instrumental analysis chem 4811 chapter 2 dr. augustine ofori agyeman assistant professor of...
TRANSCRIPT
INSTRUMENTAL ANALYSIS CHEM 4811
CHAPTER 2
DR. AUGUSTINE OFORI AGYEMANAssistant professor of chemistryDepartment of natural sciences
Clayton state university
CHAPTER 2
INTRODUCTION TO SPECTROSCOPY
DEFINITIONS
Spectroscopy- The study of the interactions of electromagnetic radiation
(radiant energy) and matter (molecules, atoms, or ions)
Spectrometry- Quantitative measurement of the intensity of one or more
wavelengths of radiant energy
Spectrophotometry- The use of electromagnetic radiation to measure
chemical concentrations(used for absorption measurements)
Spectrophotometer - Instrument used for absorption measurements
Optical Spectrometer- Instrument that consists of prism or grating dispersion devise,
slits, and a photoelectric detector
Photometer- Instrument that uses a filter for wavelength selection instead
of a dispersion device
DEFINITIONS
ELECTROMAGNETIC RADIATION
- Also known as radiant heat or radiant energy
- One of the ways by which energy travels through space
- Consists of perpendicular electric and magnetic fields that are also perpendicular to direction of propagation
Examplesheat energy in microwaves
light from the sunX-ray
radio waves
Gamma rays
X rays Ultr-violet
Infrared Microwaves Radio frequency FM Shortwave AM
Vis
ible
Visible Light: VIBGYORViolet, Indigo, Blue, Green, Yellow, Orange, Red
400 – 750 nm
- White light is a blend of all visible wavelengths
- Can be separated using a prism
Wavelength (m)
Frequency (s-1)
10-11 103
1020104
ELECTROMAGNETIC RADIATION
one second
λ1
λ3
λ2
ν1 = 4 cycles/second
ν2 = 8 cycles/second
ν3 = 16 cycles/second
amplitude
peak
trough
ELECTROMAGNETIC RADIATION
node
Wavelength (λ) - Distance for a wave to go through a complete cycle
(distance between two consecutive peaks or troughs in a wave)
Frequency (ν)- The number of waves (cycles) passing a given point
in space per second
Cycle- Crest-to-crest or trough-to-trough
Speed (c)- All waves travel at the speed of light in vacuum (3.00 x 108 m/s)
ELECTROMAGNETIC RADIATION
ELECTROMAGNETIC RADIATION
Plane Polarized Light- Light wave propagating along only one axis (confined to one plane)
Monochromatic Light- Light of only one wavelength
Polychromatic Light- Consists of more than one wavelength (white light)
Visible light- The small portion of electromagnetic radiation to which
the human eye responds
- Inverse relationship between wavelength and frequency
λ α 1/ν
c = λ ν
λ = wavelength (m)
ν = frequency (cycles/second = 1/s = s-1 = hertz = Hz)
c = speed of light (3.00 x 108 m/s)
ELECTROMAGNETIC RADIATION
ELECTROMAGNETIC RADIATION
- Light appears to behave as waves and also considered as stream of particles (the dual nature of light)
- Is sinusoidal in shape
- Light is quantized
Photons- Particles of light
h = Planck’s constant (6.626 x 10-34 J-s)
ν = frequency of the radiation
λ = wavelength of the radiation
E is proportional to ν and inversely proportional to λ
)(mwavenumberλ
1ν~ 1
ν~hcλ
hchν)(E photon one ofEnergy photon
ELECTROMAGNETIC RADIATION
- Takes place in many ways
- Takes place over a wide range of radiant energies
- Is not visible to the human eye
- Light is absorbed or emitted
- Follows well-ordered rules
- Can be measured with suitable instruments
INTERACTIONS WITH MATTER
INTERACTIONS WITH MATTER
- Atoms, molecules, and ions are in constant motion
Solids- Atoms or molecules are arranged in a highly ordered array (crystals)
orarranged randomly (amorphous)
Liquids- Atoms or molecules are not as closely packed as in solids
Gases- Atoms or molecules are widely separated from each other
INTERACTIONS WITH MATTER
Molecules
Many types of motion are involved- Rotation - Vibration
- Translation (move from place to place)
- These motions are affected when molecules interact with radiant energy
- Molecules vibrate with greater energy amplitude when they absorb radiant energy
INTERACTIONS WITH MATTER
