iffichanically agitated gas-liquid contactors by a …
TRANSCRIPT
EFFECT OF HORIZONTAL BAFFLES ON THE PERFORIIANCE OF
IffiCHANICALLY AGITATED GAS-LIQUID CONTACTORS
by
HIMANSHU ZINZUWADIA, B. of Ch.E.
A THESIS
IN
CHEMICAL ENGINEERING
Submitted to the Graduate Faculty of Texas Tech University in Partial Fulfillment of the Requirements for
the Degree of
MASTER OF SCIENCE
IN
CHEMICAL ENGINEERING
Approved
Accepted
May, 1987
%06
' ' ' ACKNOWLEDGEMENTS
The author wishes to express his deep appreciation to Dr. Uzi
Mann for his support and encouragement during the course of this
work and his assistance in writing this thesis.
The author would like to thank Dr. P.E. Fischer and Dr. R.R.
Rhinehart for serving on the thesis committee and for giving
valuable suggestions.
Special thanks are due to Mr. R.M. Spruill for his help and
suggestions in constructing the experimental system. The author
sincerely appreciates the constant support and help of fellow
students Mr. Vijay Shirsat and Mr. Rajiv Doshi.
Financial support from the National Science Foundation and a
donation of some equipment by Mixing Equipment Company is
gratefully acknowledged.
11
TABLE OF C0NTENT5
ACKNOWLEDGEMENTS
ABSTRACT
LISTOF FIGURES
LIST OF SYMBOLS vi
LISTOFTABLES
CHAPTER
1. INTRODUCTION
1.1 Background
1.2 Literature Review
2. THEORETICAL CONSIDERATIONS :
2.1 Determination of Volumetric Mass-Transfer Coefficient and Interfacial Area :
2.2 Measurement of Power Requirement
3. EQUIPMENT AND PROCEDURE
3.1 Descríption of the Equipment
3.2 Operating Procedure
3.3 Analytical Procedure
4. RESULTS ANDDISCUSSION
4.1 Presentation of Data
4.2 Normalîzation of Data
4.3 Validity of the Model and the Accuracv of Experimental Data
iii
4.4 Comparison of Results for Horízontal and Vertical Baf f les 60
4.5 Analysis of Results 77
5. CONCLUSIONS AND RECOMMENDATIONS 87
5.1 Conclusions 87
5.2 Recommendations 88
LITERATURECITED 93
APPENDICES
A. A SAMPLE CALCULATION OF
MASS-TRANSFER COEFFICIENT 93 B. EXPERIMENTAL DATA 100 C. STRAIN MEASUREMENT WITH
A RESISTANCE STRAIN GAUGE 1 1 0
IV
ABSTRACT
The objective of this investigation was to study the effect of
horizontal baffles on the performance of a mechanically agitated
gas-liquid contactor in comparison to its performance with vertical
baff les. The performance cr i ter ia were: the overall average
volumetric mass-transfer coefficient, k^a, and the agitation power.
The investigation was conducted on a lab-scale tank 1 f t . in
diameter and 2 f t . height agitated with a 3" Rushton Turbine. Air
was injected into a 0.25M Sodium Sulfite solution, and the oxygen
consumption rate was determined. The primary independent
variables were the impeller speed (changed from 500 to 1450 rpm.)
and the rate of air injection (changed from zero to 0.67 l./sec).
The horizontal baffles improved the contactor performance
over that when vertical baffles were used. The values of kjâ were
80% tol30% higher, while the agitation power required was as much
as 33% lower at certain operating conditions. The improvement was
more pronounced at impeller speeds of 800-1450 rpm.
The experimental data indicated that the improvement in kjâ
was identical to the increase of the gas holdup in the liquid. It was
seen that the horizontal baffles induce entrainment from the upper
surface of the liquid (surface aeration) and the increase of k a is
due to a higher entrainment of gas in the liquid and not due to
smaller size bubbles.
The results of this investigation suggested that there is a
potential for using horizontal baffles on an industrial scale.
LISTOFFIGURES
1.1 Schematic of a Mechanically Agitated Gas-Liquid Contactor
1.2 Power Curves for non-Aerated Liquid (Rushton [ 1950])
1.3 Development of Cavities with Increasing Aeration Number
1.4 Appearance of Different Cavity Regimes in an Agitated Contactor wíthout Recirculation (Van't Reit et al.. [ 1976])
1.5 Reduction in Power of Agitatíon with and without Gas
Recirculation (Van't Reit et a l . í l976] )
2. l Concentration Profiles near the Interface
2.2 Stress and Strain Relations in a Shaft
3.1 Schematic Diagram of the Experimental System
3.2 Tank Assembly with Vertical Baffles
3.3 Tank Details
3.4 Details of Vertical Baffles and Tank Lid
3.5 Details of Horizontal Baffles
3.6 Tank Assembly with Horizontal Baffles
3.7 Rushton Turbine Impeller (R-100) of 3" Size
4.1 Average Interfacial Area for Vertical Baffles
4.2 Average Interfacial Area for Horizontal Baffles
4.3 Effect of Gas Flow Rate on Average Interfacial Area
for Vertical Baffles
4.4 Effect of Gas Flow Rate on Average Interfacial Area
for Horízontal Baffles
4.5 Correlation of Average Interfacial Area for
Vertical Baffles
vi
4.6 Correlation of Average Interfacial Area for Horizontal Baffles
4.7 Comparison of Average Interfacial Area for Vertical and Horizonta) Baffles
4.8 Gas Holdup for Vertical Baffles
4.9 Gas Holdup for Horizontal Baffles
4.10 Comparison of Power of Agitation without Aeration for Vertical and Horizontal baffles
4.1 1 Plots of Pg versus Ny for Vertical and Horizontal Baf fles (20 rps.)
4.1 2 Plots of P3 versus N^ for Vertical and Horizontal Baffles (21.7 rps.)
4.1 3 Plots of Pg versus N/ for Vertícal and Horizontal Baffles (23.3 rps.)
4.14 Power of Agitation with Aeration to Gas Holdup Ratio for Vertical and Horizontal Baffles
4.15 Mean Bubble Diameter for Vertical Baf f les
4. l 6 Reduction in Power for Vertical Baffles as a Function of Effectíve Aeration Number
4.17 Rates of Gas Recirculation in the Contactor without Gas Injection for Horizontal Baffles
Vll
LISTOF SYMBOLS
a Average gas-Liquid interfacial area per unit volume of
dispersion, mVm^
a' Average gas-Liquid interfacial area per unit volume of liquid.
mVm^
â'i Average gas-Liquid interfacial area per unit volume of liquid
f o rH /T= l
C Concentration
C/ j Concentration of gaseous solute at the gas-liquid interface,
kmole/m ^
C^b Concentration of gaseous solute in the bulk of the liqid,
kmole/m^
d Mean bubble díameter, m.
D Impeller diamter, m
D Diffusivity of the solute gas, mVsec.
E Modulus of elasticity, N/m^
g Acceleration due to gravity, m/sec^
H Liquid height,m.
H/ Dispersion height, m.
Hj Impeller height of f the bottom of the tank, m.
He Henry's law constant, kPa.mVkmole.
k Pseudo-first order reaction constant, sec-1.
k| Mass transfer coefficient, m/sec.
ki° Mass transfer coefficient for absorption without chemical
reaction, m/sec.
viii
kjâ Overall average volumetric mass transfer coefficient, sec"
n Rotational speed, rps.
Ny Aeration number.
Ny ^ Effective aeration number.
Npp Froude number.
Np Power number.
Npø Reynolds number.
NQ Impeller discharge coefficient in liquid.
NQ(J Impeller díscharge coefficient in dispersion.
P3 Power of agitation under aerated condition, Watts.
Po Power of agitation under non-aerated condition, Watts.
PAO. PA
Partial pressure of gaseous solute, kPa.
P/ j Partial pressure of gaseous solute at the interface, kPa.
Qp Rate of gas entrainment from surface or surface aeration,
mVsec
Qq FI0W rate of sparged gas, mVsec.
Qp Pumping rate of the impeller, mVsec.
r^ Local Rate of reaction of gaseous solute A, kmoles/mVsec.
r^ Overall Rate of reaction of gaseous solute A, kmol/mVsec.
R Universal gas law constant
T Tank diameter, m.
V5 Linear or superficial velocity of the gas, m/sec.
V Volume , m^
IX
Greek svmbols:
e Temperature, °K.
f Tensile strain along the principal axis of the shaft,
S Film thickness according to film theory, m.
e Gas hold-up, m gas/m^ of dispersion.
[1 Viscosity of the liquid, kg.m/sec.
p Density of the liquid, kg/m^
SubscriDts:
L,l Liquid
G,g Gas
d Dispersion
LISTOFTABLES
1.1 Correlations for Reduction in Power Consumption in Presence of a Gas f or a Rushton Turbine
1.2 Mass-Transfer Correlations
3.1 Summary of Important Contactor Dimensions and Ratios 4
4.1 Scheme of Experiments 5
4.2 Reproducibility of Experimental Data 5
B. 1 Inlet Conditions and Other Data for Set 2 to Set 7 1C
B.2 Power of Agitation without Gas Injection for Vertical Baffles IC
B.3 Power of Agitation without Gas Injection for Horizontal Baffles
B.4 Observations and Calculations for Set 2
B.5 Observations and Calculations for Set 3
B.6 Observations and Calculations for Set 4
B.7 Observations and Calculations for Set 5
B.8 Observations and Calculations for Set 6
B.9 Observations and Calculations for Set 7
XI
CHAPTER 1
INTRODUCTION
1.1 Backaround
Many chemical processes involve contacting a gas and a liquid.
For good contacting, the gas must be very well dispersed in the
liquid. Such a dispersion can be achieved by injecting the gas into
the liquid through one or several orifices (spargers, perforated
rings, perforated plates, etc.) placed under a mechanical agitator
which breaks the gas bubbles into smaller ones and disperses them
throughout the liquid phase. Because of the increase in the surface
area due to bubble breakage and better dispersion of the bubbles,
the mass-transfer rate between the gas and liquid is greatly
improved with mechanical agitation. Such an arrangement for
contacting gas and liquid is generally called "Mechanically Agitated
Gas-Liquid Contactor" and will be referred to as an agitated
contactor throughout this thesis.
Agitated contactors are generally used for processes involving
an absorption of a slightly soluble gas which undergoes a chemical
reaction in the liquid phase. In such processes the reaction rate is
usually much faster than the absorption rate. These contactors are
commonly used in biological fermentations (e.g., manufacture of
penicillin), hydrogenations (e.g., manufacture of margarine from
vegetable oils), and oxidations (e.g., biological oxidation of chemical
and municipal waste). The design of such contactors is reliable and
they can be operated under wide range of operating conditions. In
1
most applications, the design of agitated contactors involves two
major considerations: (i) the overall gas-liquid mass-transfer rate
and, (ii) the agitator power requirement. Besides these, there are
secondary considerations such as, blending of the reactor contents,
heat-transfer, and solids suspension, which are usuaHy not critical
because they are automatically provided for when the design is
satisf actory f or mass transf er consideration (Uhl and Gray [ 1966] ).
The aim of a good design then is to obtain maximum mass-transfer
rate, and at the same time keep the power consumption to a
minimum.
A schematic diagram of a typical agitated contactor is shown
in Fig.1.1. Major dimensions are shown relative to the tank
diameter (T). The important internal elements are; a flat bladed
disk turbine impeller for agitation (also referred to as a Rushton
turbine), a sparger ring below the impeller to inject the gas and,
four vertical baffles extending throughout the depth of liquid. It
may be noted that the vertical baffles are offset from the tank waV
to minimize formation of stagnant zones near the baffles.
The four vertical baffles serve to convert rotational motion oi
the liquid to vertical motion and hence prevent the formation of Í
vortex. It is necessary to prevent the formation of a vortex in ar
agitated contactor because, when a vortex develops the fluid swirl^
around as one element and there is no turnover of the 1iqui(
contents and the dispersion of the gas in the liquid is poor.
The use of such a design sometimes leads to high values o
shear rates in the vicinity of the impeller and hence high powe
Gas Outlet-
10
0.5 T
t=s
i^
0 0
0 0 0
0 ° 0 0 ^ 0
0
0
0 o
0 ^ .D
O
< •
0
0 0 0
0 0 0
0
0 0
0
0
0
0
b o 0
0
o©-o 0
^ O ^ 0 . 0 0
< •
T
împeller Shaft
Gas Inlet
Baffle
1.5 T
t mpeller (D/T = 0.33)
Sparger
Figure 1.1. Schematic of a Mechanically Agitated Gas-Liquid Contactor
4
requirement for agitation. For reducing the power requirement to
get the same or higher mass-transfer rates, a different type of
baffle arrangement called horizontal baffles was examined. From
preliminary visual observations with this type of baffles it was
seen that they served effectively to prevent the formation of a
vortex and the power requirement for agitation was much less
compared to that f or vertical baf f les.
The objective of this investigation then was to determine
whether an improvement in the performance of an agitated
gas-liquid contactor could be achieved by using horizontal baffles
instead of the commonly-used vertical baffles. The performance
criterion to be examined was overall mass-transfer rate for a given
power input.
For these purposes an experimental system of the type shown
in Fig.1.1 was constructed. Experimental methods for determining
the power of agitation and the gas-liquid interfacial area were
devised. The details of the system configuration and the
experimental procedure are given in Chapter 3.
1.2 Literature Review
1.2.1 Dimensionless Numbers Used in Mixing Analysis
Agitation Revnolds Number (NR^). This group represents the
ratio of inertial forces to viscous forces. It is commonly defined as
pvL/|JL, where, p is the density of the liquid, \i is the viscosity of
the liquid, v denotes a characteristic velocity and L denotes a
characteristic length. For agitation L is taken as the diameter of
the impeller, D, and v is taken as the tip speed of the impeller, nD,
where n is the rotational speed of the impeller. Substitution gives,
D np N Re
M-
Agitation Reynolds number determines whether the flow is laminar
or turbulent and hence it is a critical group in correlating power.
Froude Number (Npp). This group represents the ratio of
inertial to gravitational forces and is defined as vVLg, where g is
the gravitational acceleration. Substituting the characteristic
length and velocity for agitation as before the Froude number for
agitation becomes,
n D Ncn = Fr =
g
In most fluid flow problems, gravitational effects are unimportant
and the Froude number is not a significant factor. The reason it is
included in the dimensional analysis of agitated tanks is that most
agitation operations are carried out with a free liquid surface in the
tank. The shape of the surface and, therefore, the flow pattern in
the vessel, are affected by the gravitational field. This is
particularly noticeable in unbaffled tanks where vortex formation
occurs; the shape of the vortex represents a balancing of
gravitational and inertial forces.
Power Number (Np). The power number is the ratio of the
power applied by agitation (P) to the pumping produced by the
rotat ing impeller and is defined as P/mv^ where P is the power
applied by mechanical agitation, m is the mass of the liquid
displaced by the impeller per unit time and v is a characteristic
velocity. The quantity m can be expressed as pnD^ As before, the
characteristic velocity is the impeller tip speed nD. Thus power
number for agitation can be writ ten as.
