Page 1: GE Logiq 200 Pro Service

TechnicalPublications

2262352Revision 1

LOGIQt200 PRO SeriesInSite Option Installation Instructions

CopyrightE 1999, 2000 By General Electric Co.used by GEMS- -Europe, GEMS- -Asia and GEMS- -Americas

Operating Documentation

GE Medical Systems

Page 2: GE Logiq 200 Pro Service

SAMSUNG GE MEDICAL SYSTEMS, LTD.SERVICE TEAM

64--3, Sangdaewon--dong, Chungwon--ku, Sungnam--siKyunggido, KOREA

Page 3: GE Logiq 200 Pro Service

REPORT ON TECHNICAL PUBLICATION -- RT(OMISSIONS,ERRORS,SUGGESTIONS)

PRODUCT PUBLICATION #

PAGE # PAGE REV DATA REPORTED

(DO NOT TEAR) FOLD HERE AND SEAL (DO NOTTEAR)

SERVICE RECORD SUBMITTED YES NO

PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS

FOLD FOLD

IF ADDITIONAL PAGES ARE INCLUDED, FOLD AND STAPLE TO LOWER PART OF THISSHEET BEFORE MAILING.

SERVICE REPPHONEFAXCOUNTRYCITY

REV0

Stick along this line

2184996

Page 4: GE Logiq 200 Pro Service

R & D Support Group Of SGMSFAX Number: 82--342--742--0436

REV. 02184997

Page 5: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS

2262352

LOGIQTM 200 InSite Option--Installation Instructions

A

REV 1

LIST OF EFFECTIVE PAGES

REV DATE PRIMARY REASON FOR CHANGE

0 NOV. 22, 1999 Initial release. . . . . . .1 MAR. 1, 2000 System V3.1 release. . . . . . . .

PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV

Title page 1. . . . . . . . . .GE Logo page --. . . . . .P9030CB 2. . . . . . . . . . .P9030CC 2. . . . . . . . . .P9030CD 0. . . . . . . . . .

A 1. . . . . . . . . . . . . . . . . .1 to 16 1. . . . . . . . . . . . .

Page 6: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION1

REV 1

1. InSite Option Installation

1--1 INTRODUCTION

These Service Instructions describe the installation of the InSite Service Capability for LOGIQ 200 PRO system.

1--2 TOOLS REQUIRED

Standard field Service Tool kitLOGIQ 200 PRO Service Manual (2235374)LOGIQ 200 PRO Proprietary Manual (2242594)

1--3 TIME REQUIRED

Approximately 2 hours plus travel time

1--4 PROCEDUREREAD and UNDERSTAND these instructions thoroughly before proceeding with the installation. Perform each step insequence and check it off when completed. If a problem occurs after completion, go back through the procedure andcheck for implementation errors before contacting your Ultrasound Region Engineer.Check each step as it is completed.

Functional Check--outPerform functional checks to verify that the system is functioning properly. Refer to all Functional Checks inChapter 4 of the LOGIQ 200 PRO Service Manual, 2235374.

NoteThe LOGIQ 200 PROsystem supports 2 kinds of modem -- US Robotics and MultiTech.Installation procedures are described separately in the following chapters of this insturction.

Page 7: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION2

REV 1

2. InSite Installation for US Robotics Modem

2--1 PARTS REQUIRED

ITEM PART NAME QTY PART # REMARKS1. Modem 1 2135816--2 US Robotics V.Everything2. RS232C Cable 1 local3. LOGIQ 200 PRO InSite Option 1 2262352

Installation Instructions

NoteThe LOGIQ 200 PRO Modem includes the preparatory RS232C cable to validate the InSitefunction. Never use this cable when scanning patient to prevent the patient injury.

Page 8: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION3

REV 1

2--2 Setting-Up Modem

1. Make sure that the DIP switches located at the rear side of the modem are set as shown in NO TAG.

1 2 3 4 5 6 7 8 9 10

quadOFF

ON

DIP SWITCH SETTINGSILLUSTRATION 1--1

2. Connect the power cord with the modem and plug it with a wall outlet. Refer to NO TAG.

3. Power the modem ON.

4. Power the modem OFF to connect the modem to LOGIQ 200 PRO.system.

2--3 Connecting Modem

1. Make sure that both the LOGIQ 200 PRO system and the modem are powered OFF.

2. Plug the D--sub 25--pin Connector (System side) into the Service port of the LOGIQ 200 PRO Rear Panel.

3. Plug the D--sub 25--pin Connector (Modem side) into the RS232 port of the US Robotics V.EverythingModem.

4. Connect the Modular Audio Cable between the phone jack ( ) of the modem and the dedicated analogtelephone line. Refer to ILLUSTRATION 1--1.

