instrumental analysis chem 4811 - meerutcollege.org · instrumental analysis chem 4811 chapter dr....

79
INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut U.P.

Upload: others

Post on 10-Nov-2020

14 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INSTRUMENTAL ANALYSISCHEM 4811

CHAPTER

DR. Neelam KumariAssOCIATE professor

DEPARTMENT of chemistryMeerut college meerut

U.P.

Page 2: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

CHAPTER 4

INFRARED SPECTROSCOPY

(IR)

CHAPTER 4

INFRARED SPECTROSCOPY

(IR)

Page 3: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

IR SPECTROSCOPY

- Used for qualitative identification of organic andinorganic compounds

- Used for checking the presence of functional groups in molecules

- Can also be used for quantitative measurements of compounds

- Each compound has its unique IR absorption pattern

- Wavenumber with units of cm-1 is commonly used

- Wavenumber = number of waves of radiation per centimeter

- Used for qualitative identification of organic andinorganic compounds

- Used for checking the presence of functional groups in molecules

- Can also be used for quantitative measurements of compounds

- Each compound has its unique IR absorption pattern

- Wavenumber with units of cm-1 is commonly used

- Wavenumber = number of waves of radiation per centimeter

Page 4: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

IR SPECTROSCOPY

- The IR region has lower energy than visible radiation andhigher energy than microwave

The IR region is divided into

Near-IR (NIR): 750 nm – 2500 nm

Mid-IR: 2500 nm – 20000 nm

Far-IR: 20000 nm – 400000 nm

- The IR region has lower energy than visible radiation andhigher energy than microwave

The IR region is divided into

Near-IR (NIR): 750 nm – 2500 nm

Mid-IR: 2500 nm – 20000 nm

Far-IR: 20000 nm – 400000 nm

Page 5: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

IR ABSORPTION BY MOLECULES

- Molecules with covalent bonds may absorb IR radiation

- Absorption is quantized- Molecule shows change in dipolemoment during vibrational transitions

- Molecules move to a higher energy state

- IR radiation is sufficient enough to cause rotation and vibration

- Radiation between 1 and 100 µm will cause excitation tohigher vibrational states

- Radiation higher than 100 µm will cause excitation tohigher rotational states

- Molecules with covalent bonds may absorb IR radiation

- Absorption is quantized- Molecule shows change in dipolemoment during vibrational transitions

- Molecules move to a higher energy state

- IR radiation is sufficient enough to cause rotation and vibration

- Radiation between 1 and 100 µm will cause excitation tohigher vibrational states

- Radiation higher than 100 µm will cause excitation tohigher rotational states

Page 6: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

IR ABSORPTION BY MOLECULES

- Absorption spectrum is composed of broad vibrationalabsorption bands

- Molecules absorb radiation when a bond in the molecule vibratesat the same frequency as the incident radiant energy

- Molecules vibrate at higher amplitude after absorption

- A molecule must have a change in dipole moment duringvibration in order to absorb IR radiation

- Absorption spectrum is composed of broad vibrationalabsorption bands

- Molecules absorb radiation when a bond in the molecule vibratesat the same frequency as the incident radiant energy

- Molecules vibrate at higher amplitude after absorption

- A molecule must have a change in dipole moment duringvibration in order to absorb IR radiation

Page 7: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

IR ABSORPTION BY MOLECULES

Absorption frequency depends on

- Masses of atoms in the bonds

- Geometry of the molecule

- Strength of bond

- Other contributing factors

Absorption frequency depends on

- Masses of atoms in the bonds

- Geometry of the molecule

- Strength of bond

- Other contributing factors

Page 8: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DIPOLE MOMENT (µ)

µ = Q x r

Q = charge and r = distance between charges

- Asymmetrical distribution of electrons in a bond renders thebond polar

- A result of electronegativity difference

- µ changes upon vibration due to changes in r

- Change in µ with time is necessary for a molecule to absorbIR radiation

µ = Q x r

Q = charge and r = distance between charges

- Asymmetrical distribution of electrons in a bond renders thebond polar

- A result of electronegativity difference

- µ changes upon vibration due to changes in r

- Change in µ with time is necessary for a molecule to absorbIR radiation

Page 9: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DIPOLE MOMENT (µ)

- The repetitive changes in µ makes it possible for polar moleculesto absorb IR radiation

- Symmetrical molecules do not absorb IR radiation since theydo not have dipole moment (O2, F2, H2, Cl2)

- Diatomic molecules with dipole moment are IR-active(HCl, HF, CO, HI)

- Molecules with more than two atoms may or may not beIR active depending on whether they have permanent

net dipole moment

- The repetitive changes in µ makes it possible for polar moleculesto absorb IR radiation

- Symmetrical molecules do not absorb IR radiation since theydo not have dipole moment (O2, F2, H2, Cl2)

- Diatomic molecules with dipole moment are IR-active(HCl, HF, CO, HI)

- Molecules with more than two atoms may or may not beIR active depending on whether they have permanent

net dipole moment

Page 10: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

PRINCIPAL MODES OF VIBRATION

Stretching

- Change in bond length resulting from change ininteratomic distance (r)

Two stretching modes- Symmetrical and asymmetrical stretching

- Symmetrical stretching is IR-inactive (no change in µ)

Stretching

- Change in bond length resulting from change ininteratomic distance (r)

Two stretching modes- Symmetrical and asymmetrical stretching

- Symmetrical stretching is IR-inactive (no change in µ)

Page 11: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

PRINCIPAL MODES OF VIBRATION

Bending

- Change in bond angle or change in the position of a group ofatoms with respect to the rest of the molecule

