Page 1: Fresenius 2008H Dialysate Delivery Machine...Fresenius 2008H Dialysate Delivery Machine Hemant Surgical Industries Ltd. SPECIFICATIONS - Fresenius 2008H Dimensions Floor space Approx

Fresenius 2008H Dialysate Delivery Machine

Hemant Surgical Industries Ltd.

Page 2: Fresenius 2008H Dialysate Delivery Machine...Fresenius 2008H Dialysate Delivery Machine Hemant Surgical Industries Ltd. SPECIFICATIONS - Fresenius 2008H Dimensions Floor space Approx

SPECIFICATIONS - Fresenius 2008H DimensionsFloor space Approx. 50 cm x 50 cmHeight 133 cmTotal weight Approx. 70 kg

Power Supply- MainSingle phase AC 220 V ±10% 50 Hz

Water SupplyWater Pressure Min. 20 psi; Max. 105 psi

0 0Water Temperature Min. 10 C; Max. 25 CQuality AAMI Standard, RD5, 1992

BicarbonateNormal Setting Selectable 1:27.6 / 1:19.1 / 1:31.1 – volumetricAdjustment Range 25 to 40 meq/L Bicarbonate

Monitoring Conductivity Temperature compensated conductivity display with automatically set alarm

windows ±0.5 mS/cm around calculated conductivity.

Range of Display 10.0 to 17.00 mS/cm.

Dialysate Heating 0 0 0 34 to 39 C, ( selectable in 0.1 C steps )

Heat Disinfection 0 0 83 ± 5 C

Chemical Disinfection0Temperature 37 C (set point applicable)

Time Between 10 and 60 minutes (internally selectable)

Rinsing 0 Temperature 37 C. Flow rate 620 ml/min. Time between 10 and 60 minutes

(internally selectable)

Blood Pump 0 to 600 ml/min.

Heparin PumpAdministration Rate 0.5 to 10 ml/hrMonitoring Internal monitoring with time alarm

Arterial Pressure Monitor Pre-pump monitor range (-300 to 280 mmHg) or post-pump monitor range (-80 to 500 mmHg)

Venous Pressure Monitor -80 to +500 mmHg with automatically

TMP Monitor +60 to -520 mmHg with automatically

Level Detector Ultrasonic impulse sensor

Blood Leak Detector Two color light source transmitter / sensor

Ultrafiltration Control Volumetric Control, 0-4000 ml/hr.UF Time Digital Display (0 - 9 hrs. 59 min.) UF Goal Digital display (0 to 9990 ml). Remaining Time of Dialysis (RTD) 0-9:59 hours auto transfer from UF time,UF Removed Display Digital display max 9999 mlAdditional Monitoring Alarm in case of power failure. Alarm in case of water shortage

Optional Internal Blood Pressure Monitor (IBP)

Imported & Marketed By :

Hemant Surgical Industries Ltd.647, Girgaum Road, Avijah Building, Shop No. 6 & 7, Near Metro Cinema,

Mumbai-400 002. • Tel. : 2203 0935, 2203 0936 • Fax : 91-22-2207 7585