5. Connect one side of the power cord (supplied with the Modem Set) to the Modem as shown in NO TAG.

6. Connect another side of the power cord to a wall outlet.

Page 9: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION4

REV 1

2--3 Connecting Modem (Continued)

Rear Panel ofLOGIQ 200

RS232C Connector[System side](supplied with RS232C Cable Set)

RS232C Connector[Modem side](supplied with RS232C Cable Set)

Rear Panel OfUS Robotics V. Everything

Modular Audio Cable(supplied with Modem Set)

AC/DC Adaptor

Dedicated Analog Telephone Line

Wall Outlet

NoteThe type of power plug is different depending on the region.

MODEM CONNECTIONILLUSTRATION 1--2

Page 10: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION5

REV 1

2--4 Setting-Up LOGIQ 200 PRO Series

1. Power the LOGIQ 200 PRO system ON.

2. Power the modem ON after completing the LOGIQ 200 PRO power-up sequence.

3. Press Setup Menu key to go directly to the Setup/Top Menu.

4. In the Setup Top menu, Using the Trackball or arrow keys, choose General System 2 Setup and press Set.Refer to NO TAG .

Set to go to General System Setup Menu

SETUP MENU : Top Menu 01/13/99 11:12:35CATEGORY : OBSTETRICS PRESET :1 Ver. 3.0COMMAND : EXIT SAVE DELETE

General System 1 Setup*General System 2 SetupProbe Parameter 1 SetupProbe Parameter 2 SetupImage Display & Application SetupBody Pattern SetupComment SetupMeasurement SetupPatient Entry SetupUser Define Category & Key SetupUser Utility

Activating InSite Option--Service Port Setting (Setup Top Menu)ILLUSTRATION 2--1

Page 11: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION6

REV 1

2--4 Setting-Up LOGIQ 200 PRO Series (continued)

5. The General System 2 Setup Menu will appear on the monitor as shown in NO TAG.

6. In the General System 2 Setup Menu, Using the Track ball or Arrow Up/Down keys choose “Service Port”.

7. Choose “4:Insite”by inputing the number and press Set .

SETUP MENU : General System 2 01/13/99 11:14:16

COMMAND : EXIT SAVE DELETE

Record 1 :NoneRecord 2 :NonePrinter-A Port :1 (1.None 2.Printer)Printer-B Port :1 (1.None 2.Printer 3.DMC)

* Service Port :4 (1.None 2.Printer 3.DMC 4.InSite)Start OB Report page :1 (1.Summary Page 2.Graph Page)OB Trend Graph Page :1 (1.1graph 2.5graphs)EFW GP :1 (1:Hadlock 2:William 3:Brenner)EFW USA :AutoEFW Europe :1 (1.Shp/Wrs 2.Rch/Brk 3.Hdlck 4.Grm 5.Merz)EFW Tokyo :1 (1.TYO 2.TYO-S1 3.TYO-S2 4.TYO-S3)OB Report Form :1 (1.Tokyo 2.Osaka 3.USA 4.Europe)Hadlock table :1 (1:’84 2:’82)OB User Table : Sub menuBSA Type :1 (1.Oriental 2.Occidental)Display Body pattern Menu :1 (1.Yes 2.No)Display Comment Menu :1 (1.Yes 2.No)Auto Patient data Search :1 (1.Yes 2.No)New Patient Menu Display :1 (1.Yes 2.No)

Activating InSite Option Setting (General System 2 Setup)ILLUSTRATION 2--2

Page 12: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION7

REV 1

2--5 Activating InSite Option

1. Enter the service software after completing the power--up sequence. For the service software, refer to LO-GIQ 200 PRO Series Proprietary Manual, 2242594. The Service Utility Menu:Top Menu will be displayedas shown in NO TAG. Refer to the latest service note for the password.

