horno secado dhg9140a, luzeren
DESCRIPTION
Manual de uso para equipo de laboratorio Horno de secado modelo DHG9140A, marca LUZERENTRANSCRIPT
Operation Manual of Drying Oven(Version 2013.01)
Written for
Model DHG9140A
THIS OPERATION MANUAL SHOULD BE CAREFULLY READ AHEAD OF USING
~lUZEREN
Summary
Provided to industrial and mining enterprises,laboratories and scientific resear
ch institutes for desiccation, torrefaction, wax-melting and sterilization.
Primary Technical CharacteristicsDHG9070A DHG9140A DHG9240A DHG9420A DHG9620A DHG9920A
Interior dimension:400*350*500 450*550*550 500*600*750 600*550*1300 800*600*1300 1000*600*1600
W*D*H(mm)Shelves(adjustable) 2PCS 2PCS 2PCS 3PCS 4PCS 5PCS
Temperature range RT+1O-250°C
Temperature0.1 °C/±0.5°C
precision/stability
Electrical110V 60Hz 380V 50Hz
requirementPower 1250W 1500W 2100W 4000W 4500W 6000W
Install and Adjust
1. Don't collide the body and make an effort on its glass in the middle of
transportation.
2. The equipments should be placed steady after falling to the ground.
3. Don't use the instrument in the place where is sunlight or heat damp.
4. The equipments should keep off the electromagnetism interference source,
and connect equipments a ground of line connect ground effectively.
5. The equipments while circulating normally, product inside the box's putting
to put should not run into inside the wall.
6. The power cable never mind depends a behind, also not let the instrument
or other products press on-line in the power supply in order to prevent
hurts power cable.
Usage
1. Open the door and need the processing articíe to put to put plank into the
box up, close the door.
2. Connect power supply, insert the power supply plug into the AC outlet, the
power switch of the front-panel place in the position of "open", the gauge
appears the numeral manifestation at this time, meaning the equipments
enters work appearance.
3. Pass the operation temperature controller, set you must of temperature
inside the box.
4. The instrument begins to work, temperature inside the box atlains initial
value gradually, time through the dry processing need after, handle a work
completion.
5. Close power supply, need temperature inside the box to open to put the
spirit valve after near to the environment temperature. Open the door, take
out article.
(1) Hint specially
1. Product didn't carry on correction generally through strict test before the
factory, is bad such as the environment of the hour of usage, the
environment temperature outran feat scope, will cause the temperature
manifestation value and actual temperature error margin inside the box, if
outrun the technique index sign scope of, can revise, concretely the step
according to the temperature controller operating instructions to press to
need to carry on correction.
2. Instrument is under the normal work appearance, if open a box of time to
lead to grow, after closing a box of door temporarily temperature inside the
box is some fluctuations, this is the normal phenomenon.
3. Atlention of testing: The thermometer that test inside the box use wants to
use 1.0 mercury thermometer of the accuracies, and put mercury to the
center in the box.
2
Controller Details
Panel FunctionCD PV display unitDisplays measured value or the
various characters depending the
statue of the instrument.Configuration of the Instrument Panel® SV display unitDisplays the set value or the timing
and rated parameters.
® IndicatorsRUN/AT
PV _11....----1 Working indicator (Green); Flashes
during auto-tuming execution.sv
OUTo O OOUT MK ALM
ORUN/AT
Heating output indicator (Green)
Tum on when outputs operate.
MK[YJ ~DEC/Restart INC/Mute
GJShift
~Set
lhis machine hasn't this kind of
function. (Door Control Function)
ALMALMalarming indicator (Red)
Tumed on when alarming output
operate and the buzzer sounds.
Operating
When it is switched on, the upper row displays "1 nP" and the low row
displays the range of temperature. After about 3 seconds, it starts running.
Press the temperature menu button "Set ", the controller runs into the
temperature I time setting state, the under window of controller displays the
symbol of temperature setting "SP", and the over window of the displays the
set value of temperature, the user can adjust the value according to the needs.
Re-press the button, the controller runs into the time setling state, the under
window of controller displays the symbol of time setting "SI:", the over window
displays the set value of time, the user can adjust the value according to
needs (0-9999) (If the working time is set as 0000, it indicates the timer is
inoperative, the controller will run continuously.) Press the button again, then
it returns to the running state. Press the menu button of temperature for 3
seconds, it runs into the internal value setting state; if press another 3 seconds,
it will return to the running state. In the above mentioned 2 states, if there is not
any operation for 30s, the controller will auto quit the above mentioned states
and the adjusted value will not be saved. When the runtime is over, the over
3
window of controller will display "End", the buzzer will sound for 30s, then it
can be muted by pressing the button " • ". When press the button "~"for 4s at
the time, the program will restart then.