Np = 3 n 5 pn^D
The physical significance of the Power number is similar to the drag
coefficient in fluid flow analysis.
Aeration Number (N/v ). This dimensionless number is the ratio
of the gas flow rate, Qg, to the impeller pumping capacity which is
proportional to nD^ Hence the Aeration number for agitation is
defined as,
Qg
nD'
Aeration number comes into play only in the case where a gas is
introduced into the liquid. It is useful in correlating the reduction
in power consumption in gas-liquid dispersion.
Weher Number (N\A/^). Weber number represents the effect of
surface forces on the bubbles size. It is the ratio of inertial
7
forces to surface tension forces and is defined as pv^L/cj.
Substituting for v and L as before, the Weber number for agitation
can be wri t ten as,
pN^D' Nwe = •
o
It appears in many correlations where the surface tension effects
are important.
1.2.2 Agitation Power for Contactors in Absence of Gas Injection
The subject of power requirement in agitated contactors with
Newtonian liquids is extensively covered in the chemical engineering
l i terature. A comprehensive analysis was prepared by Rushton £t.
a]., (1950 a,b). More recently, the subject was reviewed by Holland
and Chapman (1966), Uhl and Gray (1966), Nagata (1975), and
Midoux and Charpentier (1 984).
The power requirement is usually expressed as curves of the
power number, Np, versus the Reynolds number, Np^. Each
combination of impeller type and vessel geometry has i ts unique
power curve; however, the same power curve can be used for
di f ferent size vessels if the geometric similarity is preserved.
Some typical curves for the Rushton turbine impeller that is used in
this investigation are shown in Figure. 1.2. When the vessel does
not have baffles or has inadequate baffling, a vortex is formed, and
Froude number, Npp, must be introduced for Np^ greater than 300.
Most power curves available in the l i terature were developed for
Newtonian liquids.
8
Q
4b
aff
les,
b/T
= 0
.1
Q
No
ba
ffle
s
0 1
© • — — — —
.
o o
q d
o o
o o o o o o
o o o o o
o o o o
o o o
o o
CD
^
t _ CD
o E Z5
z: 00 ;a o c CD
cr
CD
Aer
at
1
c ^ o . c o (_ LO O (J\
**— — cn '— CD C > O ( _ - i - j
Z5 J C LJ CO
~^
kS í^ ^ ^ O "O
^ 5 CM 1 ]
Fig
ure
N 'jaquunN JOMOCJ
1.2.3 Hydrodynamics and Agitation Power for Contactor in Presence of Gas Injection
When gas is introduced into an agitated contactor the power
requirement is reduced. In a comprehensive investigation of
hydrodynamic conditions, Van't Reit et a1.. (1973,1974,1976) and
Bruij in et a!.. (1974) have found that the formation of stable
cavities behind the impeller reduces form drag and hence the power
requirement.
Various regimes of cavity formation for a six blade Rushton
turbine operating with increasing aeration number are shown in
Fig.1.3. According to Van't Reit et al.,(1976) the appearance of
cavit ies wi th increasing Aeration number follows the following
order:
a. Six pairs of vortex cavities; one pair of cavity on each blade.
b. Six clinging or adherent cavities.
c. Three large cavities alternately positioned with respect to
three clinging cavities, the 3-3 configuration.
d. Two groups of three large cavities with larger and smaller on
successive blades.
e. Three bridging cavities arranged symmetrically and fiHing the
space between alternate blades, which is also called the flooding of
the impeller.
According to Rruijin et al.. (1974), it is possible to isolate the
effect of cavities on the agitation power if the recirculation of the
gas at the agitator level is prevented, (this was done by placing the
impeller in a large square vessel where the gas was sparged below
10 lOi
y . -V
(a) 5 Vortex Cavities (b) 6 Clinging Cavities
(c) 3 Clinging and 3 Large Cavities
(d) 6 Large Cavities of different size
(e) Brídging Cavities or flooding
Figure 1.3. Development of Cavities with Increasing Aeration Number
11
the impeller, passed through the impeller only once and then
escaped to the surface). In that case the reduction in power,
Pa^^o, with increasing aeration number depends only on the type and
size of the agitator. Fig.1.4 shows the appearance of various
cavity regimes on the plot of Pg/Po vs. Aeration Number, N^. On the
other hand if recirculation of the gas at the agitator level is
permitted, which is usually the case in the operation of most
agitated contactors, then the relationship of Pa/Po vs. N^ not only
depends on the impeller type but also on various other factors such
as tank geometry, baffle type, and on the speed of rotation of the
impeller. Power reduction curves for a typical agitated contactor
of the type shown in Fig. 1.1 are shown in Fig. 1.5.
According to Roustan (1978), these curves can be brought
together to give a unique power reduction curve for a given impeller
by the following considerations: The recirculating flow of gas, 0^',
can be expressed as the product of the pumping flow from the
impeller, Qp, and the holdup of the gas in the tank as below:
Qc' = Qp.e. (1.1)
Then, an effective aeration number, N/^^, can be defined as,
Qg + Qc' N A ^ = —r- = NA - N Q ^ . C . (1.2)
where NQ^J is defined as,
Qp NQH = — — . (1.3)
^^ nD^
Nûd ^ considered to be the discharge coefficient of the impeller in
z
lO
o
'T o <D
ro c o
O î
O
hO
Q c
II o < o :z
(T) CD
-•—» (\5 > (^ CJ
CT) c - r —
O^ c *"" o • o
c ca CD
(_ (^
•
• . o
. .
(f) C D
-»—> (\3 > ca o -í-»
c CD (_ CD XZ TD CO
CT) C
Jyí O O
o
CD
* ""
cn CD
•4->
(^ > (^ o CD cr>
• o CD
4—» (^
•*->
'S>^ < û c r-( a>
. c , ,
• • —
cn ^ ra
- ^ -•—' o> <i> .Sí -^ Û : •T
CD
4—» • • — ( J
> -^ ^ (J ^ - » — '
c "- CD C
CD — « 4 - - ^ - '
««_ C^
C ) ==5 O
t: í-O — ^ o CD <u
^ cr c (^ - ^ l_ 13 (^ O CD J C C2L •>->
JD
cn CD -»—' ( > ( o X CD
O >
C^
cz CL <
CD
O
=3 O
O O
c\ o
CD vû 6 o
( o Û.
13
00 o
o
O Q c
O
o
CM
o
o d
ÍD
in
CN
O o
co o
• o
II c c o
• ^ —
- 1 — »
(^ f '
Z3 o (_
•r^~ O CD
o (^ -»-» c o o • o <D -*-» (^
-4—»
o> co
o o
<D > l _ ZJ ( J
o -C
• o c ; : i i , (^ vû
-•-» CJ>
^ : ^ ^
c r
(^ CD
cr>+-' < .S^ o ^ L_ ' c CD (^
^ >
o ^ ^ c c o
(^ c o
ÍS • o o CD CD
Û: cr ^ ÍS
CD
o>
d 00
d d d d •^ d d
C\J
d o d
(^ æ
14
the presence of a gas. Since it is difficult to measure the
velocity and flow rates of a dispersion, the value of \^Q^ has not
been found for gas-liquid dispersions by any investigator so far, but
according to the hypothesis of Cooper and Wolf (1958), which
assumes that the presence of gas does not appreciably modify the
pumping flow from the impeller, the discharge coefficient of the
impeller in presence of gas remains the same as the discharge
coefficient of the impeller, NQ, in liquid alone. Fairly accurate
values of NQ for a11 the common impellers are given in the
literature. The value of NQ for a Rushton turbine can be estimated
according to the following relation (See for example, Uhl and Gray
[1955]);
NQ = 0.98 (T/D) (1.4)
Using the approach outlined above, the relations between the
impeller power reduction and Aeration number for a given impeller
rotating in vessels of different geometry can be brought together
on a common basis. Though the above approach is the most
fundamental one available to date, the understanding of
hydrodynamic mechanism was expanded only during the last ten
years.
It is a common practice in the existing literature to correlate
the reduction in power in the form of empirical dimensional
correlations. Several investigators have correlated the ratio of
power of agitation with gas injection, P^, and the power of
agitation without gas injection, Po, with various parameters such as
15
impeller diameter, D, impeller speed, n, and volumetric gas flow
rate, Qg. The usual group used for correlation is the Aeration
number, N^. Some correlations also include dimensionless groups of
the physical properties of the two phases. Table 1.1 gives the
summary of correlations and their range of applicability. In Table
1.1, T is the tank diameter, D is the impeller diameter, H is the
liquid height, \-\^ is the impeller height off the bottom of the tank,
and Vg is the linear or the superficial gas velocity and is
obtained by dividing the volumetric gas flow rate, Qg, by the tank
cross sectional area.
Hughmark (1980) made a stat ist ical analysis of most of the
reported data and correlations and found the correlations of Michael
and Miller (1952) and Luong and Volesky (1979) to be the most
suitable for correlating the aerated power data over a wide range of
variables. Hughmark's analysis indicates that the effect of surface
tension is not significant in correlating the reduction in power
consumption to the system variables. Calculated values from
correlation of Loung and Volesky (1979) show a i21 .4% from the
experimental data. Hughmark (1980) has also revised the
correlation of Loung and Volesky (1979), and showed that using the
liquid volume, V^, in the contactor in place of D in the correlation
improves the correlation coefficients. He also used a modified
Froude number to obtain a correlation that showed an average
deviation of i8.5% between calculated and reported experimental
data.
16
CD C
c o
4—>
CL
E CO c o (J
CD C
g ^ — C_ • ^ o o C_ • o í^ CD ^
Û : ^ £_ C^ O , ^
^ CD
1 ->-» CD (\3 CO
o -h ^ o c u —
CD
" O CD
cn
opo
(_ C L
tio
n
( ^
CD c_ í _ O U
<<—
Ran
ge
Pro
p
o £ - <D O CT)
cn ^ V \ E E _ o> æ a >
Q • _
X h-
X
\—
Q
h-
tem
co > . co
CD
•^en
c
CD
CD
Û :
i n r o O
CD
V cn
O
m
O 1 c
( N
1 —
II
cr-, cf r
_ : o>4 1
I '
d r-II »
1 O
X
00
V c n
O
r o r o r o r o
d d i n _
o d
cu
1 (_ <
c
i r co
(VJ O ^
i n r o O
d A
c n O
r^ r > - i
CTI c i n <ID
1 1
CN vû
d II
ro o C 1 CL
00 CN
1
i n d II
i l
cr>
o o
" ^ o
• t—*
i n •<q-
d
n^ c
CN o Q.
vD i n
^ o . O
o II
03 CL
i n g
•~ CN 1
' 1 cn
a^ CN o o " ^
II " _,
O
^ o
d II O)
>
r o _ • — o
o^ — r^
d o o ^ ^ ro -^ <^ ro d o c>
o o — ro
d d
O CN —
u < e 1 1 OJ 1 u ^ l_ l_ > »
< < < 5 (N
ra _
"rã <1J ^
u .—
f H
1 - c n
cx ' ã. O) o
O 1 c ^
E'"^^-^ ^
Q T
Q
^ ^ c _ ^
CN
d 1
.^ II
° E o oo'
r o
II o
TO o " ^ 1 c^ c
~ • ^ r o
' 00
O ^ d
II "
UJ CN o O
f ^ i n o II II cn
>
—
r o T
d d
L') CTi — CN
d d
cn c Z) o
1 ? ^ 5J zj
- cr o < ra tn
"O c
c ° , r^ t\! c r^ j i -Q " ^ ro o I Q : ^
OJ c o l _
L-
o
CN
d
ll
c
E
—
1
00
II
o^ l Q O 1 c
1 ^ - ^^^
c£~ Q to l ^ *
O
II
^ fO o
(."M
r o i n 1 LD _ " ' L,0 I 00
CN d
II "
1
—
r o r o
d
CN
d " 0 0 ,
i ^ 0 0 ,
1 c c
< <
"O
cr> in 2 c Qj r
0 0 _ _ ) > >.
' j i í_ "O 0 =;
>— - ^ • ^ • -
^ S 0 g II t c ^
- ^
cr
0 II
0
00' ^~
d 1 II
Ol
û . c
- 1
^
^ \
'—
D
t :
cr>
c
CL
c 0
r
17
Based on statistical analysis of reported data, the correlation
of Hughmark (1980) is the most recommended one for Newtonian
liquids (Joshi et al.. [ 1982] ).
1.2.4 Gas-Liquid Mass-Transfer
Mass-transfer rate (r) in agitated tanks is usually expressed
as the amount of gas absorbed per unit volume of the gas-liquid
dispersion per unit time. This quantity depends on the gas-liquid
in ter fac ia l area per unit volume of dispersion, 3_, and the
concentration driving force (CQ-C|_)- The overall mass-transfer
coefficient, K|_, is defined by the relationship;
r = Ki_a . (CQ-C[_)
where CQ denotes the concentration of a solute A in the gas phase
and C|_ denotes the concentration of the solute A in the bulk of the
absorbing liquid. Thus, the mass-transfer coefficient is defined as
the rate of mass transfer per unit driving force per unit interfacial
area. It has the units of m/sec. The gas-liquid interfacial area can
be estimated by using physical methods which involve actually
observing the bubble sizes and their distribution, but the physical
methods are tedious and expensive. By using a chemical method in
which rate of absorption of a gas is measured for a known value of
driving force, the product K|_a can be estimated. This is a
convenient quantity and it is extensively used to characterize
mass-transfer rate. This product is commonly referred to as
volumetric average mass-transfer coefficient. Usually it is diff icult
to estimate the interfacial area per unit volume in most chemical
18
methods; however, under some conditions the value of 3 can be
deduced from the measurements of a chemical method. One such
method is used in this investigation and is described in Chapter 2.
When a chemical method is used for calculating K^a. or a.,
information on the behavior of the gas phase is needed to calculate
the average driving force. In most applications, the gas phase is
assumed to be backmixed. Mehta and Sharma (1971) showed that
the gas phase can be considered fully backmixed under the usual
operating conditions in a laboratory-scale mixer.
Most correlations of mass-transfer involve either the overall
mass transfer coefficient, K^â, or the effective interfacial area, a.
All these correlations are in the form of empirical dimensional
relationships between the relevant variables. All the correlations
are of the following general type:
K[_aora is proportional to,
Vs^.nPor v s " . [ ^ ^
In Table 1.2 therefore, the values of the exponents a , p and y for a
turbine wi th six f lat blades are given, together wi th the most
important dimensions of the vessels investigated. The large
variations of the exponent values for different vessel geometries
and different systems indicates that this correlation is not general,
and depends on vessel geometries and the type of gas-liquid
system.
