greg miller harmonization norcal aacc 041812

Upload: fbaxy

Post on 05-Oct-2015

8 views

Category:

Documents


0 download

DESCRIPTION

Greg Miller Harmonization NorCal AACC

TRANSCRIPT

  • From chaos to order: harmonization of clinical laboratory test results

    Greg Miller, PhD Virginia Commonwealth University

    Richmond, VA, USA

    H rm nzati on ia o .net

  • Outline

    Why do we need harmonized results

    How to achieve results traceable to a reference system

    Barriers to harmonized results

    AACCs global harmonization initiative

  • Primary reasons for testing

    To identify and manage diseases in an individual patient

    To develop data from which to establish clinical practice guidelines for managing diseases

  • Good laboratory medicine requires:

    Total error of measurement is small enough that a result reflects a patients biological condition

    Comparable results that are independent of

    where and when a test was performed

    the measurement procedure used

  • Total Error

    Calibration bias to an accepted reference (includes calibrator lot-to-lot variability)

    Imprecision in a measurement procedure

    Sample specific influences

    non-specificity for the measurand

    interfering substances

  • Why do we need comparable results

    Clinical practice guidelines are established for interpretation of lab test results

    If different measurements give different results for the same patient sample:

    The interpretive guidelines become less useful

    Patients may receive incorrect treatment

  • How to achieve comparable results

    Calibration of all measurement procedures is traceable to a common reference system

    Consistent performance is monitored and maintained by surveillance using PT or EQA

  • ISO 17511

    In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials

  • Category Reference

    measurement procedure

    Primary (pure substance) reference material

    Secondary (value

    assigned) reference material

    Examples

    1 Yes Yes Possible Electrolytes,

    glucose, cortisol

    2 Yes No Possible Enzymes

    4 No No Yes Proteins,

    tumor markers, HIV

    5 No No No Proteins, EBV, VZV

    3 Yes No No Hemostatic factors

    S

    t

    a

    n

    d

    a

    r

    d

    i

    z

    a

    t

    i

    o

    n

    Harmonization

    Traceability categories from ISO 17511

  • Terminology

    Standardization: results are uniform among measurement procedures

    traceability is established to SI and a reference measurement procedure

    Harmonization: results are uniform among measurement procedures

    NO reference measurement procedure and no pure substance reference material exists

  • Primary Reference Material

    (pure substance)

    SI unit

    Traceability (based on ISO 17511)A reference system (ideal)

    Primary Reference Measurement Procedure

    (e.g. gravimetry)Primary Calibrator

    Secondary Reference Material

    (matrix)

    Secondary Reference Measurement Procedure

    (e.g. IDMS)

  • Primary Reference Material

    (NIST SRM 917b crystalline glucose)

    SI unit (glucose, mmol/L)

    Traceability (based on ISO 17511)A reference system for glucose

    Primary Reference Measurement Procedure(gravimetry, calibrated with

    NIST mass standards)Primary Calibrator(glucose in water, 1, 3, 6, 11 mmol/L)

    Secondary Reference Material(NIST SRM 965b glucose in frozen human serum)

    Secondary Reference Measurement Procedure

    (IDMS)

  • Reference Procedure

    Primary Reference Material

    (pure substance)

    (e.g. IDMS)

    SI unit

    Traceability (based on ISO 17511)

    Routine Procedure

    Patient sample result

    T

    R

    A

    C

    E

    A

    B

    I

    L

    I

    T

    Y

    P

    a

    t

    i

    e

    n

    t

    s

    a

    m

    p

    l

    e

    r

    e

    s

    u

    l

    t

    s

    a

    r

    e

    e

    q

    u

    i

    v

    a

    l

    e

    n

    t

    t

    o

    t

    h

    e

    r

    e

    f

    e

    r

    e

    n

    c

    e

    p

    r

    o

    c

    e

    d

    u

    r

    e

    r

    e

    s

    u

    l

    t

    s

    Secondary Reference Material

    (matrix)Mfr Selected Procedure

    Mfr Product Calibrator

    Mfr Working Calibrator

    Mfr Standing Procedure

    (calibrator)

  • Reference Procedure

    Primary Reference Material

    (pure substance)

    (e.g. IDMS)

    SI unit

    Traceability (based on ISO 17511)

    Routine Procedure

    Patient sample result

    T

    R

    A

    C

    E

    A

    B

    I

    L

    I

    T

    Y

    P

    a

    t

    i

    e

    n

    t

    s

    a

    m

    p

    l

    e

    r

    e

    s

    u

    l

    t

    s

    a

    r

    e

    e

    q

    u

    i

    v

    a

    l

    e

    n

    t

    t

    o

    t

    h

    e

    r

    e

    f

    e

    r

    e

    n

    c

    e

    p

    r

    o

    c

    e

    d

    u

    r

    e

    r

    e

    s

    u

    l

    t

    s

    Mfr Selected Procedure

    Mfr Product Calibrator

    Mfr Working Calibrator

    Mfr Standing Procedure

    (calibrator)Panel of patient

    samples

  • Traceability requires commutable calibration

    materials

    Commutable means that values measured for a calibration material and for native clinical samples have the same relationship between two, or more, measurement procedures for the same measurand.

