User Manual Special Notes to Operators and Maintenance Managers Before using this system,be sure to throughly read this manual and familiarize yourself with the system. After reading this manual, keep it near the system,in a handy location where you can refer to it easily. Q2E-EA1222-3 opyright © itachi edical orporation. 2008 - 2010. ll rights reserved. Tokyo, Japan

Upload: phamcong

Post on 19-May-2018

1.580 views

Category:

Documents


360 download

TRANSCRIPT

Page 1: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

User Manual

Special Notes to Operators and Maintenance Managers

★ Before using this system,be sure to throughly read this manual and familiarize yourself with the system.

★ After reading this manual, keep it near the system,in a handy location where you can refer to it easily.

Q2E-EA1222-3

Copyright © Hitachi Medical Corporation. 2008 - 2010. All rights reserved. Tokyo, Japan

Page 2: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed.

About the ultrasound scanner display

Q2E-EA1222

Page 3: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

A table of contents

Before starting examinationTurning the Power ON or OFF

Panel Key Description

Examination screen (Screen buttons)

Selecting the Examination Conditions

ID Input

Examination of each modeB Mode

M Mode

Color Mode

Doppler Mode

Real-time Tissue Elastography ※

Display

Strain Ratio Measurement

ECG and Physiologic Module ※

BodyMarkComments and AnnotationsBasic Measurement

FilingMedia Handling

Formatting

Recording and Loading Images

Image Search and Play

Copy to media and Edit

1

2

3

4

5

6

9

10

11

1213

14

15

16

17

20

21

22

23

24

( i ) Q2E-EA1222

※option

Page 4: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Contrast Echo ※

Startup and Exit

Adjustment

MTI・TIC

Advanced MeasurementIMT ※

%WT ※Color Tissue Tracking ※

LA Tracking ※

WideView ※

Picture in Picture (PinP) ※

A table of contents

25

26

27

28

293134

36

37

※option

( ii ) Q2E-EA1222

4D (Real-time 3D) ※

Preparation

Scanning

Function

Volume Data Save and Load

HI DEF 3D (Scanning)

HI DEF 3D (Function)

HI DEF 3D (Multi Slice mode)

HI DEF 3D (CFM 3D mode)

38

39

40

41

42

43

44

45

Page 5: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

③ Cancel the FREEZE key to start examination

① Turn the Breaker switch ON

② Press the ON/STANDBY key

① Press the ON/STANDBY key

Start-up screen

Examination ready screen

The operation panel lights up, and the device starts up

② Click the OK button in the message

Breaker switch

Before turning the Breaker switch off, make sure that the system has completely shut down.

The light of the operation panel goes out, and system shutdown processing begins.

The LCD monitor goes completely black when the device shuts down

* Alphanumeric keyboard is option.

1 Q2E-EA1222

Turning the Power ON

Turning the Power OFF

Before starting examination Turning the Power ON or OFF

③ Turn the Breaker switch OFF

Page 6: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

『ON/STANDBY』 KeyPowers on or shuts down the device.

『PROBE』 KeyDisplays the probe switching dialog box.

『End Exam』 KeyPrints data, notifies the DICOM work list server of the completion of examination, and deletes measurement results.

『Patient』 KeyDisplays dialog boxes for entering or modifying patient information.

『PSS』 Knob Switches between the image adjustment parameter settings (data registered from the PSS setup menu).

『Multi encoder』 knobOperates the function selected from the screen button area.

『Image Menu』 KeyReflects the image adjustment parameters on screen buttons.

『Function1』・『Function2』 KeyActivates the preset option functions or displays a list of options in the form of screen buttons.

『Elasto/TDI Mode』 Knob / KeyActivates Elasto or TDI mode.Note that the activated mode depends on the settings.

『Fine Flow / CFA Mode』 KeySwitches Fine Flow or CFA on or off.Note that the mode depends the settings.

『HI Zoom』 KeyTurns high-definition zoom on or off.

『PAN Zoom』 KeyTurns PAN zoom on or off.

『READ』 KeyStarts the Filing window.

『UPDATE (NEXT)』 KeyInitiates the sweep of Doppler mode or M mode.

Panel key colors

2

Illuminated in orange: The key is selected.Illuminated in white: The key can be selected.Not illuminated: The key cannot be selected.

Before starting examination Panel Key Description

Q2E-EA1222

Page 7: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

You can use the screen buttons in any of the following ways:

- Use the pointer by means of the trackball and the ENTER key.

Area to display tab menu

Area to display thumbnail

Screen button area

Turns the selected function on or off.

Refreshes the screen buttons so that they reflect the selected function menu.

You can operate the selected function using the Multi encoder.

Scrolls through functions when the button is pressed.

The actual buttons displayed vary depending on the examination mode.

Executes the selected function when pressed.

Multi encoderPerforms operations related to the Encoder Func. button selected from the screen button area.

Screen button keys

Displays the action triggered by the Multi encoder.

OFFON

Menu

Encoder

Rotation

Push

- Use the Screen button keys on the operation panel.

3

How to use the screen buttons Types of screen buttons

System Information Display

Before starting examination Examination screen (Screen buttons)

Q2E-EA1222

*Possible to operate by screen buttons as well.

Page 8: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Use the trackball to move the pointer on the window.

You can add a new setting to the applications registered for each probe of the equipment, or edit the registered application settings.For further details, see the instruction manual.

Press the PROBE key to display the Select Probe/Application window.

① Select the probe to be used.② Select the application, and then press the ENTER

key.

The default application functionality allows you to assign frequently used application settings for each probe.For details on how to assign application settings for each probe, see the instruction manual for the device.

① Use the trackball to move the pointer on the window.② Place the pointer over the default application for the

desired probe, and then press the ENTER key to select it.

The selected application is activated, and the probe is directly selected.You are ready to scan the patient.

Present application is displayed here.

4

Selecting the probe and application

The selected application is activated, and the probe is selected.You are ready to scan the patient.

Selecting from the default application

Creating a new application (Editing)

Before starting examination Selecting the Examination Conditions

1 2

Q2E-EA1222

Page 9: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

After you enter the necessary information or made the necessary changes, place the pointer over the Start Exam button, and then press the ENTER key.

② Enter the ID from the keyboard.

① Press the Patient key to display the Patient Information window

The item to be entered varies depending on the display format

① Press the Patient key.

② Press the Start Exam key without inputting ID.

③ Press the OK button from Confirmation screen. ID number is automatically assigned.

Enter patient information, and then click the Save button to save the data. The next time you don’t have to enter ID, you can select from the list of IDs displayed in the lower part of the window.

Standard

OB

Use the trackball to move the pointer on the window, and then press the ENTER key to select the next item to be entered.

5

Entering patient information

Automatic assignment of ID number

Before starting examination ID Input

Q2E-EA1222

Page 10: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Focus Set.

#Focus Step

#Focus Width

6

Darker

Brighter

Press the center of the B Mode Knob.

Turn the Knob to adjust the gain.

Use the HI-Support key to optimize the B gain and depth-specific gain.

Shallower display depth

Deeper display depth

Use the Depth Knob to change the display size.

Higher frequency

Lower frequency

* Frequency indication is not displayed according to the mode used.

Shallower

Deeper

Increase width

Decrease width

Note: Enabled only if the focus width is greater than 1.

To display the focus-related screen buttons, press the Focus Set. button.To go back to the general menu, press the Focus Set. button again.

