high frequency radiography & fluoroscopy x-ray … · 2020. 4. 21. · genware service soware...

2
Chest Stand Control Unit Table Centre Column Center Spot Film Device Tube Column Stand X- Ray Examination Room 400 cm 450 cm All wall of the Examination Room are 23 cm (9”) thick made of bricks. Single Leaf Door lined with 2.0 mm lead. X-Ray Tube Head MPB With Lead Glass viewing Window of 2.0 mm Lead equivalence Examination Table Unit Electronics SALIENT FEATURES GENUINE X-RAY & RADIOLOGICAL EQUIPMENTS PVT. LTD. Plot No. 3, Radhe Industrial Estate, Block No. 497 (P), Tajpur Road, Sarkhej - Bavla Road, Changodar, Ta. Sanand, Ahmedabad - 382 213. Gujarat, India Phone : +91 - 2717 - 298934 E-mail : Website : genuinexray@rediffmail.com [email protected] www. genuinexray.com REVOLUTION IN RADIOLOGY Autho. Dealer : All Machine Design / Technical Specificaons shown in catalogue are subject to change without any prior noce as standard procedure and constant improvement. CLINICAL OUTPUT ROOM Dimensions as per AERB Guidelines - L 4.5 x B 4.0 x H 2.75 meters. CONTROL PANEL GOVERNMENT OF INDIA AERB APPROVED IS:7620 PART-1 CM/L - 3950571 REVOLUTION IN RADIOLOGY Certified & Approved by : RADIOGRAPHY & FLUOROSCOPY X-RAY SYSTEMS 50 kW, 65 kW, 80 kW, 100 kW HIGH FREQUENCY Cost effective for all Radiographic applications (Hospitals, Medical Clinics, Chiropractic and Veterinary applications) Field upgradable for your Future Imaging requirements. Small, Compact & Light Weight. Anatomical Programs available in many languages that are easily edited by the Operator Membrane Control Console. Technique Selection : 1. kV with AEC 2. kV / mAs 3. kV / mA / Time 4. Patient / Body Part Thickness (Membrane Control Console) Low kV ripple - maximizes image quality by providing constant potential performance. Automatic Tube Calibration Service Software allows fast and easy set up as well as system diagnostics for off-site analysis Collimator Power Supply

Upload: others

Post on 26-Jan-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

  • Chest

    Stand

    ControlUnit

    Table Centre

    Column Center

    Spot Film Device

    Tube ColumnStand

    X- Ray Examination Room

    400 cm

    450 cm

    All wall of the Examination Roomare 23 cm (9”) thick made of bricks.

    Single Leaf Doorlined with 2.0 mm lead.

    X-RayTube Head

    MPB With Lead Glassviewing Window of 2.0mm Lead equivalence

    ExaminationTable

    UnitElectronics

    SALIENT FEATURES

    GENUINE X-RAY & RADIOLOGICALEQUIPMENTS PVT. LTD.

    Plot No. 3, Radhe Industrial Estate, Block No. 497 (P),Tajpur Road, Sarkhej - Bavla Road, Changodar, Ta. Sanand,Ahmedabad - 382 213. Gujarat, IndiaPhone : +91 - 2717 - 298934

    E-mail :

    Website :

    [email protected]@genuinexray.com

    www. genuinexray.com

    REVOLUTION IN RADIOLOGY

    Autho. Dealer :

    All Machine Design / Technical Specifica�ons shown in catalogue are subject to change without any prior no�ce as standard procedure and constant improvement.

    CLINICAL OUTPUT

    ROOM Dimensions as per AERB Guidelines - L 4.5 x B 4.0 x H 2.75 meters.

    CONTROL PANEL

    GOVERNMENT OF INDIA

    AERB APPROVEDIS:7620

    PART-1CM/L - 3950571 REVOLUTION IN RADIOLOGY

    Certified & Approved by :

    RADIOGRAPHY & FLUOROSCOPY X-RAY SYSTEMS50 kW, 65 kW, 80 kW, 100 kW

    HIGH FREQUENCY

    Cost effective for all Radiographic applications (Hospitals, Medical Clinics, Chiropractic and Veterinary applications)

    Field upgradable for your Future Imaging requirements.

    Small, Compact & Light Weight.

    Anatomical Programs available in many languages that are easily edited by the Operator Membrane Control Console.

    Technique Selection :

    1. kV with AEC

    2. kV / mAs

    3. kV / mA / Time

    4. Patient / Body Part Thickness

    (Membrane Control Console)

    Low kV ripple - maximizes image quality by providing constant potential performance.

    Automatic Tube Calibration

    Service Software allows fast and easy set up as well as system diagnostics for off-site analysis

    Collimator Power Supply

  • HF Generator Series offers advanced high quality Radiographic capabilities with unmatched performance and reliability resulting in the lowest cost of ownership over the product lifetime. This Generator is suitable for Film and CR - Based Radiographic systems and features leading edge micro-processor based controls to minimize patient doze and maximize image quality while demonstrating excellent reproducibility with user-friendly operator controls.

    GXL-HF 50 / GXL-HF 65 / GXL-HF 80 / GXL-HF 100

    LED MONITOR

    TECHNOLOGY

    -15 Degree

    RADIOGRAPHY & FLUOROSCOPY X-RAY SYSTEMS

    45 Degree 90 Degree

    HIGH FREQUENCY RADIOGRAPHY & FLUOROSCOPY X-RAY SYSTEMS

    POSITIONS

    All R & F Systems come with Floor to Ceiling Column Tube Stand fully counter balanced.

    Optional Ceiling Free Column Tube Stand available on demand. Control Unit Control Panel

    X - Ray Tube

    H.V. Cable

    OPTIONALX- RAY GENERATOR

    Vertical Bucky

    GENERAL SPECIFICATION 100 kW

    100 kW

    100 kW

    80 kW

    80 kW

    80 kW

    65 kW

    65 kW

    65 kW

    50 kW

    50 kW

    50 kW

    Generator Type

    Rotor Supply

    Input Phase / Voltage

    Power De-Ra�ng

    Compa�ble X-Ray Tubes

    Anatomical Programs (APR)

    Image Receptors

    Technique Selec�on

    Communica�on Ports

    Auxiliary Collimator Power Supply

    GenWare Service So�ware

    RADIOGRAPHY

    kVp Range

    kVp Accuracy

    High Voltage Ripple

    Rise Time (10% - 90%)

    mA Range / Steps (1 mA / 0.1 mA

    Steps op�onal)

    mA Accuracy

    Exposure Timer Range

    Exposure Timer Accuracy

    (measured at 75% points of kV waveform)

    mAs Range (non-AEC) / Steps

    mAs Accuracy

    FLUOROSCOPY

    kVp Range

    kVp Accuracy

    High Voltage Ripple

    mA Range / Steps

    mA Accuracy

    High Frequency Output (Maximum 400 kHz)

    Low Speed Starter (Op�onal Dual Speed Starter)

    Ф3 / 400 - 480 VAC

    300 tube models

    1024 / 20,000 + techniques

    Up to 6

    kV / AEC, kV / mAs, kV / mA / ms or kV / cm thickness (requires standard console)

    USB, CAN, Ethernet, RS485, RS232, RS422

    Standard Feature

    WEB - based diagnos�c and technical support

    40 - 150 kV in 1 kV increments

    ±2% for 90 - 110 kV; ±(5% + 1kV) for 40 - 150 kV