  • Model ZR22G, ZR402G, and ZR202GDirect In Situ Zirconia Oxygen Analyzersand High temperature Humidity Analyzers

    GS 11M12A01-E-H 2nd Edition

    The basic analyzer consists of a direct insertion type probe and aconverter that provides an analog output for control or monitoringpurposes. In addition a digital signal is superimposed on thecurrent output (HART) for optimization of maintenance. Theconverter controls also the (semi) automatic calibration when usedwith the calibration units.The detector and the converter are combined in one unit (ZR202)for simplified piping and wiring or the converter is mountedseparately (ZR402)Various accessories are available for optimal installation in difficultapplications ( dust filter, flame arrester, probe protector, hightemperature probe adapter, probe supporter) and for ease ofmaintenance and calibration.The analyzer is most commonly used to monitor the Oxygen levelin combustion air to allow optimal burner control to avoid airpollution and save energy. The direct insertion installation isespecially attractive from a maintenance point of view: samplingsystems are not required. The analyzer can also be used to monitor the Humidityconcentration at high temperatures in air with water as found inhumidifying processes like Bakeries or in indirectly heated dryers asfound in Paper machines.

    FEATURES Direct measurement: No sampling required Field replaceable unique Zirconia cell, made to last. Heavy duty sensor heater Rebuildable probe construction Various methods for reference air to optimize accuracy

    versus ease of installation Automatic compensation of duct pressure possible. Remote maintenance and set-up by digital communication

    (HART)

    GeneralSpecifications

    SYSTEM CONFIGURATION

    ZR22G ZR402G

    ZR202G

    Instrument air

    Calibration gas pressure regulator

    Air Set

    Needle valveflowmeter

    Reference gas

    Calibration gas

    Check valve ~

    Zero gas cylinder

    Model ZR40H Auto Calibration unit

    Model ZR22G Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Detecor

    Model ZR402G Converter

    Calibration gasunit case

    EXA ZR402G

    Conduit *1

    *2

    *3

    *4

    100 to 240 V ACContact inputAnalog output, contact outputDigital output (HART)

    Signal (6-core shield cable)

    Heater (2-core shield cable)

    ~ 100 to 240 V ACAuto Calibration unitZ R20H

    Reference gas

    Calibration gas

    Contact inputAnalog output, contact outputDigital output (HART)

    Air Set

    Instrument air

    Calibration gasunit case

    Calibration gas pressure regulator

    Zero gas cylinder

    Model ZR202G Intgrated type Zirconia Oxygen/ High temperature Humidity Analyzerwith auto calibration (ZR202G-hhh-h-h-A-h-h-h-h-A)

  • 2

    GS 11M12A01-E-H

    Process gas temperature 0 to 600 C Process gas temperature 0 to 1400 C

    MountingInsertionlength

    General-use Probe Application

    Horizontalto

    vertical

    Horizontal

    Vertical

    0.4to2m

    3m

    Boiler Heating furnace Low dust level

    For pulverized coil boiler with gas flow velocity 10 m/s or more Cement kiln High dust level

    3m

    0.4to2m

    Horizontalto

    vertical

    1.0to2m

    Horizontalto

    vertical

    Boiler Heating furnace Low dust level

    Black liquid recovery boiler Cement kiln High dust level

    Vertical 3m

    High temperature detector

    Heating furnace 0 to 5.0kPa

    - 0 to 5.0kPa

    General-purposedetector ZR22G-

    General-purpose detector with probe supporter ZR22G- with ZO21V-

    General-purpose detectorwith protector and filterwith: ZR22G-

    ZO21R-L-K9471 UA

    General-purpose detectorwith filterwith: ZO21D-L

    E7042UQ

    ZR22G-015ZO21P-HNeedle valvenot supplied

    ZA22G-015ZO21P-HE7046EC

    * Note: Downward oriented mounting preferred

    *

    Examples of system component selection based on sample gas condition

  • 3

    GS 11M12A01-E-H

    STANDARD SPECIFICATIONS (Oxygen analyzer)

    General specificationsMeasurement range : 0.01 to 100-vol% O2Output signal : 4- 20 mA DC, maximum load 550 OhmOutput span : 0-5 vol% and for zero suppression any

    span with minimum of 0.3 x concentration at 4 mA

    Digital Communication: Frequency Shift Key following HART

    protocol; 250- 550 Ohm, depending onnumber of field devices (multi drop mode)

    Warm-up time : 20 minutes

    Performance specificationsRepeatability : 0.5% FS for set range 25 vol%Linearity : 1% FS for set range 2.0 m) is onlypossible using probe supporter or probeprotector. (

  • 4

    GS 11M12A01-E-H

    4. ZO21P-H High Temperature Probe Adapterfor ZR22G detector

    Description : The ZO21P-H high temperature probeadapter consists of a Stainless steel TEEpiece with JIS5k32AFF flange forconnection to the ZR22G-015 detector.The process side of the TEE is blinded offwith a flange DIN PN10DN50 with an R _thread for mounting of the sampling probe.The other side of the TEE is blinded offwith a flange and supplies a R1/4 or 1/4"FNPT connection for ejector or valve.

    Sample temperature : 0-1400 C for SiC sampling probe0- 1000 C for SUS 310S sampling probe

    Sample pressure : -0.5 to + 5.0 kPa. For negative samplepressure an auxiliary ejector assembly isnecessary.

