FLEXIBLE AC CURRENT PROBE FlexProbe ® 24-3001 24-3002 ENGLISH User Manual ® INSTRUMENTS CAT III 1000V 300A 3000A OFF ON OL FlexProbe ® 24-3002

Upload: hoangdan

Post on 16-Nov-2018

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

FLEXIBLE AC CURRENT PROBE

FlexProbe®

24-3001 24-3002

E N G L I S H User Manual

®INS TRUMEN TS

CAT III1000V

300A

3000A

OFF

ONOL

FlexProbe ®

24-3002

Page 2: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

Statement of Compliance

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.

We guarantee that at the time of shipping your instrument has met its published specifications.

An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.

The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.

Serial #: ________________________________

Catalog #: 2120.81 / 2120.82

Model #: 24-3001 / 24-3002

Please fill in the appropriate date as indicated:

Date Received: _________________________________

Date Calibration Due: _______________________

Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com

Page 3: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

Table of Contents

1. INTRODUCTION ............................................................................... 21.1 International Electrical Symbols ................................................21.2 DefinitionofMeasurementCategories .....................................31.3 ReceivingYourShipment ..........................................................31.4 OrderingInformation .................................................................3

1.4.1 AccessoriesandReplacementParts ............................3

. PRODUCT FEATURES ...................................................................... 42.1 Description ................................................................................42.2 GeneralFeatures ......................................................................52.3 FlexProbe®24-3001Features ..................................................62.4 FlexProbe®24-3002Features ..................................................7

3. SPECIFICATIONS............................................................................. 83.1 ElectricalSpecifications ............................................................83.2 EnvironmentalSpecifications ....................................................93.3 MechanicalSpecifications ........................................................93.4 SafetySpecifications ..............................................................10

4. OPERATION .................................................................................. 114.1 MakingMeasurementswiththeFlexProbe® ...........................114.2 TipsforMakingPreciseMeasurements..................................12

5. MAINTENANCE ............................................................................. 14Warning ...........................................................................................145.1 ChangingtheBattery ..............................................................145.2 Cleaning ..................................................................................14

RepairandCalibration ...........................................................................15TechnicalandSalesAssistance ............................................................15Limited Warranty ...................................................................................16WarrantyRepairs ...................................................................................16

Page 4: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

2 FlexProbe® - Flexible AC Current Probe

CHAPTER 1

INTRODUCTION

WARNING • Read the instruction manual completely and follow all the safety

informationbeforeattemptingtouseorservicethisinstrument.• Wearprotectiveclothingandglovesasrequired.• Use caution on any circuit: potentially high voltages and currents

maybepresentandmayposeashockhazard.• Read the safety specifications section prior to using the current

probe.Neverexceedthemaximumvoltageratingsgiven.• Safetyistheresponsibilityoftheoperator.TheFlexProbe®mustbe

used only by qualified personnel using applicable safety precau-tions.

• ALWAYS de-energize the circuit before wrapping the FlexProbe® aroundbareconductors,busbars,ornearliveparts.Donotwraponliveconductors.

• ALWAYSconnecttheelectronicmoduletothedisplaydevicebeforewrappingtheFlexProbe®aroundthesamplebeingtested.

• ALWAYSinspectthemodule,sensor,sensorcable,andoutputtermi-nalspriortouse.Replaceanydefectivepartsimmediately.Useonlyfactoryparts.

• NEVER use the FlexProbe® on electrical conductors rated above1000V Cat. III and 600V Cat. IV.

1.1 International Electrical SymbolsThis symbol signifies that the instrument is protected by double or reinforced insulation.This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and product damage may result.Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.This symbol refers to a type B current sensor. Do not apply around or remove from HAZARDOUS LIVE conductors without additional protective means (de-energizing the circuit or wearing protective clothing suitable for high voltage work).

