high-resolution mapping of scar-related atrial...
TRANSCRIPT
DOI: 10.1161/CIRCEP.114.002737
1
High-Resolution Mapping of Scar-Related Atrial Arrhythmias Using Smaller
Electrodes with Closer Interelectrode Spacing
Running title: Anter et al.; High resolution scar mapping
Elad Anter, MD; Cory M. Tschabrunn, CEPS; Mark E. Josephson, MD
Harvard-Thorndike Electrophysiology Institute, Cardiovascular Division,
Department of Medicine, Beth Israel Deaconess Medical Center &
Harvard Medical School, Boston, MA
Correspondence:
Elad Anter, MD
Harvard-Thorndike Electrophysiology Institute
Beth Israel Deaconess Medical Center
185 Pilgrim Rd, Baker 4
Boston, MA 02215
Tel: 617-632-9209
Fax: 617-632-7620
E-mail: [email protected]
Journal Subject Codes: [22] Ablation/ICD/surgery, [106] Electrophysiology, [124] Cardiovascular imaging agents/techniques
osephson, MD
H d Th dik El t h i l I tit t C di l Di i iHaaarvrvrvarararddd-T-T-Thoooornrrnr dike Electrophysiology IInsnsnstitute, Cardiovaaaascssscular Division,
Departmeeennt oooof MeMeMeMedidididicccineneee,,, BBBetth IIsrrraell DDeacacaca onnnneeess MeMM dididicaal CeCeCeC ntererer &&&
HaHaHaH rvararararddd d MeMMM dididid cacacacal lll ScScSchohohoololl,,, BoBoBoBostststs onononon, MAMAMAMA
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
2
Abstract:
Background - The resolution of mapping is influenced by electrode size and interelectrode
spacing. Smaller electrodes with closer interelectrode spacing may improve mapping resolution,
particularly in scar. The aims of this study were to establish normal electrogram criteria in the
atria for both 3.5mm electrode tip linear catheters (Thermocool®) and 1mm multielectrode-
mapping catheters (Pentaray®) and to compare their mapping resolution in scar-related atrial
arrhythmias.
Methods and Results - Normal voltage amplitude cutoffs for both catheters were validated in 10
patients with structurally normal atria. In 20 additional patients with scar-related atrial
arrhythmias, similar sequential mapping with both catheters was performed. Normal bipolar
voltage amplitude was similar between 3.5mm and 1mm electrode catheters with a 5th percentile
of 0.48mV and 0.52mV, respectively (p=0.65). In patients with scar-related atrial arrhythmias,
the total area of bipolar voltage <0.5mV measured using 1mm electrode catheters was smaller
than that measured using 3.5mm catheter (14.7cm2 vs 20.4cm2; p=0.02). The mean bipolar
voltage amplitude in this area of low voltage was significantly higher with 1mm electrode
catheters (0.28mV and 0.17mV, p=0.01). Importantly, 54.4% of all low voltage data points
recorded with 1mm electrode catheter had distinct electrograms that allowed annotation of local
activation time compared with only 21.4% with 3.5mm electrode tip catheters (p=0.01).
Overdrive pacing with capture of the tachycardia from within the area of low voltage was more
frequent with 1mm electrode catheters (66.7 vs 33.4; p=0.01).
Conclusions - Mapping with small closely spaced electrode catheters can improve mapping
resolution within areas of low voltage.
Key words: ablation, mapping, electrophysiology, arrhythmia
ororrrmemememed.d.d.d. NNNNorororormamamamal lll bbbbipopopopolalaaa
heters s ss wiwiwiwithththth aaaa 5555thth pepepepercr
m
ea of bipolar voltage <0.5mV measured using 1mm electrode catheters was smal
m r
m
0.28mV and 0.17mV, p=0.01). Importantly, 54.4% of all low voltage data points
with 1mm electrode catheter had distinct electrograms that allowed annotation of
annnnd d d 0.000 525222mVmmV, rererer spectively (p=0.65). In papapatit ents with scccar-rerererelall ted atrial arrhythm
eeeae oooof bipolar rr vovov ltagagagagee ee <0<0<0.5.5.5. mVmVmVV mmmeaasurereed uusiing g g g 1m1m1m1mmm m elelelleeectrtrrododo e cacacacaththththeteteterererrs wawawaasss s smsmsmsmal
measasasuruu ed usisisis ngg 33.5mmmm cccatatathehhh teeer rrr ((((14.777cmmm2 vs 20202020.4cmccmc 2; pp=p=p 0.0202). TTTThehhh mmmmeeeann bbipoolllar
mplitudududde in tttthihh s ararareaeaea of low www voltaggge wawawaw s sisisisigngngnnificananantltt y yy highgg ererere with 1m1m11mm mmm electrode
0.28m888 VV V andd d 000.17777mVVV, , p=pp 00.0 0001).).). IIImppportantlylyly,, 5545 44.44%%%% offff aallllllll lllowwww v lolltagegg ddddata aaa popopopoints
wwitith h 1m1mm m elelellecectrtrt ododdde e caacaththththeteterer hhhhaddada dddisiisi titititincnnct t t ellellecectrtrtt ogograaaramsms tthahhah t t tt alallllolol wewed dd d anannonotatat titionon oof f by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
3
Introduction
Atrial fibrillation (AF) ablation is an acceptable therapeutic option for patients with symptomatic
AF refractory to medications.1 Pulmonary vein isolation (PVI) is the cornerstone of the procedure
and is associated with a reasonable clinical outcome in patients with paroxysmal AF. However, in
patients with persistent AF, PVI is less effective and additional substrate ablation is frequently
performed.2, 3 This approach often results in the development of post-ablation, scar-related,
organized atrial tachyarrhythmias. 4-6
The mechanism of these arrhythmias is usually reentry involving preexisting or iatrogenic
ablation-related scar tissue.7, 8 These circuits are typically challenging to map because of significant
scar coupled with fractionated and multicomponent electrograms, limiting local time annotation. In
addition, entrainment and post-pacing interval mapping techniques may also be difficult to perform
and interpret due to high output pacing and/or lack of capture in these areas of low voltage.
The standard catheter for mapping these arrhythmias is a linear catheter with a 3.5mm
distal electrode separated by 2mm from a proximal 2mm electrode, resulting in a center-to-center
interelectrode spacing of 4.75mm. As such, each bipolar electrogram represents an underlying
tissue diameter ranging from 3.5mm to 7.5mm, depending on the angle of incidence (from
perpendicular to parallel to the tissue, respectively). Catheters with 1mm electrodes, 2mm
interelectrode spacing, and 3mm center-to-center interelectrode spacing record electrograms
from a significantly smaller underlying tissue diameter, ranging from 1mm to 4.0mm (also
dependent on catheter orientation relative to the surface). These catheters may have advantages
for mapping scar-related arrhythmias, including: 1) higher mapping resolution that can identify
heterogeneity within the area of low voltage, localizing “channels” of surviving bundles; 2)
smaller electrodes with closer interelectrode spacing are subjected to less signal averaging and
gg ppreexistingg or iatrtrrrogogogo e
map p bebbbecacaususee ofof ssigigigignnnini
ed with fractionated and multicomponent electrograms, limiting local time annotatio
n r
e
e m
rode separated b 2mm from a pro imal 2mm electrode res lting in a center to
ed wiwiiwithththth fffrararaactctctionanananated and multicomponent eeelelell cctrograms, limi itininini gg g local time annotatio
ntttrt aaaia nment andd pposttt-ppaciiingnn inttnteervavval mappppinngg teechchchc nininiiqquesss mmmayy aalsooo bbbe didiiifffffficculult tooo ppper
et dueeee tooto hhhhigigh hhh outttptp ttut pacinnngggg and/ororrr llllacacckkkk offf capaptututturererere inininin theheheh seseses areas offf f lololoow voltltltage.
e standddarddd catatathehhetet r ffof r mapppppping gg thhhese ar hhrhytytythmhh iiias iisi a llinininear rr ccca hththeter wiiti h hh aaa 3.33 5m
r dod ttedd bb 22 ff iim lal 22 lle tct dde ltltiin iin tte tto
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
4
cancellation effects, and may thus record higher bipolar voltage amplitude with shorter
electrogram duration, allowing more accurate time annotation; 3) pacing with capture at lower
output due to increased electrical density.
The aims of this study were to: 1) establish normal voltage amplitude cutoffs in the atria
for both 3.5mm electrode tip catheters and 1mm multielectrode-mapping catheters; 2) compare
their mapping resolution in scar-related organized atrial tachyarrhythmias (AT).