Focus step

Increase steps

Decrease steps

B Mode image display Changing frequency

Display depth Focus position

Focus settings

Focus width

Examination of each mode B Mode ①

Q2E-EA1222

Page 11: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Examination of each mode B Mode ②

Compound imaging by multidirectional scanning

HI Com ON/OFF are switched with each pressing of HI Com button.

Use of HI Com

dTHI-W-R: High resolution (Standard) mode

dTHI-W-P: High resolution(Deep scan) mode

dTHI-R: Standard mode

dTHI-P: Deep scan mode

HdTHI-R: High definition (Standard) mode

HdTHI-P: High definition(Deep scan) mode

ON

OFFHI Com

Up/Down Shift

HI Com

dTHI

Display Set.

7

*Different by probes.

To display the display-related screen buttons, press the Display Set. button.To go back to the general menu, press the Display Set. button again.

Left/right inversion

To invert the displayed image horizontally, press the Invert L/R button.

When dTHI is on, use the REF FREQ key to shift through the following modes:

Up

Down

Changing the depth

Deeper

Shallower

To change the depth, activate the Up/Down Shiftbutton, and then use the trackball.

To change the compound direction, use the OBLIQUE key.

Display-related settings

Selecting Harmonic Mode

Oblique B Mode image (for linear probe only)

Trapezoid

To switch the trapezoid display mode on or off, press the Trapezoid button.

Trapezoid display (for linear probe only)

Spatial Compound (HI Com)

B ModeImage

HI Com is off.

Left RightSteering OFF

Steering HI Com

HI Com

HI Com

HI Com is on.

Q2E-EA1222

Page 12: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Press the CINE key after freezing the image.

Cine memory images can be played back by frame with a trackball.

8

This feature allows you to enlarge only the ROI (region of interest) and render it in high definition. This will also increase the frame rate.

Press the HI Zoom key.ROI is displayed.

Use the trackball to move the ROI to the desired location.

To change the size of the ROI, press the ENTER key. A dot ( ) appears at the lower right corner of the ROI. Use the trackball to change the size of the ROI.

If you press the ENTER key again, or do not touch the trackball for over five seconds, the mark disappears, and the ROI can be moved again.

Press the UPDATE (NEXT) keyto the zoom mode..Press the HI Zoom key again to cancel the zoom mode.

You can enlarge an image and then move it to change its display position.

Press the PAN Zoom key.

Reduce

Enlarge

Use the trackball to change the display position of the image.Press the PAN Zoom key again to cancel the zoom mode.

To switch between single-window and dual-window mode, use the DUAL/SINGLE key.

To switch between the left and right windows, use the L(U) and R(D) keys.

Real-time Frozen Real-time

HI Zoom PAN Zoom

Frame-by-frame playback of cine memory Dual display

Frozen

Examination of each mode B Mode ③

Q2E-EA1222

Page 13: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Press the M Mode Knob.Press the M Mode Knob again to exit M Mode.

The ODM image indicated bythe ODM Symbols mark .

The ODM image indicated by the ODM Symbols mark .

ODM ODM

Dual ODM

Independent

Independent

9

Turn the Knob to adjust the gain.

To play the M Mode cine memory images frame by frame, freeze the display, and then turn the M Mode Knob.(If this causes the gain to change, press the center of M Mode Knob and then turn it.)

Moving the M Mode cursor

Use the trackball to move the cursor right or left.

Dual ODM display

To switch between Single Directional ODM and Dual Directional ODM, press the Dual ODM button.

Use the UNDO key to select the ODM cursor to be moved.The selected cursor is highlighted in yellow.

During Dual ODM you can use screen buttons to switch between the two cursor line control options.

Both cursors have their own rotation center, and therefore move separately.

Both cursors share the same rotation center, and therefore move together.

Rotating and moving the ODM cursor line

Turn the Angle Knob so that the cursor line rotates around the mark.Press the center of the Knob to turn the cursor line 45 degrees at a time.Use the trackball to move the cursor line position.

To switch ODM Mode ON or OFF, press the ODMbutton displayed during M Mode.

Normal M Mode ODM Mode

M-Mode display Omni-directional M Mode (ODM Mode) display

Brighter

Darker

Examination of each mode M Mode

Q2E-EA1222

* option

Page 14: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Fine Flow and CFA are switched by each pressing of a screen button.

Dual CFM

To CFA

Dual CFM

To Fine Flow

10

Press the center of the CFI Mode Knob.To exit CFI mode, press the CFI Mode Knob again.

Turn the Knob to adjust the gain.

Press the Fine Flow/CFA Mode Knob.To exit Fine Flow mode, press the Fine Flow/CFA Mode Knob again.

Turn the CFI Mode Knob to adjust the gain.

Fine Flow mode is selected.(High-definition blood flow power display)

CFA is selected.

Higher velocity

Tilt the OBLIQUE key in the direction to which you want to tilt the ROI.

Press the Elasto/TDI Mode Knob to switch TDI Mode ON or OFF.

(Turn the Knob to adjust the image.)

* Depending on the probe or application being executed, this might cause the Elasto to be activated. For details, see the Instruction Manual.

Use the trackball to move the color box.

To change the size of the ROI, press the ENTERkey. A dot ( ) appears at the lower right corner of the ROI. Use the trackball to change the size of the ROI.

To switch the parallel display on or off, press the Dual CFM button.

Dual CFM OFF Dual CFM ON

Color Doppler (CFI Mode) display Tissue Doppler image (TDI Mode) display

Switching between the Power Doppler (CFA mode) and Fine Flow modes

Changing the position and size of the color box

Parallel display of colored and black-and-white images

Changing the velocity range

Oblique color box (for linear probe only)

Brighter

Darker

Lower velocity

Examination of each mode Color Mode

Q2E-EA1222

Page 15: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Simult. ON

Frozen

Frozen

Real-time

Simult. OFF

Simult.

#Sound Volume

11

To activate Pulse Doppler, press the PW Mode Knob.To activate Continuous Wave Doppler, press the CW Mode key.To exit the current mode, press the same Knob or key again.

When you cannot get the Doppler wave form just by Doppler cursor, press “UPDATE (NEXT)” key to start the sweep of Doppler wavform.

To adjust the gain, turn the PW Mode Knob.

Use the trackball to move the Doppler cursor.

If the Doppler cursor is not yellow, press the TRACKBALL key to turn it yellow, and then move the cursor.

Greater PRF

Less PRF

You can change the signal removal level.

Remove higher frequency

Remove lower frequency

Wider

Narrower

Higher volume

Lower volumeSelect the Sound Volume button, and then use the Multi encoder to adjust the volume.

Press the Knob to bring up the angle bar. Slightly turn the Knob to adjust the bar so that it is aligned with the direction of blood flow.

Counterclockwise

Clockwise

blood flow.

Doppler Mode display

Pulse repetition frequency (PRF)adjustment

Sample gate adjustment

Wall filter Volume adjustment

Switching of simultaneous real-time display

Angle correction

Examination of each mode Doppler Mode

Real-time

Real-time

Real-time

Every time the Knob is pushed, the display in registered angle, reversing, and erasing of the angle bar are repeated.

Q2E-EA1222

Page 16: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

NextPrev.Page

E-Dyn. MapL MapR Frame Rej. Noize Rej.Frame Rata Persistence Density

Real-time Tissue ElastographyDisplay

To turn the Elasto mode on or off, press the Elasto/TDI Mode Knob.* Depending on the probe or application that is being executed, the TDI is activated.* When using mechanical compression unit (Option) for transrectal probe, select “E_Balloon_Urology” in

“Select Probe with Application” screen, and boot up the Elasto mode.