2. Move the cursor to the item “Option Set”and press Set key. Option Set Menu will be displayed.

Service Utility MENU : Top Menu 99/03/12 09:16:37Current Software Version is 3.0

System DiagnosisBoard/Program VersionSystem ResetKey Operation History

* Option SetInitialize :1 (1:America 2:Europe 3:Asia 4:Japan)System Clock AdjustSoftware/Probe data upgradeExit

SERVICE TOP MENUILLUSTRATION 2--3

Page 13: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION8

REV 1

2--5 Activating InSite Option (continued)

3. The Option Set Menu will appear on the monitor as shown in the NO TAG. The status of each option will beindicated.

Service UtilitY MENU : Option Set 99/03/12 09:16:37Current Software Version is 3.0

CINE :1 (1:Not Install 2:Install)ATO :1 (1:Not Install 2:Install)MOD Driver :1 (1:Not Install 2:Install)Biopsy Guideline :1 (1:Not Install 2:Install)MGOB Calculation :1 (1:Not Install 2:Install)Cardiac Calculation :1 (1:Not Install 2:Install)Urology Calculation :l (1:Not Install 2:Install)Serial Printer :l (1:Not Install 2:Install)Paralled Printer :1 (1:Not Install 2:Install)Data Management Center:l (1:Not Install 2:Install)

* InSite :2 (1:Not Install 2:Install)

Exit

OPTION SET MENUILLUSTRATION 2--4

4. Using the Arrow up/down keys or Trackball , select the InSite option item and then Input the number 2:Install.

5. To activate the Option, exit the menu and reset the System.

6. This completes the InSite option installation of the LOGIQ 200PRO Series.

Page 14: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION9

REV 1

3. InSite Installation for MultiTech Modem

3--1 PARTS REQUIRED

ITEM PART NAME QTY PART # REMARKS1. Modem 1 2245773 MultiTech2. RS232C Cable 1 local3. LOGIQ 200 PRO InSite Option 1 2262352

Installation Instructions

NoteThe LOGIQ 200 PRO Modem includes the preparatory RS232C cable to validate the InSitefunction. Never use this cable when scanning patient to prevent the patient injury.

Page 15: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION10

REV 1

3--2 Setting-Up Modem

1. Connect the power cord with the modem and plug it with a wall outlet. Refer to NO TAG.

2. Power the modem ON.

3. Power the modem OFF to connect the modem to LOGIQ 200 PRO.system.

3--3 Connecting Modem

1. Make sure that both the LOGIQ 200 PRO system and the modem are powered OFF.

2. Plug the D--sub 25--pin Connector (System side) into the Service port of the LOGIQ 200 PRO Rear Panel.

3. Plug the D--sub 25--pin Connector (Modem side) into the RS232 port of the MultiTech Modem.

4. Connect the Modular Audio Cable between the phone jack of the modem and the dedicated analogtelephone line. Refer to ILLUSTRATION 1--3.

5. Connect one side of the power cord (supplied with the Modem Set) to the Modem as shown inILLUSTRATION 1--3.

6. Connect another side of the power cord to a wall outlet.

Page 16: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION11

REV 1

3--3 Connecting Modem (Continued)

Rear Panel ofLOGIQ 200

RS232C Connector[System side](supplied with RS232C Cable Set)

RS232C Connector[Modem side](supplied with RS232C Cable Set)

Rear Panel OfMultiTech Modem

Modular Audio Cable(supplied with Modem Set)

AC/DC Adaptor

Dedicated Analog Telephone Line

Wall Outlet

NoteThe type of power plug is different depending on the region.

MODEM CONNECTION .ILLUSTRATION 1--3

Page 17: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION12

REV 1

3--4 Setting-Up LOGIQ 200 PRO Series

1. Power the LOGIQ 200 PRO system ON.

2. Power the modem ON after completing the LOGIQ 200 PRO power-up sequence.

3. Press Setup Menu key to go directly to the Setup/Top Menu.

4. In the Setup Top menu, Using the Trackball or arrow keys, choose General System 2 Setup and press Set.Refer to NO TAG .