Another correlation method was proposed by Westerterp ei.
al., (1953). They found that the effect of v^ is significant only at a
19
CD C L > >
CZL
£
CD C
j Q C_
o -I—» CD
O i5 O
CD C
O (_
o - » — > <D
O
o
CD C
o c_ o
-t—'
CD
o i5 o
O o (\) C L
<D
o 03
LO
cn c o
- t — »
i5 CD c_ c_ o (_J c_ < D
»<—
cn c ( ^ c_ ^7 cn cn ( ^
21
C l (^ h-
o > c o
- l — »
i5 CD C_ í_ o (-J CD
cn -i_» c CD c o CL X
^-
CÛ_
b
CD O
c CD c_
CD
CD
cn
d LO
d
LD
q CsÍ
co
d
-i—' L_ ^ <D ^
cr> o
<D O
' c co
~ LD I , " O^
o d IT)
d
d
o CN
d
CD Q C
> O
OJ
d ro
d ro d
o
d
o LT)
d
o c
Q CD C P ^ r5 >. o X cû
O IT) O cj> fO —
00
d ro
LO ro
o
d
( l
>
CÛ
c CD <D
20
low value of n. Above a certain value of n (called critical impeiler
speed) they found that Ki_a is independent of v^. This critical
impeller speed, n^, found by considering the circulation of the gas
bubbles and bubble coalescence, is given as,
noD ÍT = A + B —
(a g / p)0-25 l D
where A and B are correlation coefficients with values of 1.22 and
1.25 for a turbine impeller with six flat blades. In the regime above
the critical impeller speed the specific interfacial area is
correlated as follows:
aH = C (n-n^) D vTpTTã')
(1-e) where,
C = (0.79 1 0 . 1 6 ) [i i[i\n cP)
1.2.5 Surface Aeration
Entrainment of the gas into the liquid from the freeboard liquid
surface is called surface aeration. Several types of agitated
contactors, particularly in fermentation industry, are specifically
designed to enhance surface aeration. Calderbank (1959) has
pointed out that surface aeration leads to an increase in the
hold-up of the gas in the contactor and hence to an increase in the
gas-liquid interfacial area. For this reason, surface aeration is an
important consideration in the design of an agitated contactor. The
correlations discussed in the preceding sections ignore the effect
of surface aeration. Very few studies on the effects of surface
2
aeration on mass-transfer and power requirement have been
published.
Matsumara et al.. (1977) have studied surface aeration rates
wi th a Rushton turbine in a tank with four vertical baffles. They
have proposed the following implicit correlation for estimating the
rate of surface aeration (Q^);
-2.2 -n
(l-Tl)
where
= 1.913x10"^ InD
z\ nD
\^ J
0.1 ^n^D^pA
V
1.38 0.07 6.4 n^D'i f D ^
T.
Ti = 0^/(0^+0^)
where Qg is the volumetric flow rate of sparged gas, and Q^ is the
volumetric flow rate of entrained gas. Matsumara et al..( 1978)
have also measured the power consumption in the presence of
surface aeration using a Rushton turbine impeller. The following
correlation has been proposed.
= 0.25 ívo inDj
-0.2 r n D"* 0-05
ig ;
There is not much information in the l i terature about
mass-transfer coefficients in the presence of surface aeration.
Mehta (1970) has obtained the values of effective interafacial area,
a, and liquid side mass transfer coefficient, k|_â, using a six-bladed
disk turbine as the impeller. The values of a and k^â were found to
vary between 125-325 mVm^ and 0.08-0.2 sec" ^ respectively,
when Pa/V|_ was varied in the range of 0.017 to 41.1 1 kW/m^
CHAPTER 2
THEORETICAL C0N5IDERATI0NS
As stated above, the objective of this investigation is to
compare the performance of an agitated contactor with horizontal
baff les to one wi th vert ical baff les. The cr i ter ia used for
comparing the performance are, the mass-transfer rate and the
agitat ion power requirement. In the following sections the
theoretical basis for the experimental methods and the calculation
procedure used in this investigation are developed.
2.1 Determination of Volumetric Mass-Transfer Coefficient and Interfacial Area
A chemical method for determining overall volumetric
mass-transfer coefficient, K\_3, and interfacial area, 3, is used in
this investigation. Below it is shown how these quantities can be
determined from measurable quantities. For a complete analysis,
refer to the original publications on theory of mass-transfer with
chemical reaction by Danckwerts and Sharma (1966,1970). This
theory is a commonly-accepted theory for chemical methods of
measurement of Ki_i and 3.
Consider a system of a gaseous solute A being absorbed in a
liquid B. The concentration profile near the gas-liquid interface at
steady state is shown schematically in Figure 2.1 . In this figure P3
and PAJ denote the partial pressure of the gaseous solute A (which
is a measure of concentration in the gas phase) in the bulk of the
22
»3
gas and that at the interface respectively, whereas, C|_ and C^^
denote the concentration of A in the bulk of the liquid, and at the
interface, respectively.
Now according to the theory of Lewis and Whitman (1923), the
interface is always saturated; therefore, the liquid concentration
and gas concentration at the interface are related by.
^A i = PA
RøHe (2.1)
<D d O o. E o o
cn
cz o -i—>
(^
d OJ <-)
o (_;
C L
Liquid Phase
Distance
Figure2.1. Concentration Profiles near the Interface
24
where He is the Henry's law constant which represents the
equilibrium constant between concentration in the gas phase and in
the liquid phase (when the solubility of the gaseous solute is small
[e.g. oxygen in water] and ideal gas law applies), ø denotes the
temperature, and R denotes the gas law constant. Several
mass- t ransfer coeff ic ients can be defined depending on the
concentrat ions used for the driving force. Mass-transfer
coefficients for the gas side (kQ°) and for the liquid side (k^^) are
defined as the ratios of flux, Ny , of gaseous species A and the
respective driving forces; (PG"PAÍ^ ^^^ ^^^ 9^^ side, and (C/!^J-CL)
for the liquid side. An overall mass-transfer coefficient K^ is
defined based on the overall driving force from gas to liquid
expressed in terms of concentration units. Then at steady state,
the flux of A can be expressed in terms of these different driving
forces as follows;
N A = ^G ° (PG-PAÍ ) = ki ° ( C A Í - C L ) = KL (CG-CL) . (2.2)
Using the equalities in (2.2), i t can be shown that overall
mass-transfer coefficient based on liquid side, K^, is related to
individual mass-transfer coefficients on the gas and the liquid side
as,
1 1 1 (2.3)
KL He kG° kL°
For relatively insoluble gases the value of He is very high and hence
the f i r s t term on the right hand side in (2.3) may be considered
negligible in comparison to the second term. In that case the
25
overall resistance to mass-transfer can be considered to be
entirely in the liquid phase and the flux of A can be wri t ten as.
NA = kL°(CArCL)- (2.4)
The physical significance of the mass-transfer coefficient k ^ is
discussed in the following paragraphs.
According to the well known fi lm theory of mass-transfer
developed by Hiegbie (1935), the resistance to mass-transfer from
gas to liquid can be located in a gas film and liquid film surrounding
the interface. This means that any drop in concentration due to
mass-transfer occurs in a narrow region around the interface. Also
the concentration profiles near the interface are approximated as
straight lines which are shown as dotted lines in Fig. 2.1 above.
Assuming that the gaseous solute A is transfered in the liquid
phase by diffusion, the flux of A in the liquid is expressed as,
dCA NA =-D
dx x = 0. (2.5)
where, D is the d i f fus iv i ty of the gas and dC^/dx is the
concentration gradient at the interface. If the concentration
prof i le in the f i lm is assumed to be linear and a depth of
penetration, 5, defined by Hiegbie's film theory, then the slope of
the concentration profile at steady state is constant and is given
as. dCA
dx
CAI - C L
x = 0 (2.6)
Substituting this result in (2.5),
26
N A = D ^Ai - ^L
(2.7)
From (2.4) and (2.7), it can be seen that according to the film
theory the mass-transfer coeff ic ient, k^^, is the ratio of the
diffusivity to the depth of penetration. Therefore,
k, ° = D
(2.8)
Consider now, a situation where the gaseous solute A reacts
wi th the liquid B to form a liquid product C according to the
reaction,
A(g) + B(L) - C(L).
The reaction is a f i r s t order reaction with a rate constant, k,
defined as,
(-r^)] = kC^. (2.9)
where (-rA)] is the rate of disappearance of A per unit volume of
the liquid. Now if the reaction is much faster compared to
absorption rate so that most of the diffusing gas A is consumed
inside the liquid fi lm then the concentration profi le near the
interface can no longer be considered as a straight line as before.
Danckwerts and Sharma (1963) derived the equation for
concentration profile for absorption accompanied by a fast chemical
reaction, using the film theory, they found the value of the slope of
the concentration profile at x = 0 to be.
dCA
dx x = 0
^Ai - C
VÔTk L
(2.10)
27
Substituting (2.10) into (2.5), an equation for rate of absorption of
gas in presence of a fast f i rs t order chemical reaction is,
H^ = 4W ( C A J - C L ) = kL° VkD / kL° (CAÍ - CL) . (2.11)
Comparing (2.1 1) to (2.4), the term VkD may be considered as the
mass- t ransfer coeff ic ient for gas absorption wi th chemical
reaction and is denoted as k^. Thus,
NA = kL(CAi - C L ) , (2.12)
where, k^ = VkD.
Note that the derivation given above for the mass-transfer
coefficient with a fast chemical reaction in the liquid film applies
only when the following condition is satisfied;
ki VkD (0 = — ^ = > 2 . (2.13)
kL° kL°
In the above equation, cp is commonly called the enhancement
factor. It is the ratio of mass-transfer coefficient with chemical
reaction to mass-transfer coefficient without chemical reaction.
Now, when the reaction rate is much faster than the
absorption rate, the concentration of the solute in the bulk of the
liquid, CL, is very small in comparison to C^j. Therefore, (2.1 2) can
be wr i t ten as,
NA = k L . C A i - (2.14)
For a weak solute and assuming the gas phase behaves as ideal gas,
Henry's law can be used to relate the concentration at the interface
to the partial pressure of the gaseous solute. Equation 2.14 then
J
28
becomes,
NA = k 'Ai
L R eHe
(2.15)
It was shown above (following Equation 2.3) that the partial
pressure of the gaseous solute at the interface, Py j, is equal to its
partial pressure in the bulk of the gas, P^, for negligible gas phase
resistance.
The above analysis is now extended to a gas-liquid reaction in
an agitated contactor in which gas carrying the solute A is
dispersed in a pool of liquid B. The mixture of gas bubbles and liquid
wi1l be referred to as the "dispersion".
Consider a small element of volume dVp in the contactor as
shown below.
This element consists of gas bubbles carrying solute gas A,
dispersed in liquid B. The fractional volume of bubbles is denoted by
£. Thus the volume of the liquid, dV^, in the volume dVp is
(1 -c)dVp. The interfacial area per unit volume of the dispersion is
denoted by a. At steady state, using (2.14), the rate of absorption
in the volume element is therefore.
29
NA a dVn = f P
'A R 1 A
,R GHe, a dV R. (2.16)
As discussed earlier, the local rate of absorption of A is equal to
the local rate of reaction of A. Since the local rate of reaction of
A, ( - P A ) ] , is expressed on the basis of liquid volume, Equation 2.16
can be wr i t ten as,
(-rA)i ( l -£ ) dVp = kL
Upon rearrangement (2.17) becomes.
'A
, ReHe , a dV R-
(-rA)i dVp = k^a
(1-c)
'A
ReHe dV R'
(2.17)
(2.18)
The above equation can be integrated over the entire tank.
/ (- ^ A ) ! dVR VR
1 Í P
(l-£)He VR
A
. Rô kia dV R, (2.19)
The left hand side of (2.19) is the overall rate of reaction of the
gas which is a measurable quantity. To integrate on the right hand
side, the variation of all the quantities inside the integral with the
volume should be known. In an agitated contactor, He is a constant
for isothermal operation. The value of k] for a fast chemical
reaction is given by VkD . For constant temperature, both k and D
are constants. Also, the gas phase is assumed to be fully mixed,
i.e., the concentration of A in the gas is uniform throughout the
tank. It has been shown by Mehta and Sharma (1971) that under
30
usual operating conditions in the laboratory, there is considerable
recirculation of the gas inside the tank and therefore the gas phase
can be considered backmixed. Therefore, a representative gas
phase concentration within the contactor can be taken as a
constant referred to as (PA^^^^av- "^^^^ ef fect ive average
concentration is calculated by considering the oxygen material
balance and the variation of pressure from the bottom of the tank
to the top. (See next section). Under the above assumptions (2.19)
can be wr i t ten as;
/ ,(-r^A)ldVR VR
•1 ^ P A ^
He L R e J av V
(1-e) dVp. (2.20)
R Dividing both sides of the above equation by dispersion volume VR
J- /,(-rA)ldVR VR VR
(P.
He Re av V R V
(1-e) dVR . (2.21)
R Now defining an average interfacial area per unit volume of liquid as
a' = - / k dV VD V D ( 1 - £ )
R (2.22) R R
Eqn. (2.21) reduces to.
( -^Ah kiâ'
He
'A
l R e J (2.23)
av
In the above equation (-rA^d is the overall average rate of
uptake of oxygen per unit volume of the dispersion which is a
measurable quantity. Furthermore, when data are available for the
chemical reaction rate constant, k, and diffusion coeff icient, D,
31
then the constant k^ can be computed as VkD (see Equation 2.11).
The value of He is known for the system used. In that case the
value of 3 is given by.
He (-rA)H ^ ^ . (2.24)
ki (PA/RT)av
Equation 2.24 is the basis for calculating the average interfacial
area per unit volume of the liquid which is used to characterize the
mass-transfer performance.
The reaction system chosen for this investigation was
oxygen-sodium sulfite solution because the reactants are readily
available and inexpensive. This system is commonly used in studies
of this type. (See for example Westerterp et al..í 19631). Sodium
sulfite is oxidized to sodium sulfate in presence of a cupric ion
catalyst according to,
2 S 0 3 ~ + O2 - 2 S 0 4 ~ .
Westerterp et a1., (1963) have shown that this reaction is a fast
chemical reaction of second order in the liquid film with the rate
given by,
( T A ) I = ^' [O2] [503~] . (2.25)
Note that in a solution this second order reaction can be
approximated as a pseudo first order reaction because the sulfite
concentration is in excess of the oxygen concentration and
therefore [SO3--] can be assumed constant. Thus, the overall
average rate of reaction per unit volume of liquid can be written as,
32
(-^A)I = 1 d[C5o-- ]
V L dt = k^O^]. (2.26)
The rate of change of concentration of sulphite is measured
experimentally. If this rate is denoted by d^C^o —]/dt then the
overall average rate of reaction of O^, (~A)d ' ^^ defined before, is
given by.
dC ( - rA)d = - 1 / 2
SO.
dt
^ L
VR (2.27)
Note that here the basis for rate of reaction was changed from unit
volume of the liquid to unit volume of the dispersion. The procedure
for measurement of sulfite oxidation rate is described in Chapter 3.
Westerterp el_al.,( 1963) have also estimated the value of kj
and i ts variation with temperature for the oxygen-sodium sulphite
system to be.
k = 4467.63 f -16500
exp Re
0.5 sec -1 (2.28)
where R is in ca1/gmo1e/°K and e in °K.
