human growth factor panel - legendplex™
TRANSCRIPT
®
LEGENDplex™Mul�-Analyte Flow Assay Kit
For Accurate Quantification of Multiple Human Th(T helper Cell) Cytokines from Cell Culture Supernatant,
Serum, Plasma and Other Biological Samples
Please read the entire manual before running the assay
BioLegend.com
®
LEGENDplex™Mul�-Analyte Flow Assay Kit
Cat. No. 740180
Human Growth Factor Panel(13-plex )
Please read the entire manual before running the assay.
BioLegend.com
For Research Purposes Only. Not for use in diagnostic or therapeutic procedures. Purchase does not include or carry the right to resell or transfer this product either as a stand-alone product or as a component of another product. Any use of this product other than the permitted use without the express written authorization of BioLegend is strictly prohibited.
It is highly recommended that this manual be read in its entirety before using this product. Do not use this kit beyond the expiration date.
biolegend.com 1
LEGENDplex™ Human Growth Factor Panel
Table of Contents Page
Chapter 1: KIT DESCRIPTION..................................................
Introduction……………………………………………..........................
PrincipleoftheAssay……………………....……………....….…......
BeadsUsage...........................................………..……………...
StorageInformation…………………………………….......…..........
MaterialsSupplied………………….....……………….................…
MaterialstobeProvidedbytheEnd-User……...........……...
Precautions.................................……………………................
Chapter 2: ASSAY PREPARATION.............................................
SampleCollectionandHandling…………………………............
ReagentPreparation…………………………………………...............
StandardPreparation..........................................................
SampleDilution…………...........…….......................................
Chapter 3: ASSAY PROCEDURE..................................................
PerformingtheAssayUsingaFilterPlate……………….........
PerformingtheAssayUsingMicrotubes……………….............
Chapter 4: FLOW CYTOMETER SETUP.......................................
Setup Procedure for FACSCalibur with Dual Laser..............
Setup Procedure for FACSCalibur with a Single Laser........
Setup Procedure for BD FACSAriaTM, FACSCantoTM and LSR Series..........................................................................
Setup Procedure for Other Flow Cytometers ...................
Chapter 5: DATA ACQUISITION AND ANALYSIS.........................
DataAcquisition..................................................................
Data Analysis.....................................................................
3
3
3
4
6
6
7
8
9
9
9
10
11
12
12
15
18
19
23
30
34
35
35
36
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
2
Chapter 6: FREQUENTLY ASKED QUESTIONS...........................
Chapter 7: ASSAY CHARACTERIZATION..........................................
RepresentativeStandardCurve.………………………………........
37
41
41
41
42
42
43
44
45
46
50
53
54
AssaySensitivity...……………………………………………………..…..
Cross-Reactivity……………………………………………………..........
Accuracy.............................................................................
LinearityofDilution………………………………………………..........
Intra-AssayPrecision……………………………………...................
Inter-AssayPrecision……………………………………...................
BiologicalSamples…………………………………………….………....
TROUBLESHOOTING...........................……………………………………...
PLATE MAP...............……………………………………………………...............
RACK MAP....................................................................................
biolegend.com 3
LEGENDplex™ Human Growth Factor Panel
Chapter 1: KIT DESCRIPTION
Introduction
Growthfactorsactivatecellproliferationanddifferentiationandregulatemanycellular processes including angiogenesis, hematopoiesis, tumorigenesis, me-tabolism,inflammationandtissuerepair.Theycouldalsobepotentialthera-peutictargetsanddiseasebiomarkers.Theaccuratemeasurementofgrowthfactorsiscriticalforabetterunderstanding of disease progression and related cellular processes.
The LEGENDplexTM Human Growth Factor Panel is a bead-based multiplexassay,usingfluorescence–encodedbeadssuitableforuseonvariousflowcytometers.Thispanelallowssimultaneousquantificationof13humangrowthfactors:An-giopoietin-2(Ang-2),EGF,EPO,FGF-basic,G-CSF,GM-CSF,HGF,M-CSF,PDGF-AA,PDGF-BB,SCF,TGF-αandVEGF.Thispanelprovideshighersensitivityandbroad-erdynamicrangethantraditionalELISAmethods.Thepanelhasbeenvalidatedfor use on serum, plasma and cell culture supernatant samples.
The Human Growth Factor Panelisdesignedtoallowflexiblecustomizationwith-in the panel. For mix and match within the panel, please visit www.biolegend.com/legendplex.
This assay is for research use only.
Principle of the Assay
BioLegend’s LEGENDplexTM assays are bead-based immunoassays using the same basic principle as sandwich immunoassays.
Beadsaredifferentiatedbysizeandinternalfluorescenceintensities.Eachbeadsetisconjugatedwithaspecificantibodyonitssurfaceandservesasthecap-turebeadsforthatparticularanalyte.Whenaselectedpanelofcapturebeadsismixedandincubatedwithasamplecontainingtargetanalytesspecifictothecaptureantibodies,eachanalytewillbindtoitsspecificcapturebeads.Afterwashing,abiotinylateddetectionantibodycocktailisadded,andeachdetec-tionantibodyinthecocktailwillbindtoitsspecificanalyteboundonthecap-turebeads,thusformingcapturebead-analyte-detectionantibodysandwiches.Streptavidin-phycoerythrin(SA-PE)issubsequentlyadded,whichwillbindtothebiotinylateddetectionantibodies,providingfluorescentsignalintensitiesinproportiontotheamountofboundanalytes.
Sincethebeadsaredifferentiatedbysizeandinternalfluorescenceintensityonaflowcytometer,analyte-specificpopulationscanbesegregatedandquanti-fiedbythePEfluorescentsignal.Theconcentrationofaparticularanalyteisdetemined by a standard curve generated in the same assay.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
4
Beads Usage
The Human Growth Factor Panel includes two sets of beads. Each set has a uniquesizethatcanbeidentifiedonflowcytometerbasedontheirforwardscatter(FSC)andsidescatter(SSC)profiles(BeadsAandBeadsB,Figure1).Eachbeadsetcanbefurtherresolvedbasedontheirinternalfluorescenceintensities.TheinternaldyecanbedetectedusingFL3,FL4,orAPCchannel,dependingonthetypeofflowcytometerused.ThesmallerBeadsAconsistsof6beadpopulationsandthelargerBeadsBconsistsof7beadpopulations(Figure2-3).
Usingatotalof13beadpopulationsdistinguishedbysizeandinternalfluores-centdye,theHumanGrowthFactorPanelallowssimultaneousdetectionof13growthfactorsinonesampletest.Eachanalyteisassociatedwithaparticularbeadsetasindicated(Figures2-3andTable1).
Figure1.BeadsDifferentiatedbySize
Beads A = smaller beads
Beads B = larger beads
Figure2.BeadsAClassificationbyFL4
A5 A7 A8
A4
A6
A10
biolegend.com 5
LEGENDplex™ Human Growth Factor Panel
Figure3.BeadsBClassificationbyFL4
For Beads usage in the panel, please refer to Table 1 below:
Table1.BeadsIDandTargetInformation
Target Bead ID Top Standard Concentrations(ng/mL)
HumanAngiopoietin-2 A4 50
Human EGF A5 10
Human EPO A6 50
Human FGF-basic A7 50
Human G-CSF A8 50
Human GM-CSF A10 50
Human HGF B2 50
Human M-CSF B3 10
Human PDGF-AA B4 50
Human PDGF-BB B5 50
Human SCF B6 10
HumanTGF-α B7 50
HumanVEGF B9 50
*BeadIDisusedtoassociateabeadpopulationtoaparticularanalyteintheLEGENDplexTMDataAnalysisSoftware.TheassociationofanalyteandbeadIDwillbedefinedduringthegatingstepofthedataanalysis.
WhenenteringanalyteandbeadIDinfomationduringthegatingstep,alwaysenterinthesequentialorderofthebeadID(e.g,A4,A5,A6...B2,B3,B4...).Please refer to the LEGENDplexTMDataAnalysisSoftwareUserGuideandOnlineHelpfordetails(www.biolegend.com/legendplex).
B4 B5
B6 B7
B3
B9
B2
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
6
StorageInformation
Recommendedstorageforalloriginalkitcomponentsisbetween2°Cand8°C.DONOTFREEZEPre-mixedBeads,DetectionAntibodiesorSA-PE.
• Oncethestandardshavebeenreconstituted,immediatelytransfercon-tents into polypropylene vials. DO NOT STORE RECONSTITUTED STAN-DARDSINGLASSVIALS.
• Uponreconstitution,leftoverstandardandMatrixBshouldbestoredat≤-70°Cforusewithinonemonth.Avoidmultiple(>2)freeze-thawcycles.Discardanyleftoverdilutedstandards.
Materials Supplied
The LEGENDplexTMkitcontainsreagentsfor100testslistedinthetablebelow.Whenassayedinduplicate,thisisenoughforan8-pointstandardcurveand40samples.
Kit Components Quantity Volume Part #
Setup Beads 1: FITC Beads 1 vial 1 mL 77840
Setup Beads 2: PE Beads 1 vial 1 mL 77842Setup Beads 3: Raw Beads 1 vial 2 mL 77844Human Growth Factor Panel Pre-mixed Beads 1bottle 3.5 mL 76289
Human Growth Factor Panel Detec-tionAntibodies 1bottle 3.5 mL 76303
Human Growth Factor Panel Stan-dardCocktail,Lyophilized 1 vial lyophilized 76318
LEGENDplexTM SA-PE 1bottle 3.5 mL 77743LEGENDplexTMMatrixB,Lyophilized 1 vial lyophilized 77549LEGENDplexTMAssayBuffer 1bottle 25 mL 77562LEGENDplexTMWashBuffer,20X 1bottle 25 mL 77564Filter plate 1 plate 76187Plate Sealers 4 sheets 78101DataAnalysisSoftwareDongle 1 21217Human Growth Factor Panel Manual 1 76319
biolegend.com 7
LEGENDplex™ Human Growth Factor Panel
Materials to be Provided by the End-User
• Aflowcytometerequippedwithtwolasers(e.g.,a488nmbluelaseror532nmgreenlaseranda633-635nmredlaser)capableofdistinguishing575 nm and 660 nm oraflowcytometerequippedwithonelaser(e.g.,488nmbluelaser)capableofdistinguishing575nmand670nm.
Partiallistofcompatibleflowcytometers:
Flow Cytometer
Reporter Channel
ChannelEmission
ClassificationChannel
Channel Emission
Compensa-tionneeded?
