hygiene plan for kavo estetica e80 and e70 treatment units

2
KaVo ESTETICA ® E80 / E70 Hygiene plan for KaVo ESTETICA E80 and E70 treatment units. When ? What ? How ? In the morning n Rinsing of instrument tubes (inside) and u Activate rinsing program water supply system n Flushing of suction tubes (inside) u AQUAmat function: Suck water from and suction system AQUAmat tank station n Cleaning of suction system u Activate HYDROclean function and spittoon After each treatment KaVo Dental GmbH · D 88400 Biberach/Riß · Telefon +49 7351 56-0 · Fax +49 7351 56-1488 · www.kavo.com n Flushing of suction tubes (inside) u AQUAmat function: Suck water from and suction system AQUAmat tank station u If required: Activate Tubewashing with Dekamat function n Cleaning of suction system and spittoon u Activate HYDROclean function n Cleaning and disinfecting of device u Manual cleaning and disinfecting by wiping surfaces, upholstery and operating light n Cleaning and disinfecting of handles at u White handles are removable and thermo- dentist element and operating light disinfectable, grey handles are removable and KAVOLUX. sterilisable n Cleaning and disinfecting of spittoon u Manual cleaning and disinfecting by wiping inside and filter cover and outside. Spittoon is removable. Porcellain spittoon is thermodisinfectable. Filter inserts are removable and thermodisinfectable. In the evenings As required n Control suction filter inserts u Control filter, if required change filter n Cleaning and disinfecting of suction u Activate Tubewashing with Dekamat function tubes (inside), suction system and spittoon n Suction tubes (outside) and instrument u Manual cleaning and disinfecting by wiping tubes (outside) n Suction hose connectors u Connecting pieces are removable and thermodisinfectable n Rinsing of instrument tubes (inside) u Prepare and start rinsing program and water supply system n Cleaning and disinfecting of: u Activate intensive disinfection program (simultaneously run of Tubewashing with – instrument tubes (inside) Dekamat function) – water supply system – suction tubes (outside) – suction system – spittoon n Control return air filter of turbine u Control filter, if required change filter n Silicone pads u Silicone pads are removable and thermo- disinfectable n Cup filler and cup rest u Cup filler and cup rest are removable Manual cleaning and disinfecting by wiping Cup filler is thermodisinfectable n Intrument holder u Instrument holder is removable Manual cleaning and disinfecting by wiping n Amalgam separator u Change / deplete collection container and removal of waste according manufacturing instruction Each Friday and after long downtimes 1.007.2708 rs 20100708-02 en functions. They do not replace the instructions for use included with a delivery. They must be followed to avoid malfunctions and injury. Please use only KaVo approved disinfectants! This hygiene plan only contains the essential operating Allowed disinfectants: - Mikrozid Liquid (Schülke & Mayr) – FD 322 (Dürr) - Incidin Liquid (Ecolab)

Upload: others

Post on 15-Nov-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Hygiene plan for KaVo ESTETICA E80 and E70 treatment units

KaVo ESTETICA® E80 / E70

Hygiene plan for KaVo ESTETICA E80 and E70 treatment units.

When ? What ? How ?

In the morning n Rinsing of instrument tubes (inside) and u Activate rinsing program water supply system

n Flushing of suction tubes (inside) u AQUAmat function: Suck water from and suction system AQUAmat tank station

n Cleaning of suction system u Activate HYDROclean function and spittoon

After each treatment

KaVo Dental GmbH · D 88400 Biberach/Riß · Telefon +49 7351 56-0 · Fax +49 7351 56-1488 · www.kavo.com

n Flushing of suction tubes (inside) u AQUAmat function: Suck water from and suction system AQUAmat tank station u If required: Activate Tubewashing with Dekamat functionn Cleaning of suction system and spittoon u Activate HYDROclean functionn Cleaning and disinfecting of device u Manual cleaning and disinfecting by wiping surfaces, upholstery and operating light n Cleaning and disinfecting of handles at u White handles are removable and thermo- dentist element and operating light disinfectable, grey handles are removable and KAVOLUX. sterilisablen Cleaning and disinfecting of spittoon u Manual cleaning and disinfecting by wiping inside and filter cover and outside. Spittoon is removable. Porcellain spittoon is thermodisinfectable. Filter inserts are removable and thermodisinfectable.