In conformity with WEEE 2002/96/EC

Page 5: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

FlexProbe® - Flexible AC Current Probe 3

1.2 Definition of Measurement CategoriesCat. I: Formeasurementsoncircuitsnotdirectlyconnected to theAC

supplywall outlet such as protected secondaries, signal level,andlimitedenergycircuits.

Cat. II: Formeasurements performed on circuits directly connected totheelectricaldistributionsystem.Examplesaremeasurementsonhouseholdappliancesorportabletools.

Cat. III: For measurements performed in the building installation atthe distribution level such as on hardwired equipment in fixedinstallationandcircuitbreakers.

Cat. IV: For measurements performed at the primary electrical supply(<1000V) such as on primary overcurrent protection devices,ripplecontrolunits,ormeters.

1.3 Receiving Your ShipmentUponreceivingyourshipment,makesure that thecontentsareconsis-tentwith thepacking list.Notifyyourdistributorofanymissing items. Iftheequipmentappearstobedamaged,fileaclaimimmediatelywiththecarrierandnotifyyourdistributoratonce,givingadetaileddescriptionofanydamage.Savethedamagedpackingcontainertosubstantiateyourclaim.

1.4 Ordering InformationFlexProbe® 24-3001 (300/3000AAC,24",10mV/A/1mV/A,withlead) ....Cat. #2120.81

FlexProbe® 24-3002 (300/3000AAC,24",10mV/A/1mV/A) ................Cat. #2120.82Both models include a flexible probe with electronic module, 9V battery, user manual and a product warranty and registration card.

1.4.1 Accessories and Replacement PartsBanana(Female)/BNC(Male)Adapter............................. Cat. # 2118.46ForconnectionofFlexProbe®toBNCterminalsonscopesandotherdis-playinginstruments.

Order Accessories and Replacement Parts Directly OnlineCheck our Storefront at www.aemc.com/store for availability

Page 6: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

4 FlexProbe® - Flexible AC Current Probe

CHAPTER 2

PRODUCT FEATURES

2.1 DescriptionTheFlexProbe®isaflexibleACcurrenttransformercomposedofaflexiblesensorandanelectronicmodule.Theflexiblesensorpermitsmeasure-mentsonconductorswherestandardclamp-onprobescouldnotbeused.Inparticular, it canbe installed in tight spaces,aroundbreakerpanels,aroundcablebundles,aroundwideor largebusbars,orevenwrappedaroundirregularshapes.

TheShapeMemory featureenables theuser to “pre-shape” thesensorbeforeinsertingitbetweenoraroundconductors.Thisfeaturefacilitatesclosing,enhancesusersafety,andalleviatesthedroopingeffecttypicallyassociatedwithflexiblesensors.

TheFlexProbe®islightweight.Itdoesnotusemagneticcoreslikestandardtransformers.Thetransformationprincipleisbasedonanaircore.Itpres-entsvirtuallynoloadtothesystemundertest,hasalowphaseshiftandexcellentfrequencyresponse,andcannotbedamagedbyoverloads.

Thesensorassemblyiswaterproofandinsulatedfor1000VCat.III;600VCat.IV.TheFlexProbe®meetsEN61010,isCEmarked,andisdesignedwithmaterialsandcomponentstomeetUL3111,andinternationalagencyapprovals.

TheFlexProbe® hasamVoutput proportional to the currentmeasuredfordirectreadingsonDMMs,dataloggers,oscilloscopes,andpowerorharmonicmeters.TRMSmeasurementsaretakenwhenconnectedtoaTRMSmeter.TheFlexProbe® isinsensitivetoDCcurrentsandonlytheACcomponentofthemeasuredsignalismeasured.

Thelengthof theflexiblesensorcanbeselected in lengthsof24",36",and48"inlength.Consultthefactoryforcustomlengths,rangesand/orfeatures.