Molecules
- Bonding electrons move to higher energy levels when molecules interact with visible or UV light
- Changes in motion or electron energy levels result inchanges in energy of molecules
Transition- Change in energy of molecules
(vibrational transitions, rotational transitions, electronic transitions)
INTERACTIONS WITH MATTER
Atoms or Ions
- Move between energy levels or in space but cannot rotate or vibrate
The type of interactions of materials with radiant energy are affected by- Physical state
- Composition (chemical nature)- Arrangement of atoms or molecules
INTERACTIONS WITH MATTER
Light striking a sample of matter may be- Absorbed by the sample
- Transmitted through the sample- Reflected off the surface of the sample
- Scattered by the sample
- Samples can also emit light after absorption (luminescence)
- Species (atoms, ions, or molecules) can exist in certain discrete states with specific energies
Transmission- Light passes through matter without interaction
Absorption- Matter absorbs light energy and moves to a higher energy state
Emission- Matter releases energy and moves to a lower energy state
Luminescence- Emission following excitation of molecules or atoms by
absorption of electromagnetic radiation
INTERACTIONS WITH MATTER
Ene
rgy
Absorption Emission
Excitedstate
Groundstate
INTERACTIONS WITH MATTER
Ground State: The lowest energy state
Excited state: higher energy state (usually short-lived)
INTERACTIONS WITH MATTER
- Change in state requires the absorption or emission of energy
λ
hchνE)(energyinChange
- Matter can only absorb specific wavelengths or frequencies
- These correspond to the exact differences in energy between the two states involved
Absorption: Energy of species increases (ΔE is positive)
Emission: Energy of species decreases (ΔE is negative)
INTERACTIONS WITH MATTER
- Frequencies and the extent of absorption or emission of species are unique
- Specific atoms or molecules absorb or emit specific frequencies
- This is the basis of identification of species by spectroscopy
Relative energy of transition in a moleculeRotational < vibrational < electronic
- The are many associated rotational and vibrational sublevels for any electronic state (absorption occurs in
closely spaced range of wavelenghts)
INTERACTIONS WITH MATTER
Absorption Spectrum- A graph of intensity of light absorbed versus frequency
or wavelength
- Emission spectrum is obtained when molecules emit energy by returning to the ground state after excitation
Excitation may include- Absorption of radiant energy
- Transfer of energy due to collisions between atoms or molecules- Addition of thermal energy
- Addition of energy from electrical charges
ATOMS AND ATOMIC SPECTROSCOPY
- The electronic state of atoms are quantized
- Elements have unique atomic numbers (numbers of protons and electrons)
- Electrons in orbitals are associated with various energy levels
- An atom absorbs energy of specific magnitude and a valence electron moves to the excited state
- The electron returns spontaneously to the ground state and emits energy
ATOMS AND ATOMIC SPECTROSCOPY
- Emitted energy is equivalent to the absorbed energy (ΔE)
- Each atom has a unique set of permitted electronic energy levels(due to unique electronic structure)
- The wavelength of light absorbed or emitted are characteristic of a specific element
- The absorption wavelength range is narrow due to the absence of rotational and vibrational energies
- The wavelength range falls within the ultraviolet and visible regions of the spectrum (UV-VIS)
ATOMS AND ATOMIC SPECTROSCOPY
- Wavelengths of absorption or emission are used for qualitative identification of elements in a sample
- The intensity of light absorbed or emitted at a given wavelength is used for the quantitative analysis
Atomic Spectroscopy Methods- Absortion spectroscopy- Emission spectroscopy
- Fluorescence spectroscopy- X-ray spectroscopy (makes use of core electrons)
Gamma rays X rays Ultr-
