Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
![Page 2: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/2.jpg)
2
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
There are three International Standards of measurement for Perimetry:
ASBASB
dBdBFT/LFT/L
![Page 3: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/3.jpg)
3
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
ASBASB
AApopoSStiltilBB
0 (Dimmest) – 10,000 (Brightest)
A measurement of light, dealing with the brightness of the
surface of the bowl
International StandardInternational Standard
![Page 4: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/4.jpg)
4
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
FT/LFT/L
Foot LambertFoot Lambert
HFA Range
0 (dimmest) – 929 (Brightest)
FT/L
Higher number = Brighter light
International StandardInternational Standard
![Page 5: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/5.jpg)
5
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
dBdB
DecibelDecibel
HFA Range
0.1 dB (Brightest) – 51dB (Dimmest)
Higher number = Dimmer Spot
International StandardInternational Standard
![Page 6: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/6.jpg)
6
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
International Standards Comparison
Dimmest Brightest
0 ASB 10,000 ASB
0 FT/L 929 FT/L
51 dB 0.1 dB
![Page 7: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/7.jpg)
7
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
All Perimeters use the same standards
These standards are based on the
Goldmann Perimeter
ASBASB
ASB
ASB ASBASB
AS
BA
SB
ASBASB
FT
/LF
T/L
FT/LFT/L
FT/LFT/L
FT/LFT/L
FT/LFT/L
FT/L
FT/L
dBdB
dBdB
dBdB
dBdB
dBdB
dBdB
![Page 8: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/8.jpg)
8
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
The Goldmann Perimeter, which defined the Standards of Perimetry. These standards are still in use today.
![Page 9: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/9.jpg)
9
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
The standards of Perimetry are:
Spot IntensitySpot Intensity
Background IlluminationBackground Illumination
Spot DurationSpot Duration
Spot SizeSpot Size
Spot SpeedSpot Speed
![Page 10: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/10.jpg)
10
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Background Illumination
The standard value is 31.5 ASB.The standard value is 31.5 ASB.
The illumination in the bowl must remain The illumination in the bowl must remain constant throughout the test.constant throughout the test.
![Page 11: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/11.jpg)
11
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Why 31.5 ASB?
Originally used by Goldmann Perimeter
Adopted as standard by International Peremetric Society (International Council of Ophthalmology 1979)
Approximates minimum level for photopic or daylight vision
Photopic vision relies on retinal cone function instead of rods
Cones – Object Contrast Rods – Absolute Brightness
Small changes in pupil size or clarity of media do not have an effect on Contrast, so have little effect on test results.
![Page 12: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/12.jpg)
12
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Spot IntensitySpot Intensity
Glass Wedge/Film WedgeGlass Wedge/Film Wedge
The Spot intensity is controlled by filter wheelsThe Spot intensity is controlled by filter wheels
The spot intensity is directly related to the The spot intensity is directly related to the bowl intensity.bowl intensity.
![Page 13: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/13.jpg)
13
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Spot Size
Size V spot is the largestSize V spot is the largest
The size of the spot can be size I,II,III,IV,or VThe size of the spot can be size I,II,III,IV,or V
The default spot size is size IIIThe default spot size is size III
Mr. Default Mr. Default to you!to you!
![Page 14: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/14.jpg)
14
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Spot size 3 is .43 degrees
Spot Size is smaller in the HFA II bowl than in the HFA I, because the bowl
radius is smaller.
![Page 15: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/15.jpg)
15
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Spot Duration
The default time duration is 200mS The default time duration is 200mS (milliseconds) + or – 10mS(milliseconds) + or – 10mS
How long the spot is displayedHow long the spot is displayed
Duration can be changed to 500 mS for older Duration can be changed to 500 mS for older patients.patients.
![Page 16: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/16.jpg)
16
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
The principle of temporal summation holds that for very short durations, the visibility of a stimulus increases with duration; when a stimulus lasts more than about 0.5 seconds, on the other hand, its visibility is basically independent of duration.
200 ms – long enough for visibility to not be affected by small variations in duration. But less than …
250 ms – latency for voluntary eye movement
![Page 17: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/17.jpg)
17
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Spot Speed
This only applies to This only applies to KineticKinetic testing. testing.
