera head are fully functional

and in good physical condition. The system produces good quality

images.

455 Olympus Maintenance

Unit MU-1 x2

Both units power up ok, the pumps are working fine. Fair physical

condition.

468 Datex Ohmeda

monitor

The Light monitor has no mains adapter and the printer assembly

is missing. The front indicator cover is also missing. Not tested.

The Cardiocap II was tested with patient simulator for ECG and

Page 26: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

SpO2 and all the parameter s and values were within limit. The

Start/Stop soft key membrane has worn out. Fair physical

condition.

471 Storz Telecam SL The video processor powers up but a camera head is not included

to perform function check.

478 Eschmann Little

Sister 3 Steam

Autoclave.

The unit power ok, the display and cycle selector switch are

working. Only 134 degrees C with and without drying are

activated. The door micro-switch and water level sensor are

working. The chamber and door seal is in fair condition. Test

cycle completed with no error. The temperature and pressure

maintained during sterilization phase. The printer is working but

needs new ribbon. Cycle count: 990. Unit is in fair physical

condition.

482 Eschmann TD

Electrosurgical Unit

Unit power up ok, the numerical display and RF power selector

keys are working fine. The ESU was tested using an ESU

analyser. The bipolar RF outputs were well within the set values.

However, the monopolar mode has no RF power output. The unit

is in average physical condition and monopolar and bipolar

footswitches are included. The build-in suction unit and light

source are working fine.

483 Stack Trolley with

Sony Trinitron

Monitor, Storz

Hysteroflater, Downs

Camera Controller

and 1457 Camera

head and Downs Light

Source

Light source and camera head processor powers up with its

function buttons and numerical display in working order. Sony

monitor powers up however an image is failed to obtain due to

faulty connections. No further tests were carried out. Camera head

not tested. Hysteroflator powers up. Overall the stack is in

reasonable physical condition.

490 Olympus CLV-S30

Light Source

Unit power up ok, Xenon lamp is working fine. Brightness

control, auto and manual mode fully functional. The backup

halogen lamp is also working. Unit is in good condition.

491 Eschmann TD850

Electrosurgical Unit

Unit power up ok, the numerical display and RF power selector

knob is working fine. The monopolar mode was tested using an

ESU analyser and some of the RF power measured were slightly

low than the set values for cut; COAG was ok. The bipolar RF

outputs were within the set values. The unit is in fair physical

condition and monopolay and bipolar footswitches are included.

492 Philips Heartstart XL

Defib x 2 with Patient

Both defibrillators power up and is able to pass through its initial

self-test successfully. Function buttons, selector knob and display

Page 27: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Lead x 2 are in working order. Units do not include ECG leads but one was

used to test. Units showed to be functional in both AED and

manual modes when tested with ECG simulator. Both are able to

charge and deliver shock but not fully tested. Audio voice

prompts and printer were functional. Battery not included. Both

units are in good physical condition.

500 Valleylab Force 2

Electrosurgical Unit

The ESU power up ok and the numerical displays and output

selector keys are working fine. The monopolar mode was tested

with an ESU analyser but no RF outputs were recorder on the

analyser. The unit is in fair physical condition but no RF output.

503 HP M3046A Viridia

M4 Patient Monitor,

HP Module Rack with

Modules, HP

Omnicare 24C

Monitor

M3046A unit powers up ok. No modules included but one was

used for testing. No error messages were displayed. ECG was

within limit. NIBP seems to be working but not fully tested. The

display and the touch buttons are good working order. Unit is in

fair physical condition with cracked case and scratches on the

display.

HP Omnicare unit powers up ok. Module rack and modules (x8)

included. ECG was tested with test leads and simulator and was

within limit. NIBP and SpO2 are working. No leads or accessories

included. Recorder included but has broken paper door. Overall in

satisfactory physical condition.

505 Doctors Friend DF-

750 Suction Unit x 1,

SAM 12 Suction Unit

x 1 and Weyer Mini-

vac ACU Paed Suction

Unit x 1 - 6672, 16744,

80021108

Weyer mini-vac and DF 750 suction units powers up ok.

