greg miller harmonization norcal aacc 041812
DESCRIPTION
Greg Miller Harmonization NorCal AACCTRANSCRIPT
-
From chaos to order: harmonization of clinical laboratory test results
Greg Miller, PhD Virginia Commonwealth University
Richmond, VA, USA
H rm nzati on ia o .net
-
Outline
Why do we need harmonized results
How to achieve results traceable to a reference system
Barriers to harmonized results
AACCs global harmonization initiative
-
Primary reasons for testing
To identify and manage diseases in an individual patient
To develop data from which to establish clinical practice guidelines for managing diseases
-
Good laboratory medicine requires:
Total error of measurement is small enough that a result reflects a patients biological condition
Comparable results that are independent of
where and when a test was performed
the measurement procedure used
-
Total Error
Calibration bias to an accepted reference (includes calibrator lot-to-lot variability)
Imprecision in a measurement procedure
Sample specific influences
non-specificity for the measurand
interfering substances
-
Why do we need comparable results
Clinical practice guidelines are established for interpretation of lab test results
If different measurements give different results for the same patient sample:
The interpretive guidelines become less useful
Patients may receive incorrect treatment
-
How to achieve comparable results
Calibration of all measurement procedures is traceable to a common reference system
Consistent performance is monitored and maintained by surveillance using PT or EQA
-
ISO 17511
In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
-
Category Reference
measurement procedure
Primary (pure substance) reference material
Secondary (value
assigned) reference material
Examples
1 Yes Yes Possible Electrolytes,
glucose, cortisol
2 Yes No Possible Enzymes
4 No No Yes Proteins,
tumor markers, HIV
5 No No No Proteins, EBV, VZV
3 Yes No No Hemostatic factors
S
t
a
n
d
a
r
d
i
z
a
t
i
o
n
Harmonization
Traceability categories from ISO 17511
-
Terminology
Standardization: results are uniform among measurement procedures
traceability is established to SI and a reference measurement procedure
Harmonization: results are uniform among measurement procedures
NO reference measurement procedure and no pure substance reference material exists
-
Primary Reference Material
(pure substance)
SI unit
Traceability (based on ISO 17511)A reference system (ideal)
Primary Reference Measurement Procedure
(e.g. gravimetry)Primary Calibrator
Secondary Reference Material
(matrix)
Secondary Reference Measurement Procedure
(e.g. IDMS)
-
Primary Reference Material
(NIST SRM 917b crystalline glucose)
SI unit (glucose, mmol/L)
Traceability (based on ISO 17511)A reference system for glucose
Primary Reference Measurement Procedure(gravimetry, calibrated with
NIST mass standards)Primary Calibrator(glucose in water, 1, 3, 6, 11 mmol/L)
Secondary Reference Material(NIST SRM 965b glucose in frozen human serum)
Secondary Reference Measurement Procedure
(IDMS)
-
Reference Procedure
Primary Reference Material
(pure substance)
(e.g. IDMS)
SI unit
Traceability (based on ISO 17511)
Routine Procedure
Patient sample result
T
R
A
C
E
A
B
I
L
I
T
Y
P
a
t
i
e
n
t
s
a
m
p
l
e
r
e
s
u
l
t
s
a
r
e
e
q
u
i
v
a
l
e
n
t
t
o
t
h
e
r
e
f
e
r
e
n
c
e
p
r
o
c
e
d
u
r
e
r
e
s
u
l
t
s
Secondary Reference Material
(matrix)Mfr Selected Procedure
Mfr Product Calibrator
Mfr Working Calibrator
Mfr Standing Procedure
(calibrator)
-
Reference Procedure
Primary Reference Material
(pure substance)
(e.g. IDMS)
SI unit
Traceability (based on ISO 17511)
Routine Procedure
Patient sample result
T
R
A
C
E
A
B
I
L
I
T
Y
P
a
t
i
e
n
t
s
a
m
p
l
e
r
e
s
u
l
t
s
a
r
e
e
q
u
i
v
a
l
e
n
t
t
o
t
h
e
r
e
f
e
r
e
n
c
e
p
r
o
c
e
d
u
r
e
r
e
s
u
l
t
s
Mfr Selected Procedure
Mfr Product Calibrator
Mfr Working Calibrator
Mfr Standing Procedure
(calibrator)Panel of patient
samples
-
Traceability requires commutable calibration
materials
Commutable means that values measured for a calibration material and for native clinical samples have the same relationship between two, or more, measurement procedures for the same measurand.