Parallel display of the Elasto Mode and B Mode images

The Dual Elasto screen button switches the display.

① Make sure that the mark is selected and illuminated in yellow in the trackball priority section in the lower right of the examination window.

The ROI size is fixed. Use the trackball to move the ROI to the desired location.

If the mark appears in the lower right corner of the ROI, you can change the size of the ROI. Use the trackball to adjust the size of the ROI.

② Use the trackball to move the ROI or change its size. To switch between the two functions, press the ENTER key.

The ROI can be adjusted not only in real time image but also in frozen image.

Press the ROI ReConfig. button, and then set the ROI as described in steps 1 and 2 above. Pressing the ROI ReSetbutton returns the ROI to its state immediately after the freeze. Canceling the freeze state causes the ROI to return to the location before the freeze.

Use the trackball to select the frame.

To adjust the Elasto frequency, use the REF FREQ key.(Depending on the probe, the frequency might be fixed.)

High:Image quality priority

Low:Sensitivity priority

To make other adjustments, press the Image Menu key, and then use the displayed screen buttons.

If not selected, press the TRACKBALL key to select the trackball.

Dual ElastoDual Elasto

Frozen

ROI ReConfig.

option

12

Display Elasto Mode

ROI adjustment

Image quality adjustment

Q2E-EA1222

Page 17: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

ShapeChange;All

ShapeChange;PartROI Shape;Rect

ROI Shape;Ellipse

A …The mean strain within the region surrounded by ROI-AB …The mean strain within the region surrounded by ROI-BB/A…The comparison of strain within ROI-A to ROI-B (strain ratio)

Real-time Tissue ElastographyStrain Ratio Measurement

The strain ratio measurement allows you to use an Elasto mode image to measure the “strain” and “compare the strain”.

For the breast, setting the ROI as shown in the left figure allows you to estimate the strain in both the fat and lesion regions and then calculate the FLR, or the fat-lesion ratio.

* The strain is expressed as a percentage (%). For example, if a 100-mm object has a strain of 1 mm, the strain is expressed as 1%.

Fat

Lesion

Measurement results

① After pushing the FREEZE key, select the appropriate Elasto mode image, and then start “Str.Ratio” from Measurement Menu.

ROI size change and relocationTo adjust the ROI size, use the ENTER key and the trackball to move the control point. The control point is displayed by placing the cursor on top of the ROI frame. If ROI Shape; Ellipse is selected, the size adjustment mode varies depending on the shape of the control point.

Selecting the ROI shapeUse the screen buttons to select one of the following shapes:

Press the ROI Shape button to switch between elliptical and rectangular shapes.

Press the Shape Change button to switch between overall and partial adjustment. Partial adjustment allows you to form the ROI into an asymmetric shape.

:Size change without changing the shape+ :Size change in the diagonal direction

:Partial size change (available only if Shape Change; Part is selected)

ROI-A

ROI-B

To move the ROI, move the cursor to inside the ROI. The ROI can be moved after the cursor changes to a four-directional arrow.

* This button is enabled only if ROI Shape;Ellipse is selected.

② Move the cross cursor displayed in the center of the screen to the desired starting point of ROI-A, and then press the ENTER key.

③ Move the next displayed cross cursor to the desired end point of the ROI-A, and then press the ENTER key to draw the ROI. The measurement results are displayed when ROI-A is set.

④ A cross cursor appears outside ROI-A. Repeat steps 2 and 3 above to set ROI-B.* In Dual Elasto mode, ROI-A and ROI-B can be specified in either the left or right window.

13

About strain ratio measurement

Procedure for measuring strain ration

Strain ratio measurement and ROI adjustment

option

Q2E-EA1222

Page 18: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Case of Clip electrode;Red (R) Black (RF)Green (LF)

To adjust the amplitude of the waveform, press the ECG Gain button, and then turn the Multi encoder.

LargeSmall

Up Down * You can change the position even after a freeze.

Press the ECG Invert button, then ECG waveform inverted.

To change the sweep speed of the waveform, press the SWEEP SPEED key.

Press the PHYSIO key to display the ECG screenbutton.

To adjust the position of the waveform, press the ECG Pos. button, and turn the Multi encoder.

Fast

Slow

ECG and Physiologic Module

Case of disposable electrode;RedBlackGreen

ECG

Press the ECG A. Gain button, then ECG waveform automatically adjust.

14

Attach an electrode to the patient's body

Control of the waveform

Displaying an ECG trace

ECG Invert

#ECG Pos.

#ECG Gain

option

: right hand

: left leg: right leg

: right waist: right shoulder

: left waist

Q2E-EA1222

Page 19: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

BodyMark

Mark Move : Allows you to move the display position of the body mark by using the trackball.

BodyMark : Displays the body mark-related screen buttons.

Mark Display : Clears body marks displayed in the examination window.

: Scroll through the pages in the variable area.

When using dual-window mode, you can use the L or R key to select the window in which the body mark (probe mark) is to be set.

Turn the PROBE MARK Knob to rotate the probe mark.

① Make sure the mark is selected in the trackball priority section in the lower part of the examination window. If not selected, press the PROBE MARK Knob to select this mark.

② Use the trackball to move the probe mark to the desired location.

Rotate

Move

① Clicking the BodyMark screen button displays screen buttons representing the pre-registered body marks. ② Select the target BodyMark.

You can also select a body mark by pressing the PROBE MARK Knob.You can scroll through the pre-registered body marks by pressing the Knob.

Selecting the BodyMark

(Example)

Screen buttons

BodyMark Region :Displays the category name indicating a specific section of the body.Abdomen / Breast / Head/Neck / Urology / Cardiovascular / Organ / Upper limb / Lower limb / Obstetrics / Gynecology

Variable area (displays body marks as well as category names indicating the specific section of the body)

Clear

15

Adjusting the probe mark

Changing the BodyMark

Q2E-EA1222

Page 20: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Comments and Annotations

A comment can be entered directly from the keyboard. An annotation can be selected from the menu showing a list of registered text entries.

Activation and input

Press the Char. key from the keyboard to display a cursor. Use the trackball to move the cursor to the target location, and then enter the desired text.Press the Comment key again to finish entering the comment.

* If an alphanumeric keyboard (optional) is not attached, use the on-screen keyboard.For details of the usage, see the instruction manual HI VISION Avius-operation-.

cursor

ABCD

Moving a comment Setting the home position of the comment cursor

Inputting arrows by using the keyboard

Place the cursor over the entered text, and then press the ENTER key. A yellow frame appears around the text. The text can then be moved.

ABCD ABCD

ABCD

Move the cursor to any given location, and then press the Home Posi. screen button to set the home cursor location. From the next time on, the cursor will automatically appear in this location.

To input an arrow, hold down the Ctrl key and press one of the following alphanumeric keys:

Arrow

NBVCLKOIA/N Key

Deleting all text Deleting a single line

Press the Annotation key on either the operation panel or the keyboard.From the displayed list, select the target text, use the trackball to move it to the desired location, and then press the ENTER key.Press the ANNOTATION key again to finish entering the annotation.

or

Press the CLEAR key from either the operation panel or the keyboard to delete all comments (annotations) from the examination window.

To delete a single line of text, place the cursor over the text to be deleted, and then press the Shift and Del keys together.