Manometer and suction out is working but not verified. Suction

regulator is functional. DF 750 is relatively new (with box,

surgical connecting tube and manual) and is in good physical

condition. SAM 12 unit has its top plug cut off therefore not

tested. Both mini-vac and SAM 12 are in satisfactory physical

condition.

506 Assistina 301 Plus

Units x 3 - 19931,

19811, 12516

All three units are in satisfactory physical conditions. Unable to

test functionally. One unit is not complete.

515 Penlon Prima SP

Anaesthetic Machine

with AV900 Ventilator

and Sevoflurane

Vaporisers x 2

Unit powers up ok. The pressure gauges and flow meters are

working fine. O2 flush is functional however a major leak is

detected. Basal flow and anti-hypoxic feature is working. The

AV900 ventilator powers up and passed the initial self-test

successfully. The touch screen and display are working. The re-

breathing circuit and absorber are all intact. The ventilator is

functional. All the set parameters and alarm system is working.

The medical gas hose is included for the O2, Air and N2O. AGSS

hose included. Overall the lot is in fair physical condition with

scuffs and marks.

Page 28: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Software version: 3.32

522 Acuson Sequoia C512

Ultrasound with

Probe x 2 (7V3C,

3V2C)

Unit powers up ok. Display, control panel and function buttons

are in working order. Both probes are in good condition with

minor drop out to the crystals. One of the probe produced good

grey scale and fair resolution with good caliper measurements and

second one showed to have average grey scale and poor

resolution, when tested with phantom. The unit seemed functional

in D, PW, CW and M modes. ECG leads not included and

therefore not tested. Sony video cassette recorder and Sony video

graphic printer powers up ok. Overall unit is in reasonable

physical condition.

532 Datex Ohmeda ADU

Carestation

Anaesthetic Machine

with Isoflurane

Vaporiser

Unit powers up ok. The pressure gauges are working fine. O2

flush is working. The ventilator powers up and passed the initial

self-test successfully. The re-breathing circuit is disconnected and

absorber is missing. The ventilator is working. All the set

parameters and alarm system is functional. The medical gas hose

is included for the O2, Air and N2O. AGSS hose included. Patient

monitor power up and completed its initial self-test. Its display

and function buttons are in working order. SpO2 was tested and is

working. Multi parameter module, NIBP module and gas module

included but not tested. SpO2 probe, NIBP trunk cable and ECG

trunk cable included. Overall the lot is in fair physical condition

with scuffs and marks.

535 Eschmann Little

Sister 2 Steam

Autoclave.

The unit power ok, but the overheat indicator is on. No further test

was conducted. Cycle count: 16979. Unit is in poor physical

condition.

536 B Braun Dialog Plus

Dialysis Machine x 4 -

35209, 21220, 23115,

23152

All units are able to power up, with their function buttons and

touchscreen displays functional. All menus accessible. The units

are not fully tested. BP not tested. All of the unit’s double blood

pumps are intact except for one with missing pump, as are the

heparin pumps. Units also include one air detectors which again

are also intact. Does not include water and drainage hoses.

Overall the dialysis machines are in average physical condition.

Version: 8.28 (x4)

Operation hours: 35136, 35209, 28877, 31762

539 Drager Phototherapy

Lamp 800

The unit power up ok, the fan and lamp are working fine.

Operating hours: 33882hrs. Average physical condition.

Page 29: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

540 Hill-Rom Birth Bed All motorised positions are working ok (height, back rest and leg

adjustment). Castors, brake and mattress in good condition. The

bed is not working on battery.

541 Hill-Rom Birth Bed All motorised positions are working ok (height, back rest and leg

adjustment). Castors, brake and mattress in good condition. The

bed is not working on battery.

542 Hill-Rom Birth Bed All motorised positions are working ok (height, back rest and leg

adjustment). Castors, brake and mattress in good condition. The

bed is not working on battery.