BD FACSCaliburTM
(single laser)FL2 575 nm FL3 670 nm Yes
BD FACSCaliburTM
(dual laser)FL2 575 nm FL4 660 nm No*
BD FACSArrayTM Yellow 575 nm Red 660 nm No*
BD FACSCantoTM
BD FACSCantoTM IIPE 575 nm APC 660 nm No*
BDTMLSR,LSRIIBD LSRFortessaTM PE 575-585
nm APC 660 nm No*
BD FACSAriaTM PE 575 nm APC 660 nm No*
*Compensationisnotrequiredforthespecifiedflowcytometerswhensetupproperly,butisrecommendedforconsistentresults.
Forsettinguptheaboveflowcytometers,pleasefollowtheFlow Cytom-eter Setup guide in this manual or visit: www.biolegend.com/legendplex.
Forflowcytometersnotlistedhere,theend-userneedstosetupthemachine following similar guidelines. Please refer to Setup Procedure for Other Flow CytometerssectioninChapter4.
• Multichannelpipettescapableofdispensing5μLto200μL
• Reagentreservoirsformultichannelpipette
• Polypropylenemicrofugetubes(1.5mL)
• Laboratory vortex mixer
• Sonicatorbath(e.g.,BransonUltrasonicCleanermodel#B200,orequiva-lent)
• Aluminum foil
• Absorbent pads or paper towels
• Plateshaker(e.g.,Lab-LineInstrumentsmodel#4625,orequivalent)
• Tabletopcentrifuges(e.g.,Eppendorfcentrifuge5415C,orequivalent)
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
8
Iftheassayisrunusingthefilterplateprovided(recommended),
• Avacuumfiltrationunit(MilliporeMultiScreen®HTSVacuumManifold,catalog#MSVMHTS00orequivalent).Instructionsonhowtousethevacuum manifold can be found at www.biolegend.com/legendplex.
• Avacuumsource(minivacuumpumporlinevacuum,e.g.,MilliporeVacuumPump,catalog#WP6111560,orequivalent)
Iftheassayisruninmicrotubes,V-orU-bottom96-wellplate(optional),
• 1.1mLpolypropylenemicroFACStubes,in96-tuberack(e.g.,NationalScientificSupplyCo,catalog#TN0946-01R,orequivalent).
• Centrifugewithaswingingbucketadaptorformicrotiterplatesormicro-tuberacks(e.g.,BeckmanCoulterAllegraTM 6R Centrifuge with MICROPLUS CARRIERadaptorforGH3.8andJS4.3Rotors).
• 96-well,V-bottom,clearpolypropylenemicroplate(lowproteinbinding,e.g.,grenierbio-one,Catalog#651201orequivalent).
• Precautions
• All blood components and biological materials should be handled as poten-tiallyhazardous.FollowuniversalprecautionsasestablishedbytheCenterforDiseaseControlandPreventionandbytheOccupationalSafetyandHealthAdministrationwhenhandlinganddisposingofinfectiousagents.
• Sodiumazidehasbeenaddedtosomereagentsasapreservative.Al-thoughtheconcentrationsarelow,sodiumazidemayreactwithleadandcopperplumbingtoformhighlyexplosivemetalazides.Ondisposal,flushwithalargevolumeofwatertopreventazidebuild-up.
• Matrix B for LEGENDplexTMkitscontainscomponentsofhumanoriginandshouldbehandledaspotentiallyhazardous.TherawmaterialhasbeenscreenedforinfectiousdiseasesandisnegativeforHIV,HBVandHCVusingFDA-approved test methods.
• Donotmixorsubstitutereagentsfromdifferentkitsorlots.Reagentsfromdifferentmanufacturersshouldnotbeusedwiththiskit.
• Donotusethiskitbeyonditsexpirationdate.
• SA-PEandPre-mixedBeadsarelight-sensitive.Minimizelightexposure.
biolegend.com 9
LEGENDplex™ Human Growth Factor PanelChapter 2: ASSAY PREPARATION
SampleCollectionandHandling
PreparationofSerumSamples:
• Allow the blood to clot for at least 30 minutes and centrifuge for 10 min-utes at 1,000 x g.
• Remove serum and assay immediately or aliquot and store samples at ≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
• Whenusingfrozensamples,itisrecommendedthatsamplesarethawedcompletely,mixedandcentrifugedtoremoveparticulatespriortouse.
PreparationofPlasmaSamples:
• PlasmacollectionusingEDTAasananti-coagulantisrecommended.Centri-fuge for 10 minutes at 1,000 x gwithin30minutesofbloodcollection.
• Remove plasma and assay immediately, or aliquot and store samples at ≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
• Whenusingfrozensamples,itisrecommendedthatsamplesarethawedcompletely,mixedwellandcentrifugedtoremoveparticulates.
PreparationofTissueCultureSupernatant:
• Centrifuge the sample to remove debris and assay immediately. If not pos-sible,aliquotandstoresamplesat≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
ReagentPreparation
PreparationofAntibody-ImmobilizedBeads
SonicatethePre-mixedBeadsbottlefor1minuteinasonicatorbathandthen vortex for 30 seconds prior to use. If no sonicator bath is available, in-creasethevortexingtimeto1minutetocompletelyresuspendthebeads.
PreparationofWashBuffer
• Bringthe20XWashBuffertoroomtemperatureandmixtobringallsaltsintosolution.
• Dilute25mLof20XWashBufferwith475mLdeionizedwater.Storeun-usedportionsbetween2°Cand8°Cforuptoonemonth.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
10
PreparationofMatrixB(forSerumorPlasmaSamplesOnly)
•Add 5.0 mL LEGENDplexTMAssayBuffertothebottlecontaininglyophilized MatrixB.Allowatleast15minutesforcompletereconstitution.Vortexto
mixwell.LeftoverreconstitutedMatrixBshouldbestoredat≤-70°Cforupto one month.
StandardPreparation
1. Priortouse,reconstitutethelyophilizedHumanGrowthFactorPanelStan-dardCocktailwith250µLAssayBuffer.
2. Mix and allow the vial to sit at room temperature for 10 minutes, and then transfer the standard to an appropriately labeled polypropylene microfuge tube. This will be used as the top standard C7.
Note:Analytesinthispanelhavedifferenttopstandardconcentrations(seetable1,page5fordetails).
3. Label 6 polypropylene microfuge tubes as C6, C5, C4, C3, C2 and C1, re-spectively.
4. Add75µLofAssayBuffertoeachofthesixtubes.Prepare1:4dilutionofthetopstandardbytransferring25µLofthetopstandardC7totheC6tube and mix well. This will be the C6 standard.
5. Inthesamemanner,performserial1:4dilutionstoobtainC5,C4,C3,C2andC1standards(see the table below).AssayBufferwillbeusedasthe0pg/mLstandard(C0).
Tube/Standard
ID
Serial Dilution
Assay Buffertoadd (µL)
Standard to add
Final Conc.
(pg/mL) *
Final Conc. (pg/mL) **
C7 -- -- -- 10,000 50,000
C6 1:4 75 25 µLofC7 2,500 12,500
C5 1:16 75 25 µLofC6 625 3,125
C4 1:64 75 25 µLofC5 156.3 781.3
C3 1:256 75 25 µLofC4 39.1 195.3
C2 1:1024 75 25 µLofC3 9.8 48.8
C1 1:4096 75 25 µLofC2 2.4 12.2
C0 -- 75 -- 0 0
*TopstandardconcentrationofEGF,M-CSFandSCF is 10 ng/mL; **TopstandardconcentrationofAngiopoietin-2,EPO,FGF-basic,G-CSF,GM-CSF,
HGF,PDGF-AA,PDGF-BB,TGF-α,VEGFis50ng/mL.
biolegend.com 11
LEGENDplex™ Human Growth Factor Panel
SampleDilution
• Serumorplasmasamplesmustbediluted2-foldwithAssayBufferbeforetesting(e.g.dilute50µLofsamplewith50µLofAssayBuffer).
Iffurthersampledilutionisdesired,dilutionshouldbedonewithMatrixBto ensure accurate measurement.
Addingserumorplasmasampleswithoutdilutionwillresultinlowas-sayaccuracyandpossibly,cloggingofthefilterplate.
• For cell culture supernatant samples, the levels of analyte can vary greatly fromsampletosample.Whilethesamplescanbetestedwithoutdilu-tions,apreliminaryexperimentmayberequiredtodeterminetheappro-priatedilutionfactor.
Ifsampledilutionisdesired,dilutionshouldbedonewithcorrespondingfresh cell culture medium to ensure accurate measurement.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
12
Chapter 3: ASSAY PROCEDURE
The LEGENDplexTM assaycanbeperformedeitherinafilterplate,inmicrotubes,orinaV-orU-bottommicroplate.
• Thein-filterplateassayprocedureishighlyrecommendedduetoitsgoodsample to sample consistency, assay robustness and ease of handling. This procedurerequiresavacuumfiltrationunitforwashing(seeMaterials to beProvidedbytheend-user,Page7).IfyouhaveperformedaLuminex®-basedmultiplexassaybefore,yourlabshouldalreadyhavethevacuumfiltrationunitsetup.
• Ifthein-filterplateassayprocedureisnotpossible,orifyouprefer,theassaycanbeperformedinmicrotubesorinaV-orU-bottommicroplate.Forin-tubeassay,werecommendusingmicroFACStubes(seeMaterials to be Providedbytheend-user,Page7).
Performing the Assay Using a Filter Plate
• Allowallreagentstowarmtoroomtemperature(20-25°C)beforeuse.
• Setthefilterplateonaninvertedplatecoveratalltimesduringassaysetupandincubationstepssothatthebottomoftheplatedoesnottouchanysurface.Touchingasurfacemaycauseleakage.
• Keeptheplateuprightduringtheentireassayprocedure,includingthewashing steps, to avoid losing beads.
• Theplateshouldbeplacedinthedarkorwrappedwithaluminumfoilforallincubationsteps.
• Standards and samples should be run in duplicate and arranged on the plateinaverticalconfigurationconvenientfordataacquisitionandanaly-sis(asshowninattachedPLATEMAP,Page53-54).Besuretoloadstan-dardsinthefirsttwocolumns.Ifanautomationdeviceisusedforread-ing,theorientationandreadingsequenceshouldbecarefullyplanned.