    Insertion length : 1.0, 1.5 mProcess connection : DIN PN10 DN50 flanged connectionMaterial : SUS 316 (adapter), SUS 304 (flange), SiC

    or SUS310S (sampling probe)Mounting : Vertically downward sampling probe with

    zero 5 degree from vertical (SiC samplingprobe. Horizontal mounting with SUS310Ssampling probe is possible

    Weight : 7 kg for 1m version; 8 kg for 1.5m version

    5. E7046EC/E7046EN Auxiliary ejector forZo21P-H probe adapter

    Ejector assembly : Inlet air pressure 29 to 68 kPa, airconsumption 30- 40 Nl/min, sample flowrate 3- 7 Nl/min, material SUS304,connection Rc 1/4 or 1/4" NPT, tubeconnection compression fitting for 6 mmOD copper or SS tubing.

    Pressure gauge : 0- 100 kPa, materials in contact with gasSUS316, case material coated Aluminum.

    Needle valve : Material SUS316, connection R 1/4 or 1/4"NPT

    6. ZO21R Probe Protector for Zirconia OxygenAnalyzer

    Used when sample gas flow velocity is approx. 10m/sec or moreand dust particles wears the detector in cases such as pulverizedcoal boiler of fluidized bed furnace (or burner) to protect thedetector from wearing by dust particles. When probe insertionlength is 2.5m or more and horizontal installation, specify theZO21R-L-uuu-u*B to reinforce the probe.

    Insertion Length : 1.05m, 1.55m, 2.05m.Flange : JIS 5K 65A FF equivalent. ANSI CLASS

    150-4-FF (without serration) equivalent orDIN PN10-DN50-A equivalent. However,flange thickness is different.

    Material : SUS316, SUS304 (Flange)Weight : 1.05m; Approx. 6/10/8.5kg (JIS/ANSI/DlN),

    1.55m; Approx. 9/13/11.5kg(JIS/ANSI/DIN), 2.05m; Approx.12/16/14.5kg (JIS/ANSI/DIN)

    Installation : Bolts, nuts, and washers are provided fordetector, probe adapter and process-sideflange.

    7. Filter for Oxygen Analyzer K9471UAThis filter is used to protect the cell from corrosive dustcomponents or high velocity dust in recovery boilers and cementkiln. Measured gas flow rate is needed to be 1m/sec or more toreplace gas inside zirconia sensor.

    Mesh : 30 microns Material : Carborundum (Filter), SUS316Weight : Approx. 0.2 kg

    8. L9852CB/K9471UN Stop ValveThe stop valve is mounted on the calibration gas line.It is attached when the suffix code (/SV) is selected for theZirconia Oxygen Analyzer/High-temperature Humidity Analyzerprove ZR22G or the Zirconia Oxygen Analyzer/High-temperatureHumidity Analyzer ZR202G.

    Connection : Rc 1/4 or 1/4 FNPT Material : SUS/BrassWeight : Approx. 80 g

    9. K9292DN/K9292DS Check Valve This is used to prevent entry of process gas into calibration gasline. Purpose is the same as stop valve, but is convenient, as itdoes not need to be opened or closed for calibration.Mount directly on calibration gas inlet of detector in place of stopvalve. However as source pressure of 50 kPa G or more isneeded.When option code /CV of the ZR22G or the ZR202G isspecified, check valve is provided.

    Connection : Rc1/4 or 1/4FNPTMaterial : SUS316Pressure : Between 70 kPa G or more 350 kPa G or

    lessWeigh : Approx. 40g

    *10. ZA8F Flow Setting Unit Used when instrument air is provided. This unit controls flowrates of calibration gas and reference gas and consists offlowmeter and flow rate control valve.

    FIowmeter : Calibration gas; 0.1 to 1.0 l/min. Reference

    air; 0.1 to 1.0 l/min.

    Construction : Dust-proof and rainproof construction

    Case Material : SPCC, Dark-green (Munsell 2.0 GY 3.1/0.5

    or equivalent)

    Painting : Baked epoxy resin

    Tube Connections : Rc1/4 or 1/4FNPT

    Reference Air pressure: Clean air supply of measured gas

    pressure+approx. 70 kPa G (pressure at

    inlet of the auto-calibration unit)

    Air Consumption : Approx. 1.5 l/min

    Weight : Approx. 2kg

    Note : Used instrument air for span calibration

    gas, if without instrument air is used,

    contact YOKOGAWA.

  • 5

    GS 11M12A01-E-H

    *11. ZR40H Auto-calibration Unit (for Separatetype)

    This unit is used when the instrument air is provided and theauto-calibration unit is attached. Solenoid valves are provided as standard equipment.

    Reference air pressure: Clean air supply of measured gas

    pressure+approx. 70 kPa G (pressure atinlet of the auto-calibration unit)

    Air consumption : Approx. 1.5 l/minWeight : Approx. 3.5 kg

    *12. ZR20H Auto-calibration Unit (for Integratedtype

    Used when automatic calibration is specified for the integratedtype and instrument air is provided.Equipped with the analyzer when automatic calibration isspecified in the suffix code of the ZR202G Integrated type byselecting either -A (Horizontal mounting) or -B (Verticalmounting). The ZR20H should be arranged when auto-calibration is to be required after the ZR202H has been installed.Ask Yokogawa service station for its mounting.

