EOPM Tide System High Accuracy Barometrically Compensated Tide Measurement System
with Excell ® Oceanographic Pressure Module and Internal Data Logging
User Manual
March 2005
P/N A177-030
Falmouth Scientific, Inc.1400 Route 28A
Cataumet, MA 02534
Tel: (508) 564-7640Fax: (508) 564-7643
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 2/24
EOPM Tide System User Manual March 2005
ii
The information, descriptions and illustrations in this manual are the property of
Falmouth Scientific, Inc. (FSI), and may be disclosed only to those persons or
organizations specifically authorized by FSI. Duplication of any part of this
manual is strictly prohibited unless authorized in writing by FSI.
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 3/24
P/N A177-030
iii
About This ManualThis manual provides a general overview and the installation and setup
instructions for the Falmouth Scientific EOPM Tide System. The EOPM Tide
System provides accurate, barometrically compensated remote measurements of tide by integrating a Falmouth Scientific Excel ® Oceanographic Pressure Module
(EOPM) with a barometer. Acquired data are stored in the EOPM and can be
downloaded at any time to a computer or transmitted over an RF link.
For additional information about the EOPM instrument, refer to the “EOPM
Operation Manual” included with the system.
Customer ServiceFSI welcomes your feedback. Please contact FSI customer service to offer any
comments or suggestions or to request technical support. FSI can be contactedusing any of the following means:
FSI, Inc.1400 Route 28A
Cataumet, MA 02534
Telephone or Facsimile
Tel: (508) 564-7640Fax: (508) 564-7643
For more information about Falmouth Scientific, please visit our Web site at
www.falmouth.com.
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 4/24
EOPM Tide System User Manual March 2005
iv
Table of ContentsAbout This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
SECTION 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
EOPM Main External Components . . . . . . . . . . . . . . . . . . . . . . 1-1
NEMA Enclosure Main Components . . . . . . . . . . . . . . . . . . . . 1-3
SECTION 2: Unpacking and Setup . . . . . . . . . . . . . . . . . . . . . . . 2-1
Unpacking the EOPM Tide System . . . . . . . . . . . . . . . . . . . . . 2-1
Standard Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Optional Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Installing and Starting CTDPro . . . . . . . . . . . . . . . . . . . . . . . . . 2-2Installing CTDPro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Starting CTDPro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Communication Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
RS-232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
RS-485 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Setting up and Connecting the EOPM TideSystem Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Mounting the NEMA Enclosure . . . . . . . . . . . . . . . . . . . . . . 2-4
Connecting the EOPM Tide System with theRS-232 Serial Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Connecting the EOPM Tide System with theRS-485 Serial Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
SECTION 3: Acquiring, Downloading and Viewing Data . . . . . . 3-1
SECTION 4: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
NEMA Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
EOPM Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
SECTION 5: Connector Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 5/24
1-1
P/N A177-030
SECTION 1: Introduction
he Falmouth Scientific EOPM Tide System integrates a FalmouthScientific Excel ® Oceanographic Pressure Module (EOPM) with a
barometer to acquire accurate tide data in all atmospheric pressure
conditions. The EOPM, which is deployed underwater within a specified depth
range, acquires pressure data using a highly accurate, highly stable and fully
temperature compensated micromachined silicon diaphragm pressure sensor. The
barometer is contained, along with a battery and battery charger, in a remote,
surface mounted NEMA enclosure.
The main components of the system are shown in Figure 1-1 and consist of the
following;
• EOPM
• NEMA enclosure
• Serial cable
• Instrument cable
The EOPM connects to the NEMA enclosure with a 20-m instrument cable. The
NEMA enclosure is AC powered and supplies DC power to the EOPM and
includes a battery and battery charger for power backup should the AC power be
disconnected. The EOPM acquires and stores in 128 MB (256 MB optional)
memory both pressure and barometric data in separate data files. The data arerecorded continuously or at preset intervals at a sample rate of 20 Hz. For each
sample the time and date are also recorded. The data can be downloaded to a
computer at any time over an RS-232 or RS-485 serial interface at baud rates up to
115200 (over a short instrument cable), or output on an RF link. The computer
connects directly to the NEMA enclosure with a 2-m serial cable.
