Page 1: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

M. Saiful Huq, PhD, FAAPM, FInstP

Dept. of Radiation Oncology, University of Pittsburgh Cancer Institute and UPMC Hillman Cancer Center, Pittsburgh, Pennsylvania, USA

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small

static fields used in external beam radiotherapy

Page 2: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Collaborators

• Yongqian Zhang, Ph.D.

• Min-sig Huang, Ph.D.

• Troy Teo, Ph.D.

• Kevin Fallon, M.S.

• Cihat Ozhasoglu, M.S.

• Ron Lalonde, Ph.D.

Page 3: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Contributors to IAEA TRS-483

Missing: Ahmed Meghzifene and Stan Vatnitsky

Page 4: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• The Code of Practice addresses the reference and

relative dosimetry of small static fields used for

external beam photon radiotherapy of energies with

nominal accelerating potential up to 10 MV. It does

not address other radiotherapy modalities such as

electron, proton and orthovoltage beams

TRS-483 CoP

Page 5: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• It provides a CoP for machine specific reference (msr)

dosimetry in a clinical high energy photon beams. It is

based on the use of a ionization chamber that has been

calibrated in terms of absorbed dose to water ND,w,Qo or

ND,w,Qmsr in a standard’s laboratory’s reference beam of

quality Qo or Qmsr.

• It also provides guidance for measurements of field

output factors and lateral beam profiles at the

measurement depth

TRS-483 CoP

Page 6: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Radiation generators: 10 cm x 10 cm field can be set

• Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Radiation generators: 10 cm x 10 cm ref field cannot be

set

• Define machine specific reference (msr) field, fmsr

• Dimension of fmsr field

• Should be as close as possible to the conventional reference

field

Reference dosimetry: msr field

Page 7: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Machine type msr field CyberKnife 6 cm diameter fixed collimator

TomoTherapy 5 cm x 10 cm field

GammaKnife 1.6 cm or 1.8 cm diameter collimator helmet, all sources simultaneously out

BrainLab microMLC add on For example 9.8 cm x 9.8 cm or 9.6 cm x 10.4 cm

SRS cone add-ons The closest to a 10 cm x 10 cm equivalent square msr field achievable

• fmsr should extend at least a distance rLCPE beyond the

outer boundaries of the reference ionization chamber

FWHM ≥ 2rLCPE + d

msr fields for common radiotherapy machines

Page 8: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• How were the field sizes for the msr dosimetry

arrived at?

• Ask: What is the size restriction on an ionization

chamber for msr dosimetry?

msr fields: selection of chambers

Page 9: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• CPE conditions exist when one of the edges of the field extends at least a distance rLCPE beyond the outer boundaries of the ionization chamber. If the size of the detector is d, the FWHM of the field has to fulfil the condition:

FWHM ≥ 2 rLCPE + d

rLCPE (in cm) = 0.07797•%dd(10,10)x – 4.112

rLCPE (in cm) = 8.369•TPR20,10(10) – 4.382 d

rLCPE

CPE condition for field size

Page 10: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Consider a 6 MV beam. It’s TPR20,10(10) = 0.677 • rLCPE = 8.369 0.677 – 4.382 = 12.8 mm • PTW 30013 Farmer type chamber:

• cavity length l = 23mm • cavity radius r = 3.1 mm • wall thickness twall= 0.057 g/cm2 • With ρ (PMMA) = 1.19 g/cm3, twall= 0.48mm • In the longitudinal direction, the chamber outer size will be dl = l + twall =

23.48 mm (say 23.5mm) • In the radial direction dr = 2( r + twall) = 7.2mm • As dl > dr, the largest detector size is dl

• Eq. for FWHM yields a FWHM = 2 x 12.8 + 23.5 = 49.1 mm

For a PTW 30013 chamber: FWHM ≥ 4.9 cm

Example

Page 11: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Chambers must meet specifications for reference class

ionizations chambers. Table 3 in the CoP

• Refers to chamber settling time, polarity effect, leakage,

recombination correction, chamber stability, chamber material

• fmsr ≥ 6 cm x 6 cm

• Chambers listed in Table 4 meet this criteria

msr fields: selection of chambers

Page 12: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Farmer type chambers: • WFF beams: Farmer type chambers listed in Table 4 meets this

criteria

• FFF beams use a chamber with a length shorter than the length of Farmer type chambers

• If you have to use a Farmer type chamber a correction for the non-uniformity of the beam profile should be used. For 6 MV beam this can be about 1.5%

msr fields: selection of chambers

Page 13: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• For field sizes smaller than 6 cm x 6 cm similar

analysis led to the chambers listed in Table 5

(including Gamma Knife)

• These are chambers with volumes smaller than 0.3

cc (chamber length 7 mm)

msr fields: selection of chambers

Page 14: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• For reference dosimetry in msr fields, you will need to determine “equivalent square msr field” sizes. For non-square field sizes, the corresponds to the field for which the phantom scatter is the same.

