comparison of a novel single lead atrial sensing (a+)-icd...
TRANSCRIPT
Comparison of a novel single lead atrial sensing (A+)-ICD system with a
dual chamber ICD system in patients without antibradycardia pacing
indications: Results of a randomized study
Christian Sticherling, MD*, Markus Zabel, MD*, Sebastian Spencker, MD, Udo Meyerfeldt,
MD, Lars Eckardt, MD, Steffen Behrens, MD, Michael Niehaus, MD,
for the ADRIA investigators
University Hospital Basel, Switzerland, University Hospital Göttingen, Germany, Charité, Campus Benjamin Franklin, Berlin, Germany, Hospital Villingen-Schwenningen, Germany,
University Hospital Münster, Germany, Vivantes Hospital Berlin, Germany, University Hospital Hannover, Germany
* CS and MZ contributed equally to the study
ADRIA= Belos A+ versus DR Clinical Investigation of Arrhythmia Discrimination
Word count: 5’032
Subject code: [22] Ablation/ICD/surgery
Address for correspondence: Prof. Christian Sticherling, MD, FESC University Hospital Basel Division of Cardiology
Petersgraben 4 4031 Basel, Switzerland. E-mail: [email protected] Phone: 0041-61-265 5526 Fax: 0041-61-265 4598
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
2
Abstract
Background: Supraventricular tachyarrhythmias (SVT) are the main cause for inappropriate
therapy by implantable cardioverter-defibrillators (ICD). For better rhythm discrimination, an
atrial electrogram (AEGM) is helpful and usually obtained from an additional atrial lead, even
in the absence of sinus node or atrioventricular nodal disease. An A+- ICD system with
integrated atrial sensing rings mounted 15-18 cm from the tip of an ICD lead may obviate the
need to implant a separate atrial lead. The aim of the study was to compare the novel A+- ICD
and a conventional dual chamber DR-ICD.
Methods and Results: Two-hundred forty nine patients with standard ICD indications but no
requirement for antibradycardia pacing were randomized to receive an A+-ICD (n=124) or
DR-ICD (n=125). Implantation details, need for ICD system revision, long-term sensing,
documented arrhythmia episodes and the respective rhythm discrimination during follow-up
were analyzed.
The implantation time was significantly shorter in the A+ ICD group (67±30 vs. 79±30 min,
P=0.003). Mean P-wave amplitudes were 3.5±0.8 mV (A+) and 3.2±0.6 mV (DR) and
remained stable over the follow-up period of 12 months. Surgical revision was necessary in
13 patients in the DR and 10 in the A+ group. All 593 ventricular tachyarrhythmia (VT)
episodes were correctly discriminated. Sensitivity and specificity of SVT discrimination were
not different between the study groups.
Conclusion: The novel A+- ICD system can be implanted faster and is equivalent to a
standard DR-ICD with regard to the detection of VT and SVT. It represents a useful
alternative to obtain atrial sensing.
Clinical Trial Registration Information: ClinicalTrials.gov number,NCT00324662
Key words: sudden cardiac death, implantable defibrillator, tachyarrhythmias
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
3
Introduction
Therapy with an implantable cardioverter-defibrillators (ICD) is established for secondary and
primary prevention of sudden cardiac death 1-5. Most patients receive their ICD for primary
prevention and only a minority requires antibradycardia pacing. In patients who do not need
cardiac resynchronization therapy (CRT) in combination with ICD treatment, pacing has
untoward effects 6. This notwithstanding, many patients without pacing requirements receive
an atrial lead in order to obtain an atrial electrogram for improved atrial rhythm monitoring
and for better discrimination of supraventricular tachyarrhythmias (SVT) and ventricular
fibrillation (VF) or ventricular tachycardia (VT). Addition of an atrial lead, however, may
increase the risk of complications and may prolong implantation and fluoroscopy time.
The need to implant a separate atrial lead may be obviated by a novel A+-single coil
defibrillation lead with two integrated rings mounted 15-18 cm from the lead tip. The rings
function as a sensing dipole that floats within the right atrium. In conjunction with a dedicated
atrial input stage that filters and amplifies the atrial signal up to 4-fold, the A+-ICD system
has been proven safe and operational 7-10. This prospective study aimed to compare the A+
ICD with the conventional dual chamber ICD (DR-ICD) concept in a large patient cohort at
implantation and over 12 months of follow-up.
Methods
The Belos A+ versus DR Clinical Investigation of Arrhythmia Discrimination (ADRIA) study
enrolled 265 patients in 10 centers in Germany and Switzerland (see Appendix). The study
protocol was approved by the institutional review boards. The aim of the study was to
evaluate potential advantages and disadvantages of an A+ ICD system in comparison to a
conventional DR-ICD. The primary objective of this study was to determine whether there is
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
4
a difference in the specificity of SVT detection between the two systems. Secondary
objectives were the comparison of the sensititivies of the detection enhancement algorithm, of
the complication rates, the implantation and fluoroscopy times and the lead performance over
time.
The study was sponsored by the manufacturer of the devices (Biotronik SE, Berlin,
Germany). Before publication, the data was reviewed but the manuscript was not influenced
by Biotronik.
Patient Selection
Patients with a standard ICD indication were enrolled in the study 3. Exclusion criteria were
the need for antibradycardia pacing, permanent atrial fibrillation, life expectancy under 6
months, pregnancy, and age below 18 years. All patients provided written informed consent.