Methods
Patients
This study included 30 patients. In order to define normal bipolar electrograms in the atria
(voltage amplitude and electrogram duration), we studied 10 patients with paroxysmal AF
undergoing index PVI. In attempt to include only patients with structurally normal atria, eligible
patients were 65 year-old, had paroxysmal AF with arrhythmia duration
valvular disease, and all underwent cardiac magnetic resonance imaging (CMR) that
demonstrated non-scarred atria as evidenced by absence of late gadolinium enhancement. In
order to compare mapping resolution in scar, 20 additional patients with history of prior catheter
and / or surgical ablation and recurrent persistent atrial arrhythmias were also studied. Patients
were enrolled prospectively, however data was analyzed retrospective at the end of the study.
The institutional review board of Beth Israel Deaconess Medical Center approved the study
protocol and patients provided written informed consent prior to the clinical procedure.
Mapping Protocol
All procedures were performed under general anesthesia with the use of jet ventilation.
Electroanatomic mapping was performed with Carto 3 mapping system (Carto 3 ® system,
Biosense Webster). The linear mapping catheter was an open irrigated mapping/ablation catheter
included 30 patients. In order to define normal bipolar electrograms in the atria
m
g index PVI. In attempt to include only patients with structurally normal atria, eli
e
sease, and all underwent cardiac magnetic resonance imaging (CMR) th
incluuudededed d d 303030 ppatataatients. In order to define noonormrr al bipolar electctctctror grams in the atria
mmmpm lilililitude and eeeleectttrorrr grgrgrgraaaam m m m dddudurarararatiitionoon), wweee stududieeedd dd 010100 pppatattieeentnts s wwithhh ppparoxoxoxxysysysmamamaallll AFAFAFA
g indndnddexexexex PPPPVIVIVIVI. InInInI aaaattttttt ememememptpp ttto ooo inininnclclclcluddddeeee onononnlylylyy ppppaata ieieieentntntnts s wiwiwiwithhhh ssstrtrtrt ucucucuctututurararaalll y yy nonononormrmrmrmalalala atrtrtrriaiaiaa,,, elelele i
ere 6565656 yyyeaeaearrr-o-o-oldldldld,,,, hahahad dd papaparororoxyxxyxysmmmmalalal AAAFF F F wiwiiiththth aaarrrrrrr hyhyhyhythththhmimimiaa a a dudududurararatititiononoon
seseasase,e, aandnddd aallllllll uundndddere wewwentnt ccara didididiaacac maggagneetititit c c reesosonaaaancce e imiimi agagininii g g (C(C(C(CMRMRMRMR) )) ) ththththatat by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
5
(Thermocool®, Biosense Webster, Diamond Bar, CA – “linear”) and the multielectrode mapping
catheter was a 20-pole steerable mapping catheter arranged in 5 soft radiating spines covering a
diameter of 3.5cm (Pentaray®, Biosense Webster; Diamond Bar; interelectrode spacing 2-6-2mm
- “multielectrode-mapping ”). The linear catheter has a 3.5mm distal electrode separated by 2mm
from a 2mm proximal electrode, resulting in 4.75mm center-to-center spacing. The
multielectrode-mapping catheter has 1mm electrodes with interelectrode spacing of either 2 or 6
mm. For the purpose of this study, we recorded bipolar pairs with interelectrode spacing of 2mm
resulting in 3mm center-to-center interelectrode spacing. In order to examine the effect of
interelectrode spacing on mapping resolution in scar, a subset of 10 patients with atrial scar
underwent similar mapping with both the linear and the multielectrode-mapping catheters.
However, in addition, during mapping with the multielectrode-mapping catheter, electrogram
were recorded using both bipolar pairs separated by 2 and 6mm. All bipolar electrograms were
filtered between 30 and 500 Hz. Long sheaths were used to support both mapping catheters in
the right and left atrium.
Mapping of the substrate
Mapping of the chamber (right and left atrium) was performed in sinus rhythm. Initial mapping
was performed with the linear mapping catheter in point by point fashion. In all 10 patients with
structurally normal atria in whom normal values were established, the linear mapping catheter
was Thermocool Smart Touch® that allows force sensing and confirmation of tissue contact.
Electrograms were accepted only within a contact force range between 5-25 grams. Thermocool
Smart Touch® was also used in 11 of the 20 patients with scar-related ATs, using similar criteria.
The remaining 9 patients were mapped before contact force catheters were approved for
commercial use in the US, and therefore an ablation catheter without contact force sensing was
xamine the effect oooof f ff
atientttts s wiiiwithththth aattrtrt iiaiai l l scscscscara
i a
d w
w s
nd left atri m
simimiimilalalal rrr mamamam ppppinining with both the linear annnd ddd ththe multielectctc rodededed ---m- apping catheters.
innn n aaada dition, duuriing mmappipipiing wwithhh tthe mmmulttieelectctctrorororodedded -mmaaappingng ccacc ttht eteeree , elleectrogogogra
ded ususuusinininingggg boththth bbbbipipii lllolar paiaiaiirsrsrsrs separrrratatatateddeded bbby y 222 andddd 6m6m6m6 mmmm. AlAlAlA l l bibibipolar elelelelectrograms w
ween 333000 and dd 505055 000 0 HHHz. LLLonggg shehhh athhsh were useddd tto suppppppororrrt t bobobothththh mappppipipinggg ccccatatatatheheheheters
dd lleftft ttrii
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
6
used. To ensure detailed and uniform mapping of the entire chamber, we set the mapping filling
10mm (allowing interpolation between points to be as a requirement for a
complete map. In addition, each map made with either the linear or the multielectrode-mapping
catheters was registered with CMR whenever imaging was available. After completion of the
electroanatomical map with the linear catheter, a new shell of the chamber was created with the
multielectrode-mapping catheter
the multielectrode mapping catheter does not have contact force sensing technology, we used an
internal point filter software available on the mapping system to limit data acquisition to within
5mm from the original shell made with contract force technology. This algorithm, although not
an equivalent to tissue contact, can potentially reduce mapping of cavitary structures. In the
group of patients with scar-related ATs, similar sequential mapping with both catheters was
performed in sinus rhythm in order to characterize the substrate and particularly the zone of low
voltage and scar. Mapping of one or both atria was performed based on the electrocardiographic
characteristics of the AT.
Mapping of the arrhythmia
In the group of patients with scar related atrial arrhythmias, 14 patients had 26 ATs while 6 had
AF. We attempted to map all ATs with both catheters. Electrograms were selected for local time
annotation if their bipolar voltage amplitude was >0.06mV (2-fold higher than the noise level in
our lab) and had distinct and reproducible near-field potentials. Once activation maps were
completed and the putative circuit identified, overdrive pacing was performed in order to entrain
the tachycardia and further characterize the circuit. Entrainment pacing was performed at a cycle
length 10-25 msec faster than the tachycardia. The pacing site was considered part of the circuit
if the post-pacing interval measured from the stimulation artifact to the return atrial electrogram
data acqquisition ttoooo wiwww
s algogoririiiththththm,m aaltltltlthohooougugugughh
e h
a s
in sinus rhythm in order to characterize the substrate and particularly the zone o
d a
tics of the AT
ent t t t totototo ttisisisi susususuee coooonntnn act, can potentially redddduucu ee mapping off cavivivitary structures. In th
attttieeeents with sccarr-reelaated dd AATAA s, simmmilarr sssequeuentiaiaiaiall mammam ppppinnng wwitth bbbbooth ccacc thtt eteters wwwa
in sinininnususus rrhhyhyththththm iniii orddder tttoooo hhcharacacccteteteteriririze ttthhhe sububububststs rarararattetete aaandndndn ppartittt culalalalarlrlrlrly y thhhhe zone o
d scar. MMaM ppppppinininng gg ffof one or bbobb thhhh at iriia was pepp rfffformed dd babb seseeedd d onnn thehhh elelll ctrocacaaardrdrdr iiioi grgg a
ttiic fof tthhe AATT
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
7
on the ablation catheter was within 25 msec of the tachycardia cycle length (TCL). Pacing at
multiple sites was performed with both the linear and multielectrode-mapping catheters.
Comparison of pacing output threshold
Overdrive pacing of an arrhythmia is often performed in attempt to determine its mechanism
and/or localize its origin. However, pacing with capture is not always possible, particularly in
areas of low voltage and scar. In addition, pacing at high output can result in significant artifact
precluding its interpretation. As catheters with smaller electrodes and closer interelectrode
spacing have higher current density, we hypothesized they will have lower pacing threshold. We
compared pacing output threshold between the linear and the multielectrode-mapping catheters
during tachycardia. Pacing with the multielectrode-mapping catheter was performed after
completion of the activation map and at multiple atrial sites. Each pacing site was tagged on the
electroanatomical shell and pacing output threshold was recorded. Pacing was then repeated with
the linear catheter at similar anatomical sites and cycle length. Pacing sites were considered
similar if they were within 5mm from one another. Titration of pacing output was performed by
changing the pacing output strength within a range from 2.0 to 20 mA at a constant pulse
duration of 2.0 msec. Failure to capture was defined as lack of capture at 20mA at 2ms.