• Tables 15-17 tells you how to do this. This is needed to calculate TPR20,10(10) or %dd(10,10)x using Palman’s equation.

Equivalent square msr field sizes

Page 15: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

15

Table 15 (Tables 16 & 17 are for FFF beams)

Page 16: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Choice of an appropriate detector for small field dosimetry measurements depends on the parameter to be measured.

• Note: NO ideal detector exists for measurements in small fields

• Use two or three different types of suitable detectors so that redundancy in results can provide more assurance that no significant errors in dosimetry are made

Relative dosimetry: Detectors

Page 17: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Assume that detectors used for large field dosimetry will not perform well in small fields

• Ion chambers: major issues are volume averaging and substantial perturbations in the absence of LCPE, signal to background ratio for small volume ionization chambers

• Below certain field sizes, volume averaging effects become unacceptably large. Below these field sizes only liquid ion chamber and solid state detectors are suitable for dosimetry, but even those exhibit substantial perturbations for the smallest field sizes

Relative dosimetry: Detectors

Page 18: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Output correction factors are given as a function of

the size of the square fields. For non-square fields,

one determines a “equivalent square small field” for

which output corrections are the same

Output correction factor

Page 19: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Field output correction factors are given as a

function of Collimator setting for CyberKnife and

Gamma Knife (Tables 23 and 25) and as a function of

“equivalent square” for Tomotherapy, MLC and SRS

cones for 6 and 10 MV beams in Tables 24, 26 and 27

Output correction factor

Page 20: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Table 23: Output correction factors for CyberKnife

Page 21: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Table 26: Output correction factors for 6MV in WFF and FFF beams

Page 22: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• For relative dosimetry ensure placement of the

detector in the center of the radiation field

Practical considerations

Page 23: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Correct and incorrect orientations of detectors

for measurements of beam profile

Practical considerations

Page 24: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• The absorbed dose to water at the reference depth zref in water

for the fmsr field in a beam of quality as Qmsr and in the absence

of the ionization chamber is given by:

• is the chamber reading corrected for influence quantities

• is the absorbed dose to water calibration coefficient of the chamber at beam quality Qmsr

msr

msr

msr

msr

msr

msr

fQwD

fQ

fQw NMD ,,, ⋅=

Chamber calibrated specifically for the msr field

Formalism: Preferred option

Page 25: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• The absorbed dose to water at the reference depth zref in water for the fmsr field

in a beam of quality as Qmsr and in the absence of the ionization chamber is given

by:

• is the chamber reading corrected for influence quantities • is the absorbed dose to water calibration coefficient of the chamber at

beam quality Q0 in the ref field fref = 10x10 cm2

• is a correction factor that accounts for the differences between the

response of an ionization chamber in the field fref and beam quality Qo and the field fmsr and beam quality Qmsr

Chamber calibrated in a conventional reference field and generic values of beam quality corrections factors available

Formalism: Option b

Page 26: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Table 12: vs %dd(10,10)x and TPR20,10(10) for WFF beams

Page 27: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Table 13: vs% dd(10,10)x and TPR20,10(10) for FFF beams

Page 28: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Table 14: for GammaKnife

Page 29: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Beam quality

TPR20,10(S) %dd(10,S)x

Page 30: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Equations for beam quality in non-standard reference fields

(Palmans 2012 Med Phys 39:5513)

0.55

0.60

0.65

0.70

0.75

0.80

0.85

2 4 6 8 10 12

s / cm

TP

R20

,10(

s)

(b)4 MV

10 MV8 MV

6 MV

5 MV

25 MV21 MV18 MV15 MV12 MV

Beam quality

C=(16.15 ± 0.12) x 10 -3

Page 31: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Field output factor relative to reference field (ref stands here for

a conventional reference or msr field)

where is the so-called output correction factor, which

can be determined as a directly measured value, an

experimentally generic value or a Monte Carlo calculated

generic value

Field output factor

Page 32: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Method 1: Field output factor relative to msr field is given by

• Method 2: Field output factor relative to msr field using

intermediate field or ‘daisy chaining’ method

where

Assumed to be unity

Field output factor

Page 33: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Field output correction factor

Presenter
Presentation Notes
Data from Francesco. Just an illustration.
Page 34: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

34

Field output correction factor

Page 35: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Rectangular small fields with uneven in-plane and cross-plane FWHM, the equivalent square field size is given by

Sclin = √ A.B 0.7 < A/B < 1.4 • For circular small fields with FWHM radius r Sclin = r√ π = 1.77r

Equivalent square field size

Page 36: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Application of TRS-483

What did we measure?