Set to go to General System Setup Menu

SETUP MENU : Top Menu 01/13/99 11:12:35CATEGORY : OBSTETRICS PRESET :1 Ver. 3.0COMMAND : EXIT SAVE DELETE

General System 1 Setup*General System 2 SetupProbe Parameter 1 SetupProbe Parameter 2 SetupImage Display & Application SetupBody Pattern SetupComment SetupMeasurement SetupPatient Entry SetupUser Define Category & Key SetupUser Utility

Activating InSite Option--Service Port Setting (Setup Top Menu)ILLUSTRATION 2--5

Page 18: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION13

REV 1

3--4 Setting-Up LOGIQ 200 PRO Series (continued)

5. The General System 2 Setup Menu will appear on the monitor as shown in NO TAG.

6. In the General System 2 Setup Menu, Using the Track ball or Arrow Up/Down keys choose “Service Port”.

7. Choose “4:Insite”by inputing the number and press Set .

SETUP MENU : General System 2 01/13/99 11:14:16

COMMAND : EXIT SAVE DELETE

Record 1 :NoneRecord 2 :NonePrinter-A Port :1 (1.None 2.Printer)Printer-B Port :1 (1.None 2.Printer 3.DMC)

* Service Port :4 (1.None 2.Printer 3.DMC 4.InSite)Start OB Report page :1 (1.Summary Page 2.Graph Page)OB Trend Graph Page :1 (1.1graph 2.5graphs)EFW GP :1 (1:Hadlock 2:William 3:Brenner)EFW USA :AutoEFW Europe :1 (1.Shp/Wrs 2.Rch/Brk 3.Hdlck 4.Grm 5.Merz)EFW Tokyo :1 (1.TYO 2.TYO-S1 3.TYO-S2 4.TYO-S3)OB Report Form :1 (1.Tokyo 2.Osaka 3.USA 4.Europe)Hadlock table :1 (1:’84 2:’82)OB User Table : Sub menuBSA Type :1 (1.Oriental 2.Occidental)Display Body pattern Menu :1 (1.Yes 2.No)Display Comment Menu :1 (1.Yes 2.No)Auto Patient data Search :1 (1.Yes 2.No)New Patient Menu Display :1 (1.Yes 2.No)

Activating InSite Option Setting (General System 2 Setup)ILLUSTRATION 2--6

Page 19: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION14

REV 1

3--5 Activating InSite Option

1. Enter the service software after completing the power--up sequence. For the service software, refer to LO-GIQ 200 PRO Series Proprietary Manual, 2242594. The Service Utility Menu:Top Menu will be displayedas shown in NO TAG. Refer to the latest service note for the password.

2. Move the cursor to the item “Option Set”and press Set key. Option Set Menu will be displayed.

Service Utility MENU : Top Menu 99/03/12 09:16:37Current Software Version is 3.0

System DiagnosisBoard/Program VersionSystem ResetKey Operation History

* Option SetInitialize :1 (1:America 2:Europe 3:Asia 4:Japan)System Clock AdjustSoftware/Probe data upgradeExit

SERVICE TOP MENUILLUSTRATION 2--7

Page 20: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION15

REV 1

3--5 Activating InSite Option (continued)

3. The Option Set Menu will appear on the monitor as shown in the NO TAG. The status of each option will beindicated.

Service UtilitY MENU : Option Set 99/03/12 09:16:37Current Software Version is 3.0

CINE :1 (1:Not Install 2:Install)ATO :1 (1:Not Install 2:Install)MOD Driver :1 (1:Not Install 2:Install)Biopsy Guideline :1 (1:Not Install 2:Install)MGOB Calculation :1 (1:Not Install 2:Install)Cardiac Calculation :1 (1:Not Install 2:Install)Urology Calculation :l (1:Not Install 2:Install)Serial Printer :l (1:Not Install 2:Install)Paralled Printer :1 (1:Not Install 2:Install)Data Management Center:l (1:Not Install 2:Install)

* InSite :2 (1:Not Install 2:Install)

Exit

OPTION SET MENUILLUSTRATION 2--8

4. Using the Arrow up/down keys or Trackball , select the InSite option item and then Input the number 2:Install.

5. To activate the Option, exit the menu and reset the System.

6. This completes the InSite option installation of the LOGIQ 200PRO Series.

Page 21: GE Logiq 200 Pro Service

GE MEDICAL SYSTEMS LOGIQ 200 PRO Series InSite Option Installation Instructions

2262352

INSTALLATION16

REV 1

This page is left blank intentionally.