1. Pre-wettheplatebyadding100μLofLEGENDplexTM1XWashBuffertoeach well and let it sit for 1 minute at room temperature. To remove the excess volume, place the plate on the vacuum manifold and apply vacuum. Donotexceed10”Hgofvacuum.Vacuumuntilwellsaredrained(5-10seconds).BlotexcessWashBufferfromthebottomoftheplatebypress-ingtheplateonastackofcleanpapertowels.Placetheplateontopoftheinverted plate cover.
For measuring cell culture supernatant samples:
• Add25µLoffreshcellculturemedium(nottheTCsamples)tostan-dard wells.
biolegend.com 13
LEGENDplex™ Human Growth Factor Panel
• Add25µLofAssayBuffertosamplewells.• Add25µLofeachstandardtothestandardwells.• Add25µLofeachsampletothesamplewells(SeeDilutionof
Samples)
For measuring serum or plasma samples:
• Add25µLofMatrixBtothestandardwells.• Add 25 µLofAssayBuffertothesamplewells.• Add25µLofeachstandardtothestandardwells.• Add25µLofeachdilutedserumorplasmasampletothesample
wells(SeeDilutionofSamples).
2. Vortexthepre-mixedbeadsbottlefor30seconds.Add25μLofthepre-mixedbeadstoeachwell.Thevolumeshouldbe75μLineachwellafterbeadsaddition.(Note:Duringadditionofthebeads,shakethepre-mixedbeadsbottleintermittentlytoavoidbeadsettling).
3. Seal the plate with a plate sealer. Toavoidplateleaking,donotapplypositivepressuretothesealerwhensealingtheplate.Wraptheentireplate, including the inverted plate cover, with aluminum foil. Place the plateonaplateshaker,secureitandshakeatapproximate500rpmfor2hours at room temperature.
4. Do not invert the plate! Place the plate on the vacuum manifold and apply vacuumasbeforeinStep1.Add200µLof1XWashBuffertoeachwell.RemoveWashBufferbyvacuumfiltration.BlotexcessWashBufferfromthebottomoftheplatewithanabsorbentpadorpapertowels. Repeat this washing step once more.
5. Add25µLofDetectionAntibodiestoeachwell.
6. Sealtheplatewithafreshplatesealer.Wraptheentireplate,includingtheinvertedplatecover,withaluminumfoil.Placetheplateonaplateshakerandshakeatapproximately500rpmfor1houratroomtemperature.
7. Do not vacuum!Add25µLofSA-PEtoeachwelldirectly.
8. Sealtheplatewithafreshplatesealer.Wraptheentireplate,includingtheinvertedplatecover,withaluminumfoil.Placetheplateonaplateshakerandshakeatapproximate500rpmfor30minutesatroomtemperature.
9. Repeat step 4 above.
10. Add200µLof1XWashBuffertoeachwell(or150µLifplateistobereadbyanautosampler.Highervolumemayresultinleakingofthefilterplateduring reading. Besuretosetthesamplevolumetobeanalyzedto70uLor less so that sample can be read again if needed).Resuspendthebeadsonaplateshakerfor1minute.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
14
11. Readsamplesonaflowcytometer,preferablywithinthesamedayoftheassay (Note:Prolongedsamplestoragecanleadtoreducedsignal)
Iftheflowcytometerisequippedwithanautosampler,readtheplatedirectly using the autosampler. The probe height may need to be adjusted when using an autosampler.
If an autosampler is not available, the samples need to be transferred from thefilterplatetoFACStubesandreadmanually.Inthiscase,thesamplevolumemayneedtobeincreasedfrom200µLto300µLbyaddingextra100µLof1XWashBuffertoeachtubetoavoidsamplerunningdrywhenreadonaflowcytometer.
Assay Procedure Summary for Filter PlateAdd 100 μL 1X Wash Bu�er to �lter plate wells
Vacuum to remove excess bu�er
Incubate 2 hours, RT, shaking
Capture beads
Biotinylated Detection Antibody
Analytes
Without washing, add 25 μL SA-PEIncubate 30 min, RT, shaking
Wash 2 times using vacuum �ltration unitAdd 25 μL Detection AntibodiesIncubate 1 hr, RT, shaking
For cell culture supernatant samples, Add to the plate:25 μL Cell Culture Medium to standard wells25 μL Assay Bu�er to sample wells25 μL diluted standard to standard wells25 μL sample to sample wells25 μL pre-mixed beads to all wells
Wash 2 times using vacuum �ltration unitAdd 200 μL (or 150µL) of 1x Wash Bu�er Read on a �ow cytometer
For serum and plasma samples,Add to the plate:25 μL Matrix B to standard wells25 μL Assay Bu�er to sample wells25 μL diluted standard to standard wells25 μL sample to sample wells25 μL pre-mixed beads to all wells
BA
C
A B C
A B C
biolegend.com 15
LEGENDplex™ Human Growth Factor Panel
Performing the Assay Using Microtubes or a 96-Well V- or U- bot-tom Microplate
• Allowallreagentstowarmtoroomtemperature(20-25°C)beforeuse.
• Determine the number of assays to run. Standards and samples should be runinduplicateandarrangedintheorderconvenientfordataacquisitionand analysis (see below). Ifanautomationdeviceisusedforreading,theorientationandreadingsequenceshouldbecarefullyplanned.
Ifusingtubes,labelalltubesandarrangetubesonarack(asshowninat-tached RACK MAP, assuming using micro FACS tubes and a 96-microtube rack,e.g.,NationalScientificSupplyCo.,Catalog#:TN0946-01R).A96-mi-crotuberackisrecommendedbecauseitallowspipettingwithamultichan-nelpipette.
If using a microplate, arrange the standard and samples according to the PLATEMAPattached.
1. For measuring cell culture supernatant samples:• Add25µLoffreshcellculturemedium(nottheTCsamples) to stan-
dardtubes/wells.• Add25µLofAssayBuffertosampletubes/wells.• Add25µLofeachstandardtothestandardtubes/wells.• Add25µLofeachsampletothesampletubes/wells(See Dilutionof
Samples).• Add25µLofthepre-mixedbeadstoalltubes/wells.• Add25µLDetectionAntibodiestoalltubes/wells.
For measuring serum or plasma samples:
• Add25µLofMatrixBtothestandardtubes/wells.• Add25µLofAssayBuffertosampletubes/wells.• Add25µLofeachstandardtostandardtubes/wells.• Add25µLofeachdilutedserumorplasmasampletosampletubes/
wells(See DilutionofSamples).• Add25µLofthepre-mixedbeadstoalltubes/wells.• Add25µLDetectionAntibodiestoalltubes/wells.
Note: Thevolumeshouldbe100μLineachtube/well.Shakebeadsbottleintermittentlyduringtheadditiontoavoidbeadsettling.
2. Covertheentirerack/platewithaluminumfoiltoprotectthetubes/platefrom light. Shake(approximate1,000rpmforin-tubeassayand600rpmforin-plateassay)onaplateshakerfor2hoursatroomtemperature.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
16
3. Withoutwashingthetubes/plate,add25µLSA-PEtoeachtube/wells.
4. Covertheentirerack/platewithaluminumfoiltoprotectthetubes/platefrom light. Shake(approximate1,000rpmforin-tubeassayand600rpmforin-plateassay)onaplateshakerfor30minutesatroomtemperature.
5. Centrifugethetubes/plateat1,000xg for 5 minutes, using a swinging bucketrotorwithmicroplateadaptor(Please refer to Materials to be Pro-videdbytheend-user,Page7).
6. Removethesupernatantusingamicrochannelpipette.Becarefulnottoremove beads, but to remove as much liquid as possible.
7. Add200µLof1XWashBuffertoalltubes/well.Resuspendthebeadsbyvortexing(forin-tubeassay)orshakingonaplateshakerfor1minute(forin-plateassay).Centrifugethetubes/plateat1,000xg for 5 minutes, using aswingingbucketrotorwithmicroplateadaptor.Removethesupernatant.
8. Add200µLof1XWashBuffertoalltube/well(or 150 µL if plate is to be readwithanautosampler.Highervolumemayresultinleakingofthefil-terplateduringreading.Besuretosetthesamplevolumetobeanalyzedto 70 uL or less so that sample can be read again if needed).Resuspendthebeadsbyvortexing(forin-tubeassay)orpippeting(forin-plateassay).
9. Readsamplesonaflowcytometer,preferablywithinthesamedayoftheassay(Note:Prolongedsamplestoragecanleadtoreducedsignal).
Iftheflowcytometerisequippedwithanautosampler,thesamplescanbereadeitherdirectly(forin-plateassay)orafterbeingtransfferedtoa96-wellplate(forin-tubeassay).The probe height may need to be adjusted when using an autosampler.
If an autosampler is not available, the samples can be read directly in FACS tubesorafterbeingtransferredfromthemicroplatetomicroFACStubes.Inthiscase,thesamplevolumemayneedtobeincreasedfrom200µLto300µLbyaddingextra100µLof1XWashBuffertoeachtube,toavoidsamplerunningdrywhenanalyzedonaflowcytometer.
biolegend.com 17
LEGENDplex™ Human Growth Factor Panel
Assay Procedure Summary for Tubes (or V-Bottom Plate)
Arrange the number of tubes needed on a rack (or set up the plate)
Incubate 2 hours, RT, shaking
Without washing, add 25 μL SA-PEIncubate 30 min, RT, shaking
For cell culture supernatant samples, Add to the tubes/wells:25 μL Cell Culture Medium to standard tubes/wells25 μL Assay Bu�er to sample tubes/wells25 μL diluted standard to standard tubes/wells25 μL sample to sample tubes/wells25 μL pre-mixed beads to all tubes/wells25 μL Detection Antibodies to all tubes/wells
Spin down beadsWash one timeAdd 200 μL (or 150µL) of 1x Wash Bu�erRead on a �ow cytometer
For serum and plasma samples, Add to the tubes/well:25 μL Matrix B to standard tubes/wells25 μL Assay Bu�er to sample tubes/wells25 μL diluted standard to standard tubes/wells25 μL sample to sample tubes/wells25 μL pre-mixed beads to all tubes/wells25 μL Detection Antibodies to all tubes/wells
A B
C
A
B
C
Capture beads
Detection Antibody
Analytes
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
18
Chapter 4: FLOW CYTOMETER SETUP
Inordertogeneratereliabledata,theflowcytometermustbesetupproperlybeforedataacquisition.Thefollowingsectionswilladdressmachinesetupfortheflowcytometerslistedbelow.