    Construction : Dust-proof and rainproof construction:NEMA4X/IP67 (excluding flowmeter)

    Mounting : Mounted on ZR202G, no vibrationMaterials : Body: Aluminum alloy, Piping: SUS316,

    SUS304, Flowmeter: MA (Methacrylateresin)

    Finish : Polyurethane corrosion-resistance coating,Case: Frosty white CC21, Cover: Deep-seamoss green CC32

    Piping Connection : Refer to Model and Suffix CodesPower Supply : 24V DC (from ZR202G), Power

    consumption: 1.3 WReference Air Pressure

    : Sample gas pressure + Approx. 70 to 100kPa (690 kPa max.), (Pressure at inlet ofauto-calibration unit)

    Air Consumption : Approx. 1.5 l/minWeight : Approx. 2 kgAmbient Temperature

    : -20 to +55C, no condensing and freezingAmbient Humidity : 0 to 95%RHStorage Temperature: -30 to +65C

    * Recommended calibration gas = 1% Oxygen in NitrogenFlow Rate 0.6 0.1 Nl/min

    13. Dust Protector ZH21BThis protector is designed to protect the detector Tip fromingress of dust (i.e., to prevent combustible materials fromentering the detector cell) where humidity measurements aremade under dusty environements.

    Insertion length : 0.416 mFlange : JIS 5 K 80 A FF equivalent or ANSI 4 B

    150 LB FF equivalentMaterials of Flange : SUS 316, SUS 304, Stainless SteelWeight : approximately 6/8.5 kg (JIS/ANSI)Mounting : mounted on the detector, and process

    flange with bolts and associated nuts andwashers

  • 6

    GS 11M12A01-E-H

    Configurations : All configurations are possible, but insertion length of 0.4m is most common

    Options : ZH21B Dust protector for high dust environment. Insertion length 0.416 m flange JIS 5k-80 or ANSI 150# 4",material SUS 316, Carborundum (SiC), SUS 304 (flange), weight 6 kg

    Standard specifications for use as HumidityAnalyzer

    Description : The Zirconia Oxygen analyzer can also beused to monitor the absolute humidity in anenvironment where standard air is mixedwith water vapor. Standard air consists of79% Nitrogen and 21% oxygen. If the air ismixed with Water vapor, the vol% ofoxygen decreases as the vol% of Waterincreases. The Zirconia based Oxygenanalyzer is therefore very useful as ananalyzer for moisture content of the samplegas. Both the ZR402 and the ZR202 canbe software switched from Oxygen analysisto Humidity analysis.

    Sample composition : The sample gas may only contain a mixtureof Oxygen, Nitrogen and Water.

    Performance : The performance specifications aredetermined by the primary measurement:the measurement of Oxygen. The possibleerrors in humidity units can be calculatedfrom the possible errors in the Oxygenanalysis.

    Range : 1. Percent by volume: range 0- 100%,minimum span 0- 25 vol%2. Weight Ratio: kg H2O/ kg dry air: range1- 1.00, min span 0- 0.200 kg/kg dry air3. Dew point: range 0- 100 C4. Relative Humidity: range 0- 100%

    Calculation

    air

    0% 100%water

    100 %

    O2

    H2O

    N2

    0%100%

    0%

    x

    a

    Calculation Dew Point, calculation rH

  • 7

    GS 11M12A01-E-H

    ZR402G

    Converterthread

    -M M20x1.5 mm

    Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Converter

    Model Suffix code Optioncode Description

    Certificate

    Tag plates

    *1 For humidity measurements, be sure to specify /HS options.*2 Specify either /SCT or /PT option code.*3 Will be avalable in the near future.

    -E-G

    Display

    -A

    -E

    /HS

    /Q

    /H

    Instruction manual

    Options

    /SCT/PT

    Set for Humidity Analyzer (*1)

    QIS/QIC

    Hood (*3)

    English

    EnglishGerman

    Always -A

    Stainless steel tag plate (*2)Printed tag plate (*2)

    MODEL AND SUFFIX CODES

    1. Separate type Zirconia Oxygen / Hightemperature Humidity Analyzer, Converter

    ZR202G

    ANSI CLASS150-4-RF (Equivalent)DIN PN10-DN50-A (Equivalent)

    -C-E

    -040-070-100-150-200

    -S-C

    0.4 m0.7 m1.0 m1.5 m2.0 m

    -C-E

    Rc 1/4-R

    M20x1.5 mm-M

    -A

    -E

    /C

    Set for Humidity Analyzer (*4)/HS

    /CV/SV

    Hood (*8)/H

    QIS/QIC/Q

    /F1

    /SCT/PT

    Auto Calibration No auto calibration unit mountedAuto calibration unit mounted (*8)Vertical mounting

    -N-A-B

    integrated type Zirconia Oxygen/ High temperature Humidity Analyzer

    Model Suffix code Optioncode Description

    Length

    Wetted material SUS316Stainless steel with Inconel calibration gas tube

    Flange(*2)

    Reference air Natural convectionPressure compensation

    Gas Thread

    Connection box thread

    Instruction manual English

    Always -A

    Options

    Inconel bolt (*3)

    Check valve (*5)Stop valve (*5)

    Dust Filter (*6)

    Stainless steel tag plate (*7)Printed tag (*7)

    Durethane coating/D

    *1 For the horizontal installed probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify ZO21R-L-hhh-h. Specify the flange suffix code either -C or -K.*2 The thickness of the flange depends on its dimensions.*3 Inconel probe bolts are used. Use this option for high temperature use (ranging from 600 to 700 8C). *4 For humidity measurements, be sure to specify /HS options. Pressure compemsation of reference air can not be selected.*5 Specify either /CV or /SV option code.*6 Not used with the high-temperature humidity analyzer.*7 Specify either /SCT or /PT option code.*8 Will be avalable in the near future.