violetInfrared Microwaves Radio frequency
FM Shortwave AMV
isib
le
10-11 103
1020104
Bon
d br
eaki
ngan
d io
niza
tion
Ele
ctro
nic
exci
tati
on
vibr
atio
n
rota
tion
Molecular Processes Occurring in Each Region
MOLECULES AND MOLECULAR SPECTROSCOPY
MOLECULES AND MOLECULAR SPECTROSCOPY
- Energy states are quantized
Rotational Transitions- Molecules rotate in space and rotational energy is associated
- Absorption of the correct energy causes transition to a higher energy rotational state
- Molecules rotate faster in a higher energy rotational state
- Rotational spectra are usually complex
MOLECULES AND MOLECULAR SPECTROSCOPY
Rotational Transitions
- Rotational energy of a molecule depends on shape,angular velocity, and weight distribution
- Shape and weight distribution change with bond angle
- Molecules with more than two atoms have many possible shapes
- Change in shape is therefore restricted to diatomic molecules
- Associated energies are in the radio and microwave regions
MOLECULES AND MOLECULAR SPECTROSCOPY
Vibrational Transitions
- Atoms in a molecule can vibrate toward or away from each other at different angles to each other
- Each vibration has characteristic energy associated with it
- Vibrational energy is associated with absorption in the infrared (IR region)
Increase in rotational energy usually accompanies increase in vibrational energy
MOLECULES AND MOLECULAR SPECTROSCOPY
Vibrational Transitions
- IR absorption corresponds to changes in both rotational and vibrational energies in molecules
- IR absorption spectroscopy is used to deduce the structure of molecules
- Used for both qualitative and quantitative analysis
MOLECULES AND MOLECULAR SPECTROSCOPY
Electronic Transitions
- Molecular orbitals are formed when atomic orbitals combine to form molecules
- Absorption of the correct radiant energy causes an outer electron to move to an excited state
- Excited electron spontaneously returns to the ground state (relax) emitting UV or visible energy
- Excitation in molecules causes changes in the rotational and vibrational energies
MOLECULES AND MOLECULAR SPECTROSCOPY
Electronic Transitions
- The total energy is the sum of all rotational, vibrational, and electronic energy changes
- Associated with wide range of wavelengths (called absorption band)
- UV-VIS absorption bands are simpler than IR spectra
MOLECULES AND MOLECULAR SPECTROSCOPY
Molecular Spectroscopy Methods
- Molecular absorption spectroscopy- Molecular emission spectroscopy
- Nuclear Magnetic Resonance (NMR)- UV-VIS
- IR- MS
- Molecular Fluorescence Spectroscopy
ABSORPTION LAWS
Radiant Power (P)- Energy per second per unit area of a beam of light
- Decreases when light transmits through a sample(due to absorption of light by the sample)
Intensity (I)- Power per unit solid angle
- Light intensity decreases as light passes through an absorbing material
Transmittance (T)
- The fraction of incident light that passes through a sample
Io I
oI
IT
0 < T < 1
Io = light intensity striking a sampleI = light intensity emerging from sample
ABSORPTION LAWS
Transmittance (T)
- T is independent of Io
- No light absorbed: I = Io and T = 1
- All light absorbed: I = 0 and T = 0
Percent Transmitance (%T)
0% < %T < 100%
100%xI
I%T
o
ABSORPTION LAWS
Absorbance (A)
- No light absorbed: I = Io and A = 0
Percent Absorbance (%A) = 100 - %T
- 1% light absorbed implies 99% light transmitted
- Higher absorbance implies less light transmitted
logTI
Ilog
I
IlogA
o
o
ABSORPTION LAWS
Beer’s Law
A = abc
A = absorbance
a = absorptivitya = ε [molar absorptivity (M-1cm-1) if C is in units of M (mol/L)]
b = pathlength or length of cell (cm)
c = concentration
ABSORPTION LAWS
Beer’s Law
- I or T decreases exponentially with increasing pathlength
- A increases linearly with increasing pathlength
- A increases linearly with increasing concentration
- More intense color implies greater absorbance
- Basis of quantitative measurements (UV-VIS, IR, AAS etc.)
