Page 1: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

New Immunity Basic Standards: IEC 61000-4-31 and -4-39

Andreas KlinkAmetek CTS Europe GmbH Berlin

Page 2: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Standard overview

Page 3: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

IEC 61000-4-31

Testing and measurement techniques AC mains ports broadband conducted disturbance immunity (IEC 61000-4-31)

Page 4: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

IntroductionSo far covered with IEC 61000-4-6: Disturbances coming from intended radio-frequency (RF) transmitters.Increase of conducted disturbances caused by broadband signal sources on AC mains ports

Intended:Power line telecommunications (PLT)

Conducted broadband disturbance

Unintended:Other noises generated by connected electrical and electronic equipment to the AC mains network

IEC 61000-4-31

Page 5: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Coupling/Decouping Network for DM (CDND)

CMDM

Brodband noise DM = Differential modeCM = Common mode

Page 6: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Coupling NetworksCoupling of the broadband test signal Common mode / differential mode impedanceDecoupling function to other devices are not under testCDNDs (D stands for differential) for AC ports and CDNs (IEC 61000-4-6) for all other ports

Page 7: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

CDND according IEC 61000-4-31Frequency range 150 kHz to 80 MHzLongitudinal conversion loss (LCL) of 16 dB in order to couple the common mode signal as well as the differential mode signal simultaneously to the EUT line

UCM (common modeasymm.mode)

UDM (diff.-mode, balanced or symm. mode)

UDM

Balanced pair

LCL Example:

LCL = 0 dB, UDM = UCM

LCL = 16 dB, 1/6.3 x UDM = UCM

LCL = 20 dB, 1/10 x UDM = UCM

LCL = 40 dB, 1/100 x UDM = UCM

LCL = 60 dB, 1/1000 x UDM = UCM

LCL = 80 dB, 1/10000 x UDM = UCM𝐿𝐿𝐿𝐿𝐿𝐿 = 20 ∗log ( UDM/UCM)

Page 8: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Coupling Network

CMDM

Brodband noise

LCL = 16 dB, UDM /6.3 = UCM

UDM = 6.3 x UCM

Page 9: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

CDND according IEC 61000-4-31Common mode (CM) impedance 25 ΩDifferential mode (DM) impedance 100 ΩDecoupling >15 dB

Page 10: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Test levels

Page 11: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Disturbance signal

Page 12: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

GeneratorWhite noise source 150 kHz to 80 MHzPulse modulation 1 Hz and 2 Hz (50 % duty cycle)Broadband power amplifierLow-pass filter (LPF), and/or a high-pass filter (HPF) to avoid interference with some types of EUTSolution: Teseq NSG 4031

Page 13: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Block diagram of NSG 4031

Page 14: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Example of setup for level setting

Page 15: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Test setup

Page 16: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Test example: PLT adapter

Page 17: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Standard overview

Page 18: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

IEC 61000-4-39

Testing and measurement techniques – Radiated fields in close proximity - immunity test (IEC 61000-4-39)

Page 19: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

IntroductionIncrease of RF sources in the near fieldChange form broadcast transmitting antennas on the hill to the phone in every pocket

Page 20: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

IntroductionExamples of RF sources in the near fieldThe possibility is given that such a RF source comes to a close distance and could influence the environment. Typical situations are in air planes, cars and in the medical environment.The advice to switch off the devices is often not possible.The way out is to make the devices immune against near field disturbances.

Page 21: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

IEC 61000-4-39

Testing IEC 61000-4-39

H Field150 kHz – 26 MHz100 mm Loop

RF Field380 MHz – 6 GHzTEM Horn

RF Field 26 MHz – 380 MHzTBD

H Field9 kHz – 150 kHz120 mm Loop

Page 22: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Test levels

Frequency range 9 kHz to 150 kHz 150 kHz to 26 MHz 380 MHz to 6 GHzLevel Test field strength Test field strength Test field strength

A/m A/m V/m1 1 0,1 102 3 0,3 303 10 1 1004 30 3 300X Special Special Special

Page 23: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Modulation

Frequency range 9 kHz to 150 kHz 150 kHz to 26 MHz 380 MHz to 6 GHz

Modulation AM PM PM

Frequency 1 kHz 2 Hz, 1 kHz 2 Hz, 217 Hz, 1 kHz

Parameter 80% 50 % duty cycle 50 % duty cycle

on/off ratio --- 20 dB 20 dB

Level setting CW CW CW

Page 24: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Equipment

Frequency range 9 kHz to 150 kHz 150 kHz to 26 MHz 380 MHz to 6 GHzField magnetic magnetic RF

Radiating loop Radiating loop TEM hornDiameter 120 mm 100 mm ---

Number o turns 20 3 ---Distance to EUT 50 mm 50 mm 100 mm

SetupCurrent probe or

shunt ---Directional coupler,

power meterSensor Field sensor probe Field sensor probe Field strength probe

Diameter 40 mm 40 mm ---Number o turns 51 1 ---

Page 25: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

ExampleRadiating Loop Antenna9 kHz – 150 kHz120 mm

Radiating Loop Antenna150 kHz – 26 MHz100 mm

TEM Horn380 MHz – 6 GHz

Spacer Spacer

TEM JIG

Field Sensor Loop9 kHz – 150 kHz40 mm, 51 turns

Field Sensor Loop150 kHz – 26 MHz40 mm.1 turn

Field Probe380 MHz – 6 GHz

Page 26: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

TEM horn

Page 27: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

TEM hornOne TEM horn for the frequency range 600 MHz to 6 GHzExtendable with matching networks for the TETRA 400 and GMRS 460 / FRS 460 applicationsHigh efficiency

Page 28: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Verification of the field uniformity

TEM Horn Antenna

Field probe(single axis)

100m ± 5mm

0 to -4dBuniform area

100mm

Page 29: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Field uniformity examples

Page 30: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Test execution

Page 31: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Polarisation

Page 32: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

Test equipment

Page 33: New Immunity Basic Standards: IEC 61000 -4-31 and -4-39 · (IEC 61000-4-6) for all other ports. CDND according IEC 61000-4-31 Frequency range 150 kHz to 80 MHz Longitudinal conversion

The End Any Questions?