Bending Modes- Scissoring and Rocking

- In-plane bending modes (atoms remain in the same plane)

- Wagging and TwistingOut-of-plane (oop) bending modes (atoms move out of plane)

Bending

- Change in bond angle or change in the position of a group ofatoms with respect to the rest of the molecule

Bending Modes- Scissoring and Rocking

- In-plane bending modes (atoms remain in the same plane)

- Wagging and TwistingOut-of-plane (oop) bending modes (atoms move out of plane)

Page 12: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

PRINCIPAL MODES OF VIBRATION

3N-6 possible normal modes of vibration

N = number of atoms in a molecule

Degrees of freedom = 3N

H2O for example- 3 atoms

- Degrees of freedom = 3 x 3 = 9- Normal modes of vibration = 9-6 = 3

3N-6 possible normal modes of vibration

N = number of atoms in a molecule

Degrees of freedom = 3N

H2O for example- 3 atoms

- Degrees of freedom = 3 x 3 = 9- Normal modes of vibration = 9-6 = 3

Page 13: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

PRINCIPAL MODES OF VIBRATION

Linear Molecules

- Cannot rotate about the bond axis

- Only 2 degrees of freedom describe rotation

3N-5 possible normal modes of vibration

CO2 for example- 3 atoms

- Normal modes of vibration = 9-5 = 4

Linear Molecules

- Cannot rotate about the bond axis

- Only 2 degrees of freedom describe rotation

3N-5 possible normal modes of vibration

CO2 for example- 3 atoms

- Normal modes of vibration = 9-5 = 4

Page 14: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

TRANSITIONS

Fundamental

- Excitation from the ground state Vo to the first excited state V1

- The most likely transition and have strong absorption bands

Fundamental

- Excitation from the ground state Vo to the first excited state V1

- The most likely transition and have strong absorption bands

Vo

V1

V2

V3

Page 15: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

FUNDAMENTAL TRANSITIONS

Overtone

- Excitation from ground state to higher energy states V2, V3, ….

- Result in overtone bands that are weaker than fundamental

- Frequencies are integral multiples of fundamental absorption

- Fewer peaks are seen than predicted on spectra due to IR-inactivevibrations, degenerate vibrations, weak vibrations

- Additional peaks may be seen due to overtones

Overtone

- Excitation from ground state to higher energy states V2, V3, ….

- Result in overtone bands that are weaker than fundamental

- Frequencies are integral multiples of fundamental absorption

- Fewer peaks are seen than predicted on spectra due to IR-inactivevibrations, degenerate vibrations, weak vibrations

- Additional peaks may be seen due to overtones

Page 16: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

COUPLING

- The interaction between vibrational modes

- Two vibrational frequencies may couple to produce a newvibrational frequency

Combinational Bandν1 + ν2 = ν3

Difference Bandν1 - ν2 = ν3

- The interaction between vibrational modes

- Two vibrational frequencies may couple to produce a newvibrational frequency

Combinational Bandν1 + ν2 = ν3

Difference Bandν1 - ν2 = ν3

Page 17: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

VIBRATIONAL MOTION

- Consider a bond as a spring

μf

2π1ν~

c

c = speed of light (cm/s)f = force constant (dyn/cm; proportional to bond strength)

f for a double bond = 2f for a single bondf for a triple bond = 3f for a single bond

c = speed of light (cm/s)f = force constant (dyn/cm; proportional to bond strength)

f for a double bond = 2f for a single bondf for a triple bond = 3f for a single bond

- M1 and M2 are masses of vibrating atoms connecting the bond

21

21

MM

MMgramsinmassreducedμ

Page 18: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INSTRUMENTATION

Components

- Radiation source

- Sample holder

- Monochromator

- Detector

- Computer

Components

- Radiation source

- Sample holder

- Monochromator

- Detector

- Computer

Page 19: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RADIATION SOURCES

Intensity of radiation should- Be continuous over the λ range and cover a wide λ range

- Constant over long periods of time- Have normal operating temperatures between 1100 and 1500 K

- Have maximum intensity between 4000 and 400 cm-1

(mid-IR region)

- Modern sources include furnace ignitors, diesel engine heaters

- Sources are usually enclosed in an insulator to reduce noise

Intensity of radiation should- Be continuous over the λ range and cover a wide λ range

- Constant over long periods of time- Have normal operating temperatures between 1100 and 1500 K

- Have maximum intensity between 4000 and 400 cm-1

(mid-IR region)

- Modern sources include furnace ignitors, diesel engine heaters

- Sources are usually enclosed in an insulator to reduce noise

Page 20: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RADIATION SOURCES

Mid-IR Sources

Nernst Glowers- Heated ceramic rods

- Cylindrical bar- Made of zirconium oxide, cerium oxide, thorium oxide

- Heated electrically to 1500 K – 2000 K- Resistance decreases as temperature increases

- Can overheat and burn out

Mid-IR Sources

Nernst Glowers- Heated ceramic rods

- Cylindrical bar- Made of zirconium oxide, cerium oxide, thorium oxide

- Heated electrically to 1500 K – 2000 K- Resistance decreases as temperature increases

- Can overheat and burn out

Page 21: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RADIATION SOURCES

Mid-IR Sources

Globar- Silicon carbide bar

- Heated electrically to emit continuous IR radiation

- More sensitive than the Nernst glower

Mid-IR Sources

Globar- Silicon carbide bar

- Heated electrically to emit continuous IR radiation

- More sensitive than the Nernst glower

Page 22: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RADIATION SOURCES

Mid-IR Sources

Heated Wire Coils- Heated electrically to ~ 1100 oC

- Similar in shape to incandescent light bulb filament- Nichrome wire is commonly used