2.1.1. Calculation of the Average
Driving Force
The average driving force in the contactor is calculated by
assuming a well mixed gas phase and using a material balance over
oxygen. Consider molar balance of oxygen over the control volume
shown in the diagram on below.
At steady state, an oxygen molar balance over the control
volume gives.
33
îontrol Volume
g in
(rate of O^ in)- (rate of O^ out) = rate of consumption of Oo inside the control volume.
n symbolic form this may be wri t ten as.
n, in • 0,1 out (-^o)dVR. (2.29)
Note here that the subscript A in the previous equations is now
change to denote oxygen. The molar flow rate, n^ ,on the left hand
side of the above equation can be expressed as a product of
volumetric flow rate Q^ and molar concentration CQ . Also the gas
concentration can be expressed in terms of partial pressure of the
gas by using ideal gas law; In that case, (2.29) becomes.
Q gin ^02
, Rø J in + Q gout . Re J out
= (-r^o)dVR. (2.30)
In thisequation the outlet concentration may be taken as the
average concentration in the tank because the assumption that the
gas phase is well mixed. Therefore, the outlet concentration can be
expressed as.
34
í ^o^] (^gin i^oA . (- o,_)d^R (^.31) in í ^02 \ ^ i Re Jin Re Jav Qgo^t l Re j in Qgo^t
Equation 2.31 gives the method for calculating the average driving
force as follows: Here the absorption rate per unit volume of the
dispersion, (-ro^^d, is determined experimentally. Both the
dispersion volume, VR, and the inlet gas flow rate, Qgin, ai e
measured experimentally. The outlet flow rate, Qgout' ^^^ ^e
calculated from the knowledge of inlet flow rate by considering the
change in pressure from inlet to outlet due to hydrostatic head and
change in number of moles due to water evaporation into the gas
and the change of oxygen concentration due to reaction. The
diff iculty here is that the concentration of oxygen at the outlet
cannot be calculated without knowing the outlet flow rate. A trial
and error procedure is used in which the outlet flow rate is f irst
guessed as being equal to the inlet flow rate and the concentration
of oxygen at the outlet is calculated. The outlet flow rate is then
corrected by using this concentration of oxygen. The trial and
error is repeated until a constant value of oxygen concentration is
obtained. A summary of a typical calculation is given in Appendix A.
7 9 Measurement of Power pfiquirement
The agitation power is determined by measuring the shaft
torque and its rotational speed. The following equations summarize
the relationships used in this work. A detailed discussion may be
35
found in the book by Oldshue (1966). The power delivered to a
rotating shaft is given by the following equation.
2TmTe P = =- .
33,000
Where P is the power in horsepower, T^ is the torque in 1bf-in and n
is the rotational speed in rpm. Torque of a circular shaft can be
calculated from the knowledge of shear stress perpendicular to the
axis of the shaft by the foHowing relation;
16To f = r s
TTd^
Here, fg is the maximum shear stress in the cross section of the
shaft and d is the shaft diameter.
The shear stress, f^, is the average of the tension and
compression stresses introduced along the principal axes of the
shaft. The tension and compression stresses are in turn related to
the strains along the principal axes via the elastic relation. These
relationships are;
fg = 1/2 (Tensile stress + Compressive stress)
= 1/2(E. e - E.(ve)).
where E denotes the Modulus of Elasticity, v denotes the Poisson's
ratio, and f denotes the tensile strain along the principal axis.
Figure 2.2. shows these stresses and strains in a rotating shaft.
To determine f^ a measure of the strain is needed. This is
done by using a resistance strain gauge attached to the shaft. They
work on the principle that the resistance of a thin metallic wire
36
<r
Compressive strain
t
Shear stresses
Tensile strain
Figure 2.2. Stress and Strain Relations ín a Shaft
37
varies linearly with the strain acting on it. The details regarding
the construction and use of strain gauges are given in Appendix C.
CHAPTER 3
EQUIPMENT AND PRQCEDURE
The experimental system used in this investigation was
designed to allow measurements of the mass-transfer coefficients
and power requirements under various conditions of mechanical
agitation. The system was designed under two cirteria: (i) to
employ existing equipment and, (ii) ease of operation. The design of
the system was based on the configuration for Mechanically
Agitated Contactors recommended by Qldshue (1966).
3 • 1 DescriDtion of the Eauipment
Fig. 3.1 shows a schematic diagram of the experimental
system. For ease of description the system is divided into three
main sections:
• contactor (or tank),
• agitator assembly, and
• gas supply.
The heart of the system is the contactor. A diagram of the
contactor with drive assembly is given Fig. 3.2. For convinience,
identification codes are given in parenthesis for each part.
A detailed drawing of the tank (TA), is shown in Fig. 3.3. The
tank was constructed of a polyacrylic tube 30.5 cm diameter and
3.2 mm thickness. The bottom of the tank was closed by a
polyacrylic sheet of 6.4 mm thickness. A flange was mounted on
the top of the tank to support the cover (C) which was fastened to
38
39
CD í_ <D d Û -cn o E
•*->
( 3
<
Á
<
c (\3
CO
<D C
cc
CD 4-J <D
E (\3
O
cr
T >
<D >
(^ >
TD <D <D
î
<D
E (\3 C_ cn (\î •.— O CJ
- i —
- » — » (V3
i-<D C. ( ) co
'
ro <D í_ Z5 CD
E <D
-t—'
cn • ^ co , (\1
4—»
c <D
F i _ CD C L X
LU
CL a. Z3
co
<
42
the tank by means of six C-clamps. A sample line (SL) was mounted
near the bottom on the side wall of the tank. A rubber gasket (G)
was used to seal the space between the flange and the cover. The
top cover (C) shown in Fig. 3.4 was made of a polyacrylic sheet of
1 2.7 mm thickness. A bush bearing (BB) was mounted in the center
of the top cover through which the shaft (S) entered the tank. The
cover also had openings for entry of the sparge tube (SP) and
venting the gas. The cover supported four vertical baffles, as
shown in Fig. 3.4, set at ninety degrees to one another. The Four
vertical baffles were made of polyacrylic strips of 25 mm width and
2 mm thick and were suspended from the top cover by means of
screws. The baffles hung from the top cover wi th a 25 mm
clearance from the wall, and extended to the bottom of the tank.
The baff les were fastened to each other at the bottom by
connecting diametrically opposite baffles wi th polyacrylic resin
strips of 3.2 mm thickness.
A set of the horizontal baffles (H) used in this study are shown
in Fig. 3.5. It consisted of four baffles connected to a central hub
to form a cruciform shape, and they extended radially to the tank
walls. The impeller shaft (S) passed through the hub in the center.
The whole unit of horizontal baffles slipped very closely in the tank
and could be supported at any height by means of two set screws
(SS). Fig. 3.6 shows the assembly of the contactor with horizontal
baffles in place.
The agitator assembly was a LIGHTNIN model CV-3 (See Fig.
3.2). The motor (M) had a power of 1 / 4 hp rated at maximum speed
43
12.5 mm
2 mm •#• <^^-
25 mm :^
I •'A'.i. ' . ' ." . ' . ' . . . ." ' U ' "•' " ' " ' " " • ' • ' • •' •.•.'.'.•••••.'.'.'.'•.'.'.'.'.'.?r?j i
o
290 mm
Figure 3.4 Details of Vertical Baffles and Tank Lid
44
SS
(a) Top View
305 mm •
• V ,
r
49 mm 7
(b) Cross Section A-A
Figure 3.5 Details of Honzontal Barfles
46
of 1700rpm. Its maximum rated current was 4 amps. The agitator
drive (A) was a direct drive provided with a ball bearing and a chuck
to connect the impeller shaft. The impeller shaft (S) was a 1 /2 inch
diameter solid circular shaft. The impeller (I) was a six-blade
Rushton turbine impeller (LIGHTNIN type R-100). A detailed
drawing of the impeller is shown in Fig. 3.7.
The power consumption was measured by a resistance strain
gauge (SG) mounted on the shaft below the drive. (See Fig. 3.2). A
MICRO-MEASUREMENTS strain gauge (CEA-1 3-1 87UV-1 20) was
used. It was pasted on the shaft with a special adhesive supplied
by the manufacturer.
The strain gauge was connected to the external circuit by
means of slip rings (SB) made of Inconel alloy and spring loaded
carbon brushes of the type used in small electric motors. (See Fig.
3.2 ).
Air was supplied to the system f rom a central air line. The air'
passed through a pressure reducer (PR) along wi th a pressure
indicator (P) ( U.S.Gage Co. Model N l -1 1 ). Further downstream a
needle valve (V) was used to control the flow rate. (See Fig. 3.1 ).
The flow rate was measured by a previously-caliberated rotameter
(R) (Brooksmodel 1 1 10-06 FIA).
The air was injected into the tank through a copper sparge
tube (SP), one end of which was made into a sparger ring right
underneath the impeller. The sparger ring had a diameter of 5
inches and ten 1/16" diameter holes were dri l led along the
periphery of the sparger ring.
Top View
0.3 mm
2 mm R 19.1 mm^
U . ^ ^ M . l . L i A . ^ ^ ^ ^ ^ ^ ^ L H L > ^ V ^ V k ^ ^ ^ ^ i . k ^ V V V ^ ^ . U A . ^ . ^ ^ g g
56.2 mm
ssa J'
47
2.7 mm
76.2 mm
Section A-A
Figure 3.7 Rushton Turbine Impeller (R-100) of 3" Size
48
3.2 Oneratinn Prorfidurft
Several runs were made to determine the power requirement
and mass-transfer coefficients for different tank configurations.
For each run, the tank configuration, i.e., the baffle arrangement,
impeller height, and liquid height, were selected in advance. Table
3.1> summarizes the important dimensions in the contactor
configuration.
Once the contactor was set up to the desired configuration,
the following procedure was foHowed for each run:
The strain gauge was connected to a bridge circuit for torque
determination and initial reading for value of voltage across the
bridge circuit was taken. (Refer Appendix C).
A known amount of sodium sulphite was then dumped into the
liquid in the tank so as to get a 0.25M solution. The agitation was
started slowly and continued ti l l a11 the sodium sulphite dissolved.
The rpm of the agitator was then adjusted to the desired value.
Once again, the reading for voltage across the bridge circuit under
strained conditions was taken. From the values of these voltage
readings, and the rpm, agitation power without gas injection could
be calculated.
The gas flow was started and the flow rate was adjusted to
the desired value. Since the motor of the agitator was not of
constant-torque type, the rpm of the agitator increased upon
sparging of the gas. The agitator speed was brought down to the
desired value by adjusting the speed controller on the motor. The
voltage across the bridge circuit was measured for operation with
49
Table 3.1 Summary of Important Contactor Dimensions and Ratios
Important Dimensions
Tank diameter, T (m)
Impeller diameter, D (m)
Impeller Height off the tank bottom, H Í (m)
Liquid height, H (m)
0.305
0.0762
0.15
0.45
Vertical baffle dimensions
Baffle width, B (m)
Baffle clearance from the wall, B^ (m)
0.025
0.025
Horizontal baffle dimensions
Height of each baffle, H^ (m)
Distance from the tank bottom to the middle of a horizontal baffle , B j (m)
0.05
0.45
Important Ratios
D/T
Hi /H
B/T
Hp/H
Bd/H
0.25
0.33
0.08
0.1 1
1
50
gas injection. From this va1ue the agitation power requirement with
gas injection was calculated.
The reaction was started by adding a known amout of cupric
sulphate catalyst so as to have a concentration of 10"^ M. The
color of liquid in the tank f i rs t turned turbid and then deep green.
About 10 minutes were allowed to elapse for steady state to be
reached. This time for steady state was decided from experience
on rate measurements under different conditions. The steady state
was decided based on constant value of the rate of reaction. The
mass-transfer coefficient determination was then commenced by
withdrawing the f i rs t sample from the tank via the sampling line
(SL). About 30 m1 of liquid was allowed to run to waste before the
actual sample was taken. The sample analysis was started
immidiately a f ter withdrawing the sample. Five samples at
intervals of 5 to 10 minutes were taken.
3.3 Analvtical Procedure
Elseworth et_ai.,( 1957) described the foHowing procedure for
determination of sulf i te concentration. For any sample a measured
excess (usually 20 m1.) of 0.1 N-iodine was added to 50 ml of boiled
disti l led water and 5 m1 of 30% by volume glacial acetic acid in a
500-m1 conical flask. The sample (10 m1. usually) was pipetted into
the iodine solution by immersing the tip below the liquid surface and
swir l ing the f lask. The excess iodine was t i t ra ted wi th 0.1
N-thiosulfate using starch as an indicator.
The values of the thiosulfate t i t rat ion reading (TT m1) were
plotted against time and a straight line drawn to f i t the points. If p
is the slope of this line, then, the suphite oxidation rate
(dCso^—/dt) is given by 150 x p. The derivation is as follows,
Let T j be the volume of lodine neutralized, and TT^ be the
Thiosulfate t i ter at a time point t p Then, if excess lodine used is
20 ml we may wr i te, for t i t rat ion with thiosulphate,
N,^x(20-T i ) = NThiosu1fateXTTi,
where, Ni^^Njhiosulfate = 0-lN.
Therefore, (20-T1) = TTp
or, Ti = 2 0 - T T | .
From Scott and Furman (1965), 100 ml. of 0.1 N lodine neutralized
= 0.6303 grams of Sodium Sulfite
Concentration of Sodium Sulfite at time t ] for a sample size of 10
m1 is then given by
0.6303 X (20-TT i ) grams/liter
Therefore molar concentration of Sodium Sulfite is then given by,
0.6303 X (20-TT | ) /126 gram moles/liter
The concentrat ion of Sul f i te ion therefore is half of the
concentration of Sodium Sulfite (Na^SOs). The concentration of
Sulfite ion at time t ] is therefore,
Ci = 0.31515 x (20 -TT i ) / 126
= 0.0025 X (20-TTi) gram moles/liter
= 2 .5x (20 -TT i ) milimoles/liter
Similarly the concentration of Sulfite ion at another time t^ , when
the t i t rat ion reading is TT^, can be wr i t t ten as.
52
C2 = 2.5 X (20 -TT^)
Then, if t is in minutes, the rate of change of concentration is given as,
(C2-C 1 ) / ( t2- t 1) = 2.5 X (TT i -TT^^/^t^ - 11)
= 150x(TT1-TT2) / ( t2 - t l )
= 150xp milimoles/lit/hr.
The quantity (TT^ -TT^^ / ^ t ^ - t i ) is the slope of the plot of
Thiosulfate titration reading (TT m1.) vs. time (t min).
CHAPTER 4
RESULTS AND DISCU5SI0N
The use of horizontal baffles instead of vertical baffles proved
to be beneficial in terms of lesser power requirement and higher
mass-transfer rates under the range of conditions investigated. In
this chapter the results for horizontal and vertical baffles are
compared and discussed.