Device Specifications
Following enrollment, patients were randomized to receive either a standard DR-ICD (Belos
DR, Lexos DR, or Lumos DR-T) that required implantation of two electrodes, or an A+-ICD
(Belos A+ or Lexos A+) with a single A+ lead (Kainox A+ or Kentrox A+). All ICDs and A+
leads were manufactured by Biotronik SE, Berlin, Germany. Other leads could be selected at
the discretion of the operator.
The A+ lead combines a single ventricular shock coil, bipolar sensing and pacing in the
right ventricle, and true bipolar sensing in the right atrium with the aid of two rings spaced 15
mm apart (figure1). The position exactly in the middle between the rings is located 15 or 17
cm from the lead tip. All A+ leads had a passive ventricular fixation mechanism and silicon
lead insulation. The older Kainox A+ lead had a significantly larger surface area of the
ventricular tip electrode (6.0 mm2) than the successor Kentrox A+ lead (1.8 mm2).
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
5
The A+ ICD systems differ from the DR-ICD systems in three aspects: they have no
atrial pacing capability, use a pre-amplifier for the atrial sensing channel and contain only one
right ventricular shock coil. The devices were equal with respect to programmability,
diagnostic memory characteristics, and the employed tachycardia discrimination algorithm
(SMART®). The latter classifies all ventricular events as SVT or VT depending on the RR
and PP interval, and its respective regularity and multiplicity, as described in literature11.
When programmable SVT or VT detection criteria are met, the episode is stored, including
one minute of dual chamber intracardiac electrogram (IEGM) preceding the detection. In case
of SVT without a coexistent VT or VF, antitachycardia therapy is withheld.
Implantation
Implantations were performed according to institutional standards. The total implantation
time, the time between the beginning of lead insertion and lead(s) connection to the ICD (lead
implantation time), and the total fluoroscopy time were recorded in each patient. The
investigators also measured ventricular pacing threshold at 0.5 ms pulse duration, ventricular
pacing impedance, and P- and R-wave amplitudes.
Device Programming
The study protocol recommended the use of lower cut-off rate of 120 or 130 bpm for a VT-1
monitoring zone without therapy. The detection of as many SVT episodes as possible was
thus facilitated in order to compare the efficacy of the SMART detection enhancement
algorithm between the two groups.
The investigators were free to select the cut-off rates for ICD therapy zones. The tachycardia
discrimination algorithm was activated in all VT zones. The VT zone was programmed 50
bpm above the monitor zone with a detection duration of 14, onset of 20%, stability of 12%
and SMART detection and redetection on. The detection for the VF-zone was programmed to
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
6
8 of 12, while the rate cut-off could be programmed individually. The post-ventricular atrial
blanking (PVAB) was 16ms before to 64ms after a ventricular event.In order to avoid
ventricular pacing, pacing was programmed to VVI 40 beats per minute.
Follow-up
During the follow-up visits at 1, 3, 6 and 12 months after device implantation, ICD
interrogation was performed, all documented episodes of SVT or VT/VF were collected and
the need for ICD system revisions documented if applicable.
Analysis of Arrhythmia Episodes
All spontaneous tachyarrhythmia episodes stored by the device were classified by authorized
participants of each study center as VF, VT or SVT. Pre-defined subtypes of SVT were atrial
flutter, atrial fibrillation, sinus tachycardia, and other SVT (figures 2a and 2b). Episodes of
VT or VF with concurrent SVT were classified as VT/VF. Episodes other than VT/VF or
SVT were excluded from analysis. From all episodes atrial and ventricular IEGM were
documented.
Blinded to the classification of the on-site investigator, members of the steering
committee (C.S:, M.Z., M.N.) of the ADRIA study reviewed a sample of 370 randomly
chosen episodes from 85 patients. The concordance of the classification for VT/VF or SVT
between the investigators and the committee was 96.5%. The committee thereafter
recommended that the on-site rhythm classifications should be generally accepted, but
reserved the right to overrule on-site decisions if errors were noticed during analysis of the
documentation for the episodes.
Specificity and sensitivity of the SMART discrimination algorithm were calculated in all
episodes documented in the VT-zone. The specificity of the discrimination algorithm was
defined as the proportion of correctly classified SVT episodes: True negative / (True negative
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
7
+ false positive). The sensitivity was defined as the proportion of correctly classified VT/VF
episodes: True positive / (True positive + false negative).
To account for the bias in deriving of specificity and sensitivity from multiple episodes of the
same type in the individual patients, the results were corrected by the generalized equation
estimation (GEE) 12, 13. Briefly, GEE-method adjusts the effective number of episodes of
every individual patient depending on the correlation of the patients’ episodes, which are not
necessarily independent. The GEE-method is applied before the individual specificity figures
are pooled, and is generally accepted to provide the best estimate of the true specificity and
sensitivity.
Statistics
Statistical analysis of study results was performed using the software packages SPSS 16.0 and
(for GEE) PASW Statistics 17.0 (SPSS Inc., Chicago, IL, USA). Continuous variables are
reported as mean ± SD. The two-sided t-test was used for normally distributed data, and the
Wilcoxon-Mann-Whitney (U) rank test was used for data that were not distributed normally.
Categorical data were compared using a χ2 test according to Pearson or Fisher’s exact test as
appropriate. In all cases, statistical significance was established with p <0.05.