Page 7: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

FlexProbe® - Flexible AC Current Probe 5

2.2 General Features• Modelstomeasurefrom0.5to30000Arms

• Accuracy1%ofReading

• TRMSmeasurementswhenconnectedtoaTRMSinstrument

• Nocoresaturationordamageifoverloaded

• OverrangeLEDformeasurementcircuitry

• EN61010;1000VCat.III;600VCat.IV

• Waterproofsensor

• 9Vbatteryfortypical150hourcontinuousoperations

• ShapeMemory™forcustompre-shapingofsensorbeforeuse (nodrooping)

• Highfrequencyresponse

• Lowphaseshiftforpowermeasurements

• InsensitivetoDC,measuresonlyACcomponentonDC+ACsig-nals

• Excellent linearity

• Lightweight

Page 8: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

6 FlexProbe® - Flexible AC Current Probe

2.3 FlexProbe® 24-3001 Features

®

INS TRUMEN TS

CAT III1000V

300A

3000A

OFF

ONOL

FlexProbe ®

24-3002

1

2

3

6

4

5

8

7

1. Flexiblesensor(Ø0.5",12.5mm)

2. Rangeselectionswitch

3. PowerONindicator(GreenLED)

4. Overload indicator (Red LED)

5. Electronicmodule

6. Doubleinsulated14"(36cm)leadwithsafetybananaplugs

7. FlexProbe®connector/latch

8. Sensor lead

Page 9: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

FlexProbe® - Flexible AC Current Probe 7

2.4 FlexProbe® 24-3002 Features

®

INS TRUMEN TS

CAT III1000V

300A

3000A

OFF

ONOL

FlexProbe ®

24-3002

1

2

3

4

5

6

7

9

8

1. Flexiblesensor(Ø0.5",12.5mm)

2. Rangeselectionswitch

3. PowerONindicator(GreenLED)

4. Common:Blackbananajack(-)

5. Positive:Redbananajack(+)

6. Overload indicator (Red LED)

7. Electronicmodule

8. FlexProbe®connector/latch

9. Sensor lead

Page 10: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

8 FlexProbe® - Flexible AC Current Probe

CHAPTER 3

SPECIFICATIONS

Reference Conditions: 41°F ± 5°F (23°C ± 3°C), 20 to 70% RH, 50Hz, 1 minute warm-up, battery at 9V ± 0.1V, external magnetic field < 40A/m, no DC component, no external current carrying conduc-tor, 10MΩ load, conductor center.

3.1 ElectricalCurrent Range: 300/3000AAC

Output Signal:mVoutput(4000mVpeakmax)

Three position slide switch: 10mV/A 5A to 300Arms 1mV/A 50A to 3000Arms ON/OFF

AC Current Accuracy (after calibration and for one year):1%ofReading±500mA

Crest Factor: 4.5

Phase Shift:<0.7°at50or60Hz

Frequency Range:40to5kHz

Residual Noise When I Measured = 0: 200mArms

Working Voltage: 1000Vrms max (Cat. III)

Common Mode Voltage: 600Vrms max

Influence of Frequency: Add1%tothereferenceaccuracybetween40and2000Hz,andupto5%at5kHz

Influence of Adjacent Conductor: <2%

Influence of Conductor Position in Jaw: <4%

Power Source:9Valkaline(NEDA1604A,IEC6LR61)

Battery Life: 150hourstypical(continuoususe) 300hourswithLithiumbattery

Page 11: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

FlexProbe® - Flexible AC Current Probe 9

Low Battery:GreenLEDOffwhen≥7V

Overload Indication: RedLEDindicatesinputgreaterthanselectedrange

3.2 EnvironmentalOperating Temperature Range: 14°to131°F(-10°to+55°C)

Storage Temperature Range: -40°to158°F(40°to+70°C)

Influence of Temperature:Sensor:-14°to212°F(-10°to100°C);<0.5%per10°CModule:-14°to131°F(-10°to55°C);<0.5%per10°C

Operating Relative Humidity:50°to86°F(10°to30°C);85±5%RH(withoutcondensation)104°to122°F(40°to50°C);45±5%RH(withoutcondensation)