Liver

Liver Ior

Liver

Both the cursor and the menu disappear, and entering of the annotation ends.

or

Press the ENTER key again to fix the location.

* This will also cause the measurements to be deleted.

16

Comments

Annotations

Deleting text

Q2E-EA1222

Page 21: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Basic Measurement ①

The basic measurement methods are caliper measurement and trace measurement.Other special measurements are available by either pressing the MEAS. MENU key or by selecting one of the screen buttons.

① Press the CALIP key to display the caliper mark .② Press the ENTER key to set the start point, and then move the next caliper mark to the end point.

Press the UNDO key to set the end point and go back to moving the start point again. ③ Press the ENTER key to set the caliper. After a caliper mark has been set, a new caliper mark appears so

that the next measurement can be specified.

The measurement to be executed

Measurement procedure

In Doppler mode, press the CALIP key to switch between speed measurement and time measurement. (Measurement mode can also be changed by using the screen button.)

Manual trace procedure

Ellipse trace procedure

In B mode, press the Trace key to switch between manual and Ellipse area measurement method. (Measurement mode can also be changed by using the screen button.)

① Press the TRACE key to display the caliper mark .② Press the ENTER key to fix the start point, and then use the trackball to trace the target region.

Press the UNDO key to erase the trace line in reverse. ③ Press the ENTER key to close the trace line and end the measurement. After an area has been set, a new

caliper mark appears so that the next measurement can be specified.

① Press the TRACE key to display the caliper mark.② Press the ENTER key to set the caliper, move the

next caliper mark to the next desired location, and then press the ENTER key to set the caliper mark. Adjust the frame size by using the trackball.Press the UNDO key to readjust (a), (b), (c), and (d) in the figure to the right.

③ Press the ENTER key to set the ellipse and end the measurement. After an area has been set, a new caliper mark appears so that the next measurement can be specified.

(a) Frame size(b) Location of the 1st caliper mark(c) Frame size(d) Location of the 2nd caliper mark

Measurement that is executed

Measurement result

Measurement result

Caliper;Time Caliper;Velocity

Trace;Manual Trace;Ellipse

(a)

(b)

(c)

(d)

17

B-Mode : DistanceM-ModeD-Mode : Speed (pressure gradient) and time

B-Mode : AreaM-ModeD-Mode : Average flow velocity

Caliper measurement

Trace measurement

: Distance, circumference, and area

: Depth difference, time, and slope

Q2E-EA1222

Page 22: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Press TRACKBALL key to change the display of trackball function.Operate screen buttons when mark is selected (yellow highlighted) as trackball priority (displayed in the lower right of examination window).

Basic Measurement ②

Threshold : Adjusts the threshold so that the trace line fits the waveform.

Smooth Line : Performs smoothing processing in the direction of temporal axis.

Dir. : Available trace direction options include up, down, and up/down.

Select the measurement result frame containing the measurement you want to redo. The color of the frame will change to yellow. Press the ENTER key to redo the measurement.

After a new caliper for the next measurement appears, press the UNDO key, and then select a previously measured item.

Press the CLEAR key to delete the measurements displayed in the examination window.

Starting and exiting the real-time measurement

Editing a real-time trace

If you press the TRACE key while observing the Doppler waveform, trace processing starts up, updating, in real-time, the measurement results.

Press the CLEAR key to end the processing.

You can also perform real-time measurement by selecting measurement from the MEAS.MENU. The preset measurement items are displayed.

*This will also cause the comments (annotations) to be deleted.

Measurement result

Real-time Real-timeReal-time

Measurement correction

Deleting measurements

Real-time Doppler measurement

Average. : Display measurement result by averaging heartbeats.

*Possible to adjust by screen buttons as well.

18 Q2E-EA1222

Page 23: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

*Possible to adjust by screen buttons as well.

Basic Measurement ③

Section trace

If trace measurement is selected from measurement menu, press UPDATE(NEXT) key or Store button.

Continuation of Real-time Doppler Measurement

Auto trace

To average heartbeats, press Average of screen button, and designate average heartbeat.

To average heartbeats, press Average of screen button, and designate average heartbeat.Circle the waveform of designated heartbeat.

Display the last one heartbeat just after freezing the image.Turn PW mode knob to move measurement cursor.Use TRACKBALL to move cine memory.

Selecting the waveform after a frozen

Set trace type of screen as Section.Set the displayed caliper on the contraction starting point of subject waveform, and press ENTER key. Move the line displayed next to end diastolic period, and circle 1 waveform.Select up and down direction of wave form by + mark. Press ENTER key to plot Vp (peak velocity) and Vd (end diastolic velocity) and display measurement value.

If trace measurement is selected from measurement menu, press UPDATE(NEXT) key or Store button.

19 Q2E-EA1222

Page 24: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Filing Media Handling

Manual Edit Patient Mode Adjust Media Remove Har・・・ Simplified T・・・ DVD-RAM USB Memory USB HDD Filing Cine Memory Display Account Menu Physiology Video/ORIG PinP

Application

Tool M

easureA

nnotation

Supported Media

Supported media: DVD-RAM, DVD-R*, USB memory devices, and USB hard disk (USB-HDD).The use of reliable DVD-RAM media is recommended. (For details on the model names of the recommended media, see the instruction manual HI VISION Avius ~Preparation~.)

Disk ejection

Press the EJECT key.

① From the tab menu on the left side of the screen, select Tool and then Remove Hardware.* The Safely Remove Hardware window is displayed.

② Select the hardware to be removed, and then press the Stop button.* The Stop a Hardware device window is displayed.

③ Click the OK button.* Hardware can be removed.

④ Click the Close button.

USB (memory or hard disks) removal

* Do not eject the disk by pressing the eject button on the drive.

* Note that no additional data can be written to a DVD-R disk.

20 Q2E-EA1222

Page 25: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

FilingFormatting

Formatting media

④ Follow the following message to start formatting.

Formatting a DVD-RAM

You can also format media by selecting Tool and then Media from the tab menu. There is no need to format a DVD-R. For details on how to copy to DVD-R , see the instruction manual HI VISION Avius ~Operation~.

Click OK if you agree with the message, and then execute formatting.

Formatting USB memory (USB-HDD)

Click the Start button.Click the Start button.

When the Finished message indicating the completion of formatting is displayed, click the Close button.

① Select an image in the filing window, and then click Copy. The menu shown in the figure to the right is displayed.

② Click under Destination, select the media, and then select Format. A message is displayed.

If you want to proceed to copying images, click OK in the copy menu window to initiate image transfer.

Select the media to be formatted.③ Click the Yes button if you agree with the message.

【Copy menu window】

1) Enter the name of DVD into Volume label.

*The volume label can use only the alphanumeric only by 11 characters.

2) Select the check box.

2) Enter the name of USB into Volume label.

1) Select FAT32.

3) Clear the check box.

【Format window】【Format window】

21 Q2E-EA1222

Page 26: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Filing Recording and Loading Images

* While recording the live image, the transfer progress is displayed in the lower part of the screen (system information display area).

Recording still and live image

Press the REC2 key during observation.

Prolonged live image recording

Stopping prolonged live image recording

Stopping live image recording

Recording live images

Recording live image real-time (option)

Displaying the Filing window① Press the READ key during the freeze to display the Filing window.

② Select Location. ③ Select Patient. ④ Select the Study Date.

* Start searching for images from 2, 3, and 4 in the Filing window above.