545 Braun SCT 2

Tournequet x3

All three tourniquets were tested with air pressure and the gauges

and regulator are working fine. Units are in fair physical

condition.

553 Philips Ambulatory

ECG Recorders x 7 -

3839, 3833, 3823,

3820, 3832, Innervator

NS252 Stimulator and

Nonin 2120 Monitor -

1191, 505080

Nonin 2120 power up ok. Error E17 is displayed. No further

tested conducted. ECG recorders M2601B (x7) are in fair physical

conditions. Nerve simulator is in poor condition with worn out

buttons and broken parts.

556 Dinamap Pro 400/

Criticare Poet Plus

8100 monitor

The Dinamap power up ok, display and function keys working.

NIBP tested ok. However the unit is in poor cosmetic condition

with mold on the NIBP tubing and SpO2 probe. The Criticare

monitor power up ok and in better condition, not fully tested.

558 ASP Tournequet SCT

2

Unit is in below average physical condition, the regular and

manometers seem to be ok. Unit not tested with medical air.

563 Agilent Pagewriter

100 ECG Machine on

Trolley with Leads

Unit powers up ok and all the LEDs are lit up. Function buttons

are working. ECG waveforms can be acquired using the included

leads and the analyser. The recorder is working and the print is

clear. Unit is in satisfactory physical condition with scuffs and

marks.

577 Drager Phototherapy

Lamp x2

Both units power up ok, the fan and lamp are working fine.

Operating hours: 07271 & 28331hrs. Fair physical condition.

579 Satelec X Mind

System Dentak X Ray

system on mobile

stand

Unit power up and all the indicators, selector keys and timer

numerical display are working fine. The unit was tested and x ray

was present at the output. The power ON/OFF switch is slightly

loose but functional and there two missing castors on the mobile

stand. Unit is in fair condition.

Page 30: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

580 Smith & Nephew

Stack System, 350

CCD Camera

Processor with

Camera Head x 3

(One Incomplete) ,

Xenon XL Light

Source, Sony Colour

Video Printer and

Sony Trinitron CRT

Monitor

All units within the stack system are in good working order. The

video processor and camera head produces good quality images

and all 3 camera heads are functional (one of the camera heads do

not include the lens/adaptor). The light source is fully functional

and the printer also appears to be working (out of paper). Overall

the stack system is in reasonable physical condition.

598 GE Dash 4000 Patient

Monitor x 2

One of the monitors power up successfully and the display

panel/function keys are in working order. The LCD is missing a

few lines of pixels and the front frame has been damaged. The

other monitor powers up but fails to complete the start up process

or reach the home screen. The casing has few small cracks and the

front panel buttons have been worn out. No accessories included

with units.

607 Penlon Sigma

Isoflorane Vapouriser

Vapouriser is in very good condition and calibrated 24/04/15.

Calibration certificate attached.

608 Penlon Sigma

Isoflorane Vapouriser

Vapouriser is in very good condition and calibrated 24/04/15.

Calibration certificate attached.

609 Datex Ohmeda

Critical care monitors

Only one of the monitor was setup and power up ok. The lot

consists of x4 module racks (no modules), x4 LCD monitor with

only one control with DVI cable; mains adapter included, bags

containing NIBP hose & cuff, SpO2 finger probes and ECG

cables. Lot is in fair physical condition.

618 Drager Infant

Incubator 8000 IC

Power up and self-test ok. The control for air temperature mode,

oxygen monitor and humidity are working. The air mode was

tested and the set temperature was maintained accordingly and the

external thermometer reading was within limit. Humidity control

is also functional and glass water bottles are fitted at the back.

The oxygen sensor completes calibration process. The canopy,

mattress and incubator frame is in good physical condition. The

motorised height adjustment is fully functional. Oxygen hose

attached.

619 Drager Infant

Incubator 8000 IC

Power up and self-test ok. The control for air temperature mode,

oxygen monitor and humidity are working. The air mode was

tested and the set temperature was maintained accordingly and the

Page 31: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

external thermometer reading was within limit. Humidity control

is also functional and glass water bottles are fitted at the back.