Radiofrequency ablation
Ablation of ATs was guided by activation and/or entrainment mapping. In reentrant circuits,
radiofrequency ablation was performed in attempt to disrupt the circuit with a line applied
between two anatomical barriers (i.e., mitral annulus to left inferior pulmonary vein in mitral
annular flutter) or at the earliest activation site of focal non-reentrant tachycardias. Ablation was
performed using the 3.5mm open irrigated catheter with energy of 20-40W. Energy was titrated
to achieve a 10-15 Ohms impedance decrease. If the AT terminated during radiofrequency
ower pap cingg thresshohohoholdl
ctrodededd -m-mapap iipipingngg cacaaaththththe
h
n n
t
c d
he ere ithin 5mm from one another Titration of pacing o tp t as performe
hyccccarararardidididiaa. PPPPaacinnngg gg with the multielectrode-m-m-maapping cathetee er wwwwas performed after
n offf f the activattioon mmmap ananand atttt muuulltiplle atrriaial siiiiteteeessss.... EEachchch paccinng sisisite wwwwasss tttagggededed on
tomiiicacaccallll shshshshell a ddndd ppacinii g ouououutptptput thrhrhrhreeese hohohoh ldddd was rerererecococordrdrdrdedddd. PPPa iicing wasasasas ttthen repe ttate
cathehhh ter at simimimiilili ar anatomicii al siiites andd d cyyy llcle lllenggg hthth. PaPaPaP cicicicinggg sssiiiti es were consnsnsnsidididdered
hhe itithihi 55m ffr thth TiTittr tatiio fof iin ttp tt ffo e
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
8
ablation, rapid atrial pacing was performed in attempt to re-induce the arrhythmia. If it remained
non-inducible, the ablation was considered successful. When linear ablation was used across an
isthmus, activation mapping and/or differential pacing on either side of the ablation line was
performed to confirm bidirectional block.
Electrogram measurement protocol
In order to define the lower normal voltage cutoffs in the healthy right and left atria, we analyzed
data from 10 control patients with structurally normal atria and determined the 5th percentile of
the bipolar voltages amplitude. In order to measure bipolar signal duration in normal atria with
both catheters, data was analyzed offline with electronic calipers on the Prucka CardioLab (GE
Healthcare, Waukesha, Wisconsin) recording system using uniform lead gain and paper speed of
200 mm/s. The local atrial signal duration was measured from the earliest initial deflection from
the isoelectric line to the time the last signal component returned to the isoelectric line. In order
to compare differences in electrograms characteristics between the catheters, we analyzed signals
for presence of fractionation (defined as signals with multiple intrinsic deflections), split
potentials (defined as electrograms separated by 30msec by an isoelectric interval), and late
potentials (defined as those displaying an additional signal separated from the local atrial
electrogram by 50msec).
Statistical analysis
Descriptive statistics are reported as mean ± SD for continuous variables and as absolute
frequencies and percentages for categorical variables. Normality of the continuous variables
distributions was assessed using one sample Kolmogorov-Smirnov test and comparisons between
the linear and multielectrode-mapping data were performed using the unpaired Student’s t test or
Mann-Whitney U test as appropriate. Categorical variables were compared using the Fisher’s
ation in normal attriririria a aa w
he Prrucuckkakak CCCCarardidididioLoLLLabababab
e
t o
e differences in electrograms characteristics between the catheters, we analyzed s
ce of fractionation (defined as signals ith m ltiple intrinsic deflections) split
, WaWaWaWaukukukukesessshahahaha, WiWWW sconsin) recording systeeemmm using uniformrr leaeaeaead gain and paper spe
ThThThThe local atriall sigggnnal dududuuratiiiiono waww s memm assurredddd fffrorororom mmm thhheee earrlieestttt iiinitiaalaa dddeefllecttttiooon
tric lililiinenene tttto thhhhe tititime ttthhhhe lassstttt iisisignalll ccccomomompoponent rrreteteteturrrnnnned ddd totototo thththt e isiii oelelelelectcctctric lililine. IIIn o
e diffffffef rences iiin nnn ellllectrogggrams chhhah ra tcteris iitics bbbbettween tthehehee catheheheetet rs, we anaaaalylylylyzezzz d s
fof ff tctiio titi ((ddefifi dd iig lls itithh ltltiiplle ii tnt iri iic dd feflle tctiio )) lplitit
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
9
exact test. A p value <0.05 was considered statistically significant. Analyses were conducted
using SPSS Statistics 20.0 (Chicago, Illinois).
Results
Patient Characteristics and Previous Ablations
Patient characteristics are depicted in table 1. In the derivation group of 10 patients with
structurally normal atria, the mean age was 55±9 years and the mean left atrial diameter was
40±6mm. The mean ejection fraction was 60±10%. All patients had CMR without evidence late
gadolinium enhancement (indicative of scar) in the atria. In the group of patients with scar-
related ATs, 16 had prior percutaneous ablation procedures (median of 2 prior procedures per
patient) and 4 additional patients had history of surgical maze. Six patients had previously
undergone mitral valve replacement surgery. The recurrent arrhythmia was persistent ATs in 14
patients and persistent AF in 6 patients. The mean time from the prior ablation or surgical
procedure to the “current” ablation procedure was 13.3±9.2 months (range, 3 to 23 months).
Mapping density
In the derivation group of 10 patients with structurally normal atria, the number of electrograms
acquired with the multielectrode-mapping catheter was significantly greater than that obtained
with the linear catheter (right atrium 1286±336 [median 1220] vs. 462±216 [median 446],
p<0.01; left atrium 1554±374 [median 1335] vs. 366 ±273 [median 352], p<0.01). The mapping
time was shorter using the multielectrode-mapping catheter (13±7min [median 12] vs. 21±
12mm [median 20], p=0.03). This is likely because each “beat” acquired with the multielectrode-
mapping catheter can record up to 10 overlapping bipolar electrograms compared with only 1
bipolar electrogram recorded with a linear catheter. Similarly, in the group of patients with scar-
related atrial arrhythmias, mapping density was higher with the multielectrode-mapping catheter
off paaaatitititienenenentstststs wwwwitititith h h h scscscscarararar--
s, 16 had prior percutaneous ablation procedures (median of 2 prior procedures p
d
mitral valve replacement surgery. The recurrent arrhythmia was persistent ATs
d
to the “current” p .3±9.2 months (range, 3 to 23 months)
s, 16 hahahad d d prprprior rrr pepp rcutaneous ablation procccedee ures (median oooof ff f 2 prior procedures p
dddd 44 additionaaal lll ppatititienenenentstststs hhhhadadadad hhhhisisisistototot ryrry of fff ssurggiccal mmmazzzzee.e SSSSixxx pppatatatieentstststs hhhhadadadd pppprereeviviviououououslslsly yyy
mimimitrtrtrtralalal vvvalalala vevvv rrrrepepepplalalalacecececememmm ntntntnt sssururururggeryryryy. ThThThThe e e e rererer cucuuurrrrrrrrenenene t t t araara rhrhrhrhytytyty hmhmhmhmiiai wwwwasasas pppperererersisisis sttttenenentt t t ATATATATs sss
d peeeersrssisisisteeentntnt AAAFF FF inininin 666 pppatatatieieientntntntss.s TTTThehehe mmmmeaeaeae n tititimememee ffffrororoommmm thththhee e e prprprprioioior rr abababa lalalalatititionononn ooor rr sususus rgrgrggiciciccal
toto tthee “cuucurrr ene t”t”t”” abbabblallal titttiono prp ococeddede uru e e wawwas s 1313133.33.3±9±9±9±9.22.22 mono thhthhs s ((r(r( ana gege,, 3333 totot 22223333 moontthss)) by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
10
(1858±975 [median 1766] vs. 472±171 [median 452], p<0.01) despite shorter mapping times
(15±6min [median 15] vs. 23± 12min [median 22], p=0.02).
Normal cutoff values
Normal voltage cutoffs for each catheter were established in 10 patients with structurally normal
atria during sinus rhythm. In the right atrium, a total of 4382 electrograms were recorded with
the linear catheter and a total of 11,956 with the multielectrode-mapping catheter. The mean
bipolar electrogram amplitude was similar between the linear and the multielectrode-mapping
catheters (2.6±1.8mV vs. 2.9±2.4mV; p=0.76). The fifth percentile of the normal bipolar voltage
distribution was also similar between the linear and multielectrode-mapping catheters (0.48mV
and 0.52mV, respectively; p=0.68). In the left atrium, a total of 3852 electrograms were recorded
with the linear catheter and a total 16,435 electrograms with the multielectrode-mapping
catheter. The mean bipolar electrogram amplitude recorded was similar between the linear and
the multielectrode-mapping catheters (2.6±1.6mV vs. 2.8±3.2mV; p=0.84). The fifth percentile
of normal bipolar voltage distribution was also similar between the linear and multielectrode-
mapping catheters (0.50mV and 0.52mV, respectively; p=0.80). Using this data, we defined the
catheters.