- Rhodium is also used- May easily fracture and burn out

Mid-IR Sources

Heated Wire Coils- Heated electrically to ~ 1100 oC

- Similar in shape to incandescent light bulb filament- Nichrome wire is commonly used

- Rhodium is also used- May easily fracture and burn out

Page 23: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RADIATION SOURCES

Near-IR Sources

Quartz halogen lamp- Contains tungsten wire filament

- Also contains iodine vapor sealed in a quartz bulb

- Evaporated tungsten is redeposited on the filamentincreasing stability

- Output range is between 25000 and 2000 cm-1

Near-IR Sources

Quartz halogen lamp- Contains tungsten wire filament

- Also contains iodine vapor sealed in a quartz bulb

- Evaporated tungsten is redeposited on the filamentincreasing stability

- Output range is between 25000 and 2000 cm-1

Page 24: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RADIATION SOURCES

Far-IR Sources

High pressure Hg discharge lamp- Made of quartz bulb containing elemental Hg

- Also contains inert gas and two electrodes

Far-IR Sources

High pressure Hg discharge lamp- Made of quartz bulb containing elemental Hg

- Also contains inert gas and two electrodes

Page 25: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RADIATION SOURCES

IR Laser Sources

- Excellent light source for measuring gases

Two typesGas-phase (tunable CO2) laser and solid-state laser

Laser- Light source that emits monochromatic radiation

IR Laser Sources

- Excellent light source for measuring gases

Two typesGas-phase (tunable CO2) laser and solid-state laser

Laser- Light source that emits monochromatic radiation

Page 26: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

MONOCHROMATORS

- Salt prisms and metal gratings are used as dispersion devices

- Mirrors made of metal with polished front surface

- Spectrum is recorded by moving prism or grating suchthat different radiation frequencies pass through

the exit slit to the detector

- Spectrum obtained is %T verses wavenumber (or frequency)

- Salt prisms and metal gratings are used as dispersion devices

- Mirrors made of metal with polished front surface

- Spectrum is recorded by moving prism or grating suchthat different radiation frequencies pass through

the exit slit to the detector

- Spectrum obtained is %T verses wavenumber (or frequency)

Page 27: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

FT SPECTROMETERS

- Based on Michelson interferometer

- Employs constructive and destructive interferences

- Destructive interference is a maximum when two beams are180o out of phase

- An FT is used to convert the time-domain spectrum obtained intoa frequency-domain spectrum

- The system is called FTIR

- Based on Michelson interferometer

- Employs constructive and destructive interferences

- Destructive interference is a maximum when two beams are180o out of phase

- An FT is used to convert the time-domain spectrum obtained intoa frequency-domain spectrum

- The system is called FTIR

Page 28: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

FT SPECTROMETERS

- Has higher signal-to-noise ratio

- More accurate and precise than dispersive monochromators(Conne’s advantage)

- Much greater radiation intensity falls on the detector dueto the absence of slits

(throughput or Jacquinot’s advantage)

- Has higher signal-to-noise ratio

- More accurate and precise than dispersive monochromators(Conne’s advantage)

- Much greater radiation intensity falls on the detector dueto the absence of slits

(throughput or Jacquinot’s advantage)

Page 29: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

FT SPECTROMETERS

Focusing Mirrors of Interferometer

- Made of metal and polished on the front surface

- Gold coated to prevent corrosion

- Focus source radiation onto the beam splitter

- Focus light emitting from sample onto detector

Focusing Mirrors of Interferometer

- Made of metal and polished on the front surface

- Gold coated to prevent corrosion

- Focus source radiation onto the beam splitter

- Focus light emitting from sample onto detector

Page 30: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

FT SPECTROMETERS

Beam Splitter of Interferometer

- Made of Ge or Si coated onto a highly polishedIR-transparent substrate

- Ge coated on KBr is used for mid-IR

- Si coated on quartz is used for NIR

- Mylar film is used for far-IR

- Transmits and reflects 50% each of beam (ideally)

Beam Splitter of Interferometer

- Made of Ge or Si coated onto a highly polishedIR-transparent substrate

- Ge coated on KBr is used for mid-IR

- Si coated on quartz is used for NIR

- Mylar film is used for far-IR

- Transmits and reflects 50% each of beam (ideally)

Page 31: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DETECTORS

Two classes

Thermal detectors

and

photon-sensitive detectors

Two classes

Thermal detectors

and

photon-sensitive detectors

Page 32: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DETECTORS

Thermal Detectors

- Bolometers

- Thermocouples

- Thermistors

- Pyroelectric devices

Thermal Detectors

- Bolometers

- Thermocouples

- Thermistors

- Pyroelectric devices

Page 33: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DETECTORS

Thermal Detectors

Bolometers- Very sensitive electrical resistance thermometer

- Older ones were made of Pt wire(resistance change = 0.4% per oC)

- Modern ones are made of silicon placed in a wheatstone bridge

- Is a few micrometers in diameter

- Fast response time

Thermal Detectors

Bolometers- Very sensitive electrical resistance thermometer

- Older ones were made of Pt wire(resistance change = 0.4% per oC)

- Modern ones are made of silicon placed in a wheatstone bridge

- Is a few micrometers in diameter

- Fast response time

Page 34: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DETECTORS

Thermal Detectors

Thermocouples- Made up of two wires welded together at both ends

- Two wires are from different metals- One welded joint (hot junction) is exposed to IR radiation

- The other joint (cold junction) is kept at constant temperature- Hot junction becomes hotter than cold junction