    Flame Arrester/A

    2. Integrated type Zirconia Oxygen / High temperature Humidity Analyzer, Converter

  • 8

    GS 11M12A01-E-H

    3. Separate type Zirconia Oxygen / High Temperature Humidity Analyzer, Detectors

    Model

    ZR22G

    Flange(*3)

    ANSI CLASS150-4-RF (Equivalent)DIN PN10-DN50-A (Equivalent)JIS 5K-32-FF (Equivalent) (for high temperature use) (*4)

    -C-E-Q

    Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Detector

    Length

    Wetted material

    -015-040-070-100-150-200

    -S-C

    0.15 m (for high temperature use) (*1)0.4 m0.7 m1.0 m1.5 m2.0 m

    SUS316Stainless steel with Inconel calibration gas tube

    Suffix code Optioncode Description

    Reference air Natural convectionPressure compensation

    -C-P

    Gas Thread Rc 1/4-R

    Connection box thread

    Instruction manual

    M20 x1.5 mm-M

    Always -A-A

    English-E

    Options

    Valves

    Certificate

    Filter

    Tag plates

    *1 Used with the ZO21P High temperature Probe Adapter. Select flange (-Q).*2 When installing horizontally the probe whose insertion length is 2.5 meters or more, use the Probe Protector. Be sure to specify ZO21R-L-hhh-h. Specify the flange suffix code either -C or -K.*3 The thickness of the flange depends on its dimensions.*4 Not used in conjunction with P (pressure compensation) for reference air.*5 Inconel probe bolts are used. Use this option for high temperature use (ranging from 600 to 700 8C). *6 Specify either /CV or /SV option code.*7 Not used with the high temperature humidity analyzer.*8 Specify either /SCT or /PT option code.*9 No water proof,avoid rain. Operating maximun temperature is 808C.Note: ZR22G can not be conected to Model ZA8C, AV8C, HA400 converter.

    Inconel bolt (*5)/C

    Durethane coating/D

    Check valve (*6)Stop valve (*6)

    /CV/SV

    Dust Filter (*7)/F1

    Flame Arrester/A

    QIS/QIC/Q

    Stainless steel tag plate (*8)Printed tag plate (*8)

    /SCT/PT

  • 9

    GS 11M12A01-E-H

    Joint: RC 1/4, Material: SUS316

    Joint: 1/4 NPT, Material: BS

    DescriptionPart No.

    L9852CB

    K9471UN

    Joint: RC 1/4, Material: SUS304

    Joint: 1/4 NPT, Material: SUS304

    Part No.

    K9292DN

    K9292DS

    Description

    ZR40H

    -R

    -M

    Rc 1/4

    Model Suffix code Option code Description

    Automatic calibration unit for ZR402G

    Gas piping connection

    Wiring connection

    - -A

    20 mm (M20 x 1.5)

    Always -A

    ZR20H

    -R

    -E

    Rc 1/4

    Model Suffix code Option code Description

    Automatic calibration unit for ZR202G*1

    Gas piping connection

    Reference air *2

    - -A

    Instrument air

    Always -A

    -P Pressure compensated

    Mounting -A Horizontal mounting

    -B Vertical mounting

    *1 Ask Yokogawa service station for additional mounting of ZR20H to the preinstalled ZR202G.*2 Select the appropriate reference air of ZR20H according to the one of ZR202G.

    11. Automatic Calibration Unit for Separatetype Analyzer (Needs instrument air.)

    12. Automatic Calibration Unit for Integratedtype Analyzer (Needs instrument air.)

    8. Stop Valve for Calibration-gas line

    9. Check Valve for Calibration-gas line

    1.0 m1.5 m

    DIN PN10-DN50-A equivalentANSI Class 150-4-RF

    ZO21P

    SiCSUS 310S

    -100-150

    *A

    -E-A

    -H

    -A-B

    High Temperature Probe Adapter

    Material

    Insertionlength

    Flange

    Style AStyle code

    Model Suffix code Optioncode Description

    ZH21B

    * The flange thickness varies. Specify the probe ZR22G-040.

    DescriptionOption codeSuffix codeModel

    Dust protector (0 to 6008C)

    0.409 mInsertion length -040

    Flange

    Style code *B

    -A ANSI Class 150-4B FF equivalent *

    Style B

    ZA8F

    -J-A

    Model Suffix code Option code Description

    Flow setting unit

    JointWith 1/4" NPT adapter

    Style A

    Rc 1/4

    Style code *A

    Part No.