Electrogram duration during sinus rhythm was shorter with the multielectrode-mapping
catheter. When mapping with the multielectrode-mapping catheter, the mean bipolar electrogram
duration was 32msec +/- 22 (median 30msec; range 18-54ms) and 95% of all electrograms had
duration <46ms. In comparison, when mapping the linear catheter, the mean bipolar electrogram
duration was 44msec +/-24 (median 42msec; range 26-68ms) and 95% of all electrograms had a
duration <58msec.
the normal bippolarararar vvvvo
appingngg cacaththththettteterers s (0(0(0(0.444.488
V c
n
he mean bipolar electrogram amplitude recorded was similar between the linear
e n
bipolar oltage distrib tion as also similar bet een the linear and m ltielectrod
V, rerererespspsps ececcctitititivvelylyly;;;; p=0.68). In the left atriuuuum,m,m a total of 388852 eeeelell ctrograms were rec
neaeaeaearrr r catheter aandd a tootal 11116,433335555 eeeleectrrooograamms wiwwiw thththh thhe mmmultiielleccctrrode-m-m-mappppinggg
he memememeananan bbbbippolllar elelll ttctrogrgramamamam amppplilililitututudededed recordededededdd d wawawawas sisisisimimimim lall r bebbb twtwwweeeeeee n ththhhe lililil near
ectrodedd -mappppipipiinggg cathhheters (2(2(2.6666 1±1±1 66.6mV vs. 2222 8.8±3±3±3± 22.2mVmVVV;;; ; p=pp 000.84848484).).).) TTTThhehh ffffifififi thththth pppeeere cen
bibi lla ltlt didi tst iribb titi lls isi imilla bbett thth lili dnd lltiti lel ttr dod
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
11
Low voltage (“scar”) mapping
We compared mapping resolution between the two catheters in low voltage and scar tissue
during sinus rhythm. In patients with scar-related atrial arrhythmias, the total area of low bipolar
voltage (defined as <0.5mV) measured using the multielectrode-mapping catheter was
significantly smaller than that measured using the linear catheter (14.7cm2 [median 15.2; range
7-48cm2] vs. 20.4cm2 [median 22.4; range 9-55cm2]; p=0.02). In addition to smaller low voltage
“scar” area, the bipolar voltage distribution within this area of low voltage was higher using
multielectrode-mapping catheters with a mean bipolar voltage of 0.28+/- 1.2mV compared with
0.17+/- 1.0mV measured with the linear catheter (p=0.01). This resulted in improved delineation
of the low voltage zones (Table 2 and Figures 1 and 2).
Late potentials within the area of low voltage (<0.5mV) were more frequently recorded
with the multielectrode-mapping catheter (28.2+/-14.5% vs. 13.6+/-7.8% of total low voltage
signals; p=0.01). Electrogram fractionation within the area of low voltage was also more
frequently recorded with multielectrode-mapping catheters (68.8+/-22.6% vs. 38.6+/-12.4% of
total low voltage signals; p<0.01). These signals were essentially seen only in the group of
patients with scarred atria. The frequency of split potentials was similar between the catheters.
Determinants of mapping resolution within scar
We examined the individual contribution of 1) mapping density; 2) electrode size; and 3)
interelectrode spacing on catheter mapping resolution within the zone of low voltage.
In order to examine the relationship between mapping density and mapping resolution,
each pair of maps was reanalyzed and density adjusted by excluding data points from the map of
greater original density in order to achieve similar data density with homogenous point
distribution. In order to minimize bias, all data points were exported to a Matlab platform and
+/- 1.2mV comppararrrededede
ed inn iiiimpmprorovevedd dd dedededelilililinne
v
r
u a
recorded ith m ltielectrode mapping catheters (68 8+/ 22 6% s 38 6+/ 12 4%
vollltatatatagegegeg zzzzononono esss ((((TaTT ble 2 and Figures 1 annnd ddd 2)222 .
eeee ppopop tentials wwitthinnn ttthhhe aaarrrea ofofoff lowoow vooltttagee (<0000.5.5.5mVmmVm ) weww re mmorrre freqqqqueueueu nntlyly recececor
ultielelelelecttctrrororode-mapappipiing caththththetteteter (28282828 222.2r +/+/+/+/ 11-14.44 5%5%5%5% vvvvs. 11133.33 6666+/+/+/+/ 7-777 88.8%%% % ofofoff tttotoototal llllow v lllolttta
0.0111).).). ElElElectrogogoggram frff actiiionatioi n iwii hththin thhhe area offf loww vollltatatagegegeg was alllso momommorrrer
ddedd iithth lltiti lel ttr dod iin thth tet ((6868 88+/+/ 2222 6%6% 3388 6+6+// 1212 44%%
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
12
distances between points were calculated based on the x, y and z coordinates. Exclusion of points
was performed blindly based on data point coordinates and aimed to create a similar point data
density between maps. The resultant mean mapping density was 12±3mm between points with a
bipolar voltage <0.5mV measured with the multielectrode-mapping catheter remained
significantly smaller than that measured with linear catheter (15.4cm2 [median 12.8; range 7-
32.6cm2] vs. 22.4cm2 [median 21.8; range11-44cm2]; p=0.03). The mean bipolar voltage
amplitude within this area of low voltage remained higher when using the multielectrode-
mapping catheter (0.26+/-0.12mV vs. 0.17+/-0.09mV; p=0.01). This data suggests that mapping
density is not a significant determinant in the differential resolution between the catheters.
In order to examine the relationship between electrode size and bipolar voltage
amplitude, we first attempted to control for the other two variables, mapping density and
interelectrode spacing. Mapping density was controlled as described above with all paired maps
having similar mapping density. As the catheters do not have similar interelectrode spacing, we
reanalyzed maps made with the multielectrode-mapping catheter to only include data from
bipolar electrode pairs separated by 6mm rather than 2mm. In this configuration, the relationship
of interelectrode spacing between the catheters had reversed with the multielectrode-mapping
catheter having a larger center-to-center interelectrode spacing compared with the linear catheter
(7 vs. 4.75mm). The total area of bipolar voltage <0.5mV measured with the multielectrode-
mapping catheter remained smaller than that measured with the linear catheter (18.6cm2 [median
15.2; range 9-24.8cm2] vs. 22.4cm2 [median 23.1; range 11-36.6cm2]; p=0.04). The mean bipolar
voltage amplitude recorded with the multielectrode-mapping catheter remained higher (0.22+/-
0.10mV vs. 0.17+/-0.09mV; p=0.04). This data suggests that smaller size electrodes are at least
the multielectrodededee---
data susuggggesesttstts ttthahahh t t mammmap
n
o
ode spacing. Mapping density was controlled as described above with all paired m
milar mapping densit As the catheters do not ha e similar interelectrode spacing
nottt aaaa ssssigigigninininifififificaantntntnt determinant in the differrrenene tial resolutiooon bebebettwt een the catheters.
orrrrdeeeer to examiine thehehe relatatationsshhhih ppp bbetwweeenn eeleccctrtrtrrodododdee siiizeee annd bipipippoolo ar vvvvoooltttagge
we fifififirsrrsttt attemptttted ddd ttto contrtrtrtrollolol for ttthehehehe oooththther tttwowo vvvvarriaiaaiablblblblesesess, mapping g gg dedededensititity anddd
ode spapp iciinggg. MMMaM ppppppinii g gg dedd nsiiti y yy was controllllll deddd as deddd sc iiribebeeed dd abababooovo e iiwii hthh alllll ppppaiaiaia rrrer d m
milil ipi dd isitt AAs tthhe tathhette dd tt hha iimiilla iintte lle tct dde ici
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
13
partially responsible for the differential resolution between the catheters.
Lastly, we examined the relationship between interelectrode spacing and bipolar voltage
amplitude. A subset of 10 patients with scar-related atrial arrhythmias underwent mapping with
the multielectrode-mapping catheter recording both bipolar electrode pairs separated by 2 and
6mm. This data was then separated into two datasets, one containing only data recorded using
bipolar electrodes separated by 2mm and a second containing only data recorded using bipolar
electrode separated by 6mm. The total low voltage area and mean bipolar voltage amplitude was
calculated for each separate map. The total area of bipolar voltage <0.5mV measured with
interelectrode spacing of 2mm was smaller than that measured with interelectrode spacing of
6mm catheter (15.4cm2 [median 14.0; range 8-23.5cm2] vs. 19.2cm2 [median 18.6; range 10-
28.9cm2]; p=0.03). The mean bipolar voltage amplitude within this area of low voltage was
higher when mapping with interelectrode spacing of 2mm (0.26+/-0.11mV vs. 0.22+/-0.16mV;
p=0.03). This data suggest that interelectrode spacing is a determinant of mapping resolution
within scar tissue.