The oxygen sensor completes calibration process. The canopy,

mattress and incubator frame is in good physical condition. The

motorised height adjustment is fully functional. Oxygen hose

attached.

622 Dinamap Pro

300/300V2/100 &

compact TS

All four NIBP monitor power up ok, display and functions keys

are working; a few of the soft keys has worn out but functional.

The NIBP was tested on all the units and was working fine. The

Compact TS unit SpO2 is working ok but the finger probe spring

is damaged. Lot is in fair physical condition.

625 Keeler Cyromatic

Cryounit with Foot

Pumps and

Connection Leads

Powers up and the display panel is functional. Unable to carry out

any further checks. The unit is in good physical and cosmetic

condition. Footswitch and 2 probes/handpieces are included. The

footswitch connector is slightly damaged.

642 Zeiss ENT Operating

Microscope

The light source power up ok, bulb is working. Good optical

system with eye pieces and objective lens. The motorised zoom &

focus control and joystick is damaged. Average physical

condition.

643 HGM Elite

Ophthalmic Laser

System with Slit Lamp

The Slit lamp power supply, bulb, fixation light and brightness

control are in good working order. All the slit lamp movement

and lock were functional. The motorised table height adjustment

is working fine. The Elite laser system power up ok the display

and laser parameter selector are working. The unit was tested on

thermal paper and there was indication of laser present at the

output. However the laser output power was not verified. Safety

goggles x3, Laser Key, footswitch and carry case included. Unit is

in good physical condition..

644 Leica CM 1100-

Kryostat on KE

Standard Supertrans

Demtruk Trolley

Leica CM 1100 powers up ok. Its control unit, hand wheel and its

lever, coarse feed wheel and the chamber lid is in good condition.

The clamping lever, knife holder, waste tray and section waste

tray are included. No further tests were carried out. Temperature

and defrost time can be adjusted. The compressor is slightly

noisy. Overall lot is in good physical condition.

646 GE Capnostat CO2

Sensors

Lot is in good physical condition with adult x3 & neonatal x2

away adapters. Sensor not tested.

649 GE E-Prestn

Multiparameter

Modules x 2 - 016985,

E-PRESTN (x2) modules are in satisfactory physical condition

with scuffs and marks. SpO2, temperature, pressure, NBP and

ECG parameters were not tested.

Page 32: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

016987 YOM: 2007 (x2)

672 Olympus Stack system

with processor CV-

230 & Light source

CLV-U40

The light source power up ok, the xenon lamp is on and

brightness, air pump and auto/manual working fine. The processor

was tested with a camera head but no scope to determine the

picture quality. The Sony monitor is not working properly, the

video channel fail to switch over or change. Average overall

physical condition.

689 Lifepak 20 & 12 with

charger and leads

All Lifepak 20 power up and self-test completed successfully. The

LCD and all function buttons are in good working order. The

defibrillator was tested using a defibrillator analyser and the ECG

traces and delivered energy was within the set limit. Both AED

and manual defibrillation modes are working and so is the

monitor. The voice prompt is also working fine. Lot is in good

overall condition. Backup battery is not working and there is no

pacer option or paddles. The service light is on.

All Lifepak 12 power up and self-test completed successfully. The

LCD and all function buttons are in good working order. The

manual defibrillator was tested using a defibrillator analyser and

the ECG traces and delivered energy was within the set limit. The

monitor is also working fine. Lot is in good overall condition. No

battery packs or paddles included.

The two Medtronic Battery Support System 2 power up ok and

self-test completed. Units are in good condition but not fully

tested. The lot also come with AED therapy cables x8 and ECG

cable (3 wires) x7

704 Philips Heartstart

AED

All AED power up and self-test completed successfully. The LCD

and function buttons are in good working order. The AED was

tested using a defibrillator analyser and the ECG traces and

delivered energy was within the set limit. The voice prompt is

also working fine. Lot is in good overall condition. Battery packs

are not included.