* Images selected from 3 and 4 are displayed in 5 (List View).

Press the REC1 key.

* When setting up for the Cardiac region, the last three heartbeats immediately before the freeze are automatically loaded as a live image.

* The assignment of keys and the method of loading might vary depending on the settings. (For details, see the instruction manual HI VISION Avius ~Operation~.)

To stop recording, either press the REC2key again or press the FREEZE key.

Recording live image

Press the assigned REC# key.

Press the assigned REC# key again to stop recording.

* In prolonged live image recording mode, you can record live images even during the freeze.

* The image can be recorded either directly to the DVD-RAM or to the device's HDD.

* The assigned key might vary depending on the settings.

① Press the REC2 key after the freeze.

② Follow the instructions. Use the trackball to select the first frame you want to display as a live image, and then press the ENTER key. Next, select the last frame.

Indications displayed while recording still and live images

still images live images

live images

2 3 4

5

2 3 4⑤ Images are displayed

in List View.

22 Q2E-EA1222

Page 27: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Filing Image Search and Play

Selecting and playing images

Press the screen button to switch views.

Switching between the Retrieve window and Examination window

Searching images

Playing images

Switching between multi-window and single-window views

① Press the READ key during the freeze to display the Filing window.

Press either the READ or FREEZE key to go back to the examination window.

Press the Filing button to go back to the Filing window.

Playing and stopping live images (AVI/MPEG2)

Use the screen button to play or stop the live image.

Play Stop

① Using the trackball and the ENTER key, select the thumbnail representing the image you want to play.

Switching the images (pages)

Press the screen button to switch images(pages).

To the next image(page)

To the previous image(page)

① Enter the keyword.

② Click the Search button.② Click in the Search field of the Filing window to select the search method.

(Example) How to search by keyword

③A thumbnail representing the corresponding data is displayed.

* Press the ENTER key to select the image. A check mark is displayed on the selected image.

② Click the Retrieve button to play the image.

Switching between the Retrieve window and Filing window

Filing

To MultiTo Single

Filing

Prev. Page Next Page

23 Q2E-EA1222

Page 28: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Filing Copy to media and Edit

⑥ Press the Trim button, and then select the transfer destination to save as a new file.

Copying an image (examination data) to media

Masking patient information

Press the READ key during the freeze to display the Filing window.

④ Select the Yes button if you agree with the message.

① From the Filing window, select the images whose information you want to masking, and then click the Convert button.

Trimming live image

③ Click the Select Start button.

⑤ Click the File Operation button.

① Play the live image to be edited in the Filing window in single-window mode.

② Using the trackball, display the first frame of the segment to be cut out.

④ Similarly, display the last frame, and then press the Select End button.

① Select the image or examination data to be transferred.

② Click the Copy button. ③ Select the transfer destination and click the OK button.

③ Select the Without personal information check box,andthen select the OK button. A message is displayed.

② Click under Destination, select the media.

Select Start

Select End

File Operation▲

Trim

Select Start Select End

24 Q2E-EA1222

Page 29: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Contrast EchoStartup and Exit

Starting up contrast option

For SonazoidⓇ, Hitachi recommends selecting an application whose name starts with Contrast-Mid. For SonoVueⓇ, Hitachi recommends selecting an application whose name starts with Contrast-Low.

When both the Contrast button and the Color Mode Knob are turned ON, the mode becomes CWPI (color contrast mode).

Window display in the contrast modedCHI-W-#: Wideband Pulse Inversion (black and white contrast images) CWPI-#: Color Wideband Pulse Inversion (color contrast images) FI: Fundamental Image (fundamental wave images, monitor images)

Press the Contrast button to change to the contrast harmonic mode.

Press the F1(Function 1) or F2(Function 2) key to display the following menu, and contrast echo operations become possible:

Selecting an application

1 2 Select an application for contrast. ① Select a probe ② Select an application. (There is no need to press OKbutton.)

ContrastContrast

Contrast harmonic ONContrast harmonic OFF

If the above menu is not displayed, select the Contrast button to display the menu.

The description in # varies depending on the frequency.

Flash

MTI

UCI

I.FlashAlternateContrast To FI

MFrame Rate

M

Starting up contrast harmonic

Menu;Contrast

User Define

Mesurement

Body Mark T.StartContrast

Body Mark

Mesurement

T.Start User Define

Menu;Contrast

* Make the screen button OFF to turn OFF contrast harmonic mode.

* Display of screen buttons would change.Contents of buttons differ according to the situation.

Press same keys as used for starting up (F1 or F2) to exit Contrast Echo Option.

Exiting Contrast Echo Option

* In case Contrast Echo Option is not exited, press the Contrast button.

25

Contrast

option

Q2E-EA1222

Page 30: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Contrast EchoAdjustment

This is a useful function for monitoring the time elapsed after the injection of a contrast agent.

Press the T.Start(Timer Start) button to display the elapsed time in the bottom right of the window.

Starting up the timer

T.Start

Press the T.Clear(Timer Clear) button to reset the counter to 00:00.

You can independently adjust MI and gain for the left (contrast image) and the right (monitor image).

Choices during dual-window display (alternate mode)

① Select the window you want to adjust by using the L(U) key or the R(D) key.

② Adjust MI and gain by using their respective keys.

Contrast image Monitor image

【Alternate Window】

How to adjust MI

Reduce MI

Increase MI

Changing contrast harmonic frequency

dCHI-W-R: Priority given to high resolution

dCHI-W-P: Priority given to depth sensitivity

You can change frequencies by using the REF FREQ key.

You can change MI by using the US POWER key.

【Alternate Window】

This is a function for displaying, in real-time and in parallel, both a contrast harmonic image that effectively extracts the contrast agent and a fundamental image with low MI (monitor image) that extracts the tissues of a living body.

Changing Alternate mode (monitor mode)

Press the Alternate button.

Alternate OFF Alternate ON

Alternate Alternate

T.Start

T.Clear

T.Clear

Contrast image

Fundamental image

(Monitor image)

* Setting like MI, gain, etc of each window can be checked at the bottom of examination window. (Refer to the right image)

26

option

Q2E-EA1222

Contrast image

Fundamental image

(Monitor image)

MI gain of Contrast image

MI gain of Monitor image

Page 31: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Contrast EchoMTI・TIC

MTI (Micro Bubble Trace Imaging)

MTI improves the visibility of the effects of a contrast agent and the visibility of where blood vessels run by displaying, per frame and through superposition, the shadowed parts of a contrast image.

Press the MTI button to start superposition.

MTI MTI

MTI OFF MTI ON

TIC (Time Intensity Curve)

① On the Filing window, play an AVI file for TIC.

TIC is a function for drawing, based on recorded an AVI file, a time brightness level curve that shows the relationship between contrast brightness level and time.It is helpful in observing the changes in brightness level over time.

Flash

Press the Flash button to destroy the contrast agent, clear the superposed images, and observe the reperfusion by performing superposition again.

② Select the TIC button.

③ Select an ROI shape from the following and place an ROI where you want to watch the changes in brightness level .

④ Select the Analysis button in the bottom right of the window.

TIC

ダミー

Image playback window

TIC ROI setting window

【Filing window】 【Image playback window】 【TIC ROI setting window】

【TIC analysis window】【TIC ROI setting window】

27

option

Q2E-EA1222

Page 32: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

When three point method is selected in measurement setting, measurement marker like the right one is displayed.Move the cursor over the detection mark and press the ENTER key, and the detection mark becomes movable. Move the detection mark to any position and press the ENTER key again to fix that position.