Arrhythmia Mapping
Twenty patients with scar-related atrial arrhythmias had 26 ATs, however only 23 had TCL
200ms (mean 274±56ms) that allowed mapping. Five ATs were not mapped because they
stopped spontaneously before mapping and did not recur. Activation mapping was performed
with the multielectrode-mapping catheter in all 18 sustained ATs and with the linear catheter in
13 of the 18 (72.2%). In effort to limit bias, the sequence of mapping alternated between the
catheters, with the linear catheter used first in 8 patients and the multielectrode-mapping catheter
used first in 5 patients. The mechanism of the tachycardia was defined as macroreentry or focal
microreentry on the basis of the activation map. In macroreentrant atrial tachycardias, the circuits
5mV measured wwitititith h hh
terelelel tctctrorodededd sspapapp cicicicingngngng o
e 2 2 2 0
p s
e 6
h o
r tiss e
eterrrr (((1515151 4.44cmcmcmc 2 [m[m[m[median 14.0; range 8-23.555cmcmc 2] vs. 19.2cmmm2 [m[m[m[median 18.6; range 10
pp=0=0=0=0.03). The memem annn bbbipooolalalaar voollltl agaagee ammmpplittude wiwiwiw thththhinn ttthihihis arreaa ooof ff low vovoolttaage waww s
en maaaappppppininining g wiiiththth iii ttnterelectrtrtrtroddodode sppacacaca iniining offf 2222mmmmmmm ((((00.00 2626262 +/+/+/ 0-00.11111m1 V VV vvvsss.s 0.2222222+/+/+/+ -000 11.116
his ddad ta suggggegegesssts tthahh t iini terellell ctrodedd spapp ciiinggg iiis a ddded termininnnant ofofofo mappppppinii g gg rerereresososos lllul tio
titi
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
14
traversed large portions of the atrium while in focal atrial tachycardias a point source with
centrifugal activation from this center was identified. The mechanism was macroreentry in 16 of
18 (88.9%) and focal in 2 of 18 (11.1%). The most common atrial tachycardia was a
macroreentrant circuit involving the mitral isthmus (8 of 16; 50%) followed by macroreentrant
circuit involving the left atrial roof (4 of 16; 25%). The third most common tachycardia (3 of 18;
17%) involved the septum, and there was 1 right-sided atrial tachycardia.
Activation mapping with the multielectrode-mapping catheter included a mean of
1172±346 points (median 1220) with an average mapping time of 13±8 minutes (median 14).
The activation map constituted an average of 92±8% of the TCL. All circuits involved the area
of low voltage as defined with the multielectrode-mapping catheter during sinus rhythm. In 12 of
the 16 macroreentrant ATs mapped with the multielectrode-mapping catheter, activation
mapping was also performed using the linear catheter. The mean number of activation points per
map was smaller (241±73; median 228; p=0.02) despite similar mapping time of 15±11 minutes
(median 15; p=0.72). In comparison to mapping with the multielectrode-mapping catheter,
activation maps constituted an average of 68±32% of the TCL (p=0.01). In 4 ATs that occurred
in patients with limited atrial scar, mapping with either the linear or the multielectrode-mapping
catheter resulted in nearly complete activation map constituting >90% of the TCL. However, in 8
ATs that occurred in patients with extensive atrial scarring, mapping with the linear catheter was
associated with limited activation mapping constituting only 53±11% of the TCL compared with
90±8% when the same AT was mapped with the multielectrode-mapping catheter (Figure 3 and
4). In these cases, mapping with the linear catheter demonstrated significantly lower mean
voltage amplitude that was coupled with highly fractionated and non-distinctive electrograms,
limiting accurate activation time annotation (Figure 5). Overall, 54.4% of all low voltage data
8 minutes ((medianannn 114
circuiuiiitststt iiiinvnvololllveved d ththththeeee a
t n
c
was also performed using the linear catheter. The mean number of activation poin
m n
5; p 0 72) In comparison to mapping ith the m ltielectrode mapping catheter
tagegegege aaaass dededeefifififinned dd wiww th the multielectrode-mmmapapping catheteeer dududurrring sinus rhythm. In
rrror rrrer entrant ATTss mammapppeddedd witttthhh h thhhe muultttielecctrooodedede-m-m-mmapppping caathhhhettter, acacactitt vavationnn
was allllsososso pppperformedddd u iising ttthehehehe lllineaarrrr cacacaththththettter. ThThThTheeee memememean nnnnumbbber offff aaaactcctctiivatttiiion poiinii
malllllel r (2(2(2414141±7±7±77333;3 medddiiai n 222222288;8 ppp 00=0.0002222) ) ) despppiiti e siiii iimiillal r mamappppppinininng g g g ititime offf 15151515±1±1±1±11111 min
55 00 7272)) II iis tt iin iithth tthhe ltltiielle tct dde ipi thth tet
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
15
points recorded with multielectrode-mapping catheters had distinct electrograms that allowed
annotation of local activation time compared with only 21.4% of all low voltage points recorded
with linear catheters (p=0.02). Ablation was successful in 15 of the 16 macroreentrant circuits
and in the 2 focal arrhythmias. In 1 patient with macroreentrant circuit involving the interatrial
septum ablation was unsuccessful.
Atrial Pacing Threshold
We compared the pacing threshold during tachycardia between the linear and multielectrode-
mapping catheters at similar atrial sites. In 13 patients in whom the arrhythmia was mapped with
both catheters, overdrive pacing of the tachycardia was performed with both the linear and the
multielectrode-mapping catheters (n=54 and 47, respectively). Pacing with capture was more
frequent with multielectrode catheters (66.7% vs 33.4%; p=0.01). Moreover, the pacing
threshold was lower using multielectrode-mapping catheters as 57.7% of pacing sites had
threshold compared with only 25.5% when pacing from a linear catheter (p=0.02).
Discussion
Major findings
This study established the normal bipolar voltage distributions in the atria for both a linear
catheter with 3.5mm distal electrode and 1mm multielectrode-mapping catheter. In addition, this
study compared the mapping resolution in low voltage and scar with both catheters.
The major findings are: 1) bipolar voltage amplitude in healthy atria is similar between
3.5mm and 1mm electrode catheters with a 5th percentile of 0.48mV and 0.52mV, respectively;
2) mapping resolution within areas of low voltage and scar is enhanced with 1mm electrode
catheters; 3) electrode size and interelectrode spacing were major determinants of mapping
resolution within areas of low voltage and scar; 4) lastly, arrhythmia mapping using activation
rrhyty hmia was mapappppepepp
h botottth hhh thththhee lilililineneaar aaaandndndnd
o
i
w
=
odedede-m-m-mapapappipipipinng cacacacatheters (n=54 and 47, reeespspspeectively). Pacacacing ggg wwwith capture was mo
itttth multielectrrode caathetetteteeers (6(6(6(66.777%% vvs 33.44%;;;; pppp=0=0=00.01). Mooreeovvveer, thhhhe papaccinggg
was lololooweewewerr usiniii g multltltieii lectctcttrorororodddde-mmapapapappiipipingng caththththetetteterererers aasasas 5555777.7 7%7%7%% of fff paacicicicinngngn sitititi es hhhh dddad
cccoc mpppareddd wiititi h onlylyly 2222555.5 5%%%% whehhh n papp cinggg ffffrom mm aaa a lilililinear caththththeteteteterererer (((p=pprr
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
16
and/or overdrive pacing techniques was more accurate using 1mm electrode catheters.
Three-dimensional (3D) electroanatomic mapping systems have been developed to assist
with mapping and ablation of cardiac arrhythmias. These systems have become an essential tool
for mapping complex arrhythmias and are frequently used to assess the underlying substrate for
presence of low voltage areas, so called “scar”. Accurate detection and characterization of atrial
scar with electroanatomic mapping is essential for catheter ablation of atrial arrhythmias. Normal
electrographic criteria in the ventricle were defined by Cassidy and Marchlinski,9, 10 however
limited data exist regarding the normal voltage distribution in the atria. Kapa et al have defined
the normal bipolar voltage distribution in the left atrium using 3.5mm Thermocool® catheter. In
10 patients with paroxysmal AF, the mean bipolar voltage amplitude was 1.44±1.27mV, and
95% of all points demonstrated a bipolar voltage amplitude >0.45mV. Our study confirms this
observation and extend its finding also to the right atrium. We found that 95% of all
electrograms in the right and left atria had a bipolar voltage amplitude >0.48mV. In addition, we
measured the distribution of electrograms duration in healthy atria and found that 95% of all
electrograms had a duration < 58ms.
Iatrogenic AT are unfortunately common after AF ablation, particularly in patients with
persistent AF where additional ablation sets often include linear lesions and ablation at sites of
complex fractionated electrograms. The mechanism of these ATs is typically macroreentry but
also sometimes microreentry involving preexisting or iatrogenic ablation-related scar tissue.