ABSORPTION LAWS
Absorption Spectrum of 0.10 mM Ru(bpy)32+
λmax = 452 nm
ABSORPTION LAWS
λmax = 540 nm
Absorption Spectrum of 3.0 mM Cr3+ complex
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
350 400 450 500 550 600
Wavelength (nm)
Abs
orba
nce
ABSORPTION LAWS
Maximum Response (λmax)
- Wavelength at which the highest absorbance is observed for a given concentration
- Gives the greatest sensitivity
ABSORPTION LAWS
Deviations from Beer’s Law
- Deviations from linearity at high concentrations
- Usually used for concentrations below 0.01 M
- Deviations occur if sample scatters incident radiation
- Error increases as A increases (law generally obeyed when A ≤ 1.0
ABSORPTION LAWS
Calibration
- The relationship between the measured signal (absorbance in this case) and known concentrations of analyte
- Concentration of an unknown analyte can then be calculated using the established relationship and
its measured signal
CALIBRATION METHODS
Calibration with External Standards
- Solutions containing known concentrations of analyte are called standard solutions
- Standard solutions containing appropriate concentration range are carefully prepared and measured
- Reagent blank is used for instrumental baseline
- A plot of absorbance (y-axis) vs concentration (x-axis) is made
CALIBRATION METHODS
Calibration with External Standards
CALIBRATION METHODS
Calibration with External Standards
- Equation of a straight line in the form y = mx + z is established
m = slope = ab
z = intercept on the absorbance axis
- Concentration of unknown analyte should be within working range (do not extrapolate)
- Must measure at least three replicates and report uncertainty
CALIBRATION METHODS
Method of Standard Additions (MSA)
- Known amounts of analyte are added directly to the unknown sample
- The increase in signal due to the added analyte is used to establish the concentration of unknown
- Relationship between signal and concentration of analyte must be linear
- Analytes are added such that change in volume is negligible
CALIBRATION METHODS
Method of Standard Additions (MSA)
- Different concentrations of analyte are added to different aliquots of sample
- Nothing is added to the first aliquot (untreated)
- Concentrations in increments of 1.00 is usually used for simplicity
- Plot of signal vs concentration of analyte is made
CALIBRATION METHODS
curvencalibratioofslope
sampleuntreatedtoduesignalsampleunknownofionConcentrat
Method of Standard Additions (MSA)
Useful
- In emergency situations
- When information about the sample matrix is unknown
- For elimination of certain interferences in the matrix
CALIBRATION METHODS
Internal Standard Calibration
- Signal from internal standard is used to correct for interferences in an analyte
- The selected internal standard must not be already present in all samples, blanks, and standard solutions
- Internal standard must not interact with analyte
Internal Standard- Known amount of a nonanalyte species that is added to all
samples, blanks, and standards
CALIBRATION METHODS
Internal Standard Calibration
- For an analyte (A) and internal standard (S)
Signal ratio (A/S) is plotted against concentration ration (A/S)
Concentration ratio (A/S) of unknown is obtained from the linear equation
CALIBRATION METHODS
standard of (A/S) ratio signal
sampleunknownof(A/S)ratiosignal
standardof(A/S)ratioionConcentrat
sampleunknownof(A/S)ratioionConcentrat
Internal Standard Calibration
Corrects errors due to
- Voltage fluctuations
- Loss of analyte during sample preparation
- Change in volume due to evaporation
- Interferences
CALIBRATION METHODS
- Indeterminate (random) errors are associated with all spectroscopic methods
Examples- Noise due to instability of light source
- Detector instability- Variation in placement of cell in light path
- Finger prints on cells
ERRORS ASSOCIATED WITH BEER’S LAW
EVALUATION OF ERRORS
TlogT
T0.434
c
ΔcionconcentratinerrorRelative
- ΔT is the error in transmittance measurement
- The relative error is high when T is very high or very low
- For greatest accuracy, measurements should be within 15% - 65% T or 0.19 - 0.82 A
- Samples with high concentration (A > 0.82) should be diluted and those with low concentrations (A < 0.19) should