The investigation was divided into eight sets of experiments
listed in Table 4.1. For each set, the baffle type and the injected
gas flow rate were kept constant. Six runs for each set were made
with different impeller speeds. For each run in which gas was
injected from the sparger, the rate of oxygen uptake, the gas
holdup, and the power of agitation were measured. The power of
agitation and the gas holdup were also measured for the runs
without gas injection. From the measured oxygen uptake rates, the
values of volumetric average mass-transfer coefficient, and the
values of average interfacial area per unit volume of the liquid were
calculated according to the procedure described in Chapter 2. The
observations and results are shown in Appendix B.
4.1 Presfintation of Data
The mass-transfer rates in this investigation can be
characterised by the average interfacial area per unit volume of the
liquid, a'. As discussed in Section 1.2.4, the mass-transfer rate in
agitated contactors is usually reported in terms of the volumetric
53
Table 4.1 Scheme of Experiments
54
Set Number
1
2
3
4
5
6
7
8
Gas Flowrate
mVs
0
1.54E -04
3.68 E -04
6.81 E -04
1.54E -04
3.68 E -04
6.81 E -04
0
Baffle Type
Vertical
Vertical
Vertical
Vertical
Horizontal
Horizontal
Horizontal
Horizontal
55
average mass-transfer coefficient, k^a, which is the product of the
average interfacial area per unit volume of the dispersion, 3, and
the mass-transfer coefficient, kj. As shown in Chapter 2, for a
fast chemical reaction, like the oxidation of sulfite used in this
investigation, the value of k is VkD (Eqn. 2.1 1), where k is the
reaction rate constant and D is the molecular diffusivity of oxygen
in the liquid. Both these quantities are physical properties which
are independent of agitation conditions. Hence kj is constant at a
given temperature. Consequentially, the average interfacial area,
a, can be used to evaluate the mass-transfer rate.
Two types of average interfacial area may be defined; one
defined on the basis of area per unit volume of the dispersion, 3,
and the other on the basis of area per unit volume of the liquid, 3.
They are related according to.
a = l -£
where e is the fractional gas holdup. An examination of the
experimental data revealed that when 3 was correlated with the
impeller speed and the gas flow rate, the spread was smaller
compared to the case when a was correlated. Thus it was decided to
report the experimental results in terms of a'.
It is common to report the data of mass-transfer performance
in terms of dimensionless correlations. A dimensional analysis
indicated that a.' can be correlated according to the following
dimensionless relation.
56
â'D = f (NR^, Npp, N^e, N^, H/To, Hi/H.. other geometric ratios).
The values of a11 the geometric ratios except H/T were constant for
all the runs, so their effect could not be considered. The same fluid
type was used in all the runs, therefore the fluid physical
properties (e.g., viscosity, surface tension, density) were fixed and
their effect could not be studied independently. Therefore, the
effect of Reynolds number and Weber number which involve fluid
physical properties could not be studied. Also, the impeller
diameter was constant, so the effect of Froude number could not be
considered. This left only the impeller speed and the gas flow rate
as the two independent variables. Thus in this investigation, the
results of mass-transfer rates are reported in terms of the
impeller speed and the gas flow rate directly.
4.2 Normalization of Data
During different runs in a set, the liquid volume in the
contactor diminished gradually due to sample taking and therefore
the liquid height H varied from one run to another. Because the
values of interfacial area depend on the liquid height, before
attempting to compare and discuss the results of different runs,
this effect of differing liquid height had to be considered.
According to Westerterp et a1.. (1963), the relation between a' and
H/T is given by,
H (a*)H/T=i = a' — = a'i
57
This relationship was tested in order to bring the results to a
common basis of H / T = l . A procedure similar to the one used by
Westerterp et a1. ( l 963) was foHowed.
In each set, one run was selected to investigate the influence
of H/T. This was done by varying the liquid height while keeping a11
the other conditions (i.e. gas flow rate, rotational speed, and the
contactor configuration) constant, calculating the values of a' at
di f ferent H/T values, and plotting them against H/T on a log-log
paper. A linear regression line was drawn through the points and
the value of the slope, q, was determined. The individual values of
q varied between -0.7 and -1.2 with an average value of about - 1 .
Dependency of q on the gas flow rate, Qg, or the rotational speed,
n, could not be determined due to the limited data.
Taking q = - 1 , the values of 3 were then converted to a common
basis of H/T = 1. These values are denoted by â ' i , and are given by
H a'i = (a " )H /T= i = a' — •
Most of the data in l i terature are reported on this basis (See for
example, Midoux and Charpentier [ 1984]).
4.3 Validity of the Model and the Accuracv of Experimental Data
Before discussing the results, it is important to address the
issues of experimental data reproducibility and the validity of
di f ferent assumptions used in the calculation procedure described
in Chapter 2.
58
To check the reproducibility, six randomly-selected runs
were reproduced. Table 4.2 shows the results of these runs.
The dâta indicate that the values of â'i were reproducible within
115% accuracy.
Now the validity of the assumptions made in developing the
calculation procedure is examined. Recall here that following main
assumptions were made in the derivation 1n Chapter 2:
• Henry's law relates the interfacial concentration of gaseous
solute in the gas phase and the liquid phase.
• Resistance to mass-transfer on the gas side is negligible.
• Film theory applies.
• The gas in the liquid is fully mixed.
The maximum partial pressure of oxygen in the inlet air was
22.09 kPa (See Table B. 1 in Appendix B). For these conditions,
Henry's law can be used for predicting the equilibrium solubility of
oxygen, in water and electrolytic solutions, in this range of oxygen
partial pressure. (See for example, Hougen, Watson and Ragatz
[1970]).
The assumptions of negligible mass transfer resistance in the
gas side and the applicability of film theory are Justified by the fact
that the experimental results obtained by different investigators
using the chemical method same as the one used in this
investigation, were within ±20% accuracy of the results of
interfacial area obtained from physical methods. (See for example
Midoux and Charpentier [ 1 980].)
59
C\3
(\3 Q
CD
E í_ CD û . X
o
o TD O £_ Û . CD
cr CN
CD
JZl ( ^
c o
l^ <D Q
TJ
£ B \ "O CM Q
E t-C:L
- 1 ^ (^l ÛC
E - • ^ c (M .,—
E -E i_
-< i °
— cn (^ û . C í_ o X3 -o ( ^ CD -»-> CD O CL
cr co CD
- t—>
(^ cr ^ o O <D !— cn \
( E
No.
c =3 Û:
6 -z. • \ - >
<D CO
+
00
o
r ( j ^
CJ o
o (N
^ ' o
X
^ LO —
ro
.—
d -f
^
m .—
ro •.—
ro M
ro CN
^ ' o
X
CX) V£)
ro
I D
CNI
*£)
1 1
6 ho
in o ^ ro
ro ro ro
^
' o X
— 00
vû
CN
ro
00
•>r
l ^ CN
ro
00
^ 00 CN
hO ro ro CN
^ ' o
X
'sT IT)
^
in
^
q 00 1
CN
r 00
d CN
ro ro 00
'sT ' o
X
00 \D
ro
LT)
CN
CN -— 1
CN
CNÍ c:^ CN
in CN
ro ro
r
CN
^
' o
X
— 00
^
VD
vû
60
The assumption of fuHy mixed gas phase in the liquid for runs
wi th vertical baffle configuration can be Justified based on the
experimental results of Mehta and Sharma (1971) who used agitated
tanks of similar scale and geometry to the one used in this
investigation operated at similar conditions. They compared their
results wi th measurements of interfacial area by physical methods
and showed that the assumption of a fully mixed gas phase gives
close agreement (within i 15%) between the results from chemical
method and those of physical methods. Also, in this investigation,
visual observations indicated that there was a very high
recirculation of the gas within the contactor. This also suggests
that the gas phase was close to being fully mixed. The assumption
of fu l ly mixed gas phase for runs wi th horizontal baff le
configuration was tested from independent measurements of power
of agitation the details of which can be found in Section 4.5.
4.4 Comparison of Results for Horizontal and Vertical Baffles
4.4.1. Average Interfacial Area
Figure 4.1 shows the plots of the average interfacial area per
unit volume of liquid, a' i , vs. impeller speed , n, wi th, the injected
gas flow rate, Qg, as a parameter for vertical baffle configuration.
Two dist inct regions could be identified: One at impeller speed
lower than 13.3 rps. and another at impeller speed greater than
1 3.3 rps. The former (n < 13.3 rps.) is a region with l i t t le agitation
e f f ec t ; where the average inter facia l area increases w i th
increasing gas flow rate and increases very slightly with impeller
co <D
•nco- o C\J
û
o
( l fM I—
c
<D
co LLI _J Li . LL < m _ j < o h-DC LU >
o CD to
o o
o VJ
• ^ co — LO CO CO •r- CO <0
o o o II II II
C3) C3) 0»
o o o
o • ^
o CNJ
o o
- 4 - ^
n o<d CL
o <
- - LO
o 00
o <o
o •^
o C\J
( ^ CD í_ <
o
í_ <D
O czn 03 t_ <D > <
CD c_ Z3
cn
62
speed. The latter (n >13.3 rps.) is a region with strong agitation
effect; where the average interfacial area seems to be independent
of gas flow rate and increases linearly with increasing impeller
speed.
These observations are consistent with those reported by
Westerterp et_al.,(1963), Mehta and Sharma (1971), and others. It
is generally agreed that there is no effect of agitation on the
interfacial area below a certain critical speed, and therefore, there
is no sense in operating an agitated contactor below this critical
impeller speed.
No attempt was made in this investigation to determine the
exact crit ical speed, because in practice most agitators are
operated at much higher speeds in the region with strong agitation
effect. Therefore, the effect of gas flow rate, Qg, and impeHer
speed, n, in that region was examined in more detail for both
horizontal and vertical baff 1e configurations.
Figure 4.2 shows the plots of 3\ vs. n at different gas flow
rates for horizontal baffle configuration. Here too, two distinct
regions are identified with critical speed between 13.3 rps. and 20
rps.
Figure 4.3 shows the plots of a'i vs. Qg with n as a parameter
for vertical baffle configuration. It is seen from these plots that
there is l i tt le effect of Qg on a'i above the impeller speed of 13.3
rps. Unfortunately, there are too few data points to conduct a
meaningful statistical analysis. Figure 4.4 shows the plots of a'|
63
co
35
0
30
0
25
0
20
0 o LO
o o
• ^
E <M
E
o IT)
cn CD
CÛ
" (^ •i—i
c o
CD c_ <
o
o .._)
CD
( ^ L_ O >
<
Csi -^' CD í_ Z3 O^
Li_
64
co LLI _ l Li. LL < CÛ _J <
o l -cr LU >
<3
< C
<3
D O
n c
O
>
H »
H
Q . W W Ui ͱ CL Q. CL _ W "- — ^- f^ Q. . _
co tí- rv- co CO -< ^ co co j« co o • ^
• » - c j ^- f^ CO CNJ CsJ
II II II II II c c c c c ' H o a <
1 t 1
M
1
1
o o o
co o o o o
LO
o o o o
•^ o <—>
0.0
co o o o d
OJ o o o d
CT)
l ^
E cn
O
CD czn
Ave
ra
fles
C "^ o c C D ^
o: .y .t
o <i> [ Z > ( ^ w_
^ (^ w CD O (_ ^ < O — CD (^
3. E
fl e
rfa
c ^ c O) — (_ Z3 cr>
o o o
o o o o
o o CsJ
o o o
00 o co
o o cg
( l E
E
65
LU
< CQ _ J
o M d o X
•4
\
•4
•4 B
1
•
•
•
•
M
M ( / ) ( / ) (/) Q. CX CL Q. ^ »- —
o d r«* co ro co •- co co ^ co o —' co OD — C\J CsJ C4 II II II II II c c c c c
* M • • -4
1 1
1
H •
o o o
o o o o
m o o o
o o o (f)
E CD
o co o o o
CM o o o d
o o o
o o o o
o m
o o co
o
CsJ
o o CNJ
o in
o o o
CD cn C\3 l _ CD
C/) <D
> ^
^ " «*_
c ( O CÛ CD (\3 ( 3
Û : c o
^ - C ^
o c
c o C D < ^ "•— (\J O CD o < CD ^
«•— (\3 <«— — LLJ O
( ^
^ t: <D C l_ — cr>
C3| E
(M
E
66
vs. Qg for horizontal baffle configuration. Here too litt le effect of
Qg on â' 1 could be observed for impeller speed above 1 3.3 rps.
Based on the results above, it was assumed that for impeller
speed above the critical speed, a'i is independent of Qg. In that
case a linear regression line was drawn through all the data points
with n > 13.3 rps. for vertical baffle configuration. Figure 4.5
shows these points and the line obtained by linear regression. The
correlation of Westerterp et a1.. (1963) is also shown for
comparison. Figure 4.6 shows the data points with n > 13.3 rps. and
the f i t ted line for horizontal baffle configuration.
The lines for the two cases are compared in Figure 4.7. It is
seen that for an impeller speed in the range of 20 to 24.2 rps. (a
range of practical importance) the measured values of a.'i when
horizontal baffles were used were 80% to 130% higher than the
values when vertical baffles were used.
4.4.2 Gas Holdup
Figure 4.8 shows the plots of gas holdup, e, vs. impeller speed
at different gas flow rates for vertical baffle configuration.
Figure 4.9 shows these plots for horizontal baffle configuration. It
is seen by comparing Figs. 4.8 and 4.9 that the percent improvement
in the gas holdup for horizontal baffles over vertical baffles
corresponds very nearly to that observed in average interfacial
area, a' i . Also, a closer examination of the trend of e vs. n in each
case (vertical and horizontal) shows that each is very similar to the
trend of â'i vs. n. (Compare 4.1 with 4.8 and 4.2 with 4.9). These
67 7 7
in Cvj
c\j (\J
co CVJ
<>J CsJ
cn CL { _
c
o ra u— í_ CD
•t—t
cz cn
CD CD CD —
> CQ
v_ O <=> 3 c c o ^ •-: >
<D
c_ o (..J
LO M-CD c_ ZJ o^
ro CD c_ <
( l
68
Lf) Cvl
C\J
o C\J
<7)
( l E
E
ro
CO CNJ
cn C l t _
cvj C CM
C\J
o fO
«^. l _ <D
.<_» c= cn
<D CD — C D ^ ^ ra o cû > • : ; ;
^ j _ »
v_ C o o c - ^ o ^ — o ^ = 1 = -T - C_ O o í_ ^ ^ c_ O f^
(_J <D c_
vû < ^
CD c_ =3 Cf^
69
in CsJ
CsJ
<7)
i5 O cn
<D
<^ CJ
CM CsJ
»— t\J
o cg
cn CL
c_ C
<D >— - ^ (\J c CÛ
Ave
rage
or
izon
tal
s — . —
o T:?
c c
Com
pari
so
Ve
rtic
a a . c_
Fig
ure
4.7
Are
a fo
o ^ co
o <3) CsJ
O • ^
C\J
o <3) *"
o • ^
^"
o <3)
(^3|
co LU _ l LL Li_ < CQ _J < O I -G: LU >
co o
<n <
o n o
LO CsJ
-n:o- o Csj
o o to
o <D (/)
O o to
•^ ao ^ LO o co ^ co co o o o
II II II C3) O) CTî
o o o < o n
L.O
n oj co
o <
70
cn CD
( ^
CÛ
fO
o CD
>
Cl. Z)
o X co
iD c6
CD (_ Z3 czn
m
in o o
co o
c\j o o
c6
o o
c o
CJ o (^ í_
71
co i i j _ j u_ Ll_ < CÛ
O M
O
1 B
B
< B
1
o o o o o o ( / ) < / ) ( / >
TT co — LO <0 CO ^ co <o O <b C)
II II II O) (J> o>
o o o -4 • n
1
•
•
i—
•
B -
.