The EOPM is configured over the serial interface connection to the NEMA
enclosure using CTDPro, a Microsoft Windows® based software program
included with the system. With CTDPro you can also download, save and view
water pressure and barometric pressure data, processed tide data, and time and date
information. The data can also be exported to spreadsheet programs.
EOPM Main External ComponentsThe main external components of the EOPM are shown in Figure 1-1. The
pressure sensor is exposed to ambient pressure through a pressure port and an
oil-filled, removable capillary. The oil provides the required pressure coupling and
T
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 6/24
1-2 SECTION 1 Introduction
EOPM Tide System User Manual March 2005
Figure 1-1: The EOPM Tide System Components
Instrument cable
Pressure sensor portCapillary tube
(Capillary tube removed)
8-pin bulkhead
Dummy plug
connector
NEMA enclosure
AC cordSerial cable
EOPM
Instrumentconnector
Serialconnector
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 7/24
1-3NEMA Enclosure Main Components
P/N A177-030
is an electrical insulator that prevents corrosion due to the dissimilar metal
compositions of the transducer and other metals. Capillary action ensures the oil
remains contained in the pressure port. A single bulkhead connector on the lower
end cap provides RS-232 and RS-485 digital communications and inputs external
power. A dummy plug is provided for connecting to the bulkhead connector whenit is not in use.
NEMA Enclosure Main ComponentsThe main external components of the NEMA enclosure are shown in Figure 1-1
and are the following:
Instrument connector. An 8-pin connector that connects directly to the
EOPM with the 20-m instrument cable.
Serial connector. A 5-pin connector that connects directly to a computerwith the 2-m serial cable.
AC cord. A 3-wire, 1.5-m long cord that connects to AC power.
Instrument cable. Connects to the instrument connector and to the EOPM.
Serial cable. Connects to the serial connector and to a computer.
The main internal components of the NEMA enclosure are shown in Figure 1-2
and are the following:
Battery. Provides backup power should the AC power be disconnected and is
kept charged by the battery charger.Battery charger. Charges the battery or maintains a float charge whenever
AC power is connected.
120/240 VAC switch. Selects either 120 VAC or 240 VAC operation. The
switch is located on the battery charger under the terminal strip and is factory
set to 120 VAC. The terminal strip must be removed to access the switch.
Barometer. Measures barometric pressure and outputs the pressure as an
analog signal to the EOPM for analog-to-digital conversion and storage in
EOPM memory.
Air vent. Provides atmospheric pressure input to the barometer.
Terminal strip. Provides all the required connections to the internal
components of the NEMA enclosure, to the EOPM and to a computer or RF
link.
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 8/24
E OP M T i d e S y s t em U s er M an u al
M ar ch 2 0 0 5
Figure 1-2: The Main Internal Components of the NEMA Enclosur
1 8
Terminal strip
120/240 VAC switchlocated on batterycharger under terminal
Air vent forbarometer
strip
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 9/24
2-1
P/N A177-030
SECTION 2: Unpacking and Setup
he NEMA enclosure of the EOPM Tide System must be connected to ACpower and to the EOPM. To configure the EOPM and to download data
from it, the NEMA enclosure must also be connected to an available serial
port of a computer running CTDPro. The default serial port that CTDPro uses is
COM1, and the default baud rate is 9600 bits/sec. However, any serial port
from COM1 to COM16 can be used, and baud rates of 300, 1200 and
19200–115200 bits/sec are additionally available.
Note: For baud rates of 19200 and higher, an instrument cable that is shorter
than 20 meters is recommended.
This section provides instructions on how to unpack the EOPM Tide System, how
to install and start CTDPro on your computer, and how to set the system up and
connect it to your computer using the RS-232 or RS-485 interface.
Unpacking the EOPM Tide SystemBefore unpacking the EOPM Tide System, check the shipping container for signs
of external damage. If the container appears damaged, report the damage to FSI
and to the freight carrier.
When unpacking the EOPM Tide System, inspect all the items for any apparent
damage and verify that all the items listed in the packing list are included in theshipment. Report any damage or missing items to FSI.