be concentrated
EVALUATION OF ERRORS
Ringbom Method
(100 – %T) is plotted against log(c)
- The result is an s-shaped curve (Ringbom plot)
- The nearly linear portion of the curve (the steepest portion) is the working range where error is minimized
(100-%T)
Log(c)
OPTICAL SYSTEMS IN SPECTROSCOPY
Fundamental Concepts of Optical Measurements
- Measurement of absorption or emission of radiation
- Providing information about the wavelength of absorption or emission
- Providing information about the intensity or absorbance at the wavelength
OPTICAL SYSTEMS IN SPECTROSCOPY
Main Components of Spectrometers
- Radiation source
- Wavelength selection device
- Sample holder (transparent to radiation)
- Detector
OPTICAL SYSTEMS IN SPECTROSCOPY
- FT spectrometers do not require wavelength selector
- Radiation source is the sample if emission is being measured
- External radiation source is required if absorption is being measured
- Sample holder is placed after wavelength selector for UV-VIS absorption spectrometry so that monochromatic light falls
on the sample
- Sample holder is placed before the wavelength selector for IR, fluorescence, and AA spectroscopy
COMPONENTS OF THE SPECTROMETER
Po PLightsource
monochromator
(λ selector) sample readout detector
b
Absorption (UV-Vis)
Lightsource
monochromator
(λ selector) sample readout detector
Absorption (IR)
COMPONENTS OF THE SPECTROMETER
Source& sample
monochromator
(λ selector)
readout detector
Emission
- Sample is an integral portion of the source
- Used to produce the EM radiation that will be measured
COMPONENTS OF THE SPECTROMETER
Sourceλ selector
sample
monochromator
(λ selector)readout detector
Fluorescence
COMPONENTS OF THE SPECTROMETER
- Must emit radiation over the entire wavelength range being studied
- Intensity of radiation of the wavelength range should be high
- A reliable and steady power supply is essential to provide constant signal
- Intensity should not fluctuate over long time intervals
- Intensity should not fluctuate over short time intervals
Flicker: short time fluctuation in source intensity
RADIATION SOURCE
Two types of radiation sources
Continuum Sources and
Line Sources
RADIATION SOURCE
Continuum Sources- Emit radiation over a wide range of wavelengths
- Intensity of emission varies slowly as a function of wavelength- Used for most molecular absorption and fluorescence
spectrometric instruments
Examples- Tungsten filament lamp (visible radiation)
- Deuterium lamp (UV radiation)- High pressure Hg lamp (UV radiation)
- Xenon arc lamp (UV-VIS region)- Heated solid ceramics (IR region)
- Heated wires (IR region)
RADIATION SOURCE
Line Sources- Emit only a few discrete wavelengths of light
- Intensity is a function of wavelength- Used for molecular, atomic, and Raman spectroscopy
Examples- Hollow cathode lamp (UV-VIS region)
- Electrodeless discharge lamp (UV-VIS region)- Sodium vapor lamp (UV-VIS region)- Mercury vapor lamp (UV-VIS region)
- Lasers (UV-VIS and IR regions)
RADIATION SOURCE
RADIATION SOURCE
Tungsten Filament Lamp- Glows at a temperature near 3000 K
- Produces radiation at wavelengths from 320 to 2500 nm- Visible and near IR regions
Dueterium (D2) Arc Lamp- D2 molecules are electrically dissociated
- Produces radiation at wavelengths from 200 to 400 nm- UV region
Mercury and Xenon Arc Lamps- Electric discharge lamps
- Produce radiation at wavelengths from 200 to 800 nm- UV and Visible regions
Silicon Carbide (SiC) Rod - Also called globar
- Electrically heated to about 1500 K- Produces radiation at wavelengths from 1200 to 40000 nm
- IR region
RADIATION SOURCE
Also for IR Region
- NiChrome wire (750 nm to 20000 nm)
- ZrO2 (400 nm to 20000 nm)
RADIATION SOURCE
Laser
- Produce specific spectral lines- Used when high intensity line source is required
Can be used forUV
Visible FTIR
RADIATION SOURCE
WAVELENGTH SELECTION DEVICES
Two types
Filters and
Monochromators
FILTERS
- The simplest and most inexpensive
Two major types
Absorption Filters and
Interference Filters
FILTERS
Absorption Filters
- A piece of colored glass
- Stable, simple and cheap
- Suitable for spectrometers designed to be carried to the field