Altitude: Operating:0to2000m Non-operating:0to12000m

3.3 Mechanical SpecificationsDimensions (Sensor): 24"±1"(610x25mm)

Color (Sensor): Redwithblackconnector

Dimensions (Electronic Module):4.9x2.5x1.1"(124x64x28mm)

Color (Electronic Module): Darkgray

Weight: 1.1lb(0.5kg)

Output:

Model 24-3001:Doubleinsulated14"(36cm)leadwithsafetybananaplugs

Model 24-3002:4mmsafetybananajacks(redandblack)

Maximum Cable Diameter: 7.6"(193mm)

Cable Length from Sensor to Module:6.5ft(2m)

Drop Test:1.0mon38mmofOakonconcrete.TestaccordingtoEN61010.

Page 12: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

10 FlexProbe® - Flexible AC Current Probe

Case Protection: Sensor:IP65 ElectronicModule:IP40

Vibration: Test Per IEC 68-2-6

Frequency Range:10to55Hz

Amplitude: 0.15mm

Mechanical Shock: 100G,TestPerIEC68-2-32

Material: Module:UL94V2 CableAssembly:UL94V0

3.4 Safety SpecificationsElectrical SafetyDoubleinsulationorreinforcedinsulationbetweentheprimaryorthesec-ondaryandtheoutercaseofthehandleperEN61010.

EN61010,1000VCat.III,600VCat.IV

PollutionDegree:2

Electromagnetic Compatibility:

EN 50081-1 Class BEN50082-2 ElectrostaticdischargeIEC1000-4-2 Radiated Field IEC 1000-4-3 Fast transients IEC 1000-4-4MagneticFieldat50/60HzIEC1000-4-8

*Specifications are subject to change at any time without notice

Page 13: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

FlexProbe® - Flexible AC Current Probe 11

CHAPTER 4

OPERATION

Pleasemakesure thatyouhavealready readand fullyunderstand theWARNINGsectiononpage2.

4.1 Making Measurements with the FlexProbe®

1. ConnecttheelectronicmoduletotheACVoltrangeofyourDMMormeasuringinstrument.

2. Select the appropriate module output voltage range. If the currentmagnitudeisunknown,selectthe3000Arange.

3. Wraptheflexiblecorearoundtheconductortobetested.Selectthe300Arange toobtain thebest resolution, if the reading is less than300mV.

Do not exceed specified current range for the output. Do not use on selected range if Overload LED goes on.

4. Example:

a. ReadtheDMMdisplayedvalue:2.59V

b. DeterminetherangeandoutputusedontheFlexProbe:0.01V/Aforthe300Arange;.001V/Aforthe3000Arange

c. Dividethereadingbytheoutputtodeterminetheactualcurrentvalue:2.59÷0.01=259Aor2.59÷0.001=2590A

Forbestaccuracy,carefullycentertheconductorinsidetheflexiblecore,andavoidifpossible,theproximityofotherconductorswhichmaycreatenoiseandinterference(particularlynearthelatch).

TrueRMSmeasurementsareobtainedwhentheFlexProbe® is connected toaTrueRMSmeter.NotethattheDCcomponentisnotmeasured.

Page 14: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

12 FlexProbe® - Flexible AC Current Probe

4.2 Tips for Making Precise Measurements

• When using the FlexProbe® with a meter, it isimportant to select the range that provides thebest resolution. Failure to do this may result inmeasurementerrors.

• For best results, select the highest FlexProbe® output signal possible and the most sensitivemeterrangeforthisoutput.

• MakesuretheDMMormeasuringinstrumentcanaccurately measure mVAC. Certain inexpensiveDMM have poor resolution and accuracy whenmeasuringlowmVAC.

• Forbestaccuracy,centertheFlexProbe®aroundtheconductortobemeasured.