When temperature is beyond "SP+RL ". the ALM indicator turns on.
• The method for changing set value
Press Set key the low row displays SP. Pressé," or "~"key, the upper
row displays the needed value. Again press Set key the low row displays Stkey. Press"." or "~" key, the upper row displays the timing time needed.
Again press Set key, the instrument return to the standard display mode.
• Timing function
When the value of St is setting to zero, the timing function is canceled.
Otherwise when the value of St is not setting to zero the timing function is
available.
When the power is turned on, the timing function starts. In case of
achieving the setting timing the heating output is close. and the buzzer sounds
four times to notice the operator. When the system is starting with auto tuning,
the timing function is also canceled. When the auto tuning process is
completed, the timing function would start again. During the operating period of
the instrument, it is allowed to update the value of St online, and the time
passed can be memorized and the instrument will operate the updated timing
time. When the updated time is less than the passed accumulative time, the
heating output close immediately, the buzzer sounds four times.
• The method for changing control parameter
After press Set key over 4 seconds, find L[. Then press " ." or "~"key to
the upper displayer display 3.Again press the Set key to find the symbol
character for the needed parameter. Through press the "." or "~"key till the
upper row displays display the parameter needed is reached. Some
parameters may be setting at once. Press the Set key over 4 seconds the
instrument returned to the standard mode. If no key is press down the
instrument return to its standard mode after 1 minute.
4
• The function of the controller Auto- Tuning
After pressing ••••key 6 seconds the AT indicator flashes and the
auto-tuning press starts; after the end of auto-tuning the AT indicator then be
turned out. A set of PIO parameters can be got automatically for overcoming
temperature overshoot. The controller operates according to this set of PIO
parameters is working continuously.
In the period of auto tuning, if presses Shift key beyond 6 seconds the AT
indicator turn out and the auto tuning process is canceled. The instrument
operated according to the original PIO parameters.
• Following table lists the function parameters
ParameterDisplay Name
SettingInstruction of the function
indicator range
Lc- L[ Password key When LC=3 ,then we can enter the next Oparameters listed.
P-- P ProportionalAdjustment of proportional function. 20.0~80.0
band
AlarmingWhen temperature is beyond "SP+AL",
AL- RL the ALM indicator turns on. The buzzer O.O~20.0setting
sounds and the heater output turns off.When the zero error comparatively
Pb Zero point larger, to update this value should bePb- -9.9-9.9
adjust needed.Pb=actual value- measure value
When the full point error also
Full pointcomparatively larger, to update this value
PK- PIr should be needed. -999~999adjust
PK=1000*(actual value-measure value)/measure value
rH- rH Range ofThe range of temperature setting. O.O~400.0
temp setting
5
Electrical circuit diagram
U 1 1Heater IlOVAC
Regulation
1. Chemistry product of the combustibility and volatility isn't put absolutely
into the box.
2. If appear circumstances, such as abnormality, smell and smoke ...etc.
during the period of use, please turn off power immediately, the customer
shouldn't absolutely to fix blindly, should notify our company, it is looked
into to fix by the professional personnel.
3. The box inner part and the equipments surface want to usually wipe to
sweep by keep, the clarity of the increment glass. Don't use sour, the alkali
or other causticity aqua to wipe outward appearance noodles.
4. The equipments don't work for long-term, it should be pulled out power
cable in order to prevent the equipments hurt a persono Should press the
usage condition to circulate periodically (general the quarterly degree) for
2-3 days also to banish the dampness of the electric appliances part, avoid
a relevant spare part of damage.
Packing ListNO. Description of goods Quantity Remarks1 Drying Oven 1
6
Protection to Fix CardThank you very much for using our products
Product Name:Product Model Number:Product Trouble:
Customer Name:Company Name:Customer Phone:Customer Address:Post Codes:E-mail:
(Purchase Date)
Serial Number 52164120061
Please tell the serial number to our product engineer for repairing.
Protection to Fix ElucidationAccording to the prescription, the period for protecting to fix our company'sproduct is one year. (From the day when you purchase the product) During thisperiod, any belongs to under normal usage circumstance cause because of theproduct quality's problem of breakdown, our company will be responsible forgiving free maintain.
This couplet is considered as the protection to fix warrantees, so please reservethem carefully.
~lUZEREN