Page 37: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• TPR20,10(10) and %dd(10)x in msr fields for a 6MV

beam based on measurements in different field sizes

in a TrueBeam STx linac

• Polarity effect in a GammaKnife Perfexion machine

• Reference and relative dosimetry (Field output

factors) in GammaKnife Perfexion, CyberKnife M6

with Incise MLC and TrueBeam STx machines

Measured …..

Page 38: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

msr Field Size

(S)(cm2)

Measured TPR20,10 (S)

Calculated TPR20,10 (10)*

3x3 0.631 0.669 4x4 0.634 0.666 6x6 0.645 0.666 8x8 0.656 0.667

10x10 0.667 0.667

* Measured using CC13 chamber. Calculated using Palman’s equation

0.6

0.62

0.64

0.66

0.68

0.7

3x3 4x4 6x6 8x8 10x10

TPR

20,1

0

Field Size (cm2)

Calculated TPR20,10 (10)Measured TPR20,10 (S)

TPR20,10(10) for 6 MV beam

Page 39: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

* Measured using CC13 chamber. Calculated using Palman’s equation

%dd(10,10)x for 6 MV beam

msr Field Size (cm2)

Measured %dd (S)

Calculated %dd (10,10)X *

3x3 60.48 65.79

4x4 61.71 66.15

6x6 63.80 66.65

8x8 65.34 66.76

10x10 66.54 66.54

60

62

64

66

68

70

3x3 4x4 6x6 8x8 10x10

% d

d

Field Size (cm2)

Calculated %dd (10,10)X

Measured %dd (S)

Page 40: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• Palman’s equation accurately calculates TPR20,10(10) or

%dd(10,10)x from measured values of of TPR20,10(s) and

%dd(10,s) for various field sizes

Conclusion on Beam quality index measurements

Page 41: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Polarity effect

Co-60 beam 16 mm collimator

3 Exradin A16 chamber, 1 Exradin A14 chamber, 1 PTW 31016 chamber, 2 Capintec PR05P chamber

Page 42: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

V (volts)

Pola

rity

fact

or

0.97

0.975

0.98

0.985

0.99

0.995

1

1.005

1.01

0 100 200 300 400 500 600 700 800

A16SN100075

A16SN040907

A16SN031113

PTW31016

A14

PR05P9546

PR05P7837

Polarity effect

V (volts)

Page 43: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Reference dosimetry

Page 44: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

GammaKnife® - IconTM

Page 45: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber Serial number

Dose rate TRS-483

Dose rate TRS-398

Difference (%)

Exradin A16 040907 3.235 3.177 1.8

Exradin A16 092725 3.230 3.171 1.8

Gamma Knife (Icon)

Page 46: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

CyberKnife M6TM InciseTM MLC

Page 47: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chambers Dw(zmax)/MU TRS-483 (cGy/MU)

Dw(zmax)/MU TRS-398 (cGy/MU)

Ratio TRS-398TRS-483

Exradin A12

1.016 1.009 0.993

PTW 30013 Sl. No. 1551

1.014

1.010

0.996

PTW 30013 Sl. No. 0262

1.018

1.013

0.995

PTW 30013 Sl. No. 0343

1.017

1.012

0.995

PTW 30013 Sl. No. 0905

1.024

1.019

0.995

Avg ± sd 1.018 ± 0.4% 1.012 ± 0.4%

CyberKnife : 6X FFF

Page 48: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

TrueBeamTM STx

Page 49: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber-Serial #

Mcorrected (nC)

ND,w

109(Gy/C) %dd(10,10)x

Dw/MU at zmax

TPR20,10(10)

Dw/MU at zmax

Dw (%dd)x/Dw(TPR)

PTW 30013-1551

14.46

5.397

66.38

0.991

1.007

0.668

0.993

1.009

0.998

PTW 30013-0262

14.57

5.343

66.38

0.991

1.004

0.668

0.993

1.006

0.998

PTW 30013-0905

14.54

5.37

66.38

0.991

1.008

0.668

0.993

1.010

0.998

EXR A12 -020581

15.75 4.915 66.38 0.995 1.002 0.668 0.995 1.003 0.999

TRS-483 : 6X

Page 50: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber-Serial #

Mcorrected (nC)

ND,w

109(Gy/C) %dd(10,10)x

Dw/MU at zmax

TPR20,10(10)