ListofFlowCytometersandPossibleConfigurations
Flow CytometerReporterChannel
Channel Emission
ClassificationChannel
ChannelEmission
Compensationneeded?
BD FACSCaliburTM
(singlelaser) FL2 575 nm FL3 670 nm Yes
BD FACSCaliburTM
(duallaser) FL2 575 nm FL4 660 nm No*
BD FACSCantoTM,BD FACSCantoTM II PE 575 nm APC 660 nm No*
BDTM LSR, LSRII,BD LSRFortessaTM PE 575-585
nm APC 660 nm No*
BD FACSAriaTM PE 575 nm APC 660 nm No*
*Compensationisnotrequiredforthespecifiedflowcytometerswhensetupproperly,butisrecommendedforconsistentresults.
Forflowcytometersnotlistedhere,theend-userneedstosetupthemachinefollowing similar guidelines. Please refer to Setup Procedure for Other Flow Cytometerssectioninthischapter.
The setup process typically includes the following steps. Please see the detailed setupprocedurethatfollows,regardingyourspecificinstrument.
1).Startuptheinstrumentfollowingthemanufacturer’srecommendations.
2).Createatemplatefordataacquisitionusingyourinstrument’sdata acquisitionsoftware.Atemplateisadocumentorworksheetwithdensityplotsthatallowstheusertoperformmachinesetupanddataacquisition.
3).SetupthePMTvoltagesofeachchanneltobeusedfordataacquisition usingtheSetupBeadsprovidedinthekit.
4).DeterminewhethercompensationisneededbasedontheconfigurationofyoursystemasshowninTable1.Ifcompensationisneeded,perform compensationusingtheSetupBeadsprovidedinthekit.
biolegend.com 19
LEGENDplex™ Human Growth Factor Panel
Setup Procedure for FACSCaliburTM with Dual Lasers
For a dual laser FACSCaliburTM,useFL2forreporterandFL4forbeadsclassifica-tion.Ingeneral,thereisnoneedforcompensationbetweenthesechannelsifthe machine is set up properly, following the setup procedure described below.
1. Start up the Instrument
Performinstrumentstartupandverificationcheckfollowingthemanufac-turer’srecommendations.
2. ObtainaTemplateforDataAcquisition
A template for FACSCaliburTM is a document with density plot that allows theusertoperformmachinesetupanddataacquisition.
Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplate and proceed to Step 3.
If a template is not yet available, create a new template by following the instructionsbelow:
2.1 FromtheBDCellQuestdataacquisitionsoftware,gotoFile→new document.
2.2 CreateadotplotwithFSC(forwardscatter)forX-axisandSSC(side scatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create twogatesandlabelthemBeadsAandBeadsB(Figure4).
Figure 4.
2.3 Create4fluorescentdotplotsasshownbelow(Figure5)withFL2for X-axis,FL4orFL1forY-axis.Forthefluorescentdotplots,gateon BeadsA(dotplotsontheleftpanelbelow)andBeadsB(dotplotson therightpanelbelow).The dot plots should be in log mode.
02
00
40
06
00
80
01
00
0S
SC
-H
00 200 400 600 800 1000FSC-H
Acquisition Dot Plot
Beads A
Beads B
10
01
01
10
21
03
10
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
Page 1
Gated on Beads BGated on Beads A
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
20
2.4 Save the new document as “LEGENDplex Template for FACSCalibur Dual Laser” and proceed to the next step of the setup.
Figure 5.
3. Set up the PMT Voltages
The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannel(FL4/APC)andchannelFL1.TheSetupBeads2:PE
BeadsareusedtosetupthePMTvoltageofthereporterchannel(FL2/PE)The Setup Beads 1: FITC Beads are not needed for this setup.
FollowtheinstructionsbelowforsettingupthePMTsettings:
3.1 VortexthevialoftheRawBeadsfor30secondstoresuspendthe beads.
3.2 Transfer400μLoftheRawBeadstoafreshFACStube.
3.3 Settheflowcytometerflowratetolow.Insetupmode,runthe RawBeads.AdjustthesettingsforFSCandSSCsothatbothbead populationsarevisible(Figure6).
Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadpopulationsafteradjustingsettings.
biolegend.com 21
LEGENDplex™ Human Growth Factor Panel Figure 6.
3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>200.
3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate
(Figure6).
3.6 AdjusttheFL1settingsothattheFL1signalsforallbeadsare between 1x100 and 1x101(Note:Thisstepisnotrequired,butis
recommended).
3.7 AdjusttheFL4settingsothattheFL4signalsforallbeadsarebe- tween 1x101 and 5x103(Figure7).
Figure 7.
3.8 VortexthevialofSetupBeads2:PEBeadsfor30secondstoresus- pend the beads.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
22
3.9 Transfer400μLofthePEBeadstoafreshFACStube.
3.10ReplacetheRawBeadstubefromtheflowcytometerwiththe PE beads tube.
3.11Settheflowcytometerflowratetolow.Insetupmode,runthePE Beads.Note:PEbeadsareonlyofsmallsize,fallingintheBeadsA
gate(Figure8).
3.12AdjusttheFL2settingsothatthemedianfluorescenceintensityof thePEbeadsfallsbetweenthelot-specificrangelabeledonthePE Beadsvial(Figure8,gateR3).
Figure 8.
3.13 Save the document again for future use.
3.14Theflowcytometerisnowreadyforsampleacquisition.
biolegend.com 23
LEGENDplex™ Human Growth Factor Panel
Setup Procedure for FACSCaliburTM with a Single Laser
For a single laser FACSCaliburTM,useFL2forreporterandFL3forbeadsclassifi-cation.Compensationisneededtoproperlysetuptheinstrument.
1. Start up the Instrument
Performinstrumentstartupandverificationcheckfollowingthemanufac-turer’srecommendations.
2. ObtainaTemplateforDataAcquisition
A template for FACSCaliburTM is a document with density plot that allows theusertoperformmachinesetupanddataacquisition.
Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplate and proceed to Step 3.
If a template is not yet available, create a new template by following the instructionsbelow:
2.1 FromtheBDCellQuestdataacquisitionsoftware,gotoFile→new document.
2.2 CreateadotplotwithFSC(forwardscatter)forX-axisandSSC(side scatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create twogatesandlabelthemBeadsAandBeadsB(Figure9).
Figure 9.
0200
400
600
800
1000
SS
C-H
00 200 400 600 800 1000FSC-H
Acquisition Dot Plot
Beads A
Beads B
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
Page 1
Gated on Beads BGated on Beads A2.3 Create4fluorescentdotplotsasshownbelow(Figure10)withFL2 forX-axis,andFL3orFL1forY-axis.Forthefluorescentdotplots,gate onBeadsA(dotplotsontheleftpanelbelow)andBeadsB(dotplots ontherightpanelbelow).The dot plots should be in log mode.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
24
Figure 10.
0200
400
600
800
1000
SS
C-H
00 200 400 600 800 1000FSC-H
Acquisition Dot Plot
Beads A
Beads B
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
Page 1
Gated on Beads BGated on Beads A
2.4 Setallcompensationstozero.
2.5 Save the new document as “LEGENDplex Template for FACSCalibur Single Laser” and proceed to the next step of the setup.
biolegend.com 25
LEGENDplex™ Human Growth Factor Panel3. SetupPMTVoltagesandCompensation
The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannelFL3andchannelFL1.TheSetupBeads2:PEBeadsare used to set up the PMT voltage of the reporter channel FL2. All three setupbeadsareneededforsettingupcompensation.
FollowtheinstructionsbelowforsettingupthePMTsettingsandcompen-sation:
3.1 VortexthevialoftheRawBeadsfor30secondstoresuspendthe beads.
3.2 Transfer400μLoftheRawBeadstoafreshFACStube.
3.3 Settheflowcytometerflowratetolow.Insetupmode,runtheRaw Beads.AdjustthesettingsforFSCandSSCsothatbothbeads populationsarevisible(Figure11).
Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadpopulations,afteradjustingsettings.
Figure 11.
3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>200.
3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate (Figure11).
3.6 AdjusttheFL1settingsothattheFL1signalsforallbeadsarebe- tween 1x100 and 1x101 (Figure12).
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
26
Figure 12.
3.7 AdjusttheFL2settingssothattheFL2signalsforallbeadsarebe- tween 1x100 and 1x101(Figure12).
3.8 AdjusttheFL3settingsothattheFL3signalsforallbeadsarebe- tween 1x101 and 5 x 103(Figure12).
3.9 VortexthevialoftheSetupBeads1:FITCBeadsfor30secondsto resuspend the beads.
3.10Transfer200μLoftheFITCBeadstoafreshFACStube.Add200μL ofRawBeadsandmixwell(thiswillgenerateFITC-positiveand FITC-negativepopulationsofbeadsandisneededforproper compensation).
3.11 In setup mode, run the mixed FITC and Raw Beads.
3.12OntheFL1vsFL2dotplot(Figure13),thebeadswilldisplayasFITC- negativeandFITC-positivepopulations(indicatedbyanarrow).
Note:FITCbeadsareonlyofsmallsize,fallingintheBeadsAgate.
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
0200
400
600
800
1000
SS
C-H
00 200 400 600 800 1000FSC-H
Acquisition Dot Plot
Beads A
Beads B
Page 1
Gated on Beads BGated on Beads A
biolegend.com 27
LEGENDplex™ Human Growth Factor Panel Figure 13.
3.13AdjusttheFL2-%FL1compensationsetting(e.g.,FL2-FL1=20%)so thattheFITC-negativeandFITC-positivepopulationshavesimilar meanFL2fluorescenceintensities(Figure14).
Figure 14.
3.14VortexthevialofPEBeadsfor30secondstoresuspendthebeads.
3.15Transfer200μLofthePEBeadstoafreshFACStube.Add200μLof RawBeadsandmixwell(ThiswillgeneratePE-positiveand PE-negativepopulationsofbeadsandisneededforpropercompen- sation).
3.16 In setup mode, run the mixed PE and Raw Beads.
3.17OntheFL1vsFL2dotplot(Figure15),thebeadswilldisplayasPE- negativeandPE-positivepopulations(indicatedbyanarrow).Note: PEbeadsareonlyofsmallsize,fallingintheBeadsAgate.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
28
Figure 15.
3.18AdjusttheFL2settingsothatthemedianfluorescenceintensityofthe PEbeadsfallsbetweenthelot-specificrangelabeledonthePEBeads vial(Figure16).