    E7046AL

    E7046BB

    E7046AP

    E7046AQ

    SiC, insertion length 1.0 m

    SiC, insertion length 1.5 m

    SUS310S, insertion length 1.0 m

    SUS310S, insertion length 1.5 m

    Description

    RC 1/4 6 / 4 TUBE joint: SUS304

    Part No.

    E7046EC

    Description

    ZO21R

    DIN PN10-DN50-A equivalent

    -L Probe Protector(0 to 700 8C)

    -100-150-200

    1.05 m (3.5ft)1.55 m (5.1ft)2.05 m (6.8ft)

    Insertionlength

    Flange ( *1)

    *1 Thickness of flange depends on dimensions of flange.

    ANSI Class 150-4-RF equivalent

    Model Suffix code Optioncode Description

    *B Style BStyle code

    -E-A

    Part No.

    K9471UA

    Description

    Filter

    4. Adapter for High temperature Probe ofseparate type Oxygen Analyzer

    Note: For this high-temperature use probe adapter, be sure tospecify the ZR22G probe of its insertion length 0.15meters.

    High temperature Probes (Spare Parts)

    5. Auxiliary Ejector for High-temperature Useof separate type Oxygen Analyzer

    7. Filter for Zirconia Oxygen Analyzers

    13. Dust Protector for High-temperatureHumidity Analyzers

    10. Flow setting unit for manual calibration(Needs instrument air.)

    6. Probe Protector for Zirconia Oxygen Analyzer

  • 10

    GS 11M12A01-E-H

    1 to 6 (Panel Thickness)

    EXA ZR402G 4 - [ 6 holesfor Wall mounting

    4-R8 to R10or4-C5 to C8

    4 - [ 6 holes

    8-G1/2, 8-1/2NPT etc(Wiring connection)

    ( for wall mounting)

    F08_01.EPS

    2-inch mounting pipe

    Wall mounting Panel Cut-out

    100

    108228

    280

    126.

    5

    57.3

    136.3 54.7

    111

    36

    120.

    2

    (1/2

    NP

    T)

    190 183+20

    274

    +2

    0

    126.

    5

    1. Model ZR402G Separate type Zirconia Oxygen/ High temperature Humidity Analyzer, Converter

  • 11

    GS 11M12A01-E-H

    153 to 164

    125

    L

    170252 to 265 49

    48.5

    338 to 351

    25

    t

    122

    L= 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 (m)

    4-G1/2,2-1/2NPT etc.Cable connection port

    Display side

    15663

    25664 49

    34264

    25

    t

    PIPING:A

    PIPING:B

    L

    L= 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 (m)

    4-G1/2,2-1/2NPT etc.Cable connection port

    [50

    .8

    Rc1/4 or 1/4NPTReference air inlet

    Rc1/4 or 1/4NPTCalibration gas inlet

    Reference air outlet

    Stopvalve

    PIPINGABBABBAABBBBA

    ANSI CLASS150-2-RF equivalentANSI CLASS150-3-RF equivalentANSI CLASS150-4-RF equivalentDIN PN10-DN50-A equivalentDIN PN10-DN80-A equivalentDIN PN10-DN100-A equivalentJIS 5K-65-FFJIS 10K-65-FFJIS 10K-80-FFJIS 10K-100-FFJPI CLASS150-4-RF equivalentJPI CLASS150-3-RF equivalentWestinghouse

    152.4190.5228.6165200220155175185210229190155

    120.6152.4190.5125160180130140150175

    190.5152.4127

    4 - [194 - [198 - [194 - [188 - [188 - [184 - [154 - [198 - [198 - [198 - [194 - [19

    4 - [11.5

    A B C19242418202014181818242414

    tFlange

    [ 5

    0.8

    Rc1/4 or 1/4NPTReference air inlet

    Rc1/4 or 1/4NPTCalibration gas inlet

    [123

    Terminal side

    C

    Flange

    Flange

    [B

    C

    Flange

    [A

    [A

    [B

    2. Model ZR202G Integrated type Zirconia Oxygen/ High temperature Humidity Analyzers

    Model ZR202G...-P (with pressure compensation) Integrated type Zirconia Oxygen / High temperatureHumidity Analyzers

  • 12

    GS 11M12A01-E-H

    25

    t285 64

    [50

    .8

    [12

    7

    85

    L

    48

    15663 69

    L=0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 3.6, 4.2, 4.8, 5.4 (m)

    C

    Reference gas outlet

    25 48

    153 to 164 69

    Rc1/4 or 1/4NPTReference air inlet

    2-G1/2,2-1/2NPT etc.Cable connection port

    Rc1/4 or 1/4NPTCalibration gas inletFlange

    PIPING : A

    PIPING:B

    Rc1/4 or 1/4NPTReference air inlet

    2-G1/2, 2-1/2NPT etc.Cable connection port

    [B

    [A

    C

    Flange

    Flange

    [B

    [A

    Rc1/4 or 1/4NPTCalibration gas inlet

    Flange

    Stop Valve

    EXTERNAL DIMENSIONS

    3. Model ZR22G Separate type Zirconia Oxygen / High temperature Humidity Analyzer, Detectors

    Model ZR22G...-P (with pressure compensation) Separate type Zirconia Oxygen / High temperatureHumidity Analyzer, Detectors