  • 02

    4

    6

    8

    10

    0 2 4 6 8 10Measurement Procedure 1

    M

    e

    a

    s

    u

    r

    e

    m

    e

    n

    t

    P

    r

    o

    c

    e

    d

    u

    r

    e

    2

    Clinical SamplesReference Materials

    Commutable: same relationship for clinical samples and reference materials

  • 02

    4

    6

    8

    10

    0 2 4 6 8 10Measurement Procedure 1

    M

    e

    a

    s

    u

    r

    e

    m

    e

    n

    t

    P

    r

    o

    c

    e

    d

    u

    r

    e

    2

    Clinical SamplesReference Materials

    Non-commutable: different relationship for clinical samples and reference materials

  • Use of a non-commutable material for calibration traceability will cause:

    Incorrect value assignment for a routine (field) measurement procedure calibrator

    Incorrect results for patient samples

    Miller, Myers, Rej. Why commutability matters. Clin Chem 2006; 52: 553-4 .

  • cholesterolcreatinine

    glucose

    uricacidhomocysteine

    urea ALTAST HbA1c

    HCG

    TSH PSA

    Troponin

    I

    EpsteinBarrvirus

    Picking the low-hanging fruit !omics

  • What happens when there is no reference measurement procedure

  • Mfr Selected Procedure

    Mfr Product Calibrator

    Mfr Working Calibrator

    Mfr Standing Procedure

    Traceability (based on ISO 17511)

    Routine Procedure

    Patient sample result

    T

    R

    A

    C

    E

    A

    B

    I

    L

    I

    T

    Y

    P

    a

    t

    i

    e

    n

    t

    s

    a

    m

    p

    l

    e

    r

    e

    s

    u

    l

    t

    s

    a

    r

    e

    t

    r

    a

    c

    e

    a

    b

    l

    e

    t

    o

    a

    r

    e

    f

    e

    r

    e

    n

    c

    e

    m

    a

    t

    e

    r

    i

    a

    l

    Secondary Reference Material

    (matrix)

    Value assignment Commutability

  • Value assignment when there is no reference measurement procedure

    International conventional reference material

    Arbitrary e.g. U/L

    Bioassay for hormone activity

    An arbitrary designated comparison procedure

  • Traceable to an international conventional reference material

    The true value is not known

    The goal of harmonization is to have results that are comparable irrespective of the measurement procedure used

    Clinical guidelines can still be implemented

  • Examples: traceable to a reference material

    Human chorionic gonadotropin

    Prostate-specific antigen

    Thyroid stimulating hormone

    Human immunodeficiency virus

    (no reference measurement procedure)

  • What happens when there is both:

    no reference measurement procedure

    no reference material

  • Mfr Product Calibrator

    Mfr Working Calibrator

    Mfr Standing Procedure

    Traceability (based on ISO 17511)

    Routine Procedure

    Patient sample result

    T

    R

    A

    C

    E

    A

    B

    I

    L

    I

    T

    Y

    P

    a

    t

    i

    e

    n

    t

    s

    a

    m

    p

    l

    e

    r

    e

    s

    u

    l

    t

    s

    a

    r

    e

    n

    o

    t

    t

    r

    a

    c

    e

    a

    b

    l

    e

    t

    o

    a

    n

    y

    i

    n

    t

    e

    r

    n

    a

    t

    i

    o

    n

    a

    l

    r

    e

    f

    e

    r

    e

    n

    c

    e

    There is no coordination among manufacturers

    Reference interval concept is used to interpret results

  • Examples: traceable to a manufacturers working calibrator

    B-type natriuretic peptide

    Epstein-Barr virus

    Varicella zoster virus

    (no reference material nor reference measurement procedure)

  • ISO 17511 category 5 needs practical procedures to achieve harmonization

    Possibilities:

    traceable to an all methods mean (outliers removed) of a panel of patient samples?

    traceable to a designated measurement procedure (arbitrary, but with good correlation with clinical outcome)?