- Potential difference is a function of IR radiation- Slow response time

- Cannot be used for FTIR

Thermal Detectors

Thermocouples- Made up of two wires welded together at both ends

- Two wires are from different metals- One welded joint (hot junction) is exposed to IR radiation

- The other joint (cold junction) is kept at constant temperature- Hot junction becomes hotter than cold junction

- Potential difference is a function of IR radiation- Slow response time

- Cannot be used for FTIR

Page 35: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DETECTORS

Thermal Detectors

Thermistors- Made of fused mixture of metal oxides

- Increasing temperature decreases electrical resistancewhich is a function of IR radiation

- Resistance change is about 5% per oC

- Slow response time

Thermal Detectors

Thermistors- Made of fused mixture of metal oxides

- Increasing temperature decreases electrical resistancewhich is a function of IR radiation

- Resistance change is about 5% per oC

- Slow response time

Page 36: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DETECTORS

Thermal Detectors

Pyroelectric Detectors- Made of dielectric materials (insulators), ferroelectric materials,

or semiconductors

- A thin crystal of the material is placed between two electrodes

- Change in temperature changes polarization of materialwhich is a function of IR radiation exposed

- The electrodes measure the change in polarization

Thermal Detectors

Pyroelectric Detectors- Made of dielectric materials (insulators), ferroelectric materials,

or semiconductors

- A thin crystal of the material is placed between two electrodes

- Change in temperature changes polarization of materialwhich is a function of IR radiation exposed

- The electrodes measure the change in polarization

Page 37: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DETECTORS

Photon Detectors

- Made of materials such as lead selenide (PbSe),indium gallium arsenide (InGaSa), indium antimonide (InSb)

- Materials are semiconductors

- S/N increases as temperature decreases (cooling is necessary)

- Very sensitive and very fast response time

- Good for FTIR and coupled techniques

Photon Detectors

- Made of materials such as lead selenide (PbSe),indium gallium arsenide (InGaSa), indium antimonide (InSb)

- Materials are semiconductors

- S/N increases as temperature decreases (cooling is necessary)

- Very sensitive and very fast response time

- Good for FTIR and coupled techniques

Page 38: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RESPONSE TIME

- The length of time that a detector takes to reach a steadysignal when radiation falls on it

- Response time for older dispersive IR spectrometerswere ~ 15 minutes

- Response time is very slow in thermal detectors due to slowchange in temperature near equilibrium

- The length of time that a detector takes to reach a steadysignal when radiation falls on it

- Response time for older dispersive IR spectrometerswere ~ 15 minutes

- Response time is very slow in thermal detectors due to slowchange in temperature near equilibrium

Page 39: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RESPONSE TIME

Semiconductors

- Are insulators when no radiation falls on them but conductorswhen radiation falls on them

- There is no temperature change involved

- Measured signal is due to change in resistance uponexposure to IR radiation ( is rapid)

- Response time is the time required to change the semiconductorfrom insulator to conductor (~ 1 ns)

Semiconductors

- Are insulators when no radiation falls on them but conductorswhen radiation falls on them

- There is no temperature change involved

- Measured signal is due to change in resistance uponexposure to IR radiation ( is rapid)

- Response time is the time required to change the semiconductorfrom insulator to conductor (~ 1 ns)

Page 40: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

SATURATION

- Very high levels of light can saturate detector

- Linearity is very important especially inthe mid-IR region

- Very high levels of light can saturate detector

- Linearity is very important especially inthe mid-IR region

Page 41: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

TRANSMISSION (ABSORPTION) TECHNIQUE

- Can be used for both qualitative and quantitative analysis

- Provides very high sensitivity at relatively low cost

- Sample cell material must be transparent to IR radiation(NaCl, KBr)

- Samples should not contain water since materials aresoluble in water

- Can be used for both qualitative and quantitative analysis

- Provides very high sensitivity at relatively low cost

- Sample cell material must be transparent to IR radiation(NaCl, KBr)

- Samples should not contain water since materials aresoluble in water

Page 42: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

TRANSMISSION (ABSORPTION) TECHNIQUE

Solid Samples

Three sampling techniquesMulling, pelleting, thin film

Mulling- Samples are ground to power with a few drops of viscous

liquid like Nujol (size < 2 µm)

- 2-4 mg sample is made into a mull (thick slurry)

- Mull is pressed between salt plates to form a thin film

Solid Samples

Three sampling techniquesMulling, pelleting, thin film

Mulling- Samples are ground to power with a few drops of viscous

liquid like Nujol (size < 2 µm)

- 2-4 mg sample is made into a mull (thick slurry)

- Mull is pressed between salt plates to form a thin film

Page 43: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

TRANSMISSION (ABSORPTION) TECHNIQUE

Solid Samples

Three sampling techniquesMulling, pelleting, thin film

Pelleting- 1 mg ground sample is mixed with 100 mg of dry KBr powder

- Mixture is compressed under very high pressure

- Small disk with very smooth surfaces forms (looks like glass)

Solid Samples

Three sampling techniquesMulling, pelleting, thin film

Pelleting- 1 mg ground sample is mixed with 100 mg of dry KBr powder

- Mixture is compressed under very high pressure

- Small disk with very smooth surfaces forms (looks like glass)

Page 44: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

TRANSMISSION (ABSORPTION) TECHNIQUE

Solid Samples

Three sampling techniquesMulling, pelleting, thin film

Thin Film- Molten sample is deposited on the surface of KBr or NaCl plates

- Sample is allowed to dry to form a thin film on substrate

- Good for qualitative identification of polymers but notgood for quantitative analysis

Solid Samples

Three sampling techniquesMulling, pelleting, thin film

Thin Film- Molten sample is deposited on the surface of KBr or NaCl plates

- Sample is allowed to dry to form a thin film on substrate

- Good for qualitative identification of polymers but notgood for quantitative analysis