1
•
1
»
a>
Lrt CNJ
o C\J
LO
co
LT)
CsJ
<3
00 O
CD O
•<r o CJ
o o o
cn <D
(^ CÛ
-l—>
C
o INI
o zr. co (^
CJ
- ^ CD (_ z: CD
c o
Cc) O fO
72
indicators wil l be examined in detail in the section on analysis of
results.
4.4.3 Power of Agitation
Power of agitation, PQ, for runs without gas injection for both
horizontal baffle configuration and the vertical baffle configuration
is shown as commonly used curves of Power number, Np , vs.
Reynolds number, NR^ , in Fig. 4.10. It is seen that at the same
value of impeller speed, the same fluid and impeller, the Power
number when horizontal baffles are used was 25% to 31% lower
than that when vertical baffles are used.
The power of agitation when gas is injected from the sparger,
Pg, depends on the impeller speed and the flow rate of the injected
gas. It is common to present this dependency in the form of curves
of P3 vs. the Aeration number, N^. These curves for the vertical
baff le configuration and the horizontal baffle configuration are
shown in Figs. 4.1 1 to 4.13. In general it is seen that at low values
of gas flow rate (0.154 1/sec) and high values of impeller speed
(23.3 rps.) the Pg is 33% lower for horizontal baffle configuration
than that for vertical baffle configuration. At increasing values of
injected gas flow rate and decreasing values of rotational speed,
the d i f ference in P^ for ver t ica l and horizontal baf f le
configurations becomes smaller until at injected gas flow rate of
0.68 1/sec. and the impeller speed of 20 rps. the difference is
hardly 3%. It is obvious from the above discussion, that the
dependency of Pg on N ^ is not unique but differs from one impeller
73
C3) O CO
cd
• ^
o CM CO
m
LM T —
^ o in
00 co co • ^
00 <y>
co
co • ^
m co
co c\j co
co
•
•
•
•
• •
•
•
1
o
o
c o
1
o o
o o
>
1
• V
ort
ica
l b
aíl
los
o lio
rizo
nta
l D
alll
as
>
CVJ
in
LO
o LT)
o
ry> o
•^
00
• f
o 00
in o in co o
o o o
o
a. ^ZL cn o
in LO
o LO
•^
o J C
- cn ^ Sí c '^ .2 '^ (^ — ro < ^ ^ o s» - Kl o -(- o
o "ri c y O -UJ co c_ . _ o ca - ^ CD. { _
E o o **" LJ C
O
d> '^ — (T3 ' ^ O CD < c_ Z) cr>
74
20 r
ps.
II
c
1 <u
— ra
a ã
.y .§ o o > n: <3 -<
- < 1 —
...
• 1
<1
<
<
1
•4
1
<<
CO
o
r«-o o
o o o
LT)
__ (O O - ^ .<_> l_ CD >
^^ t- co o o . **— <_
< C \ |
z ^ CO (O -J <D
d < t - w_ —^ ^
o
o
CO o o
CM
<^ (\3 > CQ
c ^ l ^
o o 00 i:^ " ^ (— o o s: X
• o
—— c — ( ^
^ CD (_ 13 cr>
o cb
o o
o • ^
o co
o C\J
o o
o <:)
o GO
o
(^ Û .
75
co Q .
CM
II C
1
^ V
ertic
al
Baf
íles
^ H
oriz
onta
l B
affle
s
~ ~~ * j - i r-
<3
î ^
< Q
1
<1 -4
M
<
i
«*
1
1 1
00
o
r>. o o
o o
• o
i n o
__ (^ CJ
• * '™
.!_> { _
<D > ^ t - C2L O l_ «•_
r ^ < —
2 CN cn ^ 13 cn CO <D
o < ^.ZZ zz.
• ^
o C3
CO
o o'
Cvj o
> ' ( ^ (oûQ
CL — (T3
"-- ^ o C co o ..-. M 2. "C d. o
. X Csl - o — c ^ ca
<D t_ Z5 CT> "'
o (=>
o o
o co o o
Cû o in
o • ^
o <o
o C\J
o o o
CJ) o 00
(^ CL
76
o
o) _a)
"<5 CQ
15 o
>
_a)
C
!2 "c o
O
X
< <
zJ
to Q .
cr> co OJ 1! c
CO
o
LO
o
< ( - 7 -
• 'J '
o
co o
oj o
o o
(\J o
-^" 4 - » l _ CD
> t _
o * t
cn Q . t _
ro < f - ^
ZL cn => CO c_ <1) >
(VI CL (*_
o co
- 1 — '
o Q.
ro -
^
<D t _ 13 CD
CNI • —
co <D
^ ^ « • —
(^ CÛ ^—. (^
J_»
c o rsj t _
o X • o
c (^
- < I — I - o o
o Csj C\J
o o CVJ
o 00
o co
o •«í- o
cg
o o o
00
(\3
77
speed to another. The way to bring these curves together is to
somehow incorporate the gas holdup, e, in the relation. This is
commonly done by plotting P^/{]-e) vs. the Aeration number, N^.
However, it was found that when P^/e was plotted vs. N^ the
spread of the data points was smaller.
The physical significance of Pa/(l-£) is that it is the power
input per unit volume of the liquid; whereas, the physical
significance of P^/e is that it is the ratio of power input per unit
volume of liquid to the volume of the gas in the contactor per unit
volume of liquid.
P^/e = 'Pa ] / (e ^ L 1 - £ . [\-e}
Thus the power of agitation in tank with injection of gas is
represented as curves of Pg/e vs. Aeration Number (Qg/nD^).
Figure 4.1 4 shows the plots of Pg/e vs. N/ and their corresponding
linear regression f i ts for tanks vertical and horizontal baffles. It
is seen from figure 4.14 that the power for an agitated contactor
with horizontal baffles to have the same gas holdup as in the case
of vertical baffles is 50 to 57% less than that required for vertical
baffles at a given Aeration number.
4.5 AnaJvsis of Results
It has been indicated above that when horizontal baffles were
used the average interfacial area, â'i / was 80% to 130% higher
than when vertical baffles were used. Also, it was found that the
power of agitation for runs without gas injection was 25-31% lower
78
<
CL ZD
TD
O X cn ra tD O
4—»
C O
.t—t
ra t _ o <
Ll • ! _ >
• — ^
c O
•t—t
(TJ
— LJ^ <
<•— O
(_ o ^ o
CL
^
^
CD t _ Z3 cn
cn CD
V — 1..
(YJ CÛ
(YJ 4 _ >
c o rsi
t _ o X TD c ro
— (^ ( )
*—> t_ o
> t _
o **—
o 4 _ *
Û :
( 3 CJ
79
for horizontal baffle configuration than that for vertical baffle
configuration. It was also seen that for runs with gas injection,
the power of agitation per unit gas holdup, Pg/e, was 50 to 57%
lower for horizontal baffle configuration than that with vertical
baffles. In this section an attempt is made to identify the causes
of these results.
It was indicated in section 4.4.2. that the gas holdup, e, in the
horizontal baffle configuration and that in the tank with vertical
baffles change with rotational speed in a manner similar to the
variation of 3, in the respective cases (see Fig. 4.1 and 4.8 and 4.2
and 4.9). This observation was scrutinized further by the following
analysis:
Since â'i snd e are measured independently, then from these
measurements a mean gas bubble diameter can be deduced from the
following relation.
_ 5 e d =
a' 1 1 -e
Fig. 4.15 shows plots of d vs. n for vertical baf f 1e configuration. 11
is seen that for n > 13.3 rps., the mean bubble diameter remains
essentially constant. A linear regression line was drawn through
a11 the points for n > 13.3 rps. and the slope of this line was found.
to be close to zero. This suggests that the mean bubble diameter,
d, is independent of n. The average value of bubble diameter for
vertical baffle configuration was found to be 2.53 x 10"^ m.
Similar analysis for horizontal baffle configuration indicated that
80
< cc
ID
U-
O o LU —1 LL LL < C _J <
o I -DC LU >
cc o LL DC LLI H LU < Q LU _ l CQ CQ Z) CÛ
ol
I o
in C\J
o n
o C>J
n<3 o
o o o 0) o o W) (/) (/> ^ ia : : Í
• ^ c o • ' -in co 00 T - CO CD
O (6 O II II II
O ) C7) C7)
o o o
in ^ co
CIL
LO
o co
tn Cvj
o CNÍ
in LO
Q
—r CD o c>
cn <D
Cû
"(6 o
<D >
CD 4 _ >
CD
E Q
.Cl
o Z3
CÛ
c CD
21 LO
CD t _ Z3 CZT)
I
o
• o
81
the bubble diameter remained constant in this case too. The
average value of bubble diameter was f ound to be 2.51x10"^.
From this analysis, it can be concluded that the mean bubble
diameter is the same for vertical and horizontal baffles. This
implies that the increase in average interfacial area observed when
horizontal baf f les were used was probably due to a higher number of
gas bubbles in the dispersion and not due to smaller gas bubbles.
Next the plausible causes of a higher number of bubbles in the
dispersion and the lower agitator power requirement when
horizontal baffles are used are examined. It appears that the main
reason is surface aeration when horizontal baffles were used. In
the runs with horizontal baffles when no gas was sparged it was
observed visually that gas bubbles penetrated into the liquid from
the liquid surface and circulated in the liquid. This is called surface
aeration. No surface aeration was observed in runs with vertical
baffles. Below an attempt is made to analyze this phenomenon in
some detail.
The rates of gas recirculation in the tank are related to the
power of agitation as discussed in Section 1.2.3. Recall that
Roustan (1978) developed a technique for getting a unique power
reduction curve for a given impeller by plotting the ratio of power in
presence of gas to power in absence of gas, Pa/Po, vs. the effective
Aeration Number, H^^. Values of N^^ were calculated by Eqns. 1.2
and 1 -4. The curve developed in this manner is unique for a given
impeller and does not depend on the tank geometry or the baffle
arrangement. Figure 4.16 shows a plot of Pg/Po vs. N/^^ for
82
o co
in CVI
o
o CvJ C3
* <n
cn CD » «•— w í_ (^ <D Û JD
-;;; ^ (^ ZJ
- z t ^ CD O
> 4 - .
o ^ t" < í_ CD CD ^ >
< CL CJ "• 2 c . ^ o
o
0.1
in o
. _ * M —
C UJ .2 V-4-» O
•D .2 < D -4—» Û: O
c í£ ^ " >
- í^ <1J (^ l_ ZJ cn • — LJ_
o o
ra o CL
83
vertical baffles and a third order polynomial regression curve
through the points.
Consider the scheme of gas flow rate in the contactor shown in
the diagram on below:
Here Q^ denotes the surface aeration rate, Q^ the internal gas
recirculation rate (i.e gas recircualting within the liquid), Qc' is the
total gas recirculation rate (Q^ + Q^). The total gas recirculation
rate, Q^', in absence of gas injection can be estimated from
experimentally measured quantities (Power of agitation and Gas
Holdup) as follows:
Q E QE+Qg ^
?# <b
Q: i wT"wwtmw
P tttWWTWfH IHiMM oo
fcllMINIIMIMIMIMIIIIIIimilllMMIIII ^ s V s s • • .w.y.;:
^^táÊÊÊiitÊtmHtiÊiáiiiíiliii
g
To estimate the rates of total gas recirculation, QQ', in the
absence of gas injection, it was assumed that the decrease in the
Power number with increasing Reynolds number observed for
horizontal baffles in Fig. 4.10, was solely due to surface aeration.
If there had been no surface aeration, the Power number for
84
horizontal baffles would have remained constant at a value of 5.5,
which is the extrapolated value from the Np -Np^ curve for
horizontal baffle configuration in Fig. 4.10. Using this value of
Power number the value of an hypothetical power requirement, PQ',
without surface aeration was calculated for each run with
horizontal baffle configuration. The power reduction PQ/PO' was
then obtained for each run using the value of PQ' calculated above
and the observed values of the power of agitation without gas
injection, PQ.
Now as discussed above, the curve of P /Po vs. N^^ shown Fig.
4.16 does not depend on the baffle configuration. Therefore, it is
assumed that this curve can also be used for operation with
horizontal baffle configuration. tJsing the curve in Fig. 4.16 the
effective Aeration Number, N^^, required to obtain a given power
reduction can be obtained. Now noting that,
N A " = nD
then, when Qg = 0, which is the case here,
Qc' = ^A^ ^^^'^
Figure 4.1 7 shows the values of OQ' thus obtained as a function
of impeller speed. It is seen from fig. 4.17 that the total rates of
gas recirculation are 2 to 15 times higher than the rates of injected
gas flow rates, Qg, which were in the common range (0.1 to 0.7
1/sec.) of operating conditions for the range of impeller speeds
used. This observation supports the assumption of fully mixed gas
phase for horizontal baff 1e configuration.
85 co C\í
CNJ C\i
o CM
CO
CO
•«r
cn CL í_
c
o
he C
onta
B
affle
s
4—»
latio
n in
or
izon
tal
=> T
Rec
irc
on f
or
cn — (\3 - ^
•o '?
cr ^ . =3
!^S CD
CD
^ c
, O
O
86
The rates of gas recirculation, QQ', thus obtained are related
to surface aeration rates, Q^, as follows: The recirculation of the
gas in an agitated contactor can occur either within the contactor,
QQ, or the exit gas may be reincorporated by surface aeration, Q ,
as shown in the diagram on the page before, which means that Q ' =
Qc - QE-
No method was available in this investigation to measure either
QQ or Q^ individually. Therefore the value of Q^ cannot be calculated
directly. However, in the extreme case when Qc = 0, 0£ = QQ. This
of course represents the highest limit on the value of Q .
CHAPTER 5
CQNCLU5I0NS AND RECOMMENDATIONS
5.1 Condusions
The ef fect of horizontal baff les on the performance of
mechanically agitated contactor was tested. The results indicated
that under the range of conditions investigated, the horizontal
baffles proved to be better compared to the conventional vertical
baffles in terms of lesser power requirement and higher values of
average interfacial area per unit liquid volume. The following list
summarises the conclusions:
1. Under the range of conditions investigated, the values of
average interfacial area are 80-130% higher when horizontal
baffles were used than when vertical baffles were used.
2. Power of agitation without gas sparging was 20-30% less
for operation wi th horizontal baffles than that wi th vert ical
baffles.
3. Power of agitation to gas holdup ratio for agitation in
presence of gas sparging was 50-57% less for operation wi th
horizontal baffles than that with vertical baffles. This implies
that the power of agitation required for a given value of gas holdup
at agiven impeller speed and gas flow rate, is 50-57% less for
operation with horizontal baffles than that with vertical baffles.
