RESULTS | AS012
Fully Automated Mycotoxin Analysis
in the ppq Range FREESTYLE ThermELUTE™
2
RESULTS | AS012
The robotic system FREESTYLE ThermELUTE™ processes your samples fully automated from the raw extract to the ready chromatogram with highest sensitivity, e.g. for Aflatoxin M1 even in the range of ppq. The system analyses up to 500 samples per week in combination with any HPLC system using the unique technology of thermal denaturation. After washing, the immunoaffinity column is heated, which causes the bond between toxin and antibody to break up. The toxins, in form of a large-volume, aqueous eluate, are directly eluted into the HPLC sample loop via partial filling and analysed by the HPLC. In the FREESTYLE ThermELUTE™ LCTech’s SMART columns are used. These immunoaffinity columns have a size of only 3 cm but their performance is comparable to the standard 1 mL and 3 mL immunoaffinity columns. The SMART columns require significant less sample through miniaturisation of the overall process, less solvent and less time and thus save costs. In the following we show some chromatograms processed using the protocol described below:
Extraction and dilution are corresponding to the protocol of the SMART columns (incl. dilution of the sample with PBS-puffer before loading onto the column)
Processing via FREESTYLE ThermELUTE™
Loading volume: 20 mL (milk / milk powder; extracted) with a flow rate of 1 mL/min. For the example shown below 10 mL milk were loaded onto the column.
Washing volume: 4 mL with a flow rate of 1 mL / min. Heating time: 8 min with 98 °C
The following chromatograms were obtained with a Dionex 3400 RS detector. Please note that the achieved results always depend on the HPLC and detector used, respectively.
Unique Automation FREESTYLE ThermELUTE™
3
RESULTS | AS012
0.0 2.0 4.0 6.0 8.0 10.0 12.0-10,000
20,000
40,000
60,000
80,000
100,000 counts
min
M1
Chromatograms Aflatoxin M1 cleaned-up with AflaCLEAN M1 SMART
AflaCLEAN M1 SMART loaded with AflaCLEAN M1 SMART loaded with 100 pg/20 mL = represents 10 ppt 10 pg/20 mL = represents 1 ppt
AflaCLEAN M1 SMART loaded with AflaCLEAN M1 SMART loaded with 8.8 pg/20 mL = represents 0.9 ppt 7.5 pg/20 mL = represents 0.8 ppt
0.0 2.0 4.0 6.0 8.0 10.0 12.0-10,000
20,000
40,000
60,000
80,000
100,000counts
min
M1
0.0 2.0 4.0 6.0 8.0 10.0 12.0 -20,000
0
20,000
40,000
60,000
80,000
100,000 counts
min
M1
0.0 2.0 4.0 6.0 8.0 10.0 12.0 -70,000
500,000
1,000,000
1,500,000
2,000,000counts
min
M1
4
RESULTS | AS012
AflaCLEAN M1 SMART loaded with 5 pg/20 mL = represents 0.5 ppt = 500 ppq Aflatoxins B/G cleaned-up with AflaCLEAN SMART
AflaCLEAN SMART loaded with AflaCLEAN SMART loaded with 0.05 ng/10 mL total toxin = 0.005ng/10mL total toxin = 17.9 ppt G2/B2 (referred to matrix); 1.8 ppt G2/B2 (referred to matrix); 71.4 ppt B1/G1 7.1 ppt B1/G1
0.0 2.0 4.0 6.0 8.0 10.0 12.0-10,000
20,000
40,000
60,000
80,000
100,000 counts
min
M1
0.0 2.0 4.0 6.0 8.0 10.0 12.0-10,000
50,000
100,000
150,000
200,000
250,000
300,000 counts
min
G2
G1
B2
B1
0.0 1.3 2.5 3.8 5.0 6.3 7.5 8.8 10.0 12.0-7,000
0
5,000
10,000
15,000
23,000 counts
min
G1
B2
B1
G2
5
RESULTS | AS012
-10,000
40,000 counts M1
-10,000
40,000 counts M1
0.0 1.3 2.5 3.8 5.0 6.3 7.5 8.8 10.0 12.0 -10,000
40,000 counts
min
M1
AflaCLEAN SMART loaded with 0.0025 ng/10 mL total toxin = 0.9 ppt G2/B2 (referred to matrix), 3.6 ppt B1/G1
Reproducibility To show the reproducibility of the results we consecutively processed three independent samples.
Aflatoxin M1 with AflaCLEAN M1 SMART Aflatoxin B/G with AflaCLEAN SMART
Comparison of three chromatograms from three different samples. These comparisons show the excellent reproducibility even in this very low measurement range.
0.0 1.3 2.5 3.8 5.0 6.3 7.5 8.8 10.0 12.0-5,000
0
10,000
20,000
30,000
35,000 counts
min
G2
G1
B2
B1
-10,000
40,000 counts G1
B2 B1
-10,000
40,000 counts G1
B2 B1
0.0 1.3 2.5 3.8 5.0 6.3 7.5 8.8 10.0 12.0-10,000
40,000 counts
min
G1
B2 B1
G2
G2
G2
6
RESULTS | AS012
Sensitivity In order to prove the sensitivity we cleaned-up a blind sample and a spiked sample and compared these two samples. Aflatoxin M1 with AflaCLEAN M1 SMART
Comparison of non spiked (blind sample), black and spiked sample with 500 ppq (red). very high sensitivity of M1.
-10,000
20,000
40,000 counts
0.0 1.3 2.5 3.8 5.0 6.3 7.5 8.8 10.0 12.0
-10,000
20,000
40,000 counts
min
M1
7
RESULTS | AS012