These arrhythmias can be challenging to map using standard linear catheters because of significant
scar coupled with fractionated and multicomponent electrograms, limiting activation and
entrainment mapping. Patel and colleagues have reported the feasibility to rapidly map and
successfully ablate these arrhythmias using the Pentaray® multielectrode-mapping catheter in post-
a. Kappa et al havee ddddefefee
Thermrmococoooollll® ® ®® cacaththththetetetete
with paroxysmal AF, the mean bipolar voltage amplitude was 1.44±1.27mV, an
points demonstrated a b olar voltage amplitude >0.45mV. Our study confirms t
n
ms in the right and left atria had a bipolar voltage amplitude >0.48mV. In additio
he distrib tion of electrograms d ration in health atria and fo nd that 95% of a
wiwiwiththhth pppararrroxoxoxo ysssmmamm l AF, the mean bipolar vvvololtage amplituuude wwwwas 1.44±1.27mV, an
ppppoiiiints demonsttratededed a bbbipipipipolarar volooltagge ampmplituuudedede >>>>00.44545mVV. OuOuuur studdddy coonnfirmsmm t
n anddd d exexextetetete dnd iiittts ffffinii diididing alslslssooo to thehehehe rrigiigighhhth atttriiiui m.mmm WWWWeeee ffounununund ddd thththat 995%5%5%5% of allllll
ms iiiin hhthe righghght tt anddd d llleffft atriaiii hadddd a bbbipipipolar v lollltat gegg ampppliitutututudedd >>>>0.00 4848484 mVVV. InInInIn aaaadddddddditio
hhe ddiisttribib ttiio fof lle tct dd titi ii hhe lalthth tat iri dd ffo dd thth tat 995%5% ff a
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
17
PVI patients with ATs. 11
In this study, we established the normal atrial electrogram characteristics of 2 commonly
used catheters: a standard mapping/ablation catheter (Thermocool®) and a multielectrode-mapping
catheter (Pentaray®). While the resolution of mapping between the catheters was similar in healthy
atrial tissue, they significantly diverge in low voltage and scar tissue. The resolution of electrical
mapping is influenced by multiple parameters including electrode size, interelectrode spacing, angle
of incidence (catheter orientation relative to the surface), the vector of wave propagation, and
filtering. Their combined effect is associated with significant variations in the recorded bipolar
voltage amplitude at any single recording point. While these changes may not be as clinically
important in the normal healthy tissue whereas a bipolar voltage amplitude variation between 0.5-
4.0mV represents normal tissue, such variations in areas of low voltage and scar are significant and
determine feasibility to identify surviving myocardial channels and “isthmuses” often present in
zones of low voltage. In this study we found that smaller electrodes with closer interelectrode
spacing provide higher resolution in scar mapping. Maps made with multielectrode-mapping
catheters usually contained more data points, and points acquired with increased variability in the
angle of incidence and the relationship to the vector of propagation. They may thus be less
subjective to the individual confounding effects of bipolar voltage amplitude measurement. We
found that mapping density alone is not responsible for the differences in mapping resolution within
scar and that smaller electrodes and closer interelectrode spacing are significant determinants of
mapping resolution. Smaller electrodes have smaller electrical fields of view and smaller antenna,
and as such are subjected to less averaging and cancellation effects. A bipolar electrode pair with
closer interelectrode spacing record data from smaller tissue diameters and is therefore more
sensitive to detect surviving myocardial fibers in zones of generally low voltage “channels”. It also
in the recorded bipipppololololaa
ay notott bbbbee asas cclilililininicaaaallllllllyy
n the normal healthy tissue whereas a bipolar voltage amplitude variation between
r n
feasibility to identify surviving myocardial channels and “isthmuses” often present
w
o ide higher resol tion in scar mapping Maps made ith m ltielectrode mapping
n tthehehe nnnnororormamamm l hehehealaa thy tissue whereas a bippppololololaar voltage ampmm lituuuuddde variation between
reeese eeene ts normall tiissueuue, suuuchchchh varriiiai tiiioons ininn areeaas oooff ff lololow ww w voooltttage anand ddd sscar aaraa ee e ssiggniffficccanaa
feasibibibbillililititityyy to iiiideddd tttntififfify survivivivivinining myyocococo arararddidid alll chhhah nnnnnnnn lels ss aaaand ddd “i“i“i“ sttthmhh usessss””” ofo ten presentt t
w v lolltagegg . InII ttthihhih s st duddy yy we fffounddd thahh tt sm llallell r lelecttr doddess wwwwitii h hh clccc oser iiiinterelelelelecececectrtrrtroode
idid hihi hgh lol ttiio iin iin MM dde itithh ltltiielle tct dde iin
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
18
has a more distinct electrogram with shorter electrogram duration that allows more accurate
annotation of activation time. Lastly, the pacing output threshold within the zone of low voltage and
scar is lower with small electrode catheters, likely due to the increased electrical “current” density at
the electrode-tissue interface.
In addition, pacing from a multielectrode-mapping catheter placed in or around a
macroreentrant circuit can be used to determine the mechanism of the tachycardia. If pacing
performed during AT from a site that demonstrates later activation relative to its neighbors results in
orthodromic capture of these earlier “upstream” neighboring sites, the mechanism must be reentry.
Barbhaiya and colleagues have recently reported the utility of such approach to define the
mechanism of atrial tachycardias. 12
Study limitations
This study includes a relatively small cohort of patients. While mapping data with the linear catheter
was confirmed by contact force measurement, such data was not available for maps made with the
multielectrode-mapping catheter. However, although differences in tissue contact can affect bipolar
voltage amplitude, it is unlikely to account for the differences between the catheters. First, all maps
performed with multielectrode-mapping catheters had similar volumes compared to maps made
with linear catheters. Second, points were only accepted if they were within 5mm from the original
shell made with the linear catheter. Lastly, in the derivation cohort of patients with normal atria, the
bipolar voltage amplitude was similar between the catheters. In patients with ATs, only 13 of the 18
ATs mapped with the multielectrode-mapping catheter were also mapped with the linear
catheter. Although the number of ATs mapped with both catheters was small, the data was
highly consistent. Electrogram duration was measured using fixed electrogram gain; however
data acquisition was obtained using a standard amplifier with variable gain. The clinical impact
mechanism must be e rerereree
oachh tttto o dddedefifififinene tttthhhehe
m 12
t
i
med by contact force measurement, such data was not available for maps made with
ode mapping catheter Ho e er altho gh differences in tiss e contact can affect b
m offf atatatatririririalalall tatatatachcc ycycycycardias. 12
taaaatiiiions
incluludedededesss a relalll tititivelylyll smallllllll ccccohhohohort ofofofof ppatattatiiiei ttnts. WWWWhihihihillele mmmmappppipipip ngng ddddata wiwiwiwithththth thhhe lililinear c
med ddd bybyby contactctctt fforce measurement, such dddata was not avvaiaiaia lablblble eee fofff r mapspp mmmmadadadade eee with
dde iin thth tet HHo altlthho hh didiffff iin ttiis tnt tt fafffe tct bb
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
19
of mapping with small electrode catheters on long-term clinical outcome is outside the scope of
this study and remains unanswered. Finally, it may be economically challenging to use the
Pentaray catheter in addition to a circular mapping catheter commonly used during routine AF
ablation procedures. However, in our experience, we have been able to perform the entire
ablation procedure using just the Pentaray® catheter instead of the circular catheter, including
creating the left atrial geometry and verifying entrance and exit blocks.
Conclusions
This study established the normal bipolar voltage criteria in the atria for both a standard linear
mapping/ablation catheter with 3.5mm distal electrode (Thermocool®) and a multielectrode-
mapping catheter with 1mm electrodes (Pentaray®). In addition, this study showed that catheters
with smaller electrodes and closer interelectrode spacing have significant advantage in mapping
atrial scar. This technique can facilitate successful ablation of scar-related atrial tachycardias.
Conflict of Interest Disclosures: Elad Anter receives research grants and speaking honoraria
from Biosense Webster and Boston Scientific. Mark E. Josephson receives research grants and
speaking honoraria from Medtronic. Cory M. Tschabrunn has no relevant disclosures.