Per-protocol analysis was performed. Cross-over was not allowed. Patients who received a
device that was not compatible with their randomization scheme during follow-up were
considered drop-outs. Arrhythmia episodes and adverse events that had occurred before drop-
out were taken into account.
The necessary sample size to compare the specificities of the study groups was determined to
reject the null hypothesis of inferiority of the A+ specificity with a ∆≥7%. It was assumed
that the specificity in the both study groups is 93% (unpublished data from non-randomized
studies available during study planning). With an α=0.05 and a power of 0.8, we calculated
that we would need 181 uncorrelated SVT episodes per group including a 10% drop-out rate.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
8
We assumed an interclass correlation of 0.5 between the episodes of individual patients, and
expected 3 episodes per patient. We calculated that it would be possible to disprove the H0
with 120 patients per group.
Results
Patient Population
Of the 265 enrolled patients, 16 (6%) were excluded for violation of the randomization or
withdrawal of informed consent before implantation. Of the remaining 249 patients, 124 were
randomized to the A+- and 125 to the DR study arm. There were no significant differences in
the baseline characteristics between the two groups, except for the presence of previous
myocardial infarction (47% DR vs. 33% A+) and the use of statins (65% DR vs. 43% A+,
Table 1). Seventy-three percent of the patients had ischemic heart disease and 39% a history
of VF or VT. The medication was typical for a contemporary ICD population, with a high
percentage of beta-blocker, ACE-inhibitor and diuretic use.
Implantation
The ICD leads implanted in the A+ group were Biotronik Kainox A+ (n=58) and Biotronik
Kentrox A+ (n=66). In the DR group, the ventricular ICD leads were Biotronik Kainox (n=6),
Biotronik Kentrox (n=69), Biotronik Linox (n=38) and other ICD leads (n=12), and the atrial
leads Biotronik Setrox (n=45), Biotronik Selox (n=44), Biotronik Elox (n=3), Medtronic
Capsure 5076 (n=28) or other leads (n=5).
The mean total implantation time was significantly shorter in the A+ group (67 vs. 79 min;
P=0.003), owing to shorter lead implantation time (26 vs. 36 min; P <0.001, Table 2).
After pre-amplification of the atrial signal in the A+ ICD devices, the mean P-wave amplitude
(3.5 mV, equivalent to approximately 1.1 mV without pre-amplification) was similar to the
figure in the DR-group (3.2 mV) and remained stable in both groups over time.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
9
The mean R-waves did also not differ between the two groups (11+2 mV). Compared to the
DR-group, ventricular pacing thresholds were higher in the A+-group (1.2+1.0 V/0.5 ms vs.
0.8+0.6 V/0.5 ms) and pacing impedances were lower (Table 2). However, for the newer
generation Kentrox A+ leads, utilized in more than half of the patients in each study arm,
there were no differences in pacing thresholds and pacing impedances between the A+ ICD
(955 Ohm, 0.7 V/0.5 ms) and the DR-ICD (912 Ohm, 0.8 V/0.5 ms).
Follow-up and Adverse Events
The mean follow-up was 370 ± 104 days, with a cumulative follow-up of 251 years. Thirteen
patients (5%) died. None of these deaths were device-related. Fifteen patients (6%) were lost
to follow-up. Arrhythmia episodes and adverse events that had occurred before drop-out were
included in the analysis. There were 23 system-related adverse events requiring reintervention
(Table 3). These included high defibrillation thresholds (DFT) with the initial lead during
implantation or pre-discharge testing (A+-group n=2, DR-group n=1), atrial lead dislocation
(DR-group n=5), or repositioning of the ventricular lead because of dislocation or high pacing
thresholds (A+ group n=7, DR-group n=5). Of the 3 patients with high DFT, one patient in
the DR-arm received a high-energy device and an additional SVC coil was implanted in a
patient in the A+ group. Another patient in the A+ group received a dual-coil lead along with
a single chamber VVI-ICD and was hence counted as a drop-out.
Documented Episodes
A total of 3498 spontaneous episodes of arrhythmia were recorded. Sixty-four
supraventricular episodes were excluded because the tachycardia discrimination algorithm
had been switched off by the investigators because of small atrial signals or permanent atrial
arrhythmias (A+-group n=7, DR-group n=2; P=ns). Another 224 episodes (154 episodes in 20
patients in the DR-group and 70 episodes in 14 patients in the A+-group; P=ns) were
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
10
excluded since the investigators classified them as "other rhythm” or "unclear" (mostly due to
intermittent T-wave oversensing or premature ventricular contractions).
Of the remaining 3210 episodes, 1894 occurred in the A+ group and 1316 in the DR group.
This difference was not statistically significant after correction by the GEE-method (P=0.09).
In total, 18.5% of episodes were VT or VF, and 81.5% were SVT. The distribution of
ventricular and supraventricular arrhythmias was comparable in both groups except for a
slightly higher number of VT episodes in the A+-group. During data analysis, 21 (0.7%)
episode classifications made by the investigator were overruled by the steering committee.
Table 4 compares rhythm classification for the two randomized groups.
The proportion of the different arrhythmias at different rates is illustrated in Figure 3. The
large number of sinus tachycardia episodes (61%) was due to the intentionally low cut-off rate
for the VT-1 zone. The programmed lower cut-off rate for VT1 was equal or less than 130
bpm (as recommended in study protocol) in 58.4% of all episodes, between 131 and 140 bpm,
in 29.2% of episodes and equal or greater than 140 bpm in 12.3%.