3.19AdjusttheFL1-%FL2compensationsetting(e.g.,FL1-FL2=1.5%)so thatthePE-negativeandPE-positivepopulationshavesimilarmean FL1fluorescenceintensities(Figure16).
Figure 16.
3.20OntheFL3vsFL2dotplot(Figure17),thebeadswilldisplayasPE- negativeandPE-positivepopulations(indicatedbyanarrow).
3.21 AdjusttheFL3-%FL2compensationsetting(e.g.,FL3-%FL2=40%)so thattheFL3-lowpopulationandthePE-positivepopulationhave similarmedianFL3signal(Figure18).
biolegend.com 29
LEGENDplex™ Human Growth Factor Panel Figure 17.
Figure 18.
3.22 Save the document again for future use.
3.23Theflowcytometerisnowcompensatedandreadyforsample analysis.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
30
Setup Procedure for BD FACSAriaTM, FACSCantoTM and LSR Series
ThispartoftheguideappliestoBDdigitalflowcytometersusingFACSDivaTM softwareversion6.0andabove.
For the BD FACS machines running FACSDivaTM, use the PE channel for reporter andtheAPCchannelforbeadsclassification.Ingeneral,thereisnoneedforcompensationbetweenthesechannelsifthemachineissetupproperlyfollow-ing the setup procedure described below.
Thissetupprocedureisrequiredunderthefollowingsituations:
• YouarerunningtheLEGENDplexkitforthefirsttime.
• It has been over a month since the procedure was last performed.
• Yourflowcytometerhasbeenservicedsinceyoulastperformedthis procedure.
This setup process is not needed if you have run this experiment before and haveaccesstoasavedexperimenttemplate(Thesettingswillbesavedinthefinalstepofthissetupprocedureandanysettingssavedcanbeimportedtoanewexperiment.PleaserefertoStep2andStep3.9below).
1. Start up the Instrument
Performinstrumentstartupandverificationcheckfollowingthemanufac-turer’srecommendations.
2. ObtainaTemplateforDataAcquisition
A template for FACSDivaTMisaworksheetwithdensityplotsthatallowstheusertoperformmachinesetupanddataacquisition.
If a template is not yet available, create a new template by following the instructions.Afteratemplateiscreated,savethefileinD:\BDExport\Tem-plates\Experiment.DonotchangethenameoftheTemplatesfolder.
Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplateandproceedtoStep3.Toopenanexistingtemplate,selectEx-periment>NewExperiment.AlistoftemplatessavedinD:\BDExport\Tem-plates\Experimentwillpopup.Selectthedesiredtemplatefromthelist.
Tocreateanewtemplate,followtheinstructionsbelow:
2.1 From the BD FACSDivaTMsoftware,gotoExperiment>NewExperi- ment.
biolegend.com 31
LEGENDplex™ Human Growth Factor Panel 2.2 Intheglobalworksheet,opentheworksheet.Createadotplotwith
FSC(forwardscatter)forX-axisandSSC(sidescatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create two gates and label themBeadsAandBeadsB(Figure19).
Figure 19.
2.3 CreatetwodotplotswithPEforX-axis,APCforY-axis(Figure20), gatedonBeadsA(leftpanelbelow)andBeadsB(rightpanelbelow), respectively.CreateonedotplotwithFITCforX-axis,APCforY-axis, gatedonBeadsAandBeadsB(graphnotshown).The plots should all be in log mode.
Figure 20.
2.4 Save the document as “LEGENDplex Template for FACS Diva” inD:\BDExport\Templates\Experimentandproceedtothenextstep of setup.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
32
3. Set up PMT Voltages
The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannelAPC,reporterchannelPE,andFITCchannel.TheSetup Beads 1: FITC Beads and 2: PE Beads are not needed for this setup becausenocompensationisrequiredifthesetupproceduredescribedhereis closely followed.
FollowtheinstructionsbelowforsettingupthePMTsettings:
3.1 VortexthevialofRawBeadsfor30secondstoresuspendthebeads.
3.2 Transfer400μLoftheRawBeadstoafreshFACStube.
3.3 Settheflowcytometerflowratetolow.RuntheRawBeads.Adjust thesettingsforFSCandSSCsothatbothbeadpopulationsarevisible (Figure21).
Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadspopulationsafteradjustingsettings.
Figure 21.
3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>50(x1000).
3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate (Figure21).
3.6 AdjusttheFITCsettingsothattheFITCsignalforthemajorityof beads is between 1x101 and 1x102 (Figure22).
biolegend.com 33
LEGENDplex™ Human Growth Factor Panel Figure 22.
Beads Beads A+ Beads B
3.7 AdjustthePEsettingsothatthePEsignalforthemajorityofbeadsis between 1x101 and 1x102(Figure23).
Figure 23.
3.8 AdjusttheAPCsettingssothatthetheAPCfluorescenceintensitiesof allbeadpopulationsarebetween1x102 and 5 x 104(Figure23).
3.9 Save the document again for future use.
Tosaveyourassay-specificsettings,inthebrowser,right-click CytometerSettingsandselectSavetoCatalog.Namethefile,and thenclickOK.Toimportthesavedsettingforanewexperiment,right clickoncytometersettingsandselectimportsettings.
3.10Theflowcytometerisnowreadyforsampleacquisition.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
34
Setup Procedure for Other Flow Cytometers
Forflowcytometersnotaddressedabove,thesetupprocedurewilldifferfromone to another. It is very important for the end-user to set up the machine following similar guidelines. Inthiscase,machinecompensationbetweenthereporterandbeadsclassificationchannelsisstronglyrecommended.
Theflowcytometersetupinstructionsarealsopostedonourwebsite:www.biolegend.com/legendplex. Newflowcytometersetupinstructionswillbeadded to on our website when they become available.
biolegend.com 35
LEGENDplex™ Human Growth Factor Panel
Chapter 5: DATA ACQUISITION AND ANALYSIS
DataAcquisition
1. Beforereadingsamples,makesurethattheflowcytometerissetupprop-erly.Forflowcytometersetup,pleasefollowtheFlowCytometerSetupguide in this manual or visit: www.biolegend.com/legendplex.
2. Createanewtemplateoropenanexistingtemplate(fordetailsonhowtocreateacytometer-specifictemplate,pleaserefertotheFlowCytometerSetupGuide).
3. Vortexeachsamplefor5secondsbeforeanalysis.
4. Settheflowratetolow.Setthenumberofbeadstobeacquiredto2000-2500 on Beads A gate or Beads B gate. Do not acquire too many events (e.g.>10,000totalevents).
Note: Do not acquire too few or too many beads. Too few beads acquired mayresultinhighCVsandtoomanybeadsacquiredmayresultinslowdata analysis later.
Whenreadingsamples,settheflowcytometertosetupmodefirstandwaituntilbeadpopulationisstabilizedbeforeswitchingtoacquisitionmode.
To simplify data analysis using the LEGENDplexTMDataAnalysisSoftware,read samples in the same order as shown on the PLATE MAP or RACK MAP attachedattheendofthemanual.Foranin-plateassay,readcolumnbycolumn(A1,B1,C1...A2,B2,C2...).Foranin-tubeassay,readrowbyrow(A1,A2,A3,...B1,B2,B3...).
Whennamingdatafiles,trytousesimplenameswithaconsecutivenum-beringforeasydataanalysis(e.g.forstandards,C0.001,C0.002,C1.003,C1.004, C2.005, C2.006, C3.007, C3.008, ... C7.015, C7.016; for samples, S1.017,S1.018,S2.019,S2.020,S3.021,S3.022…)
StoreallFCSfilesinthesamefolderforeachassay.Ifrunningmultipleas-says, create a separate folder for each assay.
5. Proceed to data analysis using LEGENDplexTMDataAnalysisSoftwarewhendataacquisitioniscompleted.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
36
Data Analysis
• TheFCSfilegeneratedonaflowcytometershouldbeanalyzedusingBioLegend’s LEGENDplexTMDataAnalysisSoftwareorothercompatibledataanalysissoftware.TheLEGENDplexTMDataAnalysisSoftwarecanbedownloaded here: www.biolegend.com/legendplex.
• Afterdownloading,installitonaPC(runningWindows7orWindows8)anduseitinconjunctionwiththeDataAnalysisSoftwareDongleincludedinthiskit.Thedonglehasalicensekeystoredinitandisneededtorunthesoftware.Tousethedongle,simplyplugitintheUSBportofthecomputeronwhichthedataanalysissoftwareisinstalled,priortolaunchingthesoftware.
• TheSoftwareDonglehasafixednumberofpoints.Eachanalysiswillconsume a certain number of points. The number of points consumed dependsonassayplexsizeandnumberofsamples.Afterthepointsona dongle are consumed, a new dongle will be needed to run more data analysis.Althoughthesoftwarewillcontinuetobefunctional,thedatawillnotbesaveduntilanewdongleisavailable.Anewdongleisprovidedineach LEGENDplexTMkit.Asavedanalysiscanalwaysbere-analyzedusingthesoftwareregardlessofdonglestatus.
• Follow the LEGENDplexTMDataAnalysisSoftwareUserGuideandOnlineHelptousethesoftware(www.bioLegend.com/legendplex; or press F1 for online help at any step of the data analysis).
biolegend.com 37
LEGENDplex™ Human Growth Factor Panel
Chapter 6: FREQUENTLY ASKED QUESTIONS
Q. WhatisthedifferencebetweenLEGENDplexTM and Luminex®Assays?
BioLegend’s LEGENDplexTM assays and Luminex®-basedmultiplexassaysare both bead-based immunoassays using the same basic principle of sandwichimmunoassays.Bothsystemsusefluorescence-codedbeadstoachievemultiplexing.Themajordifferenceishowthedataisacquired.LEGENDplexTMassaysusecommonlabflowcytometersandtheirrespec-tivesoftwarefordataacquisition,whereasLuminex®-based assays use dedicatedmachinesandsoftwarefordataacquisition.Therefore,oneofthe advantages of LEGENDplexTM assays is that they can be run on common flowcytometersandnospecializedmachineisneeded.
Q. Can I use LEGENDplexTM kits on Luminex®machines?
No. LEGENDplexTMkitsshouldbeusedonaregularflowcytometerandthedatacanbeanalyzedusingtheuser-friendlydataanalysissoftwareavail-ablefordownloadatwww.biolegend.com/legendplex.