Disadvantage- Range of wavelengths transmitted is very broad (50 – 300 nm)
FILTERS
Interference Filters
- Made up of multiple layers of materials
- The thickness and the refractive index of the center layer of the material control the wavelengths transmitted
- Range of wavelengths transmitted are much smaller (1 – 10 nm)
- Amount of light transmitted is generally higher
- Transmits light in the IR, VIS, and UV regions
MONOCHROMATORS
- Disperse a beam of light into its component wavelengths- Allow only a narrow band of wavelengths to pass
- Block all other wavelengths
Components- Dispersion element
- Two slits (entrance and exit)- Lenses and concave mirrors
MONOCHROMATORS
Dispersion Element
- Disperses (spreads out) the radiation falling on it according to wavelength
Two main TypesPrisms
andGratings
MONOCHROMATORS
Prisms
- Used to disperse IR, VIS, and UV radiations
- Widely used is the Cornu prism (60o-60o-60o triangle)
ExamplesQuartz (UV)
Silicate glass (VIS or near IR)NaCl or KBr (IR)
MONOCHROMATORS
Prisms
- Refraction or bending of incident light occurs when a polychromatic light hits the surface of the prism
- Refractive index of prism material varies with wavelength
- Various wavelengths are separated spatially as they are bent at different degrees
- Shorter wavelengths (higher energy) are bent more than longer wavelengths (lower energy)
MONOCHROMATORS
Diffraction Gratings
- Consists of a series of closely spaced parallel grooves cut(or ruled) into a hard glass, metallic or ceramic surface
- The surface may be flat or concave
- Reflective coating (e.g. Al) is usually on the ruled surface
- Used for UV-VIS radiation (500 – 5000 grooves/mm) and IR radiation (50 – 200 grooves/mm)
MONOCHROMATORS
d
Top view Side view
Diffraction Gratings
MONOCHROMATORS
Diffraction Gratings
- Size ranges between 25 x 25 mm to 110 x 110 mm
- Light is dispersed by diffraction due to constructive interference between reflected light waves
- Separation of light occurs due to different wavelengths being dispersed (diffracted) at different angles
MONOCHROMATORS
Diffraction Gratings
- Constructive interference occurs when
nλ = d(sini ± sinθ)
n = order of diffraction (integer: 1, 2, 3, …)λ = wavelength of radiation
d = distance between groovesi = incident angle of a beam of light
θ = angle of dispersion of light
MONOCHROMATORS
Dispersive Resolution
Resolving Power (R): - Ability to disperse radiation
- Ability to separate adjacent wavelengths from each other
δλ
λR
λ = average of the wavelengths of the two lines to be resolved
δλ = difference between the two wavelengths
MONOCHROMATORS
Resolution of a Prism
dλ
dηtR
t = thickness of the base of the prism
dη/dλ = rate of change of the refractive index (η) with λ
- Resolving power increases with thickness of the prism and decreases at longer wavelengths
- Resolution depends on the prism material
MONOCHROMATORS
Resolution of a Grating
R = nN
n = the order
N = total number of grooves in the grating that are illuminated by light from the entrance slit (whole number)
Increased resolution results from - Longer gratings
- Smaller groove spacing- Higher order
MONOCHROMATORS
Dispersion of a Grating
dy
dλ)(D dispersion Reciprocal 1-
dλ = change in wavelength
dy = change in distance separating the λs along the dispersion axis
Units: nm/mm
MONOCHROMATORS
nF
dD 1-
Dispersion of a Grating
Spectral bandwidth (bandpass) = sD-1
s = slit width of monochromator
d = distance between two adjacent groovesn = diffraction order
F = focal length of the monochromator system
- D-1 is constant with respect to wavelength
ECHELLE MONOCHROMATOR
Echellette Grating
- Grooved or blazed such that it has relatively broad faces from which reflection occurs
- Has narrow unused faces
- Provides highly efficient diffraction grating
ECHELLE MONOCHROMATOR
- Contains two dispersion elements arranged in series
- The first is known as echelle grating
- The second (called cross-dispersion) is a low-dispersion prism or a grating
Echelle grating- Greater blaze angle
- The short side of the blaze is used rather than the long side- Relatively coarse grating
- Angle of dispersion (θ) is higher- Results in 10-fold resolution
OPTICAL SLITS