    PIPINGANSI CLASS150-2-RF equivalentANSI CLASS150-3-RF equivalentANSI CLASS150-4-RF equivalentDIN PN10-DN50-A equivalentDIN PN10-DN80-A equivalentDIN PN10-DN100-A equivalentJIS 5K-65-FFJIS 10K-65-FFJIS 10K-80-FFJIS 10K-100-FFJPI CLASS150-4-RF equivalentJPI CLASS150-3-RF equivalentWestinghouse

    152.4190.5228.6165200220155175185210229190155

    120.6152.4190.5125160180130140150175

    190.5152.4127

    4 - [194 - [198 - [194 - [188 - [188 - [184 - [154 - [198 - [198 - [198 - [194 - [19

    4 - [11.5

    A B C19242418202014181818242414

    ABBABBAABBBBA

    tFlange

  • GS 11M12A01-E-H

    13

    R1/2(Note2)

    Approx. 351

    180 Flange (Thickness)JIS 5K 32A FF equivalent

    Gasket (Thickness 3)

    Detector(ZR22G)

    Flange

    52 over

    30

    60.5

    Reference air inlet

    Calibration gas inlet

    High temperatureProbe SiC pipe(GK material)

    110

    115

    4225

    170

    App

    rox,

    215

    (Ins

    etio

    n le

    ngth

    ) (N

    ote1

    )

    App

    rox,

    100

    69

    Approx. 48

    C

    (Note 1) 1=1000 or 1500 (mm)(Note 2) Sample gas outlet

    (if the sample gas pressure is negative, connect the auxiliary ejector.)

    127

    85

    Measurement gas outlet

    Flange providedby customer

    Unit: mm

    60.5A

    B

    A

    Pipe hole (2- G1/2, 2-1/2NPT, etc)

    Needle valve

    Pressure gauge

    Nozzle (Note1)

    R1/2Ejector assembly Detector

    Rc1/4 or 1/4 NPT 6/ 4 or 1/4 inch copper tube (stainless) with ejector to connect Rc1/4

    4020Tee

    Blow

    43

    App

    rox.

    67

    App

    rox.

    88

    39

    App

    rox.

    70

    38

    Ful

    l ope

    n he

    ight

    (Note1) The connector of ejector assembly is a dedicated connector with nozzle function.

    Measurement gas inlet

    4. Model ZO21P Adapter for High temperature Probe of separate type Oxygen Analyzer

    5. Model E7046EC, E7046EN Auxiliary Ejector for High temperature Use of separate type OxygenAnalyzer

  • 14

    GS 11M12A01-E-H

    51

    32

    10

    Caborundum filter

    Increase of insertion length

    Flange

    JIS 5K-65-FFequivalent

    155

    228.6

    130

    190.5

    4- 15

    8- 19

    5

    12

    A B C t

    ANSI CLASS150-4-FF equivalent

    40

    50

    D

    D

    C

    B

    B

    Flange (with bolts, nuts, gasket and washer)

    gasket (t1.5)

    SUS316

    Dimensions ofholes on opposing surface

    A

    Gas flow

    Washer (12)Mounting nut (M12)

    l (Insert length)

    l=1050,1550,2050

    6. Model ZO21R Probe Protector for Zirconia Oxygen Analyzers

    7. Model K9471UA Filter for Oxygen Analyzer

    K9292DN : Rc 1/4(A),R 1/4(B)K9292DS : 1/4NPT(A),1/4NPT(M)(B)

    Approx. 19 Approx. 54

    66(F

    ull o

    pen

    leng

    th)

    40

    Rc1/4 or 1/4NPT

    48

    8. L9852CB/ K9471UN Stop Valve forCalibration-gas line

    9. K9292DN/ K9292DS Check Valve forCalibration-gas line

  • 15

    GS 11M12A01-E-H

    6 hole

    Reference air outlet Zero gas outlet Zero gas inlet Instrument air inlet

    4-Rc1/4

    20 2040 40

    Piping connectionport

    180

    235.

    870

    35

    268

    222.

    87

    140

    REFERENCE CHECK

    REFERENCE SPAN ZERO

    F20.EPS

    10. Model ZA8F Flow setting unit for manual calibration

  • 16

    GS 11M12A01-E-H

    Unit: mm

    OCK

    140

    250

    42

    26

    41.2 41.2

    54 71.526

    49.5

    1222

    3

    4616

    3040 47.5 25

    102

    MA

    X58

    90 116.5

    f 6.5

    *1 with four M6 screws can wall-mount

    *1 4-

    (wiring inlet is at same position on rear)

    Flowmeter

    Setting Valve forcalibration gas

    Setting Valve forreference air

    wiring inlet ; 2-G1/2,Pg13.5,M2031.5 or 1/2NPT(Female)

    calibration gas outlet Rc1/4 or 1/4 NPT(Female)

    reference air inlet

    Rc1/4 or 1/4 NPT(Female)zero gas inlet

    Rc1/4 or 1/4 NPT(Female)Rc1/4 or 1/4 NPT(Female)reference air outlet

    2B pipe mounting example

    11. Model ZR40H Automatic Calibration Unit for Separate type Analyzer

    ZR402G Converter ZR40H Automatic Calibration unit

    AC-Z

    AC-C

    AC-S

    Zero

    Span

    EV1

    EV2

    CHECKOUT

    ZERO GAS IN AIR IN

    Instrument air Approx.1.5 l/min.