Q. WhatarethecompatibleflowcytometersfortheLEGENDplex™assays?
In general, LEGENDplexTMassayscanbeusedonmostcommonflowcytom-eters, such as:
BD FACSCalibur™ BD FACSCanto™ BD FACSCanto™ II BD TM LSR I BD TM LSR II BD LSRFortessaTM
BD FACSAria™ BD FACSArrayTM
PleaserefertotheMATERIALSTOBEPROVIDEDBYEND-USERsectionfordetailsonthechannelconfigurationsofthecytometer.
Forflowcytometersnotlistedabove,theenduserneedstomakesure
thatthemachineissetupproperlybeforeuse(refertoSetup Procedure for Other Flow CytometerssectioninChapter4).Inthiscase,machinecompensationbetweenthereporterandbeadsclassificationchannelsisstrongly recommended.
TheFCSorListmodefilesfromthefollowinginstrumentsarecompatiblewith the LEGENDplex TMDataAnalysisSoftware.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
38
Vendor Instrument Model Vendor Instrument
Model
BD
FACScan™ Sony iCyt Eclipse™
FACSCalibur™ Life Technologies Attune®
FACSCanto™ II Miltenyi Biotec MACSQuant®
LSRFortessa™ Partec PAS
LSR II Stratedigm S1400
FACSAria™
BeckmanCoulter
CyAn™ ADP
FACSAria™ II FC 500
Accuri™ C6 Gallios™
ORFLO Moxi Flow™ MoFlo®Astrios™
CyteK DxP10™ MoFlo®XDP
Q. Whatistherightprocedureforrunningtheassay?
1. Performtheassayasinstructedinthiskitmanual.
2. Determinethetypeofflowcytometeryouhaveandperformmachinesetup as instructed on Flow Cytometer Setup guide in this manual or visit: www.biolegend.com/legendplex.Openanexistingorcreateamachine-specifictemplatefordataacquisition.
3. AnalyzethesamplesontheflowcytometerandsavetheFCSfilesina new folder.
4. Install the LEGENDplexTM DataAnalysisSoftwarealongwiththesoft-waredongleonaPCwithoperatingsystemWindows7orWindows8(32bitor64bit).Makesuretoinstallthecorrectversion(32bitor64bitversion),dependingonyourcomputer’soperatingsystem.
5. TransferthefoldercontainingtheFCSfilesofyourexperimenttothecomputerwherethedataanalysissoftwareisinstalled.
6. PerformdataanalysisasinstructedintheDataAnalysisSoftwareUserGuide.
Q. WhendoIneedtodomachinecompensation?
Ifaflowcytometerequippedwithasinglelaserisusedforbothreporter(FL2/PE)andclassification(FL3),thencompensationisabsolutelyrequiredto compensate the signal spill over between the two channels, especially from FL2 to FL3. Please use the Setup Beads provided and follow the FLOW CYTOMETER SETUP guide in this manual.
biolegend.com 39
LEGENDplex™ Human Growth Factor Panel Ifaflowcytometerislistedinthecompatibleflowcytometerlistinthis
manualandisequippedwithtwolasers,oneforreporter(FL2/PE)andtheotherforclassification(FL4/APC),thencompensationisnotrequirediftheFLOW CYTOMETER SETUP guide in this manual is closely followed. How-ever,compensationisrecommendedforconsistentresults.
Forotherflowcytometersnotlistedinthecompatibleflowcytometerlist, the end-user needs to set up the machine following similar guidelines (refertomanufacturer’smanualforproperinstrumentsetup).Inthiscase,machinecompensationbetweenthereporterandbeadsclassificationchannels is strongly recommended.
Q. Isthereaspecialsoftwarerequiredfordataanalysis?
Theflowcytometerrawdatafiles(FCS2.0,3.0)canbeanalyzedusingtheLEGENDplexTMDataAnalysisSoftwareorotherequivalentanalysissoft-ware. Download the LEGENDplexTM DataAnalysisSoftwareforfreehere:www.biolegend.com/legendplex.Asoftwaredonglewithlicensekeyisprovidedinthekit.
Fordataacquisition,nospecialsoftwareisneeded.Justusethedataac-quisitionsoftwarethatcomeswiththeflowcytometer,aslongasthedatageneratedisinFCSformat,whichmeetsFCSconvention2.0,3.0and3.1.
Q. CanIselecttomeasureonlysomeanalyteswithinapanel?
Yes.Thetargetswithinapanelarefullycustomizable.Thecustomercanselectanycombinationoftargetswithinapanelandorderacustomizedproduct.Pleaseuseourwebsitetargetselectiontoolfororderingcustom-izedproduct(www.biolegend.com/legendplex).
Q. CanIselecttomeasureanalytesacrosspanels?
Cross-panelcustomizationisalsopossible,aslongasthetotalplexsizeisno more than 13 and there is no overlapping beads region among targets selected.Forcross-panelcustomization,[email protected].
Q. I have run out of a component from the kit. Can I use a similar compo-nentfromadifferentkit?
TheLEGENDplex™Beads,Standards,DetectionAntibodies,MatrixB,andSA-PEarelot-specificandmustbeusedincombinationwitheachother.Donotmixthesecomponentsfromdifferentkitsorlots.
Othercomponents,suchasAssayBuffer,WashBuffer,Plate,andPlateSealerarenotlot-specificandcanthereforebeexchangedbetweendiffer-entkitsandlots.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
40
Q. My standard curve is not linear; can I use the non-linear part of the stan-dardcurveduringanalysis?
Yes. It is possible to use the non-linear part of the standard curve for calculatingtheresults.TheLEGENDplexTMDataAnalysisSoftwareusesafive-parametercurvefittingalgorithm,whichdeterminestheminimumandmaximumdetectionconcentrationsofeachtargetandreportsthem.Ifsampleconcentrationsfallabovethemaximumdetectableconcentration,thesamplewillhavetobedilutedandreanalyzed.Ifsampleconcentra-tionsfallbelowtheminimumdetectionconcentration,itisconsiderednotdetectablebytheparticularassay.
Q. Duringdataacquisition,whydothebeadpopulations(definedbyFSCandSSC)sometimesappeartobedispersedorshifted?
Thisisusuallycausedbyafastflowrateorasuddenchangeinflowrate.Therearethreeflow-throughsettings:low,medium,andhigh.Ifyouareusingalowflowrateandthenchangetomediumandhighflowrate,thesuddenchangeofflowratemaysometimesresultinadispersedorshiftedbeadpopulation.Therefore,itisnotrecommendedtochangeflowrateduringdataacquisition.Thebestwayistorunthesampleinsetupmodeusinganidealflowrate,andoncethepopulationisstable,thenchangeittoacquisitionmode.
Q. DoestheLEGENDplex™DataAnalysisSoftwarerunonAppleMacintosh?
Notyet.ThecurrentversionofthesoftwarehastorunonaPCwithoperatingsystemWindows7orWindows8(32bitor64bit).IfyourflowcytometerisconnectedtoaMac,afterdataacquisition,theentirefoldercontainingthedata(FCSfiles)shouldbetransferredtoaPCandanalyzed.
Q. Whydoeseachkitincludeasoftwaredongle?
The dongle allows you to use our LEGENDplexTMsoftwarefordataanalysisand each dongle contains 1 million points. All dongles are the same.
Q. Iranoutofpointsonmydongle.CanIgetanewoneseparately?Or,canIuseadonglefromonekitforanotherkit?
Dongle is not sold separately. Contact BioLegend if you need a new dongle. Thedonglecanbeusedacrosskits.
biolegend.com 41
LEGENDplex™ Human Growth Factor Panel
Chapter 7: ASSAY CHARACTERIZATION
RepresentativeStandardCurve
This standard curve was generated using the LEGENDplexTM Human Growth FactorPanelfordemonstrationpurposeonly.Astandardcurvemustberun with each assay.
AssaySensitivity
Theassaysensitivityorminimumdetectableconcentration(MDC)isthetheoreticallimitofdetectioncalculatedusingtheLEGENDplexTM Data AnalysisSoftwarebyapplyinga5-parametercurvefittingalgorithm.
Analyte MDC in Cell Culture Medium (pg/mL)
MDC in Serum (pg/mL)
Human Ang-2 4.9 5.4
Human EGF 2.3 2.6
Human EPO 4.6 4.7
Human FGF-basic 7.3 7.6
Human G-CSF 7.6 4.4
Human GM-CSF 3.3 4.4
Human HGF 7.6 4.8
Human M-CSF 0.9 0.7
Human PDGF-AA 6.6 11.0
1.0
10.0
100.0
1000.0
10000.0
1.0 100.0 10000.0
MFI
Concentration (pg/mL)
Angiopoietin-2 EGF EPO FGF-basic G-CSF GM-CSF HGF M-CSF PDGF-AA PDGF-BB SCF TGF-α VEGF
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
42
Human PDGF-BB 6.9 3.6
Human SCF 2.2 2.4
HumanTGF-α 6.6 5.4
HumanVEGF 9.9 11.5
Cross-Reactivity
Thefollowingrecombinantproteinsweretestedat50ng/mLusingtheLEGENDplexTM HumanGrowthFactorPanel.VEGFpaircandetectbothVEGF165andVEGF121.SCFpairdetectsfreeSCFproteins.Noornegligiblecross-reactivitywasfoundforallotheranalytes.
IL-27 IGF-II IL-13 IL-17F IL-23
IL-33 TGF-β1 IL-2 IL-4 IL-12p40
MIP-1β TGF-β2 IL-6 IL-21 IL-12p70
ENA-78 TGF-β3 IL-9 IL-22 IL-15
MCP-1 TNF-β IL-10 TSLP IL-18
RANTES PLGF-1 IFN-γ IL-1α IL-11
Ang-1 VEGF-c TNF-α IL-1β
IGF-I IL-5 IL-17A IFN-α
Groα IP-10 CCL11 MIP-1α
IL-8 TARC MIP-3α I-TAC
Accuracy (Spike Recovery)
Forspikerecoveryincellculturemedium,RPMIorDMEMwith10%FCSwasfirstdilutedtwo-foldwithAssayBufferandspikedwithtargetproteinsatthreedifferentlevelswithintheassayrange.Thespikedsampleswerethenassayed,andthemeasuredconcentrationswerecomparedwiththeexpected values.
