IEC 61000 -4-11

TESTING & MEASURING TECHNIQUES

Page 2: Iec 61000  4-11
Page 3: Iec 61000  4-11

Voltage dips Short interruption Voltage variation Interruption, voltage dips and voltage

variation causes Test level and preferred test level Test procedure 2 types test configuration Summary

CONTENT

Page 4: Iec 61000  4-11

The immunity test method and range of test level applies to electrical/electronic equipment which:

Irated ≤ 16A/phase for connection to 50Hz/60Hz ac network

IEC 61000 -4-11

Page 5: Iec 61000  4-11

TEST & MEASURE

VOLTAGE DIPS/SAG

SHORT INTERRUPTION

VOLTAGE VARIATION

Page 6: Iec 61000  4-11

Sudden reduction in voltage below specified dip threshold

short period of time, follow by recovery to the original voltage

VOLTAGE DIPS/SAG

Page 7: Iec 61000  4-11
Page 8: Iec 61000  4-11

Decreased below specified interruption threshold Consider as voltage dip to zero. Usually not more then 1 minute Followed by recovery to original voltage

SHORT INTERRUPTION

Page 9: Iec 61000  4-11
Page 10: Iec 61000  4-11

Change in voltage continuouslyVOLTAGE VARIATION

Page 11: Iec 61000  4-11

Continuously varying load connecting to network

CAUSES OF VOLTAGE VARIATION

Page 12: Iec 61000  4-11

Short circuit (eg. SLG) Sudden large change of load

CAUSES OF DIP & INTERRUPTION

Page 13: Iec 61000  4-11

CASE EXAMPLE

Page 14: Iec 61000  4-11

For equipment that has Vrated range; if the voltage range do not exceed 20% of the lower Vrated, a single voltage may be specified as test level UT

TEST LEVEL

Page 15: Iec 61000  4-11

CONSIDER ANNEX B (CLASSES) & TABLE BEFORE TESTING

Class 1 – compatibility level < public network level◦ Sensitive disturbance to ps (ups necessary)

Class 2 - compatibility level = public network level Class 3 - compatibility level > public network level

◦ Large motor frequently start◦ Welding machines are present◦ Major part of the load is fed trough converter◦ Load vary rapidly

Page 16: Iec 61000  4-11
Page 17: Iec 61000  4-11

For 3 phase short interruption, it shall be tested simultaneously

DIPS

SHORT INTERRUPTIO

N

Single phase

3 phase

Single phase

3 phase

changes in supply voltage shall occur at zero crossing

Addition angel selected from 45º, 90º, 180º,225º,270º and 315º if necessary.

Test 1 by 1 phase

Recommended to use 0º for 1 of the phases

THE TEST

Page 18: Iec 61000  4-11

ADDITIONAL ANGEL OR PHASE SHIFT

Page 19: Iec 61000  4-11
Page 20: Iec 61000  4-11

Output voltage of generator at no load: 80% of nominal value verified in maximum duration of 5s

70% and 40% of nominal value verified in maximum duration of 3s

Power source within ±2% of frated

Shortest ps cable as specified by EUT manufacturer

TEST PROCEDURE (before testing) VERIFICATION GENERATOR

Page 21: Iec 61000  4-11

TEST PROCEDURE

PLANNING PRE-ANALYSIS

TEST

•EUT designation•Possible connections (eg: plug, terminal etc)•Input power port•EUT operation representative•EUT Technical specification•Setup description

Define which system configuration (next slide) to be tested to reproduce field situation RESULT/

TEST REPORT/RECORD

EUT FULL FUNCTION

CHECK

Test report & Any degradation also record

Page 22: Iec 61000  4-11

Are classified into 4 categories:

RESULT OF IMMUNITY TESTS:

Page 23: Iec 61000  4-11

2 TYPES TEST CONFIGURATION

COMBINATION OF

TRANSFORMER TAP

IMMUNITY TEST GENERATOR

USING VARIABLE TRANSFORMER

Page 24: Iec 61000  4-11

IMMUNITY TEST GENERATOR USING VARIABLE TRANSFORMER

Page 25: Iec 61000  4-11

IMMUNITY TEST USING TAPPER TRANSFORMER

Page 26: Iec 61000  4-11

TIMING OF CONTROL SIGNAL

tp1 = signal from MPU at wrong time

tp2 = signal from MPU at right time

Page 27: Iec 61000  4-11
Page 28: Iec 61000  4-11
Page 29: Iec 61000  4-11

TEST & MEASUR

E

VOLTAGE DIPS/SAG

SHORT INTERRUPTION

VOLTAGE VARIATION

SUMMARYIEC 61000-4-11

2 TYPES TEST CONFIGURATION

COMBINATION OF

TRANSFORMER TAP

IMMUNITY TEST GENERATOR

USING VARIABLE TRANSFORMER