  • Practical Considerations

  • Panel of patient

    samples

    Reference Procedure

    Primary Ref Material

    (pure substance)(e.g. IDMS)

    SI unit

    Traceability (an application)

    Mfr Product Calibrator

    Routine Procedure

    Results equivalent to those from

    RMP

    Value is assigned to produce traceable results for patient samples

  • Panel of patient

    samples

    Reference Procedure

    Primary Ref Material

    (pure substance)(e.g. IDMS)

    SI unit

    Traceability (an application)

    Mfr Product Calibrator

    Routine Procedure

    Results equivalent to those from

    RMP

    May be non-commutable, but has a procedure-specific factor to correct

    non-commutability bias

  • 02

    4

    6

    8

    10

    0 2 4 6 8 10Ma

    n

    u

    f

    a

    c

    t

    u

    r

    e

    r

    s

    r

    o

    u

    t

    i

    n

    e

    p

    r

    o

    c

    e

    d

    u

    r

    e

    Reference procedure

    Clinical SamplesProduct Calibrator

    The known matrix-related bias can be offset in the procedure-specific value assigned to product calibrators

    so that the results for patients match those from the RMP

  • A manufacturer's product calibrator is intended for use with a specific measurement procedure

    It cannot be used with a different manufacturers measurement procedure

    Important

  • Secondary Reference Material

    (calibrator)

    Procedure nProcedure 3

    Procedure 2Procedure 1

    Patient Samples

    Traceability to a Reference Material

    Reference Material must be commutable with patient samples for all measurement procedures with which it will be used

    Results nResults 3

    Results 2Results 1

  • Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures

    The Problem

  • Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures

    Historically, commutability of reference materials was frequently not validated for use with routine clinical laboratory measurement procedures

    The Problem

  • Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures

    A manufacturers standing procedure is frequently the same as the clinical laboratory procedure but calibrated with a master lot of calibrator that may be traceable to a non- commutable reference material

    The Problem

  • Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures

    A manufacturers standing procedure is frequently the same as the clinical laboratory procedure but calibrated with a master lot of calibrator that may be traceable to a non- commutable reference material

    The Problem

    A non-commutable calibration material

    breaks the traceability chain

  • Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures

    Even though manufacturers claim traceability, the process fails to provide equivalent results for patient samples when different measurement procedures are used

    The Problem

  • Secondary reference materials may not be commutable

    A reference material may not have been validated to be suitable as a calibrator for all routine measurement procedures

    Limitation may apply to:

    National metrology institutes (e.g. NIST, IRMM)

    Other providers of standards and calibrators such as: WHO, professional organizations, commercial providers

  • TSH methods All traceable to IS 94/674 (WHO)

    Thienpont et al. Clin Chem 2010; 56: 902-911.

    Mean 95% CI for 40 patient samples

  • Calibration traceability does not ensure accuracy for an individual patient sample

    Measurement procedure may not be specific for the measurand

    Measurand may not be well defined

    Molecular form(s) of clinical interest

    Interfering substances present in a patients sample may influence the result

  • Other examples:

    Follicle stimulating hormone (Clin Chim Acta 1998;273:103-17)

    Prostate-specific antigen ( Clin Chem 2006;52:59-64)

    C-peptide (Clin Chem 2008;54:1023-6)

    Insulin (Clin Chem 2009;1011-1018, 2009)

    Human chorionic gonadotropin (Clin Chem 2009;55:1484-91)

    Cytomegalovirus (Clin Chem 2009;55:1701-10)

    Troponin I (Pathology 2010;42:402-8)

    Prostate-specific antigen (Clin Chem 2011;57:1776-7)

  • What about new biomarkers

    A new biomarker is shown to be clinically useful

    A measurement procedure is commercialized

    A 2nd procedure is commercialized

    If the 2nd and subsequent procedures do not give comparable results, the clinical trial interpretive guidelines become less useful

  • Sustainable calibration traceability needs to be part of the development process:

    Definition of the measurand

    Requirements for measurement specificity

    Reference measurement procedure

    Commutable reference materials

    New biomarkers

  • What do we do?What do we do?

  • Must change practice to require commutability validation for reference materials intended for use with:

    Manufacturers standing procedures

    Routine clinical laboratory procedures

    A guideline is available: CLSI C53-A Characterization and Qualification of Commutable Reference Materials for

    Laboratory Medicine (2010)

  • Commutability: who is responsible

    Reference material manufacturer

    Cannot know all procedures in use

    Should make a material likely to be commutable

    Can address commonly used procedures

    Measurement procedure manufacturer

    Must confirm commutability for an intended use

  • Progress has been made!