Correction of trace method

Move the cursor on the vertical line of either sides of ROI, or either of horizontal lines (top or bottom of ROI), and press ENTER key to resize the ROI.

① Move and set the ROI to the position where IMC is placed lower than the center of ROI.

② Set the positions A, B, and C of the three-point measurement to the position where the blood vessel has no noise in its inside, and the IMC and tunica externa can be distinguished clearly.

③ Press ENTER key to fix the position of ROI. Measurement result is displayed when ENTERkey is pressed again.

Press Measurement Menu, and start R-IMT/L-IMT from the items displayed.

IMT measurement is the measurement to calculate the thickness of the internal media complex (IMC) of a carotid artery through automatic trace processing.“trace method“ and “three-point method”, which calculate only the mean value of three points are available.

Moving the ROI

Resize

Rotate

Manual correction Click Manual Correction to display control points on the trace line.Move the cursor over the correction point you want to correct, push the ENTER key, and again push the ENTER key to put on that point.

Move the cursor into the ROI and press the ENTER key, and the ROI becomes movable. Press ENTER again to fix the ROI.

Turn the Angle key while the ROI is displayed with a solid line for rotation.

Three-point method

How to set an ROI

Correction of automatic trace method

L-IMT

R-IMT

optionAdvanced MeasurementIMT

Change the threshold of red line (tunica intima) and green line (tunia externa) so that they can be moved to the position users like.Number of threshold is displayed.

28 Q2E-EA1222

Page 33: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

First frame Last frame

This performs measurement by using B Mode video (moving images). Pixels of two points (left ventricular wall) specified by the user are automatically traced, and based on the results of that tracing, changes in the distance between the two points are calculated. From the distance between the two specified points and from the distance of the maximum wall thickness, the rate of wall thickness increase (%WT) will be calculated.

Press the REC2 key after the freeze. The latest R-R interval frame is automatically captured.

Press the READ key to open the Image Filing window. Select a AVI file used for measurement from the image list and press the Retrive button or directly press double click on the image.

Press the Measurement Menu key, select %WT from the displayed items, and start %WT.

This measurement is possible only for data recorded in AVI file format, and not possible for real-time or cine memory images.

Move the cursor over the position of a measurement point, and the cursor's shape changes to the following. If you press the UNDOkey, the measurement pair is deleted and new positions can be specified. (It is possible to correct multiple pairs)

Start

Set the acquiring image parameter of filing menu.

② Select R-R out, Number 1, then close a menu.

* Need a real time archive unit and ECG.

① Press TAB MENU, setup→filing→cardiac

① Specify two points for the endocardial wall and epicardial wall as a pair by using the cursor.

② Specify all the regions you want to measure and press Start button to start the analysis.

* Set measurement pairs on the top frame of the recorded video of one heartbeat.

29

Acquisition image data formeasurement (%WT/CTT/LAT)

Starting %WT measurement

Placing tracing points

Saving dynamic images

Re-measurement

Selecting an image to be measured

optionAdvanced Measurement%WT ①

Q2E-EA1222

Page 34: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

30

You can output the following measurement data in CSV file format• Distance per pixel [mm/pixel] • %WT [%] of each measurement pair • Maximum thickness [mm] of each measurement pair • Initial thickness [mm] of each measurement pair • Thickness [mm] of each frame for each measurement pair • Coordinates on the window of each frame for each measurement value

① Press the Save file button.② Select the drive on which the data is to be saved.

Outputting measurement data

The playback function allows you to evaluate the movement of measured points.

Starting playback Stopping playback

The video is automatically played back (traced and displayed) from its measurement start frame to the measurement end frame, and saved as a new video.

(This requires optional digital video recording software.)

Press Speed button to change the playback speed.

Graph

RevCapture

Saving data Saving the locus image

Displaying graphs

Checking the locus image

#Speed

Review

optionAdvanced Measurement %WT ②

Select a suitable tab to change graphs. (Distance, Strain and Data)

Change the graphs only for the List All.

Q2E-EA1222

Page 35: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Select a short axis level to be analyzed.

Color Tissue Tracking (hereafter CTT) is a function for calculating the radial strain of a cardiac short axis image. CTT automatically determines a tracing point from the ROI specified on an AVI file. Calculated strain and strain values will be displayed by using color images and time sequence graphs.

Acquire image data used for measurement

Configure the same save settings as %WT. Extract a short axis image of BASAL, MID, or APICAL and save a short axis level as close to a perfect circle as possible as an image for Bull’s eyes display.

Selecting an image to be measured

Press the READ key to open the Image Filing window.Select a moving image used for measurement from the image list and press the Retrievebutton to play back that video. For a Bull's Eye display, review the three sections simultaneously.

Press the Meas. Menu key, select CTT from the items displayed, and start CTT.

You can change the thickness of the ROI. The default value is 0.8 cm.

・ Press the ENTER key to show the ROI.・ You can set the ROI while the video is being

played back. ・ Check that the ROI is set on the top frame.

31

optionAdvanced Measurement Color Tissue Tracking ①

Q2E-EA1222

Page 36: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Adjusting the ROI

① Align the measurement cursor with the intima, click the ENTER key,align the measurement cursor also with the intima on the opposite side, and click the ENTER key.

② Move the trackball to adjust the intima elliptically and click the ENTER key.

Correcting the ROI

③ Press the Start button.

Correcting partially Click the light blue point on the ROI for correction.

Resetting

Changing the image and graph positions

Command functions

You can change the display positions of video images and graphs.

*Measurement on graphs is enabled only when graphs are displayed in large size on the right side.

Changing the %WT value and color

To change the %WT value, retype the value in the dialog box. To change the color, click the color box to display the color palette. Choose a color you prefer.

32

You can reset the ROI setting by using the Reset button.

optionAdvanced Measurement Color Tissue Tracking ②

Q2E-EA1222

Page 37: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Saving a AVI file on basic area

Saving measurement data

Saving AVI file on full area

Resetting the ROI

A CTT video (moving images) is created from the analyzed video. You can check the created video from the Image Filing window.

You can correct the ROI.The ROI returns to the pre-analysis status.

For a AVI file, you can save all frames synchronized with a graph's time bar.

Press the Save button to save the measurement result data (used for Bull's Eye) in CSV file format.

Command functions

*Use the frame bar (frame 1 to frame 45 in the figure) to specify a start point and an end point. Use the pointer to move the frame bar.

Analysis of trimming For a AVI file of two heartbeats or more, you can perform an analysis between the specific frames of each heartbeat.Use the frame bar (frame 1 to frame 45 in the figure) to specify a start point and an end point. Use the pointer to move the frame bar.

33

optionAdvanced Measurement Color Tissue Tracking ③

Q2E-EA1222

Page 38: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

* For P-P capturing, use the manual transfer to capture images by specifying any single heartbeat.

LA Tracking (hereafter LAT) is tracking system of traced line on an B-mode AVI file. LAT calculates the area and volume from the analyzed result, and displays the change graph of time.

Acquire image data used for measurement

Configure the same save settings as %WT.Extract short axis images of the apex 4-chamber (hereafter A4C) or the apex 2-chamber (A2C) and save the images for one heartbeat. For Bi plane measurement, save images in the order A2C to A4C.