How fast the spot movesHow fast the spot moves
Default speed for the HFA is 4 degrees per second.Default speed for the HFA is 4 degrees per second.
![Page 18: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/18.jpg)
18
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Goldmann vs. Humphrey Parameters
Goldmann HumphreyGoldmann Humphrey
Spot Size I,II,III,IV, V I,II,III,IV,VSpot Size I,II,III,IV, V I,II,III,IV,V
(V is the largest) (V is the largest)(V is the largest) (V is the largest)
Filters 1,2,3,4 Glass WedgeFilters 1,2,3,4 Glass Wedge
(4 is the brightest) Film Wedge(4 is the brightest) Film Wedge
a,b,c,d,ea,b,c,d,e
(e is the brightest)(e is the brightest)
![Page 19: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/19.jpg)
19
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Kinetic TestingSingle intensity; Moving Target
A light spot (Stimulus) is introduced along a particular meridian, following a straight line until a patient response (sees the light spot) is indicated.
MeridianMeridian
Patient ResponsePatient Response
![Page 20: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/20.jpg)
20
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Static Testing
Varies the intensity of the spot over the entire Hill of Vision. Patient does not see the spot move. The spots are projected at different positions on the bowl, but the instrument will return to the spots at different intensities.
Variable Intensity; Stationary Target
![Page 21: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/21.jpg)
21
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
The “Island” of Vision
A A StaticStatic test on a test on a normal eye will normal eye will produce a pattern produce a pattern similar to the one at similar to the one at right.right.
DimDim
BrightBright
Blind SpotBlind Spot
FoveaFovea
NasalNasal TemporalTemporal
![Page 22: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/22.jpg)
22
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Scotomas
There are two types ofThere are two types of ScotomasScotomas::
Definition: A defect on the Retina. An area or spot on Definition: A defect on the Retina. An area or spot on the Retina that is not as sensitive to light as it should the Retina that is not as sensitive to light as it should be.be.
1.1. Relative Relative – An area or spot on the Retina that – An area or spot on the Retina that can detect light, detect light, but not as good as when compared to a normal eye at that same but not as good as when compared to a normal eye at that same spot.spot.
2. Absolute – An area or spot on the Retina that cannot detect light, no matter how bright it is.
![Page 23: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/23.jpg)
23
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Scotoma
Is this Scotoma Relative Is this Scotoma Relative or Absolute?or Absolute?
Relative!!Relative!!
BlindspotBlindspot
![Page 24: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/24.jpg)
24
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Hey man, like, ALL my Hey man, like, ALL my relatives are Scotomas!relatives are Scotomas!
Absolutely!!!Absolutely!!!
Can you dig it??!!Can you dig it??!!
Cool!!Cool!!
![Page 25: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/25.jpg)
25
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
The Blind spot exhibits the same characteristics as an ________ ________?
Absolute Absolute Scotoma!Scotoma!
![Page 26: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/26.jpg)
26
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Thresholding and Bracketing
What is thresholding?
Threshold (or Threshold Level) is the minimum amount of light that the eye can detect at a particular point on
the retina.
Bracketing is the process of determining the Threshold value.
![Page 27: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/27.jpg)
27
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Thresholding by Bracketing
For our purposes, we will assume that the instrument begins the threshold test with a spot of 36db. The spot or stimulus is presented at a particular spot on the bowl to see if there is a patient response.
0 dB36 dB55 dB
Indicates a negative (did not see) responseIndicates a negative (did not see) response
Indicates a positive (did see) responseIndicates a positive (did see) response
DimmerDimmer BrighterBrighter
![Page 28: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/28.jpg)
28
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Thresholding by Bracketing
The patient did not respond, indicating that he/she did NOT see the spot. The instrument will now introduce a spot of 32dB (a 4dB change) at the exact same location.
0 dB0 dB36 dB36 dB55 dB55 dB
Indicates a negative (did not see) responseIndicates a negative (did not see) response
Indicates a positive (did see) responseIndicates a positive (did see) response
32 dB32 dB
DimmerDimmer BrighterBrighter
![Page 29: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/29.jpg)
29
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Thresholding by Bracketing
The patient responded positively, indicating that he/she DID see the spot. The instrument will now introduce a spot at the same exact position, at a brightness of 34dB. (A 2dB change)
00 dB dB36 dB
55 dB
Indicates a negative (did not see) response
Indicates a positive (did see) response
32 dB
DimmerDimmer BrighterBrighter
34 dB
![Page 30: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/30.jpg)
30
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Once again, the patient did not see the spot. The instrument has determined by BRACKETING that the threshold for this spot is 32dB. (NOTE: Some tests may closer define the threshold by testing at 1 dB steps.)