    *2 Needle valve is supplied as accessory with flow meter

    flow meter

    flow meter

    REFOUT

    *2 *2solenoid valveEV1,2

    EV1 EV2

  • 17

    GS 11M12A01-E-H

    zero gas inlet

    Rc1/4 or 1/4NPT(Female)

    reference air inlet

    Rc1/4 or 1/4NPT(Female)

    terminal box side

    span gas inlet

    Rc1/4 or 1/4NPT(Female)

    zero gas inlet

    Rc1/4 or 1/4NPT(Female)

    Span gas inlet

    Rc1/4 or 1/4NPT(Female)

    reference air inlet

    Rc1/4 or 1/4NPT(Female)

    166.5

    166.5

    258

    60

    244

    214 MAX44

    4516

    0

    66.540 40

    66.540 40

    180 MAX44

    Vertical mounting on the ZR202G (-B)

    Horizontal mounting on the ZR202G (-A)Unit: mm

    display side

    12. Model ZR20H Automatic Calibration Unit for Integrated type Analyzer

    t

    C

    C

    [B

    [B

    [A

    [B

    76.3

    72

    A

    180

    228.5

    B

    145

    190.5

    C

    4-[19

    8-[19

    t

    12

    12

    D

    40

    50

    Unit: mm416 (Insertion length)

    Install facing upwards.Flange

    JIS 5 K FF equivalent

    ANSI 4 B 150 LB FF

    13. Model ZH21B Dust Protector for High temperature Humidity Analyzers Unit: mm

  • 18

    GS 11M12A01-E-H

    1FG

    2AO1(+)

    3AO1(-)

    4AO2(+)

    5AO2(-)

    6CJ(+)

    7CJ(-)

    8TC(+)

    9TC(-)

    10CELL

    (+)

    11CELL

    (-)12FG

    13DI-1

    14DI-2

    15DI-C

    16AI(+)

    17AI(-)

    18AC-

    ZERO

    19AC-

    SPAN

    20AC-

    COM

    21FG

    22FG

    CELL(+)

    CELL(-)

    TC(+)

    TC(-)

    CJ(+)

    CJ(-)

    23DO-1

    24DO-1

    25DO-2

    26DO-2

    27DO-3

    28DO-3

    29DO-4

    30DO-4

    31HTR

    32HTR

    33L

    34N

    35G

    36FG HTR HTR

    100-240V AC50/60 Hz

    Model ZR402G Separate type Zirconia Oxygen Analyzer/ Converter

    Model ZR22G Separate type Zirconia OxygenAnalyzer, Detector

    Analog output 2

    Analog output 1

    4-20 mA DC4-20 mA DCDigital output

    Contact input 2

    Contact input 1

    Contact input 1

    Contact input 2

    Contact input 3

    Contact input 4

    Solenoid valve for span gas for automatic calibration

    Solenoid valve for zero gas for automatic calibration

    Flow setting unit

    *1

    *1

    1DI-1

    2DI-2

    3DI-C

    4DO-1

    5DO-1

    6DO-2

    7DO-2

    8FG

    9AO(+)

    10AO(-)

    11L

    12N

    13G

    14FG

    Model ZR202G Integrated type Zirconia Oxygen / High temperature Humidity Analyzer

    Contact input 1Contact output 2Contact input 1

    Analog output 4-20 mA DC

    100 to 240 V AC, 50 or 60 Hz

    Digital output

    Contact input 2

    *1

    *1 Ground resistence is 100 ohm or less.

    WIRING CONNECTIONS

  • 19

    GS 11M12A01-E-H

    Inquiry Sheet for Models ZR22G, ZR402G, and ZR202G Direct In Situ Zirconia Oxygen Analyzers and High temperature Humidity Analyzers

    u vol%O2,

    u kg/kg,

    u 8C,

    u kPa,

    u m/sec,

    u g/Nm3,

    u ppm,

    u ppm,

    u ppm,

    u ppm,

    u

    u Vol%H2O

    u

    u

    u

    u

    u

    u

    u

    u

    Please place checkmarks in the appropriate boxes and fill in the necessary information in the blanks.

    Customer

    Destination of delivery

    Plant name

    Measurement points

    2. Process conditions

    2.1 Measurement gas components

    2.2 Oxygen concentration

    Moisture contents

    2.3 Temperature

    2.4 Pressure

    2.5 Gas flow

    2.6 Dust type, Size

    2.7 Corrosive gas

    2.8 Combustible gas

    2.9 Others

    3. Installation site conditions

    3.1 Ambient temperature

    3.2 Vibration

    3.3 1 Probe installation location

    3.4 Instrument air supply

    3.5 Converter location

    3.6 Cable length between probe and converter

    3.7 Calibration method

    4. Quotation data

    Probe Model ZR22G General-use Probe Refer to the Probe Configuration for probe selection.Model ZO21P-H High temperature Use Probe Adapter

    Model ZR402G Separate type Analyzer, Converter

    Model E7046EC/E7046EN Auxiliary Ejector for high temperature use.

    Model ZH21B Dust Protector for high temperature humidity analyzer.

    Model ZO21R Probe Protector for Oxgen Analyzer

    1. General information

    Nor. Min. Max.