Standard Items
The following items are included with the EOPM Tide System:
• EOPM
• NEMA enclosure
• Instrument cable
• Serial cable
• CTDPro for Windows CD
• Dummy connector
• EOPM Operation Manual
• This manual
T
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 10/24
2-2 SECTION 2 Unpacking and Setup
EOPM Tide System User Manual March 2005
Optional Items
The following item is optionally available for an RS-485 connection to a
computer:
• RS-485 Interface (includes converter box, power supply, andDB25 to DB9 serial port adapter cable)
Installing and Starting CTDProCTDPro must be installed on the computer to which the NEMA enclosure will be
connected. CTDPro is used to configure the EOPM, to download and save water
pressure and barometric pressure data from the instrument, and to process this data
into tide data.
Installing CTDProTo install CTDPro:
Exit all programs and turn off any virus protection or screen saver software.
Insert the CTDPro for Windows CD into your CD-ROM drive.
Select Start➤ Control Panel ➤ Add or Remove Programs.
The Add or Remove Programs dialog box opens.
Click Add New Programs, and then click CD or Floppy and follow the
instructions to install the software.
Starting CTDPro
To start CTDPro:
Select Start➤ All Programs➤ FSI Applications.
The FSI Applications menu opens.
Choose CTDPro.
CTDPro starts and the CTDPro Main window opens.
1
2
3
4
1
2
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 11/24
2-3Communication Interfaces
P/N A177-030
Communication InterfacesThe EOPM Tide System includes two communication interfaces for configuring
and downloading data from the instrument. The instrument is normally wired for
an RS-232 serial interface but can be wired for an RS-485 serial interface. (See“Connecting the EOPM Tide System with the RS-485 Serial Interface” on page
2-6 for instructions on how to wire the system for the desired communication
interface.)
RS-232 Interface
The RS-232 serial interface requires that your computer provide true RS-232 data
levels. Some older laptop computers do not support the negative output voltages
provided by the RS-232 standard and these computers may not work properly with
the instrument.
RS-485 Interface
The RS-485 serial interface requires the use of a converter box which is provided
with the optional RS-485 Interface. (See “Optional Items” on page 2-2.) Due to
specialized timing requirements, generic converters will not work properly. The
converter box that is supplied with the option must be used with the EOPM Tide
System.
Setting up and Connecting the EOPM Tide
System ComponentsIn addition to the components supplied with the EOPM Tide System, the following
items are required to set up the system, deploy the EOPM and connect the NEMA
enclosure to your computer:
• Mounting location for the NEMA enclosure
• Male DB25 to female DB9 serial port adapter cable
• An available serial port on the computer
• A 100–125 VAC or 220–240 VAC, 50–60 Hz power source
• RS-485 Interface (includes converter box, power supply, and DB25 to DB9
serial port adapter cable), for RS-485 interface connection only
Note: If you will be fabricating your own cables for connecting the EOPM and
your computer to the NEMA enclosure, see “SECTION 5: Connector Wiring,” for
wiring information on the instrument and serial connectors.
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 12/24
2-4 SECTION 2 Unpacking and Setup
EOPM Tide System User Manual March 2005
Mounting the NEMA Enclosure
Mount the NEMA enclosure in a suitable location, indoors or out, where the AC
power source is available and where the instrument cable can be connected to the
EOPM. The overall dimensional outline and mounting hole pattern for the NEMA
enclosure is shown in Figure 2-1.
Figure 2-1: NEMA Enclosure Overall Dimensional Outline and Mounting Hole Pattern
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 13/24
2-5Setting up and Connecting the EOPM Tide System Components
P/N A177-030
Connecting the EOPM Tide System with theRS-232 Serial Interface
A EOPM Tide System is shown set up with an RS-232 serial interface in
Figure 2-2. To set up the EOPM Tide System with an RS-232 serial interfaceconnection:
Disconnect the dummy connector on the EOPM.
Connect the instrument cable to the instrument connector on the NEMA
enclosure and to the bulkhead connector on the EOPM.
Deploy the EOPM as required.
Connect the serial cable to the serial connector on the NEMA enclosure and
to the DB25 connector of the DB25 to DB9 serial port adapter cable.