Selecting an image to be measured

Press the READ key to open the Image Filing window.Select the AVI file to be used for measurement from the image list and press the Retrievebutton or double click to play back that file. For Bi plane measurements, play back the two sections simultaneously.

Press the Meas. Menu key, select LA Trackin from the items displayed, and start LAT.

When the cursor exists the image area, tracing is always possible.Click at the start point and repeat clicking at several points along the curve. Fit the trace line into the left atrium and double-click at the end point.Check that the ROI is set on the top frame.You can set two ROIs (A and B) in a single window.

For Bi plane measurement, select between A4C and A2C for the type of cross-section.

LAT measurement

34

optionAdvanced MeasurementLA Tracking ①

Q2E-EA1222

Page 39: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Press the Start button to perform LAT analysis.For Bi plane measurements, switch between A4C and A2C for AVI file.

The graph type can be changed to such types as volume, area, and volume change speed. Also, use check boxes to change the graph display type as you wish.

You can change the display positions of video images and graphs by using the icons.

For a AVI file of two heartbeats or more, you can perform an analysis between the specific frames of each heartbeat.Use the frame bar (frame 1 to frame 45 in the figure) to specify a start point and an end point. Use the pointer to move the frame bar.

Press Save button to output the measurement result data in CSV file format. Select a drive to save the data.(Data saved here can be loaded in graph form again.)

Command functions

Changing the image and graph positions

Start analyze

Analysis of trimming

Changing the graph type

Saving a AVI file on basic area

Saving measurement data

Saving AVI file on full area

A LAT moving images is created from the analyzed image. You can check the created video from the Filing window.

For a AVI file, you can save all frames synchronized with a graph's time bar.

35

optionAdvanced MeasurementLA Tracking ②

Q2E-EA1222

Page 40: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

StopStart

Cine

Wide View

Rotate

Pressing the key ( F1 or F2 ) to which the Wide View is assigned brings up the appropriate screen buttons.

Options menu

* The layout of the buttons varies depending on the configuration.

Wide View menu(for PAUSE)

Wide view can be terminated by pressing the same key ( F1 or F2 ) that was used to start up the Wide View.Just as with the startup process, if this does not directly cause the Wide View to shut down, press the screen button marked Wide View to terminate the Wide View.

PAUSE window REC window STOP window

Place the probe on the scanning beginning point.

* Care must be taken so as not to let the probe shift in the short axis direction.

When your target range has been completely scanned, press the Stop button or the FREEZE key to finish.

Enlarge or reduce

Turn the Angle/ODM Knob.

Press the PAN Zoom key to enable enlargement or reduction. Change the magnification ratio by using the DEPTH (MAG) Knob.

Playing captured original images

The captured original image is displayed to the right of the window. Use the trackball to perform playback of cine memory for the original image.

Original image

Wide View operation conditions・Freeze OFF ・Pan zoom, Hi zoom, and other optional functions not used・Linear or convex scan type probe used ・B Mode or Color Mode single-image display

* In order to correctly obtain a wide view, the details of the deepest level of an image must be precisely extracted and displayed in this window.

* Depending on the system configuration, this can either directly display the Wide View menu or require pressing the screen buttonmarked Wide View first to display the wide view menu.

Press the Start button, and scan along a range over which you want to obtain continuous images.

* When the Freeze mode is released, the window returns to PAUSE.

Stop

User DefineBodyMark

Menu;WideViewStartMeasurement Cine Review

StressContrast WideView

36

Starting the wide view and capturing an image

Exiting Wide View

Manipulating a Wide View image (when on the STOP window)

option

Q2E-EA1222

Page 41: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Picture in Picture (PinP)

Manual Edit Patient Mode Adjust Media Remove Har・・・ Simplified T・・・ DVD-RAM USB Memory USB HDD Filing Cine Memory Display Account Menu Physiology Video/ORIG PinP

Application

Tool M

easure A

nnotationS

etup

Executing Picture in Picture (PinP)

① From the tab menu on the left side of the screen, select the Tool tab.

② Select PinP to execute imaging through the S-terminal (camera images) on the Examination window as shown below.

When PinP is activated, the following screen buttons are displayed on the window:

Display position of the camera imageWhile the Size; small button is selected, press the Location; LT button to move the camera image.

Press the Size;Small button to display the size of the camera image.

Display size of the camera imagePress the Freeze button to stop the camera image only.

Hiding a camera imagePress the CCD Window button to temporarily hide the camera image.

Size;Small Size;Full Freeze

Location;LT

CCD Window

Camera image hidden

Camera image displayed

Camera image not frozen

Camera image frozen

Freeze

Location;RT

CCD Window

Menu;PinP

Size; SmallLocation;LTFreeze User Define

Mesurement

Body Mark CCD Window

* When a dual image display supporting probe is

connected, press Size;Small button to

display dual images(Size;Half).

* Each time the button is pressed, thecamera image moves around 4 corners on the examination window in turn.

LT- top left RT- top rightLB- bottom left RB- bottom right

37

Freezing the camera image

option

Q2E-EA1222

Page 42: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

4D (Real-time 3D)Preparation

Starting up the 4D software① Select a 3D-specific probe and an application.

② Press the F1 or F2 key to start the 4D software.

Press the F1 or F2 key again while the 4D software is active to exit.

Exiting the 4D software

Setting the 3D region

Extract a sagittal section of the fetus by using the 3D ROI Setting window in the figure on the right, and set the View Line and the 3D ROI.

The View Line is a useful function that lets you specify which surface to start with when observing a 3D image.

Adjust rotation

Adjust curvature

Adjusting the View Line

Set the View Line as close to the fetal nose and mouth as possible.

① Use the Angle Knob to adjust the rotation and curvature of the View Line .

② Use the trackball to adjust the position of the View Line.

Press the Angle Knob to change functions.

Select the View Line

Select the 3D ROI

Adjusting the 3D ROI

① Press the Trackball key to change the trackball function to 3D ROI.

② Adjust the position of the 3D ROI by using the trackball.

③ Press the ENTER key to enable size adjustments.

Press the ENTER key to change trackball functions.

Move the 3D ROI Resize the 3D ROI

38

If the 4D software is not started, select the 4D screenbutton to display the menu.

In case 4D Option is not exited, press 4D screen button.

3D ROI Setting window

option

Q2E-EA1222

Page 43: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

4D (Real-time 3D)Scanning

Setting the top of the 3DSelect which side of the window, right or left, is to display the head side of the fetus. A 3D image is displayed with the selected side as its top.

In the 3D ROI Setting window, select a screen button for the head side of the fetus.

Starting 4D scanningPress the UPDATE key, 4D scanning is started.

Press the UPDATE key. When the fetus has moved, returning to the 3D ROI Setting windowmakes position adjustments easy.

For a fetus age 20 weeks or less, adjustments are not usually needed.

You can also make adjustments by using the Thresholdscreen button displayed when the Image Menu key is pressed and by the Multi encoder.

Check that the B-Mode Knob priority is as shown in the figure on the left. Press the B Mode Knobto change functions.

Removing the artifact (noise) of a 3D imageTurn the B-mode Knob to the right while the real-time 3D image is being displayed to remove image signals of a certain brightness level or lower.

Press the Right key to change to the 3D single image display (3D window). 3D single image display

2D-3D window 3D ROI Setting window

2D-3D window 3D window

Switching between the 2D-3D window and the 3D ROI Setting window.

