how to fix it – top 10 support calls afhcan training module
TRANSCRIPT
![Page 1: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/1.jpg)
How to Fix It –Top 10 Support Calls
AFHCAN Training Module
![Page 2: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/2.jpg)
Number 10• “Error getting groups and providers, AFHCAN
software has timed out...”– Logon to cart as an administrator to get to the
desktop (red background)– Browse to: C:\Program Files\AFHCAN Telehealth
Client – Open the "ATSClient.exe.config" file using Notepad– Find the following 2 lines:
• <add key="ClientGui.DefaultResponseTimeOut" value="180" />
• <add key="ClientGui.QueryResponseTimeOut" value="180" />
– Verify or set the value to 180– Save and close the file– If the problem persists increase the value as needed– Contact your technical support to have them contact
AFHCAN Customer Support if necessary
![Page 3: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/3.jpg)
Number 9
• “The otoscope lightbox is not bright enough.”– Be sure the light control knob is
turned all the way to the right– Check the bulb – is it seated
exactly in the notch?– Be sure light cable is not
kinked/bent/worn out
![Page 4: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/4.jpg)
![Page 5: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/5.jpg)
![Page 6: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/6.jpg)
TW = tension wire
CB = ceramic base
EC = electrical connection
![Page 7: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/7.jpg)
Number 8
• “How do I open the web interface at my workstation?”– Obtain the IP address of the AFHCAN
server from your technical support– Launch Internet Explorer or other web
browser– Enter the IP address in the address
location bar followed by “/afhcan” - example:
• http://161.223.135.111/afhcan
– Login as normal
![Page 8: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/8.jpg)
Number 7
• “How do I send and/or reuse an archived case?”– Cannot send or reuse an archived
case– Can only view and add comments– Can reuse images in an archived
case
![Page 9: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/9.jpg)
Number 6
• “How do I re-use existing images in an archived case?”– Search archives to find your case– Click on image to bring into main
viewing screen– Right-click on image and choose “Save-as”
• Save to external media– Diskette– USB
![Page 10: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/10.jpg)
Number 5
• “I cannot print”– Check all cabling– Reset the wireless printer
• Unplug from wall, wait 15 seconds and plug back in
– Restart the cart– Known bug in v3.4 with PEFs
• Fixed in v4.x
![Page 11: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/11.jpg)
Number 4
• “An error occurred that caused the software to quit running...”
– Reset AFHCAN software• Have your IT support contact
AFHCAN Customer Support for steps involved
![Page 12: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/12.jpg)
![Page 13: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/13.jpg)
Number 3
• “Where did my case go?”– Verify case was sent to correct location
• Search all Open Cases– Remote location?
– Local location?
– Verify case is not on Alert• Sys Admin screen
– If case is on Alert• Contact your IT support group
– Check Activity Log
![Page 14: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/14.jpg)
Number 2
• “I cannot log into the AFHCAN software.”– New provider?
• Add provider info
– Existing provider?• Verify provider identity• Correct provider id?• Reset password
– Is AFHCAN software “Working Off-line”?• Contact your IT support group
![Page 15: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/15.jpg)
Number 1
• “What is my password?”– Verify provider identity– Reset password
– Login as administrator– Select Sys Admin– Find provider– Highlight password and type in new
password – you will have to verify this by typing it in twice
– Click apply
![Page 16: How to Fix It – Top 10 Support Calls AFHCAN Training Module](https://reader036.vdocument.in/reader036/viewer/2022062314/56649eff5503460f94c1549f/html5/thumbnails/16.jpg)
Questions ?