Forspikerecoveryinserum(n=8)andplasma(n=8),sampleswerefirstdilutedtwo-foldwithAssayBufferandspikedwithtargetproteinsatthreedifferentlevelswithintheassayrange.Thespikedsampleswerethenas-sayed,andthemeasuredconcentrationswerecomparedwiththeexpectedvalues.
biolegend.com 43
LEGENDplex™ Human Growth Factor Panel
Analyte% of Recovery in Cell Culture
Medium
% of Recov-ery in Serum
% of Recov-ery in Plasma
Human Ang-2 96% 45% 56%
Human EGF 98% 82% 87%Human EPO 93% 60% 67%Human FGF-basic 94% 126% 104%Human G-CSF 96% 84% 83%Human GM-CSF 96% 118% 109%Human HGF 99% 103% 84%Human M-CSF 98% 51% 56%Human PDGF-AA 98% 120% 117%Human PDGF-BB 101% 133% NAHuman SCF 100% 80% 80%HumanTGF-α 99% 78% 71%HumanVEGF 100% 75% 78%
NA=NotApplicableduetohighendogenouslevels
LinearityofDilution
Forspikelinearityincellculturemedium,RPMIorDMEMwith10%FCSwasfirstdilutedtwo-foldwithAssayBufferandspikedwithaknowncon-centrationoftargetproteins.Thespikedsampleswereseriallydiluted1:2,1:4,1:8withassaybufferandassayed.Themeasuredconcentrationsofseriallydilutedsampleswerecomparedwiththatofthespikedsamples.
Fortestinglinearityinserum(n=8)andplasma(n=8),sampleswerefirstdilutedtwo-foldwithAssayBufferandspikedwithaknownconcentrationoftargetproteins.Thespikedsampleswereseriallydiluted1:2,1:4,1:8withMatrixBandassayed.Themeasuredconcentrationsofseriallydilutedsampleswerecomparedwiththatofthespikedsamples.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
44
AnalyteLinearity in Cell
Culture Medium
Linearity in Serum
Linearity in Plasma
Human Ang-2 102% 116% 130%
Human EGF 104% 102% 110%Human EPO 105% 110% 131%Human FGF-basic 100% 83% 87%Human G-CSF 96% 96% 111%Human GM-CSF 111% 86% 89%Human HGF 102% 100% 113%Human M-CSF 92% 126% 147%Human PDGF-AA 96% 92% 114%Human PDGF-BB 104% 98% 95%Human SCF 106% 104% 102%HumanTGF-α 98% 97% 96%HumanVEGF 103% 108% 117%
Intra-Assay Precision
Twosampleswithdifferentconcentrationsoftargetproteinswereanalyzedin one assay with 16 replicates for each sample. The intra-assay precision was calculated as below.
Analyte Sample Mean (pg/mL) STDEV %CV
Human Ang-2Sample 1 38.8 1.3 3%
Sample 2 540.6 21.9 4%
Human EGFSample 1 38.3 1.8 5%
Sample 2 579.2 39.8 7%
Human EPOSample 1 38.6 3.2 8%
Sample 2 547.6 45.6 8%
Human FGF-basicSample 1 38.4 1.8 5%
Sample 2 525.6 16.5 3%
biolegend.com 45
LEGENDplex™ Human Growth Factor Panel
Human G-CSFSample 1 35.1 2.9 8%
Sample 2 565.9 46.4 8%
Human GM-CSFSample 1 36.8 1.5 4%
Sample 2 529.0 26.3 5%
Human HGFSample 1 38.2 2.1 6%
Sample 2 511.3 36.2 7%
Human M-CSFSample 1 36.3 2.3 6%
Sample 2 518.4 22.0 4%
Human PDGF-AASample 1 36.7 1.4 4%
Sample 2 485.7 24.1 5%
Human PDGF-BBSample 1 39.4 2.8 7%
Sample 2 507.7 29.0 6%
Human SCFSample 1 38.0 1.8 5%
Sample 2 482.3 25.3 5%
HumanTGF-αSample 1 38.1 2.0 5%
Sample 2 559.6 48.9 9%
HumanVEGFSample 1 37.2 1.8 5%
Sample 2 499.5 18.8 4%
Inter-Assay Precision
Twosampleswithdifferentconcentrationsoftargetproteinswereanalyzedin three independent assays with 3 replicates for each sample. The inter-assay precision was calculated as below.
Analyte Sample Mean (pg/mL) STDEV %CV
Human Ang-2Sample 1 38.1 1.3 3%
Sample 2 601.1 60.6 10%
Human EGFSample 1 41.5 2.0 5%
Sample 2 676.8 40.6 6%
Human EPOSample 1 37.4 2.6 7%
Sample 2 659.1 62.0 9%
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
46
Human FGF-basicSample 1 40.0 1.3 3%
Sample 2 583.4 24.7 4%
Human G-CSFSample 1 35.0 2.1 6%
Sample 2 668.8 59.2 9%
Human GM-CSFSample 1 39.4 2.3 6%
Sample 2 638.6 36.2 6%
Human HGFSample 1 38.9 2.2 6%
Sample 2 633.3 41.0 6%
Human M-CSFSample 1 38.5 1.5 4%
Sample 2 618.9 24.3 4%
Human PDGF-AASample 1 39.7 2.7 7%
Sample 2 569.7 41.1 7%
Human PDGF-BBSample 1 39.6 2.1 5%
Sample 2 596.7 44.2 7%
Human SCFSample 1 39.5 1.8 5%
Sample 2 586.4 52.4 9%
HumanTGF-αSample 1 38.4 1.8 5%
Sample 2 685.6 55.2 8%
HumanVEGFSample 1 37.8 1.4 4%
Sample 2 602.9 48.5 8%
Biological Samples
Serum
Normalhumanserumsamples(n=36)weretestedforendogenouslevelsofgrowthfactors.Theconcentrationsmeasuredareshownbelow:
Analyte Range (pg/mL)
% of Detectable
Mean of detectable
(pg/mL)Human Ang-2 39.2 - 2514 100% 904.2
Human EGF ND - 44.0 55% 16.2
Human EPO ND - 104.7 70% 42.7
Human FGF-basic ND - 723.5 65% 160.7
biolegend.com 47
LEGENDplex™ Human Growth Factor Panel
Human G-CSF ND - 181.1 90% 40.0
Human GM-CSF ND - 77.1 20% 33.7
Human HGF 44.8 - 614.5 100% 207.4
Human M-CSF ND - 150.0 20% 83.7
Human PDGF-AA 218.2 - 3614 100% 986.4
Human PDGF-BB 121.4 - 6091 100% 1405
Human SCF ND - 203.3 45% 50.7
HumanTGF-α ND - 857.9 40% 244.9
HumanVEGF ND - 163.8 40% 84.7
ND=NotDetectable
Plasma
Normalhumanplasmasamples(n=20)weretestedforendogenouslevelsofgrowthfactors.Theconcentrationsmeasuredareshownbelow:
Analyte Range (pg/mL) % of Detectable
Mean of detectable
(pg/mL) Human Ang-2 154.8 - 2078 100% (pg/mL)
Human EGF ND - 115.8 85% 35.9
Human EPO ND - 272.7 90% 39.3
Human FGF-basic ND - 1019 90% 132.0
Human G-CSF 15.7 - 170.9 100% 36.6
Human GM-CSF ND - 40.6 25% 23.5
Human HGF 36.3 - 1285 100% 311.1
Human M-CSF ND -268.1 15% 140.4
Human PDGF-AA 391.3 - 29802 100% 8772
Human PDGF-BB 131.5 - 72120 100% 27243
Human SCF ND - 24.6 35% 13.6
HumanTGF-α ND - 727.6 45% 177.9
HumanVEGF ND - 373.2 65% 145.1
Cell Culture Supernatant
HumanPBMC(1x106cells/mL)wereculturedundervariousconditions(PHA,10µg/mL;PMA,50ng/mL;LPS, 50 ng/mL).Supernatants were collectedafter1and4daysandassayedwiththeLEGENDplexTM Human GrowthFactorkit.Theresults(allinpg/mL)aresummarizedbelow.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
48
Analyte Control1 Day
PHA1 Day
PMA1 Day
LPS 1 Day
Human Ang-2 146 160 154 178
Human EGF 52.5 54.1 55.7 54.5
Human EPO ND ND ND ND
Human FGF-basic ND ND 15.4 14.7
Human G-CSF 34.8 2011 78.8 3297
Human GM-CSF 5.4 8.0 228 7.9
Human HGF 47.6 28.3 76.2 23.4
Human M-CSF ND 7.7 106 2.0
Human PDGF-AA 1101 1415 1454 1336
Human PDGF-BB 5103 5100 5349 5139
Human SCF ND ND ND ND
HumanTGF-α 10.8 100 9.6 115
HumanVEGF 359 165 158 459ND=NotDetectable
Analyte Control4 Day
PHA4 Day
PMA4 Day
LPS 4 Day
Human Ang-2 95.1 48.8 64.3 72.4
Human EGF 61.3 61.6 78.1 67.2
Human EPO ND ND 15.7 26.5
Human FGF-basic ND 14.7 17.2 16.0
Human G-CSF 39.1 2125 1565 4530
Human GM-CSF 14.1 5.5 1157 7.8
Human HGF 436 42.8 120 45.3
Human M-CSF 23.7 3.9 1713 3.5
Human PDGF-AA 1239 1830 3520 1576
Human PDGF-BB 6778 7891 10247 8041
Human SCF ND ND ND ND
HumanTGF-α 22.4 166 41.5 171
HumanVEGF 517 1580 177 2270ND=NotDetectable
biolegend.com 49
LEGENDplex™ Human Growth Factor PanelHumanHUVC(1x106cells/mL)wereculturedundervariousconditions(50ng/mLIFN-γ+200ng/mLTNF-α;100ng/mLLPS).Supernatants were collectedafter1dayandassayedwiththeLEGENDplexTM Human Growth Factorkit.Theresults(allinpg/mL)aresummarizedbelow.
Analyte Control IFN-γ+TNF-α LPS
Human Ang-2 8588 6119 8694
Human EGF 2128 2388 2798
Human EPO 29.4 26.7 30.5
Human FGF-basic 710 797 948
Human G-CSF 68.9 752 1836
Human GM-CSF 6.0 89.7 48.2
Human HGF 11.9 94.3 13.5
Human M-CSF 8.5 108.7 19.9
Human PDGF-AA 398 518 396
Human PDGF-BB 840 912 1288
Human SCF ND ND ND
HumanTGF-α 8.4 9.3 10.6
HumanVEGF ND ND ND
ND=NotDetectable
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
50
TROUBLESHOOTING
Problem Possible Cause Solution
Bead popula-tionshiftingupward or downward dur-ingacquisition
The strong PE signal from high concentra-tionsamplesorstan-dards may spill over to classificationChannel(e.g.,FL3/FL4/APC)and mess up the bead separation.