Page 45: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

TRANSMISSION (ABSORPTION) TECHNIQUE

Liquid Samples

- Many liquid samples are analyzed as is

- Dilution with the appropriate solvent may be necessary(CCl4, CS2, CH3Cl)

- Solvent must be transparent in the region of interest

- Salt plates are hygroscopic and water soluble (avoid water)

- Special cells are used for water containing samples (BaF2, AgCl)

Liquid Samples

- Many liquid samples are analyzed as is

- Dilution with the appropriate solvent may be necessary(CCl4, CS2, CH3Cl)

- Solvent must be transparent in the region of interest

- Salt plates are hygroscopic and water soluble (avoid water)

- Special cells are used for water containing samples (BaF2, AgCl)

Page 46: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

TRANSMISSION (ABSORPTION) TECHNIQUE

Gas Samples

- Gas cells have longer pathlengths to compensate for verylow concentrations of gas samples

- Generally about 10 cm(commercial ones are up to 120 cm)

- Temperature and pressure are controlled if quantitativeanalysis is required

Gas Samples

- Gas cells have longer pathlengths to compensate for verylow concentrations of gas samples

- Generally about 10 cm(commercial ones are up to 120 cm)

- Temperature and pressure are controlled if quantitativeanalysis is required

Page 47: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

TRANSMISSION (ABSORPTION) TECHNIQUE

Background Absorption

- Absorption from material other than sample

Two sources- Solvent

- Air in the light path

- Background spectrum is subtracted from sample spectrum

- CO2 and H2O absorb IR radiation(sealed desiccants are used to reduce CO2 and H2O absorption)

Background Absorption

- Absorption from material other than sample

Two sources- Solvent

- Air in the light path

- Background spectrum is subtracted from sample spectrum

- CO2 and H2O absorb IR radiation(sealed desiccants are used to reduce CO2 and H2O absorption)

Page 48: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

ATTENUATED TOTAL REFLECTION(ATR) TECHNIQUE

- Uses high refractive index optical element

- Called internal reflection element (IRE) or ATR crystal(germanium, ZnSe, silicon, diamond)

- Sample has lower refractive index than the ATR crystal

- Sample is placed on the ATR crystal

- When light travelling in the crystal hits the crystal-sampleinterface, total internal reflection occurs if the angle

of incidence is greater than the critical angle

- Uses high refractive index optical element

- Called internal reflection element (IRE) or ATR crystal(germanium, ZnSe, silicon, diamond)

- Sample has lower refractive index than the ATR crystal

- Sample is placed on the ATR crystal

- When light travelling in the crystal hits the crystal-sampleinterface, total internal reflection occurs if the angle

of incidence is greater than the critical angle

Page 49: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

ATTENUATED TOTAL REFLECTION(ATR) TECHNIQUE

- The beam of light penetrates the sample a very short distance

- This is known as the evanescent wave

- Light beam is attenuated (reduced in intensity) if the evanescentwave is absorbed

- IR absorption spectrum is obtained from the interactionsbetween the evanescent wave and the sample

- The beam of light penetrates the sample a very short distance

- This is known as the evanescent wave

- Light beam is attenuated (reduced in intensity) if the evanescentwave is absorbed

- IR absorption spectrum is obtained from the interactionsbetween the evanescent wave and the sample

Page 50: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

SPECULAR REFLECTANCE TECHNIQUE

- Occurs when angle of reflected light equals angle of incident light(45o angle of incidence is typically used)

- Occurs when light bounces off a smooth surface

- Nondestructive method for studying thin films

- Beam passes through thin film of sample where absorption occurs

- Beam is reflected from a polished surface below the sample

- Reflected beam passes through sample again

- Occurs when angle of reflected light equals angle of incident light(45o angle of incidence is typically used)

- Occurs when light bounces off a smooth surface

- Nondestructive method for studying thin films

- Beam passes through thin film of sample where absorption occurs

- Beam is reflected from a polished surface below the sample

- Reflected beam passes through sample again

Page 51: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

DIFFUSE REFLECTANCE TECHNIQUE

- DR OR DRIFTS

- Used to obtain IR or NIR spectrum from a rough surface

- Beam penetrates sample to about 100 µm

- Reflected light is scattered at many angles

- Scattered radiation is collected with a large mirror

- Good for powdered samples (sample is mixed with KBr powder)

- DR OR DRIFTS

- Used to obtain IR or NIR spectrum from a rough surface

- Beam penetrates sample to about 100 µm

- Reflected light is scattered at many angles

- Scattered radiation is collected with a large mirror

- Good for powdered samples (sample is mixed with KBr powder)

Page 52: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

IR EMISSION

- Certain samples are characterized using emission spectrum

- Sample molecules are heated to occupy excited vibrational states

- Radiation is emitted upon returning to the ground state

- Emitted IR radiation is directed into the spectrometer

- Radiation source is not needed (sample is the source)

- Method is nondestructive

- Emitted radiation is used to identify sample

- Certain samples are characterized using emission spectrum

- Sample molecules are heated to occupy excited vibrational states

- Radiation is emitted upon returning to the ground state

- Emitted IR radiation is directed into the spectrometer

- Radiation source is not needed (sample is the source)

- Method is nondestructive

- Emitted radiation is used to identify sample

Page 53: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

FTIR MICROSCOPY

- For forensic analysis

- Analysis of paint chips from cars (hit-and-run accidents)

- Examination of fibers, drugs, explosives

- Characterization of pharmaceuticals, adhesives and gemstones

- For forensic analysis

- Analysis of paint chips from cars (hit-and-run accidents)