• Toincreasethesensitivityormeasureonlowcur-rents, the FlexProbe® may be wrapped severaltimesaroundtheconductor.Remembertodivideyourreadingbythenumberofturnsfortheactualmeasurement(SeeFigure1).

• TheoverallmeasurementaccuracyisthesumoftheFlexProbe® accuracyandthedisplayinginstru-mentaccuracy.

• TheFlexProbe®maybedoubledaroundthecon-ductortobemeasuredtodoubletheoutput.(SeeFigures1and2toshowthedifferentvaluesonaDMMwhilemeasuring250AAC.)

Page 15: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

FlexProbe® - Flexible AC Current Probe 13

ACV DCV

COM +

500.0 MV

One turn around aconductor carrying 250AAC

Figure 1

Figure 2

3000ARange

3000ARange

250A

®INS TRUMEN TS

CAT III1000V

300A

3000A

OFF

ONOL

FlexProbe ®

24-3002

ACV DCV

COM +

250.0 MV

®INS TRUMEN TS

CAT III1000V

300A

3000A

OFF

ONOL

FlexProbe ®

24-3002

Double the turns to doublethe output in low current applicationsor for higher sensitivity

250A

Page 16: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

14 FlexProbe® - Flexible AC Current Probe

CHAPTER 5

MAINTENANCE

Warning• Formaintenanceuseonlyspecifiedreplacementparts.

• Toavoidelectricalshock,donotattempttoperformanyservicingunlessyouarequalifiedtodoso.

• Do not perform any service while the FlexProbe® is on any circuit.

• Toavoidelectricalshockand/ordamagetotheinstrument,donotgetwaterorotherforeignagentsintotheelectronicmodule.

• Alsoseewarningonpage2.

5.1 Changing the BatteryIfthepowerONindicator(greenLED)blinksordoesnotlightup,replacethebattery.

Disconnect the instrument from any source of electricity.

• To replace thebattery, open rear case, replacebattery and re-assemble.

• ThegreenLEDshouldgoonwhenthemoduleisturnedon.

5.2 Cleaning• Useasoftclothlightlydampenedwithsoapywater,thendrywith

adrycloth.

• Donotsplashwaterdirectlyontheinstrument.

• Donotusealcohol,solventsorhydrocarbons.

• Makesure thesensor,electronicmodule,andall leadsaredrybeforeanyfurtheruse.

Page 17: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

FlexProbe® - Flexible AC Current Probe 15

Repair and Calibration

Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.

For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.(Includescalibrationcertificateplusrecordedcalibrationdata).

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15 Faraday Drive Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603) 749-6434 (Ext. 360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

(Orcontactyourauthorizeddistributor)Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.areavailable.NOTE: You must obtain a CSA# before returning any instrument.

Technical and Sales Assistance

Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:

ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone:(800)343-1391 (508) 698-2115 Fax: (508)698-2118 E-mail: [email protected] www.aemc.com

NOTE: Do not ship Instruments to our Foxborough, MA address.

Page 18: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

16 FlexProbe® - Flexible AC Current Probe

Limited Warranty

TheFlexProbe®iswarrantedtotheownerforaperiodofoneyearfromthedateoforiginalpurchaseagainstdefectsinmanufacture.ThislimitedwarrantyisgivenbyAEMC®Instruments,notbythedistributorfromwhomitwaspur-chased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.

For full and detailed warranty coverage, please read the Warranty Coverage Information, which is attached to the Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the Warranty Coverage Information with your records.

What AEMC® Instruments will do:Ifamalfunctionoccurswithintheone-yearperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplacethefaultymaterial.

REGISTER ONLINE AT:www.aemc.com

Warranty Repairs

What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15FaradayDrive•Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603) 749-6434 (Ext. 360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyourreturnedmaterial.

NOTE: You must obtain a CSA# before returning any instrument.

Page 19: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating
Page 20: FlexProbe - Cole-Parmer · III and 600V Cat. IV. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or reinforced ... Operating

07/08

99-MAN100272v6

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346

www.aemc.com