705 Philips Heartstart

AED

All AED power up and self-test completed successfully. The LCD

and function buttons are in good working order. The AED was

tested using a defibrillator analyser and the ECG traces and

delivered energy was within the set limit. The voice prompt is

also working fine. One has a poor casing but the rest are in good

overall condition. Battery packs are not included.

Page 33: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

706 Philips Heartstart

AED

All AED power up and self-test completed successfully. The LCD

and function buttons are in good working order. The AED was

tested using a defibrillator analyser and the ECG traces and

delivered energy was within the set limit. The voice prompt is

also working fine. One has a poor casing but the rest are in good

overall condition. Battery packs are not included.

724 Sonosite Scanner ilook The unit power up ok, the LCD and function soft keys are

working fine. Good image obtained with the linear probe. The

modes, CPD & 2D are functional. Charger is included. Unit is in

good physical condition. A lock screw for the unit tilt is missing.

2 battery packs included.

726 Critikon Dinamap

Vital Signs Monitor

8100T

All power up & display self-test ok but follow with continuous

audible alarm. Unit not working correctly. Fair physical

condition

732 Drager Infant

Incubator 8000 SC

Power up and self-test ok. The Control for Skin and air modes is

working. The skin probe is not included there this mode was not

fully tested. The air mode was tested and the set temperature was

maintained accordingly and the external thermometer reading was

within limit. Humidity control is also functional. The canopy,

mattress and incubator frame is in good physical condition. The

motorised height adjustment is fully functional.

736 Tournequet Unit is in average physical condition. The timer comes on; the

connection for the medical air is missing. The unit was not fully

tested with air pressure. The regulator and manometers appears to

ok.

744 Sonosite Ultrasound

system 180 Plus

The unit power up ok, the LCD and keyboard is working but there

are some of the soft keys not responding. Good image obtained

with the convex and linear probe. The modes, D, CPD, M, & 2D

are functional. Charger is included. Unit is in good physical

condition.

754 GE Dinamap Pro 400

x4

All power up ok, display and function keys are working. BP was

tested and working; SpO2 function not fully tested. Lot comes

with NIBP hose and cuff, most of the cuff Velcro has worn out.

Units are in fair physical condition

755 GE Marquette 2000

Treadmill with GE

Case Stress Best Unit

with Suntech Tango

The unit powers up successfully. Unable to boot up due to no

operating system found. Treadmill powers up ok. Acquisition

module and connection cable RS232 to the treadmill and ECG

unit is not included. No further test carried out. Unit is in

Page 34: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

BP Monitor satisfactory physical condition.

780 Eschmann Little

Sister 2000 Steam

Autoclave.

The unit power ok, the display and cycle selector switch are

working. The door micro-switch and manual water fill are

working. The chamber and door seal is in fair condition. Test

cycle completed with no error. The temperature and pressure

maintained during sterilization phase. Unit is in fair physical

condition.

783 Esaote AU4 Idea

Ultrasound with

Probes x 2 (LA14-

10Mhz and LA13-

7.5Mhz)

The unit power up ok, the display and control keypad is working.

Good image obtained with two probes. The modes 2D, PW and M

are functional. No further test carried out. Unit is in good physical

condition.

789 Hill-Rom Birth Bed All motorised positions are working ok (height, back rest and leg

adjustment). Castors, brake and mattress in good condition. The

bed is not working on battery, needs charging. Good overall

condition.

790 B Braun Dialog Plus

Dialysis Machine x 3 -

22023, 22027, 21192

All units are able to power up, with their function buttons and

touchscreen displays functional. All menus accessible. The units

are not fully tested. BP not tested. All of the unit’s double blood

pumps are intact except for one with missing pump. Units also

include one air detectors which again are also intact, as are the

heparin pumps. Does not include water and drainage hoses.

Buttons for one of the unit is slightly worn out. Disinfection

documents attached. Overall the dialysis machines are in average

physical condition.