39

Top;L Top;R

Top;L

Top;R

option

Q2E-EA1222

Page 44: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

4D (Real-time 3D)Function

Press the Cine key and you can select a good image by using the trackball among the 3D cine images, immediately before the freeze.

Z directionX direction Y direction

Rotating a 4D (3D) imageUse the following Knob to rotate a 4D (3D) image:

How to find a good image among 3D cine images

Cutting out the unnecessary part of a 3D image① After the freeze, select the Cut Menu screen button to display the menu.

② Move the cursor, press the ENTER key to start tracing, specify the range of the unnecessary part by using the trackball, and then press the ENTER key to exit.

The volume data within the specified range will be cut out.

You can also remain the region surrounding the specified range by using the screen button.

How to reset the cut-out part Press the UNDO key to restore the previous state.

You can also restore the original state of the cut-out region by using a screen button.

Press the Cut Menu button again to close the menu related to cutting.

40

Cut Menu

Cut;Inside Cut;Outside

Cut Reset

Cut;Inside

Cut Menu

Cut Reset

Trace;ManualCut Menu

option

Q2E-EA1222

Page 45: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

4D (Real-time 3D)Volume Data Save and Load

Playing 3D Volume data

Saving 3D Volume dataPress the Save Volume screen button during the freeze.

You can save the currently displayed 3D Volume data.

① Press the Read key during the freeze to display the Filing window.

Hereafter, the procedure is explained based on thumbnails.

The 3D Volume data is marked with the icon on the left.

By using saved volume data, you can perform various operations such as 3D image quality adjustments, the cut function, and MPR display.

Press the Edit Volume screen button and press the Filing button.

Switching from the 3D Edit window to the Filing window

③ Press the Edit Volume screen button to edit or process the volume data.

② Select a volume data and select the Retrieve button.

41

Edit Volume

Save Volume

Edit Volume Filing

Edit Volume menu Filing menu

option

Q2E-EA1222

Page 46: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Inversion mode display

4D (Real-time 3D)HI DEF 3D (Scanning)

Image quality adjustment in the HI DEF 3D mode

HI DEF 3D is a 3D mode in which observation of still 3D images and MPR images with a higher definition than the real-time 3D mode is made possible through a slow single-swing.

Switching between the “Real-time 3D “ and ”HI DEF 3D”

【3D ROI Setting window 】

Real-time 3D mode HI DEF 3D mode

Swing angle adjustment in the HI DEF 3D mode

L High image quality

M

H High speed

From the screen buttons displayed, select Swing Angle and make adjustments by using the Multi encoder.

Starting HI DEF 3D scanningPress the HI DEF Scan screen button to start capturing.

Press the HI DEF Scan screen button while capturing (While the progress bar is displayed) to stop capturing and return to the 3D ROI Setting window.

Stopping HI DEF 3D scanning (cancel)

Before scanning During scanning

HI DEF 3D

【3D ROI Setting window】

LSwing Speed

45Swing Angle

HI DEF Scan HI DEF Scan

HI DEF Scan HI DEF Scan

HI DEF 3D

From the screen buttons displayed, select Swing Speed and make adjustments by using the Multi encoder.

From the screen buttons displayed, Select HI DEF 3D in the 3D ROI Setting window.

option

Select Rendering Menu and then Opac.Invert fromScreen buttons.

Rendering Menu Opac.Invert【Opacity mode】 【Inversion mode】

【MPR window】 【3D window】* Progress bar is displayed.

* When scanning is completed, 【3D window】, 【MPR window】, etc are displayed. Capturing can be started also by UPDATE key.

42 Q2E-EA1222

Page 47: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

4D (Real-time 3D)HI DEF 3D (Function)

You can rotate an image and save 3D Volume data in the same manner as the 4D mode.

Adjusting a cross-section on the MPR window

The MPR window's functions

Displaying a single cross-section on the MPR window

Press the Freeze key to return to the 3D ROI Setting window.

①Use the ENTER key to select a cross-section you want to change.(The selected view is enclosed in a solid frame.)

② Adjust the cross-section by using the trackball. Use the X, Y, and Z knobs to change the cross-section through 3D image rotation.

Switching between the MPR window and 3D Single Image windowPress the MPR screen button to perform switching.

Rescanning

Press the Left key to change the selected view into the single cross-section display.

Use the screen buttons to change cross-sections.

Press the Image Menu key and use the screen button and the Multi encoder to change.

Adjusting the brightness of an MPR cross-sectional image

MPR Display MPR Display

MPR_Z MPR_X

MPR_Y

2MPR Bright

option

MPR_X

43 Q2E-EA1222

Page 48: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

Option4D(Real-time 3D)

HI DEF 3D(Multi Slice mode)

Multi Slice modeSwitch of MPR window and Multi Slice window

Press screen button Multi Slice Display to switch the window.

Multi Slice DisplayMulti Slice Display

【MPR window】 【Multi Slice window】

Adjust the position of slice line with a track ball.

Switch of the section to observe (X, Y, Z section)

Single window display of Multi Slice window

Every time screen button Reference ;# is pushed, the section to observe is switched.

The section with blue color selected by Enter keyis displayed in single window.

Press Slice Setting screen button, and adjust by Slice interval screen button and Multi encoder.

Slice interval adjustment

Increase interval

Images can be rotated with each knob similarly to 4D.

2.0Slice IntervalSlice Setting

Reference;X

Slice position adjustment

Press Screen Shift screen button, and shift window by window with Multi encoder.

Slice window display shift

To + side

Unit of value is mm.

Screen;Multiple Screen;Single

Reference;Y Reference;Z

If track ball priority is different, make adjustment after selecting the right mark with TRACK BALL key.

Press ENTER key or TRACKBALL key to fix slice line.

Select the Slice Line

Assigned trackball function

If track ball priority is different, make adjustment after pressing TRACK BALL key to switch the function to image selection.

Select the Slice Line

Select the Image Select

Assignedtrackball function

Screen Shift

44 Q2E-EA1222

Page 49: HI VISION Avius User Manual - tdpmedical.com · Note that the ultrasound scanner display referred to in this User manual might differ from the display actually displayed. About the

②Adjust color gain, color box, velocity range (PRF), etc

4D(Real-time 3D)HI DEF 3D (CFM 3D mode)

Blood flow 3D image (CFM 3D mode)①Press CFI mode knob or Fine Flow/CFA key.

Before the start of scan During the scan

HI DEF Scan HI DEF Scan

Capturing can be started also by UPDATE key.③Press HI DEF Scan to start capturing.

Every time screen button is pressed, Fine Flow and CFA (Power mode) are switched.

CFA mode(High sensitivity

blood flow display)

Fine Flow mode(High definition

blood flow display)

To CFA To Fine Flow

Switch of B/W & blood flow 3D image and Blood flow 3D image

* Progress bar is displayed.

Press B mode knob.

Blend adjustment of B/W & Blood flow 3D image Press Image Menu key, select CFM 3D Blendto make adjustment.

Increase50

CFM 3D Blend

When the blend is high, Color 3D imageis displayed preferentially.

【3D ROI Setting Window】

B/W & blood flow 3D image Blood flow 3D image

Plane Cut;Both Plane Cut;BW

Press UNDO key to restore the cut plane.

Select the object to cut by screen button, and move trackball pressing ENTER key to cut the 3D image from a plane.

Plane Cut function

E

* When scanning is completed, 【3D window】, 【MPR window】, etc are displayed.

It is also possible to turn the color signal on and off similarly pushing Fine Flow/CFA mode key.