0 dB36 dB
55 dB
Indicates a negative (did not see) response
Indicates a positive (did see) response
32 dB
DimmerDimmer BrighterBrighter
34 dB34 dB
Thresholding by Bracketing
![Page 31: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/31.jpg)
31
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
What happens when you start a test?
1. One spot in each 1. One spot in each quadrant will be quadrant will be bracketed to determine bracketed to determine the Threshold level.the Threshold level.
At the same time, the At the same time, the location of the Blind location of the Blind Spot will be determined.Spot will be determined.
Bracketing begins at 24dBBracketing begins at 24dB
Blindspot
![Page 32: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/32.jpg)
32
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
What happens when you start a test?
2. Once the threshold value for a spot in each quadrant has been found, an Expected Hill of Vision is determined.
DimmerDimmer
Expected “Hill of Vision”
36dB (CEN)36dB (CEN)
![Page 33: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/33.jpg)
33
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
The “CEN” is defined asThe “CEN” is defined as the Central ReferenceCentral Reference LevelLevel, or the, or the Expected Expected Foveal SensitivityFoveal Sensitivity, or the, or the Expected Foveal Expected Foveal Threshold.Threshold.
Expected “Hill of Vision”
DimmerDimmer
36dB (CEN))
What happens when you start a test?
![Page 34: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/34.jpg)
34
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
What happens when you start a test?
3. As the test is run, the patients “Actual” readings are plotted.
Expected “Hill of Vision”Expected “Hill of Vision”
DimmerDimmer
36dB36dB
Actual Patient plotsActual Patient plots
Blind Spot
![Page 35: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/35.jpg)
35
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Types of Tests
1. Screening Tests – A QUICK overview of the patient’s field of vision. Most screening tests tell only if the patient did or did not see the spot. In general, screening tests do not quantify. That is, they do not determine how bad a scotoma is.
Now you see it……..Now you see it…….. Now you don’t!!!!!!!!!Now you don’t!!!!!!!!!
![Page 36: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/36.jpg)
36
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Types of Tests
2. 2. Threshold TestsThreshold Tests – – QuantifiesQuantifies each spot. That is, it finds the threshold each spot. That is, it finds the threshold level. Determines how bad the scotoma is by calculating the level. Determines how bad the scotoma is by calculating the exactexact light light level the patient can see at a particular spot on the Retina.level the patient can see at a particular spot on the Retina.
I don’t see it…
I don’t see it…
Now I see it!!!!!
Now I see it!!!!!
![Page 37: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/37.jpg)
37
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Short Wavelength Automated Perimetry (SWAP)
Blue Goldmann Size 5 Stimulus on bright Yellow background
Yellow Background reduces responsiveness of the red and green cone system
Blue Stimuli are seen primarily by the Blue Cone System
SWAP can detect progression of field loss earlier in patients than standard White on White perimetry
SWAP eliminates the redundancy in the visual system by reducing responsiveness from non blue cones and the rod system
![Page 38: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/38.jpg)
38
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Reliability Factors
After a Field Test has been completed, a After a Field Test has been completed, a determination must be made if the test is accurate determination must be made if the test is accurate
and therefore reliable. Several Reliability and therefore reliable. Several Reliability Indicators appear on the test printout:Indicators appear on the test printout:
Fixation LossesFixation Losses False Negative ErrorsFalse Negative Errors
False Positive ErrorsFalse Positive Errors
Gaze TrackingGaze Tracking
![Page 39: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/39.jpg)
39
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Fixation Losses
During the test, the patient is told to Fixate (Stare) at a central LED. The instrument records the number of times that the patient lost this fixation and reports it on the printout.
Excessive fixation losses can render a test invalid.
![Page 40: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/40.jpg)
40
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Fixation Monitoring
The HFA uses two systems for measuring patient fixation: the standard Heijl-Krakau blind-spot monitoring and the IR Gaze Tracking System. Both methods can be used, either together or alone, or they can both be turned off, as required.