References:
1. Calkins H, Kuck KH, Cappato R, Brugada J, Camm AJ, Chen SA, Crijns HJ, Damiano RJ, Jr., Davies DW, DiMarco J, Edgerton J, Ellenbogen K, Ezekowitz MD, Haines DE, Haissaguerre M, Hindricks G, Iesaka Y, Jackman W, Jalife J, Jais P, Kalman J, Keane D, Kim YH, Kirchhof P, Klein G, Kottkamp H, Kumagai K, Lindsay BD, Mansour M, Marchlinski FE, McCarthy PM, Mont JL, Morady F, Nademanee K, Nakagawa H, Natale A, Nattel S, Packer DL, Pappone C, Prystowsky E, Raviele A, Reddy V, Ruskin JN, Shemin RJ, Tsao HM, Wilber D, Heart Rhythm Society Task Force on C and Surgical Ablation of Atrial F. 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design: a report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation. Developed in partnership with the
or bobooothththth aaaa sssstatatatandndndndarararard d dd lilililinnnn
blation catheter with 3.5mm distal electrode (Thermocool ) and a multielectrode
a t
er electrodes and closer interelectrode spacing have significant advantage in map
a
blationonon cccataa hehh teeeer rrr with 3.5mm distal electrododode (Thermocool )))) and a multielectrode
atttht eeete er with 1m1m1m1 m m m eleelecececectrtrtrtrododododesss (((PePePePenttarayayy®). IIn adadadaddidididitititit on,, thtththisisis sttudddy y yy hshshhowowowwedededed ttthahahahat ttt ccacc t
er eeeelelelectctcttrororoodedededes ananand d dd ccclc ososososer iiiintntntn erererrelelelelecccctrtrtrt odododdee spspspspacaca ininining g g g hahahavevevev sssigigiggnininin fififif cacacantnnn aaadvdvdvdvananananttat gegegeg iiiin n nn mamamamap
Thiiiisss tetetechchchnininiquququeeee cacacac nnn fafafacicicilililitataatatetete ssssucuucuccececessssssssffuf l ll ababablalalaatitititionononn ooooff f scscsccarararr--rereelalalateteteted dd d atatatriririialalala ttacacaca hyhyhyhycacacaardia
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
20
European Heart Rhythm Association (EHRA), a registered branch of the European Society of Cardiology (ESC) and the European Cardiac Arrhythmia Society (ECAS); and in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), the Asia Pacific Heart Rhythm Society (APHRS), and the Society of Thoracic Surgeons (STS). Endorsed by the governing bodies of the American College of Cardiology Foundation, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, the Asia Pacific Heart Rhythm Society, and the Heart Rhythm Society. Heart Rhythm. 2012;9:632-696 e21.
2. Bhargava M, Di Biase L, Mohanty P, Prasad S, Martin DO, Williams-Andrews M, Wazni OM, Burkhardt JD, Cummings JE, Khaykin Y, Verma A, Hao S, Beheiry S, Hongo R, Rossillo A, Raviele A, Bonso A, Themistoclakis S, Stewart K, Saliba WI, Schweikert RA and Natale A. Impact of type of atrial fibrillation and repeat catheter ablation on long-term freedom from atrial fibrillation: results from a multicenter study. Heart Rhythm. 2009;6:1403-1412.
3. Tzou WS, Marchlinski FE, Zado ES, Lin D, Dixit S, Callans DJ, Cooper JM, Bala R, Garcia F, Hutchinson MD, Riley MP, Verdino R, Gerstenfeld EP. Long-term outcome after successful catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2010;3:237-242.
4. Deisenhofer I, Estner H, Zrenner B, Schreieck J, Weyerbrock S, Hessling G, Scharf K, Karch MR, Schmitt C. Left atrial tachycardia after circumferential pulmonary vein ablation for atrial fibrillation: incidence, electrophysiological characteristics, and results of radiofrequency ablation. Europace. 2006;8:573-582.
5. Gerstenfeld EP, Marchlinski FE. Mapping and ablation of left atrial tachycardias occurring after atrial fibrillation ablation. Heart Rhythm. 2007;4:S65-72.
6. Jais P, Sanders P, Hsu LF, Hocini M, Sacher F, Takahashi Y, Rotter M, Rostock T, Bordachar P, Reuter S, Laborderie J, Clementy J, Haissaguerre M. Flutter localized to the anterior left atrium after catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2006;17:279-285.
7. Gerstenfeld EP, Callans DJ, Sauer W, Jacobson J, Marchlinski FE. Reentrant and nonreentrant focal left atrial tachycardias occur after pulmonary vein isolation. Heart Rhythm. 2005;2:1195-1202.
8. Chugh A, Oral H, Lemola K, Hall B, Cheung P, Good E, Tamirisa K, Han J, Bogun F, Pelosi F, Morady F. Prevalence, mechanisms, and clinical significance of macroreentrant atrial tachycardia during and following left atrial ablation for atrial fibrillation. Heart Rhythm.2005;2:464-472.
9. Cassidy DM, Vassallo JA, Marchlinski FE, Buxton AE, Untereker WJ, Josephson ME. Endocardial mapping in humans in sinus rhythm with normal left ventricles: activation patterns and characteristics of electrograms. Circulation. 1984;70:37-42.
oooooooopepepeper r r r JMJMJMJM,,, BaBaBaBalalalala RRRR, GaGaGaGaoutcocomeme aaftftftfterer ssucucccccceces010;3:3:3:3:232323237777 242424242222
ofer I, Estner H, Zrenner B, Schreieck J, Weyerbrock S, Hessling G, Scharf K, Kr
Eu
feld EP, Marchlinski FE. Mapping and ablation of left atrial tachycardias occurri
ofofofferrerr I, Estnnnnererer HHHH, ZrZrZrenenennenenen r B,B,B,B SSSSchhchrerereieeeckckck J, WeW yeyeyerbrbrbrbrororor ckkkk SSSS, , HeHeHeessss lingngngng GGGG, Sccchahahaarfrfrf KKKK, Kittttt CCC.C Left atriall taccchhhycaaardrdrdr ia aaffftf eeer circucuumfferrennntititit alalal puulmomomonaryry veieieie n abblalalal titt oonn foorrr aaatr: ininincicicidence,e,e,e, elecctrooopphysssioioiololl gigigicaccc lll cchc arararaaacteeriristicicicicsss, aaaannd rrressssultss oof rarararadid offffrererequueencyyy uropapaacecece. 222000006;66 88:88 575757333-3 582.222
feldddd EEEPP,P MMMarchchchhliiinskikiki FFFE.EE MMMMappppipipinggg and ablblblatiiioi n ffoff lllefffftt atatatat irii lall ttttaaaca hyhyhyh cardddiaii s ococococcuccc rrifibrillationononon aaaablblbblatatatatioioioion.nn.n HeHeHeHeararara t t t RhRhRhRhytytytythmhmhmhm. 202020200707007;4;4;4;4:S:S::S65656565-7-7-772.2.2. by guest on June 23, 2018
http://circep.ahajournals.org/D
ownloaded from
DOI: 10.1161/CIRCEP.114.002737
21
10. Marchlinski FE, Callans DJ, Gottlieb CD, Zado E. Linear ablation lesions for control of unmappable ventricular tachycardia in patients with ischemic and nonischemic cardiomyopathy. Circulation. 2000;101:1288-1296.
11. Patel AM, d'Avila A, Neuzil P, Kim SJ, Mela T, Singh JP, Ruskin JN, Reddy VY. Atrial tachycardia after ablation of persistent atrial fibrillation: identification of the critical isthmus with a combination of multielectrode activation mapping and targeted entrainment mapping. Circ Arrhythm Electrophysiol. 2008;1:14-22.
12. Barbhaiya CR, Kumar S, Ng J, Tedrow U, Koplan B, John R, Epstein LM, Stevenson WG , Michaud GF. Overdrive pacing from downstream sites on multielectrode catheters to rapidly detect fusion and to diagnose macroreentrant atrial arrhythmias. Circulation. 2014;129:2503-2510.
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
22
Table 1: Baseline Patient Characteristics
Variable Validation group(n=10)
Scar-related atrial arrhythmias(n=20)
Age, y 55.2±9.3 64.2±11.3
Gender, male 7 (70) 14 (70)
LA diameter, mm 40.4±6.1 56.7±11.5
LVEF, % 60.3±9.7 48.4±8.4
No. failed AADs 1 (0-2) 2 (1-3)
Mitral Regurgitation 27 (67.5) 45 (70.3)
0 (0) 11 (55)
Failed previous ablations 0 (0) 2 (0-4)
Surgical MAZE 0 (0) 4 (20)
Surgical MV replacement 0 (0) 6 (30)
Presenting arrhythmia
Paroxysmal AF 10 (100) 0 (0)
Persistent AF 0 (0) 6 (30)
AT 0 (0) 14 (70)
Values are expressed as mean ± SD, median (range) or n (%).AADs=antiarrhythmic drugs; AF=atrial fibrillation; AT=atrial tachycardia LA=left atrium; LVEF=left ventricular ejection fraction; MV=mitral valve
2 2 22 (1(1(11-3-3-3-3))))
45454545 ((((70707070 3333))))
v
M
M
0 00 (0(0(0) 1111111 (555555))))
viououousss ablatiiitionooo ss 0 (0(0(0)) ) 2 (0(0(0( -44)
MAZZZEEEE 0 0 0 (0(0(0( ) 4 444 (2(2(220)0)0)0)
MV replacccememememenenenentttt 0 000 (0(0(0(0) ))) 6 666 (3(3(3(30)000 by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
23
Table 2: Comparison of low voltage and scar surface area
Linear catheter Multielectrode-mapping catheter P (linear vs. multielectrode-mapping catheter)
Total low voltage area <0.5mV (cm2) (median, 25%, 75%)
20.4cm2 [9-55cm2] 22,4, 18.1, 50.6
14.7cm2 [7-48cm2] 15.2, 9.2, 43.6 0.02
Intermediate area 0.25 - 0.5mV (cm2) (median, 25%, 75%)
6.1±5.9 5.2, 3.8, 9.9
10.2 ±4.3 8.8, 4.7, 11.6 0.02
Dense scar <0.25mV (cm2) (median, 25%, 75%)
14.2±8.1 16.6, 10.2, 18.5
4.3±3.6 4.0, 2.1, 6.8 0.01
Values are expressed as mean ± SD, range [ ]
9.2, 43.6
6 1±5 9 10 2 ±4 36.1±5.9 5.2,2,2,2, 333.88,,,, 9.9.9.9.9 9 9
10000.2.22.2 ±4.3 88.8.8 8,8,8,, 4444.7, 11111111.6...