    Nor.

    Nor.

    Nor.

    Nor.

    Nor.

    Min.

    Min.

    Min.

    Min.

    Min.

    Max.

    Max.

    Max.

    Max.

    u No gas u Gas , quantity

    mm quantity

    u No gas u Gas , quantity

    , quantity

    , quantity

    1. Around Probe temp. from to 8C, 2. Around Converter temp. from to 8C

    u No vibration u Vibration

    u Furnace u Stack u Others

    2 Probe position u Horizontal u Vertical u Others

    u Indoor u Outdoor u Covered

    3 Probe insertion length (m) u 0.4, u 0.7, u 1.0, u 1.5, u 2.0, u 2.5, u 3.0, u Others

    4 Flange u JIS u ANSI u Others

    u Cannot be used. u Can be used. kPa

    u Indoor u Outdoor u Covered (under roof)

    meters

    u Manual u Automatic

    Quotation Qantity Description

    Options (for general use)

    Model K9471UA Filter

    Model ZR202G Integrated type Zirconia Oxygen / High temperature Humidity Analyzer

    Model ZO21S Standard Gas Unit Select any one of Model ZO21S, ZA8F, ZR40H .

    Model ZA8F Flow Setting Unit

    Model ZR40H Automatic Calibration Unit Not required if probe options are specified.

    L9852CB/K9471UN Stop Valve

    K9292DN/K9292DS Check Valve

    G7011XF/E7040EL Air Set

    G7001ZC Zero Gas Cylinder

    G7013XF/G7014XF Pressure Regulator

    Type of analyzer : u Oxygen Analyzer u High temperature Humidity Analyzer

    u Sparate type u Integrated type

    Object : u indication u record u control u alarm

    Fuel : u gas u oil u coal u

    Power requirements V AC Hz

  • GS 11M12A01-E-HSubject to change without notice Printed in The Netherlands, 02-103 (A) QCopyright

    EUROPEAN HEADQUARTERSYokogawa Europe B.V.Databankweg 203821 AL AMERSFOORTThe NetherlandsTel. +31-33-4641 611Fax +31-33-4641 610E-mail: [email protected]

    THE NETHERLANDSYokogawa Nederland B.V.Hoofdveste 113992 DH HOUTENTel. +31-30-635 77 77 Fax +31-30-635 77 70

    AUSTRIAYokogawa Austria Ges.m.b.H.Franzensbrckenstrasse 26A-1021 WIENTel. +43-1-2165 043 0Fax +43-1-2165 043 33

    AUSTRALIAYokogawa Australia Pty Ltd.Private mail bag 24Centre Court D325-27 Paul Street NorthNORTH RYDE, N.S.W. 2113Tel. +61-2-805 0699Fax +61-2-888 1844

    SINGAPOREYokogawa Engineering Asia Pte. Ltd.11, Tampines Street 92SINGAPORE, 528872Tel. +65-783 9537Fax +65-786 2606

    JAPANYokogawa Electric Corporation2-9-32, Nakacho, Musashino-shiTOKYO, 180Tel. +81-422 52 5617Fax +81-422 52 0622

    SOUTH AFRICAYokogawa South Africa (Pty) ltd.67 Port Road, RobertshamSouthdale 2135, JOHANNESBURGTel. +27-11-680-5420Fax +27-11-680-2922

    BELGIUMYokogawa Belgium N.V./S.A.Minervastraat 161930 ZAVENTEMTel. +32-2-719 55 11Fax +32-2-725 34 99

    FRANCEYokogawa France S.A. Vlizy Valley18-20 Rue Grange Dame Rose78140 VELIZY VILLACOUBLAYTel. +33-1-39 26 10 00Fax +33-1-39 26 10 30

    GERMANYYokogawa Deutschland GmbHBerliner Strasse 101-103D-40880 RATINGENTel. +49-2102-4983 0Fax +49-2102-4983 22

    HUNGARYYokogawa Hungaria Ltd.Alkotas Center 39 C1123 BUDAPESTTel. +36-1-355 3938Fax +36-1-355 3897

    ITALYYokogawa Italia S.r.l.Vicolo D. Pantaleoni, 420161 MILANOTel. +39-02-66 24 11Fax +39-02-645 57 02

    SPAINYokogawa Espaa S.A.C/Francisco Remiro, N2, Edif. H28028 MADRIDTel. +34-91-724 20 80Fax +34-91-355 31 40

    UNITED KINGDOMYokogawa United Kingdom Ltd.Stuart Road, Manor Park,RUNCORNCheshire WA7 1TRTel. +44-1-928 597100Fax +44-1-928 597101

    UNITED STATES OF AMERICAYokogawa Corporation ofAmerica2 Dart RoadNEWNAN, GA 30265-1040Tel. +1-770-253 70 00Fax +1-770-251 28 00

    CENTRAL/EAST REGIONVia Yokogawa Austria: Czechia, Slovakia, Poland,Croatia, Slovenia, Jugoslavia,Bulgaria, Romania, Macedonia,Bosnia & Herzegovina

    Distributors in:Denmark, Finland, Greece,Norway, Portugal, RussianFederation, Sweden, Switzerlandand Turkey.

    Block 02, 10-00

    ISO 9001

    CERTIFICATEDFIRM