-
02
4
6
8
10
0 2 4 6 8 10Measurement Procedure 1
M
e
a
s
u
r
e
m
e
n
t
P
r
o
c
e
d
u
r
e
2
Clinical SamplesReference Materials
Commutable: same relationship for clinical samples and reference materials
-
02
4
6
8
10
0 2 4 6 8 10Measurement Procedure 1
M
e
a
s
u
r
e
m
e
n
t
P
r
o
c
e
d
u
r
e
2
Clinical SamplesReference Materials
Non-commutable: different relationship for clinical samples and reference materials
-
Use of a non-commutable material for calibration traceability will cause:
Incorrect value assignment for a routine (field) measurement procedure calibrator
Incorrect results for patient samples
Miller, Myers, Rej. Why commutability matters. Clin Chem 2006; 52: 553-4 .
-
cholesterolcreatinine
glucose
uricacidhomocysteine
urea ALTAST HbA1c
HCG
TSH PSA
Troponin
I
EpsteinBarrvirus
Picking the low-hanging fruit !omics
-
What happens when there is no reference measurement procedure
-
Mfr Selected Procedure
Mfr Product Calibrator
Mfr Working Calibrator
Mfr Standing Procedure
Traceability (based on ISO 17511)
Routine Procedure
Patient sample result
T
R
A
C
E
A
B
I
L
I
T
Y
P
a
t
i
e
n
t
s
a
m
p
l
e
r
e
s
u
l
t
s
a
r
e
t
r
a
c
e
a
b
l
e
t
o
a
r
e
f
e
r
e
n
c
e
m
a
t
e
r
i
a
l
Secondary Reference Material
(matrix)
Value assignment Commutability
-
Value assignment when there is no reference measurement procedure
International conventional reference material
Arbitrary e.g. U/L
Bioassay for hormone activity
An arbitrary designated comparison procedure
-
Traceable to an international conventional reference material
The true value is not known
The goal of harmonization is to have results that are comparable irrespective of the measurement procedure used
Clinical guidelines can still be implemented
-
Examples: traceable to a reference material
Human chorionic gonadotropin
Prostate-specific antigen
Thyroid stimulating hormone
Human immunodeficiency virus
(no reference measurement procedure)
-
What happens when there is both:
no reference measurement procedure
no reference material
-
Mfr Product Calibrator
Mfr Working Calibrator
Mfr Standing Procedure
Traceability (based on ISO 17511)
Routine Procedure
Patient sample result
T
R
A
C
E
A
B
I
L
I
T
Y
P
a
t
i
e
n
t
s
a
m
p
l
e
r
e
s
u
l
t
s
a
r
e
n
o
t
t
r
a
c
e
a
b
l
e
t
o
a
n
y
i
n
t
e
r
n
a
t
i
o
n
a
l
r
e
f
e
r
e
n
c
e
There is no coordination among manufacturers
Reference interval concept is used to interpret results
-
Examples: traceable to a manufacturers working calibrator
B-type natriuretic peptide
Epstein-Barr virus
Varicella zoster virus
(no reference material nor reference measurement procedure)
-
ISO 17511 category 5 needs practical procedures to achieve harmonization
Possibilities:
traceable to an all methods mean (outliers removed) of a panel of patient samples?
traceable to a designated measurement procedure (arbitrary, but with good correlation with clinical outcome)?