Dw/MU at zmax

Dw (%dd)x/Dw(TPR)

PTW 30013-1551

13.77

5.397

63.89

0.995

1.007

0.632

0.995

1.007

1.000

PTW 30013-0262

13.88

5.343

63.89

0.995

1.005

0.632

0.995

1.005

1.000

PTW 30013-0905

13.85

5.37

63.89

0.995

1.008

0.632

0.995

1.008

1.000

EXR A12 -020581

14.98 4.915 63.89 0.998 1.001 0.632 0.998 1.001 1.000

TRS-483 : 6X FFF

Page 51: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber-Serial #

Mcorrected (-nC)

ND,w

109(Gy/C) %dd(10,10)x

Dw/MU at zmax

TPR20,10(10)

Dw/MU at zmax

Dw (%dd)x/Dw(TPR)

PTW 30013-1551

15.97

5.397

73.15

0.979

1.004

0.740

0.980

1.004

0.999

PTW 30013-0262

16.13

5.343

73.15

0.979

1.004

0.740

0.980

1.004

0.999

PTW 30013-0905

16.05

5.37

73.15

0.979

1.004

0.740

0.980

1.004

0.999

EXR A12 -020581

17.33 4.915 73.15 0.985 0.998 0.740 0.986 0.999 0.999

TRS-483 : 10X

Page 52: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber-Serial #

Mcorrected (-nC)

ND,w

109(Gy/C) %dd(10,10)x

Dw/MU at zmax

TPR20,10(10)

Dw/MU at zmax

Dw (%dd)x/Dw(TPR)

PTW 30013-1551

15.33

5.397

71.39

0.985

1.005

0.707

0.987

1.007

0.998

PTW 30013-0262

15.48

5.343

71.39

0.985

1.005

0.707

0.987

1.007

0.998

PTW 30013-0905

15.38

5.37

71.39

0.985

1.004

0.707

0.987

1.005

0.998

EXR A12 -020581

16.60 4.915 71.39 0.991 0.997 0.707 0.992 0.998 0.999

TRS-483 : 10X FFF

Page 53: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber-Serial #

TPR20,10(10) Dw/MU at zmax

Dw(TRS398)/ Dw(TRS483)

PTW 30013-1551

0.668

0.992

0.993

1.008

1.009

0.999

PTW 30013-0262

0.668

0.992

0.993

1.005

1.006

0.999

PTW 30013-0905

0.668

0.992

0.993

1.008

1.010

0.999

EXR A12 -020581

0.668 0.995 0.995 1.003

1.003 1.000

TRS-483 vs TRS-398 : 6X

Page 54: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber-Serial #

TPR20,10(10) Dw/MU at zmax

Dw(TRS398)/ Dw(TRS483)

PTW 30013-1551

0.632

0.996

0.995

1.008

1.007

1.001

PTW 30013-0262

0.632

0.996

0.995

1.006

1.005

1.001

PTW 30013-0905

0.632

0.996

0.995

1.009

1.008

1.001

EXR A12 -020581

0.632 0.998 0.998 1.001

1.001 1.000

TRS-483 vs TRS-398 : 6X FFF

Page 55: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber-Serial #

TPR20,10(10) Dw/MU at zmax

Dw(TRS398)/ Dw(TRS483)

PTW 30013-1551

0.740

0.980

0.980

1.004

1.004

1.000

PTW 30013-0262

0.740

0.980

0.980

1.004

1.004

1.000

PTW 30013-0905

0.740

0.980

0.980

1.004

1.004

1.000

EXR A12 -020581

0.740 0.986 0.986 0.999 0.999 1.000

TRS-483 vs TRS-398 : 10X

Page 56: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Chamber-Serial #

TPR20,10(10) Dw/MU at zmax

Dw(TRS398)/ Dw(TRS483)

PTW 30013-1551

0.707

0.987

0.987

1.006

1.007

0.999

PTW 30013-0262

0.707

0.987

0.987

1.006

1.007

0.999

PTW 30013-0905

0.707

0.980

0.980

1.005

1.005

0.999

EXR A12 -020581

0.707 0.98 0.987 0.997 0.998 0.999

TRS-483 vs TRS-398 : 10X FFF

Page 57: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

Relative dosimetry

Field output factor

Page 58: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

CyberKnife

Page 59: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

CyberKnife

Page 60: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

TrueBeam STx

Page 61: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

61

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

6X

mean = 0.713 ∆uncorr = - 2%

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Unc

orre

cted

ratio

of r

eadi

ngs

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

6X

mean = 0.728

TrueBeam STx : 6X

Page 62: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

62

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

6X (IMF)

mean = 0.715 ∆uncorr = - 2%

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Unc

orre

cted

ratio

of r

eadi

ngs

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

6X

mean = 0.728

TrueBeam STx : 6X

Page 63: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

63

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

UncorrectedCorrectedCorrected (IFM)