Version: 8.28 (x3)

Operation hours: 38900, 22847, 32134

794 Stryker X7000

Lightsource, Stryker

1088i HD Camera

Control Unit with

Camera Head,

Stryker Flow Control

Arthroscopy Pump

and Stryker Core

Powered Instrument

Driver, Sony Printer -

508661, 015215,

1088i HD camera processor powers up ok and is able to produce

good image when tested with the camera head and the monitor in

lot 879. The camera control buttons are operational. X7000 light

source is able to power up but lamp is missing therefore unable to

fully test. Instrument driver, arthroscopy pump and Sony video

graphic printer all power up ok. Not fully tested. The touch screen

and function buttons of the instrument driver is working in order.

No footswitches or hand pieces included. The knob for locking

the optic cable on the light source is missing. Monitor not

included. Overall the lot is in good physical condition.

Page 35: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

015205, 508665,

015210

812 Datex Ohmeda S/5

Monitor

Only two of the monitor power up. Self-test completed and the

LCD is working fine. But the second one has a green bar on the

left hand side of the display. The complete working was tested

with patient simulator and the ECG and Spo2 was both working

fine. X2 with front panel damage and missing knobs.

816 Dinamap Pro Datex

Cardiocap II Patient

Monitor

The monitor power up ok, the CRT display is working fine. ECG

& SpO2 was tested using a patient simulator and all the reading

and traces displayed were with set limits. There was an error

message ‘memory circuit failure’ flashing on the display. Unit is

in fair physical condition.

830 Eschmann Little

Sister 3 Steam

Autoclave.

The unit power ok, the display and cycle selector switch are

working. Only 134 degrees C with and without drying are

activated. The door micro-switch and water level sensor are

working. The chamber and door seal is in fair condition. Test

cycle completed with no error. The temperature and pressure

maintained during sterilization phase. Cycle count: 2511. Unit is

in fair physical condition with a bit of rust on the door frame.

831 Eschmann Little

Sister 3 Steam

Autoclave.

The unit power ok, the display and cycle selector switch are

working. Only 134 degrees C with and without drying are

activated. The door micro-switch and water level sensor are

working. The chamber and door seal is in fair condition. Test

cycle completed with no error. The temperature and pressure

maintained during sterilization phase. Cycle count: 8264. Unit is

in fair physical condition with a bit of rust on the door frame.

837 Leica Microscope The unit has no eye pieces or objective lens. The power supply is

missing as well. Poor condition.

838 Leica Microscopes x2 The microscopes are in poor condition with the optics dirty. One

of them with a corroded stage. Come with some objective lens

and eye pieces.

847 Datex Cardiocap II

Patient Monitor

The unit failed to power up.

868 Drager Infant

Incubator 8000 SC

Power up and self-test ok. The Control for Skin and air modes is

working. The skin probe is not included there this mode was not

fully tested. The air mode was tested and the set temperature was

maintained accordingly and the external thermometer reading was

within limit. Humidity control is also functional. The canopy,

mattress and incubator frame is in good physical condition. The

Page 36: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

motorised height adjustment is fully functional.

873 GE Vingmed System

Five Performance

Ultrasound with 5Mhz

x 1, 3.5Mhz x 1,

2.5Mhz x 1 and

10Mhz Probe x 1,

Footswitch and VCR -

YZ7112

Ultrasound powers up but an error message is displayed and the

system fails to proceed to main screen. No further tests carried

out. Printer and videocassette recorder powers up. Reasonable

physical condition.

876 Operating Table

accessories

Accessories in fair physical condition consisting of Anetic Aid leg

support x2 of different shape, Maquet Beta leg section, spare

cushion & one arm support.

879 Stryker Stack Trolley

with SV-2 HD

Monitor

Monitor is able to power up ok. The screen is slightly scraped at

the middle with few scratches but does not seem to affect the

image when tested with processor in lot 794. The stack and the

monitor are in satisfactory physical condition.