Sensitivity and Specificity
All 593 spontaneous episodes classified by the investigators as VF or VT were correctly
discriminated, resulting in a sensitivity of 100% for both groups before and after GEE-
correction.
Of the 2617 SVT episodes, 2564 were detected within the VT rate range and were therefore
evaluated by the tachycardia discrimination algorithm and included in our specificity
calculations in Table 5. The GEE-corrected specificity did not differ between the A+-group
(61.8%) and the DR-group (66.2%, P=0.22), however, the null hypothesis of inferiority of the
A+ specificity cannot be rejected (P=0.39). The overwhelming majority of SVT episodes
were sinus tachycardia. We therefore analyzed a sample of 492 false positive sinus
tachycardia episodes. False positive detection typically resulted from early premature
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
11
ventricular contractions (PVC) activating the onset criterion or intermittent atrial sensing
disturbances. Both over- and undersensing in the atrium rendered the algorithm to assume
different averaged rates in the atrium versus ventricle, precluding proper detection of a sinus
tachycardia. While the prevalence of PVC was similar in both study arms, over- and
undersensing of atrial signals was more often found in the A+ group. Episodes of over- or
undersensing were found in 23 of 63 (36%) A+ patients with sinus tachycardia, as compared
with 5 of 47 (11%) DR patients with sinus tachycardia (P=0.002) This is due to the higher
prevalence of small atrial signals before amplification in the A+-group, which can cause P-
wave undersensing or far-field R-wave oversensing and may explain the slight difference
between the specificities of the study groups. Typical examples of complete undersensing of
the atrial signal (fig. 4a) and intermittent R-wave oversensing (fig. 4b) in the monitor zone did
not result in untoward clinical consequences for the patient.
Despite misclassification of the SVT episodes, only very few patients received shock therapy
since the vast majority of episodes was in the monitor-zone, or the SVT terminated before
shock delivery. In only 46 of 1126 false-positive episodes in the VT or VF-zone, (4.1% not
GEE-corrected), shocks were delivered. Thirty-four episodes occurred in 7 patients of the
A+- and 12 episodes occured in 7 patients of the DR-group.
Discussion
This is the first study to compare a novel single lead ICD system with atrial sensing
capabilities to a conventional dual-chamber ICD. The main finding is that in patients with
standard ICD indication and no need for antibradycardia pacing, the A+ ICD system is faster
to implant and non-inferior in VT and SVT detection and therapy when compared to a DR-
ICD system. It provides an attractive alternative to an additional atrial lead if one wishes to
obtain an atrial electrogram in ICD patients.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
12
Need for atrial sensing
In the contemporary ICD-population the majority of patients do not require antibradycardia
pacing. The DAVID trial demonstrated that right ventricular pacing even increases mortality
and hospitalization rate in an ICD population and should therefore be avoided 6. In the CRT-
population, even right atrial pacing may have untoward effects 14. Although pacing should be
avoided whenever possible, sensing information from the atrium is desirable for numerous
reasons. First, the integration of the atrial sensing information can be crucial in differentiating
supraventricular from ventricular arrhythmias. Inappropriate shocks not only impair quality of
life but are associated with increased mortality and should thus be avoided by all means 15.
Currently, all manufacturers feature detection enhancement algorithms that achieve high
sensitivities and specificities and often integrate the comparison of ventricular and atrial
signals 11, 16, 17. The SMART detection algorithm used in this study has been described
elsewhere and has been shown to have a sensitivity of 100% along with a specificity of 89%
11. In our study the sensitivity was 100%, i.e. that all ventricular tachyarrhythmias were
detected as such, while the overall specificity was only 64%. The specificity of a detection
enhancement algorithm depends on numerous parameters such as the population studied, the
prevalence of supraventricular tachycardias and the device programming. Hence it is difficult
to compare these algorithms if they were not studied under the same conditions. The relatively
low specificity in the present study may be explained by the fact that the rate cut-off was
unusually low at 130 bpm in order to obtain as many episodes as possible. With faster heart
rates, the impact of PVC on triggering the onset criterion should decrease. In fact, analysis of
a non-prespecified small sample of 321 episodes recorded in 21 patients in whom the cut-off
had been set to > 140 bpm resulted in a GEE-corrected specificity of 86%, a specificity
reported earlier11. It is therefore likely, that the lower specificity stems from a clinically
unusually low study cut-off and not from a performance problem of the SMART algorithm.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
13
Another important reason for the acquisition of atrial information is the possibility to diagnose
atrial fibrillation and to initiate anticoagulant and possibly antiarrhythmic therapy. A recent
study using remote monitoring found that 13% of a contemporary pacemaker-, ICD- and
CRT-population had previously unknown atrial fibrillation18. Thus, the availability of an atrial
electrogram in combination with remote monitoring may offer accelerated treatment options
for these patients, for instance the potential to reduce strokes 19.
Implantation and complications
Not surprisingly, the use of a single-lead ICD resulted in shorter total implantation times due
to shorter lead implantation time. The use of the single-coil ICD A+ defibrillation lead did not
result in a significantly higher failure rate to defibrillate when compared with the dual-coil
leads. Although dual-coil defibrillator leads offer more flexibility in the choice of the shock
pathway in the rare case of high defibrillation threshold (DFT), randomized trials did either
show no difference 20 or only a small difference (10.2 J vs. 9.8 J) 21 in the DFT between
single- and dual-coil ICD.