4. It is shown that the increase in average interfacial area
observed for horizontal baffle configuration over that with vertical
87
88
baffle configuration is due to higher gas holdup in the system rather
than smaller size of bubbles.
5. Visual observations indicate that in operation wi th
horizontal baffle configuration, the gas penetrates into the liquid
from its surface. This is also called surface aeration.
6. A methodology to make rough estimates of the rates of
total gas recirculation, QQ', within the liquid, based on the power
reduction curve f or the given impeller was proposed.
5.2 Recommendations
Based on the the results and experience of this investigation,
the foHowing recommendations are made:
1. Additional experiments should be carried out over a wider
range of impeller speeds and gas flow rates than is used in this
investigation. It is particularly important to carry out experiments
at higher impeller speeds because in industrial practice values of
impeller speeds as high as 2000 rpm. are used. This will require a
drive with higher power than the one used in this investigation.
2. Effect of fluid properties such as viscosity, density and
surface tension should be studied to get more insight into the
results of this investigation.
3. Experiments should be carried out w i th larger size
contactors to develop scale-up relations.
4. Surface aeration was observed in the operation of the
contactor with horizontal baffle configuration and was suspected to
be the main cause of the increased average interfacial area and
89
lower power requirement the operation with vertical baffles. In
this investigation, means were not available to measure the rates
of surface aeration directly. It is strongly recommended that the
rates of surface aeration be measured independently. A possible
technique to measure rates of surface aeration has been suggested
by Matsumara et a1.. (1977).
5. Different configurations of horizontal baffles should be
examined for further improvement in the contactor performance.
These may include:
• Number of baffles.
• Location of baffles relative to the liquid surface.
• Pitch of the baffles.
• A combination of horizontal and vertical baffles.
6. The most promising potential application of horizontal
baffles appears to be in Fermentation processes. Therefore, it is
suggested that the performance of an agitated contactor with
horizontal baffles be tested for fermentation broths or simulated
fermentation broths. A possible system that simulates the
rheological behavior of fermentation broths is a paper pulp
suspension as suggested by Loucaides and McManamey (1973).
LITERATURE CITED
Brui j in, V/., van"t Reit, K.,and Smith, J.M., Trans. Instn. Chem.
Enanrs.. (London). 52, 99 (1974).
Calderbank, P. H.. Trans. Instn. Chem. Engnrs.. (London). 36, 443
(1958).
Calderbank, P. H., Trans. Instn. Chem. Engnrs. (London). 37. 173
(1959).
Cooper, R.D., and Wolf, D., Canad. J. Chem. Ena.. 94, 46 (1968).
Danckwerts, P. V., and Sharma, M. M.. Br. Chem. Enanr.. J ^ , 522
(1970).
Danckwerts, P. V., and Sharma, M. M., Chem. Enana. Sci.. 3 1 , 767
(1976).
Elseworth, R., Vv'illiams, V., and Harris-Smith, R., J. ADDI. Chem.. 7,
261 (1957).
Hassan, I. T. M., and Robinson, C. W., A.I.ChE.J.. 23, 48 (1 977).
Hiegbie, R., Trans. A.I.ChE. 12,365 (1935).
Holland, F. A., and Chapman, F.S., "Liquid Mixing and Processing,"
Reinhold, NewYork, (1966).
Hughmark, fí Ind. Enann Chem.. Proc. Des. Dev.. i 9 , 638 (1980).
Hyman, D., and Van Den Bogaerde, J. M., Ind. Engng. Chem., 52, 757
(1960).
Joshi, J.B., Pandit, A.B., and Sharma, M.M., Chem. Engng. Sci., 27
(6),813(1980)
Lewis, W.K., andWhitman, W., Ind. Engng. Chem., 16, 1215(1924).
Loung, H. T. and Volesky, B., AJ.ChE.J., 25, 893 (1979).
90
91
Loucaides, R., and McManamey, W. J . , Chem. Enana. Sci.. 28, 21 65
(1973).
Mann, U., and Rubinowitch, M., Ind. Ena. Chem.. Proc. Des. Dev.. 22.
545 (1983).
Matsumara, M., Masunga, H., and Kobayashi, J . , J. Ferm. Tech.. 55,
388 (1977).
Mehta, V. D., and Sharma, M. M., Chem. Engng. Sci.. 26. 461 (1 971).
Mehta, V. D., Ph.D.(Tech.) Thesis, Univ. Bombay, (1970).
Michael, B. J . , and Miller, S. A., A.I.ChE. J.. 8, 262 (1962).
Midoux, N., and Charpentier, J. C. Int. Chem. Enana.. 24 (2). 249
(1984).
Nagata, S., "Mixing", Wiley, New York (1966).
Nagata, S., Yamamoto, K., Hashimoto, K., Naruse, Y., Chem.Ena.
(Jaoan). 23. 565(1959).
Oldshue, J . Y., "Fluid Mixing Technology", McGraw-Hill, NewYork
(1966).
Preen, B. V., Ph.D. Thesis, University of Durham, South Afr ica
(1961).
Roustan, M., Thesis, Univ. de Toulouse, France (1978).
Rushton, J . H., Costich, E. W., and Everette, H. J . , Chem. Engng.
Proar.. 46. 395 (1950 a).
Rushton, J . H., Costich, E. W., and Everette, H. J. . Chem. Enana.
Progr.. 4^ . 467 (1950 b).
Scot t , W.W., and Furman, N.H., "Analytical Chemistry", D.Van
Nostrand Corp, New York, (1965).
92
Snijder, R.J., Hagerty, P.F., and Molstad, M.C., Ind. Enana. Chem..
49, 689(1957)
Uhl, V. W., and Gray, J. B., "Mixing, Theory and Practice", Academic
Press, New York (1966).
Van't Reit, K., Boom, J. M., and Smith, J. M.. Trans. instn. Chem.
Enars.(London). 54, 124(1976).
Van't Reit, K., and Smith, J. M., Chem. Enana. Sci.. 28, 1 031 (1973).
Van't Reit, K., and Smith, J. M., Chem. Engng. Sci.. 28, 1 093 (1974).
Westerterp, K.R., Van Dierendonck, L.L., and De Kraa, J.A., Chem.
Enana. Sci.. 18. 157(1963).
94
The procedure for calculation of mass-transfer coefficient
from experimental observations is shown below. The calculations
are carried out for Run Number 5 of Set 1 (Appendix B). For
convinience the contactor configuration, the experimental
conditions and the pertinent observations for this particular run are
also shown below.
Contactor Confiauration
• Impeller Diameter, D (m) = 0.0762.
• Tank Diameter, T(m) = 0.305.
• Impeller Height off the Bottom of the Tank, Hi (m) = 0.15.
Liquid Height, H (m) = 0.45. •
Experimental Conditons
• Gas Flow Rate at inlet, Og^p^, (mVs) = 0.000154.
• Rotational Speed, n (rps) = 23.33.
• Ambient Temperature, T (°K) = 295.522.
Atmospheric Pressure, Pgi-pn (kPa) = 100.88. •
• Moisture Content of ambient air, X^ (kmole water/kmole dry air)
= 0.0064.
Observations
• Dispersion Height, H^ (m) = 0.47.
• Sample Analysis
The samples withdrawn were anlysed according to the
procedure described under Section 3.2. Thiosulfate titration reading
(TT) varied with time interval according to the table below.
95
Time, t , (min) Thiosulfate Titration reading, TT (m1)
5 16.4
10 22.3
15 28.1
20 34.0
A plot of TT vs. t was prepared and the slope of the best
s t ra ight line through the points was determined by linear
regression.
• Slope of theTT vs. t Plot, p, = 1.174.
Calculations
Sten A: Calculation of Inlet
/Mr romnosition
Total pressure at inlet, P (kPa), 0.45 X 1.096x10'x 9.8
= Patm ^ Hpg/1000 = 100.88 — •
P = 105.71 kPa.
Molar gas flow rate at inlet, nøin' (kmol/sec),
p. Q^. 105.71 X 0.000154. ^ ^ ^in^Gin ^ = 6.626 X 1 0 - ^
RT 8.314 X 295.522
nøir^ = 6.626 x 1 0 " ' kmoles/sec.
Molar Flow Rate of Dry Air at Inlet , kmole/sec
nrin 6.626 X 1 0 " ' ^ ^^ , Giri ^ 6.584 X 1 0 - ^
(1 ^ X ^ ) (1 + 0.0064)
96
Knowing the composition of dry air (21% Oxygen and 79% Nitrogen),
the composition of inlet air was calculated and is shown below.
Component
Qxygen
Nitrogen
Water
Molar flow rate
kmole/sec
1.3826 X 10-6
5.2014X 10-6
4.2 X 10-8
Steo B: Calculation of Liauid Volume and Hold-HJU
Mole Fraction
0.2087
0.7850
0.0063
Partial Pressure kPa
22.06
82.98
0.67
Liquid Volume, V|_ (m^)
= T T / 4 X ( T O ) ' X H = 0.25 XTTX (0.305) 2 = 0.033
Vi_ = 0.033 m^
Hold-up of Gas, e ( void fraction)
H = ] - _ = 0.053.
HA
e = 0.053.
StPn C: Ca1riil?>tionof Rate of
frhsorntion of Oxvaen
According to the Analytical Procedure shown in Section 3.2, the
Rate of Reaction of Sulfite was found to be equal to;
—^ ( mmole/l i t /hr) dt
= -150 X (Slope of TT (m1) vs. t (min) plot) = -150 x p = - 1 76.1
97
According to Equation 2.26,
Rate of absorption of oxygen, i-r^)^ (mmole/dit.hr)),
1 dCso Vi = - — -— = 88.05 = 2.45 X 1 0 - ^ kmo1e/(m^sec),
2 dt VR
^"•^A^d "" 2.45 X 10-^kmo1e/(m^sec).
Step D: Calculation of Outlet Flow Rate and Average Concentration
Average gas concentration at the outlet is calculated by trial
and error procedure discussed in Section 2.1.1. The calculation
steps are as follows:
(i) Outlet flow rate Qoout ^as guessed as equal to the inlet flow
rate, QQ,^^^
(ii) [RT J
was calculated using Equation 2.29, as follows; av
^RT Jav QGoutVRTj in
(-r^A^dVR
Q Gout
22 06 2.45 X 1 0 ' X 0.033 - = 0.00373.
8.31 4 X 295.522 0.000154
(ii i) Composition of air at the outlet was calculated as follows;
Partial pressure of oxygen at outlet, (Po^) av, kPa
98
/ Po. PO2 " X R XT = 0 .00373x8.314x295.522 = 9.1645
RT J av
Total pressure at the outlet, (kPa),
Atmospheric pressure, Pgj-pp = 100.88.
Mole fraction of oxygen at the outlet,
(Po2)av = 0.09084.
atm
It was assumed that the air left the contactor saturated with
water vapor; therefore, the partial pressure of water vapor at
outlet was taken as vapor pressure of water at the constnat gas
temperature of 295.522 °K.
Partial pressure of water vapor at outlet, (kPa)
= 2.68.
Mole fraction of water vapor at outlet,
= 0.0266.
Mole fraction of inert was therefore, by difference,
= 1-0.09084-0.0266 = 0.8754.
(iv) The total molar flow rate at the outlet was found by inert
balnce; since, the number of moles of inert remained constant.
Molar flow rate at outlet, nøout' (knnoles/sec),
Molar FI0W Rate of inert at inlet 5.2014 x 1 0 - ' = = 5.942 X 1 0 - ^
Mole Fraction of inert at outlet 0.8754
99
Volumetric flow rate at the outlet, Qoout' (i^Vsec),
•^Gout X R X T 5.942 x 1 0 " ' x 8.3 1 4 x 295.522
atm 100.88 = 0.000145,
(v) Steps (ii) to (iv) were repeated with the new value of Qgout
until constant value of average concentration at the outlet was
obtained. The results of the tr ial are shown below:
Trial f P 0 2 ^
,R T J av
kmole/m^
Mole Fraction 0 Q Gout
(mVsec)
1
2
3
0.00373
0.00396
0.00397
0.09084
0.0964
0.0967
The average concentration of oxygen is therefore,
( ^ l = 0.00397 kmoles/m^ av
0.000146
0.000144
0.000144
Sten F: Calculation of Avearae
Interfacial area
According to Equation 2.24, the quantity a' is given as.
He (-rA)d
ki (Po/RT)av = 93.07 mVm^
10
• o
c c (D c o
4-» • ^ M
"O c o u <D
c "
^-"
Û
<l; I D (^
l ^ 4--> <D C/) o
•i->
(N 4 - > <1) co
o **— (^
4-.» (^ Q L-<1> CT
4-» ( ^
r
KD
l O
^
ro
CN
Set
Num
ber
296.6
29
7.7
29
6.1
29
6.6
294.
4 29
5.5
Tem
pera
ture
, T
°K
no
i
100.
7 10
0.6
101
100.
1 10
0.9
Atm
osph
eric
P
ress
ure,
P,
kPa
0.0
068
0.00
71
0.0
056
0.
0077
0.
0075
0.
0064
M
oist
ure
Con
tent
, ,m
ol
of w
ate
r/m
ol
dry
aír
105.
9 10
5.6
105.
4 10
5.8
105.
9 10
5.7
.t—>
cu c
J—>
CL> ro
(T) C
CL -
0.0
00
68
1 0.0
00368
0.0
00
15
4 0
.00
06
81
0.00
0368
0
.00
01
54
Gas
ílo
w r
ate
at
inle
t ,
Qg,
mV
s
22.0
9 2
2.0
1 22
.05
22.0
5 22
.08
22.0
6
4_>
03 <U
^ *-> ^ u' (^;^ o Q-<_ C Jxí
Q L CU CT> -
(\3 > ^ C
t::°o~
ro 'o
X
CM r-ro
ro ' o
X
CN cr> ro
ro •o
X
ro vO ro
ro 1 o
X
00 vû ro
ro 1 O
X
ro ro
ro ' o
X
00
ro
na
ss t
ran
sfe
r co
eff
icie
nt,
k ,
, m
/s
'
62
.23
65.3
2 6
4.4
3 6
3.2
3 5
9.0
8 6
0.8
3 H
enry
's l
av;
con
sta
nt,
He
02
•D O r~
4 - *
^
c o
4 - >
4 - ^
O^ < V— O c_ <D :^
o (^.
(N •
CÛ
<D
r^ 03
\ -
(f) Oí <•—
CD __
ica
4 - > c (1)
> c_ o H—
c o
4—» (J (D "•"" c cn Cs5 iD
o Q.
t_" <u
E
t_ o
o CL
m 4_> 4_) (\3
o CL
C_" o o û .
o cr
2 t_" o C!