883 Smith & Nephew

Trolley with Sony

LCD Monitor, Storz

Trolley with Sony

LCD monitor

Both Sony LCD monitors power ok, good display and menu

functional. The trolleys are in good physical condition. Sony LCD

monitor Model: MD-214MD (100-240V) & MD1950MD/HD

(24V DC), mains adapter included.

890 Olympus

EndoscopeTrolley

The trolley is in fair physical condition, the pole fitting is missing,

and castors are ok.

899 Olympus CLV-U40

Light Source

The light source power up ok, The Xenon lamp is working,

brightness control is functional and all the light output modes are

working ok. The air pump is functioning fine. Unit is in good

physical condition.

908 Eschmann Little

Sister 2000 Steam

Autoclave.

The unit power ok, the display and cycle selector switch are

working. The door micro-switch and water level sensor are

working. The chamber and door seal is in fair condition. Test

cycle completed with no error. The temperature and pressure

maintained during sterilization phase. Unit is in fair physical

condition.

937 Datex Ohmeda S/5

patient monitor

Lot consists of LCD monitor & control pad x9, Module racks x9

M-NESTPR modules x6, M-PP module x4 & M-PT modules x6.

There is also a S/5 Light Monitor. It power up ok and the display

and function keys are working. S RAM error is displayed. Some

of the setup has been lost, with the SpO2 plug in there was no

Page 37: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

response on the screen. Fair & physical condition but lot has no

interconnecting leads or patient cables.

978 Humphrey 350 Lens

Analyser - 303076

Powers up and the display panel/function keys are in good

working order. All system menus are accessible, no further checks

performed. The unit is in good physical and cosmetic condition.

Model: 350. Version: A3.

982 Datex Ohmeda S/5

Monitor with qty of

accessories

The lot consists of the following: Module rack x10, LCD

monitors x3, M-PRESTN module x10, M-NESTPR module x5,

M-CAIO x3, M-COPSv x2 & M-miniC x4. Quantity of ECG

leads, SpO2 finger probe and extension cable, inter connecting

cables, NIBP hose & cuffs. Only one of the patient monitor was

set up and tested and appears to be working fine. The ECG, SpO2

and NIBP were tested and were functional. The lot is in fair

physical condition, not all units were tested.

985 Nidek KM-500

Keratometer and

Datex M-Caio Module

Mains adaptor for the charger is not included. Unable to perform

any function check. Both the keratometer and the M-CAiO

module are in reasonable physical condition.

1003 Storz Stack trolley

with LCD monitor

The trolley power up ok and the LCD is working fine. The trolley

and monitor is in good physical condition. Keyboard and HP

printer included.

1059 Kodak Point of Care

CR140- X-Ray Printer

- 800195

X-ray printer powers up ok. No workstation included therefore

unable to fully test. No x-ray cassettes included. Overall lot is in

fair physical condition.

1076 Siemens VF13-5

Ultrasound Probe -

F13C055335

The probe is in satisfactory physical condition with damages at

the corners and a slight cut to the membrane of the probe head.

Probe is not fully tested as there is no suitable ultrasound unit to

test with.

1117 Stryker Insufflator The unit power up but with error messages – ‘Call Service’,

‘Contamination’ & ‘Device defective’. Unit not working properly.

Cylinder connection missing.

1142 Medelec Thermal

Threshold Tester -

6438, Agilent Docking

Stations x 2 and

Philips Modules x 2 -

11900, 11870, 15409,

Ardo Calypso Breast

Pump, Freewat Elite

Threshold tester power up ok. Not tested. Fair physical condition.

M3015A Philips module is in good physical condition. ECG,

Pressure, temperature, NBP and SpO2 parameters were not tested.

YOM: 2004

Page 38: Endoscopes Inspection Report Sale: 1505 Date: 12 / 05 ... Report Sale 1505.pdf · Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015 (Condition Reports for other items

Cpap, Assorted

Thermometers x 3 -

573432, 512656,

512657