-
Practical Considerations
-
Panel of patient
samples
Reference Procedure
Primary Ref Material
(pure substance)(e.g. IDMS)
SI unit
Traceability (an application)
Mfr Product Calibrator
Routine Procedure
Results equivalent to those from
RMP
Value is assigned to produce traceable results for patient samples
-
Panel of patient
samples
Reference Procedure
Primary Ref Material
(pure substance)(e.g. IDMS)
SI unit
Traceability (an application)
Mfr Product Calibrator
Routine Procedure
Results equivalent to those from
RMP
May be non-commutable, but has a procedure-specific factor to correct
non-commutability bias
-
02
4
6
8
10
0 2 4 6 8 10Ma
n
u
f
a
c
t
u
r
e
r
s
r
o
u
t
i
n
e
p
r
o
c
e
d
u
r
e
Reference procedure
Clinical SamplesProduct Calibrator
The known matrix-related bias can be offset in the procedure-specific value assigned to product calibrators
so that the results for patients match those from the RMP
-
A manufacturer's product calibrator is intended for use with a specific measurement procedure
It cannot be used with a different manufacturers measurement procedure
Important
-
Secondary Reference Material
(calibrator)
Procedure nProcedure 3
Procedure 2Procedure 1
Patient Samples
Traceability to a Reference Material
Reference Material must be commutable with patient samples for all measurement procedures with which it will be used
Results nResults 3
Results 2Results 1
-
Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures
The Problem
-
Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures
Historically, commutability of reference materials was frequently not validated for use with routine clinical laboratory measurement procedures
The Problem
-
Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures
A manufacturers standing procedure is frequently the same as the clinical laboratory procedure but calibrated with a master lot of calibrator that may be traceable to a non- commutable reference material
The Problem
-
Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures
A manufacturers standing procedure is frequently the same as the clinical laboratory procedure but calibrated with a master lot of calibrator that may be traceable to a non- commutable reference material
The Problem
A non-commutable calibration material
breaks the traceability chain
-
Many secondary reference materials are not commutable with native clinical samples for routine clinical laboratory procedures
Even though manufacturers claim traceability, the process fails to provide equivalent results for patient samples when different measurement procedures are used
The Problem
-
Secondary reference materials may not be commutable
A reference material may not have been validated to be suitable as a calibrator for all routine measurement procedures
Limitation may apply to:
National metrology institutes (e.g. NIST, IRMM)
Other providers of standards and calibrators such as: WHO, professional organizations, commercial providers
-
TSH methods All traceable to IS 94/674 (WHO)
Thienpont et al. Clin Chem 2010; 56: 902-911.
Mean 95% CI for 40 patient samples
-
Calibration traceability does not ensure accuracy for an individual patient sample
Measurement procedure may not be specific for the measurand
Measurand may not be well defined
Molecular form(s) of clinical interest
Interfering substances present in a patients sample may influence the result
-
Other examples:
Follicle stimulating hormone (Clin Chim Acta 1998;273:103-17)
Prostate-specific antigen ( Clin Chem 2006;52:59-64)
C-peptide (Clin Chem 2008;54:1023-6)
Insulin (Clin Chem 2009;1011-1018, 2009)
Human chorionic gonadotropin (Clin Chem 2009;55:1484-91)
Cytomegalovirus (Clin Chem 2009;55:1701-10)
Troponin I (Pathology 2010;42:402-8)
Prostate-specific antigen (Clin Chem 2011;57:1776-7)
-
What about new biomarkers
A new biomarker is shown to be clinically useful
A measurement procedure is commercialized
A 2nd procedure is commercialized
If the 2nd and subsequent procedures do not give comparable results, the clinical trial interpretive guidelines become less useful
-
Sustainable calibration traceability needs to be part of the development process:
Definition of the measurand
Requirements for measurement specificity
Reference measurement procedure
Commutable reference materials
New biomarkers
-
What do we do?What do we do?
-
Must change practice to require commutability validation for reference materials intended for use with:
Manufacturers standing procedures
Routine clinical laboratory procedures
A guideline is available: CLSI C53-A Characterization and Qualification of Commutable Reference Materials for
Laboratory Medicine (2010)
-
Commutability: who is responsible
Reference material manufacturer
Cannot know all procedures in use
Should make a material likely to be commutable
Can address commonly used procedures
Measurement procedure manufacturer
Must confirm commutability for an intended use
-
Progress has been made!