![Page 41: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/41.jpg)
41
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Heijl-Krakau Blind Spot Monitoring
![Page 42: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/42.jpg)
42
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze TrackingLocation of Corneal Reflex Marker
Corneal Reflex marker Location Digitized and Stored in Memory
![Page 43: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/43.jpg)
43
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
The direction of a patient's gaze is determined in two steps: 1. A reflex marker is establishedon the corneal surface.
2. The location of the pupil center is determined.
Gaze Tracking
![Page 44: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/44.jpg)
44
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
Gaze tracking is initialized in the following manner when a selected test is first started:
The patient is asked to fixate on the central illumination LED. Gaze tracking turns on the reflex gaze IR LED located just under the diamond fixation pattern and turns off eye illumination briefly . Light from the LED is reflected off the cornea, and back to the IR sensitive camera.
![Page 45: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/45.jpg)
45
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
The majority of the cornea appears black except for the reflected spot. This image is digitized and stored in memory. The reflected spot is referred to as the reflex marker. Because the corneal surface is rounded, the reflex marker will move very little even if the patient's eye rotates, and thus the marker becomes a (relatively) stationary reference point.
![Page 46: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/46.jpg)
46
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
Next, the system locates the pupil center by illuminating the entire eye with the two IR LEDs located either in the bottom of the bowl, or in the trial lens holder (when in the raised position). The iris appears bright with a dark pupil. This image is also digitized and stored in memory. It is the relationship between the location of the reflex marker on the cornea and the location of the pupil center that determines fixation.
![Page 47: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/47.jpg)
47
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
During a test, each time a spot is projected into the bowl, the locations of the reflex marker and the center of the pupil are compared to the initial images stored in memory. If the patient is fixating correctly, the positional relationship between the reflex marker and the pupil center will be the same as that of the stored images. If the patient is off fixation, the positional relationship between the reflex marker and the pupil center will be different. The greater the misalignment, the higher the mark on the Gaze Graph.
![Page 48: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/48.jpg)
48
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
Determining the Pupil Center
Patient Fixating – CornealReflex Marker and Pupil in
ProperRelationship
Patient Not Fixating –Corneal Reflex Marker and Pupil
Not in Proper Relationship
![Page 49: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/49.jpg)
49
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
Gaze Graph
Upward spikes indicate that the patient has lost fixation;• a spike that reaches the top horizontal line (or higher) indicates 10 degrees (ormore) off fixation;• a spike that extends halfway to the top line indicates 5 degrees off fixation.P Downward spikes indicate as follows:• a short spike downward indicates that the gaze at that time cannot bedetermined by the software.• a long spike downward indicates that the patient blinked at the time fixationwas checked.The absence of marks on the graph indicates proper fixation.
![Page 50: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/50.jpg)
50
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
![Page 51: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/51.jpg)
51
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
![Page 52: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/52.jpg)
52
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
Gaze Tracking
![Page 53: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/53.jpg)
53
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
False Negative Errors
Let us say that a Let us say that a StimulusStimulus (Spot) is presented at (Spot) is presented at 26dB26dB, , and the patient responds that the spot was seen. and the patient responds that the spot was seen. Later in the same test, at the same location, a brighter Later in the same test, at the same location, a brighter spot, say spot, say 22dB22dB, is presented, and the patient does not , is presented, and the patient does not respond, that is, the patient does not see the spot.respond, that is, the patient does not see the spot.
““Yessir, I see it!”Yessir, I see it!”
““Nosir, I didn’t see that one!”Nosir, I didn’t see that one!”
![Page 54: Fundamentals of hfa perimetry](https://reader035.vdocument.in/reader035/viewer/2022081422/558c0aded8b42a52568b4710/html5/thumbnails/54.jpg)
54
Humphrey Field Analyzer II/IIiFundamentals of Perimetry
Measurement
False Positive Errors
The patient responds that a spot of light was seen The patient responds that a spot of light was seen when none was presented. This sometimes can when none was presented. This sometimes can occur when the motors move on the Humphrey occur when the motors move on the Humphrey Field Analyzer, or a patient gets into a rhythm and Field Analyzer, or a patient gets into a rhythm and anticipates spots.anticipates spots.
SpotsSpots