- Examination of fibers, drugs, explosives

- Characterization of pharmaceuticals, adhesives and gemstones

Page 54: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

NEAR-IR (NIR) SPECTROSCOPY

- Region covers 750 nm – 2500 nm (13000 cm-1 – 4000 cm-1)

- Long wavelength end of IR region

- Bands occurring in this region are due to OH, NH, and CH bonds

- Bands are primarily overtone and combination bands

- Light source is tungsten-halogen lamp

- Detector is lead sulfide photodetector

- Quartz or fused silica sample cells with long pathlengths are used

- Region covers 750 nm – 2500 nm (13000 cm-1 – 4000 cm-1)

- Long wavelength end of IR region

- Bands occurring in this region are due to OH, NH, and CH bonds

- Bands are primarily overtone and combination bands

- Light source is tungsten-halogen lamp

- Detector is lead sulfide photodetector

- Quartz or fused silica sample cells with long pathlengths are used

Page 55: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

NEAR-IR (NIR) SPECTROSCOPY

Primary absorption bands seen in NIR

C−H Bands2100 – 2450 nm and 1600 – 1800 nm

N−H Bands1450 – 1550 nm and 2800 – 3000 nm

O−H Bands1390 – 1450 nm and 2700 – 2900 nm

Primary absorption bands seen in NIR

C−H Bands2100 – 2450 nm and 1600 – 1800 nm

N−H Bands1450 – 1550 nm and 2800 – 3000 nm

O−H Bands1390 – 1450 nm and 2700 – 2900 nm

Page 56: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

NEAR-IR (NIR) SPECTROSCOPY

- Used for quantitative analysis of solid and liquid samplescontaining OH, NH, CH bonds

- For quantitative characterization of polymers, food,proteins, agricultural products

- Pharmaceutical tablets can be analyzed nondestructively

- Forensic analysis of unknown wrapped powders believed to bedrugs are analyzed without destroying the wrappers

- Used for quantitative analysis of solid and liquid samplescontaining OH, NH, CH bonds

- For quantitative characterization of polymers, food,proteins, agricultural products

- Pharmaceutical tablets can be analyzed nondestructively

- Forensic analysis of unknown wrapped powders believed to bedrugs are analyzed without destroying the wrappers

Page 57: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RAMAN SPECTROSCOPY

- Light scattering technique for studying molecular vibrations

- Change in polarization is necessary for a vibration to beseen in Raman spectrum

- Implies change in distribution of electron cloud aroundvibrating atoms

- Polarization is easier for long bonds than for short bonds

- IR-inactive radiations are Raman-active

- Light scattering technique for studying molecular vibrations

- Change in polarization is necessary for a vibration to beseen in Raman spectrum

- Implies change in distribution of electron cloud aroundvibrating atoms

- Polarization is easier for long bonds than for short bonds

- IR-inactive radiations are Raman-active

Page 58: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RAMAN SPECTROSCOPY

Principles

- Part of the radiation is scattered by molecules whenthe radiation passes through sample

Three types of scattering occurs- Rayleigh scattering- Stokes scattering

- Anti-Stokes scattering

Principles

- Part of the radiation is scattered by molecules whenthe radiation passes through sample

Three types of scattering occurs- Rayleigh scattering- Stokes scattering

- Anti-Stokes scattering

Page 59: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RAMAN SPECTROSCOPY

Principles

Rayleigh scattering- Result of elastic collisions between photons and sample molecules

- Energy is same as incident radiation

Stokes scattering- Scattered photons have less energy than the incident radiation

- Results in spectral lines called Raman lines

Anti-Stokes scattering- Raman lines result from photons scattered with more energy

Principles

Rayleigh scattering- Result of elastic collisions between photons and sample molecules

- Energy is same as incident radiation

Stokes scattering- Scattered photons have less energy than the incident radiation

- Results in spectral lines called Raman lines

Anti-Stokes scattering- Raman lines result from photons scattered with more energy

Page 60: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RAMAN SPECTROSCOPY

Principles

- Stokes and anti-stokes are due to inelastic collisions

- The process is not quantized

- Raman lines are shifted in frequency from Rayleigh frequency

- Radiation measured is visible or NIR

Principles

- Stokes and anti-stokes are due to inelastic collisions

- The process is not quantized

- Raman lines are shifted in frequency from Rayleigh frequency

- Radiation measured is visible or NIR

Page 61: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

RAMAN SPECTROSCOPY

Advantages over IR

- Aqueous solutions can be analyzed

- Fewer and much sharper lines so better for quantitative analysis

Techniques- Resonance Raman Spectroscopy

- Surface-Enhanced Raman Spectroscopy (SERS)- Raman Microscopy

Advantages over IR

- Aqueous solutions can be analyzed

- Fewer and much sharper lines so better for quantitative analysis

Techniques- Resonance Raman Spectroscopy

- Surface-Enhanced Raman Spectroscopy (SERS)- Raman Microscopy

Page 62: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

APPLICATIONS OF IR SPECTROSCOPY

Quantitative

- Extent of absorption and Beer’s law can be used to determineconcentration of unknown analytes in sample

- Absorption band unique to the analyte molecule should beused for measurements

- Generally performed with samples in solutions

- Light scattering may occur with pellets whichdeviates from Beer’s law

Quantitative

- Extent of absorption and Beer’s law can be used to determineconcentration of unknown analytes in sample

- Absorption band unique to the analyte molecule should beused for measurements

- Generally performed with samples in solutions

- Light scattering may occur with pellets whichdeviates from Beer’s law

Page 63: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

APPLICATIONS OF IR SPECTROSCOPY

Quantitative

- Measure absorption intensities of standard solutions andunknown at exactly the same wavenumber