OptimizeinstrumentsettingsusingKitSetupBeads,andmakeappropriatecom-pensationbetweenchannels.
Filter plate will not vacuum or some wells clogged
Vacuumpressureisinsufficientorvacuummanifold does not seal properly.
Increase vacuum pressure such that 0.2mLbuffercanbesuctionedin3-5seconds. Clean the vacuum manifold and makesurenodebrisonthemanifold.Press down the plate on the manifold to makeagoodseal.
Samples have insoluble particlesorsampleistooviscous(e.g.,serumandplasmasamples)
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Ifsomewellsarestillcloggedduringwashing, try the following:
1).Addbuffertoallthewells,pipetteupand down the clogged wells and vacuum again.
2).Useapieceofcleanwipe,wipetheun-der side of the clogged wells and vacuum again.
3).Takeathinneedle(e.g.,insulinnee-dle),whileholdingtheplateupward,pokethelittleholeundereachofthecloggedwellsandvacuumagain.Donotpoketoohard or too deep as it may damage the filterandcauseleaking.
Filter plate was used without pre-wet.
Pre-wetplatewithwashbufferbeforerunning the assay.
biolegend.com 51
LEGENDplex™ Human Growth Factor Panel
Insufficientbead count or slow reading
Beads inappropriately prepared
Sonicate bead vials and vortex just prior toaddition.Agitatethepre-mixedbeadsintermittentlyinreservoirwhilepipettingthis into the plate.
Samples cause beads aggregationduetoparticulatematterorviscosity
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Beads were lost during washing for in-tube assay
Makesurebeadsarespundownbyvisu-allycheckthepellet(beadsareinlightblueorbluecolor).Beverycarefulwhenremoving supernatant during washing.
Probemaybepartiallyclogged
Sample probe may need to be cleaned, or if needed, probe should be removed and sonicated.
Plateleaked
Vacuumpressuresettoo high
Adjust vacuum pressure such that 0.2 mL buffercanbesuctionedin3-5seconds.Do not exceed 10” Hg of vacuum.
Plate set directly on table or absorbent tow-elsduringincubationsorreagentadditions
Set plate on plate holder or raised edge sobottomoffilterisnottouchinganysurface.
Liquid present on the under side of the plate aftervacuum
Afterwashing,pressdownplatefirmlyonastackofcleanpapertowelstodrytheunderside of the plate.
Pipettetouchinganddamagedplatefilterduringadditions
Pipettetothesideofwells.
Highback-ground
Backgroundwellswerecontaminated
Avoidcross-wellcontaminationbychang-ingtipsbetweenpipettingwhenperform-ingtheassayusingamultichannelpipette.
InsufficientwashesThebackgroundmaybeduetonon-spe-cificbindingofSA-PE.Increasenumberofwashes.
Debris(FSC/SSC)duringsample acquisi-tion
Debris or platelet may existinsamplesolution
Centrifugesamplesbeforeanalyzingsamples. Remove platelet as much as possible.
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
52
Variationbe-tweenduplicate samples
Beadsaggregation Sonicate and vortex the Beads prior to use.
Multichannelpipettemay not be calibrated or inconsistent pipet-ting
CalibratePipette.Ensuregoodpipettingpractice.Primepipettebeforeusemayhelp.
Plate washing was not uniform
Makesureallreagentsarevacuumedoutcompletely in all wash steps.
Samples may contain particulatematters.
Centrifuge samples just prior to assay setup and use supernatant. If high lipid content is present, remove lipid layer aftercentrifugation.Samplemayneeddilutioniftooviscous.
Low or poor standard curve signal
The standard was in-correctlyreconstituted,stored or diluted
Followtheprotocoltoreconstitute,storeanddilutestandard.Doublecheckyourcalculation.
Wrongorshortincuba-tiontime
Ensurethetimeofallincubationswasappropriate.
Signals too high, standard curves satu-rated
PMTvalueforFL2/PEset too high
MakesurethePMTsettingforthere-porter channel is appropriate
Plateincubationtimewas too long Useshorterincubationtime.
Sample read-ings are out of range
Samples contain no or below detectable levels of analyte
Makesuretheexperimenttogeneratethesamplesworked.Useproperpositivecontrols.
Samplesconcentrationshigher than highest standard point.
Dilutesamplesandanalyzeagain.
Standard curve was saturated at higher end of curve.
MakesurethePMTsettingforthere-porter channel is appropriate. Use shorter incubationtimeifincubationtimewastoo long
Missed beads populationsduring reading, ordistributionis unequal
Sample may cause some beads to ag-gregate.
Centrifuge samples just prior to assay setup and use supernatant. If high lipid content is present, remove lipid layer aftercentrifugation.Samplemayneeddilutioniftooviscous.
Beadspopulationsarenot mixed properly
Makesureallbeadpopulationsaremixed.and in similar numbers.
biolegend.com 53
LEGENDplex™ Human Growth Factor Panel
PLA
TE M
AP
(for
in-p
late
ass
ay)
1
2 3
4 5
6 7
8 9
10
11
12
A
C0
C4
Sa
mpl
e1
Sa
mpl
e5
Sa
mpl
e9
Sa
mpl
e 13
Sa
mpl
e17
Sa
mpl
e21
Sa
mpl
e25
Sa
mpl
e29
Sa
mpl
e33
Sa
mpl
e 37
B
C0
C4
Sa
mpl
e1
Sa
mpl
e5
Sa
mpl
e9
Sa
mpl
e 13
Sa
mpl
e17
Sa
mpl
e21
Sa
mpl
e25
Sa
mpl
e29
Sa
mpl
e33
Sa
mpl
e37
C
C1
C5
Sa
mpl
e2
Sa
mpl
e6
Sa
mpl
e10
Sa
mpl
e 14
Sa
mpl
e18
Sa
mpl
e22
Sa
mpl
e26
Sa
mpl
e30
Sa
mpl
e34
Sa
mpl
e38
D
C1
C5
Sa
mpl
e2
Sa
mpl
e6
Sa
mpl
e10
Sa
mpl
e 14
Sa
mpl
e18
Sa
mpl
e22
Sa
mpl
e26
Sa
mpl
e30
Sa
mpl
e34
Sa
mpl
e38
E
C2
C6
Sa
mpl
e3
Sa
mpl
e7
Sa
mpl
e11
Sa
mpl
e 15
Sa
mpl
e19
Sa
mpl
e23
Sa
mpl
e27
Sa
mpl
e31
Sa
mpl
e35
Sa
mpl
e39
F
C2
C6
Sa
mpl
e3
Sa
mpl
e7
Sa
mpl
e11
Sa
mpl
e 15
Sa
mpl
e19
Sa
mpl
e23
Sa
mpl
e27
Sa
mpl
e31
Sa
mpl
e35
Sa
mpl
e39
G
C3
C7
Sa
mpl
e4
Sa
mpl
e8
Sa
mpl
e12
Sa
mpl
e 16
Sa
mpl
e20
Sa
mpl
e24
Sa
mpl
e28
Sa
mpl
e32
Sa
mpl
e36
Sa
mpl
e40
H
C3
C7
Sa
mpl
e4
Sa
mpl
e8
Sa
mpl
e12
Sa
mpl
e 16
Sa
mpl
e20
Sa
mpl
e24
Sa
mpl
e28
Sa
mpl
e32
Sa
mpl
e36
Sa
mpl
e40
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
54
RA
CK
MA
P (
for
in-t
ub
e as
say)
1
2 3
4 5
6 7
8 9
10
11
12
A
C0
C0
C1
C1
C2
C2
C3
C3
C4
C4
C5
C5
B
C6
C6
C7
C7
Sam
ple
1 Sa
mp
le
1 Sa
mp
le
2 Sa
mp
le
2 Sa
mp
le
3 Sa
mp
le
3 Sa
mp
le
4 Sa
mp
le4
C
Sam
ple
5
Sam
ple
5
Sam
ple
6
Sam
ple
6
Sam
ple
7
Sam
ple
7
Sam
ple
8
Sam
ple
8
Sam
ple
9
Sam
ple
9
Sam
ple
10
Sa
mp
le
10
D
Sam
ple
11
Sa
mp
le
11
Sam
ple
12
Sa
mp
le
12
Sam
ple
13
Sa
mp
le
13
Sam
ple
14
Sa
mp
le
14
Sam
ple
15
Sa
mp
le
15
Sam
ple
16
Sa
mp
le
16
E Sa
mp
le
17
Sam
ple
17
Sa
mp
le
18
Sam
ple
18
Sam
ple
19
Sa
mp
le
19
Sam
ple
20
Sa
mp
le
20
Sam
ple
21
Sa
mp
le
21
Sam
ple
22
Sa
mp
le
22
F Sa
mp
le
23
Sam
ple
23
Sa
mp
le
24
Sam
ple
24
Sa
mp
le
25
Sam
ple
25
Sa
mp
le
26
Sam
ple
26
Sa
mp
le
27
Sam
ple
27
Sa
mp
le
28
Sam
ple
28
G
Sam
ple
29
Sa
mp
le
29
Sam
ple
30
Sa
mp
le
30
Sam
ple
31
Sa
mp
le
31
Sam
ple
32
Sa
mp
le
32
Sam
ple
33
Sa
mp
le
33
Sam
ple
34
Sa
mp
le
34
H
Sam
ple
35
Sa
mp
le
35
Sam
ple
36
Sa
mp
le
36
Sam
ple
37
Sa
mp
le
37
Sam
ple
38
Sa
mp
le
38
Sam
ple
39
Sa
mp
le
39
Sam
ple
40
Sa
mp
le
40
biolegend.com 55
LEGENDplex™ Human Growth Factor Panel
Tel: 858-768-5800
LEGENDplex™ Human Growth Factor Panel
56
biolegend.com 57
LEGENDplex™ Human Growth Factor Panel
LEGENDplex™ Kits are manufactured by BioLegend Inc. 9727 Pacific Heights Blvd.San Diego, CA 92121Tel: 1.858.768.5800Tel: US & Canada Toll-Free: 1.877.Bio-Legend (1.877.246.5343)Fax: 1.877.455.9587Email: [email protected]
For a complete list of world-wide BioLegend offices and distributors, please visit our website at: biolegend.com
®
76319_V01