EPIC Calibration Meeting, EPIC Calibration Meeting, Mallorca Mallorca Experience with combined Fe Experience with combined Fe 55 55 /Tycho SNR /Tycho SNR calibration calibration K. Dennerl, 2008 K. Dennerl, 2008 April 8 April 8 Experience with combined Fe Experience with combined Fe 55 55 / Tycho SNR / Tycho SNR calibration calibration Mallorca, Spain, 2008 April 7 - 8 Mallorca, Spain, 2008 April 7 - 8 EPIC Calibration & Operations Meeting NRCO 71

Upload: quito

Post on 14-Jan-2016

31 views

Category:

Documents


1 download

DESCRIPTION

Experience with combined Fe 55 / Tycho SNR calibration. NRCO 71. EPIC Calibration & Operations Meeting. Mallorca, Spain, 2008 April 7 - 8. Internal calibration source onboard XMM-Newton / EPIC-pn. Al-K Mn-K. 30 ks in 2000  210 ks in 2008. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Experience with combined FeExperience with combined Fe5555 / Tycho SNR calibration / Tycho SNR calibration

Mallorca, Spain, 2008 April 7 - 8Mallorca, Spain, 2008 April 7 - 8

EPIC Calibration & Operations Meeting

NRCO 71

Page 2: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Internal calibration source onboard XMM-Newton / EPIC-pn

Al-K Mn-K

Page 3: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

30 ks in 2000

210 ks in 2008

10 ks

30 ks

Page 4: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Page 5: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

General plan:

1. start with ‘normal’ Tycho observation

2. add the internal calibration source

3. finish with ‘normal’ calclosed observation

Unfortunately, this sequence was scheduled as three independent observations, performed in a non-contiguous mode

superposition with short-term changes in the absolute energy scale

no firm conclusions about potential changes of the absolute energy scale possible

however: general idea about ‘calopen’ observations

Experience with combined FeExperience with combined Fe5555 / Tycho SNR calibration / Tycho SNR calibration

does the internal calibration source disturb the Tycho observation ?

does Tycho disturb the calclosed measurement ?

Page 6: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

1) Tycho FF Medium, 2007-12-30 / 19:55 – 2007-12-31 / 03:37, 27 ks

Experience with combined FeExperience with combined Fe5555 / Tycho SNR calibration / Tycho SNR calibration

Page 7: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

2) Tycho FF Cal Medium, 2008-01-01 / 19:58 – 2008-01-02 / 09:59, 50 ks

Experience with combined FeExperience with combined Fe5555 / Tycho SNR calibration / Tycho SNR calibration

Page 8: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

3) Calclosed, 2008-01-03 / 01:06 – 2008-01-03 / 08:11, 25 ks

Experience with combined FeExperience with combined Fe5555 / Tycho SNR calibration / Tycho SNR calibration

Page 9: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

1) Tycho FF Medium, 2007-12-30 / 19:55 – 2007-12-31 / 03:37, 27 ks

Page 10: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

2) Tycho FF Cal Medium, 2008-01-01 / 19:58 – 2008-01-02 / 09:59, 50 ks

Page 11: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

3) Calclosed, 2008-01-03 / 01:06 – 2008-01-03 / 08:11, 25 ks

Page 12: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

5.75 6.00 keV

adu

Narrow-line images at Mn-K: 5.70 – 6.05 keV

ΔE = 350 eV

Page 13: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Narrow-line images at Mn-K: 5.70 – 6.05 keV

Tycho CalClosed Tycho + CalClosed

scaled to the same exposure time ΔE = 0.35 keV

Page 14: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

ΔE = 0.35 keV ΔE = 1.0 keV

ΔE = 2.0 keV ΔE = 5.0 keV

Tycho + CalClosed

all energy bands centered on Mn-K

Page 15: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

1.25 1.50 1.75 keV

adu

Narrow-line images at Al-K: 1.375 – 1.575 keV

ΔE = 200 eV

Page 16: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Narrow-line images at Al-K: 1.375 – 1.575 keV

Tycho CalClosed Tycho + CalClosed

scaled to the same exposure time ΔE = 0.20 keV

Page 17: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

calclosed

Page 18: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Tycho

Page 19: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Tycho + calclosed

Page 20: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

calc

lose

d

Mn-K

Page 21: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Tyc

ho +

cal

clos

ed

Mn-K

Page 22: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

calc

lose

d

Al-K

Page 23: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Tyc

ho +

cal

clos

ed

Al-K

Page 24: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

25 macro pixels,

20-29 singles per macro pixel

CTI determination: Illustration of the Template Cross Correlation method (EPIC pn)

Page 25: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

calclosed

CTI

Page 26: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Tycho + calclosed

CTI

Page 27: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Tycho + calclosed

CTI

Page 28: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Experience with combined FeExperience with combined Fe5555 / Tycho SNR calibration / Tycho SNR calibration

Tycho + calclosed seems ok for Mn-K, but not for Al-K

Tycho better suited for LW than for FF

Calclosed & LW no useful combination

are there celestial calibration targets which are better suited for a combined Fe55 exposure ?

Page 29: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Vela SNR

Page 30: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Vela SNR (62 ks)

Page 31: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

calclosed (25 ks)

Page 32: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Vela SNR + calclosed (scaled to 62 ks)

Page 33: Experience with combined Fe 55  / Tycho SNR calibration

EPIC Calibration Meeting, MallorcaEPIC Calibration Meeting, Mallorca Experience with combined FeExperience with combined Fe5555/Tycho SNR /Tycho SNR calibrationcalibration

K. Dennerl, 2008 April 8K. Dennerl, 2008 April 8

Experience with combined FeExperience with combined Fe5555 / Tycho SNR calibration / Tycho SNR calibration

Vela + calclosed might be an interesting alternative:

• the parts of the Vela SNR covered by the FOV are similarly extended as the internal calibration source

• the X-ray surface brightness of the Vela SNR is similar to that of the internal calibration source

• the X-ray spectrum of the Vela SNR is almost complementary to that of the internal calibration source

But the Al-K region would need to be checked in detail..

Vela + calclosed might be an interesting alternative:

• the parts of the Vela SNR covered by the FOV are similarly extended as the internal calibration source

• the X-ray surface brightness of the Vela SNR is similar to that of the internal calibration source

• the X-ray spectrum of the Vela SNR is almost complementary to that of the internal calibration source

But the Al-K region would need to be checked in detail..