- All measurements must be made from the same baseline

- Plot a calibration curve

- Use the relationship obtained to determine the concentrationof unknown

- Not as accurate as using UV-VIS spectroscopy

Quantitative

- Measure absorption intensities of standard solutions andunknown at exactly the same wavenumber

- All measurements must be made from the same baseline

- Plot a calibration curve

- Use the relationship obtained to determine the concentrationof unknown

- Not as accurate as using UV-VIS spectroscopy

Page 64: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

APPLICATIONS OF IR SPECTROSCOPY

Quantitative

- Determination of impurities in raw materials (quality control)

- Analysis of contaminations from oil or grease

- Determination of reaction rates of slow reactions

Quantitative

- Determination of impurities in raw materials (quality control)

- Analysis of contaminations from oil or grease

- Determination of reaction rates of slow reactions

Page 65: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

APPLICATIONS OF IR SPECTROSCOPY

Qualitative

- Identification of unknown samples by matching the absorptionspectra with that of known compounds

- Identification of functional groups present in a sample(classification of unknowns)

Qualitative

- Identification of unknown samples by matching the absorptionspectra with that of known compounds

- Identification of functional groups present in a sample(classification of unknowns)

Page 66: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

PREDICTING STRUCTURE OF UNKNOWN

- Identify the major functional groups from the strongabsorption peaks

- Identify the compound as aromatic or aliphatic

- Subtract the FW of all functional groups identified from the givenmolecular weight of the compound

- Look for C≡C and C=C stretching bands

- Look for other unique CH bands (e.g. aldehyde)

- Use the difference obtained to deduce the structure

- Identify the major functional groups from the strongabsorption peaks

- Identify the compound as aromatic or aliphatic

- Subtract the FW of all functional groups identified from the givenmolecular weight of the compound

- Look for C≡C and C=C stretching bands

- Look for other unique CH bands (e.g. aldehyde)

- Use the difference obtained to deduce the structure

Page 67: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Functional Group Region- Strong absorptions due to stretching from hydroxyl, amine,

carbonyl, CHx

4000 – 1300 cm-1

Fingerprint Region- Result of interactions between vibrations

1300 – 910 cm-1

Functional Group Region- Strong absorptions due to stretching from hydroxyl, amine,

carbonyl, CHx

4000 – 1300 cm-1

Fingerprint Region- Result of interactions between vibrations

1300 – 910 cm-1

Page 68: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Hydrocarbons

- Absorption bands are due to the stretching or bending ofC−H and C−C bonds

- C−C stretching vibrations are distributed across thefingerprint region (not useful for identification)

- C−C bending vibrations occur below 500 cm-1

(not useful for identification)

- Observed bands are due to C−H stretching or bending

Hydrocarbons

- Absorption bands are due to the stretching or bending ofC−H and C−C bonds

- C−C stretching vibrations are distributed across thefingerprint region (not useful for identification)

- C−C bending vibrations occur below 500 cm-1

(not useful for identification)

- Observed bands are due to C−H stretching or bending

Page 69: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Cyclic Alkanes- No peak around 1375 cm-1 due to absence of methyl groups

- Two peaks at ~ 900 cm-1 and 860 cm-1 due to ring deformation

Alkenes- Contain many more peaks than alkanes

- Peaks of interest are due to stretching and bending of C−Hand C=C bonds

- C=C band will not appear if there is symmetrical substitutionabout the C=C bond

Cyclic Alkanes- No peak around 1375 cm-1 due to absence of methyl groups

- Two peaks at ~ 900 cm-1 and 860 cm-1 due to ring deformation

Alkenes- Contain many more peaks than alkanes

- Peaks of interest are due to stretching and bending of C−Hand C=C bonds

- C=C band will not appear if there is symmetrical substitutionabout the C=C bond

Page 70: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Alkynes- C≡C peak appears around 2100 – 2200 cm-1

- Terminal alkyne ≡C−H stretch occurs near 3300 cm-1

Aromatic Hydrocarbons- C→H absorption occurs above 3000 cm-1

- Aromatic C=C ring stretching absorption around1400 – 1600 cm-1 appears as doublet

- Aromatic C↓H oop band around 690 – 900 cm-1

- Overtones around 1660 – 2000 cm-1

Alkynes- C≡C peak appears around 2100 – 2200 cm-1

- Terminal alkyne ≡C−H stretch occurs near 3300 cm-1

Aromatic Hydrocarbons- C→H absorption occurs above 3000 cm-1

- Aromatic C=C ring stretching absorption around1400 – 1600 cm-1 appears as doublet

- Aromatic C↓H oop band around 690 – 900 cm-1

- Overtones around 1660 – 2000 cm-1

Page 71: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Alcohols

- OH band in neat aliphatic alcohols is a broad band centeredat ~ 3300 cm-1 due to hydrogen bonding (3100 – 3600 cm-1)

- OH band in dilute solutions of aliphatic alcohols is a sharppeak ~ 3600 cm-1

- C−C−O stretch ~ 1048 cm-1 for primary alcohols

- Decreasing frequency by 10 cm-1 in the order 1o>2o>3o

- Methyl bending vibrations at ~ 1200 – 1500 cm-1

Alcohols

- OH band in neat aliphatic alcohols is a broad band centeredat ~ 3300 cm-1 due to hydrogen bonding (3100 – 3600 cm-1)