Connect the DB9 connector of the DB25 to DB9 serial port adapter cable to
the serial port on the computer.
Connect the AC cord to the AC power source. The black wire is AC high, the
white wire is AC neutral, and the green wire is ground.
Figure 2-2: EOPM Tide System Setup—with the RS-232 Interface
1
2
3
4
5
6
Laptopcomputer
EOPM
NEMA
AC cord
Serial cable
enclosure
Instrumentcable
DB25 to DB9serial port adaptorcable
Instrumentcable
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 14/24
2-6 SECTION 2 Unpacking and Setup
EOPM Tide System User Manual March 2005
Connecting the EOPM Tide System with theRS-485 Serial Interface
A EOPM Tide System is shown set up with an RS-485 serial interface in
Figure 2-3. When connecting the EOPM Tide System using the RS-485 serialinterface, the NEMA enclosure wiring configuration must also be changed. To set
up the EOPM Tide System with an RS-485 serial interface connection:
Open the NEMA enclosure.
Disconnect the wire from Terminal 2 of the terminal strip and connect it to
Terminal 5.
Disconnect the wire from Terminal 4 of the terminal strip and connect it to
Terminal 6.
Close the NEMA enclosure.
Disconnect the dummy connector on the EOPM.
Connect the instrument cable to the instrument connector on the NEMA
enclosure and to the bulkhead connector on the EOPM.
Figure 2-3: EOPM Tide System Setup—with the RS-485 Interface
1
2
3
4
1
2
Powersupply
Laptopcomputer
EOPM
DB25 to DB9
adaptor cableserial port
AC cord
NEMA enclosure
boxConverter
Instrumentcable
Serial cable
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 15/24
2-7Setting up and Connecting the EOPM Tide System Components
P/N A177-030
Deploy the EOPM as required.
Connect the serial cable to the serial connector on the NEMA enclosure and
to the DB25 input connector on the converter box.
Connect the DB25 output connector of the converter box to the serial port on
the computer using the DB25 to DB9 serial port adapter cable.
Connect the power supply to the converter box and to the AC power source.
Connect the AC cord to the AC power source. The black wire is AC high, the
white wire is AC neutral, and the green wire is ground.
3
4
5
6
7
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 16/24
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 17/24
3-1
P/N A177-030
SECTION 3: Acquiring, Downloading and
Viewing Data
nce the EOPM Tide System is set up, the EOPM can be configured and
data acquisition started. After data acquisition is complete, the data can
be downloaded and saved using CTDPro. CTDPro also allows viewing
of water pressure and barometric pressure data, processed tide data, and time and
date information. Data can be viewed in graphs and exported to spreadsheet
programs for tabular viewing and further processing.
To configure the EOPM and start data acquisition:
Start CTDPro as described in “Starting CTDPro” on page 2-2.
CTDPro starts and the CTDPro Main window opens.
Select EOPM from the menu bar, and then choose Configure and Deploy.
Follow the instructions provided to configure the EOPM and start data
acquisition.
To download, view and save the data acquired by the EOPM:
Start CTDPro as described in “Starting CTDPro” on page 2-2.
CTDPro starts and the CTDPro Main window opens.
Select EOPM from the menu bar, select Memory, and then choose View
Files.
Follow the instruction provided to download, view and save the data.
1
2
3
1
2
3
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 18/24
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 19/24
4-1
P/N A177-030
SECTION 4: Specifications
elow are the general specifications for the EOPM Tide System. TheNEMA enclosure and the EOPM instrument are listed separately.
Note: These specifications are subject to change without notice.