Atrial lead dislodgement occurred in 4% of the patients in the DR-group, which is slightly
lower than in previously reported large series 20. Naturally, this complication cannot occur in
the A+-group.
In the A+-group the ventricular pacing threshold was slightly higher than in the DR-
group, which was entirely due to the initial higher usage of the older-generation Kainox A+
leads. For safety purposes, this first pentapolar lead had a relatively large surface area (6.0
mm2) in order to minimize the risk of perforation. Based on the clinical experience, its
successor lead Kentrox A+ had a much smaller surface area (1.8 mm2). After exclusion of the
Kainox A+ leads, the ventricular pacing thresholds did not differ between the two groups.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
14
The 12-month mortality rate of 5.2% is consistent with the mortality rates in other ICD
studies 1.
Difference to VDD-pacemaker sensing algorithms
The concept of single-lead VDD pacing systems with a floating bipole in the atrium has long
been available but comprises only 7% of the European pacing market 21 . One reason for the
reluctance to use VDD-pacing is the concern of the development of sinus node disease with
subsequent need for pacing. Schaer and colleagues demonstrated in a large cohort of 320
patients with AV-nodal disease only 1% required an upgrade to a DDD-pacemaker during
follow-up 22. Intermittent atrial undersensing during deep breathing, change of body position
or exercise is a well known phenomenon that occurs in both VDD and DDD-pacemaker
systems21. The incidence of atrial undersensing in VDD-systems is between 5 and 7% but is
clinically silent in the vast majority of patients23-25.
During follow-up, the amplified atrial sensing signal remained stable and showed no tendency
for deterioration in the A+ group. The processing of the atrial signal differs between VDD-
pacemakers and the studied ICD, since the latter comprise a special input stage that amplifies
the atrial signal up to 4-fold.
Limitations
This study focused on atrial sensing and its contribution to rhythm discrimination in ICD. Its
impact on AV-sequential pacing has not been evaluated in this study.
Secondly, our study was not designed to investigate the influence of sensing disturbances on
the discrimination, and to give mechanistic explanation of SVT discrimination failure.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
15
Thirdly, the performance of the A+-single coil ICD lead may be different in the case of a
right-sided implant. Since all our implantions have been performed from the left side our
study does not provide any data for this setting. However, data from VDD-pacemaker studies
actually even showed better atrial sensing behaviour in right-sided implants26.
Fourthly, with the devices used in this study, atrial tachyarrhythmias were only recorded if
their ventricular rate was above the VT cut-off rate. Since atrial tachyarrhythmias with a slow
ventricular rate have been missed the information obtained from the devices studied has not
been validated for the further therapeutic management of patients with suspected atrial
tachyarrhythmias. Finally, it cannot be completely ruled out that the observed higher
incidence of atrial over- or undersensing in the A+-group can be clinically significant in
patients with slow VT and concomittant sinus tachycardia
Conclusion
The use of an A+-single coil ICD lead in defibrillator patients without pacing requirement is a
clinically elegant means to obtain atrial electrograms from only one lead. Compared to ICD
patients with an additional atrial lead this results in shorter implantation times without
sacrificing the accuracy of arrhythmia detection.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
16
Acknowledgments
The investigators wish to thank Dr. Juergen Schrader (Biotronik, Berlin, Germany) for his
support in data handling and statistical analysis.
Sources of Funding
The study was sponsored by the manufacturer of the devices (Biotronik SE, Berlin,
Germany).
Disclosures
Dr. Sticherling is consultant for Biotronik, Medtronic, and Boston Scientific. Dr. Zabel is
consultant for Biotronik, Medtronic, and Boston Scientific. Dr. Spencker is consultant for
Biotronik and Medtronic. Dr. Niehaus is consultant for Medtronic and Boston Scientific. The
other authors report no potential conflicts.
Appendix
Participating Centers and investigators
University Hopital Basel, Switzerland: Christian Sticherling, MD; Beat Schaer; MD; Heart
Center, University of Göttingen, Germany: Markus Zabel, MD; Dieter Zenker, MD; Charité,
Campus Benjamin Franklin, Berlin, Germany: Sebastian, Spencker, MD; Schwarzwald-Baar-
Klinikum Villingen-Schwenningen, Germany: Udo Meyerfeldt, MD; Hannover Medical
School, Hannover, Germany: Michael Niehaus, MD; University Hospital of Muenster,
Germany: Lars Eckardt, MD, Julia Koebe, MD; Vivantes- Humboldt Klinikum, Berlin:
Steffen Behrens, MD; St.-Johannes Hospital Dortmund, Germany: Ralf Dedner, MD; Sana
Klinikum Berlin, Germany: Olaf Göing, MD; Cardiology Practice Bonn, Germany: Thomas
Klingenheben, MD
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
17
References
1. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, Domanski M, Troutman C, Anderson J, Johnson G, McNulty SE, Clapp-Channing N, Davidson-Ray LD, Fraulo ES, Fishbein DP, Luceri RM, Ip JH. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352:225-237.
2. Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877-883.
3. Epstein AE, Dimarco JP, Ellenbogen KA, Estes NA, 3rd, Freedman RA, Gettes LS, Gillinov AM, Gregoratos G, Hammill SC, Hayes DL, Hlatky MA, Newby LK, Page RL, Schoenfeld MH, Silka MJ, Stevenson LW, Sweeney MO. ACC/AHA/HRS 2008 guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: executive summary. Heart Rhythm. 2008;5:934-955.
4. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. N Engl J Med. 1997;337:1576-1583.
5. Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med. 1999;341:1882-1890.
6. Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, Kutalek SP, Sharma A. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA. 2002;288:3115-3123.
7. Eberhardt F, Schuchert A, Schmitz D, Zerm T, Mitzenheim S, Wiegand U. Incidence and significance of far-field R wave sensing in a VDD-implantable cardioverter defibrillator. Pacing Clin Electrophysiol. 2007;30:395-403.
8. Niehaus M, de Sousa M, Klein G, Korte T, Pfeiffer D, Walles T, Raymondos K, Tebbenjohanns J. Chronic experiences with a single lead dual chamber implantable cardioverter defibrillator system. Pacing Clin Electrophysiol. 2003;26:1937-1943.
9. Niehaus M, Schuchert A, Thamasett S, Pfeiffer D, Korte T, Pichlmaier M, Panning B, Belke R, Tebbenjohanns J. Multicenter experiences with a single lead electrode for dual chamber ICD systems. Pacing Clin Electrophysiol. 2001;24:1489-1493.
10. Schuchert A, Niehaus M, Binner L, Belke R, Meinertz T. Feasibility of atrial sensing via a free-floating single-pass defibrillation lead for dual-chamber defibrillators. J Interv Card Electrophysiol. 2003;8:209-214.
11. Sinha AM, Stellbrink C, Schuchert A, Mox B, Jordaens L, Lamaison D, Gill J, Kaplan A, Merkely B. Clinical experience with a new detection algorithm for differentiation of supraventricular from ventricular tachycardia in a dual-chamber defibrillator. J Cardiovasc Electrophysiol. 2004;15:646-652.
12. Blackwelder WC. "Proving the null hypothesis" in clinical trials. Control Clin Trials. 1982;3:345-353.
13. Farrington CP, Manning G. Test statistics and sample size formulae for comparative binomial trials with null hypothesis of non-zero risk difference or non-unity relative risk. Stat Med. 1990;9:1447-1454.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
18
14. Bernheim A, Ammann P, Sticherling C, Burger P, Schaer B, Brunner-La Rocca HP, Eckstein J, Kiencke S, Kaiser C, Linka A, Buser P, Pfisterer M, Osswald S. Right atrial pacing impairs cardiac function during resynchronization therapy: acute effects of DDD pacing compared to VDD pacing. J Am Coll Cardiol. 2005;45:1482-1487.
15. Poole JE, Johnson GW, Hellkamp AS, Anderson J, Callans DJ, Raitt MH, Reddy RK, Marchlinski FE, Yee R, Guarnieri T, Talajic M, Wilber DJ, Fishbein DP, Packer DL, Mark DB, Lee KL, Bardy GH. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med. 2008;359:1009-1017.
16. Francia P, Balla C, Uccellini A, Cappato R. Arrhythmia detection in single- and dual-chamber implantable cardioverter defibrillators: the more leads, the better? J Cardiovasc Electrophysiol. 2009;20:1077-1082.
17. Friedman PA, McClelland RL, Bamlet WR, Acosta H, Kessler D, Munger TM, Kavesh NG, Wood M, Daoud E, Massumi A, Schuger C, Shorofsky S, Wilkoff B, Glikson M. Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the detect supraventricular tachycardia study. Circulation. 2006;113:2871-2879.
18. Ricci RP, Morichelli L, Santini M. Remote control of implanted devices through Home Monitoring technology improves detection and clinical management of atrial fibrillation. Europace. 2009;11:54-61.
19. Ricci RP, Morichelli L, Gargaro A, Laudadio MT, Santini M. Home Monitoring in Patients with Implantable Cardiac Devices: Is There a Potential Reduction of Stroke Risk? Results from a Computer Model Tested Through Monte Carlo Simulations. J Cardiovasc Electrophysiol. 2009;20:1244-1251.
20. Fleck T, Khazen C, Wolner E, Grabenwoger M. The incidence of reoperations in pacemaker recipients. Heart Surg Forum. 2006;9:E779-782.
21. Ovsyshcher IE, Crystal E. VDD pacing: under evaluated, undervalued, and underused. Pacing Clin Electrophysiol. 2004;27:1335-1338.
22. Schaer BA, Weinbacher M, Zellweger MJ, Sticherling C, Osswald S. Value of VDD-pacing systems in patients with atrioventricular block: experience over a decade. Int J Cardiol. 2007;122:239-243.
23. Van Campen CM, De Cock CC, Huijgens J, Visser CA. Clinical relevance of loss of atrial sensing in patients with single lead VDD pacemakers. Pacing Clin Electrophysiol. 2001;24:806-809.
24. Kuzniec J, Golovchiner G, Mazur A, Battler A, Strasberg B. Atrial signal amplitude measurements in VDD systems according to body position. Pacing Clin Electrophysiol. 2004;27:1344-1346.
25. Wiegand UK, Nowak B, Reisp U, Peiffer T, Bode F, Potratz J. Implantation strategy of the atrial dipole impacts atrial sensing performance of single lead VDD pacemakers. Pacing Clin Electrophysiol. 2002;25:316-323.