E 2 co
"ô c >> o
n rps
5.92
9.65
41000
8.33
6.08
40.58
65000
13.33
5.92
133.35
98000
o CM
5.95
151.48
102000
20.83
6.05
173.28
106000
21.67
6.13
196.61
1 10000
22.5
5.89
210.67
114000
23.33
5.86
232.89
118000
24.17
10
(T) CO iD 4—>
13 o r 4—'
^
c o
4-> •— o> < V -
o c_ <D ^ o a.
ro
cn OJ —
«•— < 4 —
CÛ , c
4—»
o N (_
o X í_ o « « - • c o
4->
í ) (D
CÛ £
<D
o CL
2 L." O
E 2 L. <I) ^ O
CL
<n 4-> 4_>
ra
o Q.
t_ O
o Q.
O
<_'
o
E
co "C o c >-o c:
n rp
s
5.5
8.
95
41
00
0 8.
33
5.25
35
.01
65
00
0
13.3
3
4.5
10
1.4
00086
o CM
4.47
1 1
3.75
10
2000
2
0.8
3
4.37
12
5.1
1060
00
21
.67
4.27
13
6.95
00
00
11
22.5
4.2
150.
2 1
1 40
00
2
3.3
3
4.17
16
5.72
1
1800
0 2
4.1
7
Table B.4 Observations and Calculations for Set 2
104
Set Number:2 Baf fle Type: Vertlcal Gas Flow Rate: 0.000154 m
Run No
Rotational speed, n.rps
Rate of oxygen uptake, - r ^
A -r kmol/mVs x 10~*
Partial Pressure of oxygen ín outlet gas, P02 out, kPa
Gas Flow rate out, -r Qg out, mVs x 10
^ l ^ . ^ 3 { \ - e )He '' ' ^
( l - Ê )
H/T
Interracial area for H/T=l
1
3] , mVm'
Gas holdup, £
Power of Agitaticn.Pg Watts.
Power reauction, f^/ PQ
Aeration number,
M X lO^
Reynolds number, x 10
Npe
1
8.33
0.002
18.74
0.161
0.342
5.99
1.5
8.93
0.0037
8.5
0.88
42
0.408
2
13.33
0.006
17.66
0.159
0.860
15.04
1.46
21.96
0.0091
32.37
0.81
26
0.653
3
20
0.019
12.52
0.149
3.96
69.42
1.^4
99.97
0.04
106.63
0.8
17
1.273
4
21.67
0.022
1 1.95
0.148
4.55
79.37
1.43
1 13.5
0.045
135.16
0.78
16
1.379
5
23.33
0.025
9.75
0.144
6.15
93.07
I.4I
151.6
0.053
153.79
0.73
15
I 485
Vs
6
22.5
0.023
1 1.55
0.147
4.54
79.33
1.4
1 1 1.07
0.044
-
-
15.5
I 432
Table B.5 Observations and Calculations for Set 3
105
SetNumber:3 Baffle Type: Vertical Gas Flow Rate: 0.000368 mVs
Run No
Rotatlonal speed, n.rps
Rate of oxygen uptake, - r ,
kmol/mVs x 10-^
Partial Pressure of oxygen in outlet gas, P02 out, kPa
Gas FIow rate out, -r Qgout .mVs x 10
\ã 3 (1 - e )he
a
(1-e)
H/T
Interfacial area for H/T=l
1
3 1 , mVm^
Gas holdup. £
Power of Agitatíon.Pg Watts.
Power reductlon, p / p„ a °
Aeration number,
N X 10"^ A
Reynolds number, x I0~^ NRe 1
1
8.33
0.005
18.91
0.388
0.66
1 1.81
1.5
17.71
0.0069
6.76
0.7
10
0.408
2
13.33
0.007
18.45
0.386
0.97
17.29
1 46
25.25
0.0095
26.33
0.65
62
0.653
3
20
0.024
16.07
0.375
3.76
67.31
1 44
96.93
0.039
92.01
0.69
42
1.273
4
21.67
0.027
15.64
0.373
4.25
87.74
1.43
121.18
0.043
109.17
0.63
39
1.379
5
23.33
0.030
15.31
0.371
4.31
93.26
1.41
131.5
0.052
126.4
0.6
36
1 485
6
20.83
0.032
14.96
0.370
5.26
77.23
1.4
103.2
0.043
-
-
40
1.326
Table B.6 Observations and Calculations for Set 4
06 6
SetNumber:4 Baff le Type: Vertlcal Gas Flow Rate: 0.000681 mVs
Run No
Rotational speed, n,rps
Rate of oxygen uptake, - r .
A -r kmo l /mVs x 10 ^
Part ial Pressure of oxygen in out let gas, P02 out , kPa
Gas Flovy rate out, -r Q g o u t , m V s x 10
lá 3 V 1 n
(1 - £ )he
a
( l - £ )
H/T
Inter facia l area for H / T = l a i , m^/m'
Gas holdup, £
Power of Agi ta t ícn,^^ Wat ts .
Power reduct ion, p^/ p„
Aerat ion number,
N X lO^ A
Reynolds number, x 10~^
NRe
1
8.33
0.007
19.39
0.710
1.02
17 45
1.5
26.18
0.01 1
5.69
0.59
13.5
0.403
2
13.33
0.010
19.21
0.709
1.36
23.32
1.46
34.05
0.012
22.32
0.55
1 1.5
0.653
3
20
0.029
17.73
0.696
4.13
71.01
1.44
102.26
0.041
72
0.54
3
1.273
4
21.67
0.032
17.52
0.694
4.56
73.2
1 4Z
1 1 1.33
0.045
90.1 1
0.52
7
1.379
5
23.33
0.037
17.07
0.690
5.46
93.76
1.41
132.2
0.052
101.12
0.43
6.6
1.485
6
24.17
0.039
17.01
0.639
5.63
97.64
1 4
136.7
0.054
-
-
6.4
1.53
Table B.7 Observations and Calculations for Set 5
SetNumbenS Baf fle Type: Horlzontal gas Flow Rate: 0.000154 mVs
Run No
Rotatlonal speed, n,rps
Rate of oxygen uptake, - r .
A -r kmol/mVs x 10 ^
Partial Pressure of oxygen in outlet gas, P02 out, kPa
Gas FIow rate out, -, Qgout ,mVs x 10
lá 3 (1 - £ )H? "" ' ^
( l - £ )
H/T
Interfacial area rcr H/T=I
1
a| , mVm^
Gas holdup. £
Power cf Agitaticn^Pg Watts.
Power reducticn, P^/ p. a -'
Aeration number,
N X 10- A
Reynolds number, x I0~^ NRe
1
8.33
0.003
19.26
0.163
0.398
7.03
1.5
10.62
0.0044
3.51
0.95
42
0.408
2
13.33
0.01
16.37
0.157
1.536
27.23
1.46
39.33
0.016
32.56
0.93
26
0.653
3
20
0.025
9.26
0.145
6.341
121.69
1 44
175.23
0.063
92.33
0.916
17
1.273
4
21.67
0.029
7.27
0.141
10.01
177.15
1 43
253.33
0.091
93.2
0.735
16
1.379
5
23.33
0.031
6.63
0.140
I 1.33
201 99
1 41
234.8
0.107
102.74
0.684
15
1.435
6
22.5
0.030
6.77
0.141
1 1.07
196.57
1.4
275.2
O.I
-
-
15.5
1.432
Table B.8 Observations and Calculations for Set 6
108
Set Number:6 Baf f le Type: Horizontal gas Flow Rate: 0.000368 mVs
Run No
Rotational speed, n.rps 8.33 3.33 20 21.67 23.33 20.83
Rate of oxygen uptake, - r ^
A -, kmol/mVs x 10~^
0.006 0.013 0.038 0.047 0.052 0.048
Partial Pressure of oxyaen in outlet aas, P02 out, kPa
Gas Flow rate out, 3 Qgout ,mVs x 10
kJJ (1 - £ )He
X 10
( l - £ )
H/T
Interracial area for H/T= l ai , mVm^
Gas holdup. £
Power of Agitaticn,^2 Watts.
Power reaucmcn, 9 / p
Aeration number,
N. X 'O^ A
9.49 18.33 13.93 12.03 1.08 1 1.93
0.395 0.339 0.369
0.334
.91
1.5
20.37
.793 6.762
29.33
I ^ô
43.63
0.C037
3.15
Reynolds number, x 10
NRe
r5
0.9
0
0.403
2.92
.44
0.362
9.747
161.97
.43
0.357
1 1.753
95.96
0.360
10.066
17971
62.6
0.013 0.063
~j 0.3
0.33
62
37.75
231.61
41
276.3
0.033
90.47
0.365
42
0.653 273
0.103
1 4
251.6
0.095
95.68
0.723 0.637
39
.379
36 40
1.435 1.326
Table B.9 Observations and Calculations for Set 7
109
Set Number: 7 Baf f le Type: Horizontal
Run No
Rotational speed, n,rps
Rate of oxygen uptake, - r ,
A -r kmo l /mVs x lO- '
Part ial Pressure of oxvgen in out let gas, P02 out , kPa
Gas Flow rate out, -r Qg cut , mVs x 10
k 1 a 3 Y 1 n
(1 - £ )^* '' ' ^
a
( l - £ )
H/T
Inter facia l area ror H / T = l
1
a i , mVm^
Gas holdup. £
Power of Agitation^Pg Wat ts .
Power reGuction, f^/ p^
Aerat ion number,
N X lO^ A
Reynolds number, x 10~^
NRe
1
8.33
0.008
19.33
0.721
I.OI
13.87
1.5
28.31
0.012
7.1
0.79
13.5
0.403
2
13.33
0.016
19.17
0.714
2.16
37.12
1.46
54.2
0.022
26.6
0.76
1 1.5
0.653
gas FIow Rate: 0.00068
3
20
0.022
18.86
0.71 1
2.67
46.04
1.44
66.3
0.065
73.92
0.681
3
1.273
4
21.67
0.061
14.95
0.673
9.99
171.24
1.43
244.33
0.102
85.19
0.73
7
1.379
5
23.33
0.073
13.39
0.669
12.61
216.52
1.41
305.3
0.127
92.37
0.613
6.6
1 435
mVs
6
24.17
0.075
13.44
0.666
13.32
237.5
1.4
332.5
0.133
-
—
6.4
1.53
11
In Section 2.3, we developed the relations for determining the
torque from measurements of strain along the principal axes of the
impeller shaft. The strain measurements are carried out using a
device called a Resistance Strain Gauge. A brief discussion about
the construction and the use of this gauge is given below. For a
detailed discussion please consult the "Pressure and Strain
Measurement Handbook and Encyclopedia" published by Omega
Engineering, Inc. This reference manual is available free .of charge
from Omega Engineering.
As mentioned cursorily in Section 2.3, Resistance Strain Gages
operate on the principal that electrical resistance of a thin metallic
wire varies in proportion to the strain applied to it. These devices
are further classified into different catogories depending on the
material of the resistance. These catagories are; (i) Piezoresistive
or Semiconductor Gauge, which is made from a semiconductor
material like Germanium or Silicon; (ii) Carbon Resistor Gauge; and
(iii) Bonded Metallic Foil Resistance Gauge.
The Bonded Metallic Foil Gauge, referred to as strain gauge
henceforth, is shown in the diagram below. This type of strain
gauge was used in this investigation because it offered ease of
handling and durability, and it is not sensitive to changes in
temperature or humidity. It consists of a grid of very thin metallic
foi l , bonded to a thin insulating backing called a carrier matrix. In
use the carrier matrix is attached to the test specimen with an
adhesive.
1 12
Foil
Carrier Matrix
Leads to connect grid in circuit
When the specimen is loaded, the strain on i ts surface is
transmitted to the grid material by the adhesive and carrier system.
The strain in the specimen can be found by measuring the change in
the electrical resistance of the grid material. It may be noted here
that the strain gauge used in this investigation actually had two
resistances aligned at right angle to each other and at a 45 ° angle
to the principal axes of the shaft. One resistance was subjected to
tensi le s t ra in (^), and the other resistance was subject to
compression strain (vf). (See Section 2.3).
Before we go on to describe the pract ica l s t ra in
measurements, it is necessary to explain a term closely associated
wi th strain gauge applications. This term is called the GAUGE
FACTOR, GF. It is defined as the ratio of the fractional change in
resistance to the fractional change in length. Expressed in equation
form as;
GF = AR/R
AL/L
AR/R
e
1 13
where f denotes the strain. It is a measure of the resistance
change with strain. Gauge Factor is a dimensionless quantity and
the larger the value the more sensitive the strain gauge. It should
be noted here that the change in resistance with strain is not just
due to the dimensional changes in the conductor, but that the
restivity of the metallic foil also changes with strain: The term
Gauge Factor applies to the strain gauge as a whole, complete with
the carrier matrix, not just to the metallic foil. The gauge factor
for each resistance of the Bonded Metallic Foil gauge used in our
investigation was 2.
From the gauge factor equation we see that it is the fractional
change of in resistance that is the important quantity rather than
absolute value of the gauge resistance. For the 120 ohm Bonded
Metallic Foil gauge with a gauge factor of 2 used in this
investigation, the gauge factor equation suggested that for 1 | i f
(microstrain, 1 x 10 '^) the change in resistance would be 240
micro-ohms. This meant that we needed to have micro-ohm
sensitivity in the measuring instrumentation.
Direct measurement of resistance was rendered very difficult
due to the high sensitivity requirement. An alternate way was to
use a Wheatstone bridge circuit, which is a sensitive way of
measuring small changes in resistance.
For our application, the strain gauge with its two resistances,
formed what is referred to as a Half Bridge Configuration in the
Pressure and Strain Measurement Handbook referenced at the
1 ,1
begining of this appendix. The circuit for this configuration is
shown in the diagram below. In the circuit below, R and R^ denote
two resistances used to complete the bride circuit. Rg denotes a
gauge resistance and R] denotes the line resistance. The bridge
cirduit was excited by a d.c. voltage equal to y^^. The output
voltage, VQy from the bridge circuit was amplified by the amplifier,
R, and measured by a voltmeter. For this configuration the strain,
Rg (^e)
Rg (-ve)
f, is given by the foHowing equation. (See "Pressure and Strain
Measurement Handbook and Encyclopedia" page no. E-59).
-4V e = R
G F [ ( 1 + V ) - 2 V R ( V - 1 ) ]
R i l 1 +
l R g^
n the above equation VR is given by the equation below
^ out '
strained VR =
[Vout^ LVin I unstrained
Thus, by measuring the ratio of Vout under strained condition, when
1 15
the shaft is rotating, and under unstrained condition, when the shaft
was at rest, the value of Vp was calculated. The line resistances
were measured by a sensitive ohmmeter.
PERMISSION TO COPY
In presenting this thesis in partial fulfillment of the
requirements for a master's degree at Texas Tech University, I agree
that the Library and my major department shall make it freely avail-
able for research purposes. Permission to copy this thesis for
scholarly purposes may be granted by the Director of the Library or
my major professor. It is understood that any copying or publication
of this thesis for financial gain shall not be allowed without my
further written permission and that any user may be liable for copy-
right infringement.
Disagree (Permission not granted) Agree (Permission granted)
Student's signature Studéfit' s slgnature
Date
|o I l^ 1 ^ ^ Date