6X, PTW60017

0.5

0.55

0.6

0.65

0.7

0.75

0 0.5 1 1.5 2

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

UncorrectedCorrectedCorrected (IF)

6X, PTW60017

mean = 0.591 ∆uncorr = - 3.6%

mean = 0.714 ∆uncorr = - 0.9%

TrueBeam STx : 6X

Page 64: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

64

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Unc

orre

cted

ratio

of r

eadi

ngs

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

6X FFF

mean = 0.743

0.5

0.6

0.7

0.8

0.9

1.0

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

6X FFF

mean = 0.728 ∆uncorr = - 2%

TrueBeam STx : 6XFFF

Page 65: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

65

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Unc

orre

cted

ratio

of r

eadi

ngs

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

6X FFF

mean = 0.743

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

6X FFF (IFM)

mean = 0.732 ∆uncorr = - 2%

TrueBeam STx : 6XFFF

Page 66: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

66

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

UncorrectedCorrectedCorrected (IF)

6XFFF, PTW60017

0.5

0.55

0.6

0.65

0.7

0.75

0 0.5 1 1.5 2

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

UncorrectedCorrectedCorrected (IF)

6XFFF, PTW60017

mean = 0.607 ∆uncorr = - 3.5%

mean = 0.730 ∆uncorr = - 0.6%

TrueBeam STx : 6XFFF

Page 67: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

67

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Unc

orre

cted

ratio

of r

eadi

ngs

Equivalent square field size, Sclin (cm)

Sun Nuclear Edge

PTW 60017

10X

mean = 0.692

0.5

0.6

0.7

0.8

0.9

1.0

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

10X

mean = 0.677 ∆uncorr = - 2%

TrueBeam STx : 10X

Page 68: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

68

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Unc

orre

cted

ratio

of r

eadi

ngs

Equivalent square field size, Sclin (cm)

Sun Nuclear Edge

PTW 60017

10X

mean = 0.692

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

Sun Nuclear EdgePTW 60017

10X (IFM)

mean = 0.678 ∆uncorr = - 2%

TrueBeam STx : 10X

Page 69: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

69

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

UncorrectedCorrectedCorrected (IF)

10X, PTW60017

0.5

0.55

0.6

0.65

0.7

0.75

0 0.5 1 1.5 2

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

UncorrectedCorrectedCorrected (IF)

10X, PTW60017

mean = 0.506 ∆uncorr = - 3.9%

mean = 0.675 ∆uncorr = - 1.5 %

TrueBeam STx : 10X

Page 70: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

70

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Unc

orre

cted

ratio

of r

eadi

ngs

Equivalent square field size, Sclin (cm)

PTW 60017

10XFFF

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

PTW 60017

10X FFF

TrueBeam STx : 10X FFF

Page 71: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

71

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Unc

orre

cted

ratio

of r

eadi

ngs

Equivalent square field size, Sclin (cm)

PTW 60017

10XFF

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

PTW 60017

10X FFF (IFM)

TrueBeam STx : 10X FFF

Page 72: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

72

0.5

0.6

0.7

0.8

0.9

1

0 2 4 6 8 10

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

UncorrectedCorrectedCorrected (IF)

10XFFF, PTW60017

0.5

0.55

0.6

0.65

0.7

0.75

0 0.5 1 1.5

Fiel

d ou

tput

fact

ors

Equivalent square field size, Sclin (cm)

UncorrectedCorrectedCorrected (IF)

10XFFF, PTW60017

mean = 0.506 ∆uncorr = - 3.9%

mean = 0.675 ∆uncorr = - 1.5 %

TrueBeam STx : 10X FFF

Page 73: Implementation of the IAEA-AAPM Code of Practice for the …€¦ ·  · 2018-05-07Implementation of the IAEA-AAPM Code ... • Follow TRS-398 CoP or AAPM TG-51 or equivalent protocol

• For the GammaKnife msr beam, differences in references dosimetry using TRS-483 and TRS-398 can be up to 2% assuming depth scaling is taken into consideration

• For linac WFF and FFF beams, the values of Dw/MU following TRS-483 are consistent within better than 1% with those obtained using TRS-398

• The small field dosimetry of certain msr (reference) and most relative (using field output factor) beams can be significantly improved when the correction factors or different detectors included in TRS-483 are appropriately incorporated into their dosimetry

Summary