1414144.2.2.2.2±8±8±8±8.1... 16161616.6.66,, 101010.2.222,, 188188.5.55
4.4.4.4.3±3±3±3±3.3.3.3.6 66 6 4.4.4 0,00, 2222.1.11,, 6.66.6 888
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
24
Figure Legends:
Figure 1: Left atrial bipolar voltage maps (0.10-0.50mV) in the posterior-anterior view recorded
with a standard linear catheter (panel A) and a multielectrode catheter (Panel B) in a patient
with structurally normal left atria. The left atrial bipolar voltage amplitude was similar between
the maps with 95% of all electrograms >0.5mV. In contrast, in a patient with scar-related atrial
tachycardia, the bipolar voltage amplitude measured with the linear catheter (Panel C) was
significantly lower than the one measured with the multielectrode catheter (Panel D).
Figure 2: Left atrial bipolar voltage maps (0.10-0.50mV) in the anterior-posterior view recorded
with a multielectrode catheter (Panel A) and a linear catheter (Panel B) in a patient with scar-
related atrial tachycardia. Mapping with the multielectrode catheter demonstrated a significantly
smaller area of low voltage and “scar” in the anterior peri-mitral area. The two catheters were
then placed in contact with the zone of presumed “scar” as recorded with the linear catheter but
not with the multielectrode catheter (Panel C). The mean bipolar voltage amplitude recorded
with the multielectrode catheter at this location was 1.15 ± 0.7mV with mean electrogram
duration of 42 ± 11msec and minimal fractionation (Panel D), suggestive of non-scar tissue.
Figure 3: In a patient with history of surgical maze, mitral valve replacement and persistent
atrial tachycardia, left atrial bipolar voltage maps were performed using both a multielectrode
catheter (Panel A) and a standard linear catheter (panel B). Mapping with the multielectrode
catheter demonstrated smaller “scar” size with an overall higher bipolar voltage amplitude.
When the bipolar voltage maps were displayed at a range of 0.1 to 025mV, the map made with
heter (P( anel D)).
L c
t c
a c
e e
d in contact ith the one of pres med “scar” as recorded ith the linear catheter
Leftftft aaaatrrtriaiaiaal lll bibbb poooolllal r voltage maps (0.10-0.5550m00mV) in the anananteriririiooroo -posterior view rec
tiiiiellllectrode cattheterrr (((Pannneelee AAAA)))) anndd a llinnnearr ccattttheheheh teteteter (PPPannel BB) inninn a patatattieieennt wwittthh sc
al tacccchyhyhycccacardiaiii . MMMaM ppinii g wiwiwwithththth the mmmmullulultitititielllecttttrododdodeeee caaaththththetetetererere ddddemonststststrararaattted a iisiigniffifificii
ea of fff lol w voltll agagagage and dd “s““ car”””” in hhthe anteriiior ppperiii- iimitrallll aaaarrrer a. TTTThhheh two cattheheheetetetetersrrr we
dd iin tta tct itithh thth ff ded ““ ”” ddedd iithth tthhe lliin thth tet
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
25
the multielectrode catheter demonstrated improved delineation of the low voltage zone, revealing
an area of relatively preserved myocardium (“channel”) that was not demonstrated in a map
made with the linear catheter (Panels C and D). When the clinical atrial flutter was mapped, this
“channel” proved to be the isthmus of the tachycardia as seen by the activation map (Panel E).
When the multielectrode catheter was placed over the isthmus, >75% of the tachycardia cycle
length was recorded at this single point acquisition (Panel F). The tachycardia terminated with
ablation at this site.
Figure 4: In a patient with macroreentrant left atrial flutter, the multielectrode catheter and the
linear catheter were placed at a similar position at the area of “early meet late”. The linear
catheter was positioned in a perpendicular orientation to the tissue with its distal electrode
applying 9gr of tissue contact force. The multielectrode catheter was positioned parallel to the
tissue (panel A). While the linear catheter recorded low, far-field, signal with amplitude of
0.12mV, the multielectrode catheter recorded sharp, near-field, signals with bipolar amplitude
range of 0.18-0.28mV that allowed both local time annotation and pacing with capture (Panel
B).
Figure 5: In a patient with history of surgical maze, two previous percutaneous left atrial
ablation procedures and mitral valve replacement, a left atrial flutter was mapped with both the
standard linear and the multielectrode catheter. A bipolar voltage map made with the linear
catheter demonstrated diffuse and homogenous scar with bipolar voltage amplitude <0.5mV,
limiting activation and entrainment mapping (Panel A right). When the multielectrode catheter
was placed at the peri-mitral area alongside the linear ablation catheter, fractionated mid diastolic
lectrorodededd cc tatatthheheh ttetet r ananananddd
e
a
gr of tissue contact force. The multielectrode catheter was positioned parallel to
n f
he m ltielectrode catheter recorded sharp near field signals ith bipolar amplit
eteeerr rr wewewew reree ppplalll cececeed ddd at a similar position at tttheheheh area of “earlr y memememeet late”. The linear
assss ppppositioned iin a pppeerpendndnddicululular orrienntaaationon to thththeee tiissueueue witthh itttss distaalaa elecctroooddde
gr offf f tititisssssssue contttacttt fffforce. ThThThThe muuultltltltieiieielelelecttttroddded cccatatatatheheheh teteteter wawawaw s positiononononedededed par llalllelll llll ttto
nel AAAA))). WhWhWhililile ththththe lililinear catthhehh ter recordeddd lllow, ffffar-fffiellllddd, sigigignananaal iwii hththh ampppliiiitutututudededede of
hh ltltiielle tct dde tathhette drd ded hha fifi leldd iig lls itithh bibi lla lplitit
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
DOI: 10.1161/CIRCEP.114.002737
26
electrograms with bipolar voltage amplitude range of 0.18-0.34mV were recorded (Panel A left).
Overdrive pacing with capture from a single electrode of the multielectrode catheter entrained
the tachycardia in orthodromic fashion and a post pacing interval that was 13ms longer than the
tachycardia cycle length. A linear ablation between the mitral valve and the left inferior
pulmonary vein terminated the tachycardia.
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from
Elad Anter, Cory M. Tschabrunn and Mark E. JosephsonCloser Interelectrode Spacing
High-Resolution Mapping of Scar-Related Atrial Arrhythmias Using Smaller Electrodes with
Print ISSN: 1941-3149. Online ISSN: 1941-3084 Copyright © 2015 American Heart Association, Inc. All rights reserved.
Dallas, TX 75231is published by the American Heart Association, 7272 Greenville Avenue,Circulation: Arrhythmia and Electrophysiology
published online March 19, 2015;Circ Arrhythm Electrophysiol.
http://circep.ahajournals.org/content/early/2015/03/19/CIRCEP.114.002737World Wide Web at:
The online version of this article, along with updated information and services, is located on the
http://circep.ahajournals.org//subscriptions/
is online at: Circulation: Arrhythmia and Electrophysiology Information about subscribing to Subscriptions:
http://www.lww.com/reprints Information about reprints can be found online at: Reprints:
document. Permissions and Rights Question and Answerinformation about this process is available in the
requested is located, click Request Permissions in the middle column of the Web page under Services. FurtherCenter, not the Editorial Office. Once the online version of the published article for which permission is being
can be obtained via RightsLink, a service of the Copyright ClearanceCirculation: Arrhythmia and Electrophysiology Requests for permissions to reproduce figures, tables, or portions of articles originally published inPermissions:
by guest on June 23, 2018http://circep.ahajournals.org/
Dow
nloaded from