- Slits are used to select radiation from the light source both before and after dispersion by the λ selector
- Made of metal in the shape of two knife edges
- Movable to set the desired mechanical width
OPTICAL SLITS
Entrance Slit
- Allows a beam of light (polychromatic) from source to fall on the dispersion element
- Radiation is collimated into a parallel beam with lenses or front-faced mirrors
- One (selected) wavelength of light (monochromatic) is focused on the exit slit after dispersion
OPTICAL SLITS
Exit Slit
- Allows only a very narrow band of light to pass through sample and detector
- The dispersed light falls on the exit slit
- The light is redirected and focused onto the detector for intensity measurements
- Slits are kept as close as possible to ensure resolution
CUVET (SAMPLE CELL)
- Cell used for spectrometry
Identical or Optically Matched Cells - Cells that are identical in their absorbance or transmittance of light
Fused silica Cells (SiO2)-Transmits visible and UV radiation
Plastic and Glass Cells- Only good for visible wavelengths
NaCl and KBr Crystals- IR wavelengths
DETECTORS
- Used to measure the intensity of radiation coming out of the exit slit
- Produces an electric signal proportional to the radiation intensity
- Signal is amplified and made available for direct display
- A sensitivity control amplifies the signal
- Noisy signal is observed when amplification is too much
- May be controlled manually or by a microprocessor(the use of dynodes)
DETECTORS
Examples- Phototube (UV)
- Photomultiplier tube (UV-VIS)- Thermocouple (IR)
- Thermister (IR)- Silicon photodiode- Photovoltaic cell
- Charge Transfer Devices (UV-VIS and IR)Charge-coupled devices (CCDs)Charge injector devices (CIDs)
SINGLE-BEAM OPTICS
- Usually used for all emission methods where sample is at the location of the source
Drift- Slow variation in signal with time
- Can cause errors in single-beam methods
Sources of Drift- Changes in Voltage which changes source intensity
- Warming up of source with time- Deterioration of source or detector with time
Single-Beam Spectrometer
- Only one beam of light
- First measure reference or blank (only solvent) as Io
Io ILightsource
monochromator(selects λ) sample computer detector
b
SINGLE-BEAM OPTICS
DOUBLE-BEAM OPTICS
- Widely used
- Beam splitter is used to split radiation into two approximately equal beams (reference and sample beams)
- Radiation may also alternate between sample and referencewith the aid of mirrors (rotating beam chopper)
- Other variations are available
- The reference cell may be empty or containing the blank
- More accurate since it eliminates drift errors
Double-Beam Spectrometer
- Houses both sample cuvet and reference cuvet
Po
PLightsource
monochromator(selects λ) sample computer detector
reference
b
DOUBLE-BEAM OPTICS
SPECTROPHOTOMETERS
Photodiode Array Spectrophotometers- Records the entire spectrum (all wavelengths) at once
- Makes use of a polychromator
- The polychromator disperses light into component wavelengths
Dispersive Spectrophotometers- Records one wavelength at a time
- Makes use monochromator to select wavelength
FOURIER TRANFORM SPECTROPHOTOMETERS
- Have no slits and fewer optical elements
Multiplex- Instrument that uses mathematical methods to interpret and
present spectrum without dispersion devices
- Wavelengths of interest are collected at a time without dispersion
- The wavelengths and their corresponding intensities overlap
- The overlapping information is sorted out in order to plot a spectrum
FOURIER TRANFORM SPECTROPHOTOMETERS
- Sorting out or deconvoluting the overlapping signals of varying wavelengths (or frequencies) is a mathematical procedure
called Fourier Analysis
- Fourier Analysis expresses complex spectrum as a sum of sine and cosine waves varying with time
- Data acquired is Fourier Transformed into the spectrum curve
- The process is computerized and the instruments employing this approach are called FT spectrometers
FOURIER TRANFORM SPECTROPHOTOMETERS
Advantages of FT Systems
- Produce better S/N ratios (throughput or Jacquinot advantage)
- Time for measurement is drastically reduced (all λs are measured simultaneously)
- Accurate and reproducible wavelength measurements