NEMA Enclosure
Power requirements: 100–125 VAC or 220–240 VAC,
50–60 Hz (factory set with switch under
terminal block; see Figure 1-2 on page
1-4)
Power connection: 3-wire AC cord, stripped end for
terminating
Battery backup time: 16 days typical
AC cord length: 1.5 m (5 ft)
Instrument connector: Subconn MCBH8M; mates with
MCIL8F on instrument cable
Instrument cable length: 20 m (65.6 ft)
Serial connector: Subconn MCBH5F; mates with
MCIL5M on serial cable
Serial cable length: 2 m (6.6 ft)
Dimensions: 28.4 cm (11.2 in.) long by 20.8 cm
(8.2 in.) wide by 14.2 cm (5.6 in.) deep
Weight: 7.45 kg (16.4 lb), including AC cord
Construction: High-impact plastic, watertight with
cover closed
Mounting: (4) 10-32 tapped holes, one each corner
Barometric pressure range: 600–1060 hPa
Data acquisition software: CTDPro running on Microsoft
Windows® 98/NT/2000/XP
B
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 20/24
4-2 SECTION 4 Specifications
EOPM Tide System User Manual March 2005
EOPM Instrument
Power requirements: 8–30 VDC max; 210 mw @ 8 VDC
Depth rating: 500 m
Bulkhead connector: Subconn MCBH8F; mates with
MCIL8M on instrument cable
Communication interface: RS-232 or RS-485
Sample rate: 1–20 S/sec, selectable
Resolution: 20 bits
Modes of operation: Continuous, Interval or Delayed Interval
Weight in air: 0.57 kg (1.26 lb)
Weight in water: 0.02 kg (0.50 lb) buoyant
Real time clock: Programmable alarm/sleep functions
Real time clock stability: ±20 ppm/year, ±5 ppm/year optional
Baud rate: 300, 1200, 9600, 19200, 38400, 57600,
and 115200; factory set to 9600
Data format: 8 data bits, 1 stop bit, no parity
Internal memory: 128 MB, 256 MB optional
Pressure sensor type: Precision micromachined silicon
diaphragm
Range: Customer specified, 50 psi (0–23 m)
range typical
Accuracy: ±0.03% of full scale
Stability: ±0.004% of full scale/month
Resolution: 0.002% of full scale
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 21/24
5-1
P/N A177-030
SECTION 5: Connector Wiring
he NEMA enclosure includes a Subconn MCBH8M circular bulkheadconnector for the instrument connector and a Subconn MCBH5F
connector for the serial connector. Figure 5-1 shows the pin orientation
for the instrument connector, and Figure 5-2 for the serial connector. Both
connectors are shown viewed from the face. The instrument connector mates with
a Subconn MCIL8F connector, and the serial cable with a MCIL5M.
To wire the NEMA enclosure to the EOPM and to your computer using your own
cables and mating connectors, refer to the instrument connector and serial
connector pinout information in Table 5-1 and Table 5-2. Also refer to the “EOPM
Operation Manual” for the bulkhead connector pinout information. A wiring
diagram of the components inside of the NEMA enclosure is shown in Figure 5-3.
Figure 5-1: NEMA Enclosure Instrument Connector—Face View Male
Figure 5-2: NEMA Enclosure Serial Connector—Face View Female
T
8
7
6
5
4
3
21
5
43
21
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 22/24
5-2 SECTION 5 Connector Wiring
EOPM Tide System User Manual March 2005
NEMA Enclosure Instrument Connector Pinouts
PININSTRUMENT
CABLE COLORLABEL FUNCTION
1 Black AGND Analog Ground
2 White TXD Transmitted Data (RS-232)
3 Red SUPPLY (+) Instrument Power
4 Green RXD Received Data (RS-232)
5 Orange RS-485A Transmitted Data (RS-485)
6 Blue RS-485B Received Data (RS-485)
7 White/Black NC No Connection
8 Red/Black VOUT Barometric Pressure Out
Table 5-1: NEMA Enclosure Instrument Connector Pinouts
NEMA Enclosure Serial Connector Pinouts
PINSERIAL CABLE
COLORLABEL FUNCTION
1 Black AGND Analog Ground
2 White TXD/RS-485A Transmitted Data (RS-232/RS-485)
3 Red NC No Connection
4 Green RXD/RS-485B Received Data (RS-232/RS-485)
5 Orange NC No Connection
Table 5-2: NEMA Enclosure Serial Connector Pinouts
7/31/2019 EOPM Tide System User Manual (Lo Res)
http://slidepdf.com/reader/full/eopm-tide-system-user-manual-lo-res 23/24
5-3
P/N A177-030
Figure 5-3: NEMA Enclosure Component Wiring
INSTRUMENT CONNECTOR SERIAL CONNECTOR