- OH band in dilute solutions of aliphatic alcohols is a sharppeak ~ 3600 cm-1

- C−C−O stretch ~ 1048 cm-1 for primary alcohols

- Decreasing frequency by 10 cm-1 in the order 1o>2o>3o

- Methyl bending vibrations at ~ 1200 – 1500 cm-1

Page 72: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Phenol

- CO→H stretch is broad band

- C→H stretch ~ 3050 cm-1

- C−C→O band ~ 1225 cm-1

- C −O−H bend ~ 1350 cm-1

- Aromatic ring C stretching between 1450 – 1600 cm-1

- Monosubstituted bands ~ 745 – 895 cm-1 and 1650 – 2000 cm-1

Phenol

- CO→H stretch is broad band

- C→H stretch ~ 3050 cm-1

- C−C→O band ~ 1225 cm-1

- C −O−H bend ~ 1350 cm-1

- Aromatic ring C stretching between 1450 – 1600 cm-1

- Monosubstituted bands ~ 745 – 895 cm-1 and 1650 – 2000 cm-1

Page 73: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Aliphatic Acids

- Broad OH band around 2900 cm-1

- C−H stretching bands from CH3 and CH2 stick out at thebottom of the broad OH band

- C=O stretch ~ 1710 cm-1

- In-plane C −O−H bend ~ 1410 cm-1 andoop C −O−H bend ~ 930 cm-1

- C −C−O stretch dimer at ~ 1280 cm-1

Aliphatic Acids

- Broad OH band around 2900 cm-1

- C−H stretching bands from CH3 and CH2 stick out at thebottom of the broad OH band

- C=O stretch ~ 1710 cm-1

- In-plane C −O−H bend ~ 1410 cm-1 andoop C −O−H bend ~ 930 cm-1

- C −C−O stretch dimer at ~ 1280 cm-1

Page 74: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Carboxylic Acids, Esters, Ketones, Aldehydes

- Characterized by very strong carbonyl (C=O) stretchingband between 1650 cm-1 and 1850 cm-1

- Fermi resonance seen in aldehydes(doublet due to resonance with an overtone of the aldehydic

C−H bend at 1390 cm-1)

Carboxylic Acids, Esters, Ketones, Aldehydes

- Characterized by very strong carbonyl (C=O) stretchingband between 1650 cm-1 and 1850 cm-1

- Fermi resonance seen in aldehydes(doublet due to resonance with an overtone of the aldehydic

C−H bend at 1390 cm-1)

Page 75: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Nitrogen-Containing Compounds

- 1o amines (NH2) have scissoring mode and lowfrequency wagging mode

- 2o amines (NH) only have wagging mode (cannot scissor)

- 3o amines have no NH band and are characterized by C−Nstretching modes ~ 1000 – 1200 cm-1 and 700 – 900 cm-1

- 1o, 2o, 3o amides are similar to their amine counterpartsbut have additional C=O stretching band

Nitrogen-Containing Compounds

- 1o amines (NH2) have scissoring mode and lowfrequency wagging mode

- 2o amines (NH) only have wagging mode (cannot scissor)

- 3o amines have no NH band and are characterized by C−Nstretching modes ~ 1000 – 1200 cm-1 and 700 – 900 cm-1

- 1o, 2o, 3o amides are similar to their amine counterpartsbut have additional C=O stretching band

Page 76: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Nitrogen-Containing Compounds

- C=O stretching called amide I in 1o and 2o amides andamide II in 3o amides

- N−H stretch doublet ~ 3370 – 3291 cm-1 for 1o amines

- 1o N−H bend at ~ 1610 cm-1 and 800 cm-1

- Single N−H stretch ~ 3293 cm-1 for 2o but absent in 3o amine

- C−N stretch weak band ~ 1100 cm-1

Nitrogen-Containing Compounds

- C=O stretching called amide I in 1o and 2o amides andamide II in 3o amides

- N−H stretch doublet ~ 3370 – 3291 cm-1 for 1o amines

- 1o N−H bend at ~ 1610 cm-1 and 800 cm-1

- Single N−H stretch ~ 3293 cm-1 for 2o but absent in 3o amine

- C−N stretch weak band ~ 1100 cm-1

Page 77: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Amino Acids [RCH(NH2)COOH]

- IR spectrum is related to salts of amines and salts of acids

- Broad CH bands that overlap with each other

- Broad band ~ 2100 cm-1

- NH band ~ 1500 cm-1

- Carboxylate ion stretch ~ 1600 cm-1

Amino Acids [RCH(NH2)COOH]

- IR spectrum is related to salts of amines and salts of acids

- Broad CH bands that overlap with each other

- Broad band ~ 2100 cm-1

- NH band ~ 1500 cm-1

- Carboxylate ion stretch ~ 1600 cm-1

Page 78: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

INTERPRETATION OF IR SPECTRA

Halogenated Compounds

- C→X strong absorption bands in the fingerprint andaromatic regions

- More halogens on the same C results in an increase in intensityand a shift to higher wavenumbers

- Absorption due to C−Cl and C−Br occurs below 800 cm-1

Halogenated Compounds

- C→X strong absorption bands in the fingerprint andaromatic regions

- More halogens on the same C results in an increase in intensityand a shift to higher wavenumbers

- Absorption due to C−Cl and C−Br occurs below 800 cm-1

Page 79: INSTRUMENTAL ANALYSIS CHEM 4811 - meerutcollege.org · INSTRUMENTAL ANALYSIS CHEM 4811 CHAPTER DR. Neelam Kumari AssOCIATE professor DEPARTMENT of chemistry Meerut college meerut

LIMITATIONS OF IR SPECTROSCOPY

- Short pathlength

- Pathlength may vary from sample to sample

- Sample cells are soluble in water

- Short pathlength

- Pathlength may vary from sample to sample

- Sample cells are soluble in water