In the evenings

As required

n Control suction filter inserts u Control filter, if required change filtern Cleaning and disinfecting of suction u Activate Tubewashing with Dekamat function tubes (inside), suction system and spittoon n Suction tubes (outside) and instrument u Manual cleaning and disinfecting by wiping tubes (outside) n Suction hose connectors u Connecting pieces are removable and thermodisinfectablen Rinsing of instrument tubes (inside) u Prepare and start rinsing program and water supply system

n Cleaning and disinfecting of: u Activate intensive disinfection program (simultaneously run of Tubewashing with – instrument tubes (inside)

Dekamat function) – water supply system – suction tubes (outside) – suction system – spittoon

n Control return air filter of turbine u Control filter, if required change filtern Silicone pads u Silicone pads are removable and thermo- disinfectable n Cup filler and cup rest u Cup filler and cup rest are removable Manual cleaning and disinfecting by wiping Cup filler is thermodisinfectablen Intrument holder u Instrument holder is removable Manual cleaning and disinfecting by wipingn Amalgam separator u Change / deplete collection container and removal of waste according manufacturing instruction

Each Friday and afterlong downtimes

1.00

7.27

08

rs

2010

0708

-02

en

functions. They do not replace the instructions for use included with a delivery. They must be followed to avoid malfunctions and injury.

Please use only KaVo approved disinfectants!

This hygiene plan only contains the essential operating

Allowed disinfectants:- Mikrozid Liquid (Schülke & Mayr)– FD 322 (Dürr) - Incidin Liquid (Ecolab)

Page 2: Hygiene plan for KaVo ESTETICA E80 and E70 treatment units

KaVo ESTETICA E50 / PRIMUS 1058 ®

Hygiene plan for ESTETICA E50 and PRIMUS 1058 treatment unit.

When ? What ? How ?

In the morning n Rinsing of instrument tubes (inside) u Activate rinsing program and water supply system

n Flushing of suction tubes (inside) u Suck cold water with HV suction tube and and suction system saliva ejector

n Cleaning of suction system and u Activate HYDROclean function spittoon

After each treatment

KaVo Dental GmbH · D 88400 Biberach/Riß · Telefon +49 7351 56-0 · Fax +49 7351 56-1488 · www.kavo.com

n Flushing of suction tubes (inside) u Suck cold water with HV suction tube and and suction system saliva ejector u If disinfection required: Suck one dose DEKASEPTOL Gel with dispensern Cleaning of suction system and u Activate HYDROclean function spittoonn Cleaning and disinfecting of device u Manual cleaning and disinfecting by wiping surfaces, upholstery and operating light n Cleaning and disinfecting of handles u White handles are removable and thermodisinfectable at the dentist element and operating Grey handles are removable and sterilisable light KAVOLUX. n Cleaning and disinfecting of u Manual cleaning and disinfecting by wiping inside spittoon and filter cover and outside. Porcelain spittoon is removable and thermodisinfectable. Glass spittoon is removable and disinfectable by wiping. Filter cover is removable and thermodisinfectable.

As required

n Control suction filter inserts u Control filter, if required change filtern Flushing of suction tubes (inside) and u Suck cold water with HV suction tube and suction system saliva ejectorn Cleaning of suction system and spittoon u Activate HYDROclean function

Cleaning and disinfecting of suction u Suck one dose DEKASEPTOL Gel with HV suction tube tubes (inside) and suction system and saliva ejector, suck once more cold water and finish with a further dose of DEKASEPTOL Gel. Replace HV suction tube and saliva ejectorn Disinfection of spittoon u Put one dose DEKASEPTOL Gel in the spittoon and clean with brushn Suction tubes (outside) and instrument u Manual cleaning and disinfecting by wiping tubes (outside)n Suction hose connectors u Connecting pieces are removable and thermodisinfectablen Rinsing of instrument tubes (inside) u Activate rinsing program and water supply system

n Control return air filter of turbine u Control filter, if required change filtern Silicone pads u Silicone pads are removable and thermodisinfectablen Cup filler and cup rest u Cup filler and cup rest are removable Manual cleaning and disinfecting by wiping Cup filler is thermodisinfectablen Intrument holder u Instrument holder is removable Manual cleaning and disinfecting by wipingn Amalgam separator u Change / deplete collection container and removal of waste according to manufacturing instruction

Each Friday and after long downtimesn Cleaning and disinfecting of instrument u Activate intensive disinfection tubes (inside) and water supply system

In the evenings

Each Friday and afterlong downtimes

Rinsing program and intensive disinfection only applicable if intensive

disinfection option is installed.

functions. They do not replace the instructions for use included with a delivery. They must be followed to avoid malfunctions and injury.

Please use only KaVo approved disinfectants!

This hygiene plan only contains the essential operating

n

Allowed disinfectants:- Mikrozid Liquid (Schülke & Mayr)– FD 322 (Dürr) - Incidin Liquid (Ecolab)

®