    However,

    The approach has been ad-hoc based on individual interests

    There is no central organizing body to coordinate the work of different organizations

    A number of organizations are addressing standardization / harmonization around the world

    IFCC has been a key leader in developing science

    Many national societies are engaged

  • JCTLM

    Credentialing function; maintains lists of reviewed and approved:

    Reference measurement procedures

    Reference laboratories

    Reference materials Commutability information is not required for listing

    Joint Committee for Traceability in Laboratory Medicine

    International Committee of Weights and Measures International Federation for Clinical Chemistry and Laboratory Medicine International Laboratory Accreditation Cooperation

  • How is calibration traceability sustained

    Certification programs

    e.g. CDC Lipids, NGSP HbA1c

    Surveillance by PT / EQA

    Requires commutable materials

    Conformance to ISO 17511

    e.g. CE mark in the EU

  • Barriers to harmonization

    Lack of a systematic process to identify and prioritize measurands

    Materials are labeled as reference materials that have not been validated to be commutable for the intended measurement procedures

    Inadequate definition of the measurand

    Inadequate analytical specificity for the measurand

    Lack of systematic procedures to implement harmonization, in particular:

    when there is no reference measurement procedure

    when there is no reference material

  • Clin Chem Lab Med 2008;46:1220-2.

    Challenged clinical laboratories to address the issues associated with harmonization

    have the courage to agree on pragmatic solutions

  • Prince Harmonization finds the Sleeping Beauty

  • Report from an AACC conference, October, 2010: Improving Clinical Laboratory Testing through

    Harmonization: An International Forum

    August 2011

  • The Roadmap

    Develop an infrastructure to coordinate harmonization activities world wide to include:

    1. Prioritization of analytes

    2. Gap analysis for what needs to be done

    3. Technical processes to achieve harmonization

    4. Surveillance of success of harmonization

  • Cooperate

    With other organizations already working to improve standardization / harmonization

    Provide a communications portal and prioritization scheme among organizations and standardization / harmonization activities

  • Focus on measurands for which no reference measurement

    procedure exists

    Measurands in ISO 17511 categories 4 and 5 have been technically more difficult to address, thus there have been few effective procedures implemented for harmonization in these categories.

  • AN INFRASTRUCTURE FOR HARMONIZATION

    Harmonization Oversight Group

    Evaluate measurand proposals

    Clinical practice groupsLaboratory practice groupsIVD manufacturers

    Metrology institutesStandards organizationsRegulatory organizations

    Solicit champion and funding Clinically affected entity Economically affected entity

    Create a Harmonization Implementation Group Technical plan Surveillance plan Implement the plans Achieve JCTLM listing

    If measurand is not being addressed, create aSpecialty Work Group

    Review priority and feasibility using a checklist

    Recommendation to Harmonization Oversight Group

    , If work underway, refer

    to that group If RMP possible, refer

    to another group

    Public health orgs

  • A Steering Committee is responsible to implement the infrastructure for harmonization

    3 Task Forces are working to develop operating procedures for the major components

    1. Establishing the harmonization oversight group

    2. Developing checklists for prioritization and gap analysis

    3. Developing a tool box of technical processes for assessment and harmonization of measurands

    Path Forward

  • Guiding Principles

    The goal is harmonized results that are fit for the intended clinical purpose. Pragmatic approaches will be required.

    An open and transparent process is required.

    Success will depend on collaboration among all stakeholders world wide.

  • www.harmonization.net

    Communication center for the project

    Status reports on measurands

    Communication with stakeholders

    Will become a general information portal for global standardization / harmonization activities

    Useful technical information

    Information on global activities

    Links to other organizations

  • H rm nzati on ia o .net

    Coming soon: International Program for Harmonization of Clinical Laboratory Results

    From chaos to order: harmonization of clinical laboratory test results Greg Miller, PhDVirginia Commonwealth UniversityRichmond, VA, USAOutlinePrimary reasons for testing Slide Number 4Slide Number 5Slide Number 6Slide Number 7Slide Number 8Slide Number 9Slide Number 10Slide Number 11Slide Number 12Slide Number 13Slide Number 14Slide Number 15Slide Number 16Slide Number 17Slide Number 18Slide Number 19Slide Number 20Slide Number 21Slide Number 22Slide Number 23Slide Number 24Slide Number 25Slide Number 26Slide Number 27Slide Number 28Slide Number 29Slide Number 30Slide Number 31Slide Number 32Slide Number 33Slide Number 34Slide Number 35Slide Number 36Slide Number 37Slide Number 38Slide Number 39Slide Number 40Slide Number 41Slide Number 42Slide Number 43Slide Number 44Slide Number 45Slide Number 46Slide Number 47Slide Number 48Slide Number 49Slide Number 50Slide Number 51Slide Number 52Slide Number 53Slide Number 54Slide Number 55Slide Number 56Slide Number 57Slide Number 58Slide Number 59Slide Number 60Slide Number 61Slide Number 62Slide Number 63