26. Haghjoo M, Arya A, Emkanjoo Z, Sadr-Ameli MA. Optimal side of implant for single-lead VDD pacing: right-sided versus left-sided implantation. Pacing Clin Electrophysiol. 2005;28:384-390.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
19
Table 1. Baseline Patient Characteristics
A+ ICD
(n=124)
DR-ICD
(n=125)
P
Female, n (%) 12 (10) 20 (16) n.s.
Mean age ± SD, years 62 ± 11 63 ± 11 n.s.
History of tachyarrhythmia, n (%)
VF 12 (10) 11 (9) n.s.
VT without VF 37 (30) 36 (29) n.s.
Atrial fibrillation 22 (18) 16 (13) n.s.
Atrial flutter 7 (6) 2 (2) n.s.
Ischemic heart disease, n (%) 86 (69) 95 (76) n.s.
Myocardial infarction 41 (33) 59 (47) 0.023
PTCA 50 (40) 56 (45) n.s.
CABG 13 (10) 21 (17) n.s.
Non-ischemic cardiomyopathy, n (%) 27 (22) 26 (21) n.s.
Valvular disease, n (%) 12 (10) 12 (10) n.s.
NYHA functional class: I/II/III/IV, n 17/47/20/0 16/38/31/4 n.s.
*Patients with information on
medication, n (%)
119 (96) 120 (96) n.s.
Beta-blocker 109 (92) 112 (93) n.s.
ACE-inhibitor 98 (82) 95 (79) n.s.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
20
Angiotensin II-antagonist 13 (11) 20 (17) n.s.
Diuretic 86 (72) 88 (73) n.s.
Amiodarone 35 (28) 38 (30) n.s.
Other antiarrhythmic drugs 2 (2) 3 (2) n.s.
Lipid-lowering therapy 51 (43) 78 (65) <0.001
Anticoagulant 22 (18) 35 (29) n.s.
Thrombocyte aggregation inhibitor 89 (75) 99 (83) n.s.
Digitalis 24 (20) 31 (26) n.s.
*Percentage of patients taking a drug was calculated among 239 patients (96%) with
information on medication. ACE: angiotensin-converting enzyme, CABG: coronary artery
bypass graft, ICD: implantable cardioverter-defibrillator, n.s.: not significant, NYHA: New
York Heart Association, PTCA: percutaneous transluminal coronary angioplasty, SD:
standard deviation, VT: ventricular tachycardia, VF: ventricular fibrillation.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
21
Table 2. Implantation Data
A+ ICD
(n=124)
DR-ICD
(n=125)
P
Total implantation time, min 67 ± 30 79 ± 30 0.003
Lead implantation time, min 26 ± 19 36 ± 19 <0.001
Fluoroscopy time, min/sec 6'15" ± 6'31" 6'59" ± 5'41" n.s.
Atrium
P-wave amplitude, mV 3.5 ± 0.8 3.2 ± 0.6 n.s.
PI, Ohm - 650 ± 133 -
Ventricle
R-wave amplitude, mV 11 ± 2 11 ± 2 n.s.
PT in all leads, [email protected] ms 1.2 ± 1.0 0.8 ± 0.6 <0.001
PT in Kentrox* leads, [email protected] ms 0.7 ± 0.5
(n=65)
0.8 ± 0.4
(n=69) n.s.
PI in all leads, Ohm 759 ± 297 880 ± 292 0.002
PI in Kentrox* leads, Ohm 955 ± 276
(n=65)
912 ± 311
(n=69) n.s.
Data are shown as mean ± standard deviation. *Newer technology, with a smaller surface area
of the ventricular tip electrode (1.8 mm2). ICD: implantable cardioverter-defibrillator, n.s.: not
significant, PI: pacing impedance, PT: pacing threshold.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
22
Table 3. Adverse Events During Implantation and Follow-up
A+ ICD
(n=124)
DR-ICD
(n=125)
P
FU duration ± SD, days 361 ± 104 379 ± 106 n.s.
Deaths, n (%) 6 (4.8) 7 (5.6) n.s.
Device related, n (%) 0 0 n.s.
Cardiac* 2 (1.6) 4 (3.2) n.s.
Non-cardiac* 4 (3.2) 3 (2.4) n.s.
Drop-out, n (%) 9 (7.3) 6 (4.8) n.s.
Late informed consent withdrawal, n (%) 0 1 (0.8) n.s.
Lost to FU or poor compliance, n (%) 5 (4.0) 5 (4.0) n.s.
†Received new device (lead or ICD), n (%) 4 (3.2) 0 n.s.
Adverse events (invasive solution), n (%) 10 13 n.s.
Failure to defibrillate (implantation, pre-discharge) 2 1 n.s.
Ventricular perforation 0 1 n.s.
Pneumothorax 1 1 n.s.
Atrial lead dislocation - 5 -
Ventricular lead dislocation or high threshold 7 5 n.s.