However,
The approach has been ad-hoc based on individual interests
There is no central organizing body to coordinate the work of different organizations
A number of organizations are addressing standardization / harmonization around the world
IFCC has been a key leader in developing science
Many national societies are engaged
-
JCTLM
Credentialing function; maintains lists of reviewed and approved:
Reference measurement procedures
Reference laboratories
Reference materials Commutability information is not required for listing
Joint Committee for Traceability in Laboratory Medicine
International Committee of Weights and Measures International Federation for Clinical Chemistry and Laboratory Medicine International Laboratory Accreditation Cooperation
-
How is calibration traceability sustained
Certification programs
e.g. CDC Lipids, NGSP HbA1c
Surveillance by PT / EQA
Requires commutable materials
Conformance to ISO 17511
e.g. CE mark in the EU
-
Barriers to harmonization
Lack of a systematic process to identify and prioritize measurands
Materials are labeled as reference materials that have not been validated to be commutable for the intended measurement procedures
Inadequate definition of the measurand
Inadequate analytical specificity for the measurand
Lack of systematic procedures to implement harmonization, in particular:
when there is no reference measurement procedure
when there is no reference material
-
Clin Chem Lab Med 2008;46:1220-2.
Challenged clinical laboratories to address the issues associated with harmonization
have the courage to agree on pragmatic solutions
-
Prince Harmonization finds the Sleeping Beauty
-
Report from an AACC conference, October, 2010: Improving Clinical Laboratory Testing through
Harmonization: An International Forum
August 2011
-
The Roadmap
Develop an infrastructure to coordinate harmonization activities world wide to include:
1. Prioritization of analytes
2. Gap analysis for what needs to be done
3. Technical processes to achieve harmonization
4. Surveillance of success of harmonization
-
Cooperate
With other organizations already working to improve standardization / harmonization
Provide a communications portal and prioritization scheme among organizations and standardization / harmonization activities
-
Focus on measurands for which no reference measurement
procedure exists
Measurands in ISO 17511 categories 4 and 5 have been technically more difficult to address, thus there have been few effective procedures implemented for harmonization in these categories.
-
AN INFRASTRUCTURE FOR HARMONIZATION
Harmonization Oversight Group
Evaluate measurand proposals
Clinical practice groupsLaboratory practice groupsIVD manufacturers
Metrology institutesStandards organizationsRegulatory organizations
Solicit champion and funding Clinically affected entity Economically affected entity
Create a Harmonization Implementation Group Technical plan Surveillance plan Implement the plans Achieve JCTLM listing
If measurand is not being addressed, create aSpecialty Work Group
Review priority and feasibility using a checklist
Recommendation to Harmonization Oversight Group
, If work underway, refer
to that group If RMP possible, refer
to another group
Public health orgs
-
A Steering Committee is responsible to implement the infrastructure for harmonization
3 Task Forces are working to develop operating procedures for the major components
1. Establishing the harmonization oversight group
2. Developing checklists for prioritization and gap analysis
3. Developing a tool box of technical processes for assessment and harmonization of measurands
Path Forward
-
Guiding Principles
The goal is harmonized results that are fit for the intended clinical purpose. Pragmatic approaches will be required.
An open and transparent process is required.
Success will depend on collaboration among all stakeholders world wide.
-
www.harmonization.net
Communication center for the project
Status reports on measurands
Communication with stakeholders
Will become a general information portal for global standardization / harmonization activities
Useful technical information
Information on global activities
Links to other organizations
-
H rm nzati on ia o .net
Coming soon: International Program for Harmonization of Clinical Laboratory Results
From chaos to order: harmonization of clinical laboratory test results Greg Miller, PhDVirginia Commonwealth UniversityRichmond, VA, USAOutlinePrimary reasons for testing Slide Number 4Slide Number 5Slide Number 6Slide Number 7Slide Number 8Slide Number 9Slide Number 10Slide Number 11Slide Number 12Slide Number 13Slide Number 14Slide Number 15Slide Number 16Slide Number 17Slide Number 18Slide Number 19Slide Number 20Slide Number 21Slide Number 22Slide Number 23Slide Number 24Slide Number 25Slide Number 26Slide Number 27Slide Number 28Slide Number 29Slide Number 30Slide Number 31Slide Number 32Slide Number 33Slide Number 34Slide Number 35Slide Number 36Slide Number 37Slide Number 38Slide Number 39Slide Number 40Slide Number 41Slide Number 42Slide Number 43Slide Number 44Slide Number 45Slide Number 46Slide Number 47Slide Number 48Slide Number 49Slide Number 50Slide Number 51Slide Number 52Slide Number 53Slide Number 54Slide Number 55Slide Number 56Slide Number 57Slide Number 58Slide Number 59Slide Number 60Slide Number 61Slide Number 62Slide Number 63