*Confirmed or assumed to have this cause. †In two cases, switch to conventional ICD lead
was required because of dislocation or high pacing threshold with the Kainox A+ lead; in one
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
23
patient, high a energy device was used during implantation; and in one patient, upgrade to
cardiac resynchronization therapy was carried out. FU: follow-up, ICD: implantable
cardioverter-defibrillator, n.s.: not significant, SD: standard deviation.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
24
Table 4. Classified Arrhythmia Episodes
A+ ICD
(n=124)
DR-ICD
(n=125)
P All patients
(n=249)
Patients with episodes, n (%) 88 (71.0) 85 (68.0) n.s. 173 (69.5)
Total number of episodes, n 1894 1316 n.s. 3210
Episodes per patient ± SD, n
Median
21.5 ± 22.3
15.5
15.5 ± 15.0
9
n.s. 18.5 ± 19.2
12
Ventricular tachyarrhythmia, n (%) 328 (17.3) 265 (20.1) n.s. 593 (18.5)
Ventricular fibrillation 62 (3.3) 47 (3.6) n.s. 109 (3.4)
Ventricular tachycardia 266 (14.0) 218 (16.5) 0.017 484 (15.1)
SVT, n (%) 1566 (82.7) 1051 (79.8) n.s. 2617 (81.5)
Atrial fibrillation 172 (9.1) 101 (7.7) n.s. 273 (8.5)
Atrial flutter 23 (1.2) 121 (9.2) n.s. 144 (4.5)
Sinus tachycardia 1223 (64.6) 735 (55.9) n.s. 1958 (61.0)
Other SVT 148 (7.8) 94 (7.1) n.s. 242 (7.5)
Patients with SVT, n (%) 67 (54.0) 60 (48.0) n.s. 127 (51.0)
ICD: implantable cardioverter-defibrillator, SVT: supraventricular tachyarrhythmia, n.s.: not
significant.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
25
Table 5. Discrimination Algorithm Results and GEE-Corrected Specificity
A+ ICD DR-ICD P All patients
Ventricular fibrillation or tachycardia, n 328 265 n.s. 593
True positive 328 265 - 593
False negative 0 0 - 0
Sensitivity, % 100% 100% n.s. 100%
Supraventricular tachyarrhythmia, n 1521 1043 n.s. 2564
True negative 830 663 - 1493
False positive 691 380 - 1071
Specificity, % 61.8 66.2 n.s. 64.4
Lower 95% confidence interval bound, % 55.7 59.4 - 60.0
Upper 95% confidence interval bound, % 67.9 72.9 - 69.1
GEE: generalized equation estimation, ICD: implantable cardioverter-defibrillator.
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
26
Figure Legends
Figure 1: Picture of the single-coil, passive-fixation Biotronik Kentrox A+ lead. The atrial
bipol can be mounted at a distance of 15 or 17 cm from the lead tip.
Figure 2: Intracardiac electrograms (EGM) depicting an episode of rapidly conducted
recorded AF by an A+ -ICD (a) and an episode of sinus tachycardia stored by an DR-ICD
(b). 25 mm/sec. Top line: annotated marker channel, middle line: atrial channel, bottom line:
ventricular channel).
Figure 3: Prevalence of adjudicated tachyarrhythmias (stratified to < 130 bpm, 130-149 bpm,
and >149 bpm) during follow-up, according to classification made by the investigators or
steering committee. AF: atrial fibrillation, Afl: atrial flutter, ST: sinus tachycardia, SVT:
supraventricular tachyarrhythmia, VF: ventricular fibrillation, VT: ventricular tachycardia.
Figure 4: EGMs recorded by an A+-ICD with examples of complete atrial undersensing (a)
and intermittent atrial R-wave oversensing (b). 25 mm/sec. Top line: annotated marker
channel, middle line: atrial channel, bottom line: ventricular channel).
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from
by guest on May 29, 2018 http://circep.ahajournals.org/ Downloaded from
by guest on May 29, 2018 http://circep.ahajournals.org/ Downloaded from
by guest on May 29, 2018 http://circep.ahajournals.org/ Downloaded from
by guest on May 29, 2018 http://circep.ahajournals.org/ Downloaded from
by guest on May 29, 2018 http://circep.ahajournals.org/ Downloaded from
by guest on May 29, 2018 http://circep.ahajournals.org/ Downloaded from
Behrens and Michael NiehausChristian Sticherling, Markus Zabel, Sebastian Spencker, Udo Meyerfeldt, Lars Eckardt, Steffen
StudySystem in Patients without Antibradycardia Pacing Indications: Results of a Randomized Comparison of a Novel Single Lead Atrial Sensing (A+)-ICD System with a Dual Chamber ICD
Print ISSN: 1941-3149. Online ISSN: 1941-3084 Copyright © 2010 American Heart Association, Inc. All rights reserved.
Dallas, TX 75231is published by the American Heart Association, 7272 Greenville Avenue,Circulation: Arrhythmia and Electrophysiology
published online December 14, 2010;Circ Arrhythm Electrophysiol.
http://circep.ahajournals.org/content/early/2010/12/14/CIRCEP.110.958397World Wide Web at:
The online version of this article, along with updated information and services, is located on the
http://circep.ahajournals.org//subscriptions/
is online at: Circulation: Arrhythmia and Electrophysiology Information about subscribing to Subscriptions:
http://www.lww.com/reprints Information about reprints can be found online at: Reprints:
document. Permissions and Rights Question and Answerinformation about this process is available in the
requested is located, click Request Permissions in the middle column of the Web page under Services. FurtherCenter, not the Editorial Office. Once the online version of the published article for which permission is being
can be obtained via RightsLink, a service of the Copyright ClearanceCirculation: Arrhythmia and Electrophysiology Requests for permissions to reproduce figures, tables, or portions of articles originally published inPermissions:
by guest on May 29, 2018
http://circep.ahajournals.org/D
ownloaded from